Physics in Nuclear Medicine
|
|
|
- Maria Page
- 10 years ago
- Views:
Transcription
1 ~.~ ~ Physics in Nuclear Medicine Second Edition James A. Sorenson, Ph.D. Director, Medical Physics Professor of Radiology Depat1meht of Radiology University of Utah Medical Center Salt Lake City, Utah Michael E. Phelps, Ph.D. Jennifer Jones Simon Professor Chief, Division of Nuclear Medicine and Biophysics Depat1ment of Radiological Sciences UCLA School of Medicine Chief, Laboratory of Nuclear Medicine Laboratory of Biomedical and Environmental Sciences, UCLA Los Angeles, California WeB. SAUNDERS COMPANY Harcourt Brace Jovanovich, Inc. Philadelphia London Toronto Montreal Sydney Tokyo
2 16 The Anger Camera: Performance Characteristics B. DESIGN AND PERFORMANCE CHARACTERISTICS OF PARALLEL-HOLE COLLIMATORS 1. Basic Limitations in Collimator Performance The collimator is a "weak link" for the performance of an Anger camera system, as indeed it is in any nuclear medicine imaging system employing the principles of absorptive collimation. Collimator efficiency, defined as the fraction of'y rays striking the collimator that actually pass through it to project the 'Y-ray image onto the detector, is typically only a few percent or less. Collimator resolution, which refers to the sharpness or detail of the 'Y-ray image projected onto the detector, also is rather poor, generally worse than the intrinsic resolution of the camera detector and electronics. Because it is a limiting factor in camera system performance, it is important that the collimator be designed carefully. Poor design can result only in poorer overall performance. Design considerations for parallelhole collimators will be discussed in this section. Design characteristics for converging and diverging collimators are similar to those of the parallel-hole type. Design characteristics of pinhole collimators will not be discussed in detail here but are described in references listed at the end of th~ chapter. The analysis to be presented for parallel-hole collimators
3 332 Physics in Nuclear Medicine is similar to that presented by Anger in reference 1, which may be consulted for a more detailed discussion. 2. Septal Thickness A primary consideration in collimator design is to ensure that septal penetration by "Y rays crossing from one collimator hole into another is negligibly small. This is essential if an accurate "Y-ray image is to be projected by the collimator onto the camera detector. No thickness of septal material is sufficient to stop all "Y rays, so the usual criteria is to accept some reasonably small level of septal penetration, e.g., -5 percent. The required septal thickness may be determined by analysis of Figure The shortest path length for "Y rays to travel from one hole to the next is w. Septal thickness t is related to w, and to the length I and diameter d of the collimator holes, by! t = 2dw/(1 - w) (16-1) If septal penetration is to be less than 5 percent, the transmission factor for the thickness w must be e-ij.w:$ 0.05 (16-2) where IJ. is the linear attenuation coefficient of the septal material. Since e , this implies IJ.w ~ 3 (16-3) w > 3/IJ. (16-4)...,.::.:-.:,:.:: :.;.',: LLIMATOR SEPTA I-d-l I-t Fig Minim~m path length w for a 'Y ray passing through the collimator septa from one hole to the next depends on length I and diameter d of the collimator holes and on septal thicknesses t.
4 The Anger Camera: Performance Characteristics 333 and thus 6d/fJ. t ~ (16-5) I - (3/fJ.) It is desirable that septal thickness t be as small as possible so that the collimator septa obstruct the smallest possible area of detector surface and collimator efficiency is maximized. This objective is realized by using a material with a large value of fj. for the collimator septa. Materials of high atomic number Z and high density p are preferred. Lead (Z = 82, p = g/cm3) is the material of choice for reasons of cost and availability; however, other materials, including tantalum (Z = 73, p = 16.6 g/cm3), tungsten (Z = 74, p = 19.4 g/cm3), and even gold (Z = 79, p = 19.3 g/cm3) have been employed in experimental applications. Attenuation coefficients of heavy elements depend strongly on 'V-ray energy in the nuclear medicine energy range (Chapter 9, Section B.l). Thus the required septal thickness also depends strongly on the 'V-ray energy for which the collimator is designed to be used. Commercially available collimators are categorized according to the maximum 'V-ray energy for which their septal thickness is considered to be adequate. Low-energy collimators generally have an upper limit of about 150 ke V arid medium-energy collimators about 400 ke V. High-energy collimators (e.g., 1 MeV) are not available commercially except by special order. Example Calculate the septal thickness required for low-energy (150 key) and medium-energy (400 key) lead collimators having hole diameters 0.25 cm and lengths 2.5 cm. Answer. The linear attenuation coefficient of lead at 150 key is fj./ = 1.89 cm2/g x g/cm3 = cm-l and at 400 key is fj./ = 0.22 cm2/g x g/cm3 = 2.49 cm-l (Appendix D). Therefore from Equation 16-5 for the low-energy collimator t~ 6 x 0.25/ and for the medium-energy collimator ~ cm ","
