Portable Waterproof ph/dissolved Oxygen Meter
|
|
|
- Godfrey Hunter
- 10 years ago
- Views:
Transcription
1 Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON Series Portable Waterproof /Dissolved Oxygen Meter WTERPROOF REDY MES % T /Dissolved Oxygen/ / F Meter /DO 300 Series ON OFF HOLD MODE L MES ENTER IP 67
2 Table of ontents 1. Introduction Display and keypad functions Display Keypad Preparation Inserting the batteries Probe information onnecting the probes alibration Important information on meter calibration Preparing the meter for calibration calibration Preparing the meter for DO calibration Dissolved oxygen calibration in % saturation Dissolved oxygen calibration in mg/l (ppm) mode Temperature calibration Measurement Taking DO readings Taking pressure/salinity compensated DO measurements Taking readings Manual temperature compensation Hold function acklit LD dvanced functions dvanced setup mode overview measurement mode overview...29 % saturation measurement mode overview...30 mg/l (ppm) measurement mode overview mode Program P1.0: Viewing previous calibration data mode Program P2.0: Viewing electrode data P2.1 Electrode offset...33 P2.2 Electrode slope mode Program P3.0: measurement configuration P3.1 REDY indicator and auto endpoint function...35 P3.2 Selecting number of calibration points...36 P3.3 Selecting or F mode Program P4.0: Resetting to factory default settings % saturation mode Program P1.0: Dissolved oxygen parameters Pressure adjustment mode % saturation mode Program P2.0: Offset adjustment % saturation mode Program P3.0: Previous calibration information % saturation mode Program P4.0: Electrode properties % saturation mode Program P5.0: Unit configuration P5.1 REDY indicator and auto endpoint function...43 P5.2 Selecting or F temperature readout % saturation mode Program P6.0: djusting LD brightness % saturation mode Program P7.0: Resetting to factory default mg/l (ppm) mode Program P1.0: Dissolved oxygen parameters P1.1 Pressure adjustment mode...47 P1.2 Salinity adjustment mode mg/l (ppm) mode Program P2.0: Previous calibration information mg/l (ppm) mode Program P3.0: Electrode properties mg/l (ppm) mode Program P4.0: Unit configuration P4.1 REDY indicator and auto endpoint function...51 P4.2 Selecting mg/l or ppm units...52 P4.3 Selecting or F temperature readout mg/l (ppm) mode Program P5.0: djusting LD brightness mg/l (ppm) mode Program P6.0: Resetting to factory default Probe care and maintenance electrode care DO probe care Membrane replacement Electrolyte solution Troubleshooting Error Messages Specifications ccessories ppendix 1: Meter theory ppendix 2: Pressure vs. ltitude table ppendix 3: Factory default settings Warranty Return of Items
3 1. Introduction Thank you for selecting an OKTON meter. This OKTON portable meter is a microprocessor-based instrument that measures, dissolved oxygen and temperature. It s completely waterproof and it floats! Your meter has many user-friendly features, all of which are accessible through the membrane keypad. Your meter includes batteries. Please read this manual thoroughly before operating your meter. 2. Display and Keypad Functions 2.1 Display The LD has a primary and secondary display. The primary display shows the measured or dissolved oxygen reading. The secondary display shows the temperature of the reading. The display also shows error messages, keypad functions and program functions Primary display REDY MES WTERPROOF /Dissolved Oxygen/ / F Meter /DO 300 Series % T REDY HOLD ON OFF MES L % ppt ppm mg/l ERR F T ON OFF HOLD MODE Secondary display 9. L MES ENTER 1. mode indicator 2. MESurement mode indicator 3. Libration mode indicator 4. indicator 5. % saturation indicator 6. Salinity input indicator 7. Oxygen solubility indicators 8. Temperature indicators 9. utomatic Temperature ompensation indicator 10. ERRor indicator 11. Low battery indicator 12. Probe indicator 13. alibration data indicator 14. Function ON/OFF indicator 15. HOLD indicator 16. REDY indicator 4 5
4 2.2 Keypad The large membrane keypad makes the instrument easy to use. Each button, when pressed, has a corresponding graphic indicator on the LD. ON/OFF...Powers and shuts off the meter. HOLD...Freezes the measured reading. To activate, press HOLD while in measurement mode. To release, press HOLD again. NOTE: When auto endpoint feature is switched on, it automatically holds reading after 5 seconds of stability. The HOLD indicator appears on the display. Press HOLD to release. MODE...Selects the measurement parameter. Press MODE to toggle between ; DO % saturation; DO mg/l (ppm). In and DO mg/l (ppm) calibration modes press MODE to access temperature calibration. NOTE: To switch between mg/l and ppm modes, see page 52. L/MES...Toggles user between alibration and Measurement mode. If you were in mode, press L/MES to enter calibration mode. If you were in DO % saturation mode, press L/MES to enter DO % saturation calibration mode. If you were in DO mg/l (ppm) mode, press L/MES to enter DO mg/l (ppm) calibration mode. NOTE: Temperature calibration is available from both and DO mg/l (ppm) calibration modes; see page 19 for directions. In advanced set-up mode: Press L/MES to return to main menu from sub menus. Press L/MES again to return to measurement mode from main menu. ENTER...Press to confirm values in alibration mode and to confirm selections in mode....press in mode to scroll through subgroups. lso lets you increment/ decrement the values in the mg/l and temperature calibration modes....press to enter mode. mode lets you customize meter preferences and defaults, and view calibration and probe data....light Press to activate backlit display. ON OFF L MES HOLD MODE ENTER 3. Preparation 3.1 Inserting the atteries Four batteries are included with your meter. 1. Use a Phillips screwdriver to remove the two screws holding the battery cover. See figure below. 2. Lift off battery cover to expose batteries. 3. Insert batteries. Follow the diagram inside the cover for correct polarity. 4. Replace the battery cover into its original position. Screw cover back into place. Remove these two screws to access battery compartment attery compartment 6 7
5 3.2 Probe information Your meter uses two probes: electrode with N connector dissolved oxygen/temperature probe with a notched 6-pin connector The temperature sensing element built into the dissolved oxygen probe will also compensate for readings as long as both probes are in your solution at the same time. You can use any standard electrode with a N connector with this meter. If you do not have the DO probe connected you will need temperature probe to provide temperature compensation for readings (not included order separately). For the dissolved oxygen/temperature probe, you must use an OKTON dissolved oxygen/temperature probe with the notched 6-pin connector. For replacement probes, see the ccessories section, pages NOTE: Keep connectors dry and clean. Do not touch connectors with soiled hands. To connect the electrode: 1. Slide the N connector of the probe over the N connector socket on the meter. Make sure the slots of the connector are in line with the posts of the socket. Rotate and push the connector clockwise until it locks. See figure below. 2. To remove probe, push and rotate the connector counterclockwise. While holding onto the metal part of the connector, pull probe away from the meter. UTION: Do not pull on the probe cord or the probe wires might disconnect. To connect the dissolved oxygen probe: NOTE: Do not substitute other DO probes. For a replacement DO probe, see the ccessories section, page Line up the notch and 6 pins on the meter with the holes in the 6-pin connector. Push down and turn the locking ring clockwise to lock into place. See figure below. 2. To remove probe, turn the locking ring counterclockwise on the probe connector. Pull probe away from the meter. UTION: Do not pull on the probe cord or the probe wires might disconnect. Electrode (with N connector) DO probe (with 6-pin connector) locking ring top view of 6-pin connector of DO probe 8 9
6 4. alibration Important Information on Meter alibration 4.1 When you recalibrate your meter, old calibration points are replaced on a point by point basis in and completely for % saturation and mg/l (ppm). For example: : if you previously calibrated your meter at 4.01, 7.00, and 10.01, and you recalibrate at 7.00, the meter retains the old calibration data at 4.01 and % saturation and mg/l (ppm): ll new calibrations automatically override existing calibration data. To view current calibration points: : Program P1.0 in the section, page 32. % saturation: Program P3.0 in the section, page 41. mg/l (ppm): Program P2.0 in the section, page 49. To completely recalibrate your meter, or when you use replacement probes, it is best to clear old calibration data by resetting the meter. To reset the meter to its factory defaults: : Program P4.0 in the section, page 38. % saturation: Program P7.0 in the section, page 46. mg/l (ppm): Program P6.0 in the section, page 55. NOTE: Resetting the meter will set meter to factory defaults., % saturation and mg/l (ppm) must be reset separately. 4.2 Preparing the Meter for alibration efore starting calibration, make sure you are in the correct measurement mode. When you switch on the meter, the meter starts up in the units last used. For example, if you shut the meter off in "" units, the meter will read "" units when you switch the meter on. Do not reuse calibration solutions after calibration. ontaminants in the solution can affect the calibration, and eventually the accuracy of the measurements. See pages for information on our high-quality calibration solutions. e sure to remove the protective rubber cap from the electrode before calibration or measurement. If the electrode has been stored dry, wet the probe in tap water for 10 minutes before calibrating or taking readings to saturate the electrode surface and minimize drift. Wash your probe in deionized water after use, and store in electrode storage solution. If storage solution is not available, use 4.0 or 7.0 buffer. The /DO 300 meter is capable of up to 5-point calibration at 1.68, 4.01, 7.00, 10.01, and to ensure accuracy across the entire range of the meter. The meter automatically recognizes and calibrates to these standard buffer values, which makes calibration faster and easier. To calibrate at 5 points you must select a 5-pt calibration in mode program P3.2 on page 36. You can select 2, 3, 4 or 5 calibration points for calibration. For directions on how to calibrate your meter: See sections on pages for calibration See section on page for % saturation calibration See section 4.4 and 4.6 on page 14, for mg/l (ppm) calibration See section 4.7 on pages 19 for Temperature alibration 10 11
