Clinical Photon Beams

Size: px
Start display at page:

Download "Clinical Photon Beams"

Transcription

1 Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida

2 Disclosures Research development grants from Philips Medical Systems, Elekta Oncology Systems, and Sun Nuclear Associates. NIH research award. Bankhead d Coley research award.

3 Learning Objectives Understanding dosimetric properties of clinical photon beams. Understanding gphysical parameters that affect dosimetric properties of clinical photon beams. Understand the need for accurate characterization of clinical photon beams in a treatment planning system.

4 Photon Beam Delivery Systems S band Linear Accelerators X band Linear Accelerators Medical Linear Accelerators: Accelerate electrons in pulses to kinetic energies from 4 to 25 MeV. Use non-conservative microwave RF fields in the frequency range from 10 3 MHz (L band) to 10 4 MHz (X band), with the vast majority running at 2856 MHz (S band). Some provide beams only in the low megavoltage range (4-6 MV), while others provide both photons and electrons at various megavoltage energies. A typical modern high-energy linac can provide 2-3 photon energies.

5 Sources of radiation that determine dosimetric characteristics of clinical photon beams Indirect (headscatter) Flattening filter Monitor Chamber Collimator jaws Electron Contamination MLC Charged particle contamination dose Secondary electrons Source Direct Output radiation or Incident radiation Primary dose Scatter dose Direct Radiation (Focal Radiation) Photon radiation generated at the target that reaches patient without any intermediate interactions. Indirect Radiation (Extra- focal Radiation): Photon radiation with a history of interaction/scattering in the head of the treatment unit with the flattening filter, collimators, or other structures in the treatment head. Contaminant electrons/positrons secondary electrons and positrons released from interactions with either the treatment head or the air column. AAPM TG74 Report

6 Sources of Direct and Indirect Radiation Direct Indirect A Monte Carlo study (Chaney et al., Med. Phys. 21,1994) Siemens MD2, 6MV

7 Characterizing Dosimetric Properties of Clinical Photon Beams Beam penetration Normalized depth dose (NDD) or tissue phantom ratio (TPR). Beam Output Total output ratio: S c,p, in-air output ratio: S c, phantom scatter factor: S p. Cross-beam profile Isodose distribution. Attenuation factors for beam modifiers hard wedges, compensators, trays, etc. With the ultimate goal of ensuring that computerized treatment With the ultimate goal of ensuring that computerized treatment plans accurately reflect the dose received by patients

8 Beam Penetration NDD d D d d, s, f, Q D dref where d is the depth of measurement on the central axis of the phantom, s is the field size at the surface of the phantom, f is the sourcesurface-distance, Q is the quality of the clinical photon beam, and D d and D dref are dose at depth d and dref respectively. Water s f d TPR data can be determined from measured NDD as follows: TPR d, s, Q NDD d, s, f, Q d 2 f d S s f dref p S p dref s d

9 Normalized Depth Dose Data Energy Dependence Buildup region 15 MV TCPE region Surface region 6MV FS = 10 x 10 cm 2

10 Normalized Depth Dose Data Field Size Dependence This depth corresponds to range of the highest energy contaminant charged particles 15 MV Photon Beam 16x16 4x4

11 Normalized Depth Dose Data Wedge/Open Comparison FS=10x10cm2 15 MV (W/O) 6 MV (W/O)

12 Normalized Depth Dose Data Wedge/Open Comparison Minima

13 Normalized Depth Dose Data Field sizes: 6x6, 10x10 and 20x20 cm 2 6 MV - Siemens -- Varian.Elekta 18 MV These data from Radiological Physics Center show that all NDD for both 6 and 18 MV photon beams at depths of 5 cm and 15 cm for different field sizes have a maximum %σ of 0.5% and this increases to 0.7% at a depth of 20 cm.

14 Monte Carlo Calculated Photon Beam Spectra The spectral shapes are somewhat similar The differences at the high-energy end are caused by the differences in the mean incident electron energies and their spread Sheikh-Bagheri & Rogers, Med. Phys., 29, 2002

15 Monte Carlo Calculated Average Energies The average energies for the same nominal accelerating potential are somewhat similar The average energies decrease at off-axis distances for all clinical beams more pronounced difference at higher energies Sheikh-Bagheri & Rogers, Med. Phys., 29, 2002

16 Beam Penetration for Irregularly- Shaped Fields Concept of Equivalent Square: The equivalent field is defined as that standard (square or circular) field which has the same centralaxis depth dose characteristics as the given non-standard field. s f d Day s Rule : S r r r S 1 e r e Water S(r) = the central axis scatter in a field of radius r, S = the central axis scatter in field of infinite it radius, λ is a scaling parameter, and μ is a dimensionless i shape parameter. They computed equivalent square fields for a complete set of rectangular fields using a value of λ=0.26 cm-1 and μ=0.5.

