CORRECTING FOR HUMIDITY EFFECTS ON TSI GENERAL PURPOSE FLOWMETERS

Similar documents
THE HUMIDITY/MOISTURE HANDBOOK

How do I measure the amount of water vapor in the air?

General Purpose Thermal Mass Flowmeter Model 4140/4143

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

Chemistry 212 VAPOR PRESSURE OF WATER LEARNING OBJECTIVES

PREMIUM CLEAN ROOM MONITOR MODELS 8630-CRM-S 8630-CRM-P

GENERAL PURPOSE THERMAL MASS FLOWMETER MODEL 4140/4143

HEATED INLET SAMPLE CONDITIONING EFFECTS ON REMOTE DUST MONITORING

QUESTIONS THERMODYNAMICS PRACTICE PROBLEMS FOR NON-TECHNICAL MAJORS. Thermodynamic Properties

Total Heat Versus Sensible Heat Evaporator Selection Methods & Application

ACFM vs. SCFM vs. ICFM Series of Technical White Papers from Ohio Medical Corporation

THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS

PERCENT OUTDOOR AIR (%OA) CALCULATION AND ITS USE

FITPRO+ SOFTWARE V3 IMPORTING DATA FROM FITPRO V2 & FITPLUS3

Rusty Walker, Corporate Trainer Hill PHOENIX

HAY MOISTURE & WEATHER:

Flow Measurement. Reference - Flow Measurement FUNDAMENTALS. ρ = fluid density (lb/ft3 or kg/m3)

CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS

EXPERIMENT 15: Ideal Gas Law: Molecular Weight of a Vapor

Open Cycle Refrigeration System

Soil Suction. Total Suction

Akton Psychrometric Chart Tutorial and Examples

Moisture Control. It s The Dew Point. Stupid! Its not the humidity.

Energy Recovery Ventilation Equipment

UNDERSTANDING REFRIGERANT TABLES

T0118 T2118 T3118. Instruction Manual

Comparing Air Cooler Ratings Part 1: Not All Rating Methods are Created Equal

EFFECT OF SMALL VARIATIONS IN TEST CONDITIONS ON THE CO2 SCRUBBER ENDURANCE TO EN14143:2003 AND NORSOK U-101 DOCUMENT NUMBER: [Filename]

Temperature and Humidity

CHAPTER 7 THE SECOND LAW OF THERMODYNAMICS. Blank

NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS

RPS High Accuracy Resonant Pressure Sensor. GE Measurement & Control Solutions. Features:

MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question.

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

K Standard Grade Ni-Cr vs Ni-Al (Chromel vs Alumel)

Turbine Flowmeters for Gas Applications

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article

Sheet 5:Chapter 5 5 1C Name four physical quantities that are conserved and two quantities that are not conserved during a process.

Online Infrared Flue Gas Analyzer Gasboard 3000

APPLICATION PROGRAM INTERFACE (API) FOR AIRASSURE WEB PM2.5 SOFTWARE

UNIT 6a TEST REVIEW. 1. A weather instrument is shown below.

Preferred SI (Metric) Units

Legacy Data Centres Upgrading the cooling capabilities What are the options?

TEMPERATURE MEASURING GALVANIC SEPARATORS CONVERTORS, PRESSURE MEASURING

ANNUBAR FLOW CALCULATION PROGRAM (For the FloBoss 103) User Manual (QER 04Q019)

THE PSYCHROMETRIC CHART: Theory and Application. Perry Peralta NC State University

User s Manual Laminar Flow Elements

PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques

EXPERIMENT 13: THE IDEAL GAS LAW AND THE MOLECULAR WEIGHT OF GASES

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection.

COMBIMASS. Technical Data COMBIMASS eco-bio +

Initial Experiments of a Novel Liquid Desiccant Dehumidifier for Industrial and Comfort Air Conditioning Systems

THE PSYCHROMETRIC CHART AND ITS USE

Humidity, Condensation, Clouds, and Fog. Water in the Atmosphere

High Altitude HVAC. Silvertip Integrated Engineering Consultants

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

ES-7A Thermodynamics HW 1: 2-30, 32, 52, 75, 121, 125; 3-18, 24, 29, 88 Spring 2003 Page 1 of 6

Energy Efficient Building Design College of Architecture Illinois Institute of Technology, Chicago. Relative Humidities RH%

The Ideal Gas Law. Gas Constant. Applications of the Gas law. P = ρ R T. Lecture 2: Atmospheric Thermodynamics

RPS/DPS High Accuracy Resonant Pressure Sensor. GE Measurement & Control. Features:

Custody Transfer Measurement. with the V-Cone Flowmeter

Vapor Pressure Curves

Physical Properties of a Pure Substance, Water

Environment-Laboratory Ambient Conditions

A. Kinetic Molecular Theory (KMT) = the idea that particles of matter are always in motion and that this motion has consequences.

High-ohmic/high-voltage resistors

Thermal environments. Tel

FUNDAMENTALS OF ENGINEERING THERMODYNAMICS

ME 538: Experiment # 1 Flow measurement and particle generation instruments

CSUS Department of Chemistry Experiment 8 Chem.1A

Kinetic Theory of Gases

Absolute and Relative Gas Concentration: Understanding Oxygen in Air

a cannonball = (P cannon P atmosphere )A cannon m cannonball a cannonball = (P cannon P atmosphere ) πd 2 a cannonball = 5.00 kg

AP Physics Course 1 Summer Assignment. Teachers: Mr. Finn, Mrs. Kelly, Mr. Simowitz, Mr. Slesinski

DETERMINING THE MOLAR MASS OF CARBON DIOXIDE

Gas Laws. The kinetic theory of matter states that particles which make up all types of matter are in constant motion.

Characteristics of Evaporators

Lecture 9 Solving Material Balances Problems Involving Non-Reactive Processes

ADU 5. Automatic Distillation Unit

Test 5 Review questions. 1. As ice cools from 273 K to 263 K, the average kinetic energy of its molecules will

Vogt Valves The Connection Bulletin for Forged Steel Flow Control Valves CB 15

Humid Air. Water vapor in air. Trace Glasses 1% Argon (A) Water vapor (H 2

Analyzing Weather Data

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Specific Volume of Liquid (Column 7). The volume per unit of mass in cubic feet per pound.

MECHANICAL ENGINEERING EXPERIMENTATION AND LABORATORY II EXPERIMENT ENGINE PERFORMANCE TEST

EVALUATION OF PERSONAL COOLING SYSTEMS FOR SOLDIERS Elizabeth A. McCullough and Steve Eckels

How to Save Over 60% on Cooling Energy Needed by a Data Center? The Roadmap from Traditional to Optimized Cooling

Equivalents & Conversion Factors 406 Capacity Formulas for Steam Loads 407 Formulas for Control Valve Sizing

CHAPTER 14 THE CLAUSIUS-CLAPEYRON EQUATION

Advances in Thermal Dispersion Mass Flow Meter Accuracy

How is Dew Point Measured Reliably?

Dew Point Mirror 373 Operation and Maintenance Manual

Power chip resistor size 2512 PRC221 5%; 2% FEATURES Reduced size of final equipment Low assembly costs Higher component and equipment reliability.

a) Use the following equation from the lecture notes: = ( J K 1 mol 1) ( ) 10 L

DATE PERFORMED: DATE DUE:

Certified according to DIN EN ISO 9001 Technical Datasheet KVO Series Vortex Flow Meter

Chemistry 13: States of Matter

OUTCOME 2 INTERNAL COMBUSTION ENGINE PERFORMANCE. TUTORIAL No. 5 PERFORMANCE CHARACTERISTICS

Name: OBJECTIVES Correctly define: WEATHER BASICS: STATION MODELS: MOISTURE: PRESSURE AND WIND: Weather

Transcription:

CORRECTING FOR HUMIDITY EFFECTS ON TSI GENERAL PURPOSE FLOWMETERS APPLICATION NOTE FLOW-001 This application note applies to all TSI flowmeters equipped with a mini-din connector Like all thermal mass flow meters, humidity affects the output of the TSI general-purpose flowmeters. The mass flow rate, thermal conductivity and viscosity all affect the amount of heat transferred from the sensor element. If water vapor is added to the gas, the total mass is increased and both the overall thermal conductivity and overall viscosity change. Dry Air Water Vapor Humidified Air On the following pages are a series of graphs that can be used to correct for the effects of water vapor. To correct the readings you need to know what percentage of the overall gas is water vapor. The percent water vapor can be determined from the relative humidity or dew point, barometric pressure, and temperature of the gas.