7 4.3 calibration NOTE: We recommend that you perform at least a 2-point calibration using standard buffers that bracket (one above and one below) the expected sample range. You can perform a 1-point calibration, but make sure that the buffer value is close to the sample value you are measuring. The temperature sensing element is in the DO probe. For temperature compensated readings and calibration, immerse the DO probe into the calibration buffer while calibrating. alibration 1. If necessary, press the MODE key to select mode. The indicator appears in the upper right hand corner of the display. MES 6. Press ENTER to confirm calibration. The meter is now calibrated to the current buffer. See figure D If you are performing multipoint calibration, go to step 7. If you are performing one-point calibration, go to step Rinse the probe with de-ionized water or a rinse solution, and place it in the next buffer. 8. Follow steps 5 to 8 for additional calibration points. 9. When calibration is complete, press L/MES to return to measurement mode. D REDY L See figure 2. Rinse the electrode and DO probe thoroughly with de-ionized water or a rinse solution. Do not wipe the probe; this causes a build-up of electrostatic charge on the glass surface. 3. Dip both probes into the calibration buffer. The end of both probes must be completely immersed into the sample. Stir the probe gently to create a homogeneous sample. 4. Press L/MES to enter calibration mode. The L indicator lights. The primary display will show the measured reading while the smaller secondary display will indicate the standard buffer solution. L T Notes To exit from alibration mode without confirming calibration, DO NOT press ENTER in step 6. Press L/MES instead. If the selected buffer value is not within ±1.0 from the measured value: the electrode and buffer icon blink and the ERR annunciator appears in the lower left corner of the display. To limit the number of buffer values available during calibration, see section P3.2 on page 36. See figure 5. Wait for the measured value to stabilize. REDY L See figure If the REDY indicator has been activated (set up program P3.1 see page 35), the REDY annunciator lights when the reading is stable
8 4.4 Preparing the meter for DO calibration efore calibrating your meter, make sure to rinse the DO probe well with de-ionized water or rinse solution. NOTE: Do not let membrane surface of the DO probe touch any other surface. The DO probe guard (the piece with holes fitted over the end of the probe) protects the membrane; make sure this is always attached to the probe while it is in use. You can calibrate for dissolved oxygen in either % saturation or mg/l (ppm). ll new DO calibrations automatically override existing DO calibration data. efore Dissolved Oxygen calibration efore starting DO calibration, make sure you are in the correct measurement mode. When you switch on the meter, the meter starts up in the units last used (either, mg/l, ppm, % saturation). For example, if you shut the meter off in mg/l units, the meter will read mg/l units when you switch the meter on. NOTE: Most users will calibrate to 100% saturation even when working in mg/l. When calibrating the meter in mg/l mode, you can make fine adjustments, typically to a midrange value between 10 and 100%. If you are calibrating to a midrange value, you need to set the 100% saturation value first. DO Probe guard 4.5 Dissolved Oxygen calibration in percent saturation You can calibrate this meter quickly and easily in air. The exact calibration value depends on barometric pressure. The meter is set to a factory default of 760 mm Hg, which results in a calibration value of 100% saturation in air. NOTE: If the barometric pressure setting has been changed from 760 mm Hg, the calibration value in air will automatically adjust to a value other than 100%. The adjusted value will be correct for the new barometric pressure setting. See page 39 to change the pressure setting. See ppendix 2 on page 67 for a chart of adjusted % saturation values. To calibrate 100 % saturation: 1. Rinse the probe well with deionized rinse water or rinse solution. For best accuracy, wrap the end of the probe in a damp cloth. Do not touch the membrane. 2. Press the MODE key to select the % saturation mode. See figure 3. Press the L key. The L indicator will appear above the primary display. The primary display shows the current value of the measurement and the secondary display will show (see NOTE above). See figure 4. Hold the probe in the air (or in damp cloth). Wait for the reading to stabilize. If the Ready indicator feature is enabled, it will appear when the reading is stable (see page 43 for Ready indicator setup). 5. Press the ENTER key. The meter automatically calibrates to 100.0% air saturation and returns to Measurement mode. See figure REDY REDY REDY MES MES L % T % % T 14 Notes The reading in the primary display in step 3 must read at 50% or above for the calibration to work correctly. Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display. You can offset your % DO calibration: see page 40 for directions. 15
9 To calibrate 0 % saturation: 4.6 Dissolved Oxygen calibration in mg/l (ppm) mode 1. Rinse the probe well with deionized rinse water or rinse solution. Do not touch the membrane. 2. Press the MODE key to select the % saturation mode. See figure D 2. Place the probe in zero oxygen solution. Stir the probe gently to homogenize the sample. Make sure that the sample is continuously flowing past the membrane sensor. 3. Press the L key. The L indicator will appear above the primary display. The primary display shows the current value of the measurement and the secondary display will show 0.0. See figure E 4. Wait for the reading to stabilize. If the Ready indicator feature is enabled, it will appear when the reading is stable (see page 43 for Ready indicator setup). 5. Press the ENTER key. The meter automatically calibrates to 0.0% saturation and returns to Measurement mode. See figure F Notes The reading in the primary display in step 3 must read at 10% or under for the calibration to work correctly. Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display. You can offset your % DO calibration: see page 40 for directions. D E F REDY REDY REDY MES MES L % T % % T alibrating the meter to 100 % in saturation mode will also calibrate the mg/l mode at the value in mg/l corresponding to 100% saturation. This should produce acceptable results in most applications. This meter also lets you make a calibration adjustment in mg/l (ppm) mode without affecting your % saturation calibration. To select between mg/l and ppm units, see page 52. To calibrate in mg/l (ppm) mode: 1. alibrate 100% saturation per section 4.5 on page Rinse the probe well with deionized rinse water or rinse solution. Wipe the outside of the probe carefully (do not touch the membrane). 3. Dip the probe into a sample of known oxygen concentration. Wait for the reading to stabilize. If the Ready indicator feature is enabled, it will appear when the reading is stable (see page 43 for Ready indicator setup). NOTE: The sample must continuously flow past the membrane for accurate readings. stirrer can keep the solution agitated; order a magnetic stirrer on page Press the MODE key to select the mg/l (ppm) mode. See figure 5. Press the L key. The L indicator will appear above the primary display. The primary display shows the current value of the measurement and the secondary display shows the temperature. See figure 6. Press the and keys to adjust the reading. REDY REDY REDY MES L L mg/l T mg/l T 7. Press the ENTER key. The meter automatically calibrates to the value you entered and returns to Measurement mode. mg/l T 16 See figure 17
10 Notes on mg/l (ppm) mode calibration During mg/l (ppm) calibration, the meter adjusts to the barometric pressure value that is programmed into the meter. It also adjusts to the salinity value that is programmed into the meter. You can change the barometric pressure value and salinity value in the mg/l (ppm) mode (see pages for directions). Whenever an error occurs during calibration, the ERR indicator appears in the lower left hand corner of the display. 4.7 Temperature alibration The built-in temperature sensor included in the DO probe is factory calibrated. alibrate your sensor only if you suspect sensor drift that may have occurred over a long period of time or if you have a replacement DO probe. Make sure DO probe is attached before you perform the temperature calibration. 1. Switch the meter on. Press MODE to select either or mg/l (ppm) Measurement mode. 2. Press the L/MES key to enter or mg/l (ppm) calibration mode. The L indicator will appear above the primary display. See figure 3. While in or mg/l (ppm) calibration mode, press the MODE key to enter temperature calibration mode. The primary display shows the last set temperature reading and the secondary display shows you the factory default temperature value. See figure 4. ompare the primary display reading to a NIST-traceable thermometer or another thermometer known to be accurate. 5. Press the or keys to adjust the primary display reading to agree with your temperature standard. See figure 6. Press the ENTER key to confirm temperature calibration and return to Measurement mode. REDY REDY REDY L L L mg/l T T T 18 Notes To exit from Temperature alibration mode without confirming calibration, DO NOT press ENTER in step 6. Press L/MES instead. Temperature calibration is restricted to ±5 from the factory default value displayed during calibration (shown in the secondary display). 19
11 5. Measurement 5.1 Taking DO readings During measurement, the probe can be: fully immersed in the solution partially immersed in the solution Do not allow the probe s membrane surface to touch anything! The probe guard (the piece with holes fitted over the end of the probe) protects the membrane; you Probe guard should leave this piece attached to the probe at all times. IMPORTNT: since the DO probe consumes oxygen from the sample, the sample must constantly flow past the membrane to achieve more accurate readings. You can use a stirrer to accomplish this (see ccessories on page 65 for a magnetic stirrer). To take measurements: 1. Rinse the probe well with deionized rinse water or rinse solution. 2. Select the appropriate measurement mode. Press the MODE key to toggle between modes: % saturation. mg/l (ppm)**, 3. Dip the probe into the sample. Stir the probe gently to homogenize the sample. Make sure that the sample is continuously flowing past the membrane sensor. 4. Note the reading on the display. If the REDY indicator is switched on, it will appear when the reading is stable. See figure REDY MES **To select between mg/l and ppm units, see page 52. The T indicator should appear in the lower right hand corner of the display. If it does not, this indicates an error. ONTINUED ON NEXT PGE mg/l T Taking measurements with REDY indicator selected on If the REDY indicator has been activated, the REDY annunciator lights when the reading is stable. Switch the REDY indicator on or off in the program P5.1 for % saturation measurements (page 43) or program P4.1 (page 51) for mg/l (ppm) measurements. Taking measurements with the auto endpoint feature selected on When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically hold the reading. The hold indicator appears on the left side of the display. Press the HOLD key to release the reading. Switch the uto endpoint feature on or off in the program P5.1 for % saturation measurements (page 43) or program P4.1 (page 51) for mg/l (ppm) measurements. 5.2 Taking pressure/salinity compensated DO measurements If necessary, you can adjust the pressure and salinity values of your measurements in the mode. The DO meter will automatically compensate for salinity and pressure based on the values entered in the setup functions. The meter is factory set at 760 mm Hg (101.3 Pascals) pressure adjustment and a factor of 0.0 salinity adjustment. See ppendix 2 on page 67 for a Pressure vs ltitude table. Pressure adjustment NOTE: This adjustment appears in both the % saturation and mg/l (ppm) measurement modes. 1. Press the MODE key to select % saturation or mg/l (ppm) mode. Press the key to enter Set Up mode. See figure 2. Press the ENTER key until the upper display shows a number and the lower display shows Hg or P. 3. Use the and keys to enter the barometric pressure. The upper display will show the value you have entered. See figure 4. Press ENTER to confirm pressure value. 5. Press L/MES to return to measurement mode, or continue with step 4 on page 22 to make a salinity adjustment [available from mg/l (ppm) mode only]
