Mass the Organic Chemistry Department. (A guide for novel users)
|
|
|
- Rosemary Ferguson
- 10 years ago
- Views:
Transcription
1 Mass the Organic hemistry Department (A guide for novel users) Peter M. van Galen Research Assistant Mass Spectrometry Organic hemistry Department Nijmegen University September 2005
2 0 Introduction Basic principles: Electron Impact (EI) and sector analysis Measurement principles Sampling The Ion source The separation of ions 'Single' focussing separation by magnetic deflection Double focusing separation Summary Quadrupole analyzer Ion trap Time-of-flight analyzer Resolution Some remarks on elemental composition calculations Error limits Double bond equivalent (unsaturation) Odd-electron and even-electron ions The nitrogen rule Isotope ratio measurements Examples Ionization Methods in Organic Mass Spectrometry Gas-Phase ionization Some general remarks on ionisation Electron Ionization (EI) hemical Ionization (I) Desorption hemical Ionization (DI) Negative-ion chemical ionization (NI) Field Desorption and Ionization Field Desorption (FD) Field Ionization (FI) Particle Bombardment Fast Atom Bombardment (FAB) Secondary Ion Mass Spectrometry (SIMS) Atmospheric Pressure Ionization (Spray Methods) Electrospray Ionization (ESI) Atmospheric Pressure hemical Ionization (API) Laser Desorption Matrix-Assisted Laser Desorption Ionization (MALDI) Some commonly used chemicals in mass spectrometry I Reagent Gases FAB matrices Location of charge and primary dissociation in molecular ions Location of charge Homolytic dissociation Butanol methyl-2-propanol N-ethyl-n-propylamine and N-(tert-butyl)-N-methylamine Ethylbenzene Heterolytic dissociation The McLafferty rearrangement The retro Diels-Alder reaction Stevenson s Rule...39
3 3.7 Further dissociation of fragment ions Remarks Loss of O from acylium ions Loss of alkenes from ethers, alcohols etcetera Formation of ion-series Literature and references Some printed literature: Tools used Appendix...44 Sample submission form...44 GMS guidelines...44
4 0 Introduction In mass spectrometry, one generates ions from a sample to be analyzed. These ions are then separated and determined. Separation is achieved on different trajectories of moving ions in electrical and/or magnetic fields. Mass-spectrometry has evolved from the experiments and studies early in the twentieth century that tried to explain the behavior of charged particles in magnetic and electrostatic force fields. Well-known names from these early days are J.J. Thompson investigation into the behavior of ionic beams in electrical and magnetic fields (1912), A.J. Dempster directional focussing (1918) and F.W. Aston energy focussing (1919). In this way a refinement of the technique was achieved that led to the gathering of important information concerning the natural abundance of isotopes. The first analytical applications then followed in the early forties when the first reliable commercial mass spectrometers were produced. This was mainly for the quantitative determination of the several components in complex mixtures of crude oil. In the beginning of the sixties the application of mass-spectrometry in identification and structure elucidation of more complex organic compounds started. Since then the technique has developed to a powerful and versatile, maybe even more then NMR, tool for this purpose. This booklet is a paraphrase of an earlier release back in When the new high-resolution mass-spectrometer was purchased in the beginning of 1999 and some years before that the purchase of a G/MS apparatus more and more the need was there to rewrite the existing manual. In this way more attention is paid to the several, partial new, ionization techniques and alternative ways to separate masses, as there are TOF (time of flight), Ion Trap and sector analysis. -1-
5 1 Basic principles: Electron Impact (EI) and sector analysis. Though the principles of a modern analytical mass-spectrometer are easily understood this doesn't account for the apparatus. A mass spectrometer especially a multi-sector instrument is one of the most complex electronic and mechanical devices one encounters as a chemist. Therefore this means high costs at purchase and maintenance besides a specialized training for the operator(s). 1.1 Measurement principles In the following figure the essential parts of an analytical mass-spectrometer are depicted. Its procedure is as follows: Ions with a large mass Ions with a small mass Recorder Vacuum Accelerating potential Magnetic Field (perpendicular to page) Ion beam Exit slit dc-amplifier ollector Anode Filament for electronbeam Sample molecules Ionisation area Sample leak Electrometer tube Figure 1: schematic reproduction of a mass-spectrometer 1. A little amount of a compound, typically one micromole or less is evaporated. The vapor is leaking into the ionization chamber where a pressure is maintained of about 10-7 mbar. 2. The vapor molecules are now ionized by an electron-beam. A heated cathode, the filament, produces this beam. Ionization is achieved by inductive effects rather then strict collision. By loss of valence electrons, mainly positive ions are produced. 3. The positive ions are forced out of the ionization chamber by a small positive charge (several Volts) applied to the repeller opposing the exit-slit (A). After the ions have left the ionization chamber, they are accelerated by an electrostatic field (A>B) of several hundreds to thousands of volts before they enter the analyzer. 4. The separation of ions takes place in the analyzer at a pressure of about 10-8 mbar. This is achieved by applying a strong magnetic field perpendicular to the motional direction of the ions. The fast moving ions then will follow a circular trajectory, due to the Lorenz acceleration, whose radius is determined by the mass/charge ratio of the ion and the -2-
6 strength of the magnetic field. Ions with different mass/charge ratios are forced through the exit-slit by variation of the accelerating voltage (A>B) or by changing the magnetic-field force. 5. After the ions have passed the exit-slit, they collide on a collector-electrode. The resulting current is amplified and registered as a function of the magnetic-field force or the accelerating voltage. The applicability of mass-spectrometry to the identification of compounds comes from the fact that after the interaction of electrons with a given molecule an excess of energy results in the formation of a wide range of positive ions. The resulting mass distribution is characteristic (a fingerprint) for that given molecule. Here there are certain parallels with IR and NMR. Massspectrograms in some ways are easier to interpret because information is presented in terms of masses of structure-components. 1.2 Sampling As already indicated a compound normally is supplied to a mass-spectrometer as a vapor from a reservoir. In that reservoir, the prevailing pressure is about 10 to 20 times as high as in the ionization chamber. In this way, a regular flow of vapor-molecules from the reservoir into the mass-spectrometer is achieved. For fluids that boil below about 150 o the necessary amount evaporates at room temperature. For less volatile compounds, if they are thermally stabile, the reservoir can be heated. If in this way sampling can't be achieved one passes onto to direct insertion of the sample. The quality of the sample, volatility and needed amount are about the same for massspectrometry and capillary gas chromatography. Therefore, the effluent of a G often can be brought directly into the ionization chamber. Use is then made of the excellent separating power of a G in combination with the power to identify of the mass-spectrometer. When packed G is used, with a much higher supply of carrier-gas, it is necessary to separate the carrier gas prior to the introduction in the mass-spectrometer (jet-separator). 1.3 The Ion source. Molecular leak Gas beam First accelerating Shield slit Heater Focus slit Second Electron accelerating slit Filament slit Repellers Ionizing region Anode Ion accelerating region Electron beam Figure 2: schematic representation of an ion-source. In figure 2, the scheme of an ionization chamber, ion-source, typically electron impact, is presented. In this chamber in several ways, ions of the compound to be investigated can be produced. The most common way is to bombard vapor-molecules of the sample with electrons of -3-
7 about 70 ev generated as described in 1.1. These electrons are generated by heating a metal wire (filament), commonly used are tungsten or rhenium. A voltage of about 70 Volts (from 5 to 100) accelerates these electrons towards the anode. During the bombardment, one or more electron can be removed from the neutral molecule thus producing positively charged molecular radicalions. Only about one in 10 3 of the molecules ABD + e ABD ABD + 2e AB + D or D + AB AB + D or D + AB A + BD or BD + A AD + B etc. etc. etc. Figure 3: possible fragmentation from a 'molecule' ABD. present in the source are ionized. The ionization probability differs among substances, but it is found that the cross-section for most molecules is a maximum for electron energies from approximately 50 to 100 ev. Most existing compilations of electron impact spectra are based on spectra recorded with approximately 70 ev electrons, since sensitivity is here close to a maximum and fragmentation is unaffected by small changes in electron energy around this value. During this ionization, the radical-ions on average gain an excess energy enough to break one or more bonds en hence producing fragment-ions. In figure 3 the possible fragmentation of a molecule ABD is presented. It should be stated here that this is a simplified presentation and that in real life a multitude of possible ways to form fragments even via re-arrangement reactions exists. 1.4 The separation of ions. There are several ways to separate ions with different mass/charge ratios, e.g. magnetic sector analyzers, quadrupole mass filters, quadrupole ion traps, time-of-flight analyzers and ion cyclotron-resonance instruments. The first two types presently account for the great majority of instruments used in organic chemistry. Ideally, when separating, it is possible to distinguish between ions with very little difference in mass/charge ratio while maintaining a high flow of ions. These conditions are not in agreement and a compromise should be reached. For some applications a nominal mass discrimination will do, for other applications a much higher resolving power is needed. For example when one needs to distinguish between ions 2 H 4 +, H 2 N +, N 2 + and O + (with respective masses of , , and amu) a resolving power of 0.01 mass units is needed. The main differences in mass-spectrometers are encountered in the way ions are separated 'Single' focussing separation by magnetic deflection. Separation in this way is effected by the application of a magnetic field perpendicular to the motion of the ions leaving the ion-source. Deflections of about 30 to 180 degrees are achieved (Fig.1). The trajectory of the ion follows from the applied forces: The Lorentz force and the centrifugal force. The Lorentz force is given by: B FL = Bzev (1) F F L Where B is the strength of the magnetic field, z the amount of charges, e the charge of one electron and v is the velocity. When traversing a radial path of m + Figure 4: the magnetic analyzer. -4-
