Ionization Methods in Organic Mass Spectrometry Contents:
|
|
|
- Martina Harvey
- 9 years ago
- Views:
Transcription
1 Ionization Methods in Organic Mass Spectrometry Contents:. Introduction. Gas-Phase Ionization. Electron Ionization (EI). Chemical Ionization (CI). Desorption Chemical Ionization (DCI). Negative-ion chemical ionization. Field Desorption and Ionization. Field Desorption (FD). Field Ionization (FI). Particle Bombardment. Fast Atom Bombardment (FAB). Secondary Ion Mass Spectrometry (SIMS). Atmospheric Pressure Ionization. Electrospray Ionization (ESI). Atmospheric Pressure Chemical Ionization (APCI). Laser Desorption. Matrix-Assisted Laser Desorption Ionization (MALDI). Appendices. CI Reagent Gases. Direct Formation of Negative Ions. FAB matrices
2 Introduction A mass spectrometer works by using magnetic and electric fields to exert forces on charged particles (ions) in a vacuum. Therefore, a compound must be charged or ionized to be analyzed by a mass spectrometer. Furthermore, the ions must be introduced in the gas phase into the vacuum system of the mass spectrometer. This is easily done for gaseous or heat-volatile samples. However, many (thermally labile) analytes decompose upon heating. These kinds of samples require either desorption or desolvation methods if they are to be analyzed by mass spectrometry. Although ionization and desorption/desolvation are usually separate processes, the term "ionization method" is commonly used to refer to both ionization and desorption (or desolvation) methods. The choice of ionization method depends on the nature of the sample and the type of information required from the analysis. So-called 'soft ionization' methods such as field desorption and electrospray ionization tend to produce mass spectra with little or no fragment-ion content. Gas-Phase Ionization These methods rely upon ionizing gas-phase samples. The samples are usually introduced through a heated batch inlet, heated direct insertion probe, or a gas chromatograph. Electron Ionization (EI) Also referred to as electron impact ionization, this is the oldest and best-characterized of all the ionization methods. A beam of electrons passes through the gas-phase sample. An electron that collides with a neutral analyte molecule can knock off another electron, resulting in a positively charged ion. The ionization process can either produce a molecular ion which will have the same
3 molecular weight and elemental composition of the starting analyte, or it can produce a fragment ion which corresponds to a smaller piece of the analyte molecule. The ionization potential is the electron energy that will produce a molecular ion. The appearance potential for a given fragment ion is the electron energy that will produce that fragment ion. Most mass spectrometers use electrons with an energy of 70 electron volts (ev) for EI. Decreasing the electron energy can reduce fragmentation, but it also reduces the number of ions formed.. heated batch inlet. heated direct insertion probe. gas chromatograph. liquid chromatograph (particle-beam interface). well-understood. can be applied to virtually all volatile compounds. reproducible mass spectra. fragmentation provides structural information. libraries of mass spectra can be searched for EI mass spectral "fingerprint" Limitations. sample must be thermally volatile and stable. the molecular ion may be weak or absent for many compounds.. Low Typically less than 1,000 Da.
4 Chemical Ionization (CI) Chemical ionization uses ion-molecule reactions to produce ions from the analyte. The chemical ionization process begins when a reagent gas such as methane, isobutane, or ammonia is ionized by electron impact. A high reagent gas pressure (or long reaction time) results in ion-molecule reactions between the reagent gas ions and reagent gas neutrals. Some of the products of these ion-molecule reactions can react with the analyte molecules to produce analyte ions. Example (R = reagent, S = sample, e = electron,. = radical electron, H = hydrogen): R + e ---> R+. + 2e R+. + RH ---> RH+ + R. RH+ + S ---> SH+ + R (of course, other reactions can occur). heated batch inlet. heated direct insertion probe. gas chromatograph. liquid chromatograph (particle-beam interface). often gives molecular weight information through molecular-like ions such as [M+H]+, even when EI would not produce a molecular ion.. simple mass spectra, fragmentation reduced compared to EI Limitations
5 . sample must be thermally volatile and stable. less fragmentation than EI, fragment pattern not informative or reproducible enough for library search. results depend on reagent gas type, reagent gas pressure or reaction time, and nature of sample.. Low Typically less than 1,000 Da. Desorption Chemical Ionization (DCI) This is a variation on chemical ionization in which the analyte is placed on a filament that is rapidly heated in the CI plasma. The direct exposure to the CI reagent ions, combined with the rapid heating acts to reduce fragmentation. Some samples that cannot be thermally desorbed without decomposition can be characterized by the fragments produced by pyrolysis DCI.. sample deposited onto a filament wire. filament rapidly heated inside the CI source.. reduced thermal decomposition. rapid analysis. relatively simple equipment Limitations. not particularly reproducible. rapid heating requires fast scan speeds