17 L Equivalent square d W s Sterling Formula: (Sterling et.al., Brit. J. Radiol. 37, 544 (1964)) LW S 2 4 A L W / P Assuming, λ = 0.26 cm-1., and μ = 0.5 L/2 W/2 SLW (, ) 4 Dxydxdy (, ) 0 0 L / W SLW (, ) / S (10,10)

18 KLEIN- NISHINA CROSS SECTION FOR THE COMPTON INTERACTION d e d Ψ 2 2 r0 h ' h h ' sin 2 h h ' h 2 PHOTONS SCATTERED AT AN ANGLE, Ψ PHOTONS SCATTERED INTO A UNIT SOLID ANGLE, Ω SOLID ANGLE AVAILABLE PER UNIT ANGLE d d 2 sin Based on the kinematics of Compton interaction, the average energy of scattered photons is less than 1Mev and is independent of the incident energy.

19 Measurement of Normalized Depth Dose data Follow AAPM TG Report # 106 recommendations: Use 4-5 mm diameter ion chamber for depth beyond 1cm. Use parallel plate or extrapolation chamber to measure data near the surface. Diodes and diamond d detectors t are appropriate as long as data measured with these detectors is scoss cross-referenced e e ced to data measured ed with an ion chamber. Prone to radiation damage and non-linear response.

20 Is depth ionization data depth dose? YES!!! With the caveat, TCPE exists at the point of measurement. the energy spectrum of incident photons does not change with the depth. fluence across the detector remains the same. These conditions are met at depths beyond the range of contaminant charged particles However at shallow depth, The contaminants and secondary electrons have energy spectra that change rapidly with depth. Results in a variation of ~10% in restricted mass stopping power ratio data for water and air. Translates into a spatial uncertainty of less than 1.5 mm in dose in the build up region

21 Beam Output f f S c S c,p 10 cm c c S p s S c, p S c s c (Derived) Water

22 In-air output Ratio Elekta: 4-18 MV clinical photon beams.

23 Monte Carlo Calculations of In- Air Output t Ratio (BEAMnrc code) In-Air Output Ratio O o Simulation Geometry (Varian 2100EX) MV measured 6 MV calculated lated 18 MV measured 18 MV calculated Side of square field /cm /tex/rof/clxyro

24 Energy spectrum of head scattered photons Mean Energy:0.5 MeV (Varian 2100C.)

25 Energy spectrum of head scattered photons (Varian 2100C.) Mean Energy:0.5 MeV

26 In-air output Ratio e: Elekta, s: Siemens, and v: Varian (for clinical photon beams ranging from 6-25 MV.

27 Monitor Back Scatter Machine MBS Publication Flattening Filter Beam Modifier (internal wedge) Monitor Chamber Varian Clinac % Kubo, Med. Phys. 16, 295 (1987) Lower Collimator Upper Collimator Therac % Hounsell, P.M.B. 43, 445 (1998) Beam Modifier (external wedge) Tertiary Collimator (Cerrobend Block or Varian MLC) Elekta SL15 <1% Yu et.al. P.M.B. (with 3 mm AL) 41, 1107(1996) 5% (without Al) Varian 600c/2100C 2-5% Lam et. al. Med. Varian 2100C Phys. 25, 334 (1998) The differences in In-Air Output Ratio for the same field size on different machines is primarily attributed to the difference in monitor back scatter

28 Measurement of In-Air Output t Ratios Mini phantom Water-equivalent materials. 4g/cm 2 diameter and 10g/cm 2 depth to maintain lateral CPE and eliminate contaminant electron. For small segment fields (c<4cm), high Z material (Brass etc.) should be used. Corrections for energy absorption coefficients and energy spectra change are needed. r 1 h TG 74 recommendations 1 2

29 Cross Beam Characteristics Affected by the radially symmetric conical high Z- material flattening filter, which Flattens the beam by differentially absorbing more photons in the center and less in the periphery unwanted consequence of flattening the beam is the differential change in beam quality at off-axis points. hardens the beam Cross beam flatness is defined as: D F 100 D max max D One flattening filter for each clinical photon beam results in a compromise of beam flatness characteristics of small and large fields. Flattening filters are designed d to give a gradually increasing i radial intensity. it This is referred to as horns on a cross-beam profile Cross beam profiles may not be radially symmetric due to non circular focal spot. Therefore, cross-beam data is characterized by a set of two orthogonal dose profiles measured perpendicular to the beam s central axis at a given depth in a phantom D min min

30 Cross Beam Profile 6 MV Photon Beam, Depth of 5.0 cm, Field size of 4x4, 10.4x10.4, and 21x21 cm 2. The flatness of photon beams is extremely sensitive to change in energy of the incident beam. A small change in the penetrative quality of a photon beam results in very large change in beam flatness.

