Application Note: Absorbance
|
|
|
- Muriel Bernice Burns
- 10 years ago
- Views:
Transcription
1 Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light attenuates exponentially as it passes through clear materials or solutions. So, naturally the reduction of transmitted light is exponentially related to the concentration of the sample and path length of light traveled (Figure 1A). %T = [(Transmitted Light) / (Incident Light)] * A 3 1B % Transmittance Log 1 (%T) Fig. 1A) % Light transmitted per Unit, 1B) relationship between absorbance and %T Beer-Lambert Law Beer-Lambert Law relates the transmittance of light to absorbance by taking the negative logarithmic function, base 1, of the transmittance observed by a sample, which results in a linear relationship to the intensity of the absorbing species and the distance traveled by light. = 2 - Log 1 (T) In short, Beer-Lambert Law states that the absorbance of a sample is directly proportional to the path length of the sample holder and the concentration of the sample. Incident Light 1 cm Light Path Transmitted light = 2 Log 1 (T) LED Filter Sample Detector Light Cuvette Source Figure 2. Absorption measurement scheme S-75 Revision A Page 1 of 5
2 Units Range and detection limits is measured in absorbance units (Au), which relate to transmittance as seen in figure 1. For example, 1.Au is equal to 1% transmittance, 2.Au is equal to 1% transmittance, and so on in a logarithmic trend. Because transmittance inherently decreases logarithmically, linearity breaks down at higher concentrations due to sample saturation (Figure 3). Since the transmitted light at 2.Au is less than 1%, any increases in concentration will reflect only slight changes in absorbance. Generally, you will never reach an absorbance of 4.Au and still maintain a linear relationship. The same applies for lower end limits. Samples being measured for absorption, which are primarily clear solutions or close to zero, are subject to the errors associated with light refraction, reflection, or scattering. For example, scattering can decrease the transmission of light through a sample and will effect the absorbance even though no absorption is occurring. These kinds of errors can be reduced in turn maximizing resolution through particle removal, which minimizes scattering of light and using glass cuvettes, which minimizes absorption of light. Applications micromolar Concentration Figure 3. Theoretical response curve of absorbances at increasing concentrations measurements can be used to measure the concentrations in different ways such as using absorption coefficients with pigment calculations or calibration curves as with nutrient analyses. For either method used the basic principal is to create a correlation factor between absorbance and concentration to be used for determination of absorbing species in natural samples. S-75 Revision A Page 2 of 5
3 Pigments Pigment concentrations are calculated by taking the absorbance reading of the largest peak in a pure pigment spectrum, as characterized by a spectrophotometer, and dividing that absorbance value by the pigments specific absorption or molar absorption coefficients. Units (Au) Concentration of Pigment (mg/l) = * 1 Specific Absorption Coefficient (L g -1 cm -1 ) Pigment Algal Group Represented Specific Absorption Coefficient (L g -1 cm -1 ) Units (Au) Calculated Concentrations (mg/l) Chlorophyll a All Algae ( = 664nm) 2.87 Fucoxanthin Diatoms ( = 443nm) 1.9 Chlorophyll b Green Algae ( = 646nm) 6.91 Table 1: Example of spectrophotometric determination of pure pigment concentrations using specific absorption coefficients (Jeffrey et al 1997) for Chlorophyll a, b, and Fucoxanthin. The derivations of these coefficients require extremely careful and highly analytical techniques but the end calculation is very simple. To obtain a specific absorption coefficient of a pigment, one would need to purify the pigment from all contaminants and either dry or crystallize the molecule. After purification, the resulting material needs to be dissolved in solvent and measured for absorbance. Keep in mind that absorbance measurements are solvent specific as well. The specific absorption coefficient obtained from a spectrum of the pigment would serve as the universal standard for quantifying this specific pigment s concentration from any spectrophotometric measurement. Specific absorption coefficients of pigments are usually given in volume per weight with respect to path length of the cuvette used (i.e. liter gram -1 centimeter -1 ). Nutrients Nutrient concentrations are simpler calculations and less tedious. The simplicity of nutrient calculations comes from a basic understanding of algebraic expressions. A calibration curve is produced, which relates the absorbance to concentration of absorbing species. S-75 Revision A Page 3 of 5