Below are the equations for converting from relative humidity to percent water vapor by volume. If the dew point is known, then the following equations can be used by substituting the dew point temperature for the gas temperature and using 100% for the relative humidity. Percent water vapor by volume is given by % Water vapor by volume = RH/100 * p ws /p where RH = percent relative humidity p ws = saturated vapor pressure, which is a function of temperature p = absolute pressure The saturation vapor pressure over liquid water for the temperature range of 32 F to 392 F is given by 1 : ln(pws) = C 8 / T + C 9 + C 10 T + C 11 T 2 + C 12 T 3 + C 13 ln(t) where C 8 = -1.0440397E +04 C 9 = -1.1294650E +01 C 10 = -2.7022355E -02 C 11 = 1.2890360E -05 C 12 = -2.4780681E -09 C 13 = 6.5459673 T = absolute temperature, R = F + 459.67, R = K*9/5, R = ( C+273.15)*9/5 pws = saturated vapor pressure (psia) 1 psia = mmhg/51.715 = Kpa/6.8947 = mbar/68.947 = cm H2O/70.308 = inches H2O/27.680 1. For the purposes of effect on flow readings the above equation can be used below 32 F within the pressure range of these meters. Reference: 1997 ASHRAE Handbook, Fundamentals, American Society of Heating, Refrigeration and Air Conditioning Engineers. -2-

List of Graphs Example 1: 414x Flowmeter- Correction to air flow rate 414x Flowmeter- Correction to total mass flow rate 404x Flowmeter- Correction to air flow rate 404x Flowmeter- Correction to total mass flow rate mass flow of just the air, not including the water vapor. Most of the correction is due to the added mass of the water vapor. combined mass flow of the air and water vapor. mass flow of just the air, not including the water vapor. Most of the correction is due to the added mass of the water vapor. combined mass flow of the air and water vapor. page 5 page 6 page 7 page 8 Example 2: A 4040 reads 50.0 Std L/min of air with 20% R.H. at one atmosphere and 21 C. If you want the mass flow rate of air without the water vapor, then use the second graph from page eight. The correction from the graph is. The mass flow rate of just the air then is 50 std L/min *(1-0.010) = 49.5 std L/min of air. A 4140 reads 5.00 std L/min of air with 30% R.H. at 0.9 atmospheres and 26 C. If you want the mass flow rate including the water vapor, then you use the equations on page two and the first graph on page five. The percent water vapor = Relative Humidity * saturated vapor pressure/absolute pressure =RH * exp(c 8/ T + C 9 + C 10 T + C 11 T 2 + C 12 T 3 + C 13 ln(t)) psia/p abs(psia) where RH = 30% and T = (26 +273.15)*9/5 =538.47 percent water vapor = 30% * exp(c 8/ T + C 9 + C 10 T + C 11 T 2 + C 12 T 3 + C 13 ln(t))psia/(0.9atm*14.7 psia/atm) = 30% * 0.4878 psia/ 13.23 psia = 1.1% water vapor Using 1.1% water vapor, 5 L/min, and the first graph on page six the correction is approximately - 1.2%. Therefore the total mass flow rate is 5.00 std L/min * (1-.012) = 4.94 std L/min -3-

Model 414x Flowmeters Humidity Correction to get Air Flow Rate Use the following graphs if you are interested in measuring the mass flow of just the air, not including the water vapor. Most of the correction is due to the added mass of the water vapor. -4.5% 0.5% 1.0% 1.5% 2.0% 0. -4.5% 0. -4-

Model 414x Flowmeters Humidity Correction to Total Mass Flow Rate Use the following graph if you are interested in measuring the total mass flow including the water vapor mass. Correction to Get Total Mass Flowrate 0.5% 1.0% 1.5% 2.0% 0. Correction to Get Total Mass Flowrate 0. -5-

Model 404x Flowmeters Humidity Correction to get Air Flow Rate Use the following graphs if you are interested in measuring the mass flow of just the air, not including the water vapor. Most of the correction is due to the added mass of the water vapor. Humidity Effects on 404X Series -4.5% 0.5% 1.0% 1.5% 2.0% Humidity Effects on 404X Series -6-

Model 404x Flowmeters Humidity Correction to Total Mass Flow Rate Use the following graph if you are interested in measuring the total mass flow including the water vapor mass. Humidity Effects on 404X Series Correction to Get True Total Mass Flowrate 0.5% 1.0% 1.5% 2.0% Humidity Effects on 404x Series Correction to Get True Total Mass Flowrate TSI Incorporated Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030 India Tel: +91 80 67877200 China Tel: +86 10 8251 6588 Singapore Tel: +65 6595 6388 FLOW-001 Rev. D 2013 TSI Incorporated Printed in U.S.A.