Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

Size: px
Start display at page:

Download "Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy"

Transcription

1 Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The availability and applications of fiber optic sampling in UV/Vis and UV/Vis/NIR spectroscopy is expanding rapidly. Fiber optics provide a remote sampling capability where sample requirements preclude use of cuvettes, or where remote sampling is required, i.e. on-line applications. Fiber optics accessories can be added to the current line of the LAMBDA spectrophotometers and are available in different lengths and wavelength ranges. All fiber optic accessories require an optrode or measurement head, as dictated by the application. This application note will describe one of the fiber optic measuring options available from PerkinElmer, the dipping probe accessory. Technical specifications, performance, and example data will be presented. Fiber optic dipping probe The fiber optic dipping probe (Figure 1) is a recent introduction to the PerkinElmer LAMBDA range of UV/Vis instruments and accessories, providing remote sampling capabilities. This probe includes a unique design allowing a single probe body to be used with different pathlength tips. The probe is coupled to the LAMBDA instrument through a transfer optic assembly, which provides light coupling to and from the probe. Energy efficiency of the dipping probe is typically between 8% to 15% relative to an open beam. The probe can be used for transmission or absorption studies in many liquids, including ones with suspended solids. Measurements can be taken over the UV/Vis wavelength range, from 200 nm to 1100 nm.

2 Fiber optic dipping probe features: Durable stainless steel design Multiple pathlengths with replaceable threaded tips High throughput Dual pass lightpath High repeatability Light shield to reduce stray light The replaceable threaded tips (Figure 2) are available in standard pathlengths of 2 mm, 5 mm, and 10 mm, with special order pathlengths to 50 mm. The 10 mm pathlength tip was used for all data acquired for this application note. The dipping probe incorporates a light shield so measurements can be conducted in normal light. The outlet and inlet fiber optics are routed directly into the sample compartment, so there is no need to lift the sample compartment cover during operation. Measurements can be taken directly on liquids in test tubes, beakers, and even large drums, making this accessory ideal for assaying the purity of incoming materials. The LAMBDA 35 running UVWinLab software can collect scan, rate reaction, wavelength programming, and concentration data using this accessory. A close-up picture of the sample compartment transfer optic is shown in Figure 4. This allows connection of the SMA connectors of the dipping probe. Performance The LAMBDA double beam series of spectrophotometers all have state-of-the-art performance specifications. Models include the LAMBDA 25/35/45 and the new research grade LAMBDA 650/850/950. The high energy throughput of these units allows the dipping optrode to be added with little loss of data quality. Corrected baselines with the LAMBDA 35 can typically be acquired to ±0.002 A or better using the dipping probe accessory. The accessory provides excellent energy throughput for a 600 micron monofiber, typically in the range of 10% to 20% transmission compared to an empty compartment. This high energy allows flat baselines to be acquired even at higher scan speeds, as shown in Figure 5. For best and consistent results, the placement of the dipping probe for sample measurement should be as close as possible to where the background correction was acquired. Shown are three overlaid corrected baselines of distilled water acquired with the dipping probe accessory, scanned at 960 nm/min, with a 2 nm bandpass. Typical noise data of the dipping probe and LAMBDA 35 system is shown in Table 2. The overall low noise over the entire wavelength range allows quality spectra with a high signal-to-noise ratio to be collected. The quality of fiber optic sampling systems has improved dramatically over the last decade. Early accessories offered for UV/Vis spectrophotometers were often plagued by low signal-to-noise and poor Figure 1. Fiber optic dipping probe accessory. Table 1. Fiber optic dipping probe specifications. Fiber type Silica/silica,.22NA Fiber size 600 microns (core) Probe body material Stainless steel 316 Diameter 6.35 mm (.25 ) Length of probe body ~150 mm (6 ), depending on pathlength Pathlength 2, 5, 10 mm replaceable tip; up to 50 mm possible Connections SMA-905 Overall length 1.5 m (4.5 ) Fiber optic cable sheathing PVC monocoil Wavelength ranges UV/Vis ( nm) Upper temperature range 85 C Figure 2. Different pathlength tips for the fiber optic dipping probe accessory. Figure 3. PerkinElmer LAMBDA 35 UV/Vis spectrophotometer fitted with dipping probe accessory. 2

3 repeatability. Often these early design accessories were evaluated and rejected by laboratories. The dipping probe accessory uses the latest generation silica monofibers, providing for high energy throughput. This allows high precision of sample readings, approaching the repeatability seen in cuvette readings. Shown in Figure 6 is an overlay of 15 scans of a yellow food dye acquired with the dipping probe accessory. The solution was measured in a flask and the fiber optic probe was removed and reinserted between scans as in laboratory use. The solution was scanned from 550 to 300 nm at 480 nm/min, and a 2 nm bandpass. The mean of the absorbance at the peak at nm was with a standard deviation of ± Concentration assays are especially easy to set up and run with the dipping probe accessory. Four concentrations of Evan s Blue dye were prepared in flasks (shown in Figure 3), and using UVWinLab 5.1 were measured with the Scanning Quant application. This application allows standards to be scanned; the software automatically determines the peak position and absorbance, and a calibration curve is plotted in real-time. Prior to measuring the standards, the dipping probe was placed in a flask of distilled water, and a corrected baseline was acquired. The experimental data is presented in Figures 7, 8, and 9. The solutions were scanned on a LAMBDA 35 at 480 nm/min and a 2 nm bandpass. In this experiment, excellent linearity was achieved, with a calculated correlation coefficient of Because no cuvettes were needed, assay time was very short. The excellent signal-to-noise of the dipping probe accessory fitted with a LAMBDA 35 allows very low concentration of analyte to be measured with precision. Low concentrations of benzene in water (25, 50, 100, and 150 ppb) were prepared in flasks and measured with the dipping probe accessory. The solutions were measured on a LAMBDA 35 from 280 nm to 230 nm, at 240 nm/min and a 2 nm bandpass. The results are shown in Figure 10. A background correction of distilled water was acquired and is shown as the 0 ppb curve. Even though the absorbances of these solutions were Experimental The dipping probe accessory provides for easy and rapid measurements of solutions in all types of containers. Table 2. Typical peak-to-peak and RMS noise of the LAMBDA 35 with dipping probe. Measured with a 4 nm slit at 0 Abs. Wavelength P/P RMS Figure 5. Expanded scale overlay of 3 corrected baselines of distilled water collected with the dipping probe accessory. Conditions were 960 nm/min, 2 nm slit. The baselines are within ± A. Figure 4. Monofiber transfer optic stage mounted in the sample compartment of the LAMBDA 35. Figure 6. Overlay of 15 scans of a yellow dye collected with the dipping probe accessory. The fiber optic probe was removed and reinserted into the flask containing the solution between scans. Conditions used were 480 nm/min and a 2 nm slit. 3

4 very low, and they were measured in clear flasks, quality spectra were obtained with little noise. Note that full-scale on the graph is 0.1 A, and the lowest concentration (25 ppb) had a measured absorbance at the peak of A. Again these measurements were rapid and easy, without the need for cuvettes, and without the need to open and close the sample compartment cover between measurements. Conclusion The dipping probe accessory is a new optional accessory available for the PerkinElmer line of LAMBDA UV/Vis and UV/Vis/NIR spectrophotometers. The accessory allows remote measurements outside of the sample compartment, and eliminates the need to use cuvettes. The high signal-to-noise of the LAMBDA 35 combined with the newest generation of very efficient silica monofibers, allows sample measurements that approach the precision of cuvette measurements. The ability to measure in any container, and in full room light, allows the purity of raw materials to be assayed rapidly. Figure 7. Four concentrations of Evan s Blue Dye (shown in Figure 3) measured in flasks with the Scanning Quant Application of UVWinLab 5.1 using the fiber optic dipping probe accessory. Figure 9. Standard curve for the Evan s Blue dye standards. Figure 8. Concentration results for Evan s Blue dye. Excellent photometric accuracy and linearity is achieved with the fiber optic dipping probe. Figure 10. The excellent signal-to-noise of the dipping probe accessory allows detection of benzene in water at very low concentrations measured in flasks. Note that the absorbance of these solutions is about 0.1 A or less. 4