4 Concentration (um) Concentration (um) 6 1uM - 5uM y = x R 2 = uM - 7uM y = x R 2 = Regression equation. R 2 value showing good linear relationship Units (Au) Units (Au) Figure 4: Example of established linear correlation between concentration of absorbing species and absorbance readings (Au) for 2 different working ranges. The equation of the regression can be used to calculate the concentrations of unknown samples by simply plugging in X values that fall within your working range. The equation of the curve is used to calculate natural sample concentrations as long as they fit within the range of the curve (Fig. 4). Accurate concentration values can be achieved from absorbance measurements if a couple of important points are kept in mind: 1) Choosing a range of absorbance units and establishing a linear calibration curve for that range 2) If measuring samples outside the range your calibration curve, making sure your samples are diluted to fit the range calibrated 3) Following procedure for correct preparation of reagents 4) Avoid cross contamination among samples 5) Keeping the optical path constant (i.e. don t change cuvettes sizes while analyzing samples) 6) Blank correcting by running a reagent blank 7) Minimizing transmission loss, due to particle scattering, by centrifugation or other means of particle removal 8) Minimizing transmission loss by using sample holders that will not absorb of light 9) Plugging x, the absorbance value from an instrument, into the regression equation, Y = mx+b will provide you with an accurate estimate of your absorbing species if you have established a good linear relationship as determined by the R 2 value or your regression (Note: Accuracy of estimates increase as your R 2 value increases) S-75 Revision A Page 4 of 5
5 References: S. W. Jeffrey, Mantoura R. F. C. and Wright S. W., Phytoplankton pigments in oceanography. UNESCO Publishing France, Paris J. D. H. Strickland and T. R. Parsons, A practical handbook of seawater analysis. Second Edition, Bulletin 167. Fisheries Research Board of Canada, Ottawa Timothy R. Parsons, Maita Y. and Lalli C. M., A manual of chemical and biological methods for seawater analysis. First Edition. Pergamon Press Ltd., Great Britain Brian M. Tissue, On-line Educational Hypermedia ( hypermedia for science education). Science Media Ltd. Irvine, Ca. S-75 Revision A Page 5 of 5
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
Measuring Protein Concentration through Absorption Spectrophotometry
Measuring Protein Concentration through Absorption Spectrophotometry In this lab exercise you will learn how to homogenize a tissue to extract the protein, and then how to use a protein assay reagent to
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.
EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare
Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law
Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine
Lab #11: Determination of a Chemical Equilibrium Constant
Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a
Using the Spectrophotometer
Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to
Chemistry 111 Lab: Intro to Spectrophotometry Page E-1
Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 SPECTROPHOTOMETRY Absorption Measurements & their Application to Quantitative Analysis study of the interaction of light (or other electromagnetic
Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs
Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
2 Spectrophotometry and the Analysis of Riboflavin
2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound
where h = 6.62 10-34 J s
Electromagnetic Spectrum: Refer to Figure 12.1 Molecular Spectroscopy: Absorption of electromagnetic radiation: The absorptions and emissions of electromagnetic radiation are related molecular-level phenomena
How to measure Chlorophyll-a
World Bank & Government of The Netherlands funded Training module # WQ - 40 How to measure Chlorophyll-a New Delhi, March 2000 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New Delhi 11 00 16
Upon completion of this lab, the student will be able to:
1 Learning Outcomes EXPERIMENT B4: CHEMICAL EQUILIBRIUM Upon completion of this lab, the student will be able to: 1) Analyze the absorbance spectrum of a sample. 2) Calculate the equilibrium constant for
Colorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
COLORIMETER. Description 0358BT. Figure 1. The Colorimeter
COLORIMETER Description 0358BT Figure 1. The Colorimeter Introduction The Colorimeter is designed to determine the concentration of a solution by analyzing its color intensity. Monochromatic light from
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES
UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this
Reaction Stoichiometry and the Formation of a Metal Ion Complex
Reaction Stoichiometry and the Formation of a Metal Ion Complex Objectives The objectives of this laboratory are as follows: To use the method of continuous variation to determine the reaction stoichiometry
Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument
Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument Technical Note INTRODUCTION Direct measurements of nucleic acid samples at OD 260 or protein samples at OD
Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:
Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T
MCAL Spectrophotometry. Spectrophotometry
