MEDICAL IMAGING 2nd Part Computed Tomography



Similar documents
Cone Beam Reconstruction Jiang Hsieh, Ph.D.

CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver /9/07) by M. Jensen and J. E.

CT Image Reconstruction. Terry Peters Robarts Research Institute London Canada

Introduction to Medical Imaging. Lecture 11: Cone-Beam CT Theory. Introduction. Available cone-beam reconstruction methods: Our discussion:

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction

Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays

EECS 556 Image Processing W 09. Interpolation. Interpolation techniques B splines

Purchasing a cardiac CT scanner: What the radiologist needs to know

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

5 Factors Affecting the Signal-to-Noise Ratio

State of Stress at Point

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

BIOMEDICAL ULTRASOUND

Polarization of Light

Fundamentals of Cone-Beam CT Imaging

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Synthetic Sensing: Proximity / Distance Sensors

Computed Tomography Instrumentation. Ian A. Cunningham. Data-Acquisition Geometries. Ian A. Cunningham

PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors

DICOM Correction Item

Image Analysis for Volumetric Industrial Inspection and Interaction

Common Core Unit Summary Grades 6 to 8

Manual for simulation of EB processing. Software ModeRTL

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. III - Medical and Industrial Tomography - W.B.Gilboy

Computerized Tomography, X-rays, and the Radon Transform

Introduction to acoustic imaging

MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

CT Traceability - Prof. Wim Dewulf, Group T - KU Leuven

Data. microcat +SPECT

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

A System for Capturing High Resolution Images

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity

Page: 1 of 6 Page: 1 of 6

Graphical displays are generally of two types: vector displays and raster displays. Vector displays

Instruction Sheet

Solving Simultaneous Equations and Matrices

Section 1.1. Introduction to R n

How To Improve Your Ct Image Quality

Waves - Transverse and Longitudinal Waves

COMP175: Computer Graphics. Lecture 1 Introduction and Display Technologies

C1 Medical Imaging Modalities & Characteristics Linwei Wang

The accurate calibration of all detectors is crucial for the subsequent data

Experimental study of beam hardening artefacts in photon counting breast computed tomography

APPLICATION OF X-RAY COMPUTED TOMOGRAPHY IN SILICON SOLAR CELLS

Powder diffraction and synchrotron radiation

Practice Final Math 122 Spring 12 Instructor: Jeff Lang

Lecture 14. Point Spread Function (PSF)

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

IMAGE PROCESSING AND DATA ANALYSIS IN COMPUTED TOMOGRAPHY

Encoders for Linear Motors in the Electronics Industry

X-ray thin-film measurement techniques

Ultrasonic Wave Propagation Review

Robot Perception Continued

X-ray Imaging Systems

Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July Reprinted from Photonics Spectra, Laurin Publishing

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

A PHOTOGRAMMETRIC APPRAOCH FOR AUTOMATIC TRAFFIC ASSESSMENT USING CONVENTIONAL CCTV CAMERA

Dispersion diagrams of a water-loaded cylindrical shell obtained from the structural and acoustic responses of the sensor array along the shell

Aliasing, Image Sampling and Reconstruction

Two years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Module 1 : Conduction. Lecture 5 : 1D conduction example problems. 2D conduction

M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy

FURTHER VECTORS (MEI)

Development of High Speed Oblique X-ray CT System for Printed Circuit Board

Exam 1 Sample Question SOLUTIONS. y = 2x

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

Resolution Enhancement of Photogrammetric Digital Images

INTRODUCTION. A. Purpose

Robotics. Lecture 3: Sensors. See course website for up to date information.

Cathode Ray Tube. Introduction. Functional principle

Specific Intensity. I ν =

ENGINEERING METROLOGY

Laue lens for Nuclear Medicine

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

Introduction to CCDs and CCD Data Calibration

Experiment 5. Lasers and laser mode structure

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Measuring of optical output and attenuation

BME Introduction to BME. Bioelectrical Engineering Part: Medical Imaging

Development and quantitative assessment of a beam hardening correction model for preclinical micro-ct

PHYS 39a Lab 3: Microscope Optics

Sound Power Measurement

IAEA HUMAN HEALTH SERIES

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

Digital radiography conquers the veterinary world

Standard Test Method for Classification of Film Systems for Industrial Radiography 1

Transcription:

MEDICAL IMAGING 2nd Part Computed Tomography

Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional projection radiograpgy the image represents the 2D projection of a 3D structure. To obtain information on the real structure the number of projections have to be increased A tomogram is an image of a slice or a plane within the body The basics of CT is to take a series of conventional cross-sectional x-rays while the patient is rotated slightly around an axis between each exposure. A series of projection data is obtained, thes data are used to reconstruct cross-sectional images.

