HEATED INLET SAMPLE CONDITIONING EFFECTS ON REMOTE DUST MONITORING
|
|
|
- Arabella York
- 10 years ago
- Views:
Transcription
1 HEATED INLET SAMPLE CONDITIONING EFFECTS ON REMOTE DUST MONITORING APPLICATION NOTE EXPMN-8 (A4) Introduction Under high ambient humidity conditions, photometric mass measurements have been reported to over-estimate mass concentration readings when compared to a reference method. This is largely attributed to water-uptake, resulting in hygroscopic growth of the aerosol particles. When the aerosol particles increase in size they scatter more light, resulting in higher mass concentration readings by the photometric instrument (McMurry and Stolzenburg, 1989; Thomas and Gebhart, 1994; Brauer, 1995; Day et al., 2). Certain types of inlet conditioners may substantially improve the correlation between particulate mass concentration derived from real-time particle sizers and filter-based samplers in humid conditions (Peters, TM, et al, 28). TSI Engineers have developed a Heated Inlet Sample Conditioner for use with the TSI DustTrak II and DustTrak DRX Aerosol Monitors to reduce the effect of humidity on photometric mass measurements. This application note summarizes findings from initial field tests demonstrating the effect on DustTrak Aerosol Monitor readings with heated inlet sample conditioning in high-humidity outdoor environments. Outdoor Urban Pollution Test Outdoor testing of urban pollution in a warm, high humidity environment was conducted in Singapore using two identical DustTrak DRX Aerosol Monitors (Model 8533) with Autozero Modules inside TSI Environmental Enclosures (Model 8535). See Figure 1 for example of sampling location. One DustTrak Monitor was configured with a Heated Inlet Sample Conditioner set to condition incoming sample air to 3% Rh. The other DustTrak Monitor had no sample conditioning. Both instruments were programmed to run side-by-side for identical sample periods. Temperature during the test averaged 28.7 o C (range 27. to 3.8 o C). Relative humidity averaged 79.7%Rh (range 72.9 to 81.4% Rh). The results demonstrate heated inlet sample conditioned photometric readings are lower than non-conditioned photometric readings (see figure 2). The average change between the heated inlet conditioned measurements and the non-conditioned sample measurements was 29.9%. TSI and TSI logo are registered trademarks of TSI Incorporated. DustTrak is a trademark of TSI Incorporated.
2 Figure 1. Co-located DustTrak Monitors inside Environmental Enclosures (Model 8535) in Singapore Singapore test data - outdoor ambient air PM Data Points (time) no conditioning HI 3%Rh Figure hr test results from PM2.5 sample in outdoor humid environment. PM2.5 no sample conditioning: Min.19mg/m 3, Max.119 mg/m 3, Avg.52 mg/m 3 PM2.5 with 3%Rh sample conditioning: Min.13mg/m 3, Max.75 mg/m 3, Avg.36 mg/m 3 Average % change between conditioned and non-conditioned: 29.99% Additional tests were conducted to observe changes in DustTrak Monitor readings with the Heated Inlet Sample Conditioner active and inactive. The intent was to see the difference in measurements between two instruments when one instrument had the Heated Inlet in place and not active, versus in place and active. These tests were run with side-by-side DustTrak DRX Aerosol Monitors with Autozero Modules inside Environmental Enclosures (Model 8535) in high humidity outdoor environments in Singapore. The co-located instruments started logging data with no Heated Inlets in place. After a period of time, the Heated Inlet was added to one of the DustTrak Monitors, but not turned on. Again, after a period of time, the Heated Inlet was turned on. The test continued for equal periods of time with the Heated Inlet turned off then another period of time with the Heated Inlet removed from the flow path. Figure 3 contains data logged during a 1-hr test with the Heated Inlet adjusted every two hours as follows: Data points 1-24: Heated Inlet not attached to the instrument. Data points 25-48: Heated Inlet in place, but not turned on. Data points 49-72: Heated Inlet turned on and active. Data points 73-96: Heated Inlet turned off, but remained in place atop instrument. Data points 97-12: Heated Inlet removed from the sample flow path. -2-
