BORON IN GLASS DETERMINATION USING WDXRF

Size: px
Start display at page:

Download "BORON IN GLASS DETERMINATION USING WDXRF"

Transcription

1 269 ABSTRACT BORON IN GLASS DETERMINATION USING WDXRF Alexander Seyfarth Bruker AXS Inc., 5465 East Cheryl Parkway, Madison, Wisconsin The application of wavelength-dispersive XRF to the analysis of boron in glass is reviewed with respect to physical, instrumental, calibration, and maintenance considerations. Advances in XRF instrumentation have enabled improvement in the analysis of boron. Reduced window thickness and high-current capabilities of X-ray tubes, as well as new analyzer crystals, produce a higher intensity boron signal. With correct determination of background and matrix interferences for successful quantification, only the sample itself limits the analysis. Observed boron migration in some glasses needs to be understood in order to establish stable, routine quantification in a process environment. INTRODUCTION Boron is an important component of glass. It exerts a strong influence on the viscosity of the melt, imparting a beneficial effect to the chemical resistance and lowering the expansion coefficient of the finished article. Boron is widely used in the production of E type glasses (7 wt%), neutral glass for pharmaceutical applications (7 to 11 wt%), and Pyrex -type borosilicate glasses (13 wt%). Because of the high volatility of boron in the melting furnace, a very accurate procedure is needed to meet the strict boron concentration specifications required for the finished product and to permit any necessary batch adjustment to be implemented almost in real time. XRF is well suited for glass analyses. Since the 1980s, XRF has been the method of choice because of its ease of sample preparation and rapid results. In the production control of soda-lime glass, the technique has already replaced wet chemical techniques because of its high precision and fast measurement time. Boron, being a very light element, poses challenges to the analyst as well as to the instrument. Demand for boron analysis in the glass industry prompted the development of new generation wavelength-dispersive X-ray fluorescence (WDXRF) spectrometers. ASPECTS OF THE ANALYSIS OF BORON To put the difficulty of the analysis of boron in glass into perspective, consider its physical and chemical properties [1 3]. Physical data for boron is given in Table I. Table I. Physical data for boron. Atomic number (Z) K shell absorption edge (ev) Kα radiation (ev) Mass absorption coefficient µλ (cm 2 /g) Fluorescence yield (Ω) , Boron s low atomic number and low K-shell energy cause various issues for XRF analysis which will be addressed further.

2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -

3 270 EXCITATION AND FLUORESCENCE YIELD The emitted intensity, I B-Kα, of the characteristic fluorescence radiation is governed by the overlap of the excitation spectrum and the absorption coefficient µ λ of boron [4]. Absorption by the tube window cuts off the low energy (long wavelength) part of the excitation tube spectrum. Because of this, boron is excited with the primary X-ray continuum and the rhodium L lines. For optimum primary continuum, the tube window must be as thin as possible. Currently window thicknesses range from 150 to 30 µm; 75 µm has become the industry standard since it balances price, performance, and durability. Fluorescence yield is another important factor in the analysis of boron. Fluorescence yield is proportional to Z 4 and is therefore close to zero for low atomic number elements. Using the highest available tube current at the lowest possible voltage, the emission of boron photons from the sample can be optimized. Starting in 1988, the possible settings for tube current have increased from 100 ma to today s maximum of 170 ma [Siemens SRS 303 (1988) to Bruker S8 TIGER 4 kw (2006)]. Figure 1. Comparison of boron intensity in 3 kw and 4 kw WDXRF systems. As can be seen in Figure 1, 40% greater boron intensity was obtained from a boronphosphosilicate glass (BPSG) sample (4 wt% B) analyzed on a 4 kw instrument in 1999 than a 3 kw instrument in 1988 [SRS 3000 with 75 µm end window X-ray tube and 3 and 4 kw generator (1999)].

