Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers
|
|
|
- Adela Martin
- 10 years ago
- Views:
Transcription
1 Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers N4690 Series, 2-port Microwave ECal Series, 2-port RF ECal N4430 Series, 4-port ECal Technical Overview Control ECal directly from the PNA or ENA 300 khz to 26.5 GHz module 10 MHz to 67 GHz module Nine connector types available Ideal calibration technique for manufacturing Mixed-connectors available (Type-N 50 ohm, 3.5 mm and 7-16)
2 Introduction Electronic calibration (ECal) is a precision, single-connection, one-, two-, three-, or four-port calibration technique for your vector network analyzer. ECal modules use fully traceable and verifiable electronic impedance standards. The modules are state-of-the-art, solid-state devices with programmable and highly repeatable impedance states. ECal modules are transfer standards that provide consistent calibrations and eliminate operator errors while bringing convenience and simplicity to your calibration routine. Consistent calibrations provide consistent measurements. ECal replaces the traditional calibration technique, which uses mechanical standards. With mechanical standards you are required to make numerous connections to the test ports for a single calibration. These traditional calibrations require intensive operator interaction, which is prone to error. With ECal, a full one- to four-port calibration can be accomplished with a single connection to the ECal module and minimal operator interaction. This results in faster and more repeatable calibrations. Accurate transfer standards The ECal modules are transfer standards capable of transferring the factory calibration accuracy to your network analyzer. They are characterized by Agilent using a precision calibration technique (similar in accuracy to TRL) that is traceable to the National Institute of Standards and Technology (NIST). Each calibration module s unique S-parameter data is stored in the module s memory. During calibration, ECal uses this data to calculate the error terms for your network analyzer. All measurements on either insertable or non-insertable devices are traceable to NIST. Faster calibration with a single connection Unlike the traditional mechanical technique, ECal only requires one connection to perform a full one- to four-port calibration from the calibration module to the test ports. By reducing the number of connections required for a calibration, you can calibrate faster, so you save time and make measurements sooner reduce the chance of operator error, for greater confidence in your calibrations reduce the wear on connectors, for lower repair costs on both the test port connectors and calibration standards Network analyzer compatibility The family of RF ECal modules provides calibration across the frequency range of the ENA and 8753 series of network analyzers. The N4690 family of microwave ECal modules provides calibrations through 67 GHz for PNA-L and PNA network analyzers. The N4430 family of four-port modules provide calibration for the ENA, PNA, and 4-port PNA-L, and any multiport solutions. Mixed-connector options are available for the 85092C, 85093C, 85098C, N4431B and N4432A. The available connectors are Type-N 50 ohms, 3.5mm, and Suggested ECal and network analyzer/firmware compatibility 1 Agilent VNA ECal module 85097B model number model number Required 8753E/ES/ET Series Y 8719D/ES/ET 2, 3 N4690 Series Y 8720D/ES/ET 2, 3 N4690 Series Y 8722D/ES/ET 2, 3 N4690 Series Y ENA Series 4 N4430 Series, Series N ENA-L Series 5 N4430 Series, Series N PNA Series 6 N4430 Series, N4690 Series N PNA-L Series 7 N4430 Series, N4690 Series N 1. For complete compatibility refer to the ECal Reference Guide (publication N ). 2. Analyzer firmware control available with firmware rev N4690 series works with firmware revision 7.74 or higher. Please note: 8719, 8720, and 8722 have been discontinued. 4. ENA series consists of E5070/1. 5. ENA-L series consists of E5061/2. 6. The PNA Series Network Analyzers consists of E8361/2/3/4. Please note: RF PNA Series E8356/7/8, E8801/2/3 and N3381/2/3 have been discontinued. 7. PNA-L series consists of N
3 ECal and Agilent Network Analyzer Configurations PNA and ENA Series ECal modules are controlled directly from the PNA and ENA Series network analyzers. No external PC is required. Simply connect the ECal module to the USB port on the network analyzer. You can control your calibration from the front panel keys of the PNA Series or automatically by your user program. 8719, 8720, 8722 and 8753 product families ECal modules are controlled manually or automatically via the 85097B interface kit. The 85097B consists of an interface unit and a power supply. The interface module is the interface between the parallel port on your VNA, the parallel port of the ECal module and the external power supply. Firmware revisions of 7.68 or higher in the 8753 and 8720 families allow for VNA control of ECal modules. The 85097B interface module, with the analyzer s internal firmware control, provides digital control and supplies power to one or two ECal modules. Calibration control is available via the front panel keys or from a user program. Firmware revision 7.74 and adapter cable ( ) are required for the N4690 Series. The adapter cable can be ordered as an option with the 85097B. Network analyzer Parallel port connection Calibration configuration using the PNA series ECal module VNA interface unit Optional ECal module VNA interface power supply to AC power Calibration configuration using the 85097B VNA interface kit 3
4 Simple Non-insertable Calibrations Most common RF and microwave components have non-insertable connectors; for example, devices with female connectors on both ports. These devices require an adapter removal calibration, which adds an uncertainty factor to the measurement. Most modern vector network analyzers use an adapter removal technique, which compensates for adaptercaused errors. The simplest and fastest noninsertable calibration method uses an ECal module with connectors that match your device, and the same calibration method as insertable devices. Simply order your ECal module with connectors that match your device under test: Option 00M, male connectors on both ports Option 00F, female connectors on both ports. Option M0F with one male and one female connector. Test set ECal module Perform adapter removal calibrations faster Some analyzers, such as later versions of the 8753 and 8720, offer adapter removal calibration for non-insertable and mixed connector measurement capability. Since this method requires two full two-port calibrations, it is often time consuming and prone to operator errors. Using ECal to perform the two-port calibrations addresses both of these concerns by reducing the calibration time and the number of connections, simplifying the overall adapter removal process. Perform a Usercharacterization Normally, when you perform a calibration with an ECal module, the error terms for a calibration are computed using the factory characterization (data) stored in the module. User-characterization allows you to change the characterization of the module in two ways: Change the connector configuration: allows you to add an adapter or fixture to the test port of the module and embed the effects into the characterization of the module. The result of the new characterization extends the reference plane from one or more of the module s test ports to those on the adapter (or fixture). Modify the state settings: allows you to specify the number of data points (1601 maximum) or other stimulus settings the module uses to perform a calibration. When you perform a user-characterization, the factory characterization data remains stored in the module s memory. At calibration, you can select the factory characterization or any of the user-defined characterizations stored in the module. The module can store up to five user-defined characterizations (in addition to the factory characterization data). User-characterization is available with PNA and ENA Series Network Analyzers. 4
5 Power limits Maximum input power 8509x N469x Minimum input power + 20 dbm +10 dbm 45 dbm Electrical characteristic corrected performance 3 (Residual e-terms) 2-port ECal modules Type-F 85099C (RF) 1 Frequency range 300 khz to 300 MHz 300 MHz to 1 GHz 1 to 3 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) C (RF) 2 Frequency range 300 khz to 300 MHz 300 MHz to 1 GHz 1 to 3 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) When mated with male connectors with a 0.77 mm (0.030 in) to 0.86 mm (0.034 in) pin diameter 2. Typical values when mated with male connectors with a 0.56 mm (0.022 in) to 1.07 mm (0.042 in) pin diameter 3. Except 10 MHz to 45 MHz on N469x modules, which are typical 5
6 Type-N (50 ohms) 85092C (RF) Frequency range 300 khz to 1 GHz 1 to 3 GHz 3 to 6 GHz 6 to 9 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) N4690B (microwave) Frequency range 300 khz to 10 MHz 10 to 500 MHz 500 MHz to 2 GHz 2 to 10 GHz 10 to 18 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) Type-N (75 ohms) 85096C (RF) Frequency range 300 khz to 300 MHz 300 MHz to 1 GHz 1 to 3 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) mm C (RF) Frequency range 300 khz to 1 GHz 1 to 3 GHz 3 to 6 GHz 6 to 9 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) N4691B (microwave) Frequency range 300 khz to 10 MHz 10 to 500 MHz 500 MHz to 2 GHz 2 to 10 GHz 10 to 20 GHz 20 to 26.5 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) mm modules have precision slotless connectors that guarantee the best calibration accuracy is transferred to your system.
7 7 mm 85091C (RF) Frequency range 300 khz to 1 GHz 1 to 3 GHz 3 to 6 GHz 6 to 9 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) N4696B (microwave) Frequency range (GHz) 300 khz to 10 MHz 10 to 500 MHz 500 MHz to 2 GHz 2 to 10 GHz 10 to 18 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) C (RF) Frequency range 300 khz to 1 GHz 1 to 3 GHz 3 to 6 GHz 6 to 7.5 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) mm N4692A (microwave) Frequency range (GHz) 10 to 45 MHz, typical 45 MHz to 2 GHz 2 to 20 MHz 20 to 30 GHz 30 to 40 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) mm N4693A (microwave) Frequency range (GHz) 10 to 45 MHz, typical 45 MHz to 2 GHz 2 to 10 MHz 10 to 20 GHz 20 to 40 GHz 40 to 50 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db)
8 1.85 mm N4694A (microwave) Frequency range (GHz) 10 to 45 MHz, typical 45 MHz to 2 GHz 2 to 20 GHz 20 to 30 GHz 30 to 40 GHz 40 to 50 GHz 50 to 60 GHz 60 to 67 GHz Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) port modules Type-N (50 ohms) N4431B (RF), option 020 Frequency range 300 khz to 1 GHz 1 to 3 GHz 3 to 6 GHz 6 to 8 GHz 8 to 9 GHz 9 to 13.5 GHz Thru paths AB, CD, AD, BC Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) Thru paths AC, BD Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) mm N4431B (RF), option 010 Frequency range 300 khz to 1 GHz 1 to 3 GHz 3 to 6 GHz 6 to 8 GHz 8 to 9 GHz 9 to 13.5 GHz Thru paths AB, CD, AD, BC Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) Thru paths AC, BD Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db)
9 Type-N (50 ohms) N4432A The characteristic performance in the following table applies to N4432A Option 020 (3.5 mm female connectors on all ports). Characteristic performance for N4432A Option 020 Frequency range 300 khz to 1 GHz 10 MHz to 5 GHz 5 to 9 GHz 9 to 13.5 GHz 13.5 to 18 GHz Parameter Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) mm N4433A The characteristic performance in the following table applies to N4433A Option 010 (3.5 mm female connectors on all ports). Characteristic performance for N4433A Option 010 Frequency range 300 khz to 1 GHz 10 MHz to 5 GHz 5 to 9 GHz 9 to 13.5 GHz 13.5 to 20 GHz Parameter Directivity (db) Source match (db) Reflection tracking (±db) Transmission tracking (±db) Load match (db) Values based on using the PNA Network Analyzer N5230A Option 240 or 245 9
10 Ordering information Select an ECal module based on the connector type required and the frequency range of your vector network analyzer (refer to table below). Notes: 1. Order the 85097B interface module if you will be using ECal with your 8719, 8720, 8722 or (Please reference the ECal and network analyzer/firmware compatibility table on page 3.) The 85097B consists of an interface module and a power supply. 2. When using the N469x ECal products with the 8720 or 8753 network analyzer families, an adapter cable ( ) is needed. This adapter cable is orderable as an option with the 85097B. ECal modules and available options 2-port Connector Frequency range ECal module Available options Type model number Type-F 300 khz to 3 GHz 85099C 00A, 00F, 00M, UK6, M0F Type-N 300 khz to 9 GHz 85092C 00A, 00F, 00M, UK6, 1A7, A6J, M0F, 50 ohms mixed-connectors Type-N 300 khz to 18 GHz N4690B 00A, 00F, 00M, UK6, 1A7, A6J, M0F 50 ohms Type-N 300 khz to 3 GHz 85096C 00A, 00F, 00M, UK6, M0F 75 ohms 3.5 mm 300 khz to 9 GHz 85093C 00A, 00F, 00M, UK6, 1A7, A6J, M0F mixed connectors 3.5 mm 300 khz to 26.5 GHz N4691B 00A, 00F, 00M, UK6, 1A7, A6J, M0F 7 mm 300 khz to 9 GHz 85091C UK6, 1A7, A6J 7 mm 300 khz to 18 GHz N4696B UK6, 1A7, A6J khz to 7.5 GHz 85098C 00A, 00F, 00M, UK6, M0F, mixed-connectors 2.92 mm 10 MHz to 40 GHz N4692A 00A, 00F, 00M, UK6, 1A7, A6J, M0F 2.4 mm 10 MHz to 50 GHz N4693A 00A, 00F, 00M, UK6, 1A7, A6J, M0F 1.85 mm 10 MHz to 67 GHz N4694A 00A, 00F, 00M, UK6, 1A7, A6J, M0F 4-port Connector Frequency range ECal module Available options Type model number 3.5 mm or 300 khz to 13.5 GHz N4431B 010, 020, UK6, 1A7, A6J, mixed-connectors Type-N 50 ohms Type-N 300 khz to 18 GHz N4432A 020, mixed-connectors 50 ohms 3.5 mm 300 khz to 20 GHz N4433A 010 Options Option 00F 00M 00A 1A7 A6J UK6 M0F Description Replace f-m connectors on ECal module(s) with f-f connectors Replace f-m connectors on ECal module(s) with m-m connectors Adds male-to-male and female-to-female adapters (also adds a 5/16 90 N-cm (8 in-lb) torque wrench to 3.5 mm modules) ISO compliant calibration ANSI Z540 compliant calibration Commercial calibration certificate with measured data f-m connectors on ECal module(s) 010 Four female, 3.5 mm connectors 020 Four female, Type-N 50 ohm connectors 10
11 Mixed-connector options 2-port (85092C/3C/8C ECal modules only) Port A option Port B option Model number Type (f) (m) Type (f) (m) Type (f) (m) 85092C Type-N 50 ohm mm C 3.5 mm Type-N 50 ohm C mm Type-N 50 ohm port (N4431B ECal module only) Connector type Port A option Port B option Port C option Port D option 3.5 mm (f) mm (m) Type-N 50 ohm (f) Type-N 50 ohm (m) (f) (m) port (N4432A ECal module only) Connector type Port A option Port B option Port C option Port D option 3.5 mm (f) mm (m) Type-N 50 ohm (f) Type-N 50 ohm (m) Limits ECal module high frequency to 7.5 GHz. 11
12 Web Resources Visit our Web sites, for additional product information and literature. Electronic calibration (ECal): PNA Series network analyzers: Test and measurement accessories: Agilent Updates Get the latest information on the products and applications you select. Agilent Direct Quickly choose and use your test equipment solutions with confidence. Agilent Open Agilent Open simplifies the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system-ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development. Agilent Technologies Test and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. We strive to ensure that you get the test and measurement capabilities you paid for and obtain the support you need. Our extensive support resources and services can help you choose the right Agilent products for your applications and apply them successfully. Every instrument and system we sell has a global warranty. Two concepts underlie Agilent s overall support policy: Our Promise and Your Advantage. Our Promise Our Promise means your Agilent test and measurement equipment will meet its advertised performance and functionality. When you are choosing new equipment, we will help you with product information, including realistic performance specifications and practical recommendations from experienced test engineers. When you receive your new Agilent equipment, we can help verify that it works properly and help with initial product operation. Your Advantage Your Advantage means that Agilent offers a wide range of additional expert test and measurement services, which you can purchase according to your unique technical and business needs. Solve problems efficiently and gain a competitive edge by contracting with us for calibration, extra-cost upgrades, out-ofwarranty repairs, and onsite education and training, as well as design, system integration, project management, and other professional engineering services. Experienced Agilent engineers and technicians worldwide can help you maximize your productivity, optimize the return on investment of your Agilent instruments and systems, and obtain dependable measurement accuracy for the life of those products. For more information on Agilent Technologies products, applications or services, please contact your local Agilent office. Phone or Fax United States: Korea: (tel) (tel) (080) (fax) (fax) (080) Canada: Latin America: (tel) (tel) (305) (fax) Taiwan: China: (tel) (tel) (fax) (fax) Other Asia Pacific Europe: Countries: (tel) (tel) (65) Japan: (fax) (65) (tel) (81) [email protected] (fax) (81) Contacts revised: 09/26/05 The complete list is available at: Windows and Windows NT are U.S. registered trademarks of Microsoft Corp. Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc Printed in USA, February 22, E
Agilent Test Solutions for Multiport and Balanced Devices
Agilent Test Solutions for Multiport and Balanced Devices Duplexer test solutions 8753ES option H39/006 During design and final alignment of duplexers, it is often necessary to see both the transmit-antenna
Agilent 8753ET/8753ES Network Analyzers
Agilent 8753ET/8753ES Network Analyzers 8753ET, 300 khz to 3 or 6 GHz 8753ES, 30 khz to 3 or 6 GHz Configuration Guide System configuration summary The following summary lists the main components required
Agilent 8510-13 Measuring Noninsertable Devices
Agilent 8510-13 Measuring Noninsertable Devices Product Note A new technique for measuring components using the 8510C Network Analyzer Introduction The majority of devices used in real-world microwave
Agilent E5063A ENA Series Network Analyzer
Agilent E5063A ENA Series Network Analyzer 100 khz to 4.5/ 8.5/18 GHz Configuration Guide Ordering Guide The following steps will guide you through configuring your E5063A. Standard furnished item Description
Agilent N2717A Service Software Performance Verification and Adjustment Software for the Agilent ESA Spectrum Analyzers Product Overview
Agilent N2717A Service Software Performance Verification and Adjustment Software for the Agilent ESA Spectrum Analyzers Product Overview Reduce your cost of ownership by minimizing time to calibrate and
Agilent U2000 Series USB Power Sensors
Agilent U2000 Series USB Power Sensors Configuration Guide U2000 Series USB Power Sensors U2000A (10 MHz to 18 GHz) U2001A (10 MHz to 6 GHz) U2002A (50 MHz to 24 GHz) U2004A (9 khz to 6 GHz) U2000H (10
Agilent Technologies E8047B Analysis Probe System for the Intel Xeon Processor Family
Agilent Technologies E8047B Analysis Probe System for the Intel Xeon Processor Family Product Overview Requires an Agilent 16700A/B Series or 16900A Series Logic Analysis System The Agilent E8047B analysis
Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet
Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers Data Sheet Specifications Specifications are only valid for the stated operating frequency, and apply over 0 C to +55 C unless otherwise
Agilent 8720 Family Microwave Vector Network Analyzers
Agilent 8720 Family Microwave Vector Network Analyzers Product Overview High-Performance Solutions for Your Measurement Challenges Now more choices for solving your measurement challenges What's new in
81110A Pulse Pattern Generator Simulating Distorted Signals for Tolerance Testing
81110A Pulse Pattern Generator Simulating Distorted Signals for Tolerance Testing Application Note Introduction Industry sectors including computer and components, aerospace defense and education all require
Agilent Split Post Dielectric Resonators for Dielectric Measurements of Substrates. Application Note
Agilent Split Post Dielectric Resonators for Dielectric Measurements of Substrates Application Note l Introduction The split post dielectric resonator (SPDR) provides an accurate technique for measuring
Agilent E5100A Network Analyzer
Agilent E5100A Network Analyzer Data Sheet These specifications are the performance standards or limits against which the instrument is tested. When shipped from the factory, the E5100A meets the specifications
E5092A Configurable Multiport Test Set
Keysight Technologies E5080A Network Analyzer 9 khz to 4.5/6.5/9 GHz E5092A Configurable Multiport Test Set Configuration Guide 02 Keysight 5080A Network Analyzer & E5092A Configurable Multiport Test Set
Agilent PNA Network Analyzers 10 MHz to 110 GHz
Agilent PNA Network Analyzers 10 MHz to 110 GHz Meeting your measurement needs today and into the future... Exceptional performance Advanced automation Flexible connectivity Easy-to-use PNA Network Analyzers
Agilent Evolution of Test Automation Using the Built-In VBA with the ENA Series RF Network Analyzers
Agilent Evolution of Test Automation Using the Built-In VBA with the ENA Series RF Network Analyzers Product Note E5070/71-2 An easy-to-learn and easy-to-use programming language 1. Introduction The Agilent
Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox. Application Note
Agilent Creating Multi-tone Signals With the N7509A Waveform Generation Toolbox Application Note Introduction Of all the signal engines in the N7509A, the most complex is the multi-tone engine. This application
MAURY. VNA Calibration Kits, Microwave Components & Adapters. Calibrate With Confidence IN THIS CATALOG:
MAURY VNA Calibration Kits, Microwave Components & Adapters IN THIS CATALOG: VNA Calibration Kits Calibration Standards Coaxial & Waveguide Adapters Connector Gage Kits Air Lines Torque Wrenches Manual
Agilent E6481A OPAS32 Engineering Information Management and Analysis Software Configuration Guide
Agilent E6481A OPAS32 Engineering Information Management and Analysis Software Configuration Guide The Agilent Technologies E6481A OPAS32 is an engineering information management and analysis software
2. The Vector Network Analyzer
ECE 584 Laboratory Experiments 2. The Vector Network Analyzer Introduction: In this experiment we will learn to use a Vector Network Analyzer to measure the magnitude and phase of reflection and transmission
Agilent 4338B Milliohm Meter
Agilent 4338B Milliohm Meter 10 µω to 100 kω Technical Overview Introduction Ideal for precise measurements of extremely low resistances using an ac test signal, the Agilent Technologies 4338B suits bench-top
Agilent GSM/EDGE Base Station Test with the E4406A VSA and ESG-D Series RF Signal Generators Product Overview
Agilent GSM/EDGE Base Station Test with the E4406A VSA and ESG-D Series RF Signal Generators Product Overview The flexible GSM/EDGE base station test solution with a 3G future... Designed for manufacturing
Agilent E3830 Series Wide-bandwidth Signal Analyzer Custom Systems 100 MHz Bandwidth Microwave Vector Signal Analysis
Agilent E3830 Series Wide-bandwidth Signal Analyzer Custom Systems 100 MHz Bandwidth Microwave Vector Signal Analysis The measurement challenge 2 Many advanced microwave applications involve complex wideband
Microwave Multiplexer Module
Microwave Multiplexer Module Form Drive s K90 OM K90 OM K90 OM Agilent 44476A Switching signals from D to 8 GHz Triple -to- microwave multiplexers N N N Drive Jumpers W W W HP876B HP876B HP876B L90 V TO
Agilent 10 Hints for Making Better Network Analyzer Measurements. Application Note 1291-1B
Agilent 10 Hints for Making Better Network Analyzer Measurements Application Note 1291-1B Contents HINT 1. Measuring high-power amplifiers HINT 2. Compensating for time delay in cable HINT 3. Improving
X-Series Signal Analysis. Future-ready instruments Consistent measurement framework Broadest set of applications and software
X-Series Signal Analysis Future-ready instruments Consistent measurement framework Broadest set of applications and software Arrive Ahead with X-Series We can t predict the future, but Agilent can help
Agilent 8762/3/4A,B,C Coaxial Switches
Agilent 8762/3/4A,B,C Coaxial Switches Technical Overview High performance switches for microwave and RF instrumentation and systems Agilent Technologies offers a ver satile line of multiport coaxial switches.
Agilent PN 89400-13 Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth
Agilent PN 89400-13 Extending Vector Signal Analysis to 26.5 GHz with 20 MHz Information Bandwidth Product Note The Agilent Technologies 89400 series vector signal analyzers provide unmatched signal analysis
Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer. Application Note 1364-1
Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer Application Note 1364-1 Introduction Traditionally RF and microwave components have been designed in packages with
Applying Error Correction to Network Analyzer Measurements. Application Note 1287-3. Table of Contents. Page
Applying Error Correction to Network Analyzer Measurements Application Note 287-3 Table of Contents Page Introduction 2 Sources of Errors and Types of Errors 3 Types of Error Correction 4 One-Port 4 The
Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters
Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters Purpose: This technical note presents single-ended insertion loss ( SE IL) and return loss ( SE RL) data generated
The Next Generation in Automated Oscilloscope Test
The Next Generation in Automated Oscilloscope Test Spanning 100 MHz to 13 GHz, and varying in height from 1U to 8U, Agilent s family of LXI compliant oscilloscopes can accommodate virtually any automated
Agilent Network Analyzer Selection Guide
Agilent Network Analyzer Selection Guide Network Analyzers to Meet Your Needs Agilent offers a variety of network analyzers with the frequency, performance, and versatility to meet your measurement needs.
February 2010 Number 201001. Vector Network Analyzer upgrade to SureCAL Power Sensor and RF Components Packages
February 2010 Number 201001 Vector Network Analyzer upgrade to SureCAL Power Sensor and RF Components Packages Overview: This Technical Note describes the addition of Vector Network Analyzer (VNA) measurement
Measurement & Modeling Device Characterization Solutions
Also Available from Maury Microwave Measurement & Modeling Device Characterization Solutions Featuring The Most Complete Selection of Load Pull Solutions in the Test & Measurement Industry Helping Your
iva Cable & Antenna Analyzer
iva Cable & Antenna Analyzer VSWR, Return Loss Measurement & Distance to Fault The iva Series Cable & Antenna Analyzer is an exciting new product from Kaelus that enables users to accurately measure VSWR/return
Agilent E4980A Precision LCR Meter 20 Hz to 2 MHz
Agilent E4980A Precision LCR Meter 20 Hz to 2 MHz Setting a new standard for low frequency impedance measurements Providing the best combination of accuracy, speed and versatility A New Standard for Impedance
Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows
LS-296 Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows T. L. Smith ASD / RF Group Advanced Photon Source Argonne National Laboratory June 26, 2002 Table of Contents 1) Introduction
Agilent P940xA/C Solid State PIN Diode Switches
Agilent P940xA/C Solid State PIN Diode Switches Operating and Service Manual Agilent Technologies Notices Agilent Technologies, Inc. 2007 No part of this manual may be reproduced in any form or by any
R&S ZCxxx Millimeter-Wave Converters Specifications
ZCxxx_dat-sw_en_3607-1471-22_v0200_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S ZCxxx Millimeter-Wave Converters Specifications 21.07.2015 15:09:16 CONTENTS Definitions... 3 General information...