Introduction 3 1917: The austrian mathematician Radon develop a way to reconstruct the density distribution of an object if the line integrals (for every direction) are available 1957 63: The physician Cormack develops a lot of theoretical work on x- rays creating the basis of CT scanning (without knowing Radon work) 1971: The first CT system has been created by Godfrey N. Hounsfield, he received the Nobel prize in 1979 (together with Cromack) 1972: The first clinical use of a CT system in London, the acquisition time of a couple of tomogram was 5 minutes

Introduction 4 1973-74: The first II Generation CT system total-body scanner has been realised in the U.S., the acquisition time for 1 tomogram was 1 minute 1976-77: III and IV Generations CT with acquisition time for 1 tomogram lower than 10 secs 1983: First electron beam CT, very expensive (scarce diffusion) 1989: First helical CT, very low acquisition time (less than 1 sec) and able to explore a large body volume 1992: Dual-slice CT scanner, able to acquire a large number of layers in a very low time 2000: Multislice CT scanner, multiple arrays of detectors; continuous development.

Computed Tomography Generations First-generation scanner Second-generation scanner 5 Single source collimated in a pencil beam and single detector Linear scan Arbitrary number of rays and angular projections can be measured, scattered radiation is mostly undetected Partial fan-beam and detectors array. Linear scan Able to complete a full scan in less time than I Generation scanners

Computed Tomography Generations Third-generation scanner Fourth-generation scanner 6 Fan-beam and detectos array. Source and detectors rotate in synchrony Dramatic decrease of scan time Fan-beam and large ring of stationary detectos array. Single rotating source Collimators cannot be used in detectors (they must receive energy from a mobile source) high detection efficiency and scattering can cause problems

Computed Tomography Generations 7 Main characteristics of the first four CT scanners generations I II III IV Scan Time (sec) 135-300 15-60 1-2 1-2 Number of detectors (for each layer) 1 10-30 250-1000 500-2000 Movement Rotation + Transl. Rotation + Transl. Rotation Rotation X-ray tube rotation 180 180 360 360 Data amount (each tomogram) 57.6 kb 57.6 kb 1 MB 2-42 MB Matrix 80x80 80x80 256x256 512x512

Computed Tomography Generations Fifth-generation scanner: electron beam computed tomography (EBCT); no moving parts, a electronically steered electron beam is used, this beam hit one of the four tugsten anode strip encircling the patient. This produces x-rays that are collimated in a fan-beam and revealed by a stationary ring of detectors. A full set of data are acquired in 50 milliseconds. Very expensive Sixth-generation scanner: helical CT addresses the need for rapid volumetric data acquisition. Conventional arrangement of source and detctors (III and IV generations) which can continuously rotate. During tube rotation patient table is slightly moved through the sourcedetectors plane. The image acquisition is not sequential (as it is for previous CT generations) but volumetric 8 A) sequential scanning B) helical scanning

Computed Tomography Generations Seventh-generation scanner: multislice CT scanners, a thick fan-beam is used and multiple detectors array are used to collect the x-rays. This technology is in continuous development, recently 64-128 (and 256) multiple detectors array has been developed. 9 With the 256-slice system the entire heart can be covered in a single rotation Helical and multislice CT has made the requirements for new development in data processing techniques even more critical Future development: multislice scanners able to interact with a cone-beam Due to the large volume that can be explored it will be possible to stop using mobile patient table, this will decrease the complexity in image reconstruction

CT Instrumentation 10 Most of the commercial medical CT scanners use just one x-ray source. X-ray tubes usually wear out in less than one year and need frequent status control and calibration. X-rays generated by the tube require collimation and filtration. Typically fan-beam geometry (having 30-60 degress of fan angle) is obtained through two pieces of lead that form a slit between them. Slit width is adjustable by the operator in order to obtain different fanbeam thickness, typically between 0.5 and 10 mm for single slice systems and 20 to 30 mm for multislice CT systems. Fan-angle 30-60 Thickness 0.5-30 mm X-rays fan-beam

CT Instrumentation 11 Different types of CT detectors: Solid state detectors contain a scintillation crystal able to interact with x-rays and produce photoelectrons that are captured by a photo-diode and converted to electric current Xenon gas detectors long thin tubes containing compressed xenon gas, when ionized the gas generates a current between anode and cathode. Less efficient than solid state detectors but highly directional Slip ring solves the mechanical problem of continuous electrical contact with rotating gantry; the gantry holds the x-ray tube (that needs high voltage to work properly) and the detectors that generate hundreds of signals to be transmitted out

