OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD
|
|
|
- Duane Wilson
- 10 years ago
- Views:
Transcription
1 OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO /4) at TSI. This method of calibration is common for all optical particle sizing instruments. In order to calculate mass from this size information there are assumptions that must be made. Standard assumptions are that the particles are spheres and by supplying a density the mass of the particles can be determined. In reality, the particles may not be spheres and the light scatter of the particles can be significantly different than that of the calibration particles (white PSL spheres). Corrections for these differences may be incorporated into an effective density which is the material density multiplied by a correction factor for the aerosol being measured. The calibration consists of three main steps: 1. Create a mass reference 2. Calculate the effective density 3. Apply the effective density to the OPS Note that the OPS Pump Always On and Dead Time Correction options should be enabled. The former is to avoid contamination of the optical chamber by the test aerosols, and the latter is to minimize coincidence error. During the test, the total particle concentrations also need to be lower than the OPS upper concentration limit which is 3000 particles/cm 3. The following steps illustrate how to calibrate the OPS for mass measurement. Once calibrated, the OPS will measure mass with reasonable accuracy as long as the aerosols of interest are the same as the aerosol used for calibration, and the size distributions remain the same. If the aerosol types or size distributions are different, a new calibration may be required. Note: Mass is a post-processing calculation. Density information can be entered either before or after data has been taken and it will be reflected in the graphs, tables and statistics. Create a Mass Reference The reference mass can be created using the OPS internal filter or an external filter. The external filter method is preferred since the mass collected by the OPS internal 37 mm internal filter may be biased by transport losses inside the OPS. Aerosol Instrument Manager, TSI, and TSI logo are trademarks of TSI Incorporated.
2 External Filter (Preferred Method) The calibration steps are outlined below: 1. Setup an external filter measurement system. To avoid sampling bias, the flow rate and sample time of the external filter measurement system should be the same or very close to the OPS flow rate and sample time. 2. Weight the new filter with a microbalance. You may need to minimize the static charge effect by exposing the filter to a bipolar ion source before weighting it. 3. Clean the filter cartridge before placing the new filter in it. 4. Co-locate the two instruments as shown in figure 1. Aerosol Souce Optical Particle Sizer OPS To Flow Control and Vacuum Figure 1. Experimental setup for gravimetric calibration 5. To avoid test aerosol filling up the optical chamber, DO NOT hook up the test aerosol to the OPS before starting the OPS pump. 6. Start the OPS. If the Pump Always On is enabled, the pump should be running after warm up. 7. Introduce the aerosol of interest into the OPS and the external filter measurement system, and start the measurement. Since the amount of mass collected by the filter depending on the aerosol density, size distribution and concentration, you need to adjust the sample time appropriately to get reliable mass measurement. 8. Stop the measurement. Take the filter out and measure it with the microbalance. 9. Using the mass collected on the external filter, calculate the mass concentration by using the equation shown below: Mass Collected Mass concentration = Flow Rate Sample Time Internal Filter The OPS has a 37-mm filter downstream of the optical chamber that can be used for mass calibration. A new filter is required for every calibration. Note that the mass collected by this filter may be biased by the transport losses inside the OPS, so it is recommended to use external filter measurement system if possible. The calibration steps are outlined below: -2-
3 1. Weight the new filter with a microbalance. You may need to minimize the static charge effect by exposing the filter to a bipolar ion source before weighting it. 2. Clean the filter cartridge before placing the new filter in it. Put the assembly back to the OPS. Make sure the assembly is secured to avoid leaking. 3. To avoid test aerosol filling up the optical chamber, DO NOT hook up the test aerosol to the OPS before the pump starts. 4. Start the OPS. If the Pump Always On is enabled, the pump should be running after warm up. 5. Introduce the aerosol of interest into the OPS, and start the measurement. Since the amount of mass collected by the filter depending on the aerosol density, size distribution and concentration, you need to adjust the sample time appropriately to get reliable mass measurement. 6. Stop the measurement. Take the filter out and measure it with the microbalance. 7. Using the mass collected on the 37 mm filter, calculate the mass concentration by using the equation shown below: Mass Collected Mass concentration = Flow Rate Sample Time The flow rate and sample time are the OPS flow rate and the OPS sample time. Calculate the effective density 1. Set the OPS particle density to 1 g/cm3, and record the integrated mass concentration. The effective density can be calculated by: Effective density = mass concentration from the filter 3 1 g/cm OPS integrated mass concentration Apply the effective density to the OPS (on instrument) 1. Set the OPS particle density to the effective density calculated above. a. On the OPS select the Setup tab at the bottom of the screen. b. Select the Sampling icon. c. Select the User Calibration icon. d. Enter your effective density in the Density box as shown in Figure
