1 Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown, Pennsylvania C. W. Magee Evans East, Inc. Plainsboro, New Jersey Fachbereieh Matsrjalwissensphaft cter Techn. Hochschule Darmstadt inv.-nr.: 3-/O New York A Wiley-Interscience Publication JOHN WILEY & SONS Chichester Brisbane Toronto Singapore

2 SUMMARY OF CONTENTS INTRODUCTION The basics of the SIMS technique are reviewed. Microelectronics processing terms used in the text are explained. 1. ANALYSIS CONDITIONS There are analysis parameters of first order and second order importance. The choice of three primary beam characteristics, species, energy, and angle of incidence, influence the selection of sputtering rate, detected area, and energy acceptance window. Knowledge of the capabilities of the instrument will help to decide which secondary ions to monitor. 1.1 Primary Beam The choice of primary ion beam depends on the elements to be analyzed and the type of information required for each element. This choice can be made with the aid of the tables of Relative Sensitivity Factor (RSF) versus element in Appendix E, and detection limit versus element in Appendix F. 1.2 Primary Beam Energy The choice of primary ion beam energy depends on the depth of the ion beam mixing that can be tolerated. Primary ion current and focusing are usually improved with higher primary ion energy. Ion beam mixing versus beam energy is illustrated with a figure showing profile distortion with increasing energy. 1.3 Angle of Incidence The primary ion angle of incidence affects the sputtering yield, the density of the primary species in the surface layer, and depth resolution. Changes in the primary energy can cause changes in the incidence angle for a magnetic sector instrument. The dependence of sputtering yield on incidence angle is shown. 1.4 Sputtering Rate Sputtering rate depends on the bombarding species and the primary current density. Sputtering rate is increased if the primary current is increased or if the rastered area is reduced. The use of a a higher sputtering rate usually improves the detection limit but may degrade depth resolution. Figures demonstrate these two effects. Aspects of primary ion current measurement are discussed. 1.5 Detected Area The detected area should be chosen to be only a small fraction of the crater area to avoid crater sidewall contributions. A very small detected area can be used to avoid defects and lateral structure and to improve depth resolution. A figure showing the effect on a profile of changing the raster size while the detected area remains constant is included.

3 1.6 Species Monitored The choice of impurity species for analysis is usually concerned with detection limit (DL), but also involves sensitivity and dynamic range. The best DL can be found with knowledge of the mass interferences and the RSFs for atomic and molecular ions. The use of rarer isotopes for analysis is shown. Reference is made to the RSF and detection limit appendices. The importance of acquiring mass spectra is demonstrated. The optimum matrix species to be monitored is also discussed. 1.7 End Point The ability to know when to stop a profile is important and depends on the analyst's knowledge of the typical background level, the sputtering rate, and the sample. 1.8 Energy Distribution The secondary ion energy distribution varies with element, with molecular versus atomic ion, and with sample charging during analysis of insulators. The choice of energy acceptance window can affect the accuracy of the profile and the detection limit. 2. PROFILE ISSUES Many characteristics affect SIMS data. Not only the properties of the materials to be analyzed but also the physics of the sputtering process play a role in the quality of the results. 2.1 Ion Beam Mixing and Depth Resolution Ion beam mixing depends on primary beam energy, species, and angle of incidence. Depth resolution is affected by ion beam mixing. Decay length and interface width are discussed. 2.2 Segregation and Charge Driven Diffusion Movement of an analyte species, such as As in S1O2, can occur during ion bombardment because of chemical gradients. Charge driven diffusion can distort the profile of some elements, such as Na. 2.3 Matrix Effects Secondary ion yields can vary with sample composition. The most frequently encountered matrix effects are caused by changes in the oxygen atom density of the sample. 2.4 Surface Effects The native oxide, the equilibration depth, and surface defects produce depth profile aberrations. The equilibration depth can be reduced with analysis at lower primary beam energy, or with the use of an oxygen leak. The equilibration depth can be zero for the analysis of oxides using an oxygen primary beam. Surface defects can be avoided by optical examination of the area to be analyzed. A plot showing the effect of surface deposits on a profile is given.