5 PerkinElmer Life and Analytical Sciences 710 Bridgeport Avenue Shelton, CT USA Phone: (800) or (+1) For a complete listing of our global offices, visit /lasoffices 2004 PerkinElmer, Inc. All rights reserved. The PerkinElmer logo and design are registered trademarks of PerkinElmer, Inc. LAMBDA and UVWinLab are trademarks and PerkinElmer is a registered trademark of PerkinElmer, Inc. or its subsidiaries, in the United States and other countries. All other trademarks not owned by PerkinElmer, Inc. or its subsidiaries that are depicted herein are the property of their respective owners. PerkinElmer reserves the right to change this document at any time without notice and disclaims liability for editorial, pictorial or typographical errors _07 Printed in USA

Agilent Cary 60 UV-Vis

Agilent Cary 60 UV-Vis Agilent Cary 60 UV-Vis Efficient. Accurate. Flexible. Specifications Introduction The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is designed to meet both current

More information

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Guaranteed specifications Design overview Double beam, ratio recording, double out-of-plane Littrow monochromator UV-Vis-NIR spectrophotometer (Agilent Cary

More information

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this

More information

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements. T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting

More information

UV-Vis spectrophotometers

UV-Vis spectrophotometers UV-Vis spectrophotometers an Advanced Digital Technology EXPERIENCE SECOMAM is proud to provide the new fourthgeneration UVIKON ultraviolet-visible spectrophotometers.with more than five decades of experience

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

INTERNATIONAL OLIVE COUNCIL

INTERNATIONAL OLIVE COUNCIL INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org

More information

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory

Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR Spectrometer and URA Accessory FIELD APPLICATION REPORT UV/Vis/NIR Spectroscopy Author: Frank Padera Shelton, CT Contributor: Chris Lynch Shelton, CT Measurement of Enhanced Specular Reflector (ESR) Films Using a LAMBDA 1050 UV/Vis/NIR

More information

VWR SPECTROPHOTOMETERS

VWR SPECTROPHOTOMETERS VWR SPECTROPHOTOMETERS Reliable Accurate Easy to Use New_Spectrophotometry_v2.indd 2 Your first choice for spectrophotometry 5/9/2014 2:22:18 PM Quality Guaranteed VWR SPECTROPHOTOMETERS Single Beam Units

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8) Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:

More information

instruments Analytical Instruments for Science

instruments Analytical Instruments for Science instruments Analytical Instruments for Science instruments Contents PAGE NO. Introduction 4 T60 UV-Vis Spectrophotometer 6 T70 UV-Vis Spectrophotometer 10 T80 UV-Vis Spectrophotometer 14 T90+ UV-Vis Spectrophotometer

More information

Lab #11: Determination of a Chemical Equilibrium Constant

Lab #11: Determination of a Chemical Equilibrium Constant Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F) Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F) Please contact Dr. Amanda Young for training requests and assistance: 979-862-6845, amandayoung@tamu.edu Hardware Our spectrophotometer is made up of

More information

High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering

High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering Power spectrum development with heterodyne technology advances biotechnology and nanotechnology measurements. M. N. Trainer

More information

Chem 131A: Absorbance of Riboflavin

Chem 131A: Absorbance of Riboflavin Chem 131A: Absorbance of Riboflavin Purpose: The purpose of this experiment is to: 1) Familiarize the student with the use of the HP 8452 diode array spectrophotometer, 2) examine the limitations of the

