FTIR Instrumentation
|
|
- Milton Pearson
- 7 years ago
- Views:
Transcription
1 FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: 1. IR Instrumentation Two types of instrumentation are used to obtain infrared spectra: Dispersive spectrophotometers, which use a monochromator to produce an infrared spectrum one resolution element at a time. Michelson interferometers, which use a moving mirror to create an interference pattern, or interferogram, from which all resolution elements are determined simultaneously. 2. FTIR Spectroscopy History Chemical infrared spectroscopy emerged as a science in the 1880s. In the 1890s, A. A. Michelson, to further his studies of the speed of light, invented the interferometer. In the early 1940s, chemical infrared spectroscopy was still an immature scientific field. But with commercial development of the optical null dispersive spectrophotometer, later that decade, chemical infrared spectroscopy came into widespread use. Dispersive instruments proved the tremendous value of infrared analysis, and soon became the mainstay of organic characterization laboratories. In 1949 astrophysicist Peter Fellgett used an interferometer to measure light from celestial bodies and produced the first Fourier transform infrared spectrum. But for many years, only a few advanced research groups with access to large, expensive computers and with personnel able to wait up to 12 hours to transform an interferogram into a spectrum used Fourier transform infrared (FTIR) spectroscopy. For two decades more, FTIR spectrometers were limited to studying problems not solvable with dispersive techniques. In the late 1960s when microcomputers able to do the Fourier transform became available, commercial FTIR spectrometers appeared. The 1966 development of the Cooley-Tukey algorithm, which quickly does a Fourier transform (the Fast Fourier Transform or FFT), was also instrumental in the commercialization of FTIR spectrometers. However, the first FTIR spectrometers were large and expensive, and were found primarily in a few well-to-do research labs. Gradually, technology reduced the cost, increased the availability, and enhanced the capabilities of FTIR spectroscopy systems. The performance-to-price ratio provided by FTIR spectrometers today was unthinkable only a decade ago. 1
2 3. Theory of Interferometer Operation The optical system in an FTIR spectrometer is very simple: the interferometer requires two mirrors, an infrared light source, an infrared detector, and a beamsplitter. fixed mirror beam splitter collimator moving mirror detector source focussing mirror Figure 1. The Michelson interferometer The beamsplitter is the heart of the interferometer. Essentially a half-silvered mirror, the beamsplitter reflects about half of an incident light beam while simultaneously transmitting the remaining half. One half of this split light beam travels to the interferometer's moving mirror while the other half travels to the interferometer's stationary mirror. The two mirrors reflect both beams back to the beamsplitter where each of the two beams is again half reflected and half transmitted. Two output beams result: one travels to the detector as the other travels to the source. When the two beams return to the beamsplitter, an interference pattern, or interferogram, is generated. This interference pattern varies with the displacement of the moving mirror, that is, with the difference in pathlength in the two arms of the interferometer. The interference pattern, detected by the infrared detector as variations in the infrared energy level, is what ultimately yields spectral information. 4. More Interferometer Theory An example of how light is affected in an interferometer further illustrates interferometer operation. Consider the interference pattern generated when 2
3 monochromatic light, such as that produced by a helium-neon (HeNe) laser, is the light source. An idealized beamsplitter: - neither generates nor absorbs radiation. - transmits half of all incident light. - reflects half of all incident light. All light energy that enters the interferometer from the source either returns back to the source or travels to the infrared detector. As the moving mirror scans back and forth, more or less than half of the total light energy that originally entered the interferometer reaches the detector. All light that does not reach the detector reaches the source instead. For example, if both interferometer mirrors are equidistant from the beamsplitter, the two light beams travel equal distances. As each beam returns to the beamsplitter, each is again split and then recombined with half of the light from the other interferometer arm. The two component beams that make up the output beam passed to the detector are completely in phase with each other and, therefore, interfere constructively. The two component beams that make up the output beam passed to the source are completely out of phase with each other and, therefore, interfere destructively. In this case, all of the original light energy goes to the detector and none goes to the source. By contrast, if the moving mirror is displaced one-quarter wavelength of the HeNe laser radiation (to produce optical retardation of one-half wavelength), light returning from the interferometer's moving mirror reaches the beamsplitter one-half wavelength out of phase with respect to the light returning from the fixed mirror. This results in complete destructive