GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
|
|
|
- Robert Haynes
- 10 years ago
- Views:
Transcription
1 ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas JSC, 8104 Bourgas, Bulgaria SUMMARY GC FID methods using two different capillary columns have been developed for the determination of benzene in gasoline samples and petroleum fractions. The methods were compared and evaluated. GC FID with a PONA column is not appropriate for routine analysis. GC FID with a TCEP column enables precise and accurate measurement of the benzene content of gasoline and petroleum fractions and can be widely used for analysis of petroleum products there is no limitation of sample composition. Results obtained from analysis of commercial samples of gasoline and petroleum fractions agree with those obtained by use of the IR spectroscopy reference method. INTRODUCTION Gasoline samples and petroleum fractions are very complex mixtures containing many hundreds of different components, of which benzene is by far the most harmful. Reduction of benzene levels in gasoline with a view toward significantly reducing damage to vehicular pollution control devices is an ongoing goal of the environmental community. European Community laws specify a maximum tolerated level of 1% (v/v) for this compound in gasoline [1]. Analytical techniques for benzene have advanced significantly in recent years, and development continues. Modern technologies include gas chromatography (GC), mass spectrometry (MS), infrared (IR), ultraviolet (UV), and fluorescence spectroscopy, and combined techniques, such as GC MS and GC Fourier transform infrared (FTIR), etc. [2 11]. Among these techniques capillary GC with flame ionization detection (FID) and capillary GC MS are most important and most widely used for analysis of benzene in gasoline [4 7,10,11]
2 The American Society for Testing and Materials (ASTM) developed a series of test methods specifically for determination of benzene in gasoline: Method D3606 for determination of benzene and toluene in finished motor and aviation gasoline by gas chromatography; Method D4420 for aromatics in finished gasoline by gas chromatography; Method D5580 for determination of benzene, toluene, ethylbenzene, p/m-xylene, o-xylene, C 9 and heavier aromatics and total aromatics in finished gasoline by gas chromatography; and Method D5769 for determination of benzene, toluene, and total aromatics in finished gasoline by gas chromatography mass spectrometry. ASTM standard test methods employ packed columns, capillary columns, or multiple column systems connected with valves, with the chromatograph coupled with thermal conductivity, hydrogen flame ionization, or mass-selective detection. GC MS methods are seldom used in modern laboratories, because they are labor-intensive, although they are often regarded as reference methods. A brief summary of ASTM, IP, and EN approved methods for benzene analysis is presented in Table I. Each method has its advantages and drawbacks. The need to design a method for the determination of benzene in gasoline was a consequence of two considerations the ASTM, EN, and IP methods could not be used in our laboratories, because of the particular instrumental requirements, and an easily applicable method was needed for determination of benzene in gasoline and petroleum fractions. This paper gives preliminary results of an investigation of the possibility of using gas chromatography for this purpose. The aim of this work was to demonstrate a new, simple, selective and accurate gas chromatographic method for determination of the benzene content of gasoline and petroleum fractions. The purpose of evaluation of the benzene content was to provide the petrochemical engineer with the information required for process control operation and product quality assurance. EXPERIMENTAL Reagents and Calibration Solutions Benzene, isooctane, and 2-butanone (all 99.5% purity) were obtained from Fluka. Gasoline samples and petroleum fractions, each with different PONA values were obtained from Lukoil Neftochim Bourgas JSC. Five calibration solutions of benzene in isooctane, with and without internal standard, were prepared. The approximate concentrations of benzene were 0.1, 0.5, 1.0, 1.5, and 2.0% (v/v). 2-Butanone was used as internal standard
