A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY

Size: px
Start display at page:

Download "A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY"

Transcription

1 A VERSATILE COUNTER FOR CONVERSION MÖSSBAUER SPECTROSCOPY I. BIBICU 1, G. NICOLESCU 2, L. CIOLACU 2, L. SERBINA 2 1 National Institute for Materials Physics, Bucharest 77125, Romania, bibicu@infim.ro 2 National Institute of Physics and Nuclear Engineering, IFIN-HH, Bucharest 77125, Romania Received June 26, 29 A multi-purpose gas-flow proportional counter for surface Mössbauer spectroscopy is described. The main improvements obtained are: the height of the detection volume can be changed in large limits from 1 to 38 mm, the detection volume can be symmetrical or not in respect with anode plan, the anode changing is easily and different anode configuration can be used. It is suitable for studies with 57 Fe, 119 Sn and 151 Eu isotopes. The performances of the detector are presented. Key words: gas-flow detector, backscatering experiment, surface studies, Fe-57, Sn-119, Eu INTRODUCTION Mössbauer spectroscopy [1] is based on the resonant recoil free absorption and emission of low energy γ-rays. The effect is significant when the nuclei are imbedded in a rigid matrix. The nuclear resonance absorption is successively followed by nuclear and atomic processes resulting in the emission of various radiations: γ-ray, conversion electrons, X-rays and Auger electrons. This secondary radiation is employed to obtain an emission-type Mössbauer spectrum by the backscattering method, in which the counts of secondary radiations increase at the velocity of nuclear resonance. The information obtained by the backscattering method is restricted to the layer to which the secondary radiation employed in the measurement can escape from the surface of sample. The detection of the three backscattered particles permits surface studies to be performed at various depths. The smallest depths can be investigated by electron detection. The conversion electron signal is quite high for the 57 Fe and 119 Sn isotopes. The purpose of this paper is to describe a new version of a gas-flow proportional detector for detection of the conversion electrons or characteristic X-ray. The improved performances of the multi-purpose detector are demonstrated by measurements on samples containing the isotopes Fe-57, Sn-119 and Eu-151. The energy of the backscattered particles is in a large range [2]. Rom. Journ. Phys., Vol. 56, Nos. 1 2, P , Bucharest, 211

2 2 Counter for conversion Mössbauer spectroscopy DESCRIPTION The sample to be studied is mounted inside the detector and the sample itself is a part of the cathode of the detector. It was designed in a cylindrical geometry. The cross section of the detector is shown in Fig. 1 and the dimensions are in mm. Fig. 1. Cross section of the detector 1 and 2, main parts of the detector body; 3 input piece; 4 sample holder; 5 teflon insulator; 6 stainless steel ring; 7 sample; 8 collimator; 9 gas connection; 1 high voltage connector; 11 mylar windows; 6, 12 tightness piece; S Mössbauer source. The body of the detector, made of aluminium, consists of only two main parts (1 and 2), in order to lower the leak rate. The background due to photoelectrons is minimised by using low-z materials as much as possible. The anode, gold-coated tungsten wire, 25 µm diameter, is supported in the center of the detector by a stainless steel ring 6. The ring can be changed easily with another, with the same dimensions. On the ring are mounted parallel anode wires. The number of wires and the distance between them can be changed. Thus for o specific geometry and measurement type can be find an optimum anode configuration. The ring diameter is higher than twice of sample (7) diameter. The investigated part of the sample has a diameter higher than 2 mm. The electric field at the sample is thus practically constant. The casing of the stainless steel ring is made of teflon insulator (5). The Mössbauer sample is mounted inside the counter by means of a sample holder 4. The holder allows an easy manipulation of the sample outside the detector and it can always be repositioned in a reproducible manner with respect to the detector body. The holder also allows performing simultaneously surface and transmission measurements for samples with suitable thickness for transmission measurements.