4 2.5 Particulates Surface or buried particulates can distort a profile. An example is presented along with a discussion of the ability of an image depth profile to detect and correct this problem by use of selective area analysis. 2.6 Crater Shape Crater shape and bottom flatness affect the depth resolution of a profile. Diagrams are included to show profilometer measurements of different types of crater distortions. 2.7 Microtopography Topography changes can occur during ion bombardment. A figure showing a depth profile in Si and SEM micrographs of the topography formation are used. The quantification and depth resolution of the profile are affected. The importance of determining the onset depth of microtopography in a given matrix is discussed. 2.8 Memory Effect The memory effect, especially for a magnetic sector instrument, is discussed in terms of origin and dynamic range limitations. The dynamic range limitation is shown in two ways. The first is the effect of analyzing a sample matrix that is the contaminant species for a subsequent profile (heteromemory). The second is the analysis of a high concentration of a species in a depth profile where a good detection limit for that same species is desired (homomemory). In both cases, the dynamic range is limited; the limit is about 5 orders of magnitude for a magnetic sector instrument. 2.0 Count Rate Saturation Faraday cup (FC) and/or electron multiplier (EM) saturation during a. profile can cause distortion of the profile. Examples are the flattening of the peak of an implant profile or a dip at the peak. Detected area versus rastered area is discussed Sample Location and Mounting The sample position for a magnetic sector instrument can affect the ion transmission through the mass spectrometer. This is illustrated by data from profiles taken at the center and edges of the sample area. The advantage of doing all analyses at one location for highest accuracy is noted. The distance from sample to analyzer is discussed. Sample preparation and mounting techniques are described Mass Interferences Mass interferences are not always easily predicted. Mass spectra are shown that indicate problems that can occur if mass spectra are not taken. The use of high mass resolution and voltage offset techniques to resolve interferences is illustrated using profiles, alon^ with a discussion of the approximate level of sensitivity that is lost. Tlie use o\ Xppeufox G. \s &\scusse<i. 3. QUANTIFICATION 3.1 Procedure The equations for quantification are presented along with a profile of raw data and a quantified profile. Calculation of relative sensitivity factor (RSF) is shown and the use of the RSF data in Appendix E is discussed. '

5 3.2 Calibration Using Ion Implantation Ion implantation criteria for a SIMS calibration standard are discussed. More than one isotope can be implanted provided the fluences of all isotopes can be determined. The choice of the best isotope should be based on interferences and isotope abundance. The energy of the implant should carry the peak and at least some of the front side of the implant profile deeper than the equilibration zone. The optimum fluence requires not implanting a fluence that will cause detection saturation, but enough fluence to allow more than one order of magnitude of detection above background. The accuracy of the implant fluence and RBS measurement of the fluence are treated. 3.3 Systematic Trends in RSFs Patterns of RSF versus ionization potential and RSF versus electron affinity are noted. An example of RSF plots for a compound matrix is presented with reference to either matrix element. RSFs for molecular ions are discussed. A method is given to determine RSFs for an unmeasured matrix. 4. SPECIFIC APPLICATIONS 4.1 Bulk Analysis Bulk analysis can be achieved by sputtering at a high rate. Mass spectra can be used to produce quantitative results with the use of RSF values. 4.2 Metals and Rough Surfaces Analysis of metals has been difficult because of microtopography formation, inclusions, nonuniform chemistry, grains, phases, and polycrystallinity, all of which affect depth resolution. Examples are provided for Al and Au layers. Sample rotation may improve the analysis of metals. Rough surfaces can be analyzed, but the depth resolution is poor. 4.3 Insulators The parameters of interest in the analysis of bulk insulators and insulating layers are discussed. Electron beam neutralization allows the analysis of a wide range of materials. The need to flood the entire ion beam rastered area with electrons is emphasized. Example profiles are used to demonstrate the analysis of.alkali elements in SiO2- High sputtering rate analysis is discussed. Differences between quadrupole and magnetic sector methods are noted. 4.4 Interfaces The methods to measure interface width, determine interface location, and detect interface contamination are discussed. The change in relative yield across an SiO^/Si interface is also treated. 4.5 Multilayers Multilayer analysis requires consideration of the changes in secondary ion yield caused by different matrix compositions, such as GaAs and AlGaAs, along with the changes in sputtering rates. An analysis of a quantum well structure is used as an illustration.

6 4.6 Residual Gas Elements Analysis of residual gas elements (H, C, N, O) with higher sputtering, rate typically produces a better detection limit. The effect of residual gas pressure on the detection limit for these species is illustrated. 4.7 Small Areas The difficulties of analysis of small areas, particularly on patterned wafers, are described. A profile is provided to show what effect occurs if part of the crater is off the pattern. A discussion is given of ion microprobe versus ion microscope solutions to this problem and of image depth profiles. 4.8 Major Elements SIMS profiling of major constituents can supplement what Auger electron spectrometry can provide, but at a faster rate. The use of Cs + bombardment and detection of positive cluster ions is discussed. 5. APPENDICES A. Guidelines for Analysis B. SIMS and Ion Implantation Data for the Elements Table Bl. Naturally Occurring Isotopes Table B2. SIMS and Ion Implantation Data for the Elements Table B3. Ionization Potentials and Electron Affinities Table B4. First Ionization Potentials of Selected Atoms and Molecules C. Matrix Densities D. Relative Sputtering Rates E. Relative Sensitivity Factors (RSFs) Table El. Table E2. Table E3. Table E4. Table E5. Table E6. Table E7. Table E8. Table E9. Table E10. Table Ell. Table E12. Table E13. Table E14. Table E15. A 2 O 3 Diamond GaAs GaP Ge HgCdTe InP LiNbO 3 Si, O 2 +,+;Cs +,- Si,O 2 +,-;Cs\ + Si, Quadrupole Si, O 2 +, Molecular Si, Cs +, Molecijlar SiO2, Quadrupole TaSi 2