MCAL Spectrophotometry Instruments include: Cary 50 UV-vis Spectrophotometer Eclipse Spectrofluorometer HPLC Diode Array and Fluorescence ICP-OES with CCD detection Spectrophotometry The instruments all
Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer. NAME: Lab Section: Date: Sign-Off:
Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer NAME: Lab Section: Date: Sign-Off: Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer Introduction Weak acids,
An Introduction to Algae Measurements Using In Vivo Fluorescence
An Introduction to Algae Measurements Using In Vivo Fluorescence Submersible fluorescence sensors enable real-time field estimates of phytoplankton that can be directly correlated to standard laboratory
Chem 131A: Absorbance of Riboflavin
Chem 131A: Absorbance of Riboflavin Purpose: The purpose of this experiment is to: 1) Familiarize the student with the use of the HP 8452 diode array spectrophotometer, 2) examine the limitations of the
2 Absorbing Solar Energy
2 Absorbing Solar Energy 2.1 Air Mass and the Solar Spectrum Now that we have introduced the solar cell, it is time to introduce the source of the energy the sun. The sun has many properties that could
Electromagnetic Radiation (EMR) and Remote Sensing
Electromagnetic Radiation (EMR) and Remote Sensing 1 Atmosphere Anything missing in between? Electromagnetic Radiation (EMR) is radiated by atomic particles at the source (the Sun), propagates through
A Beer s Law Experiment
A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40
Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document
Nucleic Acid Purity Assessment using A 260 /A 280 Ratios
Nucleic Acid Purity Assessment using A 260 /A 280 Ratios A common practice in molecular biology is to perform a quick assessment of the purity of nucleic acid samples by determining the ratio of spectrophotometric
INTERNATIONAL OLIVE COUNCIL
INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: [email protected]
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
CHEM 161: Beer s Law and Analysis of a Sports Drink
CHEM 161: Beer s Law and Analysis of a Sports Drink Introduction Although sunlight appears white, it contains a spectrum of colors. A rainbow actually shows this range of colors in visible light: violet,
2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY
2nd/3rd Year Physical Chemistry Practical Course, Oxford University 2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY (4 points) Outline Spectrometry is widely used to monitor the progress
Reaction of Blue Food Dye with Bleach
Exercise 2 Reaction of Blue Food Dye with Bleach 2 Introduction In the experiment, you will study the rate of the reaction of FD&C Blue #1 (Blue #1 is denoted by E number E133 in food stuff) with sodium
Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.
Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Introduction: The determination of protein concentration is frequently required in biochemical work. Several methods
Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry
Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry General Medicine Jiřina Crkovská (translated by Jan Pláteník) 2010/2011 1 Spectrophotometry is one of the most widely used instrumental
Beer's Law: Colorimetry of Copper(II) Solutions
Exercise 11 Page 1 Illinois Central College CHEMISTRY 130 Name: Beer's Law: Colorimetry of Copper(II) Solutions Objectives In this experiment, we will use Beer's Law to determine the unknown concentrations
The Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant, K
Copyright 1999 2010 by Mark Brandt, Ph.D. 12
Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits
INTRODUCTION TO SPECTROMETRY
Experiment 11 INTRODUCTION TO SPECTROMETRY Portions adapted by Ross S. Nord and Colleagues, Chemistry Department, Eastern Michigan University from ANAL 043, written by Donald F. Clemens and Warren A. McAllister,
Experiment C-31 Color Absorption
1 Experiment C-31 Color Absorption Objectives To understand the concepts of light waves and color. To investigate how red, green and blue liquids absorb light of different wavelengths. To learn about colorimeter
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
Infrared Spectroscopy: Theory
u Chapter 15 Infrared Spectroscopy: Theory An important tool of the organic chemist is Infrared Spectroscopy, or IR. IR spectra are acquired on a special instrument, called an IR spectrometer. IR is used
Molecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student. LBS 171L Section 9 TA: Dana October 27, 2005
How uch Phosphate is the Body Being Exposed to During a Lifetime by Showering? Lab 5: Quantitative Analysis- Phosphates in Water By: A Generous Student LBS 171L Section 9 TA: Dana October 7, 005 [Note:
Overview. What is EMR? Electromagnetic Radiation (EMR) LA502 Special Studies Remote Sensing
LA502 Special Studies Remote Sensing Electromagnetic Radiation (EMR) Dr. Ragab Khalil Department of Landscape Architecture Faculty of Environmental Design King AbdulAziz University Room 103 Overview What
Solubility Product Constants
Solubility Product Constants PURPOSE To measure the solubility product constant (K sp ) of copper (II) iodate, Cu(IO 3 ) 2. GOALS 1 To measure the molar solubility of a sparingly soluble salt in water.
