1 Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with enriched isotope spike of analyte
2
3
4 Isotope Dilution The amount of spike is selected so that the resulting ratio between spiked isotope and unspiked isotope is near unity maximizes precision Typically use the most abundant isotope as the reference -- maximizes sensitivity
5 Isotope Dilution Check isotope ratio in unspiked sample to determine if the natural ratio in the sample matches with the predicted ratio If not -- interference in acting on one or both of the isotopes Always attempt to use interference free isotopes
6 Isotope Dilution Prepare the spike to desired concentration Add spike as early as possible after equilibration of spike and sample you don t have to have complete sample recovery During any stage of the process complete equilibration is absolutely necessary
7 Isotope Dilution Analyze the solution on the ICP using many repetitive scans (to maximize precision) Need to measure isotopic ratios on standards of a known ratio in order to correct for machine mass discrimination Use previous equation to calculate concentrations!
8 Advantages: Sample Analysis Design Isotope Dilution Most accurate and precise method for quantitative elemental concentrations Partial loss of analyte during preparation is compensated for since physical and chemical interferences are not an issue -- will cancel out as they will affect each isotope identically Ideal form of internal standardization since another isotope of the same element is used in this capacity
9 Isotope Dilution Disadvantages: Generally only applicable to multiple-isotopic elements Need an enriched isotope spike for the analyte of interest - not always available or sometimes at very high cost Need two interference free isotopes VERY time consuming
10 Sample Analysis Design STEP 3 INTERNAL STANDARDIZATION & INSTRUMENT DRIFT CORRECTION
11 Internal Standard Every sample should be analyzed with an internal standard (IS) What is an internal standard (IS)? element that is added to EVERY sample/ blank/ calibration standard/qa sample/etc., that is not expected to be in the sample in appreciable quantities and is not an element of interest use IS to monitor machine drift (both short and long term) and matrix effects
12 Internal Standard Choice of IS depends upon which elements you are quantifying The IS should have similar properties in the plasma as element(s) of interest ICP-MS: similar in mass/ionization potential
13 Example: Sample Analysis Design Internal Standard attempting to quantify U - use Th attempting to quantify most transition metals - use As attempting to quantify REEs - use Re 115 In and 103 Rh are common IS for general use alternatively, you can add several IS to each sample
14 Internal Standard From previous slide, we assume that samples have little or no Th, As, or Re It s important to have an idea of what s in your sample prior to quantitative analysis Solid samples can use a naturally occurring element as IS, provided that you know the concentration in each sample
15 Procedure for IS use: Sample Analysis Design Internal Standard Calculate the concentration of the IS in each centrifuge tube the latter will contain an aliquot of your sample and an aliquot of the IS Divide the measured ion signal (CPS) by the concentration of your IS to derive the factor = CPS/ppb Divide CPS/ppb of each tube by the CPS/ppb for those measured for the blanks since these are not influenced by possible effects due to sample matrices The latter yields a dimensionless correction factor (I refer to it as a normalization factor) Use correction factor to adjust analyte counts for drift or matrix effects
16
17
18 Advantages: Sample Analysis Design Internal Standard Fluctuations are monitored in each sample/ calibration / blank Disadvantages: Assume that behavior of IS is the same as the analyte
19 Instrumental Drift Correct for instrument drift with: Internal standardization is a common procedure Use of drift corrector solutions (DCS)
20 Instrumental Drift Drift Corrector Solutions (DCS): Measure the same solution intermittently throughout the course of the analytical session Change in ion signal is assumed to be linear between each DCS measurement
21 Instrumental Drift The DCS should contain all elements of interest and can be matrix matched to samples Example: use standard reference materials (SRMs) for DCS
22 Instrumental Drift Apply a linear correction to samples between DCS solutions DCS 1 + ((DCS 2 - DCS 1 )*F) F = position dependent fraction
