TECHNICAL NOTE PHOTEK LIMITED. STREAK CAMERA TUBES



Similar documents
Physics 441/2: Transmission Electron Microscope

Experiment 5. Lasers and laser mode structure

Module 13 : Measurements on Fiber Optic Systems

Infrared Viewers. Manual

Understanding astigmatism Spring 2003

A Guide to Acousto-Optic Modulators

Slice Emittance Measurements at the SLAC Gun Test Facility*

hsfc pro 12 bit ultra speed intensified imaging

A NEW LOOK AT RISLEY PRISMS. By Craig Schwarze Senior Systems Engineer OPTRA Inc.

Acousto-optic modulator

View of ΣIGMA TM (Ref. 1)

Synthetic Sensing: Proximity / Distance Sensors

Comparing Digital and Analogue X-ray Inspection for BGA, Flip Chip and CSP Analysis

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.

Proposed experiment to test the non-locality hypothesis in transient light-interference phenomena

Rodenstock Photo Optics

Self-Guided Intense Laser Pulse Propagation in Air

FTIR Instrumentation

WHITE PAPER. P-Iris. New iris control improves image quality in megapixel and HDTV network cameras.

Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Pump-probe experiments with ultra-short temporal resolution

Improving Chromatic Dispersion and PMD Measurement Accuracy

TOF FUNDAMENTALS TUTORIAL

X-RAY TUBE SELECTION CRITERIA FOR BGA / CSP X-RAY INSPECTION

LBS-300 Beam Sampler for C-mount Cameras. YAG Focal Spot Analysis Adapter. User Notes

CHAPTER D1 ION OPTICS OF MAGNETIC/ELECTRIC SECTOR MS

Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors

Untangling the megapixel lens myth! Which is the best lens to buy? And how to make that decision!

CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL


PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

Physics 42 Lab 4 Fall 2012 Cathode Ray Tube (CRT)

Understanding Line Scan Camera Applications

Spectral Measurement Solutions for Industry and Research

Theremino System Theremino Spectrometer Technology

Cathode Ray Tube. Introduction. Functional principle

GRID AND PRISM SPECTROMETERS

Avalanche Photodiodes: A User's Guide

2) A convex lens is known as a diverging lens and a concave lens is known as a converging lens. Answer: FALSE Diff: 1 Var: 1 Page Ref: Sec.

Revision problem. Chapter 18 problem 37 page 612. Suppose you point a pinhole camera at a 15m tall tree that is 75m away.

Design and Test of Fast Laser Driver Circuits

Shutter Speed in Digital Photography

INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS

Full Waveform Digitizing, Dual Channel Airborne LiDAR Scanning System for Ultra Wide Area Mapping

FIBRE-OPTICS POWER METER INSTRUCTION MANUAL

To measure an object length, note the number of divisions spanned by the object then multiply by the conversion factor for the magnification used.

Analysis and Improvement of Mach Zehnder Modulator Linearity Performance for Chirped and Tunable Optical Carriers

A down-under undergraduate optics and photonics laboratory

THE FIBRE-OPTICS TRAINER MANUAL

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution

Optical Modeling of the RIT-Yale Tip-Tilt Speckle Imager Plus

Rodenstock Photo Optics

Recommended alignment procedure for Shaper 6_6 / Focal- Shaper 9

waves rays Consider rays of light from an object being reflected by a plane mirror (the rays are diverging): mirror object

Performance testing for Precision 500D Classical R/F System

Physics 111 Homework Solutions Week #9 - Tuesday

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

X-Rays and Magnetism From Fundamentals to Nanoscale Dynamics

Nikon f/2.8g VR II versus Nikon f/4g VR

ATOMIC SPECTRA. Apparatus: Optical spectrometer, spectral tubes, power supply, incandescent lamp, bottles of dyed water, elevating jack or block.

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Interferometers. OBJECTIVES To examine the operation of several kinds of interferometers. d sin = n (1)

Diffraction of a Circular Aperture

Interference. Physics 102 Workshop #3. General Instructions

Information about the T9 beam line and experimental facilities

Efficiency, Dispersion and Straylight Performance Tests of Immersed Gratings for High Resolution Spectroscopy in the Near Infra-red

PHYSICAL METHODS, INSTRUMENTS AND MEASUREMENTS Vol. IV Femtosecond Measurements Combined With Near-Field Optical Microscopy - Artyom A.

