User Guide. RDO Optical Dissolved Oxygen Sensor



Similar documents
User Guide Dissolved Oxygen Probes

DISSOLVED OXYGEN PROBE

Milwaukee USER MANUAL

Direct Measurement Method Method to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

HOBO Dissolved Oxygen Logger (U26-001) Manual

Orion Star Series Meter

STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance

Dissolved Oxygen Sensors

Handbook. Thermo Scientific. ph Electrode Selection Handbook

Measurement by Ion Selective Electrodes

Nitrogen, Ammonia. Known Addition ISE Method 1 Method Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL

6 H2O + 6 CO 2 (g) + energy

Professional Plus. Calibration Tips

SINGLE RANGE DS METERS Operation Instructions

User Guide. Thermo Scientific Orion ROSS Ultra, ROSS Ultra Triode, ROSS and PerpHecT ROSS Electrodes

USEPA Electrode Method Method 8156

Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter

AIRMoN, ph and Conductivity Field Measurement

Trace Dissolved Oxygen Sensor

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes

Measuring the Conductivity of Pure Water Using USP<645>

ph Measurements of Common Substances

PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN In Use Date: 2/1/2011 Discontinued Date:

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

User Manual. Thermo Scientific Orion VERSA STAR

Appendix C. Vernier Tutorial

Care and use guide. Contents. a handy solution made easy. page

Thermo Scientific Rotor Care Guide. Protect your rotor investment

Thermo Scientific Orion 2111XP Sodium Analyzer. User Guide

Care and use guide. Contents. a handy solution made easy. page

Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3

Dissolved Oxygen Probe (Order Code DO-BTA or DO-DIN)

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

Orion High-Level Chloride Monitor

YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen

Model 1811EL Low-Level Sodium Monitor

Care and use guide. Contents. page

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

Instruction Manual. Image of SP-3015 & SP Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

Quick Start Guide See Inside for Use and Safety Information

Winemaker s Guide to Chemistry Analysis

Neutralizing an Acid and a Base

Portable Waterproof ph/dissolved Oxygen Meter

For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN or ).

OxyGuard. Dissolved Oxygen Probe

Why Expanded Beam types are essential in military and other harsh environment applications

PATIENT GUIDE. Care and Maintenance Drainage Frequency: Max. Drainage Volume: Dressing Option: Clinician s Signature: Every drainage Weekly

Reasons for reissue are provided in Section 6, REVISION SUMMARY.

DISSOLVED OXYGEN METER Model : DO-5509

KODAK Developer System Cleaner and Neutralizer

MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL

PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION

UltraBasic Benchtop ph Meter Operation Manual

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

Operations Manual. YSI 550A Dissolved Oxygen Instrument

Global Water Instrumentation, Inc.

Household Acids and Bases

Instruction Manual Temp 4/5/6 Temperature Meter Series

ExStik TM ph Waterproof Meters

PART 1 - INTRODUCTION...

How to prepare the Green Cleaning Recipes for a Healthy Home pages:

& CLEANING GUIDE MAINTAIN BEAUTY & PERFORMANCE

Installation Instructions

Char-Lynn Hydraulic Motor. Repair Information Series. October, 1997

Chlorine, Total. DPD Method 1 Method to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information

EZ-PAGE Electrophoresis System USER MANUAL

Thermo Scientific Rotor Care Guide. Protect your rotor investment

2006 JUDY SERVICE GUIDE

CONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor

Chloride and Salinity

HP Scanjet G4000 series. User Guide

STANDARD OPERATING PROCEDURE

Matte Finish Paint Owner s Guide. Matte Finish Paint Owner s Guide. Matte Finish Paint Owner s Guide 3. NP150-V2013 (Rev 06/11/12) Printing xx/xx/12

SenTix 950 SenTix 980

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

Sip/Puff Switch User Guide

Bluelab Soil ph Meter Instruction Manual

MZ Series Overview. Introduction. (MZ 220 Illustrated)

Ignition Coil Sealing Instructions Revision 1.0

Toroidal Conductivity Sensor

Refer to page 6 to page 19 for detailed instructions.

BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets

Instruction Manual HI Refractometer for Refractive Index and Brix Measurements

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

Operator Quick Guide EC SENSOR

For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at

Eutech 2700 Series. Bench Meters. ph ORP ION Conductivity TDS Salinity Resistivity Dissolved Oxygen Temperature

AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance

Owner s Guide Brugervejledning Bedienungsanleitung Guía de usuario Omistajan opas Notice d utilisation Felhasználói útmutató Manuale di istruzioni

The table below lists the symbols used on the Clamp and/or in this manual. Important Information. See manual.

Coating and Extraction of Honeycomb Denuders

ph Measurement and Control

KUSTER ELECTRONIC GAUGES

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE

ClikSTAR - Important facts about your new insulin delivery device.

Transcription:

User Guide RDO Optical Dissolved Oxygen Sensor

ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion, phisa, phuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher. 1-888-pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, phuture MMS, phuture Pentrode, phuture Quatrode, phuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable ph, Universal Access are trademarks of Thermo Fisher. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher. PerpHecT meters are protected by U.S. patent 6,168,707. PerpHecT ROSS are protected by U.S. patent 6,168,707. ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results solutions are protected by US Patent 5,830,338. ROSS Ultra electrodes are protected by US patents 6,793,787. Orion ORP Standard is protected by US Patent 6,350,367. Orion NoCal electrodes are protected by US Patent 7,276,142. 2008 Thermo Fisher Scientific Inc. All rights reserved. RDO is a registered trademark of In-Situ Inc., Fort Collins, CO USA. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice. This publication supersedes all previous publications on this subject.

Table of Contents Introduction..................................... 1 Required Equipment.............................. 2 Important Notes Read Before Assembling or Operating the Equipment........................ 2 RDO Sensor Preparation.......................... 3 RDO Sensor Calibration........................... 4 Calibration Sleeve Preparation for Water-Saturated Air Calibration.......................... 4 Zero Oxygen Solution for Zero Point Calibration............. 5 Stainless Steel Protective Guard................... 6 RDO Sensor Measurements........................ 7 RDO Sensor Storage.............................. 7 RDO Sensor Maintenance......................... 8 Cleaning the RDO Cap................................ 8 Cleaning the RDO Sensor Body..........................8 Cleaning the Optical Window........................... 9 Yearly Replacement of the RDO Cap..................... 9 Troubleshooting................................. 10 Troubleshooting Chart................................ 11 Interfering Substances........................... 12 Substances That Damage the RDO Sensor............... 12 Assistance..................................... 12 Warranty....................................... 12 WEEE Compliance............................... 13 Ordering Information............................ 13 Specifications.................................. 14 RDO Optical Dissolved Oxygen Sensor User Guide

II RDO Optical Dissolved Oxygen Sensor User Guide

Introduction This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion RDO optical dissolved oxygen sensors. These dissolved oxygen sensors are compatible with a wide range of applications including rugged field applications, water and wastewater analysis, beer and wine testing and BOD (biochemical oxygen demand) analysis. All of the RDO sensors feature automatic temperature compensation and convenient press-on RDO caps. Orion Star RDO Optical Dissolved Oxygen Sensors The Orion Star RDO optical dissolved oxygen sensors have a 9 pin minidin connector that ensures a waterproof and secure connection when used with the Orion Star Plus RDO optical dissolved oxygen meters. Figure 1 RDO Optical Dissolved Oxygen Sensor Temperature Sensor O-rings Spring Contacts Thread Ring Optical Window RDO Cap RDO Optical Dissolved Oxygen Sensor User Guide 1

Required Equipment 1. Thermo Scientific Orion Star Plus RDO optical dissolved oxygen meter. 2. Thermo Scientific Orion RDO optical dissolved oxygen sensor. 3. RDO cap, Cat. No. 087001. 4. Calibration sleeve, Cat. No. 087003. 5. Stainless steel protective sensor guard (optional), Cat. No. 087002. 6. Magnetic stirrer or stirrer probe, Cat. No. 096019. The stirrer probe can be used with 3-Star, 4-Star and 5-Star Plus benchtop RDO meters. 7. Distilled or deionized water. Important Notes Read Before Assembling or Operating the Equipment Note: Do not touch the bottom of the RDO cap, optical window or spring contacts on the RDO sensor. Note: The RDO sensor has an internal clock that counts down the 365 day lifespan of a new RDO cap. The countdown begins when the RDO cap is installed on the RDO sensor, the sensor is connected to the meter and the first measurement is taken. The 365 day countdown cannot be reset or changed once the first measurement is taken. 2 RDO Optical Dissolved Oxygen Sensor User Guide

RDO Sensor Preparation 1. Remove the protective shipping cap from the end of the RDO sensor 2. Lubricate the two O-rings on the RDO sensor with a thin film of the silicone grease that is supplied with the sensor. 3. Align the arrow on the RDO cap with the arrow on the RDO sensor. See Figure 2. Figure 2 4. Slide and press the RDO cap onto the RDO sensor until the cap is flush with the sensor body. See Figure 3. Note: Do not twist the RDO cap. Once the cap is installed, the cap should not be removed from the RDO sensor until a new cap needs to be installed or troubleshooting needs to be performed. Figure 3 5. Connect the RDO sensor to the 9 pin minidin input on the meter. See Figure 4. Figure 4 Note: An Err 880 message on the meter indicates that the RDO sensor is not fully connected to the meter. If this error appears, reconnect the sensor by firmly pushing the connector into the 9 pin minidin meter input. Note: An Err 882 message on the meter indicates that the RDO cap is missing or not properly installed on the sensor. RDO Optical Dissolved Oxygen Sensor User Guide 3 Temperature Sensor Thread Cap

RDO Sensor Calibration Calibration of an RDO optical dissolved oxygen sensor can be done quickly and conveniently by performing a water-saturated air calibration. Under equilibrium conditions, the partial pressure of oxygen in air-saturated water is equal to the partial pressure of oxygen in water-saturated air. This means that a sensor calibrated in water-saturated air will correctly read the partial pressure of oxygen in water samples. When measuring low concentration samples (less than 1 ppm), a second calibration with a zero oxygen standard is recommended. Calibration Sleeve Preparation for Water-Saturated Air Calibration 1. Remove the cap from the calibration sleeve and remove the sponge from the cap. 2. Saturate the sponge with distilled water and squeeze the excess water out of the sponge. 3. Reassemble the calibration sleeve. 4. Unscrew and remove the thread ring from the RDO sensor. See Figure 5. Slide the calibration sleeve over the front of the RDO sensor and screw the sleeve onto the sensor. 5. Wait at least five minutes for the RDO sensor to equilibrate. 6. Follow the instructions in the meter user guide for performing a water-saturated air calibration. 4 RDO Optical Dissolved Oxygen Sensor User Guide

Zero Oxygen Solution for Zero Point Calibration 1. Perform a water-saturated air calibration. 2. Prepare a sodium sulfite solution by dissolving about 15.0 g of Na 2 SO 3 in about 250 ml of distilled water. Transfer the solution to a BOD bottle or flask. Note: A small amount of cobalt salt can be added to the sodium sulfite solution. The cobalt salt will act as an indicator and change color when the sodium sulfite solution no longer has a zero oxygen content. 3. Add a stir bar to the bottle or flask. Insert the RDO sensor into the bottle or flask and use parafilm to seal the open area between the bottle or flask and the sensor. 4. Place the bottle or flask on a magnetic stir plate and gently stir the solution. 5. Wait at least five minutes for the RDO sensor to equilibrate. 6. Follow the instructions in the meter user guide for performing a zero point calibration. 7. Thoroughly rinse the RDO sensor under running water and blot it dry with a lint-free tissue. Note: If the RDO sensor is sluggish or inaccurate after a zero point calibration, not all of the sodium sulfite was removed from the sensor. A very thorough soaking and rinsing of the RDO sensor in distilled water is required to remove all of the sodium sulfite solution and restore the sensor performance. Soak the RDO sensor in distilled water for 30 minutes, blot it dry with a lint-free tissue, use a wash bottle to rinse the sensor with distilled water, soak the sensor for another 30 minutes in fresh distilled water and blot it dry with a lint-free tissue. RDO Optical Dissolved Oxygen Sensor User Guide

