Thermal Anemometer Measurements

Similar documents
Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson

Thermal Dispersion Mass Flow

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Meriam Instrument LFE Data Sheet

Energy and Flow Measurement for Hydronic Systems

ACFM vs. SCFM vs. ICFM Series of Technical White Papers from Ohio Medical Corporation

Accurate Air Flow Measurement in Electronics Cooling

RITTER Multiple Sonic Nozzle Calibration System

The Fundamentals of Gas Flow Calibration

Technical & Sizing Bulletin for Meriam LFEs. Laminar Flow Elements

THE HUMIDITY/MOISTURE HANDBOOK

T U R B I N E G A S M E T E R

For Multi-Parameter Meters see mvx

CO MPa (abs) 20 C

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Valve Sizing. Te chnic al Bulletin. Flow Calculation Principles. Scope. Sizing Valves. Safe Product Selection.

SIZING AND CAPACITIES OF GAS PIPING

Mass Flow Meter Product Guide

= 800 kg/m 3 (note that old units cancel out) J 1000 g = 4184 J/kg o C

THEORY AND APPLICATION OF KURZ THERMAL CONVECTION MASS FLOW METERS INSTRUMENTS

Temperature. Number of moles. Constant Terms. Pressure. Answers Additional Questions 12.1

User s Manual Laminar Flow Elements

The Unique Accelabar Flow Meter

Pressure drop in pipes...

C. starting positive displacement pumps with the discharge valve closed.

Technical Information

Gases and Kinetic-Molecular Theory: Chapter 12. Chapter Outline. Chapter Outline

CHAPTER 12. Gases and the Kinetic-Molecular Theory

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

CONTROL VALVE PRESSURE DROP AND SIZING

PAGE 2. Figure 1: Difference between PWL ins and SPL 1m

Piping Hydraulic Line Design and Sizing Software KLM Technology Group

Advanced Differential Pressure Flowmeter Technology V-CONE FLOW METER TECHNICAL BRIEF

SIZING AND CAPACITIES OF GAS PIPING

The Gas Laws. Our Atmosphere. Pressure = Units of Pressure. Barometer. Chapter 10

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article

Pump Formulas Imperial and SI Units

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

FUNDAMENTALS OF ENGINEERING THERMODYNAMICS

Thermal Mass Flow Meter (Series MST)

CHEMISTRY GAS LAW S WORKSHEET

PULSATION REDUCTION BY ACOUSTIC FILTERS FOR METERING APPLICATIONS. Robert J. McKee Ray G. Durke

Liquid level measurement using hydrostatic pressure and buoyancy

A drop forms when liquid is forced out of a small tube. The shape of the drop is determined by a balance of pressure, gravity, and surface tension

SIZING AND CAPACITIES OF GAS PIPING (Not Adopted by the State of Oregon)

OIL & GAS PRODUCTION AUDITING

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

Lecture 3 Fluid Dynamics and Balance Equa6ons for Reac6ng Flows

Specific Volume of Liquid (Column 7). The volume per unit of mass in cubic feet per pound.

Stack Sampling Stack sampling or source sampling

Chapter 19 Purging Air from Piping and Vessels in Hydrocarbon Service

Equivalents & Conversion Factors 406 Capacity Formulas for Steam Loads 407 Formulas for Control Valve Sizing

SIZING OF WATER PIPING SYSTEM

HOW TO SIZE A PRESSURE REGULATOR

1. Standard conditions are in. Hg (760 mm Hg, psia) and 68 F (20 C).

BASIC UNDERSTANDING OF FLOW CALCULATIONS AND ESTIMATES MAKES SIZING VALVES SIMPLER

STEAM TURBINE 1 CONTENT. Chapter Description Page. V. Steam Process in Steam Turbine 6. VI. Exhaust Steam Conditions, Extraction and Admission 7

Theory overview of flow measurement using differential pressure devices based on ISO-5167 standard.

Turbine Flowmeters for Gas Applications

Online Infrared Flue Gas Analyzer Gasboard 3000

Barometric Effects on Transducer Data and Groundwater Levels in Monitoring Wells D.A. Wardwell, October 2007

Indiana's Academic Standards 2010 ICP Indiana's Academic Standards 2016 ICP. map) that describe the relationship acceleration, velocity and distance.

