The Array Scanner as Microdensitometer Surrogate: A Deal with the Devil or... a Great Deal?
|
|
|
- Buddy Simmons
- 9 years ago
- Views:
Transcription
1 The Array Scanner as Microdensitometer Surrogate: A Deal with the Devil or... a Great Deal? Don Williams Eastman Kodak Company, Rochester, NY, USA, ABSTRACT Inexpensive and easy-to-use linear and area-array scanners have frequently substituted as densitometers for lowfrequency (i.e., large-area) hard copy image metrology. A number of workers and standard s committees 1 have in fact advocated such use. Increasingly, scanners are also being tasked for high spatial frequency, image microstructure metrology, which is usually reserved for high-performance microdensitometers that use microscope optics, photomultiplier tubes (PMT), and log amps. It is hard to resist their adoption for such use given the convenience level. The high speed, large scan areas, auto-focus, discomfiting low cost, and low operator skill requirements of commercial off-the-shelf (COTS) scanners makes one question if their use for such purpose is somehow too good to be true though. To confidently judge their limitations requires a comprehensive signal and noise spatial frequency performance evaluation with respect to available driver options. This paper will outline and demonstrate evaluation techniques that use existing ISO metrology standards for modulation transfer function (MTF), noise, and dynamic range with a comparison to the performance of a Photometric Data Systems (PDS) microdensitometer. Keywords: MTF, dynamic range, microdensitometer, metrology, standards 1. INTRODUCTION Despite their half century legacy, microdensitometers continue to serve as de-facto standardized instruments for high spatial frequency photometric image metrology of both transmissive and reflective media. Microdensitometers however continue to excel in their flexibility and imaging performance. Independent aperture and sampling settings, calibration freedom, ease of operator intervention, and PMT/log amp detection make them invaluable as high-end research and development tools. For these reasons, they are unlikely to ever be completely replaced. But, costly or unattainable maintenance, high skill requirements, and slow speeds have forced many metrology centers to abandon them altogether in favor of their speedier and less-expensive brethren, the array scanner. Is this a deal with the devil though? Will the immediate convenience of rapid and low cost data collection allowed by array scanners haunt adopters with mid-grade imaging performance, cross-device inconsistency, and non-standardized driver settings? Or, is it really a great deal? Are the high dynamic range and resolution claims (summarily and incompletely specified in bits and sampling frequency ) of COTS scanners truly achievable in a scientific signal-to-noise ratio sense at such low cost? Using established scientific practices and standardized ISO protocols, a scanner imaging performance accreditation for metrology purposes will be demonstrated. While the methods described here can be applied to any device intended for quantitative metrology, the candidate COTS scanner chosen for comparison here is an Imacon 848 scanner in transmission mode. The single greatest reason for considering the Imacon 848 as a microdensitometer surrogate had been the claim of true 8000 dpi performance over a 6 frame 35mm film chop. A single frame RGB scan at this resolution will take about five minutes. On the microdensitometer the same resolution scans would take about 8 hours. Clearly, this almost 100X decrease was a factor. Three primary imaging performance metrics were measured on the Imacon 848. They are 1) Opto-Electronic Conversion Function (OECF), 2) noise and signal-to-noise ratio (SNR), and 3) modulation transfer function (MTF). All were measured according to protocols defined in ISO 14524, ISO 21550, and ISO respectively. The OECF and SNR metrics can be generalized to any scan resolution mode. The MTF results are format dependent however because the effective sampling resolution is dictated by a reduction optics design. Fortunately, the 35mm size format provides the highest potential resolution along with being a popular format for existing microdensitometer requests. So, the 35mm mode was selected for MTF characterization.
2 2. OPTO-ELECTRONIC CONVERSION FUNCTION OECF Unlike microdensitometers which usually encode signals in one of two ways - linear with transmission or density - the finished file encoding for scanners like the Imacon 848 is a function of the various settings available in the software driver. Only the effect of the gamma settings was of interest for this characterization and is described here. The generic driver settings were as follows. Highlight = 255, Shadow= 0, Gamma= variable, Sharpening = disabled, 16 bit capture The OECF results for the labeled gamma settings are shown in Figure 1. Though captured in 16 bits, the ordinate is labeled in 8 bit equivalent counts for ease of interpretation. Those wishing for the 16 bit equivalent should multiply the ordinate values by 255. The curves of Fig. 1 apply to all color channels equivalently. Figure 1 Imacon OECF curves ; gamma 1.5 thru 3.5 ; color channel equivalence The behaviors of the OECFs as a function of gamma are typical. Low gammas are characterized with high low-density slopes that quickly diminish at the high densities. This would suggest better signal detection for low density values using low gamma settings. Similarly, greater high-density slopes at high gamma settings might suggest their choice for tasks requiring high density signal detection. Qualitatively, note the increasing plateau behavior of the OECF at high gamma settings. While this would suggest a decrease in effective dynamic range, noise levels also need to be considered for such a measure. So, the OECF is a necessary but insufficient tool for evaluating a device s dynamic. The noise and SNR metrology for the Imacon 848 and PDS microdensitometer are covered in the next section. 3. NOISE AND SIGNAL-TO-NOISE RATIO (SNR) The measured noise levels for the Imacon 848 and PDS densitometer are shown in Fig. 2. The PDS noise levels remain fairly constant at rms density over its range of operation and is clearly superior to the Imacon at densities above 1.5. The Imacon 848 outperforms the PDS up to densities of 1.5 but quickly degrades beyond a 0.02 rms density at a mean density of 2.60.
3 Figure 2 RMS density for Imacon and PDS microdensitometer Note that no distinction was made for the Imacon 848 density noise levels as a function of gamma setting. Though the rms noise in counts does change as a function of gamma, in terms of equivalent density it does not. This is supported further through the incremental SNR plots of Figure 3. The SNR and thus signal detection abilities of the Imacon 848 remain essentially unchanged for gamma settings between 1.5 and 3.5. This, of course, is the beauty of using incremental SNR as outlined in ISO and It is independent of gamma setting. Figure 3 Incremental* SNR for Imacon and PDS microdensitometer * Incremental SNR uses the slope of the OECF as its signal metric and is used to measure a systems ability to distinguish incremental rather than absolute signals. For example, how well a density of X can be detected from X + δx.
4 3. MODULATION TRANSFER FUNCTION - MTF The single greatest reason for considering the Imacon as a microdensitometer surrogate had been the claim of true 8000 dpi performance over a 6 frame 35mm chop. This claim is reasonably well supported through the MTFs of Figures 4 and 5. The Imacon Colorflex driver allows resolution selection for native and non-native resolutions. Only the native resolutions were tested. Standard resolution MTFs are shown in Fig. 5 while high resolution mode MTFs are illustrated in Fig. 4. For a given sampling frequency, no directional or spectral MTF differences were found. Therefore, the illustrated responses are the same for all colors and scan directions. Figure 4 - High Resolution Imacon MTFs Figure 5 - Standard Resolution Imacon MTFs A number of the MTFs for each mode are not classically shaped optical MTFs. That is, not characterized by a monotonically decreasing shape. The unorthodox behaviors at 1575, 2100, and 3150 dpi ( Fig. 5) and 2000, 2667, and 4000 dpi (Fig. 4) suggest that some form of forced processing is occurring. For these reasons, it was not recommended that these sampling resolutions be used when performing spatial metrology grade scans with the Imacon 848. Either the 6300 dpi (low res. mode) or 8000 dpi (high res. mode) samplings are more suitable. For comparison purposes, the measured MTFs of the microdensitometer using a 10µm, 6.7 µm and 5µm diameter circular aperture are plotted in Figure 6 along with the 6300 dpi and 8000 dpi Imacon 848 MTFs. This illustrates that a 10 µm circular microdensitometer aperture response best matches the Imacon 848 at 8000 dpi. The 6300 dpi match is slightly less congruous but probably still suitable for microdensitometer tasks using a µm circular apertures. From theoretical considerations, circular aperture MTF responses are roughly equivalent to square aperture responses 20% smaller in size. ( e.g. 10 µm circular aperture response is approximately equal to an 8 µm square aperture response). The above two sampling frequencies are equivalently efficient in terms of bandwidth per pixel. This can be deduced by taking the ratio of each limiting resolution frequency 2 (10% MTF response ) to the half-sampling frequency. The half sampling frequency for 6300 dpi is 124 cy/mm. For 8000 dpi it is 157 cy/mm. These are the maximum useful frequencies that these sampling rates can support as dictated by the Nyquist criterion. The ratio of their measured limiting resolution ( maximum measured frequency), 100 cy/mm and 120 cy/mm respectively ( see Fig. 6), to their halfsampling frequencies ( maximum theoretical frequency) are roughly 80%. The equivalence and relatively high value of this ratio is a good indication that optics or image processing functions are not unduly limiting the spatial frequency potential of this scanner.
5 Figure 6 MTF comparisons between Imacon and PDS microdensitometer Similar thinking can be applied to any scanner where a continuum of sampling rates is allowed. Only the naïve among us would believe that increasing sampling rates would be followed by true unbounded spatial frequency response improvements. Take the example of the suite of MTFs for a flatbed scanner in Fig.7. Each MTF curve corresponds to a different sampling rate, as labeled, allowed by the scanner driver. MTF sampling rate progression for flatbed scanner Response dpi 900 dpi 1200 dpi 1500 and 1800 dpi dpi Frequency (cy/mm) Figure 7 MTF sampling rate progression for candidate flatbed scanner Though the sampling rates of Fig. 7 are equally incremented by 300 dpi, the frequency response improvements across sampling rates are not proportional. For instance, there is a very large MTF improvement in the dpi transition while there is barely any in the dpi transition. Indeed, there is a decrease in MTF performance in the dpi transition. A simple contour plot technique can be used to more easily evaluate these sampling rate MTF behaviors as illustrated in Fig. 8.
6 0 Equi-Response MTF Contours ( 0.1 increments) Sampling Frequency (dpi) Frequency (cycles/mm) * Dashed line indicates half-sampling rate for corresponding dpi Figure 8 - MTF contour plots for flatbed scanner Using the 10% MTF contour for demonstration, one can use Fig. 8 to easily track which dpi settings offer the most efficient bandwidth sampling. For instance, the relatively large rightward outcroppings at 600 and 1200 dpi suggest a possible native resolution detent of the scanner. Beyond 1200 dpi it is clear that the MTF performance dramatically decreases, even at the 1300 dpi level. This type graphic allows visualization of overall MTF performance for better sampling rate selection depending on the source object being measured. 4. SUMMARY For most of today s hardcopy media types, array scanners can easily substitute for micro-densitometers. But OECF, MTF, and noise characterizations are required beforehand to accredit their absolute photometric performance. It is often said, the devil is in the details. Characterizing these details with the protocols outlined in ISO 14524, ISO 21550, and ISO can make the difference between the deal with the devil and the great deal. This paper has provided examples of how these protocols can be used to map and identify the micro signal detection capabilities of array scanners. REFERENCES 1. D. René Rasmussen, et al, INCITS W1.1 Standardization for Evaluation of Perceptual Macro-Uniformity for Printing Systems, Proc. IS&T 2003 PICS Conference, IS&T, , Don Williams and P.D. Burns, Diagnostics for Digital Capture using MTF, Proc. IS&T 2001 PICS Conference, IS&T, , 2001.
Diagnostics for Digital Capture using MTF
Diagnostics for Digital Capture using MTF Don Williams and Peter D. Burns Eastman Kodak Company Rochester, NY USA Abstract The function (MTF) has long been used as a diagnostic tool for analog image capture,
Assessment of Camera Phone Distortion and Implications for Watermarking
Assessment of Camera Phone Distortion and Implications for Watermarking Aparna Gurijala, Alastair Reed and Eric Evans Digimarc Corporation, 9405 SW Gemini Drive, Beaverton, OR 97008, USA 1. INTRODUCTION
Video Camera Image Quality in Physical Electronic Security Systems
Video Camera Image Quality in Physical Electronic Security Systems Video Camera Image Quality in Physical Electronic Security Systems In the second decade of the 21st century, annual revenue for the global
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
Scanners and How to Use Them
Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction A scanner is a device that converts images to a digital file you can use with your computer. There are many different types
Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.
Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ., Raleigh, NC One vital step is to choose a transfer lens matched to your
Digital Camera Imaging Evaluation
Digital Camera Imaging Evaluation Presenter/Author J Mazzetta, Electro Optical Industries Coauthors Dennis Caudle, Electro Optical Industries Bob Wageneck, Electro Optical Industries Contact Information
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture
Characterizing Digital Cameras with the Photon Transfer Curve
Characterizing Digital Cameras with the Photon Transfer Curve By: David Gardner Summit Imaging (All rights reserved) Introduction Purchasing a camera for high performance imaging applications is frequently
Bernice E. Rogowitz and Holly E. Rushmeier IBM TJ Watson Research Center, P.O. Box 704, Yorktown Heights, NY USA
Are Image Quality Metrics Adequate to Evaluate the Quality of Geometric Objects? Bernice E. Rogowitz and Holly E. Rushmeier IBM TJ Watson Research Center, P.O. Box 704, Yorktown Heights, NY USA ABSTRACT
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL
CONFOCAL LASER SCANNING MICROSCOPY TUTORIAL Robert Bagnell 2006 This tutorial covers the following CLSM topics: 1) What is the optical principal behind CLSM? 2) What is the spatial resolution in X, Y,
Module 13 : Measurements on Fiber Optic Systems
Module 13 : Measurements on Fiber Optic Systems Lecture : Measurements on Fiber Optic Systems Objectives In this lecture you will learn the following Measurements on Fiber Optic Systems Attenuation (Loss)
Using visible SNR (vsnr) to compare image quality of pixel binning and digital resizing
Using visible SNR (vsnr) to compare image quality of pixel binning and digital resizing Joyce Farrell a, Mike Okincha b, Manu Parmar ac, and Brian Wandell ac a Dept. of Electrical Engineering, Stanford
Why use ColorGauge Micro Analyzer with the Micro and Nano Targets?