Stainless Steel Protective Guard The RDO optical dissolved oxygen sensor can be used with the stainless steel protective guard, Cat. No. 087002. The guard protects the RDO sensor from damage and adds weight to the sensor for low depth measuring applications. Installation of the Protective Guard 1. Unscrew and remove the thread ring from the RDO sensor. See Figure 5. Figure 5 Thread Ring 2. Slide the protective guard over the front of the RDO sensor and screw the guard onto the sensor. See Figure 6. Figure 6 RDO Sensor with Protective Guard Thread Cap Temperature Sensor O-rings Spring Contacts 6 RDO Optical Dissolved Window Oxygen Sensor User Guide RDO Cap

RDO Sensor Measurements For accurate measurements, the RDO sensor must be immersed in solution until the temperature sensor is covered by at least a ½ inch of the solution. If any salt or mineral deposits are observed on the RDO sensor, immediately rinse them off with distilled or deionized water. 1. Rinse the RDO sensor with distilled water, blot it dry with a lint-free tissue and insert it into the sample. 2. Follow the instructions in the meter user guide for performing a measurement. 3. Remove the RDO sensor from the sample, rinse it with distilled water, blot it dry with a lint-free tissue, insert the sensor into the next sample and repeat step 2 or refer to the RDO Sensor Storage section for the recommended storage procedures. RDO Sensor Storage Short Term Storage Between measurements and overnight, leave the RDO cap on the RDO sensor and keep the sensor in the calibration sleeve, making sure to wet the sponge with distilled water, or in a beaker with distilled water. Keep the RDO sensor away from direct sunlight during storage. Long Term Storage Leave the RDO cap on the RDO sensor and keep the sensor in the calibration sleeve, making sure to wet the sponge with distilled water, or in a beaker with distilled water. Keep the RDO sensor away from direct sunlight during storage. RDO Optical Dissolved Oxygen Sensor User Guide 7

RDO Sensor Maintenance Figure 7 RDO Sensor and RDO Cap Temperature Sensor O-rings Spring Contacts Optical Window RDO Cap Cleaning the RDO Cap 1. Make sure that the RDO cap is properly installed on the RDO sensor and rinse the cap and temperature sections of the sensor with distilled water from a squirt bottle or spray bottle and blot dry with a lint-free soft cloth. 2. Inspect the RDO cap for scratches or discoloration and replace the cap if any flaws are observed. If algal growth is present on the RDO cap, gently clean it with a soft brush. Note: Do not use solvents or soaps to clean the RDO cap and do not rub the cap with abrasive material. Use of these materials will damage the RDO cap and void the warranty. Do not let salt or mineral deposits form on the RDO cap. To remove salt or mineral deposits from the cap, refer to step 2 in the Cleaning the RDO Sensor Body section. Cleaning the RDO Sensor Body 1. Make sure that the RDO cap is properly installed on the sensor and gently scrub the sensor body with a soft brush. 2. Calcareous fouling can normally be dissolved with household vinegar. Disconnect the RDO sensor from the meter and soak the assembled sensor, with the RDO cap installed on the sensor, in vinegar overnight. If the marine growth remains, use clean cotton swabs to gently wipe the growth off after it has been softened by soaking in vinegar. 3. After cleaning the RDO sensor, thoroughly rinse the sensor in distilled water and blot it dry with a lint-free tissue before using or storing it. After cleaning the RDO sensor, a watersaturated air calibration should be performed. 8 RDO Optical Dissolved Oxygen Sensor User Guide