6 . fails for large or labile compounds Low Typically less than 1,500 Da. Negative-ion chemical ionization (NCI) Not all compounds will produce negative ions. However, many important compounds of environmental or biological interest can produce negative ions under the right conditions. For such compounds, negative ion mass spectrometryis more efficient, sensitive and selective than positive-ion mass spectrometry. Negative ions can be produced by a number of processes. Resonance electron capture refers to the capture of an electron by a neutral molecule to produce a molecular anion. The electron energy is very low, and the specific energy required for electron capture depends on the molecular structure of the analyte. Electron attachment is an endothermic process, so the resulting molecular anion will have excess energy. Some molecular anions can accommodate the excess energy. Others may lose the electron or fall apart to produce fragment anions. In negative-ion chemical ionization, a buffer gas (usually a common CI gas such as methane) is used to slow down the electrons in the electron beam until some of the electrons have just the right energy to be captured by the analyte molecules. The buffer gas can also help stabilize the energetic anions and reduce fragmentation. This is really a physical process and not a true "chemical ionization" process.. Same as for CI
7 . efficient ionization, high sensitivity. less fragmentation than positive-ion EI or CI. greater selectivity for certain environmentally or biologically important compounds Limitations. not all volatile compounds produce negative ions. poor reproducibility. Low Typically less than 1,000 Da. Field Desorption and Ionization These methods are based on electron tunneling from an emitter that is biased at a high electrical potential. The emitter is a filament on which fine crystalline 'whiskers' are grown. When a high potential is applied to the emitter, a very high electric field exists near the tips of the whiskers. There are two kinds of emitters used on JEOL mass spectrometers: carbon emitters and silicon emitters. Silicon emitters are robust, relatively inexpensive, and they can handle a higher current for field desorption. Carbon emitters are more expensive, but they can provide about an order of magnitude better sensitivity than silicon emitters. Field desorption and ionization are soft ionization methods that tend to produce mass spectra with little or no fragment-ion content.
8 Field Desorption (FD) The sample is deposited onto the emitter and the emitter is biased to a high potential (several kilovolts) and a current is passed through the emitter to heat up the filament. Mass spectra are acquired as the emitter current is gradually increased and the sample is evaporated from the emitter into the gas phase. The analyte molecules are ionized by electron tunneling at the tip of the emitter 'whiskers'. Characteristic positive ions produced are radical molecular ions and cationattached species such as [M+Na]+ and [M-Na]+. The latter are probably produced during desorption by the attachment of trace alkali metal ions present in the analyte. Direct insertion probe. The sample is deposited onto the tip of the emitter by. dipping the emitter into an analyte solution. depositing the dissolved or suspended sample onto the emitter with a microsyringe. simple mass spectra, typically one molecular or molecular-like ionic species per compound.. little or no chemical background. works well for small organic molecules, many organometallics, low molecular weight polymers and some petrochemical fractions Limitations. sensitive to alkali metal contamination and sample overloading. emitter is relatively fragile. relatively slow analysis as the emitter current is increased. the sample must be thermally volatile to some extent to be desorbed
9 . Low-moderate, depends on the sample. Typically less than about 2,000 to 3,000 Da.. some examples have been recorded from ions with masses beyond 10,000 Da. Field Ionization (FI) The sample is evaporated from a direct insertion probe, gas chromatograph, or gas inlet. As the gas molecules pass near the emitter, they are ionized by electron tunneling.. heated direct insertion probe. gas inlet. gas chromatograph. simple mass spectra, typically one molecular or molecular-like ionic species per compound.. little or no chemical background. works well for small organic molecules and some petrochemical fractions Limitations. The sample must be thermally volatile. Samples are introduced in the same way as for electron ionization (EI).. Low Typically less than 1000 Da.
