Experimental study of beam hardening artefacts in photon counting breast computed tomography
|
|
- Wesley Hopkins
- 8 years ago
- Views:
Transcription
1 Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a, R. Pani d, M.G. Quattrocchi a, P. Randaccio e, V. Rosso a, P. Russo c a Università di Pisa and INFN, Pisa, Italy b Università di Bologna and INFN, Bologna, Italy c Università di Napoli Federico II and INFN, Napoli, Italy d Università La Sapienza and INFN, Roma, Italy e Università di Cagliari and INFN, Cagliari, Italy
2 Summary Univ. Federico II Beam hardening effect Bimodal energy model Beam hardening in PMMA slabs Experimental CT set-up Beam hardening in PMMA breast phantoms Conclusions and future work
3 Motivation and beam hardening effect X-ray Computed Tomography (CT) system on the gantry of a dedicated, scintillator based single photon emission tomography (SPECT) system for breast 99m-Tc imaging (see presentation S. Vecchio at this Conference); the breast would be scanned in a pendant geometry, i.e. with the patient in a prone position and the breast uncompressed; the beam energy distribution becomes more abundant in high energy photons and this effect causes an under-estimation or cupping artefact in the reconstructed attenuation coefficient at the center of the volume sample.
4 Bimodal energy model For a polychromatic beam the X-ray attenuation in a material is described by two effective energies (E 1, E 2 ; E 2 >E 1 ) and, correspondingly, by two effective attenuation coefficients μ 1 and μ 2 (<μ1): the lower value μ 2 at the beam effective energy E 2 accounts for the effective attenuation in large material thicknesses ln(i x /I 0 )=μ 2 x + ln{[1+α]/[1+αexp(μ 2 x-μ 1 x)]} α = f(e 1 )γ(e 1 )/ f(e 2 )γ(e 2 ) Source-Detector efficiency E. Van de Casteele et al., Phys. Med. Biol. 47, (2002) 4181
5 -ln (I x /I 0 ) Bimodal energy model: measurements Y = A + B*X A=0.221+/-0.03 B=0.244+/ cm -1 R= (p<0.0001) PMMA thickness, x (cm) ln(i x /I 0 )=μ 2 x + ln(1+α) for large thickness - a stack of 1 up to 14 PMMA sheets (20 20 cm 2, 1 cm thick) - CdTe diode detector (mod. XR-100T-CdTe) Amptek Inc.
6 CdTe detector Spectra Intensity (counts s -1 mm -2 kev -1 ma -1 ) Direct Beam (51.5 cm air) After 29 cm air + 14 cm PMMA cm air E mean = 47.3 kev E mean = 51.2 kev x Photon energy (kev) I 0 I 14 cm E 2 (Kev) 51.0 E 1 (Kev) 21.3 μ 2 (cm -1 ) μ 1 (cm -1 ) X-ray attenuation in PMMA as a function of material thickness: effective attenuation coefficient μ eff = cm -1 (E eff =51.0 kev)
7 Experimental set-up W-anode X-ray tube 80 kvp 4 56 fan beam A B C 0.3 mm Si Hybrid pixel detector 256 x 256 pixels, 55 x 55 μm 2 Detector intrinsic resolution: 110 μm Sensitive area mm 2 Readout: Single photon counting Medipix2 chip* * Developed by the Medipix2 collaboration, PMMA Phantoms 14 cm thick
8 Beam hardening in PMMA cylinder phantom -3D view of the reconstructed* transaxial slice of the 14 cm diameter PMMA cylinder; - isotropic voxel side= mm; - total thickness = 7.4 mm; views on 360-2D reconstruction of a single slice (thickness = mm); *Custom algorithm implementing the filtered backprojection fan beam reconstruction algorithm
9 Beam hardening in 14 cm thick PMMA cylinder phantom Attenuation coefficient (cm -1 ) 0,35 0,30 0,25 0,20 0,15 0,10 0,05 0,00 18% Distance along a diameter (cm) the drop of the attenuation coefficient (μ edge -μ center )/μ edge =18% ( 0.33 cm cm -1 ) - low detection efficiency - the charge sharing effect of the silicon pixel detector
10 Beam hardening in PMMA ellipsoid phantom 5 mm 7.6 mm 3D view of the CT reconstruction of three different sections of the PMMA ellipsoid phantom related to three different distances from the phantom top ( nipple ) A) distance = 10.5 cm, φ = 14 cm B) distance = 4.5 cm, φ = 11.5 cm C) distance = 0.5 cm, φ = 4 cm 7.6 mm
11 Attenuation Coefficient (cm -1 ) Beam hardening in PMMA ellipsoid phantom Profile at 10.5 cm from the top, φ = 14.0 cm Profile at 4.5 cm from the top, φ = 11.5 cm Profile at 0.5 cm from the top, φ = 4.0 cm Distance along the diameter (cm) (μ edge -μ center )/μ edge = 18% (μedge -μ center )/μ edge = 4% (μ edge -μ cente r)/μ edge = 12%
12 Conclusions and future work Preliminary tests for beam hardening cupping artefact in photon counting X-ray breast CT system using PMMA phantoms and a very fine pitch silicon pixel detector have been shown Drop of the attenuation coefficient of 4% when the PMMA thickness is 4-cm and of 18% for 14-cm PMMA thick material A bimodal energy model for beam hardening artefact in CT has been shown applicable to our data and produce an estimate of 19% for the attenuation coefficient drop for the 14- cm-diameter phantom Correction of the CT data in the pre-reconstruction phase will be applied and tests will be reported of this photon counting system, in comparison with an integrating flat panel detector
13 Bimodal Energy Model Attenuation coefficient (cm -1 ) A: (4 cm, cm -1 ) drop=7% B: (11.5 cm, cm -1 ) drop=13% C: (14 cm, cm -1 ) drop=19% A μ α μ PMMA thickness (cm) B C Calculated attenuation coefficient as a function of PMMA thickness
14 Experimental set-up for PMMA attenuation coefficient evaluation Univ. Federico II W Anode 80 kvp, 0.25 ma 4.2 mm Al 14 PMMA sheets 1cm thick CdTe detector (mod. XR-100T-CdTe) 36 cm 15.5cm 51.5 cm X-ray tube: W anode with a 40 μm focal spot size (Source-Ray, Inc., mod. SB , NY, USA). 35 kvp to 80 kvp with an anode current in the range μa fan beam irradiation geometry (4 deg horizontal 56 deg vertical) CdTe diode detector (mod. XR-100T-CdTe) associated at power supply amplifier (mod. PX2T-CR) from Amptek Inc., Bedford, MA, USA
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems GE s Industrial X-ray and CT Forum, Cologne July 7 th 2015 Oliver Brunke GE Sensing & Inspection Technologies
More informationCT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
More informationThinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
More informationIndirect X-ray photon counting image sensor with 27T pixels and 15 electrons RMS accurate threshold
Indirect X-ray photon counting image sensor with 27T pixels and 15 electrons RMS accurate threshold B. Dierickx 1,2, B. Dupont 1,3, A. Defernez 1, N. Ahmed 1 1 Caeleste, Antwerp, Belgium 2 Vrije Universiteit
More informationLaue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
More informationInformation about the T9 beam line and experimental facilities
Information about the T9 beam line and experimental facilities The incoming proton beam from the PS accelerator impinges on the North target and thus produces the particles for the T9 beam line. The collisions
More informationComputed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution
More informationPerformance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using GATE
Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using Parham Geramifar 1,3,4, Mohammad Reza Ay 2,3,4, Mojtaba Shamsaie Zafarghandi 1, George Loudos 5, Arman
More informationMODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada
More informationPET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
More informationFundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
More informationCone Beam Reconstruction Jiang Hsieh, Ph.D.
Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition
More informationAmptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
More informationIrradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
More informationScan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
More informationInstruction Sheet 554 828
08/13-W07-Hund Instruction Sheet 554 828 X-ray image sensor (554 828) Precision slide X-ray image sensor (554 829) Software Computed Tomography Pro (554 820) Sensor head on precision slide with USB module
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationAn iterative approach to the beam hardening correction in cone beam CT
An iterative approach to the beam hardening correction in cone beam CT Jiang Hsieh a) Applied Science Laboratory, GE Medical Systems, Milwaukee, Wisconsin 53201 Robert C. Molthen Department of Biomedical
More informationHigh resolution X-ray CT as enabling technology for material research
High resolution X-ray CT as enabling technology for material research Veerle Cnudde Dept. of Geology & Soil Science - UGCT, Ghent University, Belgium With credits to the entire UGCT team (www.ugct.ugent.be)
More informationDevelopment of on line monitor detectors used for clinical routine in proton and ion therapy
Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary
More informationImage Analysis for Volumetric Industrial Inspection and Interaction
Image Analysis for Volumetric Industrial Inspection and Interaction Rasmus Larsen Mads F. Hansen, Hildur Olafsdottir, Thomas H. Mosbech, Lasse F. Laursen DTU Informatics (CT) reconstruction from few projections
More informationProton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
More informationMEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
More informationEvaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 15, NUMBER 5, 2014 Evaluation of a metal artifact reduction algorithm in CT studies used for proton radiotherapy treatment planning Karin M. Andersson,
More informationCT Traceability - Prof. Wim Dewulf, Group T - KU Leuven
CT Traceability - Calibration and Accuracy Prof. Wim Dewulf, Group T - KU Leuven Outline Introduction: terminology and procedures Voxel size calibration Edge offset calibration Conclusions Outline Introduction:
More informationCT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
More informationDevelopment and quantitative assessment of a beam hardening correction model for preclinical micro-ct
University of Iowa Iowa Research Online Theses and Dissertations 2012 Development and quantitative assessment of a beam hardening correction model for preclinical micro-ct Sucheta Mohapatra University
More informationStationary digital breast tomosynthesis system with a multi-beam field emission x-ray source array
Stationary digital breast tomosynthesis system with a multi-beam field emission x-ray source array Guang Yang a, Ramya Rajaram b, Guohua Cao a, Shabana Sultana b, Zhijun Liu a, David Lalush c, Jianping
More informationQuality control of CT systems by automated monitoring of key performance indicators: a two-year study
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,
More informationAutomated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging
Automated Pose Determination for Unrestrained, Non-anesthetized Small Animal Micro-SPECT and Micro-CT Imaging Shaun S. Gleason, Ph.D. 1, James Goddard, Ph.D. 1, Michael Paulus, Ph.D. 1, Ryan Kerekes, B.S.
More informationX-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
More informationJorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy
Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino
More informationDevelopment of High Speed Oblique X-ray CT System for Printed Circuit Board
計 測 自 動 制 御 学 会 産 業 論 文 集 Vol. 6, No. 9, 72/77 (2007) Development of High Speed Oblique X-ray CT System for Printed Circuit Board Atsushi TERAMOTO*, Takayuki MURAKOSHI*, Masatoshi TSUZAKA**, and Hiroshi
More informationAn Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
More informationAssessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
More informationX- and Gamma Ray Imaging Systems based on CdTe-CMOS Detector Technology
X- and Gamma Ray Imaging Systems based on dte-mos Detector Technology Konstantinos Spartiotis, Ray Durrant, nssi Leppänen, Henrik Lohman, Olli Männistö, Tuomas Pantsar, Jouni Pyyhtiä, Tom Schulman, Oy
More informationCALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY
243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution
More informationGamma Ray Detection at RIA
Gamma Ray Detection at RIA Summary Report: Physics & Functional Requirements Cyrus Baktash Physics goals Experimental tools: Techniques & Reactions Functional Requirements Physics Questions (Discussed
More informationMDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
More informationCyclotron Centre in Poland and 2D thermoluminescence dosimetry
Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of
More information(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier
(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (no PiN and pinned Diodes) Peter Fischer P. Fischer, ziti, Uni Heidelberg, Seite 1 Overview Reminder: Classical Photomultiplier
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
More informationSimulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters
C-A/AP/#122 November 23 Simulation of Proton Therapy Treatment Verification via PET imaging of Induced Positron-Emitters J. Beebe-Wang, P. Vaska, F.A. Dilmanian, S.G. Peggs and D.J. Schlyer "This paper
More informationProper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster. Jesse Garant, JG&A Metrology Center
Proper Implementation of Industrial CT Scanning to Reduce Inspection Costs & Get to Production Faster Jesse Garant, JG&A Metrology Center Traditional Metrology and Inspection Tactile Devices (Touch Probe)
More informationMicro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Micro-CT Innovation with Integrity
Micro-CT for SEM Non-destructive Measurement and Volume Visualization of Specimens Internal Microstructure in SEM Innovation with Integrity Micro-CT 3D Microscopy Using Micro-CT for SEM Micro-CT for SEM
More informationWir schaffen Wissen heute für morgen
Diffractive optics for photon beam diagnostics at hard XFELs Wir schaffen Wissen heute für morgen PSI: SLAC: ESRF: SOLEIL: APS: SACLA: EuroXFEL C. David, S. Rutishauser, P. Karvinen, Y. Kayser, U. Flechsig,
More informationR/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
More informationAPPLICATION OF X-RAY COMPUTED TOMOGRAPHY IN SILICON SOLAR CELLS
APPLICATION OF X-RAY COMPUTED TOMOGRAPHY IN SILICON SOLAR CELLS V.A. Popovich 1, W. Verwaal 2, M. Janssen 1, I. J. Bennett 3, I.M.Richardson 1, 1. Delft University of Technology, Department of Materials
More informationResults from first tests of TRD prototypes for CBM. DPG Frühjahrstagung Münster 2011 Pascal Dillenseger Institut für Kernphysik Frankfurt am Main
Results from first tests of TRD prototypes for CBM DPG Frühjahrstagung Münster 2011 Pascal Dillenseger Institut für Kernphysik Contents Overview of the CBM experiment CBM-TRD General TRD requirements The
More informationGamma and X-Ray Detection
Gamma and X-Ray Detection DETECTOR OVERVIEW The kinds of detectors commonly used can be categorized as: a. Gas-filled Detectors b. Scintillation Detectors c. Semiconductor Detectors The choice of a particular
More informationIntroduction to Imaging 3 Giulia Festa Università degli Studi di Roma Tor Vergata 27 September 2010 Contents Prompt gamma Activation Analysis (PGAA) Prompt Gamma Activation Imaging combined with Neutron
More informationRb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals
Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction
More informationDose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi
Dose enhancement near metal electrodes in diamond X- ray detectors Acknowledgements Surrey University: P.J. Sellin M. Abd-El Rahman P. Veeramani H. Al-Barakaty F. Schirru Mechanical Workshop A. Lohstroh*,
More informationI-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit
I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete
More informationLectures about XRF (X-Ray Fluorescence)
1 / 38 Lectures about XRF (X-Ray Fluorescence) Advanced Physics Laboratory Laurea Magistrale in Fisica year 2013 - Camerino 2 / 38 X-ray Fluorescence XRF is an acronym for X-Ray Fluorescence. The XRF technique
More informationExperimental and modeling studies of imaging with curvilinear electronic eye cameras
Experimental and modeling studies of imaging with curvilinear electronic eye cameras Viktor Malyarchuk, 1 Inhwa Jung, 1 John A. Rogers, 1,* Gunchul Shin, 2 and Jeong Sook Ha 2 1 Department of Materials
More informationCT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
More informationPoS(PhotoDet 2012)068
