AUTOMATIC NETWORK ANALYZER PROCEDURES FOR 5045 KLYSTRON CAVITIES * J.G. Judkins
|
|
|
- Rodger Young
- 10 years ago
- Views:
Transcription
1 Introduction SLAC-TN July 1991 AUTOMATIC NETWORK ANALYZER PROCEDURES FOR 5045 KLYSTRON CAVITIES * J.G. Judkins Stanford Linear Accelerator Center Stanford University, Stanford, CA This Note describes the results of using Automatic Network Analyzers in measuring SLAC 5045 klystron cavities. Two different analyzers were compared; the HP8753 and HP8510. Both analyzers have frequency synthesizer accuracy and stability to perform the measurement without the need for a frequency counter. The klystron has six cavities which can be put into three categories; input, gain and output. The input and output cavities require an external Q measurement (Qe) to determine coupling (p) and center frequency (fo). The gain cavities require a resonant frequency measurement only. Overview Both analyzers were suitable for measuring input and gain cavities. The HP8510 has an advanced calibration technique that enables it to measure high reflections accurately. The HP8753 does not have this technique which makes it unsuitable for output cavity measurements. This will be discussed in more detail later. The input and output cavities are greatly overcoupled (p >> 1) which results in an Sll measurement near unity around the periphery of the Smith chart. The analyzers Smith chart display lends itself to determine Qe easily for such high reflections. The gain cavities require input/output coupling probes to determine resonance. The probes are inserted side-by-side into the drift tube and the analyzer is set for transmission measurement (S21). Th e coupling between the probes is approximately -25dB. This is used as the baseline to determne the resonant frequency peaks. The resonant frequency is determined by loosely coupling the probe assembly to the appropriate cavity. Over-coupling at this point will detune the cavity. Therefore, the probes are inserted until the resonance peak is barely discernible on the LOG magnitude display of the analyzer. Most of the gain cavities have a frequency specification of f2mhz. Moving the probes in and out; you can see the effect of overcoupling on frequency determination. The output cavity requires a waveguide calibration. Two calibration techniques were tried and only one proved accurate. The Thru-Reflect-Line (TRL) method proved accurate and consistent with slotted line measurements. The TRL cal is available on the HP8510 and not the HP8753. The Cal method for the HPS753 (TN) * Work supported by the Department of Energy, contract DE-AC03-76SFOO51.5.
2 uses a short, quarter-wave short, and a load. The load is the weak part of this method. Although it works well for moderately high reflections, it is not accurate for very high or small reflections in the order of 0.98 < Sll < The quality of the load determines the accuracy of the directivity component of the calibration. A waveguide load with a return loss of 50 db would be necessary to accomplish the measurements within spec. This high quality load is attainable but fragile and difficult to maintain. The waveguide load used in the comparison tests has a return loss of 36 db (1.02:1 VSWR). Th is was not sufficient to measure the output cavity reflection magnitude of 0.99 within spec. The TRL calibration standards are lower cost, easy to maintain and more accurate. The short, quarter-wave short, load technique was tried on both the HP 8510 and HP 8753 with the same undesirable results. Summary In comparison to the slotted line, the network analyzer takes more time to calibrate but measures much faster. For consistent results the analyzer should be calibrated every four hours. Several cavities could be measured with one calibration faster than the slotted line. With the addition of a PC, documentation can be down-loaded automatically. The data stored on disc is readily available for statistical analysis. The measurement procedures and configurations can be guided by software. This speeds up the setup process, improves repeatability and consistency of test results. In summary, the HP 8510 will provide all the measurements necessary for 5045, PEP and X-band klystrons. The HP 8753 is limited to 5045 input/gain cavities and PEP tube cavities. Although the PEP tube cavities were not part of this analysis, the methods in general can be applied. The PEP tube input and output cavities have Type-N adapters which makes them suitable for either network analyzer.
3 INPUT CAVITY MEASUREMENT The input cavity has a Type-N coaxial input to the coupling loop. This connector will be treated as part of the cavity. A Type-N, Sll-1 port cal will work on either the HP 8753 or HP 8510 network analyzers. The HP 8510 requires a sliding load to characterize the directivity; whereas the HP 8753 uses a fixed load. The HP 8753 has a high precision fixed load specified to 3 GHz which covers its entire frequency range. The fixed load for the HP 8510 is specified to 2 GHz. It may be worthwhile to acquire the higher precision Type-N load to negate the need for a sliding load below 3 GHz. NOTE : The input cavity bandwidth is narrower than the output. It should not be necessary to use Frequency Subset to determine the detune short position. If a fine dot cannot be achieved then the Frequency Subset method described in the output cavity section needs to be performed. SETUP 5045 INPUT CAVITY TEST PROCEDURE FOR HP 8510 Press PRESET STIMULUS START FREQ: enter 2.84 GHz STOP FREQ: enter 2.88 GHz NUMBER OF POINTS: select 401 CAL If Cal Kit Type-N B.l is not already in CAL 1 or CAL 2; then load it from DISC or TAPE.
4 CALIBRATION ELECTRICALDELAY MEASURE : CAL Select TYPE-N Select Sll B.l 1 PORT Perform short, open, load calibratioin as indicated using Type-N Cal Kit. After calibrating, connect network analyzer Port 1 to the cavity under test and insert shorting bar. FORMAT Select INVERTED SMITH CHART Under RESPONSE MENU select ELECTRICAL DELAY. With the shorting bar in place, adjust ELECTRICAL DELAY either by key entry or knob to get a fine dot at the detuned short position (approximately 1.3 nsec). Remove the cavity shorting bar and note fo at the point where the arc crosses the real axis of the Smith Chart. Invoke markers 1,2 and 3 to measure fo and Qe. The Qe bandwidth points are determined where the Sll reflection arc intersects the inverted Smith Chart susceptance arc of jp = 1. See Figure GAIN CAVITY MEASUREMENTS FOR HP 8510 The gain cavities require a center frequency measurement only. Since there are no coupling mechanisms into these cavities, an input and output probe assembly is inserted into the drift tube. It is important to loosely couple to the cavity so as not to pull the frequency resulting in an erroneous measurement. The probe assembly is inserted just far enough to discern the resonant frequency peak from the baseline. Network analyzer calibration is not required for this measurement. The free space coupling between the probes is approximately -25 db. This is used as the baseline. SETUP Press PRESET STIMULUS
5 MEASURE START FREQ.; enter 2.8 GHz STOP FREQ.; enter 3.0 GHz NUMBER OF POINTS; select 801 PARAMETER FORMAT SELECT S21 Select LOG MAG display Connect network analyzer Port 1 and Port 2 to the dual probe assembly. Insert the probe assembly into the drift tube until a small resonant peak is noticed on the LOG MAG display. The frequency of resonance can be read by placing a marker on top of the peak. Notice that inserting the probe further will cause the peak to shift frequency. Therefore, it is important to couple as loosely as possible and still be able to discern the resonant peak. It is possible that false resonances can occur in the drift tube. The probe depth is read off the scale on the probe carriage. This will insure that the cavity is excited and not a coaxial mode in the drift tube. See Figures 2, 3 and 4.
6 5045 OUTPUT CAVITY MEASUREMENTS FOR HP 8510 Output cavity measurements can be made with or without the dual window section attached. The cavity is greatly overocupled (p >> l), which causes a very high reflection (Sll N 0.99). An Sll measurement is used to determine the characteristics of the cavity. The high reflection requires an accurate calibration to minimize errors in connections and characterization of adapters. SETUP Press PRESET STIMULUS START FREQ; enter 2.75 GHz STOP FREQ; enter 2.95 GHz NUMBER OF POINTS; select 801 SWEEP TIME > 200 msec Response Menu: Press MORE, select WAVEGUIDE DELAY; Enter GHz for CUTOFF CAL If Cal Kit WR-284 from disc or tape. Press MORE; FREQ. is not already in Cal 1 or Cal 2; then load it select SET 20; enter 1 ohm. The need for TRIM SWEEP routine is that RAMP SWEEP can be as much as 3 MHz off frequency accuracy. TRIM SWEEP corrects for this difference and allows RAMP SWEEP to approach STEP SWEEP accuracy.
7 ADJUSTTRIMSWEEP PARAMETERMENU : CALIBRATION ELECTRICALDELAY Connect Ports 1 and 2 with or without FORMAT SELECT S21 select PHASE SCALE STIMULUS DISPLAY 1 degree per devision select STEP sweep Wait for display to update with one full sweep (- select DATA-MEMORY select MATH(/). STIMULUS select RAMP sweep. Display should be a flat line. coax to waveguide adapters. 40 seconds). CAL press MORE; select TRIM SWEEP and adjust knob for flattest trace. Select WR-284 calibration Select TRL 2-Port Perform TRL calibration as indicated. There is no need to perform ISOLATION Cal. To omit, select ISOLATION; then select OMIT ISO- LATION. After calibration select Sll, INVERTED SMITH CHART. The FREQUENCY SUBSET is used to reduce sweep width in order to determine electrical delay. It is necessary to use FREQUENCY SUB- SET and TRIM SWEEP method to achieve the accuracy required to determine fo and Qe. The output cavity Specs for fo are 2853 MHz; +3MHz, -8MHz. If we allow the network analyzer to have 10% error in determining fo, then the window of error would be about 1 MHz or f0.5mhz. This tanslates into a tolerance of 1.2 psec for adjusting electrical delay. The reduced sweep width provides a finer dot to achieve the detuned short position accuracy. After determining electrical delay, the full sweep will be turned back on to perform the cavity measurements. 7
8 CAL Press MORE; Select MODIFY CAL SET; then Select FREQUENCY SUBSET. Select CENTER FREQ; enter GHz. Select SPAN; enter 20 MHz. Select CREATE & SAVE, and store in an unused CAL SET. Attach the coax to WR-284 adapter to the cavity waveguide. Place the shorting rod in the cavity. RESPONSEMENU : Select ELECTRICAL DELAY. Adjust DELAY to achieve a fine dot at the detuned short position. This is approximately 6.2 nsec with the window structure attached. After correcting for electrical delay, Select the original-full sweep-cal set to perform the cavity measurements. The electrical delay will be automatically carried over to the wider bandwidth Cal. Measure: Remove the cavity short and note fo as the point where the arc crosses the real axis. The Qe bandwidth points are determined where the Sll arc intersects the susceptance arc of jp = 1. See Figure 5. The TRIM SWEEP adjustment can be checked by switching between wide and narrow Cal sets and noting the fo for each. The fo should be within 1 MHz.
9 CH2 Sll/M : U FS A MS ps.5140 CF MHz P /,, \ / \. \ I 2 / \ \ \ \ \ \ \ I -_ r- - \ 4 Cor Del Avg a t START MHz 3 STOP MHZ 7-s Al Fig. 1
10 A2 Fig. 2, i
11 Fig
12 t markq 3 I I I I I I I I I J I I I I I I I I A4 Fig. 4
13 Sll Y REF 1.0 Units mun i t*/ ms S / I / 0\ L a \\\ N-H\ \ / h. \ START GHz STOP GHZ A5 Fig. 5
ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER
ECE 435 INTRODUCTION TO THE MICROWAVE NETWORK ANALYZER Latest revision: October 1999 Introduction A vector network analyzer (VNA) is a device capable of measuring both the magnitude and phase of a sinusoidal
Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows
LS-296 Vector Network Analyzer Techniques to Measure WR340 Waveguide Windows T. L. Smith ASD / RF Group Advanced Photon Source Argonne National Laboratory June 26, 2002 Table of Contents 1) Introduction
Agilent PN 8753-1 RF Component Measurements: Amplifier Measurements Using the Agilent 8753 Network Analyzer. Product Note
Agilent PN 8753-1 RF Component Measurements: Amplifier Measurements Using the Agilent 8753 Network Analyzer Product Note 2 3 4 4 4 4 6 7 8 8 10 10 11 12 12 12 13 15 15 Introduction Table of contents Introduction
Agilent 8510-13 Measuring Noninsertable Devices
Agilent 8510-13 Measuring Noninsertable Devices Product Note A new technique for measuring components using the 8510C Network Analyzer Introduction The majority of devices used in real-world microwave
2. The Vector Network Analyzer
ECE 584 Laboratory Experiments 2. The Vector Network Analyzer Introduction: In this experiment we will learn to use a Vector Network Analyzer to measure the magnitude and phase of reflection and transmission
Performing Amplifier Measurements with the Vector Network Analyzer ZVB
Product: Vector Network Analyzer R&S ZVB Performing Amplifier Measurements with the Vector Network Analyzer ZVB Application Note This document describes typical measurements that are required to be made
1. The Slotted Line. ECE 584 Microwave Engineering Laboratory Experiments. Introduction:
ECE 584 Microwave Engineering Laboratory Experiments 1. The Slotted Line Introduction: In this experiment we will use a waveguide slotted line to study the basic behavior of standing waves and to measure
Agilent 8720 Family Microwave Vector Network Analyzers
Agilent 8720 Family Microwave Vector Network Analyzers Product Overview High-Performance Solutions for Your Measurement Challenges Now more choices for solving your measurement challenges What's new in
Power Amplifier Gain Compression Measurements
Technical Brief Power Amplifier Gain Compression Measurements GPIB Private Bus Sweep Out Sweep In Pulse In AC Mod Out Blank/Marker Out Blanking In Overview The 1 db gain compression of an amplifier describes
Applying Error Correction to Network Analyzer Measurements. Application Note 1287-3. Table of Contents. Page
Applying Error Correction to Network Analyzer Measurements Application Note 287-3 Table of Contents Page Introduction 2 Sources of Errors and Types of Errors 3 Types of Error Correction 4 One-Port 4 The
Impedance 50 (75 connectors via adapters)
VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:
Making Spectrum Measurements with Rohde & Schwarz Network Analyzers
Making Spectrum Measurements with Rohde & Schwarz Network Analyzers Application Note Products: R&S ZVA R&S ZVB R&S ZVT R&S ZNB This application note describes how to configure a Rohde & Schwarz Network
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity
Shielding Effectiveness Test Method. Harbour s LL, SB, and SS Coaxial Cables. Designs for Improved Shielding Effectiveness
Shielding Effectiveness Test Method Harbour s LL, SB, and SS Coaxial Cables Designs for Improved Shielding Effectiveness Harbour Industries 4744 Shelburne Road Shelburne Vermont 05482 USA 802-985-3311
One Port Network Analyzer
99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com One Port Network Analyzer 5.4GHz Impendance : 50Ω(75Ωconnectors via adapters) Test
Experiments Using the HP8714 RF Network Analyzer
Experiments Using the HP8714 RF Network Analyzer Purpose: The purpose of this set of experiments is two folded: to get familiar with the basic operation of a RF network analyzer, and to gain a physical
Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term 1998-99
Engineering Sciences 151 Electromagnetic Communication Laboratory Assignment 3 Fall Term 1998-99 WAVE PROPAGATION II: HIGH FREQUENCY SLOTTED LINE AND REFLECTOMETER MEASUREMENTS OBJECTIVES: To build greater
Spectrum Analyzers And Network Analyzers. The Whats, Whys and Hows...