4 Figure 2. User Calibration screen 2. The OPS will now use this effective density in the displayed mass results both in graphs and tables. Apply the effective density to the OPS (in software) 1. If data is being collected using Aerosol Instrument Manager software, the density information is entered using the OPS Properties Dialog Box. 2. When you start the Aerosol Instrument Manager software and select the OPS 3330 Data type, the OPS 3330 Properties window opens. Select the Data Setup tab and select the Set Up Channels button. 3. If you select the Particle Density check box, you should then enter your effective density in the Density box. 4. If density varies by particle size, it is possible to do a more advanced correction and supply different densities for each size channel. If you uncheck the box in Particle Density portion of the Channel Setup window, you can enter different densities for each of the Channels. Figure 3 shows an example of Channel Setup using different densities. -4-
5 Figure 3. Channel Setup Screen with different densities 5. The Aerosol Instrument Manager software will now reflect your effective density value(s) in the mass results both in graphs, tables and statistics. -5-
6 TSI Incorporated Visit our website for more information. USA Tel: UK Tel: France Tel: Germany Tel: India Tel: China Tel: Singapore Tel: OPS-001 Rev. B 2012 TSI Incorporated Printed in U.S.A.
REAL-TIME DUST AND AEROSOL MONITORING
REAL-TIME DUST AND AEROSOL MONITORING THE DUSTTRAK II AND DRX AEROSOL MONITORS UNDERSTANDING, ACCELERATED REAL-TIME DUST MONITORING. ANY ENVIRONMENT. ANY APPLICATION. The DustTrak II and DRX Aerosol Monitors
CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS
CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS UNDERSTANDING, ACCELERATED TSI SETS THE STANDARD FOR PARTICLE COUNTING Performance Semiconductor, hard disk drive, and flat panel manufacturers
HEATED INLET SAMPLE CONDITIONING EFFECTS ON REMOTE DUST MONITORING
HEATED INLET SAMPLE CONDITIONING EFFECTS ON REMOTE DUST MONITORING APPLICATION NOTE EXPMN-8 (A4) Introduction Under high ambient humidity conditions, photometric mass measurements have been reported to
FITPRO+ SOFTWARE V3 IMPORTING DATA FROM FITPRO V2 & FITPLUS3
FITPRO+ SOFTWARE V3 IMPORTING DATA FROM FITPRO V2 & FITPLUS3 APPLICATION NOTE RFT-020 (US) T his quick reference guide provides instructions for importing data into a FitPro+ v3.0, v3.1, or v3.2 software
APPLICATION PROGRAM INTERFACE (API) FOR AIRASSURE WEB PM2.5 SOFTWARE
APPLICATION PROGRAM INTERFACE (API) FOR AIRASSURE WEB PM2.5 SOFTWARE APPLICATION NOTE PM2.5-003 (US) Introduction TSI has released an API for the AirAssure Web Software that will enable customers to extract
NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS
NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS APPLICATION NOTE ITI-089 I. Introduction The technology used in personal sampling pumps has not changed much over the past 20 years. II. There
TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS
TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS APPLICATION NOTE CC-105 (US) Introduction This application note will demonstrate how real-time viable particle count data will
Printing Fit Test Cards with an ID Card Printer
Respirator Fit Testing Printing Fit Test Cards with an ID Card Printer Application Note RFT-013 ID Card Printer Setup TSI Incorporated does not claim that the FITPRO fit test card printing features will
AEROSOL STATISTICS LOGNORMAL DISTRIBUTIONS AND dn/dlogd p
Concentration (particle/cm3) [e5] Concentration (particle/cm3) [e5] AEROSOL STATISTICS LOGNORMAL DISTRIBUTIONS AND dn/dlod p APPLICATION NOTE PR-001 Lonormal Distributions Standard statistics based on
THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS
THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS INDOOR AIR QUALITY INSTRUMENTS UNDERSTANDING, ACCELERATED BREATHE A LITTLE EASIER WITH TSI Indoor air quality is a growing concern. With the increasing
PERCENT OUTDOOR AIR (%OA) CALCULATION AND ITS USE
PERCENT OUTDOOR AIR (%OA) CALCULATION AND ITS USE APPLICATION NOTE TI-138 TSI s IAQ-CALC TM Indoor Air Quality Meter and TH-CALC TM Thermohygrometer calculate Percent Outdoor Air (%OA). In order to take
Indoor air quality monitoring made easy
Indoor air quality monitoring made easy with the Fluke 985 Particle Counter Application Note Introduction Energy savings is top of facilities owners agenda. HVAC system tends to be the largest contributor
ab204067-vacuum manifold for 96 well filter plates This product is for research use only and is not intended for diagnostic use.
