Development of a parallel plate ion chamber for radiation protection level
|
|
|
- Noah Wilkerson
- 10 years ago
- Views:
Transcription
1 Artigo Original Revista Brasileira de Física Médica.2011;5(1):9-14. Development of a parallel plate ion chamber for radiation protection level Desenvolvimento de uma câmara de ionização de placas paralelas para proteção radiológica Márcio Bottaro 1, Maurício Moralles 2 and Maurício Landi 1 1 Instituto de Eletrotécnica e Energia/Seção Técnica de Ensaios em Equipamentos Eletromédicos (STEEE), Universidade de São Paulo, SP, Brasil. 2 Instituto de Pesquisas Energéticas e Nucleares/Centro do Reator Nuclear de Pesquisas (CRPq), Comissão Nacional de Energia Nuclear, São Paulo, SP, Brasil. Abstract A new parallel plate vented ion chamber is proposed in this paper. The application of this chamber was primarily intended to the measurement of stray radiation in interventional procedures, but the energy response of about 2.6%, which was obtained in the first prototype, on the range from 40 to 150 kv using ISO narrow qualities, provided the possibility of a wide modality application on radiation protection. Primary studies with Maxwell 2D electromagnetic field simulator revealed an optimized model regarding effective volume and saturation voltage levels, which conferred to the ion chamber a dual entrance window feature. The development of this ion chamber has the main contribution of Monte Carlo calculations as a support tool to the establishment of the effective volume of the chamber and determination of the best materials for housing mounting and conductive elements, such as guard rings, electrode, and windows. Even the composition of the conductive layers, which would be neglected due to their very small thicknesses (about 35 µm), had important influence on the results and could be better understood with Monte Carlo N-Particle Transport Code System (MCNP) simulations. Keywords: ion chamber, Monte Carlo, energy response, radiation protection. Resumo Uma nova câmara de ionização de placas paralelas ventilada é proposta neste trabalho. A aplicação da câmara teve como objetivo principal a medição da radiação parasita nos procedimentos intervencionistas, porém as variações da resposta em energia de aproximadamente 2,6% na faixa de 40 a 150 kv, obtida no primeiro protótipo utilizando os feixes padrão estreitos da ISO , possibilitou uma ampla aplicação na modalidade de proteção radiológica. Estudos iniciais feitos com o simulador de campo eletromagnético Maxwell 2D revelaram um modelo otimizado em relação ao volume efetivo e tensão de saturação, os quais conferiram à câmara de ionização a característica de janela de entrada dupla. O desenvolvimento desta câmara de ionização teve como principal contribuição as simulações de Monte Carlo como uma ferramenta de suporte para o estabelecimento do volume efetivo da câmara e para a determinação dos melhores materiais para os elementos de montagem e circuito condutivo, como por exemplo, os anéis de guarda, eletrodo e as janelas. Até mesmo as composições de camadas condutivas, que seriam negligenciadas devido a sua pequena espessura (aproximadamente 35 µm), tiveram uma importante influência nos resultados, que foi melhor compreendida com as simulações realizadas com o Monte Carlo N-Particle Transport Code System (MCNP). Palavras-chave: câmara de ionização, Monte Carlo, resposta da energia, proteção radiológica. Introduction Interventional radiology equipments are extensively used in medical practice. In the last decades, the minimally invasive procedures, associated with technological improvements, resulted in the expansion of the equipment market all over the world. In Brazil, this panorama is not different and an increasing number of manufacturers are putting much effort in the development and production of interventional radiology machines to supply market necessities 1. Medical electrical equipment certification process is compulsory in Brazil since In this process, type tests of these equipments are performed according to international standards from the International Electrical Commission (IEC) series. For interventional radiology equipment, such tests are also performed and some requirements of the applied IEC standards require particular instruments for the X radiation measurements, especially leakage and stray radiation profiles in significant zones of occupancy. These requirements include special dimensions and volume chambers 2. Corresponding author: Marcio Bottaro Instituto de Eletrotécnica e Energia da Universidade de São Paulo (USP) Professor Luciano Gualberto, São Paulo (SP) Brazil [email protected] Associação Brasileira de Física Médica 9
2 Bottaro M, Moralles M, Landi M For leakage radiation measurements, the most important requirements are the entrance window area of 100 cm² and linear dimensions not exceeding 20 cm. In this modality, some commercial chambers are available and the evaluation of their time and dose response function is very important. For the stray radiation profiles in significant zones of occupancy and also for measurements of isokerma maps of scattered radiation, the most important requirement is related to the volume of the chamber, 500 cm³, and linear dimensions that cannot exceed 20 cm. For this modality, no chambers are commercially available, since the old FLUKE 96010A was discontinued. Although FLUKE 96010A regards requirements for both leakage and stray radiation, it was designed with the specific purpose of leakage radiation measurement. This paper presents a new purpose of vented parallel plate ionization chamber, according to the dimensions and volume requirements of IEC international standards series and regarding all the modalities of diagnostic X-ray equipment. It was specially designed to present energy response and sensibility necessary to the measurements of leakage radiation, stray radiation profiles in significant zones of occupancy, and scattered radiation in isokerma maps. In the next section, the design method using Maxwell 2D electromagnetic field simulator and the Monte Carlo calculation code MCNP4C is presented 3. Practical measurements and simulations of energy response, based on normalized X-ray qualities 4, are reported in Results and the fundamental contributions of Monte Carlo calculations to understand the chamber behavior are discussed in Discussion and Conclusions sections. Materials and methods Electrical field design and simulation The parallel plate ion chamber was designed with three fundamental aims: a low cost, robust, and easy manipulation detector. Such objectives lead to a first simple design with a single entrance window and a collector electrode surrounded by a guard ring. First chamber model The first chamber had polimetilmetacrilate PPMA cylindrical walls (red), windows and collector electrode (blue) made of polycarbonate, covered with conductive graphite (gray) in a way to state a chamber polarization and charge collection circuit. All the internal volume is filled with air (white) as in a vented chamber. This first model is shown in Figure 1. It was obtained from Maxwell 2D design tool and indicates a section of the central axis perpendicular to the window plane, where all the components can be seen. As with this free software, we are able only to simulate 2D electric fields, this section was chosen as a more representative plane to evaluate electrical field design and behavior. The electrical filed simulation obtained with the use of Maxwell 2D software is shown in Figure 2, and it was performed with a chamber voltage of 400 Vdc. In Figure 2 one observes that there is a large area (orange) corresponding to the effective volume of the chamber, where the charges produced by ionization are collected. There are also two types of regions that determine dead volumes, where the produced charges are not collected due to different reasons. The first type of dead volume is represented from blue to yellow, where the electric field is very weak and corresponds to the region outside of the space between the collecting electrodes. The second type of dead volume corresponds to the region between the guard ring and the window, shown in red on the lower corners, which presents a higher electric field that deviates the charges to the guard ring, and consequently they are not collected by the measuring electronics. Second chamber model With the purpose to obtain a smaller dead volume, a second model of chamber was designed, which had almost the same material components, PMMA walls but now with two polycarbonate windows and a centralized polycarbonate collector electrode. All components were covered with conductive graphite to provide the chamber polarization and the charge collection circuit. The guard rings were placed at a different position, still surrounding the collecting electrode, but dividing the window circuit and chamber design symmetrically. This confers to the second model other interesting properties: reduced saturation voltage, reduced dead volumes and bilateral capacity of measurement. This new model is shown in Figure 3. Results of Maxwell 2D calculations are illustrated in Figure 4. The electromagnetic field calculation for the second chamber was performed with 250 Vdc, almost half of the C-Graphite D - Air A - PMMA B - Polycarbonate Figure 1. Section of the first chamber model design used in the electrical field simulation. E [V/M] e e e e e e e e e e e e e e e e+011 Figure 2. First chamber model electrical field simulation. 10 Revista Brasileira de Física Médica.2011;5(1):9-14.