31 Cross Beam Profile 6 MV Photon Beam, Field Size of 10.4x10.4 cm2, Depths of 1.5, 5.0, 10.0, 15.0, and 25.0 cm. The field flatness changes with depth. This is attributed to an increase in scatter to primary dose ratio with increasing depth and decreasing incident photon energy off axis

32 Effect of Electron Steering on Beam Flatness Symmetric Tilted Displaced

33 Effect of a Dipole Magnet on Exit Beam Energy Spread Radial Displacement Radial Divergence

34 Cross Beam Symmetry S 100 Area left Area right Arealeft Area right Dosimetry and beam steering system

35 Isodose Distribution 30 cm X 30 cm 18 MV X-ray beam

36 Isodose Distributions (20 X 20 Cm 2 ) 6 MV 18 MV Note contaminant electrons contribute to dose outside the field at shallow depths. The magnitude and extent t of dose outside the geometric edge of a field at shallow depths increases with beam energy.

37 Isodose Distributions (20 X 20 Cm 2, 18 MV) Note Contaminant electrons contribute to dose outside the field at shallow depths. The magnitude and extent of dose outside the geometric edge of a field at shallow depths increases even more in the presence of beam modifiers.

38 Cross Beam Measurements What is the affect of detector size? Incorrect measurement of penumbra region Diameter Penumbra 20%~80% Diode CC04 CC13 0.8x0.8 mm 2 4 mm 6 mm 4.0 mm 6.1 mm 7.2 mm

39 Detector Size Effect on TPS Commissioning Treatment Planning System Commissioning Impact of detector size effect on dose distribution??? ib ti Yan G et. al., Med. Phys (35)., 2008

40 Extraction of True Profile

41 IMRT QA results: DTA 2%/2 mm CC13 CC04 Deconvolved

42 Measurement of Attenuation Factors for Beam Modifiers The attenuation factor for a beam modifier is defined as the ratio of the dose rate at the point of calculation for a given field with and without the modifier in place. Attenuation factors for devices such as block trays, accessories etc. are often assumed to be independent of field size, depth and SSD. These factors should be measured at a depth well beyond the maximum range of electron contamination The attenuation devices that are in contact with the patient skin (immobilization apparatus, table top, etc.) require additional considerations. These devices not only attenuate the incident beam but they introduce scatter radiation that increase the scatter to primary ratio within the patient. It is best to include such attenuation devices as a part of the patient in 3DRTPS

43 Measurement of Wedge Factors The WF is defined as the ratio of the dose rate at the reference depth for a wedged d field to that t for the same field without a wedge modifier. The field size dependency of the WF originates from a wedge-induced increase in head scatter. the field size dependence of the WF is correctly accounted for by in-air output ratios (S c ) wedge specifically measured for wedged fields These data should be measured with the chamber axis perpendicular to the gradient direction of the wedge Two sets of measurements should be made with the wedge in opposite orientations to ensure the correct placement of the chamber

44 Characterizing Clinical Photon Beams in 3DRTPS Ahnesjo et al., PMB 1999

45 Approaches to Dose Computation Algorithms Data measured in water and in air Parameterize water data Reconstitute water data Calculate inhomogeneity corrections to water data Correction Correction based methods Calculate dose directly based on beam and phantom configurations Model Model based methods Figure 8.9,The Modern Technology of Radiation Oncology; J. Van Dyk

46 Correction vs. Model Based Methods Correction Based Measured data used as basis for Dose Computation. Require measurements with buildup cap in air or in a mini-phantom. Require lots of data. Generating functions used to reduce size of data set for convenient clinical use (i.e. less storage space). Patient dose distribution obtained by first computing Dose in water from generating function, then correcting for tissue heterogeneity, patient contour, and beam modifiers. Model Based Measured data used to setup description of treatment beam. Require a parameter to estimate size of photon source at target. Require more time for tuning of model parameters. Patient dose distribution obtained by computing beam and beam transport (i.e. beam interactions in treatment head and in patient) directly.

47 Accuracy Goal in Dose Calculations Required accuracy (overall( treatment < 5%): Ahnesjo et al., PMB 1999

48 Characterizing Clinical Photon Beams in 3DRTPS MUST model the following features realistically: Finite size of source (& penumbra) Extra-focal radiation (primary collimator, flattening filter) Beam spectrum (& change in spectrum with position) Beam intensity variation across field (e.g., beam horns) Transmission through secondary collimators Scatter outside field (related to extra-focal radiation) MLC, blocks, block tray Dynamic wedge, fixed wedge, compensators (beam hardening)

49 Characterizing Clinical Photon Beams in 3DRTPS Caveats: Almost all photon dose computation with convolution models assumes kernel invariance, which requires the photon dose kernel to be constant with spatial locations in the calculation phantom. However, in clinical treatments, patient inhomogeneities, as well as beam divergence and polychromaticity, cause kernel variation in various ways. Modeling of charged particle contaminants is at best an approximation of real clinical situation Modeling of indirect radiation as a single or multiple analytical source functions, modeling of off-axis softening with a simple parametric fit, source size, etc. are best effort estimates of physical processes