ab204067-vacuum manifold for 96 well filter plates This product is for research use only and is not intended for diagnostic use. Version 2 Last Updated 06 November 2015 Table of Contents INTRODUCTION 1.
PREMIUM CLEAN ROOM MONITOR MODELS 8630-CRM-S 8630-CRM-P
PREMIUM CLEAN ROOM MONITOR MODELS 8630-CRM-S 8630-CRM-P MANUAL SUPPLEMENT Contents of this manual supplement include: Sequence of Operation Menu Structure Description of New Software Items Deleted Software
Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations
Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations B e c k m a n C o u l t e r L i f e S c i e n c e s 2 5 0 S. K r a e m e r B o u l e
Email Export. Exporting Client Email Addresses
Email Export Exporting Client Email Addresses STX has the ability to export email addresses from the Client file. Once those addresses are exported, STX has nothing else to do with getting the addresses
How To Install Safari Antivirus On A Dv8000 Dv Recorder On A Pc Or Macbook Or Ipad (For A Pc) On A Microsoft Dv8 (For Macbook) On An Ipad Or Ipa (
Using Symantec AntiVirus Corporate Edition Version 9.0 Software On a DX8000 DVR DX8000 Digital Video Recorder C1613M-A (12/04) Contents Using Symantec AntiVirus Corporate Edition 9.0 Software.....................................................................5
www.ecomonitoring.com
Ambient Air Monitoring Systems ECM ambient air monitoring systems are integrated in air conditioned thermal isolated light weight shelters allowing stationary and mobile installations. The usually monitored
Oilfield Services Measurement Solutions
Oilfield Services Measurement Solutions Superior flow and density measurement Best-in-class measurement How can I streamline my operations for efficiency without compromising environmental, safety and
PARTICLE GENERATOR MODEL 8026
PARTICLE GENERATOR MODEL 8026 OPERATION AND SERVICE MANUAL P/N 1980319, REVISION M MARCH 2016 Copyright TSI Incorporated / Revision M / 2000 2016 / All rights reserved. Address TSI Incorporated / 500
Sub-100 nm calibration for size and number
Calibration of DMS Series Fast Particulate Spectrometers Contents Introduction...1 Sub-100 nm calibration for size and number...2 Supra-100 nm calibration for size...3 Certification...4 Agglomerate Aerosols:
Coarse PM Methods Evaluation Study
Coarse PM Methods Evaluation Study Study Design and Preliminary Results Thomas Ellestad, Robert Vanderpool, Paul Solomon, and Mary Harmon USEPA/ORD/NERL Sanjay Natarajan, Chris Noble, and Bob Murdoch Research
Small Animal Respiratory Exchange Ratio (RER)
iworx Physiology Lab Experiment Experiment AMe-1 Small Animal Respiratory Exchange Ratio (RER) Note: The lab presented here is intended for evaluation purposes only. iworx users should refer to the User
Integrating CaliberRM with Software Configuration Management Tools
Integrating CaliberRM with Software Configuration Management Tools A Borland White Paper By Jenny Rogers, CaliberRM Technical Writer January 2002 Contents Introduction... 3 Enabling SCM for a Project...