3 Table I Test methods for determination of benzene in gasoline Method Technique Scope Repeatability Reproducibility 0.1 5%(v/v) for Range ASTM GC TCD finished motor and % 0.03(X) (X) aviation gasoline (v/v) ASTM 4420 ASTM 5580 ASTM 5769 IP 425 EN GC TCD GC FID GC MS 0.1 5%(v/v) for finished motor gasoline 0.1 5%(v/v) for finished motor gasoline 0.1 4% (v/v) for finished motor gasoline; gasoline containing oxygenated blending components GC FID % (v/v) GC FID % (v/v) unleaded petrol measured value Range % (v/v) Range % (v/v) Concentration Range % (v/v) % (v/v) % (v/v) Chromatographic Systems and Conditions measu -red value (X 0.65 ) (X 0.65 ) 0.046(X 0.67 ) 0.221(X 0.67 ) Results were obtained by use of two different gas chromatographs. The first was a Hewlett Packard 5890 series II model gas chromatograph (GC) equipped with a split/splitless injector, a flame ionization detector, and a 50 m 0.2 mm i.d. PONA column coated with crosslinked polydimethylsiloxane gum. The injector temperature was 200 C and the detector temperature 250 C. The GC oven temperature was programmed from 35 C at 2 min 1 to 200 C which was held for 10 min. Hydrogen was used as the carrier gas at a flow rate of 1.0 ml min 1 ; the split ratio was 1:
4 The second GC system was a Hewlett Packard model 5890 series II GC with flame ionization detector and a 50 m 0.25 mm i.d. column coated with a 0.4 µm film of CP-TCEP (Varian). The detector and injector temperatures were 275 and 250 C, respectively, and the oven temperature was maintained at 50 C for 10 min then programmed 10 min 1 to a final temperature of 115 C which was held for 10 min. Hydrogen was used as carrier gas at a flow rate of 1.0 ml min 1, the split ratio was 1:60. RESULTS AND DISCUSSION PONA Analyses This method was developed to enable determination of the concentration of benzene in gasoline. Fig. 1 depicts the separation of benzene and the other components of gasoline. Benzene was identified by GC MS. The compound coeluting with benzene, 1-methylcyclopentene, was also identified by GC MS (Fig. 2). No attempt was made to identify the other components of gasoline. As is apparent from Fig. 1, benzene and 1-methylcyclopentene are not resolved completely to baseline. For this purpose the experimental conditions were varied systematically, using the same capillary column. Isothermal chromatography was performed for 15 min at 35 C and the resolution R s was calculated by use of the equation: R s t = 2 w R2 R1 t R1 + w R2 where t R1 and t R2 are the retention times of 1-methylcyclopentene and benzene and w R1 and w R2 are the widths of the peaks at half height. The resolution R was In programmed temperature GC the temperature gradient was changed and the resolution, R, was These separation conditions were used for direct determination of benzene in gasoline samples. Quantitative results were obtained from measured areas of the recorded benzene peak, using a response factor obtained from analysis of a blend of known benzene content. The response factor for benzene was measured at a variety of concentrations. The data obtained are given in Table II. The values provided are averages from three analyses. Relative standard deviations from these replicate injections were all small (in the range %). The response factor was independent of concentration. Table III shows results obtained from analyses of three blends containing known amounts of benzene in isooctane, performed by a single
5 Fig. 1 Partial chromatogram obtained from a gasoline sample on a PONA column operator. Each blend was analyzed five times. The blends contained 0.1, 0.5, and 1.5 (v/v) benzene and were prepared separately immediately before analysis. Thus the results in Table III include variations caused by sample
6 Fig. 2 Partial total-ion chromatogram obtained from a gasoline sample, and mass spectra of peaks: 1, methylcyclopentane; 2, benzene preparation and chromatographic analysis. Average values agreed almost exactly with the prepared concentrations. Relative standard deviations were 2.1% or less
