7/22/2011. Breast MRI: Pulse Sequences, Acquisition Protocols, and Analysis. Objectives. Challenges in DCE Breast Imaging



Similar documents
7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010

MRI SEQUENCES. 1 Gradient Echo Sequence

Musculoskeletal MRI Technical Considerations

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

SITE IMAGING MANUAL ACRIN 6698

5 Factors Affecting the Signal-to-Noise Ratio

MRI for Paediatric Surgeons

GE Medical Systems Training in Partnership. Module 12: Spin Echo

Breast MRI Quality Control

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines

The File-Card-Browser View for Breast DCE-MRI Data

32-Channel Head Coil Imaging at 3T

MGH Adult Diffusion Data Scanning Protocols

Case Report: Whole-body Oncologic Imaging with syngo TimCT

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

Overview. Optimizing MRI Protocols. Image Contrast. Morphology & Physiology. User Selectable Parameters. Tissue Parameters

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact:

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

Standardized MRI Protocol for Brain Tumor Clinical Trials. Benjamin M. Ellingson, Ph.D. Assistant Professor of Radiology at UCLA

NEURO MRI PROTOCOLS TABLE OF CONTENTS

Characterization of small renal lesions: Problem solving with MRI Gary Israel, MD

Sustaining a High-Quality Breast MRI Practice

Kap 8 Image quality, signal, contrast and noise

Purchasing a cardiac CT scanner: What the radiologist needs to know

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

How To Grow With An Mri

MRI Physics for Radiologists

Proton magnetic resonance spectroscopy in the brain: Report of AAPM MR Task Group #9

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine

Electronic Supplementary Information

Spectral Adiabatic Inversion Recovery (SPAIR) MR imaging of the Abdomen

Basic Principles of Magnetic Resonance

ORIGINAL PAPER WATER FAT SEPARATION USING THE SINGLE ACQUISITION SANDWICH TYPE 3-POINT DIXON METHOD TO OPTIMIZE KNEE JOINT SCANS

THE WIDEST WIDE-BORE IN THE INDUSTRY

Generation and Detection of NMR Signals

Etude POPART'MUS MRI Component

PET/CT-MRI First clinical experience

Breast MRI: Imaging and Intervention. Jaroslaw Nicholas Tkacz, M. D.

University Children s Hospital Basel

REVIEW. Magnetic Resonance: An Introduction to Ultrashort TE (UTE) Imaging

In Practice Whole Body MR for Visualizing Metastatic Prostate Cancer

MR Imaging of Peritoneal Malignancy Russell N. Low, MD

Recommendations for cross-sectional imaging in cancer management, Second edition

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

MRI QA Technologist s Tests

In vivo stem cell tracking in the myocardium of small laboratory animals: Feasibility and acquisition strategies on clinical magnetic resonance

COST AID ASL post- processing Workshop

Diffuse infiltration in multiple myeloma treatment response assessment with "total-spine" contrast enhanced MR imaging

Glossary of MRI Terms

REVIEW. Magnetic Resonance in Medicine 00:00 00 (2014)

It takes two to connect the dots in MRI.

Master s Program in Medical Physics. Physics of Imaging Systems Basic Principles of Magnetic Resonance Imaging V. Prof. Dr.

Concepts for High-Resolution Low-Dose CT of the Breast

Clinical applications of MRI in radiation therapy. Jatta Berberat, PhD Kantonsspital Aarau

Lecture 14. Point Spread Function (PSF)

RE: Invitation for Bid (IFB) No. Q14-32 Type of Service: Magnetic Resonance Imaging System, Maintenance Agreement and Mobile MRI Unit. Addendum No.

VI. FREQUENTLY ASKED QUESTIONS CONCERNING BREAST IMAGING AUDITS

QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET

runl I IUI%I/\L Magnetic Resonance Imaging

MRI Department Goals. Effective Ways to Improve Patient Cooperation and Safety in MRI Exam. Lecture Topics

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3

MRI SCANNERS: A BUYER S GUIDE

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute

The smart approach to digital MRI

High Field MRI: Technology, Applications, Safety, and Limitations. Brief overview of high-field MRI. The promise of high-field MRI

Embedded Systems in Healthcare. Pierre America Healthcare Systems Architecture Philips Research, Eindhoven, the Netherlands November 12, 2008

Technique and Safety of. by Pierluigi Castellone, Electronics Engineer Brain Products General Manager

MRI of Bone Marrow Radiologic-Pathologic Correlation

RF Coils... They ve Come a Long, Long Way

MRI Accreditation Program Clinical Image Quality Guide

Principles of functional Magnetic Resonance Imaging

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents

Secretin Enhanced Imaging of the Pancreas

Phantom Test Guidance

MRI DATA PROCESSING. Compiled by: Nicolas F. Lori and Carlos Ferreira. Introduction

Spatial Encoding Using Multiple rf Coils: SMASH Imaging and Parallel MRI

Radiologic Science Degree Completion Program Assessment Report

The Magnetic Resonance Imaging Examination

LONI De-Identification Policy

Fundamentals of Cone-Beam CT Imaging

Magnetic Resonance Imaging

Super-Resolution Reconstruction in MRI: Better Images Faster?

FDA Guidelines for Magnetic Resonance Equipment Safety

Sports Radiology : Making a diagnosis or solving a problem

High Spatial Resolution EPI Using an Odd Number of Interleaves

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

Real Time MRI guided Focal Laser Ablation Therapy for Prostate Cancer

Lecture #7 (2D NMR) Utility of Resonance Assignments

2. MATERIALS AND METHODS

Update: MRI in Multiple sclerosis

Transcription:

Breast MRI: Pulse Sequences, Acquisition Protocols, and Analysis Objectives Ron Price Vanderbilt University Medical Center Nashville, TN 37232 1. Review background of MRI breast cancer imaging 2. Present technical challenges of breast MRI 3. Discuss typical pulse sequences 4. Describe typical image acquisition protocols 5. ACR requirements for pulse sequences and protocols 6. Discuss new approaches to image review and analysis Dynamic Contrast Enhanced (DCE) MRI of the Breast Scientific Background for Current Breast MRI Protocols Kuhl, et.al. Radiology 1999; 211:101-110 Continuous (likely benign) Kuhl, et. al. Radiology 1999; 211: 101-110 2000 Suspicious (plateau) 2005 Highly Suspicious (washout) 2007 2011 Importance of Both Lesion Enhancement and the Enhancement Pattern Kuhl, et.al. Radiology 1999; 211:101-110 Challenges in DCE Breast Imaging Degree of Enhancement Alone: Sensitivity: 91% Specificity: 37% i.e. high false positive rate Accuracy: 58% Both Enhancement and Curve Type Sensitivity: 91% Specificity: 83% Accuracy: 86% Group size: 266 cases, 101 cancers Recent study finds that false-positive rate falls following initial baseline study: 13% to 5.6% 1) Enhancing lesions result from Gd contrast agent leaking from poorly formed blood vessels within and around the malignant tumor. 2) The contrast agent shortens the T1 of the lesion relative to the surrounding normal tissues and thus may be detected as bright regions on T1-weighted images, provided there is adequate signal-to-noise (SNR). 3) The breast has significant adipose (fatty) tissues, also with short T1, thus a significant background: fat-suppression or subtraction is essential. 4) High 3D spatial resolution for small-lesion detection and shape assessment. 5) Enhancement patterns are critical to differentiation of benign and malignant masses, high temporal resolution is also essential. 6) Full simultaneous coverage of both breasts: comparison and disease extent 7) Image artifacts must be minimized: motion (cardiac and breathing), out-ofvolume wrap and non-uniform fat-suppression. Unfortunately: SNR, spatial resolution, volume coverage and imaging time all compete with one other and artifact free images may be difficult to obtain. An MR pulse sequence that can meet all of these technical requirements is a significant challenge. 1

Fat Suppression (FS) to Minimize Lesion Background Signal DCE T1-weighted images without fat suppression must utilize image subtraction to minimize background. Pre-Contrast No fat-suppressions Early Post-Contrast Later Post-Contrast T1-weighted 3DGRE without FS (pre-contrast) Pre-contrast T1-weighted 3DGRE with FS Wash-out Phase Early Post - Pre Late Post - Pre Post-contrast T1-weighted 3DGRE with FS Subtraction of Post Pre FS Approaches to Fat Suppression Frequency Selective Fat Suppression 1) Short TI inversion recovery (STIR) sequences: (Acquire at null-point of fat T1 recovery, M z = 0) 2) Sequences with frequency selective pre-pulses: (FATSat, CHEMSat, CHESS, PRESat) 3) Combined frequency selective and inversion i recovery: SPIR (Spectral Presaturation with Inversion Recovery) SPAIR (Spectral Adiabatic Inversion Recovery) 4) Phase-cycling: 3.5 ppm difference in precessional frequency (Dixon method, in-phase/out-of-phase of fat and water signal using selected echo times in gradient-echo sequences.) 5) Highly water-selective binomial RF excitation ( e.g. 1-3-3-1) RODEO* (Rotating Delivery of Excitation Off Resonance) *Harms,SE, et al Radiology 187: 493-501 (1994) 6) Subtraction of pre-contrast images from post-contrast images 1) Sequence pre-pulse at resonant frequency of fat with pulse bandwidth set for appropriate volume coverage 2) Nominal fat frequency located at 3.5 ppm below water frequency 3) Important to have homogeneous B 0 field 4) B 0 field will by affected by magnetic susceptibility of patient: Importance of good auto-shimming. For: 1.5T = (64 X 10 6 Hz) X 3.5 pp/10 6 ~ 220 Hz For: 3T ~ 440 Hz Frequency-Selective Fat Suppression Pre-pulse 1) A 90 0 pulse centered at the fat frequency re-orients the fat protons into the transverse plane, in phase. The spoiling gradients are then used to destroy (crush or scramble) the coherence of the transverse magnetization to ensure that fat does not contribute to the signal or 2) a SPAIR pre-pulse that is a 180 0 inverting pulse followed by a spoiler gradient. What spatial resolution and SNR is required? Fat-Sat pre-pulse RF excitation Basically, the answer is the best you can get and still maintain the necessary SNR and temporal sampling. RF Slice Select gradient Gradient Echo ACR established guidelines: 1) < 1.0mm X 1.0mm in-plane pixel size 2) < 3 mm slice thickness (with no slice-gap) 3) not too grainy G sl G ro Spoiler gradients Read Out G pe gradients Comparison: 1.25 X 1.25 mm pixel vs 0.6 X 0.8 pixels 1) Correctly upgraded BIRADS scores in 13 of 26 cancers 2) Correctly down-graded 10 of 28 benign lesions 2

What determines adequate temporal sampling? Enhancement Pattern for Focal Invasive Cancers (Type III Enhancement Curve) Adequate Temporal Sampling is Essential for Correct Enhancement Curve Classification Type III Type III MR Signal Time-to-peak enhancement ~ 1-3 minutes Type II? 0 1 2 3 4 5 6 7 (Minutes after contrast agent injection) 0 1 2 3 4 Temporal Sampling rate of >2 minutes (may be marginal for confident classification) 0 1 2 3 4 Temporal Sampling rate of 1 minute (accurate curve classification) Gadoliniun Contrast Agent: Rate and Volume 1) For accurate timing and consistency, power injector preferred 2) 0.1 mmol/kg (Typically, 10-20 ml volume) 3) Rate ~ 2 ml/s, w/saline flush MR Compatible Power Injectors Rapid temporal sampling with 3D pulse sequences may require parallel imaging with multi-element coils. Example of image acquisiton time with a bi-lateral 16- channel coil with parallel-imaging acceleration factor of 5.6. Note: May need to contact vendor representative for some of this information. Acquistiton time = TR X slice phase matrix X in-plane phase matrix X NSA Parallel Imaging Undersampling (Acceleration) Factors (SENSE, SMASH, GRAPPA...) Example: FOV = 250 mm, Matrix = 356 X 356 X 200 (SENSE) TE/TR/φ = 3.2 ms/6.5 ms/10 0 In-plane phase matrix = 356 (0.7mm X 0.7mm) Slice phase matrix = 200 (1.25 mm slice thickness) SPAIR (Spectrally selective Adiabatic IR) Fat-suppression Acq. Time = 0.0065 sec X 200 X 356 2.8 (phase) X 2.0 (slice) = 83 sec Image contrast is determined by when the center of k- space is acquired, one k y step each TR interval. Thus, the center of k-space (k y =0, k x =0) should be timed to coincide with the arrival of the Gd contrast agent. Center - contrast se-encoding steps phas The center of k-space occurs when the zero-strength phase-encoding gradients are applied. For 3D phase encoding is applied in two directions: slice and in-plane phase. (0,0) Sequential k-space acquisition Contrast Injection Scan Delay k-space center occurs at middle of scan Total Acquisition Time Read-out gradient Periphery - resolution Contrast Injection starts at k-space center (filled center to periphery) Mezrich R, A Perspective on k- Space: Radiology 1995; 195:297-315 Paschal CB and Morris HD: k-space in the Clinic, JMRI 19(2) 145-159(2004) Scan Delay Acquisition Time (ACR: Delay + Acquistion Time 4 minutes) Spiral or centric-elliptical (phase and slice) center-out for 3D k-space acquisition 3

RF T1 Weighted 3D Fat Suppressed Gradient Echo Sequence (sequential, spiral or 3D elliptical k space filling) Fat suppression Pre pulse (e.g. SPAIR) α Slab selective RF pulse (10 0 15 0 ) TE (1-3 ms) TR (4-6 ms) Gradient Echo Fat Suppression pulse RF Slab Select α RF pulse 3D Fat Sat Balanced Steady state Free Precession (b SSFP) Sequence (No spoiling: T1/T2 Weighted) Gradient Echo G sl SPAIR Spoiler (partition/slice) G sl (partition/slice) G ro Read Out G ro Read Out Rewinding gradients G pe In plane G pe In plane Centric-elliptic (phase and slice) Sequential (e.g. bottom-to-top) Spoiler or Crusher gradients eliminate residual transverse magnetization and mimimizes T2 weighting. Sequences will run several un recorded cycles to establish equilibrium prior to acquisition. Transverse magnetization is retained Vendor supplied 3D Fat-suppressed T1-W GE Sequences (Most 3D sequences will use centric k-space filling.) Left MIP General Electric Philips Siemens Aurora Hitachi Toshiba VIBRANT (Volume Image Breast Assessment) THRIVE (T1 High-Res Isotropic Vol Excitation) VIEWS (fl3d, 3D-FLASH) RODEO (AuroraEdge, 3DGRE) TIGRE (fast STIR, FIR) RADIANCE (FFE_3D2.5_quick) DCE pre-contrast Non-uniform fat suppression Early post-contrast Enhancing lesion Right MIP Typical Sequence Timing Parameters for T1-weighting: TE/TR/φ 1-3 ms/4-6 ms/10 0-15 0 Acquisition time: 1-3 minutes ACR: 4 minutes (sum of scan delay and single series time) T2 (fat-suppression) Subtraction: post-pre 3D Flash TE/TR/φ = 1.6 ms/4.4 ms/15 0 Slice thickness (DCE) = 1.0 mm TA = 86 sec FOV = 300X300 mm Matrix = 448 X 448 Pixel size = 0.7 X 0.7 mm Typical Breast MRI Protocol (Image acquisiton time ~15-20 minutes) ACR Breast MRI Accreditation Program Launched May 2010 1) Scout Images (~1 minute) 2) Pre-contrast (~5-7 minutes) i. T1-weighted no-fat suppression (fat/glandular morphology) ii. T2-weighted with fat suppression (bright fluid for cysts) iii. High-resolution, 3D T1-weighted fat-suppressed gradient-echo echo sequence (pre-contrast baseline image of identifying enhancing lesions) 3) Post-contrast (3-5 volume acquisitions ~ 10 minutes) Dynamic multi-phase 3D T1-weighted fat-suppressed GE sequence (Note: Pre-contrast and post-contrast images must have identical image paramaters to allow subtraction.) 4) Analysis i. Subtraction of pre-contrast and post-contrast images (identify enhancing lesions) ii. Dynamic contrast curve evaluation (enhancement pattern assessment) iii. Maximum Intensity Projection (MIP) images of subtracted images (vascular bed assessment) www.acr.org 4

ACR Accreditation Pulse Sequence Requirements www.acr.org Analysis of DCE images: CAD programs may improve consistency of breast MRI interpretation Breast MR Imaging: Computer Aided Evaluation Program for Discriminating Benign from Malignant Lesions, Williams TC, DeMartini WB, Partridge SC, Peacodk S and Lehman CD, Radiology: Vol 244 (1), pp-84-103 (2007) Acquisition time 4 minutes (delay + time to acquire a single image volume) CAD program attempt to automatically identifies lesion by: 1)Enhancement threshold 2)Persistence of enhancement 3)Initial peak enhancement Output from comercially available analysis program with automatic color-coding of wash-out curve category. Future: Kinetic Modeling for estimating Tumor extravasation rate constant: K trans Extravascular-extracellular volume fraction: V e DCE kinetic modeling for assessing breast cancer therapy response. Conclusions Pretherapy Postchemo Current imaging protocols for breast cancer assessment rely upon dynamic contrast enhanced (DCE) MRI to provide clearly detectable lesion enhancement as well as an accurate characterization of the lesion enhancement pattern. To meet these clinical i l requirements, the technical elements for breast MRI are: 1)A dedicated breast-coil array to provide high SNR images and simultaneous coverage of both breasts. 2)Fat-suppressed, T1-weighted 3D multi-phase gradient echo sequences with high in-plane spatial resolution (< 1mm X 1mm), thin slices (< 3mm) and good temporal resolution (~ 90sec) made possible by using parallel imaging. 3)Post-processing capability should provide post-contrast injection subtraction images, multi-phase time-intensity curves and maximum intensity projection (MIP) for 3D viewing and vascular maps. 5

Bi lateral Multi channel Breast Coil 6