Flow Measurement. Reference - Flow Measurement FUNDAMENTALS. ρ = fluid density (lb/ft3 or kg/m3)

Experiment 3 Pipe Friction

THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT

LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS

A LAMINAR FLOW ELEMENT WITH A LINEAR PRESSURE DROP VERSUS VOLUMETRIC FLOW ASME Fluids Engineering Division Summer Meeting

Air Eliminators and Combination Air Eliminators Strainers

Chapter 8: Flow in Pipes

WEEK 1. Engineering Calculations Processes Process Variables

Sanitary Type Electronic Flow Switches

AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS

THE KINETIC THEORY OF GASES

ANNUBAR FLOW CALCULATION PROGRAM (For the FloBoss 103) User Manual (QER 04Q019)

APPENDIX A CONTROL VALVE TESTING PROCEDURES AND EQUATIONS FOR LIQUID FLOWS

1. A belt pulley is 3 ft. in diameter and rotates at 250 rpm. The belt which is 5 ins. wide makes an angle of contact of 190 over the pulley.

FLUID FLOW Introduction General Description

Gas Laws. The kinetic theory of matter states that particles which make up all types of matter are in constant motion.

HVAC FORMULAS. TON OF REFRIGERATION - The amount of heat required to melt a ton (2000 lbs.) of ice at 32 F. 288,000 BTU/24 hr. 12,000 BTU/hr.

Rusty Walker, Corporate Trainer Hill PHOENIX

Current Staff Course Unit/ Length. Basic Outline/ Structure. Unit Objectives/ Big Ideas. Properties of Waves A simple wave has a PH: Sound and Light

Electronic Flow Switches

4.What is the appropriate dimensionless parameter to use in comparing flow types? YOUR ANSWER: The Reynolds Number, Re.

2 1/2 Pipe. 40 = height. the gauge pressure inside the vessel from the gauge pressure at the nozzle inlet as shown:

Thermatel Model TA1 Mass Flow Transmitter

Vogt Valves The Connection Bulletin for Forged Steel Flow Control Valves CB 15

Laminar Flow Elements

Sheet 5:Chapter 5 5 1C Name four physical quantities that are conserved and two quantities that are not conserved during a process.

How To Understand Fluid Mechanics

Series DFMT Digital Paddlewheel Flowmeter

Balancing chemical reaction equations (stoichiometry)

Effects of Temperature, Pressure and Water Vapor on Gas Phase Infrared Absorption by CO 2

Water hammering in fire fighting installation

PIEZOMETER AIRFLOW MEASURING RING

Head Loss in Pipe Flow ME 123: Mechanical Engineering Laboratory II: Fluids

DETERMINING THE MOLAR MASS OF CARBON DIOXIDE

In semiconductor applications, the required mass flows

Transcription:

Thermal Anemometer Measurements The Kurz thermal anemometers use two RTDs, one heated 50 to 100 C above the ambient, the other monitors the ambient. The current or power required to keep the velocity element heated is the parameter used during calibration. Mass Rate A thermal anemometer, controlled at a fixed temperature above the ambient, responds to convective heat transfer. With forced convective heat transfer, the output is proportional to the sensor s Reynolds number (Re). Looking at the Reynolds number terms we can see how it measures mass rate per unit area. It does NOT measure volumetric flow rate but a density weighted version know as standard flow rate. The thermal anemometer automatically compensates for density because it responds to the Reynolds number. Because a thermal anemometer measures the unit-area mass flow, it can be said to measure mass rate. In other words, it measures the true velocity, weighted by the actual density of the flowing gas. If the mass rate is normalized by a known density, it has velocity units, a term known as standard velocity. Mass Flow Equations Reynolds Number Lets look at the Reynolds number since it is proportional to the sensor s power or current when heated X degrees above the ambient: Re = ρvd/μ where ρ = actual density v = actual velocity d = sensor s diameter μ = gas viscosity It is this density and velocity (ρv) product that makes the thermal anemometer a mass flow meter. Density (ρ) has units of mass/volume and velocity (v) has units of length/time. So the ρv product has units of (mass/time)/area or mass rate per unit area. For example: 360209-L Rev. A. Thermal Anemometer Measurements L-1