Cleaning the Optical Window 1. Remove the RDO cap and gently wipe the optical window with a clean lens cloth. Note: Do not wet the optical window area with water or any other solution. Yearly Replacement of the RDO Cap The RDO cap must be replaced every 365 days. The meter will display an Err 881 message when the RDO cap needs to be replaced. To view the days remaining before the RDO cap must be replaced, refer to the meter user guide. 1. Turn the power to the meter off. 2. Remove the old RDO cap from the RDO sensor. Pull the cap straight off of the sensor and do not twist the cap. 3. Remove the existing two O-rings from the RDO sensor. 4. Remove any moisture from the RDO sensor body with a lint-free tissue. Do not touch the optical window or spring contacts with anything other than a lint-free tissue. Make sure that there is no moisture in the O-ring grooves. 5. Apply a thin film of silicone grease around the O-ring grooves and install the two new O-rings on the sensor. Only use the silicone grease and O-rings that are supplied with the RDO cap. 6. Lubricate the two O-rings on the RDO sensor with a thin film of silicone grease. Do not allow any silicone grease to touch the optical window or spring contacts. 7. Examine the optical window for scratches or dirt and clean it if necessary. To clean the optical window, gently wipe it with a clean lens cloth. 8. Install a new RDO cap onto the RDO sensor by aligning the arrow on the cap with the arrow on the sensor and pressing the cap onto the sensor until the cap is flush with the sensor body. Do not twist the RDO cap. Make sure that the O-rings are not pinched or rolled between the cap and the sensor. 9. Turn the power to the meter on. 10. After replacing the RDO cap, a water-saturated air calibration should be performed. RDO Optical Dissolved Oxygen Sensor User Guide

Troubleshooting The most important principle in troubleshooting is to isolate the components of the system and check each in turn. The components of the system include the meter, RDO sensor, sample and technique. Meter The meter is the easiest component to eliminate as a possible cause of error. Thermo Scientific Orion meters include an instrument checkout procedure in the meter user guide. Consult the meter user guide for complete instructions and verify that the meter operates correctly and is stable in all steps. RDO Sensor 1. Rinse the RDO sensor thoroughly with distilled water, blot it dry with a lint-free tissue and inspect the RDO cap for scratches or discoloration. 2. Remove the RDO cap from the RDO sensor and make sure that there is no water inside the cap, the optical window is clean and clear, the O-rings are intact and have a thin coating of silicone grease and the spring contacts are clean and undamaged. 3. If readings continue to be erratic and unstable, the RDO cap may need to be replaced. Refer to the RDO Sensor Maintenance section. Sample If the RDO sensor works properly in standards but not in the sample, look for possible interferences or substances in the sample that could alter the sensor response or physically damage the sensor. If possible, determine the composition of the sample and check for issues. Technique Check if the method of analysis is compatible with your sample. 10 RDO Optical Dissolved Oxygen Sensor User Guide

Troubleshooting Chart Issue Unable to calibrate the RDO sensor Dissolved oxygen measurements are unstable Dissolved oxygen measurement is too low Wrong temperature displayed Bad zero point calibration Bad reading after zero point calibration Err 880 error message Err 881 error message Err 882 error message Recommended Action Verify the calibration setup and procedure. Make sure that no water droplets are on the surface of the RDO cap. Perform RDO sensor maintenance. Refer to the Troubleshooting and RDO Sensor Maintenance sections. Measurements may take longer if the solution temperature is unstable. Perform RDO sensor maintenance. Refer to the Troubleshooting and RDO Sensor Maintenance sections. Interfering substance may be present. Salt may be present in the sample. Set the salinity factor in the meter. Perform RDO sensor maintenance. Refer to the Troubleshooting and RDO Sensor Maintenance sections. Verify that the temperature sensor on the RDO sensor is immersed in the solution. Contact Technical Support. The RDO sensor must be in an oxygenfree solution for at least five minutes. Make sure that the open area between the sensor and bottle is covered with parafilm. Soak the RDO sensor in distilled water for 30 minutes, use a wash bottle to thoroughly rinse the sensor with distilled water and then soak the sensor for another 30 minutes in fresh distilled water. The RDO sensor is not fully connected to the meter. Reconnect the RDO sensor. The RDO cap has expired. Install a new RDO cap. Verify that the RDO cap is fully seated on the sensor body. Install a new RDO cap. RDO Optical Dissolved Oxygen Sensor User Guide 11