50 Characterizing Clinical Photon Caveats (continued): Beams in 3DRTPS One can always use a set of beam modeling parameters to get the best agreement between the computed and measured beam data in a phantom.. However, that would not be a sufficient condition for robust and accurate beam modeling. The value or function used to describe a parameter should have some physical meaning. each parameter used in the dose calculation algorithm should model the physical reality it represents even if there is less than perfect agreement between measure and computed data. The observed differences often reflect limitations of the dose computation algorithm

51 Benchmark Dataset (Developed under NIH initiative) iti A collaborative effort involving Sun Nuclear Associates; the contractor, t and consultants t from: the University it of Florida; the RPC at M.D. Anderson Cancer Center; the University of Iowa; and the Vassar Brothers Hospital. Already measured a complete set of data on the new generation of Elekta (Synergy), Siemens (Oncor) and Varian (Trilogy) linear accelerators Measured data are comprehensive in beam geometries to validate dose computation for any clinical situation. data are sufficient in spatial resolution and were validated by independent measurements This benchmark datasets will be sufficient for the TPS This benchmark datasets will be sufficient for the TPS companies to compare the accuracy of their dose modeling for treatment delivery

52 Summary The dosimetric properties of a clinical photon beam are characterized by: Its ability to penetrate a tissue-like medium (water) its change in dose output with field size Its cross beam behavior Its attenuation through modifying devices (e.g., wedge, compensator etc.) The dosimetric properties of clinical photon beams from linacs depend on the photon energy fluence distribution emanating from the treatment head, on the geometry of the linac, and on the radiological properties of the medium with which it interacts.

53 Summary It is quite evident that all modern clinical linear accelerators (linacs) of a particular commercial make produce beams of very similar characteristics High quality benchmark data have already been acquired by comprehensively characterizing single linacs of each make. These benchmark data thoroughly describe the characteristics of photon beams so that treatment-planning companies and clinics throughout the United States can use it to examine the accuracy of dose-calculation algorithms.

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually

More information

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs

Advanced variance reduction techniques applied to Monte Carlo simulation of linacs MAESTRO Advanced variance reduction techniques applied to Monte Carlo simulation of linacs Llorenç Brualla, Francesc Salvat, Eric Franchisseur, Salvador García-Pareja, Antonio Lallena Institut Gustave

More information

Performance evaluation and quality assurance of Varian enhanced dynamic wedges

Performance evaluation and quality assurance of Varian enhanced dynamic wedges JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 7, NUMBER 1, WINTER 2006 Performance evaluation and quality assurance of Varian enhanced dynamic wedges Parham Alaei and Patrick D. Higgins Department

More information

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline

MLC Characteristics. Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues. Presentation Outline Treatment Delivery Systems 2 Field Shaping; Design Characteristics and Dosimetry Issues Timothy D. Solberg David Geffen School of Medicine at UCLA TU-A-517A-1 Presentation Outline MLC Characteristics TG-50

More information

Variance reduction techniques used in BEAMnrc

Variance reduction techniques used in BEAMnrc Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,

More information

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015 INDUSTRIAL DESIGN Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG VARIAN ONCOLOGY SYSTEMS 1 VARIAN ONCOLOGY SYSTEMS CERN Accelerator, May 2015 Industrial Design

More information

Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS

Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS Chapter 8 ELECTRON BEAMS: PHYSICAL AND CLINICAL ASPECTS W. STRYDOM Department of Medical Physics, Medical University of Southern Africa, Pretoria, South Africa W. PARKER, M. OLIVARES Department of Medical

More information

Challenges in small field MV photon dosimetry

Challenges in small field MV photon dosimetry Challenges in small field MV photon dosimetry Maria Mania Aspradakis Cantonal Hospital of Lucerne, Lucerne, Switzerland Cantonal Hospital of Lucerne, Lucerne, Switzerland maria.aspradakis@ksl.ch Why are

More information

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools

3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools 3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground

More information

Quality Assurance of Radiotherapy Equipment

Quality Assurance of Radiotherapy Equipment Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria

More information

Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D.

Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan M.S, Craig Stevens MD, Ph.D., Ken Forster Ph.D. Efficacy of IMRT with dot decimal compensators for radiotherapy of malignant pleural mesothelioma post extrapleural pneumonectomy Department of Radiation Oncology H. Lee Moffitt Cancer Center Khosrow Javedan

More information

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system

Independent corroboration of monitor unit calculations performed by a 3D computerized planning system JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 1, NUMBER 4, FALL 2000 Independent corroboration of monitor unit calculations performed by a 3D computerized planning system Konrad W. Leszczynski* and

More information

THE DOSIMETRIC EFFECTS OF

THE DOSIMETRIC EFFECTS OF THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,

More information

Manual for simulation of EB processing. Software ModeRTL

Manual for simulation of EB processing. Software ModeRTL 1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended

More information

Key words: treatment planning, quality assurance, 3D treatment planning

Key words: treatment planning, quality assurance, 3D treatment planning American Association of Physicists in Medicine Radiation Therapy Committee Task Group 53: Quality assurance for clinical radiotherapy treatment planning Benedick Fraass a) University of Michigan Medical

