Tuning & Mass Calibration
|
|
|
- Oswin Banks
- 10 years ago
- Views:
Transcription
1 Tuning & Mass Calibration 1 1
2 The Sample List Sample List Name Project Name 2 The sample list is the top level screen in the TurboMass Gold Software. Data storage is set up in PROJECT files and within each project file it is possible to use a number of SAMPLE lists The PROJECT name and the SAMPLE LIST name are shown at the top of the screen 2
3 The Sample List Opens existing or creates new sample list Opens existing or creates new project 3 The ILE, NEW or OPEN allows the SAMPLE LIST to be selected Sample lists allow all the information to be entered for each sample. These will typically include The data file name for each sample The TurboMass Gold Gold tune method The TurboMass Gold Gold acquisition method The GC Method Injector site and Vial number (with an autosampler) A text description for the method The ILE, OPEN PROJECT will open an existing project; PROJECT WIZARD will take the user through the steps required to generate an new project. Beware that the MS tune files, MS methods and GC methods are stored within each project folder. If you wish to use the same tune files and method files in a new project it is best to use an existing project as a template for the new project. A completely new project will not have access to old files 3
4 The Sample List 4 Using Project Wizard anew Project Name must be entered. Also set the Directory Location for the new Project. Avoid saving Projects in the c:\turbomass directory 4
5 The Sample List 5 Using Create using Current Project as Template or Create using existing Project as Template will copy methods and tune files from the old project to the new one 5
6 Project iles MS & GC Methods Data iles Quantitation Methods Integrated Peak Data Sample Lists Calibration Data 6 PROJECT ILES consist of a number of sub folders ACQUDB will contain MS methods, GC methods, MS tune files, MS calibration files CURVEDB contains calibration data for quantitative analysis DATA contains the RAW data files from each run METHDB contains the quantitative methods PEAKDB contains data on integrated peaks SAMPLEDB contains all the sample lists associated with the project It is recommended that a new project is started at least once a month to simplify file management and archiving. 6
7 Project iles 7 Each run stored in separate folder containing several files The RAW data files are folders that contain all the data collected for that particular run. Each folder will contain a number of file labelled uncxxx.dat and uncxxx.idx where XXX is a number corresponding to the function number in the MS method. The MS unction will be discussed later in MS Method generation but each function may be a full scan, for example between m/z 35 to 350, or a selected ion recording of one or a few specific ions 7
8 The Sample List Click for the Tune Page 8 The tune page is accessed from the sample list by clicking on the button shown 8
9 The Vacuum Gauges (TurboMass Gold) Penning gauge should read» 2 x 10 5 torr To access the Vacuum gauge display 9 The TUNE page controls a number of the TurboMass Gold functions. These include The vacuum pumps (Accessed through the OPTIONS drop down menu) The heater controls for both the ion source and the transfer line The settings for the ion source, lenses, quadrapoles and photomultiplier ilament current. Also displayed on the TUNE page are the vacuum gauge readouts showing the pressure in the system (accessed by clicking on the button with the red/green bars ) 9
10 The Vacuum Gauges (TurboMass) To access the Vacuum gauge display Heater controls Penning gauge should read» 2 x 10 5 torr 10 The TUNE page controls a number of the TurboMass functions. These include The vacuum pumps (Accessed through the OTHER drop down menu) The heater controls for both the ion source and the transfer line The settings for the ion source, lenses, quadrapoles and photomultiplier ilament current. Also displayed on the TUNE page are the vacuum gauge readouts showing the pressure in the system (accessed by clicking on the button with the red/green bars ) 10
11 The Heater Controls (TurboMass Gold) Heater controls 11 The TUNE page controls a number of the TurboMass Gold functions. These include The vacuum pumps (Accessed through the OPTIONS drop down menu) The heater controls for both the ion source and the transfer line The settings for the ion source, lenses, quadrapoles and photomultiplier ilament current. Also displayed on the TUNE page are the vacuum gauge readouts showing the pressure in the system (accessed by clicking on the button with the red/green bars ) 11
12 Heater Controls (TurboMass) Heater controls 12 The ion source heater control is shown near the bottom left of the TUNE page. The required temperature is entered in the white box next to SOURCE TEMPERATURE. The transfer line temperature control is accessed by clicking on HEATERS menu. Again the temperature is set in the white box. Both temperatures should be set to ensure that no condensation of the sample will occur in either the ion source or the transfer line. The transfer line will also contain the tail of the GC column and so be set at least as high as the highest oven temperature that will be used during the analysis The grey boxes on the TUNE page are the readbacks. These show the actual value for each parameter. If these are not apparent go to the OTHER drop down menu and ensure that READBACKS is turned on. 12
13 The Vacuum Gauges Click here to turn filaments on when vacuum is adequate 13 The filament in the ion source is turned on by clicking on the PRESS OR OPERATE button. Before turning on the filament check that the vacuum is adequate. The vacuum is displayed by the readout from two gauges. The Pirani gauge (the left gauge) will only display the pressure above 1.7 x 10 4 torr and is used only during the initial stages of pumpdown. The Penning gauge (the right gauge) switches in when the Turbomolecular pump turns on at about 1 x 10 4 torr. This gauge should be checked to determine if the vacuum is adequate. The vacuum will be dependent on the carrier gas flow rate, the higher the carrier gas flow the poorer the vacuum. It is possible to use TurboMass Gold with flow rates up to about 5ml/min. or a typical carrier gas flow rate of 1ml/min the vacuum should be in the order of 2 x 10 5 torr. 13