7 F. Detection Limits Table Fl: Detection Limits for Si, Ge, GaAs, GaP, InP, and InSb Table F2: Detection Limits for Diamond, SiO 2, LiNbO 3, HgCdTe/CdTe G. Mass Interferences Computer program to calculate molecular ion isotope abundances. H. Mass Spectra Magnetic Sector Quadrupole INDEX

h e l p s y o u C O N T R O L

h e l p s y o u C O N T R O L contamination analysis for compound semiconductors ANALYTICAL SERVICES B u r i e d d e f e c t s, E v a n s A n a l y t i c a l g r o u p h e l p s y o u C O N T R O L C O N T A M I N A T I O N Contamination

More information

Thomas Grehl. Improvement in TOF-SIMS Instrumentation for Analytical Application and Fundamental Research

Thomas Grehl. Improvement in TOF-SIMS Instrumentation for Analytical Application and Fundamental Research Thomas Grehl Improvement in TOF-SIMS Instrumentation for Analytical Application and Fundamental Research 2003 Experimentelle Physik Improvements in TOF-SIMS Instrumentation for Analytical Application

More information

1. Photon Beam Damage and Charging at Solid Surfaces John H. Thomas III

1. Photon Beam Damage and Charging at Solid Surfaces John H. Thomas III 1. Photon Beam Damage and Charging at Solid Surfaces John H. Thomas III 1. Introduction............................. 2. Electrostatic Charging of Samples in Photoemission Experiments............................

More information

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration

Copyright 2007 Casa Software Ltd. www.casaxps.com. ToF Mass Calibration ToF Mass Calibration Essentially, the relationship between the mass m of an ion and the time taken for the ion of a given charge to travel a fixed distance is quadratic in the flight time t. For an ideal

More information

Damage-free, All-dry Via Etch Resist and Residue Removal Processes

Damage-free, All-dry Via Etch Resist and Residue Removal Processes Damage-free, All-dry Via Etch Resist and Residue Removal Processes Nirmal Chaudhary Siemens Components East Fishkill, 1580 Route 52, Bldg. 630-1, Hopewell Junction, NY 12533 Tel: (914)892-9053, Fax: (914)892-9068

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I

The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I 5 The interaction of Cu(100)-Fe surfaces with oxygen studied with photoelectron spectroscopy. I Mg Kα excited photoemission. Abstract The oxidation of Cu(100)-Fe surfaces was studied using XPS. Surfaces

More information

bulk 5. Surface Analysis Why surface Analysis? Introduction Methods: XPS, AES, RBS

bulk 5. Surface Analysis Why surface Analysis? Introduction Methods: XPS, AES, RBS 5. Surface Analysis Introduction Methods: XPS, AES, RBS Autumn 2011 Experimental Methods in Physics Marco Cantoni Why surface Analysis? Bulk: structural function Electrical/thermal conduction Volume increases

More information

Vacuum Evaporation Recap

Vacuum Evaporation Recap Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.

More information

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Surface Characterization - Marie-Geneviève Barthés-Labrousse SURFACE CHARACTERIZATION Marie-Geneviève Centre d Etudes de Chimie Métallurgique, CNRS, Vitry-sur-Seine, France Keywords: Surface Analysis, Surface imaging, Surface composition, Surface chemical analysis,

More information

Solar Photovoltaic (PV) Cells

Solar Photovoltaic (PV) Cells Solar Photovoltaic (PV) Cells A supplement topic to: Mi ti l S Micro-optical Sensors - A MEMS for electric power generation Science of Silicon PV Cells Scientific base for solar PV electric power generation

More information

Preface Light Microscopy X-ray Diffraction Methods

Preface Light Microscopy X-ray Diffraction Methods Preface xi 1 Light Microscopy 1 1.1 Optical Principles 1 1.1.1 Image Formation 1 1.1.2 Resolution 3 1.1.3 Depth of Field 5 1.1.4 Aberrations 6 1.2 Instrumentation 8 1.2.1 Illumination System 9 1.2.2 Objective

More information

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially

More information

SIMS DEPTH PROFILING OF THIN NITRIDE- AND CARBIDE-BASED FILMS FOR HARD COATING

SIMS DEPTH PROFILING OF THIN NITRIDE- AND CARBIDE-BASED FILMS FOR HARD COATING SIMS Rev.Adv.Mater.Sci. depth profiling 15(2007) of thin nitride 253258 and carbidebased films for hard coating 253 SIMS DEPTH PROFILING OF THIN NITRIDE AND CARBIDEBASED FILMS FOR HARD COATING M. Cwil

More information

A Plasma Doping Process for 3D FinFET Source/ Drain Extensions

A Plasma Doping Process for 3D FinFET Source/ Drain Extensions A Plasma Doping Process for 3D FinFET Source/ Drain Extensions JTG 2014 Cuiyang Wang*, Shan Tang, Harold Persing, Bingxi Wood, Helen Maynard, Siamak Salimian, and Adam Brand Cuiyang_wang@amat.com Varian