More information

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y

Post Treatment Log File Based QA Varian. Krishni Wijesooriya, PhD University of Virginia. D e p a r t m e n t of R a d i a t i o n O n c o l o g y Post Treatment Log File Based QA Varian Krishni Wijesooriya, PhD University of Virginia Learning Objectives What information could be accessed via log files Scenarios where Log files could be used. How

More information

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study

Feasibility Study of Neutron Dose for Real Time Image Guided. Proton Therapy: A Monte Carlo Study Feasibility Study of Neutron Dose for Real Time Image Guided Proton Therapy: A Monte Carlo Study Jin Sung Kim, Jung Suk Shin, Daehyun Kim, EunHyuk Shin, Kwangzoo Chung, Sungkoo Cho, Sung Hwan Ahn, Sanggyu

More information

QA of intensity-modulated beams using dynamic MLC log files

QA of intensity-modulated beams using dynamic MLC log files 36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,

More information

Our Department: structure and organization

Our Department: structure and organization EORTC meeting for Radiation Therapy Technologists: RTT s role in the modernization of radiotherapy 10th October 2014, Villejuif (Grand Paris), France Elekta Stereotactic Body Frame: transmission modelled

More information

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular

More information

ACCELERATORS AND MEDICAL PHYSICS 2

ACCELERATORS AND MEDICAL PHYSICS 2 ACCELERATORS AND MEDICAL PHYSICS 2 Ugo Amaldi University of Milano Bicocca and TERA Foundation EPFL 2-28.10.10 - U. Amaldi 1 The icone of radiation therapy Radiation beam in matter EPFL 2-28.10.10 - U.

More information

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer

Evaluation of complexity and deliverability of IMRT- treatment plans. Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer Evaluation of complexity and deliverability of IMRT- treatment plans Author: Elin Svensson Supervisors: Anna Bäck and Anna Karlsson Hauer June 11 Abstract Background: Intensity modulated beams are used

More information

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada

More information

Secondary Neutrons in Proton and Ion Therapy

Secondary Neutrons in Proton and Ion Therapy Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles

More information

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department

More information

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons

90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons SLAC-PUB-7722 January 9 degrees Bremsstrahlung Source Term Produced in Thick Targets by 5 MeV to GeV Electrons X. S. Mao et al. Presented at the Ninth International Conference on Radiation Shielding, Tsukuba,

More information

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers

In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers In room Magnetic Resonance Imaging guided Radiotherapy (MRIgRT( MRIgRT) Jan Lagendijk and Bas Raaymakers : Chris Bakker Boxue Liu Alexander Raaijmakers Rajko Topolnjak Richard van de Put Uulke van der

More information

World-first Proton Pencil Beam Scanning System with FDA Clearance

World-first Proton Pencil Beam Scanning System with FDA Clearance Hitachi Review Vol. 58 (29), No.5 225 World-first Proton Pencil Beam Scanning System with FDA Clearance Completion of Proton Therapy System for MDACC Koji Matsuda Hiroyuki Itami Daishun Chiba Kazuyoshi

More information

Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations

Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations Monte Carlo Simulations: Efficiency Improvement Techniques and Statistical Considerations Daryoush Sheikh-Bagheri, Ph.D. 1, Iwan Kawrakow, Ph.D. 2, Blake Walters, M.Sc. 2, and David W.O. Rogers, Ph.D.

More information

Digital vs. Analogue Control Systems

Digital vs. Analogue Control Systems Digital vs. Analogue Control Systems Presented at the 2011 Annual Meeting of the American College of Medical Physics, Chattanooga, TN, May 1, 2011 Ivan A. Brezovich, PhD, Dept. of Rad Onc, Univ of Alabama

More information

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry

Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry Monte Carlo Simulation for Solid Angle Calculations in Alpha Particle Spectrometry John Keightley NSUF 18 May 2010 Acknowledgement: Several discussions with Stefaan Pommé, IRMM, EC-JRC, Geel Some diagrams

More information

IAEA-TECDOC-1540. Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems

IAEA-TECDOC-1540. Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems April 2007 IAEA-TECDOC-1540 Specification and Acceptance Testing of Radiotherapy Treatment Planning Systems

More information

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam

Reduction of Electron Contamination Using a Filter for 6MV Photon Beam J. Korean Soc Ther Radiol Oncol : Vol. 15, No. 2, June, 1997 Reduction of Electron Contamination Using a Filter for 6MV Photon Beam Choul Soo Lee, MPH., Myung Jin Yoo, Ph.D. and Ha Yong Yum, M.D. Department

More information

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong

RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac

More information

Chapter XIII. PDD TAR TMR Dose Free Space

Chapter XIII. PDD TAR TMR Dose Free Space Ver 2.11 Chapter XIII PDD TAR TMR Dose Free Space Joseph F. Buono RTT Allied Health Science Nassau Community College 1 Education Drive Garden City, NY 11530-6793 phone: 516-572 - 7536 oice - 9460 Secretary

