Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
|
|
|
- Colin McBride
- 10 years ago
- Views:
Transcription
1 Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada) Figure 1. Schematic of Triple Quadrupole Configuration Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Higher selectivity resulting in less interference of co-eluting compounds and matrix, thus less HPLC separation is required Better Signal-to-Noise (S/N) allowing quantitation with lower limits of quantitation More reliable identification of detected analytes using Multiple Reaction Monitoring (MRM) in comparison to Selected Ion Monitoring (SIM) Wider linear range of quantitation Better accuracy and reproducibility especially at low concentrations Overview This paper describes the use of single quadrupole and triple quadrupole Mass Spectrometry coupled to Liquid Chromatography for the analysis of 17 pesticides in drinking water and soil samples. Both detection technologies are compared with respect to selectivity, sensitivity, identification, linear range, accuracy, and reproducibility for quantitative analysis. Introduction The coupling of Liquid Chromatography (LC) and tandem Mass Spectrometry (MS/MS) is a widely used analytical technique for quantitative and qualitative analysis. Electrospray Ionization (ESI), Atmospheric Pressure Chemical Ionization (APCI), or Photo Ionization (APPI) allow the ionization of various semivolatile, thermally labile, and polar to nonpolar compounds, such as pharmaceuticals, pesticides, personal care products, steroids, explosives, drugs of abuse etc., in trace levels. Generated ions will be transferred after ionization through a vacuum interface into the mass analyzer. Quadrupoles are mass analyzers which consist of four rods with DC and RF voltages applied. An ion of a specific mass-to-charge ratio (m/z) will be stable and can pass through the quadrupole only when a specific DC/RF voltage combination is applied. Quadrupoles are therefore called mass filters. p 1
2 last m/z last m/z last SIM last SIM last MRM last MRM m/z m/z Q1/Q3 SIM 2 SIM 2 MRM 2 MRM 2 m/z 3 m/z 2 m/z 1 m/z 3 m/z 2 m/z 1 SIM 1 SIM 1 MRM 1 MRM 1 Time Time Time 1.5e8 1.0e e5 1.5e5 1.4e5 1.3e e e5 1.1e e e4 8.0e4 7.0e e5 8.00e4 6.00e4 4.00e4 2.00e e m/z, amu 0 Figure 2. Pesticide in a soil extract detected in different single and triple quadrupole scan modes (left to right): Full scan with MS spectrum (50 μg/kg), Selected Ion Monitoring (5 μg/kg), and Multiple Reaction Monitoring (5 μg/kg) Single quadrupole systems contain only one mass filtering quadrupole while triple quadrupole systems consist of three quadrupoles. Q1 and Q3 are working as mass filters while Q2 is acting as collision cell. Quadrupoles can be used in scanning or filtering mode. During a mass scan, DC and RF voltages are ramped resulting in the acquisition of full scan mass spectra. Such spectra are typically used for qualitative data analysis. However, scanning a quadrupole suffers from low sensitivity and slow scan speed. Thus, quantitative studies are performed with quadrupoles working in filtering mode. The most selective mode to use a single quadrupole MS is called Selected Ion Monitoring (SIM). Hereby, a fixed set of DC and RF voltages is applied to the quadrupole and thus only a single m/z can pass. Ions with different m/z are filtered out. Multiple Reaction Monitoring (MRM) is the most common mode of using a triple quadrupole MS/MS for quantitative analysis, allowing enhanced sensitivity and selectivity. The first quadrupole filters a specific precursor ion of interest. Ions generated in the ion source having a different m/z can not pass Q1. The collision cell is optimized to produce a characteristic product ion by collision of the precursor ion with a neutral collision gas, such as nitrogen. This process is called Collision Induced Dissociation (CID). Generated product ions are transferred into the third quadrupole where only a specific m/z is allowed to pass. All other product ions are filtered out in Q3. Thus MRM mode works like a double mass filter which drastically reduces noise and increases selectivity. The principle and resulting chromatograms of various scan modes analyzing a mix of pesticides spiked into soil is given in Figure 2. It illustrates