Spectrum Analyzers And Network Analyzers The Whats, Whys and Hows... Bertrand Zauhar, VE2ZAZ [email protected] June 2010 Today's Program Definitions of Spectrum and Network Analyzers, Differences between
Network Analysis. Specifying calibration standards for the HP 8510 network analyzer Product Note 8510-5A
Network Analysis Specifying calibration standards for the HP 8510 network analyzer Product Note 8510-5A 2 Known devices called calibration standards provide the measurement reference for network analyzer
HPM-5 Performance Testing
HPM-5 Performance Testing Test 234 JF Sawdy 7/3/02 Samples: Double-ended 2mm cable assemblies: 3, 6 and 8 meters end-to-end, using Meritec s 700219-01 28 awg twinax cable and Meritec s HPM-5 connectors
Improving Network Analyzer Measurements of Frequency-translating Devices Application Note 1287-7
Improving Network Analyzer Measurements of Frequency-translating Devices Application Note 1287-7 - + - + - + - + Table of Contents Page Introduction......................................................................
"FP", "FR", "FQ" Series Bandpass Filters
Description "FP", "FR", "FQ" Series Bandpass Filters The tuning instructions described on the following pages apply to all 7, 8.5, and 10 Bandpass, Notch, and Q circuit filters. Typical models and electrical
Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer
Basics of RF Amplifier Measurements with the E5072A ENA Series Network Analyzer Application Note Introduction The RF power amplifier is a key component used in a wide variety of industries such as wireless
Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements
Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements Nick M. Ridler 1 and Nils Nazoa 2 1 National Physical Laboratory, UK (www.npl.co.uk) 2 LA Techniques Ltd,
Agilent AN 1287-9 In-Fixture Measurements Using Vector Network Analyzers
Agilent AN 1287-9 In-Fixture Measurements Using Vector Network Analyzers Application Note Agilent Network Analysis Solutions Table of Contents 3 3 4 4 5 5 6 7 8 12 13 13 13 15 16 17 17 17 17 18 18 19 19
MATRIX TECHNICAL NOTES
200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR
14.5GHZ 2.2KW CW GENERATOR. GKP 22KP 14.5GHz WR62 3x400V
14.5GHZ 2.2KW CW GENERATOR GKP 22KP 14.5GHz WR62 3x400V UTILIZATION OF GKP 22KP GENERATOR With its characteristics of power stability whatever the load, very fast response time at a pulse, low ripple,
RF Network Analyzer Basics
RF Network Analyzer Basics A tutorial, information and overview about the basics of the RF Network Analyzer. What is a Network Analyzer and how to use them, to include the Scalar Network Analyzer (SNA),
HEWLETT PACKARD. HP 8510C Network Analyzer 45 MHz to 110 GHz. Unmatched excellence in microwave network analysis
HEWLETT PACKARD HP 8510C Network Analyzer 45 MHz to 110 GHz Unmatched excellence in microwave network analysis Excellence in network analysis with unmatched RF performance... The HP 85 IOC Microwave Network
Paul Wade, W1GHZ. 161 Center Rd Shirley, MA 01464 [email protected]. Figure 1 WR-75 waveguide to coax transition for 10 GHz. 1 Notes appear on page 16.
Rectangular Waveguide to Coax Transition Design Learn how to find the optimum dimensions for a waveguide to coax transition using an empirical approach that relies on a set of impedance measurements and
Tuning a Monopole Antenna Using a Network Analyzer
11/21/11 Tuning a Monopole Antenna Using a Network Analyzer Chris Leonard Executive Summary: When designing a monopole antenna it is important to know at which frequency the antenna will be operating at.
Revision 1.10 April 7, 2015 Method of Implementation (MOI) for 100BASE-TX Ethernet Cable Tests Using Keysight E5071C ENA Option TDR
Revision 1.10 April 7, 2015 Method of Implementation (MOI) for 100BASE-TX Ethernet Cable Tests Using Keysight E5071C ENA Option TDR 1 Table of Contents 1. Revision History... 3 2. Purpose... 3 3. References...
SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION
1 SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION By Lannes S. Purnell FLUKE CORPORATION 2 This paper shows how standard signal generators can be used as leveled sine wave sources for calibrating oscilloscopes.
Investigations on Critical Higher Order Modes in CEBAF Upgrade Cavities
Investigations on Critical Higher Order Modes in CEBAF Upgrade Cavities F. Marhauser Thanks also to H. Wang and K. Tian 27. August 2009 How HOMs are excited and why they are important - Wakefields- some
Utilizing Time Domain (TDR) Test Methods For Maximizing Microwave Board Performance
The Performance Leader in Microwave Connectors Utilizing Time Domain (TDR) Test Methods For Maximizing Microwave Board Performance.050 *.040 c S11 Re REF 0.0 Units 10.0 m units/.030.020.010 1.0 -.010 -.020
KLYSTRON TUNING PROCEDURES:
KLYSTRON TUNING PROCEDURES: Applicable to 5-Cavity C-Band and Ku-band Klystrons made by CPI Canada. General. The channel tuning mechanism, an integral part of the tube, provides for the precise tuning
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER
MEASUREMENT UNCERTAINTY IN VECTOR NETWORK ANALYZER W. Li, J. Vandewege Department of Information Technology (INTEC) University of Gent, St.Pietersnieuwstaat 41, B-9000, Gent, Belgium Abstract: Precision
Agilent AN 1287-5 Improving Throughput in Network Analyzer Applications. Application Note
Agilent AN 1287-5 Improving Throughput in Network Analyzer Applications Application Note Table of Contents 2 3 5 12 13 15 18 20 20 20 21 22 23 Introduction Throughput Considerations Sweep Speed Instrument
Experiment 7: Familiarization with the Network Analyzer
Experiment 7: Familiarization with the Network Analyzer Measurements to characterize networks at high frequencies (RF and microwave frequencies) are usually done in terms of scattering parameters (S parameters).
UNDERSTANDING NOISE PARAMETER MEASUREMENTS (AN-60-040)
UNDERSTANDING NOISE PARAMETER MEASUREMENTS (AN-60-040 Overview This application note reviews noise theory & measurements and S-parameter measurements used to characterize transistors and amplifiers at
Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz
Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz Data Sheet Accurate measurement of transmission and reflection characteristics is a key requirement in your selection of a scalar network analyzer.
Cable Analysis and Fault Detection using the Bode 100
Cable Analysis and Fault Detection using the Bode 100 By Stephan Synkule 2014 by OMICRON Lab V1.3 Visit www.omicron-lab.com for more information. Contact [email protected] for technical support.
Agilent FieldFox RF Vector Network Analyzer N9923A Quick Reference Guide
Contents Agilent FieldFox RF Vector Network Analyzer N9923A Quick Reference Guide Do You Have Everything?... 2 The Power Button and LED... 2 Battery Usage... 3 Measure S-Parameters... 4 Multi-Trace Configurations...
Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters
Comparison of Vector Network Analyzer and TDA Systems IConnect Generated S-Parameters Purpose: This technical note presents single-ended insertion loss ( SE IL) and return loss ( SE RL) data generated
AMERICAN NATIONAL STANDARD
ENGINEERING COMMITTEE Interface Practices Subcommittee AMERICAN NATIONAL STANDARD ANSI/SCTE 05 2014 Test Method for F Connector Return Loss In-Line Pair NOTICE The Society of Cable Telecommunications Engineers
Techniques for Precise Cable and Antenna Measurements in the Field
Techniques for Precise Cable and Antenna Measurements in the Field Using FieldFox handheld analyzers Application Note This application note introduces the practical aspects of cable and antenna testing,
Site Master Cable and Antenna Analyzer with Spectrum Analyzer
Maintenance Manual Site Master Cable and Antenna Analyzer with Spectrum Analyzer S331E, 2 MHz to 4 GHz S332E, 2 MHz to 4 GHz, Spectrum Analyzer, 100 khz to 4 GHz S361E, 2 MHz to 6 GHz S362E, 2 MHz to 6
Release Notes. R&S ZVH4/ZVH8 Cable and Antenna Analyzer
Release Notes Revision: 01 R&S ZVH4/ZVH8 Cable and Antenna Analyzer Firmware Release V1.50 These Release Notes describe the following models and options of the R&S Cable and Antenna Analyzer: R&S Cable
iva Cable & Antenna Analyzer
iva Cable & Antenna Analyzer VSWR, Return Loss Measurement & Distance to Fault The iva Series Cable & Antenna Analyzer is an exciting new product from Kaelus that enables users to accurately measure VSWR/return
Network Analyzer Basics- EE142 Fall 07
- EE142 Fall 07 Lightwave Analogy to RF Energy Incident Transmitted Reflected Lightwave DUT RF Why Do We Need to Test Components? Verify specifications of building blocks for more complex RF systems Create
S-PARAMETER MEASUREMENTS OF MEMS SWITCHES
Radant MEMS employs adaptations of the JMicroTechnology test fixture depicted in Figure 1 to measure MEMS switch s-parameters. RF probeable JMicroTechnology microstrip-to-coplanar waveguide adapter substrates
Techniques for precise cable and antenna measurements in the field
Techniques for precise cable and antenna measurements in the field Gustaaf Sutorius Application Engineer March 25, 2015 Agenda Techniques for precise cable and antenna measurements in the field Cable &
spinner Measurement & Calibration equipment for network analyzers
spinner Measurement & Calibration equipment for network analyzers Edition B 2011 High Frequency Performance Worldwide www.spinner-group.com SPINNER test & calibration equipment for best measurement results
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy
Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy Application Note RF & Microwave Spectrum Analyzers Table of Contents 3 3 4 4 5 7 8 8 13 13 14 16 16 Introduction Absolute versus relative
Z Analyze Software. User Manual Version 1.6 - April 2015. Frontier Technical Services 1
Frontier Technical Services 1 Z Analyze Software User Manual Version 1.6 - April 2015 Frontier Technical Services Carl Almgren Sidelinesoft Alexei Smirnov Z Analyze Software Manual Matthew Ferrand / Jared
W-band vector network analyzer based on an audio lock-in amplifier * Abstract
ARDB 179 SLAC PUB 7884 July 1998 W-band vector network analyzer based on an audio lock-in amplifier * R. H. Siemann Stanford Linear Accelerator Center, Stanford University, Stanford CA 94309 Abstract The
Agilent 10 Hints for Making Better Network Analyzer Measurements. Application Note 1291-1B
Agilent 10 Hints for Making Better Network Analyzer Measurements Application Note 1291-1B Contents HINT 1. Measuring high-power amplifiers HINT 2. Compensating for time delay in cable HINT 3. Improving
Probes and Setup for Measuring Power-Plane Impedances with Vector Network Analyzer
Probes and Setup for Measuring Power-Plane Impedances with Vector Network Analyzer Plane impedance measurement with VNA 1 Outline Introduction, Y, and S parameters Self and transfer impedances VNA One-port
MEASUREMENT SET-UP FOR TRAPS
Completed on 26th of June, 2012 MEASUREMENT SET-UP FOR TRAPS AUTHOR: IW2FND Attolini Lucio Via XXV Aprile, 52/B 26037 San Giovanni in Croce (CR) - Italy [email protected] Trappole_01_EN 1 1 DESCRIPTION...3
R&S ZNC Vector Network Analyzer Specifications
ZNC3_dat-sw_en_5214-5610-22_v0300_cover.indd 1 Data Sheet 03.00 Test & Measurement R&S ZNC Vector Network Analyzer Specifications 04.09.2012 13:39:47 CONTENTS Definitions... 3 Measurement range... 4 Measurement
EE 186 LAB 2 FALL 2004. Network Analyzer Fundamentals and Two Tone Linearity
Network Analyzer Fundamentals and Two Tone Linearity Name: Name: Name: Objective: To become familiar with the basic operation of a network analyzer To use the network analyzer to characterize the in-band
Successfully negotiating the PCI EXPRESS 2.0 Super Highway Towards Full Compliance
Successfully negotiating the PCI EXPRESS 2.0 Super Highway Towards Full Compliance Page 1 Agenda Introduction PCIe 2.0 changes from 1.0a/1.1 Spec 5GT/s Challenges Error Correction Techniques Test tool
Agilent E5100A Network Analyzer
Agilent E5100A Network Analyzer Data Sheet These specifications are the performance standards or limits against which the instrument is tested. When shipped from the factory, the E5100A meets the specifications
Transmitter Interface Program
Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following
Fast and Accurate Test of Mobile Phone Boards
Products: R&S FSP Fast and Accurate Test of Mobile Phone Boards Short test times in conjunction with accurate and repeatable measurement results are essential when testing and calibrating mobile phones
Software Utility VNA Frequency Converter Leveling Tool Getting Started
Software Utility VNA Frequency Converter Leveling Tool Getting Started (=ONÓ2) Getting Started Test & Measurement 1331.3069.02 01 2016 Rohde & Schwarz GmbH & Co. KG Mühldorfstr. 15, 81671 München, Germany
Network Analyzer Operation
Network Analyzer Operation 2004 ITTC Summer Lecture Series John Paden Purposes of a Network Analyzer Network analyzers are not about computer networks! Purposes of a Network Analyzer Measures S-parameters
Copyright 1996 IEEE. Reprinted from IEEE MTT-S International Microwave Symposium 1996
Copyright 1996 IEEE Reprinted from IEEE MTT-S International Microwave Symposium 1996 This material is posted here with permission of the IEEE. Such permission of the IEEE does not in any way imply IEEE
Solving Difficult Cable Measurements
Slide #1 Solving Difficult Introduce yourself and the topic for the next 40 minutes. Inquire if there are any customers that are currently performing cable testing and determine the level of the audience.