23 Instrumental Drift Advantages of DCS correction: all analytes are monitored for drift nothing added to sample solutions Disadvantages of DCS correction: assume change is linear cannot easily monitor matrix effects
24 Background & blanks Standard blank - blank used to monitor polyatomic ion interferences, gas peaks, and contamination from reagents; used for background subtraction Procedural blank - blank used to monitor contamination acquired during all stages of sample preparation; grinding, digestion, acidification, powdering, etc
25 Background & blanks Use of blanks during an analytical session: ALWAYS begin an analytical session with at least one standard blank Analyze standard blanks periodically throughout the course of the session in particular to monitor memory effects Process and analyze at least one procedural blank at some point during your research study; for its analysis, it s preferable to measure it early in order to avoid any potential memory effects
26 Background & blanks The more standard blanks that are run during an analytical session, the more information you will have with regards to monitoring change(s) in background levels throughout the entire session
27 Background & blanks How to determine the background : 1. just use the first standard blank 2. average all standard blanks 3. take median of all standard blanks 4. apply statistical analysis to standard blanks and select some of them
28 Outlier tests: Sample Analysis Design Background & blanks 1. I know the truth 2. Looks different 3. Statistical proof
29 Background & blanks Option 1 should be avoided - unscientific and invalid Option 2 is better but only if the measurement is repeated Option 3 is the best approach, but needs to be carried out carefully in order to avoid false negatives and positives
30 Background & blanks Huber Outlier Test take median of all values calculate absolute deviation x i - x m take mean of absolute deviations (MAD) multiply MAD by coefficient (k = 3-5) anything higher than k*mad is rejected as outlier
31 Background & blanks Calculation of Limit of Detection (DL) and Limit of Quantification (QL) Easy way: LOD = 3*STDEV blank ; LOQ = 10*STDEV blank
32 SUMMARY A good analytical method will: 1. provide the means to calculate an accurate background level 2. allow for correction of instrument drift 3. use Internal standardization to monitor matrix effects 4. provide some method for monitoring/ correcting interferences 5. Use a proper calibration strategy
33
34
35
36 Sample Analysis Design PART II
37 Sample Analysis Design Generating high quality, validated results is the primary goal of elemental abundance determinations It is absolutely critical to plan an ICP-MS analysis carefully from sample gathering to final analysis on the ICPMS Always create and follow an analysis design that shall permit you to follow the procedure for future samples, AND know what you did when looking at old data
38 Sample Analysis Design Goal: high quality, validated, quantitative determination of elemental concentrations/isotope ratios This goal can be achieved by: Specificity - only detecting isotope(s) of interest, not interferences, OR making sure that matrix effects do not play a role Sensitivity - can we differentiate the isotope of interest from the background signal? Accuracy - does the analysis represent the real value? Precision - how repeatable is the value?
39
40 Sample Analysis Design Sensitivity and Precision: generally defined by how well the instrument is tuned (assuming dilution factor is appropriate) Specificity and Accuracy: generally governed by how well you prepare your sample and set up your analysis method/ parameters
41 Sample Analysis Design Spectroscopic interferences Matrix effects
42 Sample Analysis Design Interference analytical artifact which causes the ion signal in a sample to vary from an idealized (true) signal Interferences cause inaccurate analyses Two main types: 1. Spectroscopic 2. Nonspectroscopic
43 Sample Analysis Design Spectroscopic Interferences in ICP-MS 1. Isobaric overlap 2. Polyatomic ions 3. Refractory oxides 4. Doubly charged ions
44 Sample Analysis Design Isobaric Overlap Two different elements having the same nominal mass E.g., 40 Ca (96.9%) and 40 Ar (99.6%) Monoisotopic elements: B, Na, Al, Sc, Mn, As, Nb, Y, Rh, I, Cs, Pr, Tb, Ho, Tm, Au, Th None have isobaric interferences
45 Isobaric Overlap A few multi-isotope elements have no interference free isotopes: E.g., In In 113 Cd, and 115 In 115 Sn Usually possible to choose an interference free isotope or one that is unlikely to be interfered upon strongly (either due to very low natural abundance of interfering element (isotope) or the interfering element (isotope) is not present in sample)
46 Isobaric Overlap IF an isobaric interference is unavoidable, you can correct for it by measuring the counts on a non-interfering isotope of the interfering element CPS analyte = CPS total -CPS interferent *(X a /X b ) X a = abundance of interfering isotope X b = abundance of non-interfering isotope
47 Polyatomic Interferences More serious than isobaric interferences Result from possible, short-lived combination of atomic species in the plasma or during ion transfer Common recombinants are Ar, H, and O Dominant elements in reagents also form polyatomic interferences - N, S, and Cl
48 Polyatomic Interferences Analyzing deionized water with ICP-MS instrument we have solely H and O present (within the matrix ) Thus, peaks would be visible at masses: 41 ( 40 Ar + 1 H) 41 K 56 ( 40 Ar + 16 O) 56 Fe 80 ( 40 Ar + 40 Ar) 80 Se As well as other minor peaks from the minor isotopes of Ar and O
49 Polyatomic Interferences If you acidify the deionized water with HNO 3 or H 2 O 2, these behave in the same manner as deionized water these acids are considered ideal matrices as they don t add unnecessary polyatomic interferences Acidification with HCl or H 2 SO 4 changes the situation since Cl and S can be used to form polyatomic ions
50 HCl matrix: Sample Analysis Design Polyatomic Interferences 35 Cl + 16 O = 51 V 37 Cl + 16 O = 53 Cr 40 Ar + 35 Cl = 75 As 40 Ar + 37 Cl = 77 Se
51 Polyatomic Interferences H 2 SO 4 matrix: 32 S + 16 O = 48 Ti 32 S + 16 O + 16 O, or 32 S + 32 S = 64 Zn, 64 Ni 34 S + 16 O = 50 Ti, 50 V, 50 Cr 33 S + 16 O + 16 O = 65 Cu
52 Sample Analysis Design These interferences will be present just from the gas and solvent Therefore, if possible, solutions should be prepared with a weak (1-5% v/v) HNO 3 acid matrix Why not deionized water? acid helps keep elements from sticking to sides of test tubes and tubing during transport ionization efficiency is significantly lower relative to HNO 3
53 Polyatomic Interferences The most serious polyatomic interferences are formed from the most abundant isotopes of H, C, N, O, Cl, and Ar So, if polyatomic interferences are present in plain deionized water with no other ions present, what happens in natural (complex) samples/matrices?
54 Polyatomic Interferences Need to know the most abundant elements present in sample Is (are) there any one (or more) elements present that are in very high abundances? If so - then it s likely interferences could be formed from such abundant element(s) E.g., rock samples - high Si
55 Polyatomic Interferences Typically, most polyatomic interferences are only present below mass 82 - since Ar, H, and O are by far the most abundant isotopes in the plasma, they form most interferences Counteractive measures: Proper machine settings (especially nebulizer flow rate and RF power) can minimize formation
56 Polyatomic Interferences If a polyatomic interference is unavoidable, then a correction equation like the one for isobaric interferences can be used (assuming you are certain of the species of the interference)
57 Refractory Oxides Refractory Oxides Occur because of incomplete sample dissociation or from recombination Always occur as an interference as an integral value of 16 mass units above the interfering element MO, MO 2, or MO 3 where M is the interfering element
58 Refractory Oxides Elements with high oxide bond strength are most likely to form refractory oxide interferences Severity is expressed as MO/M as a percentage MO/M should be as low as possible, typically <5% MO/M minimized by adjusting nebulizer flow rate, z-axis position, and RF power
59 Refractory Oxides Si, Ce, Zr, Ti, Sm, Mo, and P all form strong oxide bonds and may have severe oxide interferences Usually monitor the CeO/Ce ratio to make sure it s below 5%
60 Examples: Sample Analysis Design Refractory Oxides LREE on the HREE BaO on Eu MoO on Cd
61 Refractory Oxides Again, correction equations can be applied but these lead to increased error, especially if the interference is large Oxide formation tends to not be stable and the corrections accumulate error quickly If there is a severe oxide interference, it s usually better to try to separate the interference from the analyte