Data Acquisition If a thin beam of monoenergetic x-rays, having an intensity of, propagates for a distance d through an omogenic material, the intensity of the attenuated beam is: I 0 12 I = Ie µd 0 I0 µ I Being µ the linear attenuation coefficient µ is a function of material atomic number and electronic density d If the beam propagates through different omogenic materials the intensity of the attenuated beam is: d µ i ( µ 1+ µ 2+... + µ n) d i= 1 0 0 I = I e = I e n I0 x µ 1 µ 2 µ n d I

Data Acquisition 13 If the material has a continuous variable attenuation coefficient the intensity of the attenuated beam is: I = I e 0 0 D dx Given a measurement of and knowledge of, the projection measurement is defined as: p I 0 I0 I = ln( ) = I 0 D µ dx Therefore the basic measurement of a CT scanner is a line integral of the linear attenuation coefficient x I 0 x D I

Data Acquisition A CT scanner has to be calibrated in order to obtain the reference intensity I 0 Multiple projections (along parallel lines) create a so-called attenuation profile 14 CT detectors are able to measure the line integrals of bidimensional distribution of linear attenuation coefficients Through a suitable number of attenuation profiles it is possible to identify the bidimensional distribution of attenuation coefficients and it is possible to reconstruct the bidimensional image. Attenuation profile

CT Numbers 15 Different CT have different x-rays tubes, so the same object will produce differenc values of µ on different scanners (the value of µ is a function of: x-ray energy, tissue density and atomic number) In order to compare data from different scanners is not possible to calculate absolute values of linear attenuation coefficents but at each pixel is associated a numerical value called CT number A CT number is calculated in Hounsfield units (UH) and is defined as: µ w µ t µ w h = 1000 µ Being the water attenuation coefficient and the tissue attenuation coefficient A CT number is 10 times the percentual difference between µ and t µ w divided by µ w w µ t

CT Numbers 16 As an example, a tissue having a attenuation coefficient 4% bigger than water has a CT number: h t (1.04 µ w µ w) = 1000 = 40 µ w Clearly h=0 for water, that is the reference value. Since µ = 0 for air, it is possible to find that h=-1000 for air. The largest CT number found naturally in the body are for bones, 1000 HU for average bone. CT numbers can surpass 3000 HU for metal and contrast agents A typical HU range is: -1024 / 3071 (4096 levels, 12 bits each pixel)

CT Numbers Examples of typical HU values 17

CT Numbers Windowing 18

CT Numbers Conventional CT scan without contrast showing possible tumor in the liver Conventional CT scan of the same patient using contrast

Image Formation 20 CT images are reconstructed as a numerical bidimensional matrix. Each element of the matrix is called pixel The number of pixels is variable between 256x256 and 1024x1024 For each couple of x and y, the pixel represent a micro bidimensional area and f(x,y) is the functional value of this area f(x,y) is a grey level value, positive integer power of 2 NxN matrix As a matter of fact this is a bidimensional representation of a tridimensional tissue volume f(x,y)

Image Representation 21 Each pixel of the bidimensional matrix represents the properties of a tissue volume called voxel Voxel dimension is a function of the field of view (FOV) In general the voxel cross-sectional plane dimensions are lower than the one of z axis (along-ray), voxels are parallelepiped Spatial resolution along z axis is lower than the one along x and y axes Anisotropy pixel y voxel x z

Parallel Ray Reconstruction 22 The basic CT measurements is a line integral of the effective linear attenuation coefficient within a cross-section µ We desire a picture of or a CT number over the entire cross-section. Can we reconstruct a picture of given a collection of its line integrals? µ The answer is yes! A line in the (x, y) plane is given by: { } L(, l θ) = ( x, y) xcosθ + ysinθ = l The unit normal to the line is oriented at an angle relative to x axis and the line is at distance l from the origin in the direction of the unit normal θ

Parallel Ray Reconstruction 23 The line integral of function f(x,y) is defined as: gl (, θ) = f( xy, ) δ( xcosθ + ysin θ ldxdy ) Being the line impulse defined as: δl ( xy, ) = δ( xcosθ + ycos θ l) The inpulse shifting property causes the integrand to be zero everywhere except on the line For a fixed θ glθ (, ), is called a projection θ l glθ (, ) For all and, is called the 2D Radon Transform of f( x, y)