interference in the detected output beam so no light energy reaches the detector, and all light returns to the source. Figure 2. Constructive (left) and destructive (right) interference. As the moving mirror scans back and forth, the detector observes light and dark bands corresponding to partial and total constructive and destructive interference in the detected output beam. In this monochromatic light example, the interference signal the detector receives produces a cosine interferogram. This wave goes from full intensity at the point where both mirrors are equidistant from the beamsplitter to zero intensity when the moving mirror has traveled one-quarter wavelength. Full intensity is again achieved when the moving mirror has traveled one-half or multiples of one-half wavelength. Similarly, zero intensity occurs at each multiple of one-quarter wavelength of mirror travel. 3
4 An infrared source is polychromatic, emitting light over a broad range of frequencies. Each frequency still yields a unique cosine signal: the resulting interferogram represents the sum of all cosine waves generated by each individual infrared frequency. However, only at the point where both mirrors are equal distances from the beamsplitter does complete constructive interference occur--simultaneous constructive interference of all wavelengths. At any other point, only some wavelengths interfere constructively while others do not. The Fourier transform resolves the frequency and intensity of each cosine wave in the interferogram. In other words, this algorithm converts the measured.intensity-versusmirror-displacement signal (the interferogram) into a plot of intensity versus frequency (a spectrum). Δx Figure 3. Spectrum with three frequencies with different amplitudes (top), interferograms for the three separate frequencies (middle), and Interferograms for the three mixed frequencies. Δx Δx = 0 5. Advantage 5.1 Multiplex Advantage (Fellgett's Advantage) An interferometer does not separate light into individual frequencies before measurement. This means each point in the interferogram contains information from each wavelength in the input light. In other words, if 8,000 data points along the interferogram are collected, each wavelength in the input light is sampled 8,000 times. By contrast, a dispersive spectrophotometer that measures 8,000 individual points across a spectrum samples each wavelength only once. 5.2 Throughput Advantage (Jacquinot's Advantage) The simple optical path of the interferometer (no slits and fewer optical elements) means more energy gets to the sample than is possible with dispersive 4
5 spectrophotometers. This means more energy reaches the detector, increasing the spectrum's potential signal-to-noise ratio. Together the multiplex and throughput advantages allow an FTIR spectrometer to obtain a high-quality infrared spectrum in a fraction of the time needed to get the same spectrum on a dispersive instrument. Also, to increase the resolution of dispersive instruments, the slits through which light must pass are narrowed, thereby decreasing energy throughput. In an FTIR spectrometer, resolution is increased by lengthening the moving mirror stroke length with no decrease in energy throughput. As wavelength resolution increases, the advantages of interferometric versus dispersive measurements increase. 5.3 Frequency Precision (Conne's Advantage) With dispersive instruments, frequency precision and accuracy depend on: 1) calibration with external standards and 2) the ability of electromechanical mechanisms to uniformly move gratings and slits during and between scans. By contrast, the interferometer has an internal frequency standard, generally a heliumneon laser. Also, both mirror movement and detector sampling are clocked by the interferometer fringes from the laser's monochromatic light. All frequencies in the output spectrum are calculated from the known frequency of the laser light. An FTIR spectrometer easily achieves frequency precision and accuracy of better than 0.01 wavenumbers. This means that spectra collected with an FTIR spectrometer can be quantitatively compared whether they were collected five minutes or five years apart. 5.4 Constant Spectral Resolution In an FTIR spectrometer, the resolution of the measured spectrum is the same for all frequencies, not varied throughout the spectrum as is often true with dispersive instruments. 6. Resolution In a dispersive spectrophotometer, resolution depends on slit width and on the reciprocal of the grating's linear displacement. As the slits are narrowed, the range of frequencies passed to the detector decreases, resulting in higher resolution. In an FTIR spectrometer, resolution is increased by increasing optical retardation, that is, by increasing how far the moving mirror moves. A relationship between resolution and optical retardation can be empirically derived by considering a spectrum made up of two sharp bands. Each band is at a distinct frequency and generates a sinusoidal signal. The total signal at the detector is the sum of the two signals. If the two spectral frequencies are relatively far apart, the resulting signals are quite different and only a short movement of the mirror is needed to measure a complete 5
