Quantitative analysis using ATR-FTIR Spectroscopy

Similar documents
QUANTITATIVE DETERMINATION OF COMMON TYPES OF ASBESTOS BY DIFFUSE REFLECTANCE FTIR

Introduction to Fourier Transform Infrared Spectrometry

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

FTIR Instrumentation

DETECTION OF COATINGS ON PAPER USING INFRA RED SPECTROSCOPY

Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Back to Basics Fundamentals of Polymer Analysis

Components for Infrared Spectroscopy. Dispersive IR Spectroscopy

Signal, Noise, and Detection Limits in Mass Spectrometry

13C NMR Spectroscopy

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information

CHE334 Identification of an Unknown Compound By NMR/IR/MS

m/z

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Pesticide Analysis by Mass Spectrometry

Signal to Noise Instrumental Excel Assignment

Increasing Quality While Maintaining Efficiency in Drug Chemistry with DART-TOF MS Screening

Advanced Solutions for Gas Monitoring

Infrared Spectroscopy 紅 外 線 光 譜 儀

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

Organic Chemistry Tenth Edition

AMD Analysis & Technology AG

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION

Waters Core Chromatography Training (2 Days)

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

NMR and other Instrumental Techniques in Chemistry and the proposed National Curriculum.

Optical Lab Room 1122 Nexus 670 and Magna 560 Instructions Feb. 26, 2013

Raman spectroscopy Lecture

INFRARED SPECTROSCOPY (IR)

Validation of a FTIR Spectroscopy Method for Measuring and Monitoring Fuel Dilution in Lubricating Oils

How To Test For Contamination In Large Volume Water

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3

Spectrum 10 Spectroscopy Software

Agilent Cary 60 UV-Vis

Chemistry Instrumental Analysis Lecture 1. Chem 4631

TotalChrom. Chromatography Data Systems. Streamlining your laboratory workflow

DETERMINACIÓN DE ESTRUCTURAS ORGÁNICAS (ORGANIC SPECTROSCOPY) IR SPECTROSCOPY

Unit title: Chemical Laboratory Techniques

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory

Setting up a Quantitative Analysis MS ChemStation

Application Note: Absorbance

AxION edoor. Web-Based, Open-Access Mass Spectrometry Software

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Austin Peay State University Department of Chemistry Chem The Use of the Spectrophotometer and Beer's Law

Journal of the University of Chemical Technology and Metallurgy, 42, 2, ) are in C 1

Functional Data Analysis of MALDI TOF Protein Spectra

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

MB3000-PH FT-IR spectrometer for the pharmaceutical industry Designed for QA/QC, formulation, research and development laboratory applications

Experiment 11. Infrared Spectroscopy

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

Sample Analysis Design Isotope Dilution

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

LIQUID CHROMATOGRAPHY HOW MUCH ASPIRIN, ACETAMINOPHEN, AND CAFFEINE ARE IN YOUR PAIN RELIEVER? USING HPLC TO QUANTITATE SUBSTANCES (Revised: )

Evaluating System Suitability CE, GC, LC and A/D ChemStation Revisions: A.03.0x- A.08.0x

The photoionization detector (PID) utilizes ultraviolet

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics

Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2

Fourier Transform Infrared (FTIR) Based Oil Condition Monitoring for Synthetic Turbine Oils

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

The excitation in Raman spectroscopy is usually. Practical Group Theory and Raman Spectroscopy, Part II: Application of Polarization

How To Use An Atr Spectra On A Computer With A Pike Atr Atr 27 Atr Cell

Analyzing Small Molecules by EI and GC-MS. July 2014

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Nuclear Magnetic Resonance Spectroscopy

Nederland België / Belgique

Uses of Derivative Spectroscopy

Background Information

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

SUCRALOSE. White to off-white, practically odourless crystalline powder

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Infrared Spectroscopy: Theory

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis

Copyright by Mark Brandt, Ph.D. 12

X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING

Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer. NAME: Lab Section: Date: Sign-Off:

Telephone: (U.S.A.) or (worldwide) Fax: (worldwide) techsupport.analyze@thermo.

