1 FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: 1. IR Instrumentation Two types of instrumentation are used to obtain infrared spectra: Dispersive spectrophotometers, which use a monochromator to produce an infrared spectrum one resolution element at a time. Michelson interferometers, which use a moving mirror to create an interference pattern, or interferogram, from which all resolution elements are determined simultaneously. 2. FTIR Spectroscopy History Chemical infrared spectroscopy emerged as a science in the 1880s. In the 1890s, A. A. Michelson, to further his studies of the speed of light, invented the interferometer. In the early 1940s, chemical infrared spectroscopy was still an immature scientific field. But with commercial development of the optical null dispersive spectrophotometer, later that decade, chemical infrared spectroscopy came into widespread use. Dispersive instruments proved the tremendous value of infrared analysis, and soon became the mainstay of organic characterization laboratories. In 1949 astrophysicist Peter Fellgett used an interferometer to measure light from celestial bodies and produced the first Fourier transform infrared spectrum. But for many years, only a few advanced research groups with access to large, expensive computers and with personnel able to wait up to 12 hours to transform an interferogram into a spectrum used Fourier transform infrared (FTIR) spectroscopy. For two decades more, FTIR spectrometers were limited to studying problems not solvable with dispersive techniques. In the late 1960s when microcomputers able to do the Fourier transform became available, commercial FTIR spectrometers appeared. The 1966 development of the Cooley-Tukey algorithm, which quickly does a Fourier transform (the Fast Fourier Transform or FFT), was also instrumental in the commercialization of FTIR spectrometers. However, the first FTIR spectrometers were large and expensive, and were found primarily in a few well-to-do research labs. Gradually, technology reduced the cost, increased the availability, and enhanced the capabilities of FTIR spectroscopy systems. The performance-to-price ratio provided by FTIR spectrometers today was unthinkable only a decade ago. 1

2 3. Theory of Interferometer Operation The optical system in an FTIR spectrometer is very simple: the interferometer requires two mirrors, an infrared light source, an infrared detector, and a beamsplitter. fixed mirror beam splitter collimator moving mirror detector source focussing mirror Figure 1. The Michelson interferometer The beamsplitter is the heart of the interferometer. Essentially a half-silvered mirror, the beamsplitter reflects about half of an incident light beam while simultaneously transmitting the remaining half. One half of this split light beam travels to the interferometer's moving mirror while the other half travels to the interferometer's stationary mirror. The two mirrors reflect both beams back to the beamsplitter where each of the two beams is again half reflected and half transmitted. Two output beams result: one travels to the detector as the other travels to the source. When the two beams return to the beamsplitter, an interference pattern, or interferogram, is generated. This interference pattern varies with the displacement of the moving mirror, that is, with the difference in pathlength in the two arms of the interferometer. The interference pattern, detected by the infrared detector as variations in the infrared energy level, is what ultimately yields spectral information. 4. More Interferometer Theory An example of how light is affected in an interferometer further illustrates interferometer operation. Consider the interference pattern generated when 2

3 monochromatic light, such as that produced by a helium-neon (HeNe) laser, is the light source. An idealized beamsplitter: - neither generates nor absorbs radiation. - transmits half of all incident light. - reflects half of all incident light. All light energy that enters the interferometer from the source either returns back to the source or travels to the infrared detector. As the moving mirror scans back and forth, more or less than half of the total light energy that originally entered the interferometer reaches the detector. All light that does not reach the detector reaches the source instead. For example, if both interferometer mirrors are equidistant from the beamsplitter, the two light beams travel equal distances. As each beam returns to the beamsplitter, each is again split and then recombined with half of the light from the other interferometer arm. The two component beams that make up the output beam passed to the detector are completely in phase with each other and, therefore, interfere constructively. The two component beams that make up the output beam passed to the source are completely out of phase with each other and, therefore, interfere destructively. In this case, all of the original light energy goes to the detector and none goes to the source. By contrast, if the moving mirror is displaced one-quarter wavelength of the HeNe laser radiation (to produce optical retardation of one-half wavelength), light returning from the interferometer's moving mirror reaches the beamsplitter one-half wavelength out of phase with respect to the light returning from the fixed mirror. This results in complete destructive interference in the detected output beam so no light energy reaches the detector, and all light returns to the source. Figure 2. Constructive (left) and destructive (right) interference. As the moving mirror scans back and forth, the detector observes light and dark bands corresponding to partial and total constructive and destructive interference in the detected output beam. In this monochromatic light example, the interference signal the detector receives produces a cosine interferogram. This wave goes from full intensity at the point where both mirrors are equidistant from the beamsplitter to zero intensity when the moving mirror has traveled one-quarter wavelength. Full intensity is again achieved when the moving mirror has traveled one-half or multiples of one-half wavelength. Similarly, zero intensity occurs at each multiple of one-quarter wavelength of mirror travel. 3