Density Measurement. Technology: Pressure. Technical Data Sheet 00816-0100-3208 INTRODUCTION. S min =1.0 S max =1.2 CONSTANT LEVEL APPLICATIONS
Technical Data Sheet 00816-0100-3208 Density Measurement Technology: Pressure INTRODUCTION Pressure and differential pressure transmitters are often used to measure the density of a fluid. Both types of
Technical Cleaning Ó
Technical Cleaning JUSTIFICATION Cleaning of Data Centers and Communications Rooms should be performed by qualified personnel who will use techniques and equipment designed to clean without re-contaminating
Taking Attendance in Salesforce with the Attendance Tracker
Taking Attendance in Salesforce with the Attendance Tracker Contents To Access the Attendance Tracker... 2 Set up a Master Record for the Program Cycle... 3 Enrolling Participants... 4 To Add Individual
XCal-View user manual
XCal-View user manual XCal-View user manual M-9925-0107-04 1 Introduction to XCal-View About XCal-View Renishaw XCal-View software has been written as a direct replacement for the previous analysis package
How to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
LOAD BALANCING 2X APPLICATIONSERVER XG SECURE CLIENT GATEWAYS THROUGH MICROSOFT NETWORK LOAD BALANCING
SECURE CLIENT GATEWAYS THROUGH MICROSOFT NETWORK LOAD BALANCING Contents Introduction... 3 Network Diagram... 3 Installing NLB... 3-4 Configuring NLB... 4-8 Configuring 2X Secure Client Gateway... 9 About
Calibration Certificate
P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: [email protected] SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING
ILLUMINATE ASSESSMENT REPORTS REFERENCE GUIDE
ILLUMINATE ASSESSMENT REPORTS REFERENCE GUIDE What are you trying to find? How to find the data in Illuminate How my class answered each question (Response Frequency) 3. Under Reports, click Response Frequency
FACILITY MONITORING SYSTEM DESIGN RECOMMENDATIONS
FACILITY MONITORING SYSTEM DESIGN RECOMMENDATIONS APPLICATION NOTE A monitoring system is referred to by a number of different names, a Facility Monitoring System (FMS), a Continuous Monitoring System,
How to send emails triggered by events
Complete. Simple. Affordable Copyright 2014 AlienVault. All rights reserved. AlienVault, AlienVault Unified Security Management, AlienVault USM, AlienVault Open Threat Exchange, AlienVault OTX, Open Threat
Chemical Measurement Solutions
Chemical Measurement Solutions Superior flow and density measurement Best-in-class measurement I need to simultaneously increase production and plant up-time while lowering operations and maintenance costs.
SYSTEC Degasser WatchDog
SYSTEC Degasser WatchDog DataApex Clarity Control Doc. #: DI330508 Rev. A Praha, May 2013 Science Instruments and Software, s.r.o. Havlovská 1114/14 160 00 Praha 6, Czech Republic TEL. +420 246 037 483
Sample preparation for X-ray fluorescence analysis
Technical articles Sample preparation for X-ray fluorescence analysis III. Pressed and loose powder methods Gakuto Takahashi* 1. Introduction There are two main sample preparation techniques for measurement
2009 Braton Groupe sarl, All rights reserved.
D A T A B A S E M A N A G E R U S E R M A N U A L The information in this manual is not binding and may be modified without prior notice. Supply of the software described in this manual is subject to a
Environmental Monitoring
Chapter 10 Environmental Monitoring By the end of this chapter, you will be able to: l List the types of environmental testing l Name three types of checks that are carried out to ensure the isolator is
Application Note #231 Revision B November 2010. Using a VPN Connection to Remotely Access Lutron Systems. Overview. Key Terms. Before You Begin...
