1 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner CSIRO Minerals, PMB 5, Menai 2234, NSW, Australia. An XRF analyser suitable for measuring up to 15 liters of slurry has been tested in the laboratory. A range of slurries consisting of approximately 40 wt% solids and low concentrations of tantalum were obtained and measured. The accuracy of the analyser has been determined to be ±2 ppm (1 sigma). Improvements are planned that will significantly increase the speed and accuracy of the analyser and facilitate its adaptation to online / byline monitoring of mineral process streams. INTRODUCTION CSIRO Minerals is developing methods for the on-stream measurement of parts-permillion (ppm) concentrations of heavy metals in process slurries using K-shell X-ray fluorescence (XRF). As part of this ongoing development, an investigation was carried out on mineral slurries containing approximately 100 ppm of tantalum (Tantalum concentrations are reported as tantalum oxide (Ta2O5) as a proportion of the dry solids mass). MEASUREMENT OF SLURRIES WITH THE CSIRO ANALYSER Sample provision and third-party analysis CSIRO Minerals was supplied with eight samples from a tantalum production facility. Each sample consisted of at least 10 liters of slurry, containing wt% solids. The solids content was determined by drying the samples in an oven at 110 C for at least 2 days and weighing the residue. After each slurry had been dried and sub-sampled for analytical XRF analysis, the remaining dry ore was weighed and reconstituted with water for measurement in the CSIRO analyser. The original set of only eight slurries was considered too small for multivariate analysis. Therefore, additional samples were prepared based on this set. Firstly, the tantalum concentrations of two of the samples were increased in 5 ppm intervals by the addition of Ta2O5 powder. Secondly, iron oxide (Fe2O3) was added to one sample to increase its iron content to 2, 4 and finally 8 wt%. These additions were made to investigate the effects of changing matrix composition on the Ta measurement. In this way, the core set of 8 samples was expanded to 22 samples suitable for measurement.

2 This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website ICDD Website -

3 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume The tantalum content of all samples was measured independently by three separate facilities using differing techniques, namely laboratory XRF analysis, neutron activation analysis (NAA) and ICP. A Grubbs [1] intercomparison analysis was performed and the most accurate measurement technique was determined to be the XRF analytical measurements with an estimated repeatability error of 1.6 ppm. The ICP measurements were found to have systematic errors related to the difficulty of completely dissolving all the tantalum into solution, whereas the NAA results showed relatively large statistical errors and hence neither were used in any further analysis. The CSIRO XRF Analyser The CSIRO XRF slurry analyzer is illustrated in figure 1. Two counter-rotating mixers are used to ensure slurry uniformity. The rig is designed to allow a large volume of slurry (approximately 10 liters) to be measured. The radiation source is a 37 MBq (1 mci) Co- 57 radioisotope source. The resulting scattered and fluorescent X-rays are transmitted through a polycarbonate window and are detected by a high resolution X-ray detector. The characteristic tantalum K X-rays are used to estimated the tantalum content rather than the L X-rays. The K X-rays are more suitable for two reasons: Firstly, they are more highly penetrating allowing a much larger volume to be analysed; and secondly they are also much less sensitive to matrix and particle size effects. Slurry sample Mixers Polycarbonate window Incident γ-ray Scattered X-ray Collimator X-ray detector in vacuum housing γ-ray source Figure 1. Schematic drawing of laboratory rig. A high-purity, planar germanium detector (Canberra 1000 mm2 LEGe) was used to detect the scattered and fluorescent X-rays, read-out using a Aptec 5004 combination shaping amplifier/adc/mca card. This was operated at a modest count rate (limited by the intensity of the source) of approximately 30k c/s with a corresponding dead time of 25%. Each measurement took approximately 5½ hours.

4 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume Uniformity of the Slurry During Measurement The most difficult problem faced in obtaining repeatable measurements was ensuring that the slurry was homogeneous. In particular, it was not sufficient just to lift all the solids off the bottom of the tank if a density profile remained where the larger particles were found near the bottom of the tank. This is important since the rig is more sensitive to material within a few centimeters of the source and detector. In the slurry samples used, the tantalum is not uniformly distributed with respect to particle size in fact there are systematic variations! These introduce significant errors in tantalum determination if the slurry is not well mixed. Calculation of Tantalum Concentration and Solids Mass Fraction Raw XRF Spectra Figure 2 shows a typical X-ray spectrum for a slurry containing approximately 100 ppm of tantalum. Each channel is 100 ev wide. The broad peak centered at 85 kev is due to Compton scattered source γ-rays, with the smaller peaks at channels 122 and kev due to Raleigh scattered source γ-rays. 4 x e l n a h pe rc ts C oun Channel number (100 ev bins) Figure 2. An X-ray spectrum obtained from measurement of a slurry containing approximately 100 ppm of Ta2O5. Figure 3 is an enlarged view of the spectrum in the vicinity of the tantalum Ka X-ray peaks. The tantalum Ka2 and Ka1 peaks are centered on channels 563 and 575.

5 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume x e l n a h pe rc ts C oun Channel number (100 ev bins) Tantalum Determination Figure 3. An enlarged view of the measured XRF spectrum in the vicinity of the tantalum Ka peaks. A four term linear regression analysis was used to relate the tantalum content of each sample to features observed in the X-ray spectrum. The terms used were the number of counts in the tantalum peaks, the background in the vicinity of the peaks, the number of counts in the Compton scattering peak, and a small term necessary to correct for a monotonic drift in the electronics with respect to time. To avoid over-fitting on the relatively small data set used, the 22 samples measured were divided into a training set (consisting of 21 points) used to determine the regression coefficients, and a test set (1 point) used to evaluate the performance of the calibration. The calibration/training process was cycled through all 22 of the samples. The presence of tungsten at 0 10 ppm levels was revealed by the NAA data. We were able to correct for any interference satisfactorily by estimating the background as the sum of three different windows two of which straddled the tungsten Ka1 peak. Solids Fraction Determination To determine the solids loading, a multivariate linear regression analysis was used to relate information on the shape of the Compton scattering peak to the solids fraction. The Compton peak shape is expected to vary with solids fraction for two reasons: 1. Doppler broadening of Compton scattered X-rays depends on the atomic number (Z) of the element from which the X-ray scatters, with higher Z elements having broader Compton scattering peaks. Compton profile analysis [2] has been shown to provide

6 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume good discrimination between the lightest elements (such as H and O) and heavier elements present in the tantalum ore. 2. The proportion of multiply to singly Compton scattered photons decreases with increasing solids loading fraction due to photoelectric absorption. This measurement technique has the advantage that no additional apparatus is needed and that the on-line mass measurement is performed over almost an identical region of slurry to which the tantalum concentration is being measured. RESULTS Tantalum Concentration Figure 4 shows a plot of the CSIRO calculated tantalum versus the analytical XRF laboratory determined values. The horizontal error bars represent the uncertainty of the analytical determination, and the vertical error bars on each point are the combined CSIRO and analytical XRF error. Figure 4. Plot of tantalum concentration for the CSIRO feasibility rig versus the analytical XRF concentration. The RMS error of 2.3 ppm (1 sigma) is the root mean square error between the CSIRO and analytical XRF analyses. It is a measure of the total error including the estimated 1.6 ppm repeatability error of the analytical analysis. Correcting for this results in an estimated error for the CSIRO analyzer of 1.7 ppm. There is always a degree of uncertainty in apportioning errors between two measurements and a more conservative claim would be that our data is consistent with a 1 standard deviation error of approximately 2 ppm.

7 Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume The 2.3 ppm RMS error between the analytical XRF results and the CSIRO results is better than the agreement between analytical XRF and either the neutron activation analysis (about 3 ppm) or the ICP (>10 ppm) results. Whilst our results are very encouraging, it should be noted that they have been obtained on a fairly limited suite of measurements, based on just 8 discrete samples. Further measurements are required to properly evaluate other factors contributing to the analyzer accuracy, notably matrix and particle size effects. Solids Mass Fraction The on-line determinations of the solids mass fraction were compared with the true mass fractions, calculated from the masses of solid and water used to reconstitute the slurries. The measurement accuracy for the solid fraction determination was found to be about 0.3 wt%, with a repeatability error of 0.16 wt%. For a typical solids loading of about 40 wt%, this leads to a 0.7% relative error in the determination of the Ta2O5 content, expressed as a fraction of the dry mass. CONCLUSION Measurements undertaken by CSIRO Minerals at the Lucas Heights Laboratory have shown that the current XRF rig is capable of accurately measuring the tantalum content of slurries. An accuracy of approximately 2 ppm was obtained for a measurement time of 5½ hours. Several improvements to the analyzer are planned. These include: Using a strong radioisotope source together with a high-speed digital X-ray processor to increase the X-ray counting rate; Altering the geometry of the analyzer to measure slurry passing through a pipeline. This improves the signal-to-background ratio for the tantalum X-rays and minimizes the sensitivity of the analyzer to the spatial distribution of the solids fraction. It is estimated that it will be possible to achieve a statistical error on the measurement of Ta2O5 of approximately 1 ppm with a 20 minute counting time. In principle, the same technique can be used to determine other high Z elements to similar accuracy. ACKNOWLEDGEMENTS The authors would like to thank Steve Rainey and Ivan Kekic for assistance in the design and construction of the slurry analyser. REFERENCES [1] Grubbs F, 1948, On Estimating Precision of Measuring Instruments and Product Variability, J Amer. Statistical Assoc., 43, [2] Tickner J and Roach G, Characterization of coal and minerals using Compton Profile Analysis, to appear in Nuclear Instruments and Methods B

AN INNOVATED LABORATORY XAFS APPARATUS

AN INNOVATED LABORATORY XAFS APPARATUS Copyright (c)jcpds-international Centre for Diffraction Data 2002, Advances in X-ray Analysis, Volume 45. 397 AN INNOVATED LABORATORY XAFS APPARATUS TAGUCHI Takeyoshi XRD Division, Rigaku Corporation HARADA

More information

SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS

SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 38 SPECTRAL INTERFERENCE IN X-RAY FLUORESCENCE ANALYSIS OF COMMON MATERIALS Frank R. Feret and Hafida

More information

Introduction to the QuanX

Introduction to the QuanX Introduction to the QuanX Energy-Dispersive X-ray Fluorescence Spectrometry Theory Introduction Overview: Theory Presentation EDS overview Basic overview History of X-ray spectrometry Atomic Structure

More information

EDXRF of Used Automotive Catalytic Converters

EDXRF of Used Automotive Catalytic Converters EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,

More information

Nuclear Medicine Imaging Systems: The Scintillation Camera

Nuclear Medicine Imaging Systems: The Scintillation Camera Radiation Detection and Measurement Range of charged particles (e.g.,!: µm; ": mm) Range of high energy photons (cm) Two main types of interactions of high energy photons Compton scatter Photoelectric

More information

1- AND 2-DIMENSIONAL DETECTION SYSTEMS AND THE PROBLEM OF SAMPLE FLUORESCENCE IN X-RAY DIFFRACTOMETRY

1- AND 2-DIMENSIONAL DETECTION SYSTEMS AND THE PROBLEM OF SAMPLE FLUORESCENCE IN X-RAY DIFFRACTOMETRY Copyright JCPDS - International Centre for Diffraction Data 24, Advances in X-ray Analysis, Volume 47. 224 1- AND 2-DIMENSIONAL DETECTION SYSTEMS AND THE PROBLEM OF SAMPLE FLUORESCENCE IN X-RAY DIFFRACTOMETRY

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY

CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY 243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution

More information

GAMMA-RAY SPECTRA REFERENCES

GAMMA-RAY SPECTRA REFERENCES GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.

More information

X-RAY AND GAMMA RAY SPECTRA

X-RAY AND GAMMA RAY SPECTRA Experiment N3 X-RAY AND GAMMA RAY SPECTRA References: Handbook of Chemistry and Physics Nuclear Level Schemes, A=45 Through A=257, QC795.8.E5.N8 Background: As in case of atomic physics, much of what we

More information

Gamma Ray Attenuation Properties of Common Shielding Materials

Gamma Ray Attenuation Properties of Common Shielding Materials Gamma Ray Attenuation Properties of Common Shielding Materials Daniel R. McAlister, Ph.D. PG Research Foundation, Inc. 955 University Lane Lisle, IL 60532, USA Introduction In areas where people are likely

More information

EFFICIENCY FOR CLOSE GEOMETRIES AND EXTENDED SOURCES OF A P-TYPE GERMANIUM DETECTOR WITH LOW-ENERGY SENSITIVITY.

EFFICIENCY FOR CLOSE GEOMETRIES AND EXTENDED SOURCES OF A P-TYPE GERMANIUM DETECTOR WITH LOW-ENERGY SENSITIVITY. Abstract EFFICIENCY FOR CLOSE GEOMETRIES AND EXTENDED SOURCES OF A P-TYPE GERMANIUM DETECTOR WITH LOW-ENERGY SENSITIVITY R.M. Keyser (1) and T.R. Twomey (1) (1) ORTEC, 801 South Illinois Avenue, Oak Ridge,

More information

Method for Dual High Energy X-ray Imaging with Flat Panel Detectors

Method for Dual High Energy X-ray Imaging with Flat Panel Detectors 18th World Conference on Nondestructive Testing, 16-20 April 2012, Durban, South Africa Method for Dual High Energy X-ray Imaging with Flat Panel Detectors Markus FIRSCHING 1, Frank NACHTRAB 1, Theobald

More information

Quantifying Amorphous Phases. Kern, A., Madsen, I.C. and Scarlett, N.V.Y.