4 An infrared source is polychromatic, emitting light over a broad range of frequencies. Each frequency still yields a unique cosine signal: the resulting interferogram represents the sum of all cosine waves generated by each individual infrared frequency. However, only at the point where both mirrors are equal distances from the beamsplitter does complete constructive interference occur--simultaneous constructive interference of all wavelengths. At any other point, only some wavelengths interfere constructively while others do not. The Fourier transform resolves the frequency and intensity of each cosine wave in the interferogram. In other words, this algorithm converts the measured.intensity-versusmirror-displacement signal (the interferogram) into a plot of intensity versus frequency (a spectrum). Δx Figure 3. Spectrum with three frequencies with different amplitudes (top), interferograms for the three separate frequencies (middle), and Interferograms for the three mixed frequencies. Δx Δx = 0 5. Advantage 5.1 Multiplex Advantage (Fellgett's Advantage) An interferometer does not separate light into individual frequencies before measurement. This means each point in the interferogram contains information from each wavelength in the input light. In other words, if 8,000 data points along the interferogram are collected, each wavelength in the input light is sampled 8,000 times. By contrast, a dispersive spectrophotometer that measures 8,000 individual points across a spectrum samples each wavelength only once. 5.2 Throughput Advantage (Jacquinot's Advantage) The simple optical path of the interferometer (no slits and fewer optical elements) means more energy gets to the sample than is possible with dispersive 4

5 spectrophotometers. This means more energy reaches the detector, increasing the spectrum's potential signal-to-noise ratio. Together the multiplex and throughput advantages allow an FTIR spectrometer to obtain a high-quality infrared spectrum in a fraction of the time needed to get the same spectrum on a dispersive instrument. Also, to increase the resolution of dispersive instruments, the slits through which light must pass are narrowed, thereby decreasing energy throughput. In an FTIR spectrometer, resolution is increased by lengthening the moving mirror stroke length with no decrease in energy throughput. As wavelength resolution increases, the advantages of interferometric versus dispersive measurements increase. 5.3 Frequency Precision (Conne's Advantage) With dispersive instruments, frequency precision and accuracy depend on: 1) calibration with external standards and 2) the ability of electromechanical mechanisms to uniformly move gratings and slits during and between scans. By contrast, the interferometer has an internal frequency standard, generally a heliumneon laser. Also, both mirror movement and detector sampling are clocked by the interferometer fringes from the laser's monochromatic light. All frequencies in the output spectrum are calculated from the known frequency of the laser light. An FTIR spectrometer easily achieves frequency precision and accuracy of better than 0.01 wavenumbers. This means that spectra collected with an FTIR spectrometer can be quantitatively compared whether they were collected five minutes or five years apart. 5.4 Constant Spectral Resolution In an FTIR spectrometer, the resolution of the measured spectrum is the same for all frequencies, not varied throughout the spectrum as is often true with dispersive instruments. 6. Resolution In a dispersive spectrophotometer, resolution depends on slit width and on the reciprocal of the grating's linear displacement. As the slits are narrowed, the range of frequencies passed to the detector decreases, resulting in higher resolution. In an FTIR spectrometer, resolution is increased by increasing optical retardation, that is, by increasing how far the moving mirror moves. A relationship between resolution and optical retardation can be empirically derived by considering a spectrum made up of two sharp bands. Each band is at a distinct frequency and generates a sinusoidal signal. The total signal at the detector is the sum of the two signals. If the two spectral frequencies are relatively far apart, the resulting signals are quite different and only a short movement of the mirror is needed to measure a complete 5

6 cycle of the pattern made by the addition of the two waveforms. On the other hand, if the two frequencies are closer together, the interferometer needs to scan a larger distance to measure a full cycle of the addition pattern. The relationship between spectral line spacing and the interferogram is R = 1/d where R is the spectral resolution in wavenumbers and d is the optical retardation in centimeters. Thus, the length of mirror travel (retardation) must be doubled to reduce instrument resolution by half, or, in other words, to improve instrument resolution by a factor of two. Several different criteria are used to measure FTIR spectrometer resolution. These include the Rayleigh criterion and the full-width-at-half-height (FWHH) method. Measuring instrument resolution with the FWHH method involves collecting an absorption spectrum with very sharp peaks, such as that obtained from low-pressure CO at about 5 torr pressure. fixed mirror electronics beam splitter moving mirror source aperture detector sample Figure 4: Left: Outer view of the Bruker IFS 66v/S FTIR spectrometer. (A) Sample chamber, (B) Interferometer chamber, (C) Laser compartment, (D) Electronic compartment and vacuum control, (E) Detector chamber. Right: Schematic internal view. Bold lines mark the optical pathway. 6

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

Fourier Transform Infrared (FTIR) Spectroscopy

Fourier Transform Infrared (FTIR) Spectroscopy 4/8/11 Fourier Transform Infrared (FTIR) Spectroscopy Basis of many IR satellite remote sensing instruments Ground-based FTIR spectrometers also used for various applications Interference Consequence of

More information

SPECTROPHOTOMETRY. PRINCIPLE AND APPLICATIONS

SPECTROPHOTOMETRY. PRINCIPLE AND APPLICATIONS EUROPEAN UNION GOVERNMENT OF ROMANIA GOVERNMENT OF THE REPUBLIC OF SERBIA Structural Funds 2007-2013 SPECTROPHOTOMETRY. PRINCIPLE AND APPLICATIONS As.dr.ing. ADRIAN EUGEN CIOABLA WORKSHOP 05 06 September

More information

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic

More information

Components for Infrared Spectroscopy. Dispersive IR Spectroscopy

Components for Infrared Spectroscopy. Dispersive IR Spectroscopy Components for Infrared Spectroscopy Mid-IR light: 00-000 cm - (5.5 m wavelength) Sources: Blackbody emitters Globar metal oxides Nernst Glower: Silicon Carbide Detectors: Not enough energy for photoelectric

More information

Physical Properties Analysis of Petroleum Products Using FT-NIR

Physical Properties Analysis of Petroleum Products Using FT-NIR Physical Properties Analysis of Petroleum Products Using FT-NIR Darrell Kowalyk, A.Sc.T. CB Engineering LTD 515, 9945 50 Street Edmonton, AB, CA 1. Introduction Petroleum products have specifications based

More information

ORIEL. FT-IR Spectroscopy SECTION FOUR FEATURES. Glossary of Terms Introduction to FT-IR Spectroscopy

ORIEL. FT-IR Spectroscopy SECTION FOUR FEATURES. Glossary of Terms Introduction to FT-IR Spectroscopy ORIEL ORIEL PRODUCT TRAINING FT-IR Spectroscopy SECTION FOUR FEATURES Glossary of Terms Introduction to FT-IR Spectroscopy Stratford, CT Toll Free 800.74.5393 Fax 03.378.457 www.newport.com/oriel oriel.sales@newport.com

More information

Lecture PowerPoints. Chapter 24 Physics: Principles with Applications, 7th edition Giancoli

Lecture PowerPoints. Chapter 24 Physics: Principles with Applications, 7th edition Giancoli Lecture PowerPoints Chapter 24 Physics: Principles with Applications, 7th edition Giancoli This work is protected by United States copyright laws and is provided solely for the use of instructors in teaching

More information

10Spec TM 10 Degree Specular Reflectance Accessory

10Spec TM 10 Degree Specular Reflectance Accessory Installation and User Guide 10Spec TM 10 Degree Specular Reflectance Accessory The information in this publication is provided for reference only. All information contained in this publication is believed

More information

Chapter Four: Interference

Chapter Four: Interference Chapter Four Interference CHAPTER OUTLINE 4.1 Superposition of Waves 4.2 Interference 4.2.1Theory of Interference 4.2.2Intensity Distribution 4.2.3Superposition of Incoherent Waves 4.2.4Superposition of

More information

Types of Spectroscopy

Types of Spectroscopy Introduction to Infrared and Raman spectroscopy Arto Koistinen UEF / SIB-labs Optical microscopy course 2012 31.10.2012 Types of Spectroscopy = study of interaction between radiation and matter X-ray photoelectron

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

III. Light is a Wave (Physical Optics)

III. Light is a Wave (Physical Optics) III. Light is a Wave (Physical Optics) III.D. Interferometers. Michelson Interferometer Optical interferometers are combinations of partially and fully reflecting s, and sometimes es, that form an interference

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

Young s Double Slit Experiment

Young s Double Slit Experiment Young s Double Slit Experiment Apparatus optics bench laser slit film screen white paper and tape pencil metric ruler Ocean Optics spectrometer and fiber optics cable Goal In this experiment, you will

More information

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT

OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (07) 12 DEF CORR QUALIFICATION OF EQUIPMENT ANNEX 4: QUALIFICATION OF IR SPECTROPHOTOMETERS Full document title and reference

More information

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm?

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm? Test IV Name 1) In a single slit diffraction experiment, the width of the slit is 3.1 10-5 m and the distance from the slit to the screen is 2.2 m. If the beam of light of wavelength 600 nm passes through

More information

Chapter 24. Wave Optics

Chapter 24. Wave Optics Chapter 24 Wave Optics Wave Optics The wave nature of light is needed to explain various phenomena. Interference Diffraction Polarization The particle nature of light was the basis for ray (geometric)

More information

Ch 24 Wave Optics. concept questions #8, 11 problems #1, 3, 9, 15, 19, 31, 45, 48, 53

Ch 24 Wave Optics. concept questions #8, 11 problems #1, 3, 9, 15, 19, 31, 45, 48, 53 Ch 24 Wave Optics concept questions #8, 11 problems #1, 3, 9, 15, 19, 31, 45, 48, 53 Light is a wave so interference can occur. Interference effects for light are not easy to observe because of the short

More information

CHAPTER 12 INFRARED SPECTROSCOPY. and MASS SPECTROSCOPY

CHAPTER 12 INFRARED SPECTROSCOPY. and MASS SPECTROSCOPY KOT 222 ORGANIC CHEMISTRY II CHAPTER 12 INFRARED SPECTROSCOPY and MASS SPECTROSCOPY Part I Infrared Spectroscopy What is Spectroscopy? Spectroscopy is the study of the interaction of matter and electromagnetic

More information

Physics 122 Class #10 Outline. Announcements Interference of light waves Double slit Diffraction Grating Single slit Interferometer

Physics 122 Class #10 Outline. Announcements Interference of light waves Double slit Diffraction Grating Single slit Interferometer Physics 122 Class #10 Outline Announcements Interference of light waves Double slit Diffraction Grating Single slit Interferometer Reading Next Week ALL of Chapter 25 It is key to rest of course. Main

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

Answer: b. Answer: a. Answer: d

Answer: b. Answer: a. Answer: d Practice Test IV Name 1) In a single slit diffraction experiment, the width of the slit is 3.1 10-5 m and the distance from the slit to the screen is 2.2 m. If the beam of light of wavelength 600 nm passes

More information

CAUTION Laser radiation can cause retinal damage and blindness if allowed to be focused into the eye

CAUTION Laser radiation can cause retinal damage and blindness if allowed to be focused into the eye Interference and Diffraction of Light When two or more waves overlap at some point, they can add together so that the combined amplitude could be either greater or less than the amplitudes of the constituent

More information

Unit 2 Particles and Waves

Unit 2 Particles and Waves North Berwick High School Department of Physics Higher Physics Unit 2 Particles and Waves Section 5 Interference and Diffraction Section 5 Interference and Diffraction Note Making Make a dictionary with

More information

Demonstration Experiments: The Grating Spectrometer 27 September 2012

Demonstration Experiments: The Grating Spectrometer 27 September 2012 8 The Grating Spectrometer Introduction: A spectrometer is an instrument used to study the spectrum of light. The light source can be anything from a cryogenically cooled crystal to a superhot plasma or

More information

Constructive vs. destructive interference; Coherent vs. incoherent interference

Constructive vs. destructive interference; Coherent vs. incoherent interference Constructive vs. destructive interference; Coherent vs. incoherent interference Waves that combine in phase add up to relatively high irradiance. = Constructive interference (coherent) Waves that combine

More information

Module17:Coherence Lecture 17: Coherence

Module17:Coherence Lecture 17: Coherence Module7:Coherence Lecture 7: Coherence We shall separately discuss spatial coherence and temporal coherence. 7. Spatial Coherence The Young s double slit experiment (Figure 7.) essentially measures the

More information

Introduction. Chapter 12 Mass Spectrometry and Infrared Spectroscopy. Electromagnetic Spectrum. Types of Spectroscopy 8/29/2011

Introduction. Chapter 12 Mass Spectrometry and Infrared Spectroscopy. Electromagnetic Spectrum. Types of Spectroscopy 8/29/2011 Organic Chemistry, 6 th Edition L. G. Wade, Jr. Chapter 12 Mass Spectrometry and Infrared Spectroscopy Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys

More information

Standardization of FT-IR instruments

Standardization of FT-IR instruments A White Paper from FOSS Standardization of FT-IR instruments Using FOSS patented standardization procedure it is possible to reduce the need of Slope and Intercept adjustment and further increase the precision

More information

Info Note UV-VIS Nomenclature and Units

Info Note UV-VIS Nomenclature and Units Info Note 804: Ultraviolet-visible spectroscopy or ultraviolet-visible spectrophotometry (UV/VIS) involves the spectroscopy of photons in the UV-visible region. It uses light in the visible and adjacent

More information

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1) Interferometers The true worth of an experimenter consists in his pursuing not only what he seeks in his experiment, but also what he did not seek. Claude Bernard (1813-1878) OBJECTIVES To examine the

More information

Wave Optics. b. the crest from one wave overlaps with the d. darkness cannot occur as the two waves are coherent.

Wave Optics. b. the crest from one wave overlaps with the d. darkness cannot occur as the two waves are coherent. Wave Optics 1. Two beams of coherent light are shining on the same piece of white paper. With respect to the crests and troughs of such waves, darkness will occur on the paper where: a. the crest from

More information

EM Waves Practice Problems

EM Waves Practice Problems EM Waves Practice Problems PSI AP Physics B Name Multiple Choice 1. Which of the following theories can explain the bending of waves behind obstacles into shadow region? (A) Particle theory of light (B)

More information

Lab 7: Fabry-Perot Interferometer

Lab 7: Fabry-Perot Interferometer Lab 7: Fabry-Perot Interferometer 1 Introduction Refer to Appendix D for photos of the apparatus A Fabry-Perot interferometer is a device that uses multiple beam interference of light for high resolution

More information

Problem Set 3 Solutions CH332 (SP 06) 1. Skoog problem 15-1 (omit terms (j), (k) and (m)). Draw diagrams as necessary.

Problem Set 3 Solutions CH332 (SP 06) 1. Skoog problem 15-1 (omit terms (j), (k) and (m)). Draw diagrams as necessary. Problem Set 3 Solutions CH332 (SP 06) 1. Skoog problem 15-1 (omit terms (j), (k) and (m)). Draw diagrams as necessary. a) fluorescence Relaxation of an excited state by emission of a photon without a change

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

Purpose of the experiment

Purpose of the experiment Modern Physics Lab Spectroscopy Purpose of the experiment Familiarize you with advanced experimental techniques and equipment. Learn how to identify various elements by their emission spectrum. Background

More information

Accessories for Oriel MIR8025 FT Spectrometers

Accessories for Oriel MIR8025 FT Spectrometers 1332 MONOCHROMATORS Accessories for Oriel MIR8025 FT Spectrometers Versatile, modular accessories Convenient bench holding/coupling accessory Fiber coupling accessories simplifying beam collection and

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

ENGINEERING METROLOGY

ENGINEERING METROLOGY ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring

More information

PHYS General Physics II Lab Diffraction Grating

PHYS General Physics II Lab Diffraction Grating 1 PHYS 1040 - General Physics II Lab Diffraction Grating In this lab you will perform an experiment to understand the interference of light waves when they pass through a diffraction grating and to determine

More information

DIFFRACTION GRATINGS AND SPECTROSCOPY

DIFFRACTION GRATINGS AND SPECTROSCOPY Experiment 8 Name: S.N.: SECTION: PARTNER: DATE: DIFFRACTION GRATINGS AND SPECTROSCOPY Objectives To introduce and calibrate a diffraction grating, and use it to examine several types of spectra. To learn

More information

The He-Ne Laser * He-Ne Laser System. Power supply and ballast. interatomic collision. 1E-7 sec

The He-Ne Laser * He-Ne Laser System. Power supply and ballast. interatomic collision. 1E-7 sec The He-Ne Laser * I. Introduction The He-Ne laser (Figure 1) uses a low pressure (ca. 1 Torr He, 0.1 Torr Ne) mixture excited by a dc electric discharge. A ballast resistor is placed in series with the

More information

physics 112N interference and diffraction

physics 112N interference and diffraction physics 112N interference and diffraction the limits of ray optics shadow of the point of a pin physics 112N 2 the limits of ray optics physics 112N 3 the limits of ray optics physics 112N 4 this is how

More information

Raman spectroscopy Lecture

Raman spectroscopy Lecture Raman spectroscopy Lecture Licentiate course in measurement science and technology Spring 2008 10.04.2008 Antti Kivioja Contents - Introduction - What is Raman spectroscopy? - The theory of Raman spectroscopy

More information

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. Exam Name MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) A single slit forms a diffraction pattern, with the first minimum at an angle of 40 from

More information

UV/VIS SPECTROPHOTOMETER

UV/VIS SPECTROPHOTOMETER UV/VIS SPECTROPHOTOMETER MECASYS, Co., Ltd. What is UV/VIS? Electromagnetic Spectrum X-Rays UV VIS NIR IR Micro wave 200nm 400nm 700nm 2500nm 500 µm Type of Radiation X-Rays UV/VIS IR Micro Wave Type of

More information

VNA Basics. VNA Basics Errors and Calibration Examples. Spectrum Analyzer

VNA Basics. VNA Basics Errors and Calibration Examples. Spectrum Analyzer 2 Spectrum Spectrum 1 Measures S-parameters of a Device Under Test (DUT) For further reading: Agilent application note Network Basics, available at wwwagilentcom Spectrum 4 Motivation: Why Measure Amplitude?