how increasing selectivity of the mass spectrometric experiment reduces the noise of eluting background and matrix components. Single quadrupole and triple quadrupole systems allow the detection of many SIM and MRM transitions, respectively. This enables quantitation of many targeted analytes in a single experiment. Typically, additional SIM and MRM transitions have to be detected to perform identification of quantified compounds. Hereby, the most intense ion is called the quantifier and all additional ions are called qualifiers. The EU Commission Decision 2002/657/EC defined performance criteria for confirmatory methods, such as MS and MS/MS, by introducing the concept of identification points. The required number of 4 identification points can be achieved by detecting 4 SIM on a single quadrupole MS or 2 MRM transitions on a triple quadrupole MS/MS. p 2
3 Table 1. Pesticides detected in Multiple Reaction Monitoring (MRM) and Selected Ion Monitoring (SIM) Pesticide CAS MRM 1 MRM 2 SIM 1 SIM 2 SIM 3 SIM 4 t R (min) Atrazine / / Chlortoluron /72 213/ Cyanazine / / Desethylatrazine / / Diuron /72 233/ Hexazinone / / Isoproturon /72 207/ * 134* 7.9 Linuron / / Metazachlor / / * Methabenzthiazuron / * Metobromuron / / Metolachlor / / Metoxuron /72 229/ Monolinuron / / Sebuthylazine / / Simazine / / Terbuthylazine / / * Not enough HPLC separation to differentiate these in-source fragment ions Experimental HPLC An Agilent 1200 standard HPLC system with binary pump, well plate autosampler, and column oven was used. Chromatographic separation was performed on a Phenomenex Synergi 4u Fusion RP-80 (50x2 mm) column. A gradient of eluent A (water with 5 mm ammonium formate) and eluent B (methanol with 5 mm ammonium formate) from 80/20 to 10/90 (A/B) over 8min was used. The column temperature was set to 25 C. A sample volume of 10 μl was injected. MS/MS MS and MS/MS detection was performed using an API 3200 LC/MS/MS system with Turbo V source and an Electrospray ionization probe. All pesticides were detected in positive polarity using an ionization voltage of 5000V. The ion source temperature was set to 500 C. The detected quantifier and qualifier ions are listed in Table 1. A dwell time of 25 ms was used to detect 34 MRM transitions and 15ms to detect 68 SIM. Results and Discussion Selectivity Detection on a triple quadrupole MS/MS results in higher selectivity due to the double mass filtering (Figure 3). In MRM typically a single signal per analyte is detected while SIM can result in multiple signals per compound. The reason for this is the generation of identical in-source fragment ions of pesticides of the same compound class. Thus more time consuming HPLC is required to separate such compounds to allow quantitation and identification. Needed HPLC development and analysis times are then comparable to traditional UV detection. Sensitivity Detection on a MS/MS system also results in better sensitivity. Detected Signal-to-Noise (S/N) is higher resulting in lower Limits of Quantitation (LOQ). The typical difference in sensitivity observed is at least one order of magnitude as illustrated in Figure 4. p 3
4 XIC of +MRM (34 pairs): 216.1/174.0 amu from Sample 42 (QqQ 100) of Data quant Q vs QqQ.wiff (Tur... Max. 9.6e4 cps. XIC of +MRM (34 pairs): 233.1/72.0 amu from Sample 42 (QqQ 100) of Data quant Q vs QqQ.wiff (Turb... Max. 4.7e4 cps. 9.6e4 9.0e4 8.0e4 7.0e4 6.0e4 216/ / e4 3.5e4 233/72 233/ e4 1.5e4 500 XIC of +Q1 MI (68 ions): amu from Sample 41 (Q 100) of Data quant Q vs QqQ.wiff (Turbo Spray)... Max. 9.6e5 cps. XIC of +Q1 MI (68 ions): amu from Sample 41 (Q 100) of Data quant Q vs QqQ.wiff (Turbo Spray)... Max. 5.4e5 cps. 9.6e5 9.0e5 8.0e5 7.0e5 6.0e5 4.0e5 3.0e5 2.0e e5 4.5e5 4.0e5 3.5e5 3.0e5 2.5e5 2.0e5 1.5e Figure 3. Comparison of selectivity detecting Atrazine (left) and Diuron (right) in MRM and SIM mode at 100 μg/l in drinking water XIC of +MRM (34 pairs): 253.2/171.2 amu from Sample 14 (QqQ 1) of Data quant Q vs QqQ.wiff (Turb... Max cps. XIC of +MRM (34 pairs): 213.1/72.2 amu from Sample 30 (QqQ 10) of Data quant Q vs QqQ.wiff (Turb... Max cps /171 S/N= /71 S/N= /72 S/N= /46 S/N= XIC of +Q1 MI (68 ions): amu from Sample 13 (Q 1) of Data quant Q vs QqQ.wiff (Turbo Spray),... Max. 5.6e4 cps. 0 XIC of +Q1 MI (68 ions): amu from Sample 29 (Q 10) of Data quant Q vs QqQ.wiff (Turbo Spray),... Max. 7.8e4 cps. 5.6e S/N= S/N=6 85 S/N<3 71 S/N=5 7.8e4 7.0e4 6.0e4 213 S/N=55 72 S/N= S/N=3 168 S/N=6 7.8 Figure 4. Comparison of sensitivity detecting 1 μg/l Hexazinone (left) and 10 μg/l Chlortoluron (right) in MRM and SIM mode in drinking water If identification with 4 identification points is required and more than a single ion has to be detected the difference in LOQ based on quantifier MRM and SIM is significantly higher (Table 2). Linearity and Reproducibility Generally a wider linear range can be used for quantitation if a triple quadrupole MS/MS is used because of increased selectivity allowing detection at lower LOQ (Figure 5). The upper limit of quantitation of SIM and MRM is typically comparable due to saturation of the ion source or detector. The typical linear range in MRM using Electrospray Ionization is 3 to 4 orders of magnitude while SIM provides only 2 to 3 orders of magnitude. Additionally, MS/MS allows more accurate and reproducible quantitation. Table 3 compares Coefficients of Variation (%CV) of detected pesticides at different concentrations. Triple quadrupole MS/MS provides %CV below 10 at LOQ and below 5 at higher concentrations. The reproducibility on single quadrupole MS is significantly lower. p 4
5 Table 2. Limits of Quantitation (LOQ) and Limit of Identification (LOI) of detected pesticides* Pesticide LOQ (MRM 1) LOI (MRM 2) LOQ (SIM 1) LOI (SIM 4) Enhanced Quantifier Enhanced Qualifier Atrazine x 20x Chlortoluron x 20x Cyanazine x 10x Desethylatrazine x 20x Diuron x 20x Hexazinone x 20x Isoproturon N/A 10x N/A Linuron x 50x Metazachlor x 250x Methabenzthiazuron x 40x Metobromuron x 10x Metolachlor x 500x Metoxuron x 40x Monolinuron x 25x Sebuthylazine x 20x Simazine x 5x Terbuthylazine x 20x *using Multiple Reaction Monitoring (MRM) and Selected Ion Monitoring (SIM) in drinking water samples based on S/N=6 Table 3. Reproducibility (%CV) of detected pesticides in MRM and SIM at different concentrations Pesticide %CV at 1μg/L (MRM) %CV at 10μg/L (MRM) %CV at 100μg/L (MRM) %CV at 1μg/L (SIM) %CV at 10μg/L (SIM) %CV at 100μg/L (SIM) Atrazine Chlortoluron Cyanazine Desethyl-atrazine Diuron Hexazinone Isoproturon Linuron <LOQ Metazachlor Methabenzthiazuron Metobromuron <LOQ Metolachlor Metoxuron Monolinuron <LOQ Sebuthylazine Simazine Terbuthylazine p 5
6 quantitation QqQ spiked water.rdb (Linuron 1): "Linear" Regression ("1 / x" weighting): y = 1.7e+003 x (r = ) quantitation QqQ spiked water.rdb (Terbuthylazine 1): "Linear" Regression ("1 / x" weighting): y = 1.22e+004 x (r = ) 3.4e5 3.0e5 2.5e5 2.0e5 249/160 r= ( μg/L) 2.4e6 2.0e6 1.5e6 230/174 r= ( μg/L) 1.5e5 1.0e6 quantitation Q spiked water.rdb (Linuron 1): "Linear" Regression ("1 / x" weighting): y = 1.86e+004 x e+004 (r = ) quantitation Q spiked water.rdb (Terbuthylazine 1): "Linear" Regression ("1 / x" weighting): y = 1.08e+005 x e+004 (r = ) 3.5e6 3.0e6 2.5e6 249 r= (2-200μg/L) 2.0e7 1.8e7 1.6e7 1.4e7 230 r= ( μg/L) 2.0e6 1.2e7 1.0e7 1.5e6 8.0e6 1.0e6 6.0e6 4.0e6 2.0e6 Figure 5. Comparison of linearity detecting Linuron (left) and Terbuthylazine (right) in MRM (top) and SIM (bottom) mode in spiked drinking water samples Summary The detection in Multiple Reaction Monitoring (MRM) using a triple quadrupole MS/MS has a number of advantages in comparison to Selected Ion Monitoring (SIM) of a single quadrupole MS. Due to double mass filtering MS/MS detection provides much higher selectivity with less interference of co-eluting compounds and matrix components, resulting in less time consuming method development and faster analysis times. Better Signal-to-Noise allows quantitation with lower Limits of Quantitation (LOQ). Fewer ions have to be detected per compound on MS/MS in comparison to MS for confirmatory analysis. The concept of identification points introduced by the EU Commission Decision requires the detection of 2 MRM transitions and 4 SIM, respectively. If identification is required the detected qualifier signals have to be compared to specify LOQ. The difference between both technologies can be then at least two orders of magnitude. Finally, a wider linear range, higher accuracy, and reproducibility can be obtained on triple quadrupole MS/MS. For Research Use Only. Not for use in diagnostic procedures AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Publication number: Headquarters International Sales 353 Hatch Drive Foster City CA USA For our office locations please call the division Phone headquarters or refer to our website at