5 334 Physics in Nuclear Medicine t~ 6 X 0.25/ ~ cm Thickness needed for low-energy collimators are only a few tenths of a millimeter, which is in the range of lead' 'foil" thicknesses and approaches the limits of lead thicknesses that can be used without loss of necessary mechanical strength. Indeed, low-energy collimators generally are quite fragile, and their septa can be damaged easily by mechanical abuse (dropping, stacking on sharp objects, etc.). Medium-energy collimators require substantially greater thicknesses, typically a few millimeters of lead. Low-energy 'Y-ray emiters (e.g., 99mTc, 140 key) can be imaged using medium-energy collimators. This is done, however, with an unnecessary sacrifice of collimator efficiency because the collimator septa are unnecessarily thick. (See Table 16-2 for comparative efficiencies of low- and medium-energy collimators.) Low-energy collimators are used whenever possible to obtain maximum collimator efficiency. When choosing a collimator, however, one must consider not only the energy of the 'Y rays to be imaged but also the energies of any other 'Y rays emitted by the radionuclide of interest or by other radionuclides that may be present as well, e.g., residual activity from another study or radionuclide impurities. Higher-energy 'Y rays may be recorded by Compton downscatter into a lower-energy analyzer window. If the collimator septa are too thin, the collimator may be virtually transparent to higher-energy 'Y rays, causing a relatively intense "foggy" background image to be superimposed on the desired image, with a resulting loss of image contrast. Whether or not a low-energy collimator can be used when higher-energy 'Y rays are pr~sent depends on the energy and intensity of those emissions and requires experimental evaluation in specific cases. 3. Geometry of Collimator Holes Collimator performance also is affected by the geometry of the collimator holes, specifically, their shape, length, and diameter. The preferred hole shape, to maximize the exposed area of detector surface for a given septal thickness, is round or hexagonal, with the holes arranged in a close-packed hexagonal array. Square and triangular holes also have been used.
6 The Anger Camera: Performance Characteristics 335 Collimator hole length and diameter affect strongly both collimator resolution and collimator efficiency. Collimator resolution Rcis defined as the full width at half-maximum (FWHM) of the radiation profile from a point or line source of radiation projected by the collimator onto the detector (Figure 16-12). This profile is also called the point or line spread function (PSF or LSF). Collimator resolution Rc is given byt Rc = d(le + b)/le (16-6) where b is the distance from the radiation source to the collimator, and d is the diameter and Ie = 1-2~-1 the "effective length" of the collimator holes (Figure 16-12). Here ~ is the linear attenuation coefficient of the collimator material. Th'e effective length of the collimator holes is somewhat less than their actual length due to septal penetration. From Example 16-1 it can be seen that for low-energy collimators (150 key), the difference between effective and actual length is about 0.1 cm whereas for medium-energy collimators it is about 0.8 cm. t It should be noted that some versions of Equation 16-6 include additional correction terms involving the thickness of the detector crystal, reflecting the fact that the image actually is formed at some depth within the detector crystal. Because photons of different energies penetrate to different average depths within the crystal, the correction actually is photon-energy dependent, a point not noted by most authors. The correction is small and for simplicity is omitted from Equation 16-6, as well as from Equations and for the converging and diverging collimators presented later in this chapter. COLLIMATOR RESOLUTION,-, /, PROJECTED,// \\ RADIATION -I \ -FWHM PROFI LE : \\ \ I, I', ' III 1\) ~1111 tl MULTIHOLE \ :-- COLLIMATOR \\ ; d \ I \ I \ I \ I \ / \~' \ I I I II POINT OR LINE SOURCE Fig Radiation profile (point or line spread function, PSF or LSF) for a parallel-hole collimator. The FWHM (full width at half-maximum) of the profile is used to characterize collimator resolution. b
7 336 Physics in Nuclear Medicine Example Calculate the resolution (FWHM) of the low-energy collimator described in Example 16-1, at source depths b = 0 and b = 10 cm, assuming it has a septal thickness of 0.03 cm. Answer. The effective length of the collimator is Thus, for b = 0 and at b = 10 cm Ie = 2.5 cm - (2/21.43) cm = 2.4 cm Rc = 0.25( )/2.4 cm = 0.25 cm Rc = 0.25( )/2.4 cm = 1.3 cm This example illustrates the strong dependence of collimator resolution on source distance from the collimator. Collimator efficiency g, defined as the fraction of 'Y rays passing through the collimator per 'Y ray emitted by the source, is given by g = ~(d/le)2[j2/(d + t)2] (16-7) where t is septal thickness and K is a constant that depends on hole shape (~0.24 for round holes in a hexagonal array, ~0.26 for hexagonal holes in a hexagonal array, ~0.28 for square holes in a square array)! Equation 16-7 applies to a source in air and assumes no attenuation of radiation by intervening body tissues. Several aspects of Equations 16-6 and 16-7 should be noted. First, resolution improves as the ratio of hole diameter to effective length (d/l;) is made smaller. Long, narrow holes provide- image with the best resolution; however, collimator efficiency decreases approximately as the square of the ratio of hole diameter to length, (d/i;)2. Thus an approximate relationship between collimator resolution Rc and efficiency g is g oc R~ (16-8)
8 ~ The Anger Camera: Performance Characteristics 337 Example Calculate the efficiency g of the collimator described in Examples 16-1 and 16-2, assuming it has hexagonal holes in an hexagonal array. Answer. For hexagonal holes in a hexagonal array, K = 0.26 g = (0.26)2 (0.25/2.4)2 [(0.25)2 / ( )2] = (0.0676) x (0.0109) x (0.797) = 5.85 x 10-4 (photons transmitted/photons emitted) This example illustrates the relatively small fraction of emitted 'Y rays that are transmitted by a typical Anger camera collimator. Therefore, for a given septal thickness, collimator resolution is improved only at the expense of decreased collimator efficiency, and vice versa. The implications of this important tradeoff in collimator design are discussed further in Chapter 18, Section C.2. Equation 16-7 also demonstrates the effect of septal thickness on efficiency. As noted in Section B.l, medium-energy collimators have lower efficiencies than low-energy collimators because of their greater septal thicknesses. lq addition to providing low- and medium-energy collimators, manufacturers of Anger camera systems also provide a selection of collimators with different combinations of resolution and efficiency. Those with good resolution but poor efficiency generally are described as "high resolution" collimators, whereas those with the opposite characteristics are described as "high sensitivity" collimators. Those with characteristics intermediate to the extremes are referred to as "general purpose," "all purpose," or by other similar names. Equation 16-6 indicates that collimator resolution becomes poorer as source-to-collimator distance b increases. Thus structures closest to the collimator are imaged with sharpest detail. Figure shows graphically the'relationship between collimator resolution and source-to-collimator distance for three different collimators provided by one commercial manufacturer. Typically, collimator resolution deteriorates by a factor of 2 at a distance of 4-5 cm from the collimator. On the other hand, according to Equation 16-7, collimator efficiency for a source in air is independent of source-to-collimator distance b. This rather surprising result is obtained provided the counting rate for the
9 338 Physics in Nuclear Medicine. entire detector area is measured. The reason for this is illustrated by Figure As the source is moved farther away from the collimator, the efficiency with which radiation is transmitted through anyone collimator hole decreases in proportion to l/b2 (inverse-square law), but the number of holes through which radiation can pass to reach the detector increases in proportion to b2. The two effects cancel each other, with the result that total counting rate--and thus collimator efficiencydoes not change with source-to-collimator distance. Another illus,tration of this effect is shown in Figure As source-to-collimator distance increases, the maximum height of the PSF or LSF decreases, but the width increases (and resolution becomes poorer), so that the total area under the curve (total detector counting rate) does not change. lnvariance of collimator efficiency with source-to-collimator,distance applies to point sources, line sources, and uniform sheet sources in air with parallel-hole collimators; however, it applies only to uniform sheet sources with converging, diverging, or pinhole collimators (Section C). When the source is embedded at different depths in the patient, attenuation effects also must be considered. Septal penetration and scatter of photons from the walls of the collimator holes also are not considered in the above analysis. 2. ~ I. ~,. ~ LJ... I. z ~ I. t- ;:)6 I. cn ljj Q: O. Q: t- <X ~ O. -J -J 0 O. u SOURCE-TO-COLLIMATOR DISTANCE(cm) Fig Collimator resolution versus source-to-collimator distance for three different collimators. [Adapted from ref. 4: Hine GJ, Paras D, Warr CP: Recent advances in gamma-camera imaging. Proc SPIE 152:123, 1978, permis-. ~....
10 ~ The Anger Camera: Performance Characteristics 339 SINGLE-HOLE EFFICIENCY«l/b2 i DETECTOR ~(1 RADIATION EXAp~ESAED ~~-~~~--~---~~-~---/ '\ SOURCE «b 2 ~---~- ---~..- b ~ \ 'COll I MA TOR DETECTOR Fig Explanation for constant counting rate (collimator efficiency) versus source-to-collimator distance for a point source in air and a parallel-hole collimator. Efficiency for a single hole decreases as l/b2, but number of holes passing radiation (area of detector exposed) increases as b2. Table 16-2 summarizes the performance characteristics of a number of collimators provided by one commercial manufacturer. Collimator resolution is the FWHM for a source at 10 cm from the face of the collimator. Collimator efficiency g refers to the relative number of'y rays transmitted by the collimator per 'Y ray emitted by the source. Note that the approximate relationship between collimator efficiency and resolution COUNTING RATE CE-TO-COLlIMATOR ANCE DISTANCE Fig Point spread functions versus distance for a parallel-hole collimator. Area under curve is proportional to collimator efficiency and does not change with distance.
11 340 Physics in Nuclear Medicine given by Equation 16-8 is verified by these data. Note also the relatively small values of collimator efficiency. 4. System Resolution The sharpness of images recorded with an Anger camera is limited by several factors, including intrinsic resolution, collimator resolution, scattered radiation, and septal penetration. In terms of the FWHM of a point or line spread function, the most important factors are the intrinsic resolution Rj of the detector and electronics and the collimator resolution Rc. The combined effect of these two factors is to produce a system resolution Rs that is somewhat worse than either one alone. System resolution Rs (FWHM) is given by, Rs = YR1 + R~ (16-9) Because collimator resolution depends on source-to-collimator distance, system resolution also depends on this parameter. Figure shows system resolution versus source-to-collimator distance for a typical parallel-hole collimator and different values of intrinsic resolution. At a distance of 5-10 cm (typical depth of organs inside the body), system resolution is much poorer than intrinsic resolution and is determined primarily by collimator resolution. There are significant differences between system resolutions for cameras having substantially different intrinsic resolutions (e.g., 4 mm versus 8 mm), but the difference in system resolutions for cameras having small differences in intrinsic resolutions (e.g., 4 mm versus 5 mm) is minor and not clinically Table 16-2 Performance Characteristics of Some Typical Commercially Manufactured Parallel-Hole Collimators Suggested Resolution R Maximum (FWHM at 10 Collimator Type Energy (ke V) Efficiency g cm) Low energy, x mm high resolution Low energy, x mm general purpose Low energy, x 10~ mm high sensitivity Medium energy, x 10-4 c 13.4 mm high sensitivity Adapted from ref. 3: Hine GJ, Erickson JJ: Advances in scintigraphic instruments, in Hine GJ, Sorenson JA (eds): Instrumentation in Nuclear Medicine (vol 2). New York, '~"".,,~...,'., "
12 ~ The Anger Camera: Performance Characteristics E 1.8 u ~ 1.6 :I: 3: I z I (J) w 0.8 II: ~ 0.6 TYPICAL I- ORGAN ~ 0.4 DEPTHS (J) O. 2 ::::::::i::::::::::;::::::;::lli::!:::~:~:::! 0 Ii SOURCE-TO-COLLIMATOR DISTANCE (cm) Fig System resolution versus source-to-collimator distance for a typical parallel-hole collimator and for different values of intrinsic resolution. At most typical organ depths, system resolution is determined primarily by collimator resolution. significant. Small differences in intrinsic resolution may be apparent on bar pattern images or on images of very superficial structures in the patient, but they usually are not apparent on images of deeper-lying structures. System resolution also is degraded by scattered radiation. This is discussed in Chapter 18, Section B. The method for combining component resolutions to determine system resolution also is discussed in Appendix G. C. PERFORMANCE CHARACTERISTICS OF CONVERGING, DIVERGING, AND PINHOLE COLLIMATORS Equations for collimator resolution, Rc, and efficiency, g, for converging, diverging and pinhole collimators are as follows:
13 342 Physics in Nuclear Medicine CONVERGING COLLIMATOR DIVERGING COLLIMATOR Rc = [d(l~ + b)/1~][1/cos9][1 - (1~/2)/(f + l~)] (16-10). g = K2(d/l~)2[d2/(d + t)2][f2/(f - b)2] (16-11) l~ = (I - 2~-1)/cos9 (16-12) Rc = [d(l~ + b)/1~][i/cos9][ 1 +(1~f2j)] (16-13) g = K2(d/l'e)2[d2/(d+ t)2][f+ l)/(f+ I + b)] (16-14) PINHOLE COLLIMATOR Rc = de(l + b)/l (16-15) g = de cos39/16b2 (16-16) de = \/d[d + 2~-ltan(a/2)] (16-17) The parameters in these equations are as shown in Figure For the pinhole collimator, de is the" effective" pinhole diameter, accounting for penetration of the edges of the pinhole aperture. The equations for collimator resolution Rc refer to the equivalent FWHM of the point- or. ~ f 1\ f ' '~I-d! T I f -it 1! I TaT..( I ~..( WI i I~.{ + \! I : ii 1 t 0 IT!, W l iw l 1 V'8 ~, Ie\, CONVERGING DIVERGING PINHOLE Fig Parameters for collimator resolution, R (FWHM), and efficiency, g (transmitted/emitted photons), for Equations (16-17). Adapted with permission from Medical Physics Data Book, Padika1 TN, Fivozinsky SP (eds). Washington DC, National Bureau of Standards, 1982, pp
14 ..- The Anger Camera: Performance Characteristics >- 250 B u z... ~ 16 ~ z = /2 ~ NG ~ 0... ~ 100 ~ '" f- > ~ 4 ;:: 50 PINH C/) ~ I... ~ / C/) ~ SOURCE-TO-COLLIMATOR DISTANCE(cm) Fig Penonnance characteristics (left, system resolution; right, pointsource geometric efficiency in air) versus source-to-collimator distance for four different types of Anger camera collimators. [From Rollo FD, Harris CC: Factors affecting image fonnation, in Rollo FD (ed): Nuclear Medicine Physics, Instrumentation, and Agents. St.Louis, C. V. Mosby Co., Modified from Moyer RA: J Nucl. Med. 15:59, 1974]. line-source spread function, corrected for magnification or minification of the image by the collimator (Equations 15-3, 15-5, and 15-6). Thus, if the collimator projects a profile with a 2 cm FWHM measured on the detector and the image magnification factor is x2, the equivalent FWHM in the imaged plane is 1 cm. The equations above may be compared with Equations 16-6 and 16-7 for parallel-hole collimator. They are similar except for the presence of additional terms involving collimator focal lengths f and, for off-axis sources, the angle 8 between the source, the focal point (or pinhole), and the central axis of the collimator (Figure 16-7). The equations illustrate that for converging and diverging collimators, resolution is best at the center (8 = 0, cos 8 = 1). The performance characteristics of different types of collimators are compared in Figure which shows system resolution and efficiency vs. distance, including effects of camera intrinsic resolution as well as collimator magnification. Figure illustrates that resolution always is best with'the source as close as possible to the collimator. Changes in collimator efficiency with distance depend on whether the radiation source is a point source or a uniform sheet source. For a point source (Figure 16-18, right), collimator efficiency increases with increasing source-to-collimator distance with the converging collimator. Maximum efficiency is obtained at the collimator convergence point (~35 cm), where 'Y rays are transmitted through all of the collimator holes, and then
15 . 344 Physics in Nuclear Medicine decreases beyond that point. Point-source collimator efficiency decreases with distance for the diverging and pinhole collimators, more severely for the latter. For an extended, large-area sheet source, sufficiently large to. cover the entire field of view of the collimator, efficiency does not change with source-to-collimator distance for all of these collimators. Again, for sources embedded within a patient, attenuation effects also must be accounted for. " Figure illustrates that the converging collimator offers the best combination of resolution and efficiency at typical imaging distances (5-10 cm); however, the field of view is also somewhat limited at 'these distances (Equation 15-6, Example 15-2), and for this reason converging collimators are most useful with cameras having relatively large area detectors. Diverging collimators offer a larger imaging area (Example 15-1) but at the cost of both resolution and efficiency. Pinhole collimators offer very good resolution and reasonable efficiency at close distances but lose efficiency very rapidly with distance; they also have a quite limited field of view because of magnification effects at typical imaging distances (Equation 15-3). Generally they are used for imaging smaller organs that can be positioned close to the collimator, e.g., thyroid and heart. They also are useful with high-energy "{-ray emitters because they do not suffer from septal penetration problems. Differences between the resolution and field of view obtained at different source-to-collimator distances with parallel-hole, converging, diverging, and pinhole collimators are further illustrated by Figure AT COLLIMATOR FACE jj ~ } DI ~XNCE... 20cm DISTANCE...,,; c P!VERG!NcG PARALLEL HOLE CONVEcRGING.'::r Fig Bar pattern images demonstrating changing field size and resolution obtained vs. distance for three collimator types.
16 The Anger Camera: Performance Characteristics 345 It should also be noted that the distortions caused by changing magnification with depth for different structures inside the body sometimes make images obtained with the converging, diverging, and pinhole collimators difficult to interpret (see Figure 15-8). REFERENCES 1. Anger HO: Radioisotope cameras, in Hine GJ (ed): Instrumentation in Nuclear Medicine (vol 1). New York, Academic Press, 1967, chap Strand S-E, Larsson I: Image artifacts at high photon ftuence rates in single-crystal NaI(Tl) scintillation cameras. J Nucl Med 19: , Hine GJ, Erickson JJ: Advances in scintigraphic instruments, in Hine GJ, Sorenson JA (eds): Instrumentation in Nuclear Medicine (vol 2). New York, Academic Press, 1974, chap 1 4. Hine GJ, Paras D, Warr CP: Recent advances in gamma-camera imaging. Proc SPIE 152: , 1978? Additional discussions of Anger camera performance characteristics and collimator design are discussed in the following: Hine GJ, Paras P, Warr CP: Measurement of the Performance Parameters of Gamma Cameras, Part I. Rockville, MD., U.S. Dept. H.E.W., Pub!.. No. (FDA) , 1977 Richardson RL: Anger scintillation camera, in Rollo FD (ed): Nuclear Medicine Physics, Instrumentation, and Agents. St. Louis, C. V. Mosby Co., 1977, chap 6 ~,~-,'-- :..
Laue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE
ALGORITHM TO DETERMINE THE CALIBRATION PARAMETERS FOR A NDA METHOD OF INVENTORY VERIFICATION IN A DIFFUSION ENRICHMENT CASCADE Gevaldo Lisboa de Almeida, Carlos Feu Alvim Agência Brasileiro-Argentina de
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
Geometric Optics Converging Lenses and Mirrors Physics Lab IV
Objective Geometric Optics Converging Lenses and Mirrors Physics Lab IV In this set of lab exercises, the basic properties geometric optics concerning converging lenses and mirrors will be explored. The
X-ray thin-film measurement techniques
Technical articles X-ray thin-film measurement techniques II. Out-of-plane diffraction measurements Toru Mitsunaga* 1. Introduction A thin-film sample is two-dimensionally formed on the surface of a substrate,
View of ΣIGMA TM (Ref. 1)
Overview of the FESEM system 1. Electron optical column 2. Specimen chamber 3. EDS detector [Electron Dispersive Spectroscopy] 4. Monitors 5. BSD (Back scatter detector) 6. Personal Computer 7. ON/STANDBY/OFF
Reflection and Refraction
Equipment Reflection and Refraction Acrylic block set, plane-concave-convex universal mirror, cork board, cork board stand, pins, flashlight, protractor, ruler, mirror worksheet, rectangular block worksheet,
Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light
AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light Name: Period: Date: MULTIPLE CHOICE. Choose the one alternative that best completes the statement or answers the question. 1) Reflection,
Experiment 3 Lenses and Images
Experiment 3 Lenses and Images Who shall teach thee, unless it be thine own eyes? Euripides (480?-406? BC) OBJECTIVES To examine the nature and location of images formed by es. THEORY Lenses are frequently
Optical Communications
Optical Communications Telecommunication Engineering School of Engineering University of Rome La Sapienza Rome, Italy 2005-2006 Lecture #2, May 2 2006 The Optical Communication System BLOCK DIAGRAM OF
Anamorphic imaging with three mirrors: a survey
Anamorphic imaging with three mirrors: a survey Joseph M. Howard Optics Branch (Code 551), NASA Goddard Space Flight Center, Greenbelt, MD 20771 Ph: 301-286-0690 Fax: 301-286-0204 [email protected]
2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.
Revision problem Chapter 18 problem 37 page 612 Suppose you point a pinhole camera at a 15m tall tree that is 75m away. 1 Optical Instruments Thin lens equation Refractive power Cameras The human eye Combining
potential in the centre of the sphere with respect to infinity.
Umeå Universitet, Fysik 1 Vitaly Bychkov Prov i fysik, Electricity and Waves, 2006-09-27, kl 16.00-22.00 Hjälpmedel: Students can use any book. Define the notations you are using properly. Present your
EXPERIMENT 6 OPTICS: FOCAL LENGTH OF A LENS
EXPERIMENT 6 OPTICS: FOCAL LENGTH OF A LENS The following website should be accessed before coming to class. Text reference: pp189-196 Optics Bench a) For convenience of discussion we assume that the light
Endoscope Optics. Chapter 8. 8.1 Introduction
Chapter 8 Endoscope Optics Endoscopes are used to observe otherwise inaccessible areas within the human body either noninvasively or minimally invasively. Endoscopes have unparalleled ability to visualize
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
Experimental and modeling studies of imaging with curvilinear electronic eye cameras
Experimental and modeling studies of imaging with curvilinear electronic eye cameras Viktor Malyarchuk, 1 Inhwa Jung, 1 John A. Rogers, 1,* Gunchul Shin, 2 and Jeong Sook Ha 2 1 Department of Materials
Tutorial 4.6 Gamma Spectrum Analysis
Tutorial 4.6 Gamma Spectrum Analysis Slide 1. Gamma Spectrum Analysis In this module, we will apply the concepts that were discussed in Tutorial 4.1, Interactions of Radiation with Matter. Slide 2. Learning
Introduction to the Monte Carlo method
Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw
ATOMS: ATOMIC STRUCTURE QUESTIONS AND ANSWERS
ATOMS: ATOMIC STRUCTURE QUESTIONS AND ANSWERS QUESTION ONE: MODELS OF THE ATOM (2011;1) At different times scientists have proposed various descriptions or models of the atom to match experimental evidence
Convex Mirrors. Ray Diagram for Convex Mirror
Convex Mirrors Center of curvature and focal point both located behind mirror The image for a convex mirror is always virtual and upright compared to the object A convex mirror will reflect a set of parallel
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
1. You stand two feet away from a plane mirror. How far is it from you to your image? a. 2.0 ft c. 4.0 ft b. 3.0 ft d. 5.0 ft
Lenses and Mirrors 1. You stand two feet away from a plane mirror. How far is it from you to your image? a. 2.0 ft c. 4.0 ft b. 3.0 ft d. 5.0 ft 2. Which of the following best describes the image from
Data. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
GAMMA-RAY SPECTRA REFERENCES
GAMMA-RAY SPECTRA REFERENCES 1. K. Siegbahn, Alpha, Beta and Gamma-Ray Spectroscopy, Vol. I, particularly Chapts. 5, 8A. 2. Nucleonics Data Sheets, Nos. 1-45 (available from the Resource Centre) 3. H.E.
Measuring the Point Spread Function of a Fluorescence Microscope
Frederick National Laboratory Measuring the Point Spread Function of a Fluorescence Microscope Stephen J Lockett, PhD Principal Scientist, Optical Microscopy and Analysis Laboratory Frederick National
Polarization of Light
Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
Conductive and Radiative Heat Transfer in Insulators
Conductive and Radiative Heat Transfer in Insulators Akhan Tleoubaev, Ph.D. LaserComp, Inc., December 1998 Heat transfer for most thermal insulation materials occurs via both conduction and radiation.
Structural Integrity Analysis
Structural Integrity Analysis 1. STRESS CONCENTRATION Igor Kokcharov 1.1 STRESSES AND CONCENTRATORS 1.1.1 Stress An applied external force F causes inner forces in the carrying structure. Inner forces
Information about the T9 beam line and experimental facilities
Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions
Light and Sound. Pupil Booklet
Duncanrig Secondary School East Kilbride S2 Physics Elective Light and Sound Name: Pupil Booklet Class: SCN 3-11a - By exploring the refraction of light when passed through different materials, lenses
Computer Animation of Extensive Air Showers Interacting with the Milagro Water Cherenkov Detector
Computer Animation of Extensive Air Showers Interacting with the Milagro Water Cherenkov Detector Miguel F. Morales Department of Physics, University of California, Santa Cruz, CA 95064, USA We employ
Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors
Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus
2) A convex lens is known as a diverging lens and a concave lens is known as a converging lens. Answer: FALSE Diff: 1 Var: 1 Page Ref: Sec.