Image Science Associates introduces a new system to analyze images captured with our 30 patch Micro and Nano targets. Designed for customers who require consistent image quality, the ColorGauge Micro Analyzer
Lecture 14. Point Spread Function (PSF)
Lecture 14 Point Spread Function (PSF), Modulation Transfer Function (MTF), Signal-to-noise Ratio (SNR), Contrast-to-noise Ratio (CNR), and Receiver Operating Curves (ROC) Point Spread Function (PSF) Recollect
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors
Investigation of Color Aliasing of High Spatial Frequencies and Edges for Bayer-Pattern Sensors and Foveon X3 Direct Image Sensors Rudolph J. Guttosch Foveon, Inc. Santa Clara, CA Abstract The reproduction
CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY. 3.1 Basic Concepts of Digital Imaging
Physics of Medical X-Ray Imaging (1) Chapter 3 CHAPTER 3: DIGITAL IMAGING IN DIAGNOSTIC RADIOLOGY 3.1 Basic Concepts of Digital Imaging Unlike conventional radiography that generates images on film through
Measuring Line Edge Roughness: Fluctuations in Uncertainty
Tutor6.doc: Version 5/6/08 T h e L i t h o g r a p h y E x p e r t (August 008) Measuring Line Edge Roughness: Fluctuations in Uncertainty Line edge roughness () is the deviation of a feature edge (as
The Fundamentals of MTF, Wiener Spectra, and DQE. Motivation
The Fundamentals of MTF, Wiener Spectra, and DQE Robert M Nishikawa Kurt Rossmann Laboratories for Radiologic Image Research Department of Radiology, The University of Chicago Motivation Goal of radiology:
Overview. Proven Image Quality and Easy to Use Without a Frame Grabber. Your benefits include:
Basler runner Line Scan Cameras High-quality line scan technology meets a cost-effective GigE interface Real color support in a compact housing size Shading correction compensates for difficult lighting
RGB Color Managed Workflow Example
White Paper 23 Level: Basic Date: May 2008 RGB Color Managed Workflow Example Overview Today s abundance of RGB source art such as digital illustrations, photos, and digital art are routinely repurposed
The Effect of Network Cabling on Bit Error Rate Performance. By Paul Kish NORDX/CDT
The Effect of Network Cabling on Bit Error Rate Performance By Paul Kish NORDX/CDT Table of Contents Introduction... 2 Probability of Causing Errors... 3 Noise Sources Contributing to Errors... 4 Bit Error
A Comprehensive Set of Image Quality Metrics
The Gold Standard of image quality specification and verification A Comprehensive Set of Image Quality Metrics GoldenThread is the product of years of research and development conducted for the Federal
DOLBY SR-D DIGITAL. by JOHN F ALLEN
DOLBY SR-D DIGITAL by JOHN F ALLEN Though primarily known for their analog audio products, Dolby Laboratories has been working with digital sound for over ten years. Even while talk about digital movie
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners
Correcting the Lateral Response Artifact in Radiochromic Film Images from Flatbed Scanners Background The lateral response artifact (LRA) in radiochromic film images from flatbed scanners was first pointed
KODAK PLUS-X AEROGRAPHIC Film 2402
AS-45 KODAK PLUS-X AEROGRAPHIC Film 2402 KODAK PLUS-X AEROGRAPHIC Film 2402 is a panchromatic, negative camera film with medium speed, medium-high contrast, and extended red sensitivity for the reduction
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
Basler. Line Scan Cameras
Basler Line Scan Cameras High-quality line scan technology meets a cost-effective GigE interface Real color support in a compact housing size Shading correction compensates for difficult lighting conditions
Dispelling the Myths of Color Scanning
Dispelling the Myths of Color Scanning Capturing High-Quality Images at an Affordable Price Without Compromising Speed or Efficiency By Eastman Kodak Company and Tower Technology, Inc. 2 What we really
Uses of Derivative Spectroscopy
Uses of Derivative Spectroscopy Application Note UV-Visible Spectroscopy Anthony J. Owen Derivative spectroscopy uses first or higher derivatives of absorbance with respect to wavelength for qualitative