4 271 ANALYZED LAYER, SAMPLE PREPARATION, AND X-RAY BEAM PATH Because the energy of boron Kα radiation is so low (183 ev), only a thin 0.6 µm (600 nm) surface layer of the sample can be analyzed. The density and composition of the glass can, in some cases, reduce this down to 200 nm. The glass samples to be analyzed must therefore be prepared with a good surface finish. Mellott compiled and investigated the surface roughnesses of commercial glasses using Atomic Force Microscopy (AFM) and found that many are in the range from 2 to 0.2 nm [5]. Quantitative boron analysis on non-polished glasses was much less accurate and repeatable than on the same polished material (private communication in 2004 with Brian Simpkins, Alcan Packaging, formerly Wheaton Glass, Millville, New Jersey). Reducing the path length of the X-ray beam (tube-to-sample and sample-to-detector) and using thin-window flow proportional detectors, combined with a controlled vacuum of better than 0.1 mbar, enables higher yield of detected boron radiation. Comparing the published sensitivities from a 1998 dual-turret spectrometer with those from a direct-loading S4-type system using the same X-ray tube, the type of generator shows a substantial effect, attributable to the optimized beam path and window design [6 8]. Changes in the vacuum during the measurement affect the count rate of the detected boron and needs to be taken into consideration for those systems that have large sample compartments (e.g., dual-turret systems). In contrast, direct-loading systems can usually achieve a stable vacuum without changeover during the whole measurement period. These direct-loading systems feature pressure-controlled sample introduction. Measurement is started only when the pressure in the sample compartment reaches a certain level, resulting in constant pressure and higher, more stable count rates for boron. ANALYZER CRYSTALS Figure 2. Comparison of boron intensity collected with a traditional crystal versus multilayer crystal. Ricardo et al. showed in 2001 that with a new type of patented multilayer detector crystal, the boron signal can be raised by 28% [9]. The XS-B crystal depicted in Figure 2 is a LaB 4 C multilayer and is available from Incoatec GmbH (Gesthacht, Germany). Schuster et al. performed a feasibility study for the analysis of boron in a coating of BPSG, a common semiconductor material, in 1987 [10]. Using a WDXRF system equipped with a 125 µm endwindow X-ray tube, 80 ma tube current, and early multilayer crystal (OVO 160), he obtained a sensitivity of 2.7 counts/s per wt% B. The detection limit was reported to be 0.04 wt% for 600 s measurement time.

5 272 Using a 200 Å Mo-B 4 C multilayer crystal (OVO-B) and an S4 Series WDXRF system optimized for beam path, Mauser achieved a detection limit of 0.02 wt % B in 100 s and a sensitivity of 330 counts/s per wt% B on the same type of material [7]. Using the XS-B multilayer crystal on the same S4 instrument further improved the sensitivity to 420 counts/s per wt% B as was shown by Behrens in 2004 [11]. CHOICE OF COLLIMATORS AND BACKGROUND POSITIONS Multilayer crystals produce wide peak profiles and should be used with coarse collimators. The rule of thumb in WDXRF is to use the coarsest collimator for the lightest element detection. Since achieving a high count rate is paramount for the analysis, peak resolution and interference corrections are not an issue. This holds true for BPSG, but in order to create a meaningful widerrange calibration for commercial borosilicate and E glass as well, count rate needs to be balanced against resolution [12]. K Ll, Ca Ll Cl Ll, B KA1 O (3) Zr Mz Si LB/ S Ll Ln Ca Ca and K L l lines, as well as Ca Ln and Zr M lines, should be separated as shown in Figure 3. The effect of higher order reflections of O cannot be reduced much by using pulse height discrimination and needs to be minimized by collimation. With Ca and Zr present in the analyzed material, the selection of background positions becomes difficult. The approach from Feather and Willis is particularly useful in this regard, as it can be used to perform interference correction on the background [13]. CALIBRATION APPROACH < Y : 0 to KCps > Figure 3. Interference on and around the boron K signal. Obtaining a sufficient number of glass standards, as well as including enough samples to perform the interference corrections needed, pose the biggest challenge to the analyst. The quality of the boron certification by Manitol-based volumetric means is very much dependent on the experience of the analyst with that methodology. Few commercial labs now offer Manitol-based boron analysis. Inductively coupled plasma (ICP) methods are increasingly being offered instead, but these methods are time- and cost-prohibitive because of the requirement for sample digestion. Few glass samples of defined boron composition are commercially available. Many borosilicate glass samples, such as those available from Breitlaender GmbH (Hamm, Germany), are just premelt batch and not chemically certified. Borosilicate glass suppliers are the best source of sample material. They can produce a range of experimental matrix for the calibration based on past batch-melt samples. A comparison and discussion of the quality of certification techniques for boron in glass is unfortunately absent from the literature.

