Operator Quick Guide EC SENSOR
|
|
|
- Avice Parker
- 9 years ago
- Views:
Transcription
1 Operator Quick Guide EC SENSOR Revision G - 24/04/2008
2 General Information About This Guide The information in this guide has been carefully checked and is believed to be accurate. However, Hach Ultra assumes no responsibility for any inaccuracies that may be contained in this guide. In no event will Hach Ultra be liable for direct, indirect, special, incidental, or consequential damages resulting from any defect or omission in this guide, even if advised of the possibility of such damages. In the interest of continued product development, Hach Ultra reserves the right to make improvements in this guide and the products it describes at any time, without notice or obligation. Basic Principle of Operation In its simplest form, an electrochemical cell consists of a metal anode and a metal cathode dipped into an electrolyte solution contacting these electrodes. An electronic circuit is linked to the anode and cathode. Through an applied voltage, current will flow between the anode and the cathode. Gas penetrating through the membrane into the cell dissolves in the electrolyte. It undergoes a reaction at the cathode, causing a measurable electric current to flow. This current is proportional to the amount of gas entering the cell. The amount of gas entering the cell is proportional to the partial pressure of this gas in the sample, outside the cell. The result is shown as gas concentration. The sensor electronics perform four functions: Apply constant voltage to the anode Measure the current flowing through the sensor Compensate for temperature variation in the gaseous or liquid sample Convert the cell s electric current into an analog signal for sensor output Initial Sensor Cell Cleaning Your Orbisphere electrochemical sensor has been thoroughly cleaned and tested at the factory. To protect the electrodes from oxidation, the cell has been filled with electrolyte and a membrane has been installed. However, shipping and storage conditions can adversely affect electrochemical sensor cells, therefore a sensor service (cell cleaning & membrane replacement) must be performed before start up. H 2 sensors do not require a complete cleaning procedure, as dechloridization and rechloridization processes are normally not required. Positioning Information Unless the sensor is part of the Orbisphere equipment that includes it, the sensor must be installed in an Orbisphere socket or flow chamber, that allows the contact with the sample fluid to be analyzed. Check that the sensor will be mounted: perpendicular to the pipe horizontal on a horizontal pipe section (or on flowascending vertical pipe) min. 15 meters away from the pump's discharge side in a place where sample flow is stable and rapid, and as far as possible from: - valves - pipe bends - suction side of pumps - CO 2 injection system or similar There may be situations where not all the above conditions can be met. Please consult your Hach Ultra representative to appraise the situation and define the best applicable solution. Page 2
3 Membrane Removal It is important to install the electrochemical sensor standing on its base. This base offers good protection for the delicate connector socket, and at the same time provides a suitable work stand. Remove the plastic storage cap. Unscrew the protection cap, using the tool provided in the maintenance kit (blue case). Pay attention to the components inside the protection cap. Note the assembly order of each item. The illustration on the right is an example only. Your configuration may differ. Pull up the attaching ring with the tool provided in the maintenance kit. Remove the membrane and mask (if applicable). Drain the electrolyte into a sink and rinse the sensor cavity with tap water. Avoid eye or skin contact with electrolyte which can be slightly corrosive. Insert the prongs of the membrane support removal tool into the membrane support holes, and unscrew the membrane support. The membrane support is individually machined and paired with the sensor. For correct sensor operation, it is ESSENTIAL to keep the membrane support with its respective sensor. Should the membrane support require replacement, contact your Hach Ultra representative. Page 3
4 Membrane Installation Install the membrane support with the groove on the upper side (as illustrated right). The membrane support is individually machined and paired with the sensor. Therefore make sure that the correct membrane support is used on the correct sensor. Insert the prongs of the membrane support removal tool into the membrane support holes. Tighten the membrane support finger tight. Too much torque will damage the sensor electrodes. The membrane mounting surface must be clean and even. Replace the membrane O-ring on the sensor head with a new one. The Nitril O-ring can be reused if it is still in good condition. Membrane O-rings are part of the protection cap kit. Using the syringe or bottle nozzle included in the maintenance kit, fill up the sensor cavity with electrolyte. Be careful not to touch the electrodes with the needle, as a scratch on the surface may lead to loss of performance. Tilt the sensor slightly and inject into the lower hole, pushing bubbles out at the upper hole. Gently tap on the sensor side to move trapped bubbles. Return the sensor to the vertical position. The last drop of electrolyte should form a cupola on top of the sensor tip. Page 4
5 Membrane Installation (cont) In the maintenance kit, pick up the two part membrane mounting tool. Install the sleeve over the sensor head (end with shoulder downwards). Once installed, a membrane cannot be reused. Avoid touching the membrane with bare fingers, as this may affect its sensitivity. Take a few membranes out of the storage box. Using tweezers included in the kit, pick up one membrane from the stack, and gently place it on the sensor tip. Make sure it is centered, and no bubble is trapped. If a sensor mask is used, place it directly on top of the membrane. For correct membrane selection, please refer to the sensor specification table at the end of the EC Sensor Operator s Manual. Distinguish the membrane from the protection paper: - The membrane is transparent (translucent) - The protection paper is opaque The membrane diameter is larger than the sensor head diameter. This is normal, as the membrane will fold over the sensor tip. The membrane holding ring comes in two slightly different internal diameters, depending on the membrane(s) total thickness. For a correct membrane installation, be sure to use the correct holding ring for the application. Place the membrane holding ring on the installation tool tip. To avoid damaging the membrane, make sure that the tool tip is totally clean and its surface is even. Page 5
6 Membrane Installation (cont) Insert the installation tool inside the guiding sleeve. Push the installation tool firmly downwards. This clasps the mounting ring onto the sensor head, folding the membrane over the sensor tip. Remove the installation tool and guiding sleeve. Visually check for correct ring placement, try to push it down with your fingers. Check that the membrane is tight, with no wrinkles, and no bubbles are present. Rinse the sensor with tap water and wipe dry with a clean cloth. Check for electrolyte leaks. Prepare the protection cap for installation, by replacing all the parts inside the protection cap with new ones (except the grille), and place them in the order they were removed. The Tefzel washers, under the cap, should be slightly lubricated with silicone grease. The illustration on the right is an example only. Your configuration may differ. Tighten the protection cap finger tight. Hold the grille with a finger to avoid rotation. Using the tool provided in the maintenance kit (blue case), give it an extra 1/8-1/4 turn, but no more. Always store the sensor with the storage cap and base installed. Put a few drops of clean water in the storage cap to prevent the sensor cell from drying out. A sensor that has been taken apart or serviced must always be calibrated. Allow the sensor to settle for 30 minutes, before performing the sensor calibration. Page 6
7 Chemical Cleaning Cleaning and Regeneration Center The Orbisphere Cleaning and Regeneration Center (illustrated below) is a very efficient cleaning and regeneration tool for Orbisphere electrochemical sensors. This tool reverses the electrochemical process that is taking place in the sensor cell during normal operation. This removes oxidation and at the same time regenerates the electrodes' surface. In addition, the regeneration center offers a continuity tester for checking the sensor electronics. Use of this tool is recommended because the regeneration of the electrodes allows for a noticeably extended sensor life. Detailed information on how to use the Orbisphere tool is included in the related Operator's Manual. It is mandatory to use the tool for servicing electrochemical H 2 sensors. This process is the dechloridization and rechloridization of the electrodes. Oxygen and Ozone Sensor Cell Cleaning The following supposes that the sensor has been taken apart. For disassembly and assembly procedures, see the previous sections. Wear on the membrane, and chemical reactions within the sensor, requires that the sensor be serviced regularly to restore its original sensitivity. Service includes electrode cleaning and membrane replacement. A clear sign that a sensor maintenance is required is when measurements are noticeably less stable than usual, and when a calibration does not improve the situation. Method description: Electrochemical cleaning with the tool (if available) When the tool is not available or results are insufficient: Anode and cathode chemical cleaning Central electrode polishing Final rinsing To eliminate any silver residue that ammonia cleaning cannot remove, it is sometimes required to repeat the chemical cleaning using nitric acid (HNO 3, not over 70% by weight). Hydrogen Sensor Cell Cleaning The hydrogen analyzer works on the principle that hydrogen molecules, passing through the membrane, generate an electric current at the platinum anode surface. For this to take place, an extremely clean metal surface is essential. If any film, grease or other impurity covers the platinum surface, the reaction is impeded and may even be stopped. In addition, the chemical reaction that takes place on the chloridized silver cathode leads to loss of performance after a certain operation time. As a result, a sensor service must be carried out to restore its original performance. The procedure for cleaning the H 2 electrochemical sensor requires the use of the Orbisphere tool, and is explained in full in the Operator's Manual. As an overview, it consists of the following sequence of operations: Dechloridization of the cathode: This process removes the chloride film from the silver cathode surface (carried out by the Orbisphere 32301). Rechloridization of the cathode: A layer of silver chloride is grown on the cathode's surface (carried out by the Orbisphere 32301). Activation of the platinum anode: The center anode surface is polished, and treated with nitric acid. Page 7
8 Chemical Cleaning (cont) Membrane Support Cleaning Empty and rinse the electrolyte reservoir under tap water. Rinse the membrane support under water and wipe it dry. Check for the presence of any residue on the surfaces. Residue can be removed by placing the support in a container of nitric acid (HNO 3, not over 70% by weight) until it recovers its original appearance (normally within 30 seconds) Rinse one minute under tap water and check again for surface cleanliness. Nitric acid is dangerous! Please refer to the safety information from your chemical supplier. Anode and Cathode Ammonia Cleaning Fill the sensor electrolyte reservoir with a solution of 25% by weight ammonium hydroxide (NH4OH) in water and leave for 10 minutes. Rinse with tap water for at least one minute. Inspect the sensor head. The counter electrode should be a silver-white color. If the counter electrode still shows deposits, repeat the procedure. Ammonia is dangerous! Please refer to the safety information from your chemical supplier. Page 8
9 Chemical Cleaning (cont) Anode and Cathode Nitric Acid Cleaning Check for the absence of silver deposit on the central guard ring electrode walls (indicated right, by the arrows), as such deposits can make contact with the counter electrode. To eliminate any silver residue inside the sensor cell, it is sometimes required to repeat the chemical cleaning using nitric acid (HNO 3, not over 70% by weight). Also, the electrochemical cleaning does not remove deposit on the cell s plastic parts, so nitric acid cleaning may be required. This procedure is not recommended for normal maintenance, and should not be used more often than twice a year, as the acid degrades the metal of the counter electrode, thus reducing the sensor's life. Place concentrated nitric acid into the sensor electrolyte reservoir, and add 1 drop on the center electrode. Leave for no longer than 3 seconds. Quickly empty the acid and rinse thoroughly under tap water for one minute. Nitric acid is dangerous! Please refer to the safety information from your chemical supplier. Page 9
10 Chemical Cleaning (cont) Sensor Face Polishing Once the sensor has been cleaned, the face of the center electrode must be polished together with the membrane support. Install the membrane support with the groove on the upper side (as illustrated right). The membrane support is individually machined and paired with the sensor. Therefore make sure that the correct membrane support is used on the correct sensor. Insert the prongs of the membrane support removal tool into the membrane support holes. Tighten the membrane support finger tight. Too much torque will damage the sensor electrodes. Place the dish with the polishing cloth on a flat surface. Spread a little polishing powder onto the cloth. Mix with a few drops of water to get a white, milky liquid. Make sure to use the correct polishing powder for your application (refer to the spare parts table at the end of the EC Sensor Operator s Manual). Use one polishing cloth per sensor model, to prevent a possible contamination through metal particle transfer. Holding the sensor vertically, and using a circular motion, polish the sensor face for at least 30 seconds, until the electrodes are clean and shiny. This step may need to be repeated several times. Make sure to avoid skin contact with the polishing cloth; it should be kept free of dust and grease. Page 10
11 Chemical Cleaning (cont) Sensor Face Polishing (cont) Remove the membrane support with the installation tool. Rinse the support and sensor cavity with a strong jet of clean water. Use distilled water if the water quality is doubtful. Carefully inspect that the tiny groove between the center electrode and the guard ring electrode (illustrated right) is totally clean and free of any polishing residue. Clean only with a strong water spray. The edge of a paper sheet can be used to remove any sticking residue. O 3 Sensor Only: Final Center Electrode Cleaning Once the O 3 sensor has been successfully cleaned and polished, a final nitric acid treatment should be applied, as follows: Place the sensor in a vertical position on its base. Fill the electrolyte reservoir with a few drops of water, just enough to cover the outer electrode. The center electrode must be kept dry. Place a drop of nitric acid on the center electrode, covering only the electrode and guard ring. Avoid spilling acid into the water. Wait 1 minute, then rinse thoroughly under tap water. Nitric acid is dangerous! Please refer to the safety information from your chemical supplier. Page 11
12
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
DISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
Trace Dissolved Oxygen Sensor
Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION
OxyGuard. Dissolved Oxygen Probe
OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard
Volkswagen New Beetle 2.0 Liter 4-cyl General, Engine (Engine Code AEG) 17 Engine-Lubrication system (Page GR-17)
17 Engine-Lubrication system (Page GR-17) Lubrication system components, removing and installing Oil pan, removing and installing Oil pressure and oil pressure switch, checking Dynamic oil pressure warning
Volkswagen Jetta, Golf, GTI 1999, 2000 2.8 Liter VR6 2V Engine Mechanical, Engine Code(s): AFP 17 Engine-Lubrication (Page GR-17)
17 Engine-Lubrication (Page GR-17) Lubrication system components, removing and installing Oil filter housing, disassembling and assembling Oil pan, removing and installing Oil pressure and oil pressure
VC 104+ Rigid Grade / Rigid Grade Imagine VC 104 Rigid Grade Commercial Customised
Page 1 of 6 instructions VC 104+ Rigid Grade / VC 104+ Rigid Grade Imagine / (for the rest of this document referred to as VC 104) can be applied to new and used trucks and trailers with painted rigid
Instruction Manual. Alfa Laval SB Membrane Sample Valve ESE02963-EN1 2015-10. www.sks-online.com www.sks-webshop.com.