14 The Vacuum Gauges 14 Click here to turn filaments on when vacuum is adequate The filament in the ion source is turned on by clicking on the OPERATE button. Before turning on the filament check that the vacuum is adequate. The vacuum is displayed by the readout from two gauges. The Pirani gauge (the upper gauge) will only display the pressure above 1.7 x 10 4 torr and is used only during the initial stages of pumpdown. The Penning gauge (lower gauge) switches in when the Turbomolecular pump turns on at about 1 x 10 4 torr. This gauge should be checked to determine if the vacuum is adequate. The vacuum will be dependant on the carrier gas flow rate, the higher the carrier gas flow the poorer the vacuum. It is possible to use TurboMass with flow rates up to about 5ml/min. or a typical carrier gas flow rate of 1ml/min the vacuum should be in the order of 2 x 10 5 torr. 14
15 The Air / Water background Monitor air water background for a leak 4 = helium 18 = water 28 = N 2 32 = O 2 Select Tune Masses Display Green indicates that filaments are on 15 Once the vacuum is in the order of 2 x 10 5 torr select the scope display by clicking on the button (TUNE MASS parameters) and then select ions of 4 (He), 18 (H 2 O), 28 (N 2 ) and 32 (O 2 ) before turning the filament on. When turning the filament on monitor the abundance of each ion, especially the 32 ion. If the 32 ion is more than 25% of the 28 ion turn off the filaments immediately and check the system for air leaks. 15
16 The Air / Water background m/z Type of Leak , 44 None base < m/z 4 < m/z 18 < 25% m/z 28 << m/z 32 Small Leak base < m/z 4 >m/z 18 < 25% m/z 28 << m/z 32 Medium Leak > m/z 18 > 25% m/z 28 << m/z 32 Large Leak > m/z 28 << m/z 32 Very Large Leak very wide 16 Typically the 4 ion should be the largest, followed by the 18 then the 28 with 32 being approximately 25% of the 28. If the 28 and 32 ions are excessively high turn the filament off and allow the system to pump down for longer. The air/water background is best measured with the splitter on the injector open with a split flow of at least 10ml/min. If the system is connected to an ATD 400 or TurboMatrix ATD, then open the outlet split. Normally the water peak (m/z 18) is usually larger than the nitr ogen peak (m/z 28) but if a nonpolar column is installed and the system has been in use for some time then the m/z 18 intensity may be l ower than the m/z 28 intensity. If a large air leak is present the ratio of m/z 32 to m/z 28 will change with the 32 ion increasing in proportion to the 28 ion. 16
17 The Tune Page Turn Ref gas on Tuning Masses 17 To tune the TurboMass Gold access the TUNE MASS parameters and select masses 69, 131, 219 and 502 Turn the reference gas on by clicking on the button labelled RE GAS 17
18 18 18 The Reference Gas The Reference Gas N C C C C C C C C C C C C Heptacosafluorotributylamine (C N) The reference gas normally used is heptacosafluorotributylamine (C N)
19 The Reference Gas Synonyms 19 C C C C N C C C C C C C C The reference gas has a number of synonyms, the more common ones are PTBA, C43 and hepta 19
20 The Reference Gas Mass Spectrum Tuning Masses 20 The masses that are normally used to check the tune are 69, 131, 219 and may also be used to check the tune at higher masses 20
21 The Tune Page Turn Reference Gas on Turn Autotune on 21 An Autotune procedure may be used by clicking on the button highlighted above. Before commencing Autotune, however, ensure that the emission current is set to 100µa or lower. Lowe r emission currents will increase the lifetime of the filament and keep the source cleaner. However some loss in sensitivity will be observed at lower emission currents. Do not use an emission cur rent of less than 50µa Tuning should not be necessary more than once a week. However it is good practice to check that the tune is OK by turning on the reference gas each morning and checking the relative intensities of each ion before proceeding with the analysis. If the system needs to be retuned the sensitivity of the TurboMass Gold may be affected and should be recalibrated before any quantitative work is attempted. 21
22 AutoTune Autotune Parameters 22 The Autotune parameters may be accessed using the SETUP button 22
23 AutoTune Turns mass recalibration on / off Tuning Masses Mass Resolution 23 The Autotune uses to masses for the setup procedure. The default settings are 69 for low masses and 502 for the high masses. The resolution is set at low mass (69) and high mass (502). The default setting is 0.6 Daltons at ½ height which will give a peak width at base of approximately 1 Dalton If Prompt for Recalibration is ticked the system will follow the tune with a mass recalibration procedure. 23
24 AutoTune Start Autotune 24 When the Auto Tune parameters have been set press START to begin Auto Tune 24
25 The Tune Page Adjust photomultiplier so that m/z 69 peak is on scale Double click here to reduce gain Relative abundances: % % % % 614 >0.1% 25 Double click here to increase gain The Autotune procedure will set all parameters including the photomultiplier voltage. However, this is set to provide a reasonable intensity for the reference gas ions and must be set independently to achieve the sensitivity required for the samples that are to be run. To check that the Autotune has set the system up correctly adjust the photomultiplier voltage so that the 69 ion is on scale. The relative abundances of the ions should be in the order of: % % % % 614 >0.1% (when used) When making the above checks ensure that all peaks are on scale with the gain set at 1. The gain may be increased to make individual peaks larger by entering the value in the GAIN column or by double clicking in the grey box immediately above the peak display. The gain may also be reduced by double clicking beneath the peak. If the gain is set to 1(69 ion), 3 (131), 3 (219) and 30 (502) the peaks should be approximately the same height. 25