10 Particle Bombardment In these methods, the sample is deposited on a target that is bombarded with atoms, neutrals, or ions. The most common approach for organic mass spectrometry is to dissolve the analyte in a liquid matrix with low volatility and to use a relatively high current of bombarding particles (FABor dynamic SIMS). Other methods use a relatively low current of bombarding particles and no liquid matrix (static SIMS). The latter methods are more commonly used for surface analysis than for organic mass spectrometry. The primaryparticle beam is the bombarding particle beam, while the secondary ions are the ions produced from bombardment of the target. Fast Atom Bombardment (FAB) The analyte is dissolved in a liquid matrix such as glycerol, thioglycerol, m-nitrobenzyl alcohol, or diethanolamine and a small amount (about 1 microliter) is placed on a target. The target is bombarded with a fast atom beam (for example, 6 kev xenon atoms) that desorb molecular-like ions and fragments from the analyte. Cluster ions from the liquid matrix are also desorbed and produce a chemical background that varies with the matrix used.. direct insertion probe. LC/MS (frit FAB or continuous-flow FAB).. rapid. simple. relatively tolerant of variations in sampling
11 . good for a large variety of compounds. strong ion currents -- good for high-resolution measurements Limitations. high chemical background defines detection limits. may be difficult to distinguish low-molecular-weight compounds from chemical background. analyte must be soluble in the liquid matrix. no good for multiply charged compounds with more than 2 charges. Moderate Typically ~300 Da to about 6000 Da. Secondary Ion Mass Spectrometry (SIMS) This discussion refers to dynamic SIMS. Dynamic SIMS is nearly identical to FAB except that the primary particle beam is an ion beam (usually cesium ions) rather than a neutral beam. The ions can be focused and accelerated to higher kinetic energies than are possible for neutral beams, and sensitivity is improved for higher masses. The use of SIMS for moderate-size ( ,000 Da) proteins and peptides has largely been supplanted by electrospray ionization.. Same as for FAB
12 . Same as for FAB, except sensitivity is improved for higher masses (3000 to 13,000 Da). Limitations. Same as for FAB except. target can get hotter than in FAB due to more energetic primary beam. high-voltage arcs more common than FAB. ion source usually requires more maintenance than FAB. Moderate Typically 300 to 13,000 Da. Atmospheric Pressure Ionization (Spray Methods) In these methods, a solution containing the analyte is sprayed at atmospheric pressure into an interface to the vacuum of the mass spectrometer ion source. A combination of thermal and pneumatic means is used to desolvate the ions as they enter the ion source. Solution flow rates can range from less than a microliter per minute to several milliliters per minute. These methods are well-suited for flow-injection and LC/MS techniques.
13 Electrospray Ionization (ESI) The sample solution is sprayed across a high potential difference (a few kilovolts) from a needle into an orifice in the interface. Heat and gas flows are used to desolvate the ions existing in the sample solution. Electrospray ionization can produce multiply charged ions with the number of charges tending to increase as the molecular weight increases. The number of charges on a given ionic species must be determined by methods such as:. comparing two charge states that differ by one charge and solving simultaneous equations. looking for species that have the same charge but different adduct masses. examining the mass-to-charge ratios for resolved isotopic clusters. flow injection. LC/MS. typical flow rates are less than 1 microliter per minute up to about a milliliter per minute. good for charged, polar or basic compounds. permits the detection of high-mass compounds at mass-to-charge ratios that are easily determined by most mass spectrometers (m/z typically less than 2000 to 3000).. best method for analyzing multiply charged compounds. very low chemical background leads to excellent detection limits. can control presence or absence of fragmentation by controlling the interface lens potentials. compatible with MS/MS methods
14 Limitations. multiply charged species require interpretation and mathematical transformation (can sometimes be difficult). complementary to APCI. No good for uncharged, non-basic, low-polarity compounds (e.g.steroids). very sensitive to contaminants such as alkali metals or basic compounds. relatively low ion currents. relatively complex hardware compared to other ion sources. Low-high Typically less than 200,000 Da. Atmospheric Pressure Chemical Ionization (APCI) Similar interface to that used for ESI. In APCI, a corona discharge is used to ionize the analyte in the atmospheric pressure region. The gas-phase ionization in APCI is more effective than ESI for analyzing less-polar species. ESI and APCI are complementary methods.. same as for electrospray ionization. good for less-polar compounds. excellent LC/MS interface. compatible with MS/MS methods
15 Limitations. complementary to ESI.. Low-moderate Typically less than 2000 Da. Laser Desorption Laser desorption methods use a pulsed laser to desorb species from a target surface. Therefore, one must use a mass analyzer such as time-of-flight (TOF) or Fourier transform ion cyclotron resonance (FTICR) that is compatible with pulsed ionization methods. Magnetic sector mass spectrometers equipped with an array detector can also be used for the detection of ions produced by MALDI. Direct laser desorption relies on the very rapid heating of the sample or sample substrate to vaporize molecules so quickly that they do not have time to decompose. This is good for low to medium-molecular weight compounds and surface analysis. The more recent development of matrix-assisted laser desorption ionization (MALDI) relies on the absorption of laser energy by a matrix compound. MALDI has become extremely popular as a method for the rapid determination of high-molecular-weight compounds. Matrix-Assisted Laser Desorption Ionization (MALDI) The analyte is dissolved in a solution containing an excess of a matrix such as sinapinic acid or dihydroxybenzoic acid that has a chromophore that absorbs at the laser wavelength. A small
16 amount of this solution is placed on the laser target. The matrix absorbs the energy from the laser pulse and produces a plasma that results in vaporization and ionization of the analyte.. direct insertion probe. continuous-flow introduction. rapid and convenient molecular weight determination Limitations. MS/MS difficult. requires a mass analyzer that is compatible with pulsed ionization techniques. not easily compatible with LC/MS. Very high Typically less than 500,000 Da. Appendices: CI Reagent Gases Methane:
17 . good for most organic compounds. usually produces [M+H]+, [M+CH3]+ and [M+C3H5]+ adducts. adducts are not always abundant. extensive fragmentation Isobutane:. usually produces [M+H]+, [M+C4H9]+ adducts and some fragmentation. adducts are relatively more abundant than for methane CI. not as universal as methane Ammonia:. fragmentation virtually absent. polar compounds produce [M+NH4]+ adducts. basic compounds produce [M+H]+ adducts. non-polar and non-basic compounds are not ionized. Click here to view a JPEG file (~19K bytes) that shows a positive-ion CI reagent-gas mass spectrum of ammonia Direct formation of negative ions - a better approach to NCI? By directly controlling the energy of the ionizing electrons, one can select the exact electron energy required for electron capture or fragment ion formation from a specific compound. This avoids the problems associated with the NCI approach, where the ionizing plasma can contain electrons with poorly defined energies as well as neutral molecules and other ionic species. JEOL USA, Inc. is supporting research using a patented electron monochromator that has been developed by J. Laramee, M. Deinzer and coworkers at Oregon State University. This device will allow the ionizing electron energy to be precisely controlled for negative-ion mass spectrometry, leading to more sensitive, selective, and reproducible mass spectra for environmentally important compounds.