3 152 I.Bibicu et al. 3 In this case it is necessary another counter placed behind the versatile conuter. The front side of the detector is closed with an input piece 3, identically with sample holder. The parts 3 and 4 are equipped with thin aluminized mylar windows 11. In the input piece 3 may be inserted a collimator 8. The components 3 and 4 can be moved independently in respect with detector body and the space between them represents the detection volume. The height of the detection volume can be fixed, mechanically, in large limits from 1 to 38 mm, moving the two components. The detection volume can be settled symmetrical or not in respect with anode plan. By changing the volume detection and flow gas it is possible to make measurements by electron, X-ray or gamma ray detection. For conversion electron detection a 94% He + 6% CH 4 or 99% He + 1% C 4 H 1 mixtures was used and for X-ray a 9% Ar + 1% CH 4 mixture. The gas flow rate can be set in the range of 5 1 cm 3 /hour by a flowmeter working at a pressure of about.1 Mpa. It is possible to set relatively low flow rates with sufficiently small deviations, so that no fluctuations in the amplifying factor appear. The access of the gas was designed such to avoid vibration of the anode, due to the flow rate and any gas leak. We used an economical radiation shielding for counter, consisting of a combination of lead, copper and steel disks. To destroy the characteristic radiation, alternate mounting of lead, copper and steel disks was used. In order to absorb unfavourable X-rays from the sources, a plexiglas filter was placed in front of the shielding. 3. RESULTS The measurements were made, at room temperature, with a new constructed, versatile flow-gas proportional counter, suitable for studies with all mentioned isotopes. We used the following Mössbauer sources: 57 Co diffused in Rh matrix, 119m Sn diffused in CaSnO 3 matrix and 151 Sm diffused in samarium oxide. Measurements were carried out by inserting the proportional counter into an Elscint AME-5 Mössbauer spectrometer using a compatible CMCA-55 Wissel unit for data acquisition. We used as Mössbauer samples stainless steel, β-sn metallic foil and Eu 2 O 3. The thickness of the detection space was adjusted to minimum necessary required by signal/noise ratio [3], [4]. The detector shows the best performance for electrons at a voltage level of 9 11 V and for X-rays at a level voltage of V. Gas flow rate was in the range 5 1 cm 3 /hour. The parameters of the Mössbauer spectra were calculated using a computer fitting program, which assumes a Lorentzian line shape. The isomer shifts were referred to α-fe. The electron amplitude spectra of the detector with 57 Co and 119m Sn sources and stainless steel, respectively β-sn samples inside are presented in figure 2 and 3. Eu did not emit electrons by the de-excitation of the nuclear excited states [5].

4 4 Counter for conversion Mössbauer spectroscopy Fig. 2. Pulse height spectrum of electrons scattered from stainless steel sample, recorded with 57 Co source without filter; 3 mm thickness of the detecting volume; 3 minutes acquisition time Fig. 3. Pulse height spectrum of electrons scattered from β-sn sample, recorded with 119m Sn source without filter; 5 mm thickness of the detecting volume; 3 minutes acquisition time. In the case of Fe we obtained a spectrum better than the previously reported ones [6]. It has a visible structure and it is similar with that one obtained using the rhodium foil 3% enriched in 57 Fe. [7]. A good setting of the SCA window for the resonant part of the pulse-height spectrum now it is possible for iron studies. The amplitude spectrum of the detector with 119m Sn source and β-sn sample inside displays a large peak, with a good resolution corresponding to electron energies: kev [2]. The amplitude spectra of the detector for X-ray, with the mentioned sources, without filter, are shown in figure 4, 5 and 6 respectively. The amplitude spectra have a good resolution for low energy X-ray, especially for Sn low energy X-ray ( kev) and low effectiveness for Mössbauer radiation. These facts prove the performances of the detector.

5 154 I.Bibicu et al kev 12 8 Escape peak KeV 1 2 Fig. 4. Pulse height spectrum of photons recorded with 57 Co source without filter and with stainless steel sample: 17 mm thickness of the detecting volume; 6 minutes acquisition time , 25.27keV kev Fig. 5. Pulse height spectrum of photons recorded with 119m Sn source without filter and with β-sn sample; 2 mm thickness of the detecting volume; 3 minutes acquisition time kev kev Fig. 6. Pulse height spectrum of photons recorded with 151 Sm source without filter and with Eu 2 O 3 sample; 14 mm thickness of the detecting volume; 3 minutes acquisition time.