26 Ion Source (Viewed from Column exit) ilament 70 ev Lens 1 Lens 2 Repeller Electron Trap 47V ve ve 26 This schematic shows the main components of the ion source. At the top is the filament which, when turned on, generates a stream of electrons that are attracted to the Electron Trap (or collector) electrode that has 47 volts applied to it. The repeller is the larger electrode. This also has a positive voltage applied to it which has the effect of repelling the positive ions that are generated by the collision of electrons with the sample molecules eluting from the end of the column. These positive ions are repelled towards the 2 lenses mounted outside the inner source. Not shown in this schematic are two magnets mounted just below source and above the filament. The quadrapoles are mounted to the right of Lens 2 26
27 Manual Tuning Always 70 ev or TurboMass Default is 200µa. Do Always not exceed use Maintenance 100µa. Default tune value & set is to 100µaor lower 27 If the tune does not meet the required specifications then some manual tuning may be necessary. If the Autotune does not give good values for the peak intensities it is usually a sign that the source is dirty. When manual tuning ensure that the Electron energy (E Energy) is set to 70eV. All spectra in the libraries have been collected at 70eV and to achieve a good match it is essential that spectra collected are run at this setting. 27
28 Ion Source Emission Current ilament 70 ev Lens 1 Lens 2 47V ve ve 28 Emission Current (ma) Default = 100ma. Do not exceed 100µa The emission current generated by the stream of electrons from the heated filament to the collector electrode which always has 47V applied. 100µa will give best sensitivity, lower current settings will enhance the life of the filament and keep the source cleaner. 28
29 Manual Tuning Usually between 0 & 1V. High values indicate a dirty inner source. Autotune will not set a value higher than 1.5V 29 The repeller voltage is usually between 0 and 1Volt. The cleaner the ion source the lower the voltage required. If the Autotune gives a value significantly greater than 1V it suggests that the surface of the repeller electrode is badly contaminated and the source will need cleaning. Autotune will never set a repeller voltage higher than 1.5V. If the repeller is set to 1.5V after an Autotune then it is likely that the source requires cleaning. 29
30 Ion Source Effect of Repeller Voltage ilament 70 ev Repeller (ve) 0 to 1.5V Lens 1 Lens 2 47V ve ve 30 The object of the repeller is to broaden the electron beam by applying a positive voltage to the side of the beam. It also repels the positively charged ions towards Lens 1. If the surface of the repeller is dirty than a higher voltage is necessary to achieve the desired effect. 30
31 Manual Tuning Lens 1 pulls ions from source Lens 2 to focus ion stream 31 Lens 1 has a negative voltage applied to pull the positive ions produced from the ion source. The autotune will set this to 7V Occasionally a slightly better sensitivity may be seen at 6V. Lens 2 is used to focus the stream of ions into a parallel beam. The voltage applied is generally higher than for Lens 1 and is typically between 80 and 100V 31
32 Ion Source Effect of Lens Voltages ilament 70 ev Repeller (ve) 0 to 1.5V 47V ve ve 5 to 7V Lens 1 Lens 2 32 Lens 1 voltage is set to 7 volts by the Autotune to pull the positive ions from the source. Occasionally slightly better sensitivity may be achieved by reducing Lens 1 voltage to 6 volts or even 5 volts 32
33 Ion Source Effect of Lens Voltages ilament 70 ev Repeller (ve) 0 to 1.5V Lens 1 Lens 2 47V ve ve 80 to 100V 33 Lens 2 voltage is usually around 80 to 100 Volts and focuses the positive ions in a parallel beam towards the quadrapoles 33
34 Manual Tuning Usually TurboMass around uses 3a. a Higher Tungsten values indicate ilament possible dirty Typical source or values aged are filament from 4 5 amps Should be 3 to 4 times Emission current. Higher values suggest a dirty source 34 The filament current is set automatically to give the emission current and is usually set to around 3.5 amps. If it is significantly higher it suggests that the collector electrode is dirty. 34
35 Source Current normally 3 to 4 times Emission current Ion Source Effect of Lens Voltages ilament 70 ev Lens 1 Lens 2 47V ve ve 35 The source current is a measure of the electrons produced by the filament that ground on the chassis of the ion source. It should be in the order of 3 to 4 times the emission current. If it significantly higher then again it suggests that the collector electrode is dirty. 35
36 Manual Tuning Affects low mass peak width Affects high mass peak width 36 Max Signal to Noise ratio at 480 to 500 V but adjust so that chromatographic peaks are not saturated There are 2 parameters that will affect resolution; LMRes affects resolution at low masses whilst HMRes affects the resolution at high masses. In Autotune they are set up using masses 69 and 502 respectively. Changing these will affect the peak width and therefore the intensity. These may be used in manual tune to change the relative intensities of the tune peaks. A smaller value will make the peak broader and therefore increase its relative intensity. Mas ses that occur between 69 and 502 will also be affected by theses settings. If these are used to adjust the relative peak intensi ties then the gain should be increased so that the ions 70, 132, 220 and 503 can be seen along side the major ions. There should not be baseline separation between each ion pair but the valley between the 2 ions should be approximately 20 to 50% of the minor ion intensity. The ion energy and ion energy ramp affect the way the quadrapole s are ramped up. These should only be touched as a last resort but if a satisfactory tune cannot be achieved try setting the MS ion energy to 0.3 or less and the IE Ramp to 3 to 4 and then readjust the Lens 2 voltage. 36