12 Salinity adjustment NOTE: this mode appears in mg/l (ppm) measurement mode only. 1. Press the Mode key to select mg/l (ppm) mode. 2. Press the key to enter Set Up mode. See figure 4. Press the ENTER key until the upper display shows a number and the lower display shows SL. 5. Use the and keys to enter the salinity of your solution in ppt. The upper display will show the value you have entered. See figure 6. Press the ENTER key to confirm value. 7. Press L/MES to return to Measurement mode. ppt 5.3 Taking Measurements e sure to remove the protective rubber cap on the electrode before measurement. To ensure utomatic Temperature ompensation (T) for measurements, the DO probe must be plugged in and must be immersed into the sample you wish to take readings. To take readings: 1. Rinse the electrode with deionized or distilled water before use to remove any impurities adhering to the probe body. If the electrode has dehydrated, soak it for 30 minutes in OKTON electrode storage solution or a 2M 4M Kl solution. 2. Press ON to switch on meter. Press MODE key to select measurement mode. The MES annunciator appears on the top center of the LD. The T indicator appears in the lower right hand corner to indicate utomatic Temperature ompensation. If the T indicator does not appear, make sure DO probe is plugged into the meter. See figure MES T 3. Dip both the electrode and the DO probe into the sample. When dipping the electrode into the sample, the sensor or the glass bulb of the electrode must be completely immersed into the sample. Stir the probe gently in the sample to create a homogenous sample. 4. llow time for the reading to stabilize. Note the reading on the display. 5. Press the MODE key to toggle between modes: % saturation. mg/l (ppm)** **To select between mg/l and ppm units, see page 52. The T indicator should appear in the lower right hand corner of the display. If it does not, this indicates an error. ONTINUED ON NEXT PGE 22 23
13 Taking measurements with REDY indicator selected on If the REDY indicator has been activated, the REDY annunciator lights when the reading is stable. Switch the REDY indicator on or off in Set up program P3.1 see page 35 for directions. Taking measurements with the auto endpoint feature selected on When a reading is stable for more than 5 seconds, the auto endpoint feature will automatically HOLD the reading. The HOLD indicator appears on the left side of the display. Press the HOLD key to release the reading. Switch the uto endpoint feature on or off in Set up program P3.1 see page 35 for directions. 5.4 IMPORTNT: For manual compensation, you must disconnect the DO or temperature probe (see page 9). 1. Switch the meter on. Press the MODE key to select mode. 2. Press the L/MES key to enter calibration mode. The L indicator will appear above the primary display. 3. While in calibration mode, press the MODE key to enter temperature calibration mode. The primary display shows the current temperature setting and the secondary display shows the default value 25. See figure 4. heck the temperature of your sample using an accurate thermometer. 5. Press the or keys to offset the temperature to the measured value from step 4. See figure 6. Press ENTER to confirm the selected temperature and to return to the measurement mode. The meter will now compensate readings for the manually set temperature. Notes Manual Temperature ompensation To exit this program without confirming the manual temperature compensation value, DO NOT press ENTER in step 6. Press L/MES instead. L L 24 25
14 6. HOLD function 7. acklit LD This feature lets you freeze the value of the, % saturation or mg/l (ppm) and temperature readings for a delayed observation. HOLD can be used any time when in MES mode. 1. To hold a measurement, press the HOLD key while in measurement mode. The HOLD annunciator will appear on the display. See figures and 2. To release the held value, press HOLD again. ontinue to take measurements. HOLD MES T Using the backlit LD This feature lets you view readings in low light situations. 1. To activate the backlit LD feature, press the key to activate. See figure To keep the LD backlit, keep the key pressed. If the key is not pressed, the backlight automatically shuts off automatically after 20 seconds after pressing the key. MES ENT NOTE: This meter shuts off automatically after 20 minutes of nonuse. If the meter is shut off either automatically or manually, the HOLD value will be lost. NOTE: This meter has an auto endpoint feature. When this feature is switched on, the display will automatically HOLD a reading that has been stable for more than 5 seconds. The HOLD indicator appears. Press the HOLD key to release the reading. To switch on or off the auto endpoint feature, see the program P3.1 on page 35 for, P5.1 on page 43 for % saturation and P4.1 on page 51 for mg/l (ppm). HOLD MES % T djusting LD brightness This mode lets you adjust the brightness of the backlit LD. Selecting a dimmer backlighting level helps conserve batteries. This adjustment is only available in the % saturation or mg/l (ppm) modes. 1. Press the MODE key to switch to % saturation or mg/l (ppm) mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter Ld in the upper display. See figure 4. Press the ENTER key. number (0-8) appears in the upper display and Ld appears in the lower display. See figure 5. Press the and keys to select from level 0 (dimmest light) to level 8 (brightest light) Press the ENTER key to confirm selection and to return to the subgroup menu. Press the L/MES key to return to measurement mode. 27
15 8. dvanced functions 8.1 dvanced mode detailed overview The advanced set up mode lets you customize your meter s preferences and defaults. Your OKTON waterproof meter features different sub groups that organize all parameters. This meter displays different subgroups depending on which mode the meter is in: % Saturation mg/l (ppm) Press the key to enter mode. Press the and keys to scroll through sub groups. From the measurement mode L: Viewing previous calibration data View previous calibration data L: alibration data dpr: Dissolved oxygen parameters dpr: Dissolved oxygen parameters Instructions on page 32 ELE: Viewing electrode data View electrode offset View electrode slope ELE: Probe data OF: onfiguration % saturation offset adjustment L: alibration data L: alibration data ELE: Probe data Instructions on page Instructions on page OF: Unit configuration Ready indicator on or off / auto endpoint on or off number of calibration points: 2, 3, 4, or 5 Select F or rst: Reset to factory default Reset to factory default settings rst: Reset meter to factory default ELE: Probe data OF: onfiguration Instructions on page 38 OF: onfiguration Ld: acklit display Ld: acklit display rst: Reset meter to factory default 28 rst: Reset meter to factory default 29
16 From the % saturation measurement mode From the mg/l (ppm) measurement mode dpr: Dissolved oxygen parameters dpr: Dissolved oxygen parameters Instructions on page 39 Select Hg or Pa barometric pressure units Select barometric pressure OFS: % saturation offset adjustment Instructions on page Select Hg or Pa barometric pressure units Select barometric pressure Select salinity adjustment factor L: Viewing previous calibration data Set % saturation offset adjustment View previous calibration data, including date/time Instructions on page 40 L: Viewing previous calibration data Instructions on page 49 ELE: Viewing probe data Instructions on page 41 Instructions on page 42 View previous calibration data ELE: Viewing probe data View probe slope View % saturation offset View mv value equivalent to 100% saturation View mv value equivalent to 0% saturation Instructions on page 50 Instructions on page View probe slope View mv value equivalent to 100% saturation View mv value equivalent to 0% saturation OF: Unit configuration Ready indicator on or off / auto endpoint on or off Select mg/l or ppm units Select F or OF: Unit configuration Ld: acklit display Instructions on page Ready indicator on or off / auto endpoint on or off Select F or Instructions on page 54 djust brightness of backlit LD Ld: acklit display rst: Reset to factory default djust brightness of backlit LD Reset to factory default settings Instructions on page 45 Instructions on page 55 rst: Reset to factory default Reset to factory default settings Instructions on page
17 8.2 mode P1.0: Viewing previous calibration data Program P1.0 lets you recall previous calibration data, which helps you know when to recalibrate your meter. This is a view only mode. 8.3 mode P2.0: Viewing electrode data Program P2.0 has two view only options that lets you check the electrode parameters for diagnostic purposes. 1. Press the MODE key to select measurement mode. 2. Press the key to enter mode. See figure 3. Press the and keys to scroll through subgroups until you view parameter P Press the ENTER key repeatedly to view previous calibration data. P1.1 = 1.68 P1.2 = 4.01 P1.3 = 7.00 P1.4 = P1.5 = See figure 5. When you have scrolled through all calibration data, you will automatically return to the subgroup menu. Press the L/MES key to return to measurement mode. P2.1: Electrode offset 1. Press the MODE key to select measurement mode. 2. Press the key to enter Set Up mode. 3. Press the and keys to scroll through subgroups until you view parameter P2.0. See figure 4. Press the ENTER key to select parameter P The display shows the electrode offset value. the display shows the mv offset at If you have not calibrated at any buffer, the primary display shows 0.00 mv. See figure 6. Press the ENTER key to to proceed to step 4 of P2.2. Or, press the L/MES key to return to measurement mode. mv NOTES: If there is no previous calibration data at a particular point, the primary display will show See figure 32 33
18 P2.2: Electrode slope NOTE: this parameter is available in measurement mode only. 8.4 mode P3.0: measurement configuration P3.1: REDY indicator and auto endpoint function This program lets you select: 1. If necessary, press the MODE key to select measurement mode. Press the key to enter mode. 2. Press the and keys to scroll through subgroups until you view parameter P2.0. See figure D 3. Press the ENTER key twice to select parameter P The display shows electrode slope in percentage of the range you are currently measuring. Default setting is See figure E 5. Press the ENTER key to return to the subgroup menu. Press the L/MES key to return to measurement mode. D E % REDY indicator on to indicate when the reading is stable. REDY indicator off for faster meter response. uto endpoint function on. Select auto endpoint on to hold the reading when it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD key to release the display and access other functions. 1. Press the MODE key to select measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P3.0. See figure 4. Press the ENTER key to select parameter P3.1. See figure REDY ON 5. Press the and keys to select the configuration you require. OFF switches the REDY indicator off. ON switches the REDY indicator on. ON and HOLD together switches the auto endpoint feature on. 6. Press the ENTER key to confirm selection and to proceed to step 4 of P3.2. Press the L/MES key to return to measurement mode. NOTES: Meter default is set for Ready indicator on, and auto endpoint function off. REDY OFF REDY HOLD ON You can also change the Ready indicator and auto endpoint function in Program P5.1 on page 43 in % saturation mode or Program P4.1 on page 51 in mg/l (ppm) mode. ny changes you make to the Ready indicator/ auto endpoint function in mode will also change in % saturation and mg/l (ppm) modes