Agilent E5061B Network Analyzer. 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz
Agilent E5061B Network Analyzer 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz E5061B responds to various measurement needs, from LF to RF The Agilent E5061B is a member of the industry standard ENA Series network
Experiment 7: Familiarization with the Network Analyzer
Experiment 7: Familiarization with the Network Analyzer Measurements to characterize networks at high frequencies (RF and microwave frequencies) are usually done in terms of scattering parameters (S parameters).
Keysight M9485A PXIe Multiport Vector Network Analyzer. Configuration Guide
Keysight M9485A PXIe Multiport Vector Network Analyzer Configuration Guide 02 Keysight M9485A PXIe Multiport Vector Network Analyzer - Configuration Guide Ordering Guide The following steps will guide
Revision 1.10 April 7, 2015 Method of Implementation (MOI) for 100BASE-TX Ethernet Cable Tests Using Keysight E5071C ENA Option TDR
Revision 1.10 April 7, 2015 Method of Implementation (MOI) for 100BASE-TX Ethernet Cable Tests Using Keysight E5071C ENA Option TDR 1 Table of Contents 1. Revision History... 3 2. Purpose... 3 3. References...
Agilent U2000 Series USB Power Sensors
Agilent U2000 Series USB Power Sensors Single/Multi-Channel Power Measurement Simple Setup Affordable Demo Guide Table of Contents Introduction 2 The Agilent N1918A Power 3 Analysis Manager Demonstration
R3765/67 CG Network Analyzer
R3765/67 CG Network Analyzer RSE 05.03.02 1 R376XG Series Overview R3765 300kHz ~ 3.8 GHz R3767 300kHz ~ 8 GHz AG BG Basic model Built-in Bridge A/R & B/R Transmission Reflection CG Built-in S-parameter
Agilent E5063A ENA Series Network Analyzer. 100 khz to 4.5/ 8.5/18 GHz
Agilent E5063A ENA Series Network Analyzer 100 khz to 4.5/ 8.5/18 GHz The Best Balance Between Price and Performance The E5063A is a low cost network analyzer for simple passive component testing up to
Impedance 50 (75 connectors via adapters)
VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:
spinner Measurement & Calibration equipment for network analyzers
spinner Measurement & Calibration equipment for network analyzers Edition B 2011 High Frequency Performance Worldwide www.spinner-group.com SPINNER test & calibration equipment for best measurement results
Advanced Calibration Techniques for Vector Network Analyzers
Advanced Calibration Techniques for Vector Network Analyzers Presented by: Advanced VNA VNA Calibration Agilent Technologies, Inc. Inc. 2006 2006 Welcome to Advanced Calibration Techniques for Vector Network
HP 8753E RF Vector Network Analyzer
HP 8753E RF Vector Network Analyzer 30 khz to 3 or 6 GHz Fast and powerful, the HP 8753E is perfectly adapted for superior, efficient measurements 2 Impressive 300% speed improvement in measurement sweep,
Agilent E6832A W-CDMA Calibration Application
Agilent E6832A W-CDMA Calibration Application For the E6601A Wireless Communications Test Set Data Sheet The next generation of mobile phone manufacturing test. E6601A is the newest test set from Agilent
Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements. Application Note 1304-6
Loop Bandwidth and Clock Data Recovery (CDR) in Oscilloscope Measurements Application Note 1304-6 Abstract Time domain measurements are only as accurate as the trigger signal used to acquire them. Often
a 1 a 2 2 Port b 2 b 1 Multi-Port Handset Switch S-Parameters Application Note AN20 Seven-Port S-Parameter Definition Introduction Summary:
AN2 Multi-Port Handset Switch S-Parameters Introduction High-power UltraCMOS switches are the nextgeneration solution for wireless handset power amplifiers and antenna switch modules. Most multi-throw
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity
iva Cable & Antenna Analyzer
iva Cable & Antenna Analyzer VSWR, Return Loss Measurement & Distance to Fault The iva Series Cable & Antenna Analyzer is an exciting new product from Kaelus that enables users to accurately measure VSWR/return
Agilent 4339B/4349B High Resistance Meters
Agilent 4339B/4349B High Resistance Meters Technical Overview Within Budget Without Compromise Introducing the Agilent Technologies 4339B and 4349B High Resistance Meters Used for Making Ultra- High Resistance
Agilent Technologies 3000 Series Oscilloscopes
Agilent Technologies 3000 Series Oscilloscopes Data Sheet The performance and features you need at the industry s lowest price Features: 60 to 200 MHz bandwidths 1 GSa/s maximum sample rate Large 15-cm
Selecting RJ Bandwidth in EZJIT Plus Software
Selecting RJ Bandwidth in EZJIT Plus Software Application Note 1577 Introduction Separating jitter into its random and deterministic components (called RJ/DJ separation ) is a relatively new technique
Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements
Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements Nick M. Ridler 1 and Nils Nazoa 2 1 National Physical Laboratory, UK (www.npl.co.uk) 2 LA Techniques Ltd,
Agilent Television Power Consumption Testing. Application Note
Agilent Television Power Consumption Testing Application Note Introduction Today, there are many types of televisions (TVs) on the market: the cathode ray tube (CRT) TV, liquid crystal display (LCD) TV,
Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz
Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz Data Sheet Accurate measurement of transmission and reflection characteristics is a key requirement in your selection of a scalar network analyzer.
E5061B-3L5 LF-RF Network Analyzer with Option 005 Impedance Analysis Function. Data Sheet
E506B-3L5 LF-RF Network Analyzer with Option 005 Impedance Analysis Function Data Sheet E506B-3L5 with Option 005; NA plus ZA in one box To ensure the performance and reliability of electronic equipment,
Agilent U2000 Series USB Power Sensors
Agilent U2000 Series USB Power Sensors GSM Timeslot Burst Power Measurement Product Note Table of Content Introduction 2 Measuring GSM Timeslot 4 Signal Overview of Agilent U2000 5 Series USB Power Sensors
HP 8590 C/E/L/Q and EM Series Spectrum Analyzers and Accessories Configuration Guide
HP 8590 C/E/L/Q and EM Series Spectrum Analyzers and Accessories Configuration Guide HP 8591C Cable TV Analyzer (1 MHz to 1.8 GHz) HP 8591E RF Spectrum Analyzer (9 khz to 1.8 GHz) HP 8593E Microwave Spectrum
Trace Port Analysis for ARM7-ETM and ARM9-ETM Microprocessors
Trace Port Analysis for ARM7-ETM and ARM9-ETM Microprocessors Product Overview Introduction Quickly and accurately determine the root cause of your team s most difficult hardware, software, and system
Agilent Balanced Measurement Example: Differential Amplifiers
Agilent Balanced Measurement Example: Differential Amplifiers Application Note 1373-7 Introduction Agilent Technologies has developed a solution that allows the most accurate method available for measuring
Keysight Vector Network Analyzer Calibration and Connector Care. Timothy Lee Keysight Technologies Australia Pty Ltd Application Engineer
Keysight Vector Network Analyzer Calibration and Connector Care Timothy Lee Keysight Technologies Australia Pty Ltd Application Engineer Calibration Connector care Agenda Vector Network Analyzer Calibration
Techniques for Precise Cable and Antenna Measurements in the Field
Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox handheld analyzers Application Note This application note introduces the practical aspects of cable and antenna testing,
W1 Connector W1 Adapter W1 Termination
Data Sheet W1 Connector W1 Adapter W1 Termination DC to 110 GHz A complete coaxial connector system with mode-free performance to 110 GHz Features Excellent RF Performance to 110 GHz 50Ω Impedance Low
Vector Network Analyzer (VNA) Calibration: The Basics
White Paper Vector Network Analyzer (VNA) Calibration: The Basics By Michael Hiebel Note: VNA calibration has been the subject of hundreds of papers, and when discussed in terms of its mathematical derivation
Keysight E5063A ENA Series PCB Analyzer
Keysight E5063A ENA Series PCB Analyzer Technical Overview The Best Solution for PCB Manufacturing Test More accuracy and R&R More languages supported More ESD robustness Why choose the E5063A PCB Analyzer
Keysight Technologies RF and Microwave Test Accessories
Keysight Technologies RF and Microwave Test Accessories Catalog Switches Attenuators Amplifiers Attenuator/Switch Drivers Adapters & Connectors Mixers Directional Couplers Power Dividers & Splitters Power
Eye Doctor II Advanced Signal Integrity Tools
Eye Doctor II Advanced Signal Integrity Tools EYE DOCTOR II ADVANCED SIGNAL INTEGRITY TOOLS Key Features Eye Doctor II provides the channel emulation and de-embedding tools Adds precision to signal integrity
HEWLETT PACKARD. HP 8510C Network Analyzer 45 MHz to 110 GHz. Unmatched excellence in microwave network analysis
HEWLETT PACKARD HP 8510C Network Analyzer 45 MHz to 110 GHz Unmatched excellence in microwave network analysis Excellence in network analysis with unmatched RF performance... The HP 85 IOC Microwave Network
Agilent Network Analyzer Selection Guide
Agilent Network Analyzer Selection Guide Network Analyzers to Meet Your Needs Agilent offers a variety of network analyzers with the frequency, performance, and versatility to meet your measurement needs.