More information

Ion Beam Sputtering: Practical Applications to Electron Microscopy

Ion Beam Sputtering: Practical Applications to Electron Microscopy Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a

More information

Lecture 11. Etching Techniques Reading: Chapter 11. ECE 6450 - Dr. Alan Doolittle

Lecture 11. Etching Techniques Reading: Chapter 11. ECE 6450 - Dr. Alan Doolittle Lecture 11 Etching Techniques Reading: Chapter 11 Etching Techniques Characterized by: 1.) Etch rate (A/minute) 2.) Selectivity: S=etch rate material 1 / etch rate material 2 is said to have a selectivity

More information

X-ray diffraction techniques for thin films

X-ray diffraction techniques for thin films X-ray diffraction techniques for thin films Rigaku Corporation Application Laboratory Takayuki Konya 1 Today s contents (PM) Introduction X-ray diffraction method Out-of-Plane In-Plane Pole figure Reciprocal

More information

Scanning Electron Microscopy Services for Pharmaceutical Manufacturers

Scanning Electron Microscopy Services for Pharmaceutical Manufacturers Scanning Electron Microscopy Services for Pharmaceutical Manufacturers Author: Gary Brake, Marketing Manager Date: August 1, 2013 Analytical Testing Laboratory www.atl.semtechsolutions.com Scanning Electron

More information

Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?

Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common? Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction

More information

Chapter 1: Moles and equations. Learning outcomes. you should be able to:

Chapter 1: Moles and equations. Learning outcomes. you should be able to: Chapter 1: Moles and equations 1 Learning outcomes you should be able to: define and use the terms: relative atomic mass, isotopic mass and formula mass based on the 12 C scale perform calculations, including

More information

Introduction to X-Ray Powder Diffraction Data Analysis

Introduction to X-Ray Powder Diffraction Data Analysis Introduction to X-Ray Powder Diffraction Data Analysis Center for Materials Science and Engineering at MIT http://prism.mit.edu/xray An X-ray diffraction pattern is a plot of the intensity of X-rays scattered

More information

Chemical Sputtering. von Kohlenstoff durch Wasserstoff. W. Jacob

Chemical Sputtering. von Kohlenstoff durch Wasserstoff. W. Jacob Chemical Sputtering von Kohlenstoff durch Wasserstoff W. Jacob Centre for Interdisciplinary Plasma Science Max-Planck-Institut für Plasmaphysik, 85748 Garching Content: Definitions: Chemical erosion, physical

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

Molecular Spectroscopy

Molecular Spectroscopy Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended

More information

SIMS AND SCANNING ION MICROSCOPY

SIMS AND SCANNING ION MICROSCOPY SIMS AND SCANNING ION MICROSCOPY G. Allen, I. Brown To cite this version: G. Allen, I. Brown. SIMS AND SCANNING ION MICROSCOPY. Journal de Physique Colloques, 1989, 50 (C2), pp.c2-121-c2-125. .

More information

Sample Analysis Design Isotope Dilution

Sample Analysis Design Isotope Dilution Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with

More information

Mass Spectrometry. Overview

Mass Spectrometry. Overview Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2

More information

Coating Technology: Evaporation Vs Sputtering

Coating Technology: Evaporation Vs Sputtering Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information

More information

LC-MS/MS for Chromatographers

LC-MS/MS for Chromatographers LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,

More information

L. Yu et al. Correspondence to: Q. Zhang (dkwzhang@ucdavis.edu)

L. Yu et al. Correspondence to: Q. Zhang (dkwzhang@ucdavis.edu) Supplement of Atmos. Chem. Phys. Discuss., 5, 967 974, 5 http://www.atmos-chem-phys-discuss.net/5/967/5/ doi:.594/acpd-5-967-5-supplement Author(s) 5. CC Attribution. License. Supplement of Molecular transformations

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry

An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry Limits to Detection/Measurement Ionization is fundamentally a Probabilistic Process Just like radioactive decay So is transmission through the analyzer There is an intrinsic statistical uncertainty Proportional

More information

Lecture 6 Scanning Tunneling Microscopy (STM) General components of STM; Tunneling current; Feedback system; Tip --- the probe.

Lecture 6 Scanning Tunneling Microscopy (STM) General components of STM; Tunneling current; Feedback system; Tip --- the probe. Lecture 6 Scanning Tunneling Microscopy (STM) General components of STM; Tunneling current; Feedback system; Tip --- the probe. Brief Overview of STM Inventors of STM The Nobel Prize in Physics 1986 Nobel

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

4.2 Bias, Standards and Standardization

4.2 Bias, Standards and Standardization 4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard

More information

TOF FUNDAMENTALS TUTORIAL

TOF FUNDAMENTALS TUTORIAL TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] info@rmjordan.com [e-mail] This