7 Table II Response factors of benzene on PONA and on TCEP columns Concn, % (v/v) Run, N = Response factors of benzene on PONA column Ave response factor Standard deviation Relative standard deviation (%) Run, N = Response factors of benzene on TCEP column Ave response factor Standard deviation Relative standard deviation (%) Table III Precision of PONA and TCEP analysis PONA analysis TCEP analysis Standard de- standard Average Recodard de- standard Relative Stan- Relative Preparetabilitred (%, bility Repea- Prepa- Repeata Average Recovery (%) viation devia- (%, v/v) very (%) viation devia- (%, v/v) (%, v/v) (%, v/v) v/v) (%, v/v) (%, v/v) tion (%) (%, v/v) tion (%)
8 TCEP Analyses GC analysis is a rapid means of direct measurement of the benzene content of gasoline. Separation of the benzene in gasoline is depicted in Fig. 3, which shows that the proposed internal standard is, indeed, suitable for quantitative analysis. It is apparent that benzene is well separated into an individual peak. Analysis time is of the order of 26 min. Fig. 3 Gas chromatogram obtained from a gasoline sample on a TCEP column Before sample analysis the instrument was calibrated using standard solutions of benzene and 2-butanone in isooctane. Response factors for benzene were calculated relative to the internal standard. Response factors determined for quantitation of benzene by GC FID are presented in Table II. Calibration was performed by analysis of four concentration levels, each repeated five times. Relative standard deviations for analysis of the calibration mixtures were in the range %. The linearity of the method was confirmed by analysis of blends of benzene and 2-butanone in isooctane. Each sample was analyzed five times. Table III shows the results obtained. Average recovery of benzene
9 was 100%. Relative standard deviations were 1.8% or less. Precision (repeatability) was determined by use of the standard deviation calculated for five replicate determinations, using the equation: Repeatability = sd 2 Precision was %. The procedure developed in our laboratory is rapid, accurate, and highly precise. Recovery, standard deviations for replicate analyses, and repeatability are typical of those obtained for established petroleum GC analyses and satisfactory for routine use of the method for laboratory quality control. Analysis of Real Samples Ten samples of different commercial gasolines and petroleum fractions were analyzed. The petroleum fractions were obtained from two different technology installations. All samples consisted of different amounts of paraffin, olefin, naphthene, and aromatic hydrocarbons. The 1- methylcyclopentene content of the samples analyzed was also different, and was in range %. Two samples (petroleum fractions from a catalytic cracking process) seemed to contain unusually large amounts of 1-methylcyclopentene 0.80%. Data from data analysis on the GC-PONA and GC-TCEP columns are summarized in Table IV, which also includes values found by IR spectrometry. It is apparent that values from the proposed Table IV Amounts of benzene determined in real samples Sample PONA TCEP IR (%, v/v) (%, v/v) (%, v/v) Gasoline A92H ( ) Gasoline A92H ( ) Gasoline A95H ( ) Gasoline A95H ( ) Gasoline A98H ( ) Gasoline A98H ( ) Reformat ( ) Reformat ( ) Petroleum fraction of catalytic cracking process ( ) Petroleum fraction of catalytic cracking process ( )
10 methods and the reference method were comparable for analysis of gasolines and reformat sample. Data obtained by use of the PONA column for GC analysis of petroleum fractions from the catalytic cracking process did not, however, agree with those obtained by GC analysis with the TCEP column or by the IR method. Under the conditions used for PONA analysis the peaks of benzene and 1-methylcyclopentene were not separated sufficiently. CONCLUSION This study was an attempt to explore the potential of equipment in our laboratory for development of methods for determination of benzene in gasoline and petroleum fractions. Application of two capillary columns, conventional PONA and TCEP, enabled the separation of benzene from other compounds in petroleum samples. The success of GC FID with the PONA column depends on the composition of the gasoline samples and petroleum fractions. It is not appropriate for routine analysis despite the accuracy and the precision established. It was demonstrated that GC FID with the TCEP column has several advantages compared with analysis on the PONA column, including simplicity of handling, and performance highly precise and accurate measurements of benzene content. The analysis time is 26 min using a simple conventional column and equipment. It was faster and provided information on benzene content, which is the gasoline property regulated. The method will be useful for analysis of other petroleum products containing benzene. REFERENCES [1] Directive 98/70/EC of the European Parliament and of the Council, 13 October, 1998 [2] ASTM D Standard Test Method for Determination of Benzene and Toluene in Finished Motor and Aviation Gasoline by Gas Chromatography, Vol.05.02, 504, 2000 [3] ASTM D Standard Test Method for Aromatics in Finished Gasoline by Gas Chromatography, Vol.05.02, 881,
11 [4] ASTM D Standard Test Method for Determination of Benzene, Toluene, Ethylbenzene, p/m-xylene, o-xylene, C 9 and Heavier Aromatics and Total Aromatics in Finished Gasoline by Gas Chromatography, Vol.05.03, 620, 2000 [5] ASTM D Standard Test Method for Determination of Benzene, Toluene, and Total Aromatics in Finished Gasoline by Gas Chromatography/Mass Spectrometry, Vol.05.03, 760, 2000 [6] IP 425/96 Determination of Benzene Content of Petrol Gas Chromatography Method [7] EN 12177:1998 Liquid petroleum products Unleaded petrol Determination of benzene content by gas chromatography [8] EN 238:1996E Liquid petroleum products Petrol Determination of the benzene content by infrared spectrometry [9] E. Lopez Anreus, S. Garrigues, and M. de la Guardia, Anal. Chim. Acta, 333, 157 (1996) [10] G.S. Frysinger, R.B. Gaines, E.B. Ledford, J. High. Resol. Chromatogr., 22, 195 (1999) [11] G.S. Frysinger and R.B. Gaines, J. High. Resol. Chromatogr., 22, 251 (1999)
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
SULFUR COMPOUNDS IN PETROLEUM HYDROCARBON STREAMS
Petroleum & Coal ISSN 1337-7027 Available online at www.vurup.sk/petroleum-coal Petroleum & Coal 54 (1) 9-13, 2012 SULFUR COMPOUNDS IN PETROLEUM HYDROCARBON STREAMS Antoaneta Pavlova, Pavlina Ivanova,
Chemical Analysis. Natural Gas Analyzer. A Family of optimized GC Solutions. Gas Chromatography. think forward
Chemical Analysis Natural Gas Analyzer A Family of optimized GC Solutions think forward Gas Chromatography Bruker Natural Gas Analyzers Natural gas is bought and sold as a bulk commodity with price based
CHROMOCTANE O N LINE ANALYZER
GECIL Process GT instruments CHROMOCTANE O N LINE ANALYZER ASTM D6733 DHA and RON and MON on line GC for Reformer unit Automatic Chromatographic system for analysis of naphta feeds and reforming gasolines
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
Fractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
C184-E014D GC-2014. Shimadzu Gas Chromatograph System
C184-E014D GC-2014 Shimadzu Gas Chromatograph System Big Performance & Small Space GC-2014 GC-2014 Shimadzu Gas Chromatograph System High Performance and Expandability Merged at a higher level High Performance
HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.
HEXANES Prepared at the 51st JECFA (1998), published in FNP 52 Add 6 (1998) superseding specifications prepared at the 14th JECFA (1970), published in NMRS 48B (1971) and in FNP 52 (1992). ADI "limited
BTU ANALYSIS USING A GAS CHROMATOGRAPH
BTU ANALYSIS USING A GAS CHROMATOGRAPH By Joseph E. Landes Southern Petroleum Laboratories, Inc. Technical Director Hydrocarbon and Field Services 8820 Interchange Drive Houston, Texas 77054 Introduction
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
On-line Dissolved Gas Analysis
On-line Dissolved Gas Analysis accord. to IEC 567/ASTM 3612 Online-Monitoring of transformers Automatic on-line analysis of 11 gases The analysis is done in 2 steps with high selectivity and accuracy fully
Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder
Lecture Chromo-3: Gas Chromatography CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Outline Introduction Instrument overview Carrier gas Sample
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
SUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
Ferda ORHUN Mineral Research and Exploration Institute of Turkey
GAS CHROMATOGRAPHY, EXPERIMENTAL STUDIES ON THE FLOW RATE EFFECTS OF CARRIER GAS AND APPLICATION OF THE METHOD TO HYDROCARBON ANALYSIS OF SOME NATURAL GASES IN TURKEY Ferda ORHUN Mineral Research and Exploration
Process Instrumentation
Process Instrumentation Analysis by Gas Chromatography Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Process Gas Chromatography Features of Process Chromatographs Agilent 7890A
Annex to the Accreditation Certificate D PL 17640 01 00 according to DIN EN ISO/IEC 17025:2005
Deutsche Akkreditierungsstelle GmbH Annex to the Accreditation Certificate D PL 17640 01 00 according to DIN EN ISO/IEC 17025:2005 Period of validity: 05.06.2015 to 04.11.2018 Holder of certificate: Haltermann
Alignment and Preprocessing for Data Analysis
Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC)
ADVANCEMENTS IN MICRO GAS CHROMATOGRAPHY (GC) 1 Fast Analysis of C1 to C8 hydrocarbons for mud logging applications within 2 minutes using a temperature programmable Micro GC Fusion NEN GAS2015 Debbie
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:
Get the Right GC for Your Industry
Get the Right GC for Your Industry Shimadzu was founded on the basis of "Technology for Science" and we continue to take the initiative in the analytical technology industry. From our absorption spectrophotometers
GC/MS Considerations for Fracking Gases and Water. Dr. Harold McNair Virginia Tech Blacksburg, VA. [email protected]
GC/MS Considerations for Fracking Gases and Water Dr. Harold McNair Virginia Tech Blacksburg, VA. [email protected] 1 Outline What is fracking? Energy, Economics, Environmental Factors Chromatographic Considerations
Bringing Innovations to Standards AC Analytical Controls by PAC All In One Biodiesel Analyzer
Bringing Innovations to Standards AC Analytical Controls by PAC All In One Biodiesel Analyzer www.paclp.com Background Biodiesel offers an alternative to fossil diesel fuel & contributes to the reduction
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
CRYOCOOLER CONTAMINATION STUDY. S.W.K. Yuan, D.T.Kuo, and A.S. Loc. BEI Technologies, Cryocooler Group Sylmar, CA 91342 ABSTRACT
CRYOCOOLER CONTAMINATION STUDY S.W.K. Yuan, D.T.Kuo, and A.S. Loc EI Technologies, Cryocooler Group Sylmar, CA 91342 ASTRACT Eliminating or reduction of contamination is essential for the success of cryocoolers
MINERAL OIL (MEDIUM VISCOSITY)
MINERAL OIL (MEDIUM VISCOSITY) Prepared at the 76 th JECFA, published in FAO JECFA Monographs 13 (2012), superseding specifications for Mineral oil (Medium and low viscosity), class I prepared at the 59th
TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ACKNOWLEDGEMENT ABSTRACT LIST OF TABLES LIST OF FIGURES LIST OF SYMBOLS LIST OF APPENDICES
vii TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ii ACKNOWLEDGEMENT iv ABSTRACT v TABLE OF CONTENTS vii LIST OF TABLES xii LIST OF FIGURES xiv LIST OF SYMBOLS xvii LIST OF APPENDICES xviii 1 INTRODUCTION
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
AC DHA Solutions. Detailed Hydrocarbon Analysis of Light Petroleum Streams & Light End in Crude Oils. ASTM D 5134, D 6729, D 6730, D 6733 Compliance
AC DHA Solutions Detailed Hydrocarbon Analysis of Light Petroleum Streams & Light End in Crude Oils ASTM D 5134, D 6729, D 6730, D 6733 Compliance Signal Alignment Automated with LineUp TM Tool Compatible
Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD
Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD Application Note Environmental Author Xiaohua Li Agilent Technologies (Shanghai) Co., Ltd. 412 Ying Lun
Shimadzu Simulated Distillation Gas Chromatograph System C184-E030
Shimadzu Simulated Distillation Gas Chromatograph System C184-E030 Meets All Simulated Distillation Gas Chromatograph Standards - Provides Highly Accurate Analysis Results with Excellent Reproducibility-
Capillary and Packed Gas Chromatograph
Capillary and Packed Gas Chromatograph C184-E014G High Performance and Expandability Merged at a higher level Capillary and Packed Gas Chromatograph High Performance Superior Performance Improved design
Unit title: Chemical Laboratory Techniques
Unit title: Chemical Laboratory Techniques Unit code: H/601/0355 QCF level: 4 Credit value: 15 Aim This unit gives learners the opportunity to practise and become proficient in a range of practical skills
Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity
For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
Analysis of Vegetable Oils by High Performance Liquid Chromatography
Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation
IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.