Characterization of the Hamamatsu R11265 multi-anode photomultiplier tube with single photon signals Luca CADAMURO, Marta CALVI, Andrea GIACHERO,, Matteo MAINO, Clara MATTEUZZI, Gianluigi PESSINA Dipartimento
More informationConcepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
More informationkv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
More informationIndustrial X-ray for Nondestructive Testing Unrestricted Siemens AG 2014. All rights reserved
Overview of X-ray Technology and Competence offered by Corporate Technology Industrial X-ray for Nondestructive Testing Nondestructive Testing (NDT) with X-rays: Our offer at a glance High-tech X-ray lab
More informationMODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT
MODERN VOXEL BASED DATA AND GEOMETRY ANALYSIS SOFTWARE TOOLS FOR INDUSTRIAL CT C. Reinhart, C. Poliwoda, T. Guenther, W. Roemer, S. Maass, C. Gosch all Volume Graphics GmbH, Heidelberg, Germany Abstract:
More informationSpiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences mmcnittgray@mednet.ucla.edu X-Ray
More informationQUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
More informationFull-field kinematic measurements by Digital Image Correlation for material characterization at different scales
Department of Civil and Environmental Engineering (DICA), Politecnico di Milano, June 8 th, 2015 Full-field kinematic measurements by Digital Image Correlation for material characterization at different
More informationAN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
More informationPERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
More informationChapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
More informationVolume visualization I Elvins
Volume visualization I Elvins 1 surface fitting algorithms marching cubes dividing cubes direct volume rendering algorithms ray casting, integration methods voxel projection, projected tetrahedra, splatting
More informationHigh-resolution bio-imaging with liquid-metal-jet x-ray sources
High-resolution bio-imaging with liquid-metal-jet x-ray sources Hans Hertz Dept. of Applied Physics Royal Inst. of Technol. (KTH), Stockholm KTH: Ulf Lundström, Daniel Larsson, Anna Burvall, Tunhe Zou,
More informationElectron-Muon Ranger (EMR)
Electron-Muon Ranger (EMR) Status of EMR Project R.Asfandiyarov 1, P.Béné 1, R.Bloch 1, A.Blondel 1, D.Bolognini 3 F.Cadoux 1, S.Débieux 1, J-S.Graulich 1, C.Husi 1, D.Lietti 3, F.Masciocchi 1, L.Nicola
More informationHigh Quality Cone-beam CT Reconstruction on the GPU
High Quality Cone-beam CT Reconstruction on the GPU Gábor Jakab 1,2, Attila Rácz 1, Kálmán Nagy 1 1 Mediso Medical Equipment Developing and Service Ltd. Budapest, 1022 Hungary http://www.mediso.hu, research@mediso.hu
More informationTwo years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions
Two years ANR Post Doctorate fellowship Mathematics and Numerics of Dynamic Cone Beam CT and ROI Reconstructions Laurent Desbat TIMC-IMAG, UMR 5525, Grenoble University, France Laurent.Desbat@imag.fr Rolf
More informationCT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
More informationCHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
More informationFor many years, physicians have been faced with the. SPECT/CT Physical Principles and Attenuation Correction*
SPECT/CT Physical Principles and Attenuation Correction* James A. Patton 1 and Timothy G. Turkington 2 1 Vanderbilt University Medical Center, Nashville, Tennessee; and 2 Duke University Medical Center,
More informationCathode Ray Tube. Introduction. Functional principle
Introduction The Cathode Ray Tube or Braun s Tube was invented by the German physicist Karl Ferdinand Braun in 897 and is today used in computer monitors, TV sets and oscilloscope tubes. The path of the
More informationPurchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage
More informationRole of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
More informationPositron Emission Tomography: Principles, Technology, and Recent Developments
Nuclear Physics A 752 (2005) 679c 687c Positron Emission Tomography: Principles, Technology, and Recent Developments Sibylle I. Ziegler Nuklearmedizinische Klinik, Klinikum rechts der Isar der Technischen
More informationMethods for analysis and visualization of industrial XCT data of fibre reinforced composites
Methods for analysis and visualization of industrial XCT data of fibre reinforced composites Christoph Heinzl Team DI(FH) Michael Reiter Software development, simulation, dimensional measurement Andreas
