ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS

Size: px
Start display at page:

Download "ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS"

Transcription

1 ON-LINE MONITORING OF AN HADRON BEAM FOR RADIOTHERAPEUTIC TREATMENTS INFN-Laboratori Nazionali del Sud Via S. Sofia 44, Catania, Italy Patient positioned for treatment System under consideration (experimental configuration) G.A.P. CIRRONE, S.COCO, G.CUTTONE, C. DE MARTINIS, D. GIOVE, P.A. LOJACONO, M. MAURI, R. MESSINA

2 AT LABORATORI NAZIONALI DEL SUD (CATANIA), CATANA FACILITY (Centro di AdroTerapia e Applicazioni Nucleari Avanzate) employs a 62 MeV proton beam, produced by a superconducting cyclotron, for the radiotherapeutic treatment of choroid and iridal melanoma. The main requirement for a successful treament is: ACCURACY in the spatial distribution of thedosereleased to skin guaranteed by numerous experimental checks Patient positioned for the treatment

3 ROUTINE BEAM MONITORING SYSTEMS DIODE: transversal scanning of the beam by means of a silicon diode which measures the current and displays its distribution along the two coordinates of a Cartesian system, transversal to the beam direction; GAFCHROMIC FILM: determination of the lateral dose distribution released by the beam from the darkening of an irradiated gafchromic film; both systems, though very accurate, are subject to long acquisition and processing times Acquisition period ~30 minutes Acquisition period ~24 hours

4 DEVELOPED SYSTEM HADRON BEAM SCINTILLATING SCREEN 45 MIRROR monitoring system of a hadron beam based on a fluorescent screen coupled with an analog CCD camera; acquisition and processing of the image reflected by the mirror by means of a software application implemented with LabVIEW 6.1; CCD LabVIEW Acquisition period ~ 1 minute

5 TESTED SCINTILLATORS Alumina (Al2O3) Caesium Iodide (CsI) Plastic scintillator BC 400 Alumina: poor in light efficiency; Caesium Iodide: excellent light efficiency, but scarce homogeneity; BC 400: good brightness and homogeneity; BC 400 has been chosen for our experimental purposes.

6 FINAL CONFIGURATION Mirror forming 45 with the beam direction. Scintillating screen lodged on a support, perpendicularly to the beam axis. Camera forming 90 with the beam axis, framing the image reflected by the mirror.

7 DATA ANALYSIS High quality monochrome image acquisition board: IMAQ PCI-1409 (National Instruments) Acquisition and memorization of signal in the form of 10 bits matrices Spatial coordinates of each pixel of the camera Corresponding grey-scale values

8 SOFTWARE FOR DATA ANALYSIS LabVIEW Environment Spatial calibration Cartesian reference system Calibration factor: mm/pixel Selection of the Region Of Interest

9 SOFTWARE FOR DATA ANALYSIS LabVIEW Environment NOISE REDUCTION: average of N images EXPERIMENTAL SIGNAL ACQUISITION BACKGROUND SUBTRACTION

10 SOFTWARE ANALYSIS Acquisition and analysis of lateral profiles signal % W 50 % 60 H 90/50 % Psx distance from axis (mm) Pdx Remote communication, through the intranet, between the application and the accelerators console

11 FRONT PANEL OF THE VI LabView Environment

12 BLOCK DIAGRAM OF THE VI LabView Environment

13 3D-GRAPH

14 CHARACTERIZATION OF THE SYSTEM LINEARITY SIGNAL TO NOISE RATIO SNR vs sqrt(frames) Equazion: y = A*x + B A = ± B = ± R 2 = SNR Experimental data linear FIT sqrt (frames) Linearity in the treatment range (10 20 Gy/min) Castleman says: SNR (N)= N^0.5*SNR(1) Experimental results are in good agreement with theoretical considerations

15 CHARACTERIZATION OF THE SYSTEM CONTRAST SPATIAL RESOLUTION Background subtraction upgrades image contrast of a factor 2 Given: MTF = 0.3 Spatial resolution is : 2.22 mm/lp

16 EXPERIMENTAL RESULTS AND COMPARISONS signal % CCD - DIODE COMPARED PROFILES diode CCD Peculiar parameters of a lateral beam profile, calculated with the two systems,are in a good agreement CCD diode ratio 0 profile ratio distance from axis (mm) X AXIS W50% (mm) H 90/50% Psx (mm) Pdx (mm) CCD 50 FRAMES DIODE DIFF.%

17 EXPERIMENTAL RESULTS AND COMPARISONS

18 CONCLUSIONS AND EXPECTATIONS Comparable results with those obtained through currently used systems Remarkably reduced times of data acquisition and processing Potential employment of the system as a fast and accurate instrument of: beam centering (directly used by cyclotron operators) beam monitoring In the future, the use of thinner scintillating materials is expected, in order to overcome some scattering phenomena which affect spatial resolution.

19 CHARACTERIZATION OF THE SYSTEM MODULATION TRANSFER FUNCTION INPUT MODULATION OUTPUT MODULATION

20 CHARACTERIZATION OF THE SYSTEM MODULATION TRANSFER FUNCTION PATTERN METHOD IDEALLY: SINUSOIDAL WAVES OF SAME AMPLITUDE AND DIFFERENT FREQUENCY ACTUALLY THE IMAGING SYSTEM MODIFIES THE SIGNAL SINUSOIDAL WAVES OF DIFFERENT AMPLITUDE AND FREQUENCY

21 CARATTERIZZAZIONE DEL SISTEMA MODULATION TRANSFER FUNCTION FOURIER TRANSFORM GIVEN A FUNCTION: FOURIER TRANSFORM IS: EXAMPLE: FOURIER TRANSFORM IS: GIVEN THE FOURIER TRANSFORM H(f), THE INVERSE OPERATION, CALLED ANTITRANSFORMATION, GIVES THE BEGINNING FUNCTION: SUM OF INFINITE SINUSOIDAL AND COSINUSOIDAL FUNCTIONS OF DIFFERENT AMPLITUDES AND VARYING FREQUENCIES

22 CHARACTERIZATION OF THE SYSTEM MODULATION TRANSFER FUNCTION SLIT METHOD: IDEALLY: AN IDEAL SLIT CAN BE SEEN AS THE SUM OF INFINITE COSINUSOIDAL FUNCTIONS OF DIFFERENT FREQUENCIES, BUT SAME AMPLITUDE CONSTANT: ACTUALLY: GAUSSIAN CURVE: SUM OF INFINITE COSIDUSOIDS OF DIFFERENT FREQUENCIES AND AMPLITUDES DECREASING WITH FREQUENCY:

23 CARATTERIZZAZIONE DEL SISTEMA MODULATION TRANSFER FUNCTION EDGE METHOD: IDEALLY: ACTUALLY:

24 CHARACTERIZATION OF THE SYSTEM MODULATION TRANSFER FUNCTION Pattern Edge

25 CHARACTERIZATION OF THE SYSTEM CONTRAST Definizione e figura Johns

26 DEVELOPED SYSTEM HADRON BEAM MIRROR 45 SCINTILLATING SCREEN CCD LabVIEW

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation

The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:

More information

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules

High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules High Resolution Spatial Electroluminescence Imaging of Photovoltaic Modules Abstract J.L. Crozier, E.E. van Dyk, F.J. Vorster Nelson Mandela Metropolitan University Electroluminescence (EL) is a useful

More information

Application Note Noise Frequently Asked Questions

Application Note Noise Frequently Asked Questions : What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random

More information

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry

Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Cyclotron Centre in Poland and 2D thermoluminescence dosimetry Jan Gajewski Institute of Nuclear Physics, Kraków, Poland Department of Radiation Dosimetry Nuclear Physics Institute Academy of Science of

More information

Realization of a UV fisheye hyperspectral camera

Realization of a UV fisheye hyperspectral camera Realization of a UV fisheye hyperspectral camera Valentina Caricato, Andrea Egidi, Marco Pisani and Massimo Zucco, INRIM Outline Purpose of the instrument Required specs Hyperspectral technique Optical

More information

Lecture 14. Point Spread Function (PSF)

Lecture 14. Point Spread Function (PSF) Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect

More information

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy

An Overview of Digital Imaging Systems for Radiography and Fluoroscopy An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General

More information

Radiography: 2D and 3D Metrics of Performance Towards Quality Index

Radiography: 2D and 3D Metrics of Performance Towards Quality Index AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..

More information

Robot Sensors. Outline. The Robot Structure. Robots and Sensors. Henrik I Christensen

Robot Sensors. Outline. The Robot Structure. Robots and Sensors. Henrik I Christensen Robot Sensors Henrik I Christensen Robotics & Intelligent Machines @ GT Georgia Institute of Technology, Atlanta, GA 30332-0760 hic@cc.gatech.edu Henrik I Christensen (RIM@GT) Sensors 1 / 38 Outline 1

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

DATA ACQUISITION FROM IN VITRO TESTING OF AN OCCLUDING MEDICAL DEVICE

DATA ACQUISITION FROM IN VITRO TESTING OF AN OCCLUDING MEDICAL DEVICE DATA ACQUISITION FROM IN VITRO TESTING OF AN OCCLUDING MEDICAL DEVICE Florentina ENE 1, Carine GACHON 2, Nicolae IONESCU 3 ABSTRACT: This paper presents a technique for in vitro testing of an occluding

More information

MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE

MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE K. Kohm ORBIMAGE, 1835 Lackland Hill Parkway, St. Louis, MO 63146, USA kohm.kevin@orbimage.com

More information

Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE

Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE Technical Teaching Equipment Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE EDIBON SCADA System Teaching Technique used 4 5 2 Data Acquisition Board Cables and

More information

Comparisons between 2003 CMS ECAL TB data and a Geant 4 MC

Comparisons between 2003 CMS ECAL TB data and a Geant 4 MC Comparisons between 2003 CMS CAL TB data and a Geant 4 MC P. Meridiani LCG Validation Meeting 7 th July 2004 Outline CMS lectromagnetic calorimeter and 2003 TB h4sim http://cmsdoc.cern.ch/~h4sim/ (What

More information

Robot Perception Continued

Robot Perception Continued Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart

More information

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology product overview family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology scmos knowledge base scmos General Information PCO scmos cameras are a breakthrough

More information

Computer Controlled Vortex Tube Refrigerator Unit, with SCADA TPVC

Computer Controlled Vortex Tube Refrigerator Unit, with SCADA TPVC Technical Teaching Equipment Computer Controlled Vortex Tube Refrigerator Unit, with SCADA TPVC Teaching Technique used EDIBON SCADA System 2 Control Interface Box 5 Cables and Accessories 6 Manuals 3

More information

Manual for simulation of EB processing. Software ModeRTL

Manual for simulation of EB processing. Software ModeRTL 1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended

More information

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,

More information

Color holographic 3D display unit with aperture field division

Color holographic 3D display unit with aperture field division Color holographic 3D display unit with aperture field division Weronika Zaperty, Tomasz Kozacki, Malgorzata Kujawinska, Grzegorz Finke Photonics Engineering Division, Faculty of Mechatronics Warsaw University

More information

Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP

Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP Now Powered by LightField FEATURES BENEFITS Back Illuminated CCD (248B) For highest sensitivity Front illuminated CCD (248F) Affordable technology for moderate light level applications Ultra low noise

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time Module 8: IQ: Acquisition Time IQ : Acquisition Time Objectives...Describe types of data acquisition modes....compute acquisition times for 2D and 3D scans. 2D Acquisitions The 2D mode acquires and reconstructs

More information

Digital Camera Imaging Evaluation

Digital Camera Imaging Evaluation Digital Camera Imaging Evaluation Presenter/Author J Mazzetta, Electro Optical Industries Coauthors Dennis Caudle, Electro Optical Industries Bob Wageneck, Electro Optical Industries Contact Information

More information

a leap ahead in analog

a leap ahead in analog EMV Contactless Payment Systems based on AS3911 Overview and System Simulations Giuliano Manzi, PhD Mannheim, May 23 25, 2012 CST EUROPEAN USER CONFERENCE 2012 a leap ahead in analog OUTLINE AS3911 OVERVIEW

More information

APPLICATION NOTE. Basler racer Migration Guide. Mechanics. www.baslerweb.com. Flexible Mount Concept. Housing

APPLICATION NOTE. Basler racer Migration Guide. Mechanics. www.baslerweb.com. Flexible Mount Concept. Housing 62 62 APPLICATION NOTE www.baslerweb.com Basler racer Migration Guide This paper describes what to consider when replacing the Basler L100 Camera Link or the Basler runner Gigabit Ethernet (GigE) line

More information

Email: tjohn@mail.nplindia.ernet.in

Email: tjohn@mail.nplindia.ernet.in USE OF VIRTUAL INSTRUMENTS IN RADIO AND ATMOSPHERIC EXPERIMENTS P.N. VIJAYAKUMAR, THOMAS JOHN AND S.C. GARG RADIO AND ATMOSPHERIC SCIENCE DIVISION, NATIONAL PHYSICAL LABORATORY, NEW DELHI 110012, INDIA

More information

Physics 441/2: Transmission Electron Microscope

Physics 441/2: Transmission Electron Microscope Physics 441/2: Transmission Electron Microscope Introduction In this experiment we will explore the use of transmission electron microscopy (TEM) to take us into the world of ultrasmall structures. This

More information

A System for Capturing High Resolution Images

A System for Capturing High Resolution Images A System for Capturing High Resolution Images G.Voyatzis, G.Angelopoulos, A.Bors and I.Pitas Department of Informatics University of Thessaloniki BOX 451, 54006 Thessaloniki GREECE e-mail: pitas@zeus.csd.auth.gr

More information

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this

More information

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners

Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed

More information

Characterizing Digital Cameras with the Photon Transfer Curve

Characterizing Digital Cameras with the Photon Transfer Curve Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently

More information

Optical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD

Optical Metrology. Third Edition. Kjell J. Gasvik Spectra Vision AS, Trondheim, Norway JOHN WILEY & SONS, LTD 2008 AGI-Information Management Consultants May be used for personal purporses only or by libraries associated to dandelon.com network. Optical Metrology Third Edition Kjell J. Gasvik Spectra Vision AS,

More information

Assessment of Camera Phone Distortion and Implications for Watermarking

Assessment of Camera Phone Distortion and Implications for Watermarking Assessment of Camera Phone Distortion and Implications for Watermarking Aparna Gurijala, Alastair Reed and Eric Evans Digimarc Corporation, 9405 SW Gemini Drive, Beaverton, OR 97008, USA 1. INTRODUCTION

More information

Encoders for Linear Motors in the Electronics Industry

Encoders for Linear Motors in the Electronics Industry Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy

More information

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES

Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture

More information

Data Acquisition Using NI-DAQmx

Data Acquisition Using NI-DAQmx Instructor s Portion Data Acquisition Using NI-DAQmx Wei Lin Department of Biomedical Engineering Stony Brook University Summary This experiment requires the student to use NI-DAQmx to acquire voltage

More information

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film. INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries

College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 2166-Handout College on Medical Physics. Digital Imaging Science and Technology to Enhance Healthcare in the Developing Countries 13 September - 1 October, 2010 Digital Radiography Image Parameters SNR,

More information

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters

More information

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror Stephen B. Robie scintag, Inc. 10040 Bubb Road Cupertino, CA 95014 Abstract Corundum

More information

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS 1. Bandwidth: The bandwidth of a communication link, or in general any system, was loosely defined as the width of

More information

Physics testing of image detectors

Physics testing of image detectors Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal

More information

Dynamic Neutron Imaging

Dynamic Neutron Imaging Wir schaffen Wissen heute für morgen Pierre Boillat Neutron Imaging & Activation Group (NIAG) and Electrochemistry Laboratory (LEC), Paul Scherrer Institute, Switzerland Dynamic Neutron Imaging PSI, 8.

More information

E190Q Lecture 5 Autonomous Robot Navigation

E190Q Lecture 5 Autonomous Robot Navigation E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator

More information

Synthetic Sensing: Proximity / Distance Sensors

Synthetic Sensing: Proximity / Distance Sensors Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,

More information

DIMEG - University of L Aquila ITALY EXPERIMENTAL ACTIVITY ENGINE LABORATORY

DIMEG - University of L Aquila ITALY EXPERIMENTAL ACTIVITY ENGINE LABORATORY DIMEG - University of L Aquila ITALY EXPERIMENTAL ACTIVITY ENGINE LABORATORY Torre di Raffreddamento Bilan cia Combustibile DIMEG:ENGINE LABORATORY PLANTS Torre di Raffreddamento P C o o z l z d o P C

More information

Testing thermo-acoustic sound generation in water with proton and laser beams

Testing thermo-acoustic sound generation in water with proton and laser beams International ARENA Workshop DESY, Zeuthen 17th 19th of May 25 Testing thermo-acoustic sound generation in water with proton and laser beams Kay Graf Universität Erlangen-Nürnberg Physikalisches Institut

More information

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete

More information

Fig.1. The DAWN spacecraft

Fig.1. The DAWN spacecraft Introduction Optical calibration of the DAWN framing cameras G. Abraham,G. Kovacs, B. Nagy Department of Mechatronics, Optics and Engineering Informatics Budapest University of Technology and Economics

More information

Performance testing for Precision 500D Classical R/F System

Performance testing for Precision 500D Classical R/F System Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)

More information

LLRF. Digital RF Stabilization System

LLRF. Digital RF Stabilization System LLRF Digital RF Stabilization System Many instruments. Many people. Working together. Stability means knowing your machine has innovative solutions. For users, stability means a machine achieving its full

More information

A Beam Image Server For TTF2. L.Catani PCaPAC 2002

A Beam Image Server For TTF2. L.Catani PCaPAC 2002 A Beam Image Server For TTF2 L.Catani PCaPAC 2002 A Beam Image Server For TTF2 (why we keep on using LabVIEW) L.Catani PCaPAC 2002 About About evolution of the Optical Diagnostic Control System (ODCS)

More information

Experimental results for the focal waveform and beam width in the focusing lens with a 100 ps filter

Experimental results for the focal waveform and beam width in the focusing lens with a 100 ps filter EM Implosion Memos Memo 51 July, 2010 Experimental results for the focal waveform and beam width in the focusing lens with a 100 ps filter Prashanth Kumar, Carl E. Baum, Serhat Altunc, Christos G. Christodoulou

More information

AxioCam HR The Camera that Challenges your Microscope

AxioCam HR The Camera that Challenges your Microscope Microscopy from Carl Zeiss AxioCam HR The Camera that Challenges your Microscope Documentation at the edge of the visible The Camera for Maximum Success: AxioCam HR Low light fluorescence, live cell imaging,

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P4.2.1.1

Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P4.2.1.1 Electronics Operational Amplifier Internal design of an operational amplifier LD Physics Leaflets Discrete assembly of an operational amplifier as a transistor circuit P4.2.1.1 Objects of the experiment

More information

Computer Vision. Image acquisition. 25 August 2014. Copyright 2001 2014 by NHL Hogeschool and Van de Loosdrecht Machine Vision BV All rights reserved

Computer Vision. Image acquisition. 25 August 2014. Copyright 2001 2014 by NHL Hogeschool and Van de Loosdrecht Machine Vision BV All rights reserved Computer Vision Image acquisition 25 August 2014 Copyright 2001 2014 by NHL Hogeschool and Van de Loosdrecht Machine Vision BV All rights reserved j.van.de.loosdrecht@nhl.nl, jaap@vdlmv.nl Image acquisition

More information

NRZ Bandwidth - HF Cutoff vs. SNR

NRZ Bandwidth - HF Cutoff vs. SNR Application Note: HFAN-09.0. Rev.2; 04/08 NRZ Bandwidth - HF Cutoff vs. SNR Functional Diagrams Pin Configurations appear at end of data sheet. Functional Diagrams continued at end of data sheet. UCSP

More information

FREQUENCY RESPONSE ANALYZERS

FREQUENCY RESPONSE ANALYZERS FREQUENCY RESPONSE ANALYZERS Dynamic Response Analyzers Servo analyzers When you need to stabilize feedback loops to measure hardware characteristics to measure system response BAFCO, INC. 717 Mearns Road

More information

Development of on line monitor detectors used for clinical routine in proton and ion therapy

Development of on line monitor detectors used for clinical routine in proton and ion therapy Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary

More information

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p Data Transmission Concepts and terminology Transmission terminology Transmission from transmitter to receiver goes over some transmission medium using electromagnetic waves Guided media. Waves are guided

More information

Towards large dynamic range beam diagnostics and beam dynamics studies. Pavel Evtushenko

Towards large dynamic range beam diagnostics and beam dynamics studies. Pavel Evtushenko Towards large dynamic range beam diagnostics and beam dynamics studies Pavel Evtushenko Motivation Linacs with average current 1-2 ma and energy 1-2.5 GeV are envisioned as drivers for next generation

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors

Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Study of the Human Eye Working Principle: An impressive high angular resolution system with simple array detectors Diego Betancourt and Carlos del Río Antenna Group, Public University of Navarra, Campus

More information

Assessment Plan for Learning Outcomes for BA/BS in Physics

Assessment Plan for Learning Outcomes for BA/BS in Physics Department of Physics and Astronomy Goals and Learning Outcomes 1. Students know basic physics principles [BS, BA, MS] 1.1 Students can demonstrate an understanding of Newton s laws 1.2 Students can demonstrate

More information

The best lab standard. 1,4 Megapixels 2/3 inch sensor Giant pixel size 6 times optical zoom Massive 16-bit imaging for enhanced dynamic

The best lab standard. 1,4 Megapixels 2/3 inch sensor Giant pixel size 6 times optical zoom Massive 16-bit imaging for enhanced dynamic 1 AGENDA Product roadmap...3 Computer based system: PLATINUM..5 XPLORER.6 ESSENTIAL..7 Specifications Computer based system..8 Configuration Computer based system... 9 Software Computer based system...10

More information

CHAPTER 6 Frequency Response, Bode Plots, and Resonance

CHAPTER 6 Frequency Response, Bode Plots, and Resonance ELECTRICAL CHAPTER 6 Frequency Response, Bode Plots, and Resonance 1. State the fundamental concepts of Fourier analysis. 2. Determine the output of a filter for a given input consisting of sinusoidal

More information

DATA LOGGING SYSTEM FOR PRESSURE MONITORING

DATA LOGGING SYSTEM FOR PRESSURE MONITORING DATA LOGGING SYSTEM FOR PRESSURE MONITORING Georgi Todorov Nikolov, Boyanka Marinova Nikolova, Marin Berov Marinov Department of Electronics, Technical University of Sofia, Studenstki Grad, TU-Sofia, block

More information

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging

CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through

More information

Implementing and Using the EMVA1288 Standard

Implementing and Using the EMVA1288 Standard Implementing and Using the EMVA1288 Standard A. Darmont *, J. Chahiba, J.-F. Lemaitre, M. Pirson, D. Dethier Aphesa, Rue de Lorcé, 39, 4920 Harzé, Belgium ABSTRACT The European Machine Vision Association

More information

IS (Iris Security) Research, Imaging Equipment, University/Education

IS (Iris Security) Research, Imaging Equipment, University/Education IS (Iris Security) Gerardo Iovane, Francesco Saverio Tortoriello Researchers Dipartimento di Ingegneria Informatica e Matematica Applicata Università degli Studi di Salerno Research, Imaging Equipment,

More information

Lab 2 Sensor Modeling Introduction to the Wiimote and the Wiimote LABVIEW Interface

Lab 2 Sensor Modeling Introduction to the Wiimote and the Wiimote LABVIEW Interface Lab 2 Sensor Modeling Introduction to the Wiimote and the Wiimote LABVIEW Interface 1. Objective In this lab, you will: i. Become familiar with the Wiimote ii. Understand the LABVIEW Bluetooth interface

More information

FAST Fourier Transform (FFT) and Digital Filtering Using LabVIEW

FAST Fourier Transform (FFT) and Digital Filtering Using LabVIEW FAST Fourier Transform (FFT) and Digital Filtering Using LabVIEW Wei Lin Department of Biomedical Engineering Stony Brook University Instructor s Portion Summary This experiment requires the student to

More information

Introduction to CCDs and CCD Data Calibration

Introduction to CCDs and CCD Data Calibration Introduction to CCDs and CCD Data Calibration Dr. William Welsh San Diego State University CCD: charge coupled devices integrated circuit silicon chips that can record optical (and X-ray) light pixel =

More information

MATRIX TECHNICAL NOTES

MATRIX TECHNICAL NOTES 200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR

More information

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time)

A pulse is a collection of cycles that travel together. the cycles ( on or transmit time), and. the dead time ( off or receive time) chapter 2 Pulsed Ultrasound In diagnostic ultrasound imaging, short bursts, or pulses, of acoustic energy are used to create anatomic images. Continuous wave sound cannot create anatomic images. Analogy

More information

arxiv:astro-ph/0509450 v1 15 Sep 2005

arxiv:astro-ph/0509450 v1 15 Sep 2005 arxiv:astro-ph/0509450 v1 15 Sep 2005 TESTING THERMO-ACOUSTIC SOUND GENERATION IN WATER WITH PROTON AND LASER BEAMS K. GRAF, G. ANTON, J. HÖSSL, A. KAPPES, T. KARG, U. KATZ, R. LAHMANN, C. NAUMANN, K.

More information

Experiment 7: Familiarization with the Network Analyzer

Experiment 7: Familiarization with the Network Analyzer Experiment 7: Familiarization with the Network Analyzer Measurements to characterize networks at high frequencies (RF and microwave frequencies) are usually done in terms of scattering parameters (S parameters).

More information

Ultrasonic Wave Propagation Review

Ultrasonic Wave Propagation Review Ultrasonic Wave Propagation Review Presented by: Sami El-Ali 1 1. Introduction Ultrasonic refers to any study or application of sound waves that are higher frequency than the human audible range. Ultrasonic

More information

22.302 Experiment 5. Strain Gage Measurements

22.302 Experiment 5. Strain Gage Measurements 22.302 Experiment 5 Strain Gage Measurements Introduction The design of components for many engineering systems is based on the application of theoretical models. The accuracy of these models can be verified

More information

Rapid Application Development for Machine Vision A New Approach

Rapid Application Development for Machine Vision A New Approach Rapid Application Development for Machine Vision A New Approach Introduction Converging technologies, such as the PCI-bus and Intel MMX, have created so much bandwidth and computing power that automation

More information

Agilent PN 89400-13 Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth

Agilent PN 89400-13 Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth Agilent PN 89400-13 Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth Product Note The Agilent Technologies 89400 series vector signal analyzers provide unmatched signal analysis

More information

Computer Controlled Nozzle Performance Test Unit, with SCADA TFTC

Computer Controlled Nozzle Performance Test Unit, with SCADA TFTC Technical Teaching Equipment Computer Controlled Nozzle Performance Test Unit, with SCADA TFTC EDIBON SCADA System Teaching Technique used 1 Unit: TFTC. Nozzle Performance Test Unit 5 6 2 Control Interface

More information

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION 1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.

More information

Optimizing IP3 and ACPR Measurements

Optimizing IP3 and ACPR Measurements Optimizing IP3 and ACPR Measurements Table of Contents 1. Overview... 2 2. Theory of Intermodulation Distortion... 2 3. Optimizing IP3 Measurements... 4 4. Theory of Adjacent Channel Power Ratio... 9 5.

More information

LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS

LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS LEVERAGING FPGA AND CPLD DIGITAL LOGIC TO IMPLEMENT ANALOG TO DIGITAL CONVERTERS March 2010 Lattice Semiconductor 5555 Northeast Moore Ct. Hillsboro, Oregon 97124 USA Telephone: (503) 268-8000 www.latticesemi.com

More information

Physics 9e/Cutnell. correlated to the. College Board AP Physics 1 Course Objectives

Physics 9e/Cutnell. correlated to the. College Board AP Physics 1 Course Objectives Physics 9e/Cutnell correlated to the College Board AP Physics 1 Course Objectives Big Idea 1: Objects and systems have properties such as mass and charge. Systems may have internal structure. Enduring

More information

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. CHAPTER 3: OSCILLOSCOPE AND SIGNAL GENERATOR 3.1 Introduction to oscilloscope 1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal. 2. The graph show signal change

More information

Resolution Enhancement of Photogrammetric Digital Images

Resolution Enhancement of Photogrammetric Digital Images DICTA2002: Digital Image Computing Techniques and Applications, 21--22 January 2002, Melbourne, Australia 1 Resolution Enhancement of Photogrammetric Digital Images John G. FRYER and Gabriele SCARMANA

More information

High Power Programmable DC Power Supplies PVS Series

High Power Programmable DC Power Supplies PVS Series Data Sheet High Power Programmable DC Power Supplies The PVS10005, PVS60085, and PVS60085MR programmable DC power supplies offer clean output power up to 5.1 kw, excellent regulation, and fast transient

More information

Digital Mammography Update: Design and Characteristics of Current Systems

Digital Mammography Update: Design and Characteristics of Current Systems : Design and Characteristics of Current Systems Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington AAPM Annual Meeting 2009 Anaheim,

More information

Advances in scmos Camera Technology Benefit Bio Research

Advances in scmos Camera Technology Benefit Bio Research Advances in scmos Camera Technology Benefit Bio Research scmos camera technology is gaining in popularity - Why? In recent years, cell biology has emphasized live cell dynamics, mechanisms and electrochemical

More information

RF Measurements Using a Modular Digitizer

RF Measurements Using a Modular Digitizer RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.

More information

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Progress In Electromagnetics Research Symposium Proceedings, Taipei, March 5 8, 3 359 Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm Yoshito Sonoda, Takashi Samatsu, and

More information

Introduction to the EXAFS data analysis

Introduction to the EXAFS data analysis Introduction to the EXAFS data analysis Giuliana Aquilanti Elettra Sincrotrone Trieste Material almost integrally taken from Carlo Meneghini: EXAFS tutorial at Synchrotron Radiation school of Duino 2011

More information

THE DØ Luminosity Monitor (LM) consists of two array of

THE DØ Luminosity Monitor (LM) consists of two array of FERMILAB-CONF-6-469-E Radiation Damage to Scintillator in the DØ Luminosity Monitor Brendan Casey, Kayle DeVaughan, Yuji Enari, Richard Partridge, and Sahal Yacoob Abstract We report the result of evaluating

More information

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW

Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,

More information