More information

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,

More information

1. Provide clinical training in radiation oncology physics within a structured clinical environment.

1. Provide clinical training in radiation oncology physics within a structured clinical environment. Medical Physics Residency Program Overview Our Physics Residency Training is a 2 year program typically beginning July 1 each year. The first year resident will work closely with medical physicists responsible

More information

Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA

Analysis of Trajectory Log Files of TrueBeam Medical Electron Linear Accelerator for Patient Specific IMRT QA Cancer Research and Oncology: Open Access Received: Nov 16, 2015, Accepted: Jan 12, 2016, Published: Jan 15, 2016 CancerResOncol, Volume 1, Issue 1 http://crescopublications.org/pdf/crooa/crooa-1-001.pdf

More information

1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range

1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range 1. Orthovoltage vs. megavoltage x-rays. (AL) External beam radiation sources: Orthovoltage radiotherapy: 200-500 kv range The radiation from orthovoltage units is referred to as x-rays, generated by bombarding

More information

Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY

Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY Chapter 7 CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY W. PARKER, H. PATROCINIO Department of Medical Physics, McGill University Health Centre, Montreal, Quebec, Canada 7.1. INTRODUCTION

More information

MONTE CARLO MODELING OF A VARIAN 2100C 18 MV MEGAVOLTAGE PHOTON BEAM AND SUBSEQUENT DOSE DELIVERY USING MCNP5

MONTE CARLO MODELING OF A VARIAN 2100C 18 MV MEGAVOLTAGE PHOTON BEAM AND SUBSEQUENT DOSE DELIVERY USING MCNP5 MONTE CARLO MODELING OF A VARIAN 2100C 18 MV MEGAVOLTAGE PHOTON BEAM AND SUBSEQUENT DOSE DELIVERY USING MCNP5 A Thesis Presented to The Academic Faculty by Jared Hoover In Partial Fulfillment of the Requirements

More information

Status and perspective of emission imaging techniques for ion beam therapy in Lyon

Status and perspective of emission imaging techniques for ion beam therapy in Lyon 14th International Conference on Nuclear Reaction Mechanisms, Varenna 18th June 2015 Status and perspective of emission imaging techniques for ion beam therapy in Lyon on behalf of the CAS-PHABIO group

More information

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS

CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 10. ACCEPTANCE TESTS AND COMMISSIONING MEASUREMENTS JOHN L. HORTON Department of Radiation Physics University of Texas

More information

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage

3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage 3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter

More information

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Development of on line monitor detectors used for clinical routine in proton and ion therapy Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary

More information

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS

AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,

More information

Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY

Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY Chapter 12 QUALITY ASSURANCE OF EXTERNAL BEAM RADIOTHERAPY D.I. THWAITES Department of Oncology Physics, Edinburgh Cancer Centre, University of Edinburgh, Western General Hospital, Edinburgh, United Kingdom

More information

Accreditation Is Coming

Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt

More information

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed

More information

Quality Assurance in Stereotactic. Radiotherapy

Quality Assurance in Stereotactic. Radiotherapy Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy D. Solberg, Ph.D. University of Texas Southwestern

More information

TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS

TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS AAPM REPORT NO. 85 TISSUE INHOMOGENEITY CORRECTIONS FOR MEGAVOLTAGE PHOTON BEAMS Report of Task Group No. 65 of the Radiation Therapy Committee of the American Association of Physicists in Medicine Members

More information

Proton tracking for medical imaging and dosimetry

Proton tracking for medical imaging and dosimetry Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4

More information

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning

NIA RADIATION ONCOLOGY CODING STANDARD. Dosimetry Planning NIA RADIATION ONCOLOGY CODING STANDARD Dosimetry Planning CPT Codes: 77295, 77300, 77301, 77306, 77307, 77321, 77316, 77317, 77318, 77331, 77399 Original Date: April, 2011 Last Reviewed Date: November,

More information

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy

Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino

More information

Atomic and Nuclear Physics Laboratory (Physics 4780)

Atomic and Nuclear Physics Laboratory (Physics 4780) Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *

More information

The accurate calibration of all detectors is crucial for the subsequent data

The accurate calibration of all detectors is crucial for the subsequent data Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved

More information

Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth

Radiation Oncology Centers Participating in MassHealth. Daniel Tsai, Assistant Secretary and Director of MassHealth Executive Office of Health and Human s Office of Medicaid www.mass.gov/masshealth May 2015 TO: FROM: RE: Radiation Oncology Centers Participating in Daniel Tsai, Assistant Secretary and Director of (2015

More information

Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers

Acknowledgement. Prerequisites: Basic Concepts of IMRT. Overview. Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers Therapy Continuing Education Course Clinical Implementation of IMRT for Lung Cancers H. Helen Liu, PhD Department of Radiation Physics, U.T. MD Anderson Cancer Center, Houston, TX AAPM, Seattle, 2005 Acknowledgement