Studying Photosynthetic Electron Transport through the Hill Reaction
Studying Photosynthetic Electron Transport through the Hill Reaction In this lab exercise you will learn techniques for studying electron transport. While our research system will be chloroplasts and photosynthetic
Lab 2. Spectrophotometric Measurement of Glucose
Lab 2 Spectrophotometric Measurement of Glucose Objectives 1. Learn how to use a spectrophotometer. 2. Produce a glucose standard curve. 3. Perform a glucose assay. Safety Precautions Glucose Color Reagent
The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD
TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic
Phenolphthalein-NaOH Kinetics
Phenolphthalein-NaOH Kinetics Phenolphthalein is one of the most common acid-base indicators used to determine the end point in acid-base titrations. It is also the active ingredient in some laxatives.
Symmetric Stretch: allows molecule to move through space
BACKGROUND INFORMATION Infrared Spectroscopy Before introducing the subject of IR spectroscopy, we must first review some aspects of the electromagnetic spectrum. The electromagnetic spectrum is composed
Determination of the Mass Percentage of Copper in a Penny. Introduction
Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny
Measuring Manganese Concentration Using Spectrophotometry
Measuring Manganese Concentration Using Spectrophotometry Objectives To use spectroscopy to determine the amount of Manganese is an unknown sample. Scenario Your have just joined a "Green Team" at SMC
EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
Colorimetry Extinction coefficient (ε) Lambda max (λ max ) Qualitative vs. quantitative analysis
Lab Week 2 - Spectrophotometry Purpose: Introduce students to the use of spectrophotometry for qualitative (what is it) and quantitative (how much is there of it) analysis of biological samples and molecules.
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES
ZETA POTENTIAL ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this document
Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information
1 Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information Instructions for the Operation of the Cary 300 Bio UV-Visible Spectrophotometer See the Thermo OMNIC Help reference on page 49. Ultraviolet-Visible
Name Class Date. spectrum. White is not a color, but is a combination of all colors. Black is not a color; it is the absence of all light.
Exercises 28.1 The Spectrum (pages 555 556) 1. Isaac Newton was the first person to do a systematic study of color. 2. Circle the letter of each statement that is true about Newton s study of color. a.
Stage 1 Desired Results
Lesson Title: Phosphate Testing Contract Discipline Focus: Chemistry Grade level: 11-12 Length of lesson: 1.5 class periods Content Standard(s): Stage 1 Desired Results Understanding (s)/goals Students
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
UV-Visible Spectroscopy
UV-Visible Spectroscopy UV-Visible Spectroscopy What is UV-Visible Spectroscopy? Molecular spectroscopy that involves study of the interaction of Ultra violet (UV)-Visible radiation with molecules What
www.gbo.com/bioscience
Application Note UV/VIS Spectroscopy in Microplates UV-Star, µclear, MICROLON and CELLSTAR 1. Introduction to UV/VIS spectroscopy Transparent microplates (CELLSTAR, MICROLON ) or microplates with a clear
Ultraviolet Spectroscopy
Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer
- particle with kinetic energy E strikes a barrier with height U 0 > E and width L. - classically the particle cannot overcome the barrier
Tunnel Effect: - particle with kinetic energy E strikes a barrier with height U 0 > E and width L - classically the particle cannot overcome the barrier - quantum mechanically the particle can penetrated
How do scientists prepare solutions with specific concentrations of solutes?
EXERCISE 2 Name How do scientists prepare solutions with specific concentrations of solutes? Objectives After completing this exercise, you should be able to: define and correctly use the following terms:
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014
Laboratory #3 Guide: Optical and Electrical Properties of Transparent Conductors -- September 23, 2014 Introduction Following our previous lab exercises, you now have the skills and understanding to control
5.33 Lecture Notes: Introduction to Spectroscopy
5.33 Lecture Notes: ntroduction to Spectroscopy What is spectroscopy? Studying the properties of matter through its interaction with different frequency components of the electromagnetic spectrum. Latin:
Where does it come from?
What is Phosphorus? Phosphorus (P) is a nutrient that is vital to human, animal, and plant growth. It s one of the most common substances found in nature. It s found in our water, our food, and our bodies.
Lab 2 Biochemistry. Learning Objectives. Introduction. Lipid Structure and Role in Food. The lab has the following learning objectives.