More information

2 Spectrophotometry and the Analysis of Riboflavin

2 Spectrophotometry and the Analysis of Riboflavin 2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound

More information

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information 1 Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information Instructions for the Operation of the Cary 300 Bio UV-Visible Spectrophotometer See the Thermo OMNIC Help reference on page 49. Ultraviolet-Visible

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium

More information

A Beer s Law Experiment

A Beer s Law Experiment A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining

More information

SINGLEMODE OR MULTIMODE FIBER OPTIC PATCHCORDS

SINGLEMODE OR MULTIMODE FIBER OPTIC PATCHCORDS Features: SINGLEMODE OR MULTIMODE FIBER OPTIC PATCHCORDS Low insertion loss < 0.2 db Excellent repeatability FC/PC, SC, ST, LC, MU, E2000 termination available Custom ferrule termination available Designed

More information

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V.

Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V. 1234 Hach Hall 515-294-5805 www.cif.iastate.edu Cary 100 Bio UV-Vis Operating Instructions 09/25/2012 S.V. Location: Contact: 1240 Hach Hall Steve Veysey, 1234 Hach Hall Safety All researchers working

More information

Reaction of Blue Food Dye with Bleach

Reaction of Blue Food Dye with Bleach Exercise 2 Reaction of Blue Food Dye with Bleach 2 Introduction In the experiment, you will study the rate of the reaction of FD&C Blue #1 (Blue #1 is denoted by E number E133 in food stuff) with sodium

More information

Hellma. Where precision becomes an art. Calibration standards for spectrophotometers. 2008 by Hellma www.hellma-worldwide.com

Hellma. Where precision becomes an art. Calibration standards for spectrophotometers. 2008 by Hellma www.hellma-worldwide.com Hellma. Where precision becomes an art. Calibration standards for spectrophotometers 2008 by Hellma www.hellma-worldwide.com Contents Calibration of spectrophotometers 3 Traceability of the calibration

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

Reaction Stoichiometry and the Formation of a Metal Ion Complex

Reaction Stoichiometry and the Formation of a Metal Ion Complex Reaction Stoichiometry and the Formation of a Metal Ion Complex Objectives The objectives of this laboratory are as follows: To use the method of continuous variation to determine the reaction stoichiometry

More information

TotalChrom. Chromatography Data Systems. Streamlining your laboratory workflow

TotalChrom. Chromatography Data Systems. Streamlining your laboratory workflow TotalChrom Chromatography Data Systems Streamlining your laboratory workflow maximize productivity with TotalChrom CDS Acquiring, processing, reporting, reviewing and approving data is a streamlined series

More information

Using the Spectrophotometer

Using the Spectrophotometer Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to

More information

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds Overview In order for accuracy and precision to be optimal, the assay must be properly evaluated and a few

More information

Universal Data Acquisition (UDA)

Universal Data Acquisition (UDA) Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte

More information

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS 1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years

More information

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1 Experiment 13H FV 1/25/2011(2-run) THE REACTION OF RED FOOD COLOR WITH BLEACH 1 PROBLEM: Determine the rate law for the chemical reaction between FD&C Red Dye #3 and sodium hypochlorite. LEARNING OBJECTIVES:

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

UV-VIS-NIR. Varian Cary 4000, 5000 and 6000i Spectrophotometers

UV-VIS-NIR. Varian Cary 4000, 5000 and 6000i Spectrophotometers UV-VIS-NIR Varian Cary 4000, 5000 and 6000i Spectrophotometers Varian Cary At the leading edge Varian s Cary range of UV-Vis-NIR spectrophotometers has been synonymous with excellence and high performance

More information

UV-VISIBLE SPECTROPHOTOMETER

UV-VISIBLE SPECTROPHOTOMETER instruments UV-VISIBLE SPECTROPHOTOMETER PRODUCT BROCHURE Analytical Instruments for Science Contents PAGE NO. Introduction 4 T60 UV-Vis Spectrophotometer 6 T70 UV-Vis Spectrophotometer 10 T80 UV-Vis Spectrophotometer