8 curvature r through a magnetic field B this force equals the angular momentum: F mv r ion 2 = (2) The energy of the ion is expressed as: E kin zeu = 2 = 1 2 mv (3) Where U is the accelerating voltage. For an ion, that reaches the detector (1) equals (2), hence: Bzev mv r ion 2 = (4) Substitution of (4) in (3) and rearranging gives: m z 2 2 B r e 2U = (5) Note that the rearrangement of (4) to mv = Bzer demonstrates the fact that a magnetic sector is a momentum analyzer rather than a mass analyzer as is commonly assumed. Expressed in practical units the atomic mass (M) of a singly charged ion is given by: M B r U = (6) Where r is in centimeters, B is in tesla (1 tesla = 10 4 gauss), and U is in Volts. For example, a maximum field strength of 2 tesla gives a maximum mass just over Dalton for an instrument of 65 cm radius operating at an accelerating voltage of 8000V. Equation (5) shows that by varying either B or U ions of different m/z ratio, separated by the magnetic field, can be made to reach the collector. The most common form of mass scan is the exponential magnet scan, downward in mass. This has the advantage of producing mass-spectral peaks of constant width. The equations appropriate to this form of scan are: m kt = m 0 e (7) and t p t =. 303 R 2 10 (8) Where m 0 is the starting mass at time t=zero. M is the mass registered at time t. t p Is the peak width between its 5% points. t 10 Is the time taken to scan one decade in mass (e.g. m/z 500 to 50) and R is the resolving power measured by the 10% valley definition. Scanning of the accelerating voltage U apparently has advantages because of speed and ease of control, however causes defocusing and loss of sensitivity and is therefore rarely used. -5-
9 1.4.2 Double focusing separation. Mass the Organic hemistry Department Because the magnetic sector separates on basis of momentum ions with little difference in translational energy are not focussed in the same point. The spread in translational energy of the ions formed in an electron-impact source limits the resolving power. In F addition, source contamination leading to charging effects and contact potentials worsens this. Other ion sources like field desorption produce ions with an even F L r e larger spread in translational energy. In a double focusing mass spectrometer, the ions are lead through a radial electrostatic field prior to +E/2 -E/2 magnetic separation. Therefore, only ions with the M + electrical ground same kinetic energy are fed to the magnetic sector. In this way, the electrostatic analyzer acts as a 'source' Figure 5: the electrostatic analyzer. and the combination of the two sectors can be designed to have velocity-focusing properties. After acceleration, the ions possess a kinetic energy given by: E kin zeu = 2 = 1 2 mv (3) The centripetal force is given by: F mv r ion 2 = (2) The electrostatic force is given by: F L = zee (9) For ions leaving the electrostatic analyzer, (2) equals (9) so: r ion 2 mv 2 = = Ekin (10) zee zee This shows that ions are separated by the electrostatic analyzer according their kinetic energy. Substitution of (3) in (10) gives: r ion U = 2 (11) E According to (11), the radius of an ion travelling through the electrostatic analyzer is independent of charge and mass. A narrow slit placed in the image plane of an electrostatic sector can be used as ion source for a magnetic sector instrument. The energy filtering gives better resolution but gives loss of sensitivity due to the rejection of ions. By a proper choice in the combination of magnetic and electrostatic sectors, the velocity dispersion is equal and opposite in both sectors. The Nier- Johnson geometry as shown in fig. 6 is of this type and is one in which both mass and energy -6-
10 focusing occur at a single point. Most sector instruments intended for medium or highperformance work in organic analysis are based on either conventional or reversed Nier-Johnson geometry. B conversion dynode faraday cup E electron multiplier Source slit Ion source Figure 6: Scheme of a double-focusing magnetic sector instrument of Nier-Johnson geometry Summary Double-focussing magnetic sector mass analyzers are the 'classical' instruments against which other mass analyzers are compared. The characteristics are listed below. lassical shaped mass spectra Very high reproducibility Best quantitative performance of all mass spectrometer analyzers High resolution High sensitivity High dynamic range Linked scan MS/MS does not need another analyzer High energy ID (collision induced decay) MS/MS spectra are very reproducible The limitations of sector instruments can be summarized as: Not well-suited for pulsed ionization methods (e.g. MALDI) Usually larger and higher costs both in purchase and maintenance than other mass analyzers Linked scanning MS/MS gives either limited precursor selectivity with unit product-ion resolution or nominal precursor selection with poor product-ion resolution Applications All organic MS analysis methods Accurate mass measurements/peak-matching Quantitation Isotope ratio measurements -7-
11 1.4.4 Quadrupole analyzer. Mass the Organic hemistry Department The quadrupole mass filter consists of four parallel rods of hyperbolic or circular crosssection arranged symmetrically to a z-axis (Fig. 7). A voltage made up of a dc component U and a radio-frequency (r.f.) component V o cos(t) is applied to adjacent rods. Opposite rods are electrically connected. Ions injected into the filter with a very small accelerating voltage, typically V, are made to oscillate in the x and y directions by this field. The parameters a and q are defined by: Field axis ro Z ( U + V cost) + Y ( U + V cost) Figure 7: Schematic diagram of a quadrupole analyser. X 8eU a = (12) 2 m r 2 0 4eV q = (13) m r In these equations 2r 0 is the rod spacing and is the frequency of the r.f. voltage. For certain values of a and q the oscillations performed by the ions are stable, i.e. their amplitudes remain finite, but for other values of a and q these are unstable and the amplitude becomes infinite. The stability diagram, which is also known as Mathieu diagram (figure 8), shows the values of a and q for which these conditions apply. Ions with masses like m1 fall into the stable 0.2 a 0.1 y unstable m 2 m 1 Stable oscillation q a/q constant x unstable Figure 8: Stability diagram for a quadrupole analyser. oscillation area and will migrate towards the detector. Ions with masses like m2 are outside the stable oscillation area and will go lost before they reach the detector and hence mass separation is achieved. Scanning of the mass spectrum is achieved by variation of U and V 0 while maintaining the ratio U/V 0 constant. The registered mass is proportional to V 0 so a linear scan of V 0 gives an easily linear calibrated mass spectrum. Quadrupole instruments are deservedly popular because they are compact, robust, and relatively inexpensive and need little experience to operate. ompared to magnetic sector instruments there are two major advantages: ease of automated data control & handling and ease of interfacing with a variety of inlet systems. Quadrupole analyzers do not have a large mass range and don't have a high resolution as sector instruments do. They are, however, part of more sophisticated instruments as hybrid mass spectrometers and in tandem mass spectrometry. By application of other ionization techniques as electrospray, with multiple charged ions, the lack of mass range can be overcome. Benefits of a quadrupole mass analyzer can be summarized as: lassical mass spectra Good reproducibility -8-
12 Relatively small and low-cost systems Low-energy collision-induced dissociation (ID) MS/MS spectra in triple quadrupole and hybrid mass spectrometers have an efficient conversion of precursor to product Limitations of a quadrupole mass analyzer can be summarized as: Limited resolution Peak heights are variable as a function of the mass (mass discrimination). The peak height versus response must be 'tuned'. Not well suited for pulsed ionisation techniques. Low energy collision-induced dissociation (ID) MS/MS spectra in triple quadrupole and hybrid mass spectrometers depend strongly on energy, collision gas, pressure and other factors. Applications of a quadrupole mass analyzer can be summarized as: Majority of benchtop G/MS and L/MS systems. Triple quadrupole MS/MS systems Sector / quadrupole hybrid MS/MS systems Ion trap. In figure 10, a cross-section of an ion trap is shown. The three dimensional ion trap is a solid revolution of a quadrupole produced by rotation of the cross-section around the z- axis (Fig.7). The ion trap contains three cylindrically symmetrical electrodes: two endcaps, A and B, and a ring. The ring electrode is fed with an r.f. voltage (V) and some times with an additional d.c. voltage (U) relative to the end-cap electrodes. Operating parameters a z and q z, analogue to the quadrupole, can be defined for the ion trap. In this case, r 0 is the internal radius of the ring electrode, about a z m/z = 400 m/z = 300 m/z = 200 R.f. = 1000 V R.f. = 3300 V q z Instability boundary Figure 9: Stability diagram for a ion-trap. one cm, making an ion trap a small-scale device. The use of an r.f. voltage causes rapid reversals of the field direction so the ions are alternately accelerated and decelerated in the axial (z) direction and vice versa in the radial direction. Regions of stable motion, in which ions are trapped in the cell, are described by a Mathieu diagram, as for a quadrupole mass filter (figure 9). Scanning mass analysis with an ion trap is known as the mass-selective instability mode. In this mode the r.f. frequency (about 1.1 MHz) and initial amplitude with a d.c. of zero are chosen so that all ions are stored with an m/z value greater than a threshold value. Thus, first ions are generated by the electron beam and trapped in the trapping field. After a short time, the electron beam is turned off and ions R.f. supply Filament Z 0 r 0 Electron gate A B Electron multiplier Figure 10: Schematic diagram of an ion trap. -9-
13 with an m/z value below the threshold are allowed to escape from the trap. Now the threshold value of the trap is increased by increasing the amplitude of the r.f. voltage and the ions that become unstable will leave the trap through the holes in one of B the end-caps and strike the detector. A build-up of ion density within the trap can lead to space-charge effects, which will A modify the electric fields within the trap. R.f. This can be overcome by the use of an D.c. automatic gain control (AG). Here an initial Figure 11: Mass-selective instability (period B). ionization period is used to calculate the optimal ionization time, with no space-charge effects, for a second more extended ionization period. Following this second ionization period, the r.f. voltage is ramped to affect a mass scan as before. The long storage times used in the ion-trap cause ion-molecule reactions like those in chemical ionization at much lower reagent gas pressure then commonly used in conventional high pressure sources. The significant trapping times also can lead to ion-molecule reactions involving analyte molecules which result in abnormal (M+1)/M ratios under scan conditions. This only can be overcome to a certain degree by lowering the analyte concentration in the trap. Benefits of a quadrupole ion-trap mass analyzer can be summarized as: High sensitivity Multi-stage mass spectrometry (analogues to FTIR) ompact mass analyzer Limitations of a quadrupole ion-trap mass analyzer can be summarized as: Poor Quantitation Very poor dynamic range (sometimes compensated by auto-ranging) Subject to space-charge effects and ion-molecule reactions ollision energy not well defined in ID MS/MS Many parameters comprise the experiment sequence that defines the quality of the mass spectrum (e.g. Excitation, trapping, detection conditions) Applications of a quadrupole ion-trap mass analyzer can be summarized as: Benchtop G/MS, L/MS and MS/MS systems Target compound screening Ion chemistry Time-of-flight analyzer. For an ion accelerated by a voltage V, the resulting velocity is proportional to the mass-tocharge ratio. In a time-of-flight (TOF) mass spectrometer (Fig. 12), ions are separated on basis of the time t needed to travel a path L. m t = L (14) 2 zev For further reading on this topic see: Principles and Instrumentation in Time-of-Flight Mass Spectrometry, Michael Guilhaus, Journal of mass Spectrometry, Vol. 30, (1995) -10-