3 The spike at data point 37 occurred when the Heated Inlet was turned on by mistake then turned off after a few minutes. Figure 4 contains data logged during a 5-hr test following the same methodology Singapore Test Data - Outdoor ambient PM2.5 Inlet conditioning active vs inactive Data points (time) No conditioning HI 3%Rh Figure 3. Results from 1-hour test with Heated Inlet in place, removed, active, and inactive PM2.5 no sample conditioning: Min.73mg/m 3, Max.116 mg/m 3, Avg.91mg/m 3. PM2.5 with sample conditioning: Min.52mg/m 3, Max.13mg/m 3, Avg.75mg/m 3 Average % change between conditioned and non-conditioned: 3.% when Heated Inlet is active Data points 1-24 Heated Inlet not installed average % change 12.31% Data points Heated Inlet installed, but not active average % change 16.51% Data points Heated Inlet active average % change 3.% Data points Heated Inlet turned off average % change 19.11% Data points Heated Inlet removed average % change 12.52% Singapore Test Data - Outdoor ambient PM2.5 Heated Inlet conditioning active vs inactive Data points (time) no conditioning 3%Rh conditioning % change Figure 4. Result from 5-hour test with Heated Inlet in place, removed, active, and inactive. PM2.5 no sample conditioning: Min.31mg/m 3, Max.57mg/m 3, Avg.42mg/m 3 PM2.5 with sample conditioning: Min.24mg/m 3, Max.5mg/m 3, Avg.34mg/m. Average % change between conditioned and non-conditioned: 38.5% when Heated Inlet is active Data points 1-12 Heated Inlet not installed average % change 7.88% Data points Heated Inlet installed but not active average % change 14.22% Data points Heated Inlet active average % change 38.5% Data points Heated Inlet turned off average % change 26.75% Data points 49-6 Heated Inlet removed average % change 11.8% -3-
4 A TSI Channel Partner conducted a separate test of outdoor ambient pollution in the warm, humid environment along the coast of Australia. Two identical TSI DustTrak II (Model 83) Aerosol Monitors were co-located inside Environmental Enclosures (Model 8535). One DustTrak Monitor had the Heated Inlet Sample Conditioner installed and set to condition the sample to 3% Rh. Figure 5 shows the results of the test conducted in Australia Australia Field Test Data - outdoor ambient Total Dust Time No conditioning Heated inlet sample conditioning % change Figure 5. Results from Australia outdoor ambient air testing in marine environment Average % change between conditioned and non-conditioned: 35%. Results These test results indicate the Heated Inlet Sample Conditioner reduces photometric mass measurement compared to an identical photometric instrument without Heated Inlet during side-by-side sampling in outdoor humid environments. A difference of 3% to 35% was observed between data collected with the Heated Inlet active versus inactive with side-by-side identical instruments. These results have not been validated against federal reference methods and are not expected to be indicative of all aerosols. Different aerosols will be affected by humidity in different ways based on the hygroscopic properties of the aerosol. Users of the Heated Inlet Sample Conditioner should perform similar side-by-side testing to evaluate the effect on specific aerosols in their sampling locations. -4-
5 References Brauer M. Assessment of indoor aerosols with an integrating nephelometer. J Exposure Analysis Environmental Epidemiol 1995: 5(1) Day D.E., Malm W.C., and Kreidenweis S.M. Aerosol light scattering measurements as a function of relative humidity. J Air Waste Manage Association 2: 5(5); McMurry P.H., and Stolzenburg M.R. On the sensitivity of particle size to relative humidity for Los Angeles aerosols. Atmos Environ 1989: 23: Thomas A., and Gebhart J. Correlations between gravimetry and light-scattering photometry for atmospheric aerosols. Atmos Environ 1994; 28(5): Peters TM, Riss AL, Holm RL, Singh M, Vanderpool RW, Design and evaluation of an inlet conditioner to dry particles for real-time particle sizers, J Environmental Monitoring. 28 Apr;1(4): TSI Incorporated Visit our website for more information. USA Tel: UK Tel: France Tel: Germany Tel: India Tel: China Tel: Singapore Tel: EXPMN-8 A4 (6/19/214) 214 TSI Incorporated Printed in U.S.A.