19 P3.2: Selecting number of calibration points Program P3.2 lets you select the number of calibration points the meter will use in calibration mode: 2, 3, 4, or 5. The meter will automatically exit calibration mode after you have calibrated to your selected number of points. 1. Press the MODE key to select measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P3.0. See figure 4. Press the ENTER key twice to select parameter P3.2. See figure D 5. Press the and keys to select 2, 3, 4, or 5 point calibration. 6. Press the ENTER key to confirm selection and to return to the subgroup menu. Press the L/MES key to return to measurement mode. D P3.3 Selecting or F This meter lets you select between and F units for temperature readings. 1. Press the MODE key to select measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P3.0. See figure E 4. Press the ENTER key three times to select parameter P3.3. See figure F 5. Press the and keys to toggle between and F. 6. Press the ENTER key to confirm selection and to return to the subgroup menu. Press the L/MES key to return to measurement mode. E F F NOTES: You can also switch between and F in Program P5.2 on page 44 in % saturation mode or Program P4.3 on page 53 in mg/l (ppm) mode. If you switch between and F in mode, the meter will also switch in % saturation and mg/l (ppm) modes
20 8.5 mode P4.0: Resetting to factory default settings This program lets you reset all parameters to factory default settings. This clears all calibration data any other setup functions you might have changed. The following settings will remain as you have set them: Temperature unit of measure ( or F) The temperature offset calibration value ll dissolved oxygen calibration data and parameters including the LD contrast setting 1. Press the MODE key to select measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P4.0 in the lower display. See figure 4. Press the ENTER key. See figure 5. Press the and keys to toggle between NO and YES. NO retains current settings YES resets to factory default settings 6. Press the ENTER key to confirm selection and to return to measurement mode. Notes: To reset all % saturation data, see page 46. To reset all mg/l (ppm) data, see page 55. See page 68 for a table of factory default settings. MES MES 8.6 % saturation mode P1.0: Dissolved Oxygen Parameters This sub group lets you adjust the barometric pressure Pressure adjustment mode arometric pressure is vital to correct dissolved oxygen measurements. You need to enter the correct barometric pressure of the area you are measuring. This mode lets you perform two functions: Select either mm Hg or Pascal barometric pressure units djust the barometric pressure. See ppendix 2 on page 67 for a Pressure vs ltitude table. 1. Press the MODE key to select % saturation mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P1.0 in the lower display. See figure 4. Press the ENTER key twice. The upper display shows either Hg or P and the lower display shows br. 5. Press the and keys to toggle between mm Hg and Pascal units. See figure 6. Press the ENTER key to confirm selection and move to the next screen. The upper display shows the barometric pressure and the lower display shows the units selected in step 5. See figure 7. Press the and keys to adjust the barometric pressure. The pressure adjustment range is 500 to 1499 mm Hg (66.6 to kpa). 8. Press the ENTER key to confirm selection and move to step 3 on page 40. If instead you want to return to measurement mode, press L/MES twice
21 8.7 % saturation mode P2.0: Offset djustment Use the offset adjustment to act as an offset at values between 0 and 100%. It is useful to match against end-user standards (i.e. a test kit value). 8.8 % saturation mode P3.0: Previous calibration information This sub group shows you the previous calibration data, along with date and time of calibration. This is a view only parameter. In % saturation mode: calibration information is shown in % saturation units. 1. Press the MODE key to select % saturation mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P2.0 in the lower display. See figure 4. Press the ENTER key. The upper display shows the current measurement in % saturation and the lower shows OFS. See figure 5. Press the and keys to offset the % saturation measurement. 6. Press the ENTER key to confirm selection and to move back to subgroup P2.0. If you want to return to measurement mode, press L/MES. REDY MES % 1. Press the MODE key to select % saturation mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P3.0 in the lower display. See figure 4. Press the ENTER key. The upper display shows the calibration data. See figure 6. Press the ENTER key to move back to subgroup P3.0. If you want to return to measurement mode, press L/MES. L L % % NOTE: If you did not calibrate this meter in a particular mode, the screen will show. See figure 40 41
22 1. Press the MODE key to select the % saturation measurement mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P4.0 in the lower display. See figure 4. Press the ENTER key. The upper display shows the probe slope. See figure 5. Press the ENTER key. The upper display shows the % saturation offset. See directions for setting this offset on page 40. See figure 6. Press the ENTER key. The upper display shows the 100 % saturation mv value. See figure D 7. Press the ENTER key. The upper display shows the 0 % saturation mv value. See figure E 8. Press the ENTER key to move back to subgroup P4.0. If you want to return to measurement mode, press L/MES % saturation mode P4.0: Electrode properties These view only parameters show the electrode properties for diagnostic purposes: probe slope: shows the calibration slope of the probe. This mode displays slope from 0.5 to % of slope (1.0 = 100%). % saturation offset: shows the value of the % saturation offset entered in parameter P2.0 (see page 40 for instructions). 100% saturation mv value: shows the sensor s mv output corresponding to 100% saturation. 0% saturation mv value: shows the sensor s millivolt output corresponding to 0% saturation. D E K= % % mv mv 8.10 % saturation mode P5.0: Unit configuration Unit configuration mode lets you select the following parameters:. Ready indicator and auto endpoint function. Temperature in or F P5.1: REDY indicator and auto endpoint function The first program lets you select: REDY indicator on to indicate when the reading is stable. REDY indicator off for faster meter response. uto endpoint function on. Select auto endpoint on to hold the reading when it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD key to release the display and access other functions. 1. Press the MODE key to select the % saturation measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P5.0 in the lower display. See figure 4. Press the ENTER key to select parameter P5.1 rdy (Ready). See figure 5. Press the and keys to select the configuration you require. ON switches the REDY indicator on. OFF switches the REDY indicator off. ON and HOLD together switches the auto endpoint feature on. 6. Press the ENTER key to confirm selection and to move to step 3 on page 44 (P5.2). Or, press the L/MES key twice to return to measurement mode. NOTES: Meter default is set for Ready indicator on, and auto endpoint function off. You can also change the Ready indicator and auto endpoint function in Program P3.1 on page 35 in mode or P4.1 on page 51 in mg/l (ppm) mode. ny changes you make to the Ready indicator/ auto endpoint function in % saturation mode will also change in and mg/l (ppm) modes. REDY ON REDY OFF REDY HOLD ON 43
23 P5.2: Selecting or F temperature readout This meter lets you select between and F units for temperature readings. 1. Press the MODE key to select the % saturation measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P5.0 in the lower display. See figure E 4. Press the ENTER key until or F appears in the upper display. See figure F 5. Press the and keys to toggle between and F. 6. Press the ENTER key to confirm selection and to return to the subgroup menu. Press the L/MES key to return to measurement mode. E F F 8.11 % saturation mode P6.0: djusting LD brightness This mode lets you adjust the brightness of the backlit LD. Selecting a dimmer backlighting level helps conserve batteries. 1. Press the MODE key to select the % saturation measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P6.0 in the lower display. See figure 4. Press the ENTER key. number (0-8) appears in the upper display and Ld appears in the lower display. See figure 5. Press the and keys to select from level 0 (dimmest light) to level 8 (brightest light). 6. Press the ENTER key to confirm selection and to return to the subgroup menu. Press the L/MES key to return to measurement mode. NOTES: You can also switch between and F in Program P3.3 on page 37 in mode or Program P4.3 on page 53 in mg/l (ppm) mode. If you switch between and F in % saturation mode, the meter will also switch in and mg/l (ppm) modes
24 % saturation mode P7.0: Resetting to factory default settings This program lets you reset all parameters to factory default settings. This clears all calibration data, memory, and any other setup functions you might have changed. The LD contrast setting will be reset to 8 (the maximum setting). However, the following settings will remain as you have set them: temperature unit of measure ( or F) the temperature offset calibration value calibration data and set parameters 1. Press the MODE key to select the % saturation measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P7.0 in the lower display. See figure 4. Press the ENTER key. See figure 5. Press the and keys to toggle between NO and YES. NO retains current settings YES resets to factory default settings 6. Press the ENTER key to confirm selection and to return to measurement mode. Notes To reset all data, see page 38. To reset all mg/l (ppm) data, see page 55. See page 68 for a table of factory default settings mg/l (ppm) mode P1.0: Dissolved Oxygen Parameters This sub group lets you adjust the barometric pressure and salinity. P1.1 Pressure adjustment mode arometric pressure is vital to correct dissolved oxygen measurements. You need to enter the correct barometric pressure of the area you are measuring. This mode lets you perform two functions: Select either mm Hg or Pascal barometric pressure units djust the barometric pressure. See ppendix 2 on page 67 for a Pressure vs ltitude table. 1. Press the MODE key to select mg/l (ppm) mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P1.0 in the lower display. See figure 4. Press the ENTER key twice. The upper display shows either Hg or P and the lower display shows br. 5. Press the and keys to toggle between mm Hg and Pascal units. See figure 6. Press the ENTER key to confirm selection and move to the next screen. The upper display shows the barometric pressure and the lower display shows the units selected in step 5. See figure 7. Press the and keys to adjust the barometric pressure. The pressure adjustment range is 500 to 1499 mm Hg (66.6 to kpa). 8. Press the ENTER key to confirm selection and move to P1.2 on page 48 (salinity adjustment). If instead you want to return to measurement mode, press L/MES twice. NOTE: Salinity adjustment appears in mg/l (ppm) measurement mode only. 47