3 Development of a parallel plate ion chamber for radiation protection level voltage used in the fi rst chamber model. Figure 4 shows that the regions corresponding to dead volumes and low electric fi elds inside the collecting electrodes are signifi - cantly smaller than in the fi rst chamber. Prototype and Monte Carlo model Prototype The fi rst prototype model is illustrated in Figure 5. Water-based graphite was used to avoid corrosion of the PMMA, since almost all solvents in the most common graphite inks promote PMMA degradation with time. A painting procedure was developed in order to guarantee uniform ink distribution and thickness. A guiding layer was designed to facilitate the painting of the guard rings. The ion-chamber wiring was performed by pressing the conductors against the internal side of the windows and with nylon screws in guard rings and electrode. Monte Carlo model The same ion chamber was also modeled in the MCNP4C 3 code to evaluate the energy response of the detector. A simplifi ed geometry was stated, and details of wiring and internal nylon screws were neglected. In Monte Carlo calculations, null electron and photon importance was assigned to the dead volume and dead zone, because in MCNP4C the electric fi eld would not be inserted in the input code. The components of walls (PMMA), windows and collector electrode (polycarbonate) were modeled using standard compositions. The chemical composition of the graphite cover was modeled using the graphite ink manufacturer s specifi cations for dry ink: 100% graphite. The graphite thickness was based on the measurements made in the prototype chamber layers, with and without graphite coating, using a precise digital Mitutoyo micrometer model The graphite mean thickness was 35 µm. Measurement setup and qualities As the main application of the detector is related to radiation protection, ISO narrow beams were selected within diagnostic range 4. For the fi rst evaluation, four X-ray qualities were used (N60, N80, N100 and N150 kv), based on the available qualities in IEE/USP and in IPEN/CMR, which are the reference laboratories for ion chamber calibrations in São Paulo. The reference kerma/ion chamber charge ratio was used to state the energy response evaluation and Monte Carlo comparison parameter. The same qualities were implemented in STEEE/IEE, according to ISO , where a PTW Freiburg GmbH 300 cm³ TA model ion chamber was used to state the reference doses. The measurements were then performed in both laboratories and a PTW UNIDOS 457 electrometer was used to collect charge of the ion chamber prototype during tests. In both laboratories, results were taken using the reference ratio (Eq. 1): D - Air A - PMMA C-Graphite B - Polycarbonate Figure 3. Section of the second chamber model design used in the electrical fi eld simulation. Figure 5. First chamber prototype. E [V/m] e e e e e e e e e e e e e e e e+010 Reference Ratio Results 3,3E-05 3,1E-05 2,9E-05 2,7E-05 2,5E-05 2,3E-05 2,1E-05 1,9E-05 1,7E-05 1,5E IPEN MCNP STEEE Figure 4. First chamber model design section for electrical fi eld simulation. Figure 6. First results of Kerma/Charge ratio in reference and development laboratories and MCNP simulations. Revista Brasileira de Física Médica.2011;5(1):
4 Bottaro M, Moralles M, Landi M R n = C n K n (1) where: R is the dose/charge ratio; K is the laboratory reference air kerma; and C is the ionization chamber prototype collected charge. All parameters are related to their ISO narrow quality n. The same equation was employed for the results of the simulated data. The setup geometry of each laboratory was implemented in MCNP, and the input spectra were obtained from earlier simulations in GEANT4 code 5. For the MCNP simulations, the reference air kerma K was obtained in a prior simulation setup with all narrow qualities over an air volume of 500 cm³ accomplishing the air kerma defi nitions. Further simulations with the ion chamber model were performed to collect data of charge C in the predefi ned effective volume. Comparisons between experimental and simulated data are shown in the next section. 2nd Reference Ratio Results 3,3E-05 3,1E-05 2,9E-05 2,7E-05 2,5E-05 2,3E-05 IPEN 2,1E-05 STEEE 1,9E-05 MCNP 1,7E-05 1,5E Figure 7. First results of Kerma/Charge ratio in reference and development laboratories and second results of MCNP simulations. Final Reference Ratio Results 2,40E-05 2,30E-05 2,20E-05 2,10E-05 2,00E-05 1,90E-05 1,80E-05 1,70E-05 1,60E-05 1,50E Weighting Residuals 1,50E+00 1,00E+00 5,00E+01 0,00E+00-5,00E ,00E+00 MCNP STEEE Residuals Figure 8. Final results of Kerma/Charge ratio in reference and development laboratories and MCNP simulations. Results A summary of results containing charge/kerma ratio is presented in the graphic of Figure 6. A good agreement between both laboratories results is clearly seen; however, a large discrepancy with simulated results is also presented. Both window sides were tested and no relevant differences were found. Figure 6 clearly presents a detector response with considerable energy dependency. As ISO Narrow 70 kv quality was not available, an indication of the energy dependency was based on the c variation coeffi cient of the results. For the fi rst results, the