14 Ions of very high mass-to-charge (several hundreds of kd) may be recorded after an appropriate length of time. L E s E Detector Ionization region Acceleration region Field free Grid 'A' drift tube Figure 12: Schematic diagram of a time-of-flight instrument. The potentially high sensitivity of a TOF instrument results from the high transmission due to the absence of beam defining slits and the temporal separation that in contrast to spatial separation doesn't direct ions away from the detector. In contrast to continuous gas phase ionization processes, like electron impact ionization, TOF mass spectrometers are direct and simply compatible with ion formation from a surface, e.g. laser desorption (like in MALDI) and plasma desorption mass spectrometry. The pulsed character of these techniques provides a precisely defined ionization time and a small precisely defined ionization region, which is ideal for TOF-analysis. Resolution in TOF-analysis may be improved by narrowing down the time-window and the ionization-region. Resolution also may be improved by application of a reflectron. The ions leaving the source of a time-of-flight mass spectrometer have neither exactly the same starting times nor exactly the same kinetic energies. Various time-of-flight mass spectrometer designs have been developed to compensate for these differences. A reflectron is an ion optic device in which ions Target Acceleration and ion focus Pulsed laser Deflector Detector Reflectron Figure 13: schematic diagram of a reflectron. pass through a mirror or reflectron and their flight is reversed. A linear-field reflectron allows ions with greater kinetic energies to penetrate deeper into the reflectron than ions with smaller kinetic energies. The deeper penetrating ions will travel a longer time to the detector opposite to the ions with a smaller energy. A reflectron thus decreases the spread in ion flight times for a given packet of ions and therefore improves the resolution of the TOF mass spectrometer. Benefits of a time-of-flight (TOF) mass analyzer can be summarized as: Fastest available MS analyzer Well suited for pulsed ionization methods (majority of MALDI mass spectrometers is equipped with TOF) High ion transfer MS/MS information from post-source decay Highest practical mass range of all MS analyzers. -11-
15 Limitations of a time-of-flight (TOF) mass analyzer can be summarized as: Requires pulsed ionization methods or beam switching Fast digitizers used in TOF may have limited dynamic range Limited precursor-ion selectivity for most MS/MS experiments Applications of a time-of-flight (TOF) mass analyzer can be summarized as: Almost all MALDI (matrix assisted laser ionization) systems Very fast G/MS systems 1.5 Resolution. The resolution of a mass spectrometer can be defined by its ability to separate ions of adjacent mass number. Several definitions of resolution are used in mass spectrometry. It is useful to understand the distinctions between the different definitions in order to understand the characteristics of the different mass spectrometers. Figure 14: Flat topped peaks e.g. quadrupole (left, ~unit resolution at 500) and triangular/ Gaussian peaks as from a magnetic sector mass spectrometer (right, ~1000 resolution). Unit resolution means that you can separate each mass from the next integer mass. That is, one can tell the difference from masses 50 and 51 as from 1000 and This definition commonly is used for quadrupole and ion trap mass spectrometers, where the peaks usually are "flattopped". In magnetic sector mass spectrometry resolution usually is defined as: M R = (15) M That is, the difference between two masses that can be separated divided by the mass number of the observed mass. In magnetic sector instruments, peaks usually are triangular or Gaussian (see Fig.14). Peaks in magnetic sector mass spectrometers usually are called separated to a 10% valley when the overlap point is at 1/10 of the height of the higher of the two peaks. If only one peak is available, resolution is determined by the quotient of the observed mass divided by its width at the 5% level. The resolution is constant across the mass range. Fourier Transform Ion yclotron Resonance (FTIR) mass spectrometers use the same definition of resolution as magnetic sector mass spectrometers. However, the 50% valley definition is often used because of the broadening of the peaks near the baseline due to -12-
16 apodisation and Lorentzian peak shape. In addition, resolution in FTIR is inversely proportional to mass, so it is important to know at what mass the given resolution was obtained. In time-of-flight mass spectrometry, the 50 % valley definition is used. Peak shapes in TOF are Gaussian. Figure 15: Two peaks resolved to 10% valley (left) and 50% valley (right). ompare the unit resolution, as defined in quadrupole, and the resolution as defined in sector mass spectrometry. At 5000 resolution, in a magnetic sector instrument, one can distinguish between m/z and m/z or between m/z and m/z All separated to 10% valley. Unit resolution allows you to distinguish m/z 50 from m/z 51 or m/z 100 from m/z Some remarks on elemental composition calculations. The elemental composition of a molecular ion or fragment ion can be determined by accurately measuring its m/z value. Normally this is achieved by acquiring a mass spectrum at high resolution in order to measure the mass of a single species and not (partially) resolved peaks. Normally a resolution of (10% valley) is considered desirable for accurate mass spectrometry. Elemental composition calculations are used to determine all the possible elemental compositions of a given mass with specified tolerance. The calculation is usually (not always) done for m/z values obtained from high-resolution mass spectra. The total number of possible elemental compositions increases with mass and tolerance. Even for modest masses, the number can become very large. For m/z 146 ±.5 amu, containing all of, H, N and O, this gives 33 possible compositions. For the same, with a tolerance of amu, only containing and H with possible N and O the number of possible structures is two. Doubling the mass gives 286 and 9 respectively, which shows the need of other constraints. Limiting the number of possible elemental compositions can be achieved by: The accuracy of the measured mass. The set of elements and/or isotopes that are allowed in the elemental composition. The error limit for the calculated masses for the resulting elemental compositions. Upper and lower bounds on each allowed element and/or isotope. Methods of calculating that allow for odd-electron ions, even-electron ions or both Error limits. Error limits can be used to limit the number of possible elemental compositions and are defined in terms of millimass units (mmu), unified atomic mass units (u), or parts-per-million (ppm). The parts-per-million error is defined as: measured mass theoretical mass = 10 theoretical mass 6 ppm (16) -13-
17 To overcome problems that arise when the theoretical mass becomes very small or large some additional limiting is necessary. The use of a low error bound and a high error bound makes it possible to define the limiting mass. mmu10 10 ppm 3 6 Limiting mass = (17) Where mmu refers to the low or high error bound and ppm refers to the error limit as defined above. For example, the error limit is defined as 10 ppm; the low error bound as five mmu and the high error bound as 20 mmu. This gives the following error limits. 5 mmu for m/z < 500u 10 ppm for 500u S m/z < 2000u 20 mmu for m/z T 2000u When the error limit is expressed in u or mmu then no additional parameter is needed to define error limits in elemental composition calculations Double bond equivalent (unsaturation) Using the following formula, the total number of rings and double bonds (sites of unsaturation) is calculated. i = + max N i V i i ( 2) D (18) Where D is the unsaturation, i max is the total number of different elements in the composition, N i the number of atoms of element i and V i the valence of atom i. Each ring and each double bond counts as one site of unsaturation and each triple bond counts as two sites of unsaturation. Due to the division by two, the result can be either an exact integer or an integer with a remainder of a half. This indicates whether the composition is an odd or an even electron system (see following). An exact integer indicates an odd electron ion where a remainder of 0.5 indicates an even electron ion. The minimum value for D in organic chemistry is -0.5 which corresponds with a protonated saturated compound like H 3 O Odd-electron and even-electron ions. A neutral, non-ionised compound has an even number of electrons. Ions can have either an even or an odd number of electrons. alculating the unsaturation D (Eq.18) for an elemental composition gives the electronic configuration. This can be used to identify molecular ions, fragment ions and ions resulting from rearrangement reactions. EI (Electron Impact) produces odd electron ions because one electron is lost during ionization. Fragmentation in EI usually occurs through the loss of a radical so that most fragment ions will have an even-electron configuration. Odd electron fragments usually result from rearrangement reactions and therefore it is useful to identify important odd-electron ions in a mass spectrum. Soft ionization methods like FAB, I, FD or Electrospray often produce species like [M+H] + or [M+Na] + that have an even-electron configuration. These considerations can help to limit the number of possible elemental compositions assigned to an ion. Looking for molecular ions in EI -14-
18 limits the search to odd-electron compositions whereas in FAB ([M+H]+) the search is limited to even-electron ions The nitrogen rule. The nitrogen rule says that an organic compound with an even mass has an even number of nitrogen's. If the compound has an odd mass then it contains an odd number of nitrogen's. This can be explained by the fact that every element with an odd mass has an odd valence and an element with an even mass has an even valence. Nitrogen is the exception where it has an odd valence and an even mass. Therefore, when you are looking for a molecular ion with an odd mass in EI it should contain an odd number of nitrogen's. The same applies for e.g. FAB; an even mass for a protonated molecular ion accounts for an odd number of nitrogen's Isotope ratio measurements. Most of the elements present in organic compounds do have two or more stable isotopes. These isotopes generally differ one or two mass units. Apart from Si, S, l and Br the natural abundance of 13 is relatively high, especially while it is the most abundant element, besides hydrogen, in organic compounds. If a compound with mass M contains x atoms then the intensity of the first isotopic peak (M+1) should be about 1.1x% of the intensity of M. Information about isotope ratios can be used to limit the possible elemental compositions that correspond to a given measured mass. For example, a composition such as 6 H 5 l cannot be a reasonable composition for an m/z 112 ion unless there are chlorine isotope peaks at m/z 112 and m/z 114 with a relative ratio of about 3:1. Of course, the presence of interference peaks may confuse the isotopic pattern. Interference at m/z 114 might cause the relative abundance of the m/z 114 peak to be larger than the expected 3:1 ratio. Yet the m/z 112 peak could still correspond to 6 H 5 l. Therefore caution and common sense should be used in making use of isotopic information to make elemental composition assignment. The elemental composition calculation can be paired with theoretical isotope ratio calculations that allow the user to see what the expected isotopic pattern would be for any given elemental composition. This is very useful for the interpretation of the mass spectra of inorganic or organometallic compounds. Many of these compounds have distinctive isotope patterns that allow the chemist to determine how many metal atoms (for example) are present in a particular ionic species Examples Methyl stearate. Suppose that you measure the mass for a molecular ion and find it to be u, and that you expect to find carbon, hydrogen, nitrogen, and oxygen to be possible elements for the composition. If the limits for the calculation are set as displayed below, only one composition is possible for that measured mass: 19 H 38 O 2. Element Limits: 5/50 H 10/100 N 0/2 O 0/4 Tolerance: PPM Low error bound (mmu): 5.0 (for masses < 500) High error bound (mmu): 20.0 (for masses < 2000) Even or odd electron ion or both: BOTH Minimum unsaturation: -0.5 Maximum unsaturation: 10.0 Meas. mass Abund. Diff. Unsat. ompositions u ppm H38 N0 O2-15-