25 P1.2 Salinity adjustment mode 8.14 mg/l (ppm) mode P2.0: Previous calibration information Salinity correction mode lets you correct for the variations in oxygen solubility due to salt concentration in the sample. 1. Press the MODE key to select mg/l (ppm) mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P1.0 in the lower display. See figure D 4. Press the ENTER key five times. The upper display shows the salinity value and the lower shows SL. See figure E 5. Press the and keys to enter the correct salinity adjustment factor. The salinity adjustment factor range is 0.0 to 50.0 ppt. 6. Press the ENTER key to confirm selection and to move back to subgroup P1.0. If you want to return to measurement mode, press L/MES. D E ppt This sub group shows you the previous calibration data, along with date and time of calibration. This is a view only parameter. In mg/l (ppm) mode: calibration information is shown in mg/l (ppm) units. 1. Press the MODE key to select mg/l (ppm) mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P2.0 in the lower display. See figure 4. Press the ENTER key. The upper display shows the calibration data. See figure 6. Press the ENTER key to move back to subgroup P2.0. If you want to return to measurement mode, press L/MES. L L mg/l mg/l NOTE: If you did not calibrate this meter in a particular mode, the screen will show. See figure 48 49
26 Press the MODE key to select the mg/l (ppm) measurement mode. 2. Press the key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P3.0 in the lower display. See figure 4. Press the ENTER key. The upper display shows the probe slope. See figure 6. Press the ENTER key. The upper display shows the 100 % saturation mv value. See figure 7. Press the ENTER key. The upper display shows the 0 % saturation mv value. See figure D 8. Press the ENTER key to move back to subgroup P3.0. If you want to return to measurement mode, press L/MES. 50 mg/l (ppm) mode P3.0: Electrode properties These view only parameters show the electrode properties for diagnostic purposes: probe slope: shows the calibration slope of the probe. This mode displays slope from 0.5 to % of slope (1.0 = 100%). 100% saturation mv value: shows the sensor s mv output corresponding to 100% saturation. 0% saturation mv value: shows the sensor s millivolt output corresponding to 0% saturation. D K= mg/l mv mv 8.16 mg/l (ppm) mode P4.0: Unit configuration Unit configuration mode lets you select the following parameters:. Ready indicator and auto endpoint function. mg/l or ppm units. Temperature in or F P4.1: REDY indicator and auto endpoint function The first program lets you select: REDY indicator on to indicate when the reading is stable. REDY indicator off for faster meter response. uto endpoint function on. Select auto endpoint on to hold the reading when it is stable for more than 5 seconds. The display automatically freezes, and the HOLD indicator appears on the left side of the display. Press the HOLD key to release the display and access other functions. 1. Press the MODE key to select the mg/l (ppm) measurement mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P4.0 in the lower display. See figure 4. Press the ENTER key to select parameter P4.0 rdy (Ready). See figure 5. Press the and keys to select the configuration you require. ON switches the REDY indicator on. OFF switches the REDY indicator off. ON and HOLD together switches the auto endpoint feature on. 6. Press the ENTER key to confirm selection and to proceed to P4.2 on page 52. Or, press the L/MES key twice to return to measurement mode. Notes: Meter default is set for Ready indicator on, and auto endpoint function off. You can also change the Ready indicator and auto endpoint function in Program P3.1 on page 35 in mode or P5.1 on page 43 in % saturation mode. ny changes you make to the Ready indicator/ auto endpoint function in % saturation mode will also change in and % saturation modes. REDY ON REDY OFF REDY HOLD ON 51
27 P4.2: Selecting mg/l or ppm units This mode lets you select between mg/l or ppm dissolved oxygen units. P4.3: Selecting or F temperature readout This meter lets you select between and F units for temperature readings. 1. Press the MODE key to select mg/l (ppm) mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P4.0 in the lower display. See figure 4. Press the ENTER key until do appears in the upper display, P4.2 in the lower display. See figure D 5. Press the and keys to toggle between mg/l or ppm units. 6. Press the ENTER key to confirm selection and move to P4.3 on page 53. Press the L/MES key twice to return to measurement mode. D mg/l ppm 1. Press the MODE key to select mg/l (ppm) mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P4.0 in the lower display. See figure E 4. Press the ENTER key until or F appears in the upper display. See figure F 5. Press the and keys to toggle between and F. 6. Press the ENTER key to confirm selection and to return to the subgroup menu. Press the L/MES key to return to measurement mode. E F F NOTES: You can also switch between and F in Program P3.3 on page 37 in mode or Program P5.2 on page 44 in mg/l (ppm) mode. If you switch between and F in mg/l (ppm) mode, the meter will also switch in and % saturation modes
28 8.17 mg/l (ppm) mode P5.0: djusting LD brightness This mode lets you adjust the brightness of the backlit LD. Selecting a dimmer backlighting level helps conserve batteries. 1. Press the MODE key to select mg/l (ppm) mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P5.0 in the lower display. See figure 4. Press the ENTER key. number (0-8) appears in the upper display and Ld appears in the lower display. See figure 6. Press the and keys to select from level 0 (dimmest light) to level 8 (brightest light). 6. Press the ENTER key to confirm selection and to return to the subgroup menu. Press the L/MES key to return to measurement mode mg/l (ppm) mode P6.0: Resetting to factory default settings This program lets you reset all parameters to factory default settings. This clears all calibration data, memory, and any other setup functions you might have changed. The LD contrast setting will be reset to 8 (the maximum setting). However, the following settings will remain as you have set them: temperature unit of measure ( or F) the temperature offset calibration value calibration data and set parameters 1. Press the MODE key to select mg/l (ppm) mode. 2. Press key to enter mode. 3. Press the and keys to scroll through subgroups until you view parameter P6.0 in the lower display. See figure 4. Press the ENTER key. See figure 5. Press the and keys to toggle between NO and YES. NO retains current settings YES resets to factory default settings 6. Press the ENTER key to confirm selection and to return to measurement mode. NOTES: To reset all data, see page 38. To reset all % saturation data, see page 46. See page 68 for a table of factory default settings
29 9. Probe are and Maintenance 9.1 Electrode are Since your electrode is susceptible to dirt and contamination, clean it every one to three months depending on the extent and condition of use. NOTE: for specialty electrode care, consult the instruction manual included with your electrode. Reactivating the electrode If stored and cleaned properly, your electrode should be ready for immediate use. However, a dehydrated bulb may cause sluggish response. To rehydrate the bulb, immerse the electrode in a 4 buffer solution for 10 to 30 minutes. If this fails, you probably need to replace the electrode. Never touch or rub glass bulb. ontact builds up an electrostatic charge. electrode storage For best results, always keep the bulb wet. Use the rubber cap filled with OKTON electrode storage solution to store your electrode (see page for ordering information). lso, you can store in a 4 buffer with 1/100 part of saturated Kl. Other buffers are OK for storage, but NEVER use distilled water for storage. fter measuring 1. Rinse the electrode and reference junction in de-ionized water. 2. Store the electrode as recommended above in electrode storage, or as recommended by the manufacturer. 3. Prior to next use, rinse the liquid junction with de-ionized water and tap dry never wipe electrode. NOTE: If this does not restore electrode to normal response, see Reactivating the electrode section on page 57. electrode cleaning Salt deposits: dissolve the deposits by immersing the electrode in tap water for ten to fifteen minutes. Ten thoroughly rinse with distilled water. Oil/grease film: wash electrode bulb gently in some detergent and water. Rinse electrode tip with distilled water or use a general purpose electrode cleaner (see page 65 for ordering information). logged reference junction: heat a diluted Kl solution to 60 to 80. Place the sensing part of the electrode into the heated solution for about 10 minutes. llow the electrode to cool in some unheated Kl solution. Protein deposits: prepare a 1% pepsin solution in 0.1 M of Hl. Set the electrode in the solution for five to ten minutes. Rinse the electrode with distilled water. 9.2 DO Probe are The probe is a galvanic measuring element which produces an output proportional to the oxygen present in the medium in which it is placed. The galvanic probe design lets you take measurements immediately without the typical 15 minute wait of other dissolved oxygen probes. The probe consists of two parts: an upper part consisting of an anode, a cathode, and cable a lower part consisting of a membrane cap, membrane, and electrolyte solution. Under typical operating conditions, the probe should last for several years. Proper care and maintenance will help you receive the maximum probe life and ensure more accurate readings. Since any deposits on the membrane surface act as a barrier to oxygen diffusing through the membrane, the membrane must be cleaned at regular intervals to assure maximum reliability. fter using the probe, rinse the probe in clean water and wipe it with a soft cloth or paper to avoid any hardening of deposits. If growth develops on the probe, use a disinfecting chemical to clean. NOTE: lthough the membrane is strong and not easily damaged, wipe it gently while cleaning it. If the membrane is damaged or torn, the probe will no longer function. There are no special probe storage requirements. node athode Membrane cap (holds membrane and electrolyte solution) Probe guard 56 57
30 9.3 DO Membrane Replacement Replacement of the membrane is required only when you cannot calibrate the probe, or if the membrane is damaged. Typical membrane damages are punctures or wrinkles caused during measurements or cleaning. To order replacement probe components or a replacement probe, see the ccessories section on page 64. To replace the probe membrane: 1. Pull off the probe guard. 2. Unscrew the membrane cap from the probe. 3. Hold the probe under a water tap and brush away the white oxide on the cylindrical anode with a stiff plastic brush do not use metal cleaning material. 4. It the cathode has any deposits, remove them with a light scouring powder. Do not polish the cathode. NOTE: If you have purchased a replacement membrane module with preinstalled membrane, skip to step 12. ONTINUED ON NEXT PGE Temperature sensor node athode Membrane lock Membrane O-ring Membrane cap Probe guard 5. Using the installation tool, unscrew and remove the membrane lock from the membrane cap. See figure 6. Remove the membrane and O-ring. Discard both. 7. Rinse the membrane cap and membrane lock in tap water. 8. Install a new O-ring inside the membrane cap. 9. Install a new membrane. Make sure the membrane covers the O-ring all around its circumference. See figure 10. Using the installation tool, screw the membrane lock back into the cap. Tighten the lock firmly over the membrane and O-ring, but do not overtighten. 11. Inspect the membrane for wrinkles. If wrinkles exist, remove the membrane and repeat steps Fill the membrane cap with water and inspect the bottom for leaks. If water drops are leaking from the membrane, re-seat the membrane on the O-ring (repeat steps 8-11). 13. If the assembly is leak-free, empty the water and fill the membrane cap with electrolyte to the brim. 14. Screw the cap onto the probe. Excess electrolyte will drain out. Installation tool Insert notches on installation tool into slots on membrane lock. Then, unscrew membrane lock from membrane cap. arefully place O-ring, then membrane, into membrane cap. 15. Replace probe guard. 16. alibrate the probe (see section on pages 14-18) after the % saturation readings have stabilized
31 9.4 DO Electrolyte solution 10. Troubleshooting The electrolyte solution in your probe s cap will periodically evaporate and need to be replaced replacement electrolyte solution comes premixed and ready to use. However, the electrolyte mixture package has a longer shelf life in its unmixed form. If you purchase the electrolyte mixture package to make your own replacement electrolyte solution, use the following steps to prepare the solution: 1. Fill a beaker to the 400 ml mark with deionized water. 2. Pour the entire contents of the 58.5 grams electrolyte package into the beaker. 3. Stir the solution until all of the chemical is dissolved until the solution is clear. 4. Pour the solution into a clean container with a cap and keep sealed between use. Problem ause Solution Power on but a) atteries not in place. a) heck that batteries are in no display place and making good contact. b) atteries not in correct b) Reinsert batteries with polarity (+ and ). correct polarity. c) Weak batteries. c) Replace batteries. Unstable a) Insufficient electrolyte a) Fill DO probe with reference readings in DO probe. electrolyte. See page b) ir bubbles in probe. b) Tap probe to remove bubbles. c) Dirty DO probe or c) lean the probe or electrode electrode. and recalibrate. d) Probe not deep d) Make sure sample entirely enough in sample. covers the probe sensors. e) External noise pickup e) Move or switch off or induction caused by interfering motor. nearby electric motor. f) roken DO probe or f) Replace probe or electrode. electrode See page 64. Slow response a) Dirty membrane/ a) lean probe. See "Probe Dirty electrode are & Maintenance", page Not responding a) HOLD mode in operation. a) ancel HOLD mode. to key press b) Damaged key pad. b) Return to dealer. c) Internal program error. c) Reset all internal programs by reinserting batteries
32 11. Error Messages 12. Specifications LD Indicates ause Solution Display Err Unrecognized input Wrong input in Release key. Select annunciator from keypad selected mode. valid operations depending on mode. L & Err alibration error Wrong value input heck your input value, annunciators at calibration. clean probe. blink Dirty probe. See alibration sections or Probe Maintenance section. attery Low battery level Need new batteries lean battery contacts. indicator blinks or battery Replace batteries with fresh connection is bad. ones, noting polarity. * See Warranty and Return of Items on page 67 If an error message appears in the primary display (the upper row of larger digits), switching off the meter and switching it on again may eliminate the error message. If error persists, or the meter shows incorrect values, return the meter. For a complete diagram of the display, see page 5. Mode mg/l (ppm) % saturation Temperature Range -2 to to 0.0 to 200.0% 0.0 to 50.0 / mg/l (ppm) 32.0 to F Resolution mg/l (ppm) 0.1% 0.1 / 0.1 F ccuracy ±0.01 ±1.5% full scale ±1.5% full scale ±0.3 / ±0.5 F alibration up to 5 points: one or two points (100% and 0% saturation); offset up to 1.68; 4.01; plus separate mg/l (ppm) slope ±5 or ±10 F 7.00; 10.01; Temperature compensation: automatic from 0 to 50 Salinity correction: Range: 0.0 to 50.0 ppt Resolution: 0.1 ppt Method: Key in manually and automatic correction arometric pressure correction: Range: 500 to 1499 mm Hg / 66.6 to kpa Resolution: 1 mm Hg / 0.1 kpa Method: Key in manually and automatic correction DO Probe type: galvanic probe Display: dual LD acklit display: select from eight levels of brightness Operating temperature: 0 to 50 Power: four 1.5 V batteries (included) attery life: > 100 hours continuous use use of backlit display decreases battery life Dimensions: Meter: 7.5"L x 3.75"W x 2.25"H (19 x 9.5 x 5.7 cm) oxed: 9.2"L x 9.2"W x 2.75"H (23 x 23 x 7 cm) DO Probe: 6.8"L x 1.3" dia (173 mm L x 32 mm dia) with 10-ft cable Electrode: 6.0"L x 1.0" dia (152 mm L x 25 mm dia) with 10-ft cable Weight: Meter: 1.0 lb (0.5 kg) oxed: 2.0 lbs (0.9 kg) 62 63
33 13. ccessories Meters /DO 300 meter Meter only /DO 300 meter Includes DO probe with 10-ft cable ( ), electrode with 10-ft cable ( ), and electrolyte solution /DO meter kit. Includes /DO meter plus DO probe with 10-ft cable, electrode with 10-ft cable, 20 Singles pouches, 500 ml of electrolyte solution, 5 replacement membranes, membrane replacement tool, rinse bottle and foam-lined hard carrying case. Probes and probe replacement parts Replacement DO probe with 10-ft cable DO probe with 25-ft cable Replacement submersible electrode with 10-ft cable, single junction, S plastic housing Replacement DO electrolyte solution, 500 ml Electrolyte DO powder, 58.5 g. Mix your own electrolyte solution makes 400 ml. Has longer shelf life in unmixed form. WD Replacement DO membrane module. This item consists of a preassembled membrane, membrane lock, O-ring and cap. Simply screw replacement membrane module onto the probe you already have Replacement DO membranes, 5/pack Replacement DO membranes, 25/pack DO membrane installation tool. Required for membrane installation. Not required if you purchase replacement membrane module OKTON Singles calibration solution pouches solutions have ±0.01 accuracy at 25. Shpg wt 1.1 lb (510 g) per box Deionized rinse water solution pouches, 20/box calibration buffer solution pouches, 20/box alibration buffer solution pouches, 20/box alibration buffer solution pouches, 20/box ssortment pack, 5 each deionized water, 4.01, 7.00, and solution pouches. Other accessories electrode storage solution, 1 pint electrode cleaning solution, 1 pint Zero oxygen solution attery-powered magnetic stirrer To order OKTON accessories, contact your OKTON distributor. OKTON calibration solutions solutions have ±0.01 accuracy at 25. Shpg wt 1.1 lb (510 g) calibration buffer, 1 pint calibration buffer, 1 pint calibration buffer, 1 pint calibration buffer, 1 pint calibration buffer, 1 pint
34 14. ppendix 1: Meter Theory Dissolved oxygen levels in natural and waste waters depend on the physical, chemical, and biochemical activities in the water body. This meter uses a galvanic probe. It consists of a cell that contains electrolyte and that is enclosed by a selective membrane, and two metallic electrodes. The membrane is practically impermeable to water and ionic dissolved matter, but is permeable to oxygen and a few other gases. The cathode consumes the oxygen passed through the membrane, and produces an electric current in the probe. This current is proportional to the partial pressure of oxygen in the sample. Since the cathode consumes the oxygen in the sample, it is essential that the fluid must flow past the sensor to maintain accurate readings. The solubility of oxygen in water varies with barometric pressure and temperature, and decreases as salinity increases. For the most accurate DO readings, you need to compensate for these factors. This meter automatically compensates for temperature readings. It also allows you to enter a salinity correction factor and the barometric pressure to correct for this variability. 15. ppendix 2: Pressure vs. ltitude table arometric pressure affects DO readings, therefore this meter lets you enter the correct barometric pressure at your altitude. If you do not have equipment that lets you measure the exact barometric pressure at your altitude, you can estimate it using the chart below. If you change the barometric pressure setting from its factory setting (760 mm Hg), the % saturation calibration value in air will automatically adjust to a value other than 100% (see corrected % saturation value column below). The adjusted value is correct for the new barometric pressure setting. See page 39 or 47 for information on how to adjust the barometric pressure. Pressure vs ltitude ltitude (ft) Pressure (mm Hg) orrected % saturation value 0 (Sea level)
35 16. ppendix 3: Factory Default Settings Resetting the meter to factory default settings clears all calibration data and returns other setup functions to the default settings shown in the table below. However, the following settings will remain as you have set them: temperature unit of measure ( or F) The temperature offset calibration value the LD contrast setting See pages 38, 46 and 55 for directions on setting all modes of this meter to factory default settings. Program Function Options/settings Default 17. Warranty OKTON warrants this meter to be free from significant deviations in material and workmanship for a period of three years from date of purchase. OKTON warrants this probe to be free from significant deviations in material and workmanship for a period of six months from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse within the warrantied time period, please return freight prepaid and correction will be made without charge. OKTON alone will determine if the product problem is due to deviations or customer misuse. Out-of-warranty products will be repaired on a charge basis. lr Memory clear yes/no no dpr Pressure units Hg or Pa Hg arometric pressure adjust from mm Hg 760 mm Hg Salinity adjustment adjust from 0-50 ppt 0 ppt OFS Set % saturation offset up to ±10% no offset L View calibration data ELE View electrode offset 0.00 mv View electrode slope 100.0% View DO probe slope View % saturation offset 0 % offset View mv value of DO 100% saturation 37.0 mv View mv value of DO 0% saturation 0.3 mv 18. Return of items uthorization must be obtained from our ustomer Service Department before returning items for any reason. When applying for authorization, please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. We will not be responsible for damage resulting from careless or insufficient packing. restocking charge will be made on all unauthorized returns. NOTE: We reserve the right to make improvements in design, construction, and appearance of products without notice. OF Ready indicator/auto endpoint on/off Ready on only Select mg/l or ppm units mg/l or ppm mg/l Select temperature units or F retains settings Ld djust backlit display levels 0 to 8 (brightest) 8 rst Reset to factory default yes/no no 68 69
Portable Waterproof Dissolved Oxygen Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35641-Series Portable Waterproof Dissolved Oxygen Meter WTERPROOF MES T Dissolved Oxygen/ / F Data Meter
ph 300 and 310 Portable Waterproof ph/mv/ C Meter
Notes Write down the name and information of your OKTON distributor here. OPERTING INSTRUTIONS OKTON 35618-Series 300 and 310 Portable Waterproof /mv/ Meter REDY MES WTERPROOF T /mv/ Data Meter 300 Series
Instruction Manual DO 110. Hand-held Dissolved Oxygen Meter with RS 232C. 68X361305 Rev. 0-11/05. Technology Made Easy...
Instruction Manual DO 110 Hand-held Dissolved Oxygen Meter with RS 232C Technology Made Easy... 68X361305 Rev. 0-11/05 Preface This manual serves to explain the use of the hand-held series meters. The
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
DISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600
User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen
YSI Meter Protocol: Specific Conductivity, ph, & Dissolved Oxygen 1. Check the log-book to see if somebody else has already checked the meter for accuracy within the timeframes below. If the meter hasn
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance
Page 1 of 9 KEY WORDS- Dissolved Oxygen, calibration, probe preparation, meter preparation APPROVALS APPROVED BY: Kean S. Go, DATE: +/?? APPROVED BY: A Pement DATE: a// A/? f Lisa Ross, EHAP Senior Scientist
User Guide. 9708 Dissolved Oxygen Probes
User Guide 9708 Dissolved Oxygen Probes Introduction The Thermo Scientific Orion 9708 dissolved oxygen probe simplifies measurements of dissolved oxygen, particularly Biochemical Oxygen Demand (BOD).
Instruction Manual HI 9146. Portable Waterproof Microprocessor Dissolved Oxygen Meter
Instruction Manual HI 9146 Portable Waterproof Microprocessor Dissolved Oxygen Meter w w w. h a n n a i n s t. c o m Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction
Instruction Manual Temp 4/5/6 Temperature Meter Series
Instruction Manual Temp 4/5/6 Temperature Meter Series 68X243607 Technology Made Easy... Rev. 2 01/03-1 - Preface This instruction manual serves to explain the use of the Temp 4/5/6 temperature meter.
Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor
Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process
OxyGuard. Dissolved Oxygen Probe
OxyGuard Dissolved Oxygen Probe Standard type USER MANUAL D02 Standard Probe manual gb 0903 CONTENTS 1. DESCRIPTION 1.1 The OxyGuard Probe... 1 1.3 Which kind of transmitter can be used with the OxyGuard
ExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
UltraBasic Benchtop ph Meter Operation Manual
UltraBasic Benchtop Meter Operation Manual UB-5 Denver Instrument Measuring 4 4 7 10 Meter 22.2 C 4.06 UltraBasic 902388.1 Rev. A UltraBasic Meter Quick Reference 1. Connect a AC adapter to the power connection.
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL
MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL 970 350/REV A/10-03 MODEL 970 DISSOLVED OXYGEN METER OPERATING MANUAL CONTENTS Introduction 1 Specification 1 Installation 2 Displays 2-3 Controls 4 Preparation
User Manual. Thermo Scientific Orion VERSA STAR
User Manual Thermo Scientific Orion VERSA STAR Advanced Electrochemistry Benchtop Meter 68X006501 Revision A April 2014 Contents Chapter 1 Meter Introduction... 6 Meter Overview... 6 Packing List... 7
INSTRUCTION MANUAL. Handy Polaris
INSTRUCTION MANUAL Handy Polaris **This page is left blank** Table of Contents 1 General...4 2 Setup...5 2.1 Calibrate...5 2.2 Set Salinity...5 2.3 User Interface...5 2.3.1 Language...5 2.3.2 Temperature
DISSOLVED OXYGEN METER Model : DO-5509
DISSOLVED OXYGEN METER Model : DO-5509 Your purchase of this DISSOLVED OXYGEN METER marks a step forward for you into the field of precision measurement. Although this DISSOLVED OXYGEN METER is a complex
RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G
Auto-Levelling Rotary Laser Level RL HW / RL HW+ / RL HGW / RL HV / RL HVPW/RL HVPW-G 77-496 / 77-429 / 77-439 / 77-497 / 77-427/ 77-441 Please read these instructions before operating the product Auto-Levelling
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL
MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL 430 350/REV A/10-03 MODEL 430 PORTABLE ph/conductivity METER OPERATING MANUAL CONTENTS Introduction 1 Specification 2 Installation 3 Displays 3-4
Instruction Manual ph 5+ ph 6+ Ion 6+ Technology Made Easy...
Instruction Manual ph 5+ ph/ C ph 6+ ph/ C/mV Ion 6+ ph/ C/mV/Ion Technology Made Easy... 68X243633 Rev 1-1 3/2011 Part of Thermo Fisher Scientific Preface This instruction manual serves to explain the
6 H2O + 6 CO 2 (g) + energy
AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic
Care and use guide. Contents. page
Care and use guide Contents page Features Bluelab Combo Meter Introduction Preparing for use Calibration Changing nutrient and temperature display units 6 Measuring hydroponic elements 7 Battery replacement
Resolution selectable from 0.01 and 0.001 ph. Range -2.000 16.000 ph. Accuracy ± 0.002 ph
ph EC DO ph Resolution selectable from 0.01 and 0.001 ph Range -2.000 16.000 ph Accuracy ± 0.002 ph Data logging Manual log on demand manual log on stability Interval logging Temperature readout ( C or
Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso
Operating Instructions Bedienungsanleitung Mode d'emploi Instrucciones de manejo Istruzioni d'uso SevenEasy Conductivity Italiano Español Français Deutsch English Contents Contents Page 1. Introduction...
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide
Appendix C. Vernier Tutorial
C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer
User's Guide. Pinless Moisture Meter. Model MO257
User's Guide Pinless Moisture Meter Model MO257 Introduction Congratulations on your purchase of the Extech MO257 Pinless Moisture Meter. The pinless moisture sensor monitors the moisture in wood and other
Galvanic Parts Per Million Dissolved Oxygen Systems
Galvanic Parts Per Million Dissolved Oxygen Systems Model 900 Portable s Model 9100 Model 9200 Model 9300 The Royce line of PPM level dissolved oxygen (DO) analyzers is the largest, most varied line in
HOBO Dissolved Oxygen Logger (U26-001) Manual
HOBO Dissolved Oxygen Logger (U26-001) Manual The HOBO Dissolved Oxygen logger is a standalone logger that uses RDO Basic Technology to measure dissolved oxygen (DO). The logger has an optical sensor that
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:
BOECO PORTABLE PH/ORP/TEMP METER MODEL PT-380 A Simultaneous displays ph or mv and temperature. A Convenient calibration with automatic buffer recognition of European and US buffer sets. Calibration data
CAD-05 Kit GSM Auto Dialer. Owner s Manual
CAD-05 Kit GSM Auto Dialer Owner s Manual CAD-05 Kit Manual.indd 1 Warnings: This device complies with Part 15 of the FCC rules, Operation of this device is subject to the following conditions: 1. This
SMS Alarm. User Manual
SMS Alarm User Manual 1 1. Introduction... 2 2. SMSA1000 Specifications... 3 3. Parts Identifications... 4 4. Packing Contents... 5 5. Battery and SIM Card Installation... 6 5.1 Open the Back Cover...
Replacement Instructions. Warning: During this procedure, keep small parts away from children.
apple ibook G4 Memory Card Replacement Instructions Follow the instructions in this sheet carefully. Failure to follow these instructions could damage your equipment and void its warranty. Note: Written
YSI Model 5000 YSI Model 5100. Dissolved Oxygen Temperature. Operations Manual
YSI Model 5000 YSI Model 5100 Dissolved Oxygen Temperature Operations Manual Table of Contents 1. General Description...1 2. Specifications...3 3. Quick Start...5 4. Controls...6 4.1 Front Panel... 6 4.2
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features 2 Quick guide 2 To operate 3 Cleaning 4 Battery replacement 4 Calibration 5 Error messages 5 Troubleshooting guide 6 Technical specifications 6 Information about
AD-01 Slave Auto Dialer. Owner s Manual
AD-01 Slave Auto Dialer Owner s Manual AD-01 Slave Manual.indd 1 10/15/2009 10:20:44 AM 2 AD-01 Slave Manual.indd 2 10/15/2009 10:20:44 AM Features: Programmable entry/exit delay time; select up to 9 (32
Model UT713 OPERATING MANUAL
Model UT713 OPERATING MANUAL TABLE OF CONTENTS TITLE PAGE Introduction Unpacking Inspection Safety Information Turning the Calibrator On Simulating a Thermocouple Measuring a Thermocouple Simulating TC
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
http://www.coastalamusements.com
Nexus is a Registered Trademark of Coastal Amusements. OWNERS MANUAL 1935 Swarthmore Avenue, Lakewood, NJ 08701 (732)-905-6662 FAX (732)-905-6815 http://www.coastalamusements.com GAME DESCRIPTION Nexus
PR 50 Frymaster Universal Holding Cabinet (UHC)
PREP EQUIPMENT PR 50 Frymaster Universal Holding Cabinet (UHC) Models UHCP2, UHCP4, UHCPN4 Daily maintenance tasks PR 50 D1 Clean UHC Monthly maintenance tasks PR 50 M1 Calibrate UHC Annual maintenance
4-Channel Thermometer / Datalogger
USER GUIDE 4-Channel Thermometer / Datalogger RTD and Thermocouple Inputs Model SDL200 Introduction Congratulations on your purchase of the Extech SDL200 Thermometer, an SD Logger Series meter. This meter
SINGLE RANGE DS METERS Operation Instructions
SINGLE RANGE DS METERS Operation Instructions User Manual for Models 512T2 512T3 512T4 512T5 512T10 512M2 512M3 512M4 512M5 512M10 6115 Corte del Cedro Carlsbad, CA 92009-1516 USA Tel 760-438-2021 Fax
Tablet PC Tutorial for Windows 7
VIRGINIA TECH Tablet PC Tutorial for Windows 7 Getting Started Guide Instructional Technology Team, College of Engineering Last Updated: Fall 2013 Email [email protected] if you need additional assistance
AT&T Small Business System Speakerphone with Intercom and Caller ID/Call Waiting 974
974QSG_(Rev14)ml_23.5.06 5/23/06 11:28 AM Page 1 QUICK START GUIDE AT&T Small Business System Speakerphone with Intercom and Caller ID/Call Waiting 974 This Quick Start Guide provides basic instructions.
Model MC-82 Dissolved Oxygen Temp. Meter Handbook Version : 1.0
Congratulations! You have purchased the latest in Handheld Dissolved Oxygen-Temperature instrumentation. We trust that your new MC-82 will give you many years of reliable service. The MC-82 is a breeze
1-877-654-9394. UP400 Programmable Thermostat Installation Instructions & User Guide. For Installation Help. White-Rodgers.com
UP400 Programmable Thermostat Installation Instructions & User Guide For Installation Help 1-877-654-9394 White-Rodgers.com 2011 Printed in China White-Rodgers 8100 West Florissant Avenue St. Louis, MO
User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN
User's Guide Digital Psychrometer Model RH390 MAX MIN F C WBT Dew Point Introduction Congratulations on your purchase of the Extech RH390 Digital Psychrometer. This device measures Humidity, Air Temperature,
MAINTENANCE & TROUBLESHOOTING
MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens replace the projection lamp replace the batteries in the remote use the security lock feature troubleshoot the projector Cleaning
GSM AD05 Slave GSM Auto Dialer- Instruction Manual
GSM AD05 Slave GSM Auto Dialer- Instruction Manual Please read these instructions before you start the installation Features LCD display Programmable 9 x 32 digit phone numbers for each trigger. 10 second
ph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
TS1200 Intelligent Lead Free Soldering Station. Operation Manual
TS1200 Intelligent Lead Free Soldering Station Operation Manual Thank you for purchasing an Intelligent Lead Free Soldering Station. It is designed for lead free soldering. Please read this manual before
SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE
SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE Cable Processor Module overview WARNING! When tipping the Cable Processor Module back, (after removing the toggle arm pin), use extreme caution not to drop
Magic Wand Portable Scanner
Magic Wand Portable Scanner PDS-ST470-VP User Manual Table of Contents 1. Key Features... 3 2. Functional Parts... 3 3. Explanation of the Status Icons... 5 4. Using the Scanner... 5 4.1. Charging up the
To Purchase This Item, Visit BMI Gaming www.bmigaming.com 1-800-746-2255 + 1-561-391-7200
Table of Contents General Description and Play 2 Game Set Up 2 Loading Balloons 2 Technical Operation-General 3 Adjusting Balloon Sensor Wire 4 Programming Instructions 4 Default Settings Troubleshooting
Error! Bookmark not defined. Error! Bookmark not defined. Error! Bookmark not defined.
Remote Lighting Kit Table of Contents Lighting Kit Overview 2 System Requirements 3 Basic Concepts of Wireless Routers and Networks 4 Basic Concepts of Network Light Bulbs 8 Planning Your Lighting Network
Paint Test Equipment. Stargloss Gloss Meter Data Sheet
Paint Test Equipment Stargloss Gloss Meter Data Sheet Coating Thickness Gloss Porosity Adhesion Surface Roughness Surface Cleanliness Climatic Conditions Electrostatic Inspectors Accessories Complies with
Operating Instructions
Operating Instructions FiveEasyPlus ph meter FEP20 Contents Contents 1. Introduction 3 2. Safety measures 3 3. Installation 4 3.1 Installing the electrode arm (optional for FEP20) 4 4. Operating the FiveEasyPlus
Professional Plus. Calibration Tips
Professional Plus Calibration Tips Table of Contents Introduction...1 Calibration Worksheet...1 Temperature...3 Calibration Tips...3 Troubleshooting Tips...3 Conductivity...4 Calibration Tips...4 Troubleshooting
ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.
Instruction Sheet PN 51A-397/rev.F January 2011 Model 397 ph Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS MODEL 397 PERFORMANCE
T Sentry 4 Multi-Point Digital Alarm Instruction Manual
T Sentry 4 Multi-Point Digital Alarm Instruction Manual Introduction The T Sentry4 (TS4) is a microprocessor-based temperature monitoring and alarm device with user programmable high-low alarm setpoints
Professional Thermohygrometer
User Manual Professional Thermohygrometer with NIST-Traceable Calibration Model 20250-21 THE STANDARD IN PRECISION MEASUREMENT Introduction The Digi-Sense Professional Thermohygrometer (Model 20250-21)
RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer
RS Stock No. 724-4207 Instruction Manual RS-1340 Hot Wire Anemometer EN FR IT DE ES CONTENTS / EN Title CONTENTS Page 1. SAFETY INFORMATION...1 2. INTRODUCTION...2 3. SPECIFICATIONS...3 4. PARTS & CONTROLS...4
INSTALL/REMOVAL INSTRUCTIONS: WINDOW REGULATOR
REMOVAL/INSTALL OF WINDOW REGULATOR (741-306) Honda Accord 2003 07 General Tech Tips: Use painter s tape rather than duct tape to secure window. It will not damage paint or leave sticky residue. A plastic
MAINTENANCE & TROUBLESHOOTING
MAINTENANCE & TROUBLESHOOTING This section describes how to: clean the lens clean the fan intake filter replace the projection lamp replace the batteries in the remote control use the Kensington lock feature
7-SEGMENT DIGITAL CLOCK
57mm 7-SEGMENT DIGITAL CLOCK Large 57mm clock & temperature display with extra unique feature Total solder points: 263 Difficulty level: beginner 1 2 3 4 5 advanced K8089 ILLUSTRATED ASSEMBLY MANUAL H8089IP-1
Your Simple Guide to Battery. www.firstalertpro.com. Replacement. Customer Care: 1-800-852-0086. www.firstalertpro.
Previous Menu Your Simple Guide to Battery www.firstalertpro.com Replacement p e t s ts ep -by Customer Care: 1-800-852-0086 FA/1565 9/00 www.firstalertpro.com Table of Contents: page Introduction...............................
Hygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
TABLE OF CONTENTS QUICK START... 1 DIAGRAM OF THE DIGITAL PHOTO FRAME... 2 KEYPAD BUTTON FUNCTIONS... 3 REMOTE CONTROL FUNCTIONS..
TABLE OF CONTENTS QUICK START........ 1 DIAGRAM OF THE DIGITAL PHOTO FRAME... 2 KEYPAD BUTTON FUNCTIONS..... 3 REMOTE CONTROL FUNCTIONS..4 DIGITAL PHOTO FRAME NAVIGATION........5 OSD MENU FUNCTIONS..6
Wash Select II to WashPay Integration Field Retrofit Installation Guide
Wash Select II to WashPay Integration Field Retrofit Installation Guide Wash Select II Software Version 6.00 and above POS4000 Software Version 8.25 Unitec www.startwithunitec.com WASH SELECT II TO WASHPAY
How To Use A Water Quality Monitor (Wqmsw)
Water Quality Monitor WQM Host Software User s Guide The user s guide is an evolving document. If you find sections that are unclear, or missing information, please let us know. Please check our website
XP Series Balances Operation Manual
XP Series Balances Operation Manual North and South America: 6542 Fig Street Arvada, Colorado 84 1-8-321-1135 Tel: 33-431-7255 Fax: 33-423-4831 U.K. and Ireland: Denver House, Sovereign Way Trafalgar Business
Using GIGABYTE Notebook for the First Time
Congratulations on your purchase of the GIGABYTE Notebook. This manual will help you to get started with setting up your notebook. The final product configuration depends on the model at the point of your
Powers Controls TH 192 HC Heating/Cooling Room Thermostat
Powers Controls TH 192 HC Heating/Cooling Room Thermostat Technical Instructions Document No. 155-066P25 TH 192-2 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 192 HC thermostats are proportional
Triac Printed Circuit Board Replacement
Technical Service Bulletin: Triac Printed Circuit Board Replacement TRONIC 5000C Pro Models: WH17, WH27, WH36 Introduction Fig. 1 ELECTRICITY IS EXTREMELY DANGEROUS. TAKE EXTRA PRECAUTIONS AND ENSURE ALL
Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986
Electrical Stability Tester, Model 23E Instruction Manual Manual No. 102149576, Revision B Instrument No. 102130986 Fann Instrument Company Electrical Stability Tester Instruction Manual 2012 Fann Instrument
OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S
Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and
Four-Line Small Business System Speakerphone 945
945QSG_(rev22)ml23.5.06 5/23/06 11:29 AM Page 1 QUICK START GUIDE Four-Line Small Business System Speakerphone 945 This Quick Start Guide provides basic instructions. For additional installation options
User Manual I1010Q16DCZ
User Manual I1010Q16DCZ Explanation, Note and Warning *Explanation: important information, which can help you better use your tablet PC. *Note: indicates that not observing the content may result in hardware
Four-Line Small Business System Speakerphone 945
945 CIB 2ATT 3/14/05 9:58 AM Page III USER S MANUAL Part 2 Four-Line Small Business System Speakerphone 945 Please also read Part 1 Important Product Information 2005 Advanced American Telephones. All
AC-115 Compact Networked Single Door Controller. Installation and User Manual
AC-115 Compact Networked Single Controller Installation and User Manual December 2007 Table of Contents Table of Contents 1. Introduction...5 1.1 Key Features... 6 1.2 Technical Specifications... 7 2.
Installing the Video Input and TV Tuner Cards in a Compact Computer or a Dual PCI-Slot Tower Computer
Installing the Video Input and TV Tuner Cards in a Compact Computer or a Dual PCI-Slot Tower Computer This booklet describes how to install the video input and TV tuner cards in a compact Macintosh computer
How To Program An Autodialer
GJD HYL005 GSM Autodialer Instruction Manual Please read these instructions before you start the installation Features: LCD display. Programmable 9 x 32 digit phone numbers for each trigger. 10 second
Quick Start Guide See Inside for Use and Safety Information
3 rd Generation Personal 3D Printer Quick Start Guide See Inside for Use and Safety Information The USB Mass Storage Device Contains the User Guide and Quick Start Guide in other Languages Congratulations
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Phoenix Body Fat Scale Manual
Phoenix Body Fat Scale Manual Low Batteries & bad battery connections are the #1 cause of scale malfunction and inaccuracy! We test all of our scale returns from consumers. Fully 60% of consumer returns
Global Water Instrumentation, Inc.
Global Water Instrumentation, Inc. 11390 Amalgam Way Gold River, CA 95670 T: 800-876-1172 Int l: (916) 638-3429, F: (916) 638-3270 Temperature Sensor: WQ101 ph Sensor: WQ201 Conductivity Sensor: WQ301
ES&S AutoMARK. Pre-Election Day Checklist
ES&S AutoMARK Pre-Election Day Checklist Rear Access Door Touch Screen Monitor Lid Compact Flash Access Door Lid Latches Key-Operated Mode Switch Keypad Ballot Feed Tray ADA Port Headphone Jacks 1 Checking
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques
PPB Dissolved Oxygen Measurement - Calibration and Sampling Techniques Introduction The amount of dissolved oxygen in process water is continually gaining importance in many industries as a critical parameter
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
Features of Your Cisco Unified IP Phone
Cisco Unified IP Phone 8961, page 1 Cisco Unified IP Phone 9951, page 9 Cisco Unified IP Phone 9971, page 18 General Phone Information, page 27 Cisco Unified IP Phone 8961 The following sections describe
WIRELESS HOME ALARM SYSTEM (WHA1)
WIRELESS HOME ALARM SYSTEM (WHA1) IMPORTANT : PLEASE READ THIS MANUAL CAREFULLY BEFORE ATTEMPTING TO INSTALL AND OPERATE THIS ALARM SYSTEM. SAFETY Please note: Before you start to install this alarm, we
VIDEO DOOR PHONE SYSTEM
ENGLISH VIDEO DOOR PHONE SYSTEM USER MANUAL Manual covers IH-692 Master and IH-692 Slave Monitors PRECAUTIONS BEFORE INSTALLATION Connect this unit ONLY to other compatible units. Do not connect it to