FREQUENCY RESPONSE ANALYZERS
FREQUENCY RESPONSE ANALYZERS Dynamic Response Analyzers Servo analyzers When you need to stabilize feedback loops to measure hardware characteristics to measure system response BAFCO, INC. 717 Mearns Road
C.-K. Ng. Stanford Linear Accelerator Center. and. T. Weiland. University oftechnology. FB18, Schlossgartenstr. 8. D64289, Darmstadt, Germany.
SLAC-PUB-95-7005 September, 1995 Impedance of the PEP-II DIP Screen C.-K. Ng Stanford Linear Accelerator Center Stanford University, Stanford, CA 94309, USA and T. Weiland University oftechnology FB18,
Product Name Hexa-Band Cellular SMD Antenna GSM / CDMA / DCS / PCS / WCDMA /UMTS /HSDPA / GPRS / EDGE 800 MHz to 2200 MHz
Anam PA.25a Specification Part No. PA.25a Product Name Anam Hexa-Band Cellular SMD Antenna GSM / CDMA / DCS / PCS / WCDMA /UMTS /HSDPA / GPRS / EDGE 800 MHz to 2200 MHz Feature High Efficiency Multi-Band
Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers
Agilent Electronic Calibration (ECal) Modules for Vector Network Analyzers N4690 Series, 2-port Microwave ECal 85090 Series, 2-port RF ECal N4430 Series, 4-port ECal Technical Overview Control ECal directly
TECHNOTES. TD1001: The best RF protector for applications not requiring dc on the coax.
TD1001: The best RF protector for applications not requiring dc on the coax. CONTENTS Abstract 3 Introduction 3 Technology and Testing Overview 4 Test results 4 Let-through Voltage Graphs 10 Conclusions
Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer. Application Note 1364-1
Agilent De-embedding and Embedding S-Parameter Networks Using a Vector Network Analyzer Application Note 1364-1 Introduction Traditionally RF and microwave components have been designed in packages with
Capacitor Self-Resonance
Capacitor Self-Resonance By: Dr. Mike Blewett University of Surrey United Kingdom Objective This Experiment will demonstrate some of the limitations of capacitors when used in Radio Frequency circuits.
Agilent E5063A ENA Series Network Analyzer. 100 khz to 4.5/ 8.5/18 GHz
Agilent E5063A ENA Series Network Analyzer 100 khz to 4.5/ 8.5/18 GHz The Best Balance Between Price and Performance The E5063A is a low cost network analyzer for simple passive component testing up to
ELECTRICAL ENGINEERING DEPARTMENT. California Polytechnic State University SIGNAL TRANSMISSION LABORATORY EE 353
ELECTRICAL ENGINEERING DEPARTMENT California Polytechnic State University SIGNAL TRANSMISSION LABORATORY EE 353 Fall 2003 ELECTRICAL ENGINEERING DEPARTMENT California Polytechnic State University EE 353
Measurement & Modeling Device Characterization Solutions
Also Available from Maury Microwave Measurement & Modeling Device Characterization Solutions Featuring The Most Complete Selection of Load Pull Solutions in the Test & Measurement Industry Helping Your
Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform
Application Note Spectrum and Power Measurements Using the E6474A Wireless Network Optimization Platform By: Richard Komar Introduction With the rapid development of wireless technologies, it has become
Agilent Network Analysis Applying the 8510 TRL Calibration for Non-Coaxial Measurements
Agilent Network Analysis Applying the 8510 TRL Calibration for Non-Coaxial Measurements Product Note 8510-8A Discontinued Product Information For Support Reference Only Information herein, may refer to
EVAL-UFDC-1/UFDC-1M-16
Evaluation Board for Universal Frequency-to- Digital Converters UFDC-1 and UFDC-1M-16 EVAL-UFDC-1/UFDC-1M-16 FEATURES Full-Featured Evaluation Board for the Universal Frequency-to-Digital Converters UFDC-1
2398 9 khz to 2.7 GHz Spectrum Analyzer
Spectrum Analyzers 2398 9 khz to 2.7 GHz Spectrum Analyzer A breakthrough in high performance spectrum analysis, combining cost effectiveness and portability in a new lightweight instrument 9 khz to 2.7
Orthomode Transducer at 43GHz
Electromagnetic Design Orthomode Transducer at 43GHz Author: Renzo Nesti The electromagnetic (EM) design of the Orthomode transducer is based on a symmetric structure to prevent higher order mode excitation.
