Errors Due to Shared Leadwires in Parallel Strain Gage Circuits



Similar documents
THE STRAIN GAGE PRESSURE TRANSDUCER

Calibration with bridge measurements

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples

Environmental Monitoring with Sensors: Hands-on Exercise

Basic RTD Measurements. Basics of Resistance Temperature Detectors

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

Application Note. So You Need to Measure Some Inductors?

Experiment 5. Strain Gage Measurements

Applying the Wheatstone Bridge Circuit

DIRECT CURRENT GENERATORS

Errors Related to Cable Resistance Imbalance in Three Wire RTDs

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

PHYSICS 111 LABORATORY Experiment #3 Current, Voltage and Resistance in Series and Parallel Circuits

9. Force Measurement with a Strain gage Bridge

Physics 221 Experiment 5: Magnetic Fields

Objectives 200 CHAPTER 4 RESISTANCE

A Short Discussion on Summing Busses and Summing Amplifiers By Fred Forssell Copyright 2001, by Forssell Technologies All Rights Reserved

Experiment #3, Ohm s Law

Physics 3330 Experiment #2 Fall DC techniques, dividers, and bridges R 2 =(1-S)R P R 1 =SR P. R P =10kΩ 10-turn pot.

VOLTAGE REGULATOR AND PARALLEL OPERATION

Generator Stator Protection, under/over voltage, under /over frequency and unbalanced loading. Ramandeep Kaur Aujla S.NO

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

AND8008/D. Solid State Control Solutions for Three Phase 1 HP Motor APPLICATION NOTE

Pressure Transducer to ADC Application

BJT AC Analysis. by Kenneth A. Kuhn Oct. 20, 2001, rev Aug. 31, 2008

Force measurement. Forces VECTORIAL ISSUES ACTION ET RÉACTION ISOSTATISM

Series and Parallel Resistive Circuits

Subminiature Load Cell Model 8417

S-Parameters and Related Quantities Sam Wetterlin 10/20/09

V out. Figure 1: A voltage divider on the left, and potentiometer on the right.

How To Calculate The Power Gain Of An Opamp

DC GENERATOR THEORY. LIST the three conditions necessary to induce a voltage into a conductor.

The Do s and Don ts of Pressure Transducers

Common Mode Choke Filtering Improves CMRR in Ethernet Transformer Applications. Application Note. June 2011

Procon Engineering. Technical Document PELR TERMS and DEFINITIONS

The basic cascode amplifier consists of an input common-emitter (CE) configuration driving an output common-base (CB), as shown above.

Digital Energy ITI. Instrument Transformer Basic Technical Information and Application

WHAT DESIGNERS SHOULD KNOW ABOUT DATA CONVERTER DRIFT

TECH TIP # 37 SOLVING SERIES/PARALLEL CIRCUITS THREE LAWS --- SERIES CIRCUITS LAW # THE SAME CURRENT FLOWS THROUGH ALL PARTS OF THE CIRCUIT

Low Voltage, Resistor Programmable Thermostatic Switch AD22105

Application Note 142 August New Linear Regulators Solve Old Problems AN142-1

Lab 14: 3-phase alternator.

Basic Electronics Prof. Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati

Using Current Transformers with the 78M661x

13 ELECTRIC MOTORS Basic Relations

Digital Systems Ribbon Cables I CMPE 650. Ribbon Cables A ribbon cable is any cable having multiple conductors bound together in a flat, wide strip.

Reading assignment: All students should read the Appendix about using oscilloscopes.

Figure 1. Diode circuit model

Resistor Theory and Technology

School of Engineering Department of Electrical and Computer Engineering

Experiment 4: Sensor Bridge Circuits (tbc 1/11/2007, revised 2/20/2007, 2/28/2007) I. Introduction. From Voltage Dividers to Wheatstone Bridges

FREQUENCY CONTROLLED AC MOTOR DRIVE

Force on Moving Charges in a Magnetic Field

ADC-20/ADC-24 Terminal Board. User Guide DO117-5

Analog Devices Welcomes Hittite Microwave Corporation NO CONTENT ON THE ATTACHED DOCUMENT HAS CHANGED