CERTIFICATE. on Product Conformity (QAL 1) Number of Certificate: 0000034862. Gaschromategraph GC 5000 BTX Type FID for Benzene
Umwelt Bundes Amt ror our Environ ment /;;!.. TÜVRheinland Precisely Right. CERTIFICATE on Product Conformity (QAL 1) Number of Certificate: 0000034862 Certified AMS: Manufacturer: Test Institute: Gaschromategraph
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION
Gas Chromatography. Gas Chromatograph
Gas Chromatography Gas Chromatograph Chromatography: Separate analytes in a mixture with a resolution 1.5 in the shortest amount of time and detection of separated components. 1 = m.p. 6 2 4 5 3 Gas supplies
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8020 is used to determine the concentration of various aromatic volatile organic compounds. The following
Determination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS
Application Note No. 051 Determination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS by P Jüngel ¹), J Wittenzellner ²) and E Melcher ¹) ¹) Federal Research Centre
Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference
Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas
Addressing the World Helium Shortage For Gas Chromatography
Addressing the World Helium Shortage For Gas Chromatography Helium Conservation and Converting GC Methods to Nitrogen and Hydrogen Carrier Gas James D. McCurry, Ph.D. Senior Scientist Agilent Technologies
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS
SCREENING FOR THE PRESENCE OF PARA-METHYLTHIOAMPHETAMINE IN URINE BY SOME COMMERCIAL IMMUNOASSAYS AND CONFIRMATION BY GC/MS Ingrid J. BOSMAN 1, Douwe DE BOER 2, Robert A. A. MAES 1 1 Department of Human
Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x
CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana*
Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana* a Department of Pharmaceutical Analysis, Andhra University, Visakhapatnam. b Department of Pharmaceutics,
PLOT Columns. Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative
PLOT Columns Top: Bob Langford, GC Column Manufacturing Technician Bottom: Stephanie Sunner, Customer Service Representative PLOT Column Selection 93 Features & Benefits of Restek PLOT Columns Features
The Relation Between Gasoline Quality, Octane Number and the Environment
MH1 The Relation Between Gasoline Quality, Octane Number and the Environment Rafat Assi National Project Manager Jordan s Second National Communications on Climate Change (www.snc.gov.jo) Presented at
AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler
AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY Andrew Tipler Chromatography Research and Technology Manager PerkinElmer, Inc. Table of Contents Introduction 3 Fundamental
Analysing chromatographic data using data mining to monitor petroleum content in water
Analysing chromatographic data using data mining to monitor petroleum content in water Geoffrey Holmes, Dale Fletcher, Peter Reutemann and Eibe Frank Computer Science Department, University of Waikato,
Hydrogen as an alternative to Helium for gas chromatography.
Application Note Gas Chromatography. Application Note for Specialty Gases. as an alternative to for gas chromatography. Author: Peter Adam Linde AG, Linde Gas Division Seitnerstrasse 70, 82049 Pullach,
NWTPH-HCID. Hydrocarbon Identification Method for Soil and Water
NWTPH-HCID Hydrocarbon Identification Method for Soil and Water Summary This method is a qualitative and semi-quantitative procedure. It is used for groundwater or surface water, and soil/sediment from
CCME. Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method
CCME Canadian Council of Ministers of the Environment Le Conseil canadien des ministres de l environnement Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Anion chromatography using on-line recycled eluents
Journal of Chromatography A, 1089 (2005) 82 86 Anion chromatography using on-line recycled eluents Takashi Yokoyama, Hitoe Maekubo, Asami Sakai, Michio Zenki Department of Chemistry, Faculty of Science,
A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water
A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water Sjaak de Koning 1, Mark van Lieshout 2, Hans-Gerd Janssen 2 and Wil van Egmond 1 1 ATAS, De Run 4334, PO Box