6 6 Counter for conversion Mössbauer spectroscopy 155 The Mössbauer spectrum obtained by the detection of electrons from metallic Sn is presented in the Fig. 7. This spectrum evidences presence of two Mössbauer lines: one corresponding to β-sn and the other to SnO 2 oxide β Sn Relative emission SnO Velocity [mm/s] Fig. 7. Conversion electron Mössbauer spectrum of metallic β-sn foil at room temperature; mono-channel analyzer selected the peak presented in figure 3. The parameters obtained for Sn are: resonance effect (ε) = %, line width (w) =.88 mm/s. These are better than those obtained in the transmission geometry: ε = 7.92 %; w =.97 mm/s. The smaller line width is expected in the backscattering geometry due to lack of saturation broadening. The backscattering measurement evidenced presence of the SnO 2 on the sample surface; this fact was not possibly in transmission measurement. The line width is closed to theoretical value. The thickness of the SnO 2 film is estimated around 4 nm [8]. The isomer shift of SnO 2, slight different from SnO 2 reference sample, shows a change in its stoichiometry. 1.6 Relative emission Velocity [mm/s] Fig. 8. Conversion X-ray Mössbauer spectrum of metallic β-sn foil at room temperature; monochannel analyzer selected the energy range kev.

7 156 I.Bibicu et al. 7 The Mössbauer spectrum obtained by the detection of X-ray from Sn is presented in the figure 8. It shows the presence of a single line, corresponding to β-sn. The resonance effect ε = 5.98 % is lower than the value obtained in the transmission geometry (7.9%). Also the line width is lower:.92 mm/s instead.98 mm/s obtained in the transmission geometry. After our knowledge the Mössbauer spectrum obtained by detection of low energy X-ray for 119 Sn is realized for the first time in the world. Fig. 9 shows the Mössbauer spectra of the Eu 2 O 3 sample obtained by detection of backscattered X-rays. The detection of the conversion X-rays compensate for smaller resonance effect by its significant smaller line width: 2.32 mm/s instead 3.24 mm/s in transmission for 1mg/cm 2 Eu 2 O 3 sample Relative emission Velocity [mm/s] Fig. 9. Conversion X-ray Mössbauer spectrum of Eu 2 O 3 at room temperature. 4. CONCLUSIONS A versatile counter for conversion Mössbauer spectroscopy at room temperature has been described and its performances were presented. The detector design is better than previously reported [9]-[11]. The device can be applied for both surface and bulk studies. It is suitable for studies with 57 Fe, 119 Sn and 151 Eu Mössbauer isotopes. Acknowledgements. The authors thank for the financial support of the National University Research Council Grant/ competition and National Authority for Scientific Research (ANCS, under Core Research Programme PN9/29-PN45/29). REFERENCES 1. R.L. Mössbauer, Nuclear Resonance Fluorescence of gamma Radiation in 191 Ir. Z. Physik, 151, p , E. Browne, R.B. Firestone, Table of radioactive Isotopes, Ed. John Wiley & Sons, 1986.

8 8 Counter for conversion Mössbauer spectroscopy Zs. Kajcsos, Ch. Sauer, W. Zinn, Criteria for optimizing DCEMS, Hyperfine Interactions, 57, p , Zs. Kajcsos, Ch. Sauer, W. Zinn, W. Meisel, H. Spiering, M. Alflen, P.Gütlich, High-performance Mössbauer spectroscopy: criteria, possibilities, limitations, Hyperfine Interactions, 71, , I. Bibicu, M.S. Rogalski, G. Nicolescu, A detector assembly for simultaneous conversion electron conversion X-ray and transmission Mössbauer spectroscopy, Meas. Sci. Technol 7, p , I. Bibicu, M.S. Rogalski, G. Nicolescu, An improved proportional counter for conversion and transmission Mössbauer spectroscopy, Rom. J. Phys. 45, no. 1 2, p , G.P. Huffman, Applications of Mössbauer Spectroscopy, Academic Press, Inc, New-York, 198, p Y. Isozumi, D.I. Lee, I. Kadar, A new detector assembly for conversion and X-rays from Mössbauer effect, Nucl. Instrum. Methods, 12, p , A.S. Kamzin, L.A. Grigoriev, Properties of surface layers and the interior of a crystal studied by Mössbauer spectroscopy, Sov. Tech.-Phys. Lett., 16, p , I. Bibicu, M.S. Rogalski, G. Nicolescu, Toroidal proportional detector for conversion X-ray and transmission Mössbauer spectroscopy, Nucl. Instr. and Meth. in Phys. Res., B 94, p , 1994.