19 hlorpyrifos. If there are many possible heteroatoms, the number of possible elemental compositions may be very large and it may be difficult to determine which is correct without additional information. An example is the insecticide chlorpyrifos which has the composition 9 H 11 l 3 NO 3 PS. If we assume a large range for the elemental limits, then the number of compositions is very large: Element Limits: 5/20 H 5/42 N 0/5 O 0/10 l 1/4 P 0/5 S 0/5 Tolerance: 5.00 PPM Low error bound (mmu): 5.0 (for masses < 1000) High error bound (mmu): 20.0 (for masses < 4000) Even or odd electron ion or both: ODD Minimum unsaturation: -0.5 Maximum unsaturation: 10.0 Rel. abundance cutoff (percent): Meas. mass Abund. Diff. Unsat. ompositions u ppm H 5 N 5 O 7 l 2 P 0 S H 6 N 3 O 9 l 1 P 2 S H 6 N 5 O 5 l 2 P 1 S H 7 N 5 O 5 l 2 P 2 S H 8 N 5 O 3 l 2 P 3 S H 9 N 5 O 3 l 1 P 1 S H 10 N 5 O 1 l 1 P 2 S H 11 N 1 O 8 l 3 P 1 S H 11 N 3 O 4 l 4 P 0 S H 12 N 1 O 6 l 3 P 2 S H 13 N 3 O 2 l 4 P 2 S H 6 N 3 O 8 l 2 P 1 S H 6 N 5 O 4 l 3 P 0 S H 8 N 3 O 6 l 1 P 0 S H 8 N 5 O 2 l 3 P 2 S H 9 N 5 O 2 l 2 P 0 S H 10 N 3 O 4 l 1 P 2 S H 10 N 5 O 0 l 2 P 1 S H 11 N 1 O 7 l 4 P 0 S H 12 N 1 O 5 l 4 P 1 S H 12 N 3 O 2 l 1 P 4 S H 13 N 3 O 0 l 1 P 5 S H 6 N 3 O 7 l 3 P 0 S H 8 N 5 O 1 l 4 P 1 S H 9 N 1 O 7 l 1 P 1 S H 10 N 3 O 3 l 2 P 1 S H 11 N 1 O 5 l 1 P 3 S H 12 N 3 O 1 l 1 P 0 S H 12 N 3 O 1 l 2 P 3 S H 13 N 1 O 3 l 1 P 5 S H 5 N 5 O 3 l 1 P 1 S H 7 N 5 O 1 l 1 P 3 S H 9 N 1 O 6 l 2 P 0 S H 10 N 3 O 2 l 3 P 0 S H 11 N 1 O 4 l 2 P 2 S H 12 N 3 O 0 l 3 P 2 S H 13 N 1 O 2 l 1 P 1 S H 13 N 1 O 2 l 2 P 4 S H 15 N 1 O 0 l 1 P 3 S H 5 N 5 O 2 l 2 P 0 S H 6 N 3 O 4 l 1 P 2 S H 7 N 5 O 0 l 2 P 2 S H 8 N 3 O 2 l 1 P 4 S 0 ---ORRET----> H 11 N 1 O 3 l 3 P 1 S H 13 N 1 O 1 l 2 P 0 S H 13 N 1 O 1 l 3 P 3 S 0-16-
20 H 5 N 1 O 7 l 1 P 1 S H 6 N 3 O 3 l 2 P 1 S H 7 N 1 O 5 l 1 P 3 S H 8 N 3 O 1 l 1 P 0 S H 8 N 3 O 1 l 2 P 3 S H 11 N 1 O 2 l 4 P 0 S H 13 N 1 O 0 l 4 P 2 S H 5 N 1 O 6 l 2 P 0 S H 6 N 3 O 2 l 3 P 0 S H 7 N 1 O 4 l 2 P 2 S H 8 N 3 O 0 l 3 P 2 S H 9 N 1 O 2 l 1 P 1 S H 11 N 1 O 0 l 1 P 3 S H 7 N 1 O 3 l 3 P 1 S H 9 N 1 O 1 l 2 P 0 S H 7 N 1 O 2 l 4 P 0 S 0 If we take a look at the measured isotope ratios, we can put some more reasonable limits on the elements to be included. Suppose that a low resolution mass spectrum shows the following (ignoring any interferences or overlapping losses that might be found in a real mass spectrum): m/z Rel. abundance % From this, we might conclude that there are between 8 and 11 carbons (because the 13 peak should be 1.1% times the number of carbons) and three chlorines (because of the 349/351/353 isotope peaks). The maximum number of hydrogens is estimated as twice the number of carbons plus 2. These assumptions drastically reduce the number of possible compositions. Element Limits: 8/11 H 5/24 N 0/5 O 0/10 l 3/3 P 0/5 S 0/5 Tolerance: 5.00 PPM Low error bound (mmu): 5.0 (for masses < 1000) High error bound (mmu): 20.0 (for masses < 4000) Even or odd electron ion or both: ODD Minimum unsaturation: -0.5 Maximum unsaturation: 10.0 Rel. abundance cutoff (percent): Meas. mass Abund. Diff. Unsat. ompositions u ppm H 10 N 3 O 2 l 3 P 0 S H 12 N 3 O 0 l 3 P 2 S H 11 N 1 O 3 l 3 P 1 S 1 < H 13 N 1 O 1 l 3 P 3 S H 6 N 3 O 2 l 3 P 0 S H 8 N 3 O 0 l 3 P 2 S 0 It is still not easy to tell the correct composition. If we would have better mass accuracy, say, a 5 ppm error tolerance, then there are three compositions that are possible: Meas. mass Abund. Diff. Unsat. ompositions u ppm H 11 N 1 O 3 l 3 P 1 S 1 < H 6 N 3 O 2 l 3 P 0 S H 8 N 3 O 0 l 3 P 2 S 0-17-
21 The final decision about the correct composition will depend on whether we know something about the unsaturation (4 or 9 rings and/or sites of unsaturation) and whether we know anything about the number of constituing elements. Here classical elemental analysis may be helpful. With this technique, the amount of carbon, hydrogen, nitrogen and/or sulfur is determined in a sample. Through this, the purity and the elemental ratios are determined. Even it is possible, through calculus, to predict the composition of the remainder if the total of percentages does not add up to 100. The result should help in the selection of the right composition. -18-
22 2 Ionization Methods in Organic Mass Spectrometry. A mass spectrometer works by using magnetic and electric fields to exert forces on charged particles (ions) in a vacuum. Therefore, a compound must be charged or ionized to be analyzed by a mass spectrometer. Furthermore, the ions must be introduced in the gas phase into the vacuum system of the mass spectrometer. This is easily done for gaseous or heat-volatile samples. However, many (thermally labile) analytes decompose upon heating. These kinds of samples require either desorption or desolvation methods if they are to be analyzed by mass spectrometry. Although ionization and desorption/desolvation are usually separate processes, the term "ionization method" is commonly used to refer to both ionization and desorption (or desolvation) methods. The choice of ionization method depends on the nature of the sample and the type of information required from the analysis. So-called 'soft ionization' methods such as field desorption and electrospray ionization tend to produce mass spectra with little or no fragmention content. 2.1 Gas-Phase ionization. These methods rely upon ionizing gas-phase samples. The samples are usually introduced through a heated batch inlet, heated direct insertion probe, or a gas chromatograph Some general remarks on ionisation When an electron with an kinetic energy, E kin, of about 50 ev passes through a thin gasmixture, as in the ion-source, an actual collision with a neutral molecule is not likely. The repelling force of the valence electrons can only be overcome when the E kin is in the order of magnitude of 10 6 Volts. Therefore an alectron will pass in short distance of a molecule. A 50 ev electron travels with a speed of 4.2 * 10 8 cm/sec and passes a molecule, a few nanometers wide, in about seconds. Due to the passing electric field the orbit of the valence electrons gets disturbed to such an extent that it may lose one valence electron and ionisation occurs. + M + e M + 2e Energy M M + B A Internuclear distance Figure 16: Ionisation proces in a di-atomic molecule. The capture of an electron by neutral molecules is unlikely to happen. In practice it appears that in about to 10 3 to 10 4 of the formed positive ions only one negative ion is formed. -19-
23 The ionisation time of second and the fastest vibrations in organic molecules, -H stretch, indicate that the position of the First excited-state of M + nuclei may be considered unchanged during the process of ionisation. The ionised level is reached through a so Electronic ground-state of M + called Franck-ondon transition, the hatched area, arrow A, in figure 16. As a result a molecular-ion is formed with several possible M + vibrational states. Some of them in such a way, arrow figure 16, that fragnmentation occurs, because of exceeding the dissociationenergy. Also through a Franck-ondon it is possible that molecular-ions in a higher Internuclear distance electronic state are formed. These will dissociate even easier, because of a decrease Figure 17: Potential energy curves in a di-atomic of the depth in the potential energy-curve and molecule. because the minima tend to drift to higher internuclear distance (figure 17). In poly-atomic molecules the change of the potential energy curves with changing internuclear distance is much more complicated. More oscillators are present and the curves will intersect at several points. B E If we take acetone as an example and focus on the carbonyl we can distinguish three types of E B electrons namely n-, Y- and Z-electrons. Their mutual levels are depicted in figure 19. In this figure it is shown that the probability of Internuclear distance electron release during ionisation decreases from n > Y > Z. Figure 18: Intersecting potential energy curves. The ground-state of the molecular ion of acetone will be the one where a n-electron is eliminated. In the same way the first and second excited level will resemble a missing a Y- or a Z-elctron. The major part of the molecular ions of n acetone will be missing a n-lectron and subsequent formation of fragment ions has to be explained from there. Even if the primary ionisation was due to elimination of a Y or Z- Figure 19: Electronic energy levels in a carbonyl electron this is valid. These ions can t group (schematic). eliminate an excess of internal energy E (= E el + E vib + E rot ) through collisions with other molecules because of their formation in high vacuum. In the case of acetone and especially in larger molecules it is very probable that the energy curves of the first and second excited state of the molecular ion intersect. At this intersection the molecular ion can transfer, without energy loss. This radiation free transfer produces a highly excited molecular ion in the groundstate which makes it even more probable to dissociate. (See figure 18). Energy Energy -20-
24 In practice mass spectra are recorded at 70 ev electron energy. This is because the amount of formed ions increases with increasing electron energy upto about 50 ev. Above 50 ev the amount of formed ions doesn t change hardly and this is a condition to produce reproducable mass spectra (figure 20). An energy of 70 ev corresponds to 1614 kcal/mole (1 ev = 23,06 kcal/mole). This is an enormous amount when compared to the average bound energy in organic molecules (table 1). In addition the binding energy in ions will be much smaller. When you realise that electrons skim the molecule it is clear that not all 70 ev will be transferred. Relative intensity M + ( 6 H 5 N) + (M-1) + Electron energy M ++ m/e 78 (x10) m/e 79 m/e 39.5 (x10) Figure 20: Relationship between ion-intensity and energy of the ionising electrons in the mass spectrum of pyridine. Table 1: Energy content for some common bonds in organic chemistry (kcal/mol per bond) Bond Bond Single Double Triple Single Double Triple -H I O-H N O-Si O a N-H S b N-N F N-O l 81.0 S-H Br 68.0 S-S 54.0 a) ketones b) S 2 Naturally the ionisation energy is transferred, otherwise no ionisation would occur. For organic molecules this is about 7 to 10 ev (table 2, 3). Experimentally it is determined that besides the ionistaion energy not more than 3 to 5 ev is transferred to the molecule. This amount is the internal energy E (= E el + E vib + E rot ) of the molecular ion (~ kcal/mole). In ground state of the molecular ion of propane this means about 7 to 12 ev extra per bond ( 3 H 8 10 bonds). When accumulated in one bond this is enough to cause dissociation. Table 2: Ionisationpotential (ev) of some monofunctionalised aliphatic compounds H 3 H 2 H 2 X X = Potential X = Potential H OH ONO HO 9.86 l H=H H OH OAc I H NH Br
25 Table 3: As table 2 for aromatic compounds 6 H 5 X X = a b c X = a b c NO H N I HO OH OH SH H OH F NH l NHH Br N(H 3 ) a = Photo-ionisation, b = Electron Impact, c = Photo-electron spectroscopy Electron Ionization (EI) In the preceding text it is assumed that ionization takes place via EI (Electron Impact also Electron Ionization). EI is the oldest and best characterized of all the ionization methods. EI probably is the most common used ionization technique and still is of major importance. The principles are already pointed out in section 1.3. The ionization process can either produce a molecular ion which will have the same molecular weight and elemental composition of the starting analyte, or it can produce a fragment ion which corresponds to a smaller piece of the analyte molecule. After ionization, the remaining energy is about 2-7 ev per molecule, leading to a lot of fragmentation and giving much information about the original structure. The low intensity or even the total absence of the molecular ion and the complexity of the resulting spectrum are against EI. The ionization potential is the electron energy that will produce a molecular ion. The appearance potential for a given fragment ion is the electron energy that will produce that fragment ion. Most mass spectrometers use electrons with an energy of 70 electron volts (ev) for EI. Decreasing the electron energy can reduce fragmentation, but it also reduces the number of ions formed. Sample introduction heated batch inlet heated direct insertion probe gas chromatograph liquid chromatograph (particle-beam interface) Benefits well-understood can be applied to virtually all volatile compounds reproducible mass spectra fragmentation provides structural information libraries of mass spectra can be searched for EI mass spectral "fingerprint" Limitations sample must be thermally volatile and stable the molecular ion may be weak or absent for many compounds. Mass range Low Typically less than 1,000 Da. -22-