International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online

Chapter 11. Inductors ISU EE. C.Y. Lee

Experiment #5, Series and Parallel Circuits, Kirchhoff s Laws

Frequency Response of Filters

SYNCHRONOUS MACHINES

5. Measurement of a magnetic field

Balanced vs. Unbalanced Audio Interconnections

Use and Application of Output Limiting Amplifiers (HFA1115, HFA1130, HFA1135)

Building the AMP Amplifier

Basic Electrical Technology Dr. L. Umanand Department of Electrical Engineering Indian Institute of Science, Bangalore. Lecture phase System 4

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement

New High Current MOSFET Module Offers 177 µω R DS(on)

Mounting Instructions. Torque Transducer. A en

TIMING SIGNALS, IRIG-B AND PULSES

CURRENT ELECTRICITY - I

2. A conductor of length 2m moves at 4m/s at 30 to a uniform magnetic field of 0.1T. Which one of the following gives the e.m.f. generated?

PS-6.2 Explain the factors that determine potential and kinetic energy and the transformation of one to the other.

Three phase circuits

15. MODULUS OF ELASTICITY

Torque Sensor Technical Information

Motors and Generators

A wave lab inside a coaxial cable

White Paper SolarEdge Three Phase Inverter System Design and the National Electrical Code. June 2015 Revision 1.5

CALIBRATION PROCEDURE NI Contents. Software Requirements. ni.com/manuals

3. Design Requirements

Eðlisfræði 2, vor 2007

Signal Conditioning Amplifier System

Scaling and Biasing Analog Signals

4 DEDICATED SINGLE PHASE LIMA MAC GENERATOR LIGHT TOWER DUTY MODELS 250MSL1278 AND 250MSL1299 SERVICE PROCEDURE ELECTRICAL COMPONENT TESTING

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

Impedance Matching and Matching Networks. Valentin Todorow, December, 2009

Signal Conditioning Wheatstone Resistive Bridge Sensors

Chapter 24. Three-Phase Voltage Generation

Experiment NO.3 Series and parallel connection

Line Reactors and AC Drives

STUDY MATERIAL FOR CLASS Physics- CURRENT ELECTRICITY. The flow of electric charges in a particular direction constitutes electric current.

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term

Transmission Line Transformers

BIASING OF CONSTANT CURRENT MMIC AMPLIFIERS (e.g., ERA SERIES) (AN )

ECEN 1400, Introduction to Analog and Digital Electronics

EQUATIONS and INEQUALITIES

Op amp DC error characteristics and the effect on high-precision applications

APPLICATION NOTES OF SIGNAL ISOLATORS

Basic Op Amp Circuits

Transcription:

Micro-Measurements Strain Gages and Instruments Errors Due to Shared Leadwires in Parallel Strain Gage Circuits TN-516 1. Introduction The usual, and preferred, practice with multiple quarterbridge strain gage installations used for either static or combined static/dynamic measurements is to employ a separate three-wire circuit for each gage. However, if a number of such gages are connected to a multiplechannel instrument which simultaneously uses the same power supply for several channels, the associated bridge circuits (each of which contains an active and dummy gage) are effectively in parallel. This arrangement, in itself, need not cause any problems, provided the power supply has sufficient capacity to maintain a constant voltage under varying load. If the two individual currentcarrying P+ and P power supply leadwires in each circuit have the same resistance and are subjected to the same temperature, their only contribution to measurement error is the usual desensitization of the gage factor. But this error can readily be eliminated by shunt calibration. Unfortunately, the stress analyst may be motivated under certain circumstances to use a current-carrying leadwire that, as shown in Figure 1, is shared by, or common to, all the active gages. Savings of leadwire can be realized with this arrangement (sometimes called a chevron ) when the runs between gages and instrumentation are long. And, savings in manhours of installation time will be obtained when the number of installations is large. But, considering the potential problems created by the use of a common leadwire, the only valid motivations are those arising from physical and mechanical limitations. These may include the number of slip rings available for measurements on rotating equipment; the number of passthrough conductors possible in a barrier (like the wall of a pressure vessel) between the instrument and gages; and the use of multiple-grid gages with an integral bus or solder tab (common-tab rosettes and certain types of strip gages). Figure 1. Schematic of parallel Wheatstone bridge circuits with common power supply leadwire. Document Number: 1166 Revision: 1-Nov-21 For technical support, contact 163

TN-516 Micro-Measurements A fundamental problem with the use of a common leadwire is that all data are vulnerable to degradation or even to complete loss should a single gage (or grid) malfunction. These malfunctions can range from a short circuit within a gage, to a low resistance to ground, to an open circuit. Although a primary reason for avoiding common leadwire usage, the risk of data loss is not directly relevant to the following discussion, and will not be treated further. The more dangerous aspects of common leadwire usage arise from the often subtle effects that are produced when the gages function properly. Problems that can result from these include large initial resistive imbalances of the Wheatstone bridge circuits, inaccurate shunt calibration, crosstalk between gage circuits during strain measurement, and loss of leadwire temperature compensation. These are the primary subjects of the discussion that follows. 2. The Signal from Parallel Circuits The electrical output, e o e i, from each of the active gage circuits shown schematically in Figure 1 depends upon the power supply voltage, E P, and the resistances of the common leadwire (R LC ), the active gages (R Gi ), the individual return leadwires (R Li ), and the dummy gages (R Di ). The resistances of the signal leads are relatively unimportant because no significant amount of current flows through them when a modern instrument with a high impedance input circuit is used to measure the signal voltage. The resistance of the n parallel circuits between points A and C can be expressed as: R A C= n i= 1 1 1 R + R + R Gi Li Di provided that the resistance of the leadwire between the active gages (R Gi and R Gi+1 ) is negligible. Because the common leadwire has some finite resistance, it acts as a voltage divider to reduce the excitation voltage supplied to the active and dummy gages. And because it carries the sum of the currents in all the parallel circuits, the voltage drop in the common leadwire is n times as great as for individual return leadwires with the same resistance (provided all active and dummy gages have nominally the same resistance). The fraction, H, of the power supply voltage (E P ) available to the parallel circuits between points A and C for any combination of resistances in the parallel circuits is: H E A C RA C = = EP ( RA C+ RLC ) (1) (2) where E A C is the actual bridge excitation voltage across the parallel circuits (assumed to be the same for all). The significance of this expression is that the current through the common leadwire and consequently the bridge excitation voltage at any given moment between points A and C depends upon not only the resistance of the common leadwire, the individual leadwires, and the dummy resistors, but also upon the instantaneous resistances of all the independently variable active gages in the parallel network. The effect that this phenomenon produces in the bridge output will be referred to in the following discussion as crosstalk. Applying the voltage division fraction, H, to the active half-bridge term of the usual expression for output from a Wheatstone bridge, e o e i, the signal from any active gage in the parallel circuit in Figure 1 can be calculated for any combination of resistance values: ( ) H RL + R i Di eo ei = E 1 P 2 R + R + R Gi Li Di 3. Initial Imbalance Because the common leadwire does not affect the voltage across the internal half bridge, H is not applied to the 1/2 term in Equation (3). This gives rise to the problem of an initial imbalance in every circuit, even when the active and dummy gages are of the same resistance. To illustrate the magnitude of the initial imbalance, consider the case of n parallel circuits in which all active and dummy gages are of the same resistance, R G ; and all leadwires, including the common leadwire, have the same resistance, R L. If the instrument gage factor control is set at 2., the initial imbalance, in microstrain units, is: n 1 R R I L G ( ) = ( ) 2+ n+ 1 R R L G (3) 1 (4) Equation (4) is plotted in Figure 2 (on page 165) for various combinations of the parameters n and R L /R G. As demonstrated by the figure, the imbalance can easily exceed the balance range of commercial strain indicators and signal-conditioning amplifiers. 4. Calibration Errors Shunt calibration of the individual quarter-bridge circuits to adjust the instrument sensitivity would normally be done by shunting the dummy in one circuit, under the condition of zero output from the remaining parallel circuits. The use of a common leadwire causes no errors in the actual calibration process itself. However, when subsequent strain measurements are made, the strain- 6 164 Document Number: 1166 Revision: 1-Nov-21

Micro-Measurements TN-516 Imbalance (1 ) 1 9 8 7 6 5 4 3 2 1.5.1.15.2 n = 1.25.3.35.4 n = 5 n = 3 n = 2.45.5 Incremental Output ( ) 5 1 4 9 8 3 7 6 2 5 4 3 1 2 1 n = 2 n = 1 Figure 2. Initial imbalance produced by the use of a common leadwire when all active and dummy gages are of the same resistance, common and individual leadwires are of the same resistance, and the instrument gage factor setting is 2.. induced resistance changes in the individual gages produce changes in the values of R A C, H, and ultimately E A C. Consequently, the changes in E A C will cause the bridge output to vary, even when the resistance of the active gage corresponds to the calibration value. The calibration factor between resistance change and output voltage is then no longer correct for the calibrated circuit and the indicated strains will be in error. Accordingly, the calibration factor is generally correct only for the calibration conditions; namely, when the current through the common leadwire is the same as during calibration. The calibration error produced when the parallel gages are strained does not lend itself to generalization, but is symptomatic of the crosstalk between circuits treated in the following section. 5. Measurement Crosstalk Errors Crosstalk refers to changes in both sensitivity and output produced in all parallel circuits by a resistance change in any one of the circuits. As in the case with the calibration error, this occurs because the resistance changes in each of the parallel circuits affect the voltage applied to and consequently, the output from all the other circuits. Output errors can be partially generalized to yield a cross-talk sensitivity index like that shown in Figure 3. The approximate incremental output that will occur in each of the other circuits as the result of a 1 change in any one circuit is given by this graph. Note that this incremental output is not directly related to the magnitude of any other output that may be present due to strain in the affected circuits. Rather, these incremental outputs can be thought of as output shifts that are algebraically added.1.2.3.4.5 to each of the indicated strains in the affected circuits. When the extent of the incremental outputs caused by crosstalk must be known more precisely, Equation (3) should be used to calculate the outputs of all circuits under any set of instantaneous resistances. The errors produced by crosstalk are relatively small in comparison to those that can arise from the loss of leadwire temperature compensation as described in the following section. 6. Leadwire Temperature Errors In a single, isolated, three-wire, quarter-bridge circuit, if the leadwires are of the same resistance and subjected to the same temperature change, there is essentially no false circuit output due to leadwire thermal effects. Since one of the leadwires is in the active arm of the bridge, and the other is in the dummy arm, the effects of the resistance changes cancel. The remaining error, due to a small change in leadwire desensitization of the gage factor, is normally small enough to be ignored. But this situation is entirely different when a common leadwire arrangement is employed. The single aspect of common leadwire usage with the most potential for producing serious errors is the loss of leadwire temperature compensation. The copper normally used in leadwires has a rather high thermal coefficient of.6.7.8.9.1 Figure 3. Range of nominal incremental output produced in companion parallel circuits by a strain change of 1 in any one circuit when the initial resistance of all active and dummy gages is the same, common and individual leadwires are all of the same resistance, and the circuit has been shunt calibrated at 1 across the active gages, assuming a package gage factor of 2.. Document Number: 1166 Revision: 1-Nov-21 165

TN-516 Micro-Measurements resistance. When the common leadwire and the individual leadwires in each of the parallel circuits are subjected to the same change in temperature, false circuit outputs will usually result. This phenomenon occurs because the common leadwire carries the sum of the currents carried by the individual leadwires. Except for a few unique combinations of leadwire resistances, thermally induced resistance changes in the leadwires will produce a different voltage drop in the common and individual return leadwires, thus unbalancing the bridge. If, for example, the active and dummy gages all have the same initial resistance and the resistances of the common and individual return leadwires are the same, then the voltage drop in the common leadwire will be n times as great as in the individual leadwires. Unlike the case of isolated quarter-bridge circuits, these voltage changes do not cancel one another and, as a result, leadwire temperature compensation is lost. Figure 4 shows the magnitude of the false output increases dramatically with both temperature and number of parallel circuits. These outputs are essentially independent of the strain levels to which the gages are subjected. As a result, loss of temperature compensation can lead to very large percentage errors (or, in extreme cases, even the wrong sign!) when measuring typical working strains. 7. Remedial Measures By far the best remedy for problems created by a common power-supply leadwire is to avoid using it altogether. However, for those cases in which there is no alternative to the common leadwire, several different techniques can be Incremental Output (1 ) 5. 4. 3. 2. 1..5 n = 2 T = 1 F (5.6 C) 5 F 4 F 3 F 2 F Incremental Output (1 ) 5. 4. 3. 2. 1..5 n = 3 5 F 4 F 3 F 2 F T = 1 F (5.6 C ).1.2.3.4.5.6.7.8.9.1.1.2.3.4.5.6.7.8.9.1 Incremental Output (1 ) 5. 4. 3. 2. 1..5 5 F 4 F 3 F 2 F T = 1 F (5.6 C) n = 5 Incremental Output (1 ) 5. 4. 3. 2. 1..5 5 F 4 F 3 F 2 F T = 1 F (5.6 C) n = 1.1.2.3.4.5.6.7.8.9.1 Figure 4. Output increment due to leadwire resistance change with temperature when the resistance of all dummy and active gages is the same, the common and parallel leadwires are all of the same resistance, and the circuit has been shunt calibrated at 1 across the active gages, assuming a gage factor (package data) of 2.. Note that any thermal output from the strain gages is not included..1.2.3.4.5.6.7.8.9.1 166 Document Number: 1166 Revision: 1-Nov-21

Micro-Measurements TN-516 employed to minimize the difficulties. If, for example, the strain indicator can be balanced at the test temperature, and no further temperature changes occur prior to making strain measurements, there will be, of course, no leadwire temperature errors. Although this leaves only the relatively small errors due to crosstalk between parallel circuits, there may still be a problem if the initial imbalance is beyond the balance range of the instrument in use. The latter problem can be overcome by tying the instrument ends of all dummies together at a common point and inserting a fixed precision resistor in series between points D and C in Figure 1. When all leadwires, including the common leadwire, have the same resistance, a series resistor equal to R L (n 1)/n will nullify the initial imbalance (and reduce crosstalk). TABLE 1 Gauge of Common Leadwire Parallel AWG Number Diameter Ratio Circuits Reduction (Individual : Common) 2 3 steps 1: 2 2 3 5 steps 1: 3 3 5 7 steps 1: 5 5 1 1 steps 1: 1 1 1 AWG number of the common leadwire should be reduced from the AWG number of the individual leadwires by the number of integer steps above to minimize initial imbalance and loss of leadwire temperature compensation. This assumes the leadwires are all of the same length and that all dummy and active gages are of the same nominal resistance. NOTE: This procedure for circuit compensation by wire gauge selection is based on single-conductor (solid) wires, for which there is a well-defined relationship between AWG number and the conductor cross-sectional area. Since the wire is normally drawn to a specified diameter, rather than resistance, precise compensation may require measurement of the wire resistances. The relationship between wire gauge and resistance is still more variable with stranded wire, and the resistance should always be measured to accurately establish the 1/n resistance ratio between the common and individual leadwires. In either case, any minor adjustments necessary to obtain the correct resistance for temperature compensation can be made by increasing or decreasing the length of the common leadwire as required. the desired circuit performance. Selection of the common leadwire resistance in this fashion not only solves the initial imbalance and leadwire temperature problems, but also reduces the calibration and crosstalk errors. However, it should be noted that if the resistance of the common leadwire is less than 1/n times that of the individual return leadwires, then all of the problems described here will be created in reverse. One simple method of reducing all imbalances and errors is to reduce the ratio R L /R G. This can be accomplished by using higher resistance gages when available. The many problems of a single common leadwire as outlined above can be totally eliminated by using Micro- Measurements Bridge Completion Modules to move dummy resistors to the gage site, and employing common leadwires for all parallel circuits to both positive and negative terminals of the power supply. For this technique to be effective, both common leadwires must have the same resistance and be subjected to the same changes in temperature. The relatively minor problem of leadwire desensitization remains, but this can be eliminated by shunt calibrating the active or dummy gages. Unfortunately, additional wires are required for remote shunting. 8. Summary The use of a common power-supply leadwire can introduce large errors in strain measurements, along with practical instrumentation problems such as excessive initial imbalance. Unless absolutely necessary, the common power-supply leadwire should be avoided altogether. When a common leadwire must be employed, the user needs to be aware of the error magnitudes, and should always take steps to minimize these errors through selection of the proper leadwire size and, whenever practical, through numerical correction of measured data. The most effective remedy is one which eliminates the initial imbalance while maintaining leadwire temperature compensation, even under varying test temperatures. This is accomplished by selecting a common power-supply leadwire with a resistance 1/n times that of the individual return leadwires. Table 1 gives a simple procedure for determining the appropriate wire gauge number to achieve Document Number: 1166 Revision: 1-Nov-21 167