Parallel Ray Reconstruction 24 If we make the following identifications f( x, y) = µ ( x, y) Unknown function to be reconstructed by CT I gl (, θ ) = ln( ) I CT Measurements It is possible to see that this mathematical abstraction characterizes the CT measurement situation It has to be noticed that this definition of projection corresponds to a collection of line integrals of parallel lines. This correspond to the geometry of 1G scanners only. However this formalism and methods can be applied also to reconstruct fan-beam projections. 0

Sinogram 25 glθ (, ) An image of with l and θ as rectilinear coordinates is called a SINOGRAM A sinogram is a pictorial representation of the Radon transform of and represents the data that are needed to reconstruct it There is no need to calculate the sinogram for angles higher than π since these projections are redundant f( x, y)

Backprojection glθ (, ) f( x, y) Let be the 2D Radon transform of. Consider the projection at θ = θ 0. There are an infinite number of functions that could give rise to this projection; it is not possible to determine uniquely form a single projection f( x, y) glθ (, ) l = l f( x, y) 0 Ll (, θ ) If takes on a large value at, then must be large 0 over the line (or somewhere over the line) 0 0 One way to create an image exploiting this property is to simply assign every point on the value of Ll (, θ ) gl ( 0, θ0) 0 0 l If we repeat this for every, the resulting function is called backprjection image and is given by bθ ( x, y) = g( xcosθ + ysin θθ, ) The backprojection image at a angle is consistent with the projection at θ 0 but its values are assigned with no prior information about the distribution of the image intensities θ 0 26

Backprojection It is possible to add-up backprojection images for every angle, obtaining the backprojection summation image, also called laminogram (, ) π f (, ) b x y = bθ x y dθ 0 27 Not the correct way to reconstruct f( x, y) Backprojection for θ = 30 Laminogram (blurriness)

d µ r 0 I Ie µ r = d 0 Backprojection I Ie µ r = d Backprojection assumption x x x x x I = I0e µ µ µ µ µ µ = µ /5 x ( d + d + d + d + d) r Object density is distributed along the whole x-ray path Star-like Artifact 28 2 4 Backprojection (2 projections) Backprojection (4 projections)

Projection-Slice Theorem 29 Lets take the 1D Fourier transform of a projection with respect to l πρl G( ρθ, ) F gl (, θ) gl (, θ) e dl = = 1D Being { } j2 ρ the spatial frequency glθ (, ) If we substitute the analitical expression of it is possible to obtain j2πρl G( ρθ, ) = f ( x, y) δ( x cosθ+ y sin θ l) e dxdydl j2 l f( x, y) πρ δ( xcosθ ysin θ l) e dldxdy = + = f ( x, y) e j2 πρ ( xcosθ + ysin θ ) dxdy

Projection-Slice Theorem 30 The last expression is reminescent of the 2D Fourier transform of j2 ( xu yv) F( u, v) = π + f ( x, y) e dxdy Being u and v the frequency variables in the x and y directions u = ρ cosθ v = ρ sinθ G( ρ, θ) = F( ρcos θ, ρsin θ) f( x, y) Making the following identifications: and leads to a very important equivalence, called the projection-slice theorem The 1D Fourier transform of a projection is a slice of the 2D Fourier transform of the object. The 1D Fourier transform of the projection, calculated along a direction, equals the Fourier transform of the object calculated along the same direction passing through the origin

Projection-Slice Theorem Graphical representation of the theorem 31 ρ θ and can be interpretated as the polar coordinates of the 2D Fourier transform

The Fourier Method 32 A very simple image reconstruction method follows immediately from the projection-slice theorem Take the 1D Fourier transform of each projection, insert it with the corresponding correct angular orientation into the correct slice of the 2D Fourier plane, and take the inverse 2D Fourier transform f( x, y) = F G( ρ, θ ) 2D 1 { } This method is not widely used in CT due to some practical problems: Interpolating polar data onto a Cartesian grid 2D inverse Fourier transform is time-consuming

Filtered Backprojection 33 The inverse Fourier transform of F( uv, ) can be written in polar coordinates 2π 2 ( cos sin ) (, ) ( cos, sin ) j πρ f x y F e x θ + y θ = d d 0 0 ρ θ ρ θ ρ ρ θ Using the projection-slice theorem π 2 ( cos sin ) (, ) (, ) j πρ f x y G e x θ + y θ d d 0 = ρ θ ρ ρ θ The term xcosθ y being gl (, θ ) = g( l, θ + π) θ + sin is constant considering integration over ρ f x y ρ G ρθ e dρ dθ π 2 (, ) (, ) j πρ = l 0 l= xcosθ+ ysinθ