Interfering Substances Alcohols greater than 5%, hydrogen peroxide (H 2 O 2 ) greater than 3%, sodium hypochlorite (NaClO, commercial bleach) greater than 3%, gaseous sulfur dioxide (SO 2 ) and gaseous chlorine (Cl 2 ) may interfere with the dissolved oxygen measurements. If the measurements are unstable and other troubleshooting procedures do not correct the problem, there may be interfering substances present in the solution. Carbon dioxide (CO 2 ), ammonia (NH 3 ), ph, any ionic species like sulfide (S 2- ), sulfate (SO 4-2 ), chloride (Cl - ) and hydrogen sulfide (HS - ) do not interfere with the dissolved oxygen measurements. Substances That Damage the RDO Sensor Do not use the RDO sensor in solutions that contain organic solvents, such as acetone, chloroform or methylene chloride. Assistance After troubleshooting all components of your measurement system, contact Technical Support. Within the United States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit www.thermo.com/contactwater. For the latest application and technical resources for Thermo Scientific Orion products, visit www.thermo.com/waterapps. Warranty For the most current warranty information, visit www.thermo.com/water. 12 RDO Optical Dissolved Oxygen Sensor User Guide

WEEE Compliance This product is required to comply with the European Union s Waste Electrical & Electronic Equipment (WEEE) Directive 2002/96/EC. It is marked with the symbol on the left. We have contracted with one or more recycling/disposal companies in each EU Member State and this product should be disposed of or recycled through them. Further information on compliance with these directives, the recyclers in your country and information on Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS directive are available at www.thermo.com/weeerohs. Ordering Information Cat. No. Description 087010MD RDO optical dissolved oxygen sensor with 3 meter cable, RDO cap, calibration sleeve and stainless steel protective guard 087020MD RDO optical dissolved oxygen sensor with 6 meter cable, RDO cap, calibration sleeve and stainless steel protective guard 087030MD RDO optical dissolved oxygen sensor with 10 meter cable, RDO cap, calibration sleeve and stainless steel protective guard 087050MD RDO optical dissolved oxygen sensor with 15 meter cable, RDO cap, calibration sleeve and stainless steel protective guard 087100MD RDO optical dissolved oxygen sensor with 30 meter cable, RDO cap, calibration sleeve and stainless steel protective guard 087001 Replacement RDO cap 087002 Stainless steel protective guard 087003 Calibration sleeve RDO Optical Dissolved Oxygen Sensor User Guide 13

Specifications Dissolved Oxygen Range Dissolved Oxygen Accuracy Dissolved Oxygen Resolution 0 to 20 mg/l 0 to 200 % ± 0.1 mg/l up to 8 mg/l ± 0.2 mg/l from 8 to 20 mg/l 0.01 mg/l Dissolved Oxygen Response Time 90% of final value in 30 seconds, 95% of final value in 90 seconds Temperature Range 5 to 45 C Temperature Accuracy ± 0.3 C Minimum Immersion Depth Sensor Length (Tip) Sensor Diameter (Tip) Total Sensor Length Total Sensor Diameter Cable Length RDO Cap Expected Lifespan 40 mm 70 mm 20 mm 190 mm 35 mm 087010MD 3 meters 087020MD 6 meters 087030MD 10 meters 087050MD 15 meters 087100MD 30 meters 1 year Specifications are subject to change without notice 14 RDO Optical Dissolved Oxygen Sensor User Guide

Thermo Fisher Scientific Environmental Instruments Water Analysis Instruments North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 Dom. Fax: 1-978-232-6015 Int l Fax: 978-232-6031 Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 BN Nijkerk, The Netherlands Tel: (31) 033-2463887 Fax: (31) 033-2460832 257262-001 Rev. A 09-08 Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore 139949 Tel: 65-6778-6876 Fax: 65-6773-0836 www.thermo.com/water