More information

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy

An Innovative Method for Dead Time Correction in Nuclear Spectroscopy An Innovative Method for Dead Time Correction in Nuclear Spectroscopy Upp, Daniel L.; Keyser, Ronald M.; Gedcke, Dale A.; Twomey, Timothy R.; and Bingham, Russell D. PerkinElmer Instruments, Inc. ORTEC,

More information

F321 THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important are... in the nucleus of an atom

F321 THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important are... in the nucleus of an atom Atomic Structure F32 TE STRUCTURE OF ATOMS ATOMS Atoms consist of a number of fundamental particles, the most important are... Mass / kg Charge / C Relative mass Relative Charge PROTON NEUTRON ELECTRON

More information

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW

PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental

More information

Nanoelectronics 09. Atsufumi Hirohata Department of Electronics. Quick Review over the Last Lecture

Nanoelectronics 09. Atsufumi Hirohata Department of Electronics. Quick Review over the Last Lecture Nanoelectronics 09 Atsufumi Hirohata Department of Electronics 12:00 Wednesday, 4/February/2015 (P/L 006) Quick Review over the Last Lecture ( Field effect transistor (FET) ): ( Drain ) current increases

More information

Amount of Substance. http://www.avogadro.co.uk/definitions/elemcompmix.htm

Amount of Substance. http://www.avogadro.co.uk/definitions/elemcompmix.htm Page 1 of 14 Amount of Substance Key terms in this chapter are: Element Compound Mixture Atom Molecule Ion Relative Atomic Mass Avogadro constant Mole Isotope Relative Isotopic Mass Relative Molecular

More information

Silicon-On-Glass MEMS. Design. Handbook

Silicon-On-Glass MEMS. Design. Handbook Silicon-On-Glass MEMS Design Handbook A Process Module for a Multi-User Service Program A Michigan Nanofabrication Facility process at the University of Michigan March 2007 TABLE OF CONTENTS Chapter 1...

More information

Types of Epitaxy. Homoepitaxy. Heteroepitaxy

Types of Epitaxy. Homoepitaxy. Heteroepitaxy Epitaxy Epitaxial Growth Epitaxy means the growth of a single crystal film on top of a crystalline substrate. For most thin film applications (hard and soft coatings, optical coatings, protective coatings)

More information

Lecture 12. Physical Vapor Deposition: Evaporation and Sputtering Reading: Chapter 12. ECE 6450 - Dr. Alan Doolittle

Lecture 12. Physical Vapor Deposition: Evaporation and Sputtering Reading: Chapter 12. ECE 6450 - Dr. Alan Doolittle Lecture 12 Physical Vapor Deposition: Evaporation and Sputtering Reading: Chapter 12 Evaporation and Sputtering (Metalization) Evaporation For all devices, there is a need to go from semiconductor to metal.

More information

2. Deposition process

2. Deposition process Properties of optical thin films produced by reactive low voltage ion plating (RLVIP) Antje Hallbauer Thin Film Technology Institute of Ion Physics & Applied Physics University of Innsbruck Investigations

More information

Electron spectroscopy Lecture 1-21. Kai M. Siegbahn (1918 - ) Nobel Price 1981 High resolution Electron Spectroscopy

Electron spectroscopy Lecture 1-21. Kai M. Siegbahn (1918 - ) Nobel Price 1981 High resolution Electron Spectroscopy Electron spectroscopy Lecture 1-21 Kai M. Siegbahn (1918 - ) Nobel Price 1981 High resolution Electron Spectroscopy 653: Electron Spectroscopy urse structure cture 1. Introduction to electron spectroscopies

More information

Edited by. C'unter. and David S. Moore. Gauglitz. Handbook of Spectroscopy. Second, Enlarged Edition. Volume 4. WlLEY-VCH. VerlagCmbH & Co.

Edited by. C'unter. and David S. Moore. Gauglitz. Handbook of Spectroscopy. Second, Enlarged Edition. Volume 4. WlLEY-VCH. VerlagCmbH & Co. Edited by C'unter Gauglitz and David S. Moore Handbook of Spectroscopy Second, Enlarged Edition Volume 4 WlLEY-VCH VerlagCmbH & Co. KGaA IX Volume 4 Section XII Applications 6: Spectroscopy at Surfaces

More information

Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM

Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM Original Paper Lateral Resolution of EDX Analysis with Low Acceleration Voltage SEM Satoshi Hashimoto 1, Tsuguo Sakurada 1, and Minoru Suzuki 2 1 JFE-Techno research corporation, 1-1 Minamiwatarida, Kawasaki,

More information

Tutorial 4.6 Gamma Spectrum Analysis

Tutorial 4.6 Gamma Spectrum Analysis Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning

More information

BNG 331 Cell-Tissue Material Interactions. Biomaterial Surfaces

BNG 331 Cell-Tissue Material Interactions. Biomaterial Surfaces BNG 331 Cell-Tissue Material Interactions Biomaterial Surfaces Course update Updated syllabus Homework 4 due today LBL 5 Friday Schedule for today: Chapter 8 Biomaterial surface characterization Surface

More information

The Mole Concept. The Mole. Masses of molecules

The Mole Concept. The Mole. Masses of molecules The Mole Concept Ron Robertson r2 c:\files\courses\1110-20\2010 final slides for web\mole concept.docx The Mole The mole is a unit of measurement equal to 6.022 x 10 23 things (to 4 sf) just like there

More information

EXPERIMENTAL STUDY OF STRUCTURAL ZONE MODEL FOR COMPOSITE THIN FILMS IN MAGNETIC RECORDING MEDIA APPLICATION

EXPERIMENTAL STUDY OF STRUCTURAL ZONE MODEL FOR COMPOSITE THIN FILMS IN MAGNETIC RECORDING MEDIA APPLICATION EXPERIMENTAL STUDY OF STRUCTURAL ZONE MODEL FOR COMPOSITE THIN FILMS IN MAGNETIC RECORDING MEDIA APPLICATION Hua Yuan and David E. Laughlin Department of Materials Science and Engineering, Carnegie Mellon

More information

Atomic Force Microscopy. Long Phan Nanotechnology Summer Series May 15, 2013

Atomic Force Microscopy. Long Phan Nanotechnology Summer Series May 15, 2013 Atomic Force Microscopy Long Phan Nanotechnology Summer Series May 15, 2013 1 World s Smallest Movie 2 Outline What is AFM? How does AFM Work? 3 Modes: Contact mode Non contact mode Tapping mode Imaging

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

Contamination. Cleanroom. Cleanroom for micro and nano fabrication. Particle Contamination and Yield in Semiconductors.

Contamination. Cleanroom. Cleanroom for micro and nano fabrication. Particle Contamination and Yield in Semiconductors. Fe Particles Metallic contaminants Organic contaminants Surface roughness Au Particles SiO 2 or other thin films Contamination Na Cu Photoresist Interconnect Metal N, P Damages: Oxide breakdown, metal

More information

Sputtering by Particle Bombardment I

Sputtering by Particle Bombardment I Sputtering by Particle Bombardment I Physical Sputtering of Single-Element Solids Edited by R. Behrisch With Contributions by H. H. Andersen H. L. Bay R. Behrisch M. T. Robinson H. E. Roosendaal P. Sigmund

More information

Atoms, Ions and Molecules The Building Blocks of Matter

Atoms, Ions and Molecules The Building Blocks of Matter Atoms, Ions and Molecules The Building Blocks of Matter Chapter 2 1 Chapter Outline 2.1 The Rutherford Model of Atomic Structure 2.2 Nuclides and Their Symbols 2.3 Navigating the Periodic Table 2.4 The

More information

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT

NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published

More information

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,

More information

Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator

Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator R. Gupta and M.Harrison, Brookhaven National Laboratory A. Zeller, Michigan State University

More information

2 Absorbing Solar Energy

2 Absorbing Solar Energy 2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could

More information

Solid State Detectors = Semi-Conductor based Detectors

Solid State Detectors = Semi-Conductor based Detectors Solid State Detectors = Semi-Conductor based Detectors Materials and their properties Energy bands and electronic structure Charge transport and conductivity Boundaries: the p-n junction Charge collection

More information

Using Technology Equipment to Teach Chemistry Laboratory Exercises in Community Colleges

Using Technology Equipment to Teach Chemistry Laboratory Exercises in Community Colleges Using Technology Equipment to Teach Chemistry Laboratory Exercises in Community Colleges Marilyn Barger, Elizabeth McCullough, Kathleen Carvalho Hillsborough Community College/ Hillsborough Community College

More information

X Ray Flourescence (XRF)

X Ray Flourescence (XRF) X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose

More information

Optical and Auger Microanalyses of Solder Adhesion Failures in Printed Circuit Boards

Optical and Auger Microanalyses of Solder Adhesion Failures in Printed Circuit Boards Optical and Auger Microanalyses of Solder Adhesion Failures in Printed Circuit Boards K. Kumar and A. Moscaritolo The Charles Stark Draper Laboratory, Incorporated, Cambridge, Massachusetts 02139 ABSTRACT

More information

Principles of Ion Implant

Principles of Ion Implant Principles of Ion Implant Generation of ions dopant gas containing desired species BF 3, B 2 H 6, PH 3, AsH 3, AsF 5 plasma provides positive ions (B 11 ) +, BF 2+, (P 31 ) +, (P 31 ) ++ Ion Extraction

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

STM, LEED and Mass spectrometry

STM, LEED and Mass spectrometry STM, LEED and Mass spectrometry R. Schloderer, S. Griessl, J. Freund, M. Edelwirth, W.M. Heckl Introduction TDS UHV technique Preparation STM LEED QMS Concept of new UHV chamber Conclusion P. Cole, M.