18 FAB matrices glycerol. C3H8O3, m/z first and most widely-used FAB matrix.. best choice for polar compounds 1-thioglycerol. C3H8O2S, m/z similar uses to glycerol, but may enhance M+1 abundance. more volatile than glycerol, evaporates quickly 3-nitrobenzyl alcohol (NBA). C7H7NO3, m/z best choice for less polar compounds ( e.g. chlorophyll) and many organometallics 2-nitrophenyl octyl ether (NPOE). C14H21NO3, m/z only FAB matrix with no reactive hydrogen. good choice for reactive compounds such as organometallics Note: this example contains trace PEG 600 contamination triethanolamine. C6H15NO3, m/z good matrix for negative-ion FAB. enhances [M-H]- formation.
19 "magic bullet". 5:1 mixture of dithiothreitol and dithioerythritol (DL-threo- and erythro-1,4- dimercapto-2,3-butanediol). C4H10S2O2, m/z dioctyl phthalate (DOP). bis(2-ethylhexyl) phthalate. C24H38O4, m/z Similar in use to NPOE. Widespread contaminant in solvents, gives characteristic 149 peak in EI mass spectra. Use care to avoid contaminating the mass spectrometer sulfuric acid. H2SO4, m/z good for some inorganics and organometallics (e.g. copper phthalocyanine). corrosive, use care
Electrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass
Protons, neutrons and electrons Nuclear Structure particle relative charge relative mass proton 1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass Protons and neutrons make up
Getting the most from this book...4 About this book...5
Contents Getting the most from this book...4 About this book....5 Content Guidance Topic 1 Atomic structure and the periodic table...8 Topic 2 Bonding and structure...14 Topic 2A Bonding....14 Topic 2B
Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III)
Electrospray Mass Spectrometry of Tris-acetylacetonatochromium(III) Purpose: Verify the synthesis of the synthesis product, Cr(acac) 3. The presence of any urea containing byproducts will also be determined.
CHE334 Identification of an Unknown Compound By NMR/IR/MS
CHE334 Identification of an Unknown Compound By NMR/IR/MS Purpose The object of this experiment is to determine the structure of an unknown compound using IR, 1 H-NMR, 13 C-NMR and Mass spectroscopy. Infrared
Mass Spectrometry. Overview
Mass Spectrometry Overview Mass Spectrometry is an analytic technique that utilizes the degree of deflection of charged particles by a magnetic field to find the relative masses of molecular ions and fragments.2
electron does not become part of the compound; one electron goes in but two electrons come out.
Characterization Techniques for Organic Compounds. When we run a reaction in the laboratory or when we isolate a compound from nature, one of our first tasks is to identify the compound that we have obtained.
Summary of the characteristics of different mass analyzers
Summary of the characteristics of different mass analyzers All mass spectrometers combine ion formation, mass analysis, and ion detection. This discussion is concerned with how various mass analyzers are
Vacuum Evaporation Recap
Sputtering Vacuum Evaporation Recap Use high temperatures at high vacuum to evaporate (eject) atoms or molecules off a material surface. Use ballistic flow to transport them to a substrate and deposit.
Composition of the Atmosphere. Outline Atmospheric Composition Nitrogen and Oxygen Lightning Homework
Molecules of the Atmosphere The present atmosphere consists mainly of molecular nitrogen (N2) and molecular oxygen (O2) but it has dramatically changed in composition from the beginning of the solar system.
CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING
CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING Essential Standard: STUDENTS WILL UNDERSTAND THAT THE PROPERTIES OF MATTER AND THEIR INTERACTIONS ARE A CONSEQUENCE OF THE STRUCTURE OF MATTER,
Mass Spec - Fragmentation
Mass Spec - Fragmentation An extremely useful result of EI ionization in particular is a phenomenon known as fragmentation. The radical cation that is produced when an electron is knocked out of a neutral
3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
Xcalibur Versions 2.1.0 2.3.0
Thermo Xcalibur Versions 2.1.0 2.3.0 Qualitative Analysis User Guide XCALI-97211 Revision E August 2012 2012 Thermo Fisher Scientific Inc. All rights reserved. Xcalibur, Surveyor, and Accela are registered
Background Information
1 Gas Chromatography/Mass Spectroscopy (GC/MS/MS) Background Information Instructions for the Operation of the Varian CP-3800 Gas Chromatograph/ Varian Saturn 2200 GC/MS/MS See the Cary Eclipse Software
Basics of LC/MS. A Primer
Basics of LC/MS A Primer Contents Introduction..................................... 2 Instrumentation.................................. 4 Interfacing LC and MS................................. 4 Atmospheric
Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
How To Use Gc-Ms
The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1 Speakers John Hinshaw GC Dept. Dean CHROMacademy Tony Taylor Technical Director Crawford Scientific Moderator M ( g ) e M ( g ) 2e Peter
States of Matter CHAPTER 10 REVIEW SECTION 1. Name Date Class. Answer the following questions in the space provided.
CHAPTER 10 REVIEW States of Matter SECTION 1 SHORT ANSWER Answer the following questions in the space provided. 1. Identify whether the descriptions below describe an ideal gas or a real gas. ideal gas
TOF FUNDAMENTALS TUTORIAL
TOF FUNDAMENTALS TUTORIAL Presented By: JORDAN TOF PRODUCTS, INC. 990 Golden Gate Terrace Grass Valley, CA 95945 530-272-4580 / 530-272-2955 [fax] www.rmjordan.com [web] [email protected] [e-mail] This
Infrared Spectroscopy 紅 外 線 光 譜 儀
Infrared Spectroscopy 紅 外 線 光 譜 儀 Introduction Spectroscopy is an analytical technique which helps determine structure. It destroys little or no sample (nondestructive method). The amount of light absorbed
F321 THE STRUCTURE OF ATOMS. ATOMS Atoms consist of a number of fundamental particles, the most important are... in the nucleus of an atom
Atomic Structure F32 TE STRUCTURE OF ATOMS ATOMS Atoms consist of a number of fundamental particles, the most important are... Mass / kg Charge / C Relative mass Relative Charge PROTON NEUTRON ELECTRON
Chapter 1: Moles and equations. Learning outcomes. you should be able to:
Chapter 1: Moles and equations 1 Learning outcomes you should be able to: define and use the terms: relative atomic mass, isotopic mass and formula mass based on the 12 C scale perform calculations, including
Best Practice Guide. for Generating Mass Spectra. Authors: Vicki Barwick John Langley Tony Mallet Bridget Stein Ken Webb
Best Practice Guide for Generating Mass Spectra Best Practice Guide for Generating Mass Spectra Authors: Vicki Barwick John Langley Tony Mallet Bridget Stein Ken Webb BEST PRACTICE GUIDE FOR GENERATING
0 10 20 30 40 50 60 70 m/z
Mass spectrum for the ionization of acetone MS of Acetone + Relative Abundance CH 3 H 3 C O + M 15 (loss of methyl) + O H 3 C CH 3 43 58 0 10 20 30 40 50 60 70 m/z It is difficult to identify the ions
Chemical Basis of Life Module A Anchor 2
Chemical Basis of Life Module A Anchor 2 Key Concepts: - Water is a polar molecule. Therefore, it is able to form multiple hydrogen bonds, which account for many of its special properties. - Water s polarity
KINETIC MOLECULAR THEORY OF MATTER
KINETIC MOLECULAR THEORY OF MATTER The kinetic-molecular theory is based on the idea that particles of matter are always in motion. The theory can be used to explain the properties of solids, liquids,
Introduction 1 Christopher Barner-Kowollik, Jana Falkenhagen, Till Gruendling, and Steffen Weidner References 4
V Contents List of Contributors XIII Introduction 1 Christopher Barner-Kowollik, Jana Falkenhagen, Till Gruendling, and Steffen Weidner References 4 1 Mass Analysis 5 Gene Hart-Smith and Stephen J. Blanksby
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Chapter 2: The Chemical Context of Life
Chapter 2: The Chemical Context of Life Name Period This chapter covers the basics that you may have learned in your chemistry class. Whether your teacher goes over this chapter, or assigns it for you
QGA Quantitative Gas Analyser
QGA Quantitative Gas Analyser A compact bench-top system for real-time gas and vapour analysis Detailed product information / introduction catalysis studies environmental gas analysis fermentation off-gas
Properties and Classifications of Matter
PS-3.1 Distinguish chemical properties of matter (including reactivity) from physical properties of matter (including boiling point, freezing/melting point, density [with density calculations], solubility,
Analyzing Small Molecules by EI and GC-MS. July 2014
Analyzing Small Molecules by EI and GC-MS July 2014 Samples Appropriate for GC-MS Volatile and semi-volatile organic compounds Rule of thumb,
Copyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
Corrosion of Copper in Water
Corrosion of Copper in Water Digby D. Macdonald Center for Electrochemical Science and Technology Department of Materials Science and Engineering Pennsylvania State University 209 Steidle Bldg University
Amount of Substance. http://www.avogadro.co.uk/definitions/elemcompmix.htm
Page 1 of 14 Amount of Substance Key terms in this chapter are: Element Compound Mixture Atom Molecule Ion Relative Atomic Mass Avogadro constant Mole Isotope Relative Isotopic Mass Relative Molecular
Micromass LCT User s Guide
Micromass LCT User s Guide 1) Log on to MassLynx with your username & password. 2) After you have logged in, the MassLynx software will automatically run. 3) After MassLynx has come up, open your project
Eleanor Riches Waters Corporation, Manchester, UK INTRODUCTION
X E V O T Q MS W IT H AT MOS P H E RIC P R E S SU R E P HOT O IO NIZ AT IO N (A P P I) SOU RC E: T H E IONIZATION OF COMPOUNDS WITH DIVERSE STRUCTURES USING VITAMINS AS A MODEL Eleanor Riches Waters Corporation,
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis
Accurate calibration of on-line Time of Flight Mass Spectrometer (TOF-MS) for high molecular weight combustion product analysis B. Apicella*, M. Passaro**, X. Wang***, N. Spinelli**** [email protected]
High resolution mass spectrometry (HRMS*) in Graz
High resolution mass spectrometry (HRMS*) in Graz 10.2.2012 * Instruments with high resolution where exact mass measurements can be performed 1 HRMS in Graz s and Ionization Techniques 2 6 3 5 4 1 TU Graz
Classification of Chemical Substances
Classification of Chemical Substances INTRODUCTION: Depending on the kind of bonding present in a chemical substance, the substance may be called ionic, molecular or metallic. In a solid ionic compound
Chemistry Diagnostic Questions
Chemistry Diagnostic Questions Answer these 40 multiple choice questions and then check your answers, located at the end of this document. If you correctly answered less than 25 questions, you need to
Separation of Amino Acids by Paper Chromatography
Separation of Amino Acids by Paper Chromatography Chromatography is a common technique for separating chemical substances. The prefix chroma, which suggests color, comes from the fact that some of the
[ Care and Use Manual ]
PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample
ProteinChip Energy Absorbing Molecules (EAM)
ProteinChip Energy Absorbing Molecules (EAM) Instruction Manual Catalog #C30-00001, #C30-00002, #C30-00003, #C30-00004 For technical support, call your local Bio-Rad office, or in the US, call 1-800-4BIORAD
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
Coating Technology: Evaporation Vs Sputtering
Satisloh Italy S.r.l. Coating Technology: Evaporation Vs Sputtering Gianni Monaco, PhD R&D project manager, Satisloh Italy 04.04.2016 V1 The aim of this document is to provide basic technical information
Lecture 11. Etching Techniques Reading: Chapter 11. ECE 6450 - Dr. Alan Doolittle
Lecture 11 Etching Techniques Reading: Chapter 11 Etching Techniques Characterized by: 1.) Etch rate (A/minute) 2.) Selectivity: S=etch rate material 1 / etch rate material 2 is said to have a selectivity
5. Which temperature is equal to +20 K? 1) 253ºC 2) 293ºC 3) 253 C 4) 293 C
1. The average kinetic energy of water molecules increases when 1) H 2 O(s) changes to H 2 O( ) at 0ºC 3) H 2 O( ) at 10ºC changes to H 2 O( ) at 20ºC 2) H 2 O( ) changes to H 2 O(s) at 0ºC 4) H 2 O( )
CH3 Stoichiometry. The violent chemical reaction of bromine and phosphorus. P.76
CH3 Stoichiometry The violent chemical reaction of bromine and phosphorus. P.76 Contents 3.1 Counting by Weighing 3.2 Atomic Masses 3.3 The Mole 3.4 Molar Mass 3.5 Percent Composition of Compounds 3.6
5s Solubility & Conductivity
5s Solubility & Conductivity OBJECTIVES To explore the relationship between the structures of common household substances and the kinds of solvents in which they dissolve. To demonstrate the ionic nature
Basics of LC/MS. Primer
Basics of LC/MS Primer Contents Why Liquid Chromatography/Mass Spectrometry? 4 Instrumentation 6 Ion Sources 6 Electrospray ionization 7 Atmospheric pressure chemical ionization 8 Atmospheric pressure
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert1, Michel Lhermitte 1, Frederic Grisel 2 1
GENERAL UNKNOWN SCREENING FOR DRUGS IN BIOLOGICAL SAMPLES BY LC/MS Luc Humbert, Michel Lhermitte, Frederic Grisel Laboratoire de Toxicologie & Génopathologie, CHRU Lille, France Waters Corporation, Guyancourt,
Chapter 3. Mass Relationships in Chemical Reactions
Chapter 3 Mass Relationships in Chemical Reactions This chapter uses the concepts of conservation of mass to assist the student in gaining an understanding of chemical changes. Upon completion of Chapter
Chemical Sputtering. von Kohlenstoff durch Wasserstoff. W. Jacob
Chemical Sputtering von Kohlenstoff durch Wasserstoff W. Jacob Centre for Interdisciplinary Plasma Science Max-Planck-Institut für Plasmaphysik, 85748 Garching Content: Definitions: Chemical erosion, physical
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
In the box below, draw the Lewis electron-dot structure for the compound formed from magnesium and oxygen. [Include any charges or partial charges.