CHEM 51LB: EXPERIMENT 5 SPECTROSCOPIC METHODS: INFRARED AND NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY

Shimadzu UV-VIS User s Guide

APPLIED. Measuring Caffeine Concentration. Application Summary. Analytes: caffeine (C 8 H 10 N 4 O 2

Detection of Exposure Damage in Composite Materials Using Fourier Transform Infrared Technology

TABLE OF CONTENTS CHAPTER TITLE PAGE DECLARATION ACKNOWLEDGEMENT ABSTRACT LIST OF TABLES LIST OF FIGURES LIST OF SYMBOLS LIST OF APPENDICES

MassHunter for Agilent GC/MS & GC/MS/MS

Guidance for Industry

The Unshifted Atom-A Simpler Method of Deriving Vibrational Modes of Molecular Symmetries

RoHS Test Methods. Presented by: Bruce Peterson

amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

EDXRF of Used Automotive Catalytic Converters

Thermo Scientific GC-MS Data Acquisition Instructions for Cerno Bioscience MassWorks Software

Transcription:

Quantitative analysis using ATR-FTIR Spectroscopy Application Note Author Simon Boyd, Jonah Kirkwood Agilent Technologies, Inc. Introduction The use of attenuated total reflection (ATR) accessories in conjunction with Fourier transform infrared (FTIR) spectrometers is now commonplace. This accessory provides for the non-destructive measurement of samples with little or no preparation. Most samples can be directly applied to the internal reflection element (IRE) of an ATR without time-consuming dilution with matrices such as Nujol or KBr. The ATR accessory also allows for easy analysis of liquid samples with just a single drop required, applied directly to the IRE crystal. However, by the nature of their design, ATR accessories absorb infrared radiation and consequently reduce the amount of energy that reaches the infrared detector. The attenuation caused by these accessories typically varies from 70 90%. Agilent Cary 600 FTIR series spectrometers provide the ideal platform for this accessory by delivering up to four times more IR power at the sample focus than most competitive products. 1 They have been designed to offset accessory attenuation of the IR beam and to provide exceptional performance and sensitivity, which are typically compromised with the use of the accessory. Infrared spectroscopy is particularly well-suited to quantitative analyses. The determination of sample concentration is made simple with the Quant Calibration application in the Resolutions Pro software. This Application Note highlights the benefits of using Agilent s Cary 600 FTIR spectrometer series and software for simple quantitative analysis by using a model system of ethanol (EtOH) in water. The limit of detection is measured using a PIKE MIRacle single bounce diamond ATR in conjunction with an Agilent Cary 660 FTIR (mechanical-bearing) and a Cary 670 FTIR (air-bearing) spectrometer is also determined.

Instrumentation The Agilent Cary 660 FTIR and Cary 670 FTIR spectrometers were used in conjunction with a PIKE MIRacle single bounce diamond ATR accessory. The spectrometer settings used in all spectral collections are listed in Table 1. The instrument and accessory were purged for five minutes with nitrogen prior to spectral collection. Table 1. Instrument parameters used in all collections Detector Source ATR Collection Computation Instrument Parameters DLaTGS Speed (khz) 5.0 MIR PIKE Diamond MIRacle Sample scans Background scans Resolution (cm -1 ) Aperture (cm -1 ) Symmetry Apodisation type Zero filling factor UDR Materials and reagents Settings Normal 16 64 4 4 sym BH4 2 2 Standard solutions for the quantitative calibration curve were sequentially diluted from an aliquot of ethanol (5.00mL, Merck, HPLC grade) in deionized water (100.0mL, 18 MΩ cm, Millipore Milli-Q, Billerica, MA, USA). Figure 1. The PIKE MIRacle ATR used in this study. Energy-limiting accessories such as an ATR require a high energy FTIR spectrometer to offset accessory attenuation Results and discussion Quant Calibration - Ethanol in Water A calibration curve was prepared from a range of solutions of known concentration of ethanol in water. An aliquot of each solution was placed on the IRE of the ATR and spectra were recorded (with both the Cary 660 FTIR and Cary 670 FTIR) using the parameters in Table 1. The spectrum of each solution was ratioed against a background spectrum of water. Figure 2 (Appendix) is a screen capture of the Resolutions Pro software (labeled for clarity) that displays: (a) a selected spectrum of ethanol in water, (b) a calibration curve based on the prepared solutions, and (c) a customizable spectral spreadsheet of important spectral parameters. Solutions in the range 1.0 5.0 %v/v EtOH/H2O were used and the known concentration value for each solution was entered into Resolutions Pro software (Figure 2c, column 4 of spectral spreadsheet). The spectra that are included in the calibration are listed in column 3 of the spreadsheet. Multiple replicates for each concentration can also be included. A data point in the calibration curve can be included/excluded from the calibration by doubleclicking on this column to change its status. The calibration can be based on a user-defined peak height or peak area (as in this case), which can be changed by simply dragging the handles at the left and right of the peak area 2. Likewise, the baseline definition can be changed by simply dragging the handles to the left and right of the baseline with a live update to the calibration curve. The concentration of ethanol in an unknown solution with the same matrix can be readily determined from the calibration curve. Depending on the experiment, it may be advantageous to include solutions of known concentrations periodically to verify the robustness of the model as well as to ensure the quality of the predictions of unknown concentrations 3. The calibration curve can be conveniently stored for later use, meaning that a new set of calibration solutions does not need to be prepared each time a determination is required (Figure 3). 2