Physics for Scientists and Engineers, 4e (Giancoli) Chapter 33 Lenses and Optical Instruments 33.1 Conceptual Questions 1) State how to draw the three rays for finding the image position due to a thin
Overview. What is EMR? Electromagnetic Radiation (EMR) LA502 Special Studies Remote Sensing
LA502 Special Studies Remote Sensing Electromagnetic Radiation (EMR) Dr. Ragab Khalil Department of Landscape Architecture Faculty of Environmental Design King AbdulAziz University Room 103 Overview What
ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS
Copyright JCPDS - International Centre for Diffraction Data 2004, Advances in X-ray Analysis, Volume 47. 130 ABSTRACT ON-STREAM XRF ANALYSIS OF HEAVY METALS AT PPM CONCENTRATIONS G Roach and J Tickner
4. CAMERA ADJUSTMENTS
4. CAMERA ADJUSTMENTS Only by the possibility of displacing lens and rear standard all advantages of a view camera are fully utilized. These displacements serve for control of perspective, positioning
Production of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the
9/16 Optics 1 /11 GEOMETRIC OPTICS
9/6 Optics / GEOMETRIC OPTICS PURPOSE: To review the basics of geometric optics and to observe the function of some simple and compound optical devices. APPARATUS: Optical bench, lenses, mirror, target
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Optical Design Tools for Backlight Displays
Optical Design Tools for Backlight Displays Introduction Backlights are used for compact, portable, electronic devices with flat panel Liquid Crystal Displays (LCDs) that require illumination from behind.
AS COMPETITION PAPER 2008
AS COMPETITION PAPER 28 Name School Town & County Total Mark/5 Time Allowed: One hour Attempt as many questions as you can. Write your answers on this question paper. Marks allocated for each question
Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
DISCLAIMER. This document was prepared as an account of work sponsored by an agency of the United States
DISCLAIMER This document was prepared as an account of work sponsored by an agency of the United States Government. Neither the United States Government nor the University of California nor any of their
Cross section, Flux, Luminosity, Scattering Rates
Cross section, Flux, Luminosity, Scattering Rates Table of Contents Paul Avery (Andrey Korytov) Sep. 9, 013 1 Introduction... 1 Cross section, flux and scattering... 1 3 Scattering length λ and λ ρ...
Backwater Rise and Drag Characteristics of Bridge Piers under Subcritical
European Water 36: 7-35, 11. 11 E.W. Publications Backwater Rise and Drag Characteristics of Bridge Piers under Subcritical Flow Conditions C.R. Suribabu *, R.M. Sabarish, R. Narasimhan and A.R. Chandhru
Thin Lenses Drawing Ray Diagrams
Drawing Ray Diagrams Fig. 1a Fig. 1b In this activity we explore how light refracts as it passes through a thin lens. Eyeglasses have been in use since the 13 th century. In 1610 Galileo used two lenses
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM
8.2 Elastic Strain Energy
Section 8. 8. Elastic Strain Energy The strain energy stored in an elastic material upon deformation is calculated below for a number of different geometries and loading conditions. These expressions for
1051-232 Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002
05-232 Imaging Systems Laboratory II Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002 Abstract: For designing the optics of an imaging system, one of the main types of tools used today is optical
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
Physics 441/2: Transmission Electron Microscope
Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This
Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems
Characteristics of an Integrated Germanium Detector Based Gamma-Ray Spectrometer for Monitoring Systems Ronald M. Keyser, Timothy R. Twomey, Sam Hitch ORTEC 801 South Illinois Avenue Oak Ridge, TN, 37831
World-first Proton Pencil Beam Scanning System with FDA Clearance
Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi
Removing chips is a method for producing plastic threads of small diameters and high batches, which cause frequent failures of thread punches.
Plastic Threads Technical University of Gabrovo Yordanka Atanasova Threads in plastic products can be produced in three ways: a) by direct moulding with thread punch or die; b) by placing a threaded metal
Basic Nuclear Concepts
Section 7: In this section, we present a basic description of atomic nuclei, the stored energy contained within them, their occurrence and stability Basic Nuclear Concepts EARLY DISCOVERIES [see also Section
Lenses and Apertures of A TEM
Instructor: Dr. C.Wang EMA 6518 Course Presentation Lenses and Apertures of A TEM Group Member: Anup Kr. Keshri Srikanth Korla Sushma Amruthaluri Venkata Pasumarthi Xudong Chen Outline Electron Optics
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS
A METHOD OF CALIBRATING HELMHOLTZ COILS FOR THE MEASUREMENT OF PERMANENT MAGNETS Joseph J. Stupak Jr, Oersted Technology Tualatin, Oregon (reprinted from IMCSD 24th Annual Proceedings 1995) ABSTRACT The
Chapter 17: Light and Image Formation
Chapter 17: Light and Image Formation 1. When light enters a medium with a higher index of refraction it is A. absorbed. B. bent away from the normal. C. bent towards from the normal. D. continues in the
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
Radiation Detection and Measurement
Radiation Detection and Measurement June 2008 Tom Lewellen [email protected] Types of radiation relevant to Nuclear Medicine Particle Symbol Mass (MeV/c 2 ) Charge Electron e-,! - 0.511-1 Positron
Specific Intensity. I ν =
Specific Intensity Initial question: A number of active galactic nuclei display jets, that is, long, nearly linear, structures that can extend for hundreds of kiloparsecs. Many have two oppositely-directed
Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography
Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,
Questions. Strategies August/September Number Theory. What is meant by a number being evenly divisible by another number?
Content Skills Essential August/September Number Theory Identify factors List multiples of whole numbers Classify prime and composite numbers Analyze the rules of divisibility What is meant by a number
............... [2] At the time of purchase of a Strontium-90 source, the activity is 3.7 10 6 Bq.