Quantifying Amorphous Phases. Kern, A., Madsen, I.C. and Scarlett, N.V.Y. Kern, A., Madsen, I.C. and Scarlett, N.V.Y. This document was presented at PPXRD - Pharmaceutical Powder X-ray Diffraction Symposium Sponsored by The International Centre for Diffraction Data This presentation

More information

Spectral distribution from end window X-ray tubes

Spectral distribution from end window X-ray tubes Copyright (C) JCPDS-International Centre for Diffraction Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg 2 1 FORTEX - E.N.S.A.I.S. 24, bld de la Victoire,

More information

emission of light from atoms discrete line spectra - energy levels, Franck-Hertz experiment

emission of light from atoms discrete line spectra - energy levels, Franck-Hertz experiment Introduction Until the early 20 th century physicists used to explain the phenomena in the physical world around them using theories such a mechanics, electromagnetism, thermodynamics and statistical physics

More information

Introduction to Geiger Counters

Introduction to Geiger Counters Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists

More information

Basic Fundamental Parameters in X-Ray Fluorescence

Basic Fundamental Parameters in X-Ray Fluorescence 46 Spectroscopy 22(5) May 2007 Basic Fundamental Parameters in X-Ray Fluorescence The fundamental parameters approach to calibration in X-ray fluorescence is unique because it is based upon the theoretical

More information

Introduction to the Monte Carlo method

Introduction to the Monte Carlo method Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw

More information

EXPERIMENT 1. Gamma-Ray Absorption and Counting Statistics. Introduction. Note: Please read the Radiation Safety Regulations at the back of this book

EXPERIMENT 1. Gamma-Ray Absorption and Counting Statistics. Introduction. Note: Please read the Radiation Safety Regulations at the back of this book EXPERIMENT 1 Gamma-Ray Absorption and Counting Statistics Note: Please read the Radiation Safety Regulations at the back of this book Introduction The detecting medium in a scintillation counter is a solid

More information

PHY 192 Compton Effect 1. cos Θ = 1 m e E Θ

PHY 192 Compton Effect 1. cos Θ = 1 m e E Θ PHY 192 Compton Effect 1 The Compton Effect Introduction In this experiment we will study two aspects of the interaction of photons with electrons. The first of these is the Compton effect named after

More information

SH1009, Modern Fysik. X-Ray Diffraction Muhammad Yasir, Lway al Maeeni, Joakim Wahlström

SH1009, Modern Fysik. X-Ray Diffraction Muhammad Yasir, Lway al Maeeni, Joakim Wahlström SH1009, Modern Fysik X-Ray Diffraction Muhammad Yasir, Lway al Maeeni, Joakim Wahlström 24 th april, 2009 Introduction In this lab we study the x-rays and their ability to penetrate materials. We also

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Application Note 15: The use of a Near Infrared Transmission Analyser for Grape and Wine Quality Testing

Application Note 15: The use of a Near Infrared Transmission Analyser for Grape and Wine Quality Testing Application Note 15: The use of a Near Infrared Transmission Analyser for Grape and Wine Quality Testing 1. Introduction 1.1 Background The use of spectroscopy in the wine industry is well established,

More information

A HIGH-TEMPERATURE POWDER DIFFRACTION FURNACE

A HIGH-TEMPERATURE POWDER DIFFRACTION FURNACE Copyright JCPDS - International Centre for Diffraction Data 2003, Advances in X-ray Analysis, Volume 46. 50 A HIGH-TEMPERATURE POWDER DIFFRACTION FURNACE M. D. Dolan, S. Zdzieszynski, and S. T. Misture

More information

X-rays Images. Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department

X-rays Images. Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department Prof. Dr. Lucas Ferrari de Oliveira UFPR Informatics Department Discovery: German physicist Wilhelm Röntgen in 1895; "X-rays" signify an unknown quantity; X-rays were found emanating from Crookes tubes

More information

Spectroscopy II - Transmission, Absorption, Fluorescence, Beer s law and Scattering.

Spectroscopy II - Transmission, Absorption, Fluorescence, Beer s law and Scattering. Spectroscopy II - Transmission, Absorption, Fluorescence, Beer s law and Scattering. Spectrophotometry and Fluorometry In this experiment you will use the Vernier SpectroVis Plus, a small, computer controlled

More information

CHAPTER I INTRODUCTION

CHAPTER I INTRODUCTION CHAPTER I INTRODUCTION 1.0 Introduction Neutron activation analysis (NAA) was first discovered in 1936 and offers excellent sensitivities and is superior to other analytical techniques. NAA perform both

More information

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF

Quantitative Analysis Software for X-Ray Fluorescence. XRF-FP is a full-featured quantitative analysis package for XRF Quantitative Analysis Software for X-Ray Fluorescence XRF-FP XRF-FP is a full-featured quantitative analysis package for XRF APPLICATIONS X-Ray Fluorescence Thin-film Analysis RoHS/WEEE Analysis Teaching

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

HOW SENSITIVE IS THE GAMMA-RAD?

HOW SENSITIVE IS THE GAMMA-RAD? HOW SENSITIVE IS THE GAMMA-RAD? This is one of the most commonly asked questions. Sensitivity is very important, arguably the most important performance parameter in many applications of gamma-ray spectroscopy.