More information

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)

University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical

More information

Upon completion of this lab, the student will be able to:

Upon completion of this lab, the student will be able to: 1 Learning Outcomes EXPERIMENT B4: CHEMICAL EQUILIBRIUM Upon completion of this lab, the student will be able to: 1) Analyze the absorbance spectrum of a sample. 2) Calculate the equilibrium constant for

More information

Shimadzu UV-VIS User s Guide

Shimadzu UV-VIS User s Guide Shimadzu UV-VIS User s Guide 1) Push the F4 button on the UV-VIS instrument keypad. This will enable PC control. Push the F4 Button 2) Log into the UV-VIS software with your username and password. 3) After

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

Nucleic Acid Purity Assessment using A 260 /A 280 Ratios

Nucleic Acid Purity Assessment using A 260 /A 280 Ratios Nucleic Acid Purity Assessment using A 260 /A 280 Ratios A common practice in molecular biology is to perform a quick assessment of the purity of nucleic acid samples by determining the ratio of spectrophotometric

More information

Validating UV/Visible Spectrophotometers

Validating UV/Visible Spectrophotometers Technical Note Validating UV/Visible Spectrophotometers UV/Vis Spectroscopy UV/Visible spectrophotometers are widely used by many laboratories including those found in academia and research as well as

More information

Particle Characterization of UV Blocking Sunscreens and Cosmetics Using UV/Visible Spectroscopy

Particle Characterization of UV Blocking Sunscreens and Cosmetics Using UV/Visible Spectroscopy TECHNICAL NOTE UV/Visible Spectroscopy Authors: Jeffrey L. Taylor, Ph.D. Chris Lynch Jillian F. Dlugos PerkinElmer, Inc. Shelton, CT Particle Characterization of UV Blocking Sunscreens and Cosmetics Using

More information

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry General Medicine Jiřina Crkovská (translated by Jan Pláteník) 2010/2011 1 Spectrophotometry is one of the most widely used instrumental

More information

Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions

Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions Flame Direct Attach UV-VIS Integrated Sampling System Installation and Operation Instructions Description The Flame Integrated Sampling System (FLAME-DA-CUV-UV-VIS) is a snap-on, direct-attach 1 cm cuvette

More information

Phenolphthalein-NaOH Kinetics

Phenolphthalein-NaOH Kinetics Phenolphthalein-NaOH Kinetics Phenolphthalein is one of the most common acid-base indicators used to determine the end point in acid-base titrations. It is also the active ingredient in some laxatives.

More information

UV-VIS VARIAN, INC. Cary 100/300 UV-Vis. S p e c t r o p h o t o m e t e r s

UV-VIS VARIAN, INC. Cary 100/300 UV-Vis. S p e c t r o p h o t o m e t e r s UV-VIS VARIAN, INC. Cary 100/300 UV-Vis S p e c t r o p h o t o m e t e r s Unmatched quality and performance Outstanding performance, flexibility, and ease of use are what you can expect from Varian s

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

Thermo Scientific GENESYS 10S Series UV-Visible Spectrophotometers

Thermo Scientific GENESYS 10S Series UV-Visible Spectrophotometers m o l e c u l a r s p e c t r o s c o p y Thermo Scientific GENESYS 10S Series UV-Visible Spectrophotometers Powerful Performance, Affordable Price Accurate Reliable Easy-to-Use Part of Thermo Fisher Scientific

More information

Thermo Scientific NanoDrop Products. Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules

Thermo Scientific NanoDrop Products. Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules Thermo Scientific NanoDrop Products Revolutionary technology. Elegant simplicity. Instruments for microvolume analysis of biomolecules Thermo Scientific NanoDrop instruments have pre-programmed methods

More information

What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher)

What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher) What s in the Mix? Liquid Color Spectroscopy Lab (Randy Landsberg & Bill Fisher) Introduction: There is more to a color than a name. Color can tell us lots of information. In this lab you will use a spectrophotometer

More information

FT-NIR for Online Analysis in Polyol Production

FT-NIR for Online Analysis in Polyol Production Application Note: 51594 FT-NIR for Online Analysis in Polyol Production Key Words Acid Number Ethylene Oxide FT-NIR Hydroxyl Value Polyester Polyols Abstract Hydroxyl value and other related parameters

More information

Uses of Derivative Spectroscopy

Uses of Derivative Spectroscopy Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative

More information

designed to never let you down PerkinElmer Series 200 HPLC Systems

designed to never let you down PerkinElmer Series 200 HPLC Systems designed to never let you down PerkinElmer Series 200 HPLC Systems an inside look at the Series 200 HPLC systems PerkinElmer knows you need to rely on your high performance liquid chromatography (HPLC)

More information

Lantronix ETS8P Terminal Server Configuration and Troubleshooting

Lantronix ETS8P Terminal Server Configuration and Troubleshooting Lantronix ETS8P Terminal Server Configuration and Troubleshooting CHROMATOGRAPHY DATA SYSTEMS T E C H N I C A L N O T E Introduction There are a number of ways the Lantronix terminal server can be configured

More information

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)

Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Application Note: 52020 Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Timothy O. Deschaines, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words Array Automation

More information

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40

Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Absorbance Spectrophotometry: Analysis of FD&C Red Food Dye #40 Note: there is a second document that goes with this one! 2046 - Absorbance Spectrophotometry - Calibration Curve Procedure. The second document

More information

Hitachi Double Beam Spectrophotometer

Hitachi Double Beam Spectrophotometer Hitachi Double Beam Spectrophotometer Spectrophotometer simpler to use and higher in dependability thanks to mounting of a large-size color LCD. High resolution satisfying European Pharmacopoeia (spectral

More information

Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument

Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument Calculating Nucleic Acid or Protein Concentration Using the GloMax Multi+ Microplate Instrument Technical Note INTRODUCTION Direct measurements of nucleic acid samples at OD 260 or protein samples at OD

More information

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property

More information

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007

PUMPED Nd:YAG LASER. Last Revision: August 21, 2007 PUMPED Nd:YAG LASER Last Revision: August 21, 2007 QUESTION TO BE INVESTIGATED: How can an efficient atomic transition laser be constructed and characterized? INTRODUCTION: This lab exercise will allow

More information

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1

Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 Chemistry 111 Lab: Intro to Spectrophotometry Page E-1 SPECTROPHOTOMETRY Absorption Measurements & their Application to Quantitative Analysis study of the interaction of light (or other electromagnetic

More information

Determination of the Mass Percentage of Copper in a Penny. Introduction

Determination of the Mass Percentage of Copper in a Penny. Introduction Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny

More information

www.gbo.com/bioscience

www.gbo.com/bioscience Application Note UV/VIS Spectroscopy in Microplates UV-Star, µclear, MICROLON and CELLSTAR 1. Introduction to UV/VIS spectroscopy Transparent microplates (CELLSTAR, MICROLON ) or microplates with a clear

More information

INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS. Published: April 1999

INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS. Published: April 1999 INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS Published: April 1999 CONTENTS 1 Introduction 2 Test Procedures General 3 Standards and Specific Test Procedures

More information

Managing screening programs. with PerkinElmer s Specimen Gate Process Management Software

Managing screening programs. with PerkinElmer s Specimen Gate Process Management Software Managing screening programs with PerkinElmer s Specimen Gate Process Software Trusted data management is the key to Based on PerkinElmer s Specimen Gate platform, the systems installed today are modular,