6 cycle of the pattern made by the addition of the two waveforms. On the other hand, if the two frequencies are closer together, the interferometer needs to scan a larger distance to measure a full cycle of the addition pattern. The relationship between spectral line spacing and the interferogram is R = 1/d where R is the spectral resolution in wavenumbers and d is the optical retardation in centimeters. Thus, the length of mirror travel (retardation) must be doubled to reduce instrument resolution by half, or, in other words, to improve instrument resolution by a factor of two. Several different criteria are used to measure FTIR spectrometer resolution. These include the Rayleigh criterion and the full-width-at-half-height (FWHH) method. Measuring instrument resolution with the FWHH method involves collecting an absorption spectrum with very sharp peaks, such as that obtained from low-pressure CO at about 5 torr pressure. fixed mirror electronics beam splitter moving mirror source aperture detector sample Figure 4: Left: Outer view of the Bruker IFS 66v/S FTIR spectrometer. (A) Sample chamber, (B) Interferometer chamber, (C) Laser compartment, (D) Electronic compartment and vacuum control, (E) Detector chamber. Right: Schematic internal view. Bold lines mark the optical pathway. 6
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
More informationThe Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD
TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic
More informationComponents for Infrared Spectroscopy. Dispersive IR Spectroscopy
Components for Infrared Spectroscopy Mid-IR light: 00-000 cm - (5.5 m wavelength) Sources: Blackbody emitters Globar metal oxides Nernst Glower: Silicon Carbide Detectors: Not enough energy for photoelectric
More informationORIEL. FT-IR Spectroscopy SECTION FOUR FEATURES. Glossary of Terms Introduction to FT-IR Spectroscopy
ORIEL ORIEL PRODUCT TRAINING FT-IR Spectroscopy SECTION FOUR FEATURES Glossary of Terms Introduction to FT-IR Spectroscopy Stratford, CT Toll Free 800.74.5393 Fax 03.378.457 www.newport.com/oriel oriel.sales@newport.com
More informationSpectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
More informationFundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
More informationExperiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
More informationQUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
More information6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm?
Test IV Name 1) In a single slit diffraction experiment, the width of the slit is 3.1 10-5 m and the distance from the slit to the screen is 2.2 m. If the beam of light of wavelength 600 nm passes through
More informationInterferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)
Interferometers The true worth of an experimenter consists in his pursuing not only what he seeks in his experiment, but also what he did not seek. Claude Bernard (1813-1878) OBJECTIVES To examine the
More informationENGINEERING METROLOGY
ENGINEERING METROLOGY ACADEMIC YEAR 92-93, SEMESTER ONE COORDINATE MEASURING MACHINES OPTICAL MEASUREMENT SYSTEMS; DEPARTMENT OF MECHANICAL ENGINEERING ISFAHAN UNIVERSITY OF TECHNOLOGY Coordinate Measuring
More informationExperiment 5. Lasers and laser mode structure
Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,
More informationPUMPED Nd:YAG LASER. Last Revision: August 21, 2007
PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow
More informationUses of Derivative Spectroscopy
Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative
More informationRaman spectroscopy Lecture
Raman spectroscopy Lecture Licentiate course in measurement science and technology Spring 2008 10.04.2008 Antti Kivioja Contents - Introduction - What is Raman spectroscopy? - The theory of Raman spectroscopy
More informationAn Introduction to the MTG-IRS Mission
An Introduction to the MTG-IRS Mission Stefano Gigli, EUMETSAT IRS-NWC Workshop, Eumetsat HQ, 25-0713 Summary 1. Products and Performance 2. Design Overview 3. L1 Data Organisation 2 Part 1 1. Products
More informationProblem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:
Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T
More informationOptical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus
Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,
More informationEXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
More informationPassive Remote Sensing of Clouds from Airborne Platforms
Passive Remote Sensing of Clouds from Airborne Platforms Why airborne measurements? My instrument: the Solar Spectral Flux Radiometer (SSFR) Some spectrometry/radiometry basics How can we infer cloud properties
More informationRaman Spectroscopy Basics
Raman Spectroscopy Basics Introduction Raman spectroscopy is a spectroscopic technique based on inelastic scattering of monochromatic light, usually from a laser source. Inelastic scattering means that
More informationPhysics 41 Chapter 38 HW Key
Physics 41 Chapter 38 HW Key 1. Helium neon laser light (63..8 nm) is sent through a 0.300-mm-wide single slit. What is the width of the central imum on a screen 1.00 m from the slit? 7 6.38 10 sin θ.11
More informationWAVELENGTH OF LIGHT - DIFFRACTION GRATING
PURPOSE In this experiment we will use the diffraction grating and the spectrometer to measure wavelengths in the mercury spectrum. THEORY A diffraction grating is essentially a series of parallel equidistant
More informationSynthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
More informationUsing the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
More informationHolography 1 HOLOGRAPHY