Application Guide to RF Coaxial Connectors and Cables
Application Guide to RF Coaxial Connectors and Cables By: Michael J. Hannon Product Applications Engineer and Pat Malloy, Sr. Applications Engineer There is a wide variety of coaxial connectors and cables
Agilent RF and Microwave Test Accessories Catalog 2006/07
Agilent RF and Microwave Test Accessories Catalog 2006/07 About Agilent s RF and Microwave Test Accessories Catalog 2006/07 The Agilent Technologies 2006/07 RF and Microwave Test Accessories Catalog allows
Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832)
Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) Agilent E5100A/B Network Analyzer 10 khz to 180 MHz/300 MHz Product Overview Agilent s fastest analyzer improves production
Agilent PNA Microwave Network Analyzers
Agilent PNA Microwave Network Analyzers Application Note 1408-13 Introduction to Application Development Table of Contents Introduction...3 How to Use this Document...3 Basic Administration...4 Registering
RF and Microwave Accessories. CD-ROM Catalog. Find the right component for your Rohde & Schwarz test & measurement equipment
RF and Microwave Accessories CD-ROM Catalog Find the right component for your Rohde & Schwarz test & measurement equipment Product group Typical applications Adapters Interchanging of various connector
Agilent 8712ET/ES, 8714ET/ES RF Economy Network Analyzers Product Overview. Speed, accuracy, and productivity features, at a price you can afford
Agilent 8712ET/ES, 8714ET/ES RF Economy Network Analyzers Product Overview 8712ET and 8712ES, 300 khz to 1.3 GHz 8714ET and 8714ES, 300 khz to 3 GHz Speed, accuracy, and productivity features, at a price
Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer
Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer Application Note Introduction The RF power amplifier is a key component used in a wide variety of industries such as wireless
Agilent MATLAB Data Analysis Software Packages for Agilent Oscilloscopes
Agilent MATLAB Data Analysis Software Packages for Agilent Oscilloscopes Data Sheet Enhance your InfiniiVision or Infiniium oscilloscope with the analysis power of MATLAB software Develop custom analysis
Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors. Demo Guide
Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors Demo Guide Introduction This demonstration guide helps you to get familiar with the basic setup and coniguration requirements
USB Smart Power Sensor
50Ω -30 dbm to +20 dbm, 9 khz to 4000 MHz The Big Deal Low cost USB HID device compatible with 32/64 Bit operating systems Includes Measurement Application GUI (Graphical User Interface) software with
ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER
ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER Latest revision: October 1999 Introduction A vector network analyzer (VNA) is a device capable of measuring both the magnitude and phase of a sinusoidal
Agilent Compatibility of the U2000 Series USB Power Sensors with Agilent Instruments. Application Note
Agilent Compatibility of the U2000 Series USB Power Sensors with Agilent Instruments Application Note Table of Contents Introduction 2 U2000 Series USB Power 3 Sensor s Compatibility with Agilent Instruments
CellAdvisor. JD725A Dual-Port Cable and Antenna Analyzer
CellAdvisor JD725A Dual-Port Cable and Antenna Analyzer Many modern wireless base stations are a complex system of multiple RF components, such as low-noise amplifiers (LNA), duplexers, and tower-mounted
Agilent N5970A Interactive Functional Test Software: Installation and Getting Started
Agilent N5970A Interactive Functional Test Software: Installation and Getting Started Application Note The N5970A Interactive Functional Test Software along with the 8960 Series 10 (E5515C) wireless communications
One Port Network Analyzer
99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com One Port Network Analyzer 5.4GHz Impendance : 50Ω(75Ωconnectors via adapters) Test
Carrying precision into the field. 14 New FieldFox Handheld Analyzers Deliver Benchtop Accuracy, MIL-spec Durability
Carrying precision into the field 14 New FieldFox Handheld Analyzers Deliver Benchtop Accuracy, MIL-spec Durability Leverage our legacy of measurement leadership FieldFox analyzers are the handheld extension
R&S SMZ Frequency Multiplier Specifications
Test & Measurement Data Sheet 02.00 R&S SMZ Frequency Multiplier Specifications CONTENTS Definitions... 3 Specifications... 4 RF performance...4 Frequency...4 Level...4 Spectral purity...4 Connectors...5
Agilent Testing Push-to-talk Over Cellular. Application Note
Agilent Testing Push-to-talk Over Cellular Application Note Table of Contents Introduction... 3 New Network Elements... 4 Error Sources... 5 Testing the Links... 5 Simulating Cell Transitions... 5 Signaling
Agilent Mobile WiMAX R&D Test Set Solutions: Software and Technical Support Contract
Agilent Mobile WiMAX R&D Test Set Solutions: Software and Technical Support Contract Product Overview Maximize the potential of the E6651A Mobile WiMAX test set and related software with the latest software
Agilent Millimeter-Wave Network Analyzers 10 MHz to 110 GHz, with Extensions to 0.5 THz
Agilent Millimeter-Wave Network Analyzers 10 MHz to 110 GHz, with Extensions to 0.5 THz Technical Overview High Performance Bench-Top Network Analyzers Maximize your frequency coverage with a single sweep
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER W. Li, J. Vandewege Department of Information Technology (INTEC) University of Gent, St.Pietersnieuwstaat 41, B-9000, Gent, Belgium Abstract: Precision
Agilent Automotive Power Window Regulator Testing. Application Note
Agilent Automotive Power Window Regulator Testing Application Note Abstract Automotive power window regulator tests require the use of accurate data acquisition devices, as they cover a wide range of parameters
USB Smart Power Sensor
Low Power Measurement 50Ω -45 dbm to +10 dbm, 50 to 6000 MHz The Big Deal USB HID device compatible with 32/64 Bit operating systems Includes GUI with measurement applications software, simplifying complex