More information

Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy

Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy Progress in NUCLEAR SCIENCE and TECHNOLOGY, Vol. 2, pp.176-180 (2011) ARTICLE Joint Application of Perl Scripts and MCNPX in Solving the Dynamic-Geometry Related Problems in Proton Beam Radiotherapy Fada

More information

Calorimetry in particle physics experiments

Calorimetry in particle physics experiments Calorimetry in particle physics experiments Unit n. 8 Calibration techniques Roberta Arcidiacono Lecture overview Introduction Hardware Calibration Test Beam Calibration In-situ Calibration (EM calorimeters)

More information

Laue lens for Nuclear Medicine

Laue lens for Nuclear Medicine Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of

More information

CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY

CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY Review of Radiation Oncology Physics: A Handbook for Teachers and Students CHAPTER 7. CLINICAL TREATMENT PLANNING IN EXTERNAL PHOTON BEAM RADIOTHERAPY WILLIAM PARKER Department of Medical Physics McGill

More information

MOC Exam Preparation Therapy

MOC Exam Preparation Therapy MOC Exam Preparation Therapy R. Alfredo C. Siochi, PhD., DABR Overview MOC Exam Requirement Available Guidance What I did What I wished I did 1 MOC Exam Requirement Rolling Window of 10 years Must pass

More information

Introduction to the Monte Carlo method

Introduction to the Monte Carlo method Some history Simple applications Radiation transport modelling Flux and Dose calculations Variance reduction Easy Monte Carlo Pioneers of the Monte Carlo Simulation Method: Stanisław Ulam (1909 1984) Stanislaw

More information

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film. INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

More information

Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy

Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy Chapter 11 Computerized Treatment Planning Systems for External Photon Beam Radiotherapy This set of 117 slides is based on Chapter 11 authored by M.D.C. Evans of the IAEA publication (ISBN 92-0-107304-6):

More information

Algorithmes : comment calculer et rapporter la dose

Algorithmes : comment calculer et rapporter la dose Algorithmes : comment calculer et rapporter la dose N. Reynaert Medical Physics Department, Centre Oscar Lambret Lille Introduction Until 1970: No CT data available: Patient consists of water 2D planning

More information

2 Absorbing Solar Energy

2 Absorbing Solar Energy 2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could

More information

Safety Initiatives in RT The Role of Process Managment

Safety Initiatives in RT The Role of Process Managment Safety Initiatives in RT The Role of Process Managment Sasa Mutic Department of Radiation Oncology Mallinckrodt Institute of Radiology Washington University St. Louis, MO Conflict of Interest Director

More information

Use of 3D Printers in Proton Therapy

Use of 3D Printers in Proton Therapy Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology Remmes.Nicholas@mayo.edu 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing

More information

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM

More information

The feasibility of a QA program for ISIORT Trials

The feasibility of a QA program for ISIORT Trials Baveno, Italy June 22-24, 2012 The feasibility of a QA program for ISIORT Trials Frank W. Hensley 1, Don A. Goer 2, Sebastian Adamczyk 3, Falk Roeder 1, Felix Sedlmayer 4, Peter Kopp 4 1 University Clinics

More information

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Radiography: 2D and 3D Metrics of Performance Towards Quality Index AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..

More information

Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi

Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi Dose enhancement near metal electrodes in diamond X- ray detectors Acknowledgements Surrey University: P.J. Sellin M. Abd-El Rahman P. Veeramani H. Al-Barakaty F. Schirru Mechanical Workshop A. Lohstroh*,

More information

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy

Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Chapter 7: Clinical Treatment Planning in External Photon Beam Radiotherapy Set of 232 slides based on the chapter authored by W. Parker, H. Patrocinio of the IAEA publication (ISBN 92-0-107304-6): Review

More information

Monday 11 June 2012 Afternoon

Monday 11 June 2012 Afternoon Monday 11 June 2012 Afternoon A2 GCE PHYSICS B (ADVANCING PHYSICS) G495 Field and Particle Pictures *G412090612* Candidates answer on the Question Paper. OCR supplied materials: Data, Formulae and Relationships

More information

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS 1. Photons 2. Photoelectric Effect 3. Experimental Set-up to study Photoelectric Effect 4. Effect of Intensity, Frequency, Potential on P.E.

More information

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of

More information

COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS

COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS COMPARISON OF FOUR DATA ANALYSIS SOFTWARE FOR COMBINED X-RAY REFLECTIVITY AND GRAZING INCIDENCE X-RAY FLUORESCENCE MEASUREMENTS Bérenger Caby (1), Fabio Brigidi (2), Dieter Ingerle (3), Blanka Detlefs

More information

Worldwide Quality Assurance networks for radiotherapy dosimetry

Worldwide Quality Assurance networks for radiotherapy dosimetry Worldwide Quality Assurance networks for radiotherapy dosimetry Joanna Izewska Dosimetry and Medical Radiation Physics Section dosimetry@iaea.org Audit in radiotherapy dosimetry Objectives to enhance confidence

More information

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.