More information

FUSED BEADS FROM METAL DRILLINGS AND CHIPS EXAMPLE OF APPLICATION TO TIN-LEAD ALLOYS

FUSED BEADS FROM METAL DRILLINGS AND CHIPS EXAMPLE OF APPLICATION TO TIN-LEAD ALLOYS Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 405 FUSED BEADS FROM METAL DRILLINGS AND CHIPS EXAMPLE OF APPLICATION TO TIN-LEAD ALLOYS Jimmy Blanchette

More information

Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist

Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist Basic Concepts of X-ray X Fluorescence by Miguel Santiago, Scientific Instrumentation Specialist University of Puerto Rico Mayagüez Campus Department of Geology X-ray / Microanalysis Laboratory F - 304

More information

XRF MAPPING: NEW TOOLS FOR DISTRIBUTION ANALYSIS

XRF MAPPING: NEW TOOLS FOR DISTRIBUTION ANALYSIS Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 19 Copyright(C)JCPDS-International Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE

CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE 218 CHARACTERIZING PROCESS SEMICONDUCTOR THIN FILMS WITH A CONFOCAL MICRO X-RAY FLUORESCENCE MICROSCOPE Chris M. Sparks 1, Elizabeth P. Hastings 2, George J. Havrilla 2, and Michael Beckstead 2 1. ATDF,

More information

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems

Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Ronald M. Keyser, Timothy R. Twomey, Sam Hitch ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831

More information

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.

More information

Spectrophotometry. Theory

Spectrophotometry. Theory is any technique that uses light to measure chemical concentrations. Theory When a sample absorbs electromagnetic radiation it undergoes a change in energy. The interaction happens between the sample and

More information

Direct Determination of Phosphorus Content in Electroless Nickel Plating Using X-ray Fluorescence (XRF) Spectroscopy

Direct Determination of Phosphorus Content in Electroless Nickel Plating Using X-ray Fluorescence (XRF) Spectroscopy Direct Determination of Phosphorus Content in Electroless Nickel Plating Using X-ray Fluorescence (XRF) Spectroscopy Michael Haller Jim Bogert, Ryan Boyle Fischer Technology, Windsor CT, USA Volker Rößiger,

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

Analytical Method Validation

Analytical Method Validation Analytical Method Validation A. Es-haghi Ph.D. Dept. of Physico chemistry vaccine and serum research institute a.eshaghi@rvsri.ir http://www.rvsri.ir/ Introduction Test procedures for assessment of the

More information

APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS

APPROACHING A UNIVERSAL SAMPLE PREPARATION METHOD FOR XRF ANALYSIS OF POWDER MATERIALS Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 1 Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 367

More information

Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy

Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy Objectives In this lab, you will use fluorescence spectroscopy to determine the mass of riboflavin in a vitamin pill. Riboflavin fluorescence

More information

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,

More information

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9 Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation

More information

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system EDS system Most common X-Ray measurement system in the SEM lab. Major elements (10 wt% or greater) identified in ~10 secs. Minor elements identifiable in ~100 secs. Rapid qualitative and accurate quantitative

More information

Radiation Detection and Measurement

Radiation Detection and Measurement Radiation Detection and Measurement June 2008 Tom Lewellen Tkldog@u.washington.edu Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron

More information

Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy**

Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy** Analysis of Riboflavin in a Vitamin Pill by Fluorescence Spectroscopy** Objectives In this lab, you will use fluorescence spectroscopy to determine the mass and percentage of riboflavin in a vitamin pill.

More information

Outline. Exponential Attenuation. Introduction. Simple exponential attenuation. Simple exponential attenuation. Simple exponential attenuation

Outline. Exponential Attenuation. Introduction. Simple exponential attenuation. Simple exponential attenuation. Simple exponential attenuation Outline Exponential Attenuation Chapter 3 F.A. Attix, Introduction to Radiological Physics and Radiation Dosimetry Simple exponential attenuation and plural modes of absorption arrow-beam vs. broad-beam

More information

Calculating standard captured spectra of formation elements

Calculating standard captured spectra of formation elements Pet.Sci.(2012)9:463-468 1007/s12182-012-0231-8 463 Calculating standard captured spectra of formation elements Wu Wensheng, Xiao Lizhi, Zhang Lijuan and Niu Wei State Key Laboratory of Petroleum Resource

More information

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.