GAMMA-RAY SPECTRA REFERENCES

GAMMA-RAY SPECTRA REFERENCES GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

Introduction to Geiger Counters

Introduction to Geiger Counters Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists

More information

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND: Gamma Rays OBJECT: To understand the various interactions of gamma rays with matter. To calibrate a gamma ray scintillation spectrometer, using gamma rays of known energy, and use it to measure the energy

More information

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Ronald M. Keyser, Timothy R. Twomey, Sam Hitch ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

Evaluation of various planar gaseous detectors with CsI photocathodes. for the detection of primary scintillation light from noble gases

Evaluation of various planar gaseous detectors with CsI photocathodes. for the detection of primary scintillation light from noble gases Evaluation of various planar gaseous detectors with CsI photocathodes for the detection of primary scintillation light from noble gases L. Periale 1,2, V. Peskov 3, P. Carlson 3, C. Iacobaeus 4, T. Francke

More information

Radiation Strip Thickness Measurement Systems

Radiation Strip Thickness Measurement Systems Radiation Strip Thickness Measurement Systems During the past years we have increased our sales of radiometric Vollmer strip thickness measurement systems, i.e. X-ray or isotope gauges, dramatically. Now,

More information

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy Equipment Needed from ORTEC GEM10P4/CFG-PV4/DWR-30 Coaxial Detector System (Includes detector, cryostat, dewar, preamplifier, and 12-ft. cable pack); typical specifications: 10% relative efficiency, 1.75

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system EDS system Most common X-Ray measurement system in the SEM lab. Major elements (10 wt% or greater) identified in ~10 secs. Minor elements identifiable in ~100 secs. Rapid qualitative and accurate quantitative

More information

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Lectures about XRF (X-Ray Fluorescence)

Lectures about XRF (X-Ray Fluorescence) 1 / 38 Lectures about XRF (X-Ray Fluorescence) Advanced Physics Laboratory Laurea Magistrale in Fisica year 2013 - Camerino 2 / 38 X-ray Fluorescence XRF is an acronym for X-Ray Fluorescence. The XRF technique

More information

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics PART I Geiger Tube: Optimal Operating Voltage and Resolving Time Objective: To become acquainted with the operation and characteristics

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

EDXRF of Used Automotive Catalytic Converters

EDXRF of Used Automotive Catalytic Converters EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,

More information

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

History of the Atom & Atomic Theory

History of the Atom & Atomic Theory Chapter 5 History of the Atom & Atomic Theory You re invited to a Thinking Inside the Box Conference Each group should nominate a: o Leader o Writer o Presenter You have 5 minutes to come up with observations

More information

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)

Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100) Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know

More information

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF Quantitative Analysis Software for X-Ray Fluorescence XRF-FP XRF-FP is a full-featured quantitative analysis package for XRF APPLICATIONS X-Ray Fluorescence Thin-film Analysis RoHS/WEEE Analysis Teaching

More information

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF)

Advanced Physics Laboratory. XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF) Advanced Physics Laboratory XRF X-Ray Fluorescence: Energy-Dispersive analysis (EDXRF) Bahia Arezki Contents 1. INTRODUCTION... 2 2. FUNDAMENTALS... 2 2.1 X-RAY PRODUCTION... 2 2. 1. 1 Continuous radiation...

More information

ENERGY LOSS OF ALPHA PARTICLES IN GASES

ENERGY LOSS OF ALPHA PARTICLES IN GASES Vilnius University Faculty of Physics Department of Solid State Electronics Laboratory of Applied Nuclear Physics Experiment No. ENERGY LOSS OF ALPHA PARTICLES IN GASES by Andrius Poškus (e-mail: andrius.poskus@ff.vu.lt)

More information

Radioactivity III: Measurement of Half Life.

Radioactivity III: Measurement of Half Life. PHY 192 Half Life 1 Radioactivity III: Measurement of Half Life. Introduction This experiment will once again use the apparatus of the first experiment, this time to measure radiation intensity as a function

More information

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.

More information

Composite Electromagnetic Wave Absorber Made of Permalloy or Sendust and Effect of Sendust Particle Size on Absorption Characteristics

Composite Electromagnetic Wave Absorber Made of Permalloy or Sendust and Effect of Sendust Particle Size on Absorption Characteristics PIERS ONLINE, VOL. 4, NO. 8, 2008 846 Composite Electromagnetic Wave Absorber Made of Permalloy or Sendust and Effect of Sendust Particle Size on Absorption Characteristics K. Sakai, Y. Wada, and S. Yoshikado

More information

ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY. Donald L. Horrocks

ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY. Donald L. Horrocks ENERGY PER PHOTOELECTRON IN A COINCIDENCE LIQUID SCINTILLATION COUNTER AS A FUNCTION OF ELECTRON ENERGY Donald L. Horrocks Nuclear Systems Operations Beckman Instruments, Inc. SmithKlirie Beckman Irvine,

More information

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY 243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution

More information

EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction

EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction ACTA PHYSICA DEBRECINA XLVI, 25 (2012) EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES L. Csedreki 1 1 Institute of Nuclear Research of the Hungarian Academy of Sciences,

More information

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons SLAC-PUB-7722 January 9 degrees Bremsstrahlung Source Term Produced in Thick Targets by 5 MeV to GeV Electrons X. S. Mao et al. Presented at the Ninth International Conference on Radiation Shielding, Tsukuba,

More information

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION David Bernard Dage Precision Industries Inc. Fremont, California d.bernard@dage-group.com ABSTRACT The x-ray inspection of PCB assembly processes

More information

Gamma and X-Ray Detection

Gamma and X-Ray Detection Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular

More information

Spectra of pyroelectric X-ray generator

Spectra of pyroelectric X-ray generator Abstract Spectra of pyroelectric X-ray generator Nagaychenko V.I., Sanin V.M., Yegorov A.M., Shchagin A.V. Kharkov Institute of Physics and Technology, Kharkov 6118, Ukraine The construction of X-ray generator

More information

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the

More information

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions Description The LS-1 Series of tungsten halogen light sources are versatile, white-light lamps optimized for use in the

More information

Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland Cements on Mass Attenuation Coefficient

Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland Cements on Mass Attenuation Coefficient Journal of Materials Science and Engineering A 3 (12) (2013) 838-842 D DAVID PUBLISHING Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland s on Mass Attenuation Coefficient Abd

More information

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

Spectral distribution from end window X-ray tubes

Spectral distribution from end window X-ray tubes Copyright (C) JCPDS-International Centre for Diffraction Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg 2 1 FORTEX - E.N.S.A.I.S. 24, bld de la Victoire,

More information

CHEM 1411 Chapter 5 Homework Answers

CHEM 1411 Chapter 5 Homework Answers 1 CHEM 1411 Chapter 5 Homework Answers 1. Which statement regarding the gold foil experiment is false? (a) It was performed by Rutherford and his research group early in the 20 th century. (b) Most of

More information

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is: 4 SENSORS The modern technical world demands the availability of sensors to measure and convert a variety of physical quantities into electrical signals. These signals can then be fed into data processing

More information

Introduction to Powder X-Ray Diffraction History Basic Principles

Introduction to Powder X-Ray Diffraction History Basic Principles Introduction to Powder X-Ray Diffraction History Basic Principles Folie.1 History: Wilhelm Conrad Röntgen Wilhelm Conrad Röntgen discovered 1895 the X-rays. 1901 he was honoured by the Noble prize for

More information

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse SURFACE CHARACTERIZATION Marie-Geneviève Centre d Etudes de Chimie Métallurgique, CNRS, Vitry-sur-Seine, France Keywords: Surface Analysis, Surface imaging, Surface composition, Surface chemical analysis,

More information

Vacuum Evaporation Recap

Vacuum Evaporation Recap Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.

More information

Tutorial 4.6 Gamma Spectrum Analysis

Tutorial 4.6 Gamma Spectrum Analysis Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning

More information

Polymer growth rate in a wire chamber with oxygen, water, or alcohol gas additives

Polymer growth rate in a wire chamber with oxygen, water, or alcohol gas additives SLAC-PUB-13 June 6, 8 Polymer growth rate in a wire chamber with oxygen, water, or alcohol gas additives Adam M. Boyarski Stanford Linear Accelerator Center, M.S. 95, 575 Sand Hill Rd, Menlo Park, CA 95,