More information

External dosimetry Dosimetry in new radiotherapeutic techniques

External dosimetry Dosimetry in new radiotherapeutic techniques External dosimetry Dosimetry in new radiotherapeutic techniques Albert Lisbona Medical Physics Department CLCC René Gauducheau 44805 Saint-Herblain France a-lisbona@nantes.fnclcc.fr Objective : To describe

More information

Chapter 22: Electric Flux and Gauss s Law

Chapter 22: Electric Flux and Gauss s Law 22.1 ntroduction We have seen in chapter 21 that determining the electric field of a continuous charge distribution can become very complicated for some charge distributions. t would be desirable if we

More information

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector

Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector

More information

Quality Assurance of accelerators; the technologists responsibility

Quality Assurance of accelerators; the technologists responsibility Quality Assurance of accelerators; the technologists responsibility Christa Timmermans Radiation Technologist, Erasmus MC- Daniel den Hoed, Rotterdam, The Netherlands EORTC-ROG RT technologists Section

More information

Quality Assurance for Treatment Planning Systems

Quality Assurance for Treatment Planning Systems Quality Assurance for Treatment Planning Systems PTCOG Educational Session May 19 th 2010 Oliver Jäkel Heidelberg Ion Beam Therapy Center and German Cancer Research Center, Heidelberg Introduction Outline

More information

Anamorphic imaging with three mirrors: a survey

Anamorphic imaging with three mirrors: a survey Anamorphic imaging with three mirrors: a survey Joseph M. Howard Optics Branch (Code 551), NASA Goddard Space Flight Center, Greenbelt, MD 20771 Ph: 301-286-0690 Fax: 301-286-0204 Joseph.M.Howard@nasa.gov

More information

EXPOSURE INDICES IN RF EPIDEMIOLOGY

EXPOSURE INDICES IN RF EPIDEMIOLOGY EXPOSURE INDICES IN RF EPIDEMIOLOGY Paolo Vecchia National Institute of Health, Rome, Italy Advances in Biological effects and Dosimetry of Low Energy Electromagnetic Fields Erice, 28 March - 8 April 1981

More information

Fig.1. The DAWN spacecraft

Fig.1. The DAWN spacecraft Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics

More information

Powder diffraction and synchrotron radiation

Powder diffraction and synchrotron radiation Powder diffraction and synchrotron radiation Gilberto Artioli Dip. Geoscienze UNIPD CIRCe Center for Cement Materials single xl diffraction powder diffraction Ideal powder Powder averaging Textured sample

More information

Polarization of Light

Polarization of Light Polarization of Light References Halliday/Resnick/Walker Fundamentals of Physics, Chapter 33, 7 th ed. Wiley 005 PASCO EX997A and EX999 guide sheets (written by Ann Hanks) weight Exercises and weights

More information

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography

More information

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.

Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms

More information

Monte Carlo Simulations in Proton Dosimetry with Geant4

Monte Carlo Simulations in Proton Dosimetry with Geant4 Monte Carlo Simulations in Proton Dosimetry with Geant4 Zdenek Moravek, Ludwig Bogner Klinik und Poliklinik für Strahlentherapie Universität Regensburg Objectives of the Study what particles and how much

More information

Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation

Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation Calculation of Contra-lateral Lung Doses in Thoracic IMRT: An Experimental Evaluation Deborah Schofield, Laurence Court, Aaron Allen, Fred Hacker, Maria Czerminska Department of Radiation Oncology Dana

More information

Therapy with protons and ion beams

Therapy with protons and ion beams Therapy with protons and ion beams Biomedical Physics Lecture WiSe 2012/13 many slides from J. Willkens (TUM) Key quesions: How does ion cancer therapy work? What is the physics behind it? How can physics

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

RICHARD T ELL ASSOCIATES,INC.

RICHARD T ELL ASSOCIATES,INC. RICHARD T ELL ASSOCIATES,INC. An Evaluation of Test Measurement Data Obtained on the KW-Gard RF Protective Suit March 30, 1998 Prepared for Euclid Garment Manufacturing Company 333 Martinel Drive Kent,

More information

Optical Communications

Optical Communications Optical Communications Telecommunication Engineering School of Engineering University of Rome La Sapienza Rome, Italy 2005-2006 Lecture #2, May 2 2006 The Optical Communication System BLOCK DIAGRAM OF

More information

Clinical Rotation 4: PHYS 707 Spring 2016 (Feb. 25, 2016 to Aug. 25, 2016) COURSE INFORMATION

Clinical Rotation 4: PHYS 707 Spring 2016 (Feb. 25, 2016 to Aug. 25, 2016) COURSE INFORMATION Clinical Rotation 4: PHYS 707 Spring 2016 (Feb. 25, 2016 to Aug. 25, 2016) COURSE INFORMATION Class Days: Monday-Friday Class Times: Forty hours per week Class Location: One of the participating cancer

More information