1 Strontium-90 decays with the emission of a β-particle to form Yttrium-90. The reaction is represented by the equation 90 38 The decay constant is 0.025 year 1. 90 39 0 1 Sr Y + e + 0.55 MeV. (a) Suggest,
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier
Tunnel Effect: - particle with kinetic energy E strikes a barrier with height U 0 > E and width L - classically the particle cannot overcome the barrier - quantum mechanically the particle can penetrated
Studies on Pore Systems in Catalysts
JOURNAL OF CATALYSIS 4, 319323 (1965) Studies on Pore Systems in Catalysts V. The t Method B. C. LIPPENS* AND J. H. DE BOER From the Department of Chemical Technology, Technological University of Delft,
How do you compare numbers? On a number line, larger numbers are to the right and smaller numbers are to the left.
The verbal answers to all of the following questions should be memorized before completion of pre-algebra. Answers that are not memorized will hinder your ability to succeed in algebra 1. Number Basics
A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials
Illicit Trafficking, Sub/cross-national threats, Poster presentation A Comparison of an HPGe-based and NaI-based Radionuclide Identifier (RID) for Radioactive Materials Ronald M. Keyser, Timothy R. Twomey,
Spherical VLS grating RIXS spectrometers: From basics to the hv 2 concept
Spherical VLS grating RIXS spectrometers: From basics to the hv 2 concept Vladimir N. Strocov (ADRESS beamline, Swiss Light Source) Outline: 1. Optical design of spherical VLS grating spectrometers: -
PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR PRODUCTION ON THE PIERCED FLAT SHEET METAL USING LASER FORMING PROCESS
JOURNAL OF CURRENT RESEARCH IN SCIENCE (ISSN 2322-5009) CODEN (USA): JCRSDJ 2014, Vol. 2, No. 2, pp:277-284 Available at www.jcrs010.com ORIGINAL ARTICLE EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR
The Bending Strength of Pasta
The Bending Strength of Pasta 1.105 Lab #1 Louis L. Bucciarelli 9 September, 2003 Lab Partners: [Name1] [Name2] Data File: Tgroup3.txt On the cover page, include your name, the names of your lab partners,
CHAPTER 7 SCINTILLATION COUNTING
CHAPTER 7 SCINTILLATION COUNTING Radiation of various types is widely utilized for non-destructive inspection and testing such as in medical diagnosis, industrial inspection, material analysis and other
ALPERTON COMMUNITY SCHOOL MATHS FACULTY ACHIEVING GRADE A/A* EXAM PRACTICE BY TOPIC
ALPERTON COMMUNITY SCHOOL MATHS FACULTY ACHIEVING GRADE A/A* EXAM PRACTICE BY TOPIC WEEK Calculator paper Each set of questions is followed by solutions so you can check & mark your own work CONTENTS TOPIC
Human Exposure to Outdoor PLC System
1602 PIERS Proceedings, Marrakesh, MOROCCO, March 20 23, 2011 Human Exposure to Outdoor PLC System Vicko Doric 1, Dragan Poljak 1, and Khalil El Khamlichi Drissi 2 1 University of Split, Croatia 2 Blaise
BOUNDARY INTEGRAL EQUATIONS FOR MODELING ARBITRARY FLAW
BOUNDARY INTEGRAL EQUATIONS FOR MODELING ARBITRARY FLAW GEOMETRIES IN ELECTRIC CURRENT INJECTION NDE A. P. Ewingl,2, C. Hall Barbosa3.4, T. A. Cruse 2, A. C. Brun03 and 1. P. Wikswo, Jr.l ldepartment of
Appendix A. An Overview of Monte Carlo N-Particle Software
Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation
Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
ENERGY LOSS OF ALPHA PARTICLES IN GASES
Vilnius University Faculty of Physics Department of Solid State Electronics Laboratory of Applied Nuclear Physics Experiment No. ENERGY LOSS OF ALPHA PARTICLES IN GASES by Andrius Poškus (e-mail: [email protected])
Assessment of Camera Phone Distortion and Implications for Watermarking
Assessment of Camera Phone Distortion and Implications for Watermarking Aparna Gurijala, Alastair Reed and Eric Evans Digimarc Corporation, 9405 SW Gemini Drive, Beaverton, OR 97008, USA 1. INTRODUCTION
BIOMEDICAL ULTRASOUND
BIOMEDICAL ULTRASOUND Goals: To become familiar with: Ultrasound wave Wave propagation and Scattering Mechanisms of Tissue Damage Biomedical Ultrasound Transducers Biomedical Ultrasound Imaging Ultrasonic
Solar Energy. Outline. Solar radiation. What is light?-- Electromagnetic Radiation. Light - Electromagnetic wave spectrum. Electromagnetic Radiation
Outline MAE 493R/593V- Renewable Energy Devices Solar Energy Electromagnetic wave Solar spectrum Solar global radiation Solar thermal energy Solar thermal collectors Solar thermal power plants Photovoltaics
Application Note: Absorbance
Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed
1 of 9 2/9/2010 3:38 PM
1 of 9 2/9/2010 3:38 PM Chapter 23 Homework Due: 8:00am on Monday, February 8, 2010 Note: To understand how points are awarded, read your instructor's Grading Policy. [Return to Standard Assignment View]
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