energy dependencies were very similar for both laboratories, 13.4% in IPEN/CMR and 10.8% in STEEE/IEE. However, in MCNP simulations, the result was 6.9%. All of them were not in agreement with ISO energy dependency requirements of less than 5%. Such discrepancies were evaluated, and two their possible sources were studied: MCNP cross-section library used and composition of the dry graphite ink. While a new cross-section library was implemented, a spectroscopic characterization of the graphite ink was performed and a new chemical composition was found, which is different from the one specifi ed by the manufacturer. The main components found, other than the graphite, were: SiO 2, CaO, MgO, Al 2 O 3, and Na 2 O. With such results, a new chemical model for graphite layer was implemented in the MCNP code. The new results are shown in Figure 7. Both libraries that were used produced almost the same results, and a new graphite ink was used in a second prototype. The composition of this new ink was also evaluated by means of spectroscopy procedure. For this graphite ink, no other elements than carbon (graphite) were found and a third series of simulations was then performed. The results were very similar to that obtained in laboratories. Small adjustments in the dead volume were performed to fi t results more adequately. The new radiation protection ion chamber had its MCNP model defi ned. New data are plotted in Figure 8. The results show good agreement between experimental and simulated results. Unfortunately, no measurements could be performed in the reference laboratory of IPEN/CMR, but STEEE/IEE ISO qualities were found to be adequate, as they were validated in earlier measurements. Once again, both window sides were tested with no signifi cant differences. A good energy response was also observed for both experimental and simulated results. In Figure 8 the uncertainties are also shown within 68% of confi dence level, and weighting residuals were calculated in order to compare data. Residuals are within one standard deviation, and the new data showed satisfactory results. The energy dependence of experimental and simulated data was also calculated: 2.1% for simulations and 2.6% for STEEE/IEE laboratory measurements. These results are within ISO specifi cations of 5%. Further tests and simulations with other interest spectra and practical application in type 12 Revista Brasileira de Física Médica.2011;5(1):9-14.
5 Development of a parallel plate ion chamber for radiation protection level tests in IEE laboratories are about to be performed, in order to attest compliance of the detector to be applied in radiation protection measurements. Discussion The parallel plate vented chamber presented many successful results and its knowledge could also be improved with some additional simulations to better understand its full energy response in the diagnostic range, including mammographic applications. This is very important as new IEC standard series are required for all modalities of diagnostic equipment, including mammographic, dental, fluoroscopic, conventional, and computed tomography X-ray generators, and the determination of stray radiation profiles. The bilateral property was confirmed in the practical tests with no relevant dependence during the measurements. Nevertheless, other practical measurements and performance tests should be executed to attest its compliance with international standards and to validate this detector for use on practical type tests. Conclusions Monte Carlo calculation is a very important tool in the development of many radiation detectors. The present study showed that it can be also applied in the development of ionization chambers. The parallel plate vented chamber proposed in the present paper corresponds to the final prototype version. Based on MCNP results, studies regarding the best materials to be used in the chamber housing and main parts were performed, and a chamber model that can be further used to simulate its behaviour on other X-ray qualities was stated. A satisfactory energy response was achieved and other performance tests are under execution in the laboratory, accompanied by MCNP calculations. Acknowledgment Authors would like to thank STEEE staff, they were always ready to help and contribute during laboratorial measurements. Also, thanks to IPEN/CMR for the attention and support. References 1. Canevaro L (2009) Physical and technical aspects in Interventional Radiology. Revista Brasileira de Física Médica 3(1): Associação Brasileira de Normas Técnicas. Equipamento eletromédico Parte 1: Prescrições gerais de segurança 3. Norma Colateral: Prescrições gerais para proteção contra radiação de equipamentos de raios X para fins diagnósticos. ABNT, Rio de Janeiro, 2001 (NBR IEC ) 3. Briesmeister J. MCNP (2000) A general Monte Carlo N-particle transport code, version 4C. Los Alamos National Laboratory Report, LA M 4. International Organization for Standardization (1996) X and gamma reference radiation for calibration dosemeters and doserate meters and for determining their response as a function of photon energy Part 1: Radiation characteristics and production methods. Geneva Switzerland (ISO ) 5. Guimarães C C, Moralles M, Okuno E (2008) Performance of GEANT4 in dosimetry applications: Calculation of X-ray spectra and kerma-to-dose equivalent conversion coefficients. Rad. Meas. 43:1525 DOI /j. radmeas Revista Brasileira de Física Médica.2011;5(1):