REAL-TIME DUST AND AEROSOL MONITORING
REAL-TIME DUST AND AEROSOL MONITORING THE DUSTTRAK II AND DRX AEROSOL MONITORS UNDERSTANDING, ACCELERATED REAL-TIME DUST MONITORING. ANY ENVIRONMENT. ANY APPLICATION. The DustTrak II and DRX Aerosol Monitors
OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD
OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method
THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS
THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS INDOOR AIR QUALITY INSTRUMENTS UNDERSTANDING, ACCELERATED BREATHE A LITTLE EASIER WITH TSI Indoor air quality is a growing concern. With the increasing
Atmospheric Monitoring of Ultrafine Particles. Philip M. Fine, Ph.D. South Coast Air Quality Management District
Atmospheric Monitoring of Ultrafine Particles Philip M. Fine, Ph.D. South Coast Air Quality Management District Ultrafine Particles Conference May 1, 2006 Outline What do we monitor? Mass, upper size cut,
FITPRO+ SOFTWARE V3 IMPORTING DATA FROM FITPRO V2 & FITPLUS3
FITPRO+ SOFTWARE V3 IMPORTING DATA FROM FITPRO V2 & FITPLUS3 APPLICATION NOTE RFT-020 (US) T his quick reference guide provides instructions for importing data into a FitPro+ v3.0, v3.1, or v3.2 software
TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS
TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS APPLICATION NOTE CC-105 (US) Introduction This application note will demonstrate how real-time viable particle count data will
CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS
CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS UNDERSTANDING, ACCELERATED TSI SETS THE STANDARD FOR PARTICLE COUNTING Performance Semiconductor, hard disk drive, and flat panel manufacturers
Innovative solutions for better. health protection. Thermo Scientific Aerosol and Dust Monitoring Solutions
Innovative solutions for better health protection Thermo Scientific Aerosol and Dust Monitoring Solutions Identify. Define. Inform. The knowledge to do more than inform can make all the difference when
Coarse PM Methods Evaluation Study
Coarse PM Methods Evaluation Study Study Design and Preliminary Results Thomas Ellestad, Robert Vanderpool, Paul Solomon, and Mary Harmon USEPA/ORD/NERL Sanjay Natarajan, Chris Noble, and Bob Murdoch Research
APPLICATION PROGRAM INTERFACE (API) FOR AIRASSURE WEB PM2.5 SOFTWARE
APPLICATION PROGRAM INTERFACE (API) FOR AIRASSURE WEB PM2.5 SOFTWARE APPLICATION NOTE PM2.5-003 (US) Introduction TSI has released an API for the AirAssure Web Software that will enable customers to extract
PERCENT OUTDOOR AIR (%OA) CALCULATION AND ITS USE
PERCENT OUTDOOR AIR (%OA) CALCULATION AND ITS USE APPLICATION NOTE TI-138 TSI s IAQ-CALC TM Indoor Air Quality Meter and TH-CALC TM Thermohygrometer calculate Percent Outdoor Air (%OA). In order to take
www.ecomonitoring.com
Ambient Air Monitoring Systems ECM ambient air monitoring systems are integrated in air conditioned thermal isolated light weight shelters allowing stationary and mobile installations. The usually monitored
USB-500 Series Stand-Alone, Low-Cost Data Loggers
Stand-Alone, Low-Cost Data Loggers Features Stand-alone, remote data loggers Measure temperature, humidity, voltage, or current 1 or 2 channels Low cost, small size Battery powered Software Included, easy-to-use
Ambient Air Monitoring at Deployment Locations in the Middle East
Ambient Air Monitoring at Deployment Locations in the Middle East Presented by: John E. Kolivosky, P.E. 01 May 2015 DISCLAIMERS The views expressed in this presentation are those of the author and do not
NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS
NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS APPLICATION NOTE ITI-089 I. Introduction The technology used in personal sampling pumps has not changed much over the past 20 years. II. There
INDOOR AIR QUALITY REASSESSMENT. Sgt. William H. Carney Memorial Academy 247 Elm Street New Bedford, Massachusetts