Filtered Backprojection 34 The inner integral of the previous formula is an inverse 1D Fourier transform ρ The term makes it a filtering equation. The Fourier transform of a projection is multiplied by a frequency filter and inverse-transformed After the inverse-transform, the filtered projection is backprojected (this is done replacing l = xcosθ + ysinθ ), and then is performed a summation of all filtered projection ρ This reconstruction approach is called filtered backprojection and is faster and more flexible with respect to the Fourier method The term ρ is called ramp filter

Convolution Backprojection 35 From the convolution theorem of Fourier transforms follows π 1 { } 0 1D l= xcosθ+ ysinθ f( x, y) = F ρ g( l, θ ) d θ Defining cl () = F 1 1D { ρ } π 0 [ θ ] = cosθ+ sin f ( x, y ) = c ( l ) g ( l, ) d l x y θ θ π = gl (, θ ) cx ( cosθ + ysin θ ldld ) θ 0 Substitution of the convolution integral to obtain the equation for convolution backprojection

Convolution Backprojection 36 Performing a convolution rather than a filtering operation is generally more efficient if the inpulse response is narrow (it is so in this case) cl () does not exists since ρ is not integrable In practice ρ is windowed with suitable windowing functions Convolution backprojection algorithms use an approximate impulse response cl () = F D W 1 1 { ρ ( ρ )}

Reconstruction 37 Filtered backprojection uses three basic steps to reconstruct an image from a sinogram: filtering, backprojection and summation Convolution backprojection uses the same steps except that filtering is performed using a convolution Filtering step: each row of the sinogram is filtered through a ramp filter to obtain the filtered sinogram Ramp filter is essentially a high-pass filter and its value for f=0 is 0 High-frequency details are accentuated

Reconstruction 38 Backprojection is applied to each filtered projection, rather than to the raw data Last setp is the summation of filtered backprojections This step is performed using an accumulator concept No need to store hundreds of backprojection images

Reconstruction 39

CT Image Quality 40 CT image quality can be expressed as a function of: Spatial resolution in the transversal plane xy Longitudinal resolution and slice sensitivity profile Noise Some indicators of CT image quality will be introduced Contrast Artifacts

CT Image Quality 41 Spatial resolution is a function of the smalles details that can be identified in the image; it is stricly dependant on the detectors (and, in general, on the intruments used for data acquisition and reconstruction). Spatial resolution is measure of how closely lines can be resolved in an image, it is measured in mm or LP/cm (line pairs/cm). The clarity of the image is not the number of pixels in an image, in fact spatial resolution refers to the number of independent pixel values per unit length. Detectable lines Undetectable lines

CT Image Quality 42 Pixel dimensions can be considered the upper limit of the spatial resolution It is not possible to identify details having dimensions lower than pixel size Details having dimensions comparable to the pixel ones provides contributions to adjacent pixels

CT Image Quality 43 Slice sensitivity profile (SSP) is the impulse response of the CT system along the z axis (patient long axis). Thick slice Thin slice

CT Image Quality Image width affects contrast and noise of object Optimised slice width: imaged slice object size 44 Object 5mm 5mm Low contrast Low noise 1mm Better contrast More noise

CT Image Quality Noise is the random variation of color (grayscale), some regions of the image that should be uniform show coarseness 45 Noise is produced by the sensors and circuitry of the scanner Noise intensity can be evaluated considering a region of the image; the mean pixels intensity of the region is calcualted as well as standard deviation standard deviation Noise = % mean intensity Noise can be reduced through reconstruction filters Noise is in inverse proportion to layer thickness and radiation dose root square

CT Image Quality 46 Different kind of filters can be used in backprojection to optimize spatial resolution against noise: smooth, standard, detail, bone, etc. Smooth Filtering Low contrast Low noise Sharp Filtering Better contrast More noise

Artifacts 47 Artifacts can appear in CT images, due to different reasons Aliasing: if projections are undersampled higher frequency information will appear as lower frequency artifacts. An high number of detectors is needed, it is needed to sample at numerous angles Beam hardening: when travelling through te body the x-rays beam spectrum component with low energy is almost completely absorbed and the mean energy increases (this artifacs affect the bones) Other: electornic or system drift (miscalibration), x-ray scatter (can be prevented using collimators at expense of lower efficiency), motion (typically a scan require 1 to 10 secs) Beam herdening artifact