Uses of Derivative Spectroscopy

Size: px
Start display at page:

Download "Uses of Derivative Spectroscopy"

Transcription

1 Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative analysis and for quantification. The concept of derivatizing spectral data was first introduced in the 1950s, when it was shown to have many advantages. However, the technique received little attention primarily because of the complexity of generating derivative spectra using early UV-Visible spectrophotometers. The introduction of microcomputers in the late 1970s made it generally practicable to use mathematical methods to generate derivative spectra quickly, easily and reproducibly. This significantly increased the use of the derivative technique. In this application note we review briefly the mathematics and generation methods of derivative spectroscopy. We illustrate the features and applications using computer-generated examples. Agilent Technologies Innovating the HP Way

2 Introduction If a spectrum is expressed as absorbance, A, as a function of wavelength, l, the derivative spectra are: Zero order First order A f( λ) da f ( λ) dλ E-04 E E-04 Second order 2 da 2 f ( λ) dλ E-04 1st derivative -6.0E-04 2nd derivative E-04 3rd derivative E+06 4rd derivative Figure 1 shows a computer simulation of the effects of derivatization on the appearance of a simple Gaussian absorbance band. Derivative spectra are always more complex than zeroorder spectra. E-04 E+00 -E E E-07 E E Figure 1 and derivative spectra of a Gaussian band A first-order derivative is the rate of change of absorbance with respect to wavelength. A firstorder derivative starts and finishes at zero. It also passes through zero at the same wavelength as l max of the absorbance band. Either side of this point are positive and negative bands with maximum and minimum at the same wavelengths as the inflection points in the absorbance band. This bipolar function is characteristic of all odd-order derivatives. The most characteristic feature of a second-order derivative is a negative band with minimum at the same wavelength as the maximum on the zero-order band. It also shows two additional positive satellite bands either side of the main band. A fourth-order derivative shows a positive band. A strong negative or positive band with minimum or maximum at the same wavelength as l max of the absorbance band is characteristic of the even-order derivatives. Note that the number of bands observed is equal to the derivative order plus one.

3 Quantification If we assume that the zero-order spectrum obeys Beer s law, there is a similar linear relationship between concentration and amplitude for all orders of derivative: Zero order First order n th order l A e b c Aεb c da dε dλ dλ bc n n d A d ε d d bc n n λ λ wavelength absorbance extinction coefficient sample path length sample concentration For single component quantification the selection of wavelengths for derivative spectra is not as simple as for absorbance spectra because there are both positive and negative peaks. For the evenorder derivatives there is a peak maximum or minimum at the same l max as the absorbance spectrum but for the odd-order derivatives this wavelength is a zero crossing point. Taking the difference between the highest maximum and the lowest minimum gives the best signal-tonoise ratio but may lead to increased sensitivity to interference from other components. Obtaining derivative spectra Derivative spectra can be obtained by optical, electronic, or mathematical methods. Optical and electronic techniques were used on early UV-Visible spectrophotometers but have largely been superseded by mathematical techniques. The advantages of the mathematical techniques are that derivative spectra may be easily calculated and recalculated with different parameters, and smoothing techniques may be used to improve the signal-to-noise ratio. Optical and electronic techniques The main optical technique is wavelength modulation, where the wavelength of incident light is rapidly modulated over a narrow wavelength range by an electromechanical device. The first and second derivatives may be generated using this technique. It is popular for dedicated spectrophotometer designs used in, for example, environmental monitoring. First-derivative spectra may also be generated by a dual wavelength spectrophotometer. The derivative spectrum is generated by scanning with each monochromator separated by a small constant wavelength difference. First and higher-order derivatives can be generated using analog resistance capacitance devices. These generate the derivative as a function of time as the spectrum is scanned at constant speed (da/dts). For the first derivative: da da dλ 1 S dt Higher-order derivatives are obtained by using successive derivators. The electronic method suffers from the disadvantage that the amplitude and wavelength shift of the derivatives varies with scan speed, slit width, and resistance-capacitance gain factor. Mathematical techniques To use mathematical techniques the spectrum is first digitized with a sampling interval of Dl. The size of Dl depends on the natural bandwidth (NBW) of the bands being processed and of the bandwidth of the instrument used to generate the data. Typically, for UV-Visible spectra, the NBW is in the range 10 to 50 nm. Firstderivative spectra may be calculated simply by taking the difference in absorbance between two closely spaced wavelengths for all wavelengths : D λ+ λ/ 2 ( Aλ+ λ Aλ) λ Where the derivative amplitude, D l, is calculated for a wavelength intermediate between the two absorbance wavelengths. For the second-derivative determination three closely-spaced wavelength values are used: D λ ( Aλ λ 2Aλ+ Aλ + λ) 2 λ Higher-order derivatives can be calculated from similar expressions. This method involves simple linear interpolation between adjacent wavelengths. A better method is that proposed by Savitzky and Golay. To calculate the derivative at a particular wavelength, l, a window of ±n data points is selected and a polynomial is fitted using the least squares method: l Aλ a0 + a1λ+ Κ + a l λ

4 Features and applications An advantage of this method is that it can be used to smooth the data. If the polynomial order, l, is less than the number of data points (2n+1) in the window, the polynomial generally cannot go through all data points and thus the least squares fit gives a smoothed approximation to the original data points. This feature can be used to counteract the degradation of signal-to-noise that is inherent in the derivatization process (see below). The coefficients a 0...a l at each wavelength multiplied by the factorial of the order are the derivative values: a 1 is the first derivative, 2xa 2 the second derivative, 6xa 3 the third derivative, and so on. Savitzky and Golay developed a very efficient method to perform the calculations and this is the basis of the derivatization algorithm in most commercial instruments. Other techniques for calculating derivatives, for example, using Fourier Transforms, are available but are not commercially popular. One consequence of these mathematical methods for the calculation of derivatives is that data points at the beginning and end of the wavelength range are lost. If three data points are used for the process then one data point will be lost at each end of the range for each derivative order. If five points are used then two points will be lost and so on. It should be clearly understood that, although transformation of a UV-Visible spectrum to its first or higher derivative usually yields a more complex profile than the zero-order spectrum (see below), the intrinsic information content is not increased. In fact, it is decreased by the loss of lowerorder data such as constant offset factors. For clarity the points made in the following discussion are illustrated using computer-generated examples. In the figures, dotted lines show the baseline, dashed lines show component spectra, and solid lines show the analyte spectrum made up from the component spectra. Graphics As shown in figure 1 there is an increase in the number of bands as higher orders of derivative are calculated. This increase in the complexity of the derivative spectra can be very useful in qualitative analysis, either for characterizing materials or for identification. Spectra that are very similar in absorbance mode may reveal significant differences in the derivative mode E-06 E E-06 For example, the absorbance spectrum of the steroid testosterone has a single, broad, featureless band centered at about 330 nm but the second derivative has six quite distinctive peaks. Resolution As figure 1 shows, the derivative centroid bandwidth of the evenorder derivatives decreases with increasing order. Relative to the zero-order spectrum the derivative centroid bandwidth for a Gaussian band is observed to decrease to 53 %, 41 %, and 34 % of the original bandwidth in the second, fourth, and sixth orders respectively. This feature may be used in qualitative analysis to identify the presence of two analytes with very similar l max values that are not resolved in the absorbance spectrum. Figure 2 shows a computer simulation. Figure 2 Resolution enhancement

5 In absorbance mode, when two Gaussian bands with 40 nm NBW and separated by 30 nm, are added the result is a single band with a maximum midway between the two component bands. The two components are not resolved. In the fourth derivative the presence of these two bands is clearly visible with maxima centered close to the l max of the component bands. Although the bands have been resolved there is no indication of whether these arise from two chromophores in a single compound or in two different compounds. It is often claimed that this increased resolution and the increased differentiation between spectra in the derivative mode allows multicomponent analysis of mixtures of components with similar spectra that cannot be resolved in the absorbance mode. However, as noted above the information content of derivative spectra is, in fact, less than the absorbance spectra and it can easily be shown that the improvements in quantitative accuracy are the result of other effects as described below. Background elimination A common, unwanted effect in spectroscopy is baseline shift. This may arise either from instrument (lamp or detector instabilities) or sample handling (cuvette repositioning) effects. Because the first derivative of a constant absorbance offset is zero, using the firstderivative spectra always eliminates such baseline shifts and improves the accuracy of quantification. This is illustrated in figure 3 where a 0.1 A offset, that would cause a 10 % quantitative error for the analyte, is completely eliminated by the first derivative. Other background effects that are directly proportional to higher orders of wavelength with the general form: 1 n A a + a λ... + a n λ 0 1 can be eliminated by using higher orders of derivative but such spectral features with exactly this form are very uncommon so this effect has little practical use Figure 3 Background elimination