More information

Time out states and transitions

Time out states and transitions Time out states and transitions Spectroscopy transitions between energy states of a molecule excited by absorption or emission of a photon hn = DE = E i - E f Energy levels due to interactions between

More information

Raman spectroscopy Lecture

Raman spectroscopy Lecture Raman spectroscopy Lecture Licentiate course in measurement science and technology Spring 2008 10.04.2008 Antti Kivioja Contents - Introduction - What is Raman spectroscopy? - The theory of Raman spectroscopy

More information

Construction of an Alpha- Beta and Gamma-Sensitive Radiation Detector on the Basis of a Low-Cost PIN-Diode

Construction of an Alpha- Beta and Gamma-Sensitive Radiation Detector on the Basis of a Low-Cost PIN-Diode Construction of an Alpha- Beta and Gamma-Sensitive Radiation Detector on the Basis of a Low-Cost PIN-Diode Bernd Laquai, 12.6.2012 Encouraged by the observation, that the Am241 alpha source of an old smoke

More information

Atomic Calculations. 2.1 Composition of the Atom. number of protons + number of neutrons = mass number

Atomic Calculations. 2.1 Composition of the Atom. number of protons + number of neutrons = mass number 2.1 Composition of the Atom Atomic Calculations number of protons + number of neutrons = mass number number of neutrons = mass number - number of protons number of protons = number of electrons IF positive

More information

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.

Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation

More information

Preface Light Microscopy X-ray Diffraction Methods

Preface Light Microscopy X-ray Diffraction Methods Preface xi 1 Light Microscopy 1 1.1 Optical Principles 1 1.1.1 Image Formation 1 1.1.2 Resolution 3 1.1.3 Depth of Field 5 1.1.4 Aberrations 6 1.2 Instrumentation 8 1.2.1 Illumination System 9 1.2.2 Objective

More information

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

Experiment 5. Lasers and laser mode structure

Experiment 5. Lasers and laser mode structure Northeastern University, PHYS5318 Spring 2014, 1 1. Introduction Experiment 5. Lasers and laser mode structure The laser is a very important optical tool that has found widespread use in science and industry,

More information

X-ray Spectroscopy. 1. Introduction. 2. X-Ray Spectra of the Elements. Physics 441-442 Advanced Physics Laboratory

X-ray Spectroscopy. 1. Introduction. 2. X-Ray Spectra of the Elements. Physics 441-442 Advanced Physics Laboratory University of Michigan February 2005 Physics 441-442 Advanced Physics Laboratory X-ray Spectroscopy 1. Introduction X-rays are KeV photons. Atomic X-rays are emitted during electronic transitions to the

More information

Measurement of Germanium Detector Efficiency

Measurement of Germanium Detector Efficiency Measurement of Germanium Detector Efficiency Marcus H. Wiggs 2009 Notre Dame Physics REU Advisor: Dr. Philippe Collon Mentors: Matthew Bowers, Daniel Robertson, Chris Schmitt ABSTRACT: A possible discrepancy

More information

Coating Technology: Evaporation Vs Sputtering

Coating Technology: Evaporation Vs Sputtering Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information

More information

Sample preparation for X-ray fluorescence analysis

Sample preparation for X-ray fluorescence analysis Technical articles Sample preparation for X-ray fluorescence analysis III. Pressed and loose powder methods Gakuto Takahashi* 1. Introduction There are two main sample preparation techniques for measurement

More information

Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM

Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM Original Paper Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM Satoshi Hashimoto 1, Tsuguo Sakurada 1, and Minoru Suzuki 2 1 JFE-Techno research corporation, 1-1 Minamiwatarida, Kawasaki,

More information

Infrared Spectroscopy 紅 外 線 光 譜 儀

Infrared Spectroscopy 紅 外 線 光 譜 儀 Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed

More information

Active noise control in practice: transformer station

Active noise control in practice: transformer station Active noise control in practice: transformer station Edwin Buikema 1 ; Fokke D. van der Ploeg 2 ; Jan H. Granneman 3 1, 2, 3 Peutz bv, Netherlands ABSTRACT Based on literature and extensive measurements

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING Brian L. Riise and Michael B. Biddle MBA Polymers, Inc., Richmond, CA, USA Michael M. Fisher American Plastics Council, Arlington, VA, USA X-Ray Fluorescence

More information

How to measure absolute pressure using piezoresistive sensing elements

How to measure absolute pressure using piezoresistive sensing elements In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article

