EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
|
|
|
- Gladys Barker
- 10 years ago
- Views:
Transcription
1 Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91
2 Prepared by the Radiation Protection Branch For further information please contact: Information Officer Environment Protection Authority GPO Box 2607 Adelaide SA 5001 Telephone: (08) Facsimile: (08) Free call (country): Website: < ISBN Issued June 2009 Updated April and December 2011 Disclaimer This publication is a guide only and does not necessarily provide adequate information in relation to every situation. This publication seeks to explain your possible obligations in a helpful and accessible way. In doing so, however, some detail may not be captured. It is important, therefore, that you seek information from the EPA itself regarding your possible obligations and, where appropriate, that you seek your own legal advice. Environment Protection Authority This document may be reproduced in whole or part for the purpose of study or training, subject to the inclusion of an acknowledgment of the source and to it not being used for commercial purposes or sale. Reproduction for purposes other than those given above requires the prior written permission of the Environment Protection Authority.
3 Table of contents Abbreviations...1 Definitions Introduction Scope Conventions Precautions and limitations Test instruments Modifying an apparatus Test equipment Documentation Test protocol Owner details Apparatus details Tester information Test instrument information Stationary tube* Open ended collimator Beam limiting device* Exposure light* Audible signal* Control panel lights* Radiation shield (hand held only) Log book (hand held only) Dead man switch Exposure termination* Timer reset Zero timer setting* Trial exposure* Control parameters Mains switch and indicator Focal spot Multiple tubes Radiation Produced When Energised sign* Radiation Area sign* Good working order* Consistency* Half value layer* Tube voltage...16
4 2.28 Linearity Beam size* Leakage Operator protection* Shielding* References...22 Annex 1 Radiation signs...23 List of figures Figure 1 Operator position...19 Figure 2 Radiation Produced When Energised...23 Figure 3 Radiation symbol...23 Figure 4 Radiation symbol plus radiation warning...23 List of tables Table 1 Minimum HVL (in mm of Al) for various set kvp...16
5 Abbreviations General Al ALARA CR DR EPA HVL min TPC aluminium as low as reasonably achievable computed radiography direct radiography Environment Protection Authority half value layer minimum Third Party Certification Units and parameters A Gy h lx m s Sv Vp Ampere Gray hour lux metre seconds Sievert volts peak Prefixes n nano (10-9 ) µ micro (10-6 ) m milli (10-3 ) c centi (10-2 ) k kilo (10 3 ) 1
6 Definitions Fixed apparatus means any apparatus that is neither a mobile apparatus nor a portable apparatus. This includes any mobile apparatus that has been modified in such a manner so that it can no longer be moved from place to place or a portable apparatus that has been modified in such a manner so that it can no longer be carried manually from place to place. Fixed protective barrier means a barrier that cannot be readily removed and has a lead equivalence of 0.15 mm. Mobile apparatus means apparatus that is designed and constructed so as to be moveable from place to place for use as required but does not include a portable apparatus. Portable apparatus means any apparatus that is designed to be carried manually from place to place for use as required. Hand held apparatus means any apparatus that is designed and constructed so as to be used while held by the operator. 2
7 1 Introduction 1.1 This document details the protocol for conducting compliance testing of plain dental X-ray apparatus which includes fixed, mobile and portable apparatus. It also includes apparatus that are used in a hand held manner. It does not include orthopantomographic, cephalometric or computed tomography apparatus. 1.2 Scope The testing is performed in order to satisfy the legislative requirements of the Radiation Protection and Control Act 1982 (the RPC Act) and the Radiation Protection and Control (Ionising Radiation) Regulations 2000 (the Regulations) and can only be performed by persons authorised by the Environment Protection Authority (EPA). Such persons consist of Authorised Officers under the Act and those accredited under the Third Party Certification (TPC) program 1 administered by the EPA. 1.3 Conventions Throughout this document labelled text is used to indicate additional information as discussed below. Important Note Important information for the attention of the tester. Tip Indicates that additional information is provided to assist in the interpretation of the relevant test. Tolerance Some tests have a tolerance that can be applied to the measured result. This is to allow for potential errors in measurements. Tolerances are added and/or subtracted from the regulatory limits as indicated. * Some older apparatus may have been registered before 1 April 1986 under the Radioactive Substances and Irradiating Apparatus Regulations 1962 and as such need only fulfill the requirements of sections indicated with an asterisk (*). However, testers must still perform all tests and evaluate each one for compliance. The EPA will assess overall compliance of older apparatus prior to registration. Denotes that this test applies only to an apparatus that is used in a hand held manner. Note that the owner must have an exemption from regulation to use an apparatus in a hand held manner. 1.4 Precautions and limitations Image receptors Before any X-ray apparatus is energised ensure that any radiographic film or imaging plate is located well away from the X-ray apparatus, so as to avoid possible exposure to X-rays. Personal radiation monitoring device Ensure an appropriate personal radiation monitoring device is placed on your person prior to commencement of testing. For further information on these devices see EPA Guidelines Personal radiation monitoring devices. 1 Information regarding the TPC program is contained in the document EPA Guidelines Program requirements for compliance testing of diagnostic X-ray apparatus. 3
8 1.5 Test instruments Testing for compliance requires non-invasive measurements of radiation output, tube voltage, exposure time, and half value layer (HVL) of the primary X-ray beam and any radiation that might be present due to the leakage from the X-ray tube housing. The shielding effectiveness of a protective barrier is determined by means of radiation transmission measurements. To cater for the different requirements, test instruments come in variety of types and measure a variety of parameters, depending on the type and model. Broadly they can be divided into having ionisation chambers, or one or more semiconductor detectors. In particular the sensitivity (the lowest measurable dose) of instruments varies and for some measurements (ie shielding and leakage) doses in the µgy or ngy ranges may be required. Measurement of HVL can be performed directly by some instruments and indirectly by others by means of dose measurements with and without added aluminium (Al) filters placed in the primary beam. With this in mind, testers should ensure that the instruments they are using are appropriate to the task. 1.6 Modifying an apparatus An apparatus must be compliance tested in the condition in which it has been installed and will be used. Testers must not add any additional equipment (eg mains voltage conditioners) to the apparatus during compliance testing. 1.7 Test equipment The following test equipment may be required: radiation meter/detector Al filters as required ( 99% purity) image receptor (occlusal dental film or equivalent) lead block - 1 mm thick (minimum) x 65 mm (minimum) x 65 (minimum) tape measure (at least 5 m is recommended) vernier callipers or accurate ruler photographic developing facility (eg darkroom, chemicals, processing tank, etc) masking tape or equivalent. 1.8 Documentation Test Report For each apparatus tested the results must be documented in the form of a Test Report, a copy of which must be provided to the owner of the apparatus within 14 days of the date of testing. The preparation of the Test Report is the responsibility of the accredited tester. The form and style of the report is at the discretion of the tester but results must be clear and accurate. Note that, accredited testers need not routinely provide the EPA with Test Reports. However the EPA may request a copy of a report at any time. 4
9 Compliance Statement For each apparatus tested the tester must indicate the compliance status of the apparatus or the area in which it is installed by indicating either a Yes (compliant), No (not compliant) or N/A (not applicable) on the appropriate Compliance Statement. Testers must provide a Compliance Statement to the EPA using the prescribed form within 14 days of the date of testing. Posted and electronic copies (ie and facsimile) are accepted. If submitting a statement via or facsimile you need not also submit the original by post. The following Compliance Statements are applicable to the protocols of this document: Compliance Statement (Form 91A) Dental X-ray apparatus used for plain radiography. 5
10 2 Test protocol This test protocol must be followed by all personnel performing compliance testing. Information and results, as specified in this section, must be recorded in the Test Report. Important Note The tests specified in this protocol use non-invasive methods and do not require the tester to remove, modify or otherwise interfere with any part of an apparatus or structure relating to the environment in which the apparatus is installed. The consistency test (refer section 2.25) must be the first test performed for which the X-ray tube is to be energised. This ensures that results are reflective of a tube that is operated from cold. All other tests can be performed in an order at the discretion of the tester. 2.1 Owner details Record the name, address, and contact details (phone number and/or address) of the client. Note: the client can be a person and/or a business. 2.2 Apparatus details Make & model: record the full make and model of the apparatus. Generator/controller serial number: record the serial number of the X-ray generator/controller unit of the apparatus. If the serial number label is not present on the apparatus, simply record as unknown. Tube serial number: record the serial number of the X-ray tube of the apparatus. If the serial number label is not present on the apparatus, simply record as unknown. Tube housing serial number: record the serial number of the X-ray tube s housing of the apparatus. If the serial number label is not present on the apparatus, simply record as unknown. Apparatus location: record the room name or number (eg Surgery 1) in which the apparatus is located. Filtration (mm of Al): an X-ray tube will have an inherent filtration built into the tube itself and possibly additional filtering added to the tube housing. The material used is generally aluminium, though other materials (eg copper) are possible. Record the filtration expressed in the appropriate unit (eg mm of Al). 2.3 Tester information Contact details: record the name, address, and contact details (phone number and/or address) of the person conducting the compliance test. Test date(s): record the date or dates on which the tests are performed. 2.4 Test instrument information Meter/detector: record the make and model of the instrument s meter and its detector. Tip Test instruments can consist of a single module, such as an ionisation chamber and a meter read-out in one. Alternatively they may have separate detectors (eg ionisation chamber and kvp sensor) connected to a meter. Record the make and model of all modules used. 6
11 Serial Numbers: record the serial number or numbers of the instrument s meter and detector. If not known simply record as unknown. Calibration: all test instruments used must be calibrated at least once every two years to ensure that measurement accuracy is maintained. Record the date the last time a calibration of the instrument was performed. 2.5 Stationary tube* To eliminate the possibility of the tube moving once it has been correctly positioned. This minimises the possibility of retakes and of an unnecessary exposure of radiation to the patient. Place the tube housing in positions that would be typically used in dental radiography. For each position visually check whether or not the tube head moves. The tube housing must remain stationary. Regulation 91(3). 2.6 Open ended collimator To ensure that the apparatus is fitted with the correct style of beam limiting device. Visually examine the collimator (ie cone). The collimator must be open ended. Regulation 91(5). 2.7 Beam limiting device* To ensure the length of the beam limiting device (ie cone) is sufficient to limit the radiation dose to the patient. Tip The cone length, in part, determines the source-to-skin distance and hence the skin entrance dose the patient receives. The amount of scattered radiation is also reduced by the presence of a cone. 7