62 Typically, Sample Analysis Design Refractory Oxides MO > MO 2 > MO 3 > MO 4
63 Doubly Charged Ions Form when the 2nd ionization potential is less than the first ionization potential of Ar Typical of alkaline earths, a few transition metals, and some REE Low nebulizer flow rates increase doubly charged ion formation Interference signal is always M/2 where M is the element becoming doubly charged
64 How to avoid spectral MS interferences? Attempt to dry solutions prior to introduction to plasma - remove O and H gets rid of many polyatomic species (i.e. use a desolvating introduction system e.g. DSN 100) Optimize instruments settings so that formation is minimized and corrections are not severe Use simple matrices whenever possible Process samples prior to analysis to isolate elements of interest or remove potential interfering elements; e.g. ion exchange chromatography
65 How to avoid spectral MS interferences? Alternatively - use higher mass resolutions, MR medium resolution or HR high resolution Most overlaps are not exact using medium and high resolution modes Can completely separate the isotope of interest from interfering species However, consequence is loss of sensitivity
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
More informationStandard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
More informationChem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with
More informationChem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers
Key Questions & Exercises Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers 1. The atomic weight of carbon is 12.0107 u, so a mole of carbon has a mass of 12.0107 g. Why doesn t a mole of
More informationRegional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
More informationMulti-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
More informationElemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
More informationHow To Use The Nexion 300
APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of
More informationPesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
More informationMass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
More informationNANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
More informationHow To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
More informationCLASS TEST GRADE 11. PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change
CLASS TEST GRADE PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change MARKS: 45 TIME: hour INSTRUCTIONS AND INFORMATION. Answer ALL the questions. 2. You may use non-programmable calculators. 3. You may
More informationMass Spectrometry Signal Calibration for Protein Quantitation
Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of
More informationApplication of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist
Application of ICP-MS as a Tool for Environmental Monitoring Bill Spence Environmental Market Specialist Introduction Knowledge of the concentration of metallic species in the environment has been important
More informationOLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
More informationREVIEW QUESTIONS Chapter 8
Chemistry 101 ANSWER KEY REVIEW QUESTIONS Chapter 8 Use only a periodic table to answer the following questions. 1. Write complete electron configuration for each of the following elements: a) Aluminum
More informationGraphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
More informationB I N G O B I N G O. Hf Cd Na Nb Lr. I Fl Fr Mo Si. Ho Bi Ce Eu Ac. Md Co P Pa Tc. Uut Rh K N. Sb At Md H. Bh Cm H Bi Es. Mo Uus Lu P F.
Hf Cd Na Nb Lr Ho Bi Ce u Ac I Fl Fr Mo i Md Co P Pa Tc Uut Rh K N Dy Cl N Am b At Md H Y Bh Cm H Bi s Mo Uus Lu P F Cu Ar Ag Mg K Thomas Jefferson National Accelerator Facility - Office of cience ducation
More informationEDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
More informationAll answers must use the correct number of significant figures, and must show units!
CHEM 10113, Quiz 2 September 7, 2011 Name (please print) All answers must use the correct number of significant figures, and must show units! IA Periodic Table of the Elements VIIIA (1) (18) 1 2 1 H IIA
More informationELECTRON CONFIGURATION (SHORT FORM) # of electrons in the subshell. valence electrons Valence electrons have the largest value for "n"!
179 ELECTRON CONFIGURATION (SHORT FORM) - We can represent the electron configuration without drawing a diagram or writing down pages of quantum numbers every time. We write the "electron configuration".
More informationElectronegativity and Polarity
and Polarity N Goalby Chemrevise.org Definition: is the relative tendency of an atom in a molecule to attract electrons in a covalent bond to itself. is measured on the Pauling scale (ranges from 0 to
More informationDetermination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
More informationAn example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry
Limits to Detection/Measurement Ionization is fundamentally a Probabilistic Process Just like radioactive decay So is transmission through the analyzer There is an intrinsic statistical uncertainty Proportional
More informationEXPERIMENT 7 Reaction Stoichiometry and Percent Yield
EXPERIMENT 7 Reaction Stoichiometry and Percent Yield INTRODUCTION Stoichiometry calculations are about calculating the amounts of substances that react and form in a chemical reaction. The word stoichiometry
More information3. What would you predict for the intensity and binding energy for the 3p orbital for that of sulfur?
PSI AP Chemistry Periodic Trends MC Review Name Periodic Law and the Quantum Model Use the PES spectrum of Phosphorus below to answer questions 1-3. 1. Which peak corresponds to the 1s orbital? (A) 1.06
More informationVALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol
Page 1 van 36 pages VALIDATION REPORT Investigation of the major interferences for proposed Horizontal ICP AES method Identification code Validation in accordance with Reference method Scope Reference
More informationNET IONIC EQUATIONS. A balanced chemical equation can describe all chemical reactions, an example of such an equation is:
NET IONIC EQUATIONS A balanced chemical equation can describe all chemical reactions, an example of such an equation is: NaCl + AgNO 3 AgCl + NaNO 3 In this case, the simple formulas of the various reactants
More informationASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
More information100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals.
2.21 Ionic Bonding 100% ionic compounds do not exist but predominantly ionic compounds are formed when metals combine with non-metals. Forming ions Metal atoms lose electrons to form +ve ions. Non-metal
More informationDimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
More informationFinnigan ELEMENT2. High Performance High Resolution ICP-MS
m a s s s p e c t r o m e t r y Finnigan ELEMENT2 High Performance High Resolution ICP-MS Analyze Detect Measure Control Finnigan ELEMENT2 High Performance High Resolution ICP-MS Finnigan ELEMENT2 High
More informationEXPERIMENT 4 The Periodic Table - Atoms and Elements
EXPERIMENT 4 The Periodic Table - Atoms and Elements INTRODUCTION Primary substances, called elements, build all the materials around you. There are more than 109 different elements known today. The elements
More information4.2 Bias, Standards and Standardization
4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard
More informationThe Empirical Formula of a Compound
The Empirical Formula of a Compound Lab #5 Introduction A look at the mass relationships in chemistry reveals little order or sense. The ratio of the masses of the elements in a compound, while constant,
More informationVALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
More informationStoichiometry and Aqueous Reactions (Chapter 4)
Stoichiometry and Aqueous Reactions (Chapter 4) Chemical Equations 1. Balancing Chemical Equations (from Chapter 3) Adjust coefficients to get equal numbers of each kind of element on both sides of arrow.
More informationChemistry 122 Mines, Spring 2014
Chemistry 122 Mines, Spring 2014 Answer Key, Problem Set 9 1. 18.44(c) (Also indicate the sign on each electrode, and show the flow of ions in the salt bridge.); 2. 18.46 (do this for all cells in 18.44
More informationSignal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
More informationUniversal Data Acquisition (UDA)
Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte
More informationACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
More informationCopyrighted by Gabriel Tang B.Ed., B.Sc.
Chapter 8: The Periodic Table 8.1: Development of the Periodic Table Johann Dobereiner: - first to discover a pattern of a group of elements like Cl, Br, and I (called triads). John Newland: - suggested
More informationPortable X-ray fluorescence Spectroscopy. Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE
Portable X-ray fluorescence Spectroscopy Michael A. Wilson Research Soil Scientist USDA-NRCS National Soil Survey Center Lincoln, NE OBJECTIVES Background of the method Features of the instrument Applications
More informationChapter 8 Atomic Electronic Configurations and Periodicity
Chapter 8 Electron Configurations Page 1 Chapter 8 Atomic Electronic Configurations and Periodicity 8-1. Substances that are weakly attracted to a magnetic field but lose their magnetism when removed from
More informationFrom Quantum to Matter 2006
From Quantum to Matter 006 Why such a course? Ronald Griessen Vrije Universiteit, Amsterdam AMOLF, May 4, 004 vrije Universiteit amsterdam Why study quantum mechanics? From Quantum to Matter: The main
More informationElectrochemistry - ANSWERS
Electrochemistry - ANSWERS 1. Using a table of standard electrode potentials, predict if the following reactions will occur spontaneously as written. a) Al 3+ + Ni Ni 2+ + Al Al 3+ + 3e - Al E = -1.68
More informationICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
More informationChapter 12: Oxidation and Reduction.
207 Oxidation- reduction (redox) reactions Chapter 12: Oxidation and Reduction. At different times, oxidation and reduction (redox) have had different, but complimentary, definitions. Compare the following
More informationFind a pair of elements in the periodic table with atomic numbers less than 20 that are an exception to the original periodic law.
Example Exercise 6.1 Periodic Law Find the two elements in the fifth row of the periodic table that violate the original periodic law proposed by Mendeleev. Mendeleev proposed that elements be arranged
More informationChapter 11. Electrochemistry Oxidation and Reduction Reactions. Oxidation-Reduction Reactions. Oxidation-Reduction Reactions
Oxidation-Reduction Reactions Chapter 11 Electrochemistry Oxidation and Reduction Reactions An oxidation and reduction reaction occurs in both aqueous solutions and in reactions where substances are burned
More information3 - Atomic Absorption Spectroscopy
3 - Atomic Absorption Spectroscopy Introduction Atomic-absorption (AA) spectroscopy uses the absorption of light to measure the concentration of gas-phase atoms. Since samples are usually liquids or solids,
More informationCHAPTER 8 PRACTICE TEST QUESTIONS (END OF CHAPTER 7 TOO)
CHAPTER 8 PRACTICE TEST QUESTIONS (END OF CHAPTER 7 TOO) Information that most likely will be on the front cover of your exam: h i Z 2 ΔE = @ 2.18 x 10 @ 18 f Z 2 f J j @ k n f 2 n i 2 1. Which of the
More informationEnvironmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
More informationMODERN ATOMIC THEORY AND THE PERIODIC TABLE
CHAPTER 10 MODERN ATOMIC THEORY AND THE PERIODIC TABLE SOLUTIONS TO REVIEW QUESTIONS 1. Wavelength is defined as the distance between consecutive peaks in a wave. It is generally symbolized by the Greek
More informationAnalysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
More informationB) atomic number C) both the solid and the liquid phase D) Au C) Sn, Si, C A) metal C) O, S, Se C) In D) tin D) methane D) bismuth B) Group 2 metal
1. The elements on the Periodic Table are arranged in order of increasing A) atomic mass B) atomic number C) molar mass D) oxidation number 2. Which list of elements consists of a metal, a metalloid, and
More informationORTEC DET-SW-UPG. Latest Software Features. Ease of Use. Source Location with the Detective V3 Software
ORTEC DET-SW-UPG Latest Software Features Three Search Modes: Gamma/Neutron total count rate. SNM search mode. Sliding average "monitor" mode. (NEW) User choice of identification schemes: Classify mode
More informationNUTC R304. Use of Absorption Mechanisms to Decrease Heavy Metal Mobility
Use of Absorption Mechanisms to Decrease Heavy Metal Mobility by Jianmin Wang Honglan Shi Joe G. Burken NUTC R304 A National University Transportation Center at Missouri University of Science and Technology
More informationLead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
More informationCHEM 101/105 Numbers and mass / Counting and weighing Lect-03
CHEM 101/105 Numbers and mass / Counting and weighing Lect-03 Interpretation of Elemental Chemical Symbols, Chemical Formulas, and Chemical Equations Interpretation of an element's chemical symbol depends
More informationLab #11: Determination of a Chemical Equilibrium Constant
Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a
More informationLecture 5. elements (Figure 1). In addition, there are many ways of classifying trace elements.
Lecture 5 Nomenclature for Trace Element Classification We have already grouped elements into two classes, major elements and trace elements (Figure 1). In addition, there are many ways of classifying
More informationSUS. Company Profile. Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0)208/658535 Fax 0049(0)208/658536
SUS Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0208/658535 Fax 0049(0208/658536 Company Profile SUS was founded in 1986 in Oberhausen in the Ruhr area (close to Düsseldorf, in order to meet
More informationLC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
More informationValidation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
More informationAPPENDIX B: EXERCISES
BUILDING CHEMISTRY LABORATORY SESSIONS APPENDIX B: EXERCISES Molecular mass, the mole, and mass percent Relative atomic and molecular mass Relative atomic mass (A r ) is a constant that expresses the ratio
More informationIB Chemistry. DP Chemistry Review
DP Chemistry Review Topic 1: Quantitative chemistry 1.1 The mole concept and Avogadro s constant Assessment statement Apply the mole concept to substances. Determine the number of particles and the amount
More informationAlignment and Preprocessing for Data Analysis
Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise
More informationUSEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required
More informationChemistry - Elements Electron Configurations The Periodic Table. Ron Robertson
Chemistry - Elements Electron Configurations The Periodic Table Ron Robertson History of Chemistry Before 16 th Century Alchemy Attempts (scientific or otherwise) to change cheap metals into gold no real
More informationUNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
More informationInstrumentation. (Figure 2)
X-Ray Fluorescence Lab Report Nydia Esparza Victoria Rangel Physics of XRF XRF is a non destructive analytical technique that is used for elemental and chemical analysis. X-Ray Fluorescence Spectroscopy
More informationHonors Chemistry: Unit 6 Test Stoichiometry PRACTICE TEST ANSWER KEY Page 1. A chemical equation. (C-4.4)
Honors Chemistry: Unit 6 Test Stoichiometry PRACTICE TEST ANSWER KEY Page 1 1. 2. 3. 4. 5. 6. Question What is a symbolic representation of a chemical reaction? What 3 things (values) is a mole of a chemical
More informationAnalysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
More information12. REDOX EQUILIBRIA
12. REDOX EQUILIBRIA The electrochemical series (reference table) 12.1. Redox reactions 12.2. Standard electrode potentials 12.3. Calculations involving electrochemical cells 12.4. Using Eʅ values to predict
More informationstoichiometry = the numerical relationships between chemical amounts in a reaction.
1 REACTIONS AND YIELD ANSWERS stoichiometry = the numerical relationships between chemical amounts in a reaction. 2C 8 H 18 (l) + 25O 2 16CO 2 (g) + 18H 2 O(g) From the equation, 16 moles of CO 2 (a greenhouse
More informationChem101: General Chemistry Lecture 9 Acids and Bases
: General Chemistry Lecture 9 Acids and Bases I. Introduction A. In chemistry, and particularly biochemistry, water is the most common solvent 1. In studying acids and bases we are going to see that water
More informationAMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
More informationName period AP chemistry Unit 2 worksheet Practice problems
Name period AP chemistry Unit 2 worksheet Practice problems 1. What are the SI units for a. Wavelength of light b. frequency of light c. speed of light Meter hertz (s -1 ) m s -1 (m/s) 2. T/F (correct
More informationUniversity of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
More informationWord Equations and Balancing Equations. Video Notes
Word Equations and Balancing Equations Video Notes In this lesson, you will: Use the law of conservation of mass and provide standard rules for writing and balancing equations. Write and balance equations
More informationCHM 1311: General Chemistry 1, Fall 2004 Exam #1, September 8, 2004. Name (print) SSN
CHM 1311: General Chemistry 1, Fall 2004 Exam #1, September 8, 2004 Name (print) SSN Pledge: I have neither given nor received aid on this exam: Signature For ALL problems: SHOW ALL WORK TO GET FULL CREDIT
More informationMETHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
More informationDetermination of elements by ICP-AES and ICP-MS
September 2003 HORIZONTAL - 19 Determination of elements by ICP-AES and ICP-MS Henk J van de Wiel National Institute of Public Health and the Environment (RIVM) Bilthoven, The Netherlands 1 HORIZONTAL
More informationSecondary Ion Mass Spectrometry
Secondary Ion Mass Spectrometry A PRACTICAL HANDBOOK FOR DEPTH PROFILING AND BULK IMPURITY ANALYSIS R. G. Wilson Hughes Research Laboratories Malibu, California F. A. Stevie AT&T Bell Laboratories Allentown,
More informationUPLC-MS/MS Analysis of Aldosterone in Plasma for Clinical Research
UPLC-MS/MS Analysis of in Plasma for Clinical Research Dominic Foley and Lisa Calton Waters Corporation, Wilmslow, UK APPLICATION BENEFITS Analytical selectivity improves reproducibility through removal
More informationGuidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
More informationSoil Chemistry Ch. 2. Chemical Principles As Applied to Soils
Chemical Principles As Applied to Soils I. Chemical units a. Moles and Avogadro s number The numbers of atoms, ions or molecules are important in chemical reactions because the number, rather than mass
More informationStandard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry
Standard Operation Procedure Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
More informationLAB 5 - PLANT NUTRITION. Chemical Ionic forms Approximate dry Element symbol Atomic weight Absorbed by plants tissue concentration
LAB 5 PLANT NUTRITION I. General Introduction All living organisms require certain elements for their survival. Plants are known to require carbon (C), hydrogen (H), oxygen (O), nitrogen (N), phosphorus
More informationCoating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
More informationaurora ICP-MS Innovation with Integrity The Science of Sensitivity ICP-MS
aurora ICP-MS The Science of Sensitivity Innovation with Integrity ICP-MS Powers the Business of Quantification aurora ICP-MS enhanced value package Bruker s ICP-MS is famous for its innovative and unique
More informationCh 20 Electrochemistry: the study of the relationships between electricity and chemical reactions.
Ch 20 Electrochemistry: the study of the relationships between electricity and chemical reactions. In electrochemical reactions, electrons are transferred from one species to another. Learning goals and
More informationNomenclature and Formulas of Ionic Compounds. Section I: Writing the Name from the Formula
Purpose: Theory: Nomenclature and Formulas of Ionic Compounds 1. To become familiar with the rules of chemical nomenclature, based on the classification of compounds. 2. To write the proper name of the
More informationX-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING
X-RAY FLUORESCENCE SPECTROSCOPY IN PLASTICS RECYCLING Brian L. Riise and Michael B. Biddle MBA Polymers, Inc., Richmond, CA, USA Michael M. Fisher American Plastics Council, Arlington, VA, USA X-Ray Fluorescence
More informationX Ray Flourescence (XRF)
X Ray Flourescence (XRF) Aspiring Geologist XRF Technique XRF is a rapid, relatively non destructive process that produces chemical analysis of rocks, minerals, sediments, fluids, and soils It s purpose
More informationElectrospray Ion Trap Mass Spectrometry. Introduction
Electrospray Ion Source Electrospray Ion Trap Mass Spectrometry Introduction The key to using MS for solutions is the ability to transfer your analytes into the vacuum of the mass spectrometer as ionic
More informationElemental Scientific. Stability. Elemental Scientific
Elemental Scientific prepfast ICP-MS: USP Oral Drugs by Dilution Automated Inline Calibration and Sample Dilution Authors: Kyle Uhlmeyer and Paul Field (field@icpms.com) Brief The prepfast inline
More informationNomenclature of Ionic Compounds
Nomenclature of Ionic Compounds Ionic compounds are composed of ions. An ion is an atom or molecule with an electrical charge. Monatomic ions are formed from single atoms that have gained or lost electrons.
More information