ρ is kg/m 3, v is m/s so ρv is (kg/s)/m 2 The sensor is sensitive to the energy that each gas molecules hitting it takes away(convective heat flow) in the form of heat. This energy is proportional to the size and number of molecules that hit the sensor. It does not know about density and velocity, just the number of collision with a cooler gas per unit time all contributing to heat loss. Small light gas molecules like hydrogen (H 2 ) are more efficient at transferring the vibrational heat energy of the sensor s surface than large heavy molecules like Argon (Ar). Standard Velocity is the ρv product normalized to a standard density. Standard Velocity = ρv/ρ s where ρ s is the standard gas density. For air this is 0.07387 lb/ft 3 at 77 F and 29.92 in Hg (or 1.184 kg/m3 at 25 C, 760 mm Hg) Note: the density units cancel and you are left with velocity (m/s). Typical units are: Standard Feet Per Minute (SFPM) or Standard Meters Per Second (SMPS). If the gas density doubled (you went from 100 kpa to 200 kpa) at the same actual velocity, the standard velocity would double. This also means that if the process gas is at the same temperature and pressure as the standard condition the standard velocity and actual velocity are identical. Standard Volumetric Flow is the ρv product multiplied by an area (like a pipe cross section), normalized to a standard density Standard Volumetric Flow = Area x (Standard Velocity) = Aρv/ρ s where A is the area: The units here are volume/time (m 3 /s). Typical Displayed units are: SCFM, Standard Cubic Feet-per-Minute SCMM, Standard Cubic Meters-per-Minute SCFH, Standard Cubic Feet-per-Hour SCMH, Standard Cubic Meters-per-Hour Mass Flow is obtained by simply multiplying the Standard Volumetric Flow by the Standard Density. 360209-L Rev. A. Thermal Anemometer Measurements L-2

Mass Flow = (Standard Volumetric Flow) x ρ s =Aρv The units here are mass/time (kg/s). Typical units are: PPH, Pounds-per-Hour KGH, Kilograms-per-Hour Standard Density. Different gases have different standard densities. This is often described using the gas s molecular weight (molar fraction or volumetric % sum of all elements) ρ s =ρ air (MW/MW air ) Then: Mass Flow = (Standard Volumetric Flow) x ρ air (MW/MW air ) = A(vρ/ρ s ) ρ air (MW/MW air ) Air which is about 78 % N 2 (MW = 28) and 21% O 2 (MW =32) plus some trace is 0.78 x 28 = 21.84 0.21 x 32 = 6.72 = 28.56 when we add the trace elments (CO2, Ar, He) we get the established number for Air which is MW air = 28.96. On the other hand a pure gas like methane, CH 4 (C=12 + H=1 x4) MW = 16.04 This is alternately described as a reference density (Air) multiplied by a specific gravity (sg). ρ s =ρ air sg Then: Mass Flow = (Standard Volumetric Flow) x ρ air sg = A(vρ/ρ s ) ρ air sg Conversion between Actual and Standard Flow Conversion of Standard Velocity or Standard Volumetric Flow to actual requires only scaling the result for the gas density according to the ideal gas law. or V a = V s (P s /P a )(T a /T s ) 360209-L Rev. A. Thermal Anemometer Measurements L-3

F a = F s (P s /P a )(T a /T s ) Where V a is actual velocity, V s is standard velocity F a is actual volumetric flow, F s is standard volumetric flow P s is the standard pressure in absolute units P a is the actual pressure in absolute units T a is the actual temperature in absolute units (Kelvin or Rankin) T s is the standard temperature in absolute units (Kelvin or Rankin) Note: K = C + 273.16, R = F + 459.67 A more accurate conversion also accounts for the compressibility (Z) of the gas (deviation from the ideal gas law). This version looks like this. V a = V s (P s /P a )(T a /T s )(Z a /Z s ) Where Z a is the compressibility of the gas under actual conditions Z s is the compressibility of the gas under standard conditions. All other terms are the same The compressibility of a gas is a function of temperature and pressure. High pressures and low temperatures cause the largest deviations from the ideal gas law and the value of Z= 1.0. For small changes in reference conditions, the Z values are very close and can be ignored. 360209-L Rev. A. Thermal Anemometer Measurements L-4