6 273 Using over 40 glass samples of standard 32 mm size and a small 28 mm aperture, a calibration for boron was established from 0.00 to wt%. The calibration has a standard error of calibration (SEC) of wt%. The sensitivity is 94 counts/s per wt% B 2 O 3. An empirical calculation using a concentration-based correction approach produced better results than any fundamental parameters (FP) model, even with variable influence coefficients. FP for boron, as taken from the NIST-compiled tables, are not well characterized and would need further improvements. Of greatest importance is interference correction of both the background and peak position to obtain a good net intensity for boron. MAINTENANCE OF CALIBRATION In order to maintain the calibration, drift correction is needed to account for changes in detector gas, foil, and tube signal that can influence the intensity and therefore the analysis. A very interesting study by Guadagnino, Sundberg, Michiels, and Brochot showed that some glass types exhibit a migration of boron when measured repeatedly [14]. Figure 4 shows the change in B 2 O 3 content in Pyrex-type and E-type glass as a function of the number of WDXRF measurements. Borosilicate glass should therefore be used as a drift correction and QC sample only after it has been conditioned. Figure 4. Long-term stability of boron in E-type and Pyrex-type glass

7 274 CONCLUSION The analysis of boron in glass using WDXRF has dramatically improved since its first application as a process control tool in the late 1980s. Nevertheless, it is still one of the most challenging applications, since the analyst must gather enough reference material and optimize conditions to fit the spectrometer capabilities. The choice of the right collimator, background and peak counting positions, correct pulse height analysis settings, and interference correction method greatly influences the range and durability of the resulting calibration. Maintenance of the calibration and a daily check of the system and calibration need to take the stability of the glass standards into account. ACKNOWLEDGMENTS The author thanks Brian Simpkins (Alcan Packaging, formerly Wheaton Glass, Millville, New Jersey) for his time performing the scans and tests on the borosilicate glass samples, as well as for evaluating the XS-B and OVO-B crystals side-by-side in a process environment. Many thanks as well to my colleagues at Bruker AXS Larry Arias, Kai Behrens, Karl Mauser, and Karen Roscoe for contributing to the presentation and article. REFERENCES [1] Bearden, J. A. Rev. Mod. Phys. 1967, 39, [2] Krause, M. O. J. Phys. Chem. Ref. Data 1979, 8, [3] White, E. W.; Johnson Jr., G. G. X-ray Emission and Absorption Wavelengths and Two- Theta Tables (ASTM Data Series DS 37A); American Society for Testing and Materials: Philadelphia, [4] Shiraiwa, T.; Fujino, N. Jpn. J. Appl. Phys. 1966, 5, [5] Mellott, N. P.; Brantley, S. L.; Hamilton, J. P.; Pantano, C. G. Surf. Interface Anal. 2001, 31, [6] Mauser, K. E.; Uhlig, S. SRS 3400 Boron in BPSG. Bruker AXS GmbH Application Note, [7] Mauser, K. E. S4 PIONEER Boron in BPSG. Bruker AXS GmbH Application Note, [8] Buehler, A.; Seyfarth, A.; Schlotz, R. Introduction to XRF. Bruker AXS GmbH Training Manual, [9] Ricardo, P.; Wiesmann, J.; Nowak, C.; Michaelsen, C.; Bormann, R. Appl. Opt. 2001, 40, [10] Schuster, M.; Müller, L.; Mauser, K. E.; Straub, R. Thin Solid Films 1988, 157, [11] Behrens, K. New Applications with the XS-B; S4 PIONEER. Bruker AXS GmbH Application Note, [12] Uhlig, S.; Müller, L. Boron Analysis in Ceramic and Glass Industries using Wavelength Dispersive X-ray Fluorescence Analysis. Siemens Analytical Application Notes 326, [13] Feather, C. E.; Willis, J. P. X-Ray Spectrom. 1976, 5, [14] Guadagnino, E.; Sundberg, P.; Michiels, D.; Brochot, D. A collaborative study into the determination of boron in glass using X-ray fluorescence (XRF) spectroscopy. ICG/TC2 report, 2006.

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS

ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner

More information

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system EDS system Most common X-Ray measurement system in the SEM lab. Major elements (10 wt% or greater) identified in ~10 secs. Minor elements identifiable in ~100 secs. Rapid qualitative and accurate quantitative

More information

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING Brian L. Riise and Michael B. Biddle MBA Polymers, Inc., Richmond, CA, USA Michael M. Fisher American Plastics Council, Arlington, VA, USA X-Ray Fluorescence

More information

X Ray Flourescence (XRF)

X Ray Flourescence (XRF) X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose

More information

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY 243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution

More information

Spectral distribution from end window X-ray tubes

Spectral distribution from end window X-ray tubes Copyright (C) JCPDS-International Centre for Diffraction Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg 2 1 FORTEX - E.N.S.A.I.S. 24, bld de la Victoire,

More information

APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS

APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 1 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 367

More information

Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist

Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist University of Puerto Rico Mayagüez Campus Department of Geology X-ray / Microanalysis Laboratory F - 304

More information

Elemental Analyses by ICP-AES

Elemental Analyses by ICP-AES Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to

More information

Electron Microprobe Analysis X-ray spectrometry:

Electron Microprobe Analysis X-ray spectrometry: Electron Microprobe Analysis X-ray spectrometry: 1. X-ray generation and emission 2. X-ray detection and measurement X-ray energy and wavelength E=hν h : Planck's constant (6.626x10-34 Joule.sec or, 6.626x10-34

More information

OLIVÉR BÁNHIDI 1. Introduction

OLIVÉR BÁNHIDI 1. Introduction Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY

More information

BRUKER ADVANCED X-RAY SOLUTIONS GUIDE TO XRF BASICS

BRUKER ADVANCED X-RAY SOLUTIONS GUIDE TO XRF BASICS BRUKER ADVANCED X-RAY SOLUTIONS GUIDE TO XRF BASICS This guide was first published in West Germany under the title Introduction to X-ray Fluorescence Analysis (XRF). 2000-2006 Bruker AXS GmbH, Karlruhe,

More information

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum

More information

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially

More information

Vacuum Evaporation Recap

Vacuum Evaporation Recap Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.

More information

EDXRF of Used Automotive Catalytic Converters

EDXRF of Used Automotive Catalytic Converters EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,

More information

Gold Refining and Coin Manufacturing at the Royal Canadian Mint

Gold Refining and Coin Manufacturing at the Royal Canadian Mint INTERNET Sub title will go here The Application of XRF to Gold Refining and Coin Manufacturing at the Royal Canadian Mint Michael W. Hinds, Ph.D. The Business of the Mint Circulation Coins (Canada & Other

More information

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using EDS spectrum imaging

Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using EDS spectrum imaging Quantitative analysis Ceramics sample Peak deconvolution EDS map Phase analysis Application Note # EDS-10 Advanced light element and low energy X-ray analysis of a TiB 2 TiC SiC ceramic material using

More information

XRF MAPPING: NEW TOOLS FOR DISTRIBUTION ANALYSIS

XRF MAPPING: NEW TOOLS FOR DISTRIBUTION ANALYSIS Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 19 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

h e l p s y o u C O N T R O L

h e l p s y o u C O N T R O L contamination analysis for compound semiconductors ANALYTICAL SERVICES B u r i e d d e f e c t s, E v a n s A n a l y t i c a l g r o u p h e l p s y o u C O N T R O L C O N T A M I N A T I O N Contamination

More information

RoHS Test Methods. Presented by: Bruce Peterson

RoHS Test Methods. Presented by: Bruce Peterson RoHS Test Methods Presented by: Bruce Peterson Motivation for Presentation The EU RoHS directive takes effect July 1, 2006 Customer confusion about - Scope of directive How much testing needs to be performed

More information

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications

More information

Glancing XRD and XRF for the Study of Texture Development in SmCo Based Films Sputtered Onto Silicon Substrates

Glancing XRD and XRF for the Study of Texture Development in SmCo Based Films Sputtered Onto Silicon Substrates 161 162 Glancing XRD and XRF for the Study of Texture Development in SmCo Based Films Sputtered Onto Silicon Substrates F. J. Cadieu*, I. Vander, Y. Rong, and R. W. Zuneska Physics Department Queens College

More information

ANALYSIS OF LEAD IN CANDLE PARTICULATE EMISSIONS BY XRF USING UNIQUANT 4

ANALYSIS OF LEAD IN CANDLE PARTICULATE EMISSIONS BY XRF USING UNIQUANT 4 Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 539 ANALYSIS OF LEAD IN CANDLE PARTICULATE EMISSIONS BY XRF USING UNIQUANT 4 Shirley J. Wasson

More information

Introduction to Energy Dispersive X-ray Spectrometry (EDS)

Introduction to Energy Dispersive X-ray Spectrometry (EDS) Introduction to Energy Dispersive X-ray Spectrometry (EDS) 1. Introduction 1.1 Principles of the technique EDS makes use of the X-ray spectrum emitted by a solid sample bombarded with a focused beam of

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

13C NMR Spectroscopy

13C NMR Spectroscopy 13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number

More information

Raman spectroscopy Lecture

Raman spectroscopy Lecture Raman spectroscopy Lecture Licentiate course in measurement science and technology Spring 2008 10.04.2008 Antti Kivioja Contents - Introduction - What is Raman spectroscopy? - The theory of Raman spectroscopy

More information

X-Ray Study of Soft and Hard Magnetic Thin Films

X-Ray Study of Soft and Hard Magnetic Thin Films Copyright (C) JCPDS-International Centre for Diffraction Data 1999 13 X-Ray Study of Soft and Hard Magnetic Thin Films Po-Wen Wang, 390 Reed St., Stormedia, Inc., Santa Clara CA. 95050 Abstract : This

More information

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse SURFACE CHARACTERIZATION Marie-Geneviève Centre d Etudes de Chimie Métallurgique, CNRS, Vitry-sur-Seine, France Keywords: Surface Analysis, Surface imaging, Surface composition, Surface chemical analysis,

More information

Lectures about XRF (X-Ray Fluorescence)

Lectures about XRF (X-Ray Fluorescence) 1 / 38 Lectures about XRF (X-Ray Fluorescence) Advanced Physics Laboratory Laurea Magistrale in Fisica year 2013 - Camerino 2 / 38 X-ray Fluorescence XRF is an acronym for X-Ray Fluorescence. The XRF technique

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

Raman and AFM characterization of carbon nanotube polymer composites Illia Dobryden

Raman and AFM characterization of carbon nanotube polymer composites Illia Dobryden Raman and AFM characterization of carbon nanotube polymer composites Illia Dobryden This project is conducted in High Pressure Spectroscopy Laboratory (Materials Physics group) Supervisor: Professor Alexander

More information

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS 1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years

More information

Physics 441/2: Transmission Electron Microscope

Physics 441/2: Transmission Electron Microscope Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This

More information

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the

More information

DATA MINING WITH DIFFERENT TYPES OF X-RAY DATA

DATA MINING WITH DIFFERENT TYPES OF X-RAY DATA 315 DATA MINING WITH DIFFERENT TYPES OF X-RAY DATA C. K. Lowe-Ma, A. E. Chen, D. Scholl Physical & Environmental Sciences, Research and Advanced Engineering Ford Motor Company, Dearborn, Michigan, USA

More information

How to measure absolute pressure using piezoresistive sensing elements

How to measure absolute pressure using piezoresistive sensing elements In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article

More information

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

Field portable XRF analysis of environmental samples

Field portable XRF analysis of environmental samples Journal of Hazardous Materials 83 (2001) 93 122 Field portable XRF analysis of environmental samples Dennis J. Kalnicky a, Raj Singhvi b, a Lockheed Martin Technology Services Group, Environmental Services/REAC,

More information

Energy Dispersive Spectroscopy on the SEM: A Primer

Energy Dispersive Spectroscopy on the SEM: A Primer Energy Dispersive Spectroscopy on the SEM: A Primer Bob Hafner This primer is intended as background for the EDS Analysis on the SEM course offered by the University of Minnesota s Characterization Facility.

More information

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS

IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS IUCLID 5 COMPOSITION AND ANALYSIS GUIDANCE DOCUMENT: IRON ORES, AGGLOMERATES [EINECS NUMBER 265 996 3, CAS NUMBER 65996 65 8] IRON ORE PELLETS INTRODUCTION Each REACH registrant is required to file its

More information

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction

More information

Quantifying Amorphous Phases. Kern, A., Madsen, I.C. and Scarlett, N.V.Y.

Quantifying Amorphous Phases. Kern, A., Madsen, I.C. and Scarlett, N.V.Y. Kern, A., Madsen, I.C. and Scarlett, N.V.Y. This document was presented at PPXRD - Pharmaceutical Powder X-ray Diffraction Symposium Sponsored by The International Centre for Diffraction Data This presentation

More information

THEORY OF XRF. Getting acquainted with the principles. Peter Brouwer

THEORY OF XRF. Getting acquainted with the principles. Peter Brouwer THEORY OF XRF Getting acquainted with the principles Peter Brouwer THEORY OF XRF Getting acquainted with the principles Peter Brouwer First published in The Netherlands under the title Theory of XRF. Copyright

More information

Analytical Testing Services Commercial Price List ManTech International Corporation January 2016

Analytical Testing Services Commercial Price List ManTech International Corporation January 2016 Analytical ing Services Commercial List ManTech International Corporation January 2016 TABLE OF CONTENTS MECHANICAL TENSILE TESTING... 1 DIFFERENTIAL SCANNING CALORIMETRY (DSC)... 2 THERMOMECHANICAL ANALYSIS

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

Copyright 1999 2010 by Mark Brandt, Ph.D. 12

Copyright 1999 2010 by Mark Brandt, Ph.D. 12 Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

Using the PDF for material identification using elemental data. from XRF and SEM EDS.

Using the PDF for material identification using elemental data. from XRF and SEM EDS. XRF and SEM EDS Using the PDF for material identification using elemental data from XRF and SEM EDS. XRF and SEM EDS What? The Powder Diffraction File contains data on pure solid state compounds of well

More information

- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier

- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier Tunnel Effect: - particle with kinetic energy E strikes a barrier with height U 0 > E and width L - classically the particle cannot overcome the barrier - quantum mechanically the particle can penetrated

More information

Instrumentation. (Figure 2)

Instrumentation. (Figure 2) X-Ray Fluorescence Lab Report Nydia Esparza Victoria Rangel Physics of XRF XRF is a non destructive analytical technique that is used for elemental and chemical analysis. X-Ray Fluorescence Spectroscopy

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

How To Understand Light And Color

How To Understand Light And Color PRACTICE EXAM IV P202 SPRING 2004 1. In two separate double slit experiments, an interference pattern is observed on a screen. In the first experiment, violet light (λ = 754 nm) is used and a second-order

More information

Neuere Entwicklungen zur Herstellung optischer Schichten durch reaktive. Wolfgang Hentsch, Dr. Reinhard Fendler. FHR Anlagenbau GmbH

Neuere Entwicklungen zur Herstellung optischer Schichten durch reaktive. Wolfgang Hentsch, Dr. Reinhard Fendler. FHR Anlagenbau GmbH Neuere Entwicklungen zur Herstellung optischer Schichten durch reaktive Sputtertechnologien Wolfgang Hentsch, Dr. Reinhard Fendler FHR Anlagenbau GmbH Germany Contents: 1. FHR Anlagenbau GmbH in Brief

More information

Solid State Detectors = Semi-Conductor based Detectors

Solid State Detectors = Semi-Conductor based Detectors Solid State Detectors = Semi-Conductor based Detectors Materials and their properties Energy bands and electronic structure Charge transport and conductivity Boundaries: the p-n junction Charge collection

More information

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün N.B. A shorter version of this article was published in the ICOM-CC Ethnographic Conservation Newsletter, Number 30, January 2009, pp. 5-8. The Limitations of Hand-held XRF Analyzers as a Quantitative

More information

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.

More information

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical

More information

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control

More information

Coating Technology: Evaporation Vs Sputtering

Coating Technology: Evaporation Vs Sputtering Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information

More information

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF) Advanced Physics Laboratory XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF) Bahia Arezki Contents 1. INTRODUCTION... 2 2. FUNDAMENTALS... 2 2.1 X-RAY PRODUCTION... 2 2. 1. 1 Continuous radiation...

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY I. BIBICU 1, G. NICOLESCU 2, L. CIOLACU 2, L. SERBINA 2 1 National Institute for Materials Physics, Bucharest 77125, Romania, bibicu@infim.ro 2

More information

The Physics of Energy sources Renewable sources of energy. Solar Energy

The Physics of Energy sources Renewable sources of energy. Solar Energy The Physics of Energy sources Renewable sources of energy Solar Energy B. Maffei Bruno.maffei@manchester.ac.uk Renewable sources 1 Solar power! There are basically two ways of using directly the radiative

More information

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and

Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Lecture 9, Thermal Notes, 3.054

Lecture 9, Thermal Notes, 3.054 Lecture 9, Thermal Notes, 3.054 Thermal Properties of Foams Closed cell foams widely used for thermal insulation Only materials with lower conductivity are aerogels (tend to be brittle and weak) and vacuum

More information

The Basics of Scanning Electron Microscopy

The Basics of Scanning Electron Microscopy The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single

More information

WAVES AND ELECTROMAGNETIC RADIATION

WAVES AND ELECTROMAGNETIC RADIATION WAVES AND ELECTROMAGNETIC RADIATION All waves are characterized by their wavelength, frequency and speed. Wavelength (lambda, ): the distance between any 2 successive crests or troughs. Frequency (nu,):

More information

Acousto-optic modulator

Acousto-optic modulator 1 of 3 Acousto-optic modulator F An acousto-optic modulator (AOM), also called a Bragg cell, uses the acousto-optic effect to diffract and shift the frequency of light using sound waves (usually at radio-frequency).

More information

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF Quantitative Analysis Software for X-Ray Fluorescence XRF-FP XRF-FP is a full-featured quantitative analysis package for XRF APPLICATIONS X-Ray Fluorescence Thin-film Analysis RoHS/WEEE Analysis Teaching

More information

Introduction to Elemental Analysis by ED-XRF Justin Masone Product Specialist 3 June 2015

Introduction to Elemental Analysis by ED-XRF Justin Masone Product Specialist 3 June 2015 Introduction to Elemental Analysis by EDXRF Justin Masone Product Specialist 3 June 2015 1 / 9 Shimadzu Corporation Established in 1875. Headquartered in Kyoto, Japan Ranked Top 5 Instrument Providers

More information

2. Deposition process

2. Deposition process Properties of optical thin films produced by reactive low voltage ion plating (RLVIP) Antje Hallbauer Thin Film Technology Institute of Ion Physics & Applied Physics University of Innsbruck Investigations

More information

X-ray Techniques: Overview. Ron Jenkins

X-ray Techniques: Overview. Ron Jenkins X-ray Techniques: Overview Ron Jenkins in Encyclopedia of Analytical Chemistry R.A. Meyers (Ed.) pp. 13269 13288 John Wiley & Sons Ltd, Chichester, 2000 X-RAY TECHNIQUES: OVERVIEW 1 X-ray Techniques:

More information

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly

More information

Laser sintering of greens compacts of MoSi 2

Laser sintering of greens compacts of MoSi 2 Laser sintering of greens compacts of MoSi 2 G. de Vasconcelos 1, R. Cesar Maia 2, C.A.A.Cairo 3, R. Riva 2, N.A.S.Rodrigues 2, F.C.L.Mello 3 Instituto de Estudos Avançados 1, Instituto Tecnológico de

More information

Oxford Instruments Analytical technical briefing. Wavelength Dispersive X-ray Microanalysis

Oxford Instruments Analytical technical briefing. Wavelength Dispersive X-ray Microanalysis Oxford Instruments Analytical technical briefing Wavelength Dispersive X-ray Microanalysis Oxford Instruments Analytical technical briefing Introduction Electron probe X-ray microanalysis techniques (Wavelength