Holography 1 HOLOGRAPHY Introduction and Background The aesthetic appeal and commercial usefulness of holography are both related to the ability of a hologram to store a three-dimensional image. Unlike
More informationToday. next two weeks
Today Temporal and spatial coherence Spatially incoherent imaging The incoherent PSF The Optical Transfer Function (OTF) and Modulation Transfer Function (MTF) MTF and contrast comparison of spatially
More informationModern Classical Optics
Modern Classical Optics GEOFFREY BROOKER Department of Physics University of Oxford OXPORD UNIVERSITY PRESS Contents 1 Electromagnetism and basic optics 1 1.1 Introduction 1 1.2 The Maxwell equations 1
More informationModule 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
More informationAP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
More informationTime out states and transitions
Time out states and transitions Spectroscopy transitions between energy states of a molecule excited by absorption or emission of a photon hn = DE = E i - E f Energy levels due to interactions between
More informationInterference. Physics 102 Workshop #3. General Instructions
Interference Physics 102 Workshop #3 Name: Lab Partner(s): Instructor: Time of Workshop: General Instructions Workshop exercises are to be carried out in groups of three. One report per group is due by
More informationRealization of a UV fisheye hyperspectral camera
Realization of a UV fisheye hyperspectral camera Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM Outline Purpose of the instrument Required specs Hyperspectral technique Optical
More informationInfrared Spectroscopy 紅 外 線 光 譜 儀
Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed
More informationANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION
Chem 306 Section (Circle) M Tu W Th Name Partners Date ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION Materials: prepared acetylsalicylic acid (aspirin), stockroom samples
More informationDETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY
DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY Eduard Gilli 1,2 and Robert Schennach 1, 2 1 Graz University of Technology, 8010 Graz, Austria 2 CD-Laboratory for Surface Chemical and Physical
More informationDETECTION OF SUBSURFACE DAMAGE IN OPTICAL TRANSPARENT MATERIALSS USING SHORT COHERENCE TOMOGRAPHY. Rainer Boerret, Dominik Wiedemann, Andreas Kelm
URN (Paper): urn:nbn:de:gbv:ilm1-2014iwk-199:0 58 th ILMENAU SCIENTIFIC COLLOQUIUM Technische Universität Ilmenau, 08 12 September 2014 URN: urn:nbn:de:gbv:ilm1-2014iwk:3 DETECTION OF SUBSURFACE DAMAGE
More informationEXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab
EXPERIMENT O-6 Michelson Interferometer Abstract A Michelson interferometer, constructed by the student, is used to measure the wavelength of He-Ne laser light and the index of refraction of a flat transparent
More informationOptical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD
2008 AGI-Information Management Consultants May be used for personal purporses only or by libraries associated to dandelon.com network. Optical Metrology Third Edition Kjell J. Gasvik Spectra Vision AS,
More informationMB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC
Measurement & Analytics Measurement made easy MB3600 Versatile FT-NIR analyzer designed for your industry The most reliable FT-NIR specifically designed for QA/QC MB3600 FT-NIR spectrometer for your industry
More informationDETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY
DETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY Hermenegildo García Gómez Departamento de Química Instituto de Tecnología Química Universidad Politécnica de Valencia 46022
More informationRaman vs. FTIR Process Analyzers for In-Line Blending Use
Raman vs. FTIR Process Analyzers for In-Line Blending Use Ara Barsamian Refinery Automation Institute, LLC Jabarsa@refautom.com: Tel: +1-973-644-2270 This paper was provoked by recent interest in using
More informationDoppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19
Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and
More informationRaman Scattering Theory David W. Hahn Department of Mechanical and Aerospace Engineering University of Florida (dwhahn@ufl.edu)
Introduction Raman Scattering Theory David W. Hahn Department of Mechanical and Aerospace Engineering University of Florida (dwhahn@ufl.edu) The scattering of light may be thought of as the redirection
More informationLaboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control
More informationSignal to Noise Instrumental Excel Assignment
Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a
More informationReprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July 2001. Reprinted from Photonics Spectra, Laurin Publishing
Reprint (R22) Avoiding Errors in UV Radiation Measurements By Thomas C. Larason July 2001 Reprinted from Photonics Spectra, Laurin Publishing Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1
More information13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
More informationFibre Bragg Grating Sensors An Introduction to Bragg gratings and interrogation techniques
Fibre Bragg Grating Sensors An ntroduction to Bragg gratings and interrogation techniques Dr Crispin Doyle Senior Applications Engineer, Smart Fibres Ltd. 2003 1) The Fibre Bragg Grating (FBG) There are
More information0 10 20 30 40 50 60 70 m/z
Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions
More informationATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.