Known Flow Measurement Error Sources A short description of known error sources is provided for reference: Gas property induced errors Flow Profile control Acoustic noise sources Gas species decomposition (not verified yet) Gas Property Induced Errors Here are secondary effects which are know cause mistracking of the idea thermal anemometer. Pressure changes will affect the calibration for some gasses. For example, N 2 has a large 2.5% /100 psi shift in its viscosity which changes its mass flow reading the same amount. By contrast He has nearly no viscosity change with pressure. The gas compressibility Z can also a factor for some gases. Looking at CH 4 for example: 20 C, 101.3 kpa, Z= 0.9975 20 C, 800 kpa or 7.9 bar, Z= 0.9875 or 1% deviation. Temperature changes will affect the gas thermal conductivity and viscosity so the calibration will drift. This is typically 2.5% /100 C. The minimum drift occurs near 3000 SFPM where the dynamic temperature compensation is performed. The use of velocity temperature mapping (VTM) or multiple calibration curves for different gas temperatures, largely eliminates this source of error. Temperature profiles in the pipe will produce flow errors. This is caused by using uninsulated pipe upstream of the sensor where the gas is above or below the ambient temperature. Low flow free convective heat transfer forces compete with forced convective and conductive heat transfer forces for power. This causes measurable errors (depending on gas type, temperature, pressure, and orientation of sensor to both flow and gravity) starting at about 300 SFPM and becomes significant down at about 100 SFPM. Wet vs. Dry Flow Rate: The thermal anemometer responds to all gas molecules which hit it. In the case of water vapor(h 2 0) dissolved in Air, it reads what is know as wet standard volumetric flow or WSCFM. For intake combustion processes you want to know the dry standard volumetric flow or DSCFM which is 21% O 2 so your 360209-L Rev. A. Thermal Anemometer Measurements L-5

fuel air ratio can be properly computed. Knowing the specific humidity ratio ω you can use the following equation up to 5% ω and get results within 1%: DSCFM = WSCFM x 0.622/(0.622 + ω) For example. In Monterey California, a typical dew point of 14 C corresponds to 70 grains/lb of Air based on the Psychrometric chart. Since there are 7000 grains per pound this is 1% specific humidity (ω = 0.01). Therefore DSCFM = WSCFM x 0.622/(0.622 +0.01) = WSCFM x 0.984. In other words, 1.6% if the Air by volume is water vapor. For a derivation of the above equation see the technical note 364018. Flow Profiles And Correction Factors At low velocity, a laminar velocity profile develops across the pipe cross section as shown in the figure. Note that the peak velocity is about 30% higher than the velocity average (V average). At higher flow rates, a flatter velocity profile develops where the peak velocity is closer to the average. So depending on where the sensor is located, it will read a different fraction of the average velocity. It is the average velocity multiplied by the cross sectional area that will obtain the total flow. Correction Factors 360209-L Rev. A. Thermal Anemometer Measurements L-6

The use of a velocity dependent correction factor can convert the local velocity measurement to average velocity. Flow = V local *Area*cf(V local ) The above correction factor curve was measured from a 4" ID pipe with a 2" welded support, triple sting CD sensor. For other sized ducts, the data can be scaled by the 1 Correction Factor (Vpeak/Vaverage) cf( v) 0.9 0.8 0.7 0.6 0.5 Reynolds Number. 0 2000 4000 6000 8000 1 10 4 1.2 10 4 v Sensor local velocity (SFPM) Acoustic, sound or pulsing pressure induced errors. Pneumatic control valves, reciprocating pumps, sonic nozzles, pressure regulators etc. all tend to induced pressure pulsations in to the piping or ducting. At high enough amplitude, the instantaneous gas velocity will reverse direction when compared to the average flow. This causes a thermal sensor to loose heat to this gas on both the forward and reverse flow rate. The heat loss to the fluid is then higher than it was when calibrated on a gas moving in one direction at the same net flow rate. So the sensor will read higher flow than is actually flowing. The source of this pressure wave can be upstream or downstream of the sensor. In some plumbing configurations the pulsations will be sympathetic with the standing wave sound reflections in the pipes which will allow the amplitude to grow very high. Starting at the fundamental and all the harmonics of the standing wave pattern this phenomena can aggravate a flow measurement with a thermal anemometer. Mitigation of acoustic induced errors 360209-L Rev. A. Thermal Anemometer Measurements L-7

To prevent a standing wave of sound in the piping, we simply need to introduce enough sound absorption to prevent the buildup of a large sound wave with each reflection from each end of the piping section where the flow meter is in. Elimination of these acoustic induced errors can be controlled using expansion tanks, mufflers etc. to attenuate the sound waves and return the piping/duct to a smooth one-way flow field so the thermal sensor does not produce false high readings. Inserting a tank between the noise source and sensor is often all that is required to correct this error source. Even a branching tee with a short stub to a tank will attenuate the sound enough to correct most applications which suffer from high acoustic noise. In some applications, a control valve and shut off valve can be swapped so the control valve is not in the same sound-resonator as the sensor. Gas species decomposition from the velocity sensor heat (Not experimentally verified yet) Some gas species, like ozone (O 3 ) can be induced to decompose from the heated velocity sensor. As it requires heat to split up a molecule, the thermal sensor will read high under this condition. The heat of recombination, O combines with O to make O 2 occurs down stream of the sensor and is not detected. So while the net chemical reaction from O 3 to O 2 is exothermic giving off heat, the first step is endothermic. In this case, the only mitigation is to reduce the application temperature or the sensor overheat to below the activation energy needed to initiate this decomposition. In the case of O 3, this is 80 to 100 C. The opposite issue where the sensor surface could at as a catalyst for a recombination and the sensor picks up too much heat, so it reads low, has not been observed but is possible. Higher temperature applications are more likely to observe catalytic heat absorption. Use Of The Flow Equations In The Kurz Mass Flow Computer Single Point Insertion Flow Elements like the 410, 450, 452 and 454FT flow transmitter are calibrated as velocity devices in gas X. You can display standard velocity, or with application specific information you can display standard volumetric flow and mass flow: Area, Sensor and probe support blockage Correction factor (velocity profile) Gas specific gravity when reading mass flow 360209-L Rev. A. Thermal Anemometer Measurements L-8