Name: 1) Which molecule is nonpolar and has a symmetrical shape? A) NH3 B) H2O C) HCl D) CH4 7222-1 - Page 1 2) When ammonium chloride crystals are dissolved in water, the temperature of the water decreases.
Chemistry. The student will be able to identify and apply basic safety procedures and identify basic equipment.
Chemistry UNIT I: Introduction to Chemistry The student will be able to describe what chemistry is and its scope. a. Define chemistry. b. Explain that chemistry overlaps many other areas of science. The
Identification of Surfactants by Liquid Chromatography-Mass Spectrometry
Identification of Surfactants by Liquid Chromatography-Mass Spectrometry James Little, Eastman Chemical Company, Kingsport, TN American Society for Mass Spectrometry 2004 Nashville, TN 1 verview Surfactant
Introduction to Geiger Counters
Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
L. Yu et al. Correspondence to: Q. Zhang ([email protected])
Supplement of Atmos. Chem. Phys. Discuss., 5, 967 974, 5 http://www.atmos-chem-phys-discuss.net/5/967/5/ doi:.594/acpd-5-967-5-supplement Author(s) 5. CC Attribution. License. Supplement of Molecular transformations
Study the following diagrams of the States of Matter. Label the names of the Changes of State between the different states.
Describe the strength of attractive forces between particles. Describe the amount of space between particles. Can the particles in this state be compressed? Do the particles in this state have a definite
amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS
amazon SL Setting New Standards in Performance, Simplicity and Value Innovation with Integrity Ion Trap Best-In-Class Ion Trap Mass Spectrometer for Routine Analysis The amazon SL entry-level system is
Review - After School Matter Name: Review - After School Matter Tuesday, April 29, 2008
Name: Review - After School Matter Tuesday, April 29, 2008 1. Figure 1 The graph represents the relationship between temperature and time as heat was added uniformly to a substance starting at a solid
Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System
Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System Carlton G. Slough TA Instruments, 109 Lukens Drive, New Castle DE 19720, USA ABSTRACT The use of
Structure and Properties of Atoms
PS-2.1 Compare the subatomic particles (protons, neutrons, electrons) of an atom with regard to mass, location, and charge, and explain how these particles affect the properties of an atom (including identity,
Phase diagram of water. Note: for H 2 O melting point decreases with increasing pressure, for CO 2 melting point increases with increasing pressure.