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Diffusion and Fluid Flow
Diffusion and Fluid Flow What determines the diffusion coefficient? What determines fluid flow? 1. Diffusion: Diffusion refers to the transport of substance against a concentration gradient. ΔS>0 Mass
Transformer Oil Gas Analysis
Get on the Analyzer FAST TRACK Transformer Oil Gas Analysis Helping customers protect the environment! 1 Why Analyzers Value to Customers Alleviates Resource Pressure Facilitates Configuration Use Advanced
Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction
Method development for analysis of formaldehyde in foodsimulant. melamine-ware by GC-MS and LC-MS/MS. Internal Technical Report
of melamine-ware by GC-MS and LC-MS/MS Page 1 of 15 Method development for analysis of formaldehyde in foodsimulant extracts of melamine-ware by GC-MS and LC-MS/MS December 2012 Contact Point: Chris Hopley
Chemical analysis service, Turner s Green Technology Group
Chemical analysis service, Turner s Green Technology Group Hourly costs, Academic collaborations leading to co-publications: 627 SEK/hr (excl. VAT) Hourly costs, Industry/Academic collaborations, no publications:
Clarus 580 Gas Chromatograph
S P E C I F I C A T I O N S Gas Chromatography Clarus 580 Gas Chromatograph The PerkinElmer Clarus 580 Gas Chromatograph (GC) is a fully automated gas chromatograph. The system offers the capabilities
How To Use An Acquity Qda Detector
Mass-Directed Isolation of a Synthetic Peptide Using the ACQUITY QDa Detector Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY QDa Detector
GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs
GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs Kinet C, De Pauw E, Focant JF CART, Mass Spectrometry Laboratory, Chemistry Department, University of Liège, Allée de la Chimie 3,
Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
Chemistry 4631. Instrumental Analysis Lecture 1. Chem 4631
Chemistry 4631 Instrumental Analysis Lecture 1 Syllabus Chemistry 4631 Spring 2016 Lecture: MWF 9:00 9:50 a.m. Chem 352 Attendance is expected. Instructor: Dr. Teresa D. Golden Chem 279, 565-2888, [email protected].
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
Expectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA
An Automated Method for the Fractionation of Extractable Petroleum Hydrocarbons (EPH) from Water and Soil Extracts Using the Gilson GX-274 ASPEC System Keywords: Aliphatic Hydrocarbons, Aromatic Hydrocarbons,
MASS SPECTROMETRIC IDENTIFICATION OF SOME SULPHUR CONTAINING PHENALKYLAMINE DESIGNER DRUGS *
MASS SPECTROMETRIC IDENTIFICATION OF SOME SULPHUR CONTAINING PHENALKYLAMINE DESIGNER DRUGS * Ingrid J. BOSMAN 1, Douwe DE BOER 2, Edwin B. SIDERIUS 1, Lesseps J. A. L. DOS REYS 2, Robert A. A. MAES 1 1
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
ICS 75.160.20 Supersedes EN 228:2004. English Version. Automotive fuels - Unleaded petrol - Requirements and test methods
EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM EN 228 July 2008 ICS 75.160.20 Supersedes EN 228:2004 English Version Automotive fuels - Unleaded petrol - Requirements and test methods Carburants pour
A RAPID TLC METHOD FOR ANALYSIS OF EXTERNAL FLAVONOID AGLYCONES IN PLANT EXUDATES
ACTA CHROMATOGRAPHICA, NO. 14, 2004 A RAPID TLC METHOD FOR ANALYSIS OF EXTERNAL FLAVONOID AGLYCONES IN PLANT EXUDATES M. Nikolova*, S. Berkov, and S. Ivancheva Department of Applied Botany, Institute of
Thermal Conductivity Detector
Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)
The Use of Hydrogen Carrier Gas for GC/MS
Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
Theory and Instrumentation of GC Introduction
Theory and Instrumentation of GC Introduction i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference
Determination of polyols derived from sugars and residual sugars found in dry wines by means of gas chromatography (Resolution Oeno 9/2006)
Method OIV-MA-AS311-06 Type IV method Determination of polyols derived from sugars and residual sugars found in dry wines by means of gas chromatography (Resolution Oeno 9/2006) 1. Scope Simultaneous determination
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for