More information

GRETA. I-Yang Lee. For the GRETA Steering Committee. Workshop on the Experimental Equipment for RIA March 18-22, 2003, Oak Ridge, Tennessee

GRETA. I-Yang Lee. For the GRETA Steering Committee. Workshop on the Experimental Equipment for RIA March 18-22, 2003, Oak Ridge, Tennessee GRETA I-Yang Lee For the GRETA Steering Committee Workshop on the Experimental Equipment for RIA March 18-22, 2003, Oak Ridge, Tennessee Gamma-ray Tracking Concepts Compton Suppressed Ge Veto Ν det = 100

More information

3 - Atomic Absorption Spectroscopy

3 - Atomic Absorption Spectroscopy 3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,

More information

AS COMPETITION PAPER 2008

AS COMPETITION PAPER 2008 AS COMPETITION PAPER 28 Name School Town & County Total Mark/5 Time Allowed: One hour Attempt as many questions as you can. Write your answers on this question paper. Marks allocated for each question

More information

Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise

Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise TDK EMC Technology Product Section Clamp Filters that Suppress Emission Noise Provide Immunity Against Surge Noise TDK Shonai Corporation Satoru Saito Reduce Emission Noise from Cables Even if an electronic

More information

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the discussion on the major components of the x-ray machine. Today

More information

ABSORPTION OF BETA AND GAMMA RADIATION

ABSORPTION OF BETA AND GAMMA RADIATION ABSORPTION OF BETA AND GAMMA RADIATION The purpose of this experiment is to understand the interaction of radiation and matter, and the application to radiation detection and shielding Apparatus: 137 Cs

More information

0 10 20 30 40 50 60 70 m/z

0 10 20 30 40 50 60 70 m/z Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions

More information

- thus, the total number of atoms per second that absorb a photon is

- thus, the total number of atoms per second that absorb a photon is Stimulated Emission of Radiation - stimulated emission is referring to the emission of radiation (a photon) from one quantum system at its transition frequency induced by the presence of other photons

More information

The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I

The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I 5 The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I Mg Kα excited photoemission. Abstract The oxidation of Cu(100)-Fe surfaces was studied using XPS. Surfaces

More information

NMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE. 4 3 2 1 0 δ PUBLISHING

NMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE. 4 3 2 1 0 δ PUBLISHING A N I N T R O D U T I O N T O... NMR SPETROSOPY NULEAR MAGNETI RESONANE 4 3 1 0 δ Self-study booklet PUBLISING NMR Spectroscopy NULEAR MAGNETI RESONANE SPETROSOPY Origin of Spectra Theory All nuclei possess

More information

Gamma Ray Detection at RIA

Gamma Ray Detection at RIA Gamma Ray Detection at RIA Summary Report: Physics & Functional Requirements Cyrus Baktash Physics goals Experimental tools: Techniques & Reactions Functional Requirements Physics Questions (Discussed

More information

Ion Beam Sputtering: Practical Applications to Electron Microscopy

Ion Beam Sputtering: Practical Applications to Electron Microscopy Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a

More information

Infrared Spectroscopy: Theory

Infrared Spectroscopy: Theory u Chapter 15 Infrared Spectroscopy: Theory An important tool of the organic chemist is Infrared Spectroscopy, or IR. IR spectra are acquired on a special instrument, called an IR spectrometer. IR is used

More information

Physics 30 Worksheet # 14: Michelson Experiment

Physics 30 Worksheet # 14: Michelson Experiment Physics 30 Worksheet # 14: Michelson Experiment 1. The speed of light found by a Michelson experiment was found to be 2.90 x 10 8 m/s. If the two hills were 20.0 km apart, what was the frequency of the

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier

(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (no PiN and pinned Diodes) Peter Fischer P. Fischer, ziti, Uni Heidelberg, Seite 1 Overview Reminder: Classical Photomultiplier

More information

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment.