12 The focal spot is indicated on the tube housing usually by a small dot or square. Measure the distance from the focal spot (in mm) to the end of the cone. For an apparatus first registered after 1 April 1986 the minimum distance from the outer end of the cone to the focal point of the X-ray tube must be at least 200 mm. For apparatus first registered before 1 April 1986 this distance is 100 mm. Regulation 91(6). 2.8 Exposure light* To ensure that the operator is given a visual warning that the X-ray tube is energised. Energise the X-ray tube and take note of any visual warning signals. Note the colour of the warning signal. A red or amber light must be incorporated into the apparatus to indicate that the X-ray tube is energised. Tip If an apparatus was first registered before 1 April 1986 it only needs to comply with this test; but only when it does not comply with the Audible Signal test in section 2.9. Regulation 91(8)(a). 2.9 Audible signal* To ensure there is an audible warning that the X-ray tube is energised. Energise the X-ray tube and take note of any audible warning signals. An audible signal must be incorporated into the apparatus to indicate that the X-ray tube is energised. Tip If an apparatus was first registered before 1 April 1986 and it complies with this test it need not comply with the Exposure Light test in section 2.8 and the Control Panel Lights test in section Regulation 91(8)(b). 8
13 2.10 Control panel lights* To ensure the operator has a clear and visible indication that X-rays are being produced. Take note of the colour of any lights present on the controller of the apparatus. The lights on the controller must not be the same as the exposure light in section 2.8. Tip If an apparatus was first registered before 1 April 1986 it only needs to comply with this test; but only when it does not comply with the Audible Signal test in section 2.9. Regulation 91(9) Radiation shield (hand held only) Important Note This test applies only to the NOMAD Portable X-Ray apparatus. For an owner to use an apparatus in a hand held manner requires and exemption from regulation. Testers are advised to contact the EPA for advice. To ensure that a radiation shield is fitted to the apparatus and that it cannot be removed from the cone. Inspect the shield and note if the shield is capable of being detached from the end of the cone. The shield must be located near the end of the cone and must not be capable of being detached. A shield that slides along the cone is permissible. Registration Condition 1002(3) Log book (hand held only) Important Note This test applies only to the NOMAD Portable X-Ray apparatus. For an owner to use an apparatus in a hand held manner requires and exemption from regulation. Testers are advised to contact the EPA for advice. To ensure that appropriate information is kept to provide a reference for any radiation exposure related issues. Inspect the log book to ensure the required information is being kept. 9
14 A log record shall be maintained for each apparatus, containing at least the following information: the date of the radiographic procedure the name of the person taking a radiograph the location of the radiographic procedure the name and identification number of the patient the number of radiographs taken the exposure time the configuration of the apparatus (ie a hand held or on a stand). Registration Condition 1002(7) Dead man switch To ensure that an X-ray exposure will terminate when the exposure switch is released. This is known as the dead man function. Position the dose meter under the cone. Set the exposure time to at least 0.5 s and press the exposure switch. Release it prior to the end of the exposure time. The X-ray apparatus must stop producing radiation when the exposure switch is released. This can be verified by checking that the measured dose is less than for the full exposure. Regulation 91(13) Exposure termination* To ensure that an operator can energise the X-ray tube for no longer than is set by the timer of the apparatus. This can be tested in the course of doing other measurements (eg Consistency, Linearity, etc). Simply confirm that the apparatus has a timer that after the period defined by the timer s settings terminates the exposure. A timer must be provided that, after the period defined by the timer s setting, terminates the exposure either by a set exposure time or by the product of its tube current and exposure time (ie mas). 10
15 Regulation 91(14) Timer reset To ensure that, after an exposure has been taken, there is no possibility that the X-ray tube can be energised for a period differing from what was initially set by the operator. This can be tested in the course of doing other measurements (eg Consistency, Linearity, etc). Confirm that when an exposure is taken, the timer resets itself to its initial setting or zero. When an exposure terminates the timer must automatically reset to its initial setting or to zero. Regulation 91(14)(a) Zero timer setting* To ensure that there is no possibility that an operator can energise the X-ray tube when the timer of the apparatus is set to zero. If the timer has a zero setting, select that time. Position the dose meter under the primary beam and press the exposure switch. Note whether or not there was any audible signal, visual warning or dose meter reading. It must not be possible to energise the X-ray tube if the timer is set to zero. Note: if the timer does not have a zero setting then the apparatus is considered compliant with this regulation. Regulation 91(14)(b) Trial exposure* To ensure that the operator is not required to make a trial exposure (as that would be an unnecessary exposure of radiation to the patient) to establish the correct tube voltage, tube current or exposure time. 11
16 Confirm (for an apparatus with variable tube voltage, tube current or exposure time) the apparatus can have its tube voltage, tube current or exposure time set such that it does not require a trial exposure. If an apparatus is capable of having its tube voltage, tube current or exposure time settings varied these control settings must not require a trial exposure in order to set them. Regulation 91(15)(a) Control parameters To ensure that an exposure of radiation to the patient or operator is set by control parameters (kvp, ma and exposure time) whose values are accurately reflected by the labelling on the apparatus. This allows the operator to be fully informed of the nominal values of the control parameters. Confirm, for an apparatus with fixed control parameters (ie kvp, ma or exposure time), a label is affixed to the tube housing or control panel indicating the values. Tip kvp must be measured to confirm that it accurately reflects displayed values. Refer section 2.27 for details. If an apparatus has fixed kvp, ma or exposure time, the apparatus must have labels affixed to its tube housing or control panel indicating the value of these control parameters. Regulation 91(15)(b) Mains switch and indicator To ensure that when only the mains power is applied to the apparatus the X-ray tube is not automatically energised and that the operator can clearly see that the apparatus has mains power applied to it by means of an indicator light or lights. Note whether or not the apparatus has a mains switch fitted that only controls the supply of mains to the apparatus and that there is some form of indicator alight when the mains power is applied. The apparatus must have a mains switch fitted that only controls the supply of mains power to the apparatus and there must be a visual indication the mains is in the ON position. 12
17 Tip the mains indicator light need not be a dedicated light and can include indicators such as a digital display or similar. Regulation 91(16) Focal spot To ensure that the position of the focal point of the X-ray tube is known. Note whether or not the focal spot is clearly indicated on the X-ray tube housing. The position of the focal spot must be clearly indicated on the X-ray tube housing. Tip The precise location of the X-ray tube s focal point is not often known, but is usually indicated near or at the centre of the tube s housing. However if there is no focal spot indicated or the spot is not sufficiently clear, the apparatus is not considered compliant with the relevant regulation. Regulation 91(17) Multiple tubes To ensure that if multiple X-ray tubes can be operated from a single control panel, the incorrect tube cannot be inadvertently energised and risk the unintended radiation exposure. Note: if the apparatus only has a single tube this test is not applicable. 1. Position the dose meter under the cone on one of the X-ray apparatus under test. 2. Use the control panel to select the apparatus that will be energised. There must be indication that the correct apparatus has been selected. 3. Press the exposure switch from the control panel. Confirm that there is any dose indicated by the meter. Note also if the tube energised indicator light (near the tube housing) of the relevant apparatus is illuminated. 4. Repeat for any other X-ray tubes that are operated from the control panel. When more than one X-ray tube can be operated from a single control panel, it must not be possible to energise more than one X-ray tube at the same time and there must be an indication at or near each tube housing and on the control panel showing which X-ray tube is selected. Regulation 91(18). 13
18 2.22 Radiation Produced When Energised sign* Ensures that an appropriate warning sign is displayed to indicate that the apparatus will emit radiation when it is energised. Note whether or not an appropriate sign is displayed on or near the controller of the apparatus. An appropriate sign is one that complies with Australian Standard Safety Signs for the Occupational Environment as it applies to warning signs. It must bear the radiation symbol, the words RADIATION PRODUCED WHEN ENERGISED or words to that effect, and be clearly legible at 2 m. Refer Figure 2 for an example. Tip As some apparatus have a controller that is too small to fit the required warning label (for example a hand held type), it can be displayed in a position that is near to where the controller is normally placed. If this is impractical, it is acceptable to have a sign that is displayed somewhere else on the apparatus, such as its tube housing or arm. Regulation Radiation Area sign* To ensure that an appropriate warning sign is displayed, at each entrance to the room, in which an X-ray apparatus is installed (for fixed apparatus) or normally kept or used (for mobile apparatus). Confirm that an appropriate sign is displayed at or near each entrance to the room where a fixed apparatus is installed. For a mobile apparatus a warning sign must be placed at each entrance to the room where it is normally kept or used. In the case of an open area installation, a sign at each walkway or access route to the installation is required. An appropriate sign is one that complies with Australian Standard Safety Signs for the Occupational Environment as it applies to warning signs, bears the radiation symbol, has a total surface area of not less than 4500 mm 2, and is clearly legible at 2 metres. In addition to the radiation symbol the sign may use wording. If this is the case, it must bear the words RADIATION AREA or X-RAYS or words to that effect. Refer to Figure 3 and Figure 4 for examples. Tip A doorway which leads to an adjacent room (eg a storeroom, office, etc) that cannot be entered from any other room, is not regarded as an entrance to the apparatus room and hence the relevant regulation does not apply to such an entrance. Regulation
19 2.24 Good working order* To ensure that the X-ray apparatus is in good working condition. Record any abnormal performance or condition that was detected in the course of conducting the compliance tests. This includes but is not limited to ensuring that the collimator housing is not loose, the tube housing is adequately fastened, the exposure switch does not operate intermittently or any other abnormality that is not covered by another regulation. No abnormalities are detected. Regulation 9(2) Consistency* Important Note The measurements must be taken without any warm-up exposures. To determine the consistency of radiation output so as to avoid an unnecessary variation in radiation output and therefore an unnecessary variation in patient dose. 1. Take at least five consecutive measurements of radiation output (ie dose) at a fixed tube voltage, tube current and exposure time. 2. Calculate the coefficient of variation of the measurements of radiation output by dividing the standard deviation of the radiation measurements by the average of the radiation measurements. The coefficient of variation must be less than or equal to Regulation 91(19) Half value layer* To ensure that the beam is of adequate quality so as to minimise the dose to the patient. Using dose measurements and filters 1. Insert into the beam an Al filter with a thickness greater than or equal to the thickness specified in Table 1 for selected set kvp. 15
20 2. Take an exposure and measure the radiation output at a fixed distance from the focal point. 3. Repeat the measurement but without an Al filter in the beam. 4. Calculate the transmission ratio by dividing the filtered radiation output by the unfiltered radiation output. or direct measurement Take an exposure and measure the HVL. Note for this method Al filters are not required. Table 1 Minimum HVL (in mm of Al) for various set kvp Set kvp HVL in mm of Al The radiation transmitted through the Al filter must be at least half the unfiltered radiation. That is, a transmission ratio 0.5. Regulation 91(7) Tube voltage To ensure that the indicated tube voltage, as specified on the apparatus labelling (refer section 2.18), is representative of the measured tube voltage. A tube voltage that is too low will increase patient dose and one that is a too high can result in poor image quality. Place the detector at the end of the cone of the apparatus. Measure the tube voltage for all set kvp values that the apparatus is capable of generating. Tip Some test instruments are dependent on the filtration of the apparatus and the instrument itself. To correct for filtration, additional Al filters can be added to improve the accuracy of the measurement. 16
21 Measured tube voltages must be indicative of tube voltages shown on the apparatus labelling. Therefore, all measured tube voltages must be within 5% (plus a tolerance) of the labelled tube voltage. If not, the X-ray apparatus may need to be adjusted, if adjustments are possible, or relabelled. Tolerance It is permissible to add a tolerance to a tube voltage measurement, as different test instruments measure tube voltage in different ways affecting their relative accuracy. The EPA recommends that no more than 3% tolerance is added. Using the square root of the sum of the squares method provides an adequate overall tolerance. Therefore the allowable deviation of each tube voltage is: ( ) 6%. Regulation 91(15)(b) Linearity To ensure the X-ray unit generates a predictable and proportional radiation output. 1. Measure the radiation output for at least five different exposure times over the range 0.1 s to 1.0 s and at any set tube voltage. For an apparatus with a variable set tube current include at least one measurement for each tube current. 2. Normalise the measurements by dividing each dose measurement (mgy) by the product of the set tube current (ma) and the set exposure time (s) to obtain the ratio mgy/mas. 3. Calculate the coefficient of variation by dividing the standard deviation of the mgy/mas ratios by the average of the mgy/mas ratios. The coefficient of variation must be less than or equal to 0.1. Regulation 91(20) Beam size* To determine the size of the X-ray beam at the end of the beam applicator. This ensures that patient dose does not increase unnecessarily due to an oversized beam. Tape or press the beam applicator against an image receptor and make a radiation exposure. Process the radiograph (or equivalent) and measure the maximum dimension of the X-ray beam. Important Note The use of fluoroscopic equipment to measure beam size may not be acceptable if the persistence of the image is not sufficient for an accurate measurement of beam size. 17
22 The maximum dimension of the X-ray beam must be less than or equal to 60 mm. Tolerance It is permissible to add a tolerance to the beam size measurement, to account for the inaccuracy of the measurement instrument used. The EPA recommends that no more than 1.0 mm is added. Therefore the maximum allowable beam size is 61.0 mm. Regulation 91(5) Leakage To ensure that unacceptable levels of radiation cannot leak through the tube casing and the beam applicator. With at least 1 mm of lead over the end of the beam applicator, select the highest tube voltage and tube current. Position the dose meter around 250 mm from the focal spot and make a measurement of leakage radiation for each orientation of the tube housing (ie the two sides, the top and bottom, and the rear of the tube housing). The equivalent dose rate for the maximum rated tube voltage and current must be less than or equal to 1 msv per hour at a distance of 1 metre from the focal spot 2. Tip Some instruments can measure dose rate directly. In this case simply compare the measured dose rate with the specified requirement. Regulation 91(2) Operator protection* Important Note An apparatus that has an exposure control switch on the control panel, that switch must be disabled if the control panel is not located at least 2 metres from the X-ray tube and the patient. If the apparatus is capable of being used in a hand held manner, the owner may have been granted an exemption from regulation 91(10). Testers are advised to contact the EPA for advice. To ensure that the operator is afforded adequate protection from radiation exposure. 1. Identify the position the operator normally stands to take radiographs. 2. From the normal operator position, establish that the operator can locate themself: a outside the primary X-ray beam; and 2 For X-rays an equivalent dose rate of 1 msv/h is equal to a dose rate of 1 mgy/h. 18
23 b at least 2 metres from the X-ray tube and from the patient; or c behind an appropriate fixed protective barrier (refer to step 3). Tip The possible areas the primary beam is capable of being directed are dependent on the range of positions the X-ray tube is used for dental radiography and its degrees of freedom. A good rule of thumb is to assume that if the cone is physically capable of being pointed in the direction of the operator then that position will require shielding or the operator must not position themselves there. As a guide, the area in which the primary beam might be directed is represented by an arc approximately 270 behind the head patient as shown in Figure 1. The remaining area, at the foot of the patient, is generally a zone in which the primary beam cannot usually be directed as hence does not require shielding as long as the 2 m rule is observed. 3. If an operator cannot position themself outside the primary X-ray beam and at least 2 m from the tube/patient then the operator must be shielded behind a fixed protective barrier. To confirm that a barrier is a protective barrier use the following procedure. a Set the tube voltage and tube current to maximum available values. Use an exposure time of at least 0.5 s. b c Measure the radiation output in air (I 1 ) at a known distance (d 1 ) from the focal spot. Measure the radiation output through the barrier (I 2 ) at a known distance (d 2 ) from the focal spot. d Calculate the transmission ratio (T X ) using the inverse square law: T X = (I 2 /I 1 ) x (d 2 /d 1 ) 2. e Multiple the transmission ratio by 100 to obtain the amount of radiation transmitted as a percentage. Tip Typically operators in dental clinics do not have dedicated fixed screens within the room, but instead use a structural radiation barrier that is already built into a wall or a door. Barriers made of solid brick, 36 mm plasterboard, 1 mm of steel sheet, 20 mm plate glass or lead glass will be at least 0.15 mm lead equivalence. 270 Shielding required in this direction Tube/patient to operator distance 2 m Shielding not required in this direction Operator Figure 1 Operator position The exposure control switch must be arranged so that the operator can remain: outside the primary X-ray beam; and at least 2 metres from the X-ray tube and from the patient; or behind an appropriate fixed protective barrier. 19
24 Tip A protective barrier is one that has a lead equivalence of 0.15 mm. If the transmitted radiation is no more than 10% of the incident radiation the barrier is deemed to comply with this test. A fixed barrier is a one that cannot be readily removed. The EPA accepts that, while a shielded door loses its shielding capacity by simply opening the door, the door is fixed as it cannot be readily removed. There is no regulatory requirement governing the height or width of a protective barrier. But the barrier must be of sufficient dimensions so that the operator cannot be exposed to the primary beam. While the EPA recommends to owners and operators that the operator should be able to view the patient during a radiation exposure, this is not a regulatory requirement and not mandatory under this protocol. For an apparatus used in a hand held manner, the owner may have been granted an exemption from Regulation 91(10). Please contact the EPA for advice. Regulation 91(10) Shielding* Important Note When evaluating a room that may have been previously assessed (ie by a accredited tester or an Officer of the EPA), do not assume that the shielding is still adequate as the room s configuration and shielding may have changed since its last inspection. To ensure members of the public (ie any person not operating the apparatus) are appropriately shielded from the radiation from an energised X-ray apparatus. 1. Floor plan: make a sketch of the layout of the area in which the apparatus is installed. The sketch should include detail of: a b c d e f g the X-ray apparatus the patient chair windows (where applicable) doors (including all entrances used to directly access the area) the normal position of the operator adjacent areas (eg hallway, reception area, offices, staff room, store room, dark room, adjacent business, adjacent surgeries, external car parks, external walk ways) the approximate dimensions of important features (eg size of room and distances from the apparatus to the operator position). Note that the sketch need not be to a precise scale. 2. Normally occupied: identify areas where the primary beam of the apparatus is likely to be directed that are: a b normally occupied by a person: and are less than 5 metres from the X-ray tube. For each area that is normally occupied annotate on your sketch accordingly. Tip an area is normally occupied when a person is located in the area for a period other than a short period of time. Examples of normally occupied areas are a reception area, a patient waiting area, a patient recovery area, a 20
25 Practice Manager s office, busy staff rooms, an adjacent surgery, an adjacent business of any kind. Conversely areas such as hallways, infrequently used offices, infrequently used staff rooms, store rooms, dark rooms, external car parks, and external walk ways are not generally considered normally occupied. 3. Radiation barriers: for each area that has been identified as being normally occupied confirm that between the area of concern and the X-ray tube there is a fixed protective barrier. Often the only way to confirm this is by a performing a transmission measurement. Use the following procedure. a Set the tube voltage and tube current to maximum available values. Use an exposure time of at least 0.5 s. b c Measure the radiation output through air (I 1 ) at a known distance (d 1 ) from the focal spot. Measure the radiation output through the barrier (I 2 ) at a known distance (d 2 ) from the focal spot. d Calculate the transmission ratio (T X ) using the inverse square law. That is: T X = (I 2 /I 1 ) x (d 2 /d 1 ) 2. e f Multiple the transmission ratio by 100 to obtain the amount of radiation transmitted as a percentage. Annotate on the sketch the points (ie wall, door, etc) at which shielding measurements are performed. Tip Barriers made of solid brick, 36 mm plasterboard, 1 mm of steel sheet, 20 mm plate glass or lead glass will be at least 0.15 mm lead equivalence. If the primary beam of the apparatus is likely to be directed at an area normally occupied by a person that is less than 5 metres from the X-ray tube, a fixed protective barrier must be provided. Tip A protective barrier is one that has a lead equivalence of 0.15 mm. If the transmitted radiation is no more than 10% of the incident radiation the barrier is deemed to comply with this test. A fixed barrier is a one that cannot be readily removed. The EPA accepts that, while a shielded door loses its shielding capacity by simply opening the door, the door is fixed as it cannot be readily removed. There is no regulatory requirement governing the height or width of a protective barrier. But the barrier must be of sufficient dimensions so that a person cannot be exposed to the primary beam within 5 m of the X-ray tube. Regulations 91(11) and 91(12). 21
26 3 References Compliance Statement (Form 91A) Dental X-ray apparatus used for plain radiography, Environment Protection Authority. EPA Guidelines Program requirements for compliance testing of diagnostic X-ray apparatus, Environment Protection Authority. EPA Guideline Personal radiation monitoring devices, Environment Protection Authority, Safety Signs for the Occupational Environment, Australian Standard < 22
27 Annex 1 Radiation signs Figure 2 Radiation Produced When Energised Figure 3 Radiation symbol Figure 4 Radiation symbol plus radiation warning 23
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION
Elements of required physical infrastructures: space, schielding, and patient flow..
Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto
portable x-ray survey report
department of HealtH and Human services centers for medicare & medicaid services form approved omb no. 0938-0027 portable x-ray survey report provider number H1 date surveyed (1) initial (2) survey W H2
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)
GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
Agency for Health Care Administration
Page 1 of 20 FED - H0000 - INITIAL COMMENTS Title INITIAL COMMENTS CFR Type Memo Tag FED - H0005 - COMPLIANCE WITH FED/STATE/LOCAL LAWS Title COMPLIANCE WITH FED/STATE/LOCAL LAWS CFR 486.100 Type Condition
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.
X-Mind. Instinct for perfection
X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in
CODE OF PRACTICE & SAFETY GUIDE
CODE OF PRACTICE & SAFETY GUIDE R a d i a t i o n P r o t e c t i o n i n D e n t i s t r y RADIATION PROTECTION SERIES Radiation The Radiation is published by the Australian Radiation and Nuclear Safety
Radiation Safety Characteristics of the NOMAD Portable X-ray System
Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality
Radiation Safety Committee College of William and Mary
Radiation Safety Committee College of William and Mary Policy for the operation of Analytical X-ray Equipment The following rules govern the use of analytical X-ray equipment at the College of William
X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581
X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers
Safety and Radiation Protection Office Working with X-Ray Equipment
Safety and Radiation Protection Office Working with X-Ray Equipment Radiation Safety Rules May 2016 Radiation Safety Rules X-Ray Equipment General Rules for the protection of persons exposed to ionising
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
RADIATION SAFETY STANDARD
Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession
Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25
Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25 CHAPTER 227. RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT, X-RAY GAUGING EQUIPMENT, ELECTRON MICROSCOPES AND X-RAY CALIBRATION SYSTEMS Sec. 227.1.
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,
PowlVac Vacuum Integrity Tester
Instructions IB-60025B PowlVac Vacuum Integrity Tester POWELL ELECTRICAL MANUFACTURING COMPANY 8550 MOSLEY DRIVE HOUSTON, TEXAS 77075 USA PHONE (713) 944-6900 FAX (713) 947-4453 www.powellelectric.com
Required RS Training Info
C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness
X-ray (Radiography) - Chest
Scan for mobile link. X-ray (Radiography) - Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray produces images of
X-ray (Radiography), Chest
X-ray (Radiography), Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray makes images of the heart, lungs, airways,
HSE information sheet Industrial radiography - managing radiation risks Ionising Radiation Protection Series No. 1(rev 1)
HSE information sheet Industrial radiography - managing radiation risks Ionising Radiation Protection Series No. 1(rev 1) Introduction This information sheet is for clients and main or principal contractors
Radiation Safety Issues for Radiologic Technologists
Radiation Safety Issues for Radiologic Technologists Greg Sackett, M.S., CHP Medical Physicist Radiation Worker Risks? 1 Patient Risks? Acute Effects? Delayed Effects? Patient Questions? Radiation Dose
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
BS5839 Part 1:2002 Overview
BS5839 Part 1:2002 Overview 1 Contents Introduction So why is the current standard being revised? The brief for the new standard Changes to structure of current Standard Enhanced certification of Fire
USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY
GUIDE ST 5.3 / 4 MAY 2007 USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY 1 Ge n e r a l 3 2 Safety licences and exemption from licensing 3 3 Safety licence-exempt use of radiation 3
Standard the the Raising
Raising the Standard Introduction to BS 5839-1:2013 An Overview of the BS 5839 1:2013 for professionals who are responsible for the provision of fire detection and alarm systems for buildings Agenda Introduction
Integrated Management & Educational Consultancy
Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:[email protected]
Radiation Rule Guide. Veterinary Facilities. Hawaii Administrative Rules, Chapter 11-45, Radiation Control. Hawaii State Department of Health
Radiation Rule Guide Veterinary Facilities Hawaii Administrative Rules, Chapter 11-45, Radiation Control Hawaii State Department of Health MAY 2009 Radiation Rule Guide for Veterinary Facilities Hawaii
X-ray Safety Manual 2013
X-ray Safety Manual 2013 X-ray Machines and Other Ionizing Radiation Producing Devices The guidelines in this manual, when followed will ensure that appropriate protective and regulatory actions required
MALAYSIAN STANDARD CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) Copyright 2007
MALAYSIAN STANDARD MS 838:2007 CODE OF PRACTICE FOR RADIATION PROTECTION - MEDICAL X-RAY DIAGNOSIS (FIRST REVISION) ICS: 13.280 Descriptors: code of practice, medical diagnosis, radiation protection, x-ray
SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration
Implementation Page #: Page 1 of 10 Last Reviewed/Update 1. Purpose and Scope The purpose of this document is to describe the Medical Physics and Radiation Safety program at Boston University (BU) and
8001782 Owner s Manual
8001782 Digital Infrared Thermometer Owner s Manual Introduction This instrument is a portable, easy to use compact-size digital thermometer with laser sighting designed for one hand operation. The meter
X-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT
RP&AD /Remote QA/ 01 PROFORMA FOR ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT BHABHA ATOMIC RESEARCH CENTRE Radiological Physics & Advisory Division CT & CRS Building,
Digital Dental X-ray. Midmark Animal Health Products
Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't
X-RAY REGULATORY GUIDE
Minnesota Department of Health Radiation Control, X-ray Unit Protecting, maintaining and improving the health of all Minnesotans by promoting radiation safety through guidance and collaboration with the
Walking and Working Surface Checklist:
Walking and Working Surfaces Purpose: This information is presented to improve the level of safety in our operations, and to inform you that we will comply with the OSHA Standard 29 CFR 1910.23. Responsibility:
Regulatory Guide. Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0
Regulatory Guide Radiation Safety in Industrial Radiography (FANR-RG-019) Version 0 Federal Authority for Nuclear Regulation (FANR) P.O. Box 112021, Abu Dhabi, United Arab Emirates [email protected]
PART H RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT. Definitions. As used in this Part, the following definitions apply:
PART H RADIATION SAFETY REQUIREMENTS FOR ANALYTICAL X-RAY EQUIPMENT Sec. H.1 Purpose and Scope. This Part provides special requirements for analytical x-ray equipment. The requirements of this Part are
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
HDA -----------------------------------------------------------
----------------------------------------------------------- Lightstrips ---------------------------------------------------------------------- Ceiling system for heating and cooling LIGHTSTRIPS LIGHTSTRIPS
GUIDELINES. for oversize and overmass vehicles and loads MAY 2006. Government of South Australia. Department for Transport, Energy and Infrastructure
ESCORTING GUIDELINES ESCORTING GUIDELINES SCORTING GUIDELINES ESCORTING GUIDELINES ESCORTING ESCORTING GUIDELINES for oversize and overmass vehicles and loads Government of South Australia Department for
Technical Notes NDT 001 Specific Requirements for Non- Destructive Testing Laboratories
ACCREDITATION SCHEME FOR LABORATORIES Technical Notes NDT 001 Specific Requirements for Non- Destructive Testing Laboratories Technical Notes NDT 001, Apr 14 The SAC Accreditation Programme is managed
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric
In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the
In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the discussion on the major components of the x-ray machine. Today
QUALITY ASSURANCE IN DENTAL RADIOGRAPHY. Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000
QUALITY ASSURANCE IN DENTAL RADIOGRAPHY Part of the Ionising Radiation Regulations Section 4 (3) IR(ME)R 2000 Quality Assurance (QA) Ensure consistently adequate diagnostic information whilst keeping doses
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
The effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.
RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer
PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS
Sec. B.1 - B.4 SSRCR Volume I - February 2009 PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS Sec. B.1 - Purpose and Scope. a. This
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
Checklist for submission of application form for Registration
Checklist for submission of application form for Registration *Incomplete submission may cause delay in processing of the application. Name of the Hospital/Institution: City State Sr.No. Checkpoints Status
LASER SAFETY PROGRAM FOR CANISIUS COLLEGE (Department of Chemistry and Biochemistry) Effective: October 1, 2007 Revised: June 3, 2010
LASER SAFETY PROGRAM FOR CANISIUS COLLEGE (Department of Chemistry and Biochemistry) Effective: October 1, 2007 Revised: June 3, 2010 1. BACKGROUND Procedures and guidelines for working with lasers can
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
Application Note. ipix A Gamma imager to support various applications
Application Note ipix A Gamma imager to support various applications Introduction ipix is a unique gamma imager that quickly locates low level radioactive sources from a distance and estimates the dose
Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards
Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards Presented: August 3, 2011 Daniel Kassiday United States Food and Drug Administration Center for Devices and Radiological
LASER SAFETY MANUAL UNIVERSITY OF KENTUCKY POLICIES AND PROCEDURES FOR LASER USERS REVISED EDITION 2012 ISSUED BY
LASER SAFETY MANUAL UNIVERSITY OF KENTUCKY POLICIES AND PROCEDURES FOR LASER USERS REVISED EDITION 2012 ISSUED BY UNIVERSITY OF KENTUCKY RADIATION SAFETY OFFICE TABLE OF CONTENTS POLICY AND PURPOSE 2 LASER
Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral
Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come
Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology
Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment
FIRE ALARM SYSTEM TECHNICAL SPECIFICATIONS Page 1 of 10
TECHNICAL SPECIFICATIONS Page 1 of 10 FIRE DETECTION AND ALARM SYSTEM Scope Furnish a complete 24V DC conventional, electrically supervised, zone annunciated, fire detection and alarm system as specified
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
Production of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
County of Riverside Building and Safety Department
County of Riverside Building and Safety Department Mike Lara Director Photovoltaic Permitting Guidelines The information provided in this document is general and is intended only as a guide. Each project
Bold items are particular to the City of Euless
EULESS FIRE DEPARTMENT FIRE MARSHAL S OFFICE INFORMATION LINE: Revised 2/12 Fire Chief Wes Rhodes Fire Marshal Paul Smith EFD-FMO 3-2 2009 International Fire & Building Code as Amended NFPA Standards Adopted
P R E A M B L E. Facilitated workshop problems for class discussion (1.5 hours)
INSURANCE SCAM OPTICS - LABORATORY INVESTIGATION P R E A M B L E The original form of the problem is an Experimental Group Research Project, undertaken by students organised into small groups working as
90 Degree PIR Wireless Security Sensor. Installation Instructions 5420/90F
90 Degree PIR Wireless Security Sensor Installation Instructions 5420/90F Copyright Clipsal Australia P/L 2007. All rights reserved. Clipsal is a registered trademark of Clipsal Australia Pty Ltd. The
City of Riverside Building & Safety Division Phone: (951) 826-5697 www.riversideca.gov
City of Riverside Building & Safety Division Phone: (951) 826-5697 www.riversideca.gov PHOTOVOLTAIC PERMITTING GUIDELINES The information provided in this document is general and is intended only as a
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
FAQs-Main switchboard design criteria
FAQs-Main switchboard design criteria Q: What is the Australian standard for main switchboards? Current Australian standard is AS/NZS 3439.1: 2002 originating from IEC 60439. The new series of standard
Personal Dosimetry in Dental Radiology
Personal Dosimetry in Dental Radiology Position Statement Personal Dosimetry in Dental Radiology Radiological Protection Institute of Ireland May 2011 3 Clonskeagh Square, Dublin 14 Background The RPII
Wireless Home Security Alarm System AM 500
Wireless Home Security Alarm System AM 500 12 MONTH GUARANTEE Installation & Operating Instructions INTRODUCTION The AM500 is a simple self-contained alarm system. It protects the home by sounding a siren
Radiation safety in dental radiography
Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum
INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS. Published: April 1999
INTERNATIONAL HOCKEY FEDERATION PERFORMANCE REQUIREMENTS AND TEST PROCEDURES FOR HOCKEY BALLS Published: April 1999 CONTENTS 1 Introduction 2 Test Procedures General 3 Standards and Specific Test Procedures
Renewable Energy Test Station (RETS) TEST PROCEDURES FOR SOLAR TUKI
Renewable Energy Test Station (RETS) TEST PROCEDURES FOR SOLAR TUKI March 2007 A. Test Procedures for Solar Tuki Lamp S. No. Test Parameters Technical Requirements Instruments Required Test Methods A.
Material test chamber with mechanical convection
FP series 400 Material test chambers Material test chamber with mechanical convection A BINDER material test chamber with mechanical convection of the FP series provides reliably short drying times and
NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF DENTAL RADIOLOGIC TECHNOLOGY
NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF DENTAL RADIOLOGIC TECHNOLOGY This document contains the Board s accreditation standards for schools of
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the
Staff Doses & Practical Radiation Protection in DEXA
Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /
Fire Safety Requirements for Child Care Centre
Fire Safety Requirements for Child Care Centre A. Restrictions on Location Child care centre shall not be located : - (a) (b) in any premises designed for residential use; in any industrial building, warehouses,
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
SWIMMING POOL, SPA, & HOT TUB GUIDELINES
SWIMMING POOL, SPA, & HOT TUB GUIDELINES Pool permit cost is based on the total value of the project. See Permit Fee Schedule Table 1A. A. Adopted construction codes and installation requirements 1. 2015
VCE VET ENGINEERING STUDIES
Victorian Certificate of Education 2013 SUPERVISOR TO ATTACH PROCESSING LABEL HERE STUDENT NUMBER Letter Figures Words VCE VET ENGINEERING STUDIES Written examination Section Wednesday 20 November 2013
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
1.1.2 Polypropylene The polypropylene must be a white opaque film, 1 2 mil thick. Biaxial orientation is preferred.
Library of Congress Preservation Directorate Specification Number 700 704 09 Specifications for Pressure Sensitive Adhesive Labels For Application to Single Paper Sheets and Text Pages of Bound Books For
Page: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
INTEL S SAFETY WARNING LABEL REQUIREMENTS
INTEL S SAFETY WARNING LABEL REQUIREMENTS INTRODUCTION: Direction is provided on Intel s Safety Warning Label requirements to ensure suppliers provide equipment with hazards identified and the safety precautions
Frequently Asked Questions: The following regulations are provided to answer some of the many questions that are asked of this office.
Steven L. Beshear Governor CABINET FOR HEALTH AND FAMILY SERVICES DEPARTMENT FOR PUBLIC HEALTH Division of Public Health Protection and Safety Radiation Health Branch 275 East Main Street, HS 1C-A Frankfort,