More informationThe disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More informationISTITUTO NAZIONALE DI FISICA NUCLEARE
ISTITUTO NAZIONALE DI FISICA NUCLEARE Sezione di Catania INFN/TC-11/02 February 28, 2011 PRELIMINARY TESTS OF A SCINTILLATOR-BASED MINI-STATION FOR EXTENSIVE AIR SHOWERS MEASUREMENTS S.Aiola 1, P. La Rocca
More informationThis document is the author s post-print version of this article, i.e. the final draft version after review. The final document will be printed in
1 2 3 4 5 6 7 This document is the author s post-print version of this article, i.e. the final draft version after review. The final document will be printed in Nuclear Instruments and Methods in Physics
More informationDENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
More informationSpectral distribution from end window X-ray tubes
Copyright (C) JCPDS-International Centre for Diffraction Data 1999 393 Spectral distribution from end window X-ray tubes N. Broll 1, P. de Chateaubourg 2 1 FORTEX - E.N.S.A.I.S. 24, bld de la Victoire,
More informationDICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
More informationX-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
More informationIGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
More informationINFN Rome Silicon Microstrip Detector for SBS F. De Persio S. Kiprich - F. Meddi G.M. Urciuoli
INFN Rome Silicon Microstrip Detector for SBS F. De Persio S. Kiprich - F. Meddi G.M. Urciuoli 1 Summary 1) About us: Presentation of the JLAB12 INFN Roma group. 2) Silicon Detector Advantages on SBS:
More information7üVNORD. TÜV NORD Röntgentechnik
.. TÜV NORD Röntgentechnik Niedersachsen Nordrhein-Westfalen Hamburg Schleswig-Holstein Baden-Württemberg Mecklenburg-Vorpommern Bremen Sachsen-Anhalt Berlin Brandenburg Sachsen' Thüringen' Bayern' Hessen
More informationTrack Trigger and Modules For the HLT
CMS L1 Track Trigger for SLHC Anders Ryd for the CMS Track Trigger Task Force Vertex 2009 Sept. 13-18, 2009 L=1035 cm-2s-1 Outline: SLHC trigger challenge Tracking triggers Track trigger modules Simulation
More informationCMS Tracker module / hybrid tests and DAQ development for the HL-LHC
CMS Tracker module / hybrid tests and DAQ development for the HL-LHC S. Mersi, G. Auzinger georg.auzinger@cern.ch 1 Outline Reminder: the Ph2 CMS Tracker upgrade pt Modules: principle, elements, electronics
More informationT(CR)3IC Testbed for Coherent Radio Cherenkov Radiation from Cosmic-Ray Induced Cascades
T(CR)3IC Testbed for Coherent Radio Cherenkov Radiation from Cosmic-Ray Induced Cascades R. Milinčić1, P. Gorham1, C. Hebert1, S. Matsuno1, P. Miočinović1, M. Rosen1, D. Saltzberg2, G. Varner1 1 University
More informationAutomation of a CT Acquisition: A System-based User-Support. 1. Introduction. Bärbel KRATZ 1, Frank HEROLD 1, Malte KURFISS 1
11th European Conference on Non-Destructive Testing (ECNDT 2014), October 6-10, 2014, Prague, Czech Republic Automation of a CT Acquisition: A System-based User-Support More Info at Open Access Database
More information3-D Printing Artificial Reservoir Rocks to Test Their Petrophysical Properties*
3-D Printing Artificial Reservoir Rocks to Test Their Petrophysical Properties* Sergey Ishutov 1 and Franciszek Hasiuk 1 Search and Discovery Article #41427 (2014) Posted August 29, 2014 *Adapted from
More informationMedical Applications of radiation physics. Riccardo Faccini Universita di Roma La Sapienza
Medical Applications of radiation physics Riccardo Faccini Universita di Roma La Sapienza Outlook Introduction to radiation which one? how does it interact with matter? how is it generated? Diagnostics
More informationDesign and Implementation of an Institution-Wide Patient-Specific Radiation Dose
Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Monitoring Program for Computed Tomography, Digital Radiography, and Nuclear Medicine by Olav Christianson Medical Physics
More informationThe Timing Counter of the MEG experiment: calibration and performances
The Timing Counter of the MEG experiment: calibration and performances INFN Pavia On behalf of MEG TC group 12th seminar on IPRD Siena 1 MEG: signal and background background signal µ e γ physical e +
More informationJoint Research Centre
Joint Research Centre Open Source Monitoring Tools and Applications emm.newsbrief.eu Serving society Stimulating innovation Supporting legislation Open Source Monitoring - Overview EMM Introduction Custom
More information