Helium-Neon Laser. Figure 1: Diagram of optical and electrical components used in the HeNe laser experiment. Helium-Neon Laser Experiment objectives: assemble and align a 3-mW HeNe laser from readily available optical components, record photographically the transverse mode structure of the laser output beam,

More information

Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser

Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser V.Nassisi #, G.Caretto #, A. Lorusso #, D.Manno %, L.Famà %, G.Buccolieri %, A.Buccolieri %, U.Mastromatteo* # Laboratory of Applied

More information

Production of X-rays and Interactions of X-rays with Matter

Production of X-rays and Interactions of X-rays with Matter Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage

More information

Proton Nuclear Magnetic Resonance Spectroscopy

Proton Nuclear Magnetic Resonance Spectroscopy Proton Nuclear Magnetic Resonance Spectroscopy Introduction: The NMR Spectrum serves as a great resource in determining the structure of an organic compound by revealing the hydrogen and carbon skeleton.

More information

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection Scott A Speakman, Ph.D Center for Materials Science and Engineering at MIT speakman@mit.edu

More information

CHAPTER 7 SCINTILLATION COUNTING

CHAPTER 7 SCINTILLATION COUNTING CHAPTER 7 SCINTILLATION COUNTING Radiation of various types is widely utilized for non-destructive inspection and testing such as in medical diagnosis, industrial inspection, material analysis and other

More information

On the first Townsend coefficient at high electric field

On the first Townsend coefficient at high electric field 1 On the first Townsend coefficient at high electric field Yu.I. Davydov Joint Institute for Nuclear Research, 141980, Dubna, Russia arxiv:physics/0409156v2 [physics.ins-det] 18 Sep 2006 Abstract Based

More information

MONTE CARLO SIMULATION OF THE FULL MULTIPLICITY DISTRIBUTIONS MEASURED WITH A PASSIVE COUNTER

MONTE CARLO SIMULATION OF THE FULL MULTIPLICITY DISTRIBUTIONS MEASURED WITH A PASSIVE COUNTER International Conference on Mathematics, Computational Methods & Reactor Physics (M&C 9) Saratoga Springs, New York, May 3-7, 9, on CD-ROM, American Nuclear Society, LaGrange Park, IL (9) MONTE CARLO SIMULATION

More information

Radiation and the Universe Higher Exam revision questions and answers

Radiation and the Universe Higher Exam revision questions and answers Radiation and the Universe Higher Exam revision questions and answers Madeley High School Q.The names of three different processes are given in List A. Where these processes happen is given in List B.

More information

Energy Transport. Focus on heat transfer. Heat Transfer Mechanisms: Conduction Radiation Convection (mass movement of fluids)

Energy Transport. Focus on heat transfer. Heat Transfer Mechanisms: Conduction Radiation Convection (mass movement of fluids) Energy Transport Focus on heat transfer Heat Transfer Mechanisms: Conduction Radiation Convection (mass movement of fluids) Conduction Conduction heat transfer occurs only when there is physical contact

More information

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this

More information

PoS(PhotoDet 2012)068

PoS(PhotoDet 2012)068 Characterization of the Hamamatsu R11265 multi-anode photomultiplier tube with single photon signals Luca CADAMURO, Marta CALVI, Andrea GIACHERO,, Matteo MAINO, Clara MATTEUZZI, Gianluigi PESSINA Dipartimento

More information

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading:

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading: Applied Spectroscopy Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading: Banwell and McCash Chapter 7 Skoog, Holler Nieman Chapter 19 Atkins, Chapter 18 Relaxation processes We need

More information

where h = 6.62 10-34 J s

where h = 6.62 10-34 J s Electromagnetic Spectrum: Refer to Figure 12.1 Molecular Spectroscopy: Absorption of electromagnetic radiation: The absorptions and emissions of electromagnetic radiation are related molecular-level phenomena

More information

High resolution and two-dimensional focal plane detector for intermediate energy heavy ions

High resolution and two-dimensional focal plane detector for intermediate energy heavy ions INS Rep.-1088 INSTITUTE FOR NUCLEAR STUDY Jan. 1995 UNIVERSITY OF TOKYO T u- T i,an INS 1088 Tanashi, Tokyo 188 jp9503014 Japan High resolution and two-dimensional focal plane detector for intermediate

More information

Atomic Structure: Chapter Problems

Atomic Structure: Chapter Problems Atomic Structure: Chapter Problems Bohr Model Class Work 1. Describe the nuclear model of the atom. 2. Explain the problems with the nuclear model of the atom. 3. According to Niels Bohr, what does n stand

More information

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation? From lowest energy to highest energy, which of the following correctly

More information