Absorbance this translated to an LOD of 0.04% on a Cary 670 FTIR and 0.08% on a Cary 660 FTIR. Figure 3. The simplicity of storing a calibration curve (and associated spectra) permits the determination of unknown concentrations of relevant samples at a later date Using this method, the concentration of a solution, made up to be 0.500% v/v EtOH/H2O (see materials and reagents) was determined to be 0.504% v/v EtOH/H2O from the average of three replicate measurements. This represents an accuracy of better than 0.5%, demonstrating the performance of the Agilent Cary 670 FTIR spectrometer, even at low concentrations of solute. Limit of Detection - Agilent Cary 660 FTIR with a PIKE MIRacle Diamond ATR Accessory Instrument sensitivity can be defined by the limit of detection (LOD). The LOD is normally defined as the point at which the peak height of a sample is three times that of the peak-to-peak (p-p) noise in an adjacent region. This is a better representation of instrument sensitivity than a S/N specification number, as it provides practical experimental conditions based on measurements in the spectral region of interest. For instruments incorporating an ATR accessory this can be quoted as % volume of solute divided by % volume of solvent, in this case, solution of ethanol (solute) diluted in water (solvent), %(v/v) EtOH/H2O. To validate the above calculations, a solution of 0.04% v/v EtOH/H2O was made up and measured on an Agilent Cary 670 FTIR spectrometer. Figure 4 shows the IR spectrum overlaid with a spectrum of a 1.0% v/v EtOH/H2O solution for reference. Even at this very low concentration the C-O stretch of ethanol at 1045 cm -1 is visible. The height of this peak was compared with the p-p noise in the region 1220 1120 cm -1 using the Noise application within the Resolutions Pro software package (Table 2). This data shows the peak height to be three times that of the noise in the adjacent region, indicating the limit of detection for this spectrometer/atr configuration. 0.0036 0.0030 0.0024 0.0018 0.0012 0.0006 0.0000 Figure 4. Overlaid spectra of 1.0% (v/v ) and 0.04% (v/v) EtOH/H2O acquired using a Agilent Cary 670 FTIR spectrometer with a PIKE MIRacle ATR showing the C-O stretch of EtOH at 1045cm -1 Table 2. Measured LOD for two Agilent Cary 600 FTIR series spectrometers configured with a PIKE MIRacle Diamond ATR accessory LOD % v/v EtOH/H O Peak Height 1045 cm 1.0 % 0.03 0.04% % P-P Noise 1220-1120 cm Cary 670 FTIR 0.04 2.5 x 10-4 abs 0.8 x 10-4 abs Cary 660 FTIR 0.08 4.6 x 10-4 abs 1.4 x 10-4 abs 1044.045 1280 1240 1200 1160 1120 1080 1040 1000 960 Wavenumber (cm -1 ) The p-p noise of water was measured between 1220 1120 cm -1 on an Agilent Cary 660 FTIR and Cary 670 FTIR spectrometer (Table 2). Using the commonly accepted definition of LOD, a sample of EtOH in H2O is detectable if it has a peak height ~3 times the p-p noise. Using the calibration curve previously collected, 3

Conclusion The PIKE MIRacle ATR accessory, in conjunction with an Agilent Cary 670 FTIR or 660 FTIR spectrometer achieves a limit of detection of 0.04% and 0.08% v/v EtOH/H2O respectively. These results reflect the high energy throughput of the Agilent Cary 600 FTIR series of FTIR spectrometers and demonstrate the instruments to be a powerful tool in the identification and quantitation of even the most dilute samples. The high energy throughput means that the performance of the instrument is maintained even when using energy limiting accessories such as an ATR or when analyzing highly attenuating samples. This permits faster spectral acquisition times as the number of co-added scans required to meet signal-tonoise requirements is significantly reduced. Coupled with the convenience of using an ATR accessory, where little or no sample preparation is required, the Agilent Cary 600 FTIR series allows users to obtain high-quality spectra faster than ever before. In addition, the Quant Analysis application within the Resolutions Pro software package offers a simple, convenient and accurate means of quantifying the concentration of infrared-active compounds. References 1. Boyd S. and Kansiz M. Highest Available Signal- To-Noise Performance, Delivering Superior Sensitivity and Analytical Performance, Advantage Note SI-01353 2. Resolutions Pro Software Help files. 3. Harris Daniel C., Quantitative Chemical Analysis, W. H. Freeman, 2006, ISBN 071677041 4

Appendix A B C Figure 2. Resolutions Pro software illustrating the Quant Calibration application. Agilent s FTIR software simplifies quantitative analysis 5

www.agilent.com/chem Agilent Technologies, Inc., 2008, 2011 Published March, 2011 Publication Number si-01374