More information

Uses of Derivative Spectroscopy

Uses of Derivative Spectroscopy Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative

More information

An Introduction to the MTG-IRS Mission

An Introduction to the MTG-IRS Mission An Introduction to the MTG-IRS Mission Stefano Gigli, EUMETSAT IRS-NWC Workshop, Eumetsat HQ, 25-0713 Summary 1. Products and Performance 2. Design Overview 3. L1 Data Organisation 2 Part 1 1. Products

More information

Guide for the Measurements of Absolute Quantum Yields of Liquid Samples

Guide for the Measurements of Absolute Quantum Yields of Liquid Samples ` T e c h n i c a l N o t e Page 1 Guide for the Measurements of Absolute Quantum Yields of Liquid Samples This Guide is designed to help you to determine the Fluorescence Quantum Yields (QY) of liquid

More information

Automated wavelength and intensity calibration routines significantly improve accuracy of recorded spectra

Automated wavelength and intensity calibration routines significantly improve accuracy of recorded spectra Automated wavelength and intensity calibration routines significantly improve accuracy of recorded spectra Wavelength calibration Calibration of dispersive spectral instruments has long been problematic

More information

Ray Optics Minicourse COMSOL Tokyo Conference 2014

Ray Optics Minicourse COMSOL Tokyo Conference 2014 Ray Optics Minicourse COMSOL Tokyo Conference 2014 What is the Ray Optics Module? Add-on to COMSOL Multiphysics Can be combined with any other COMSOL Multiphysics Module Includes one physics interface,

More information

Spectrophotometer measurement laboratory practice

Spectrophotometer measurement laboratory practice Spectrophotometer measurement laboratory practice The operating principle of a Spectrophotometer is based on dispersion of light (Fig.1.) Fig 1. Dispersion of light A spectrophotometer is an instrument

More information

Raman Spectroscopy Basics

Raman Spectroscopy Basics Raman Spectroscopy Basics Introduction Raman spectroscopy is a spectroscopic technique based on inelastic scattering of monochromatic light, usually from a laser source. Inelastic scattering means that

More information

Modern Classical Optics

Modern Classical Optics Modern Classical Optics GEOFFREY BROOKER Department of Physics University of Oxford OXPORD UNIVERSITY PRESS Contents 1 Electromagnetism and basic optics 1 1.1 Introduction 1 1.2 The Maxwell equations 1

More information

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance: Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T

More information

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,

More information

Passive Remote Sensing of Clouds from Airborne Platforms

Passive Remote Sensing of Clouds from Airborne Platforms Passive Remote Sensing of Clouds from Airborne Platforms Why airborne measurements? My instrument: the Solar Spectral Flux Radiometer (SSFR) Some spectrometry/radiometry basics How can we infer cloud properties

More information

Diffraction. Today Single-slit diffraction Diffraction by a circular aperture Use of phasors in diffraction Double-slit diffraction

Diffraction. Today Single-slit diffraction Diffraction by a circular aperture Use of phasors in diffraction Double-slit diffraction Diffraction Today Single-slit diffraction Diffraction by a circular aperture Use of phasors in diffraction Double-slit diffraction Diffraction by a single slit Single slit: Pattern on screen Bright and

More information

Using the Spectrophotometer

Using the Spectrophotometer Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to

More information

Chapter 13 Mass Spectrometry and Infrared Spectroscopy

Chapter 13 Mass Spectrometry and Infrared Spectroscopy Chapter 13 Mass Spectrometry and Infrared Spectroscopy Copyright 2011 The McGraw-Hill Companies, Inc. Permission required for reproduction or display. 1 Overview of Mass Spectrometry Mass spectrometry

More information

Measuring the Refractive Index of Infrared Materials by Dual-Wavelength Fabry-Perot Interferometry. A Senior Project. presented to

Measuring the Refractive Index of Infrared Materials by Dual-Wavelength Fabry-Perot Interferometry. A Senior Project. presented to Measuring the Refractive Index of Infrared Materials by Dual-Wavelength Fabry-Perot Interferometry A Senior Project presented to the Faculty of the Physics Department California Polytechnic State University,

More information

Physics 41 Chapter 38 HW Key

Physics 41 Chapter 38 HW Key Physics 41 Chapter 38 HW Key 1. Helium neon laser light (63..8 nm) is sent through a 0.300-mm-wide single slit. What is the width of the central imum on a screen 1.00 m from the slit? 7 6.38 10 sin θ.11

More information

Using a Spectrophotometer

Using a Spectrophotometer Introduction: Spectrophotometry Using a Spectrophotometer Many kinds of molecules interact with or absorb specific types of radiant energy in a predictable fashion. For example, when while light illuminates

More information

EXPERIMENT 4. Microwave Experiments. Introduction. Experimental Procedure. Part 1 : Double Slit

EXPERIMENT 4. Microwave Experiments. Introduction. Experimental Procedure. Part 1 : Double Slit EXPERIMENT 4 Microwave Experiments Introduction Microwaves are electromagnetic radiation in the centimeter range of wavelengths. As such, they, like light, will exhibit typical wave properties like interference

