1 Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo Clinic, Rochester MN
2 Disclosures NIH: Research Support: EB DK DK EB RR Siemens Healthcare Off Label Usage None
3 Disclosures Our team currently supports 25 CT systems. Presently they are from only two manufacturers (GE and Siemens) 10 distinct scanner models (all multi-slice) 12 distinct multi-slice models in our practice since 1998 Examples come from the systems I know best Am an equal opportunity critic, so if I know about weaknesses (or strengths) of other systems, I m happy to share those also.
4 The first thing you need is a dictionary
5
6 Clarity, Transparency, and Uniformity
7 Collimation vs. Slice Width
8
9
10 The next thing you need is data When is a 5-mm not a 5-mm? Which pitch values give best images? Which collimations are most/least dose efficient? Which reconstruction algorithms/kernels have special features or alter CT number accuracy? Which bowtie is used for which scan modes? Which focal spot is used when? etc.
11 Measured width of 5 mm image Full vs. Plus reconstruction option
12 Full or Plus Mode Full mode will retain the prescribed slice thickness Plus mode will give you a thicker slice thickness than prescribed (about 20% thicker, e.g. a 5 mm becomes a 6 mm). Correspondingly, noise level will be about 10% lower.
13 Pitch vs. Image Quality In spiral CT, image noise is dependent on pitch To compensate, mas must be changed as pitch is changed Relationship is linear on some systems, but not all Relationship is different for cardiac reconstructions Noise is INDEPENDENT of pitch in cardiac CT Image width can be affected by pitch Windmill and cone beam artifacts affected by pitch Windmill artifacts discussed in talks by Gupta, Morin
14 Study to evaluate cone beam artifacts A thin-walled object with edges at an angle to the scan plane Rate of change of funnel shape is constant along the z-axis Scanned in air, the funnel has high contrast (~ 500 HU) Courtesy D. Platten et al. ImPACT (RSNA 2003)
15 Courtesy D. Platten et al. ImPACT (RSNA 2003) Example images Single slices through the funnel appear as rings MIP image of many slices results in a wider ring If perfect the images should be uniform
16 Cone-beam algorithm on and off Low pitch (0.5), Siemens Sensation 16 Standard Cone-beam (AMPR) Courtesy D. Platten et al. ImPACT (RSNA 2003)
17 Cone-beam algorithm on and off High pitch (1.5), Siemens Sensation 16 Standard Cone-beam (AMPR) Courtesy D. Platten et al. ImPACT (RSNA 2003)
18 Cone-beam algorithm on and off High pitch (1.5), Philips Mx8000 IDT Standard Cone-beam (COBRA) Courtesy D. Platten et al. ImPACT (RSNA 2003)
19 Cone-beam algorithm on and off High pitch (1.5), Toshiba Aquilion 16 Standard Cone-beam (TCOT) Courtesy D. Platten et al. ImPACT (RSNA 2003)
20 Cone-beam algorithm with pitch GE LightSpeed 16, cone-beam reconstruction always on Courtesy D. Platten et al. ImPACT (RSNA 2003)
21 Clinical relevance Standard Courtesy D. Platten et al. ImPACT (RSNA 2003) Cone-beam
22 Inclined (60 ) Teflon rod High pitch (1.5), Siemens Sensation 16 Standard Cone-beam (AMPR) Standard 13 cm off center Cone-beam 13 cm (AMPR) off center Courtesy D. Platten et al. ImPACT (RSNA 2003)
23 Dose Efficiency vs. Collimation Siemens Sensation 16
24 GE Recon Algorithms Soft Standard Detail Lung Bone Edge Bone Plus
25 CT Number Accuracy Some edge-enhancing algorithms/kernels can alter CT numbers E.g. GE Lung and Bone Plus Boedeker et al. Emphysema: Effect of reconstruction algorithm on CT imaging measures. Radiology 2004 Zhang, McCollough, et al. Selection of Appropriate Computed Tomographic Image Reconstruction Algorithms for a Quantitative Multicenter Trial of Diffuse Lung Disease. JCAT 2008
26 Boone: Presampled MTF in CT (Med Phys 2000)
27 Siemens Recon Kernels B10 B90 Body H10 H90 Head U30 U90 Ultra High Resolution T20 T81 Topogram Lower number smoother Higher number sharper Multiples of 10 are the basic kernels In between values are special kernels
28 Siemens Recon Kernels B18 B20 B25/B26 - cardiac B30 B31 finer grain noise B35/36 - calcium B40 B41 finer grain noise B45 B46 - cardiac/lung B50 B70 H30 H31 finer grain noise H32 no PFO H37 GE like H40 H41 finer grain noise H42 no PFO H47 GE like H48 GE like but sharper U70