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection
Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection Application Note Food Testing & Agriculture Authors Edgar Naegele and Thomas
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS
Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS D. Loeffler 1, M. Ramil 1, T. Ternes 1, M. Suter 2, R. Schönenberger 2, H.-R. Aerni 2, S, König, A. Besa,
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
Cliquid Drug Screen & Quant Software for Routine Forensic Toxicology. great interest in forensic, toxicological and clinical research laboratories.
Application Note Cliquid Drug Screen & Quant Software A Fast and Sensitive LC/MS/MS Method for the Quantitation and Confirmation of 30 Benzodiazepines and Nonbenzodiazepine Hypnotics in Forensic Urine
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples
MultiQuant Software Version 3.0 for Accurate Quantification of Clinical Research and Forensic Samples Fast and Efficient Data Review, with Automatic Flagging of Outlier Results Adrian M. Taylor and Michael
Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS
Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers
Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants
Metabolomic Profiling of Accurate Mass LC- Data to Identify Unexpected Environmental Pollutants André Schreiber 1, David Cox 1, Nadia Pace 1, Christopher Borton 2 1 AB SCIEX, Concord, ntario, Canada; 2
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics
DRUG METABOLISM. Drug discovery & development solutions FOR DRUG METABOLISM
DRUG METABLISM Drug discovery & development solutions FR DRUG METABLISM Fast and efficient metabolite identification is critical in today s drug discovery pipeline. The goal is to achieve rapid structural
Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS
Application note Explosives by Direct Injection LC/MS/MS Quantitative Analysis of Explosives in Surface Water Comparing Off-Line Solid Phase Extraction and Direct Injection LC/MS/MS API 5000 LC/MS/MS System
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report
PRODUCT BULLETIN Cliquid ChemoView 3.0 Software for Routine Screening and Quantitation Cliquid ChemoView 3.0 Software Simple automated analysis, from sample to report KEY FEATURES Secure user login that
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification
Improving the Metabolite Identification Process with Efficiency and Speed: LightSight Software for Metabolite Identification Overview LightSight Software for Metabolite Identification is a complete environment
Authenticity Assessment of Fruit Juices using LC-MS/MS and Metabolomic Data Processing
Authenticity Assessment of Fruit Juices using LC-MS/MS and Metabolomic Data Processing Lukas Vaclavik 1, Ondrej Lacina 1, André Schreiber 2, and Jana Hajslova 1 1 Institute of Chemical Technology, Prague
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
Ultra Fast UHPLC-LCMSMS Method Development in Clinical Drug Monitoring
PO-CON1359E Ultra Fast UHPLC-LCMSMS Method Development in HPLC 2013 MASS-09 Anja Grüning 1 ; Brigitte Richrath 1 ; Klaus Bollig 2 ; Sven Vedder 1 ; Robert Ludwig 1 1 Shimadzu Europa GmbH, Duisburg, Germany;
Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay
PO-CON1484E Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching ASMS 014 TP85 David R. Baker 1, John Warrander 1, Neil Loftus 1, Simon Hird
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
Guide to achieving reliable quantitative LC-MS measurements
Guide to achieving reliable quantitative LC-MS measurements Guide to achieving reliable quantitative LC-MS measurements RSC Analytical Methods Committee Editor Mike Sargent First Edition 2013 Contributors
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
AppNote 9/2013. Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides KEYWORDS ABSTRACT
AppNote 9/2013 Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides Oscar G. Cabrices Gerstel, Inc., 701 Digital Dr. Suite J, Linthicum, MD 21090, USA