6
IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 IMPROVEMENT OF GAMMA
DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 DEVELOPMENT AND IMPLEMENTATION
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level Eduardo de Lima Corrêa 1, Vitor Vivolo 1, Maria da Penha A. Potiens 1 1 Instituto de Pesquisas Energéticas e Nucleares
Lasers and Specturbility of Light Scattering Liquids - A Review
Artigo Original Revista Brasileira de Física Médica.2011;5(1):57-62. Reproducibility of radiant energy measurements inside light scattering liquids Reprodutibilidade de medidas de energia radiante dentro
DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1
TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT
DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DOSIMETRIC CHARACTERIZATION
Evaluation of the image quality in computed tomography: different phantoms
Artigo Original Revista Brasileira de Física Médica.2011;5(1):67-72. Evaluation of the image quality in computed tomography: different phantoms Avaliação da qualidade de imagem na tomografia computadorizada:
PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 PERFORMANCE EVALUATION
Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom
Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Fernanda B. C. Nonato, Claudinei T. Cescon and Linda V. E. Caldas, Instituto de Pesquisas Energéticas e Nucleares,
How To Test Water Penetration On Concrete Block And Mortar
WATER PENETRATION TEST ON CONCRETE BLOCK MASONRY Vilató, Rolando Ramirez 1 1 PhD, Prof., Mackenzie Presbyterian University, School of Engineering, São Paulo, [email protected] This paper considers
3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
Calculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 Associação Brasileira de Energia Nuclear - ABEN ISBN: 978-85-99141-02-1 VoxelDose: A SOFTWARE
Development of on line monitor detectors used for clinical routine in proton and ion therapy
Development of on line monitor detectors used for clinical routine in proton and ion therapy A. Ansarinejad Torino, february 8 th, 2010 Overview Hadrontherapy CNAO Project Monitor system: Part1:preliminary
APMP/TCRI Key Comparison Report of Measurement of Air Kerma for Medium-energy X-Rays (APMP.RI(I)-K3)
APMP/TCRI Key Comparison Report of Measurement of Air Kerma for Medium-energy X-Rays (APMP.RI(I)-K3) J. H. Lee g, W. S. Hwang g, L. H. Kotler b, D. V. Webb b, L. Büermann l, D.T. Burns d, M. Takeyeddin
CFD SIMULATION OF IPR-R1 TRIGA SUBCHANNELS FLUID FLOW
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 CFD SIMULATION OF IPR-R1 TRIGA
A COMPUTATIONAL PHANTOM OF HEAD AND NECK - SISCODES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 A COMPUTATIONAL PHANTOM
RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?
Congreso Internacional Conjunto Cancún 2004 LAS/ANS-SNM-SMSR/International Joint Meeting Cancun 2004 LAS/ANS-SNM-SMSR XV Congreso Anual de la SNM y XXII Reunión Anual de la SMSR/XV SNM Annual Meeting and
Test Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 Test Section for Experimental
Jorge E. Fernández Laboratory of Montecuccolino (DIENCA), Alma Mater Studiorum University of Bologna, via dei Colli, 16, 40136 Bologna, Italy
Information technology (IT) for teaching X- and gamma-ray transport: the computer codes MUPLOT and SHAPE, and the web site dedicated to photon transport Jorge E. Fernández Laboratory of Montecuccolino
Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY
Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY A. Mrozik, B. Marczewska, P. Bilski, W. Gieszczyk Institute of Nuclear Physics PAN,
7üVNORD. TÜV NORD Röntgentechnik
.. TÜV NORD Röntgentechnik Niedersachsen Nordrhein-Westfalen Hamburg Schleswig-Holstein Baden-Württemberg Mecklenburg-Vorpommern Bremen Sachsen-Anhalt Berlin Brandenburg Sachsen' Thüringen' Bayern' Hessen
EVALUATION OF ON-SITE DIELECTRIC RESPONSE METHODS FOR NON- DESTRUCTIVE TESTING OF WATER TREED MV XLPE CABLES
EVALUATION OF ON-SITE DIELECTRIC RESPONSE METHODS FOR NON- DESTRUCTIVE TESTING OF WATER TREED MV XLPE CABLES Sverre Hvidsten*, Peter Werelius**, Jørgen Christensen*** * SINTEF Energy Research (SEfAS),
CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY
243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland Cements on Mass Attenuation Coefficient
Journal of Materials Science and Engineering A 3 (12) (2013) 838-842 D DAVID PUBLISHING Effect of Gamma Ray Energies and Addition of Iron Slag by weight to Portland s on Mass Attenuation Coefficient Abd
National Laboratory for Metrology of Ionizing Radiation (LNMRI) 1
National Laboratory for Metrology of Ionizing Radiation (LNMRI) 1 1. Introduction. The Institute for Radiation Protection and Dosimetry (IRD) has its origin in a small dosimetry Laboratory set up in the
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES
ELECTRIC FIELD LINES AND EQUIPOTENTIAL SURFACES The purpose of this lab session is to experimentally investigate the relation between electric field lines of force and equipotential surfaces in two dimensions.
QUESTIONÁRIOS DE AVALIAÇÃO: QUE INFORMAÇÕES ELES REALMENTE NOS FORNECEM?
QUESTIONÁRIOS DE AVALIAÇÃO: QUE INFORMAÇÕES ELES REALMENTE NOS FORNECEM? Grizendi, J. C. M [email protected] Universidade Estácio de Sá Av. Presidente João Goulart, 600 - Cruzeiro do Sul Juiz de Fora
3D SCANNER. 3D Scanning Comes Full Circle. Your Most Valuable QA and Dosimetry Tools
3D SCANNER 3D Scanning Comes Full Circle Your Most Valuable QA and Dosimetry Tools BETTER OUTCOMES THROUGH TECHNOLOGY The 3D SCANNER is different by design. 3D SCANNER has been developed from the ground
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
ESTIMATION OF EFFECTIVE DOSE FROM PEDIATRIC RADIOGRAPHY OF THE CHEST IN MEDELLÍN-COLOMBIA
Revista Facultad de Ciencias Universidad Nacional de Colombia, Sede Medellín V 4 N 1 enero-junio 2015 ISSN 0121-747X / ISSN-e 2357-5749 Research Paper Páginas 11 a 16 ESTIMATION OF EFFECTIVE DOSE FROM
Laue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
Neutron Detection Setups proposed for
Neutron Detection Setups proposed for DESPEC D. Cano-Ott on behalf of the WG members CIEMAT, IFIC, LNL, FYL, UPC, UU, UW Motivation GOAL: to measure neutron emission probabilities and energies for neutron
Dose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi
Dose enhancement near metal electrodes in diamond X- ray detectors Acknowledgements Surrey University: P.J. Sellin M. Abd-El Rahman P. Veeramani H. Al-Barakaty F. Schirru Mechanical Workshop A. Lohstroh*,
ABSORBED DOSE DETERMINATION IN EXTERNAL BEAM RADIOTHERAPY
TECHNICAL REPORTS SERIES No. 398 Absorbed Dose Determination in External Beam Radiotherapy An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water Sponsored by the
Characterisation of two new ionisation chamber types for use in reference electron dosimetry in the UK
NPL REPORT DQL-RD001 Characterisation of two new ionisation chamber types for use in reference electron dosimetry in the UK Julia A D Pearce September 2004 September 2004 CHARACTERISATION OF TWO NEW IONISATION
Projects and R&D activities
Projects and R&D activities J.M. Jimenez On behalf of the Vacuum, Surface and Coatings Group (VSC) Vacuum, Surface and Coatings group* Mandate Design, construction, operation, maintenance and upgrade of
Risk management in scientific research:
Risk management in scientific research: a proposal guided in Project Management Book of Knowledge and Failure Mode and Effects Analysis Pollyana Notargiacomo Mustaro Graduate Program in Electrical Engineering
Experiment #5: Qualitative Absorption Spectroscopy
Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions
Ra, 232 Th and 40 K analysis in sand samples from some beaches of Great Vitória, Espírito Santo, Brazil: Preliminary results.
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 226 Ra, and
Energy and Buildings
Energy and Buildings 59 (2013) 62 72 Contents lists available at SciVerse ScienceDirect Energy and Buildings j our na l ho me p age: www.elsevier.com/locate/enbuild Experimental thermal characterization
The current situation of sanitary landfills in Brazil and the application of economic models
The current situation of sanitary landfills in Brazil and the application of economic models Raul Oliveira Neto, Ph.D. Mining technology course Federal University of Pampa - UNIPAMPA Avenue Anunciação
QUALITY KNOWLEDGE INTEGRATION: A BRAZILIAN COMPARISON ANALYSIS
QUALITY KNOWLEDGE INTEGRATION: A BRAZILIAN COMPARISON ANALYSIS Úrsula Maruyama [email protected] CEFET/RJ, Departamento de Ensino e Administração (DEPEA) Av. Maracanã 229, Maracanã CEP 20.271-110
How To Study The Scientific Production In Brazilian Dentistry
252 Braz Dent J (2008) 19(3): 252-256 A.C. Scarpelli et al. ISSN 0103-6440 Academic Trajectories of Dental Researchers Receiving CNPq s Productivity Grants Ana Carolina SCARPELLI Fernanda SARDENBERG Daniela
SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
TLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 TLD ALBEDO DOSIMETER
Radiation Safety Characteristics of the NOMAD Portable X-ray System
Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT 84097 USA, www.aribex.com 2 Quality
DEVELOPMENT OF A QUALITY MANAGEMENT SYSTEM FOR BRAZILIAN NUCLEAR INSTALLATIONS
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DEVELOPMENT OF A QUALITY
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
THE INTEGRATED CAD/CAM SOLUTION FOR THE WOOD INDUSTRY THE INTEGRATED CAD/CAM/ERP SOLUTION
THE INTEGRATED CAD/CAM SOLUTION FOR THE WOOD INDUSTRY THE INTEGRATED CAD/CAM/ERP SOLUTION ONE CAD/CAM SOLUTION AND ONE PROFESSION: WOOD TopSolid Wood is both a very powerful CAD/CAM solution, and a dedicated
Integration of a fin experiment into the undergraduate heat transfer laboratory
Integration of a fin experiment into the undergraduate heat transfer laboratory H. I. Abu-Mulaweh Mechanical Engineering Department, Purdue University at Fort Wayne, Fort Wayne, IN 46805, USA E-mail: [email protected]
Rem: Revista Escola de Minas ISSN: 0370-4467 [email protected] Escola de Minas Brasil
Rem: Revista Escola de Minas ISSN: 0370-4467 [email protected] Escola de Minas Brasil Queiroz, Gilson; Carvalho, Hermes; Rodrigues, Francisco; Pfeil, Michèle Estimation of friction contribution in the
MONTE CARLO SIMULATION OF THE FULL MULTIPLICITY DISTRIBUTIONS MEASURED WITH A PASSIVE COUNTER
International Conference on Mathematics, Computational Methods & Reactor Physics (M&C 9) Saratoga Springs, New York, May 3-7, 9, on CD-ROM, American Nuclear Society, LaGrange Park, IL (9) MONTE CARLO SIMULATION
Manual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS
A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS W. M. Torres, P. E. Umbehaun, D. A. Andrade and J. A. B. Souza Centro de Engenharia Nuclear Instituto de Pesquisas Energéticas e Nucleares
Diagnostic Exposure Tracking in the Medical Record
Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand
The electrical field produces a force that acts
Physics Equipotential Lines and Electric Fields Plotting the Electric Field MATERIALS AND RESOURCES ABOUT THIS LESSON EACH GROUP 5 alligator clip leads 2 batteries, 9 V 2 binder clips, large computer LabQuest
INJECTION MOLDING COOLING TIME REDUCTION AND THERMAL STRESS ANALYSIS
INJECTION MOLDING COOLING TIME REDUCTION AND THERMAL STRESS ANALYSIS Tom Kimerling University of Massachusetts, Amherst MIE 605 Finite Element Analysis Spring 2002 ABSTRACT A FEA transient thermal structural
HARVESTING AND WOOD TRANSPORT PLANNING WITH SNAP III PROGRAM (Scheduling and Network Analysis Program) IN A PINE PLANTATION IN SOUTHEAST BRAZIL 1
Harvesting and Wood Transport Planning with SNAP III... 831 HARVESTING AND WOOD TRANSPORT PLANNING WITH SNAP III PROGRAM (Scheduling and Network Analysis Program) IN A PINE PLANTATION IN SOUTHEAST BRAZIL
Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 7 Contact: Mr S C Sparks Solent Industrial Estate Tel: +44 (0)1489 790296 Hedge End Fax: +44 (0)1489 790294 Southampton E-Mail: [email protected]
STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE
STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE Antônio C. O. Barroso*, José R. Maiorino, João M. L. Moreira **, José L. Bastos, José E. R. da Silva, Fernando R. de A. Lima, Mitsuo Yamaguchi,
Modern Physics Laboratory e/m with Teltron Deflection Tube
Modern Physics Laboratory e/m with Teltron Deflection Tube Josh Diamond & John Cummings Fall 2010 Abstract The deflection of an electron beam by electric and magnetic fields is observed, and the charge
QUANTITATIVE INFRARED SPECTROSCOPY. Willard et. al. Instrumental Methods of Analysis, 7th edition, Wadsworth Publishing Co., Belmont, CA 1988, Ch 11.
QUANTITATIVE INFRARED SPECTROSCOPY Objective: The objectives of this experiment are: (1) to learn proper sample handling procedures for acquiring infrared spectra. (2) to determine the percentage composition
Study of the Natural Circulation Phenomenon for Nuclear Reactors
Study of the Natural Circulation Phenomenon for Nuclear Reactors Thadeu das Neves Conti, Gaianê Sabundjian, Walmir M. Torres, Luiz A. Macedo, Delvonei A. Andrade, Pedro E. Umbehaun and Roberto N. Mesquita
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND
DOING PHYSICS WITH MATLAB COMPUTATIONAL OPTICS RAYLEIGH-SOMMERFELD DIFFRACTION INTEGRAL OF THE FIRST KIND THE THREE-DIMENSIONAL DISTRIBUTION OF THE RADIANT FLUX DENSITY AT THE FOCUS OF A CONVERGENCE BEAM
Energy Generation with Landfill Biogas
Artigo publicado e apresentado no evento RIO 6 - World Climate & Energy Event, 17a 18 de novembro de 2006, Rio de Janeiro. Energy Generation with Landfill Biogas Coelho, Suani Teixeira 1 ; E-mail: [email protected]
SECTION 13. Multipliers. Outline of Multiplier Design Process:
SECTION 13 Multipliers VMI manufactures many high voltage multipliers, most of which are custom designed for specific requirements. The following information provides general information and basic guidance
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric
THE RESTRUCTURING OF THE QUALITY MANAGEMENT SYSTEM OF CDTN
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 THE RESTRUCTURING OF THE
Appendix A. An Overview of Monte Carlo N-Particle Software
Appendix A. An Overview of Monte Carlo N-Particle Software A.1 MCNP Input File The input to MCNP is an ASCII file containing command lines called "cards". The cards provide a description of the situation
Frequency-dependent underground cable model for electromagnetic transient simulation
Frequency-dependent underground cable model for electromagnetic transient simulation Vasco José Realista Medeiros Soeiro Abstract A computation method for transients in a three-phase underground power-transmission
TEACHING CALCULUS USING E-LEARNING IN A MOODLE PLATFORM
TEACHING CALCULUS USING E-LEARNING IN A MOODLE PLATFORM Caridade C.M.R., Faulhaber M.C., Rosa P.M., Silva P.M., Baeta N.S. Coimbra Institute of Engineering [email protected]; [email protected]; [email protected];
IMPLEMENTATION OF A MANAGEMENT SYSTEM FOR OPERATING ORGANIZATIONS OF RESEARCH REACTORS
IMPLEMENTATION OF A MANAGEMENT SYSTEM FOR OPERATING ORGANIZATIONS OF RESEARCH REACTORS Eduardo Kibrit 1, Afonso Rodrigues de Aquino 2 and Désirée Moraes Zouain 3 1 Centro Tecnológico da Marinha em São
Calorimetry in particle physics experiments
Calorimetry in particle physics experiments Unit n. 8 Calibration techniques Roberta Arcidiacono Lecture overview Introduction Hardware Calibration Test Beam Calibration In-situ Calibration (EM calorimeters)
Pre-Compliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions
PreCompliance Test Method for Radiated Emissions of Automotive Components Using Scattering Parameter Transfer Functions D. Schneider 1*, S. Tenbohlen 1, W. Köhler 1 1 Institute of Power Transmission and
Open Source and Low Cost Remote Monitoring Prototype for Grid-Connected Photovoltaic Generation Systems
1 Open Source and Low Cost Remote Monitoring Prototype for Grid-Connected Photovoltaic Generation Systems Gabriel de Andrade Torelli, André Mendes Martins, Rômulo Mello Alves, Tuliana Pinto Martins, Ivani
One approach to the use of the practices of CMMI- DEV V1.3 level 2 in a process of development of Embedded Systems
One approach to the use of the practices of CMMI- DEV V1.3 level 2 in a process of development of Embedded Systems Magda A. Silvério Miyashiro IMT Instituto Mauá de Tecnologia São Caetano do Sul, Brasil