Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter
|
|
|
- Catherine Hodges
- 10 years ago
- Views:
Transcription
1 Instruction Manual HI 9146 Portable Waterproof Microprocessor Dissolved Oxygen Meter w w w. h a n n a i n s t. c o m
2 Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with the necessary information for a correct use of the instrument, as well as a precise idea of its versatility. If you need more technical information, do not hesitate to us at [email protected] This meter is in compliance with directives. TABLE OF CONTENTS Preliminary Examination... 3 General Description... 3 Functional Description - Probe... 4 Functional Description - Meter... 5 Specifications... 6 Probe Preparation... 7 Calibration... 8 Measurement Altitude Compensation Salinity Compensation Probe & Membrane Maintenance Battery Replacement Accessories Warranty CE Declaration of Conformity PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it to make sure that no damage has occurred during shipping. If there is any damage, notify your Dealer. The meter is supplied complete with: HI 76407/4F DO probe with 4 m cable and protective sleeve for HI HI 76407/10F DO probe with 10 m cable & protective sleeve for HI membranes (HI 76407A) with O-rings protective cap 30 ml electrolyte solution (HI 7041S) 4 x 1.5V AA batteries instruction manual rugged carrying case Note: Save all packing material until you are sure that the instrument functions correctly. Any defective items must be returned in the original packing with the supplied accessories. GENERAL DESCRIPTION The Hanna HI 9146 is a water-resistant, microprocessor-based, auto-calibrating Dissolved Oxygen meter with ATC. It has been developed for dissolved oxygen and temperature measurement in water and wastewater as well as other applications such as fish farming. Dissolved oxygen is indicated in hundredths of parts per million (ppm=mg/l) or in % of saturation. The temperature range is indicated in centigrade from 0 to 50 C with 0.1 C resolution. The ppm and the % saturation are both compensated for changes in solubility of oxygen in water and for permeability of the membrane as well as the temperature effect. 2 3
3 Salinity compensation in water allows determination of mg/l of dissolved oxygen in salty waters and the altitude compensation readjusts for the altitude variance. Four 1.5V AA alkaline batteries provide power and make the instrument completely portable. HI 9146 is also designed to be used with a battery recharger or a 12 Vdc power supply. A 12 Vdc input jack is incorporated into the housing. A microprocessor ensures an accurate and rapid calibration and measurement. The case rugged and water-resistant for maximum protection in the field as well as in the laboratory. The dissolved oxygen probe has a membrane covering the polarographic sensor elements and a built-in thermistor for temperature measurement and compensation. The thin permeable membrane isolates the sensor elements from the testing solution, but allows oxygen to enter. When a voltage is applied across the sensor, oxygen that passes through the membrane causes a current to flow from which the oxygen concentration is determined. FUNCTIONAL DESCRIPTION - PROBE 1. D.O. Probe 2. Protective Cap 3. Watertight Shielded Cable 4. Polypropylene Probe Body 5. Temperature Sensor 6. O-Ring Seal 7. AgCl Anode (sensor element) 8. Platinum Cathode (sensor element) 9. Oxygen Permeable PTFE Membrane 10. Membrane Protective Cap FUNCTIONAL DESCRIPTION - METER 1) Battery Compartment 2) Probe Connector 3) Liquid Crystal Display 4) ON/OFF key 5) CAL key (to enter/exit calibration mode) 6) FACTOR key (to select altitude F1 and salinity F2) 7) UP and DOWN arrow keys (to select F1 and F2 levels) 8) HOLD key (to freeze displayed value) 9) RANGE key (to select ppm or % of saturation) 10) Power socket for 12 Vdc adapter 19) LOW BAT indicator 20) SAMPLE indicator (to indicate measurement mode) 21) F1 indicator (altitude factor) 22) F2 indicator (salinity factor) 23) % or ppm mode indicator 24) Temperature and factor display 4 5
4 SPECIFICATIONS Range 0.00 to mg/l O to 300 %O to 50.0 C Resolution 0.01 mg/l O %O C Accuracy ±1.5% Full Scale mg/l O 2 (@25 C/77 F) ±1.5% Full Scale %O 2 ±0.5 C Typical EMC ±0.3 mg/l O 2 / ± %O 2 Deviation ±0.5 C Calibration Automatic in saturated air Temperature Automatic from 0 to 50 C Compensation (32 to 122 F) Altitude 0 to 4 km Compensation 100 m resolution Salinity 0 to 80 g/l Compensation 1 g/l resolution Environment 0 to 50 C (32 to 122 F) RH 100% Power supply 4x1.5V AA batteries; 200 hours of continuous use; auto-off after 4 hours. 12 Vdc adapter Dimensions 196 x 80 x 60 mm (7.7 x 3.1 x 2.4") Weight 500 g (1.1 lb.) PROBE PREPARATION All probes are shipped dry by Hanna Instruments. To hydrate the probe and prepare it for use, connect it to the meter and proceed as follows. 1 Remove the red and black plastic cap which is for shipping purposes and can be discarded. 2 Wet the sensor by soaking the bottom 2½ cm (1") of the probe in electrolyte (HI 7041S) for 5 minutes. 3 Take a membrane (HI 76407A supplied with the meter) and Make sure that the rubber O-ring sits properly inside the membrane cap. 4 Rinse the membrane with electrolyte while shaking it gently. Refill with clean electrolyte. SECOND FILL THIRD TAP THEN SCREW BACK ON 5 Gently tap the sides of the membrane cap with your finger tip to ensure that no air bubbles remain trapped. To avoid damaging to the membrane, do not tap the membrane directly on the bottom. 6 With the sensor facing down, screw the cap clockwise. Some electrolyte will overflow. When not in use, protect the membrane by placing the protective cap on it. Shipping cap black red FIRST INSERT O RING 6 7
5 CALIBRATION PROBE POLARIZATION The probe is under polarization with a fixed voltage of approximately 800 mv. Probe polarization is essential for stable measurements with the same recurring degree of accuracy. With the probe properly polarized, oxygen is continually "consumed" by passing through the sensitive diaphragm and dissolving in the electrolyte solution contained inside the probe. If this operation is interrupted, the electrolyte solution continues to be enriched with oxygen until it reaches an equilibrium with the surrounding solution. Whenever measurements are taken with a non-polarized probe, the oxygen level indicated is that of the test solution as well as any oxygen present in the electrolyte solution. This reading is obviously incorrect. The Hanna oxygen meter shown here automatically polarize the probe when they are switched on. Calibration is simple and is recommended every time the meter is switched on. Make sure the probe is ready for measurement (see page 5), i.e. the membrane is filled with electrolyte and probe is connected to the meter and properly polarized. Switch the meter on. "COND" appears on the display to inform you that ON/OFF the probe is in auto-conditioning (automatic polarization) mode. Once "COND" disappears the probe is polarized and instrument can be calibrated. For an accurate calibration, it is recommended to wait an additional 5 or 10 minutes to ensure optimum conditioning of the probe. Remove the protective cap. Press CAL. "----" appears on the larger LCD together with "CAL" to indicate that CAL the instrument is in calibration mode. The instrument will automatically standardize itself CAL to the actual saturation value. After approx. 1 minute it will show "100%" SAMPLE % on the LCD and a small C "SAMPLE" to indicate that the calibration is complete. Press FACTOR and ensure F1 and F2 are set to the appropriate altitude and salinity values (pages 10-11). Notes: The instrument must be calibrated whenever the probe, membrane or the electrolyte is changed. To exit the calibration mode during calibration, press CAL. The display may be switched from readings in % saturation to mg/ CAL L without recalibration, by simply pressing RANGE. RANGE 8 9
6 MEASUREMENT Make sure the meter has been calibrated and the protective cap has been removed. Immerse the tip of the probe in the sample to be tested and ensure that the temperature sensor is also immersed. To display values in % saturation, press RANGE. RANGE For accurate dissolved oxygen measurements, a water movement of at least 30 cm (12 )/sec is required. This is to ensure that oxygen depleted on the membrane surface is constantly replenished. A moving stream will provide adequate circulation. During field measurements, this condition may be obtained by manually stirring the probe. Accurate readings are not possible while the liquid is stationary. During laboratory measurements, the use of a magnetic stirrer to ensure a certain agitation of the fluid is recommended. This way, any errors due to the presence of air bubbles on the membrane surface are minimized. If the sample contains significant salinity or if the measurement is taken at a higher altitude than sea level, the read-out values must be corrected (by taking into account the lower degree of oxygen solubility in such conditions - see below). Set the altitude and/or the salinity before calibration and taking mg/l measurements. The meter will automatically compensate for these factors. % O 2 SATURATION READINGS The % O 2 reading provides the rate of oxygen saturation with reference to 100.0% at sea level. TEMPERATURE READINGS The lower part of the display SAMPLE % will show the measured C temperature in Celsius degrees. Allow the probe to reach thermal equilibrium with the sample before taking any measurement. The greater the difference between the ambient temperature and the temperature of the sample, the longer it will take the probe to acclimatized itself to the sample. For an accurate measurement, allow sufficient time for thermal equilibrium between the probe and the measurement sample (a few minutes if the temperature difference is several degrees). mg/l READINGS The mg/l readings make it possible to read the concentration of the dissolved oxygen directly in ppm
7 ALTITUDE COMPENSATION SALINITY COMPENSATION Press FACTOR and F1 will be displayed. Use the UP and the DOWN keys to set the altitude from 0 and km (4000 m), in steps of 0.1 km (100 m). SAMPLE FACTOR ppm F1 or Press FACTOR twice and F2 will be displayed. Press UP and DOWN to set the salinity between 0 and 80 g/l. Press FACTOR again to display the temperature. SAMPLE FACTOR ppm F2 or The table provides an idea of the error that can be introduced at different altitudes and the quantity to be subtracted to correct the reading. C Altitude above Sea Level (in meters) C Salinity (g/l) at Sea Level 0 g/l 10 g/l 20 g/l 30 g/l 35 g/l F
8 PROBE & MEMBRANE MAINTENANCE The oxygen probe body is made of reinforced polypropylene for maximum durability. A thermistor sensor measures temperature of the sample. It is recommended that the protective cap be always kept on the probe when the probe is not in use. To replace the membrane or refill it with electrolyte, proceed as follows: Remove the protective cap by gently twisting and pulling it off the probe (see fig. 1). Unscrew the membrane by turning it counterclockwise (see fig.2) Wet the sensor by soaking the bottom 2½ cm (1") of the probe in electrolyte (HI 7041S) for 5 minutes. Rinse a new membrane (HI 76407A) with electrolyte while shaking it gently. Refill with clean electrolyte. Gently tap the sides of the membrane with your finger tip to ensure that no air bubbles remain trapped. Do not directly tap the bottom as this may cause irreparable damage to the membrane. TWIST AND PULL fig. 1 Make sure that the rubber O-ring is seated properly inside the membrane cap. With the sensor facing down, screw the membrane cap clockwise. Some electrolyte will overflow. The Platinum cathode sensor (#8 in the Functional Description on page 2) should always be bright and untarnished. If it is tarnished or stained, which could be due to contact with certain gases or extended use with a loose or damaged membrane, the cathode sensor should be cleaned. Use a clean, lint-free cardboard or cloth. Rub the cathode very gently side to side 4-5 times. This will be enough to polish and remove any stains without damaging the platinum tip. Then, rinse the probe with deionized or distilled water. Install a new membrane and fill it with fresh electrolyte following the instructions above. Recalibrate the instrument. Important: In order to have accurate and stable measurements, it is important that the surface of the membrane is in perfect condition. This permeable membrane isolates the sensor elements from the environment but allows oxygen to enter. If any dirt is deposited on the membrane, rinse it carefully with distilled or deionized water. If any imperfections are observed, or any damage is evident (such as wrinkles, tears or holes), the membrane should be replaced. Make sure that the O-Ring is properly seated in the membrane cap. UNSCREW fig
9 BATTERY REPLACEMENT When the batteries are run LOBAT ppm SAMPLE down "LOBAT" is displayed on the Liquid Crystal Display. C This is to inform the user that the display will be shut-off after about 4 hours of use to prevent erroneous measurements due to low voltage. Battery replacement must only take place in a non-hazardous area using alkaline batteries. In order to replace run down batteries, simply remove the two screws on the rear cover of the instrument (#1 in the Functional Description on page 3) and replace the four 1.5V AA batteries with new ones, paying attention to the correct polarity. A 12 Vdc power source can also be used to power the unit (see accessories). Simply unscrew the protective cap on the top of the instrument (#10 page 3) and plug the power supply into the socket. Note: The instrument uses the following configuration. Use only the Hanna voltage adapters (such as HI or HI see below) with proper polarity configuration. The meter can however also run on other adapters. In this case, check the correct polarity of your adapter before connecting it to the meter. HI 7041S ACCESSORIES Electrolyte solution for D.O. probe, 30 ml HI 76407/4F D.O. probe with 4 m (13 ) cable and protective sleeve HI 76407/10F D.O. probe with 10 m (33 ) cable and protective sleeve HI 76407A/P D.O. membrane (5 pcs) HI HI Vac to 12 Vdc power adapter, US plug 230 Vac to 12 Vdc power adapter, European plug
10 WARRANTY CE DECLARATION OF CONFORMITY All Hanna meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to the instructions. Probes are warranted for six months. This warranty is limited to repair or replacement free of charge. Damages due to accidents, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact the dealer from whom you purchased the instrument. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charge for repair or replacement. If the instrument is to be returned to Hanna Instruments, please obtain a Return Goods Authorization from the Customer Service Department and then send it with shipment cost prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. Hanna Instruments reserves the right to modify design, construction and appearance of its products without advance notice. Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of this instrument in residential area could cause unacceptable interference to radio and TV equipment, requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instrument's EMC performance. To avoid electrical shock, do not use these instruments when voltage at the measurement surface exceeds 24VAC or 60VDC. To avoid damage or burns, do not perform any measurement in microwave ovens
11 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) Fax (03) China: Tel. (10) Fax (10) Egypt: Tel. & Fax (02) Germany: Tel. (07851) Fax (07851) Greece: Tel. (210) Fax (210) Indonesia: Tel. (21) Fax (21) Japan: Tel. (03) Fax (03) Korea: Tel. (02) Fax (02) Malaysia: Tel. (603) Fax (603) Singapore: Tel Fax South Africa: Tel. (011) Fax (011) Taiwan: Tel Fax Thailand: Tel Fax United Kingdom: Tel. (01525) Fax (01525) USA: Tel. (401) Fax (401) MAN9146R1 05/05 For contacts and complete list of Sales and Technical offices, please see
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
DISSOLVED OXYGEN METER Model : DO-5509
DISSOLVED OXYGEN METER Model : DO-5509 Your purchase of this DISSOLVED OXYGEN METER marks a step forward for you into the field of precision measurement. Although this DISSOLVED OXYGEN METER is a complex
User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600
User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously
DISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
Instruction Manual HI 96801 Refractometer for Sucrose Measurements
Instruction Manual HI 96801 Refractometer for Sucrose Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation
Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements
Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction
Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor
Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process
Thermistor Thermometer MODEL NO. 93210-00
Thermistor Thermometer MODEL NO. 93210-00 Cole-Parmer Instrument Co. 625 East Bunker Court Vernon Hills, Illinois U.S.A. 60061-1844 (847) 549-7600 (847) 247-2929 (Fax) 800-323-4340 www.coleparmer.com e-mail:
Portable Waterproof ph/dissolved Oxygen Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35632-Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ /
OxyGuard. Dissolved Oxygen Probe
OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard
ExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
User Guide. 9708 Dissolved Oxygen Probes
User Guide 9708 Dissolved Oxygen Probes Introduction The Thermo Scientific Orion 9708 dissolved oxygen probe simplifies measurements of dissolved oxygen, particularly Biochemical Oxygen Demand (BOD).
Operations Manual. YSI 550A Dissolved Oxygen Instrument
YSI 550A Dissolved Oxygen Instrument Operations Manual CONTENTS CONTENTS...iii GENERAL DESCRIPTION...1 WARRANTY...2 FEATURES OF THE YSI 550A...3 PRINCIPLES OF OPERATION...5 PREPARING THE PROBE...5 DISSOLVED
STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance
Page 1 of 9 KEY WORDS- Dissolved Oxygen, calibration, probe preparation, meter preparation APPROVALS APPROVED BY: Kean S. Go, DATE: +/?? APPROVED BY: A Pement DATE: a// A/? f Lisa Ross, EHAP Senior Scientist
Microwave Meter. Instruction Manual
Microwave Meter 840046 Instruction Manual Microwave Meter 840046 Copyright 2009 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA The contents of this manual may not be reproduced or transmitted
Global Water Instrumentation, Inc.
Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 Temperature Sensor: WQ101 ph Sensor: WQ201 Conductivity Sensor: WQ301
Instruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X361305 Rev. 0-11/05. Technology Made Easy...
Instruction Manual DO 110 Hand-held Dissolved Oxygen Meter with RS 232C Technology Made Easy... 68X361305 Rev. 0-11/05 Preface This manual serves to explain the use of the hand-held series meters. The
Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
Instruction Manual Temp 4/5/6 Temperature Meter Series
Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn
INSTRUCTION MANUAL. Handy Polaris
INSTRUCTION MANUAL Handy Polaris **This page is left blank** Table of Contents 1 General...4 2 Setup...5 2.1 Calibrate...5 2.2 Set Salinity...5 2.3 User Interface...5 2.3.1 Language...5 2.3.2 Temperature
Model MC-82 Dissolved Oxygen Temp. Meter Handbook Version : 1.0
Congratulations! You have purchased the latest in Handheld Dissolved Oxygen-Temperature instrumentation. We trust that your new MC-82 will give you many years of reliable service. The MC-82 is a breeze
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide
Care and use guide. Contents. page
Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4
OXIGENO DISUELTO. Dissolved Oxygen. Install & forget. Sistema autolimpiable. Autoclean system
Sistema autolimpiable Dissolved Oxygen Autoclean system Applications: - water treatment - activated sludge - de-nitrification - fish pond Install & forget For special applications, the D.O monitor can
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
SINGLE RANGE DS METERS Operation Instructions
SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax
Thermocouple Thermometer Models
Technology Made Easy... Thermocouple Thermometer Models 60010-00 (Type J) 60010-10 (Type K) 60010-20 (Type T) 68X309905 Rev 1 04/04 TABLE OF CONTENTS 1. INTRODUCTION 4 2. SAFETY PRECAUTIONS 6 3. SPECIFICATIONS
Galvanic Parts Per Million Dissolved Oxygen Systems
Galvanic Parts Per Million Dissolved Oxygen Systems Model 900 Portable s Model 9100 Model 9200 Model 9300 The Royce line of PPM level dissolved oxygen (DO) analyzers is the largest, most varied line in
mg/l YSI Model 85 Handheld Oxygen, Conductivity, Salinity, and Temperature System Operations Manual
mg/l C YSI Model 85 Handheld Oxygen, Conductivity, Salinity, and Temperature System Operations Manual CONTENTS SECTION 1 Introduction...1 SECTION 2 Preparing The Meter...3 2.1 Unpacking...3 2.2 Warranty
DO-6. Dissolved Oxygen Meter. Newport Electronics, Inc. Operator s Manual
Dissolved Oxygen Meter Operator s Manual Newport Electronics, Inc. Additional products from Newport Electronics, Inc. THE NEW STANDARD FOR QUALITY Clock/Timers Counters Flow Sensors Frequency Meters Multimeters
Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986
Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986 Fann Instrument Company Electrical Stability Tester Instruction Manual 2012 Fann Instrument
USEPA Electrode Method Method 8156
ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method
Light Meter (Foot Candles)
Light Meter (Foot Candles) 840021 Instruction Manual SPER SCIENTIFIC LTD. TABLE OF CONTENTS 1. INTRODUCTION... 2 2. PANEL DESCRIPTION... 3 3. MEASUREMENT PROCEDURES... 4 4. ZERO ADJUSTMENT & CALIBRATION...
Hygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
Global Water Instrumentation, Inc.
Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 6 Tipping Bucket: RG200 02/05/10 01-343 - 1 - Congratulations on your
Trace Dissolved Oxygen Sensor
Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION
Model ATD-36750 Refrigerant Gas Leak Detector. User Manual. Detects all CFC, HFC, HCFC, HFO Refrigerants including blends
Model ATD-36750 Refrigerant Gas Leak Detector Detects all CFC, HFC, HCFC, HFO Refrigerants including blends User Manual Design certified by ASCI and Intertek to meet SAE J2791, J2913 & EN14624 INTRODUCTION
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
CappController Pipette Controller. Operation Manual Version: 1.2. www.capp.dk
CappController Pipette Controller Operation Manual Version: 1.2 www.capp.dk skal vises her. Table of Content 1. Safety Precautions... 1 2. General Description... 2 3. Getting Started... 4 3.1. Recharging
Portable Waterproof Dissolved Oxygen Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35641-Series Portable Waterproof Dissolved Oxygen Meter WTERPROOF MES T Dissolved Oxygen/ / F Data Meter
Dissolved Oxygen Probe (Order Code DO-BTA or DO-DIN)
Dissolved Oxygen Probe (Order Code DO-BTA or DO-DIN) The Dissolved Oxygen Probe can be used to measure the concentration of dissolved oxygen in water samples tested in the field or in the laboratory. You
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
6 H2O + 6 CO 2 (g) + energy
AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features 2 Quick guide 2 To operate 3 Cleaning 4 Battery replacement 4 Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about
Portable digital microprocessor thermometers with interchangeable probes
JUMO GmbH & Co. KG Delivery address:mackenrodtstraße 14, 36039 Fulda, Germany Postal address: 36035 Fulda, Germany Phone: +49 661 6003-0 Fax: +49 661 6003-607 e-mail: [email protected] Internet: www.jumo.net
Model UT33A: OPERATING MANUAL Table of Contents (1)
Table of Contents (1) Title Overview Unpacking Inspection Safety Information Rules For Safe Operation International Electrical Symbols The Meter structure Functional Buttons LCD Display Measurement Operation
Installation and Operating Instructions (for chargers shown below)
Installation and Operating Instructions (for chargers shown below) For additional information please call our Technical Support Group 800.742.2740 PRO CHARGING SYSTEMS, LLC 1551 Heil Quaker Boulevard,
SenTix 950 SenTix 980
SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the
ISO2. Isolated dissolved oxygen meter and oxygen electrode INSTRUCTION MANUAL. www.wpiinc.com. Serial No. World Precision Instruments
Isolated dissolved oxygen meter and oxygen electrode www.wpiinc.com INSTRUCTION MANUAL Serial No. 070715 World Precision Instruments CONTENTS ABOUT THIS MANUAL... 1 INTRODUCTION... 1 Parts List... 2 Unpacking...
Duct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
Bluelab Soil ph Meter Instruction Manual
Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for
Wireless Home Security Alarm System AM 500
Wireless Home Security Alarm System AM 500 12 MONTH GUARANTEE Installation & Operating Instructions INTRODUCTION The AM500 is a simple self-contained alarm system. It protects the home by sounding a siren
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques Introduction The amount of dissolved oxygen in process water is continually gaining importance in many industries as a critical parameter
Weather Radio Alarm Clock
1200093 User s Guide Weather Radio Alarm Clock Thank you for purchasing your Weather Radio Alarm Clock from RadioShack. Please read this user s guide before installing, setting up, and using your new weather
MEMO ALARM CLOCK Operating Manual
MEMO ALARM CLOCK Operating Manual INTRODUCTION Congratulations on purchasing this state-of-the-art Alarm clock with radio controlled time. The operation of this product is simple and straightforward and
OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S
Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and
Cable Tester Pro ORDERCODE D1909
Cable Tester Pro ORDERCODE D1909 Congratulations! You have bought a great, innovative product from DAP Audio. You can rely on DAP Audio, for more excellent audio products. We design and manufacture professional
Soil Sensor Reader PRODUCT MANUAL
Soil Sensor Reader PRODUCT MANUAL Item # 6466 CONTENTS Basic Operation 3 Power Up/Down 3 Sensor Reader Modes 3 Taking Readings 5 Connecting Sensors 5 Calibrating The EC Sensor 6 Changing the Battery 7
Getting started with
PART NO. CMA113 MADE IN CHINA 1. Measuring CAT II 2. Max. voltage 250V ~ 3. Max. current 71 Amp Getting started with Electricity consumption monitoring single phase for homes and some smaller light commercial
Periodic Verification System Model 775PVS
Periodic Verification System Model 775PVS Instruction Manual Contents 1 Description... 1 775PVS System... 1 Model 775 Fieldmeter... 2 Model 775C Charger... 2 Model 775P Plate Assembly... 3 2 Operation...
Hand-held thermometer Model CTH7000
Calibration technology Hand-held thermometer Model CTH7000 WIKA data sheet CT 55.50 Applications Precision thermometer for very accurate temperature measurements in a range of -200... +962 C Reference
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
D7050/D7050TH. Installation Instructions. Multiplex Photoelectric Smoke Detectors
D7050/D7050TH EN Installation Instructions Multiplex Photoelectric Smoke Detectors D7050/D7050TH Installation Instructions.0 Description Notices These instructions cover the installation of the D7050/D7050TH
User's Guide. True RMS Industrial Multimeter
User's Guide 97650 True RMS Industrial Multimeter Ω C ã F ã 10A V µ 10A V ã ã ma A Introduction This meter measures AC/DC Voltage, AC/DC Current, Resistance, Capacitance, Frequency (electrical & electronic),
Refer to page 6 to page 19 for detailed instructions.
CODE:GZ0000297065 Refer to page 6 to page 19 for detailed instructions. Preface This manual describes the operation of the COMPACT CONDUCTIVITY METER, B-771. Be sure to read this manual before using the
el-720 Carbon Dioxide Gas Leak Detector User s Guide
el-720 Carbon Dioxide Gas Leak Detector User s Guide Congratulations on the purchase of the AccuTools el-720 Carbon Dioxide Gas Leak Detector, the most technologically advanced instrument of its kind.
Installation and Operating Instructions (for chargers shown below)
Installation and Operating Instructions (for chargers shown below) For additional information please call our Technical Support Group 800.742.2740 PRO CHARGING SYSTEMS, LLC 1551 Heil Quaker Boulevard,
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
AIRMoN, ph and Conductivity Field Measurement
AIRMoN, ph and Conductivity Field Measurement Items needed: Sample to be measured (sample volume 50 ml) Field Observer Form (FOF) for the sample to be measured conductance cell ph vials, 4 conductivity
ETC TWO STAGE ELECTRONIC TEMPERATURE CONTROL
RANCO INSTALLATION INSTRUCTIONS ETC TWO STAGE ELECTRONIC TEMPERATURE CONTROL Relay Electrical Ratings PRODUCT DESCRIPTION The Ranco ETC is a microprocessor-based family of electronic temperature controls,
Gold Century GC-1023 Metal Detector
Gold Century GC-1023 Metal Detector Discriminating Pin Pointing Metal Detector OWNER S MANUAL And Warranty With your GC-1023 metal detector, you can hunt for coins, relics, jewelry, gold, and silver just
UltraBasic Benchtop ph Meter Operation Manual
UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.
ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:
BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data
YSI Model 5000 YSI Model 5100. Dissolved Oxygen Temperature. Operations Manual
YSI Model 5000 YSI Model 5100 Dissolved Oxygen Temperature Operations Manual Table of Contents 1. General Description...1 2. Specifications...3 3. Quick Start...5 4. Controls...6 4.1 Front Panel... 6 4.2
CoolDrive 5. User s Manual
CoolDrive 5 User s Manual Table of Contents Chapter 1 Introduction Chapter 2 Guarantee Information Chapter 3 Component Guide Chapter 4 Installation Chapter 5 Getting started with CoolDrive 5 Chapter 1
AutoRanging Digital MultiMeter
Owner's Manual AutoRanging Digital MultiMeter Model No. 82139 CAUTION: Read, understand and follow Safety Rules and Operating Instructions in this manual before using this product. Safety Operation Maintenance
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421
VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 General Specifications...1 2-2 Electrical Specifications... 2 3. FRONT PANEL DESCRIPTION... 4 3-1
aseries A13B Mini Bullet Camera User Manual
aseries A13B Mini Bullet Camera User Manual Thank you for purchasing our product. If there are any questions, or requests, please do not hesitate to contact the dealer. This manual applies to the MicroView
National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104
National Optical & Scientific Instruments Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR MODELS 106, 106-L 107, 107-L 108, 108-L 109-L ELEMENTARY
Multi Tube Rotator. Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11
Multi Tube Rotator Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11 Table of Contents Safety Information...3 Alert Signals...3 Description...4 Specifications...5 Power Requirements...5 Shaker Motion...5
DX-395. FM/AM/SW1/SW2 PLL Synthesized Receiver OWNER S MANUAL. Cat. No. 20-225. Please read before using this equipment.
20-225.fm Page 1 Wednesday, August 4, 1999 10:07 AM Cat. No. 20-225 OWNER S MANUAL Please read before using this equipment. DX-395 FM/AM/SW1/SW2 PLL Synthesized Receiver 20-225.fm Page 2 Wednesday, August
Ultrasonic support removal tank
Ultrasonic support removal tank User guide and safety manual See inside for use and safety information. The Ultrasonic support removal tank automatically removes PLA support material from a printed ABS
UT202A Operating Manual. Contents
Title Contents Page Overview Unpacking Inspection Safety Information Rules for Safe Operation International Electrical Symbols The Meter Structure Functional Buttons and auto power off Display Symbols
National Optical & Scientific Instrument Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104
National Optical & Scientific Instrument Inc. 11113 Landmark 35 Drive San Antonio, Texas 78233 Phone (210) 590-9010 Fax (210) 590-1104 INSTRUCTIONS FOR STEREOSCOPIC MICROSCOPES MODEL NUMBERS 400 400TL
DM Series CONTENTS. 1. Features 3. 2. Applications 3. 3. General 4. 4. Calibration 5. 5. Dimensions 6. 6. Specifications 7. 7. Calibration procedure 7
DM Series CONTENTS 1. Features 3 2. Applications 3 3. General 4 4. Calibration 5 DM-509-(M/T/S)X-(01/02) & DM-509-(H3/H5)-(01/02) 5. Dimensions 6 6. Specifications 7 7. Calibration procedure 7 The DM Series
DICKSON PW4 DICKSON. Pressure Chart Recorders. Applications & Specifications. Product. Getting Started, Charts & Accessories.
PW4 Pressure Chart Recorders Contents: Applications and Useful Features Operating Instructions / Getting Started Calibrations Troubleshooting Warranty / Factory Service & Applications The PW4 Weather Resistant
Instruction Manual. 2in1 LAN Tester & Multimeter. Model: LA-1011
Instruction Manual 2in1 LAN Tester & Multimeter Model: LA-1011 1 Contents Introduction... Features... Safety Precautions.. Meter Description... Electrical Specification... Operation.. AutoRanging Multimeter.
Operating Manual Electronic Wheelchair Scale with cable display M50610
Operating Manual Electronic Wheelchair Scale with cable display M50610 ADE GmbH & Co. Hammer Steindamm 27 29 D 22089 Hamburg Germany +49(40)432776-0 Fax +49(40)432776-10 E-mail: [email protected] Contents
EzFlex Combustible Gas Leak Detector Model EZ40
User's Guide EzFlex Combustible Gas Leak Detector Model EZ40 Introduction Congratulations on your purchase of the Extech EZ40 Combustible Gas Leak Detector. The EZ40 detects the presence of combustible
ER102. Emergency Crank Weather Radio Owner s Manual. www.midlandradio.com
ER102 Emergency Crank Weather Radio Owner s Manual Table of Contents Welcome to the World of Midland Electronics Major ER102 Features ER102 Controls and Functions ER102 Power Sources - Built-in Rechargeble