More information

Preface Light Microscopy X-ray Diffraction Methods

Preface Light Microscopy X-ray Diffraction Methods Preface xi 1 Light Microscopy 1 1.1 Optical Principles 1 1.1.1 Image Formation 1 1.1.2 Resolution 3 1.1.3 Depth of Field 5 1.1.4 Aberrations 6 1.2 Instrumentation 8 1.2.1 Illumination System 9 1.2.2 Objective

More information

Introduction to X-Ray Powder Diffraction Data Analysis

Introduction to X-Ray Powder Diffraction Data Analysis Introduction to X-Ray Powder Diffraction Data Analysis Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray An X-ray diffraction pattern is a plot of the intensity of X-rays scattered

More information

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES

Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES ICP OPTICAL ATOMIC EMISSION SPECTROSCOPY Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES HORIBA Scientific Longjumeau, France Keywords: metallurgy, non-ferrous, halogens 1 Introduction

More information

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT speakman@mit.edu

More information

CHEM 1411 Chapter 5 Homework Answers

CHEM 1411 Chapter 5 Homework Answers 1 CHEM 1411 Chapter 5 Homework Answers 1. Which statement regarding the gold foil experiment is false? (a) It was performed by Rutherford and his research group early in the 20 th century. (b) Most of

More information

1. Photon Beam Damage and Charging at Solid Surfaces John H. Thomas III

1. Photon Beam Damage and Charging at Solid Surfaces John H. Thomas III 1. Photon Beam Damage and Charging at Solid Surfaces John H. Thomas III 1. Introduction............................. 2. Electrostatic Charging of Samples in Photoemission Experiments............................

More information

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale

Nano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale Nano-Spectroscopy Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale Since its introduction in the early 80 s, Scanning Probe Microscopy (SPM) has quickly made nanoscale imaging an affordable

More information

5.33 Lecture Notes: Introduction to Spectroscopy

5.33 Lecture Notes: Introduction to Spectroscopy 5.33 Lecture Notes: ntroduction to Spectroscopy What is spectroscopy? Studying the properties of matter through its interaction with different frequency components of the electromagnetic spectrum. Latin:

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

The study of structural and optical properties of TiO 2 :Tb thin films

The study of structural and optical properties of TiO 2 :Tb thin films Optica Applicata, Vol. XXXVII, No. 4, 2007 The study of structural and optical properties of TiO 2 :Tb thin films AGNIESZKA BORKOWSKA, JAROSLAW DOMARADZKI, DANUTA KACZMAREK, DAMIAN WOJCIESZAK Faculty of

More information

Experiment: Crystal Structure Analysis in Engineering Materials

Experiment: Crystal Structure Analysis in Engineering Materials Experiment: Crystal Structure Analysis in Engineering Materials Objective The purpose of this experiment is to introduce students to the use of X-ray diffraction techniques for investigating various types

More information

Back to Basics Fundamentals of Polymer Analysis

Back to Basics Fundamentals of Polymer Analysis Back to Basics Fundamentals of Polymer Analysis Using Infrared & Raman Spectroscopy Molecular Spectroscopy in the Polymer Manufacturing Process Process NIR NIR Production Receiving Shipping QC R&D Routine

More information

New Portable X-Ray Diffraction/X-Ray Fluorescence Instrument (XRD/XRF)

New Portable X-Ray Diffraction/X-Ray Fluorescence Instrument (XRD/XRF) 1 New Portable X-Ray Diffraction/X-Ray Fluorescence Instrument (XRD/XRF) Introduction The primary goal of most analyses of art objects is to identify the material composition of the object. This may lead

More information

For Touch Panel and LCD Sputtering/PECVD/ Wet Processing

For Touch Panel and LCD Sputtering/PECVD/ Wet Processing production Systems For Touch Panel and LCD Sputtering/PECVD/ Wet Processing Pilot and Production Systems Process Solutions with over 20 Years of Know-how Process Technology at a Glance for Touch Panel,

More information

DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY

DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY Eduard Gilli 1,2 and Robert Schennach 1, 2 1 Graz University of Technology, 8010 Graz, Austria 2 CD-Laboratory for Surface Chemical and Physical

More information