1 ATOMIC SPECTRA Objective: To measure the wavelengths of visible light emitted by atomic hydrogen and verify the measured wavelengths against those predicted by quantum theory. To identify an unknown
More informationProposed experiment to test the non-locality hypothesis in transient light-interference phenomena
Proposed experiment to test the non-locality hypothesis in transient light-interference phenomena Masanori Sato Honda Electronics Co., Ltd., 20 Oyamazuka, Oiwa-cho, Toyohashi, Aichi 441-3193, Japan Abstract
More informationNuclear Magnetic Resonance Spectroscopy
Nuclear Magnetic Resonance Spectroscopy Nuclear magnetic resonance spectroscopy is a powerful analytical technique used to characterize organic molecules by identifying carbonhydrogen frameworks within
More informationPolarization of Light
Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights
More informationCopyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
More informationDispersion in Optical Fibers
Dispersion in Optical Fibers By Gildas Chauvel Anritsu Corporation TABLE OF CONTENTS Introduction Chromatic Dispersion (CD): Definition and Origin; Limit and Compensation; and Measurement Methods Polarization
More informationFunction Guide for the Fourier Transformation Package SPIRE-UOL-DOC-002496
Function Guide for the Fourier Transformation Package SPIRE-UOL-DOC-002496 Prepared by: Peter Davis (University of Lethbridge) peter.davis@uleth.ca Andres Rebolledo (University of Lethbridge) andres.rebolledo@uleth.ca
More informationInfrared Spectroscopy: Theory
u Chapter 15 Infrared Spectroscopy: Theory An important tool of the organic chemist is Infrared Spectroscopy, or IR. IR spectra are acquired on a special instrument, called an IR spectrometer. IR is used
More informationFiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy
Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The availability and applications of fiber optic sampling in UV/Vis
More information5.33 Lecture Notes: Introduction to Spectroscopy
5.33 Lecture Notes: ntroduction to Spectroscopy What is spectroscopy? Studying the properties of matter through its interaction with different frequency components of the electromagnetic spectrum. Latin:
More informationSpherical Beam Volume Holograms Recorded in Reflection Geometry for Diffuse Source Spectroscopy
Spherical Beam Volume Holograms Recorded in Reflection Geometry for Diffuse Source Spectroscopy Sundeep Jolly A Proposal Presented to the Academic Faculty in Partial Fulfillment of the Requirements for
More informationOrganic Chemistry Tenth Edition
Organic Chemistry Tenth Edition T. W. Graham Solomons Craig B. Fryhle Welcome to CHM 22 Organic Chemisty II Chapters 2 (IR), 9, 3-20. Chapter 2 and Chapter 9 Spectroscopy (interaction of molecule with
More informationDIODE LASER BASED PHOTOACOUSTIC SYSTEM FOR ATMOSPHERIC WATER VAPOR MEASUREMENTS
SUMMARY OF PHD THESIS DIODE LASER BASED PHOTOACOUSTIC SYSTEM FOR ATMOSPHERIC WATER VAPOR MEASUREMENTS MIKLÓS SZAKÁLL Consultants: Dr. Gábor Szabó professor, member of the Hungarian Academy of Science Dr.