INDOOR AIR QUALITY REASSESSMENT Sgt. William H. Carney Memorial Academy 247 Elm Street New Bedford, Massachusetts Prepared by: Massachusetts Department of Public Health Bureau of Environmental Health Indoor
Federation of European Heating, Ventilation and Air-conditioning Associations
Federation of European Heating, Ventilation and Air-conditioning Associations Address: Rue Washington 40 1050 Brussels Belgium www.rehva.eu [email protected] Tel: +32 2 514 11 71 Fax: +32 2 512 90 62 Use of
Annual Report on the Air Resources Board s Fine Particulate Matter Monitoring Program
Annual Report on the Air Resources Board s Fine Particulate Matter Monitoring Program January 2014 California Environmental Protection Agency Air Resources Board State of California California Environmental
Scanning Mobility Particle Sizer
PARTICLE INSTRUMENTS Highly accurate, real-time nanoparticle sizing systems you can rely on for years Features and Benefits Model 3936N88 High resolution data: up to 167 channels Broad size range: from
National Environment Protection (Ambient Air Quality) Measure. Appendix 6
SOCO National Environment Protection (Ambient Air Quality) Measure Report of the Risk Assessment Taskforce 2 Appendix 6 Pb NO 2 Possible use of Health Risk Assessment in the Review of NEPM Pollutants Specified
Cree XLamp XR-E LED. Table of Contents. CLD-DS05 Rev 16B. Product family data sheet
Cree XLamp XR-E LED Product family data sheet CLD-DS05 Rev 16B Product Description The XLamp XR-E LED is leading the LED lighting revolution with its unprecedented lighting-class brightness, efficacy,
Fog and Cloud Development. Bows and Flows of Angel Hair
Fog and Cloud Development Bows and Flows of Angel Hair 1 Ch. 5: Condensation Achieving Saturation Evaporation Cooling of Air Adiabatic and Diabatic Processes Lapse Rates Condensation Condensation Nuclei
Electric Vehicle Charging Station Products EVoReel Charge Cable Management Solutions
Self-Retracts 30 ft. Charge Cable Length Electric Vehicle Charging Station Products EVoReel Charge Cable Management Solutions Safe, Convenient & Clean EV Charging The EVoReel cable management solutions
Certificate: 0000040216 / 29 April 2014
Test report: 936/21209885/G of 20 September 2013 Initial certification: 01 April 2014 Date of expiry: 31 March 2019 Publication: BAnz AT 01 April 2014 B12, chapter IV, No. 7.3 Approved application The
IntesisBox. KNX SAMSUNG Air Conditioning. Gateway for integration of Samsung air conditioners into KNX control systems. RS485.
IntesisBox KNX SAMSUNG Air Conditioning Gateway for integration of Samsung air conditioners into KNX control systems. Integrate Samsung Air Conditioners into KNX. KNX EIB Bus IntesisBox RS485 MIM-B13x/B04A
DUST EMISSION MONITORING SYSTEM
Applications Measurement of dust concentration in dry or moist, steamsaturated and corrosive stack gases Monitoring of municipal and hazardous waste and sewage sludge incinerators Monitoring of power plants
Ambient Air Quality Division Air Quality and Noise Management Bureau Pollution Control Department
The Ambient Air Quality Monitoring Network in Thailand by Mr. Vitsanu Wingpud Ambient Air Quality Division Air Quality and Noise Management Bureau Pollution Control Department Outline Introduction P C
Metrological features of a beta absorption particulate air monitor operating with wireless communication system
NUKLEONIKA 2008;53(Supplement 2):S37 S42 ORIGINAL PAPER Metrological features of a beta absorption particulate air monitor operating with wireless communication system Adrian Jakowiuk, Piotr Urbański,
EL-USB-2-LCD FEATURES WINDOWS CONTROL SOFTWARE DISPLAY FUNCTIONS. www.lascarelectronics.com Page 1 of 5
This standalone data logger measures and stores up to 16,379 relative humidity and 16,379 temperature readings over 0 to 100%RH and -35 to +80 C (-31 to +176 F) measurement ranges. The user can easily
Free Cooling for Data Centers
Free Cooling for Data Centers Release 1, September 2015 Contents 1. Abstract... 3 2. Introduction... 3 3. Data Center Temperatures and Free Cooling... 4 4. Direct Free Cooling... 4 5. Indirect Free Cooling...