Phase diagram of water Note: for H 2 O melting point decreases with increasing pressure, for CO 2 melting point increases with increasing pressure. WATER Covers ~ 70% of the earth s surface Life on earth
TECHNICAL BULLETIN. ProteoMass Peptide & Protein MALDI-MS Calibration Kit. Catalog Number MSCAL1 Store at Room Temperature
ProteoMass Peptide & Protein MALDI-MS Calibration Kit Catalog Number MSCAL1 Store at Room Temperature TECHNICAL BULLETIN Description This kit provides a range of standard peptides and proteins for the
Performing Calculatons
Performing Calculatons There are three basic units for measurement in the organic laboratory mass, volume, and number, measured in moles. Most of the other types of measurements are combinations of them,
hij GCSE Additional Science Chemistry 2 Higher Tier Chemistry 2H SPECIMEN MARK SCHEME Version 1.0
hij GCSE Additional Science Chemistry 2 Higher Tier Chemistry 2H SPECIMEN MARK SCHEME Version.0 Copyright 20 AQA and its licensors. All rights reserved. The Assessment and Qualifications Alliance (AQA)
AP Chemistry A. Allan Chapter 8 Notes - Bonding: General Concepts
AP Chemistry A. Allan Chapter 8 Notes - Bonding: General Concepts 8.1 Types of Chemical Bonds A. Ionic Bonding 1. Electrons are transferred 2. Metals react with nonmetals 3. Ions paired have lower energy
Nuclear Physics. Nuclear Physics comprises the study of:
Nuclear Physics Nuclear Physics comprises the study of: The general properties of nuclei The particles contained in the nucleus The interaction between these particles Radioactivity and nuclear reactions
Chapter 2 Atoms, Ions, and the Periodic Table
Chapter 2 Atoms, Ions, and the Periodic Table 2.1 (a) neutron; (b) law of conservation of mass; (c) proton; (d) main-group element; (e) relative atomic mass; (f) mass number; (g) isotope; (h) cation; (i)
Chapter 3: Water and Life
Name Period Chapter 3: Water and Life Concept 3.1 Polar covalent bonds in water result in hydrogen bonding 1. Study the water molecules at the right. On the central molecule, label oxygen (O) and hydrogen
An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry
Limits to Detection/Measurement Ionization is fundamentally a Probabilistic Process Just like radioactive decay So is transmission through the analyzer There is an intrinsic statistical uncertainty Proportional
IB Chemistry. DP Chemistry Review
DP Chemistry Review Topic 1: Quantitative chemistry 1.1 The mole concept and Avogadro s constant Assessment statement Apply the mole concept to substances. Determine the number of particles and the amount
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
Certificate of Analysis
The Japan Society for Analytical Chemistry Certificate of Analysis Certified Reference Material JSAC 0602-3 Plastics (tip form) for Chemical Analysis of Hazardous Metals This certified reference material
Secondary Ion Mass Spectrometry
Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown,
Ultraviolet Spectroscopy
Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer
Physical Changes and Chemical Reactions
Physical Changes and Chemical Reactions Gezahegn Chaka, Ph.D., and Sudha Madhugiri, Ph.D., Collin College Department of Chemistry Objectives Introduction To observe physical and chemical changes. To identify
Chapter 11 Properties of Solutions
Chapter 11 Properties of Solutions 11.1 Solution Composition A. Molarity moles solute 1. Molarity ( M ) = liters of solution B. Mass Percent mass of solute 1. Mass percent = 1 mass of solution C. Mole
Mass spectrometry. Dr. Kevin R. Tucker
Mass spectrometry Dr. Kevin R. Tucker What is mass spectrometry? A versatile analytical technique used to determine the composition of chemical samples either at the atomic or molecular level. Ionic forms
Chapter 4: Structure and Properties of Ionic and Covalent Compounds
Chapter 4: Structure and Properties of Ionic and Covalent Compounds 4.1 Chemical Bonding o Chemical Bond - the force of attraction between any two atoms in a compound. o Interactions involving valence
Part B 2. Allow a total of 15 credits for this part. The student must answer all questions in this part.
Part B 2 Allow a total of 15 credits for this part. The student must answer all questions in this part. 51 [1] Allow 1 credit for 3 Mg(s) N 2 (g) Mg 3 N 2 (s). Allow credit even if the coefficient 1 is
Name Class Date. In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question.
Assessment Chapter Test A Chapter: States of Matter In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question. 1. The kinetic-molecular
Elements, Atoms & Ions
Introductory Chemistry: A Foundation FOURTH EDITION by Steven S. Zumdahl University of Illinois Elements, Atoms & Ions Chapter 4 1 2 Elements Aims: To learn about the relative abundances of the elements,
Organic Chemistry Calculations
Organic Chemistry Calculations There are three basic units for measurement in the organic laboratory mass, volume, and number, measured in moles. Most of the other types of measurements are combinations
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
Ion Beam Sputtering: Practical Applications to Electron Microscopy
Ion Beam Sputtering: Practical Applications to Electron Microscopy Applications Laboratory Report Introduction Electron microscope specimens, both scanning (SEM) and transmission (TEM), often require a
@ Oxford Fajar Sdn. Bhd. (008974-T) 2012. Matter. 1.1 Atoms and Molecules 1.2 Mole Concept 1.3 Stoichiometry
1 Matter 1.1 Atoms and Molecules 1.2 Mole Concept 1.3 Stoichiometry 2 Chemistry for Matriculation Semester 1 1.1 Atoms and Molecules LEARNING OUTCOMES Describe proton, electron and neutron in terms of
CHEM 120 Online Chapter 7
CHEM 120 Online Chapter 7 Date: 1. Which of the following statements is not a part of kinetic molecular theory? A) Matter is composed of particles that are in constant motion. B) Particle velocity increases
SCION SQ Series GC-MS
SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative