Agilent AN Applying Error Correction to Network Analyzer Measurements

Similar documents
Applying Error Correction to Network Analyzer Measurements. Application Note Table of Contents. Page

Agilent Measuring Noninsertable Devices

Agilent Test Solutions for Multiport and Balanced Devices

Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers

Agilent 8753ET/8753ES Network Analyzers

Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows

MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER

Agilent 8720 Family Microwave Vector Network Analyzers

Network Analyzer Basics- EE142 Fall 07

Agilent AN Improving Throughput in Network Analyzer Applications. Application Note

81110A Pulse Pattern Generator Simulating Distorted Signals for Tolerance Testing

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

February 2010 Number Vector Network Analyzer upgrade to SureCAL Power Sensor and RF Components Packages

Agilent PN In-fixture Microstrip Device Measurements Using TRL* Calibration. Product Note

Agilent AN In-Fixture Measurements Using Vector Network Analyzers

Experiment 7: Familiarization with the Network Analyzer

2. The Vector Network Analyzer

Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer. Application Note

Impedance 50 (75 connectors via adapters)

Agilent PN RF Component Measurements: Amplifier Measurements Using the Agilent 8753 Network Analyzer. Product Note

Agilent 10 Hints for Making Better Network Analyzer Measurements. Application Note B

Agilent Balanced Measurement Example: Differential Amplifiers

Agilent Split Post Dielectric Resonators for Dielectric Measurements of Substrates. Application Note

Agilent Evolution of Test Automation Using the Built-In VBA with the ENA Series RF Network Analyzers

Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements. Application Note

Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox. Application Note

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

Agilent N2717A Service Software Performance Verification and Adjustment Software for the Agilent ESA Spectrum Analyzers Product Overview

Agilent Technologies E8047B Analysis Probe System for the Intel Xeon Processor Family

Calculating Antenna System Return Loss As Viewed Through The RF Path

Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters

Michael Hiebel. Fundamentals of Vector Network Analysis

Agilent Network Analysis Applying the 8510 TRL Calibration for Non-Coaxial Measurements

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER

Advanced Calibration Techniques for Vector Network Analyzers

Power Amplifier Gain Compression Measurements

Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements

Category 8 Cable Transmission Measurements Comparative Study between 4-port single wire measurements and 2-port balun measurements

0HDVXULQJWKHHOHFWULFDOSHUIRUPDQFH FKDUDFWHULVWLFVRI5),)DQGPLFURZDYHVLJQDO SURFHVVLQJFRPSRQHQWV

Network Analysis. Specifying calibration standards for the HP 8510 network analyzer Product Note A

Application Note Noise Frequently Asked Questions

R&S ZCxxx Millimeter-Wave Converters Specifications

RF and Microwave Accessories. CD-ROM Catalog. Find the right component for your Rohde & Schwarz test & measurement equipment

GigaSPEED X10D Solution How

1. The Slotted Line. ECE 584 Microwave Engineering Laboratory Experiments. Introduction:

R3765/67 CG Network Analyzer

Scalar Network Analysis with the HP 8590 Series Spectrum Analyzers Product Overview

Vector Network Analyzer (VNA) Calibration: The Basics

spinner Measurement & Calibration equipment for network analyzers

DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD OF INSERTION LOSS MEASUREMENT

Agilent PN Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth

Designing the NEWCARD Connector Interface to Extend PCI Express Serial Architecture to the PC Card Modular Form Factor

Understanding Power Splitters

Cable Analysis and Fault Detection using the Bode 100

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

Utilizing Time Domain (TDR) Test Methods For Maximizing Microwave Board Performance

AUTOMATIC NETWORK ANALYZER PROCEDURES FOR 5045 KLYSTRON CAVITIES * J.G. Judkins

Agilent 8509C Lightwave Polarization Analyzer

Agilent Specifying Calibration Standards for the Agilent 8510 Network Analyzer

Improving Network Analyzer Measurements of Frequency-translating Devices Application Note

Understanding Mixers Terms Defined, and Measuring Performance

Agilent 8762/3/4A,B,C Coaxial Switches

Keysight Technologies Understanding the Fundamental Principles of Vector Network Analysis. Application Note

Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note

Selecting RJ Bandwidth in EZJIT Plus Software

The front end of the receiver performs the frequency translation, channel selection and amplification of the signal.

UNDERSTANDING NOISE PARAMETER MEASUREMENTS (AN )

Understanding Power Splitters

"FP", "FR", "FQ" Series Bandpass Filters

Understanding the Fundamental Principles of Vector Network Analysis. Application Note Table of Contents. Page

Agilent E5063A ENA Series Network Analyzer. 100 khz to 4.5/ 8.5/18 GHz

Agilent E5100A Network Analyzer

HEWLETT PACKARD. HP 8510C Network Analyzer 45 MHz to 110 GHz. Unmatched excellence in microwave network analysis

Improving Chromatic Dispersion and PMD Measurement Accuracy

Agilent Stripline TRL Calibration Fixtures for 10-Gigabit Interconnect Analysis. White Paper

Balun Parameter Definitions & Measurement May 2004

Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer

Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz

Performing Amplifier Measurements with the Vector Network Analyzer ZVB

Cable Impedance and Structural Return Loss Measurement Methodologies

Laboratory #5: RF Filter Design

HP 8970B Option 020. Service Manual Supplement

Capacitor Self-Resonance

a 1 a 2 2 Port b 2 b 1 Multi-Port Handset Switch S-Parameters Application Note AN20 Seven-Port S-Parameter Definition Introduction Summary:

Agilent Advanced impedance measurement capability of the RF I-V method compared to the network analysis method. Application Note

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term

Measurement of Inductor Q with the MSA Sam Wetterlin 3/31/11. Equation 1 Determining Resonant Q from Inductor Q and Capacitor Q

Agilent E6832A W-CDMA Calibration Application

Shielding Effectiveness Test Method. Harbour s LL, SB, and SS Coaxial Cables. Designs for Improved Shielding Effectiveness

HIGH RELIABILITY POWER SUPPLY TESTING

S-Parameters and Related Quantities Sam Wetterlin 10/20/09

DDX 7000 & Digital Partial Discharge Detectors FEATURES APPLICATIONS

Agilent GSM/EDGE Base Station Test with the E4406A VSA and ESG-D Series RF Signal Generators Product Overview

Noise Figure Measurement Accuracy The Y-Factor Method. Application Note 57-2

RF Network Analyzer Basics

Agilent E6481A OPAS32 Engineering Information Management and Analysis Software Configuration Guide

High Power Amplifier Measurements Using Agilent s Nonlinear Vector Network Analyzer

MEASUREMENT OF CRYOGENIC PERFORMANCE OF 4-8 GHz PAMTECH ISOLATORS S/N Juan Daniel Gallego Isaac López Fernández Carmen Diez González

Transcription:

Agilent AN 1287-3 Applying rror Correction to Network Analyzer Measurements Application Note 2 3 4 4 5 6 7 8 10 11 12 12 13 13 14 Table of Contents Introduction Sources and Types of rrors Types of rror Correction One-Port Calibration The ffects of Adapters Two-Port rror Correction lectronic Calibration stimating Measurement Uncertainty Performing a Transmission Response Calibration nhanced-response Calibration for Transmission Measurements Full Two-Port Calibration TRL Calibration Calibrating Noninsertable Devices Swap-qual-Adapters Method Adapter-Removal Calibration

Introduction Only perfect test equipment would not need correction. Imperfections exist in even the finest test equipment and cause less than ideal measurement results. Some of the factors that contribute to measurement errors are repeatable and predictable over time and temperature and can be removed, while others are random and cannot be removed. The basis of network analyzer error correction is the measurement of known electrical standards, such as a through, open circuit, short circuit, and precision load impedance. The effect of error correction on displayed data can be dramatic (Figure 1). Without error correction, measurements on a bandpass filter show considerable loss and ripple. The smoother, error-corrected trace produced by a twoport calibration subtracts the effects of systematic errors and better represents the actual performance of the device under test (). This application note describes several types of calibration procedures, including the popular Short-Open-Load-Through (SOLT) calibration technique, and Through-Reflect-Line (TRL). The effectiveness of these procedures will then be demonstrated in the measurement of high-frequency components such as filters. Calibrations will also be shown for those cases requiring coaxial adapters to connect the test equipment,, and calibration standards. Measuring filter insertion loss CH1 S 21 &M log MAG 1 db/ RF 0 db CH2 MM log MAG 1 db/ RF 0 db Cor After two-port calibration After response calibration Uncorrected Cor x2 1 2 START 2 000.000 MHz STOP 6 000.000 MHz Figure 1. Response versus Two-Port Calibration 2

Sources and Types of rrors All measurement systems, including those employing network analyzers, can be plagued by three types of measurement errors: Systematic errors Random errors Drift errors Systematic errors (Figure 2) are caused by imperfections in the test equipment and test setup. If these errors do not vary over time, they can be characterized through calibration and mathematically removed during the measurement process. Systematic errors encountered in network measurements are related to signal leakage, signal reflections, and frequency response. There are six types of systematic errors: Directivity and crosstalk errors relating to signal leakage Source and load impedance mismatches relating to reflections Frequency response errors caused by reflection and transmission tracking within the test receivers (The full two-port error model includes all six of these terms for the forward direction and the same six (with different data) in the reverse direction, for a total of twelve error terms. This is why two-port calibration is often referred to as twelve-term error correction.) R Directivity A Crosstalk B Frequency response reflection tracking (A/R) transmission tracking (B/R) Source Mismatch Load Mismatch Six forward and six reverse error terms yields 12 error terms for two-port devices Figure 2. Systematic Measurement rrors Random errors vary randomly as a function of time. Since they are not predictable, they cannot be removed by calibration. The main contributors to random errors are instrument noise (e.g., sampler noise, and the IF noise floor), switch repeatability, and connector repeatability. When using network analyzers, noise errors can often be reduced by increasing source power, narrowing the IF bandwidth, or by using trace averaging over multiple sweeps. 3

Drift errors occur when a test system s performance changes after a calibration has been performed. They are primarily caused by temperature variation and can be removed by additional calibration. The rate of drift determines how frequently additional calibrations are needed. However, by constructing a test environment with stable ambient temperature, drift errors can usually be minimized. While test equipment may be specified to operate over a temperature range of 0 C to +55 C, a more controlled temperature range such as +25 C ± 5 C can improve measurement accuracy (and reduce or eliminate the need for periodic recalibration) by minimizing drift errors. Types of rror Correction There are two basic types of error correction response (normalization) corrections, and vector corrections. Response calibration is simple to perform, but corrects for only a few of the 12 possible systematic error terms (namely, reflection and transmission tracking). Response calibration is a normalized measurement in which a reference trace is stored in the network analyzer s memory, and the stored trace is divided into measurement data for normalization. A more advanced form of response calibration for reflection measurements, called open/short averaging, is commonly found on scalar analyzers and averages two traces to derive a reference trace. Vector error correction is a more thorough method of removing systematic errors. This type of error correction requires a network analyzer capable of measuring (but not necessarily displaying) phase as well as magnitude, and a set of calibration standards with known, precise electrical characteristics. The vector-correction process characterizes systematic error terms by measuring known calibration standards, storing these measurements within the analyzer s memory, and using this data to calculate an error model which is then used to remove the effects of systematic errors from subsequent measurements. This calibration process accounts for all major sources of systematic errors and permits very accurate measurements. However, it requires more standards and more measurements than response calibration. The two main types of vector error correction are the one-port and two-port calibrations. One-Port Calibration A one-port calibration can measure and remove three systematic error terms (directivity, source match, and reflection tracking) from reflection measurements. These three error terms are derived from a general equation which can be solved in terms of three simultaneous equations with three unknowns. To establish these equations, three known calibration standards must be measured, such as an open, a short, and a load (the load value is usually the same as the characteristic impedance of the test system, generally either 50 or 75 ohm). Solving the equations yields the systematic error terms and makes it possible to derive the s actual reflection S-parameters. 4

When measuring two-port devices, a one-port calibration assumes a good termination on the unused port of the. If this condition is met (by connecting a load standard, for example), the one-port calibration is quite accurate. However, if port two of the is connected to the network analyzer and its reverse isolation is low (for example, filter passbands or low-loss cables), the assumption of a good load termination is often not valid. In this case, two-port error correction can provide significantly better results than one-port correction. An amplifier is a good example of a two-port device in which the load match presented by the network analyzer does not affect measurements of the amplifier s input match, because the reverse isolation of the amplifier allows one-port calibration to be effective. In Figure 3, a reflection measurement is shown with and without a one-port calibration. Without error correction, the classic ripple pattern appears, which is caused by systematic errors interfering with the test signal. The error-corrected trace is much smoother and better represents the device s actual reflection performance. 0 Data Before rror Correction 2.0 Return Loss (db) 20 40 1.1 1.01 VSWR 60 Data After rror Correction 1.001 6000 12000 MHz Figure 3. Before and After One-Port Calibration The ffects of Adapters Ideally, reflection calibrations should be performed with a calibration kit having the same type connectors as the. If adapters are necessary to make connections, the effects of these adapters must then be considered as part of the measurement uncertainty. An adapter added to a network analyzer test port after a calibration has been done may cause errors that add to or subtract from the desired signal from the (Figure 4). This error is often ignored, which may not be acceptable. Worst-case effective directivity in this case is the sum of the corrected directivity and the reflection (r) of the adapter. An adapter with a VSWR of 1.5:1 for example, will reduce the effective directivity of a test coupler to about 14 db, even if the coupler itself has infinite directivity. So with an ideal load on the output of the adapter, the reflected signal appearing at the coupled port will be only 14 db less than the reflection from a short or open circuit. Stacking multiple adapters compounds the problem. If adapters cannot be avoided, the highest-quality types are always the best choice in order to reduce degradation of system directivity. rror correction can mitigate the effects of adapters on the test port, but the test system will be slightly more susceptible to drift because of degraded raw (uncorrected) directivity. 5

leakage signal Coupler directivity = 40 db reflection from adapter desired signal r total = r adapter + Adapter Termination r has SMA (f) connectors Worst-case System Directivity Adapting from APC-7 to SMA (m) APC-7 calibration done here 28 db APC-7 to SMA (m) SWR:1.06 17 db 14 db APC-7 to N (f) + N (m) to SMA (m) SWR:1.05 SWR:1.25 APC-7 to N (m) + N (f) to SMA (f) + SMA (m) to (m) SWR:1.05 SWR:1.25 SWR:1.15 Figure 4. Adapter Considerations Two-Port rror Correction Two-port error correction yields the most accurate results because it accounts for all of the major sources of systematic error. The error model for a two-port device reveals the four S-parameters measured in the forward and reverse directions (Figure 5). Once the system error terms have been characterized, the network analyzer utilizes four equations to derive the actual device S-parameters from the measured S-parameters. Because each S-parameter is a function of all four measured S-parameters, a network analyzer must make a forward and reverse test sweep before updating any one S-parameter. When performing a two-port calibration, the part of the calibration that characterizes crosstalk (isolation) can often be omitted. Crosstalk, which is signal leakage between test ports when no device is present, can be a problem when testing high-isolation devices such as a switch in the open position, and highdynamic-range devices such as filters with a high level of rejection. Forward model Reverse model Port 1 X Port 2 Port 1 Port 2 RT a 1 b 1 D S S 21 A S 11 S A 22 A TT L a 2 b 2 a 1 b 1 L TT S21 A S11 S22 D A A S S 12 A b 2 a 2 RT S 12 A X = Fwd Directivity D = Fwd Source Match S RT = Fwd Reflection Tracking D = Rev Directivity S = Rev Source Match RT = Rev Reflection Tracking = Fwd Load Match L = Fwd Transmission Tracking TT X = Fwd Isolation = Rev Load Match L TT = Rev Transmission Tracking X = Rev Isolation S RT S L RT TT a S RT S L L RT TT S TT RT a S L RT RT TT 11 = TT S12 S11m - D - S m D X S - S - m X 22 21 m 12 ( )( 1 + - ) ( )( ) - S D - TT S m D X S - - m X 11 m 22 21 m 1 + ( )( 1 + - ) ( )( ) 21 = S21m X - - S m D 22 ( )( 1 + S - L ( )) S - D - TT S L S m D X S - S - m X 11 m 22 21 m 12 1 + ( )( 1 + - ) ( )( ) Notice that each actual S-parameter is a function of all four measured S-parameters Analyzer must make forward and reverse sweep to update any one S-parameter TT RT S RT S L L RT TT S S L L RT TT - - S S 12 11 m X m S12 a = D ( 1 + - S L )( ( )) - S D - TT - S m D X S - S m X 11 m 22 21 m 12 1 + ( )( 1 + - ) ( )( ) - S22 m D - D RT S RT TT 22 a = TT S11 m - D S22 m - D S21 m - X S12 m S L ( )( 1+ - ) ( )( S S - - S X m X 11 m 21 m 12 ) RT TT - X ( 1 + )( 1 + - ) ( )( ) Figure 5. Two-Port rror Correction 6

Unfortunately, a crosstalk calibration can add noise to the error model because measurements are often made near the analyzer s noise floor. If the isolation calibration is deemed necessary, it should be performed with trace averaging to ensure that the test system s crosstalk is not obscured by noise. In some network analyzers, crosstalk can be minimized by using the alternate sweep mode instead of the chop mode (the chop mode makes measurements on both the reflection (A) and transmission (B) channels at each frequency point, whereas the alternate mode turns off the reflection receiver during the transmission measurement). The best way to perform an isolation calibration is by placing the devices that will be measured on each test port of the network analyzer, with terminations on the other two device ports. Using this technique, the network analyzer sees the same impedance versus frequency during the isolation calibration as it will during subsequent measurements of the. If this method is impractical (in test fixtures, or if only one is available, for example), than placing a terminated on the source port and a termination on the load port of the network analyzer is the next best alternative (the and termination must be swapped for the reverse measurement). If no is available or if the will be tuned (which will change its port matches), then terminations should be placed on each network analyzer test port for the isolation calibration. A network analyzer can be used for uncorrected measurements, or with any one of a number of calibration options, including response calibrations and one- or two-port vector calibrations. A summary of these calibrations is shown in Figure 6). UNCORRCTD RSPONS ON-PORT FULL TWO-PORT SHORT SHORT SHORT thru OPN OPN OPN Convenient Generally not accurate No errors removed Other errors: Random (Noise, Repeatability) Drift asy to perform Use when highest accuracy is not required Removes frequency response error NHANCD-RSPONS Combines response and 1-port Corrects source match for transmission measurements LOAD For reflection measurements Need good termination for high accuracy with two-port devices Removes these errors: Directivity Source match Reflection tracking LOAD thru LOAD Highest accuracy Removes these errors: Directivity Source, load match Reflection tracking Transmission tracking Crosstalk Figure 6. rrors and Calibration Standards lectronic Calibration Although Figure 6 shows mechanical calibration standards, Agilent Technologies offers a solid-state calibration solution which makes two-port calibration fast, easy, and less prone to operator errors. The various impedance states in the calibration modules are switched with PIN-diode or FT switches, so the calibration standards never wear out. The calibration modules are characterized at the Agilent factory using a TRL-calibrated network analyzer, making the Cal modules transfer standards (rather than direct standards). Cal provides excellent accuracy, with results generally better than SOLT calibration, but somewhat less than a properlyperformed TRL calibration. 7

stimating Measurement Uncertainty Figure 7 shows which systematic error terms are accounted for when using analyzers with transmission/reflection test sets (Agilent 8712T family, the 8753T and the 8720T family), and S-parameter test sets (Agilent 8753S, 8720S, Agilent 8510 family and the PNA Series). Some straightforward techniques can be used to determine measurement uncertainty when evaluating two-port devices with a network analyzer based on a transmission/reflection test set. For example, Figure 8 shows measurement of the input match of a filter after a one-port calibration has been performed. The filter has 16 db of return loss and 1 db of insertion loss. The raw load match of an 8712T network analyzer is specified to be 18 db (although it s often significantly better than this). The reflection from the test port connected to the filters output port is attenuated by twice the filter loss in this case, only 2 db. This value is not adequate to sufficiently suppress the effects of this error signal, which illustrates why low-loss devices are difficult to measure accurately. Reflection Test Set (cal type) T/R S-parameter (one-port) (two-port) SHORT Reflection tracking Directivity Source match Load match error can be corrected error cannot be corrected * Agilent 8712T enhanced response cal can correct for source match during transmission measurements Transmission Transmission Tracking Crosstalk Source match Load match OPN * ( ) LOAD Test Set (cal type) T/R S-parameter (response, (two-port) isolation) Figure 7. Calibration Summary Load match: 18 db (.126) Directivity: 40 db (.010) 16 db RL (.158) 1 db loss (.891).158 (.891)(.126)(.891) =.100 Measurement uncertainty: 20 * log (.158 +.100 +.010) = 11.4 db ( 4.6dB) 20 * log (.158.100.010) = 26.4 db (+10.4 db) Figure 8. Reflection xample Using a One-Port Cal 8

To determine the measurement uncertainty of this example, it is necessary to add and subtract the undesired reflection signal (with a reflection coefficient of 0.100) with the signal reflecting from the (0.158) (to be consistent with the next example, we will also include the effect of the directivity error signal). The measured return loss of the 16-dB filter may appear to be anywhere from 11.4 db to 26.4 db, allowing too much room for error. In production testing, these errors could easily cause filters which met specification to fail, while filters that actually did not meet specification might pass. In tuning applications, filters could be mistuned as operators try to compensate for the measurement error. When measuring an amplifier with good isolation between output and input (i.e., where the isolation is much greater than the gain), there is much less measurement uncertainty. This is because the reflection caused by the load match is severely attenuated by the product of the amplifier s isolation and gain. To improve measurement uncertainty for a filter, the output of the filter must be disconnected from the analyzer and terminated with a high-quality load, or a high-quality attenuator can be inserted between the filter and port 2 of the analyzer. Both techniques improve the analyzer s effective load match. As an example (Figure 9), if we placed a 10 db attenuator with a SWR of 1.05 between port 2 of the network analyzer and the filter used in the previous example, our effective load match would improve to 28.6 db. This value is the combination of a 32.3 db match from the attenuator and a 38 db match from the network analyzer (since the error signal travels through the attenuator twice, the analyzer s load match is improved by twice the value of the attenuator). Our worst-case uncertainty is now reduced to +2.5 db, 1.9 db, instead of the +10.4 db, 4.6 db we had without the 10 db attenuator. While not as good as what could be achieved with two-port calibration, this level of accuracy may be sufficient for manufacturing applications. Load match: 18 db (.126) Measurement uncertainty: 20 * log (.158 +.039) = 14.1 db ( 1.9 db) 20 * log (.158.039) = 18.5 db (+2.5 db) Directivity: 40 db (.010).158 10 db attenuator (.316) SWR = 1.05 (.024) 16 db RL (.158) 1 db loss (.891) (.891)(.316)(.126)(.316)(.891) =.010 (.891)(.024)(.891) =.019 Worst-case error =.010 +.010 +.019 =.039 Low-loss bi-directional devices generally require two-port calibration for low measurement uncertainty Figure 9. Reflection xample using a One-port Cal plus an Attenuator 9

Performing a Transmission Response Calibration Response calibrations offer simplicity, but with some compromise in accuracy. In making a filter transmission measurement using only response calibration, the first step is to make a through connection between the two test ports (with no in place). For this example, test port specifications for the Agilent 8712T network analyzer will be used. The ripple caused by this amount of mismatch is calculated as ±0.22 db, and is now present in the reference data (Figure 10). It must be added to the uncertainty when the is measured in order to compute worst-case overall measurement uncertainty. The same setup and test port specifications for the 8712T can be used to determine the measurement uncertainty with the in place. There are three main error signals caused by reflections between the ports of the analyzer and the (Figure 11). Higher-order reflections can be neglected because they are small compared to the three main terms. One of the error signals passes through the twice, so it is attenuated by twice the insertion loss of the. A worst-case condition occurs when all of the reflected error signals add together in phase (0.020 + 0.020 + 0.032 = 0.072). In that case, measurement uncertainty is +0.60/ 0.65 db. Total measurement uncertainty, which must include the 0.22 db of error incorporated into the calibration measurement, is about ±0.85 db. RL = 14 db (.200) RL = 18 db (.126) Thru calibration (normalization) builds error into measurement due to source and load match interaction Calibration Uncertainty = (1 ± r r ) S L = (1 ± (.200)(.126) = ± 0.22 db Figure 10. Transmission xample Using a Response Cal Filter measurement with response cal Source match = 14 db (.200) 1 db loss (.891) 16 db RL (.158) Load match = 18 db (.126) 1 (.126)(.158) =.020 (.126)(.891)(.200)(.891) =.020 Total measurement uncertainty: +0.60 + 0.22 = + 0.82 db -0.65-0.22 = - 0.87 db (.158)(.200) =.032 Measurement uncertainty = 1 ± (.020+.020+.032) = 1 ±.072 = + 0.60 db - 0.65 db Figure 11. Transmission xample (continued) 10

Another test example is an amplifier with a port match of 16 db. The test setup and conditions remain essentially the same as in the first two cases (Figure 12), except now the middle error term is no longer present because of the amplifier s reverse isolation. This reduces the measurement error to about ±0.45 db and the total measurement uncertainty to about ±0.67 db (compared to ±0.85 db for the filter). Source match = 14 db (.200) 16 db RL (.158) Load match = 18 db (.126) 1 (.126)(.158) =.020 (.158)(.200) =.032 Total measurement uncertainty: +0.44 + 0.22 = +0.66 db 0.46 0.22 = 0.68 db Measurement uncertainty = 1 ± (.020+.032) = 1 ±.052 = + 0.44 db 0.46 db Figure 12. Measuring Amplifiers with a Response Calibration nhanced-response Calibration for Transmission Measurements A feature of the 8712T/S family of economy network analyzers is their ability to perform an enhanced-response calibration. This calibration requires the measurement of short, open, load, and through standards for transmission measurements. The enhanced-response calibration combines a one-port calibration and a response calibration to allow correction of source match during transmission measurements, something a standard response calibration cannot do. The enhanced-response calibration (Figure 13) improves the effective source match during transmission measurements to about 35 db, compared to 14 db for normal response calibrations with the 8712T. This reduces the calibration error from ±0.22 db to ±0.02 db, and greatly reduces the two measurement error terms that involve interaction with the effective source match. The total measurement error is ±0.24 db instead of the previous value of ±0.85 db for a standard response calibration. While not as good as full two-port error correction, this represents a significant improvement over a standard response calibration and may be sufficient for many applications. Source match = 35 db (.0178) ffective source match = 35 db! 1 db loss (.891) 16 db RL (.158) Load match = 18 db (.126) (.126)(.158) =.020 (.126)(.891)(.0178)(.891) =.0018 Calibration Uncertainty = (1 ± r r ) S L = (1 ± (.0178) (.126) = ±.02 db 1 (.158)(.0178) =.0028 Measurement uncertainty = 1 ± (.020+.0018+.0028) = 1 ±.0246 = + 0.211 db 0.216 Total measurement uncertainty: 0.22 +.02 = ± 0.24 db Figure 13. Transmission Measurements using the nhanced-response Calibration 11

Full Two-Port Calibration In an example that calculates the measurement error after a two-port calibration (Figure 14), the worst-case measurement errors for the filter have been reduced to about ±0.5 db for reflection measurements and ±0.05 db for transmission measurements. Phase errors are similarly small. Corrected error terms: (8753D 1.3-3 GHz Type-N) Directivity = 47 db Source match = 36 db Load match = 47 db Refl. tracking =.019 db Trans. tracking =.026 db Isolation = 100 db 1 db loss (.891) 16 db RL (.158) Reflection uncertainty S 11m = S 11a ± ( D + S 11a 2 s + S 21a S 12a L + S 11a (1 RT )) = 0.158 ± (.0045 + 0.158 2 *.0158 + 0.891 2 *.0045 + 0.158*.0022) = 0.158 ±.0088 = 16 db + 0.53 db, 0.44 db (worst-case) Transmission uncertainty S 21m = S 21a ± S 21a ( I / S 21a + S 11a S + S 21a S 12a S L + S 22a L + (1 TT )) = 0.891± 0.891 (10 6 / 0.891 + 0.158*.0158 + 0.891 2 *0.158*.0045 + 0.158*.0045 +.003) = 0.891 ±.0056 = 1 db ± 0.05 db (worst-case) Figure 14. Calculating Measurement Uncertainty after a Two-Port Calibration TRL Calibration Following SOLT in popularity, the next most common form of two-port calibration is called a Through-Reflect-Line (TRL) calibration. It is primarily used in noncoaxial environments, such as testing waveguide, using test fixtures, or making on-wafer measurements with probes. TRL uses the same 12-term error model as a SOLT calibration, although with different calibration standards. TRL has two variants: True TRL calibration, which requires a network analyzer with four receivers TRL* calibration, developed for network analyzers with only three receivers Other variations of TRL are based on Line-Reflect-Match (LRM) calibration standards or Through-Reflect-Match (TRM) calibration standards. In differentiating TRL and TRL*, the latter assumes that the source and load match of a test port are equal that there is true port-impedance symmetry between forward and reverse measurements. This is only a fair assumption for a three-receiver network analyzer. TRL* requires 10 measurements to quantify 8 unknowns. True TRL requires four receivers (two reference receivers plus one each for reflection and transmission) and 14 measurements to solve for 10 unknowns. Both techniques use identical calibration standards. The Agilent 8720S network analyzer, which is normally equipped with only three samplers, can perform TRL calibrations when outfitted with Option 400 (which adds a fourth sampler). In noncoaxial applications, TRL achieves better source match and load match corrections than TRL*, resulting in less measurement error. In coaxial applications, SOLT is usually the preferred calibration technique. While not commonly used, coaxial TRL can provide more accuracy than SOLT, but only if very-high quality coaxial transmission lines (such as beadless airlines) are used. 12

Calibrating Noninsertable Devices When performing a through calibration, normally the test ports mate directly. For example, two cables with the appropriate connectors can be joined without a through adapter, resulting in a zero-length through path. An insertable device may substituted for a zero-length through. This device has the same connector type on each port but of the opposite sex, or the same sexless connector on each port, either of which makes connection to the test ports quite simple. A noninsertable device is one that can not be substituted for a zero-length through. It has the same type and sex connectors on each port or a different type of connector on each port, such as waveguide at one end and a coaxial connector on the other end. There are a few calibration choices available for noninsertable devices. The first is to use a characterized through adapter (electrical length and loss specified), which requires modifying the calibration kit definition. This will reduce (but not eliminate) source and load match errors. A high-quality through adapter (with good match) should be used since the match of the adapter cannot be characterized. Swap-qual-Adapters Method The swap-equal-adapters method requires the use of two precision matched adapters that are equal in performance but have connectors of different sexes. To be equal, the adapters must have the same match, characteristic impedance, insertion loss, and electrical delay. Many of Agilent s calibration kits include matched adapters. The first step in the swap-equal-adapters method is to perform a transmission calibration with the first adapter (Figure 15). Following this, the first adapter is removed and the second adapter is placed on port 2. This second adapter then becomes the effective test port. The reflection calibration is then performed on both test ports. Following this, the is measured with the second adapter in place. The errors remaining after calibration are equal to the difference between the two adapters. The technique provides a high level of accuracy, but not quite as high as the more complicated adapter-removal technique. Port 1 Port 2 Accuracy depends on how well the adapters are matched loss, electrical length, match and impedance should all be equal Port 1 Adapter Port 2 A 1. Transmission cal using adapter A. Port 1 Adapter Port 2 B 2. Reflection cal using adapter B. Port 1 Adapter Port 2 B 3. Measure using adapter B. Figure 15. Swap-qual-Adapters Method 13

Adapter-Removal Calibration Adapter-removal calibration provides the most complete and accurate calibration procedure for noninsertable devices (Figure 16). This method uses a calibration adapter that has the same connectors as the noninsertable. The electrical length of the adapter must be specified within one-quarter wavelength at each calibration frequency. Type-N, 3.5-mm, and 2.4-mm calibration kits for the Agilent 8510 network analyzer contain adapters specified for this purpose. Two full two-port calibrations are needed for an adapter-removal calibration. In the first calibration, the precision calibration adapter is placed on the analyzer s port 2 and the test results are saved into a calibration set. In the second calibration, the precision calibration adapter is placed on the analyzer s port 1 and the test data is saved into a second calibration set. Pressing the adapter-removal calibration softkey causes the network analyzer to use the two sets of calibration data to generate a new set of error coefficients that remove the effects of the calibration adapter. At this point, the adapter can be removed and the vector analyzer is ready to measure the. Uses adapter with same connectors as Adapters electrical length must be specified within 1/4 wavelength adapters supplied with Type-N, 3.5-mm, and 2.4-mm cal kits are already defined for other adapters, measure electrical length and modify cal-kit definition Calibration is very accurate and traceable See product note 8510-13 for more details Port 1 Port 2 Cal Port 1 Adapter Adapter Port 2 B Cal Set 1 Cal Port 1 Adapter Adapter Port 2 B Cal Set 2 [CAL] [MOR] [MODIFY CAL ST] [ADAPTR RMOVAL] 1. Perform two-port cal with adapter on port 2. Save in cal set 1. 2. Perform two-port cal with adapter on port 1. Save in cal set 2. 3. Use ADAPTR RMOVAL to generate new cal set. Port 1 Adapter B Port 2 4. Measure without cal adapter. Figure 16. Adapter-Removal Calibration Suggested Reading Understanding the Fundamental Principles of Vector Network Analysis, Agilent application note 1287-1. xploring the Architectures of Network Analyzers, Agilent application note 1287-2. Network Analyzer Measurements: Filter and Amplifier xamples, Agilent application note 1287-4. Specifying Calibration Standards for the Agilent 8510 Network Analyzer, Agilent product note 8510-5A. Applying the Agilent 8510 TRL Calibration for Non-Coaxial Measurements, Agilent product note 8510-8A. 14

15

Agilent mail Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. Agilent Technologies Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. very instrument and system we sell has a global warranty. Support is available for at least five years beyond the production life of the product. Two concepts underlie Agilent s overall support policy: Our Promise and Your Advantage. Our Promise Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you use Agilent equipment, we can verify that it works properly, help with product operation, and provide basic measurement assistance for the use of specified capabilities, at no extra cost upon request. Many self-help tools are available. Your Advantage Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-of-warranty repairs, and on-site education and training, as well as design, system integration, project management, and other professional engineering services. xperienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products. By internet, phone, or fax, get assistance with all your test and measurement needs. Online assistance: www.agilent.com/find/assist Phone or Fax United States: (tel) 1 800 452 4844 Canada: (tel) 1 877 894 4414 (fax) (905) 282 6495 China: (tel) 800 810 0189 (fax) 1 0800 650 0121 urope: (tel) (31 20) 547 2323 (fax) (31 20) 547 2390 Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599 Japan: (tel) (81) 426 56 7832 (fax) (81) 426 56 7840 Korea: (tel) (82 2) 2004 5004 (fax) (82 2) 2004 5115 Taiwan: (tel) 080 004 7866 (fax) (886 2) 2545 6723 Other Asia Pacific Countries: (tel) (65) 375 8100 (fax) (65) 836 0252 mail: tm_asia@agilent.com Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2002 Printed in USA, March 27, 2002 5965-7709 16