More information

CALIBRATION OF A HIGH TEMPERATURE X-RAY DIFFRACTION STAGE BY DIFFERENTIAL THERMAL EXPANSION A.R. Drews

CALIBRATION OF A HIGH TEMPERATURE X-RAY DIFFRACTION STAGE BY DIFFERENTIAL THERMAL EXPANSION A.R. Drews Copyright(c)JCPDS-International Centre for Diffraction Data 2001,Advances in X-ray Analysis,Vol.44 44 CALIBRATION OF A HIGH TEMPERATURE X-RAY DIFFRACTION STAGE BY DIFFERENTIAL THERMAL EXPANSION A.R. Drews

More information

Design of a Nuclear Level Switch using MCNP code, and comparison with experimental results

Design of a Nuclear Level Switch using MCNP code, and comparison with experimental results Design of a Nuclear Level Switch using MCNP code, and comparison with experimental results Mitra Ansari, Majid Shahriari Department of Radiation Application, Shahid Beheshti University Tehran, Iran, P.O.Box:1983963113

More information

PRACTICE EXAM IV P202 SPRING 2004

PRACTICE EXAM IV P202 SPRING 2004 PRACTICE EXAM IV P202 SPRING 2004 1. In two separate double slit experiments, an interference pattern is observed on a screen. In the first experiment, violet light (λ = 754 nm) is used and a second-order

More information

AP Physics 1 and 2 Lab Investigations

AP Physics 1 and 2 Lab Investigations AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks

More information

Analytical Method Validation for Assay, Related substances & Dissolution. N A Y A N K S H A T R I 1 8 / 0 1 /

Analytical Method Validation for Assay, Related substances & Dissolution. N A Y A N K S H A T R I 1 8 / 0 1 / Analytical Method Validation for Assay, Related substances & Dissolution. N A Y A N K S H A T R I 1 8 / 0 1 / 2 0 1 3 Definitions The objective of validation of an analytical procedure is to demonstrate

More information

The number of scatters per unit volume n is estimated as the molar mass times Avogadro s #. n = (! A )N A. ! di I

The number of scatters per unit volume n is estimated as the molar mass times Avogadro s #. n = (! A )N A. ! di I The Energy Loss of Particles in Matter I. Cross Section As a particle traverses a matter it has a probability to react by scattering, absorption, or to interaction in the material. The reaction probability

More information

Pt to N distance in Nδ binding site

Pt to N distance in Nδ binding site Supplementary Material Table S1 Unit cell parameter evaluation of the four 300 K data sets, with the standard deviations against the unit-cell parameters used given also. a (Standard deviation) Å c (Standard

More information

NEUTRON BASED TECHNIQUES FOR DETECTION OF ILLICIT MATERIALS AND EXPLOSIVES

NEUTRON BASED TECHNIQUES FOR DETECTION OF ILLICIT MATERIALS AND EXPLOSIVES Final Report on CRP Entitled, NEUTRON BASED TECHNIQUES FOR DETECTION OF ILLICIT MATERIALS AND EXPLOSIVES Research contract 13497 R. M. Megahid *, A. M. Osman and, W. A. Kansouhfg Laboratories for Detection

More information

ORTEC AN34 Experiment 12 X-Ray Fluorescence

ORTEC AN34 Experiment 12 X-Ray Fluorescence EQUIPMENT NEEDED FROM ORTEC 4001A/4002D NIM Bin and Power Supply 142 PC Preamplifier 480 Pulser 672 Spectroscopy Amplifier 556 High Voltage Power Supply 659 0 5kV Detector Bias Supply TRUMP-PCI-2K MCA

More information

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum

More information

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino

More information

AN INTRODUCTION TO THE RADOS XRF ORE SORTER. RS Fickling Metanza Mineral Processors (Pty) Ltd

AN INTRODUCTION TO THE RADOS XRF ORE SORTER. RS Fickling Metanza Mineral Processors (Pty) Ltd 6 th Southern African Base Metals Conference 2011 AN INTRODUCTION TO THE RADOS XRF ORE SORTER RS Fickling Metanza Mineral Processors (Pty) Ltd 1. ABSTRACT RADOS has developed ore sorters in Russia over

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

DRAFT Applicability of X-Ray Fluorescence (XRF) For Rattlesnake Creek Prepared for U.S. Army Corps of Engineers Buffalo District Buffalo, New York

DRAFT Applicability of X-Ray Fluorescence (XRF) For Rattlesnake Creek Prepared for U.S. Army Corps of Engineers Buffalo District Buffalo, New York DRAFT Applicability of X-Ray Fluorescence (XRF) For Rattlesnake Creek Prepared for U.S. Army Corps of Engineers Buffalo District Buffalo, New York by Environmental Assessment Division Argonne National

More information

Limitations of Handheld XRF Instruments as Quantitative Tools for Analyzing Heavy Metal Pesticides on Organic Art Objects

Limitations of Handheld XRF Instruments as Quantitative Tools for Analyzing Heavy Metal Pesticides on Organic Art Objects Limitations of Handheld XRF Instruments as Quantitative Tools for Analyzing Heavy Metal Pesticides on Organic Art Objects IARC Symposium Native Collections and Pesticides: Testing, Analysis and Mediation

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

Basic hands-on gamma calibration for low activity environmental levels

Basic hands-on gamma calibration for low activity environmental levels Basic hands-on gamma calibration for low activity environmental levels Iolanda Osvath Environment Laboratories Marine Environment Laboratories, Monaco Radiometrics Laboratory International Atomic Energy

More information

X-Ray Fluorescence (XRF): Understanding Characteristic X-Rays

X-Ray Fluorescence (XRF): Understanding Characteristic X-Rays X-Ray Fluorescence (XRF): Understanding Characteristic X-Rays What is XRF? X-Ray Fluorescence is defined as The emission of characteristic "secondary" (or fluorescent) X-rays from a material that has been

More information

Introduction to Nuclear Radiation 9/04. Purpose of the Experiment

Introduction to Nuclear Radiation 9/04. Purpose of the Experiment Modern Physics Lab Introduction to Nuclear Radiation 9/04 Purpose of the Experiment - become familiar with detectors for radioactive decay products - apply statistical analysis techniques to data - understand

More information

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE

Portable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications

More information

Study of α-particle Scattering By Means of The Rutherford Experiment

Study of α-particle Scattering By Means of The Rutherford Experiment Study of α-particle Scattering By Means of The Rutherford Experiment G.S. Sagoo With Laboratory Partner: Nilesh Agnihotri University College London 15 th December 009 Abstract: This experiment is a modernised

More information

Detailed simulation of mass spectra for quadrupole mass spectrometer systems

Detailed simulation of mass spectra for quadrupole mass spectrometer systems Detailed simulation of mass spectra for quadrupole mass spectrometer systems J. R. Gibson, a) S. Taylor, and J. H. Leck Department of Electrical Engineering and Electronics, The University of Liverpool,

More information

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün

The Limitations of Hand-held XRF Analyzers as a Quantitative Tool for Measuring Heavy Metal Pesticides on Art Objects. By Özge Gençay Üstün N.B. A shorter version of this article was published in the ICOM-CC Ethnographic Conservation Newsletter, Number 30, January 2009, pp. 5-8. The Limitations of Hand-held XRF Analyzers as a Quantitative

More information

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon

More information

Calibration of HPGe detector for in situ measurements of 137 Cs in soil by peak to valley method

Calibration of HPGe detector for in situ measurements of 137 Cs in soil by peak to valley method Calibration of HPGe detector for in situ measurements of 137 Cs in soil by peak to valley method M. Fülöp Institute of Preventive and Clinical Medicine, Limbova 14, 83101 Bratislava, Slovak Republic INTRODUCTION

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

BST THIN FILM EVALUATION USING X-RAY FLUORESCENCE AND REFLECTIVITY METHODS

BST THIN FILM EVALUATION USING X-RAY FLUORESCENCE AND REFLECTIVITY METHODS Copyright(C)JCPDS-nternational Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42 109 Copyright(C)JCPDS-nternational Centre for Diffraction Data 2000, Advances in X-ray Analysis, Vol.42

More information

T e c h n i c a l N o t e s

T e c h n i c a l N o t e s What are x-rays and how are they generated? What are x-rays? X-rays are part of the electromagnetic radiation spectrum. This spectrum includes radio waves, microwaves, infrared, the visible spectrum, ultra

More information

X-Ray Fluorescence - Data Collection and Analysis

X-Ray Fluorescence - Data Collection and Analysis XRF User Manual 1 X-Ray Fluorescence - Data Collection and Analysis Matthias Zeller, Jerry Hunter, Sept. 19 th 2008 A) Data Collection Logging In: The login and password to log into the program as a normal

More information

Energy Dispersive X-ray X Analysis in the TEM. Lecture 19

Energy Dispersive X-ray X Analysis in the TEM. Lecture 19 Energy Dispersive X-ray X Analysis in the TEM Lecture 19 Example x-ray x spectra (EDS) Several examples of EDS spectra Note: Relative # of counts Energy range of each spectrum Pure Ge Al film on Si Silica

More information

DEPARTMENT of INFRASTRUCTURE, ENERGY and RESOURCES, TASMANIA GENERAL SPECIFICATION. G4 - COMPACTION ASSESSMENT March 1995

DEPARTMENT of INFRASTRUCTURE, ENERGY and RESOURCES, TASMANIA GENERAL SPECIFICATION. G4 - COMPACTION ASSESSMENT March 1995 DEPARTMENT of INFRASTRUCTURE, ENERGY and RESOURCES, TASMANIA GENERAL SPECIFICATION G4 - COMPACTION ASSESSMENT March 1995 INDEX Page G4.1 SCOPE 2 G4.2 DEFINITIONS 2 G4.3 TEST METHODS 3 G4.4 FIELD TESTING

More information

Balance Accuracy. FlexiWeigh

Balance Accuracy. FlexiWeigh FlexiWeigh The balance is an integral part of the FlexiWeigh workstation. The role and limitations of the balance are important to understand in this environment. This technical note describes the different

More information

Lectures about XRF (X-Ray Fluorescence)

Lectures about XRF (X-Ray Fluorescence) 1 / 38 Lectures about XRF (X-Ray Fluorescence) Advanced Physics Laboratory Laurea Magistrale in Fisica year 2013 - Camerino 2 / 38 X-ray Fluorescence XRF is an acronym for X-Ray Fluorescence. The XRF technique

More information

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based

More information

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy

ORTEC AN34 Experiment 7 High-Resolution Gamma-Ray Spectroscopy Equipment Needed from ORTEC GEM10P4/CFG-PV4/DWR-30 Coaxial Detector System (Includes detector, cryostat, dewar, preamplifier, and 12-ft. cable pack); typical specifications: 10% relative efficiency, 1.75

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

2008 USPC Official 12/1/07-4/30/08 General Chapters: <1225> VALIDATION OF C...Page 1 of 10

2008 USPC Official 12/1/07-4/30/08 General Chapters: <1225> VALIDATION OF C...Page 1 of 10 2008 USPC Official 12/1/07-4/30/08 General Chapters: VALIDATION OF C...Page 1 of 10 1225 VALIDATION OF COMPENDIAL PROCEDURES Test procedures for assessment of the quality levels of pharmaceutical

More information

Photon Counting and Spectroscopy in the Gamma Ray Energy Range

Photon Counting and Spectroscopy in the Gamma Ray Energy Range Chapter 7 Photon Counting and Spectroscopy in the Gamma Ray Energy Range 7.1 Objective Determine whether a radioactive source or set of sources obey Poisson statistics, and whether the radiation is isotropic.

More information

MASS DEFECT AND BINDING ENERGY

MASS DEFECT AND BINDING ENERGY MASS DEFECT AND BINDING ENERGY The separate laws of Conservation of Mass and Conservation of Energy are not applied strictly on the nuclear level. It is possible to convert between mass and energy. Instead

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Application Note: Absorbance

Application Note: Absorbance Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light

More information

NIU Radiation Safety Research X-ray Safety Fundamentals

NIU Radiation Safety Research X-ray Safety Fundamentals NIU Radiation Safety Research X-ray Safety Fundamentals Goals of this training presentation: 1. Explain what are X-rays. 2. Explain the hazards of X-ray devices used in NIU research. 3. Explain NIU requirements

More information

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND:

Gamma Rays OBJECT: READINGS: APPARATUS: BACKGROUND: Gamma Rays OBJECT: To understand the various interactions of gamma rays with matter. To calibrate a gamma ray scintillation spectrometer, using gamma rays of known energy, and use it to measure the energy

More information

Mini-trampoline vibration exciter- Force measurements

Mini-trampoline vibration exciter- Force measurements Mini-trampoline vibration exciter- Force measurements Leonard L. KOSS 1 ; Vincent ROUILLARD 2 1 Monash University, Australia 2 Victoria University, Australia ABSTRACT A mini-trampoline has been previously

More information

BORON IN GLASS DETERMINATION USING WDXRF

BORON IN GLASS DETERMINATION USING WDXRF 269 ABSTRACT BORON IN GLASS DETERMINATION USING WDXRF Alexander Seyfarth Bruker AXS Inc., 5465 East Cheryl Parkway, Madison, Wisconsin 53711-5373 The application of wavelength-dispersive XRF to the analysis

More information

TWO-DIMENSIONAL X-RAY DIFFRACTION

TWO-DIMENSIONAL X-RAY DIFFRACTION TWO-DIMENSIONAL X-RAY DIFFRACTION BOB B. HE GQ WILEY ,. "'! :~! CONTENTS~\.-.~..). Preface 1. Introduction 1.1 X-Ray Technology and Its Brief History, 1.2 Geometry of Crystals, 2 1.2.1 Crystal Lattice

More information

Alkali Content of Concrete Mix Water and Aggregates

Alkali Content of Concrete Mix Water and Aggregates Alkali Content of Concrete Mix Water and Aggregates Introduction Alkali silica reaction (ASR) is a chemical reaction between some types of aggregate and the highly alkaline pore water present in all Portland

More information

Figure 1. A typical Laboratory Thermometer graduated in C.

Figure 1. A typical Laboratory Thermometer graduated in C. SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES

More information

LYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD

LYNXEYE XE-T. < 380 ev. Innovation with Integrity. Energy. Resolution. High-Resolution Position Sensitive Detector with Superb Energy Resolution XRD Energy < 380 ev Resolution LYNXEYE XE-T High-Resolution Position Sensitive Detector with Superb Energy Resolution The LYNXEYE XE-T is the next generation "Compound Silicon Strip" detector with superb energy

More information