1 AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements.. To understand the basics for DQE and edqe measurement techniques. 3. To relate physical metrics of performance to clinical image quality. Medical imaging clinically meaningful information Quality of information image quality Image quality universally-appreciated universally-illusive
2 What is image quality? Image esthetics Subjective and qualitative Not necessarily reflective of diagnostic performance Diagnostic accuracy Expensive Impractical given the extent and variability of technologies Physical metrics of image quality Convenient and practical Effective to the extent they can be related to diagnostic performance Applicable under qualified conditions: Detector vs system Incorporation of scattered radiation Incorporation of anatomical attributes Incorporation of task attributes System-based vs image-based metrics Resolution Ability to resolve distinct features of an image from each other Resolution Best characterized by the modulation transfer function (MTF): The efficiency of an imaging system in reproducing subject contrast at various spatial frequencies MTF F{LSF(x)} Low resolution High resolution 7 8
3 High MTF Low MTF 9 10 Noise Unwanted signals that interfere with interpretation Noise Best characterized by the noise power spectrum (NPS): The variance of noise in an image in terms of the spatial frequencies Low resolution High resolution High res/ high noise 11 1 ACF ( x ) L L i( x x' ) i( x' ) dx' NPS( f ) { ACF ( x )} 1 3 3
4 Image without noise Uncorrelated noise Correlated noise Noise and Resolution > SNR Threshold contrast and diameter ~ 1/SNR (Rose model) Higher the SNR > Features w/ smaller C and D can be detected NEQ( f ) SNR actual MTF( f ) NNPS( f )
5 SNR efficiency Measurement method Best characterized by the detective quantum efficiency (DQE): Efficiency of a detector to utilize the maximum possible SNR provided by the finite number of x-ray photons forming the image Beam quality and magnitude kvp Filtration Exposure Access to linear or linearized data DQE( f ) SNR actual SNR ideal MTF( f ) NNPS( f ) E 0 q Al purity Aluminum versus Copper IEC recommendation: >99.9% purity RQA5, 99.0% purity RQA5, 99.99% purity RQA5, 70 kvp 1 mm Al 99.0% purity RQA5, 70 kvp 0.5 mm Cu, 1 mm Al Ranger et al, Med Phys, July Murphy et al, AAPM/COMP
6 MTF methodology: Edge Edge test device ESF deduced from image data MTF from ESF via differentiation and Fourier trans. High precision, particularly at low frequencies Simple and quick alignment Established method (endorsed by the IEC) Edge image processing 1. Finding the edge angle Least square fits Differentiation and Hough or Radon transform 1 Edge image processing railabs.org/resources. D projection to obtain ESF 3. ESF differentiation and smoothing to obtain LSF 4. Fourier transformation of LSF to get the presampled MTF 5. Normalization of the MTF at 0 frequency 6. Rebinning the MTF (0.05 mm -1 per IEC)
7 Effect of method (GE XQ/i) Diffs <6% overall <8% single f Opaque edge Slit Trans edge Effect of method (ideal) Diffs <0.3% for slit <5.% for edge Edge Slit True MTF 5 6 Effect of method (w/ glare) Diffs <10% for slit <% for edge Edge Slit True MTF DQE methodology DQE( f ) SNR actual SNR ideal MTF( f ) NNPS( f ) E 0 q MTF > Measured NNPS > Measured E 0 > Measured q (SNR ideal/e 0 ) > Estimated
8 Detector DQE Adding magnification, scatter, grid Grid Tube Patient 9 30 Adding magnification, scatter, grid edqe Ped CXR Adult CXR XL Adult CXR DQE( f ) SNR actual SNR ideal MTF( f ) NNPS( f ) E 0 q edqe( f m ) emtf( f m ) ( 1- SF) ennps( f m ) TF nb E 0 q Samei et al, Med Phys,
9 Resolution and contrast transfer From DQE to IQ Image noise: magnitude and texture Attributes of image feature of interest d' ò edqe( f m ).W( f ) fdf Model observers Fisher-Hotelling observer (FH) ' ( d FH ) òò MTF (u,v)w Task (u,v) dudv NPS(u,v) Non-prewhitening observer (NPW) ' [ òò MTF ( d NPW ) (u,v)w Task (u,v) dudv] MTF òò (u,v)w Task (u,v) NPS(u,v)dudv 0 NPW observer with eye filter (NPWE) ' [ òò MTF ( d NPWE ) (u,v)w Task (u,v) E (u,v)dudv] MTF òò (u,v)w Task (u,v) Eye filter NPS(u,v)E 4 (u,v) + MTF (u,v)w Task (u,v)n i dudv Validation of model observers edqe applied to acquisition optimization Richard, Li, Samei, SPIE 011 Boyce et al, Med Phys,
10 10 10 From DQE to IQ and dose: Effective Dose Efficiency (ede) Summary ede( f m ) edqe( f m ) q ES ED æ SSD ö ç è SID ø Samei et al, PMB, 011 medicalphysicsweb.org, July 9, MTF can be robustly measured using edge method DQE provides a meaningful measure of detector performance correlated with human performance DQE can be readily extended to system performance (via edqe) DQE can readily be normalized by dose (ede) for acquisition optimization 38
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries
2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
Quality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
Quality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM [email protected] 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Measuring Line Edge Roughness: Fluctuations in Uncertainty
Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS
ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS INFN-Laboratori Nazionali del Sud Via S. Sofia 44, Catania, Italy Patient positioned for treatment System under consideration (experimental
PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational
Digital Mammography Update: Design and Characteristics of Current Systems
: Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE
MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE K. Kohm ORBIMAGE, 1835 Lackland Hill Parkway, St. Louis, MO 63146, USA [email protected]
A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose
AAPM 2005-Continuing Education Course in Radiographic and Fluoroscopy Physics and Technology A Comparison of Screen/Film and Digital Imaging: Image Processing, Image Quality, and Dose Ralph Schaetzing,
Digital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
Lecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.
AN INNOVATIVE & CUTTING-EDGE DIGITAL RADIOGRAPHY SYSTEM
SAMSUNG SMART DIGITAL RADIOGRAPHY 2012 Samsung Electronics All rights reserved. Samsung Electronics reserves the right to modify the design, packaging, specifications and features shown herein, without
MATRIX TECHNICAL NOTES
200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR
High-resolution bio-imaging with liquid-metal-jet x-ray sources
High-resolution bio-imaging with liquid-metal-jet x-ray sources Hans Hertz Dept. of Applied Physics Royal Inst. of Technol. (KTH), Stockholm KTH: Ulf Lundström, Daniel Larsson, Anna Burvall, Tunhe Zou,
IQWorks progress to date and how you can get involved. Ed McDonagh, Andrew Reilly
IQWorks progress to date and how you can get involved Ed McDonagh, Andrew Reilly Progress since Nottingham Open source DICOM module Consolidation Evolution of algorithms Particularly edge MTF and NPS Display
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination
Moving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
Computed Radiography: Acceptance Testing and Quality Control
Computed Radiography: Acceptance Testing and Quality Control J. Anthony Seibert, Ph.D. University of California Davis Department of Radiology Sacramento, California Introduction CR is the primary means
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Digital Radiography Image Quality: Image Acquisition
Digital Radiography Image Quality: Image Acquisition Mark B. Williams, PhD a, Elizabeth A. Krupinski, PhD b, Keith J. Strauss, MS c, William K. Breeden, III, MS d, Mark S. Rzeszotarski, PhD e, Kimberly
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)
AngiX III FD Boost your confidence beyond the clinical. gehealthcare.com
AngiX III FD Boost your confidence beyond the clinical gehealthcare.com Boost your confidence beyond the clinical If you re considering investing in a system to do cardiovascular procedures, think beyond
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
Performance evaluation of computed radiography systems
Performance evaluation of computed radiography systems Ehsan Samei a) Department of Radiology, Duke University Medical Center, DUMC Box 3302, Durham, North Carolina 27710 J. Anthony Seibert Department
High Quality Image Magnification using Cross-Scale Self-Similarity
High Quality Image Magnification using Cross-Scale Self-Similarity André Gooßen 1, Arne Ehlers 1, Thomas Pralow 2, Rolf-Rainer Grigat 1 1 Vision Systems, Hamburg University of Technology, D-21079 Hamburg
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Various Technics of Liquids and Solids Level Measurements. (Part 3)
(Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,
Avalanche Photodiodes: A User's Guide
!"#$%& Abstract Avalanche Photodiodes: A User's Guide Avalanche photodiode detectors have and will continue to be used in many diverse applications such as laser range finders and photon correlation studies.
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics
Jeff Thomas Tom Holmes Terri Hightower Learn RF Spectrum Analysis Basics Learning Objectives Name the major measurement strengths of a swept-tuned spectrum analyzer Explain the importance of frequency
Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences [email protected] X-Ray
Variance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,
Optical Design for Automatic Identification
s for Automatic Identification s design tutors: Prof. Paolo Bassi, eng. Federico Canini cotutors: eng. Gnan, eng. Bassam Hallal Outline s 1 2 3 design 4 design Outline s 1 2 3 design 4 design s : new Techniques
The Exposure Index and Its Standardization
The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany [email protected] Content The Exposure Index (EI) What is it? What is it good
Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
Electron Microscopy 3. SEM. Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts
Electron Microscopy 3. SEM Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts 3-1 SEM is easy! Just focus and shoot "Photo"!!! Please comment this picture... Any
EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information
Towards large dynamic range beam diagnostics and beam dynamics studies. Pavel Evtushenko
Towards large dynamic range beam diagnostics and beam dynamics studies Pavel Evtushenko Motivation Linacs with average current 1-2 ma and energy 1-2.5 GeV are envisioned as drivers for next generation
Assessment of Camera Phone Distortion and Implications for Watermarking
Assessment of Camera Phone Distortion and Implications for Watermarking Aparna Gurijala, Alastair Reed and Eric Evans Digimarc Corporation, 9405 SW Gemini Drive, Beaverton, OR 97008, USA 1. INTRODUCTION
Improved glare (halos and scattered light) measurement for post-lasik surgery
Improved glare (halos and scattered light) measurement for post-lasik surgery Stanley Klein, Sam Hoffmann, Adam Hickenbotham, Sabrina Chan, and many more optometry students School of Optometry, Univ. of
Clinical Photon Beams
Dosimetric Characteristics of Clinical Photon Beams Jatinder R Palta PhD University of Florida Department of Radiation Oncology Gainesville, Florida Disclosures Research development grants from Philips
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
Metrics of resolution and performance for CD-SEMs
Metrics of resolution and performance for CD-SEMs David C Joy a,b, Yeong-Uk Ko a, and Justin J Hwu a a EM Facility, University of Tennessee, Knoxville, TN 37996 b Oak Ridge National Laboratory, Oak Ridge,
Improving digital image quality for larger patient sizes without compromises
Improving digital image quality for larger patient sizes without compromises Christiaan Kok, Technology Physicist Grids Smit Röntgen, Best, The Netherlands 7.5 7 6.5 fiber grids: 25 lp/cm with variable
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
Froedtert Hospital School of Radiology Curriculum Analysis Grid. Clinical Practice
Professional Curriculum Clinical Practice Procedural Performance Clinical Competency Basic Principles of Digital Radiography Image Acquisition Image Acquisition Errors Fundamental Principles of Exposure
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
Cone Beam Reconstruction Jiang Hsieh, Ph.D.
Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition
CONCEPT OF DETERMINISTIC ION IMPLANTATION AT THE NANOSCALE
CONCEPT OF DETERMINISTIC ION IMPLANTATION AT THE NANOSCALE Daniel Spemann Jan Meijer 1, Jürgen W. Gerlach, Paul Räcke 1, Susann Liedtke, Stephan Rauschenbach 2, Bernd Rauschenbach 1 University of Leipzig,
Learning Objectives: Quality Assurance (QA) is. Quality Assurance Procedures for Digital Radiography. Alternate definition of Quality Assurance (QA)
Quality Assurance Procedures for Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas
Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D.
Fundamentals of Digital Mammography: Physics, Technology and Practical Considerations Andrew P. Smith, Ph.D. Introduction Screen-film image receptors have been the standard detector used in conventional
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
Reprint (R22) Avoiding Errors in UV Radiation Measurements. By Thomas C. Larason July 2001. Reprinted from Photonics Spectra, Laurin Publishing
Reprint (R22) Avoiding Errors in UV Radiation Measurements By Thomas C. Larason July 2001 Reprinted from Photonics Spectra, Laurin Publishing Gooch & Housego 4632 36 th Street, Orlando, FL 32811 Tel: 1
Preview of Period 3: Electromagnetic Waves Radiant Energy II
Preview of Period 3: Electromagnetic Waves Radiant Energy II 3.1 Radiant Energy from the Sun How is light reflected and transmitted? What is polarized light? 3.2 Energy Transfer with Radiant Energy How
ACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
The Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO
The Whys, Hows and Whats of the Noise Power Spectrum Helge Pettersen, Haukeland University Hospital, NO Introduction to the Noise Power Spectrum Before diving into NPS curves, we need Fourier transforms
X-ray area detectors for synchrotron experiments Characteristics and Technologies
X-ray area detectors for synchrotron experiments Characteristics and Technologies Cyril Ponchut ESRF Instrumentation Services and Development Division ESI school 16-20 May 2011 X-ray area detectors for
Digital Camera Imaging Evaluation
Digital Camera Imaging Evaluation Presenter/Author J Mazzetta, Electro Optical Industries Coauthors Dennis Caudle, Electro Optical Industries Bob Wageneck, Electro Optical Industries Contact Information
Data Communications Competence Center
Importance of Cable Balance For Improving Noise Susceptibility Data Communications Competence Center DCCC03101702 July 11, 2007 Summary In a study of category 5e and category 6 UTP cables, a strong correlation
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
Modelling, Extraction and Description of Intrinsic Cues of High Resolution Satellite Images: Independent Component Analysis based approaches
Modelling, Extraction and Description of Intrinsic Cues of High Resolution Satellite Images: Independent Component Analysis based approaches PhD Thesis by Payam Birjandi Director: Prof. Mihai Datcu Problematic
Jitter Measurements in Serial Data Signals
Jitter Measurements in Serial Data Signals Michael Schnecker, Product Manager LeCroy Corporation Introduction The increasing speed of serial data transmission systems places greater importance on measuring
X- and Gamma Ray Imaging Systems based on CdTe-CMOS Detector Technology
X- and Gamma Ray Imaging Systems based on dte-mos Detector Technology Konstantinos Spartiotis, Ray Durrant, nssi Leppänen, Henrik Lohman, Olli Männistö, Tuomas Pantsar, Jouni Pyyhtiä, Tom Schulman, Oy
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution
Signal to Noise Instrumental Excel Assignment
Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems GE s Industrial X-ray and CT Forum, Cologne July 7 th 2015 Oliver Brunke GE Sensing & Inspection Technologies
A Guide to Acousto-Optic Modulators
A Guide to Acousto-Optic Modulators D. J. McCarron December 7, 2007 1 Introduction Acousto-optic modulators (AOMs) are useful devices which allow the frequency, intensity and direction of a laser beam
Encoders for Linear Motors in the Electronics Industry
Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy
Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations
Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,
Print quality requirements for single-pass inkjet printing the whole picture
Print quality requirements for single-pass inkjet printing the whole picture Mark Ritchie, Product Manager, Xaar plc The subject of print quality is an area which is very complex, subjective and difficult
Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology
Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment