1 User Guide Thermo Scientific Orion ROSS Ultra, ROSS Ultra Triode, ROSS and PerpHecT ROSS Electrodes
2 ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion, phisa, phuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS and TURBO2 are registered trademarks of Thermo Fisher pHAX-ION, A+, All in One, Aplus, AQUAsnap, AssuredAccuracy, AUTO-BAR, AUTO-CAL, AUTO DISPENSER, Auto-ID, AUTO-LOG, AUTO-READ, AUTO-STIR, Auto-Test, BOD AutoEZ, Cable-Free, CERTI-CAL, CISA, DataCOLLECT, DataPLUS, digital LogR, DirectCal, DuraProbe, Environmental Product Authority, Extra Easy/Extra Value, FAST QC, GAP, GLPcal, GLPcheck, GLPdoc, ISEasy, KAP, LabConnect, LogR, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, One-Touch, One-Touch Calibration, One-Touch Measurement, Optimum Results, Orion Star, Pentrode, phuture MMS, phuture Pentrode, phuture Quatrode, phuture Triode, Quatrode, QuiKcheK, rf link, ROSS Resolution, SAOB, SMART AVERAGING, Smart CheK, SMART STABILITY, Stacked, Star Navigator 21, Stat Face, The Enhanced Lab, ThermaSense, Triode, TRIUMpH, Unbreakable ph, Universal Access are trademarks of Thermo Fisher. Guaranteed Success and The Technical Edge are service marks of Thermo Fisher. PerpHecT meters are protected by U.S. patent 6,168,707. PerpHecT ROSS electrodes are protected by U.S. patent 6,168,707. ORION Series A meters and 900A printer are protected by U.S. patents 5,198,093, D334,208 and D346,753. ionplus electrodes and Optimum Results solutions are protected by U.S. patent 5,830,338. ROSS Ultra electrodes are protected by U.S. patent 6,793,787. ORP standard is protected by U.S. patent 6,350,367. No Cal electrodes are protected by U.S. patent 7,276, Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice. This publication supersedes all previous publications on this subject.
3 Table of Contents Introduction ROSS Electrode Families Required Equipment Electrode Preparation Preparation for Refillable ROSS Electrodes Preparation for ROSS Sure-Flow Electrodes Preparation for Gel-Filled ROSS Ultra Triode Electrodes Sample Requirements Measuring Hints Electrode Calibration General Calibration Procedure One Buffer Calibration Two Buffer Calibration ph Measurements Electrode Storage Electrode Maintenance Electrode Cleaning Procedures General Cleaning Cleaning Solutions Electrode Characteristics Temperature Effects Sure-Flow Electrode Leak Rate PerpHecT ROSS Micro ph Electrode ROSS Ultra ph/atc Triode Electrodes PerpHecT ROSS Electrode Operation with PerpHecT ph Meters Electrode Interferences Troubleshooting Assistance Warranty RoHS Statement Ordering Information Specifications Thermo Scientific Orion ROSS Electrode User Guide
4 II Thermo Scientific Orion ROSS Electrode User Guide
5 Introduction This user guide contains information on the preparation, operation and maintenance of the Thermo Scientific Orion ROSS Ultra, ROSS and PerpHecT ROSS electrodes. All ROSS ph electrodes have a ph range of 0 to 14, ph precision of 0.01 and temperature range of 0 to 100 C. ROSS electrodes can be used in samples that contain TRIS, sulfides and proteins since they do not contain silver or mercury. ROSS electrodes incorporate the patented ROSS internal reference system that provides superior measurement stability, faster response, greater accuracy and more reproducible results than conventional electrodes. ROSS electrodes have a double junction reference, so the fill solution can be modified to more closely match the sample composition in high ph, low ph or non-aqueous solutions. ROSS electrodes provide readings stable to 0.01 ph in less than 30 seconds, even in the extreme case of samples varying from one another by 50 C or more. Results are three to five times more precise than those obtained with conventional electrodes. Electrode drift is less than ph per day, so recalibration is minimized. All electrodes with a waterproof BNC, BNC or U.S. standard connector have one meter cable lengths, except for the 8107UWMMD and 8157UWMMD electrodes that have a three meter cable length. All electrodes with a screw cap connection require a separate cable. Electrodes with a waterproof BNC connector can be used on any ph meter with a BNC connection. The U ROSS Ultra reference half-cell and ROSS reference half-cell can be used with the 8101BNWP and 8101SC ROSS ph half-cell electrodes or 8411BN ROSS sodium half-cell electrode. Thermo Scientific Orion ROSS Electrode User Guide 1
6 ROSS Electrode Families ROSS Ultra Electrodes ROSS Ultra electrodes offer the ultimate measurement stability and the least long-term drift plus a two year warranty. ROSS Ultra ph/atc Triode Electrodes ROSS Ultra ph/atc Triodes have a built-in temperature probe, so ph and temperature are measured with one electrode. Gel-filled ROSS Ultra ph/atc Triode electrodes have an 18 month warranty and refillable ROSS Ultra ph/atc Triode electrodes have a two year warranty. ROSS Electrodes ROSS electrodes provide precise ph measurements and measurement stability regardless of temperature changes or sample composition. ROSS Sure-Flow Electrodes ROSS Sure-Flow electrodes incorporate a free-flowing junction that is easy to clean and never clogs. These electrodes are ideal for dirty or viscous samples that clog traditional electrodes. PerpHecT ROSS Electrodes PerpHecT ROSS electrodes are designed to be used with PerpHecT meters to provide both ph and temperature measurements. The PerpHecT meters use patented LogR technology that allows temperature to be measured using the resistance of the PerpHecT ROSS electrode. 2 Thermo Scientific Orion ROSS Electrode User Guide
7 Required Equipment 1. Thermo Scientific Orion ph meter, such as the 3-Star ph meter, 4-Star ph/ise meter, 4-Star ph/do meter, 4-Star ph/ conductivity meter or 5-Star ph/ise/do/conductivity meter. ROSS electrodes can be used on any ph meter with a BNC or U.S. standard connection. The electrodes can also be used on meters with a variety of inputs when an adapter cable is used. Visit for details. 2. Thermo Scientific Orion ROSS Ultra, ROSS or PerpHecT ROSS electrode. The 8302BNUMD, 8107BNUMD, 8107UWMMD, 8157BNUMD and 8157UWMMD ph/atc Triodes have a temperature connector that is compatible with the Star series ph meters. The 8107BNU and 8157BNU ph/atc Triodes have a temperature connector that is compatible with the A+ series meters. The 8207BNU and 8257BNU ph/atc Triodes have a temperature connector that is compatible with the PerpHecT meters. 3. ROSS fill solution, Cat. No Note: Do not use any fill solution that contains silver, as silver will damage ROSS electrodes. Note: The 8107BNU, 8107BNUMD, 8107UWMMD and 8207BNU electrodes are gel-filled and do not require any fill solution. 4. ROSS ph electrode storage solution, Cat. No , or standard ph electrode storage solution, Cat. No For enhanced electrode performance, use ROSS ph electrode storage solution. 5. ph buffers at least two ph buffers are recommended for precise measurements. One buffer should be near ph 7 and buffers should be one to four ph units apart. 6. Beakers, plastic or glass. 7. Magnetic stirrer or Thermo Scientific Orion stirrer probe, Cat. No The stirrer probe can be used with 3-Star, 4-Star and 5-Star benchtop meters. 8. Distilled or deionized water. Thermo Scientific Orion ROSS Electrode User Guide 3
8 Electrode Preparation Preparation for Refillable ROSS Electrodes 1. Remove the shipping cap, sleeve or bottle from the sensing bulb and save it for later storage of the electrode. 2. Clean any salt deposits from the exterior of the electrode by rinsing it with distilled water. Handle the 8220BNWP PerpHecT ROSS micro ph electrode with great care. 3. Uncover the fill hole by removing the tape and add ROSS fill solution, Cat. No , to the electrode up to the fill hole. To maintain an adequate flow rate, the level of fill solution must always be above the reference junction and at least one inch above the sample level. The fill hole should remain open whenever the electrode is in use. 4. Place the electrode in an electrode holder and suspend the electrode in air for 5 minutes to thoroughly wet the reference junction. The 8220BNWP electrode and ROSS Ultra Triode electrodes do not need to be suspended in air. 5. Shake the electrode downward (similar to a clinical thermometer) to remove any air bubbles. Do not shake the 8220BNWP electrode unless there are air bubbles in the fill solution and then shake gently and with great care. 6. Soak electrode in ROSS ph electrode storage solution, Cat. No , or standard ph electrode storage solution, Cat. No , for 30 to 60 minutes. ROSS Ultra Triode electrodes do not need to be soaked prior to use, since they are shipped in a storage bottle with storage solution. 7. Connect the electrode to the meter. Fill hole ROSS internal reference ph sensing bulb Fill solution chamber Reference junction ROSS Refillable Combination ph Electrode 4 Thermo Scientific Orion ROSS Electrode User Guide
9 Preparation for ROSS Sure-Flow Electrodes 1. Remove the shipping cap, sleeve or bottle from the sensing bulb and save it for later storage of the electrode. 2. Clean any salt deposits from the exterior of the electrode by rinsing it with distilled water. 3. Uncover the fill hole by removing the tape and add ROSS fill solution, Cat. No , to the electrode up to the fill hole. To maintain an adequate flow rate, the level of fill solution must always be above the reference junction and at least one inch above the sample level. The fill hole should remain open whenever the electrode is in use. 4. Wet the junction by pressing down on the electrode cap to let a few drops of the ROSS fill solution flow out of the electrode Fill hole and then replenish any lost fill solution. 5. Shake the electrode downward (similar to a clinical thermometer) to remove any air bubbles. Fill 6. Soak ROSS electrode in ROSS solutionph electrode storage solution, Cat. internal chamber No , or standard ph electrode storage solution, reference Cat. No , for 30 to 60 minutes. For enhanced electrode performance, Reference use ROSS ph electrode junction storage solution. ph sensing bulb 7. Connect the electrode to the meter. Fill hole ROSS internal reference ph sensing bulb Fill solution chamber Sure-Flow reference junction ROSS Sure-Flow Combination ph Electrode Thermo Scientific Orion ROSS Electrode User Guide 5
10 Preparation for Gel-Filled ROSS Ultra Triode Electrodes 1. Remove the electrode storage bottle from the sensing bulb and save it for later storage of the electrode. Note: Unscrew the cap on the electrode storage bottle before removing the electrode. 2. Clean any salt deposits from the exterior of the electrode by rinsing it with distilled water. 3. Connect the electrode to the meter. 4. If the electrode will not be used immediately, soak it in ROSS ph electrode storage solution, Cat. No , or standard ph electrode storage solution, Cat. No ROSS internal reference Gel solution chamber Reference junction ph sensing bulb ROSS Gel-filled Combination ph Electrode 6 Thermo Scientific Orion ROSS Electrode User Guide
11 Sample Requirements A benefit of the ROSS ph electrodes is that the fill solution composition may be changed depending on the sample requirements. The ROSS ph electrode fill solution, Cat. No , is 3 M KCl. For solutions that precipitate in the presence of chloride ions, the electrode can be filled with 10% KNO 3, Cat. No Electrodes with an epoxy body should only be used in aqueous solutions. Electrodes with a glass body may be used in non-aqueous solutions and solutions that contain organic solvents. A minimum of 20% water must be present in the sample for the best measurement results. If there is a great deal of drift when using the ROSS electrode filled with ROSS fill solution, try filling the electrode with a mixture of methanol and water saturated with KCl. The ratio of methanol to water should be similar to the sample composition. If using the 8220BNWP PerpHecT ROSS micro ph electrode, great care must be taken to thoroughly rinse out the reference chamber and junction before changing the reference solutions. Thermo Scientific Orion ROSS Electrode User Guide 7
12 Measuring Hints Always use fresh buffers for calibration. Choose buffers that are one to four ph units apart. Check the electrode slope daily by performing a two buffer calibration. The slope should be 92 to 102%. Except as noted in the Sample Requirements section, only use ROSS fill solution, Cat. No Do not use any fill solution that may contain silver. If a refillable electrode is used, always remove the fill hole cover during measurements to ensure a uniform flow of fill solution. If a ROSS Ultra Triode is used, immerse the electrode at least 2 inches in solution for best temperature response. Between measurements, rinse the electrode with distilled water and then with the next solution to be measured. Stir all buffers and samples at a uniform rate. Place a piece of insulating material, such as Styrofoam or cardboard, between the magnetic stir plate and beaker to prevent measurement errors from the transfer of heat to the sample. Since ROSS electrodes respond quickly to temperature changes, ph changes due to temperature variations will be noticed. To reduce the chance of error due to polarization, avoid rubbing or wiping the sensing bulb. Use a lint-free tissue and gently blot the sensing bulb. If a refillable electrode is used in dirty or viscous samples or the electrode response becomes sluggish, empty the electrode completely and hold the junction under running water. Empty any water from the electrode and refill it with fresh ROSS fill solution. For a thorough cleaning method, refer to the Electrode Cleaning Procedures section. If a Sure-Flow electrode is used in dirty or viscous samples or the electrode response becomes sluggish, empty the electrode completely and hold the junction open under running water. Empty any water from the electrode and refill it with fresh ROSS fill solution. Press down on the electrode cap to let a few drops of the ROSS fill solution flow out of the electrode and then replenish any lost solution. For a thorough cleaning method, refer to the Electrode Cleaning Procedures section. Handle the 8220BNWP PerpHecT ROSS micro ph electrode with care. Do not touch the ph bulb and stem against the bottom or walls of the sample containers. 8 Thermo Scientific Orion ROSS Electrode User Guide
13 Filling hole should be uncovered Fresh ph buffer Stir all samples and buffers Filling solution level must be higher than sample level Reference junction must be immersed Place insulation between stirrer and beaker Measuring Hints Flat surface ROSS ph electrodes may be used on any moist surface or in liquids. Using a Flat Surface Electrode Thermo Scientific Orion ROSS Electrode User Guide 9
14 Electrode Calibration General Calibration Procedure For detailed instructions on ph calibration, manual ph calibration and temperature compensation, consult your meter user guide. When using PerpHecT ROSS electrodes with a PerpHecT ph meter, refer to the PerpHecT meter user guide for instructions on temperature calibration and LogR temperature compensated ph measurements. Note: If the meter is set to automatically recognize the ph buffer at the measured temperature and an ATC probe is used, you do not need to manually adjust the ph values of the buffers. One Buffer Calibration 1. Choose a buffer near expected sample ph. 2. The buffer should be at same temperature as the sample. If the buffer and samples are at varying temperatures, temperature compensation is recommended. 3. Prepare the meter according to the meter user guide. 4. Rinse the electrode first with distilled water and then with the buffer being used for calibration. 5. Place the electrode into the buffer. When the reading is stable, set the meter to the ph value of the buffer at the measured temperature. Refer to the meter user guide for a detailed procedure. Table 1 provides ph buffer values at various temperatures. 6. Proceed to the ph Measurement section. Table 1 ph Values of Buffers at Various Temperatures Nominal Temperature Buffer Value at 25 C 0 C 5 C 10 C 20 C 30 C Thermo Scientific Orion ROSS Electrode User Guide
15 Two Buffer Calibration This procedure is recommended for precise measurements. 1. Select two buffers that bracket the expected sample ph. The first buffer should be near the electrode isopotential point (ph 7) and the second should be near the expected sample ph (ph 4 or ph 10). Buffers should be one to four ph units apart. 2. The buffers should be at same temperature as the sample. If the buffers and samples are at varying temperatures, temperature compensation is recommended. 3. Rinse the electrode first with distilled water and then with the first buffer. 4. Place the electrode into the first buffer. When the reading is stable, set the meter to the ph value of the buffer at the measured temperature. Refer to the meter user guide for a detailed procedure. Table 1 provides ph values at various temperatures. 5. Rinse the electrode first with distilled water and then with the second buffer. 6. Place the electrode into the second buffer. When the reading is stable, set the meter to the ph value of the buffer at the measured temperature. Refer to the meter user guide for a detailed procedure. Table 1 provides ph buffer values at various temperatures. 7. Proceed to the ph Measurement section. Temperature 40 C 50 C 60 C 70 C 80 C 90 C Thermo Scientific Orion ROSS Electrode User Guide 11
16 ph Measurements 1. Calibrate the electrode as described in the Electrode Calibration section. 2. Rinse the electrode with distilled water and then with the sample. 3. Place the electrode into the sample. 4. When the reading is stable, record the ph and temperature of the sample. Electrode Storage To ensure a quick response and free flowing junction, the sensing bulb and reference junction must not dry out. Short Term Storage (up to one week) Soak the electrode in ROSS ph electrode storage solution, Cat. No , or standard ph electrode storage solution, Cat. No For enhanced electrode performance, use ROSS ph electrode storage solution. To minimize the chance of breakage, the 8220BNWP PerpHecT ROSS micro ph electrode should be attached to an electrode stand and suspended in a beaker that contains storage solution. The electrode should not touch the side or bottom of the beaker. Long Term Storage (more than one week) For refillable ph electrodes, fill the reference chamber and securely cover the fill hole. Cover the sensing bulb and reference junction with the protective cap, sleeve or storage bottle containing a few drops of storage solution. Before returning the electrode to use, prepare it as a new electrode. Electrode Maintenance 1. Inspect the electrode for scratches, cracks, salt crystal buildup or membrane/junction deposits. 2. Rinse off any salt buildup with distilled water. Remove any membrane/junction deposits as directed in the General Cleaning section. 3. If a refillable electrode is used, drain the reference chamber, flush it with fresh ROSS fill solution and refill the chamber with fresh ROSS fill solution. 12 Thermo Scientific Orion ROSS Electrode User Guide
17 Electrode Cleaning Procedures General Cleaning 1. a. For general cleaning, prepare a 0.1 M hydrochloric acid (HCl) or 0.1 M nitric acid (HNO 3 ) solution. b. For cleaning oil or grease, prepare a 0.1 to 0.5 % liquid detergent solution mixed with hot water. c. For cleaning bacterial contaminants, prepare a 1:10 dilution of household laundry bleach. 2. a. Soak the electrode in the 0.1 M HCl or HNO 3 for half an hour and stir the solution at a moderate rate. b. Soak the electrode in the 1:10 dilution of household laundry bleach or the 0.1 to 0.5% liquid detergent solution mixed with hot water for 15 minutes and stir the solution at a moderate rate. 3. If a refillable electrode is used, drain the reference chamber and refill the electrode with fresh ROSS fill solution. 4. Soak the electrode in ROSS ph electrode storage solution, Cat. No , or standard ph electrode storage solution, Cat. No , for at least one hour. Cleaning Solutions Cat. No ph cleaning solution A for removing protein contaminants. Cat. No ph cleaning solution B for removing bacterial contaminants. Cat. No ph cleaning solution C for general cleaning. Cat. No ph cleaning solution D for removing oil and grease contaminants. Cat. No ph cleaning solution kit, includes cleaning solutions A, B, C and D. Thermo Scientific Orion ROSS Electrode User Guide 13
18 Electrode Characteristics Temperature Effects The most common cause of error in ph measurements is temperature. There are at least five ways that temperature variations can affect ph: electrode slope, buffers, samples, reference element drift and temperature sensor errors Electrode Slope Changes The electrode slope will change with variations in temperature. Slope changes may be compensated manually, automatically with an automatic temperature compensation (ATC) probe or with LogR technology using a PerpHecT meter and electrode. Thermo Scientific Orion ph meters calculate the slope based on the measured temperature and automatically adjust the ph value based on the temperature. Buffer and Sample ph Changes Buffer and sample ph values change with temperature because of their temperature dependent chemical equilibria. The ph electrode should be calibrated with buffers that have known ph values at different temperatures. Buffer values at different temperatures are given in Table 1. Thermo Scientific Orion ph meters automatically calibrate with the correct ph buffer values based on the measured temperature. All ph meters are unable to correct sample ph values back to a reference temperature because every sample has a unique ph value versus temperature relationship. Therefore, calibration and measurements should be performed at the same temperature and ph values should be reported with temperature. Reference Element Drift Drift can occur when the internal reference elements inside the ph and reference portions of the electrode are reaching thermal equilibrium after a temperature change. Long-term drift or slow response can last until the sample and electrode are at the same temperature. ROSS electrodes use a patented internal reference to minimize the equilibration time. 14 Thermo Scientific Orion ROSS Electrode User Guide
19 Temperature Sensor Errors When a ph and temperature probe are placed into a sample that varies significantly in temperature, the readings can drift for two reasons. First, the temperature response of the electrode and temperature probe may not be similar, which prolongs equilibration and drift. Second, a sample may not have a uniform temperature. Therefore, the ph electrode and temperature probe are responding to different environments. Using LogR technology, PerpHecT meters sense the temperature directly from the PerpHecT ROSS ph electrodes. The ph and temperature response is identical and both measurements occur at the sensing bulb. Drift is minimized and errors due to environmental differences are eliminated. Sure-Flow Electrode Leak Rate The junction of the ROSS Sure-Flow electrodes has been designed to leak at a rate of approximately 1 ml per 24 hours. To keep the junction from clogging, the leak rate is slightly higher than conventional electrodes. During the average ph measurement time the loss of fill solution is minimal and the fill solution does not contaminate the sample. PerpHecT ROSS Micro ph Electrode The 8220BNWP PerpHecT ROSS micro ph electrode has been designed to measure extremely small samples. The measuring tip is 3 mm in diameter and 40 mm long. The minimum depth of immersion is 4.5 mm and the minimum sample size is 15 µl in 384 well plates. ROSS Ultra ph/atc Triode Electrodes The ROSS Ultra Triode electrode operating temperature range is 0 to 100 C and the temperature accuracy is ± 1.0 C at 2 minutes when immersed at least 2 inches in solution. The temperature equilibration due to the rate of heat transfer can take up to 2 minutes in a stirred, aqueous solution. Thermo Scientific Orion ROSS Electrode User Guide 15
20 PerpHecT ROSS Electrode Operation with PerpHecT ph Meters When PerpHecT ROSS electrodes are used with a PerpHecT ph meter, enhanced temperature compensation is achieved without the need of a separate ATC probe. Using LogR technology, the temperature of the solution is measured through the resistance value of the ph electrode s sensing bulb. PerpHecT ROSS ph electrodes are manufactured to meet the PerpHecT meter specifications, so optimum performance and accuracy are achieved in the LogR mode. Each PerpHecT ROSS ph electrode must be calibrated for temperature before a ph measurement is performed using LogR technology. Refer to the PerpHecT meter user guide for details. For maximum precision, a three point temperature calibration is recommended. Do not perform a one point temperature calibration if measured solutions will be below 20 C. The following tables illustrate the expected ph compensation error for one, two and three point temperature calibrations. The accuracy values are valid only when the temperature calibration is performed within the stated temperature range. The higher temperature range data will apply to measurements made above that temperature range, provided that calibration points are within 20 C of each other. When highly accurate ph results are desired, a separate ATC probe is recommended. One Point Temperature Calibration Electrode Average temp. error C Average temp. error C Average ph error C Average ph error C 8202BN BN BN BN Two Point Temperature Calibration Electrode Average temp. error 0-25 C Average temp. error C Average ph error 0-25 C Average ph error C 8202BN BN BN BN Three Point Temperature Calibration Electrode Average temp. error 0-25 C Average temp. error C Average ph error 0-25 C Average ph error C 8202BN BN BN BN Thermo Scientific Orion ROSS Electrode User Guide
21 Electrode Interferences Sodium ions are the principal interference of the ph electrode and cause increasing error at higher ph (lower hydrogen ion activities) and at higher temperatures. Because the ph membrane is composed of special low sodium error glass, error due to sodium is negligible when measuring at ph values less than 12. When measuring at ph values greater than 12, add the correction value from the nomograph below to the observed ph reading. Typical Sodium Error Sample ph reading ph correction to be added C M M 1 M sample sodium concentration 2 M M 0.5 M 1 M 2 M 5 M 20 C M Example: ph reading Sodium concentration 0.5 M Temperature 50 C Correction 0.01 Corrected ph reading Thermo Scientific Orion ROSS Electrode User Guide 17
22 Troubleshooting Follow a systematic procedure to isolate the problem. The ph measuring system can be divided into four components: ph meter, electrode, sample/application and technique. ph Meter The meter is the easiest component to eliminate as a possible cause of error. Thermo Scientific Orion ph meters include an instrument checkout procedure and shorting cap for convenience in troubleshooting. Refer to the ph meter user guide for directions. Electrode To test electrode operation: 1. Connect the electrode to a working meter that has a mv measuring mode. Set the meter to the mv mode. 2. Rinse the electrode with distilled water and then insert the electrode into fresh ph 7 buffer. When the reading is stable, record the mv value of the ph 7 buffer. The mv value should be in the range of -30 to +30 mv. 3. Rinse the electrode with distilled water and then insert the electrode into fresh ph 4 buffer. When the reading is stable, record the mv value of the ph 4 buffer. The mv value should be in the range of +150 to +210 mv. 4. Calculate the absolute mv difference between the two buffers. The mv difference should be 160 to 180 mv. The actual mv values will change as the electrode ages, but the mv difference between the two buffers should always be 160 to 180 mv. 5. If the electrode fails this procedure, clean the electrode thoroughly as directed in the Electrode Maintenance section and repeat the procedure. If the electrode response is slow or drifting and the electrode is refillable, drain and refill the electrode with fresh ROSS fill solution, Cat. No Replace the electrode if cleaning and maintenance fail to rejuvenate it. 18 Thermo Scientific Orion ROSS Electrode User Guide
23 Sample/Application The electrode and meter may operate with buffers, but not with the sample. In this case, check the sample composition for interferences, incompatibilities or temperature effects. Refer to the Interferences section. Technique If trouble persists, review operating procedures. Review the calibration and measurement sections to be sure proper technique has been followed. Assistance After troubleshooting all components of your measurement system, contact Technical Support. Within the United States call and outside the United States call or fax In Europe, the Middle East and Africa, contact your local authorized dealer. For the most current contact information, visit Warranty For the most current warranty information, visit Thermo Scientific Orion ROSS Electrode User Guide 19
24 RoHS Statement 部 件 名 称 Part Name 传 感 器 体 带 电 缆 装 配 Sensor body with cable assembly 塑 料 硬 件 Plastic hardware 部 件 名 称 Part Name 传 感 器 体 带 电 缆 装 配 Sensor body with cable assembly 塑 料 硬 件 Plastic hardware 有 毒 或 有 害 物 质 和 元 素 Toxic or Hazardous Substances and Elements 铅 Lead (Pb) 汞 Mercury (Hg) 镉 Cadmium (Cd) X 有 毒 或 有 害 物 质 和 元 素 Toxic or Hazardous Substances and Elements 六 价 铬 Hexavalent Chromium (Cr(VI)) 多 溴 联 苯 Polybrominated biphenyls (PBB) 多 溴 二 苯 醚 Polybrominated diphenyl ethers (PBDE) O: 表 明 本 部 件 的 所 有 均 相 材 料 内 该 有 毒 或 有 害 物 质 的 含 量 低 于 根 据 EIP-A, EIP-B, EIP-C 在 SJ/T 里 的 限 制 规 定 Indicates that this toxic or hazardous substance contained in all homogeneous materials for this part, according to EIP-A, EIP-B, EIP-C is below the limit requirement in SJ/T X: 表 明 本 部 件 有 至 少 一 样 均 相 材 料 内 该 有 毒 或 有 害 物 质 的 含 量 高 于 根 据 EIP-A, EIP-B, EIP-C 在 SJ/T 里 的 限 制 规 定 Indicates that this toxic or hazardous substance contained in at least one of the homogeneous materials used for this part, according to EIP- A, EIP-B, EIP-C is above the limit requirement in SJ/T Thermo Scientific Orion ROSS Electrode User Guide
25 Ordering Information Cat. No. Description 8102BNUWP 8302BNUMD 8103BNUWP 8104BNUWP 8107BNU 8107BNUMD 8107UWMMD 8115BNUWP 8135BNUWP 8156BNUWP 8157BNU 8157BNUMD 8157UWMMD U ROSS Ultra combination ph electrode with glass body, waterproof BNC connector ROSS Ultra refillable ph/atc Triode with glass body, waterproof BNC and 8 pin minidin connectors ROSS Ultra combination ph electrode with glass body and semi-micro tip, waterproof BNC connector ROSS Ultra combination ph electrode with glass body and rugged bulb, waterproof BNC connector ROSS Ultra gel-filled ph/atc Triode with epoxy body, BNC and 8 pin DIN connectors ROSS Ultra gel-filled ph/atc Triode with epoxy body, waterproof BNC and 8 pin minidin connectors ROSS Ultra gel-filled ph/atc Triode with epoxy body, waterproof BNC and 8 pin minidin connectors and 3 meter cable ROSS Ultra combination ph electrode with epoxy body and semi-micro tip, waterproof BNC connector ROSS Ultra combination ph electrode with epoxy body and flat surface tip, waterproof BNC connector ROSS Ultra combination ph electrode with epoxy body, waterproof BNC connector ROSS Ultra refillable ph/atc Triode with epoxy body, BNC and 8 pin DIN connectors ROSS Ultra refillable ph/atc Triode with epoxy body, waterproof BNC and 8 pin minidin connectors ROSS Ultra refillable ph/atc Triode with epoxy body, waterproof BNC and 8 pin minidin connectors and 3 meter cable ROSS Ultra reference half-cell electrode with glass body, pin tip connector ROSS Sure-Flow reference half-cell electrode with glass body, pin tip connector Thermo Scientific Orion ROSS Electrode User Guide 21
26 Cat. No. Description 8101BNWP 8101SC 8102BN ROSS ph half-cell electrode with glass body, waterproof BNC connector ROSS ph half-cell electrode with glass body, screw cap connector ROSS combination ph electrode with glass body, BNC connector ROSS combination ph electrode with glass body, U.S. standard connector 8102SC 8103BN ROSS combination ph electrode with glass body, screw cap connector ROSS combination ph electrode with glass body and semi-micro tip, BNC connector ROSS combination ph electrode with glass body and semi-micro tip, U.S. standard connector 8103SC 8104BN ROSS combination ph electrode with glass body and semi-micro tip, screw cap connector ROSS combination ph electrode with glass body and rugged bulb, BNC connector ROSS combination ph electrode with glass body and rugged bulb, U.S. standard connector 8104SC 8115BN 8115SC 8135BN 8135SC ROSS combination ph electrode with glass body and rugged bulb, screw cap connector ROSS combination ph electrode with epoxy body and semi-micro tip, BNC connector ROSS combination ph electrode with epoxy body and semi-micro tip, screw cap connector ROSS combination ph electrode with epoxy body and flat surface tip, BNC connector ROSS combination ph electrode with epoxy body and flat surface tip, screw cap connector ROSS combination ph electrode with epoxy body, BNC connector ROSS combination ph electrode with epoxy body, U.S. standard connector 8155SC ROSS combination ph electrode with epoxy body, screw cap connector 22 Thermo Scientific Orion ROSS Electrode User Guide
27 Cat. No. 8162SC 8163BNWP 8163SC 8165BNWP 8165DN 8172BNWP 8172DN 8175BNWP 8202BN 8203BN 8207BNU 8220BNWP 8235BN 8256BN 8257BNU 8272BN Description ROSS combination ph electrode with glass body and 14/15 standard taper, screw cap connector ROSS combination ph electrode with glass body and spear tip, waterproof BNC connector ROSS combination ph electrode with glass body and spear tip, screw cap connector ROSS Sure-Flow combination ph electrode with epoxy body, waterproof BNC connector ROSS Sure-Flow combination ph electrode with epoxy body, E DIN waterproof connector ROSS Sure-Flow combination ph electrode with glass body, waterproof BNC connector ROSS Sure-Flow combination ph electrode with glass body, E DIN waterproof connector ROSS Sure-Flow combination ph electrode with epoxy body, semi-micro tip and waterproof BNC connector PerpHecT ROSS combination ph electrode with glass body and BNC connector PerpHecT ROSS combination ph electrode with glass body, semi-micro tip and BNC connector PerpHecT ROSS Ultra gel-filled ph/atc Triode with epoxy body and BNC and 3.5 mm phono jack connectors PerpHecT ROSS combination ph electrode with glass body, micro tip and waterproof BNC connector PerpHecT ROSS combination ph electrode with epoxy body, flat surface tip and BNC connector PerpHecT ROSS combination ph electrode with epoxy body and BNC connector PerpHecT ROSS Ultra refillable ph/atc Triode with epoxy body and BNC and 3.5 mm phono jack connectors PerpHecT ROSS Sure-Flow combination ph electrode with glass body and BNC connector Thermo Scientific Orion ROSS Electrode User Guide 23
28 Cat. No. Description ROSS ph electrode fill solution, 5 x 60 ml bottles ROSS ph electrode storage solution, 475 ml bottle ph electrode storage solution, 475 ml bottle All-in-One ROSS ph buffer kit, includes 475 ml bottle each of ph 4.01, 7.00 and buffers, 475 ml bottle of ROSS ph electrode storage solution, 30 ml bottle of ph electrode cleaning solution C and 12 mm electrode storage bottle All-in-One ph buffer kit, includes 475 ml bottle each of ph 4.01, 7.00 and buffers, 475 ml bottle of ph electrode storage solution and 12 mm electrode storage bottle ph 1.68 buffer, 475 ml bottle ph 4.01 buffer, 475 ml bottle ph 5.00 buffer, 475 ml bottle ph 6.86 buffer, 475 ml bottle ph 7.00 buffer, 475 ml bottle ph 9.18 buffer, 475 ml bottle ph buffer, 475 ml bottle ph buffer, 475 ml bottle ph electrode cleaning solution kit, includes 1 x 30 ml bottle each of cleaning solution A and C, 1 x 60 ml bottle each of cleaning solution B and D, pipette and beaker ph electrode cleaning solution A, includes 4 x 30 ml bottles, pipette and beaker ph electrode cleaning solution B, includes 4 x 60 ml bottles, pipette and beaker ph electrode cleaning solution C, includes 4 x 30 ml bottles, pipette and beaker ph electrode cleaning solution D, includes 4 x 60 ml bottles, pipette and beaker Bulb guard for glass ph electrodes, pack of mm electrode storage bottle, pack of mm electrode storage bottle, pack of mm electrode storage bottle, pack of 3 Visit for additional accessories, buffers and buffer sizes. 24 Thermo Scientific Orion ROSS Electrode User Guide
29 Specifications ph Range 0 to 14 ph ph Precision ± 0.01 ph Temperature Range 0 to 100 C ROSS Ultra Triode Temperature Accuracy ± 1.0 C at 2 minutes when immersed at least 2 inches in solution PerpHecT ROSS Temperature Accuracy (Requires PerpHecT Meter) ± 0.5 C Slope 92 to 102% theoretical Nernst slope Isopotential Point ph 7 Internal Reference ROSS Fill Solution (Gel-filled electrodes do not require fill solution) 3 M KCI, Cat. No Accuracy of Measuring ph 6.86 Buffer After Calibration at 25 C Accurate within 0.03 ph for buffer at 0 to 100 C using automatic temperature compensation Response Speed in 6.86 Buffer Between 25 C to 75 C Response stable to 0.01 ph within 30 seconds Response Speed Between 6.86 and 4.01 Buffers at 25 C Response stable to ph within 15 seconds Thermo Scientific Orion ROSS Electrode User Guide 25
30 Specifications (continued) Dimensions 8102BNUWP, 8302BNUMD, 8104BNUWP, 8107BNU, 8107BNUMD, 8107UWMMD, 8135BNUWP, 8156BNUWP, 8157BNU, 8157BNUMD, 8157UWMMD, U, 8101BNWP, 8102BN, 8104BN, 8135BN, , 8162SC, 8165BNWP, 8172BNWP, 8202BN, 8207BNU, 8235BN, 8256BN, 8257BNU, 8272BN 120 mm (length) 12 mm (diameter) Dimensions 8103BNUWP, 8103BN, , 8103SC, 8203BN 165 mm (length) 95 mm (tip length) 6 mm (tip diameter) Dimensions 8115BNUWP, 8115BN, 8115SC, 8175BNWP 165 mm (length) 95 mm (tip length) 8 mm (tip diameter) Dimensions 8163BNWP, 8163SC 110 mm (length) 20 mm (tip length) 4.5 mm (tip diameter) Dimensions 8220BNWP 155 mm (length) 40 mm (tip length) 3 mm (tip diameter) 26 Thermo Scientific Orion ROSS Electrode User Guide
31 Water Analysis Instruments North America 166 Cummings Center Beverly, MA USA Toll Free: Tel: Dom. Fax: Int l Fax: : Europe P.O. Box 254, 3860 AG Nijkerk Wallerstraat 125K, 3862 CN Nijkerk, Netherlands Tel: (31) Fax: (31) Asia Pacific Blk 55, Ayer Rajah Crescent #04-16/24, Singapore Tel: Fax: Rev. A
Handbook. Thermo Scientific. ph Electrode Selection Handbook
Handbook Thermo Scientific ph Electrode Selection Handbook Table of Contents ph Theory... 4 ph Electrode Theory... 4 ph versus Temperature Theory... 5 ph Electrode Selection Guide... 6 Thermo Scientific
User Guide. RDO Optical Dissolved Oxygen Sensor
User Guide RDO Optical Dissolved Oxygen Sensor ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT, PerpHecTion,
Orion Star Series Meter
Analyze Detect Measure Control Orion Star Series Meter User s Guide ROSS and the COIL trade dress are trademarks of. AQUAfast, Cahn, EZ Flash, Ionalyzer, ionplus, KNIpHE, No Cal, ORION, perphect, PerpHecT,
User Guide. 9708 Dissolved Oxygen Probes
User Guide 9708 Dissolved Oxygen Probes Introduction The Thermo Scientific Orion 9708 dissolved oxygen probe simplifies measurements of dissolved oxygen, particularly Biochemical Oxygen Demand (BOD).
Installation Guide Universal Wireless-n Adapter GWU627 PART NO. M1161
Installation Guide Universal Wireless-n Adapter GWU627 PART NO. M1161 Table of Contents Package Contents 4 System Requirements 5 Product Overview 6 Installation 8 Installation without WPS - Windows XP
PROCEDURE COVERSHEET. Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011. In Use Date: 2/1/2011 Discontinued Date:
PROCEDURE COVERSHEET Procedure Title: ph (Urine, Body Fluid, Platelet Samples) Document Section: CHEMISTRY Author: JOANNE WONG Date Authored: JAN 2011 Laboratory Director: TED KURTZ, MD (PARNASSUS & CHINA
UltraBasic Benchtop ph Meter Operation Manual
UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
Installation Guide 2-Port HD Cable KVM Switch with Audio GCS62HU PART NO. M1130
Installation Guide 2-Port HD Cable KVM Switch with Audio GCS62HU PART NO. M1130 Table of Contents Package Contents 4 System Requirements 5 Overview 6 Standard Installation 7 Advanced Installation 9 LED
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
DISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Measuring the Conductivity of Pure Water Using USP<645>
Measuring the Conductivity of Pure Water Using USP Water Analysis Instruments, Thermo Fisher Scientific Application Note 002 Key Words Water conductivity, ultrapure water (UPW), USP, water quality,
USEPA Electrode Method Method 8156
ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method
SenTix 950 SenTix 980
SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
ph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide
AIRMoN, ph and Conductivity Field Measurement
AIRMoN, ph and Conductivity Field Measurement Items needed: Sample to be measured (sample volume 50 ml) Field Observer Form (FOF) for the sample to be measured conductance cell ph vials, 4 conductivity
Installation Guide 2/4-Port DVI KVMP Switch with Cables
Installation Guide 2/4-Port DVI KVMP Switch with Cables GCS1102/GCS1104 PART NO. M1108/M1109 Table of Contents Package Contents 4 System Requirements 5 GCS1102 Overview 6 GCS1104 Overview 8 Installation
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
Model 123A24. Rocket motor ICP pressure sensor, 1000 psi, 5 mv/psi, Helium bleed, Installation and Operating Manual
Model 123A24 Rocket motor ICP pressure sensor, 1000 psi, 5 mv/psi, Helium bleed, Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free:
ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.
Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE
Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor
Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process
Compact ph Measuring Instruments
testo 205, 206, 230 Compact ph Measuring Instruments With innovative probe engineering ph C The new ph measuring instruments with innovative probe engineering Measurement of the ph value plays an important
Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S
Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and
User Manual. Thermo Scientific Orion VERSA STAR
User Manual Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 68X006501 Revision A April 2014 Contents Chapter 1 Meter Introduction... 6 Meter Overview... 6 Packing List... 7
Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3
Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration
For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588).
User Bulletin ABI PRISM 310 Genetic Analyzer 01/2007 SUBJECT: Syringe Care and Maintenance Overview Safety Information Warranty Information In this User Bulletin This user bulletin describes the care,
Instruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter
Instruction Manual HI 9146 Portable Waterproof Microprocessor Dissolved Oxygen Meter w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4
ELECTROCHEMISTRY. Theory and Practice
ELECTROCHEMISTRY Theory and Practice Dr. Axel Bier Lab Product Application Manager - Electrochemistry For more information, please visit: www.hach.com/smartprobes P a g e 2 CONTENT INTRODUCTION 3 HOW TO
A Potentiometric Analysis of Fluoride Ion in Toothpaste
CHEM 311L Quantitative Analysis Laboratory Revision 2.0 A Potentiometric Analysis of Fluoride Ion in Toothpaste In this laboratory exercise, we will analyze a toothpaste sample for its Fluoride Ion (F
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
Instruction Manual Temp 4/5/6 Temperature Meter Series
Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.
CONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor
CONDUCTIVITY SENSOR Description D0382 Figure 1. The Conductivity Sensor Short Description The Conductivity Sensor can be used to measure either solution conductivity or total ion concentration of aqueous
MOISTURE (Karl Fischer, Buffered)
MOIST.03-1 MOISTURE (Karl Fischer, Buffered) PRINCIPLE SCOPE The sample is dissolved in a mixture of methanol and formamide (50:50 v/v) and then titrated with standardized Karl Fischer reagent. The titration
Trace Dissolved Oxygen Sensor
Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
Care and use guide. Contents. page
Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement
Ion Selective Electrodes
Ion Selective Electrodes OAKTON has a large selection of Ion Selective Electrodes () to suit a wide variety of applications. Each electrode has a typical response time of 20 to 30 seconds but will vary
Features, Limitations & Potential Errors
MINIMIZING USER ERRORS IN ph MEASUREMENTS Features, Limitations & Potential Errors of the Tools of ph Measurements Advanced Sensor Technologies, Inc. 603 North Poplar Street Orange CA 92868-1011 USA General
Operator Quick Guide EC SENSOR
Operator Quick Guide EC SENSOR Revision G - 24/04/2008 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra
Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...
Instruction Manual ph 5+ ph/ C ph 6+ ph/ C/mV Ion 6+ ph/ C/mV/Ion Technology Made Easy... 68X243633 Rev 1-1 3/2011 Part of Thermo Fisher Scientific Preface This instruction manual serves to explain the
ph Measurements of Common Substances
Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation
ExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements
Instruction Manual HI 96800 Refractometer for Refractive Index and Brix Measurements www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction
Winemaker s Guide to Chemistry Analysis
Thermo Scientific Orion Product and Application Guide Winemaker s Guide to Chemistry Analysis Recommended electrochemistry instrumentation and detailed application notes for winemaking Chemistry Analysis
Orion High-Level Chloride Monitor
Orion 1817HL Orion High-Level Chloride Monitor INSTRUCTION MANUAL rifos noglib rifos dome glid off a b c d Pure Water MONITORS 1800 series release latches to access Analyze Detect Measure Control AQUAfast,
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features 2 Quick guide 2 To operate 3 Cleaning 4 Battery replacement 4 Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about
Eutech 2700 Series. Bench Meters. ph ORP ION Conductivity TDS Salinity Resistivity Dissolved Oxygen Temperature
Eutech 2700 Series Bench Meters ph ORP ION Conductivity TDS Salinity Resistivity Dissolved Oxygen Temperature Features Comprehensive Backlit Display All Measurements Include Temperature Display Up To 500
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
1. The Determination of Boiling Point
1. The Determination of Boiling Point Objective In this experiment, you will first check your thermometer for errors by determining the temperature of two stable equilibrium systems. You will then use
Gel-filled ph Electrode
51935-88 Gel-filled ph Electrode Models 51935-00, 51935-11, 51935-22 Hach Company, 1999-2001. All rights reserved. Printed in the U.S.A. ap/dh 2/99 1ed te/dk 02/01 2ed 2 TABLE OF CONTENTS SPECIFICATIONS...5
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
Thermo Scientific Orion 2111XP Sodium Analyzer. User Guide
Thermo Scientific Orion 2111XP Sodium Analyzer User Guide ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific Inc. U.S. patent 6,793,787. AQUAfast, Cahn, ionplus, KNIpHE, No Cal, ORION,
Physical Properties of a Pure Substance, Water
Physical Properties of a Pure Substance, Water The chemical and physical properties of a substance characterize it as a unique substance, and the determination of these properties can often allow one to
Bluelab Soil ph Meter Instruction Manual
Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for
6 H2O + 6 CO 2 (g) + energy
AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic
9. Analysis of an Acid-Base Titration Curve: The Gran Plot
9. Analysis of an Acid-Base Titration Curve: The Gran Plot In this experiment, you will titrate a sample of pure potassium hydrogen phthalate (Table 10-4) with standard NaOH. A Gran plot will be used to
Hygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
Laboratory Electrochemistry Products. ph Conductivity Dissolved Oxygen
Laboratory Electrochemistry Products ph Conductivity Dissolved Oxygen Spoilt for choice Small footprint Adjustable viewing angle Wall mountable New style electrode holder Contents page Overview................................3
Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC
Common Problems with Online Water Quality Analyzers Greg Macy Director AquatiPro LLC Objectives Diagnosing Colorimetric Analyzers Identifying Common Problems with Sodium Analyzers Troubleshooting ph Analyzers
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
INSTRUCTIONS 56-1190-98. Edition AC
Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography
Acid Dissociation Constants and the Titration of a Weak Acid
Acid Dissociation Constants and the Titration of a Weak Acid One of the most important applications of equilibria is the chemistry of acids and bases. The Brønsted-Lowry acid-base theory defines an acid
Model 1811EL Low-Level Sodium Monitor
Model 1811EL Low-Level Sodium Monitor INSTRUCTION MANUAL test values C mv slope E 0 sample values ppm ppb error stable enter/ done sample test program cal. begin cal two pt. cal. complete error add std.
Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes
Tech Note Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes Probe Care and Maintenance Periodic cleaning and membrane changes are required to keep your sensors operating properly. Follow
Experiment 17: Potentiometric Titration
1 Experiment 17: Potentiometric Titration Objective: In this experiment, you will use a ph meter to follow the course of acid-base titrations. From the resulting titration curves, you will determine the
Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
Operating Instructions
Operating Instructions FiveEasyPlus ph meter FEP20 Contents Contents 1. Introduction 3 2. Safety measures 3 3. Installation 4 3.1 Installing the electrode arm (optional for FEP20) 4 4. Operating the FiveEasyPlus
The Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
Portable Waterproof ph/dissolved Oxygen Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35632-Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ /
Relative Humidity Calibration Kit
Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration
ph 200 ph-/redox-/temperature Meter Operating instructions
ph 200 ph-/redox-/temperature Meter Operating instructions 2 Table of Contents Page 1. Scope of delivery...4 2. System description...4 3. Connections...5 4. Display...7 5. Keypad...7 6. Start up...8 7.
TEMPERATURE SENSOR USER MANUAL
TEMPERATURE SENSOR USER MANUAL 1 Temperature sensor user manual Copyright 2012 Unisense A/S Version October 2012 TEMPERATURE SENSOR USER MANUAL UNISENSE A/S TABLE OF CONTENTS WARRANTY AND LIABILITY....8
Experiment 8 - Double Displacement Reactions
Experiment 8 - Double Displacement Reactions A double displacement reaction involves two ionic compounds that are dissolved in water. In a double displacement reaction, it appears as though the ions are
Professional Plus. Calibration Tips
Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting
Bose SoundSport. in-ear headphones. for use with Samsung Galaxy devices. Owner s Guide
Bose SoundSport in-ear headphones for use with Samsung Galaxy devices Owner s Guide Important Safety Information Please read this owner s guide carefully and keep for future reference. WARNINGS: DON T
PART I SIEVE ANALYSIS OF MATERIAL RETAINED ON THE 425 M (NO. 40) SIEVE
Test Procedure for PARTICLE SIZE ANALYSIS OF SOILS TxDOT Designation: Tex-110-E Effective Date: August 1999 1. SCOPE 1.1 This method covers the quantitative determination of the distribution of particle
Calibration of Volumetric Glassware
Chemistry 119: Experiment 2 Calibration of Volumetric Glassware For making accurate measurements in analytical procedures, next in importance to the balance is volumetric equipment. In this section volumetric
15. Acid-Base Titration. Discover the concentration of an unknown acid solution using acid-base titration.
S HIFT INTO NEUTRAL 15. Acid-Base Titration Shift into Neutral Student Instruction Sheet Challenge Discover the concentration of an unknown acid solution using acid-base titration. Equipment and Materials
SINGLE RANGE DS METERS Operation Instructions
SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax
General Purpose ph/orp Sensor
Instruction Sheet General Purpose ph/orp Sensor Sensors LIQ_MAN_ABR_3900 October 2014 General Purpose ph/orp Sensor Specifications Figure 1. 3900 Insertion/Submersion Sensor Sensor Type: General purpose
General Purpose ph/orp Sensors
General Purpose ph/orp Sensors Product Data Sheet LIQ_PDS_3900 / Rev. F December 2014 General Purpose ph/orp Sensors SMART preamplifier-stores calibration and other data. Extended sensor life provided
OxyGuard. Dissolved Oxygen Probe
OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard
ph units constitute a scale which allows scientists to determine the acid or base content of a substance or solution. The ph 0
ACID-BASE TITRATION LAB PH 2.PALM INTRODUCTION Acids and bases represent a major class of chemical substances. We encounter them every day as we eat, clean our homes and ourselves, and perform many other
Lab 7 Soil ph and Salinity OBJECTIVE INTRODUCTION Soil ph active
Lab 7 Soil ph and Salinity OBJECTIVE In this lab you will learn the effect of the concentration of hydrogen ions (ph) and various salts on the fertility of a soil. You will perform some tests which are
ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.
Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2
Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986
Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986 Fann Instrument Company Electrical Stability Tester Instruction Manual 2012 Fann Instrument
Experiment 12- Classification of Matter Experiment
Experiment 12- Classification of Matter Experiment Matter can be classified into two groups: mixtures and pure substances. Mixtures are the most common form of matter and consist of mixtures of pure substances.
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn
Hands-On Labs SM-1 Lab Manual
EXPERIMENT 4: Separation of a Mixture of Solids Read the entire experiment and organize time, materials, and work space before beginning. Remember to review the safety sections and wear goggles when appropriate.
Silicon Pyranometer Smart Sensor (Part # S-LIB-M003)
(Part # S-LIB-M003) The smart sensor is designed to work with the HOBO Weather Station logger. The smart sensor has a plug-in modular connector that allows it to be added easily to a HOBO Weather Station.
ph: Measurement and Uses
ph: Measurement and Uses One of the most important properties of aqueous solutions is the concentration of hydrogen ion. The concentration of H + (or H 3 O + ) affects the solubility of inorganic and organic