PREMIUM CLEAN ROOM MONITOR MODELS 8630-CRM-S 8630-CRM-P
PREMIUM CLEAN ROOM MONITOR MODELS 8630-CRM-S 8630-CRM-P MANUAL SUPPLEMENT Contents of this manual supplement include: Sequence of Operation Menu Structure Description of New Software Items Deleted Software
Modular Data Center Cooling. Ultra High Efficiency. Oasis Optimum DCIE
Modular Data Center Cooling Ultra High Efficiency Oasis Optimum DCIE Oasis Optimum Modular Cooling Compact, Economic, Greener, Ultra High Efficiency Munters Oasis Optimum DCIE data center cooling solution
Climate Models: Uncertainties due to Clouds. Joel Norris Assistant Professor of Climate and Atmospheric Sciences Scripps Institution of Oceanography
Climate Models: Uncertainties due to Clouds Joel Norris Assistant Professor of Climate and Atmospheric Sciences Scripps Institution of Oceanography Global mean radiative forcing of the climate system for
How is Dew Point Measured Reliably?
How is Dew Point Measured Reliably? Air compression increases vapor pressure and thus dew point. = water vapor Vapor pressure Frost point Saturation - 0 C + Dew point Monitoring with alarming and real
White Paper: Free Cooling and the Efficiency of your Data Centre
AIT Partnership Group Ltd White Paper: Free Cooling and the Efficiency of your Data Centre Using free cooling methods to increase environmental sustainability and reduce operational cost. Contents 1. Summary...
Indoor-outdoor environmental coupling and exposure risk to extreme heat and poor air quality during heat waves
Indoor-outdoor environmental coupling and exposure risk to extreme heat and poor air quality during heat waves David J. Sailor 1, Huafen Hu 1, Olga Wilhelmi 2, and Deborah Banerjee 3 1. Portland State
Energy from Waste Combined Heat and Power Facility, North Yard, Devonport. Community Ambient Air Quality Monitoring Programme Report Quarter 4, 2015
Energy from Waste Combined Heat and Power Facility, North Yard, Devonport Community Ambient Air Quality Monitoring Programme Report Quarter 4, 2015 Overview of Monitoring Programme MVV started ambient
Performance Assessment of EAGLE 5000 in Tobacco Smoke Environment. REPORT
Performance Assessment of EAGLE 5000 in Tobacco Smoke Environment. REPORT Reference 943/78357 Carried out by: Ainhoa Mendivil Martínez IAQ CONSULTANT October 2006 Borrador 02 ÍNDEX 1. INTRODUCTION... 3
ThermoSorb Desiccant Dryers
ThermoSorb Desiccant Dryers Sierra Why dry compressed air? Contamination reduces efficiency The air we breathe contains contamination in the form of water vapour and airborne particles. During the compression
Gree Solar Hybrid Hi Wall Inverter Air Conditioner
Gree Solar Hybrid Hi Wall Inverter Air Conditioner Utilize the natural power of the sun to heat or cool your home. Find out more at www.greeonline.com Why Choose a Solar Hybrid Inverter? The new Gree Solar
TÜV RHEINLAND ENERGIE UND UMWELT GMBH
TÜV RHEINLAND ENERGIE UND UMWELT GMBH Report on performance testing of the Air Pollution Monitor 2 (APM-2) measuring system manufactured by Comde-Derenda GmbH for the components suspended particulate matter
Air Pollution in Ulaanbaatar: Public Health Impacts and Research Opportunities
Air Pollution in Ulaanbaatar: Public Health Impacts and Research Opportunities Ryan W. Allen, PhD Associate Professor Faculty of Health Sciences Simon Fraser University Burnaby, BC, Canada [email protected]
Particulate matter sources in Helsinki area and effects on air quality - an overview
Particulate matter sources in Helsinki area and effects on air quality - an overview Risto Hillamo 1, Topi Rönkkö 2 and Liisa Pirjola 3 1 Research and Development, Finnish Meteorological Institute 2 Department
Online Infrared Flue Gas Analyzer Gasboard 3000
Online Infrared Flue Gas Analyzer Gasboard 3000 O 2 CO CO2 SO 2 NO NO x Use of measurement methods Proprietary Infrared NDIR Detectors - Micro-flow in ppm range: CO, NO, SO2, CO2 - Dual beam in % volume
Reducing your Total Building Emissions using High Efficiency Energy Recovery :
Reducing your Total Building Emissions using High Efficiency Energy Recovery : By Duncan Goodall and James Fisher Email: [email protected]/[email protected] Mobile Duncan: 07595567015
Accessories SUNNY REMOTE CONTROL
Accessories SUNNY REMOTE CONTROL Mounting Instructions SRC20-MO-IEN121010 IMEN-SRC20 Version 1.0 EN SMA Solar Technology AG 1 Information on this Document 1 Information on this Document Validity This
Fundamentals of Mass Flow Control
Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of
IntesisBox Modbus Server SAMSUNG Air Conditioners
IntesisBox Modbus Server SAMSUNG Air Conditioners Gateway for monitoring and control of Samsung NASA compatible Air Conditioning Systems from any Modbus master device TCP or RTU (BMS, PLC, SCADA, HMI,
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
Air Pollution Control Division. Technical Services Program
Page 1 of 10 Air Pollution Control Division Technical Services Program APPENDIX PM4 STANDARD OPERATING PROCEDURE FOR THE DETERMINATION OF PM 10 AND PM 2.5 IN AMBIENT AIR USING A GRIMM EDM 180 Page 2 of
AEROSOL STATISTICS LOGNORMAL DISTRIBUTIONS AND dn/dlogd p
Concentration (particle/cm3) [e5] Concentration (particle/cm3) [e5] AEROSOL STATISTICS LOGNORMAL DISTRIBUTIONS AND dn/dlod p APPLICATION NOTE PR-001 Lonormal Distributions Standard statistics based on
CenterMind P+ Intelligent Power Management for your Data Center
CenterMind P+ Intelligent Power Management for your Data Center Intelligent Power Management for your Data Center monitor and manage your data center power consumption and connectivity efficiently The
DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL 854060
DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL 854060 (FOR DUSTTRAK II AND DRX AEROSOL MONITOR MODELS 8540 AND 8543) OPERATION AND MAINTENANCE MANUAL P/N 6008416, REVISION A MAY 2015 Copyright TSI Incorporated
CenterMind G+ Smart and Proactive Environment Monitoring
CenterMind G+ Smart and Proactive Environment Monitoring Smart and Proactive Environment Monitoring real-time visibility into the state of your computer room or data center environment, 24/7 RiT CenterMind
To suit Daikin FDY "F" Series and FD "F" Series. (and other models incorporating BRC1B62/52 Remote Controller)
OPERATION MANUAL Split System Ducted Air Conditioner To suit Daikin FDY "F" Series and FD "F" Series AIR CONDITIONER A hr o C (and other models incorporating BRC1B62/52 Remote Controller) Dear Owner, Thank
Cooling Small Server Rooms Can Be. - Jim Magallanes Computer Room Uptime: www.cruptime.com Uptime Racks: www.uptimeracks.com
Cooling Small Server Rooms Can Be Inexpensive, Efficient and Easy - Jim Magallanes Computer Room Uptime: www.cruptime.com Uptime Racks: www.uptimeracks.com Server Rooms Description & Heat Problem Trends
Near Real-time Monitoring of Diesel Exhaust Particulate (DEP)
Near Real-time Monitoring of Diesel Exhaust Particulate (DEP) 30 th April 2015 James Forder [email protected] Introduction What is Diesel exhaust particulate (DEP)? Constituent of diesel engine
PRODUCT CONFORMITY CERTIFICATE
PRODUCT CONFORMITY CERTIFICATE This is to certify that the 1. SWAM 5a 24 hour Dual Channel Monitor with both PM 10 and PM 2.5 pre-separators; 2. SWAM 5a 24 hour Single Channel Monitor with PM 10 preseparator;
USER S MANUAL FH052EAV1 FH070EAV1. System Air Conditioner (Cooling and Heating) ENGLISH ESPAÑOL FRANÇAIS ITALIANO PORTUGUÊS DEUTSCH E HNIKA
USER S MANUAL FH052EAV1 FH070EAV1 E HNIKA PORTUGUÊS ENGLISH ESPAÑOL ITALIANO DEUTSCH FRANÇAIS System Air Conditioner (Cooling and Heating) E S F I P D G DB98-29263A(1) Safety Precautions Register your
HUS-IPS-5100S(D)-E (v.4.2)
Honeywell s HUS-IPS-5100S(D)-E is a controller-based IP SAN unified storage appliance. Designed for centralized mass data storage, this IP SAN solution can be used with the high performance streaming server
IntesisBox KNX LG Air Conditioning
IntesisBox KNX LG Air Conditioning Gateway for integration of LG Air Conditioners with KNX control systems. 1. Main Features Direct connection to KNX TP-1 (EIB) bus. Direct connection to LG outdoor unit
USER S MANUAL HSC-24A
AIRREX AIR CONDITIONER USER S MANUAL HSC-24A Thank you for purchasing an AIRREX AIR CONDITIONER. BEFORE operation please read this user s manual carefully. Keep this manual readily available. It is ESSENTIAL
TDA2822 DUAL POWER AMPLIFIER SUPPLY VOLTAGE DOWN TO 3 V LOW CROSSOVER DISTORSION LOW QUIESCENT CURRENT BRIDGE OR STEREO CONFIGURATION
TDA2822 DUAL POER AMPLIFIER SUPPLY VOLTAGE DON TO 3 V. LO CROSSOVER DISTORSION LO QUIESCENT CURRENT BRIDGE OR STEREO CONFIGURATION DESCRIPTION The TDA2822 is a monolithic integrated circuit in 12+2+2 powerdip,
WEATHER STATION FOR SOLAR FARM MONITORING
WEATHER STATION FOR SOLAR FARM MONITORING SOLAR FARM MONITORING SYSTEM: Measures global, horizontal, & background irradiance. Measures wind speed, wind direction, ambient temperature, and relative humidity.
USB-500/600 Series Low-Cost Data Loggers and Accessories
Low-Cost Data Loggers and Accessories Features Stand-alone, remote data loggers and portable logger assistant Measure temperature, humidity, voltage, current, or event/state change 1 or 2 channels Low
Causes of High Relative Humidity Inside Air Conditioned Buildings. Roger G.Morse AIA, Paul Haas CSP, CIH Morse Zehnter Associates
Causes of High Relative Humidity Inside Air Conditioned Buildings Roger G.Morse AIA, Paul Haas CSP, CIH Morse Zehnter Associates S. FLORIDA AVE. AMBIENT 91 F db / 78 F wb 16,000 HOURS ANNUALLY THAT wb
Carson Dorn, Inc. December 18, 2012. Patrick Holmes Associate Manager Juneau I, LLC 645 G Street, Suite 100-604 Anchorage, AK 99501. Dear Mr.
Carson Dorn, Inc. 712 West 12th Street Juneau, Alaska 9981 December 18, 212 Patrick Holmes Associate Manager Juneau I, LLC 645 G Street, Suite 1-64 Anchorage, AK 9951 Dear Mr. Holmes, The renovation of
2.1 The performance of bus air conditioning system is evaluated in the following operating conditions
1.0 Scope:- This test procedure is applicable for testing the Air Conditioning system performance at hot ambient and very hot ambient condition on intracity urban bus in a climatic hot chamber. This procedure
Solar chilled drinking water sourced from thin air: modelling and simulation of a solar powered atmospheric water generator
20th International Congress on Modelling and Simulation, Adelaide, Australia, 1 6 December 2013 www.mssanz.org.au/modsim2013 Solar chilled drinking water sourced from thin air: modelling and simulation
City of San Diego Urban Runoff Management Program. Appendix XI. Minimum BMPs for Mobile Businesses
City of San Diego Urban Runoff Management Program This Page Left Intentionally Blank. Minimum Best Management Practices (BMPs) for Mobile Businesses No. BMP Title Description and Examples Justification
Smart. Scalable. Simple.
Thermo Scientific Wireless Monitoring Solution Smart. Scalable. Simple. providing the protection you need with the simplicity you deserve Thermo Scientific Wireless Monitoring Solution Optimized Sample
CASE STUDY: COOLERADO CORPORATION
CASE STUDY: COOLERADO CORPORATION Coolerado uses Weather Analytics climate data to demonstrate to clients the energy-efficiency of their air conditioners in places around the globe. 02 Overview Coolerado
Continuous Monitoring of Flue and Combustible or Hazardous Gases and Vapors
Continuous Monitoring of Flue and Combustible or Hazardous Gases and Vapors www.static-control.in 1 Introduction The detection or monitoring of flue gases and or hazardous gases has been a complex subject
ADVANCED TREATMENT OF RAINWATER USING METAL MEMBRANE COMBINED WITH OZONATION
ADVANCED TREATMENT OF RAINWATER USING METAL MEMBRANE COMBINED WITH OZONATION Ree Ho Kim 1*, Sangho Lee 1, Seog-Ku Kim 1, and Jong-Oh Kim 2 1 Korea Institute of Construction Technology, 2311 Deahwa-Dong,
Engineering White Paper UTILIZING ECONOMIZERS EFFECTIVELY IN THE DATA CENTER
Engineering White Paper UTILIZING ECONOMIZERS EFFECTIVELY IN THE DATA CENTER SUMMARY A proper balance of temperature and humidity control is essential for efficient data center operation, and integral
SPECTRACOOL AIR CONDITIONERS PENTAIR EQUIPMENT PROTECTION MCLEAN IS NOW HOFFMAN COOLING
SPECTRACOOL AIR CONDITIONERS PENTAIR EQUIPMENT PROTECTION MCLEAN IS NOW HOFFMAN COOLING HOFFMAN SPECTRACOOL INDOOR/OUTDOOR AIR CONDITIONERS ENERGY-EFFICIENT, FILTERLESS AIR CONDITIONERS NOW AVAILABLE WITH
HR Heated HB Blower. Smart Solutions for Drying Compressed Air. Dual Tower Heat Reactivated Desiccant Air Dryer
Smart Solutions for Drying Compressed Air Dual Tower Heat Reactivated Desiccant Air Dryer HR Heated HB Blower 800.943.794 www.engineeredairproducts.com Engineered Air Products Moisture in compressed air
S300 Series Dust Monitors
S301 DustGuard S303 Trend Monitor S304 Emission Monitor S305 Stack Monitor Intrinsically Safe (ATEX) Dust Monitors Split Architecture (Remote)Models Data Logging Software S300 Series Dust Monitors Dust
Monitoring of air pollution spread on the car-free day in the city of Veszprém
INTERNATIONAL JOURNAL OF ENERGY AND ENVIRONMENT Volume 5, Issue 6, 2014 pp.679-684 Journal homepage: www.ijee.ieefoundation.org Monitoring of air pollution spread on the car-free day in the city of Veszprém
Climate solutions. evaporative cooling, adiabatic humidification and programmable controllers. energy saving solutions for data centers. carel.
Climate solutions evaporative cooling, adiabatic humidification and programmable controllers energy saving solutions for data centers carel.com Data center air-conditioning Data center air-conditioning
Ted Wittemann P.G., Sales and Support Specialist
Ted Wittemann P.G., Sales and Support Specialist Introduction What is Web Based Remote Monitoring Applications Real World Experiences Pine GTS Environet Website Demo Wrap-up and Question 20 years of providing
Waste Heat Recovery through Air Conditioning System
International Journal of Engineering Research and Development e-issn: 2278-067X, p-issn : 2278-800X, www.ijerd.com Volume 5, Issue 3 (December 2012), PP. 87-92 Waste Heat Recovery through Air Conditioning
CONTROLLED ATMOSPHERE Application for Milling Industry
IAOM 4 th Annual Southeast Asia District Conference & Expo CONTROLLED ATMOSPHERE Application for Milling Industry NGUYEN VAN PHAT Technical Manager EcO 2 Vietnam [email protected] October 2013 EcO 2 BV The
General Purpose Thermal Mass Flowmeter Model 4140/4143
Mass Flowmeters General Purpose Thermal Mass Flowmeter Model 4140/4143 Operation and Service Manual 1980383, Revision D January 2009 Mass Flowmeters General Purpose Thermal Mass Flowmeter Model 4140/4143
Theoretical Study on Separate Sensible and Latent Cooling Air-Conditioning System
Purdue University Purdue e-pubs International Refrigeration and Air Conditioning Conference School of Mechanical Engineering 2008 Theoretical Study on Separate Sensible and Latent Cooling Air-Conditioning
37-7498. Professional Contractor INSTALLATION GUIDE
37-7498 Professional Contractor INSTALLATION GUIDE Wireless Setup Connect Exit Next Exit 3 EASY STEPS Installing Sensi thermostat isn t that different than installing a non Wi-Fi thermostat. 1. Install