26 2.1.3 hemical Ionization (I) Mass the Organic hemistry Department hemical ionization uses ion-molecule reactions to produce ions from the analyte. The chemical ionization process begins when a reagent gas such as methane, isobutane, or ammonia is ionized by electron impact. A high reagent gas pressure (or long reaction time) results in ion-molecule reactions between the reagent gas ions and reagent gas neutrals. Some of the products of these ion-molecule reactions can react with the analyte molecules to produce analyte ions. Example (R = reagent, S = sample, e = electron, = radical electron, H = hydrogen): R + e ---> R + + 2e R + + RH ---> RH + + R RH + + S ---> SH + + R (of course, other reactions can occur) Sample introduction heated batch inlet heated direct insertion probe gas chromatograph liquid chromatograph (particle-beam interface) Benefits often gives molecular weight information through molecular-like ions such as [M+H] +, even when EI would not produce a molecular ion. simple mass spectra, fragmentation reduced compared to EI Limitations sample must be thermally volatile and stable less fragmentation than EI, fragment pattern not informative or reproducible enough for library search results depend on reagent gas type, reagent gas pressure or reaction time, and nature of sample. Mass range Low Typically less than 1,000 Da Desorption hemical Ionization (DI) In this variation on chemical ionization, the analyte is placed on a filament that is rapidly heated in the I plasma. The direct exposure to the I reagent ions, combined with the rapid heating acts to reduce fragmentation. Some samples that cannot be thermally desorbed without decomposition can be characterized by the fragments produced by pyrolysis DI. Sample introduction sample deposited onto a filament wire filament rapidly heated inside the I source. Benefits reduced thermal decomposition rapid analysis relatively simple equipment Limitations not particularly reproducible -23-
27 rapid heating requires fast scan speeds fails for large or labile compounds Mass the Organic hemistry Department Mass range Low typically less than 1,500 D Negative-ion chemical ionization (NI) Not all compounds will produce negative ions. However, many important compounds of environmental or biological interest can produce negative ions under the right conditions. For such compounds, negative-ion mass spectrometry is more efficient, sensitive and selective than positive-ion mass spectrometry. Negative ions can be produced by a number of processes. Resonance electron capture refers to the capture of an electron by a neutral molecule to produce a molecular anion. The electron energy is very low, and the specific energy required for electron capture depends on the molecular structure of the analyte. Electron attachment is an endothermic process, so the resulting molecular anion will have excess energy. Some molecular anions can accommodate the excess energy. Others may lose the electron or fall apart to produce fragment anions. A buffer gas (usually a common I gas such as methane) is used to slow down the electrons in the electron beam until some of the electrons have just the right energy to be captured by the analyte molecules. The buffer gas can also help stabilize the energetic anions and reduce fragmentation. This is really a physical process and not a true "chemical ionization" process. Sample introduction Same as for I Benefits efficient ionization, high sensitivity less fragmentation than positive-ion EI or I greater selectivity for certain environmentally or biologically important compounds Limitations not all volatile compounds produce negative ions poor reproducibility Mass range Low Typically less than 1,000 D. 2.2 Field Desorption and Ionization These methods use very strong electric fields to extract ions from a substrate in vacuum. They provided one of the earliest means by which molecular ions could be generated from non-volatile and/or thermally unstable materials. Field ionization (FI) and field desorption (FD) can be distinguished by the way samples are supplied to the emitter. In FI, samples are introduced in the gas-phase from another inlet system. In FD the samples is coated on the emitter prior to analysis. The intense electric fields used in both FI and FD are produced with wire-based emitters. These emitters have multipoint micro-needles (dendrites, Ø ~ 3Å) on their surface. The field strength at the tip of a micro-needle is given by: -24-
28 F = V 1 2 (1 ) 4d r ln( ) r + r (19) Where V is the applied voltage, r is the radius of the micro-needle, d is the distance to the counter electrode and is a shape factor (~0,0064). The energy transferred in FI and FD is only a fraction of an electronvolt, EI (~7, <10) and I (<5), so fragmentation is not or hardly introduced Field Desorption (FD) The sample is deposited onto the emitter that is biased to a high potential (several kilovolts). A current is passed through the emitter to heat up the filament. Mass spectra are acquired as the emitter current is gradually increased and the sample is evaporated from the emitter into the gas phase. The analyte molecules are ionized by electron tunneling at the tip of the emitter 'whiskers'. haracteristic positive ions produced are radical molecular ions and cation-attached species such as [M+Na]+ and [M-Na]+. The latter are probably produced during desorption by the attachment of trace alkali metal ions present in the analyte. Sample introduction Direct insertion probe. The sample is deposited onto the tip of the emitter by dipping the emitter into an analyte solution depositing the dissolved or suspended sample onto the emitter with a micro-syringe Benefits simple mass spectra, typically one molecular or molecular-like ionic species per compound. little or no chemical background works well for small organic molecules, many organometallics, low molecular weight polymers and some petrochemical fractions Limitations sensitive to alkali metal contamination and sample overloading emitter is relatively fragile relatively slow analysis as the emitter current is increased the sample must be thermally volatile to some extent to be desorbed Mass range Low moderate, depends on the sample. Typically less than about 2,000 to 3,000 D. some examples have been recorded from ions with masses beyond 10,000 D Field Ionization (FI) The sample is evaporated from a direct insertion probe, gas chromatograph, or gas inlet. As the gas molecules pass near the emitter, they are ionized by electron tunneling. Sample introduction heated direct insertion probe gas inlet gas chromatograph -25-
29 Benefits simple mass spectra, typically one molecular or molecular-like ionic species per compound. little or no chemical background works well for small organic molecules and some petrochemical fractions Limitations The sample must be thermally volatile. Samples are introduced in the same way as for electron ionization (EI). Mass range Low Typically less than 1000 D. 2.3 Particle Bombardment In these methods, the sample is deposited on a target that is bombarded with atoms, neutrals, or ions. The most common approach for organic mass spectrometry is to dissolve the analyte in a liquid matrix with low volatility and to use a relatively high current of bombarding particles (FAB or dynamic SIMS). Other methods use a relatively low current of bombarding particles and no liquid matrix (static SIMS). The latter methods are more commonly used for surface analysis than for organic mass spectrometry. The primary particle beam is the bombarding particle beam, while the secondary ions are the ions produced from bombardment of the target Fast Atom Bombardment (FAB) The analyte is dissolved in a liquid matrix such as glycerol, thio-glycerol, m-nitrobenzyl alcohol, or diethanolamine and a small amount (about 1 microlitre) is placed on a target. The target is bombarded with a fast atom beam (for example, 6-8 kev Xenon atoms), that desorbs molecularlike ions and fragments from the analyte. luster-ions from the liquid matrix are desorbed also, and produce a chemical background. Sample introduction direct insertion probe L/MS (frit FAB or continuous-flow FAB). Benefits rapid simple relatively tolerant of variations in sampling good for a large variety of compounds strong ion currents -- good for high-resolution measurements Limitations high chemical background defines detection limits may be difficult to distinguish low-molecular-weight compounds from chemical background analyte must be soluble in the liquid matrix no good for multiply charged compounds with more than 2 charges Mass range Moderate: Typically ~300 D to about 6000 D. -26-
30 2.3.2 Secondary Ion Mass Spectrometry (SIMS) This discussion refers to dynamic SIMS. Dynamic SIMS is nearly identical to FAB except that the primary particle beam is an ion beam (usually cesium ions) rather than a neutral beam. The ions can be focused and accelerated to higher kinetic energies than are possible for neutral beams, and sensitivity is improved for higher masses. The use of SIMS for moderate-size ( ,000 D) proteins and peptides has largely been supplanted by electrospray ionization. Sample introduction Same as for FAB Benefits Same as for FAB, except sensitivity is improved for higher masses (3000 to 13,000 Da). Limitations Same as for FAB except target can get hotter than in FAB due to more energetic primary beam high-voltage arcs more common than FAB ion source usually requires more maintenance than FAB Mass range Moderate: Typically 300 to 13,000 D. 2.4 Atmospheric Pressure Ionization (Spray Methods) In these methods, a solution containing the analyte is sprayed, at atmospheric pressure, through an interface into the vacuum of the mass spectrometer ion source. A combination of thermal and pneumatic means is used to desolvate the ions as they enter the ion source. Solution flow rates can range from less than a microliter per minute to several milliliters per minute. These methods are well-suited for flow-injection and L/MS techniques Electrospray Ionization (ESI) The sample solution is sprayed across a high potential difference (a few kilovolts) from a needle into an orifice in the interface. Heat and gas flows are used to desolvate the ions existing in the sample solution. Electrospray ionization can produce multiply charged ions with the number of charges tending to increase as the molecular weight increases. The number of charges on a given ionic species must be determined by methods such as: comparing two charge states that differ by one charge and solving simultaneous equations looking for species that have the same charge but different adduct masses examining the mass-to-charge ratios for resolved isotopic clusters Sample introduction flow injection L/MS typical flow rates are less than 1 microliter per minute up to about a milliliter per minute -27-
31 Benefits good for charged, polar or basic compounds permits the detection of high-mass compounds at mass-to-charge ratios that are easily determined by most mass spectrometers (m/z typically less than 2000 to 3000). best method for analyzing multiply charged compounds very low chemical background leads to excellent detection limits can control presence or absence of fragmentation by controlling the interface lens potentials compatible with MS/MS methods Limitations multiply charged species require interpretation and mathematical transformation (can sometimes be difficult) complementary to API. No good for uncharged, non-basic, low-polarity compounds (e.g. steroids) very sensitive to contaminants such as alkali metals or basic compounds relatively low ion currents relatively complex hardware compared to other ion sources Mass range Low-high Typically less than 200,000 D Atmospheric Pressure hemical Ionization (API) Similar interface to that used for ESI. In API, a corona discharge is used to ionize the analyte in the atmospheric pressure region. The gas-phase ionization in API is more effective than ESI for analyzing less-polar species. ESI and API are complementary methods. Sample introduction same as for electrospray ionization Benefits good for less-polar compounds excellent L/MS interface compatible with MS/MS methods Limitations complementary to ESI. Mass range Low-moderate Typically less than 2000 D. 2.5 Laser Desorption Laser desorption methods use a pulsed laser to desorb species from a target surface. Therefore, one must use a mass analyzer such as time-of-flight (TOF) or Fourier transform ion cyclotron resonance (FTIR) that is compatible with pulsed ionization methods. Magnetic sector mass spectrometers equipped with an array detector can also be used for the detection of ions produced by MALDI. Direct laser desorption relies on the very rapid heating of the sample or sample substrate to vaporize molecules so quickly that they do not have time to decompose. This is good for low to -28-
32 medium-molecular weight compounds and surface analysis. The more recent development of matrix-assisted laser desorption ionization (MALDI) relies on the absorption of laser energy by a matrix compound. MALDI has become extremely popular as a method for the rapid determination of high-molecular-weight compounds Matrix-Assisted Laser Desorption Ionization (MALDI) The analyte is dissolved in a solution containing an excess of a matrix such as sinapinic acid or dihydroxybenzoic acid that has a chromophore that absorbs at the laser wavelength. A small amount of this solution is placed on the laser target. The matrix absorbs the energy from the laser pulse and produces a plasma that results in vaporization and ionization of the analyte. Sample introduction direct insertion probe continuous-flow introduction Benefits rapid and convenient molecular weight determination Limitations MS/MS difficult requires a mass analyzer that is compatible with pulsed ionization techniques not easily compatible with L/MS Mass range Very high Typically less than 500,000 Da. 2.6 Some commonly used chemicals in mass spectrometry I Reagent Gases Methane: good for most organic compounds usually produces [M+H] +, [M+H 3 ] + depending on partial pressure [M+ 2 H 5 ] + adducts adducts are not always abundant extensive fragmentation Isobutane: usually produces [M+H] +, [M+ 4 H 9 ] + adducts and some fragmentation adducts are relatively more abundant than for methane I not as universal as methane Ammonia: fragmentation virtually absent polar compounds produce [M+NH 4 ] + adducts basic compounds produce [M+H] + adducts non-polar and non-basic compounds are not ionized -29-
33 2.6.2 FAB matrices Glycerol 3 H 8 O 3, m/z first and most widely-used FAB matrix. best choice for polar compounds Figure 21: Positive ion-mode FAB spectrum of Glycerol thioglycerol 3 H 8 O 2 S, m/z similar uses to glycerol, but may enhance M+1 abundance more volatile than glycerol, evaporates quickly Figure 22: Positive ion-mode FAB spectrum of thioglycerol. -30-
34 nitrobenzyl alcohol (NBA) 7 H 7 NO 3, m/z best choice for less polar compounds ( e.g. chlorophyll) and many organometallics can produce radical molecular ions as well as protonated or cationized molecular ions Figure 23: Positive ion-mode FAB spectrum of NBA nitrophenyl octyl ether (NPOE) 14 H 21 NO 3, m/z only FAB matrix with no reactive hydrogen good choice for reactive compounds such as organometallics Figure 24: Positive ion-mode FAB spectrum of nitro-phenyl-octyl-ether -31-
35 triethanolamine 6 H 15 NO 3, m/z good matrix for negative-ion FAB enhances [M-H] - formation. 100 HO N OH OH Figure 25: EI mass spectrum of triethanolamine dioctyl phthalate (DOP) bis(2-ethylhexyl) phthalate 24 H 38 O 4, m/z Similar in use to NPOE Widespread contaminant in solvents, gives characteristic 149 peak in EI mass spectra Use care to avoid contaminating the mass spectrometer O O O OH Figure 26: EI mass spectrum of di-octyl phtalate. -32-
36 3 Location of charge and primary dissociation in molecular ions. 3.1 Location of charge. The idea of charge location, i.e. that after ionization the charge will be located on the element with the smallest ionization potential, is very helpful in understanding why molecular ions dissociate via a very characteristic pattern. Besides this idea of charge-location, the structure of the molecule plays an essential role in dissociation. Here we have to emphasize again that the molecular ion is a radical cation (un-paired electron system). One of the most important underlying motives in dissociation is the formation of the most stable cation (with a paired electron system). The corresponding peak in the mass spectrum will have a very high intensity compared to the rest of the spectrum. The two ways molecular ions have to form cations are homolytic- and heterolytic dissociation. 3.2 Homolytic dissociation. Homolytic dissociation means the cleavage of a bond X-Y, where one of the binding electrons is paired with the free electron on element Z adjacent to either X or Z. A so called one-electronshift. 100 O H 3 O 50 H 3 58 H Figure 27: Mass spectrum of acetone An example of homolytic dissociation is the dissociation of the molecular ion of acetone. Due to charge location this molecular ion will miss a n-electron and through homolytic dissociation it will lose a methyl-radical as confirmed by the mass spectrum. In mass spectrometry homolytic dissociation is called _-cleavage. H 3 O H 3 -H 3 H 3 O H 3 O Figure 28: Loss of a methyl-radical by acetone. R 1 H 2 X H 2 R 2 -R 2 R 1 H 2 X H 2 With asymmetric molecules, there are more possible ways for homolytic dissociation (Fig.29). One rule here is that the fragment resulting from the dissociation of the bigger radical has a higher intensity than the fragment resulting from the opposite. This only if it is possible to compare the dissociated radicals as in the example. This is caused by the fact that the biggest radical R 1 H 2 X R 1 H 2 X H 2 H 2 R 2 -R 1 H 2 X H 2 X H 2 R 2 H 2 R 1,R 2 = H, alkyl; X = O, S, N etc. Figure 29: ompetetive homolytic dissociation. R 2-33-
37 can absorb more vibrational and rotational energy than the smaller radical can. In the first, the remaining ion will be less tending to further dissociation than the ion in the second case. Several examples of competitive homolytic dissociation exist and in the following, five will be presented Butanol 100 OH H 3 H 50 OH OH 25 0 H 2 H H H H H H H Figure 30: Mass spectrum of 2-butanol. H 3 OH H H 3 H 2 From the molecular ion we expect, in decreasing order, the dissociation of an ethyl-radical, a methyl-radical or a hydrogen atom (Fig.30) methyl-2-propanol HO H 3 H H H 3 H Figure 31: Mass spectrum of 2-methyl-2-propanol (t-butanol). H O H 3 H 3 H 3 The spectrum of the isomer of 2-butanol, 2-methyl-2-propanol or t-butanol, differs drastically from the former. This is because trough _-cleavage from t-butanol only methyl-radicals can be split off. -34-
38 3.2.3 N-ethyl-n-propylamine and N-(tert-butyl)-N-methylamine NH 2 H H 2 NH H 3 58 H 2 NH H 2 H 3 H 2 HN H 3 H 2 H 2 H 2 H 3 H 2 NH H Figure 32: Mass spectrum of 1-Propanamine, N-ethyl- H H 3 H 3 H H H 3 Figure 33: Mass spectrum of N-tert-Butylmethylamine. NH H 3 72 H 3 H 3 H 3 NH H 3 For these compounds the same applies as for 2-butanol. Here also, in descending order, an ethyl-radical, a methyl-radical or a hydrogen atom is split off Ethylbenzene Here one expects the molecularion to miss a -electron resulting in the dissociation of a methylradical or a hydrogen atom. Elaborate investigation has proven that the formed ions (m/z 91 and m/z 105) do posess a tropylium and a methyl-tropylium structure. These ions are very stable because they obey The Hückel rule for (4n+2 conjugated electrons) aromatic systems. They both have 6 conjugated electrons m/z = H -H 3 H 3 50 H -H m/z = Figure 34: Mass spectrum and partial fragmentation scheme of ethylbenzene. H
39 3.3 Heterolytic dissociation Mass the Organic hemistry Department Heterocyclic cleavage of a bond R-X means that that both electrons of the bond are either transferred to fragment X or fragment R. The spectra of chloro-ethane and bromoethane (below) serve as an example. Here the base-peak is found at mass 29 due to heterolytic cleavage. R X R + X X = l, Br, I or another stable radical Figure 35: Heterolytic dissociation H 5 m/z = 29 -l l 2 H Figure 36: Mass spectrum of chloro-ethane H 5 -Br Br 2 H 5 m/z = Figure 37: Mass spectrum of bromo-ethane. This heterolytic cleavage also appears with ethers and thio-ethers. In the mass spectrum of dimethyl ether besides the base-peak at m/z 45, formed by -cleavage, also a strong peak at m/z 15 appears. 100 H H 3 O H 3 -H 3 O H 3 29 H 3 O H -H 2 H 3 O H 2 H 3 O H Figure 38: Mass spectrum of dimethyl ether. -36-
40 3.4 The McLafferty rearrangement. When we compare the mass spectra of 3-methyl-2-butanone and 2-pentanone the biggest difference is the presence of peak at m/z 58 in the latter (fig 39, 40) Figure 39: Mass spectrum of 3-methyl-2-butanone H 2 H O H 2 H O - 2 H 4 H O 50 H 2 H 2 H 3 H 2 H 2 H 3 H 2 H m/z 86 m/z 86 m/z Figure 40: Mass spectrum of 2-pentanone. The peaks at m/z 71 and m/z 43 in both cases are explained easily by -cleavage. The most remarkable about m/z 58 is that it has an even mass. This only occurs if there are two bonds broken. When breaking 1 single bond in an ordinary compound containing, H and O, always fragments are formed with an odd mass. McLafferty showed that the fragment at m/z 58 is formed from the molecular-ion off 2-pentanone in the way shown in Fig. 39. In this mechanism the H atom (6-position) moves to the oxygen-atom in the 1 position. H 2 H 2 H H 2 Y X Z Examples of X,Y & Z O H O OR O R O NH 2 Figure 41: Examples of the McLafferty rearangement. Followed by homolytic cleavage of the - -bond ethene is eliminated. This reaction only occurs in the presence of a H-atom, is quite common and also appears in compounds with structures similar to the ones in Fig s 41 and 42. N O OH O O -37-
41 R H O H OH H R H OH O m/z 44 m/z 58 H 3 H 3 R H O NH 2 OH NH 2 m/z 59 R H O OH OH OH m/z 60 R H O OH 3 OH OH 3 m/z 74 H R H H m/z 92 Figure 42: Some characteristic fragments formed in the McLafferty rearrangement. 3.5 The retro Diels-Alder reaction In the in the organic chemistry well-known R R Diels-Alder reaction an unsaturated 6- membered ring is formed. With radical-ions [1] containing such an unsaturated ring this reaction can easily proceed in the oposite R R direction and is called the retro Diels-Alder. [2] The fragment that will bear the charge depends on the composition of R. When R=H the intensity of [2] will be smaller than 5%. Figure 43: retro Diels-Alder mechanism. When R= 6 H 5 [2] will be the base-peak (100%) in the spectrum. In general both fragments will be found as can been seen in the example of 2-norbornene (fig. 44) H H 4 + H 2 H 2 H 2 H H H H Figure 44: Mass spectrum and fragmentation of 2-Norbornene. -38-
