Transmission electron microscopy - Wikipedia, the free encyclopedia

Similar documents
Physics 441/2: Transmission Electron Microscope

Preface Light Microscopy X-ray Diffraction Methods

Scanning Electron Microscopy: an overview on application and perspective

Theremino System Theremino Spectrometer Technology

Science In Action 8 Unit C - Light and Optical Systems. 1.1 The Challenge of light

Introduction to microstructure

View of ΣIGMA TM (Ref. 1)

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Chapter 4. Microscopy, Staining, and Classification. Lecture prepared by Mindy Miller-Kittrell North Carolina State University

Nanoelectronics 09. Atsufumi Hirohata Department of Electronics. Quick Review over the Last Lecture

Lenses and Apertures of A TEM

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

Ion Beam Sputtering: Practical Applications to Electron Microscopy

Chapter 18: The Structure of the Atom

Copyright by Mark Brandt, Ph.D. 12

The photoionization detector (PID) utilizes ultraviolet

PHOTOELECTRIC EFFECT AND DUAL NATURE OF MATTER AND RADIATIONS

From lowest energy to highest energy, which of the following correctly orders the different categories of electromagnetic radiation?

Near-field scanning optical microscopy (SNOM)

EDS system. CRF Oxford Instruments INCA CRF EDAX Genesis EVEX- NanoAnalysis Table top system

Electron Microscopy SEM and TEM

How To Understand Light And Color

Experiment: Crystal Structure Analysis in Engineering Materials

Electron Microscopy 3. SEM. Image formation, detection, resolution, signal to noise ratio, interaction volume, contrasts

Nanometer-scale imaging and metrology, nano-fabrication with the Orion Helium Ion Microscope

History of the Atom & Atomic Theory

CSCI 4974 / 6974 Hardware Reverse Engineering. Lecture 8: Microscopy and Imaging

EXPERIMENT #1: MICROSCOPY

Usage of Carbon Nanotubes in Scanning Probe Microscopes as Probe. Keywords: Carbon Nanotube, Scanning Probe Microscope

THE NATURE OF LIGHT AND COLOR

Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES

Study Guide for Exam on Light

Coating Technology: Evaporation Vs Sputtering

Reflection Electron Microscopy and Spectroscopy for Surface Analysis

Forensic Science: The Basics. Microscopy

Vacuum Evaporation Recap

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

Polarization Dependence in X-ray Spectroscopy and Scattering. S P Collins et al Diamond Light Source UK

X-ray thin-film measurement techniques

Defects Introduction. Bonding + Structure + Defects. Properties

waves rays Consider rays of light from an object being reflected by a plane mirror (the rays are diverging): mirror object

2) A convex lens is known as a diverging lens and a concave lens is known as a converging lens. Answer: FALSE Diff: 1 Var: 1 Page Ref: Sec.

Raman spectroscopy Lecture

Displays. Cathode Ray Tube. Semiconductor Elements. Basic applications. Oscilloscope TV Old monitors. 2009, Associate Professor PhD. T.

Demonstration of sub-4 nm nanoimprint lithography using a template fabricated by helium ion beam lithography

X-Ray Diffraction HOW IT WORKS WHAT IT CAN AND WHAT IT CANNOT TELL US. Hanno zur Loye

The Focused Ion Beam Scanning Electron Microscope: A tool for sample preparation, two and three dimensional imaging. Jacob R.

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

Technology White Papers nr. 13 Paul Holister Cristina Román Vas Tim Harper

Physics 30 Worksheet # 14: Michelson Experiment

TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 Instructions

PHYSICS PAPER 1 (THEORY)

Problem Set 6 UV-Vis Absorption Spectroscopy Express the following absorbances in terms of percent transmittance:

Conductivity of silicon can be changed several orders of magnitude by introducing impurity atoms in silicon crystal lattice.

First let us consider microscopes. Human eyes are sensitive to radiation having wavelengths between

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

It has long been a goal to achieve higher spatial resolution in optical imaging and

7. advanced SEM. Latest generation of SEM SEM

Light as a Wave. The Nature of Light. EM Radiation Spectrum. EM Radiation Spectrum. Electromagnetic Radiation

Looking through the fish-eye the Electron Ronchigram. Duncan T.L. Alexander CIME seminar May 24, 2012

Cathode Ray Tube. Introduction. Functional principle

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Basic principles and mechanisms of NSOM; Different scanning modes and systems of NSOM; General applications and advantages of NSOM.

ILLUSTRATIVE EXAMPLE: Given: A = 3 and B = 4 if we now want the value of C=? C = = = 25 or 2

Electromagnetic Radiation (EMR) and Remote Sensing

Using the Spectrophotometer

Crystal Optics of Visible Light

8.1 Radio Emission from Solar System objects

Automatic and Objective Measurement of Residual Stress and Cord in Glass

SILA Sistema Integrato di Laboratori per l Ambiente. CENTRE FOR MICROSCOPY AND MICROANALYSIS Scientific coordinator: Prof.ssa Rosanna De Rosa

Scanning Electron Microscopy tools for material characterization

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

Usage of AFM, SEM and TEM for the research of carbon nanotubes

Physical Science Study Guide Unit 7 Wave properties and behaviors, electromagnetic spectrum, Doppler Effect

Acousto-optic modulator

Phase Characterization of TiO 2 Powder by XRD and TEM

Atomic Force Microscopy. Long Phan Nanotechnology Summer Series May 15, 2013

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

UNIT I: INTRFERENCE & DIFFRACTION Div. B Div. D Div. F INTRFERENCE

Tuesday 20 May 2014 Morning

Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm?

Microscope Lab Introduction to the Microscope Lab Activity

CREOL, College of Optics & Photonics, University of Central Florida

Diffraction Course Series 2015

Atomic Calculations. 2.1 Composition of the Atom. number of protons + number of neutrons = mass number

Holographically corrected microscope with a large working distance (as appears in Applied Optics, Vol. 37, No. 10, , 1 April 1998)

After a wave passes through a medium, how does the position of that medium compare to its original position?

Keywords: Planar waveguides, sol-gel technology, transmission electron microscopy

The Basics of Scanning Electron Microscopy

Optical Microscope; Scanning Electron Microscope (SEM); Transmission Electron Microscope (TEM);

Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.

The Wide Field Cassegrain: Exploring Solution Space

3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY

DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND

Microimaging. Abstract. 1 A view into the invisible world. 2 Rays and lenses. Imaging bees, polen and atoms March 23, 2006, Thomas Wendler

Advancements in High Frequency, High Resolution Acoustic Micro Imaging for Thin Silicon Applications

Evaluation of Devices and Materials by Transmission Electron Microscopy

h e l p s y o u C O N T R O L

COMPARISON OF TEXTURE IN COPPER AND ALUMINUM THIN FILMS DETERMINED BY XRD AND EBSD *

Transcription:

1 of 5 10/24/2006 12:55 PM Transmission electron microscopy From Wikipedia, the free encyclopedia (Redirected from Transmission electron microscope) It has been suggested that Selected area diffraction be merged into this article or section. (Discuss) Transmission electron microscopy (TEM) is an imaging technique whereby a beam of electrons is focused onto a specimen causing an enlarged version to appear on a fluorescent screen or layer of photographic film (see electron microscope), or to be detected by a CCD camera. The first practical transmission electron microscope was built by Albert Prebus and James Hillier at the University of Toronto in 1938 using concepts developed earlier by Max Knoll and Ernst Ruska. In the past, light microscopes have been used mostly for imaging due to their relative ease of use. However, the maximum resolution that one can image is determined by the wavelength of the photons that are being used to probe the sample. In the early days of microscopy nothing smaller than the wavelength being used could be resolved, whereas nowadays the law of RESOLFT sets the limit for optical microscopes employing such concepts (see microscope). Visible light has wavelengths of 400 700 nanometers; larger than many objects of interest. Ultraviolet could be used, but soon runs into problems of absorption. Even shorter wavelengths, such as X-rays, exhibit a lack of interaction: both in focusing (nothing interacts strongly enough to act as a lens) and actually interacting with the sample. Bacterial cells of Staphylococcus aureus captured by a transmission electron microscope, magnified 50000x. Like all matter, electrons have both wave and particle properties (as theorized by Louis-Victor de Broglie), and their wave-like properties mean that a beam of electrons can in some circumstances be made to behave like a beam of radiation. The wavelength is dependent on their energy, and so can be tuned by adjustment of accelerating fields, and can be much smaller than that of light, yet they can still interact with the sample due to their electrical charge. Electrons are generated by a process known as thermionic discharge in the same manner as the cathode in a cathode ray tube, or by field emission; they are then accelerated by an electric field and focused by electrical and magnetic fields onto the sample. The electrons can be focused onto the sample providing a resolution far better than is possible with light microscopes, and with improved depth of vision. Details of a sample can be enhanced in light microscopy by the use of stains; similarly with electron microscopy, compounds of heavy metals such as lead or uranium can be used to selectively deposit heavy atoms in the sample and enhance structural detail, the dense electron clouds of the heavy atoms interacting strongly with the electron beam. The electrons can be detected using a photographic film, or fluorescent screen among other technologies. An additional class of these instruments is the electron cryomicroscope, which includes a specimen stage capable of maintaining the specimen at liquid nitrogen or liquid helium temperatures. This allows imaging specimens prepared in vitreous ice, the preferred preparation technique for imaging individual molecules or macromolecular assemblies. Another type of TEM is the scanning transmission electron microscope (STEM), where the beam can be rastered across the sample to form the image.

2 of 5 10/24/2006 12:55 PM In analytical TEMs the elemental composition of the specimen can be determined by analysing its X-ray spectrum or the energy-loss spectrum of the transmitted electrons. Modern research TEMs may include aberration correctors, to reduce the amount of distortion in the image, allowing information on features on the scale of 0.1 nm to be obtained (resolutions down to 0.08 nm have been demonstrated, so far). Monochromators may also be used which reduce the energy spread of the incident electron beam to less than 0.15 ev. Contents 1 Applications of the TEM 2 Imaging in the TEM 3 Limitations 4 See also 5 External links Applications of the TEM The TEM is used heavily in both material science/metallurgy and the biological sciences. In both cases the specimens must be very thin and able to withstand the high vacuum present inside the instrument. For biological specimens, the maximum specimen thickness is roughly 1 micrometre. To withstand the instrument vacuum, biological specimens are typically held at liquid nitrogen temperatures after embedding in vitreous ice, or fixated using a negative staining material such as uranyl acetate or by plastic embedding. Typical biological applications include tomographic reconstructions of small cells or thin sections of larger cells and 3-D reconstructions of individual molecules via Single Particle Reconstruction. In material science/metallurgy the specimens tend to be naturally resistant to vacuum, but must be prepared as a thin foil, or etched so some portion of the specimen is thin enough for the beam to penetrate. Preparation techniques to obtain an electron transparent region include ion beam milling and wedge polishing. The focused ion beam (FIB) is a relatively new technique to prepare thin samples for TEM examination from larger specimens. Because the FIB can be used to micro-machine samples very precisely, it is possible to mill very thin membranes from a specific area of a sample, such as a semiconductor or metal. Furthermore, a less time consuming and far more common technique is the deposition of a dilute sample containing the specimen on copper or nickel grids just to name a few. These grids consist of thin films of the respective elements which rest atop sturdier grid bars and act as relatively transparent supports for specimens. Samples are normally desposited as a suspension in a volatile solvent such as ethanol.

3 of 5 10/24/2006 12:55 PM The imaging techniques explained below are particularly important in materials science. Faults in crystals affect both the mechanical and the electronic properties of materials, so understanding how they behave gives a powerful insight. By carefully selecting the orientation of the sample, it is possible not just to determine the position of defects but also to determine the type of defect present. If the sample is orientated so that one particular plane is only slightly tilted away from the strongest diffracting angle (known as the Bragg Angle), any distortion of the crystal plane that locally tilts the plane to the Bragg angle will produce particularly strong contrast variations. However, defects that produce only displacement of atoms that do not tilt the crystal to the Bragg angle (i.e. displacements parallel to the crystal plane) will not produce strong contrast. SEM image of a thin TEM sample milled by FIB. The thin membrane is suitable for TEM examination; however, at ~300 nm thick, it would not be suitable for high-resolution TEM without further milling. SEM image of a thin TEM sample milled by FIB. The thin membrane is suitable for TEM examination; however, at ~300 nm thick, it would not be Furthermore, the HRTEM technique (see below) allows the direct observation of suitable for high-resolution TEM crystal structure and therefore has an advantage over other methods e.g. there is no without further milling. displacement between the location of a defect and the contrast variation caused in the image. However, it is not always possible to interpret the lattice images directly in terms of sample structure or composition. This is because the image is sensitive to a number of factors (specimen thickness and orientation, objective lens defocus, spherical and chromatic aberration), and although quantitative interpretation of the contrast shown in lattice images is possible, it is inherently complicated and may require extensive simulation of the images. Computer modeling of these images has added a new layer of understanding to the study of crystalline materials. Imaging in the TEM The contrast in a TEM image is not like the contrast in a light microscope image. A crystalline material interacts with the electron beam mostly by diffraction rather than absorption, although the intensity of the transmitted beam is still affected by the volume and density of the material through which it passes. The intensity of the diffraction depends on the orientation of the planes of atoms in a crystal relative to the electron beam at certain angles the electron beam is diffracted strongly, sending electrons away from the axis of the incoming beam, while at other angles the beam is largely transmitted. Modern TEMs are often equipped with specimen holders that allow the user to tilt the specimen to a range of angles in order to obtain specific diffraction conditions, and apertures placed below the specimen allow the user to select electrons diffracted in a particular direction. A high contrast image can therefore be formed by blocking electrons deflected away from the optical axis of the microscope by placing the aperture to allow only unscattered electrons through. This produces a variation in the electron intensity that reveals information on the crystal structure, and can be viewed on a fluorescent screen, or recorded on photographic film or captured electronically. This technique (known as Bright Field or Light Field) is particularly sensitive to extended crystal lattice defects in an otherwise ordered crystal. As the local distortion of the crystal around the defect changes the angle of the crystal plane, the intensity of the scattering will vary around the defect. As the image is formed by the distortion of the crystal planes around the defect, the contrast in these images does not normally coincide exactly with the defect, but is slightly to one side. It is also possible to produce an image from electrons deflected by a particular crystal plane. By either moving the aperture to the position of the deflected electrons, or tilting the electron beam so that the deflected electrons pass through

4 of 5 10/24/2006 12:55 PM the centred aperture, an image can be formed of only deflected electrons, known as a Dark Field image. In the most powerful diffraction contrast TEM instruments, crystal structure can also be investigated by High Resolution Transmission Electron Microscopy (HRTEM), also known as phase contrast imaging as the images are formed due to differences in phase of electron waves scattered through a thin specimen. Resolution of the HRTEM is limited by spherical and chromatic aberration, but a new generation of aberration correctors has been able to overcome spherical aberration. Software correction of spherical aberration has allowed the production of images with sufficient resolution to show carbon atoms in diamond separated by only 0.89 Ångström units and atoms in silicon at 0.78 Ångströms (one Ångström is 0.000 000 000 1 of a meter) at magnifications of 50 million times. Improved resolution has also allowed the imaging of lighter atoms that scatter electrons less efficiently lithium atoms have been imaged in lithium battery materials. The ability to determine the positions of atoms within materials has made the HRTEM an indispensable tool for nanotechnology research and development in many fields, including heterogeneous catalysis and the development of semiconductor devices for electronics and photonics. Limitations There are a number of drawbacks to the TEM technique. Many materials require extensive sample preparation to produce a sample thin enough to be electron transparent, which makes TEM analysis a relatively time consuming process with a low throughput of samples. The structure of the sample may also be changed during the preparation process. Also the field of view is relatively small, raising the possibility that the region analysed may not be characteristic of the whole sample. There is potential that the sample may be damaged by the electron beam, particularly in the case of biological materials. See also Scanning electron microscope Electron microscope Transmission Electron Aberration-corrected Microscope Energy filtered transmission electron microscopy Wikibooks has more about this subject: Nanowiki External links The National Center for Electron Microscopy, Berkeley California USA (http://ncem.lbl.gov) The National Center for Macromolecular Imaging, Houston Texas USA (http://ncmi.bcm.tmc.edu) http://nsm1.fullerton.edu/~skarl/em/microscopy/temmicroscopy.html http://www.unl.edu/cmracfem/em.htm http://www.matter.org.uk/tem/ http://www.microscopy.ethz.ch Nanotechnology, Institute of Physics Publishing (http://www.nanotechnology.iop.org) Retrieved from "http://en.wikipedia.org/wiki/transmission_electron_microscopy" Categories: Articles to be merged since October 2006 Microscopes This page was last modified 03:06, 9 October 2006.

5 of 5 10/24/2006 12:55 PM All text is available under the terms of the GNU Free Documentation License. (See Copyrights for details.) Wikipedia is a registered trademark of the Wikimedia Foundation, Inc.