6 Discrimination Probably the most important effect of the derivative process is that broad bands are suppressed relative to sharp bands and this suppression increases with increasing derivative order. This arises from the fact that the amplitude, D n, of a Gaussian band in the n th derivative is inversely proportional to the original bandwidth, W, raised to the n th degree: D n 1 n W Scattering is inversely proportional to the fourth (Rayleigh, small particles) or second (Tyndall, larger particles) power of the wavelength. Because the relationship is inverse, the use of derivatives will not eliminate the scattering component from the spectrum as has been claimed in some publications. However, because the scattering component resembles a very broad absorbance, using derivatives discriminates against it and reduces its effect on quantification. Thus for two coincident bands of equal intensity but different bandwidth in the zero order, the n th derivative amplitude of the sharper band, X, is greater than that of the broader band, Y, by a factor that is dependent on the relative bandwidth and the derivative order: n n DX WY n n D W Y Figure 4 shows the effect of taking derivatives of two bands, one with 160 nm NBW and one with 40 nm NBW. In absorbance mode they have equal amplitude, in first derivative the narrower band has four times greater amplitude and in the second derivative it has sixteen times the amplitude. This property is used to improve the accuracy of quantification of a narrow band component in the presence of a broad band component and to reduce error caused by scattering. X st derivative 2nd derivative Scattering is a common problem in biological analyses resulting from the measurement of small particulates present in the sample. -10 Figure 4 Discrimination against broad bands

7 Figure 5 shows an absorbance band with 40 nm NBW and the same band in the presence of a scattering background. Without any correction, the amplitude at 500 nm is 920 A instead of A because of the scattering contribution. Quantification at this wavelength results in an error of 9.2 %. Using the first derivative the contribution from the scattering component is reduced such that, using peak maximum to minimum, the signal in the presence of scattering is 2992 instead of 3024, that is a quantification error of only -1.1 %. This ability to discriminate against scattering components is widely used in the analysis of biological fluids that contain a high level of particulates, and in pharmaceutical analyses where particulate excipients in tablets and capsules cause quantification errors. 1-1 Figure 5 Scatter elimination Matrix suppression The analytical problem is often not simply scattering, baseline shift, or unwanted broad absorbing components. It is a combination of two or more of these that results in a broad absorbing background matrix. In qualitative analyses, derivatization often allows the detection and positive identification of trace levels of a component in the presence of a strongly absorbing matrix. This is illustrated in Figure 6. A trace, 1 A, of a 40 nm NBW component with l max at 500 nm was added to a synthetic matrix. The matrix comprised offset, second and fourth-order scatter, and 320 nm NBW components with l max at 300 and 600 nm. In absorbance mode the presence of this component is virtually undetectable. In secondorder derivative mode its presence is obvious. 6.0E E E-05 E E-05 Figure 6 Matrix supression

8 In quantitative analyses, derivatization improves the accuracy of quantification in the presence of interference caused by a broad absorbing component, matrix, or scattering. Thus in the example given above, quantification of the analyte in the absorbance mode without any correction results in an error of nearly 5000 % (absorbance of 02 A instead of 1 A). Using the baseline-to-valley signal of the second-order derivative the error is -2.1% (2.37 x 10-5 instead of 2.42 x 10-5 A/l 2 ). An example of discrimination against a broad absorbing matrix is the quantification of caffeine in soft drinks. Soft drinks generally contain a mixture of natural and synthetic products with added colorants, resulting in a broad featureless absorbance over a wide wavelength range. In absorbance mode, quantification of caffeine is inaccurate because of the matrix effect but good accuracy can often be achieved using the second-derivative spectra. Instrument considerations Virtually all current UV-Visible spectrophotometers generate derivative spectra by mathematical means so instrument considerations for generation of derivative spectra by optical and electronic techniques are not discussed. Instrument requirements for derivative spectroscopy are, in general, similar to those for conventional absorbance spectroscopy but wavelength reproducibility and signal-to-noise are of increased importance. The increased resolution of derivative spectra puts increased demands on the wavelength reproducibility of the spectrophotometer. Small wavelength errors can result in much larger signal errors in the derivative mode than in the absorbance mode. The negative effect of derivatization on signal-to-noise also puts increased demands on low noise characteristics of the spectrophotometer. It is an advantage in this case, if the spectrophotometer can scan and average multiple spectra before derivatization to improve further the signal-to-noise ratio. For the derivatization process it is important to be able to control the degree of smoothing that is applied in order to adapt to differing analytical problems. In the case of the Savitzky-Golay method this means being able to vary the order of polynomial and the number of data points used. Signal-to-noise ratio An unwanted effect of the derivatization process is that the signal-to-noise ratio decreases as higher orders of derivatives are used. This follows from the discrimination effect and the fact that noise always contains the sharpest features in the spectrum. Thus, if the spectral data used in the derivative calculation is at 2 nm intervals, the noise has a 2 nm bandwidth. If the analyte band has a bandwidth of 20 nm then the signal-to-noise ratio of the first derivative is ten times worse than the zero-order spectrum. The decrease in signal-to-noise ratio can be reduced by using the smoothing properties of the Savitzky-Golay polynomial smoothing technique but great care must be taken as too high a degree of smoothing distorts the derivative spectrum. Alternative techniques, such as using a reference wavelength or full spectrum multicomponent analysis with a scattering spectrum as standard, may often be used to achieve the same analytical goals but without the reduced signal-to-noise penalty. Anthony J. Owen is product manager at Agilent Technologies, Waldbronn, Germany. For the latest information and services visit our world wide web site: Copyright 1995 Agilent Technologies All Rights Reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed under the copyright laws. Publication Number E Agilent Technologies Innovating the HP Way

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

FTIR Instrumentation

FTIR Instrumentation FTIR Instrumentation Adopted from the FTIR lab instruction by H.-N. Hsieh, New Jersey Institute of Technology: http://www-ec.njit.edu/~hsieh/ene669/ftir.html 1. IR Instrumentation Two types of instrumentation

More information

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD

The Fundamentals of Infrared Spectroscopy. Joe Van Gompel, PhD TN-100 The Fundamentals of Infrared Spectroscopy The Principles of Infrared Spectroscopy Joe Van Gompel, PhD Spectroscopy is the study of the interaction of electromagnetic radiation with matter. The electromagnetic

More information

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions.

EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. EXPERIMENT 11 UV/VIS Spectroscopy and Spectrophotometry: Spectrophotometric Analysis of Potassium Permanganate Solutions. Outcomes After completing this experiment, the student should be able to: 1. Prepare

More information

Copyright 1999 2010 by Mark Brandt, Ph.D. 12

Copyright 1999 2010 by Mark Brandt, Ph.D. 12 Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits

More information

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance: Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T

More information

Reaction Stoichiometry and the Formation of a Metal Ion Complex

Reaction Stoichiometry and the Formation of a Metal Ion Complex Reaction Stoichiometry and the Formation of a Metal Ion Complex Objectives The objectives of this laboratory are as follows: To use the method of continuous variation to determine the reaction stoichiometry

More information

Signal to Noise Instrumental Excel Assignment

Signal to Noise Instrumental Excel Assignment Signal to Noise Instrumental Excel Assignment Instrumental methods, as all techniques involved in physical measurements, are limited by both the precision and accuracy. The precision and accuracy of a

More information

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law

Austin Peay State University Department of Chemistry Chem 1111. The Use of the Spectrophotometer and Beer's Law Purpose To become familiar with using a spectrophotometer and gain an understanding of Beer s law and it s relationship to solution concentration. Introduction Scientists use many methods to determine

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs

Spectroscopy. Biogeochemical Methods OCN 633. Rebecca Briggs Spectroscopy Biogeochemical Methods OCN 633 Rebecca Briggs Definitions of Spectrometry Defined by the method used to prepare the sample 1. Optical spectrometry Elements are converted to gaseous atoms or

More information

The Measurement of Sensitivity in Fluorescence Spectroscopy

The Measurement of Sensitivity in Fluorescence Spectroscopy The Measurement of Sensitivity in Fluorescence Spectroscopy Among instrumental techniques, fluorescence spectroscopy is recognized as one of the more sensitive. In fluorescence, the intensity of the emission

More information

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES

UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES UV/VIS/IR SPECTROSCOPY ANALYSIS OF NANOPARTICLES SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published this

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information

Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information 1 Ultraviolet-Visible (UV-Vis) Spectroscopy Background Information Instructions for the Operation of the Cary 300 Bio UV-Visible Spectrophotometer See the Thermo OMNIC Help reference on page 49. Ultraviolet-Visible

More information

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry

Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab

More information

The accurate calibration of all detectors is crucial for the subsequent data

The accurate calibration of all detectors is crucial for the subsequent data Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved

More information

Using the Spectrophotometer

Using the Spectrophotometer Using the Spectrophotometer Introduction In this exercise, you will learn the basic principals of spectrophotometry and and serial dilution and their practical application. You will need these skills to

More information

Making OSNR Measurements In a Modulated DWDM Signal Environment

Making OSNR Measurements In a Modulated DWDM Signal Environment Making OSNR Measurements In a Modulated DWDM Signal Environment Jack Dupre Jim Stimple Making OSNR measurements in a modulated DWDM signal environment May 2001 In a DWDM spectrum, it is desirable to measure

More information

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Application Note RF & Microwave Spectrum Analyzers Table of Contents 3 3 4 4 5 7 8 8 13 13 14 16 16 Introduction Absolute versus relative

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Introduction to Fourier Transform Infrared Spectrometry

Introduction to Fourier Transform Infrared Spectrometry Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,

More information

The front end of the receiver performs the frequency translation, channel selection and amplification of the signal.

The front end of the receiver performs the frequency translation, channel selection and amplification of the signal. Many receivers must be capable of handling a very wide range of signal powers at the input while still producing the correct output. This must be done in the presence of noise and interference which occasionally

More information

Upon completion of this lab, the student will be able to:

Upon completion of this lab, the student will be able to: 1 Learning Outcomes EXPERIMENT B4: CHEMICAL EQUILIBRIUM Upon completion of this lab, the student will be able to: 1) Analyze the absorbance spectrum of a sample. 2) Calculate the equilibrium constant for

More information

Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT)

Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT) Page 1 Electronic Communications Committee (ECC) within the European Conference of Postal and Telecommunications Administrations (CEPT) ECC RECOMMENDATION (06)01 Bandwidth measurements using FFT techniques

More information

Module 13 : Measurements on Fiber Optic Systems

Module 13 : Measurements on Fiber Optic Systems Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)

More information

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.

QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11. QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition

More information

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy

Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy Fiber Optic Sampling by UV/Vis and UV/Vis/NIR Spectroscopy UV/VIS AND UV/VIS/NIR SPECTROSCOPY A P P L I C A T I O N N O T E Introduction The availability and applications of fiber optic sampling in UV/Vis

More information

INTRODUCTION. The principles of which are to:

INTRODUCTION. The principles of which are to: Taking the Pain Out of Chromatographic Peak Integration Shaun Quinn, 1 Peter Sauter, 1 Andreas Brunner, 1 Shawn Anderson, 2 Fraser McLeod 1 1 Dionex Corporation, Germering, Germany; 2 Dionex Corporation,

More information

Supporting Information

Supporting Information S1 Supporting Information GFT NMR, a New Approach to Rapidly Obtain Precise High Dimensional NMR Spectral Information Seho Kim and Thomas Szyperski * Department of Chemistry, University at Buffalo, The

More information

2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY

2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY 2nd/3rd Year Physical Chemistry Practical Course, Oxford University 2.02 DETERMINATION OF THE FORMULA OF A COMPLEX BY SPECTROPHOTOMETRY (4 points) Outline Spectrometry is widely used to monitor the progress

More information

AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis

AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis Analysis of Amino Acid Standards Label free analysis using the HPCE-512 ABSTRACT Capillary electrophoresis using indirect UV detection

More information

5 Numerical Differentiation

5 Numerical Differentiation D. Levy 5 Numerical Differentiation 5. Basic Concepts This chapter deals with numerical approximations of derivatives. The first questions that comes up to mind is: why do we need to approximate derivatives

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm?

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm? Test IV Name 1) In a single slit diffraction experiment, the width of the slit is 3.1 10-5 m and the distance from the slit to the screen is 2.2 m. If the beam of light of wavelength 600 nm passes through

More information

UV-Visible Spectroscopy

UV-Visible Spectroscopy UV-Visible Spectroscopy UV-Visible Spectroscopy What is UV-Visible Spectroscopy? Molecular spectroscopy that involves study of the interaction of Ultra violet (UV)-Visible radiation with molecules What

More information

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR

Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Agilent Cary 4000/5000/6000i Series UV-Vis-NIR Guaranteed specifications Design overview Double beam, ratio recording, double out-of-plane Littrow monochromator UV-Vis-NIR spectrophotometer (Agilent Cary

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

Blue Laser Technology Applied to the Microtrac Unified Scatter Technique for Full- Range Particle Size Measurement.

Blue Laser Technology Applied to the Microtrac Unified Scatter Technique for Full- Range Particle Size Measurement. Blue Laser Technology Applied to the Microtrac Unified Scatter Technique for Full- Range Particle Size Measurement. Philip E. Plantz, PhD Application Note SL-AN-25 Revision A Provided By: Microtrac, Inc.

More information

5.33 Lecture Notes: Introduction to Spectroscopy

5.33 Lecture Notes: Introduction to Spectroscopy 5.33 Lecture Notes: ntroduction to Spectroscopy What is spectroscopy? Studying the properties of matter through its interaction with different frequency components of the electromagnetic spectrum. Latin:

More information

2 Spectrophotometry and the Analysis of Riboflavin

2 Spectrophotometry and the Analysis of Riboflavin 2 Spectrophotometry and the Analysis of Riboflavin Objectives: A) To become familiar with operating the Platereader; B) to learn how to use the Platereader in determining the absorption spectrum of a compound

More information

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1

ANALYZER BASICS WHAT IS AN FFT SPECTRUM ANALYZER? 2-1 WHAT IS AN FFT SPECTRUM ANALYZER? ANALYZER BASICS The SR760 FFT Spectrum Analyzer takes a time varying input signal, like you would see on an oscilloscope trace, and computes its frequency spectrum. Fourier's

More information

Spectrophotometric Determination of the pka of Bromothymol Blue

Spectrophotometric Determination of the pka of Bromothymol Blue Spectrophotometric Determination of the pka of Bromothymol Blue INRODUCION cidbase indicators are compounds that are simply weak acids (or bases) that exhibit different colors depending on whether they

More information

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION 1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.

More information

Characterizing Digital Cameras with the Photon Transfer Curve

Characterizing Digital Cameras with the Photon Transfer Curve Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently

More information

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

Function Guide for the Fourier Transformation Package SPIRE-UOL-DOC-002496

Function Guide for the Fourier Transformation Package SPIRE-UOL-DOC-002496 Function Guide for the Fourier Transformation Package SPIRE-UOL-DOC-002496 Prepared by: Peter Davis (University of Lethbridge) peter.davis@uleth.ca Andres Rebolledo (University of Lethbridge) andres.rebolledo@uleth.ca

More information

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements.

Reflectance Measurements of Materials Used in the Solar Industry. Selecting the Appropriate Accessories for UV/Vis/NIR Measurements. T e c h n i c a l N o t e Reflectance Measurements of Materials Used in the Solar Industry UV/Vis/NIR Author: Dr. Jeffrey L. Taylor PerkinElmer, Inc. 710 Bridgeport Avenue Shelton, CT 06484 USA Selecting

More information

High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering

High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering High-Concentration Submicron Particle Size Distribution by Dynamic Light Scattering Power spectrum development with heterodyne technology advances biotechnology and nanotechnology measurements. M. N. Trainer

More information

A Beer s Law Experiment

A Beer s Law Experiment A Beer s Law Experiment Introduction There are many ways to determine concentrations of a substance in solution. So far, the only experiences you may have are acid-base titrations or possibly determining

More information

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION

ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION Chem 306 Section (Circle) M Tu W Th Name Partners Date ANALYSIS OF ASPIRIN INFRARED (IR) SPECTROSCOPY AND MELTING POINT DETERMINATION Materials: prepared acetylsalicylic acid (aspirin), stockroom samples

More information

Sampling Theorem Notes. Recall: That a time sampled signal is like taking a snap shot or picture of signal periodically.

Sampling Theorem Notes. Recall: That a time sampled signal is like taking a snap shot or picture of signal periodically. Sampling Theorem We will show that a band limited signal can be reconstructed exactly from its discrete time samples. Recall: That a time sampled signal is like taking a snap shot or picture of signal

More information

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies

Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies Activitity (of a radioisotope): The number of nuclei in a sample undergoing radioactive decay in each second. It is commonly expressed in curies (Ci), where 1 Ci = 3.7x10 10 disintegrations per second.

More information

Maximizing Receiver Dynamic Range for Spectrum Monitoring

Maximizing Receiver Dynamic Range for Spectrum Monitoring Home Maximizing Receiver Dynamic Range for Spectrum Monitoring Brian Avenell, National Instruments Corp., Austin, TX October 15, 2012 As consumers continue to demand more data wirelessly through mobile

More information

Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox. Application Note

Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox. Application Note Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox Application Note Introduction Of all the signal engines in the N7509A, the most complex is the multi-tone engine. This application

More information

where h = 6.62 10-34 J s

where h = 6.62 10-34 J s Electromagnetic Spectrum: Refer to Figure 12.1 Molecular Spectroscopy: Absorption of electromagnetic radiation: The absorptions and emissions of electromagnetic radiation are related molecular-level phenomena

More information

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry

Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry Spectrophotometry Practical Lesson on Medical Chemistry and Biochemistry General Medicine Jiřina Crkovská (translated by Jan Pláteník) 2010/2011 1 Spectrophotometry is one of the most widely used instrumental

More information

Extraction Heights for STIS Echelle Spectra

Extraction Heights for STIS Echelle Spectra Instrument Science Report STIS 98-09 Extraction Heights for STIS Echelle Spectra Claus Leitherer and Ralph Bohlin March 1998 ABSTRACT The optimum extraction height (h) for STIS echelle spectra of 7 pixels

More information

Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer. NAME: Lab Section: Date: Sign-Off:

Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer. NAME: Lab Section: Date: Sign-Off: Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer NAME: Lab Section: Date: Sign-Off: Chapter 5 -- The Spectrophotometric Determination of the ph of a Buffer Introduction Weak acids,

More information

MATRIX TECHNICAL NOTES

MATRIX TECHNICAL NOTES 200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR

More information

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1. Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used

More information

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz

Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Author: Don LaFontaine Making Accurate Voltage Noise and Current Noise Measurements on Operational Amplifiers Down to 0.1Hz Abstract Making accurate voltage and current noise measurements on op amps in

More information

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19 Doppler Doppler Chapter 19 A moving train with a trumpet player holding the same tone for a very long time travels from your left to your right. The tone changes relative the motion of you (receiver) and

More information

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property

More information

Improving Chromatic Dispersion and PMD Measurement Accuracy

Improving Chromatic Dispersion and PMD Measurement Accuracy Improving Chromatic Dispersion and PMD Measurement Accuracy White Paper Michael Kelly Agilent Technologies Signal transmission over optical fibers relies on preserving the waveform from transmitter to

More information

Lab #11: Determination of a Chemical Equilibrium Constant

Lab #11: Determination of a Chemical Equilibrium Constant Lab #11: Determination of a Chemical Equilibrium Constant Objectives: 1. Determine the equilibrium constant of the formation of the thiocyanatoiron (III) ions. 2. Understand the application of using a

More information

Reaction of Blue Food Dye with Bleach

Reaction of Blue Food Dye with Bleach Exercise 2 Reaction of Blue Food Dye with Bleach 2 Introduction In the experiment, you will study the rate of the reaction of FD&C Blue #1 (Blue #1 is denoted by E number E133 in food stuff) with sodium

More information

VWR SPECTROPHOTOMETERS

VWR SPECTROPHOTOMETERS VWR SPECTROPHOTOMETERS Reliable Accurate Easy to Use New_Spectrophotometry_v2.indd 2 Your first choice for spectrophotometry 5/9/2014 2:22:18 PM Quality Guaranteed VWR SPECTROPHOTOMETERS Single Beam Units

More information

Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2

Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2 ENERGY RESEARCH at the University of Oulu 117 Gas emission measurements with a FTIR gas analyzer - verification of the analysis method Kari Pieniniemi 1 * and Ulla Lassi 1, 2 1 University of Oulu, Department

More information

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8) Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:

More information

Transmission Lines. Smith Chart

Transmission Lines. Smith Chart Smith Chart The Smith chart is one of the most useful graphical tools for high frequency circuit applications. The chart provides a clever way to visualize complex functions and it continues to endure

More information

Molecular Spectroscopy

Molecular Spectroscopy Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended

More information

BOTDR Measurement Techniques and Brillouin Backscatter Characteristics of Corning Single-Mode Optical Fibers

BOTDR Measurement Techniques and Brillouin Backscatter Characteristics of Corning Single-Mode Optical Fibers BOTDR Measurement Techniques and Brillouin Backscatter Characteristics of Corning Single-Mode Optical Fibers WP4259 Issued: January 2015 Brillouin Optical Time Domain Reflectometry The Brillouin Optical

More information

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics

Jeff Thomas Tom Holmes Terri Hightower. Learn RF Spectrum Analysis Basics Jeff Thomas Tom Holmes Terri Hightower Learn RF Spectrum Analysis Basics Agenda Overview: Spectrum analysis and its measurements Theory of Operation: Spectrum analyzer hardware Frequency Specifications

More information

COLORIMETER. Description 0358BT. Figure 1. The Colorimeter

COLORIMETER. Description 0358BT. Figure 1. The Colorimeter COLORIMETER Description 0358BT Figure 1. The Colorimeter Introduction The Colorimeter is designed to determine the concentration of a solution by analyzing its color intensity. Monochromatic light from

More information

Scanners and How to Use Them

Scanners and How to Use Them Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types

More information

Ultraviolet Spectroscopy

Ultraviolet Spectroscopy Ultraviolet Spectroscopy The wavelength of UV and visible light are substantially shorter than the wavelength of infrared radiation. The UV spectrum ranges from 100 to 400 nm. A UV-Vis spectrophotometer

More information

Fiber Optics: Fiber Basics

Fiber Optics: Fiber Basics Photonics Technical Note # 21 Fiber Optics Fiber Optics: Fiber Basics Optical fibers are circular dielectric wave-guides that can transport optical energy and information. They have a central core surrounded

More information

DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED

DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED DYNAMIC LIGHT SCATTERING COMMON TERMS DEFINED Abstract: There are a number of sources of information that give a mathematical description of the terms used in light scattering. However, these will not

More information

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS

TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS 1. Bandwidth: The bandwidth of a communication link, or in general any system, was loosely defined as the width of

More information

7. DYNAMIC LIGHT SCATTERING 7.1 First order temporal autocorrelation function.

7. DYNAMIC LIGHT SCATTERING 7.1 First order temporal autocorrelation function. 7. DYNAMIC LIGHT SCATTERING 7. First order temporal autocorrelation function. Dynamic light scattering (DLS) studies the properties of inhomogeneous and dynamic media. A generic situation is illustrated

More information

Raman Spectroscopy Basics

Raman Spectroscopy Basics Raman Spectroscopy Basics Introduction Raman spectroscopy is a spectroscopic technique based on inelastic scattering of monochromatic light, usually from a laser source. Inelastic scattering means that

More information

PHYS 331: Junior Physics Laboratory I Notes on Noise Reduction

PHYS 331: Junior Physics Laboratory I Notes on Noise Reduction PHYS 331: Junior Physics Laboratory I Notes on Noise Reduction When setting out to make a measurement one often finds that the signal, the quantity we want to see, is masked by noise, which is anything

More information

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff

Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction

More information

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F)

Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F) Hitachi U-4100 UV-vis-NIR spectrophotometer (341-F) Please contact Dr. Amanda Young for training requests and assistance: 979-862-6845, amandayoung@tamu.edu Hardware Our spectrophotometer is made up of

More information

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS 1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years

More information

Fraunhofer Diffraction

Fraunhofer Diffraction Physics 334 Spring 1 Purpose Fraunhofer Diffraction The experiment will test the theory of Fraunhofer diffraction at a single slit by comparing a careful measurement of the angular dependence of intensity

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

NRZ Bandwidth - HF Cutoff vs. SNR

NRZ Bandwidth - HF Cutoff vs. SNR Application Note: HFAN-09.0. Rev.2; 04/08 NRZ Bandwidth - HF Cutoff vs. SNR Functional Diagrams Pin Configurations appear at end of data sheet. Functional Diagrams continued at end of data sheet. UCSP

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Measuring Protein Concentration through Absorption Spectrophotometry

Measuring Protein Concentration through Absorption Spectrophotometry Measuring Protein Concentration through Absorption Spectrophotometry In this lab exercise you will learn how to homogenize a tissue to extract the protein, and then how to use a protein assay reagent to

More information

Application Note: Absorbance

Application Note: Absorbance Units Units Theory of absorbance Light absorption occurs when atoms or molecules take up the energy of a photon of light, thereby reducing the transmission of light as it is passed through a sample. Light

More information

INTERNATIONAL OLIVE COUNCIL

INTERNATIONAL OLIVE COUNCIL INTERNATIONAL OLIVE COUNCIL COI/T.20/Doc. No 19/Rev. 3 February 2015 ENGLISH Original: ENGLISH Príncipe de Vergara, 154 28002 Madrid España Telef.: +34 915 903 638 Fax: +34 915 631 263 - e-mail: iooc@internationaloliveoil.org

More information

HD Radio FM Transmission System Specifications Rev. F August 24, 2011

HD Radio FM Transmission System Specifications Rev. F August 24, 2011 HD Radio FM Transmission System Specifications Rev. F August 24, 2011 SY_SSS_1026s TRADEMARKS HD Radio and the HD, HD Radio, and Arc logos are proprietary trademarks of ibiquity Digital Corporation. ibiquity,

More information

Optimizing IP3 and ACPR Measurements

Optimizing IP3 and ACPR Measurements Optimizing IP3 and ACPR Measurements Table of Contents 1. Overview... 2 2. Theory of Intermodulation Distortion... 2 3. Optimizing IP3 Measurements... 4 4. Theory of Adjacent Channel Power Ratio... 9 5.

More information

Introduction to Receivers

Introduction to Receivers Introduction to Receivers Purpose: translate RF signals to baseband Shift frequency Amplify Filter Demodulate Why is this a challenge? Interference (selectivity, images and distortion) Large dynamic range

More information

Elasticity. I. What is Elasticity?

Elasticity. I. What is Elasticity? Elasticity I. What is Elasticity? The purpose of this section is to develop some general rules about elasticity, which may them be applied to the four different specific types of elasticity discussed in

More information

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number

1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number 1) Write the following as an algebraic expression using x as the variable: Triple a number subtracted from the number A. 3(x - x) B. x 3 x C. 3x - x D. x - 3x 2) Write the following as an algebraic expression

More information

AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE

AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE Atmel 8-bit and 32-bit Microcontrollers AVR127: Understanding ADC Parameters APPLICATION NOTE Introduction This application note explains the basic concepts of analog-to-digital converter (ADC) and the

More information