Color holographic 3D display unit with aperture field division

Imaging Systems Laboratory II. Laboratory 4: Basic Lens Design in OSLO April 2 & 4, 2002

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

A More Efficient Way to De-shelve 137 Ba +

GAMMA-RAY SPECTRA REFERENCES

E190Q Lecture 5 Autonomous Robot Navigation

Line Monitoring and Control in Subsea Networks

Endoscope Optics. Chapter Introduction

PHYS 39a Lab 3: Microscope Optics

Development of Optical Wave Microphone Measuring Sound Waves with No Diaphragm

6/2016 E&M forces-1/8 ELECTRIC AND MAGNETIC FORCES. PURPOSE: To study the deflection of a beam of electrons by electric and magnetic fields.

Optical laser beam scanner lens relay system

Digital Mammography Update: Design and Characteristics of Current Systems

The Array Scanner as Microdensitometer Surrogate: A Deal with the Devil or... a Great Deal?

X-RAY IMAGING Emerging Digital Technology - CMOS Detectors

Cargo Container X-ray Inspection Systems

Boardworks AS Physics

ARTIFICIAL NEURAL NETWORKS IN THE SCOPE OF OPTICAL PERFORMANCE MONITORING

Application Note #503 Comparing 3D Optical Microscopy Techniques for Metrology Applications

AMPLIFIED HIGH SPEED FIBER PHOTODETECTOR USER S GUIDE

Challenges in DWDM System Spectral Analysis By Laurent Begin and Jim Nerschook

Scanning Near Field Optical Microscopy: Principle, Instrumentation and Applications

P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)

A Laser Scanner Chip Set for Accurate Perception Systems

Diffraction of Laser Light

Scanners and How to Use Them

The quadrature signals and the index pulse are accessed through five inch square pins located on 0.1 inch centers.

6) How wide must a narrow slit be if the first diffraction minimum occurs at ±12 with laser light of 633 nm?

Data Sheet. AEDR-8000 Series Reflective Surface Mount Optical Encoder. Description. Features. Applications

Telescope Types by Keith Beadman

3D TOPOGRAPHY & IMAGE OVERLAY OF PRINTED CIRCUIT BOARD ASSEMBLY

Transcription:

. STREAK CAMERA TUBES INTRODUCTION Streak camera tubes are used to study ultrafast optical phenomena. Streak cameras are sometimes called optical oscilloscopes, because a streak camera functions in much the same way as an oscilloscope. Light incident on the photocathode is converted to an electron beam, which is then deflected by electrostatic deflection voltage as shown in the schematic diagram below. This deflection voltage provides the time base. Since the deflection sensitivity can be as high as 100 volts/cm, it can be seen that a drive pulse with rise time of 2000 volts/ns gives rise to a time base of 50 ps/cm. (The maximum deflection speed is approximately the speed of light.) The readout system typically an image intensified CCD camera can clearly resolve 100 microns or less, giving an overall time resolution of 1 ps, or less. SPATIAL INFORMATION The streak image can contain spatial information. In a typical application the spatial information could be spectra, so the image shows intensity/time information over a spectral range of interest. 1

BASIC CONCEPTS 1. Basic Design Concept of Streak Tubes A good introduction to streak tube and camera design was published by Bruce Johnson, Applied Optics, Vol. 19, No. 20 page 3491, October 15, 1980. The following paragraphs attempt to summarise this and a host of more detailed studies: 1.1 Slit Width and Readout Pixel Size In an ideal tube, the light from a slit is imaged onto the photocathode, and faithfully reproduced on a readout medium, almost universally a CCD. Clearly, if for example the CCD pixel is 11 microns, the slit needs to be 11 microns or less to take full advantage of this. If magnification is not one, but say two, then the optimum slit width is smaller (5.5 microns). Generally, the use of slit width of much less than 20 microns leads to optical diffraction patterns and are therefore counter productive. The minimum pixel size at the phosphor screen is therefore 20-30 microns multiplied by electron optical magnification say 20 100 micron range. Realistically the slit image should be covered by 2-3 CCD pixels to define precisely when an event happens, so the CCD pixel size should be in the range of 15-25 microns (factored by the fibre optic taper or optical magnification, if it is not directly coupled). This leads to a fundamental time resolution given by the pixel size and deflection speed of so many units of length/second. 1.2 Tube Spatial Resolution Clearly there is nothing to be gained by making the slit width and readout pixel size much smaller than the inherent spatial resolution of the tube. Limiting resolution is generally 30-60 lp/mm at the photocathode, with substantial MTF in the range 15-20 lp/mm. This corresponds to slit width in the range 25-30 microns. 1.3 Magnification and Deflection Speed Increasing the distance between the deflection plates and phosphor screen increases deflection speed and magnification in a linear manner. It also increases magnification in a similar way. Clearly, a long deflector region could have high deflection sensitivity, but also high magnification resulting in very large pixel size. Conversely, a small magnification can result in a poor deflection speed, and a need for very small CCD pixels to compensate. 2