More information3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
More informationInterferometric Measurement of Dispersion in Optical Components
Interferometric Measurement of Dispersion in Optical Components Mark Froggatt, Eric Moore, and Matthew Wolfe Luna Technologies, Incorporated, 293-A Commerce Street, Blacksburg, Virginia 246 froggattm@lunatechnologies.com.
More informationSymmetric Stretch: allows molecule to move through space
BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed
More informationIt has long been a goal to achieve higher spatial resolution in optical imaging and
Nano-optical Imaging using Scattering Scanning Near-field Optical Microscopy Fehmi Yasin, Advisor: Dr. Markus Raschke, Post-doc: Dr. Gregory Andreev, Graduate Student: Benjamin Pollard Department of Physics,
More informationILLUSTRATIVE EXAMPLE: Given: A = 3 and B = 4 if we now want the value of C=? C = 3 + 4 = 9 + 16 = 25 or 2
Forensic Spectral Anaylysis: Warm up! The study of triangles has been done since ancient times. Many of the early discoveries about triangles are still used today. We will only be concerned with the "right
More informationElectronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)
Page 1 Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT) ECC RECOMMENDATION (06)01 Bandwidth measurements using FFT techniques
More informationExtraction Heights for STIS Echelle Spectra
Instrument Science Report STIS 98-09 Extraction Heights for STIS Echelle Spectra Claus Leitherer and Ralph Bohlin March 1998 ABSTRACT The optimum extraction height (h) for STIS echelle spectra of 7 pixels
More informationUV-Vis spectrophotometers
UV-Vis spectrophotometers an Advanced Digital Technology EXPERIENCE SECOMAM is proud to provide the new fourthgeneration UVIKON ultraviolet-visible spectrophotometers.with more than five decades of experience
More informationO6: The Diffraction Grating Spectrometer
2B30: PRACTICAL ASTROPHYSICS FORMAL REPORT: O6: The Diffraction Grating Spectrometer Adam Hill Lab partner: G. Evans Tutor: Dr. Peter Storey 1 Abstract The calibration of a diffraction grating spectrometer
More informationSpectroscopy and Regions of the Spectrum
Basics 9 Spectroscopy and Regions of the Spectrum Different regions of the spectrum probe different types of energy levels of an atomic or molecular system. It is not uncommon to refer to a spectroscopic
More informationChemistry 111 Lab: Intro to Spectrophotometry Page E-1
Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 SPECTROPHOTOMETRY Absorption Measurements & their Application to Quantitative Analysis study of the interaction of light (or other electromagnetic
More informationANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1
WHAT IS AN FFT SPECTRUM ANALYZER? ANALYZER BASICS The SR760 FFT Spectrum Analyzer takes a time varying input signal, like you would see on an oscilloscope trace, and computes its frequency spectrum. Fourier's
More informationINTERFERENCE OF SOUND WAVES
1/2016 Sound 1/8 INTERFERENCE OF SOUND WAVES PURPOSE: To measure the wavelength, frequency, and propagation speed of ultrasonic sound waves and to observe interference phenomena with ultrasonic sound waves.
More informationAgilent Cary 4000/5000/6000i Series UV-Vis-NIR
Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Guaranteed specifications Design overview Double beam, ratio recording, double out-of-plane Littrow monochromator UV-Vis-NIR spectrophotometer (Agilent Cary
More informationA down-under undergraduate optics and photonics laboratory
A down-under undergraduate optics and photonics laboratory Barry Perczuk and Michael Gal School of Physics, The University of New South Wales, Sydney, NSW 2052, Australia ABSTRACT Our senior undergraduate
More informationAustin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
More informationGRID AND PRISM SPECTROMETERS
FYSA230/2 GRID AND PRISM SPECTROMETERS 1. Introduction Electromagnetic radiation (e.g. visible light) experiences reflection, refraction, interference and diffraction phenomena when entering and passing
More informationMeasurement & Analytics Measurement made easy
Measurement & Analytics Measurement made easy MB3600-PH Versatile FT-NIR analyzer for the life sciences and pharmaceutical industries Designed for QA/QC, research and development and at-line PAT applications
More informationWaveforms and the Speed of Sound
Laboratory 3 Seth M. Foreman February 24, 2015 Waveforms and the Speed of Sound 1 Objectives The objectives of this excercise are: to measure the speed of sound in air to record and analyze waveforms of
More informationSpectroscopic Methods
Chapter 10 Spectroscopic Methods Chapter Overview Section 10A Overview of Spectroscopy Section 10B Spectroscopy Based on Absorption Section 10C UV/Vis and IR Spectroscopy Section 10D Atomic Absorption
More informationNano-Spectroscopy. Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale
Nano-Spectroscopy Solutions AFM-Raman, TERS, NSOM Chemical imaging at the nanoscale Since its introduction in the early 80 s, Scanning Probe Microscopy (SPM) has quickly made nanoscale imaging an affordable
More informationWir schaffen Wissen heute für morgen
Diffractive optics for photon beam diagnostics at hard XFELs Wir schaffen Wissen heute für morgen PSI: SLAC: ESRF: SOLEIL: APS: SACLA: EuroXFEL C. David, S. Rutishauser, P. Karvinen, Y. Kayser, U. Flechsig,
More informationSpectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
More information- the. or may. scales on. Butterfly wing. magnified about 75 times.