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS. Mikaël LEVI Shimadzu France
Sensitive steroid analysis on a new High Speed Triple Quadrupole LC/MS/MS Mikaël LEVI Shimadzu France Welcome Seminar Barcelona 2013 Interest in Steroids assay by LC-MS/MS Steroids measurement is essential
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP and TripleTOF 5600 System Technology Targeted peptide quantification is a rapidly growing application
Waters Core Chromatography Training (2 Days)
2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with the sensitivity and selectivity
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests AB SCIEX TOF/TOF 5800 System with DynamicExit Algorithm and ProteinPilot Software for Robust Protein
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
SCREENING FOR DRUGS IN SERUM AND URINE BY LC/ESI/CID-MS AND MS/MS WITH LIBRARY SEARCHING *
SCREENING FOR DRUGS IN SERUM AND URINE BY LC/ESI/CID-MS AND MS/MS WITH LIBRARY SEARCHING * Wolfgang WEINMANN 1, Natalie LEHMANN 1, Michaela RENZ 1, Alexander WIEDEMANN 1, Michal SVOBODA 2 1 Institute of
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Discovery of Pesticide Protomers Using Routine Ion Mobility Screening
Michael McCullagh, 1 David Eatough, 1 Vincent Hanot, 2 and Séverine Goscinny 2 1 Waters Corporation, Wilmslow, UK 2 Wetenschappelijk Instituut Volksgezondheid Institut Scientifique de Santé Publique, Brussels,
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
Instrumentation. Sample pretreatment
Analysis of Propofol (2,6-Diisopropylphenol) and its Metabolites in One Injection using the DuoSpray Ion Source AB SCIEX QTRAP 500 LC-MS/MS System and Gerstel, Inc. MultiPurpose Sampler Adrian M. Taylor,
High Sensitivity MS Determination of Carryover in a New Autosampler Design
High Sensitivity MS Determination of Carryover in a New Autosampler Design 1 William Hedgepeth*, Sue Steinike Shimadzu Scientific Instruments, Columbia, MD USA Mauro Aiello, Robert Ellis, Andre Schreiber,
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
AB SCIEX TOF/TOF 4800 PLUS SYSTEM. Cost effective flexibility for your core needs
AB SCIEX TOF/TOF 4800 PLUS SYSTEM Cost effective flexibility for your core needs AB SCIEX TOF/TOF 4800 PLUS SYSTEM It s just what you expect from the industry leader. The AB SCIEX 4800 Plus MALDI TOF/TOF
GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005
CAC/GL 56-2005 Page 1 of 6 GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005 CONFIRMATORY TESTS When analyses are
Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software
Application Note # LCMS-62 Walk-Up Ion Trap Mass Spectrometer System in a Multi-User Environment Using Compass OpenAccess Software Abstract Presented here is a case study of a walk-up liquid chromatography
Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform
Utilization of Rapid LC-MS for Screening and Quantitative Analysis of Pesticides in Food Matrix using an Exactive Plus Benchtop Orbitrap Platform Charles Yang, Leo Wang, Dipankar Ghosh, Jonathan Beck,
INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS
AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
Biocides main groups
Analytical challenges for the analysis of biocides in aqueous and solid environmental matrices Arne Wick, Kathrin Broeder, Michael Schluesener, Thomas Ternes Federal Institute of Hydrology (BfG), Koblenz,
MS/MS analysis of Polyphenols
Purdue-UAB Botanicals Center for Age-Related Disease MS/MS analysis of Polyphenols Jeevan Prasain Ph.D. Pharmacology & Toxicology UAB Polyphenols H Caffeic acid Phenolic acids and derivatives H EGC (Flavanol)
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS
Summary: Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS UCT Part Numbers: CUNAX226 - Clean-Up C8+NAX, 2mg/6mL BETA-GLUC- ml Beta-Glucuronidase Enzyme, liquid form SLAQ1ID21-3UM -
A Streamlined Workflow for Untargeted Metabolomics
A Streamlined Workflow for Untargeted Metabolomics Employing XCMS plus, a Simultaneous Data Processing and Metabolite Identification Software Package for Rapid Untargeted Metabolite Screening Baljit K.
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS-1603. No. SSI-GCMS-1603. Shilpi Chopra, Ph.D.
Gas Chromatograph Mass Spectrometer No. GCMS-1603 Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS Shilpi Chopra, Ph.D. Introduction Organophosphorus (OP) pesticides are a class
UPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
Bruker ToxScreener TM A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS ToxScreener - Get the Complete Picture Forensic laboratories are frequently required
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS. Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2
Advances in Steroid Panel Analysis with High Sensitivity LC/MS/MS Kenneth C. Lewis 1, Lisa St. John-Williams 1, Changtong Hao 2, Sha Joshua Ye 2 1 OpAns LLC, Durham, NC, USA; 2 Ionics Mass Spectrometry
Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES
Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES AGILENT ENHANCED MATRIX REMOVAL LIPID ARE LIPIDS WEIGHING YOUR SAMPLES DOWN? Interference from lipids is a problem for labs measuring
Analyst 1.6 Software. Software Reference Guide
Analyst 1.6 Software Software Reference Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment.
amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS
amazon SL Setting New Standards in Performance, Simplicity and Value Innovation with Integrity Ion Trap Best-In-Class Ion Trap Mass Spectrometer for Routine Analysis The amazon SL entry-level system is
ProteinPilot Report for ProteinPilot Software
ProteinPilot Report for ProteinPilot Software Detailed Analysis of Protein Identification / Quantitation Results Automatically Sean L Seymour, Christie Hunter SCIEX, USA Pow erful mass spectrometers like
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
S olid Ph ase Extraction (SP E) Method:
Solid Phase Extraction of US EPA 535 for Chloroacetanilide and Acetamide Degradates in Drinking Water Samples Prior to Liquid Chromatography/ Tandem Mass Spectrometry Toni Hofhine, Horizon Technology,
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model?
What Do We Learn about Hepatotoxicity Using Coumarin-Treated Rat Model? authors M. David Ho 1, Bob Xiong 1, S. Stellar 2, J. Proctor 2, J. Silva 2, H.K. Lim 2, Patrick Bennett 1, and Lily Li 1 Tandem Labs,
How To Develop A Recipe Card
TSNA Method Development Dr Michael Intorp CORESTA Cooperation Center for Scientific Research Relative to Tobacco FDA Center for Tobacco Products Scientific Workshop Tobacco Product Analysis 30 31 July
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System Olaf Scheibner, Maciej Bromirski, Thermo Fisher Scientific, Bremen, Germany
Introduction. Yuan Yuan Zhao 1, Xin Liu 2, Jessica M. Boyd 1, Feng Qin 1, Jianjun Li 2, and Xing-Fang Li 1, * 1. Abstract
Identification of N-Nitrosamines in Treated Drinking Water Using Nanoelectrospray Ionization High-Field Asymmetric Waveform Ion Mobility Spectrometry with Quadrupole Time-of-Flight Mass Spectrometry Yuan
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS
Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS Thorsten Bernsmann, Peter Fürst Chemical and Veterinary Analytical Institute Münsterland-Emscher-Lippe Joseph-König-Str. 40,
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications
TECHNICAL NOTE Toxicology Making the Leap to LC/MS/MS: Enhancing and Accelerating Clinical Research and Forensic Toxicology Applications Introduction The resolving power of chromatography combined with
Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs
Liquid Chromatography Mass Spectrometry SSI-LCMS-068 Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs LCMS-8050 Summary By utilizing the LCMS-8050 s ultrafast scan
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit
Rapid and Reproducible Amino Acid Analysis of Physiological Fluids for Clinical Research Using LC/MS/MS with the atraq Kit Fast, simple and cost effective analysis Many areas of biochemical research and
Analyst 1.6.1 Software
MAX cover image size 4.8 inches X 7.8 inches In Body page mode, draw a graphic frame to match this section and then go to the next step. Delete this text from the Front Cover Master page. View>Master Pages,
Analyst 1.6.1 Software. Advanced User Guide
This document is provided to customers who have purchased AB Sciex equipment to use in the operation of such AB Sciex equipment. This document is copyright protected and any reproduction of this document
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Solid Phase Extraction for Water Analysis. presented by
Solid Phase Extraction for Water Analysis presented by Dr. Bernd Thierfelder Outline 1. Introduction in SPE 2. Retention Mechanisms / Sorbent Selectivity 3. strata -X for Water Analysis 4. Applications
Analysis of Liquid Samples on the Agilent GC-MS
Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report
of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley