Spatial Encoding Using Multiple rf Coils: SMASH Imaging and Parallel MRI

Size: px
Start display at page:

Download "Spatial Encoding Using Multiple rf Coils: SMASH Imaging and Parallel MRI"

Transcription

1 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI 1 Spatial Encoding Using Multiple rf Coils: SMASH Imaging and Parallel MRI Daniel K. Sodickson Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA, USA 1 INTRODUCTION The speed of MR image acquisition has increased dramatically since the 1980s through a combination of technological and methodological advances. Nevertheless, many clinical applications of MRI continue to require motion compensation in some form. In body regions containing moving structures such as the heart or diaphragm, serious artifacts can arise if scan times exceed the characteristic time scales of physiologic motion. Accurate tracking of dynamic processes, such as cardiac contraction or the arrival and uptake of intravenous contrast agents, may require high temporal resolutions without undue sacri ces in spatial resolution. Meanwhile, the fastest imaging sequences and the most state-of-the-art scanners are approaching certain basic limits of imaging speed. These limits, which have both a technological and a physiological component, are related to the maximum switching rates of magnetic eld gradients and rf pulses. Most of the fast imaging sequences now in use ± echo planar imaging (EPI), fast low angle shot (FLASH), turbo spin echo (TSE), spiral, or BURST, for example ± achieve their high speeds by optimizing the strengths, the switching rates, and the patterns of applied gradients and pulses. Beyond a certain threshold, however, rapidly switched eld gradients are known to produce neuromuscular stimulation, while excessively dense rf pulse trains can lead to unacceptable levels of rf energy deposition and heating of tissue. One common feature of fast imaging sequences is that they all acquire data in a sequential fashion. Regardless of the particular sequence the acquisition follows, the MR signal is always acquired one point and one line at a time, with each separate line of data requiring a separate application of eld gradients and/or rf pulses. Therefore, imaging speed is generally limited by the maximum switching rates compatible with scanner technology and patient safety. Recently, a new paradigm of parallel MRI has been used to increase imaging speeds beyond the basic limits just described. The term `parallel MRI' may be used to describe any MRI strategy in which multiple MR signal data points are acquired simultaneously, rather than one after the other. Parallel imaging strategies in general require the use of multiple distinct detectors, with each detector providing some component of distinct spatial information to the image. (A many-detector CCD camera is a familiar optical example of a parallel imaging device while a FAX machine is an example of a sequential line-scanning device.) For MRI in particular, some of the burden of spatial encoding traditionally accomplished by eld gradients may be shifted instead to arrays of rf coils. Recent work has shown that coil arrays may be used, in combination with appropriate image reconstruction strategies, to encode and detect multiple MR signal or image components simultaneously and thereby to multiply the speed of existing imaging sequences without increasing gradient switching rate or rf power deposition. 2 THE HISTORY OF PARALLEL MRI Radiofrequency coil arrays were rst developed for use in MRI in the late 1980s. Prior to that time, single volume or surface coils were typically used for signal detection, or else pairs of coils were arranged in quadrature to increase signal-to-noise ratio (SNR). Work by Roemer and co-workers demonstrated that arrays of suitably decoupled surface coils could be used to achieve further substantial SNR increases 1,2 (see Whole Body Machines: NMR Phased Array Coil Systems). When used in combination with traditional gradient-encoding sequences, coil arrays increased the achievable SNR for any given eld of view (FOV) in any given imaging time, but ultimate imaging speed was still governed by the imaging sequence and the gradient hardware. The fact that spatial information from coil arrays might also be used directly for the encoding and decoding of images was realized in principle relatively early in the development of MRcompatible arrays. The simplest manifestation of the parallel imaging principle was the use of spatially separated coils to image distant and nonoverlapping body regions simultaneously. More challenging was the prospect of imaging a continuous FOV rapidly using detectors overlapping in space and/or sensitivity. An early theoretical proposal by Carlson suggested that the Fourier coef cients of signal voltages in multiple coils disposed around a cylinder could be used to calculate multiple k- space lines for magnetization within that cylinder. 3 Hutchinson and Raff suggested in 1988 that a large array of narrow loops surrounding an object could, in principle, be used to acquire an image without the use of any phase-encoding gradients at all. 4 Other proposals followed, and these proposals fall into two general categories: massively parallel and partially parallel strategies. In massively parallel strategies, the number of detectors approaches the number of data points or lines. These techniques aim to replace gradient encoding entirely, with resulting dramatic improvements in imaging speed. The original massively parallel proposal of Hutchinson and Raff was intended to show theoretical feasibility, and no direct practical implementation was suggested. 4 A subsequent proposal for massively parallel imaging was made by Kwiat, Einav, and Navon in Their study included a more detailed investigation of practical issues, though no images were presented. A preliminary array design based on the general principles of this proposal was presented in In partially parallel imaging strategies, the number of detectors is signi cantly smaller than the number of data points. Spatial encoding using multiple rf coils is used to supplement the spatial encoding normally accomplished using magnetic eld gradients. Kelton, Magin, and Wright, 7 and later Ra and Rim 8,9 described a `subencoding' 9 approach by which aliased For References see p. 11

2 2 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI Table 1 Parallel MRI techniques Technique Sensitivity calibration Image reconstruction Partially parallel k-space techniques Carlson and Minemura (1993) 10 Known volume coil sensitivities k-space series expansion SMASH a (1996) 11 Surface coil sensitivity reference k-space linear combination (spatial harmonics) (phantom, in vivo, AUTO-SMASH) Image domain techniques Kelton, Magin, and Wright (1989) 7 Pixel-by-pixel sensitivity reference Pixel-by-pixel matrix inversion (subencoding) Ra and Rim (1991, 1993) 8,9 Pixel-by-pixel image reference Pixel-by-pixel matrix inversion (subencoding) SENSE a (1997) 12 Pixel-by-pixel sensitivity extraction from full reference images Massively parallel Hutchinson and Raff (1988) 4 Not discussed Inverse source Kwiat, Einav, and Navon (1991) 5 Point source references Inverse source SMASH, simultaneous acquisition of spatial harmonics; SENSE, sensitivity encoding technique. a In vivo results have been published for this technique. Pixel-by-pixel matrix inversion (subencoding) component coil images acquired rapidly with reduced phase encoding may be `unaliased' using information about component coil sensitivities. Carlson and Minemura proposed using nested volume coils with differing sensitivity patterns to approximate multiple k-space lines from a series expansion. 10 In 1997, SiMultaneous Acquisition of Spatial Harmonics (SMASH) 11 was introduced, and the rst accelerated in vivo MR images using a parallel imaging strategy were obtained. The SMASH technique uses linear combinations of component coil signals from a surface coil array to replace time-consuming gradient steps directly. Following the introduction of SMASH, the subencoding principle was revisited and re ned in the SENSitivity Encoding technique (SENSE) of PruÈssmann, Weiger, and co-workers, 12 which has also been used recently to obtain accelerated in vivo images. At the present time, various other research groups both in academia and in industry have also begun to investigate parallel imaging strategies. Table 1 lists the various proposed parallel MRI techniques, further categorized by coil sensitivity calibration and image reconstruction method. The principal message of this table is that there are numerous ways to extract spatial information from an array of rf detectors. We will now discuss some of these methods in greater detail, emphasizing approaches for which in vivo results have already been published, and which are likely to be well suited for clinical MRI. We will begin with the SMASH technique, 11 for which the broadest range of in vivo results have been achieved to date. We will then discuss the image-domain subencoding reconstruction techniques. The function of phase-encoding gradients is to impose sinusoidal modulations of magnetization across the image plane. The MR signal integrated against these sinusoids then corresponds to spatial Fourier components of the image, or the familiar k-space lines. Figure 1 illustrates schematically this well-known effect. In the gure, sinusoidal modulations of varying spatial frequency, resulting from spin evolution in phase-encoding gradients of varying strength, are shown next to their associated k-space lines. In the SMASH technique, some of these sinusoidal modulations, or `spatial harmonics', are generated by manipulations of component coil sensitivities, rather than by gradient-induced modulations of magnetization. An array of rf coils contains spatial information in the form of its component coil sensitivities (Figure 2). In a linear surface coil array with adjacent components, each coil j has a distinct but overlapping sensitivity C j (x, y). By forming appropriate linear combinations of component coil signals (Figure 3), we may generate composite sensitivity pro les C tot which oscillate in 3 SMASH IMAGING 3.1 Theory The SMASH technique exploits sensitivity variations in a surface coil array to substitute for spatial modulations normally produced by phase-encoding gradients. 11 This use of coil encoding in place of gradient encoding allows the whole of k- space to be traversed using a reduced number of phaseencoding gradient steps, thereby reducing image acquisition times. Figure 1 A k-space schematic, indicating the spatial modulations resulting from phase-encoding gradients. Gradient steps on either side of the central k = 0 line correspond to various harmonics of spin modulation across the image plane For list of General Abbreviations see end-papers

3 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI 3 Figure 2 The rf sensitivities of a linear coil array. Each of the three overlapping sensitivity distributions corresponds to one of the three component coils in the array pictured at the top. (The conductor paths of adjacent component coils are also typically overlapped, as shown in this example, to minimize inductive coupling between coils.) Only sensitivity magnitudes are shown here. In practice, coil sensitivities may also have spatially varying phase distributions much the same way as the gradient-induced modulations of Figure 1: C tot x; y ˆX n j C j x; y exp imk y y 1 j where n j are complex weight factors, m is an integer, and k y =2/FOV is the minimum k-space interval corresponding to the desired FOV. If a composite sensitivity pro le generated in this way forms an accurate spatial harmonic pattern, the same k-space step is produced as would have resulted from a traditional gradient step. 11 In other words, each combined signal S tot, generated from linear combinations of component coil signals S j using the weights n j from Equation (1), is shifted in k-space by an amount ( mk y ): S j k x ; k y ˆ d x d yc j x; y x; y expf ik x x ik y yg S tot k x ; k y ˆX n j S j k x ; k y j ˆ d x d y X j n j C j x; y x; y expf ik x x ik y yg ˆ d x d yc tot x; y x; y expf ik x x ik y yg d x d y x; y expf ik x x i k y mk y yg ~ k x ; k y mk y 2 Figure 3 Multiple spatial harmonic pro les derived by linear combination of component coil sensitivities in an eight-element array where (x, y) represents the spatial distribution of spin density in the image plane and ~ is its spatial Fourier transform. This k-space shift is precisely the same shift as would be produced by evolution of spins with gyromagnetic ratio for time t y in a y gradient of magnitude G y, with G y t y = mk y. Such a coil-encoded k-space step combines naturally with any gradient-encoded k-space steps. When a coil array with multiple elements is used, multiple harmonics may be generated from a single data set (Figure 3). The result is that a reduced number of phase-encoded lines (thick lines in Figure 4) may be acquired in a reduced acquisition time, and the remaining lines of k-space (thin lines in Figure 4) may be reconstructed using linear combinations of component coil signals. If a total of M spatial harmonics are generated, then M lines of k-space may be reconstructed for each application of a phase-encoding gradient. The full signal matrix may, therefore, be generated in a fraction 1/M of the usual acquisition time. 3.2 Implementation An important rst step in a practical SMASH implementation is to measure the rf sensitivities of the various array elements. These sensitivities may be extracted in a straightforward manner from images of homogeneous phantoms, since any intensity variations in such images may be traced to variations in coil sensitivity. Typically, intensity pro les across an image plane of interest are extracted from a stored phantom data set and are used as sensitivity references. (Often, intensity pro les along a single central line of the image plane suf ce for calibration.) For cases in which phantom sensitivity reference measurements are cumbersome or inaccurate (for example when exible coil arrays are used, the particular conformations of which for any given patient are not known a priori), sensitivity pro les may also be obtained in vivo by examining component coil images in regions of comparatively uniform spin density such as the spine. Alternative in vivo sensitivity calibration procedures, such as the self-calibrating approach For References see p. 11

4 4 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI Figure 4 Schematic k-space trajectory for a partially parallel acquisition using ve spatial harmonics. Thick lines represent k-space lines corresponding to applied phase-encoding gradient steps, while thin lines represent additional k-space lines reconstructed using the linear combinations shown in Figure 3. The additional reconstructed lines substitute for omitted phase-encoding gradient steps called AUTO-SMASH, 13 may be used for imaging in regions of heterogeneous spin density such as the thorax, where large differences in signal between the heart and lungs preclude straightforward estimation of rf sensitivity. In AUTO-SMASH, a small number of reference k-space lines are added to the acquisition, and the relation between these reference lines and the usual MR signal data lines are used to `train' SMASH reconstructions directly in k-space. Following sensitivity calibration, the sensitivity pro les are tted to the desired spatial harmonic functions, using a numerical optimization algorithm with the complex weight factors n j as tting parameters. For favorable array geometries, the resulting ts can, in practice, be quite as good as the schematic ts shown in Figure 3. The remaining steps in the SMASH reconstruction procedure are summarized in Figure 5. The left-hand side of Figure 5 shows a k-space schematic, and the right-hand side shows image data from a water phantom at each of the corresponding stages of reconstruction. With the necessary weights in hand, MR signal data are acquired simultaneously in the coils of the array. A fraction 1/M of the usual number of phase-encoding steps are applied, with M times the usual spacing in k-space (Figure 5A, left). The component coil signals acquired in this way correspond to images with a fraction 1/M of the desired FOV (Figure 5A, right). With 1/M times fewer phase-encoding steps, only a fraction 1/ M of the time usually required for this FOV is spent on data collection. Next, the appropriate M linear combinations of the component coil signals are formed to produce M shifted composite signal data sets (Figure 5B). The composite signals are then interleaved to yield the full k-space matrix (Figure 5C, left), which is Fourier transformed to give the reconstructed image (Figure 5C, right). The schematic summary in Figure 5 shows a SMASH reconstruction with acceleration factor M = 2 using a 3-element rf coil array. Substantially larger factors are possible, however, when coil arrays with larger numbers of elements are used. In fact, for favorable image plane and coil array geometries, the maximum achievable SMASH acceleration factor M is equal to the number of independent component coils in the array, since a maximum of M distinct harmonics may be generated using a total of M independent coils. Since generation of spatial harmonics does not depend upon how the gradient-encoded k-space lines were generated, the SMASH reconstruction is to a large extent sequence independent. Nearly all existing rapid imaging sequences may be accelerated in this manner, and, to date, SMASH has been successfully tested with a wide range of sequence types. Both two-dimensional and three-dimensional acquisitions are amenable to acceleration using SMASH, provided distinct coil sensitivity information is available along one or more phase-encoding directions. 3.3 In Vivo Results The improvements in imaging ef ciency afforded by a parallel imaging strategy may be put to use in a number of ways. The following examples demonstrate some of the applications for which SMASH imaging has been used to increase imaging speed and improve image quality. Reductions in breath-hold duration for scans requiring breath-holding can be achieved, which increases patient comfort and compliance and allows scans free of respiratory motion artifact in patients incapable of prolonged breath-holds. Figure 6 demonstrates a twofold reduction in breath-hold duration for abdominal MR imaging. Improvements in spatial resolution can be achieved in any given imaging time: images of increased spatial resolution may For list of General Abbreviations see end-papers

5 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI 5 Figure 5 Schematic representation of the SMASH reconstruction procedure (left: k-space cartoon, right: corresponding phantom images). (A) Acquisition of data with reduced phase encoding. (B) Formation of shifted data sets using spatial harmonic combinations. (C) Interleaving of shifted data sets to generate a full signal matrix, corresponding to an image with full FOV be generated in a given acquisition time by carrying the faster SMASH acquisition farther out in k-space. Figure 7 shows the use of SMASH for spatial resolution enhancement in a cardiac scan. Additional resolution bene ts can result from the use of SMASH or other partially parallel techniques in single-shot imaging sequences such as HASTE (half-fourier single-shot turbo spin echo), EPI, or BURST. 14 Indeed, single-shot images with both reduced acquisition time and increased spatial resolution compared with the corresponding reference images have been obtained. This is possible because a reduced aquisition time also entails reduced relaxation, and hence reduced attenuation of high spatial frequencies in single-shot sequences. 14 Improvements in temporal resolution (i.e., reductions in image-acquisition interval for gated or ungated scans) minimize undesired effects of physiologic motion while allowing accurate tracking of time-dependent phenomena. Figure 8 illustrates a twofold and a fourfold reduction in acquisition window at xed spatial resolution in cardiac MR images. These progressive increases in temporal resolution result in progressively reduced motional blurring of the right coronary artery and other cardiac structures in the SMASH images. SMASH has also been used to increase true frame rate in real-time cardiac MR scans up to and beyond current two-dimensional echocardiographic frame rates. Reductions in the overall duration of long MR scans increase patient comfort and compliance and also increase the throughput of clinical MR scanners and the cost-effectiveness of MR diagnosis. Noncontrast MR coronary angiograms obtained using SMASH are shown in Figure 9. Reduction in overall acquisition time in these navigator-gated scans has the further advantage of reducing long-term diaphragmatic drift over the course of the scans. 4 IMAGE-DOMAIN SUBENCODING TECHNIQUES 4.1 Theory A subencoding image reconstruction begins at the same starting point as a SMASH reconstruction ± namely, with a set of component coil signals acquired using a reduced number of phase-encoding gradient steps. Fourier transformation of these signal sets results in aliased component coil images like those shown in Figure 5A. From that point on, the subencoding reconstruction operates entirely in the image domain. The basis of the technique lies in the fact that each pixel in an aliased image is in fact a superposition of multiple pixels For References see p. 11

6 6 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI Figure 6 Turbo spin echo (TSE) images taken along the thoracic and lumbar spine of a healthy adult volunteer, demonstrating the use of SMASH imaging for reduction of breath-hold times in vivo (TR 700 ms, TE 32 ms, 9 echoes per excitation, in-plane spatial resolution 1.9 mm1.0 mm). These images were obtained using four active elements of a six-element circularly polarized spine array on a 1.5 T Siemens Vision imaging system. The reference image was obtained in a breath-hold lasting 22 s and reconstructed using a standard sum of squares combination of component coil images. 1 The SMASH image with the same nal spatial resolution and FOV was obtained in only 11 s from a corresponding full unaliased image (Figure 10). In other words, as a result of Nyquist aliasing, an M-times aliased image is related to the full image I full as follows: x; y ˆI full x; y I full x; y y I full x; y 2y XM 1 ˆ I full x; y my mˆ0 When is acquired using a single coil, this superposition cannot be `unfolded' without a priori knowledge of the full image. The situation changes when an array of coils is used. The full image Ij full in each coil j is actually made up of two pieces: the spin density, and the coil sensitivity function C j : 3 I full j x; y ˆC j x; y x; y 4 and in an array, each component coil j has a different sensitivity C j. Therefore, we now have multiple `views' of the aliasing that can be used to deduce just how much of each aliased pixel belongs at any position in the full image. Substituting Equation (4) into Equation (3) gives j x; y ˆXM 1 mˆ0 ˆ XM 1 mˆ0 Ij full x; y my C j x; y my x; y my For any particular aliased pixel (x,y), this may be written as follows: j ˆ XM 1 mˆ0 I full jm ˆ XM 1 mˆ0 C jm m where Ijm full I j full (x,y + my), C jm :C j (x,y + my), and m : (x,y + my). Let us study the particular example (illustrated in Figure 11) in which a four-coil array is used with a factor of three aliasing. We may write 5 6 For list of General Abbreviations see end-papers

7 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI 7 Figure 7 Coronal cardiac images showing the use of SMASH for increased spatial resolution (2D segmented-k-space FLASH, 9 echoes per segment, TR 14.4 ms, TE 7.3 ms, custom-designed four-element array, Siemens Vision scanner). The reference image has a matrix size of , corresponding to an in-plane resolution of 2.2 mm1.2 mm. The SMASH image has double the spatial resolution in both dimensions ( matrix size, in-plane resolution 1.1 mm0.6 mm). Both images were obtained in a breath-hold lasting 16 cardiac cycles. A long segment of the right coronary artery (thick black arrows) may be seen in both images but is notably sharper in the higher-resolution SMASH image. Branches of the left coronary system (thick white arrow) may also be discerned in the SMASH image, whereas they are not seen in the reference image. Finally, internal mammary arteries (thin white arrow), invisible in the reference image, may just be discerned running down the center of the SMASH image 1 ˆ C 11 1 C 12 2 C ˆ C 21 1 C 22 2 C ˆ C 31 1 C 32 2 C ˆ C 41 1 C 42 2 C 43 3 This equation may be rewritten in matrix form as or, in other words, C 11 C 12 C ˆ C 21 C 22 C C 31 C 32 C C 41 C 42 C 3 43 ˆ C As long as the number of coils N c is greater than or equal to the aliasing factor M (as in our exemplary case for which N c =4, M = 3), Equation (9) may be inverted: ˆ C and the unaliased spin density over the full FOV may be determined. This `unaliasing' approach bears some kinship with the principle of computed tomography, in the sense that multiple different `views' or projections are used to extract full twodimensional image information. 4.2 Implementation Subencoding implementations differ from SMASH implementations primarily in their approaches to sensitivity calibration and in the nature and numerical stability of their image reconstruction algorithms. Sensitivity estimates at each pixel of the full FOV are generally required to perform the pixel-by-pixel subencoding matrix inversion of Equation (10). Ra and Rim describe the use of an in vivo sensitivity reference in the form of full-fov component-coil images of the target image plane, manipulated in the reconstruction in such a way that the spin density divides out. 9 The SENSE technique incorporates a different in vivo sensitivity calibration method in which full-fov component coil images are divided by an additional full-fov body coil image, and the quotient images are then subjected to several stages of interpolation, ltering, and thresholding. 12 Phantom sensitivity references can also, in principle, be used for subencoding reconstructions. For References see p. 11

8 8 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI Figure 8 Coronal cardiac images from a 3D data set demonstrating the use of SMASH to increase temporal resolution at xed spatial resolution (3D turbo eld echo, TR 8.8 ms, TE 2.4 ms, in-plane spatial resolution 1.5 mm1.5 mm, with fat suppression, T 2 preparation, electrocardiogram triggering, and navigator gating/correction using a 6-element linear coil array on a 1.5 T Philips NT scanner). The reference image (sum of squares reconstruction) has an acquisition window of 200 ms per cardiac cycle. The twofold- and fourfold-accelerated SMASH images show progressively reduced motional blurring and improved visualization of the right coronary artery running near the midline of the images Whereas the quality of SMASH reconstructions is governed by the `goodness' of spatial harmonic ts, the principal algorithmic concern of subencoding reconstructions is the numerical stability of the inverse C 1. For N c >M, the inverse may be implemented, for example, as a Moore±Penrose pseudoinverse, resulting in a least squares solution to the overdetermined problem for each pixel. For N c = M, a standard matrix inverse may be used. The use of a pixel-by-pixel inversion affords the advantage of ne regional control over the reconstruction: it does not, for example, require a global spatial harmonic t. A disadvantage of the pixel-by-pixel approach, however, is that in regions of low actual or apparent coil sensitivity, the matrix C may be poorly conditioned, and error propagation through the inverse may amplify the effects both of noise and of sensitivity miscalibrations. 4.3 In Vivo Results Image-domain subencoding techniques can be used for many of the same applications as were demonstrated with SMASH in Section 3.3. The resulting images will resemble accelerated SMASH images, with some particular differences in image artifacts and noise distribution resulting from the different approaches to sensitivity calibration and image reconstruction. In vivo SENSE images in the brain and the heart with two- to threefold accelerations have been presented. 12,15 An in vivo sensitivity calibration method similar to that suggested by Ra and Rim has also been used by the SMASH group to obtain accelerated subencoding images for direct comparison with SMASH reconstructions of the same data sets. Figures 12 and 13 compare reference, SMASH, and subencoding images from contrast-enhanced angiography and real-time cardiac MR imaging studies. 5 A RECIPE FOR PRACTICAL PARALLEL IMAGING The following is a list of essential elements needed for effective implementations of partially parallel imaging strategies such as SMASH or subencoding. 1. rf coil array. Arrays should be designed with overall dimensions appropriate to the desired FOV for imaging applications of interest. Inductive decoupling of array elements is helpful for improved SNR and for increased spatial selectivity of the sensitivity pro les, although some degree of coupling between array elements can generally be tolerated. The imaging system must be equipped with a suf cient number of receiver channels to allow simultaneous data acquisition in the various component coils of the array. Alternative data-reception schemes such as time-domain multiplexing may also be used. 16,17 2. Sensitivity calibration. Accurate coil sensitivity calibration is crucial for all parallel imaging techniques. A great deal of effort has been devoted over the years to the understanding and calibration of gradient waveforms in MR scanners. Since parallel imaging strategies use rf coils in place of gradients for spatial encoding, it is not surprising that some For list of General Abbreviations see end-papers

9 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI 9 Figure 9 Noncontrast MR coronary angiograms obtained in reduced total scan time using SMASH (same basic sequence, coil array, and MR scanner as in Figure 8, with an acquisition window of 70 ms and an inplane resolution of 0.7 mm1.0 mm). Double oblique image planes were planned along the major axis of the right coronary artery (RCA). (A) A 7.5 cm length of the native RCA in a healthy adult volunteer is seen in this high-resolution image. The 3D data set from which this image was taken was obtained during free breathing in a total of 11 min using a SMASH acceleration factor of two. (B) The RCA is interrupted by a region of signal dropout from a coronary stent in this local maximum intensity projection of a twofold-accelerated 3D SMASH image set (11 min free-breathing acquisition) in a patient with coronary artery disease the raw MR signal data. Details of software and/or hardware implementations will depend upon which parallel reconstruction strategy is elected. Occurring as they do on either side of the Fourier transform, the k-space and the image-domain reconstructions bear a theoretical kinship, but their practical speci cations differ. The choice of reconstruction strategy may be in uenced by a number of factors: (a) Ease of implementation. SMASH uses a small number of weight factors (a minimum of one per component coil per spatial harmonic), and operates in k-space prior to Fourier transformation. Consequently, it is particularly amenable to inline implementations in which the coil-encoded k-space data are generated as soon as each gradient-encoded data point is read out. By contrast, the large numbers of weight factors in subencoding (one per component coil per image pixel) provide ner pixel-by-pixel control at the expense of increased calibration and reconstruction time. In the future, compromises between these two extremes may be anticipated, using various extrapolation schemes. (b) Image artifacts. In SMASH, residual aliasing artifacts result from errors in sensitivity calibration or spatial harmonic tting. 11 In subencoding images, localized pixel-by-pixel artifacts are more commonly seen as a result of pixel-by-pixel errors in the coil sensitivity references. (These local artifacts may be accompanied by global aliasing artifacts when there are systematic errors in sensitivity calibration.) In some circumstances, however, both techniques can actually lead to artifact reduction. Not only can accelerated acquisitions reduce motion artifacts, but they have also been shown to reduce geometrical distortions in single-shot SMASH EPI images through reduced degree of analogous effort must go into the calibration of coil sensitivities. 3. Scan planning. After a target image plane and FOV have been selected, all that is required on most MR scanners to plan a SMASH or subencoding scan is a simple FOV reduction in the phase-encode direction. The need to perform spatial encoding with the rf coil array places some limitations on the choice of image plane position and orientation. For example, phase encoding must be performed along a direction in which the sensitivities of different array elements are suf ciently distinct. Substantial angulations between the image plane and the coil array are often possible (see, for example, the double-oblique images in Figures 9 and 13), though some penalty in SNR may be incurred. 4. Image reconstruction. Signal and image processing requirements are modest, but some capabilities for variable component coil signal combinations are required, whether in the MR scanner itself or in of ine processors with access to Figure 10 Superposition of pixels (white squares) in an aliased image For References see p. 11

10 10 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI Figure 11 Superposition of pixels (white squares) in an aliased or subencoded image set from an array, corresponding to Equation (7). The inversion of this equation, and hence the unraveling of pixel-by-pixel superpositions, forms the basis of the image-domain subencoding techniques accumulation of phase discrepancies in the shortened acquisition time. (c) SNR. The SNR in a partially parallel image reconstruction is a balance between coil-speci c, reconstruction-speci c, and sequence-speci c effects. Theoretical and experimental studies of SNR in SMASH imaging have been reported. 18 Generally speaking, both SMASH and subencoding techniques are bound by the well-known limit that SNR scales as the square root of the acquisition time for any given coil and sequence. With either class of technique, then, there is some expected loss of SNR compared with optimal combinations 1 of traditionally acquired full data sets in the same array, though parallel acquisitions may still show improved SNR when compared with sequential acquisitions using single surface coils spanning the same FOV. For certain imaging sequences, furthermore, some additional SNR beyond the square root limit may be recovered through reduced relaxation in the shortened acquisition times. 6 THE FUTURE OF PARALLEL MRI Since the initial demonstration in 1997 of twofold-accelerated in vivo images using SMASH, new in vivo work has shown up to fourfold increases in acquisition speed, and as much as eightfold improvements have been achieved in phantoms using specialized rf hardware. 17 What future progress may be expected, and what are the key technical and methodological hurdles to further development? Coil array design and rf sensitivity calibration top the list of developments that will be crucial for future improvements in parallel imaging, since array geometry and sensitivity characteristics determine the degree of spatial encoding which may be accomplished in any particular region of interest. The design of tailored arrays and the availability of increased numbers of receiver channels will greatly facilitate the optimization of image quality and SNR, and the maximization of achievable acceleration factors. In-line reconstruction hardware will be of bene t for real-time imaging applications. When the necessary weight factors are known in advance, the use of onboard analog signal combiners prior to digitization may even allow large `smart' arrays to be interfaced to existing MR scanners with limited numbers of receivers. Methodological advances further in the future may also allow a return to the principle of massively parallel imaging. The potential gains of massively parallel imaging approaches are dramatic (they include the possibility of acquiring an entire image in a single echo, for example), but serious technological and theoretical hurdles remain to be overcome in areas such as electric decoupling of large arrays, massively parallel data reception, and SNR. For list of General Abbreviations see end-papers

11 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI 11 Figure 12 Maximum intensity projections of contrast-enhanced 3D MR angiograms of the abdominal aorta and renal arteries in a healthy adult volunteer (3D T 1 -weighted rf-spoiled gradient-echo imaging sequence, TE 1.5 ms, TR 7.0 ms, in-plane spatial resolution 2.7 mm1.4 mm, 6- element array, Philips NT scanner). The reference image set was obtained in a 23 s breath-hold following gadolinium injection, whereas the twofold accelerated data set used for SMASH and subencoding reconstruction was obtained in 11.5 s In the meantime, clinical implementation of partially parallel imaging is possible now using existing arrays and receiver systems. Indeed, clinical studies in selected patient populations are now being initiated using parallel acquisition techniques. Particularly for applications with stringent requirements on imaging speed, parallel imaging can be a useful tool to enhance image quality, to improve imaging ef ciency, and, in general, to overcome the acquisition speed limit in magnetic resonance imaging. 7 RELATED ARTICLES Image Formation Methods; Radiofrequency Systems and Coils for MRI and MRS; Spin Warp Data Acquisition; Surface and Other Local Coils for In Vivo Studies; Whole Body Machines: NMR Phased Array Coil Systems. 8 REFERENCES 1. P. B. Roemer, W. A. Edelstein, C. E. Hayes, S. P. Souza, and O. M. Mueller, Magn. Reson. Med., 1990, 16, C. E. Hayes and P. B. Roemer, Magn. Reson. Med., 1990, 16, J. W. Carlson, J. Magn. Reson., 1987, 74, M. Hutchinson and U. Raff, Magn. Reson. Med., 1988, 6, D. Kwiat, S. Einav, and G. Navon, Med. Phys., 1991, 18, D. Kwiat and S. Einav, Med. Eng. Phys., 1995, 17, J. R. Kelton, R. L. Magin, and S. M. Wright, Proc. VIIIth Annu. Mtg Soc. Magn. Reson. Med., Amsterdam, 1989, p J. B. Ra and C. Y. Rim, Proc. Xth Annu. Mtg Soc. Magn. Reson. Med., San Francisco, 1991, p J. B. Ra and C. Y. Rim. Magn. Reson. Med., 1993, 30, J. W. Carlson and T. Minemura, Magn. Reson. Med., 1993, 29, D. K. Sodickson and W. J. Manning, Magn. Reson. Med., 1997, 38, K. P. Pruessmann, M. Weiger, M. B. Scheidegger, and P. Boesiger, Proc. VIth Sci. Mtg (Int.) Soc. Magn. Reson. Med., Sydney, 1998, p P. M. Jakob, M. A. Griswold, R. R. Edelman, and D. K. Sodickson, MAGMA, 1998, 7, M. A. Griswold, P. M. Jakob, Q. Chen, J. W. Goldfarb, W. J. Manning, R. R. Edelman, and D. K. Sodickson, Magn. Reson. Med., 1999, 41, M. Weiger, K. P. Pruessmann, M. B. Scheidegger, and P. Boesiger, Proc. VIth Sci. Mtg (Int.) Soc. Magn. Reson. Med., Sydney, 1998, p J. R. Porter, S. M. Wright, and N. Famili, Magn. Reson. Med., 1994, 32, D. K. Sodickson, J. A. Bankson, M. A. Griswold, and S. M. Wright, Proc. VIth Sci. Mtg (Int.) Soc. Magn. Reson. Med., Sydney, 1998, p D. K. Sodickson, M. A. Griswold, P. M. Jakob, R. R. Edelman, and W. J. Manning, Magn. Reson. Med., 1999, 41, Biographical Sketch Daniel K. Sodickson. b B.Sc. (Physics) and B.A. (Humanities), 1988, Yale college, Ph.D., 1994, Massachusetts Institute of Technol- For References see p. 11

12 12 SPATIAL ENCODING USING MULTIPLE RF COILS: SMASH IMAGING AND PARALLEL MRI ogy, M.D., 1996, Harvard Medical School. Graduate work in solidstate NMR involved methodological developments for molecular structure determination and studies of the classical and quantum physics underlying spin diffusion in lattices. Appointed Research Associate, Beth Israel Deaconess Medical Center, Boston, MA, Appointed Instructor in Medicine, Harvard Medical School, Developed the SMASH imaging technique at Beth Israel Deaconess Medical Center in Approx. 10 papers and 1 patent in areas related to NMR. Primary research interests: parallel acquisition strategies in MRI, rapid MRI, cardiovascular MRI. For list of General Abbreviations see end-papers

5 Factors Affecting the Signal-to-Noise Ratio

5 Factors Affecting the Signal-to-Noise Ratio 5 Factors Affecting the Signal-to-Noise Ratio 29 5 Factors Affecting the Signal-to-Noise Ratio In the preceding chapters we have learned how an MR signal is generated and how the collected signal is processed

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider

THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI. Douglas C. Noll* and Walter Schneider THEORY, SIMULATION, AND COMPENSATION OF PHYSIOLOGICAL MOTION ARTIFACTS IN FUNCTIONAL MRI Douglas C. Noll* and Walter Schneider Departments of *Radiology, *Electrical Engineering, and Psychology University

More information

MRI SEQUENCES. 1 Gradient Echo Sequence

MRI SEQUENCES. 1 Gradient Echo Sequence 5 An MRI sequence is an ordered combination of RF and gradient pulses designed to acquire the data to form the image. In this chapter I will describe the basic gradient echo, spin echo and inversion recovery

More information

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research.

ParaVision 6. Innovation with Integrity. The Next Generation of MR Acquisition and Processing for Preclinical and Material Research. ParaVision 6 The Next Generation of MR Acquisition and Processing for Preclinical and Material Research Innovation with Integrity Preclinical MRI A new standard in Preclinical Imaging ParaVision sets a

More information

32-Channel Head Coil Imaging at 3T

32-Channel Head Coil Imaging at 3T 32-Channel Head Coil Imaging at 3T Thomas Benner Athinoula A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital and Harvard Medical School, Boston, MA, USA

More information

runl I IUI%I/\L Magnetic Resonance Imaging

runl I IUI%I/\L Magnetic Resonance Imaging runl I IUI%I/\L Magnetic Resonance Imaging SECOND EDITION Scott A. HuetteS Brain Imaging and Analysis Center, Duke University Allen W. Song Brain Imaging and Analysis Center, Duke University Gregory McCarthy

More information

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R

GE 3.0T NPW,TRF,FAST,F R NPW,TRF,FAST,F R GE 3.0T 3.0T WRIST Invivo 8CH Wrist Coil Sequence Ax T2 Cor PD Cor PDFS Cor T1 Cor PD (Small FOV) FOV (mm) 80 80 80 80 40 Matrix 384x224 384x256 320x256 384x320 320x192 Phase Direction RL RL RL RL RL #

More information

Basic Principles of Magnetic Resonance

Basic Principles of Magnetic Resonance Basic Principles of Magnetic Resonance Contents: Jorge Jovicich [email protected] I) Historical Background II) An MR experiment - Overview - Can we scan the subject? - The subject goes into the magnet -

More information

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents Preface Symbols and fundamental constants 1. What is spectroscopy? A semiclassical description of spectroscopy Damped harmonics Quantum oscillators The spectroscopic experiment Ensembles and coherence

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

Kap 8 Image quality, signal, contrast and noise

Kap 8 Image quality, signal, contrast and noise 4/5/ FYS-KJM 474 contrast SNR MR-teori og medisinsk diagnostikk Kap 8 Image qualit, signal, contrast and noise resolution vailable MRparameters speed Main source of noise in MRI: Noise generated within

More information

GE Medical Systems Training in Partnership. Module 12: Spin Echo

GE Medical Systems Training in Partnership. Module 12: Spin Echo Module : Spin Echo Spin Echo Objectives Review the SE PSD. Review the concepts of T, T, and T*. Spin Echo PSD RF Gz Gy 90 80 Gx Spin Echo - SE Spin echo is a standard pulse sequence on Signa MRi/LX and

More information

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Excelart Vantage Magnet Configuration: Ultra-short-bore Strength (or W x H): 1.5 T Homogeneity, ppm V-RMS: Dimensions of maximum

More information

MGH Adult Diffusion Data Scanning Protocols

MGH Adult Diffusion Data Scanning Protocols MGH Adult Diffusion Data Scanning Protocols Structural scans SIEMENS MAGNETOM ConnectomA syngo MR D11 \\USER\INVESTIGATORS\Default\AAHScout_64 TA:0:14 PAT:3 Voxel size:1.6 1.6 1.6 mm Rel. SNR:1.00 :fl

More information

Purchasing a cardiac CT scanner: What the radiologist needs to know

Purchasing a cardiac CT scanner: What the radiologist needs to know Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage

More information

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT

GUIDE TO SETTING UP AN MRI RESEARCH PROJECT GUIDE TO SETTING UP AN MRI RESEARCH PROJECT Formal requirements and procedures OVERVIEW This document is intended to help a principle investigator set up a research project using magnetic resonance imaging

More information

CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS

CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS CALIBRATION OF A ROBUST 2 DOF PATH MONITORING TOOL FOR INDUSTRIAL ROBOTS AND MACHINE TOOLS BASED ON PARALLEL KINEMATICS E. Batzies 1, M. Kreutzer 1, D. Leucht 2, V. Welker 2, O. Zirn 1 1 Mechatronics Research

More information

Magnetic Resonance Imaging

Magnetic Resonance Imaging Magnetic Resonance Imaging What are the uses of MRI? To begin, not only are there a variety of scanning methodologies available, but there are also a variety of MRI methodologies available which provide

More information

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications

Table 11: Pros and Cons of 1.5 T MRI vs. 3.0 T MRI; Safety and Technical Issues, and Clinical Applications Safety Issue 3.0 T MRI Pro 3.0 T MRI Con Immediate fringe field surrounding magnet A ferromagnetic object inadvertently brought into the scan room will experience a sharp increase in attraction toward

More information

Musculoskeletal MRI Technical Considerations

Musculoskeletal MRI Technical Considerations Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T

More information

How To Fix Out Of Focus And Blur Images With A Dynamic Template Matching Algorithm

How To Fix Out Of Focus And Blur Images With A Dynamic Template Matching Algorithm IJSTE - International Journal of Science Technology & Engineering Volume 1 Issue 10 April 2015 ISSN (online): 2349-784X Image Estimation Algorithm for Out of Focus and Blur Images to Retrieve the Barcode

More information

Sampling Theorem Notes. Recall: That a time sampled signal is like taking a snap shot or picture of signal periodically.

Sampling Theorem Notes. Recall: That a time sampled signal is like taking a snap shot or picture of signal periodically. Sampling Theorem We will show that a band limited signal can be reconstructed exactly from its discrete time samples. Recall: That a time sampled signal is like taking a snap shot or picture of signal

More information

A System for Capturing High Resolution Images

A System for Capturing High Resolution Images A System for Capturing High Resolution Images G.Voyatzis, G.Angelopoulos, A.Bors and I.Pitas Department of Informatics University of Thessaloniki BOX 451, 54006 Thessaloniki GREECE e-mail: [email protected]

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine

Medical Imaging. MRI Instrumentation, Data Acquisition, Image Reconstruction. Assistant Professor Department of Radiology, NYU School of Medicine G16.4426/EL5823/BE6203 Medical Imaging MRI Instrumentation, Data Acquisition, Image Reconstruction Riccardo Lattanzi, Ph.D. Assistant Professor Department of Radiology, NYU School of Medicine Department

More information

Automotive Applications of 3D Laser Scanning Introduction

Automotive Applications of 3D Laser Scanning Introduction Automotive Applications of 3D Laser Scanning Kyle Johnston, Ph.D., Metron Systems, Inc. 34935 SE Douglas Street, Suite 110, Snoqualmie, WA 98065 425-396-5577, www.metronsys.com 2002 Metron Systems, Inc

More information

Development and Application of Efficient Strategies for Parallel Magnetic Resonance Imaging

Development and Application of Efficient Strategies for Parallel Magnetic Resonance Imaging Development and Application of Efficient Strategies for Parallel Magnetic Resonance Imaging Dissertation zur Erlangung des naturwissenschaftlichen Doktorgrades der Bayerischen Julius-Maximilians-Universität

More information

Generation and Detection of NMR Signals

Generation and Detection of NMR Signals Generation and Detection of NMR Signals Hanudatta S. Atreya NMR Research Centre Indian Institute of Science NMR Spectroscopy Spin (I)=1/2h B 0 Energy 0 = B 0 Classical picture (B 0 ) Quantum Mechanical

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters

International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in

More information

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus

More information

MRI for Paediatric Surgeons

MRI for Paediatric Surgeons MRI for Paediatric Surgeons Starship David Perry Paediatric Radiologist Starship Children s Hospital CHILDREN S HEALTH What determines the brightness of a pixel in MRI? i.e. What determines the strength

More information

Etude POPART'MUS MRI Component

Etude POPART'MUS MRI Component TECHNICAL SURVEY Dear Investigators, This document is a Technical Survey which provides the teams of THERALYS and of the Pierre Wertheimer Neurological Hospital of Lyon with an overview of your site s

More information

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference.

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference. Pulsed Fourier Transform NR The rotating frame of reference The NR Eperiment. The Rotating Frame of Reference. When we perform a NR eperiment we disturb the equilibrium state of the sstem and then monitor

More information

Advances in scmos Camera Technology Benefit Bio Research

Advances in scmos Camera Technology Benefit Bio Research Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

Breast MRI Quality Control

Breast MRI Quality Control Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation

More information

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound

High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound High-accuracy ultrasound target localization for hand-eye calibration between optical tracking systems and three-dimensional ultrasound Ralf Bruder 1, Florian Griese 2, Floris Ernst 1, Achim Schweikard

More information

Electronic Supplementary Information

Electronic Supplementary Information Electronic Supplementary Material (ESI) for Physical Chemistry Chemical Physics. This journal is the Owner Societies 2016 Electronic Supplementary Information Achieving High Resolution and Controlling

More information

High Spatial Resolution EPI Using an Odd Number of Interleaves

High Spatial Resolution EPI Using an Odd Number of Interleaves Magnetic Resonance in Medicine 41:1199 1205 (1999) High Spatial Resolution EPI Using an Odd Number of Interleaves Michael H. Buonocore* and David C. Zhu Ghost artifacts in echoplanar imaging (EPI) arise

More information

FDA Guidelines for Magnetic Resonance Equipment Safety

FDA Guidelines for Magnetic Resonance Equipment Safety FDA Guidelines for Magnetic Resonance Equipment Safety Loren A. Zaremba, Ph.D. Center for Devices and Radiological Health Food and Drug Administration Outline I. Introduction II. Static Magnetic Field

More information

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected].

Medical Image Processing on the GPU. Past, Present and Future. Anders Eklund, PhD Virginia Tech Carilion Research Institute andek@vtc.vt. Medical Image Processing on the GPU Past, Present and Future Anders Eklund, PhD Virginia Tech Carilion Research Institute [email protected] Outline Motivation why do we need GPUs? Past - how was GPU programming

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi

Diffusione e perfusione in risonanza magnetica. E. Pagani, M. Filippi Diffusione e perfusione in risonanza magnetica E. Pagani, M. Filippi DW-MRI DIFFUSION-WEIGHTED MRI Principles Diffusion results from a microspic random motion known as Brownian motion THE RANDOM WALK How

More information

Part-Based Recognition

Part-Based Recognition Part-Based Recognition Benedict Brown CS597D, Fall 2003 Princeton University CS 597D, Part-Based Recognition p. 1/32 Introduction Many objects are made up of parts It s presumably easier to identify simple

More information

Encoders for Linear Motors in the Electronics Industry

Encoders for Linear Motors in the Electronics Industry Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy

More information

Advanced MRI methods in diagnostics of spinal cord pathology

Advanced MRI methods in diagnostics of spinal cord pathology Advanced MRI methods in diagnostics of spinal cord pathology Stanisław Kwieciński Department of Magnetic Resonance MR IMAGING LAB MRI /MRS IN BIOMEDICAL RESEARCH ON HUMANS AND ANIMAL MODELS IN VIVO Equipment:

More information

Super-Resolution Reconstruction in MRI: Better Images Faster?

Super-Resolution Reconstruction in MRI: Better Images Faster? Super-Resolution Reconstruction in MRI: Better Images Faster? Esben Plenge 1,, Dirk H. J. Poot 1, Monique Bernsen 2, Gyula Kotek 2, Gavin Houston 2, Piotr Wielopolski 2, Louise van der Weerd 3, Wiro J.

More information

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed

More information

High Quality Image Magnification using Cross-Scale Self-Similarity

High Quality Image Magnification using Cross-Scale Self-Similarity High Quality Image Magnification using Cross-Scale Self-Similarity André Gooßen 1, Arne Ehlers 1, Thomas Pralow 2, Rolf-Rainer Grigat 1 1 Vision Systems, Hamburg University of Technology, D-21079 Hamburg

More information

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D.

4D Scanning. Image Guided Radiation Therapy. Outline. A Simplified View of the RT Process. Outline. Steve B. Jiang, Ph.D. 4D Scanning Steve B. Jiang, Ph.D. Department of Radiation Oncology [email protected] http://gray.mgh.harvard.edu/ Outline Problems with free breathing 3D scanning What is 4D CT? How does it work?

More information

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: [email protected]

Cirrus 0.2T. MRI for Everyone. North America, Asia, Europe. contact: kturek@mri-tech.pl Cirrus 0.2T MRI for Everyone North America, Asia, Europe contact: [email protected] MRI-TECH inc. Cirrus MRI system for all your needs: Low costs Low maintenance High quality Open geometry Imaging of

More information

Image Analysis for Volumetric Industrial Inspection and Interaction

Image Analysis for Volumetric Industrial Inspection and Interaction Image Analysis for Volumetric Industrial Inspection and Interaction Rasmus Larsen Mads F. Hansen, Hildur Olafsdottir, Thomas H. Mosbech, Lasse F. Laursen DTU Informatics (CT) reconstruction from few projections

More information

7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010

7/16/2010. Pulse Sequences and Acquisition Techniques for Breast MRI. Objectives. ACR Breast MRI Accreditation Program Launched May 2010 Pulse Sequences and Acquisition Techniques for Breast MRI ACR Breast MRI Accreditation Program Launched May 2010 Ron Price Vanderbilt University Medical Center Nashville, TN 37232 Information available:

More information

ASSESSMENT OF VISUALIZATION SOFTWARE FOR SUPPORT OF CONSTRUCTION SITE INSPECTION TASKS USING DATA COLLECTED FROM REALITY CAPTURE TECHNOLOGIES

ASSESSMENT OF VISUALIZATION SOFTWARE FOR SUPPORT OF CONSTRUCTION SITE INSPECTION TASKS USING DATA COLLECTED FROM REALITY CAPTURE TECHNOLOGIES ASSESSMENT OF VISUALIZATION SOFTWARE FOR SUPPORT OF CONSTRUCTION SITE INSPECTION TASKS USING DATA COLLECTED FROM REALITY CAPTURE TECHNOLOGIES ABSTRACT Chris Gordon 1, Burcu Akinci 2, Frank Boukamp 3, and

More information

University Children s Hospital Basel

University Children s Hospital Basel Cooperation Siemens International Reference Center Pediatric MRI www.siemens.com/skyra University Children s Hospital Basel Siemens International Reference Center Pediatric MRI Answers for life. Partnering

More information

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS 1. Bandwidth: The bandwidth of a communication link, or in general any system, was loosely defined as the width of

More information

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.

R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1. R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction

More information

How To Grow With An Mri

How To Grow With An Mri Growth with Body MRI Body MRI H.-Hartziekenhuis Lier 40 oz/d tussen 8 en 21 uur 3-5 body mri/d Aparte protocols voor leverletsels / pancreas, mrcp rectum, ovarium, cervix / uterus, pelvimetrie anus / fistula

More information

Resolution Enhancement of Photogrammetric Digital Images

Resolution Enhancement of Photogrammetric Digital Images DICTA2002: Digital Image Computing Techniques and Applications, 21--22 January 2002, Melbourne, Australia 1 Resolution Enhancement of Photogrammetric Digital Images John G. FRYER and Gabriele SCARMANA

More information

E190Q Lecture 5 Autonomous Robot Navigation

E190Q Lecture 5 Autonomous Robot Navigation E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Solving Simultaneous Equations and Matrices

Solving Simultaneous Equations and Matrices Solving Simultaneous Equations and Matrices The following represents a systematic investigation for the steps used to solve two simultaneous linear equations in two unknowns. The motivation for considering

More information

PHOTOGRAMMETRIC TECHNIQUES FOR MEASUREMENTS IN WOODWORKING INDUSTRY

PHOTOGRAMMETRIC TECHNIQUES FOR MEASUREMENTS IN WOODWORKING INDUSTRY PHOTOGRAMMETRIC TECHNIQUES FOR MEASUREMENTS IN WOODWORKING INDUSTRY V. Knyaz a, *, Yu. Visilter, S. Zheltov a State Research Institute for Aviation System (GosNIIAS), 7, Victorenko str., Moscow, Russia

More information

ELECTRON SPIN RESONANCE Last Revised: July 2007

ELECTRON SPIN RESONANCE Last Revised: July 2007 QUESTION TO BE INVESTIGATED ELECTRON SPIN RESONANCE Last Revised: July 2007 How can we measure the Landé g factor for the free electron in DPPH as predicted by quantum mechanics? INTRODUCTION Electron

More information

Broadband Networks. Prof. Dr. Abhay Karandikar. Electrical Engineering Department. Indian Institute of Technology, Bombay. Lecture - 29.

Broadband Networks. Prof. Dr. Abhay Karandikar. Electrical Engineering Department. Indian Institute of Technology, Bombay. Lecture - 29. Broadband Networks Prof. Dr. Abhay Karandikar Electrical Engineering Department Indian Institute of Technology, Bombay Lecture - 29 Voice over IP So, today we will discuss about voice over IP and internet

More information

MRI Physics for Radiologists

MRI Physics for Radiologists Alfred L. Horowitz MRI Physics for Radiologists A Visual Approach Second Edition With 94 Illustrations Springer-Verlag New York Berlin Heidelberg London Paris Tokyo Hong Kong Barcelona Budapest Alfred

More information

BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR

BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR BARE PCB INSPECTION BY MEAN OF ECT TECHNIQUE WITH SPIN-VALVE GMR SENSOR K. Chomsuwan 1, S. Yamada 1, M. Iwahara 1, H. Wakiwaka 2, T. Taniguchi 3, and S. Shoji 4 1 Kanazawa University, Kanazawa, Japan;

More information

It takes two to connect the dots in MRI.

It takes two to connect the dots in MRI. It takes two to connect the dots in MRI. www.siemens.com/healthcare Tim+ Tim (Total imaging matrix) technology has made fast, flexible and accurate scans the rule, rather than the exception. But we know

More information

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint

Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint Wireless Radio Frequency Coil for Magnetic Resonance Image of Knee Joint Tomohiro Sahara, Hiroshi Tsutsui, Shigehiro Hashimoto, Shuichi Mochizuki, Kenichi Yamasaki, Hideo Kondo, Teruo Miyazaki Department

More information

Nuclear Magnetic Resonance

Nuclear Magnetic Resonance Nuclear Magnetic Resonance Practical Course M I. Physikalisches Institut Universität zu Köln May 15, 2014 Abstract Nuclear magnetic resonance (NMR) techniques are widely used in physics, chemistry, and

More information

Case Report: Whole-body Oncologic Imaging with syngo TimCT

Case Report: Whole-body Oncologic Imaging with syngo TimCT Case Report: Whole-body Oncologic Imaging with syngo TimCT Eric Hatfield, M.D. 1 ; Agus Priatna, Ph.D. 2 ; John Kotyk, Ph.D. 1 ; Benjamin Tan, M.D. 1 ; Alto Stemmer 3 ; Stephan Kannengiesser, Ph.D. 3 ;

More information

Quantum Algorithms in NMR Experiments. 25 th May 2012 Ling LIN & Michael Loretz

Quantum Algorithms in NMR Experiments. 25 th May 2012 Ling LIN & Michael Loretz Quantum Algorithms in NMR Experiments 25 th May 2012 Ling LIN & Michael Loretz Contents 1. Introduction 2. Shor s algorithm 3. NMR quantum computer Nuclear spin qubits in a molecule NMR principles 4. Implementing

More information

Twelve. Figure 12.1: 3D Curved MPR Viewer Window

Twelve. Figure 12.1: 3D Curved MPR Viewer Window Twelve The 3D Curved MPR Viewer This Chapter describes how to visualize and reformat a 3D dataset in a Curved MPR plane: Curved Planar Reformation (CPR). The 3D Curved MPR Viewer is a window opened from

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p Data Transmission Concepts and terminology Transmission terminology Transmission from transmitter to receiver goes over some transmission medium using electromagnetic waves Guided media. Waves are guided

More information

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines

3/30/2013. Disclosure. Advanced Neuro MRI: Imaging Techniques and Protocol Optimization. MRI, 35 year ago. MRI Today. Outlines http://www.magnet.fsu.edu Disclosure Advanced Neuro MRI: Imaging Techniques and Protocol Optimization Research funding provided by Siemens Healthcare. Chen Lin, PhD DABR Indiana University School of Medicine

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

Synthetic Aperture Radar: Principles and Applications of AI in Automatic Target Recognition

Synthetic Aperture Radar: Principles and Applications of AI in Automatic Target Recognition Synthetic Aperture Radar: Principles and Applications of AI in Automatic Target Recognition Paulo Marques 1 Instituto Superior de Engenharia de Lisboa / Instituto de Telecomunicações R. Conselheiro Emídio

More information

Optical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD

Optical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD 2008 AGI-Information Management Consultants May be used for personal purporses only or by libraries associated to dandelon.com network. Optical Metrology Third Edition Kjell J. Gasvik Spectra Vision AS,

More information

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1 WHAT IS AN FFT SPECTRUM ANALYZER? ANALYZER BASICS The SR760 FFT Spectrum Analyzer takes a time varying input signal, like you would see on an oscilloscope trace, and computes its frequency spectrum. Fourier's

More information

High speed 3D capture for Configuration Management DOE SBIR Phase II Paul Banks [email protected]

High speed 3D capture for Configuration Management DOE SBIR Phase II Paul Banks Paul.banks@tetravue.com High speed 3D capture for Configuration Management DOE SBIR Phase II Paul Banks [email protected] Advanced Methods for Manufacturing Workshop September 29, 2015 1 TetraVue does high resolution 3D

More information

Phantom Test Guidance

Phantom Test Guidance Phantom Test Guidance for the ACR MRI Accreditation Program The American College of Radiology 1891 Preston White Dr Reston, VA 20191-4397 www.acr.org Contents 0.0 INTRODUCTION............................................................................

More information

Digital Signal Controller Based Automatic Transfer Switch

Digital Signal Controller Based Automatic Transfer Switch Digital Signal Controller Based Automatic Transfer Switch by Venkat Anant Senior Staff Applications Engineer Freescale Semiconductor, Inc. Abstract: An automatic transfer switch (ATS) enables backup generators,

More information

Image Compression through DCT and Huffman Coding Technique

Image Compression through DCT and Huffman Coding Technique International Journal of Current Engineering and Technology E-ISSN 2277 4106, P-ISSN 2347 5161 2015 INPRESSCO, All Rights Reserved Available at http://inpressco.com/category/ijcet Research Article Rahul

More information

Optical Encoders. K. Craig 1. Actuators & Sensors in Mechatronics. Optical Encoders

Optical Encoders. K. Craig 1. Actuators & Sensors in Mechatronics. Optical Encoders Any transducer that generates a coded reading of a measurement can be termed an encoder. Shaft Encoders are digital transducers that are used for measuring angular displacements and velocities. Relative

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

The Olympus stereology system. The Computer Assisted Stereological Toolbox

The Olympus stereology system. The Computer Assisted Stereological Toolbox The Olympus stereology system The Computer Assisted Stereological Toolbox CAST is a Computer Assisted Stereological Toolbox for PCs running Microsoft Windows TM. CAST is an interactive, user-friendly,

More information

Basics of Digital Recording

Basics of Digital Recording Basics of Digital Recording CONVERTING SOUND INTO NUMBERS In a digital recording system, sound is stored and manipulated as a stream of discrete numbers, each number representing the air pressure at a

More information

" {h k l } lop x odl. ORxOS. (2) Determine the interplanar angle between these two planes:

 {h k l } lop x odl. ORxOS. (2) Determine the interplanar angle between these two planes: Textures and Microstructures, 1991, Vols 14-18, pp. 273-278 Reprints available from the publisher Photocopying permitted by license only (C) 1991 Gordon and Breach Science Publishers SA Printed in the

More information

A Laser Scanner Chip Set for Accurate Perception Systems

A Laser Scanner Chip Set for Accurate Perception Systems A Laser Scanner Chip Set for Accurate Perception Systems 313 A Laser Scanner Chip Set for Accurate Perception Systems S. Kurtti, J.-P. Jansson, J. Kostamovaara, University of Oulu Abstract This paper presents

More information

Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T 2 *-weighted echo-planar imaging

Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T 2 *-weighted echo-planar imaging Magnetic Resonance Imaging 18 (2000) 479 484 Technical Note Artifacts caused by transcranial magnetic stimulation coils and EEG electrodes in T 2 *-weighted echo-planar imaging J. Baudewig a,b, W. Paulus

More information