Using a VPN Connection to Remotely Access Lutron Systems Application Note #231 Revision B November 2010 Overview This application note describes one method of configuring remote access to a Lutron system
DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL 854060
DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL 854060 (FOR DUSTTRAK II AND DRX AEROSOL MONITOR MODELS 8540 AND 8543) OPERATION AND MAINTENANCE MANUAL P/N 6008416, REVISION A MAY 2015 Copyright TSI Incorporated
Snap 9 Professional s Scanning Module
Miami s Quick Start Guide for Using Snap 9 Professional s Scanning Module to Create a Scannable Paper Survey Miami s Survey Solutions Snap 9 Professional Scanning Module Overview The Snap Scanning Module
Image Lab Software for the GS-900 Densitometer
Image Lab Software for the GS-900 Densitometer Quick Start Guide Catalog # 170-9690 Bio-Rad Technical Support For help and technical advice, please contact the Bio-Rad Technical Support department. In
Trimble CFX-750 Display with the Yield Monitoring System
Trimble CFX-750 Display with the Yield Monitoring System Quick Reference Card This Quick Reference Card is intened to help you with the basic operation and calibration of the Trimble CFX-750 display with
INPEX CONTRACT MANAGEMENT SYSTEM
INPEX CONTRACT MANAGEMENT SYSTEM Contractors Manual - Contract Management Document No.: PROCON-00276 Document no.: PROCON-00276 Page 1 NOTICE All information contained with this document has been classified
11th International Conference on Urban Drainage, Edinburgh, Scotland, UK, 2008 MONITORING EQUIPMENT AND RESULTS The flow monitoring and sampling programme was conducted over two one week periods, with
Quick Start Guide Using OneDisk with the Tappin Service
Quick Start Guide Using OneDisk with the Tappin Service Copyright 2013, Tappin, Inc. All rights reserved. Tappin and the Tappin logo are trademarks of Tappin, Inc. All other trademarks are the property
Scanning Mobility Particle Sizer
PARTICLE INSTRUMENTS Highly accurate, real-time nanoparticle sizing systems you can rely on for years Features and Benefits Model 3936N88 High resolution data: up to 167 channels Broad size range: from
Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System
Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Carlton G. Slough TA Instruments, 109 Lukens Drive, New Castle DE 19720, USA ABSTRACT The use of
FiberOptix TM. AP Sensor Zeroing and Calibration Guide INTRA-AORTIC BALLOON PRODUCT SUPPORT HOTLINE. U.S. and Canada: 1-800-447-IABP (4227)
FiberOptix TM AP Sensor Zeroing and Calibration Guide INTRA-AORTIC BALLOON PRODUCT SUPPORT HOTLINE U.S. and Canada: 1-800-447-IABP (4227) Worldwide: 617-389-8628 Caution: U.S. Federal Law limits this device
Activity P13: Buoyant Force (Force Sensor)
Activity P13: Buoyant Force (Force Sensor) Equipment Needed Qty Equipment Needed Qty Economy Force Sensor (CI-6746) 1 Mass and Hanger Set (ME-9348) 1 Base and Support Rod (ME-9355) 1 Ruler, metric 1 Beaker,
RITTER Multiple Sonic Nozzle Calibration System
RITTER Multiple Sonic Nozzle Calibration System »Wouldn t it be great to eliminate doubtful measurement results by using proven measurement technology? Providing the most precise results ensures and increases
Quality Assurance for the Analysis of Steel by Gas Component Analysis
Quality Assurance for the Analysis of Steel by Gas Component Analysis Alain Salaville, Jérôme Barraqué, JOBIN YVON SAS, Longjumeau, FRANCE Keywords: quality assurance, steel 1 Overview Quality assurance
Air Pollution Control Division. Technical Services Program
Page 1 of 10 Air Pollution Control Division Technical Services Program APPENDIX PM4 STANDARD OPERATING PROCEDURE FOR THE DETERMINATION OF PM 10 AND PM 2.5 IN AMBIENT AIR USING A GRIMM EDM 180 Page 2 of
ADVANCES IN AUTOMATIC OPTICAL INSPECTION: GRAY SCALE CORRELATION vs. VECTORAL IMAGING
ADVANCES IN AUTOMATIC OPTICAL INSPECTION: GRAY SCALE CORRELATION vs. VECTORAL IMAGING Vectoral Imaging, SPC & Closed Loop Communication: The Zero Defect SMD Assembly Line Mark J. Norris Vision Inspection
Report: Salesforce and Microsoft Dynamics CRM Outlook Integration Compared
Report: Salesforce and Microsoft Dynamics CRM Outlook Integration Compared CRM Switch CRM Assessment & Selection Services www.crmswitch.com 800 799 4539 Page 1 Table of Contents Introduction Functional
QIAsymphony RGQ Protocol Sheet
QIAsymphony RGQ Protocol Sheet Settings to run the artus CT/NG QS-RGQ Kit (Rotor-Gene Q software.) Check availability of new electronic labeling revisions at www.qiagen.com/products/artusctngqsrgqkitce.aspx
DomainSherpa.com: The Domain Name Authority. Google AdWords Keyword Planner for Domain Name Investors
Google AdWords Keyword Planner for Domain Name Investors Watch the full video at: http://www.domainsherpa.com/google-adwords-keyword-planner/ The Google Adwords Keyword Tool is dead. Long live the Google
Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document 74-461
Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/0/202 Adapted from National Bureau of Standards Document 74-46 The purpose of calibrating glassware is to determine the volume of
Standard Operating Procedure for Training for Staff Working on the PM 2.5 Chemical Speciation Program
PMz.s-Staff Training Page 101'6 Standard Operating Procedure for Training for Staff Working on the PM 2.5 Chemical Speciation Program Environmental and Industrial Sciences Division RTI International* Research
Clearview Customer Web Access
Clearview Customer Web Access This document outlines the Clearview Process flows for Logging in to Customer Web Access Service Listing o Service Order Desktop o Service Tasks o Viewing Service Schedules
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,
FACILITY MONITORING SYSTEM DESIGN RECOMMENDATIONS
FACILITY MONITORING SYSTEM DESIGN RECOMMENDATIONS APPLICATION NOTE CC-114 (US) Introduction A monitoring system is referred to by a number of different names, a Facility Monitoring System (FMS), a Continuous
Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested
DOP Solutions Technical Document: Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested Issue 1: Prepared by Neil Stephenson and Tim Triggs and edited by
On the re-calibration of Ni-foil reference standards
Fischer Traceability Report 2012 07 On the re-calibration of Ni-foil reference standards The production of a 4 new pure Ni primary standards characterized by gravimetry /1/ is the basis of this re-calibration.