More information

THERMO NORAN SYSTEM SIX ENERGY DISPERSIVE X- RAY SPECTROMETER. Insert Nickname Here. Operating Instructions

THERMO NORAN SYSTEM SIX ENERGY DISPERSIVE X- RAY SPECTROMETER. Insert Nickname Here. Operating Instructions THERMO NORAN SYSTEM SIX ENERGY DISPERSIVE X- RAY SPECTROMETER Insert Nickname Here Operating Instructions Table of Contents 1 INTRODUCTION Safety 1 Samples 1 2 BACKGROUND Background Information 3 References

More information

Module 7 Wet and Dry Etching. Class Notes

Module 7 Wet and Dry Etching. Class Notes Module 7 Wet and Dry Etching Class Notes 1. Introduction Etching techniques are commonly used in the fabrication processes of semiconductor devices to remove selected layers for the purposes of pattern

More information

III. Wet and Dry Etching

III. Wet and Dry Etching III. Wet and Dry Etching Method Environment and Equipment Advantage Disadvantage Directionality Wet Chemical Solutions Atmosphere, Bath 1) Low cost, easy to implement 2) High etching rate 3) Good selectivity

More information

Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope

Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope andras@nist.gov Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope Bin Ming, András E. Vladár and Michael T. Postek National Institute of Standards and Technology

More information

SILA Sistema Integrato di Laboratori per l Ambiente. CENTRE FOR MICROSCOPY AND MICROANALYSIS Scientific coordinator: Prof.ssa Rosanna De Rosa

SILA Sistema Integrato di Laboratori per l Ambiente. CENTRE FOR MICROSCOPY AND MICROANALYSIS Scientific coordinator: Prof.ssa Rosanna De Rosa CENTRE FOR MICROSCOPY AND MICROANALYSIS Scientific coordinator: Prof.ssa Rosanna De Rosa 0 The Centre for Microscopy and Microanalysis (CM2) is an interdisciplinary service centre, a comprehensive suite

More information

View of ΣIGMA TM (Ref. 1)

View of ΣIGMA TM (Ref. 1) Overview of the FESEM system 1. Electron optical column 2. Specimen chamber 3. EDS detector [Electron Dispersive Spectroscopy] 4. Monitors 5. BSD (Back scatter detector) 6. Personal Computer 7. ON/STANDBY/OFF

More information

Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe. Keywords: Carbon Nanotube, Scanning Probe Microscope

Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe. Keywords: Carbon Nanotube, Scanning Probe Microscope International Journal of Arts and Sciences 3(1): 18-26 (2009) CD-ROM. ISSN: 1944-6934 InternationalJournal.org Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe Bedri Onur Kucukyildirim,

More information

Getting the most from this book...4 About this book...5

Getting the most from this book...4 About this book...5 Contents Getting the most from this book...4 About this book....5 Content Guidance Topic 1 Atomic structure and the periodic table...8 Topic 2 Bonding and structure...14 Topic 2A Bonding....14 Topic 2B

More information

STANDARD OPERATING PROCEDURES

STANDARD OPERATING PROCEDURES PAGE: 1 of 33 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS

More information

How To Use Gc-Ms

How To Use Gc-Ms The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter

More information

Wafer Manufacturing. Reading Assignments: Plummer, Chap 3.1~3.4

Wafer Manufacturing. Reading Assignments: Plummer, Chap 3.1~3.4 Wafer Manufacturing Reading Assignments: Plummer, Chap 3.1~3.4 1 Periodic Table Roman letters give valence of the Elements 2 Why Silicon? First transistor, Shockley, Bardeen, Brattain1947 Made by Germanium

More information

Nanotechnologies for the Integrated Circuits

Nanotechnologies for the Integrated Circuits Nanotechnologies for the Integrated Circuits September 23, 2015 Dr. Bertrand Cambou Professor of Practice NAU, Cybersecurity School of Informatics, Computing, and Cyber-Systems Agenda The Market Silicon

More information

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Carlton G. Slough TA Instruments, 109 Lukens Drive, New Castle DE 19720, USA ABSTRACT The use of

More information

Analyzing Small Molecules by EI and GC-MS. July 2014

Analyzing Small Molecules by EI and GC-MS. July 2014 Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,

More information

Chapter Test B. Chapter: Measurements and Calculations

Chapter Test B. Chapter: Measurements and Calculations Assessment Chapter Test B Chapter: Measurements and Calculations PART I In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1.

More information

ON-LINE ANALYSIS - WHAT HAS BEEN DONE? Mr P de Waal Blue Cube Systems (Pty) Ltd

ON-LINE ANALYSIS - WHAT HAS BEEN DONE? Mr P de Waal Blue Cube Systems (Pty) Ltd Pieter de Waa! ON-LINE ANALYSIS - WHAT HAS BEEN DONE? Mr P de Waal Blue Cube Systems (Pty) Ltd Synopsis: The beneficiation equipment under the control of plant metallurgists exploits the differences in