Instruction Manual Alfa Laval SB Membrane Sample Valve ESE02963-EN1 2015-10 Original manual Table of contents The information herein is correct at the time of issue but may be subject to change without
Installation Instructions 4508 4508S
SYMPHONY Spread Lavatory Faucet with Speed Connect Drain Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American Standard faucets. Speed Connect
81000 Series Aluminum Gate Valve Installation and Repair Manual
High Vacuum Valves Leader in Quality and Value 81000 Series Aluminum Gate Valve Installation and Repair Manual Dear Customer, Thank you for choosing HVA as your vacuum valve supplier. Your valve will give
FASCINATION 700 HVLP TANNING PRO SYSTEM USER MANUAL
FASCINATION 700 HVLP TANNING PRO SYSTEM USER MANUAL Congratulations on choosing the Fascination 700 HVLP Tanning Pro System! Your system includes the following items: 1 Fascination 700 HVLP Tanning Pro
How compact discs are made
How compact discs are made Explained by a layman for the laymen By Kevin McCormick For Science project at the Mountain View Los Altos High School Abstract As the major media for music distribution for
For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/.
Instruction Sheet PN 51A-499AOZ/rev.H September 2010 Model 499A OZ Ozone Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS APPLICATION
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
Scotchgard Multi-Layer Protective Film 1004
Scotchgard Multi-Layer Protective Film 1004 3 Technical Bulletin September, 2008 Scotchgard Multi-Layer Protective Film 1004 Application Kit (3M Part No. 70-0710-4666-1) includes: (clockwise) Heavy Duty
Bydureon 2 mg powder and solvent for prolonged-release suspension for injection
INSTRUCTIONS FOR THE USER Your step by step guide Bydureon 2 mg powder and solvent for prolonged-release suspension for injection If you have questions about taking Bydureon Refer to the Common questions
Safety, Cleaning, and Chemical Disposal Procedures
Safety, Cleaning, and Chemical Disposal Procedures 1. Using Acids At many points in the fabrication process strong acids are used as etchants. These cause severe burns if kept in contact with your skin
2" & 3" Poly Wet Seal Pump Instruction Manual
Always place the pump as close to the liquid to be pumped as possible. Keep the suction line short and with few bends. Keep the pump and engine on a level foundation. A poor foundation and a heavy suction
Lenntech. 1000 psi End port pressure vessels 2.5. User s Guide to: Phoenix Vessel Technology Limited. Model number: 1503
1 User s Guide to: Phoenix Vessel Technology Limited 1000 psi End port pressure vessels 2.5. Model number: 1503 Lenntech [email protected] Tel. +31-152-610-900 www.lenntech.com Fax. +31-152-616-289 Phoenix
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
Volkswagen Jetta, Golf, GTI 1999, 2000 Brake System 47 Brakes - Hydraulic Components (Page GR-47)
47 Brakes - Hydraulic Components (Page GR-47) FS III front brake calipers, servicing Front brake caliper piston, removing and installing FN 3 front brake calipers, servicing Front caliper piston, removing
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
BACKPACK SPRAYERS. MODEL NOS: KSP16 & KSP20 Part Nos: 3402270 & 3402275 OPERATING & MAINTENANCE INSTRUCTIONS GC04/12
BACKPACK SPRAYERS MODEL NOS: KSP16 & KSP20 Part Nos: 3402270 & 3402275 OPERATING & MAINTENANCE INSTRUCTIONS GC04/12 INTRODUCTION Thank you for purchasing this CLARKE Sprayer, designed for use only with
Ozone Sensor. Model 499A OZ CAUTION CAUTION. Instruction Sheet PN 51A-499AOZ/rev.F January 2007
Instruction Sheet PN 51A-499AOZ/rev.F January 2007 Model 499A OZ Ozone Sensor For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/.
SC, FC and ST * Hot Melt Fiber Optic Connectors
SC, FC and ST * Hot Melt Fiber Optic Connectors Instructions September 1998 78-8073-7873-8 2 Contents 1.0 Safety and Helpful Hints...4 2.0 Field Termination Kit Contents...4 3.0 STÑInstallation on 900
MDU SC Connector Termination Procedure
MDU SC Connector Termination Procedure Series: 718-NANN-NNNNN B Update the stripping length 11/24/09 A Initial Release 4/30/09 Version Revision History Summary Issue Date I INTRODUCTION This document describes
The Drink-Aide The NEW look of independence
The Drink-Aide The NEW look of independence Introducing Drink-Aide What is Drink-Aide? The Drink-Aide is a water bottle that attaches easily to a wheelchair for use by persons with little or no upper
STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance
Page 1 of 9 KEY WORDS- Dissolved Oxygen, calibration, probe preparation, meter preparation APPROVALS APPROVED BY: Kean S. Go, DATE: +/?? APPROVED BY: A Pement DATE: a// A/? f Lisa Ross, EHAP Senior Scientist
Back Pack Sprayer. Operator's Manual MODELS MS - 40 MS - 50
Back Pack Sprayer Operator's Manual MODELS MS - 40 MS - 50 WARNING DANGER Read rules for safe operation and all instructions carefully. ECHO provides this Operator's Manual which must be read and understood
FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE. Battery operation instructions:
FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE Battery operation instructions: 1. The removing and inserting of batteries is to be conducted by the adults only. Unscrew the screw holding the battery pack s cover
Determining Equivalent Weight by Copper Electrolysis
Purpose The purpose of this experiment is to determine the equivalent mass of copper based on change in the mass of a copper electrode and the volume of hydrogen gas generated during an electrolysis reaction.
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
Rebuild Instructions for 70001 and 70010 Transmission
Rebuild Instructions for 70001 and 70010 Transmission Brinn, Incorporated 1615 Tech Drive Bay City, MI 48706 Telephone 989.686.8920 Fax 989.686.6520 www.brinninc.com Notice Read all instructions before
Recovery of Elemental Copper from Copper (II) Nitrate
Recovery of Elemental Copper from Copper (II) Nitrate Objectives: Challenge: Students should be able to - recognize evidence(s) of a chemical change - convert word equations into formula equations - perform
Instructions for Use. Components of the GENOTROPIN PEN 12
Instructions for Use Important Note Please read these instructions completely before using the GENOTROPIN PEN 12. If there is anything you do not understand or cannot do, call the toll-free number listed
BUTTERFLY SHOWER DOOR INSTALLATION INSTRUCTIONS
BUTTERFLY SHOWER DOOR INSTALLATION INSTRUCTIONS IMPORTANT DreamLine reserves the right to alter, modify or redesign products at any time without prior notice. For the latest up-to-date technical drawings,
1.8 CRANKSHAFT OIL SEALS
SERIES 60 SERVICE MANUAL 1.8 CRANKSHAFT OIL SEALS An oil seal is fitted between each end of the crankshaft and the bores of the flywheel housing and gear case cover to retain the lubricating oil in the
EEG Cap Instructions
EEG Cap Instructions Supplies You will need to have the following supplies at hand before applying the cap, all of which can be found in the daily preparation supplies cabinet: NuPrep mildly abrasive skin
ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.
Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE
User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600
User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously
POSEIDON 2-29, 2-25 & 2-22 POSEIDON 2-29, 2-25 & 2-22 XT
POSEION 2-29, 2-25 & 2-22 POSEION 2-29, 2-25 & 2-22 XT Repair Manual Index A. Safety precautions 3 B. Technical data 4 C. Structure 5-6. Service / Repair 7-23 E. Tools 24 F. Function 25-26 G. Electric
GENUINE PARTS INSTALLATION INSTRUCTIONS
GENUINE PARTS INSTALLATION INSTRUCTIONS DESCRIPTION: Illuminated Kick Plate APPLICATION: Rogue (2011) PART NUMBER: 999G6 GX010 KIT CONTENTS: Item A B C G H QTY 1 1 1 D 1 E 1 F 3 15 6 Description Kick Plate,
ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST
ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST Model QT66 Model QT12 Model QT6 LORTONE 12130 Cyrus Way Mukilteo, WA 98275 (425) 493-1600 SETTING UP YOUR MACHINE Install The Barrel Guide: Using the machine
For safety information, see the ABI PRISM 310 Genetic Analyzer User s Manual (PN 903565 or 4317588).
User Bulletin ABI PRISM 310 Genetic Analyzer 01/2007 SUBJECT: Syringe Care and Maintenance Overview Safety Information Warranty Information In this User Bulletin This user bulletin describes the care,
Petri Dish Electrolysis Electrolysis Reactions
elearning 2009 Introduction Petri Dish Electrolysis Electrolysis Reactions Publication No. 95008 Electrolysis is defined as the decomposition of a substance by means of an electric current. When an electric
Front brakes (FN- 3), servicing
j a t Front brakes (FN- 3), servicing 46-1 Front brakes, servicing Note: Install complete repair kit. After replacing brake pads and before moving vehicle, depress brake pedal several times firmly to properly
3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE
4 3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE A. Changing the Control Board B. Replacing a Heating Element C. Thermistor Replacement D. High Limit Switch Replacement E. Level Detector Replacement
INSTALLATION & OPERATING INSTRUCTIONS
INSTALLATION & OPERATING INSTRUCTIONS WARNING RISK OF ELECTRIC SHOCK. CONNECT ONLY TO A CIRCUIT PROTECTED BY A GROUND-FAULT CIRCUIT-INTERRUPTER. THE UNIT SHOULD BE INSTALLED BY A QUALIFIED SERVICE REPRESENTATIVE.
Rear wheel brakes, servicing. Стр. 1 из 45. Note:
Volkswagen Touareg - Rear wheel brakes, servicing Стр. 1 из 45 46-2 Rear wheel brakes, servicing Rear brakes, FN 44 brake caliper, servicing Note: After replacing brake pads, depress brake pedal firmly
Munchkin Heat Exchanger Maintenance
Munchkin Heat Exchanger Maintenance By: Mike Granahan *Refer to the appropriate boiler manual for your specific model number. Adjustment procedures will vary. Boiler shown below is an HTP Munchkin 80M
For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com.
For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at www.rockshox.com. 2 0 0 5 D U K E A I R X C / S L / R A C E S E R V I C E G U I D E Information
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Austin Peay State University Department of Chemistry CHEM 1111. Empirical Formula of a Compound
Cautions Magnesium ribbon is flammable. Nitric acid (HNO 3 ) is toxic, corrosive and contact with eyes or skin may cause severe burns. Ammonia gas (NH 3 ) is toxic and harmful. Hot ceramic crucibles and
NITROUS TRANSFER PUMP INSTRUCTIONS
NITROUS TRANSFER PUMP INSTRUCTIONS SAFETY TIPS Never directly inhale nitrous oxide. When inhaled in large quantities, nitrous oxide can cause respiratory ailments or in extreme cases, death by suffocation.
BILLET HEADLAMP SHELL
-J008 REV. 007-07- BILLET HEADLAMP SHELL GENERAL Kit Number 770-0 Models For model fitment information, please see the P&A Retail Catalog or the Parts and Accessories section of www.harleydavidson.com
Reasons for reissue are in Section 6, REVISION SUMMARY.
Figure 1 This instruction sheet covers the application of OPTIMATE FSMA Fiber Optic Connector Types 905 and 906 for data and telecommunications applications. Base part numbers which apply to each type
1000-LB. TRAILER JACK OWNER S MANUAL
1000-LB. TRAILER JACK OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in serious
CAST IRON THE BASICS. Heatline - Cast Iron Radiators SMOOTH FLAT FILE TO REMOVE ANY SWARF. ONE TIME. ASSEMBLY. JOINTS SHOULD BE TIGHTENED.
CAST IRON THE BASICS 1. DO NOT LIFT ON YOUR OWN. 2. ONLY LIFT THE RADIATOR VERTICALLY. 3. DO NOT LIFT MORE THAN 8/10 SECTIONS AT ANY ONE TIME. 4. POSITION THE RADIATOR BEFORE FINAL ASSEMBLY. 5. THIS PRODUCT
OWNER S MANUAL Table Tennis Table Patent Pending
OWNER S MANUAL Table Tennis Table Patent Pending Be sure to write your model number and serial number here for future reference. You can find these numbers printed on the bottom of the table. MODEL # T8179
AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS
AKRON EDUCTORS TROUBLESHOOTING GUIDE OPERATION & THEORY OF EDUCTORS GENERAL OPERATING AND MAINTENANCE INSTRUCTIONS Products Include: 60 gpm eductors Style 3060, 3061, 3062 & 3070 95 gpm eductors Style
Electronics and Soldering Notes
Electronics and Soldering Notes The Tools You ll Need While there are literally one hundred tools for soldering, testing, and fixing electronic circuits, you only need a few to make robot. These tools
4-600, 1000 psi. User s Guide to: Phoenix Vessel Technology Limited. End port pressure vessels Model numbers: 1386, 1384.
User s Guide to: Phoenix Vessel Technology Limited 4-600, 1000 psi End port pressure vessels Model numbers: 1386, 1384. Phoenix Vessel Technology Limited Unit 2, The Old Bakery, Lower Tuffley Lane, Gloucester
SunMaxx Solar Filling Station Operating Instructions
SunMaxx Solar Filling Operating Instructions Content 1. Declaration of conformity... 2 2. Introduction... 2 3. Transportation and unpacking... 4 4. Mounting and commissioning... 5 5. End of operation...
Private Water Supplies Sampling Manual. A Field Guide
Private Water Supplies Sampling Manual A Field Guide Foreword This sampling manual details standard procedures for the collection, storage and transportation of samples at private water supplies in accordance
Char-Lynn Spool Valve Hydraulic Motors. Repair Information. W Series Geroler Motors
Char-Lynn Spool Valve Hydraulic Motors Repair Information W Series Geroler Motors with Parking Brake 004 Nut Key Ring, Retaining Bearing Ring, Retaining Ring, Retaining Washer (Thick), Pressure Washer,
INGROUND, ROPED, AND HOLELESS JACK SEAL REPLACEMENT
D. L. Martin Company INGROUND, ROPED, AND HOLELESS JACK SEAL REPLACEMENT TYPICAL HYDRAULIC JACK HEAD ASSEMBLY 1 TOOLS REQUIRED: 1) Container to hold hydraulic fluid. 2) Clean rags and protective covers
ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST
ROTARY TUMBLER INSTRUCTIONS AND PARTS LIST Model 3-1.5B Model 45C Model 33B LORTONE 12130 Cyrus Way Mukilteo, WA 98275 (425) 493-1600 SETTING UP YOUR MACHINE For 45C Unit Only (33B & 3-1.5B already assembled)
IMPORTANT DOCUMENTATION DO NOT DISCARD!
PART NO.: 6441-263C SERIES GRT 3 JAW PARALLEL GRIPPERS INFORMATION SHEET IMPORTANT DOCUMENTATION DO NOT DISCARD! Use this information sheet to assist with gripper installation and setup. File with maintenance
Manual for GlobePharma Mini-Press II Rotary Tablet Press
1 of 13 Preparing the Rotary Press 1. Make sure the rotary press is unplugged. 2. Open the bottom cabinet of the rotary press and take out the grey tool kit, and the beige box of punches and dies. 3. Take
Dry Ice Color Show Dry Ice Demonstrations
elearning 2009 Introduction Dry Ice Color Show Dry Ice Demonstrations Publication No. 95016 Add a small piece of solid carbon dioxide to a colored indicator solution and watch as the solution immediately
Determination of a Chemical Formula
1 Determination of a Chemical Formula Introduction Molar Ratios Elements combine in fixed ratios to form compounds. For example, consider the compound TiCl 4 (titanium chloride). Each molecule of TiCl
SOLAR PV SYSTEM MAINTENANCE GUIDE
SOLAR PV SYSTEM MAINTENANCE GUIDE GUYANA HINTERLANDS STAND-ALONE SOLAR PV INSTALLATIONS IMPROVING HEALTH FACILITY INFRASTRUCTURE (IHFI) GUYANA CONTRACT NO. EPP-I-00-03-00008-00, TASK ORDER 07 APRIL 2013
Model MS-41BP 4G/ 15L. WARNING Carefully Read These Instructions Before Use 013529 R0514
Backpack Sprayer Use and Care Manual Model MS-41BP 4G/ 15L WARNING Carefully Read These Instructions Before Use 013529 R0514 WARNING WARNING: Improper use or failure to follow instructions can result in
Auto-belay Cable Replacement Process
Auto-belay Cable Replacement Process Version 2.00 WARNING: The air pressure in the auto-belay system is what causes the cable to be retracted when releasing the cable or climbing the wall with the cable
ClearView 100. Instruction Manual
ClearView 100 Instruction Manual Important Safeguards This appliance is not intended for use by children or infirm persons without supervision. Young children should be supervised to ensure that they do
SCION tc 2005 - DASH APPLIQUE Preparation
Preparation Part Number: PTS10-21051 Kit Contents Item # Quantity Reqd. Description 1 1 HVAC Center Vent Trim 2 1 Radio Cover Assy (2 pieces) 3 1 HVAC Control Panel (3 pieces) 4 1 Shift Surround 5 1 Driver
RAUFOSS EXPLOSIVE COMPRESSION FITTINGS
INSTRUCTION MANUAL FOR RAUFOSS EXPLOSIVE COMPRESSION FITTINGS Box 7 2831 Raufoss Norway Tlf.: +47 61 15 17 87 Fax: +47 61 15 25 56 25 January 2013, rev 2 Available on www.vpmetall.no Page 1 CONTENTS GENERAL
Installation, Operating and Maintenance Instructions for AEON Metal Seated Gate Valve (MSGV) Water
1 Foreword 1.1 Thank you for choosing one of the range of AEON's BS standard valves. For safety purposes, the user shall read this instruction to fully understand the operation of the product. 2 AEON Design
EXCAVATOR SAFETY TRAINING
EXCAVATOR SAFETY TRAINING INSPECTION CHECKLIST INSPECTION AREA INSPECTION RESULTS Sat. Unsat. N/A Comments Carrier & Car Body Rotation system Tracks Rollers Frame, welds, bolts Drive system Upper Structure
Instructions for Using the Watch Works Tool Kit to Change a Watch Battery
Instructions for Using the Watch Works Tool Kit to Change a Watch Battery Click on this link http://www.allamericanwatches.com/site/626101/product/e2306-a to purchase the Watch Battery Replacement Tool
Oxidation States of Copper Two forms of copper oxide are found in nature, copper(i) oxide and copper(ii) oxide.
The Empirical Formula of a Copper Oxide Reading assignment: Chang, Chemistry 10 th edition, pp. 55-58. Goals The reaction of hydrogen gas with a copper oxide compound will be studied quantitatively. By
tire inflator with pressure gauge
tire inflator with pressure gauge Model 95583 Assembly And Operation Instructions Due to continuing improvements, actual product may differ slightly from the product described herein. 3491 Mission Oaks
SenTix 950 SenTix 980
SenTix 950 SenTix 980 ph combination electrodes with refillable liquid reference system Operating manual ba75843e01 08/2009 SenTix The latest version of the present operating manual can be found on the
STATFlow. Direct-to-Drain System. For use with STATIM 2000 and STATIM 5000 Autoclaves. Installation Guide and Operator s Manual
STATFlow Direct-to-Drain System For use with STATIM 2000 and STATIM 5000 Autoclaves Installation Guide and Operator s Manual STATFlow Installation Guide and Operator s Manual 95-113193 Rev 2.0. Copyright
758 Heavy-duty Ratchet Guy Wire Cutter
INSTRUCTION MANUAL 758 Heavy-duty Ratchet Guy Wire Cutter Read and understand all of the instructions and safety information in this manual before operating or servicing this tool. Register this product
Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3
Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration
300 SERIES 331, 332, 333, 344, 356 AND 367 MODELS
Section: MOYNO 500 PUMPS Page: 1 of 8 Date: March 1, 1998 SERVICE MANUAL MOYNO 500 PUMPS 300 SERIES 331, 332, 333, 344, 356 AND 367 MODELS Mechanical Seal Models Packing Gland Models MODELS DESIGN FEATURES
BATTERIES 2010, 2003 by David A. Katz. All rights reserved. Permission for academic use with original copyright retained.
BATTERIES 2010, 2003 by David A. Katz. All rights reserved. Permission for academic use with original copyright retained. Introduction The first electrical storage device, the Leyden Jar was invented by
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation
The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation Introduction The importance of keeping cooling tower water in proper chemical
HUMULIN 70/30 KwikPen
1 Instructions for Use HUMULIN 70/30 KwikPen (70% human insulin isophane suspension 30% human insulin injection [rdna origin]) Read the Instructions for Use before you start taking HUMULIN 70/30 and each
National- Spencer Inc.
9-27-2010 National- Spencer Inc. 19.2V HEAVY DUTY GREASE GUN PRODUCT SPECIFICATION Charger Input Power 110 VAC Battery Output Power 19.2V Battery Capacity 1500 MAH Battery Pack Charge Time 1 Hour Maximum
Operating Instructions & Maintenance Guidelines for Precision Vaporizers
Operating Instructions & Maintenance Guidelines for Precision Vaporizers This page intentionally blank. Examination and Preparation for Use We want you to be completely satisfied with your vaporizer. Please
EXPRESS STILL WARNING It is legal in New Zealand to own and operate a still for the purpose of producing alcohol for your own consumption.
THE EXPRESS STILL This still represents the next generation in home distillation equipment. It eliminates the problem of having to spend all day processing a 25 litre wash. Taking 1 hour to heat up, then
The Enbrel SureClick autoinjector is a single-use prefilled autoinjector. It contains one 50-mg dose of Enbrel.
Instructions for Use Welcome! The Enbrel SureClick autoinjector is a single-use prefilled autoinjector. It contains one 50-mg dose of Enbrel. Your healthcare provider has prescribed Enbrel SureClick autoinjector
Lunette 2 Series. Curved Fixed Frame Projection Screen. User s Guide
Lunette 2 Series Curved Fixed Frame Projection Screen User s Guide Important Safety and Warning Precautions Please follow these instructions carefully to ensure proper maintenance and safety with your
Beautifully Traditional
Arcade 800mm Single Sliding Door Quadrant - nickel Arcade 900mm Single Sliding Door Quadrant - nickel ARC48 ARC49 INSTALLATION INSTRUCTIONS A www.arcadebathrooms.com IMPORTANT - Please read before installation
100 Series Maxi Internal Filters. Instruction Manual
100 Series Maxi Internal Filters Instruction Manual Important Safeguards This appliance is not intended for use by children or infirm persons without supervision. Young children should be supervised to
13. REAR WHEEL/BRAKE/SUSPENSION
13. REAR WHEEL/BRAKE/SUSPENSION 13 3.5~4.5kg-m 8.0~10.0kg-m 0.8~1.2kg-m 3.0~4.0kg-m 2.4~3.0kg-m 3.5~4.5kg-m 6.0~8.0kg-m 13-0 13. REAR WHEEL/BRAKE/SUSPENSION 13 REAR WHEEL/BRAKE/SUSPENSION SERVICE INFORMATION...
How To Prepare and Give a Prefilled Syringe Injection
Diablo Valley Onc&Hem Med Grp, Inc Phone Number: (925)6775041 How To Prepare and Give a Prefilled Syringe Injection Patient Education Quick Reference Guide The following are instructions for use of prefilled
Record Cleaning Machine
Record Cleaning Machine Owner s Manual UK December 2010 Photo 1 Photo 2 IMPORTANT! READ THIS OWNER S MANUAL CAREFULLY, BEFORE USING THE MACHINE! Introduction You are now the proud owner of an Okki Nokki