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

Hitachi Spectrophotometer

Hitachi Spectrophotometer Hitachi Spectrophotometer Spectrophotometer Meeting a Wide Range of Analytical Needs from Liquid to Solid Sample Measurements Single monochromator U-3900 Stray light : 0.015% or less Photometric range

More information

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters

More information

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy

Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy Heavy Metal Analysis for the Toy Industry by Inductively Coupled Plasma Optical Emission Spectroscopy In support of the toy industry, many laboratories are being called upon to analyze a wide variety of

More information

FIBER OPTIC ISOLATORS

FIBER OPTIC ISOLATORS Features: >0W Optical power handling capability Polarization Sensitive and Insensitive versions Product offerings over 3 000 nm wavelength range High isolation levels and low return loss Low Insertion

More information

Infrared Optical Fiber. Datasheets and Price list. JTIngram Sales and Marketing www.jtingram.com

Infrared Optical Fiber. Datasheets and Price list. JTIngram Sales and Marketing www.jtingram.com Infrared Optical Fiber Datasheets and Price list JTIngram Sales and Marketing www.jtingram.com Polycrystalline InfraRed (PIR-) & Chalcogenide InfraRed (CIR-) Fibers for spectral range of 1-18µm (10.000-550cm-1)

More information

Colorimetry Extinction coefficient (ε) Lambda max (λ max ) Qualitative vs. quantitative analysis

Colorimetry Extinction coefficient (ε) Lambda max (λ max ) Qualitative vs. quantitative analysis Lab Week 2 - Spectrophotometry Purpose: Introduce students to the use of spectrophotometry for qualitative (what is it) and quantitative (how much is there of it) analysis of biological samples and molecules.

More information

BY UV SPECTROPHOTOMETRY

BY UV SPECTROPHOTOMETRY 1 of 9 1. PURPOSE This Standard Operating Procedure describes the method for measuring the concentration of viral particles using UV/Vis spectrophotometry at 260 and 280 nm. 2. SCOPE This procedure is

More information

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.

Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Introduction: The determination of protein concentration is frequently required in biochemical work. Several methods

More information

Enzyme Kinetics: Properties of â-galactosidase

Enzyme Kinetics: Properties of â-galactosidase Enzyme Kinetics: Properties of â-galactosidase Preparation for Laboratory: Read the introduction to this laboratory before doing the Web Tutorial - Beta Galactosidase. Additional background: Freeman, skim

More information

Determination of the Rate Law for Food Dye Bleaching with Hypochlorite

Determination of the Rate Law for Food Dye Bleaching with Hypochlorite This is an example report of an investigation performed in General Chemistry lab. Pay attention to format and content, not on the results or the experiment itself. The report is best explored on screen

More information

CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001)

CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001) CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001) I. Introduction Calibration experiments or procedures are fairly common in laboratory work which involves any type of instrumentation. Calibration

More information

Appendix 5 Overview of requirements in English

Appendix 5 Overview of requirements in English Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

Determination of the Sensitivity Range of Biuret Test for Undergraduate Biochemistry Experiments

Determination of the Sensitivity Range of Biuret Test for Undergraduate Biochemistry Experiments 77 Determination of the Sensitivity Range of Biuret Test for Undergraduate Biochemistry Experiments Gerardo Janairo * ; Marianne Linley Sy; Leonisa Yap; Nancy Llanos-Lazaro; Julita Robles Chemistry Department

More information

Use of Micropipettes

Use of Micropipettes Use of Micropipettes Prior to lab you should understand: The function of micropipettes in the laboratory Basic parts of micropipette What volumes are measured with P, P and P1 micopipettors How to read

More information

Single Photon Counting Module COUNT -Series

Single Photon Counting Module COUNT -Series Description Laser Components COUNT series of s has been developed to offer a unique combination of high photon detection efficiency, wide dynamic range and ease of use for photon counting applications.