More information

STATE UNIVERSITY OF NEW YORK COLLEGE OF TECHNOLOGY CANTON, NEW YORK COURSE OUTLINE CHEM 150 - COLLEGE CHEMISTRY I

STATE UNIVERSITY OF NEW YORK COLLEGE OF TECHNOLOGY CANTON, NEW YORK COURSE OUTLINE CHEM 150 - COLLEGE CHEMISTRY I STATE UNIVERSITY OF NEW YORK COLLEGE OF TECHNOLOGY CANTON, NEW YORK COURSE OUTLINE CHEM 150 - COLLEGE CHEMISTRY I PREPARED BY: NICOLE HELDT SCHOOL OF SCIENCE, HEALTH, AND PROFESSIONAL STUDIES SCIENCE DEPARTMENT

More information

Graduate Student Presentations

Graduate Student Presentations Graduate Student Presentations Dang, Huong Chip packaging March 27 Call, Nathan Thin film transistors/ liquid crystal displays April 4 Feldman, Ari Optical computing April 11 Guerassio, Ian Self-assembly

More information

Simultaneous data fitting in ARXPS

Simultaneous data fitting in ARXPS NOTE Please find the following more updated document Resolving overlapping peaks in ARXPS data: The effect of noise and fitting method. J. Muñoz-Flores, A. Herrera-Gomez, J. Electron Spectrosc. Relat.

More information

Preprocessing, Management, and Analysis of Mass Spectrometry Proteomics Data

Preprocessing, Management, and Analysis of Mass Spectrometry Proteomics Data Preprocessing, Management, and Analysis of Mass Spectrometry Proteomics Data M. Cannataro, P. H. Guzzi, T. Mazza, and P. Veltri Università Magna Græcia di Catanzaro, Italy 1 Introduction Mass Spectrometry

More information

Dry Etching and Reactive Ion Etching (RIE)

Dry Etching and Reactive Ion Etching (RIE) Dry Etching and Reactive Ion Etching (RIE) MEMS 5611 Feb 19 th 2013 Shengkui Gao Contents refer slides from UC Berkeley, Georgia Tech., KU, etc. (see reference) 1 Contents Etching and its terminologies

More information

Use the BET (after Brunauer, Emmett and Teller) equation is used to give specific surface area from the adsorption

Use the BET (after Brunauer, Emmett and Teller) equation is used to give specific surface area from the adsorption Number of moles of N 2 in 0.129dm 3 = 0.129/22.4 = 5.76 X 10-3 moles of N 2 gas Module 8 : Surface Chemistry Objectives Lecture 37 : Surface Characterization Techniques After studying this lecture, you

More information

Conductivity of silicon can be changed several orders of magnitude by introducing impurity atoms in silicon crystal lattice.

Conductivity of silicon can be changed several orders of magnitude by introducing impurity atoms in silicon crystal lattice. CMOS Processing Technology Silicon: a semiconductor with resistance between that of conductor and an insulator. Conductivity of silicon can be changed several orders of magnitude by introducing impurity

More information

Atoms and Elements. Outline Atoms Orbitals and Energy Levels Periodic Properties Homework

Atoms and Elements. Outline Atoms Orbitals and Energy Levels Periodic Properties Homework Atoms and the Periodic Table The very hot early universe was a plasma with cationic nuclei separated from negatively charged electrons. Plasmas exist today where the energy of the particles is very high,

More information

QGA Quantitative Gas Analyser

QGA Quantitative Gas Analyser QGA Quantitative Gas Analyser A compact bench-top system for real-time gas and vapour analysis Detailed product information / introduction catalysis studies environmental gas analysis fermentation off-gas

More information

Titan: The Solar System s Abiotic Petroleum Factory

Titan: The Solar System s Abiotic Petroleum Factory Titan: The Solar System s Abiotic Petroleum Factory J. Hunter Waite, Ph.D. Institute Scientist Space Science & Engineering Division Southwest Research Institute Titan: The Solar System s Abiotic Petroleum

More information

13C NMR Spectroscopy

13C NMR Spectroscopy 13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number

More information

EDXRF of Used Automotive Catalytic Converters

EDXRF of Used Automotive Catalytic Converters EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,

More information

Bruce B. Weiner, Walther W. Tscharnuter, David Fairhurst Brookhaven Instruments Corporation Holtsville, NY 11742 US

Bruce B. Weiner, Walther W. Tscharnuter, David Fairhurst Brookhaven Instruments Corporation Holtsville, NY 11742 US Zeta Potential: A New Approach by Bruce B. Weiner, Walther W. Tscharnuter, David Fairhurst Brookhaven Instruments Corporation Holtsville, NY 11742 US A paper presented at the Canadian Mineral Analysts

More information

Spectroscopic Ellipsometry:

Spectroscopic Ellipsometry: Spectroscopic : What it is, what it will do, and what it won t do by Harland G. Tompkins Introduction Fundamentals Anatomy of an ellipsometric spectrum Analysis of an ellipsometric spectrum What you can

More information

Waters Core Chromatography Training (2 Days)

Waters Core Chromatography Training (2 Days) 2015 Page 2 Waters Core Chromatography Training (2 Days) The learning objective of this two day course is to teach the user core chromatography, system and software fundamentals in an outcomes based approach.

More information