37 Manual Tuning (TurboMass) Affects low mass peak width Affects low mass peak width 37 Max Signal to Noise ratio at 480 to 500 V but adjust so that chromatographic peaks are not saturated On the MS page there are 2 parameters that will affect resolution; LMRes affects resolution at low masses whilst HMRes affects the resolution at high masses. In Autotune they are set up to masses 69 and 502 respectively. Changing these will affect the peak width and therefore the intensity. These may be used in manual tune to change the relative intensities of the tune peaks. A smaller value will make the peak broader and therefore increase its relative intensity. Masses that occur between 69 and 502 will also be affected by theses settings. If these are used to adjust the relative peak intensities then the gain should be increased so that the ions 70, 132, 220 and 503 can be seen along side the major ions. There should not be baseline separation between each ion pair but the valley between the 2 ions should be approximately 20 to 50% of the minor ion intensity. The ion energy and ion energy ramp affect the way the quadrapoles are ramped up. These should only be touched as a last resort but if a satisfactory tune cannot be achieved try setting the MS ion energy to 0.3 or less and the IE Ramp to 3 to 4 and then readjust the Lens 2 voltage. 37
38 Manual Tuning Reinitialise Noise levels Photomultiplier Voltage 38 Once a satisfactory tune has been achieved the photomultiplier voltage must be adjusted to obtain the best sensitivity for the samples to be run and then the noise level must be reinitialised by clicking on the button shown. Noise reinitialisation must be done whenever the photomultiplier voltage is changed. Lastly turn off the reference gas and then turn on PUMP OUT REERENCE GAS accessed from the GAS menu. It is recommended that the PUMP OUT REERENCE GAS is left turned on as it reduces any background interferences and this is done by saving the tune file with PUMP OUT REERENCE GAS turned on. Caution: If PUMP OUT REERENCE GAS is not turned off before the reference gas is turned on again the reference gas vial will be emptied very quickly! 38
39 Mass Calibration 39 39
40 Mass Calibration Automatically after Autotune Automatic Mass Calibration selected after Autotune completed 40 Mass calibration is used to ensure that the ions detected from the reference gas are assigned the correct masses. This may be set up as part of the tune procedure if the PROMPT OR RECALIBRATION has been ticked. Mass calibration should not need to be done at the same frequency as tuning, maybe once a month. NOTE Mass Calibration will require the reference gas is turned on 40
41 Mass Calibration Accessing manually Manual mass calibration from here 41 Alternatively mass calibration may be accessed manually from the control panel 41
42 Mass Calibration TurboMass Accessing manually Manual mass calibration from here 42 Alternatively mass calibration may be accessed manually from the control panel 42
43 Mass Calibration TurboMass Accessing manually Mass calibrate from instrument drop down menu 43 rom the control panel page Mass Calibration is under the Instrument menu 43
44 Mass Calibration The Reference ile Heptacosaflurotributylamine Directory for mass calibration 44 During Mass calibration each peak is compared to a text file containing the ion masses and intensities for the reference gas. The file to use for heptacosafluorotributylamine is called heptaref. As heptacosafluorotributylamine contains very few ions below 69 air peaks may also be used to calibrate a low masses. If this is ticked then 4, 18 and 28 will also be matched but not
45 Mass Calibration The Reference ile Access to the text file Includes air peaks (4=He, 18=H 2 O, 28=N 2 ) 45 The text file used for heptacosafluorotributylamine is called heptaref. As heptacosafluorotributylamine contains very few ions below 69 air peaks may also be used to calibrate a low masses. On the TurboMass Gold or the Clarus 500MS the air peaks are included in the heptaref file. 45
46 Mass Calibration TurboMass The Reference ile Heptacosaflurotributylamine Directory for mass calibration Also uses m/z 4,18 and 28 for low mass calibration 46 During Mass calibration each peak is compared to a text file containing the ion masses and intensities for the reference gas. The file to use for heptacosafluorotributylamine is called heptaref. As heptacosafluorotributylamine contains very few ions below 69 air peaks may also be used to calibrate a low masses. If this is ticked then 4, 18 and 28 will also be matched but not
47 Mass Calibration Automatic Calibration Static Calibration Scanning Calibration Scan Speed compensation dwelling on individual m/z across mass range to correct for different scan speeds To set Mass Calibration parameters 47 When START has been clicked a number of options appear. There are three types of calibration that may be performed. A static calibration, a scanning calibration and the a scan speed compensation calibration performed at a faster scan speed than the first scanning calibration. Under Process, Acquire and calibrate recalibrates the mass axis and Acquire and Verify checks that the calibration is correct. The button ACQUISITION PARAMETERS allows the calibration parameters to be adjusted. 47
48 Mass Calibration Acquisition Parameters 48 Acquisition parameters determine the scan range and the run duration for each run. The DATA TYPE must always be set to Centroid when using GC/MS. Scan Parameters determine how each mass calibration scan is performed. The default values are usually adequate but if very fast scanning rates are to be used it may be necessary to perform the Scan speed compensation mass calibration at a faster speed than normally used. 48
49 Mass Calibration Calibration Display 49 The display above shows a typical mass calibration run. In this example the fast scan verification is shown. The top display shows the recalibrated masses whilst the second display shows the reference file. The third display shows the residuals errors after recalibration. 49
50 50 50
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Analysis of Liquid Samples on the Agilent GC-MS
Analysis of Liquid Samples on the Agilent GC-MS I. Sample Preparation A. Solvent selection. 1. Boiling point. Low boiling solvents (i.e. b.p. < 30 o C) may be problematic. High boiling solvents (b.p. >
How To Use Gc-Ms
The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter
MassLynx 4.1 Getting Started Guide
MassLynx 4.1 Getting Started Guide 71500113203/Revision A Copyright Waters Corporation 2005. All rights reserved. Copyright notice 2005 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA AND IRELAND.