Multi-point Insertion Flow Elements (K-BAR) are also calibrated as a velocity devices in gas X. You can display standard velocity or with application specific information you can display standard volumetric flow and mass flow: Area Sensor and probe support blockage Correction factor (velocity profile). This tends to be automatic since the velocity is measured across the duct at equal area locations. Gas specific gravity or MW, when reading mass flow. In-line Flow Elements (510, 502,, 522UHP, 532 534) and the 504 and 534FT flow transmitter are calibrated as standard volumetric flow devices in gas X. You can display standard volumetric flow or with application specific information it will display standard velocity or mass flow: Area, Sensor and probe support blockage Correction factor (velocity profile) Gas specific gravity or MW when reading mass flow To maintain the Factory calibration on in-line units requires adherence to the recommended L/D upstream and downstream criteria. This ensures the long pipe run velocity profile when used in the field. Example L/D criteria: L/D is from the heated sensor to the disturbance Model 502-16 or 504FT-16 90 Elbow at x L/D 10 11 % 20 2.5 % Calibration Error Built-in Flow Conditioners. 30 < 0.5 % These lengths are reduced significantly [factors of 16, area-ratio 2. Area-ratio = Free pipe area/(venturi throat area)] for flow elements and transmitters that have a venturi flow conditioner like the 534FT and 542. You can close couple elbows and line size changes to the 534FT with less than a 2% change in the reading. Of course, with metering runs before and after the meter you avoid this flow profile induced error source. 360209-L Rev. A. Thermal Anemometer Measurements L-9

Problems: Air flow of 100,000 lb/hr through a 3' x 3' square duct, 901F, 20 PSIG a. What is mass flow in SCFM b. What is velocity in SFPM c. The actual velocity is d. What range does Kurz calibrate to Nitrogen flow of 10 ACFM through a 3" Schedule 40 pipe, 1101F, 50 PSIG e. What is the area f. What is the flow rate in SCFM g. The velocity in SFPM is h. The calibration max. range is Answers on the Next Page: 360209-L Rev. A. Thermal Anemometer Measurements L-10

Worked out answers to the previous questions. Kurz Standard Reference Conditions. Air density: 0.0739 lb/ft 3 at 77 F and 14.69 psia, a: 100,000 lb/hr x (1 hr/60 min) x 1 ft 3 /0.0739 lb = 22,553 ft 3 /min which is 22,553 SCFM since our density was at standard conditions. b: Velocity = (Volume rate) /Area = 22,553/ (3 x 3) = 2506 SFPM. c: V a = V s (P s /P a )(T a /T s ) = 2506 (14.69/{20 + 14.69})(460 + 90)/(460 +77) = 1087 FPM. d: Calibrated rates are about 1.5 times the expected maximum. So 2506 SFPM x 1.5 ~ 4000 SFPM. e: ID of schedule 40 3" pipe is ~ 3" Area = 3.14 D 2 /4 = 3.14(3/12) 2 /4 = 0.0491 ft 2. f: Same as C above but the other way around with flow. F s = F a (P a /P s )(T s /T a ) = 10 (50 + 14.69)/14.69 (460 +77)/(460 +110) = 41.49 SCFM. g: Volumetric Rate / Area = Velocity = 41.49 SCFM / 0.0491 ft 2 = 845 SFPM. h: Calibrated rates are about 1.5 times the expected maximum. So 41.49 SCFM x 1.5 ~ 60 SCFM. 360209-L Rev. A. Thermal Anemometer Measurements L-11

Chapter L Kurz Instruments Inc October 4, 2007 360209-L Rev. A. Thermal Anemometer Measurements L-12