More information

Absorption of IR Light

Absorption of IR Light Absorption of IR Light Absorption of IR light causes changes in the vibrational motions of a molecule. The different vibrational modes available to a molecule include stretching and bending modes. The

More information

Real World Emissions Testing Based on FTIR Technology PEMS Conference 2014 Ron Tandy

Real World Emissions Testing Based on FTIR Technology PEMS Conference 2014 Ron Tandy Real World Emissions Testing Based on FTIR Technology PEMS Conference 2014 Ron Tandy Acknowledgements (Data Provided): Hong Kong EPD Carol Wong West Virginia University Arvind Thiruvengadam Contents What

More information

Refractive Index and Dispersion: Prism Spectrometer

Refractive Index and Dispersion: Prism Spectrometer Refractive Index and Dispersion: Prism Spectrometer OBJECTIVES: The purpose of this experiment is to study the phenomenon of dispersion i.e. to determine the variation of refractive index of the glass

More information

Objectives 450 CHAPTER 10 LIGHT AND OPTICAL SYSTEMS

Objectives 450 CHAPTER 10 LIGHT AND OPTICAL SYSTEMS Objectives Use wave properties to explain interference and diffraction of light. Explain how double slits, a diffraction grating, a single slit, and an aperture produce interference patterns. Use measurements

More information

Holography 1 HOLOGRAPHY

Holography 1 HOLOGRAPHY Holography 1 HOLOGRAPHY Introduction and Background The aesthetic appeal and commercial usefulness of holography are both related to the ability of a hologram to store a three-dimensional image. Unlike

More information

Chapter 24 Wave Optics. Diffraction Grating Interference by Thin Films Polarization. sinθ=mλ/d or dsinθ=mλ

Chapter 24 Wave Optics. Diffraction Grating Interference by Thin Films Polarization. sinθ=mλ/d or dsinθ=mλ Chapter 24 Wave Optics Diffraction Grating Interference by Thin Films Polarization d Θ Θ Extra distance mλ sinθ=mλ/d or dsinθ=mλ m=0,1,2,3,... Constructive inference m=1/2,3/2,5/2,... Destructive inference

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

MICROWAVE OPTICS THE MEASUREMENTS OF THE WAVELENGTH OF THE MICROWAVES BASED ON THE INTERFERENTIAL METHODS.

MICROWAVE OPTICS THE MEASUREMENTS OF THE WAVELENGTH OF THE MICROWAVES BASED ON THE INTERFERENTIAL METHODS. MICROWAVE OPICS HE MEASUREMENS OF HE WAVELENGH OF HE MICROWAVES BASED ON HE INERFERENIAL MEHODS. BASIC HEORY Microwaves belong to the band of very short electromagnetic waves. he wavelengths of the microwaves

More information

Infrared Sensor Application Note 1 A Background to Gas Sensing by Non-Dispersive Infrared (NDIR)

Infrared Sensor Application Note 1 A Background to Gas Sensing by Non-Dispersive Infrared (NDIR) Infrared Sensor Application Note 1 A Background to Gas Sensing by Non-Dispersive Infrared (NDIR) INFRARED GAS ABSORPTION SPECTRA It is well known that many gas molecules can absorb radiation in the infrared

More information

10.3 The Diffraction Grating

10.3 The Diffraction Grating diffraction grating a device with a large number of equally spaced parallel slits that produces interference patterns 10.3 The Diffraction Grating It is difficult to measure the wavelength of light accurately

More information

Pre-Lab Assignment: Interference, Measuring Wavelengths, and Atomic Spectra

Pre-Lab Assignment: Interference, Measuring Wavelengths, and Atomic Spectra Name: Lab Partners: Date: Pre-Lab Assignment: Interference, Measuring Wavelengths, and Atomic Spectra (Due at the beginning of lab) Directions: Read over the lab handout and then answer the following questions

More information

DIFFRACTION OF LIGHT

DIFFRACTION OF LIGHT Laboratory Exercise 4. DIFFRACTION OF LIGHT Diffraction Gratings. Determining the Wavelength of Laser Light Using a Diffraction Grating. Refraction. Observation of Atomic Spectra. Theoretical background:

More information

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR

Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Detection of trace contamination on metal surfaces using the handheld Agilent 4100 ExoScan FTIR Ensuring ultimate cleanliness for maximum adhesion Application Note Author John Seelenbinder Agilent Technologies,

More information

8. Fourier transform spectroscopy (cont.)

8. Fourier transform spectroscopy (cont.) 8. Fourier transform spectroscopy (cont.) Complex representations of Fourier transformations In cases where the time domain signal is neither perfectly even nor perfectly odd (as all of the previous examples

More information

Chapter 22. Wave Optics. Chapter 22. Wave Optics. What was the first experiment to show that light is a wave? Chapter 22.