Setting up a Quantitative Analysis MS ChemStation
Setting up a Quantitative Analysis MS ChemStation Getting Ready 1. Use the tutorial section "Quant Reports" on the MSD Reference Collection CD-ROM that came with your ChemStation software. 2. Know what
Expectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
Micromass LCT User s Guide
Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
SCION SQ Series GC-MS
SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative
Agilent G1701DA MSD ChemStation
Agilent G1701DA MSD ChemStation 5973 MSD / ChemStation Quick Reference Agilent Technologies What s New in Revision D.01.00 A new toolbar has been added to the Top view and an expanded set of toolbars have
Tutorial 9: SWATH data analysis in Skyline
Tutorial 9: SWATH data analysis in Skyline In this tutorial we will learn how to perform targeted post-acquisition analysis for protein identification and quantitation using a data-independent dataset
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
Thermal Conductivity Detector
Thermal Conductivity Detector Troubleshooting Tips The thermal conductivity detector (TCD) detects the difference in thermal conductivity between column effluent flow (carrier gas + sample components)
Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) (PCS# 600000-1621-RFQ)
Request for Quotes Goods/Trade Services RFQ for Gas Chromatograph Mass Spectrometer (GCMS) Closing Date and Time: Wednesday, April 13, 2016 at 5:00 p.m. (Pacific Time) Contact: Aaron Galloway Fax: 541-888-3250
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.
Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide
Q-TOF User s Booklet. Enter your username and password to login. After you login, the data acquisition program will automatically start.
Q-TOF User s Booklet Enter your username and password to login. After you login, the data acquisition program will automatically start. ~ 1 ~ This data acquisition window will come up after the program
Dash 18X / Dash 18 Data Acquisition Recorder
75 Dash 18X / Dash 18 Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 3.1 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH
How To Use An Ionsonic Microscope
Sample Analysis Design Element2 - Basic Software Concepts Scan Modes Magnetic Scan (BScan): the electric field is kept constant and the magnetic field is varied as a function of time the BScan is suitable
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software
Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software Mark Belmont and Alexander N. Semyonov, Thermo Fisher Scientific, Austin, TX, USA Ming Gu, Cerno Bioscience,
FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK. VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS
FORENSIC TOXICOLOGY LABORATORY OFFICE OF CHIEF MEDICAL EXAMINER CITY OF NEW YORK VOLATILES ANALYSIS by GCMS HEADSPACE ANALYSIS PRINCIPLE This method confirms the identity of volatile compounds in biological
Universal Simple Control, USC-1
Universal Simple Control, USC-1 Data and Event Logging with the USB Flash Drive DATA-PAK The USC-1 universal simple voltage regulator control uses a flash drive to store data. Then a propriety Data and
Cathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Residual Gas Analyzers
Residual Gas Analyzers RGA100/200/300 100 amu, 200 amu and 300 amu systems SRS Residual Gas Analyzers 100, 200 and 300 amu systems Better than 1 amu resolution 6 decades of dynamic range 5 10 14 Torr detection
Analyst 1.6 Software. Software Reference Guide
Analyst 1.6 Software Software Reference Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment.
Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter
INL/EXT-08-15056 Measurement of Specific Heat Capacity Using Differential Scanning Calorimeter J. E. Daw November 2008 The INL is a U.S. Department of Energy National Laboratory operated by Battelle Energy
Quick Start Guide Chromeleon 7.2
Chromeleon 7.2 7229.0004 Revision 1.0 July 2013 Table of Contents 1 Introduction... 1 1.1 About this Document... 1 1.2 Other Documentation... 2 2 Using Chromeleon... 3 2.1 Overview... 3 2.2 Starting Chromeleon...