Digital Halftoning Techniques for Printing
IS&T s 47th Annual Conference, Rochester, NY, May 15-20, 1994. Digital Halftoning Techniques for Printing Thrasyvoulos N. Pappas Signal Processing Research Department AT&T Bell Laboratories, Murray Hill,
International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in Office Applications Light Measurement & Quality Parameters
www.led-professional.com ISSN 1993-890X Trends & Technologies for Future Lighting Solutions ReviewJan/Feb 2015 Issue LpR 47 International Year of Light 2015 Tech-Talks BREGENZ: Mehmet Arik Well-Being in
Experimental Evaluation of Museum Case Study Digital Camera Systems
Experimental Evaluation of Museum Case Study Digital Camera Systems Erin P. M. Smoyer, Lawrence A. Taplin, and Roy S. Berns Munsell Color Science Laboratory, Rochester Institute of Technology Rochester,
Introduction to Fourier Transform Infrared Spectrometry
Introduction to Fourier Transform Infrared Spectrometry What is FT-IR? I N T R O D U C T I O N FT-IR stands for Fourier Transform InfraRed, the preferred method of infrared spectroscopy. In infrared spectroscopy,
We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement
MAS 40 Mini-Array Spectrometer light measurement Features at a glance Cost-effective and robust CCD spectrometer technology Standard USB interface Compatible with all Instrument Systems measuring adapters
SR2000 FREQUENCY MONITOR
SR2000 FREQUENCY MONITOR THE FFT SEARCH FUNCTION IN DETAILS FFT Search is a signal search using FFT (Fast Fourier Transform) technology. The FFT search function first appeared with the SR2000 Frequency
Coating Thickness and Composition Analysis by Micro-EDXRF
Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing
Application Note Noise Frequently Asked Questions
: What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random
SENSITOMETRIC CHARACTERISTICS OF THE EOS C300 DIGITAL CINE CAMERA
WHITE PAPER SENSITOMETRIC CHARACTERISTICS OF THE EOS C300 DIGITAL CINE CAMERA Updated June 19, 2012 Written by Larry Thorpe Professional Engineering & Solutions Division, Canon U.S.A., Inc. For more info:
Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
PCM Encoding and Decoding:
PCM Encoding and Decoding: Aim: Introduction to PCM encoding and decoding. Introduction: PCM Encoding: The input to the PCM ENCODER module is an analog message. This must be constrained to a defined bandwidth
Determination of source parameters from seismic spectra
Topic Determination of source parameters from seismic spectra Authors Michael Baumbach, and Peter Bormann (formerly GeoForschungsZentrum Potsdam, Telegrafenberg, D-14473 Potsdam, Germany); E-mail: [email protected]
Optimizing IP3 and ACPR Measurements
Optimizing IP3 and ACPR Measurements Table of Contents 1. Overview... 2 2. Theory of Intermodulation Distortion... 2 3. Optimizing IP3 Measurements... 4 4. Theory of Adjacent Channel Power Ratio... 9 5.
Digitization of Old Maps Using Deskan Express 5.0
Dražen Tutić *, Miljenko Lapaine ** Digitization of Old Maps Using Deskan Express 5.0 Keywords: digitization; scanner; scanning; old maps; Deskan Express 5.0. Summary The Faculty of Geodesy, University
The effect of temperature on the bandwidth of telecommunications cabling in commercial buildings. Focus on the IBDN System 4800LX from NORDX/CDT
The effect of temperature on the bandwidth of telecommunications cabling in commercial buildings Focus on the IBDN System 4800LX from NORDX/CDT Table of Contents Introduction... 3 Bandwidth and Information
MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE
MODULATION TRANSFER FUNCTION MEASUREMENT METHOD AND RESULTS FOR THE ORBVIEW-3 HIGH RESOLUTION IMAGING SATELLITE K. Kohm ORBIMAGE, 1835 Lackland Hill Parkway, St. Louis, MO 63146, USA [email protected]
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
RF Measurements Using a Modular Digitizer
RF Measurements Using a Modular Digitizer Modern modular digitizers, like the Spectrum M4i series PCIe digitizers, offer greater bandwidth and higher resolution at any given bandwidth than ever before.
GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3
GAFCHROMIC DOSIMETRY MEDIA TYPE MD-V3 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures
Film Scanner The term Film Scanner can refer to a dedicated slide and negative film scanner or to a capture type scanner.
Film Scanner The term Film Scanner can refer to a dedicated slide and negative film scanner or to a capture type scanner. Negative Scanner The term Negative Scanner can refer to a dedicated slide and negative
White paper. In the best of light The challenges of minimum illumination
White paper In the best of light The challenges of minimum illumination Table of contents 1. Introduction 3 2. The puzzle of light sensitivity 3 3. Do not be fooled! 5 4. Making the smarter choice 6 1.
GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2
GAFCHROMIC DOSIMETRY MEDIA, TYPE HD-V2 WARNING: Store below 25ºC Store away from radiation sources Do not expose film to sunlight Handle film carefully, creasing may cause damage Do not expose to temperatures
Author: Dr. Society of Electrophysio. Reference: Electrodes. should include: electrode shape size use. direction.
Standards for Reportin ng EMG Data Author: Dr. Roberto Merletti, Politecnico di Torino, Italy The Standards for Reporting EMG Data, written by Dr. Robertoo Merletti, are endorsed by the International Society
Using the Normalized Image Log-Slope
T h e L i t h o g r a p h y E x p e r t (Winter 2001) Using the Normalized mage Log-Slope Chris A. Mack, FNLE Technologies, A Division of KLA-Tencor, Austin, Texas Projection imaging tools, such as scanners,
Physics Lab Report Guidelines
Physics Lab Report Guidelines Summary The following is an outline of the requirements for a physics lab report. A. Experimental Description 1. Provide a statement of the physical theory or principle observed
AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis
AMINO ACID ANALYSIS By High Performance Capillary Electrophoresis Analysis of Amino Acid Standards Label free analysis using the HPCE-512 ABSTRACT Capillary electrophoresis using indirect UV detection
Color Management Terms
Written by Jonathan Sachs Copyright 2001-2003 Digital Light & Color Achromatic Achromatic means having no color. Calibration Calibration is the process of making a particular device such as a monitor,
Encoders for Linear Motors in the Electronics Industry
Technical Information Encoders for Linear Motors in the Electronics Industry The semiconductor industry and automation technology increasingly require more precise and faster machines in order to satisfy
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry
Spectrophotometry and the Beer-Lambert Law: An Important Analytical Technique in Chemistry Jon H. Hardesty, PhD and Bassam Attili, PhD Collin College Department of Chemistry Introduction: In the last lab
QUALITY CONTROL OF THE IUE FINAL ARCHIVE
QUALITY CONTROL OF THE IUE FINAL ARCHIVE N. Loiseau 1, E. Solano 1,M.Barylak 2 1 INSA/ESA IUE Observatory, Apdo. 50727, Villafranca del Castillo, 28080 Madrid (Spain). 2 ESA IUE Observatory, Apdo. 50727,
Glencoe. correlated to SOUTH CAROLINA MATH CURRICULUM STANDARDS GRADE 6 3-3, 5-8 8-4, 8-7 1-6, 4-9
Glencoe correlated to SOUTH CAROLINA MATH CURRICULUM STANDARDS GRADE 6 STANDARDS 6-8 Number and Operations (NO) Standard I. Understand numbers, ways of representing numbers, relationships among numbers,
CS171 Visualization. The Visualization Alphabet: Marks and Channels. Alexander Lex [email protected]. [xkcd]
CS171 Visualization Alexander Lex [email protected] The Visualization Alphabet: Marks and Channels [xkcd] This Week Thursday: Task Abstraction, Validation Homework 1 due on Friday! Any more problems
Optical Fibres. Introduction. Safety precautions. For your safety. For the safety of the apparatus
Please do not remove this manual from from the lab. It is available at www.cm.ph.bham.ac.uk/y2lab Optics Introduction Optical fibres are widely used for transmitting data at high speeds. In this experiment,
HP Smart Document Scan Software compression schemes and file sizes
Technical white paper HP Smart Document Scan Software compression schemes and file s Table of contents Introduction 2 schemes supported in HP Smart Document Scan Software 2 Scheme Types 2 2 PNG 3 LZW 3
Plastic Film Texture Measurement With 3D Profilometry
Plastic Film Texture Measurement With 3D Profilometry Prepared by Jorge Ramirez 6 Morgan, Ste156, Irvine CA 92618 P: 949.461.9292 F: 949.461.9232 nanovea.com Today's standard for tomorrow's materials.
Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices
IT Guidance Documents Picture archiving and communication systems (PACS) and guidelines on diagnostic display devices Second edition Board of the Faculty of Clinical Radiology The Royal College of Radiologists
AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE
Atmel 8-bit and 32-bit Microcontrollers AVR127: Understanding ADC Parameters APPLICATION NOTE Introduction This application note explains the basic concepts of analog-to-digital converter (ADC) and the
Outline. Quantizing Intensities. Achromatic Light. Optical Illusion. Quantizing Intensities. CS 430/585 Computer Graphics I
CS 430/585 Computer Graphics I Week 8, Lecture 15 Outline Light Physical Properties of Light and Color Eye Mechanism for Color Systems to Define Light and Color David Breen, William Regli and Maxim Peysakhov
Synthetic Sensing: Proximity / Distance Sensors
Synthetic Sensing: Proximity / Distance Sensors MediaRobotics Lab, February 2010 Proximity detection is dependent on the object of interest. One size does not fit all For non-contact distance measurement,
Multi-Channel Radiochromic Film Dosimetry. Adapted from A.Micke Spain, April 2014
Multi-Channel Radiochromic Film Dosimetry Adapted from A.Micke Spain, April 2014 Single Channel Film Dosimetry Calibration Curve X=RR ave = R ave (D) D R = D R (R ave ) Color channels X=RGB D X = D( X
Basler scout AREA SCAN CAMERAS
Basler scout AREA SCAN CAMERAS VGA to 2 megapixels and up to 120 fps Selected high quality CCD and CMOS sensors Gigabit Ethernet and FireWire-b interfaces Perfect fit for a variety of applications - extremely
Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network
Recent Advances in Electrical Engineering and Electronic Devices Log-Likelihood Ratio-based Relay Selection Algorithm in Wireless Network Ahmed El-Mahdy and Ahmed Walid Faculty of Information Engineering
Basler beat AREA SCAN CAMERAS. High-resolution 12 MP cameras with global shutter
Basler beat AREA SCAN CAMERAS High-resolution 12 MP cameras with global shutter Outstanding price / performance ratio High speed through Camera Link interface Flexible and easy integration Overview Convincing
Overview. Raster Graphics and Color. Overview. Display Hardware. Liquid Crystal Display (LCD) Cathode Ray Tube (CRT)
Raster Graphics and Color Greg Humphreys CS445: Intro Graphics University of Virginia, Fall 2004 Color models Color models Display Hardware Video display devices Cathode Ray Tube (CRT) Liquid Crystal Display
Basler pilot AREA SCAN CAMERAS
Basler pilot AREA SCAN CAMERAS VGA to 5 megapixels and up to 210 fps Selected high quality CCD sensors Powerful Gigabit Ethernet interface Superb image quality at all Resolutions and frame rates OVERVIEW
Color holographic 3D display unit with aperture field division
Color holographic 3D display unit with aperture field division Weronika Zaperty, Tomasz Kozacki, Malgorzata Kujawinska, Grzegorz Finke Photonics Engineering Division, Faculty of Mechatronics Warsaw University
Centrifugal Fans and Pumps are sized to meet the maximum
Fans and Pumps are sized to meet the maximum flow rate required by the system. System conditions frequently require reducing the flow rate. Throttling and bypass devices dampers and valves are installed
Untangling the megapixel lens myth! Which is the best lens to buy? And how to make that decision!
Untangling the megapixel lens myth! Which is the best lens to buy? And how to make that decision! 1 In this presentation We are going to go over lens basics Explain figures of merit of lenses Show how
Digital Image Basics. Introduction. Pixels and Bitmaps. Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color
Written by Jonathan Sachs Copyright 1996-1999 Digital Light & Color Introduction When using digital equipment to capture, store, modify and view photographic images, they must first be converted to a set
RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA
RANDOM VIBRATION AN OVERVIEW by Barry Controls, Hopkinton, MA ABSTRACT Random vibration is becoming increasingly recognized as the most realistic method of simulating the dynamic environment of military
Avalanche Photodiodes: A User's Guide
!"#$%& Abstract Avalanche Photodiodes: A User's Guide Avalanche photodiode detectors have and will continue to be used in many diverse applications such as laser range finders and photon correlation studies.
SHEET 1 COMMUNICATION NETWORKS COMPUTER AND TELEPHONE SYSTEMS
STRUCTURED CABLING SHEET 1 COMMUNICATION NETWORKS COMPUTER AND TELEPHONE SYSTEMS INTRODUCTION In recent years, communications have taken on a crucial role in development for businesses. Communications
Big Ideas in Mathematics
Big Ideas in Mathematics which are important to all mathematics learning. (Adapted from the NCTM Curriculum Focal Points, 2006) The Mathematics Big Ideas are organized using the PA Mathematics Standards
The Optical Design of the High Resolution Imaging Channel for the SIMBIO-SYS experiment on the BepiColombo Mission to Mercury
Mem. S.A.It. Suppl. Vol. 12, 77 c SAIt 2008 Memorie della Supplementi The Optical Design of the High Resolution Imaging Channel for the SIMBIO-SYS experiment on the BepiColombo Mission to Mercury G. Marra
Manual Analysis Software AFD 1201
AFD 1200 - AcoustiTube Manual Analysis Software AFD 1201 Measurement of Transmission loss acc. to Song and Bolton 1 Table of Contents Introduction - Analysis Software AFD 1201... 3 AFD 1200 - AcoustiTube
Quick start guide! Terri Meyer Boake
Film Editing Tutorial Quick start guide! Terri Meyer Boake 1. Preparing yourself and your files: This information is valid for all film editing software: FCP, Premiere (the version of FC being used is
TDS5000B, TDS6000B, TDS/CSA7000B Series Acquisition Modes
TDS5000B, TDS6000B, TDS/CSA7000B Series Acquisition Modes Tektronix oscilloscopes provide several different acquisition modes. While this gives the user great flexibility and choice, each mode is optimized
Gran Via Corts Catalanes, 401 08015 BARCELONA Tel. 933 257 611 Fax 934 263 029 [email protected]
GRAU LUMINOTECNIA, S.A. www.grauluminotecnia.com Gran Via Corts Catalanes, 401 08015 BARCELONA Tel. 933 257 611 Fax 934 263 029 [email protected] Martín Machío, 13 28002 MADRID Tel. 914 158 604
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff
Specifying Plasma Deposited Hard Coated Optical Thin Film Filters. Alluxa Engineering Staff December 2012 Specifying Advanced Plasma Deposited Hard Coated Optical Bandpass and Dichroic Filters. Introduction
Four Wave Mixing in Closely Spaced DWDM Optical Channels
544 VOL. 1, NO. 2, AUGUST 2006 Four Wave Mixing in Closely Spaced DWDM Optical Channels Moncef Tayahi *, Sivakumar Lanka, and Banmali Rawat Advanced Photonics Research lab, Department of Electrical Engineering
On the Operational Quality of Fingerprint Scanners
BioLab - Biometric System Lab University of Bologna - ITALY http://biolab.csr.unibo.it On the Operational Quality of Fingerprint Scanners Davide Maltoni and Matteo Ferrara November 7, 2007 Outline The
Fraunhofer Diffraction
Physics 334 Spring 1 Purpose Fraunhofer Diffraction The experiment will test the theory of Fraunhofer diffraction at a single slit by comparing a careful measurement of the angular dependence of intensity
KIP 3000 Series MULTIFUNCTION SIMPLICITY
KIP 3000 Series MULTIFUNCTION SIMPLICITY KIP 3000 SERIES SYSTEMS The KIP 3000 is a highly hl advanced d wide format digital copy system that provides the ability to easily add network printing and scan-to-file
Determining the Resolution of Scanned Document Images
Presented at IS&T/SPIE EI 99, Conference 3651, Document Recognition and Retrieval VI Jan 26-28, 1999, San Jose, CA. Determining the Resolution of Scanned Document Images Dan S. Bloomberg Xerox Palo Alto
PIXEL-LEVEL IMAGE FUSION USING BROVEY TRANSFORME AND WAVELET TRANSFORM
PIXEL-LEVEL IMAGE FUSION USING BROVEY TRANSFORME AND WAVELET TRANSFORM Rohan Ashok Mandhare 1, Pragati Upadhyay 2,Sudha Gupta 3 ME Student, K.J.SOMIYA College of Engineering, Vidyavihar, Mumbai, Maharashtra,
TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS
TCOM 370 NOTES 99-4 BANDWIDTH, FREQUENCY RESPONSE, AND CAPACITY OF COMMUNICATION LINKS 1. Bandwidth: The bandwidth of a communication link, or in general any system, was loosely defined as the width of
Digital Imaging for Photographic Collections Foundations for Technical Standards
Digital Imaging for Photographic Collections Foundations for Technical Standards Franziska S. Frey James M. Reilly Image Permanence Institute Rochester Institute of Technology Sponsored by the National