Chapter 22. Wave Optics. Chapter 22. Wave Optics. What was the first experiment to show that light is a wave? Chapter 22. Chapter 22. Wave Optics Chapter 22. Wave Optics Light is an electromagnetic wave. The interference of light waves produces the colors reflected from a CD, the iridescence of bird feathers, and the technology

More information

PRACTICE Q6--Quiz 6, Ch15.1 &15.2 Interference & Diffraction

PRACTICE Q6--Quiz 6, Ch15.1 &15.2 Interference & Diffraction Name: Class: Date: ID: A PRACTICE Q6--Quiz 6, Ch5. &5. Interference & Diffraction Multiple Choice Identify the choice that best completes the statement or answers the question.. The trough of the sine

More information

CHAPTER 2 DRUG SUBSTANCES AND INSTRUMENTS EMPLOYED

CHAPTER 2 DRUG SUBSTANCES AND INSTRUMENTS EMPLOYED CHAPTER 2 DRUG SUBSTANCES AND INSTRUMENTS EMPLOYED 2.1 Preparation of Pure polymorphic forms Lamivudine polymorphic Form I and Form II drug substances were prepared in Aurobindo Pharma Limited Research

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

Wave Phenomena. Constructive and Destructive Interference

Wave Phenomena. Constructive and Destructive Interference Wave Phenomena INTERFERENCE PATTERN OF WATER WAVES DIFFRACTION OF LIGHT OFF A COMPACT DISC Constructive and Destructive Interference Constructive interference produces maxima, where crests meet crests

More information

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

WAVELENGTH OF LIGHT - DIFFRACTION GRATING PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant

More information

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,

More information

Light, Light Bulbs and the Electromagnetic Spectrum

Light, Light Bulbs and the Electromagnetic Spectrum Light, Light Bulbs and the Electromagnetic Spectrum Spectrum The different wavelengths of electromagnetic waves present in visible light correspond to what we see as different colours. Electromagnetic

More information

Polarization of Light

Polarization of Light Polarization of Light Introduction Light, viewed classically, is a transverse electromagnetic wave. Namely, the underlying oscillation (in this case oscillating electric and magnetic fields) is along directions

More information

CH 35. Interference. A. Interference of light waves, applied in many branches of science.

CH 35. Interference. A. Interference of light waves, applied in many branches of science. CH 35 Interference [SHIVOK SP212] March 17, 2016 I. Optical Interference: A. Interference of light waves, applied in many branches of science. B. The blue of the top surface of a Morpho butterfly wing

More information

Today. next two weeks

Today. next two weeks Today Temporal and spatial coherence Spatially incoherent imaging The incoherent PSF The Optical Transfer Function (OTF) and Modulation Transfer Function (MTF) MTF and contrast comparison of spatially

More information

Chapter 34 The Wave Nature of Light; Interference. Copyright 2009 Pearson Education, Inc.

Chapter 34 The Wave Nature of Light; Interference. Copyright 2009 Pearson Education, Inc. Chapter 34 The Wave Nature of Light; Interference Units of Chapter 34 Waves versus Particles; Huygens Principle and Diffraction Huygens Principle and the Law of Refraction Interference Young s Double-Slit

More information

Experiment #3: Interference and Diffraction

Experiment #3: Interference and Diffraction Experiment #3: Interference and Diffraction EYE HAZARD: never look directly into a laser! Starting with experiment #4, Please always bring a formatted high-density PC diskette with you to the lab. Purpose:

More information

VISIBLE SPECTROSCOPY

VISIBLE SPECTROSCOPY VISIBLE SPECTROSCOPY Visible spectroscopy is the study of the interaction of radiation from the visible part (λ = 380-720 nm) of the electromagnetic spectrum with a chemical species. Quantifying the interaction

More information

2. Do Not use the laser without getting instructions from the demonstrator.

2. Do Not use the laser without getting instructions from the demonstrator. EXPERIMENT 3 Diffraction Pattern Measurements using a Laser Laser Safety The Helium Neon lasers used in this experiment and are of low power (0.5 milliwatts) but the narrow beam of light is still of high

More information

Ultraviolet - Visible Spectroscopy (UV)

Ultraviolet - Visible Spectroscopy (UV) UV Ultraviolet - Visible Spectroscopy (UV) Introduction to Ultraviolet - Visible Spectroscopy 1 (UV) Background Theory UV Absorption of ultraviolet and visible radiation Absorption of visible and ultraviolet

More information

Lecture 2: Interference

Lecture 2: Interference Lecture 2: Interference λ S 1 d S 2 Lecture 2, p.1 Today Interference of sound waves Two-slit interference Lecture 2, p.2 Review: Wave Summary ( ) ( ) The formula y x,t = Acos kx ωt describes a harmonic

More information