More information

Asian Journal of Food and Agro-Industry ISSN 1906-3040 Available online at www.ajofai.info

Asian Journal of Food and Agro-Industry ISSN 1906-3040 Available online at www.ajofai.info As. J. Food Ag-Ind. 008, (0), - Asian Journal of Food and Agro-Industry ISSN 906-00 Available online at www.ajofai.info Research Article Analysis of NIR spectral reflectance linearization and gradient

More information

2C: One in a Million. Part 1: Making solutions. Name: Section: Date: Materials

2C: One in a Million. Part 1: Making solutions. Name: Section: Date: Materials Name: Section: Date: 2C: One in a Million Drinking water can contain up to 1.3 parts per million (ppm) of copper and still be considered safe. What does parts per million mean? Both living things and the

More information

Spectrophotometer Water Quality Analysers

Spectrophotometer Water Quality Analysers Spectrophotometer Water Quality Analysers Provisional 5 June 2015 Measurements You Trust Photonic Measurements spectrophotometers for water quality measurements provides a rich user interface, that makes

More information

POLYDIMETHYLSILOXANE

POLYDIMETHYLSILOXANE POLYDIMETHYLSILOXANE Prepared at the 37th JECFA (1990), published in FNP 52 (1992) superseding specifications prepared at the 29th JECFA (1985), published in FNP 34 (1986). Metals and arsenic specifications

More information

limit to percentage levels, a limited working range translates

limit to percentage levels, a limited working range translates Figure 1. The AutoPrep 50 Automatic Dilution System. Figure 2. Schematic diagram of the AutoPrep 50. Flame atomic absorption spectroscopy (flame AA) is still the most frequently used instrumental technique

More information

21 CFR Part 11 Implementation Spectrum ES

21 CFR Part 11 Implementation Spectrum ES 21 CFR Part 11 Implementation Spectrum ES INFRARED SPECTROSCOPY T E C H N I C A L N O T E Introduction Compliance with 21 CFR Part 11 is mandatory for pharmaceutical companies and their suppliers to sell

More information

QUANTITATIVE DETERMINATION OF COMMON TYPES OF ASBESTOS BY DIFFUSE REFLECTANCE FTIR

QUANTITATIVE DETERMINATION OF COMMON TYPES OF ASBESTOS BY DIFFUSE REFLECTANCE FTIR MATERIALS ANALYSIS QUANTITATIVE DETERMINATION OF COMMON TYPES OF ASBESTOS BY DIFFUSE REFLECTANCE FTIR Solutions for Your Analytical Business Markets and Applications Programs Solution Note Materials Author(s)

More information

73 Series Spectrophotometers

73 Series Spectrophotometers 73 Series Spectrophotometers The 73 series introduces four new spectrophotometers to the extensive Jenway range. These spectrophotometers use icon driven software and have an improved navigation system

More information

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,

More information

Field Test to Demonstrate Real-Time In-Situ Detection of Volatile Organic Compounds Hazmat Spill Center, Nevada Test Site September 19-25, 2001

Field Test to Demonstrate Real-Time In-Situ Detection of Volatile Organic Compounds Hazmat Spill Center, Nevada Test Site September 19-25, 2001 Field Test to Demonstrate Real-Time In-Situ Detection of Volatile Organic Compounds Hazmat Spill Center, Nevada Test Site September 19-2, 21 Clifford K. Ho Sandia National Laboratories Albuquerque, NM

More information

The Determination of an Equilibrium Constant

The Determination of an Equilibrium Constant The Determination of an Equilibrium Constant Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium constant, K

More information

Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana*

Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana* Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana* a Department of Pharmaceutical Analysis, Andhra University, Visakhapatnam. b Department of Pharmaceutics,

More information

The Measurement of Sensitivity in Fluorescence Spectroscopy

The Measurement of Sensitivity in Fluorescence Spectroscopy The Measurement of Sensitivity in Fluorescence Spectroscopy Among instrumental techniques, fluorescence spectroscopy is recognized as one of the more sensitive. In fluorescence, the intensity of the emission

More information