A Guide to Calibrating Your Spectrum Analyzer
A Guide to Calibrating Your Application Note Introduction As a technician or engineer who works with electronics, you rely on your spectrum analyzer to verify that the devices you design, manufacture,
R&S ZNBT8 Vector Network Analyzer Specifications
ZNBT8_dat-sw_en_3606-9727-22_v0200_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S ZNBT8 Vector Network Analyzer Specifications 20.05.2014 08:39:42 CONTENTS Definitions... 3 Measurement range... 4
Part 2 Designing Combline Filters with Free Software Paul Wade W1GHZ 2014
Part 2 Designing Combline Filters with Free Software Paul Wade W1GHZ 2014 Combline filters have more variables than most other filter types, so design software is more complicated. For instance, interdigital
R&S ZCxxx Millimeter-Wave Converters Specifications
ZCxxx_dat-sw_en_3607-1471-22_v0200_cover.indd 1 Data Sheet 02.00 Test & Measurement R&S ZCxxx Millimeter-Wave Converters Specifications 21.07.2015 15:09:16 CONTENTS Definitions... 3 General information...
Sentinel I28 and Blackbelt Mass Flow Calibration and Verification
Service Bulletin: #001 Date: August 2012 Sentinel I28 and Blackbelt Mass Flow Calibration and Verification Summary This Application Bulletin will give a short synopsis of the purpose of a calibration and
APPLICATION NOTES POWER DIVIDERS. Things to consider
Internet Copy Rev A Overview Various RF applications require power to be distributed among various paths. The simplest way this can be done is by using a power splitter/divider. Power dividers are reciprocal
R&S ZNC Vector Network Analyzer Solid performance on a future-oriented platform
Test & Measurement Product Brochure 01.00 R&S ZNC Vector Network Analyzer Solid performance on a future-oriented platform R&S ZNC Vector Network Analyzer At a glance High reliability, outstanding ease
The Design & Test of Broadband Launches up to 50 GHz on Thin & Thick Substrates
The Performance Leader in Microwave Connectors The Design & Test of Broadband Launches up to 50 GHz on Thin & Thick Substrates Thin Substrate: 8 mil Rogers R04003 Substrate Thick Substrate: 30 mil Rogers
Scalar Network Analysis with the HP 8590 Series Spectrum Analyzers Product Overview
This literature was published years prior to the establishment of Agilent Technologies as a company independent from Hewlett-Packard and describes products or services now available through Agilent. It
Agilent Specifying Calibration Standards for the Agilent 8510 Network Analyzer
Agilent Specifying Calibration Standards for the Agilent 8510 Network Analyzer Application Note 8510-5B Discontinued Product Information For Support Reference Only Information herein, may refer to products/services
OMNIYIG. .5 TO 2 GHz, 1 TO 4 GHz, 6 to 18 GHz Thin Film YIG-TUNED OSCILLATORS OUTLINE DIMESIONS MOUNTING SURFACE GND +15V + TUNE - HTR HTR
. TO Hz, 1 TO Hz, to 1 Hz Thin Film YITUNED OSCILLATORS Highly Reliable StateoftheArt ThinFilm Technology s. to 1 Hz oscillators employ thinfilm technology, coupled with aas FET transistors, and were designed
MAURY. VNA Calibration Kits, Microwave Components & Adapters. Calibrate With Confidence IN THIS CATALOG:
MAURY VNA Calibration Kits, Microwave Components & Adapters IN THIS CATALOG: VNA Calibration Kits Calibration Standards Coaxial & Waveguide Adapters Connector Gage Kits Air Lines Torque Wrenches Manual
Field Calibration Software
SIGNAL HOUND Field Calibration Software User s Manual Version 1.1.0 7/8/2016 This information is being released into the public domain in accordance with the Export Administration Regulations 15 CFR 734
Introduction. Description of Measurement Techniques
Introduction The capability to measure the dielectric properties of various materials has been developed in the Electromagnetic Properties Measurement Laboratory (EPML) of the Electromagnetics Research
Measuring RF Parameters of Networks Bill Leonard NØCU
Measuring RF Parameters of Networks Bill Leonard NØCU NAØTC - 285 TechConnect Radio Club http://www.naøtc.org/ What is a Network? A Network is a group of electrical components connected is a specific way