PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT



Similar documents
NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

Checklist for submission of application form for Registration

HyperAccess Access Control System

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter

User's Guide. Pinless Moisture Meter. Model MO257

HDR Brachytherapy 1: Overview of QA. Disclosures: Learning Objectives 7/23/2014

Lessons learned from the radiological accident in Mayapuri, New Delhi, India

Monitoring of Intravenous Drip Rate

Accreditation Is Coming

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

REMOTE AFTERLOADING TECHNOLOGY

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

Temperature & Humidity SMS Alert Controller

X-Mind. Instinct for perfection

Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material)

MS9000SAN Rework Station

MINIMAT-EC-Servo Screwdriver Spindles

TABLE 1: Wiring Terminals. Connect to... 1C 1H 2C 2H 1H1C 2H1C 2H2C 3H2C

SERVICE INSTRUCTION R410A. WALL MOUNTEDtype INVERTER SPLIT TYPE ROOM AIR CONDITIONER. Models Indoor unit Outdoor unit

TS510 & TS500. Installation & User Guide. Compatible Equipment

ODYSSEY. Security System Owner s Manual. Kit No. 08E51-SHJ E55-SHJ American Honda Motor Co., Inc. - All Rights Reserved.

Cronos -1 Gamma Object / Tool Monitors

FLOW CALCULATOR INSTRUCTION MANUAL MESURES BAMOPHOX M1 02 E MES FLOW CALCULATOR /1

Regulatory Guide. Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0

Operation Manual. Electrical Release Control System E-CO2. for Gas Fire Extingushing Systems. according FSS-Code Chapter 5

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

Programmable Room Thermostat 7 Day (5-2 Day) Models: / Power Supply: Battery / Mains

Portable Air Conditioner

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25

RVL470. Heating Controller. Building Technologies HVAC Products. Series B

Material test chamber with mechanical convection

3-Component Measuring System

BB-18 Black Body High Vacuum System Technical Description

Atomic Alarm Clock Instruction Manual

Radiation Monitoring System RMS

EXTERNAL DIMENSIONS AND PARTS

Drayton Digistat +2RF/+3RF

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

SOLARCARE SERIES PRODUCT AND APPLICATION GUIDE

API Marine. BMS Presentation. Ballast Measurement Systems

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Homologation Directive for TT04 Vehicle Tracking Systems

Hygro-Thermometer + InfraRed Thermometer Model RH101

COMBIMASS. Technical Data COMBIMASS eco-bio +

Transponder reader CÉLIANE TECHNICAL GUIDE

Model SETR-50 and SETR-51 Trim Tab Control

Applicable Models ME511L, ME551i2, MS31i2, MS33i2, MS51i2, MS53i2

Selection, use and maintenance of portable monitoring instruments

T7079A,B Solid State Remote Temperature Controllers

APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMMAND: PERMIT NUMBER: TYPE OF ACTION:

Vehicle Tracking Service & Fleet Management

User Manual. Video Doorphone CDV-1020AQ

HERZ-Thermal Actuators

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

CruzPro VAF110. AC Volts, Amps, Frequency, kw Monitor

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

Auto Dialer. Manual E-921APQ E-921GPQ

Electro-Pneumatic Bus Door Control Mechanism

Centro-Matic Automatic Lubrication Systems System Controls

SATEX SA Manual. Weegtechniek holland b.v. PRECISION BALANCE. Patroonsweg DA Zeewolde HOLLAND

BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEMS BURGLAR ALARM SYSTEM EURO ALARM WIRELESS BURGLAR ALARM SYSTEM OASIS

WEATHER STATION FOR SOLAR FARM MONITORING

Daily Operator Checks. Monthly Maintenance

Quality Assurance Testing of AtomLab Dose Calibrators

SERVICE INSTRUCTION R410A. WALL MOUNTEDtype SPLIT TYPE ROOM AIR CONDITIONER INVERTER. Models Indoor unit Outdoor unit AOU 9RLFW AOU12RLFW AOU15RLS

SECTION DETECTION AND ALARM

User Manual of WT 493

Wespot. SecNurse A NEW GENERATION TECHNICAL AIDS USER S GUIDE

Captus 3000 Thyroid Uptake And Well System

OPERATING INSTRUCTIONS SECURITY SYSTEM KIT NO.: 08E51-EP American Honda Motor Co., Inc. - All Rights Reserved. 1

CITY OF PHILADELPHIA DEPARTMENT OF LICENSES AND INSPECTIONS ANNUAL CERTIFICATION FOR FIRE ALARM SYSTEMS

GSM AD05 Slave GSM Auto Dialer- Instruction Manual

contents Leading the way in intelligent automation INELS Wireless? Why Click Accessories Complete control

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

DS1104 R&D Controller Board

Incubator with mechanical convection

PREPARATION. INTERNAL CALIBRATION (No external weight needed.) SPAN CALIBRATION

Model 5511 Filler Controller User s Manual Version 1.1 October 2011

3-phase, bidirectional energy meter

AD425S Automatic door operator

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

ICP-CC488. Installation Guide ICP-CC488 Control Panel

GSM HOME SECURITY SYSTEM

Users Manual FA106 / FA306 version 1.0. Field Analyzer. FA106 and FA306. Users Manual

9452/9453 Installation and User Guide

User's Guide. Integrating Sound Level Datalogger. Model Introduction

How To Program An Autodialer

Galleon 6-10K Online UPS. Voltronic Power Technology Corporation

WA Manager Alarming System Management Software Windows 98, NT, XP, 2000 User Guide

Operating Manual for the Electronic Built-in Interval Timer. Micro II (Countdown Timer)

Machine Safety Switches

Integrated Custom Systems

SENTRON Switching and Protection Devices Air Circuit Breakers

Transcription:

RP&AD /Remote QA/ 01 PROFORMA FOR ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT BHABHA ATOMIC RESEARCH CENTRE Radiological Physics & Advisory Division CT & CRS Building, Anushaktinagar Mumbai - 400094 2001

1 1. General Information ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT 1.1 Name and Address of the Institution: (Include Pin Code) Fax: Tel.: e-mail: 1.2 Staff (a) Radiotherapist(s) Sr. No. Name Qualification(s) Experience (b) Medical Physicist(s) / Radiological Safety Officer: Sr. No. Name Qualification(s) Experience 1.3 Treatment Unit Type : LDR / MDR / HDR / PDR Make, Model & Number : Maximum capacity of the Unit : MBq/GBq (mci/ci) Radionuclide : 137 Cs / 192 Ir / 60 Co 1

2 Physical dimensions of : Length (mm): ; Dia. (mm): source(s) No. of sources : Total & individual activity : µgy/h at 1m (mci) of sources (as quoted by the supplier) : µgy/h at 1m (mci) Dummy sources provided? : Yes / No No. of channels : Max. no. of channels used : Simultaneously Method of source transfer : Pneumatic / Drive cable / Stepping Motor/ between applicator & safe Treatment mode : Fixed / Programmable/ Pulsed /... Method of source retraction : Gravity/ Back-up Battery/ Pneumatic / in the event of power failure Special features (if any) : Nurse's station provided? : Yes / No Plan approved on (date) : Source(s) loaded on (date) : Source(s) loaded by : 2

3 2. QA Tests : Mechanical 2.1 Functioning of sensors like : Satisfactory / No pressure, torque, optical 2.2 Coupling between guide- : Satisfactory / No tube and unit 2.3 Coupling between guide- : Satisfactory / No tube and applicator 2.4 Coupling between drive : Satisfactory / No cable and source 2.5 Time taken to drive the source to : --------- sec "ON" position 2.6 Time taken to withdraw the source to : --------- sec "OFF" position 2.7 Source guide free of kinks? : Yes/No 2.8 Mechanical integrity of applicators & : Acceptable / Not acceptable position of shield, if any 3. QA Tests : Electrical 3.1 Interlocks 3.1.1 Treatment Room Door(s) 3.1.2 Guide Tube & Treatment Unit 3.1.3 Applicator & Guide Tube 3.1.4 Emergency stop button to interrupt the irradiation 3.2 Source Safe Display 3.2.1 Treatment "OFF" indicator 3.2.2 Treatment "ON" indicator 3

4 3.3 Control Console Display 3.3.1 Treatment "OFF" indicator 3.3.2 Treatment "ON" indicator 3.3.3 Source ON/OFF indicator 3.3.4 Door interlock indicator 3.3.5 Retrieval of the source indicator 3.3.6 Retrieval of the dummy indicator 3.4 Control Console Function 3.4.1 Power switch 3.4.2 Parameter reset switch 3.4.3 Treatment `ON switch 3.4.4 Treatment `OFF switch 3.4.5 Emergency `OFF switch 3.4.6 Treatment and elapsed time display counters 3.4.7 Multi-channel indexer function and sequence displays 3.4.8 Back-up system(s) 4

5 4. QA Tests : Radiation (attach relevant data for evaluation) 4.1 Leakage and Contamination ( Limit: 200 Bq) (Using a wet swipe measure the maximum removable activity on the surface of the source / source holder) 4.1.1 Removable activity on Source / Source holder : Bq 4.1.2 Contamination check done on (date) : 4.2 Positioning Accuracy and Uniformity (Tolerance: ± 1 mm) (use autoradiograph / radiograph wherever required) 4.2.1 Coincidence between dummy : ------------- mm and active source positions 4.2.2 Accuracy of source position within the : -------------- mm applicator 4.2.3 Reproducibility of dummy source position : -------------- mm 4.3 Temporal Accuracy (Tolerance: ± 1 %) 4.3.1 Timer error : -------------- 4.3.2 Timer linearity : --------------- 4.3.3 End error : --------------- sec ('or' %) 4.4 Source Strength Verification 4.4.1 Quoted source strength (as on ) : ----------- µgy. h -1 at 1m ( GBq / Ci ) 4.4.2 Measured source strength (as on ) : ---------- µgy. h -1 at 1m ( GBq / Ci ) 4.4.3 Deviation between specified : ----------- % and measured source strength (Tolerance: ± 3%) 4.4.4 Deviation between measured : ------------- % mean and individual source strength (Tolerance: ± 5%) In a separate sheet explain the procedure and the instrument(s) used. In case of multiple sources, give the average source strength and the variation among individual sources. 5

6 5. QA Tests: Treatment Planning System 5.1 Digitizer and Localization Software: (Tolerance: 3% or 1mm) 5.1.1 Accuracy of digitization of : Point co-ordinates 5.1.2 Accuracy of data transfer from : orthogonal radiographs 5.2 Calculation Algorithm: 5.2.1 Source specification required for TPS : mci / mg.ra.eq./ MBq / Ref. Air. Kerma Rate 5.2.2 Initial activity quoted by the supplier : Current activity (with date) : 5.2.3 Factor used to convert quoted source strength for TPS input : 5.2.4 Agreement of source decay correction : 5.2.5 Agreement between TPS and : published/manual calculation for single source, at relevant points 5.2.6 Agreement for multiple source loading : (Specify the procedure and the result) (Tolerance: 3% or 1 mm) 6. QA Tests: Radiation Safety 6.1 Gamma Zone Monitor: Provided? Working? : Yes / No : Yes / No 6.2 CCTV Working? : Yes / No 6.3 Unit Leakage Measure the dose rate on top, bottom and four sides of the container at 5 cm from the surface and one meter from the centre of source storage for source 'OFF condition: 1 2 3 4 5 6 Measured dose rate (µgy.h -1 ) at 5cm : from the surface of the source storage (Tolerance: 100 µgy.h -1 ) Measured dose rate (µgy.h -1 ) at 1 m : from the centre of the source storage (Tolerance: 10 µgy.h -1 ) Note: The above tolerance limits are applicable when the unit is loaded with the maximum capacity. 6

7 6.4. Survey of the Installation: In the Figure below, mark the occupancy on all sides, above and below the installation and measure the maximum dose rates at these locations, with the applicator loaded with maximum source activity and kept at the centre of the treatment table. Minimum locations recommended for radiation protection survey of the installation are: A, A, B, B Just outside the four walls of the installation at the level of source 'S' C Roof of the installation, at the level of source 'S' D Just outside entrance door E &F At the level of air conditioners G - Entrance to control panel (S : Source position ; CP : Control Panel ) F E A.C. A A.C. B S B F l o o r L e v e l R o o f L e v e l C D C.P A Section Across AA G Fig. 1- Typical Layout of a Remote Afterloading Brachytherapy Facility 7

8 6.4.1 Source Activity (GBq/Ci) during survey : 6.4.2 Instrument(s) used for survey : 6.4.3 Measured dose rate (µgyh -1 ) at positions marked in Fig. 1: A A B B C D E F G Note: The quoted dose rate should be the maximum measured in the specified direction. 7. Instruments Used for Measurement Sr.No. Type Make &Model Cal. Factor (& Ambient conditions) Date of Calibration Calibrated at Checked by Measured by Signature : Signature : Name : Name : Designation : Designation : Date : Date : : Instrument(s) used for survey : 8