Lecture Notes (Applications of Diffraction) Intro: - the iridescent colors seen in many beetles is due to diffraction of light rays hitting the small groovess of its exoskeleton - these ridges are only
More informationScanning Near-Field Optical Microscopy for Measuring Materials Properties at the Nanoscale
Scanning Near-Field Optical Microscopy for Measuring Materials Properties at the Nanoscale Outline Background Research Design Detection of Near-Field Signal Submonolayer Chemical Sensitivity Conclusions
More informationSpecifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction
More informationIntroduction to Optics
Second Edition Introduction to Optics FRANK L. PEDROTTI, S.J. Marquette University Milwaukee, Wisconsin Vatican Radio, Rome LENO S. PEDROTTI Center for Occupational Research and Development Waco, Texas
More informationA More Efficient Way to De-shelve 137 Ba +
A More Efficient Way to De-shelve 137 Ba + Abstract: Andrea Katz Trinity University UW REU 2010 In order to increase the efficiency and reliability of de-shelving barium ions, an infrared laser beam was
More informationDiffraction and Young s Single Slit Experiment
Diffraction and Young s Single Slit Experiment Developers AB Overby Objectives Preparation Background The objectives of this experiment are to observe Fraunhofer, or far-field, diffraction through a single
More informationThe coherence length of black-body radiation
Eur. J. Phys. 19 (1998) 245 249. Printed in the UK PII: S143-87(98)86653-1 The coherence length of black-body radiation Axel Donges Fachhochschule und Berufskollegs NTA Prof. Dr Grübler, Seidenstrasse
More informationAdvanced Solutions for Gas Monitoring
Continuous Emissions Monitoring Stack Testing Process Monitoring Quality Control Engine Exhaust Gas Monitoring Workplace Air Quality Monitoring Combustion Research Emergency Rescue Services Advanced Solutions
More informationSelf-Mixing Laser Diode Vibrometer with Wide Dynamic Range
Self-Mixing Laser Diode Vibrometer with Wide Dynamic Range G. Giuliani,, S. Donati, L. Monti -, Italy Outline Conventional Laser vibrometry (LDV) Self-mixing interferometry Self-mixing vibrometer Principle:
More information3.5.4.2 One example: Michelson interferometer
3.5.4.2 One example: Michelson interferometer mirror 1 mirror 2 light source 1 2 3 beam splitter 4 object (n object ) interference pattern we either observe fringes of same thickness (parallel light) or
More informationPhysics 441/2: Transmission Electron Microscope
Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This
More informationLab 9: The Acousto-Optic Effect
Lab 9: The Acousto-Optic Effect Incoming Laser Beam Travelling Acoustic Wave (longitudinal wave) O A 1st order diffracted laser beam A 1 Introduction qb d O 2qb rarefractions compressions Refer to Appendix
More informationA Guide to Acousto-Optic Modulators
A Guide to Acousto-Optic Modulators D. J. McCarron December 7, 2007 1 Introduction Acousto-optic modulators (AOMs) are useful devices which allow the frequency, intensity and direction of a laser beam
More informationRF Measurements Using a Modular Digitizer
RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.
More information