29 Special Body Kernels B25 and B26 are for cardiac with edge-preserving noise reduction. B35 and B36 are for Ca scoring without edge enhancement. B45 is intermediate sharpness between B40 and B50 (e.g. not very special ) B46 is designed specifically for accurate assessment of inside coronary stents with 3D edge preserving noise reduction techniques. B75 is comparable to B70 in sharpness but used 2D edge-preserving noise reduction
30 B10
31 B20
32 B25
33 B26
34 B30
35 B31
36 B35
37 B36
38 B40
39 B41
40 B45
41 B46
42 Noise and Noise Uniformity B31/41 is like B30/B40 but with finer grain noise and a milder edge enhancement. Noise more uniform over FOV.
43 Special Head Kernels H21, H31, H41 are like H20, H30, H40 but with finer grain noise and a milder edge enhancement. H22, H32, H42 don t include iterative beam hardening correction (PFO). Reconstruction speed is faster, but the reconstructed images may contain significant beam hardening artifacts. H37 is comparable to GE Soft H45 is intermediate sharpness between H40 and H50 H47 is comparable to GE Standard H48 is like H47 but a bit sharper
44 H10
45 H20
46 H21
47 H22
48 H30
49 H31
50 H32
51 H37
52 H40
53 H41
54 H42
55 H45
56 H47
57 H48
58 H50
59 H60
60 H70
61 Have I made your head spin yet?
62 GE bowtie and focal spot selection (once upon a time)
63 Moral of the story There are many good reasons to invoke special features and characteristics Manufacturers often tie these features to protocols where they make sense There are many ways to get these features when you don t want them or to not find them when you do Often users are not even educated about them Don t stop having good ideas and features But make them transparent so user knows what they do and when they are used
64 Lastly, you need deliverables To design or translate a protocol, you need to know what the final product needs to look like Scan time (total) and per image (temporal resolution) Slice width(s) and image plane(s) required The thinnest image width determines the detector configuration Coronals and sagitals require thinner collimation Image sharpness or smoothness Noise level Target anatomy/patient (pediatric, obese, cardiac, etc) Diagnostic reference level (CTDIvol)
65 Knowing terminology, operation, features (quirks) and performance of your system(s) You can translate across manufacturer (make) and model to yield the desired deliverables Usually not one way to accomplish the same results Usually not a lot of ways Evaluate options as quantitatively as possible on phantoms and then form a WIP prootcol for clinical evaluation/refinement Some differences between seemingly similar options can show up only in patients, where motion and specific diagnostic criteria (like noise texture or subtle enhancement of small structures) come into play
66 Routine Chest
67 Routine Chest
68 Routine Chest
69 Routine Chest
70 Routine Chest
71 Routine Chest
72 Routine Chest
73 Mayo CT Clinic Innovation Center and Dept. of Radiology J. Kofler, L. Yu, S. Leng, M. Bruesewitz, T. Vrieve
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
More informationMoving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
More informationPurchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage
More informationCynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905
The phantom portion of the American College of Radiology ACR Computed Tomography CT accreditation program: Practical tips, artifact examples, and pitfalls to avoid a Cynthia H. McCollough b) and Michael
More informationCT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
More informationMDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
More informationState-of-the-Art Technology in Cardiac CT
1 2 Next Step Evolution or Revolution? State-of-the-Art Technology in Cardiac CT Stefan Ulzheimer, PhD Global Director of Collaborations CT Siemens Medical Solutions Major Innovations in CT Head The 80
More informationThe disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More informationSpiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences mmcnittgray@mednet.ucla.edu X-Ray
More informationCT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
More informationFundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
More informationCone Beam Reconstruction Jiang Hsieh, Ph.D.
Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition
More informationChest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School
Chest CT protocols Mannudeep K. Kalra, MD, DNB Dianna D. Cody, PhD Massachusetts General Hospital Harvard Medical School M.D. Anderson Cancer Center Specific principles Routine chest CT Lung nodule follow
More informationAmerican College of Radiology CT Accreditation Program. Testing Instructions
American College of Radiology CT Accreditation Program Testing Instructions (Revised July 24, 2015) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission
More informationThe disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More information2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography
Today s agenda Computed Tomography Presented by: Dina Hernandez, BSRT, RT (R), CT, QM Krista Bush, RT, MBA Leonard Lucey, JD ACR Quality & Safety MIPPA overview and CMS requirements CT accreditation How
More informationThe disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More informationAssessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
More informationQUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
More informationTable of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry
Table of Contents Scan acquisition and user interface basics Dose modulation and reduction tools Multi-Slice Detector Geometry Image Reconstruction and Display Contrast Media Tools Multi-planar formats
More information3/12/2014. Disclosures. Understanding IAC CT Accreditation. Outline. Learning Objectives. Who is the IAC?
Disclosures Understanding IAC CT Accreditation Serve as one of two AAPM representatives to IAC CT Board of Directors, Serve as IAC representative to AAPM for focus group on accreditations Provide 3-hour
More informationMaster s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III. Prof. Dr. Lothar Schad.
1 12/9/2008 Page 1 Master s Program in Medical Physics Physics of Imaging Systems Basic Principles of Computer Tomography (CT) III Chair in Faculty of Medicine Mannheim University of Heidelberg Theodor-Kutzer-Ufer
More informationSiemens Computed Tomography
Siemens Computed Tomography Innovation Leader. The Cosmopolitan Information Topic Waltraud Winter Siemens Copyright Siemens AG 2011. All rights reserved. Page 1 October 2012 Waltraud Winter An Innovation
More informationLow-dose CT for Pulmonary Embolism
Low-dose CT for Pulmonary Embolism Gautham Gautham P. P. Reddy, Reddy, MD, MD, MPH MPH University University of of Washington Washington Introduction Introduction CT CT accounts accounts for for > 50%
More informationPerformance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
More informationMulti-slice Helical CT Scanning of the Chest
Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.
More informationProtocol Management and Review Strategies The MD Anderson Experience
Protocol Management and Review Strategies The MD Anderson Experience Dianna Cody, Ph.D. Professor Dept. of Imaging Physics Univ. of Texas M.D. Anderson Cancer Center Objectives Setting up a policy for
More informationMulti Detector Row CT Systems and Image- Reconstruction Techniques 1
Special Review Radiology Thomas G. Flohr, PhD Stefan Schaller, PhD Karl Stierstorfer, PhD Herbert Bruder, PhD Bernd M. Ohnesorge, PhD U. Joseph Schoepf, MD Published online before print 10.1148/radiol.2353040037
More informationCopyright March 1, 2016 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
More informationPhysics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
More informationTracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives
Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives Gilbert W. Beebe Symposium The National Academies Katharine Grant, PhD Staff Scientist 8 December 2011 For internal
More informationGE Healthcare. Revolution EVO. More than just high tech. Higher purpose.
GE Healthcare Revolution EVO More than just high tech. Higher purpose. Revolution EVO. Designed with purpose. Today s healthcare environment is about creating new solutions to pressing needs. It s about
More informationPET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
More informationProton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
More informationkv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
More informationCT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet
CT QC Under the ACR QC Manual Douglas Pfeiffer, MS, DABR Boulder Community Hospital What? There s a Manual??! YES!!! Almost! Doesn t have a pretty cover yet! Should be out by RSNA! Really seriously I mean
More informationMx8000 IDT 16 provides dose control methodologies like DoseRight ACS (Automatic Current Selection) and DoseRight DOM (dynamic dose modulation).
Philips Mx8000 IDT 16 CT Scanner Philips Medical systems Mx8000 IDT 16 delivers 16 simultaneous slices at sub-millimeter collimator, high resolution and isotropic imaging. New large area detectors allow
More informationSistemas de reducción de dosis en CT para pediatría. Nuevos avances Ángela de Pinto Siemens CT Business Manager angela.pinto@siemens.
Sistemas de reducción de dosis en CT para pediatría. Nuevos avances Ángela de Pinto Siemens CT Business Manager angela.pinto@siemens.com Copyright Siemens AG 2009. All rights reserved. An Innovation Leader
More informationQuality control of CT systems by automated monitoring of key performance indicators: a two-year study
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,
More informationAAPM Medical Physics Practice Guideline 1.a: CT Protocol Management and Review Practice Guideline
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 14, NUMBER 5, 2013 AAPM Medical Physics Practice Guideline 1.a: CT Protocol Management and Review Practice Guideline The American Association of Physicists
More informationQuality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM Melissa@TherapyPhysics.com 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
More informationThinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
More informationDENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
More informationUse of 3D Printers in Proton Therapy
Use of 3D Printers in Proton Therapy Nicholas Remmes, PhD Mayo Clinic, Rochester, MN Department of Radiation Oncology Remmes.Nicholas@mayo.edu 2015 MFMER slide-1 Disclosures/Confessions My 3D Printing
More informationBrilliance CT 64-channel confi guration
Brilliance CT 64-channel confi guration with Essence technology The Brilliance CT 64-channel confi guration is designed to help you conduct the most advanced multislice CT studies possible. These systems
More informationPediatric Hospitals Bring Low-dose CT to the Middle East
Pediatric Hospitals ring Low-dose CT to the Middle East For years, radiologists have been cognizant of the importance of limiting pediatric patients exposure to radiation dose. uilding on the LR principle,
More informationImplementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
More informationPrepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
More informationData. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
More informationDose Modulation Technique in CT
Content Dose Modulation Technique in CT Short Overview - Overview - Dose Saving Features - Take Home Point - Conclusions Senior Application specialist Siemens Limited, Taiwan For internal use only / Copyright
More informationMRI QA Technologist s Tests
AAPM 55th Annual Meeting MRI QA Technologist s Tests Guang Jia, PhD Associate Professor Department of Physics Louisiana State University Background Outline QC importance Technologist s role QC test tips
More informationDevelopment of SCENARIA New Version Software MEDIX VOL. 62 P.32 P.35
Development of SCENAIA New Version Software Takayuki Kadomura Toshio Sakamoto Kana Tanaka Takatsugu Ito Yuuko Nishimura Naomi Maekawa MEDIX VOL. 62 P.32 P.35 Development of SCENAIA New Version Software
More informationAssessment of calcium scoring performance in cardiac computed tomography
Eur Radiol (2003) 13:484 497 DOI 10.1007/s00330-002-1746-y CARDIAC Stefan Ulzheimer Willi A. Kalender Assessment of calcium scoring performance in cardiac computed tomography Received: 19 June 2002 Revised:
More informationMYOCARDIAL PERFUSION COMPUTED TOMOGRAPHY PhD course in Medical Imaging. Anne Günther Department of Radiology OUS Rikshospitalet
MYOCARDIAL PERFUSION COMPUTED TOMOGRAPHY PhD course in Medical Imaging Anne Günther Department of Radiology OUS Rikshospitalet CORONARY CT ANGIOGRAPHY (CTA) Accurate method in the assessment of possible
More informationWHERE IN THE WORLD JILL LIPOTI?
WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain
More informationThe Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO
The Whys, Hows and Whats of the Noise Power Spectrum Helge Pettersen, Haukeland University Hospital, NO Introduction to the Noise Power Spectrum Before diving into NPS curves, we need Fourier transforms
More informationMultislice CT: Current Technology and Future Developments
Multislice CT: Current Technology and Future Developments 1 Stefan Ulzheimer and Thomas Flohr Contents 1.1 Introduction 3 1.2 System Design 5 1.2.1 Gantry 6 1.2.2 X-Ray Tube and Generator 6 1.2.3 MDCT
More informationREGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
More informationM D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
More informationImage Analysis for Volumetric Industrial Inspection and Interaction
Image Analysis for Volumetric Industrial Inspection and Interaction Rasmus Larsen Mads F. Hansen, Hildur Olafsdottir, Thomas H. Mosbech, Lasse F. Laursen DTU Informatics (CT) reconstruction from few projections
More informationQuantitative Imaging In Clinical Trials Using PET/CT: Update
Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative
More informationRb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals
Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction
More informationChest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80
Revisions Effective January 2012 Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Technologist Instructions Patient must cough several times prior to scan to clear
More informationCardiac CT for Calcium Scoring
Scan for mobile link. Cardiac CT for Calcium Scoring Cardiac computed tomography (CT) for Calcium Scoring uses special x-ray equipment to produce pictures of the coronary arteries to determine if they
More informationQuality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
More informationTISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
More informationPET/CT-MRI First clinical experience
20 th April 2013, Barcelona, Sp PET/CT-MRI First clinical experience Philippe Appenzeller, MD Staff Radiologist and Nuclear Medicine Physician Department Medical Imaging, University Hospital Zurich PET/CT-MR
More informationGE Healthcare. pet/ct for simulation. Precision in motion.
GE Healthcare pet/ct for simulation. Precision in motion. Patient profile H: 5 10 W: 370 lbs BMI: 53 Each cancer patient is just as unique as his or her cancer. Approaches to cancer treatment are changing
More informationAutomated EMR Dose History Extraction and Monitoring
Automated EMR Dose History Extraction and Monitoring Aaron Sodickson MD, PhD Section Chief, Emergency Radiology Medical Director of CT, Brigham Radiology Network Brigham and Women s Hospital Harvard Medical
More informationR/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
More informationVoxar 3D TM. A suite of advanced visualization and analysis software tools
Voxar 3D TM A suite of advanced visualization and analysis software tools The power to deliver advanced visualization throughout the enterprise To effectively manage the rapid growth of large volumetric
More informationSITE IMAGING MANUAL ACRIN 6698
SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012
More informationSUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
More informationCT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
More informationTHE DOSIMETRIC EFFECTS OF
THE DOSIMETRIC EFFECTS OF OPTIMIZATION TECHNIQUES IN IMRT/IGRT PLANNING 1 Multiple PTV Head and Neck Treatment Planning, Contouring, Data, Tips, Optimization Techniques, and algorithms AAMD 2013, San Antonio,
More informationThe AAPM does not endorse any products, manufacturers, or suppliers. Nothing in this publication should be interpreted as implying such endorsement.
DISCLAIMER: This publication is based on sources and information believed to be reliable, but the AAPM, the editors, and the publisher disclaim any warranty or liability based on or relating to the contents
More information4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.
4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology jiang.steve@mgh.harvard.edu http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?
More informationThe effect of pitch in multislice spiral/helical CT
The effect of pitch in multislice spiral/helical CT Ge Wang a) and Michael W. Vannier Department of Radiology, University of Iowa, Iowa City, Iowa 52242 Received 20 November 1998; accepted for publication
More informationComputed Tomography Radiation Safety Issues in Ontario
Computed Tomography Radiation Safety Issues in Ontario Healthcare Human Factors Group Centre for Global ehealth Innovation University Health Network Toronto, ON, Canada June 16, 2006 Table of Contents
More informationCorrecting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed
More informationTemporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!
Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board
More informationTechniques and Applications of Automatic Tube Current Modulation for CT 1
Review Radiology Mannudeep K. Kalra, MD, DNB Michael M. Maher, MD, FFR (RCSI), FRCR Thomas L. Toth, DSc Bernhard Schmidt, PhD Bryan L. Westerman, PhD Hugh T. Morgan, PhD Sanjay Saini, MD Index terms: Computed
More informationPractical exercise: Effective dose estimate in CT
Practical exercise: Effective dose estimate in CT TRAINING COURCE PROGRAM 19 20 May 2011, Sofia, Bulgaria Virginia Tsapaki Medical Physics Dpt Konstantopoulio General Hospital email: virginia@otenet.gr
More informationSurveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,
More informationConcepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
More informationAcknowledgements. PMH, Toronto David Jaffray Doug Moseley Jeffrey Siewerdsen. Beaumont Hospital Di Yan Alvaro Martinez. Elekta Synergy Research Group
Cone Beam CT guided Radiotherapy Jan-Jakob Sonke Acknowledgements NKI-AVL: Marcel van Herk, Jose Belderbos, Suzanne van Beek, Anja Betgen, Josien de Bois, ~Rianne de Jong, Michel Frenay, Danny Minkema,
More informationDigital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
More informationAcknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):
Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 47 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL Refresher Course
More information3 CT Parameters that Influence the Radiation Dose
1 3 CT Parameters that Influence the Radiation Dose Hans Dieter Nagel, PhD Philips Medical Systems, Science and Technology, Roentgenstr. 24, D-22335 Hamburg, Germany hans-dieter.nagel@philips.com The radiation
More informationBasics of Clinical X-Ray Computed Tomography
Siemens 3=18-slice dual source cone-beam spiral CT (005) Basics of Clinical X-Ra Computed Tomograph EMI parallel beam scanner (197) Marc Kachelrieß Institute of Medical Phsics Universit of Erlangen-Nürnberg
More informationOverview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota
Creation of 3D printed phantoms for clinical radiation therapy Eric Ehler, PhD Assistant Professor University of Minnesota ehler 046@umn.edu Overview Background of 3D Printing Practical Information Current
More informationOptimization of image quality through online motion correction
Advanced imaging technologies a revolution in diagnosis and treatment Optimization of image quality through online motion correction Oline Vinter Olesen December 10, 2010 Medical Solutions, Siemens Institute
More informationThe Challenge of CT Dose Records
The Challenge of CT Dose Records Kimberly Applegate, MD, MS, FACR Professor of Radiology and Pediatrics Emory University *financial disclosures: -Springer Textbook contracts -AIM advisory board for patient
More informationClinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.
www.neurologica.com Critical Care Clinic Operating Room Radiology ED Trauma OR Neuro/Spine Pediatric Plastics Thoracic General Interventional Radiology ENT ED Trauma Trauma Stroke Critical Care Neuro Pediatric
More informationGUIDE TO SETTING UP AN MRI RESEARCH PROJECT
GUIDE TO SETTING UP AN MRI RESEARCH PROJECT Formal requirements and procedures OVERVIEW This document is intended to help a principle investigator set up a research project using magnetic resonance imaging
More informationCorporate Medical Policy
Corporate Medical Policy Computed Tomography to Detect Coronary Artery Calcification File Name: computed_tomography_to_detect_coronary_artery_calcification Origination: 3/1994 Last CAP Review 11/2014 Next
More informationMulti-slice CT Technology 3
41 Multi-slice CT Technology 3 Thomas Flohr and Bernd Ohnesorge Contents 3.1 Evolution from 1 to 64 Slices 41 3.2 Principles of Multi-slice CT System Design 45 3.3 Multi-slice CT Acquisition and Reconstruction
More informationLow-dose CT Imaging. Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital
Lung Cancer Screening with Low-dose CT Imaging Edgar Fearnow, M.D. Section Chief, Computed Tomography, Lancaster General Hospital Despite recent declines in the incidence of lung cancer related to the
More information3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
More informationMODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada
More information