Manual Analysis Software AFD 1201
AFD 1200 - AcoustiTube Manual Analysis Software AFD 1201 Measurement of Transmission loss acc. to Song and Bolton 1 Table of Contents Introduction - Analysis Software AFD 1201... 3 AFD 1200 - AcoustiTube
Agilent High Sensitivity DNA Kit Guide
Agilent High Sensitivity DNA Kit Guide Agilent High Sensitivity DNA Agilent Technologies Notices Agilent Technologies, Inc. 2009, 2013 No part of this manual may be reproduced in any form or by any means
www.kem-kueppers.com [email protected]
www.kem-kueppers.com [email protected] Calibration of flow meters according to DIN EN ISO/IEC 17025:2005 Introduction... 2 Calibration laboratory... 3 Description of the calibration process... 3 Calibration
Activity P13: Buoyant Force (Force Sensor)
July 21 Buoyant Force 1 Activity P13: Buoyant Force (Force Sensor) Concept DataStudio ScienceWorkshop (Mac) ScienceWorkshop (Win) Archimedes Principle P13 Buoyant Force.DS P18 Buoyant Force P18_BUOY.SWS
Air and Water Balancing - Sound and Vibration Testing - Cleanroom Services - Building Commissioning Building Envelope Testing Thermal Imaging
NEUDORFER ENGINEERS, INC. provides our customer with unbiased, professional, performance testing of Cleanrooms. We are committed to providing you, our customer, with value-added expertise in an effort
TECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES. Randall Messman SW Regional Sales Manager
TECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES Randall Messman SW Regional Sales Manager PGI International 16101 Vallen Drive Houston, TX 77041 Purpose Natural
Setting Up My Business Account - Wireless
How to Use Guide: My Business Account Setup Last updated: 8/18/2011 Overview The way you set up My Business Account determines how you manage your company's wireless accounts. You can create a company
M24LRxx/CR95HF application software installation guide
User manual M24LRxx/CR95HF application software installation guide Introduction This user manual describes the procedures to install the different software drivers required to use the DEVKIT-M24LR-A development
Experiment (13): Flow channel
Introduction: An open channel is a duct in which the liquid flows with a free surface exposed to atmospheric pressure. Along the length of the duct, the pressure at the surface is therefore constant and
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
Configuring Outlook 2016 for Windows
Configuring Outlook 2016 for Windows This document assumes that you already have Outlook 2016 installed on your computer and you are ready to configure Outlook. Table of Contents Configuring Outlook 2016
- 2 - IMPORTANT SAFETY REMINDERS
USER MANUAL IMPORTANT SAFETY REMINDERS This appliance should only be used for domestic cleaning, as described in this user guide. Please ensure that this guide is fully understood before operating the
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use Confidence in Results With Data Integrity 2 Make the World Healthier With the LC-MS Tests You Run Confidence in Test Results With Data
Accelerating Microsoft Exchange Servers with I/O Caching
Accelerating Microsoft Exchange Servers with I/O Caching QLogic FabricCache Caching Technology Designed for High-Performance Microsoft Exchange Servers Key Findings The QLogic FabricCache 10000 Series
Cell Sorting. S3e Cell Sorter Quick Guide
Cell Sorting S3e Cell Sorter Quick Guide STARTUP Checking Fluidics STARTUP There are three bulk fluid containers (Figure 1) that may require attention before starting the system. If fluidics need to be
HOW THE CONTAMINATION IS MEASURED CONTAMINATION CLASSES ACCORDING TO ISO 4406:1999
CONTAMINATION CONTROL in hydraulic systems in the hydraulic system is a very wide and complex matter; the following is just a short summary. Our Customer Service is at your disposal for any further information.
Gas Custody Transfer Calibration
Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform
Office 365 Mail: Mac Mail OS X 10.5-10.8
Office 365 Mail: Mac Mail OS X 10.5-10.8 Office 365 Mail Client Configuration Guide Mac OS X Mail Client (Mac OS X 10.5 to OS X 10.8) Required Items: Mac OS X 10.5, Mac OS X 10.6, Mac OS X 10.7, OS X 10.8,
1805 Series Relief Valves
October 2011 1805 Series Relief Valves P1026 1805G Type 1805-2 Type 1805-4 Figure 1. Typical 1805 Relief Valves Introduction The 1805 Series relief valves are designed for use in farm tap applications
Filtering Spam Using Outlook s Rule
Filtering Spam Using Outlook s Rule No one enjoys receiving spam (unsolicited junk e-mail). However, at Cuesta College we do not delete suspected spam. Instead, to assure the delivery of legitimate e-mail,
Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection
Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection A wide range of products to meet the needs of transferring animals, operators,
Jump Start: Aspen HYSYS Dynamics V7.3
A Brief Tutorial (and supplement to training and online documentation) Glenn Dissinger, Product Director, Aspen Technology, Inc. Julie Levine, Associate Product Marketing Professional, Aspen Technology,
Experiment #4 Sugar in Soft Drinks and Fruit Juices. Laboratory Overview CHEM 1361. August 2010
Experiment #4 Sugar in Soft Drinks and Fruit Juices Laboratory Overview CHEM 1361 August 2010 Gary S. Buckley, Ph.D. Department of Physical Sciences Cameron University Learning Objectives Relate density
LX650 Lamp/Filter Replacement Instruction Sheet
LX650 Lamp/Filter Replacement Instruction Sheet INTRODUCTION These instructions outline the details for Lamp and Filter replacement in the LX650 projector (Model No. 103-012100-01). The lamp is located
Agilent 4339B/4349B High Resistance Meters
Agilent 4339B/4349B High Resistance Meters Technical Overview Within Budget Without Compromise Introducing the Agilent Technologies 4339B and 4349B High Resistance Meters Used for Making Ultra- High Resistance
InfplusWin OPERATING MANUAL
InfplusWin OPERATING MANUAL Edition: October 2010. Vers. 1.70 ASSEMBLAD - Technical office - Automotive Division Inf-Win170ManGB GB 1 Index 1 INFPLUS-WIN SOFTWARE...3 1.1 INSTALLATION...3 2 INFPLUS-WIN
ValveLink Mobile Software
Quick Start Guide ValveLink Mobile Software Version 4.0 ValveLink Mobile Software ValveLink Mobile software lets you configure, calibrate, and troubleshoot FIELDVUE DVC2000, DVC6000, DVC6000f, DVC6200,
Lamp and Filter Replacement Instructions
Introduction Lamp and Filter Replacement Instructions Use the following instructions when replacing the lamps or filters in any of the projectors listed in Table 1, below. The lamp is located at the rear
PLEUR EVAC CHEST DRAINAGE SYSTEMS. Great Innovations to Enhance Chest Drainage
PLEUR EVAC CHEST DRAINAGE SYSTEMS Great Innovations to Enhance Chest Drainage 2 INTRODUCTION Teleflex Great Innovations to Enhance Chest Drainage Trusted brands make Teleflex a reliable and strong partner.
Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley. Norton 0
Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley Norton 0 Norton 1 Abstract The charge of an electron can be experimentally measured by observing an oil
JustClust User Manual
JustClust User Manual Contents 1. Installing JustClust 2. Running JustClust 3. Basic Usage of JustClust 3.1. Creating a Network 3.2. Clustering a Network 3.3. Applying a Layout 3.4. Saving and Loading