1.4 Chromatic Aberration and Space Charge Limitation Photoelectrons will be emitted into the tube with a characteristic energy (velocity) spread arising from the difference between photon energy and the work function of the photocathode. The electric field within the tube adds to this velocity, but the velocity of an electron at any position in the tube comprises the initial velocity superimposed onto that generated by applied voltages. Analytically, this chromatic time dispersion predominantly occurs when the imposed acceleration is still small that is close the photocathode. To a reasonable degree of accuracy, the time dispersion is inversely proportional to the electric field at the photocathode. Tubes designed for optimum time resolution therefore use a very high field close to the photocathode. This is typically 2 kv/mm in modern metal-ceramic tubes made by Photek, and corresponds to sub-picosecond time resolution This high field also helps overcome space charge limitations. At first glance, this seems unlikely, but for example 1000 photo-electrons, liberated in less than 1 picosecond, in a 10 micron square pixel corresponds to a current density of over 100 Amps/cm 2. This is sufficient to collapse a strong field close to the photocathode; thus increasing chromatic time dispersion, and limiting the signal than can be drawn from a bright event according to Child s Law. Light output is no longer proportional to light input, and the dynamic range, reduces as the optical pulse is made shorter. (References 1-7). 2. DESIGN FAMILIES OF STREAK TUBES 2.1 Large Format Tubes The RCA (Burle) C73435, Photonis P510, ITT4126 all belong to the same family of large format streak tubes. The photocathode can be as big as 35 x 5mm. The majority of cameras were designed to work in the nanosecond time domain, and the large image input area enables a large number of parallel spatial channels to be analysed. Magnification is typically in the range of 0.8 0.9. Photek equivalents are ST-Y, which is based on electro-optical modelling by Dr Ching Lai. (Reference 9) The ST-Y has a relatively low field between photocathode and slot, similar to the RCA tube, which limits the photoelectron throughput (Reference 10 ) The Photonis P510 was developed with a closer slot- photocathode separation giving double the electron throughput. An unwanted side effect of operating at high cathode-slot extraction field is that the magnification of the image in the spatial direction increases to 1.2 or higher. This means that although the photocathode is mechanically 35mm long, the useful length may only be 20mm.A similar high throughput version of ST-Y is available (ST- HDR). 3

Even this modification is inadequate for some applications, and the CEA designed a novel streak tube with quadrupole focus elements, and even higher electric field close to the photocathode. This tube, Photonis P820PSU has only 15mm photocathode, so again struggles to meet the demands of experiments like LMJ in France. (Reference 11) More recently, Photek in collaboration with Sydor and LLE, have developed a more radical re-design of the RCA style tube for even higher throughput using higher extraction field, and a novel second slot to improve spatial and temporal resolution. The magnification is lower, increasing the working area of the photocathode to 25mm. Photek also manufacture a scaled down version of the ST-Y tube, called ST-X, equivalent to the Photonis P900 series. 2.2 Framing Tubes EEV made a large number of streak and framing tubes based on a design by Butlov in 1958. These tubes were extensively sold in cameras made by DRS Hadland Ltd in their Imacon 500 series cameras. The original type was designated P856, but later versions of the tube incorporated an accelerator mesh close to the photocathode to improve time resolution. These were designated P855 for framing cameras such as the Hadland Imacon 675. The photocathode allowed an 8mm 4

square input image. Magnification is approximately 2:1, so in a framing mode, two rows of four images about 16 mm square can be produced on a screen of about 75/80 mm diameter. The EEV P855 used in the Imacon 675 cameras used 3 pairs of deflector plates. The first two pairs, with an aperture between them formed a method of shuttering the image and were known as shutter tubes. For an elegant explanation of this technique see New wide screen gateable image intensifier for ultra-high speed photography (Reference 17). This shutter tube was quite difficult to set-up, and it was particularly difficult to adjust exposure time without altering the frame rate. A new framing camera the Ultranac made by IMCO, a subsidiary of NAC Inc of Japan was launched in 1985/6. The tube used 3 meshes close to the photocathode to switch the tube on for desired exposure time, and off during the time taken to deflect the next image to a new position on the phosphor screen. The centre mesh was used to switch the electron beam on or off with about 100 volts bias, while the outer mesh maintains the electron lens in optimum focus condition. This new camera could reach 20 million frames a second, and exposure time could be adjusted independently of frame rate. Typical mesh transmission of 50-60% means that overall transmission is of the order of 15%. This is a significant loss of sensitivity compared to the deflector plate lock of the older cameras. It is probably no worse than the optical losses in high-speed cameras using beam splitters of various kinds. The Photochron2, made by Photek is similar to the Ultranac tube, but with only two meshes which maintains resolution, and improves transmission to about 30%.It easily replaces old tubes in Ultranac cameras, and could in principal be used to upgrade/repair Hadland streak and framing cameras. 5