SA-9600 Surface Area Software Manual
SA-9600 Surface Area Software Manual Version 4.0 Introduction The operation and data Presentation of the SA-9600 Surface Area analyzer is performed using a Microsoft Windows based software package. The
5973 inert Mass Selective Detector
5973 inert Mass Selective Detector Hardware Manual Agilent Technologies 5973 inert Mass Selective Detector Hardware Manual Notices Agilent Technologies, Inc. 2003 No part of this manual may be reproduced
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring
The Scheduled MRM Algorithm Enables Intelligent Use of Retention Time During Multiple Reaction Monitoring Delivering up to 2500 MRM Transitions per LC Run Christie Hunter 1, Brigitte Simons 2 1 AB SCIEX,
Chromatograph Maintenance and Troubleshooting Class # 5050
Chromatograph Maintenance and Troubleshooting Class # 5050 Shane Hale Business Development Manager Emerson Process Management Gas Chromatographs Division 5650 Brittmoore Drive Houston, TX 77041 USA Introduction
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
The Use of Hydrogen Carrier Gas for GC/MS
Technical Note The Use of Hydrogen Carrier Gas for GC/MS Gas Chromatography/ Mass Spectrometry Highlights Guidelines that can mitigate dangers and leverage benefits of using hydrogen Key safety factors
Mass Frontier 7.0 Quick Start Guide
Mass Frontier 7.0 Quick Start Guide The topics in this guide briefly step you through key features of the Mass Frontier application. Editing a Structure Working with Spectral Trees Building a Library Predicting
INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS
AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,
View of ΣIGMA TM (Ref. 1)
Overview of the FESEM system 1. Electron optical column 2. Specimen chamber 3. EDS detector [Electron Dispersive Spectroscopy] 4. Monitors 5. BSD (Back scatter detector) 6. Personal Computer 7. ON/STANDBY/OFF
How To Build An Amd Quas 3 Ar Mass Spectrometer
Classical and special AMD Magnetic Sector mass spectrometers (extracted from power point presentations) High Resolution Mass Spectromter AMD 402/403 S 6 High Resolution Mass Spectrometer AMD 604 S 8 Version
Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012.
Using NIST Search with Agilent MassHunter Qualitative Analysis Software James Little, Eastman Chemical Company Sept 20, 2012 Introduction Screen captures in this document were taken from MassHunter B.05.00
Lab 7: Residual Gas Analyzers
Lab 7: Residual Gas Analyzers Vacuum System Overview: Residual Gas Analyzers. A Residual Gas Analyzer (RGA) allows the scientist/engineer to measure the gases present in a low-pressure environment. This
LG External HDD Hard Disk Drive XG1
LG External HDD Hard Disk Drive XG1 - Thank you for purchasing this product. - This user manual contains instructions on how to use the product properly. - Please read this manual carefully. - Be sure
Chem2XL Workflow Quick Reference Guide How to Create Excel Reports Agilent ChemStation B.02.01 SR1 and higher
Chem2XL Workflow Quick Reference Guide How to Create Excel Reports Agilent ChemStation B.02.01 SR1 and higher INTRODUCTION Chem2XL is designed to easily create Microsoft Excel reports from your ChemStation
Dash 8Xe / Dash 8X Data Acquisition Recorder
75 Dash 8Xe / Dash 8X Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 2.0 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH
ChromQuest 5.0 Chromatography Data System
ChromQuest 5.0 Chromatography Data System User Guide for the Surveyor LC CHROM-97257 Revision A March 2008 2008 Thermo Fisher Scientific Inc. All rights reserved. Surveyor and ChromQuest are trademarks
Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps
Application Note: ANTV10 Application of TargetView software within the food industry - the identi cation of pyrazine compounds in potato crisps Abstract In order to regulate product safety and quality
EnAlgae SOP: Biogas analysis
Table of content 1. PURPOSE... 2 2. PRINCIPLE... 2 3. REQUIREMENTS... 3 3.1. EQUIPMENT AND MATERIALS... 3 3.2. REAGENTS... 3 3.3. PREPARATION OF MEASUREMENT... 3 4. PROCEDURE... 4 4.1. SAMPLE MEASUREMENT...
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Batch Eligibility Long Term Care claims
Hewlett Packard Enterprise Provider Electronic Solutions software lets Connecticut Medical Assistance Program providers verify patient s eligibility and submit and correct claims for services all electronically.
Creating an Email Archive in Outlook
Creating an Email Archive in Outlook Setting up AutoArchive Auto Archiving is the suggested method for archiving mail in Outlook. Mailbox items are AutoArchived on last modification date, not the date
Specifications. MassHunter WorkStation for Agilent 7700 Series ICP-MS. Specifications: MassHunter Workstation (G7211A) for the Agilent 7700 Series
Specifications MassHunter WorkStation for Agilent 7700 Series ICP-MS Specifications: MassHunter Workstation (G7211A) for the Agilent 7700 Series The ICP-MS MassHunter Workstation for the Agilent 7700 Series
Lab 3: Introduction to Data Acquisition Cards
Lab 3: Introduction to Data Acquisition Cards INTRODUCTION: In this lab, you will be building a VI to display the input measured on a channel. However, within your own VI you will use LabVIEW supplied
Royal Mail Despatch Manager Online Printer Installation Guide
Royal Mail Despatch Manager Online Printer Installation Guide Getting you started Thank you for opting to use Royal Mail s Despatch Manager Online (DMO) system. To help get you started successfully please
Quick Start Guide. Highly customizable automated trading Automate your trades according to rules and models you create.
POWER E*TRADE PRO EXCEL MANAGER Quick Start Guide We are pleased to announce the launch of Excel Manager, an exciting new feature in Power E*TRADE Pro that leverages the flexibility of Microsoft Excel
E/M Experiment: Electrons in a Magnetic Field.
E/M Experiment: Electrons in a Magnetic Field. PRE-LAB You will be doing this experiment before we cover the relevant material in class. But there are only two fundamental concepts that you need to understand.
QIAGEN Rotor Gene Q: Software V2.0
QIAGEN Rotor Gene Q: Software V2.0 Instrument Setup Instructions for RT 2 Profiler PCR Arrays Preparation Before the Experiment (Presetting the Machine will save time for your run): Please make note of
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
Fax Getting Started Guide
Fax Getting Started Guide April 2009 www.lexmark.com Contents IMPORTANT SAFETY INSTRUCTIONS...3 Getting started...4 Finding more information about the printer...4 Navigating menus using the printer control
OPERATION MANUAL. MV-410RGB Layout Editor. Version 2.1- higher
OPERATION MANUAL MV-410RGB Layout Editor Version 2.1- higher Table of Contents 1. Setup... 1 1-1. Overview... 1 1-2. System Requirements... 1 1-3. Operation Flow... 1 1-4. Installing MV-410RGB Layout
Blackbody Radiation References INTRODUCTION
Blackbody Radiation References 1) R.A. Serway, R.J. Beichner: Physics for Scientists and Engineers with Modern Physics, 5 th Edition, Vol. 2, Ch.40, Saunders College Publishing (A Division of Harcourt
AutoSpec Premier. Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements?
AutoSpec Premier Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements? Looking for sensitivity, selectivity and dynamic range? The AutoSpec Premier is the latest
TECHNICAL SERVICE DEPARTMENT Technical Service Bulletin 1-800-432-8373. Tankless Electric (RTE) Troubleshooting
Sequence of Operations 1 Power supply and field wiring block 2 Energy Cut Off (ECO) 3 Water flow plunger and cold inlet 4 Magnetic flow switch 5 Water temperature thermistor 6 Control panel and circuit
Specific Information for installation and use of the database Report Tool used with FTSW100 software.
Database Report Tool This manual contains: Specific Information for installation and use of the database Report Tool used with FTSW100 software. Database Report Tool for use with FTSW100 versions 2.01
Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.
1234 Hach Hall 515-294-5805 www.cif.iastate.edu Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V. Location: Contact: 1240 Hach Hall Steve Veysey, 1234 Hach Hall Safety All researchers working
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
Event recording (or logging) with a Fluke 287/289 Digital Multimeter
Event recording (or logging) with a Fluke 287/289 Digital Multimeter Application Note One of the major features of the Fluke 280 Series digital multimeters (DMM) with TrendCapture is their ability to record
Online Backup and Recovery Manager Setup for Microsoft Windows.
Online Backup and Recovery Manager Setup for Microsoft Windows. Backup Wizard Default Interface Recovery Wizard Online Backup and Recovery Manager Setup Backup Wizard After installing and launching the
A Navigation through the Tracefinder Software Structure and Workflow Options. Frans Schoutsen Pesticide Symposium Prague 27 April 2015
A Navigation through the Tracefinder Software Structure and Workflow Options Frans Schoutsen Pesticide Symposium Prague 27 April 2015 Kings day in The Netherlands 1 Index Introduction Acquisition, Method
Using an Automatic Back Up for Outlook 2003 and Outlook 2007 Personal Folders
Using an Automatic Back Up for Outlook 2003 and Outlook 2007 Personal Folders Part 1 Install the Personal Folder Backup Utility Note: You should close Outlook and any other applications before installing
Snap 9 Professional s Scanning Module
Miami s Quick Start Guide for Using Snap 9 Professional s Scanning Module to Create a Scannable Paper Survey Miami s Survey Solutions Snap 9 Professional Scanning Module Overview The Snap Scanning Module
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
Amino Acid Analyzer L-8900
Amino Acid Analyzer L-8900 VWR - Hitachi Your Partner in Amino Acid Analysis Hitachi has manufactured more than 1800 Amino Acid Analysers for over 40 years. The new Amino Acid Analyser Model L-8900 is
Shimadzu UV-VIS User s Guide
Shimadzu UV-VIS User s Guide 1) Push the F4 button on the UV-VIS instrument keypad. This will enable PC control. Push the F4 Button 2) Log into the UV-VIS software with your username and password. 3) After
Analyst 1.6 Software. Getting Started Guide
Analyst 1.6 Software Getting Started Guide Release Date: August 2011 This document is provided to customers who have purchased AB SCIEX equipment to use in the operation of such AB SCIEX equipment. This
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
In this note, you will learn the basic applications of McAfee VirusScan Enterprise (hereafter McAfee ). Six topics will be covered as below:
ANTIVIRUS SOFTWARE In this note, you will learn the basic applications of McAfee VirusScan Enterprise (hereafter McAfee ). Six topics will be covered as below: 1. Start McAfee VirusScan Enterprise 2. Change
ForensicDB: A Web-accessible Spectral Database
ForensicDB: A Web-accessible Spectral Database Table of Contents 1. Introduction to ForensicDB... 1 2. Accessing ForensicDB... 1 3. Setting System Properties... 5 Spectrum Properties... 5 Frame Properties...
Instructions for Setting the T560 Digital Delay Generator for the Target Delay
Instructions for Setting the T560 Digital Delay Generator for the Target Delay Important! These instructions are only to be followed by beam-line operators who have been trained to operate the target!!
MyNetFone Virtual Fax. Virtual Fax Installation
Table of Contents MyNetFone Virtual Fax MyNetFone Virtual Fax Installation... 1 Changing the SIP endpoint details for the fax driver... 11 Uninstalling Virtual Fax... 13 Virtual Fax Installation Follow
Name: Due: September 21 st 2012. Physics 7230 Laboratory 3: High Resolution SEM Imaging
Name: Due: September 21 st 2012 Physics 7230 Laboratory 3: High Resolution SEM Imaging 1. What is meant by the term resolution? How does this differ from other image variables, such as signal to noise
13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD 175 STANFORD, CA 94305-5080
Training on the ZQ Open access MS Questions? Contact Dr. Allis Chien [email protected] 650-723-0710 0710 STANFORD UNIVERSITY MASS SPECTROMETRY STANFORD UNIVERSITY MASS SPECTROMETRY 333 CAMPUS DR., MUDD
Event recording (or logging) with a Fluke 287/289 Digital Multimeter
Event recording (or logging) with a Fluke 287/289 Digital Multimeter Our thanks to Fluke Corp for allowing us to reprint the following article. One of the major features of the Fluke 280 Series digital
Transmitter Interface Program
Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following
Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835
Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that
Tips and Tricks SAGE ACCPAC INTELLIGENCE
Tips and Tricks SAGE ACCPAC INTELLIGENCE 1 Table of Contents Auto e-mailing reports... 4 Automatically Running Macros... 7 Creating new Macros from Excel... 8 Compact Metadata Functionality... 9 Copying,
Time-of-Flight Mass Spectrometry
Time-of-Flight Mass Spectrometry Technical Overview Introduction Time-of-flight mass spectrometry (TOF MS) was developed in the late 1940 s, but until the 1990 s its popularity was limited. Recent improvements
Ultrasound Condition Monitoring
Ultrasound Condition Monitoring Whitepaper Alan Bandes UE Systems, Inc. Abstract: Instruments based on airborne/structure borne ultrasound technology offer many opportunities for reducing energy waste
QUICK TUNE EFI INSTRUCTION MANUAL
QUICK TUNE EFI INSTRUCTION MANUAL Step 1: Loading the CB quick Tune Software to your PC A) Take the CB Quick Tune installation CD and load it into your PC. A window will pop up asking to have all programs
GelAnalyzer 2010 User s manual. Contents
GelAnalyzer 2010 User s manual Contents 1. Starting GelAnalyzer... 2 2. The main window... 2 3. Create a new analysis... 2 4. The image window... 3 5. Lanes... 3 5.1 Detect lanes automatically... 3 5.2
H-ITT CRS V2 Quick Start Guide. Install the software and test the hardware
H-ITT CRS V2 Quick Start Guide Revision E Congratulations on acquiring what may come to be one of the most important technology tools in your classroom! The H-ITT Classroom Response System is quite easy
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
1. Installing The Monitoring Software
SD7000 Digital Microphone Monitor Software manual Table of Contents 1. Installing The Monitor Software 1.1 Setting Up Receivers For Monitoring 1.2 Running The Application 1.3 Shutdown 2. The Detail Monitoring
LCD MONITOR TOUCH PANEL DRIVER 2 OPERATION MANUAL. for Mac. Version 1.0 PN-L703A/PN-70TA3/PN-L703B/PN-70TB3/PN-L603A/PN-60TA3/PN-L603B/PN-60TB3
LCD MONITOR TOUCH PANEL DRIVER 2 OPERATION MANUAL for Mac Version 1.0 Applicable models PN-L703A/PN-70TA3/PN-L703B/PN-70TB3/PN-L603A/PN-60TA3/PN-L603B/PN-60TB3 Contents Introduction...3 System Requirements...3
Analyzer Experiment setup guide for LSRII and Canto s
Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,
SEMTech Solutions. Leaders in Refurbished SEMs. SEMTech Solutions Windows 7 SOFTWARE CONTROL SYSTEM
SEMTech Solutions Leaders in Refurbished SEMs SEMTech Solutions Windows 7 SOFTWARE CONTROL SYSTEM Recertification Process Our Goal: Value Added Technologies Demo Outgoing Inspection Can Include: New PC
LCD MONITOR TOUCH DISPLAY DRIVER OPERATION MANUAL. for Mac. Version 1.0 PN-K322B/PN-K322BH/LL-S201A. Applicable models
LCD MONITOR TOUCH DISPLAY DRIVER OPERATION MANUAL for Mac Version 1.0 Applicable models PN-K322B/PN-K322BH/LL-S201A Contents Introduction...3 System Requirements...3 Setting up the Computer...4 Installing
COLOR TFT LCD MONITOR. User Manual
COLOR TFT LCD MONITOR User Manual GENERAL INFORMATION Thank you for choosing our TFT LCD(liquid crystal display) monitor. This product employs integrate circuits, low power consumption, and no radiation
How To Create A View Frame In 3D
12/4/2008-10:00 am - 11:30 am Room:Palazzo O-P (5th) The Secrets of Cutting Plan and Profile Sheets in AutoCAD Civil 3D Michelle Rasmussen - Application Engineer, IMAGINiT Technologies CV304-1P In this