42 3.6 Stevenson s Rule Mass the Organic hemistry Department This rule states: When cleaving a bond the positive charge preferably will end at the fragment with the lowest ionisation potential. The easiest way to explain this is shown in Fig. 45. From this picture it is obvious that the reaction with the lowest AP, the formation of A +, is the process where the charge remains on the fragment with the lowest IP, ionisation potential (I A < I B ). This is in agreement with the rule. A + + B AB + AB I(AB) A + B + AP(A + ) AP(B + ) I(A ) I(B ) D(A-B) I = ionisationpotential, D = dissociationenergy, AP = appearancepotential, the minimum energy required for formation of a fragment-ion Figure 45: Stevenson's Rule illustrated. A + B 3.7 Further dissociation of fragment ions Remarks The further fragmentation of the primarily formed fragment ions is not easily explained. We will pay attention to the most important ones only. First of all it has to be mentioned that through as well homolytic as hetrolytic cleavage from the molecular ion, a radical-cation, a cation is formed (see 3.1, 3.2 & 3.3). This cation can t lose radicals on further dissociation, but only neutral fragments. In shorthand: etc The transition: Is called a simple cleavage transition. - + Within the McLafferty rearrangement (3.4) from the molecular ion a neutral fragment is split off and hence the formed fragment is a radical cation. On further dissociation this will lose a radical first so that a cation is formed. - + etc In this case the first transition: Is called a rearrangement reaction. Some examples of dissociations of primary formed ions will be given hereafter. -39-
43 3.7.2 Loss of O from acylium ions. From the molecular ion from acetone trough homolytic cleavage an acylium ion is formed. This acylium ion can split off carbonmonoxide through heterolytic cleavage. O -H 3 H 3 O H 3 O + H 3 H Loss of alkenes from ethers, alcohols etcetera. After a homolytic cleavage of the molecular ion ether, alcohols, amines etc. can decompose further through split off of an alkene: H R 1 H 2 H 2 + X H 2 R 2 R 1 H H 2 + X H 2 + R 2 H + R 1 H2 + X H 2 X = N, O or S Alkanes also may lose alkenes, but without a proton transfer: [ H 3 H 2 H 2 + R [ H H 2 H 2 R + R + H 2 H Formation of ion-series. As stated above, alkanen can decompose by losing ethene, after which an R + -ion remains. These R + -ions themselves do decompose in the same way, resulting in other R + -ions. These R + -ions and R`+-ions also might result from other reactions, e.g. through rearrangment reactions or directly from the molecular ion (M + ). Through all these possible reactions a series of [ n H 2n+1 ] + -ions is formed. The maximum intensity for this series is found for n=3 (m/z 43) or n=4 (m/z 57). The series itself is characteristic for alkanes or compounds with a long alkane moiety. This kind of characteristic ion-series is also found with other classes of compounds. Some of the more important are listed in next table: -40-
44 Table 4: haracteristic Ion Series ompound Formula m/z values Alkanes n H 2n Aldehydes, ketones n H 2n-1 O Amines n H 2n+2 N Ethers, alcohols n H 2n+2 O Sulfides n H 2n+1 S hloro-alkanes n H 2n l Alkenes, cylo-alkanes n H 2n + Aromatics n H <n
45 Table 5: General fragments from molecular-ions m/z Fragment Found with, possible chemical class. M-1 H M-2 H 2 - M M-15 H 3 - M-16 O, NH 2 Ar-NO2, sulfoxides / ArSO2NH2, -ONH2 M-17 OH / NH 3 Alcohols / Amines M-18 H 2 O Alcohols, aldehydes, ketones etc. M-19 / 20 F / HF Fluorine derivatives M-26 2 H 2 Aromatic hydrocarbons M-27 HN Aromatic nitriles, nitrogen heterocyclics M-28 O / 2 H 4 hinones / Aromatic ethylethers, ethylesters, n-propylketones M-29 HO / 2 H 5 - / ethylketones, Ar-n-3H7 M-30 2 H 6 /H 2 O / NO - / aromayic methylethers / Aromatic NO2 M-31 OH 3 Methyl-esters M-32 H 3 OH / S Methyl-esters / - M-33 H2O + H 3 / HS - / thiols M-34 H 2 S Thiols M-41 3 H 5 Propyl-esters M-42 H 2 O / 3 H 6 Methyl-ketones, Ar-Acetates, ArNHOH3 /n- or iso-butylketones, Arpropyl ethers, Ar-n-4H9 M-43 3 H 7 / H 3 O Propyl-ketones, Ar-n-4H9 / methyl-ketones M-44 O 2 / 3 H 8 Esters (rearrangement) / anhydrides M-45 O 2 H / O 2 H 5 arbonic acids / ethyl-esters M-46 2 H 5 OH / NO 2 Ethyl-esters / Ar-NO2 M-48 SO Aromatic sulfoxides M-55 4 H 7 Butyl-esters M-56 4 H 8 t-butyl, Aromatic n,i-alkyl (n,i>=4), pentyl-ketone M-57 4 H 9 (t)-butyl-r M-58 4 H 10 - M-60 H 3 OOH Acetates -42-
46 4 Literature and references. Due to some hardware failure and mismanagement of literature archive not all of the sources can be reproduced overhere. Some of them will be partially found on the internet and a good place to start is at: I would like to state overhere that all who have made a contribution in any way are greatly acknowledged. 4.1 Some printed literature: N.M.M. Nibbering, ollege Massaspectrometrie, Amsterdam N.M.M. Nibbering, Organische Massaspectrometrie: reaktiemechanismen van ontledingen van ionen in de gasfase, Amsterdam F.W. McLafferty, Interpretation of Mass Spectra, University Sciense Books, Mill Valley, alifornia (1980). U.P. Schlunegger, Advanced Mass Spectrometry, Pergamon Press, Oxford (1983). I.Howe, D.H. Williams & R.D. Bowen, Mass Spectrometry Principles and Applications, McGraw-Hill Inc., New york (1981). R.I. Reed, Mass Spectrometry, Academic Press, London (1964) A.G. Harrison, hemical Ionisation Mass Spectrometry, R Press Inc., Florida (1983). N.M.M. Nibbering & R.H. Fokkens, Mass Spectrometry, ourse on Modern Spectroscopy, University of Twente Netherlands. N.B. ech &.G. Enke, Practical Implications of Some Recent Studies in Electrospray Ionization Fundamentals, Mass Spectrometry Reviews, 2001, 20, Steven Weinberg, Bouwstenen van het atoom, Wetenschappelijke Bibliotheek (Natuur & Techniek), D.A. Skoog & D.M. West, Principles of Instrumental Analysis, Holt, Rhinehart and Winston Inc. NY D. Henneberg und K. asper, Z. Anal. hem., 227, 2412 (1967). M. Karplus and R.N. Porter, Atoms & Molecules. John D. Roberts and Marjorie. aserio, Basic Principles of Organic hemistry, W.A. Benjamin Inc., NY Dudley H. Williams & Ian Fleming, Spektroscopische methoden zur Strukturaufklaerung, Georg Thieme Verlag Stuttgart, 1979 p Tools used Wordprocessing: WordPerfect 5.1, Word for Windows 97 & Drawing technical (fig 1-13,15-20, 45) EasyAD, export to HPGL and conversion to WPG2 format. Drawing chemical (fig figs in 3.7) hem Draw 6 Spectra (real, fig 21-24) recorded on a VG7070E and tailored with The Shrader Labs datasystem running on Windows 98SE Spectra (fig 25-27,30-40,44) taken from the Nist library within Highchem Massfrontier
47 5 Appendix Sample submission form This can be found on the internet at: > facilities MS form (PDF) or directly at (The MS home-page) GMS guidelines G/MS recipe required steps: 1. Weigh about 10 milligrams of your sample in a standard test-tube. Add 2 ml solvent (~80% height of test-tube) resulting in a standard solution of ~5 mgr./ml. 2. On a standard (split injection) G equipped with a a-polar column (e.g. HP1) inject no more then 1 microlitre using a suitable temperature program. 3. From the solution above make a proper dilution of 1 to 100 by adding 10 microlitre of it to 1 ml (preferably as a-polar as possible) solvent. The vials used for G/MS do have a volume of about 1 ml when filled to the neck. This dilution is necessary while the G/MS is equipped with on-column injection. 4. With the results acquired above and information concerning the sample (structure family, expected molar mass range etc.) contact the operator and discuss the desired analysis. 5. During the intake discussion arrangements concerning result presentation are made. Normally results are presented on floppy disk 1.4 MB pre-formatted with at least 400 kb free space. Questions and remarks? ontact Peter or Helene -44-
13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Mass Spectrometry. Overview
Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2
Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass
Protons, neutrons and electrons Nuclear Structure particle relative charge relative mass proton 1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons and neutrons make up
How To Use Gc-Ms
The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter
electron does not become part of the compound; one electron goes in but two electrons come out.
Characterization Techniques for Organic Compounds. When we run a reaction in the laboratory or when we isolate a compound from nature, one of our first tasks is to identify the compound that we have obtained.
Summary of the characteristics of different mass analyzers
Summary of the characteristics of different mass analyzers All mass spectrometers combine ion formation, mass analysis, and ion detection. This discussion is concerned with how various mass analyzers are
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis
Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis Main points of the chapter 1. Hydrogen Nuclear Magnetic Resonance a. Splitting or coupling (what s next to what) b. Chemical shifts (what type is it) c. Integration
CHE334 Identification of an Unknown Compound By NMR/IR/MS
CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
Symmetric Stretch: allows molecule to move through space
BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed
Nuclear Magnetic Resonance
Nuclear Magnetic Resonance NMR is probably the most useful and powerful technique for identifying and characterizing organic compounds. Felix Bloch and Edward Mills Purcell were awarded the 1952 Nobel
Organic Chemistry Tenth Edition
Organic Chemistry Tenth Edition T. W. Graham Solomons Craig B. Fryhle Welcome to CHM 22 Organic Chemisty II Chapters 2 (IR), 9, 3-20. Chapter 2 and Chapter 9 Spectroscopy (interaction of molecule with
Cathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
Measurement of Charge-to-Mass (e/m) Ratio for the Electron
Measurement of Charge-to-Mass (e/m) Ratio for the Electron Experiment objectives: measure the ratio of the electron charge-to-mass ratio e/m by studying the electron trajectories in a uniform magnetic
Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis
Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** [email protected]
Ultraviolet Spectroscopy
Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer
Mass Spec - Fragmentation
Mass Spec - Fragmentation An extremely useful result of EI ionization in particular is a phenomenon known as fragmentation. The radical cation that is produced when an electron is knocked out of a neutral
0 10 20 30 40 50 60 70 m/z
Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions
Infrared Spectroscopy 紅 外 線 光 譜 儀
Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed
Chapter 1: Moles and equations. Learning outcomes. you should be able to:
Chapter 1: Moles and equations 1 Learning outcomes you should be able to: define and use the terms: relative atomic mass, isotopic mass and formula mass based on the 12 C scale perform calculations, including
Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
Vacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
INFRARED SPECTROSCOPY (IR)
INFRARED SPECTROSCOPY (IR) Theory and Interpretation of IR spectra ASSIGNED READINGS Introduction to technique 25 (p. 833-834 in lab textbook) Uses of the Infrared Spectrum (p. 847-853) Look over pages
TOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] [email protected] [e-mail] This
E/M Experiment: Electrons in a Magnetic Field.
E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.
For example: (Example is from page 50 of the Thinkbook)
SOLVING COMBINED SPECTROSCOPY PROBLEMS: Lecture Supplement: page 50-53 in Thinkbook CFQ s and PP s: page 216 241 in Thinkbook Introduction: The structure of an unknown molecule can be determined using
Mass Spectrometry: Quadrupole Mass Filter
Advanced Lab, Jan. 2008 Mass Spectrometry: Quadrupole Mass Filter The mass spectrometer is essentially an instrument which can be used to measure the mass, or more correctly the mass/charge ratio, of ionized
Ionization Methods in Organic Mass Spectrometry Contents:
Ionization Methods in Organic Mass Spectrometry Contents:. Introduction. Gas-Phase Ionization. Electron Ionization (EI). Chemical Ionization (CI). Desorption Chemical Ionization (DCI). Negative-ion chemical
Amount of Substance. http://www.avogadro.co.uk/definitions/elemcompmix.htm
Page 1 of 14 Amount of Substance Key terms in this chapter are: Element Compound Mixture Atom Molecule Ion Relative Atomic Mass Avogadro constant Mole Isotope Relative Isotopic Mass Relative Molecular
Indiana's Academic Standards 2010 ICP Indiana's Academic Standards 2016 ICP. map) that describe the relationship acceleration, velocity and distance.
.1.1 Measure the motion of objects to understand.1.1 Develop graphical, the relationships among distance, velocity and mathematical, and pictorial acceleration. Develop deeper understanding through representations
QGA Quantitative Gas Analyser
QGA Quantitative Gas Analyser A compact bench-top system for real-time gas and vapour analysis Detailed product information / introduction catalysis studies environmental gas analysis fermentation off-gas
Electrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
Suggested solutions for Chapter 3
s for Chapter PRBLEM Assuming that the molecular ion is the base peak (00% abundance) what peaks would appear in the mass spectrum of each of these molecules: (a) C5Br (b) C60 (c) C64Br In cases (a) and
CHAPTER D1 ION OPTICS OF MAGNETIC/ELECTRIC SECTOR MS
Back to Basics Section D: Ion Optics CHAPTER D1 ION OPTICS OF MAGNETIC/ELECTRIC SECTOR MS TABLE OF CONTENTS QuickGuide...369 Summary...369 Preamble...371 MassAnalysisofIons...371 MagneticSector...371 ElectrostaticAnalyser(ElectricSector)...375
Secondary Ion Mass Spectrometry
Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown,
Infrared Spectroscopy: Theory
u Chapter 15 Infrared Spectroscopy: Theory An important tool of the organic chemist is Infrared Spectroscopy, or IR. IR spectra are acquired on a special instrument, called an IR spectrometer. IR is used
Organic Spectroscopy. UV - Ultraviolet-Visible Spectroscopy. !! 200-800 nm. Methods for structure determination of organic compounds:
Organic Spectroscopy Methods for structure determination of organic compounds: X-ray rystallography rystall structures Mass spectroscopy Molecular formula -----------------------------------------------------------------------------
2. Molecular stucture/basic
2. Molecular stucture/basic spectroscopy The electromagnetic spectrum Spectral region for atomic and molecular spectroscopy E. Hecht (2nd Ed.) Optics, Addison-Wesley Publishing Company,1987 Spectral regions
NMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE. 4 3 2 1 0 δ PUBLISHING
A N I N T R O D U T I O N T O... NMR SPETROSOPY NULEAR MAGNETI RESONANE 4 3 1 0 δ Self-study booklet PUBLISING NMR Spectroscopy NULEAR MAGNETI RESONANE SPETROSOPY Origin of Spectra Theory All nuclei possess
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Mass spectrometry. Dr. Kevin R. Tucker
Mass spectrometry Dr. Kevin R. Tucker What is mass spectrometry? A versatile analytical technique used to determine the composition of chemical samples either at the atomic or molecular level. Ionic forms
Determining the Structure of an Organic Compound
Determining the Structure of an Organic Compound The analysis of the outcome of a reaction requires that we know the full structure of the products as well as the reactants In the 19 th and early 20 th
DO PHYSICS ONLINE FROM QUANTA TO QUARKS QUANTUM (WAVE) MECHANICS
DO PHYSICS ONLINE FROM QUANTA TO QUARKS QUANTUM (WAVE) MECHANICS Quantum Mechanics or wave mechanics is the best mathematical theory used today to describe and predict the behaviour of particles and waves.
F321 THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important are... in the nucleus of an atom
Atomic Structure F32 TE STRUCTURE OF ATOMS ATOMS Atoms consist of a number of fundamental particles, the most important are... Mass / kg Charge / C Relative mass Relative Charge PROTON NEUTRON ELECTRON
Getting the most from this book...4 About this book...5
Contents Getting the most from this book...4 About this book....5 Content Guidance Topic 1 Atomic structure and the periodic table...8 Topic 2 Bonding and structure...14 Topic 2A Bonding....14 Topic 2B
Solving Spectroscopy Problems
Solving Spectroscopy Problems The following is a detailed summary on how to solve spectroscopy problems, key terms are highlighted in bold and the definitions are from the illustrated glossary on Dr. Hardinger
Tuning & Mass Calibration
Tuning & Mass Calibration 1 1 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within
The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD
TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic
1 The water molecule and hydrogen bonds in water
The Physics and Chemistry of Water 1 The water molecule and hydrogen bonds in water Stoichiometric composition H 2 O the average lifetime of a molecule is 1 ms due to proton exchange (catalysed by acids
1. The diagram below represents magnetic lines of force within a region of space.
1. The diagram below represents magnetic lines of force within a region of space. 4. In which diagram below is the magnetic flux density at point P greatest? (1) (3) (2) (4) The magnetic field is strongest
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS
AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
Nuclear Magnetic Resonance Spectroscopy
Nuclear Magnetic Resonance Spectroscopy Nuclear magnetic resonance spectroscopy is a powerful analytical technique used to characterize organic molecules by identifying carbonhydrogen frameworks within
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS
PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.
Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
Proton Nuclear Magnetic Resonance ( 1 H-NMR) Spectroscopy
Proton Nuclear Magnetic Resonance ( 1 H-NMR) Spectroscopy Theory behind NMR: In the late 1940 s, physical chemists originally developed NMR spectroscopy to study different properties of atomic nuclei,
Name Class Date. In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question.
Assessment Chapter Test A Chapter: States of Matter In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1. The kinetic-molecular
3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
PS-6.2 Explain the factors that determine potential and kinetic energy and the transformation of one to the other.
PS-6.1 Explain how the law of conservation of energy applies to the transformation of various forms of energy (including mechanical energy, electrical energy, chemical energy, light energy, sound energy,
Physics 9e/Cutnell. correlated to the. College Board AP Physics 1 Course Objectives
Physics 9e/Cutnell correlated to the College Board AP Physics 1 Course Objectives Big Idea 1: Objects and systems have properties such as mass and charge. Systems may have internal structure. Enduring
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
Chemistry 13: States of Matter
Chemistry 13: States of Matter Name: Period: Date: Chemistry Content Standard: Gases and Their Properties The kinetic molecular theory describes the motion of atoms and molecules and explains the properties
4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a
BASIC PRINCIPLES INTRODUCTION TO NUCLEAR MAGNETIC RESONANCE (NMR) 1. The nuclei of certain atoms with odd atomic number, and/or odd mass behave as spinning charges. The nucleus is the center of positive
CHAPTER 6 Chemical Bonding
CHAPTER 6 Chemical Bonding SECTION 1 Introduction to Chemical Bonding OBJECTIVES 1. Define Chemical bond. 2. Explain why most atoms form chemical bonds. 3. Describe ionic and covalent bonding.. 4. Explain
Elements in the periodic table are indicated by SYMBOLS. To the left of the symbol we find the atomic mass (A) at the upper corner, and the atomic num
. ATOMIC STRUCTURE FUNDAMENTALS LEARNING OBJECTIVES To review the basics concepts of atomic structure that have direct relevance to the fundamental concepts of organic chemistry. This material is essential
1. Units of a magnetic field might be: A. C m/s B. C s/m C. C/kg D. kg/c s E. N/C m ans: D
Chapter 28: MAGNETIC FIELDS 1 Units of a magnetic field might be: A C m/s B C s/m C C/kg D kg/c s E N/C m 2 In the formula F = q v B: A F must be perpendicular to v but not necessarily to B B F must be
Determination of Molecular Structure by MOLECULAR SPECTROSCOPY
Determination of Molecular Structure by MOLEULAR SPETROSOPY hemistry 3 B.Z. Shakhashiri Fall 29 Much of what we know about molecular structure has been learned by observing and analyzing how electromagnetic
Coating Technology: Evaporation Vs Sputtering
Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information
KINETIC MOLECULAR THEORY OF MATTER
KINETIC MOLECULAR THEORY OF MATTER The kinetic-molecular theory is based on the idea that particles of matter are always in motion. The theory can be used to explain the properties of solids, liquids,
5s Solubility & Conductivity
5s Solubility & Conductivity OBJECTIVES To explore the relationship between the structures of common household substances and the kinds of solvents in which they dissolve. To demonstrate the ionic nature
Time-of-Flight Mass Spectrometry
Time-of-Flight Mass Spectrometry Technical Overview Introduction Time-of-flight mass spectrometry (TOF MS) was developed in the late 1940 s, but until the 1990 s its popularity was limited. Recent improvements
PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.
PHYS 222 Spring 2012 Final Exam Closed books, notes, etc. No electronic device except a calculator. NAME: (all questions with equal weight) 1. If the distance between two point charges is tripled, the
How To Build An Amd Quas 3 Ar Mass Spectrometer
Classical and special AMD Magnetic Sector mass spectrometers (extracted from power point presentations) High Resolution Mass Spectromter AMD 402/403 S 6 High Resolution Mass Spectrometer AMD 604 S 8 Version
Application Note AN4
TAKING INVENTIVE STEPS IN INFRARED. MINIATURE INFRARED GAS SENSORS GOLD SERIES UK Patent App. No. 2372099A USA Patent App. No. 09/783,711 World Patents Pending INFRARED SPECTROSCOPY Application Note AN4
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
Modern Physics Laboratory e/m with Teltron Deflection Tube
Modern Physics Laboratory e/m with Teltron Deflection Tube Josh Diamond & John Cummings Fall 2010 Abstract The deflection of an electron beam by electric and magnetic fields is observed, and the charge
Exercises Topic 2: Molecules
hemistry for Biomedical Engineering. Exercises Topic 2 Authors: ors: Juan Baselga & María González Exercises Topic 2: Molecules 1. Using hybridization concepts and VSEPR model describe the molecular geometry
States of Matter CHAPTER 10 REVIEW SECTION 1. Name Date Class. Answer the following questions in the space provided.
CHAPTER 10 REVIEW States of Matter SECTION 1 SHORT ANSWER Answer the following questions in the space provided. 1. Identify whether the descriptions below describe an ideal gas or a real gas. ideal gas
Chapter 3. Mass Relationships in Chemical Reactions
Chapter 3 Mass Relationships in Chemical Reactions This chapter uses the concepts of conservation of mass to assist the student in gaining an understanding of chemical changes. Upon completion of Chapter
CHAPTER 6 REVIEW. Chemical Bonding. Answer the following questions in the space provided.
Name Date lass APTER 6 REVIEW hemical Bonding SETIN 1 SRT ANSWER Answer the following questions in the space provided. 1. a A chemical bond between atoms results from the attraction between the valence
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
PROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR)
PROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR) WHAT IS H-NMR SPECTROSCOPY? References: Bruice 14.1, 14.2 Introduction NMR or nuclear magnetic resonance spectroscopy is a technique used to determine
Chapter 21 Principles of Chromatography and Mass Spectrometry Problems 1,2,3,4,8,9,11,13
Chapter 21 Principles of Chromatography and Mass Spectrometry Problems 1,2,3,4,8,9,11,13 21-1 What is Chromatography Process to separate and purify compounds in a mixture by passing them trough a material
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
Chapter 2: The Chemical Context of Life
Chapter 2: The Chemical Context of Life Name Period This chapter covers the basics that you may have learned in your chemistry class. Whether your teacher goes over this chapter, or assigns it for you
History of the Atom & Atomic Theory
Chapter 5 History of the Atom & Atomic Theory You re invited to a Thinking Inside the Box Conference Each group should nominate a: o Leader o Writer o Presenter You have 5 minutes to come up with observations
Tutorial 4.6 Gamma Spectrum Analysis
Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning
Forensic Science Standards and Benchmarks
Forensic Science Standards and Standard 1: Understands and applies principles of scientific inquiry Power : Identifies questions and concepts that guide science investigations Uses technology and mathematics
Magnetic Field and Magnetic Forces
Chapter 27 Magnetic Field and Magnetic Forces PowerPoint Lectures for University Physics, Thirteenth Edition Hugh D. Young and Roger A. Freedman Lectures by Wayne Anderson Goals for Chapter 27 Magnets
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
Proton Nuclear Magnetic Resonance Spectroscopy
Proton Nuclear Magnetic Resonance Spectroscopy Introduction: The NMR Spectrum serves as a great resource in determining the structure of an organic compound by revealing the hydrogen and carbon skeleton.
Molecular Formula Determination
Molecular Formula Determination Classical Approach Qualitative elemental analysis Quantitative elemental analysis Determination of empirical formula Molecular weight determination Molecular formula determination