1 Lab 2 Biochemistry Learning Objectives The lab has the following learning objectives. Investigate the role of double bonding in fatty acids, through models. Developing a calibration curve for a Benedict
Biology 309 Lab Notebook
Name: Biology 309 Lab Notebook This is a guided lab notebook for you to keep well-organized notes about procedures and record experimental data for experiments as they are performed. It is guided because,
High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form
ISSN: 0973-4945; CODEN ECJHAO E- Chemistry http://www.e-journals.net Vol. 5, No.3, pp. 427-430, July 2008 High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage
QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1
Experiment 13H FV 1/25/2011(2-run) THE REACTION OF RED FOOD COLOR WITH BLEACH 1 PROBLEM: Determine the rate law for the chemical reaction between FD&C Red Dye #3 and sodium hypochlorite. LEARNING OBJECTIVES:
High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering
High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering Power spectrum development with heterodyne technology advances biotechnology and nanotechnology measurements. M. N. Trainer
What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher)
What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher) Introduction: There is more to a color than a name. Color can tell us lots of information. In this lab you will use a spectrophotometer
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
Project 2B Building a Solar Cell (2): Solar Cell Performance
April. 15, 2010 Due April. 29, 2010 Project 2B Building a Solar Cell (2): Solar Cell Performance Objective: In this project we are going to experimentally measure the I-V characteristics, energy conversion
Hellma. Where precision becomes an art. Calibration standards for spectrophotometers. 2008 by Hellma www.hellma-worldwide.com
Hellma. Where precision becomes an art. Calibration standards for spectrophotometers 2008 by Hellma www.hellma-worldwide.com Contents Calibration of spectrophotometers 3 Traceability of the calibration
DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED
DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED Abstract: There are a number of sources of information that give a mathematical description of the terms used in light scattering. However, these will not
Introduction: Growth analysis and crop dry matter accumulation
PBIO*3110 Crop Physiology Lecture #2 Fall Semester 2008 Lecture Notes for Tuesday 9 September How is plant productivity measured? Introduction: Growth analysis and crop dry matter accumulation Learning
BET Surface Area Analysis of Nanoparticles
OpenStax-CNX module: m38278 1 BET Surface Area Analysis of Nanoparticles Nina Hwang Andrew R. Barron This work is produced by OpenStax-CNX and licensed under the Creative Commons Attribution License 3.0
Quantifying Bacterial Concentration using a Calibrated Growth Curve
BTEC 4200 Lab 2. Quantifying Bacterial Concentration using a Calibrated Growth Curve Background and References Bacterial concentration can be measured by several methods, all of which you have studied
Biomedical Optics Theory
Introduction Biomedical Optics Theory Diffuse reflectance spectroscopy (DRS) and Laser Doppler Flowmetry (LDF) are booth optical techniques that can quantify a number of microcirculatory parameters. Prof
Light Backscatter Applications in Milk and Dairy Foods Processing
Light Backscatter Applications in Milk and Dairy Foods Processing Fred Payne a, Manuel Castillo b, Mary-Grace Danao c a Univ. of Kentucky, Biosystems and Agricultural Engineering, Lexington, KY, USA ([email protected])
EDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
Title: Standard Operating Procedure for Thermo-Miran SapphIRe portable gas analyzer
Procedure No: SOP-039 Revision No: 1.1 January 21, 2011 Page No.: 1 of 11 1. INTRODUCTION AND SCOPE This document will outline specifics with regards to portable sampling analysis utilizing and operating
OPTICAL FIBERS INTRODUCTION
OPTICAL FIBERS References: J. Hecht: Understanding Fiber Optics, Ch. 1-3, Prentice Hall N.J. 1999 D. R. Goff: Fiber Optic Reference Guide (2 nd ed.) Focal Press 1999 Projects in Fiber Optics (Applications
The photoionization detector (PID) utilizes ultraviolet
Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).
4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)
October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for
Supporting Information. Rational design of Au nanorods assemblies for highly sensitive and selective SERS detection of prostate specific antigen
Electronic Supplementary Material (ESI) for RSC Advances. This journal is The Royal Society of Chemistry 2015 Supporting Information Rational design of Au nanorods assemblies for highly sensitive and selective
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy
2-1 Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy Introduction: The kinetics of a decomposition reaction involving hydroxide ion and crystal violet, an organic dye used
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
Application Note Noise Frequently Asked Questions
: What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random
NNIN Nanotechnology Education
NNIN Nanotechnology Education Student Guide Part 1: Silver Nanoparticle Synthesis and Spectroscopy Introduction: In this lab you will synthesize silver nanoparticles one of the most commonly used nanoparticles
Coordination Compounds with Copper (II) Prelab (Week 2)
Coordination Compounds with Copper (II) Prelab (Week 2) Name Total /10 SHOW ALL WORK NO WORK = NO CREDIT 1. What is the purpose of this experiment? 2. Write the generic chemical formula for the coordination