2.3 Photochron 5 After Professor Bradley became ill in 1980, this group remained at Imperial College till 1985, when several of them including Professor W Sibbett moved to St Andrews University. Various new designs were made in small quantity, through EMI made a type 9898 Photochron 3 in significant quantities. (Reference 13). These ideas lead to Photek developing an all-metal-ceramic version of Photochron 5 for the DRS Hadland Ltd FS-300 Camera. Performance of less than 400fs was demonstrated in single shot mode (Reference 14) This tube had an internal MCP, and was designed for synchroscan operation upto 500MHz Another version of this tube with lower magnification was developed for Optronis GmbH (Reference 15) Both tubes are highly customised to suit the camera manufacturers, and are not available to the world at large, except through Optronis GmbH. (A generic Photochron 5 is commercially available to scientific users). The tube is capable of sub-picosecond time resolution, and can be operated in synchroscan to 200 MHz and beyond. www.optronis.com Summary of Photek Streak Tubes Type Photochron ST-HDR Photochron ST-X ST-Y 2 5 Availability Current Current Current Current Current Other Equivalents Cathode Size EEV P8307 EEV P855 Ultranac 8 x 9 mm maximum P 510 None Photonis P930 ITT F4157 35x4 8 mm Slit 20 x 3 35 x 5 Synchroscan 88 MHz? 200 MHz plus Limited Limited Time Resolution ~ 3 ps 7 ps 400fs 10 ps 20 ps Magnification ~ 2-2 - 4 0.7 1.5 0.7 1.5 Framing Option Overall Length Double Grid X-Y Deflectors No X-Y Deflectors 321.5 mm 250 mm 322 mm 147 mm 250 mm No No Photek have also made very large framing tubes with 80 mm photocathodes for high-speed imaging. Ref. 16. 6

REFERENCES 1. Dr C B Johnson, Applied Optics, Vol. 19 (20) pp 3491, 1980 2. Dr Hull & N J Freeman, J Phys E: Sci Instrum Vol. 13 pp 685-690 1980 3. P G May & W Sibbett, Appl Phys, Letts 43 (7) pp 624-626 1983 4. M C Adams, W Sibbett & D J Bradley Electron Optical Picosecond Streak Camera Operating at 140 MHz. Advances in E & EP 42 1979 pp 265-273 5. W Sibbett, W E Sleat & W Krause pp 171 174, Proc ESA Workshop on Space Laser 6. D J Bradley, et al. Optical Communications May 1975 7. P R Bird, D J Bradley, A G Roddie, W Sibbett, Picosecond Chronography at X-ray Wavelengths, Proc. 11 th congress on High Speed Photography 8. Mori & Watanabe, Jap J. of Applied Physics, See pages 666 to 672, 1976 9. Dr C Lai, Design and Development of a New Streak Tube, Report UCAR-10159, LLNL 10. P.A. Jaanimagi,A.Mens & J.C.Rebuffi SPIE 2549 (1995) 11. C.Cholet, D Gontier, C.Zuber&P.Brunel,SPIE Vol 7126 pp1-12 12. I G Haig, Rev. Sci. Inst., pp 141-143 1981 13. W Sibbett, et al, Rev Sci. Inst., pp 53 758-761 1982 14. D Bowley, J.Beeley, M Pittak, 22 nd International High Speed Photography, SPIE Vol 2869, pp 76-81 1996 15. J.R.Howorth, Ian Phillips & Mikhail Monastryski, IHSPP Conference Beaune, 2002 16. CC Lai, Electron optics design & performance of a new large format two frame framing tube, IHSPP Conference Beaune 2002 17. Boris Dashevsky & Alexander Surovegin (SPIE Vol 1243 pp 65-72 1990). 18. J Howorth, G Smith et al, Recent advances in framing tubes for high speed photography, ICHSIP30 2012 7