PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS
|
|
|
- Job Marshall
- 10 years ago
- Views:
Transcription
1 2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS Elaine Wirney Martins and Maria da Penha A. Potiens Instituto de Pesquisas Energéticas e Nucleares (IPEN / CNEN - SP) Av. Professor Lineu Prestes São Paulo, SP [email protected], ABSTRACT The formation of good quality image in nuclear medicine services depends on several factors, including the radiopharmaceutical activity, which must be well determined by a specific apparatus, in perfect operating condition, called activimeter. Therefore, the establishment of a quality control program for measuring the radiopharmaceuticals radioactivity before being administered to the patient is crucial to the safe and effective use of the radiopharmaceuticals used in diagnostic and therapeutical procedures. Two activimeters, belonging by the Laboratório de Calibração do Instrumentos (LCI) of Instituto de Pesquisas Energéticas e Nucleares (IPEN), were evaluated: the secondary standard system NPL-CRC radionuclide calibrator, manufactured by Southern Scientific, and the work standard system CRC -15BT with traceability to National Institute of Standard and Technology (NIST). A set of standard sources of the radionuclides 133 Ba, 57 Co and 137 Cs was used as reference to perform the quality control tests. The precision, accuracy and repeatability tests were in agreement with those established by the CNEN-NE 3.05 Brazilian standard, which recommends the variation limit of up to 10%, 5% and 5% respectively. The results obtained for the reproducibility tests presented a variation always below 1,5%, which means it is within the IEC international standard recommendation of up to 3%. Activimeters response regarding linearity showed concordance between the measured activities and the theoretical curve for both activimeters. 1. INTRODUCTION Due to the large number of procedures performed with the use of ionizing radiation in medicine, it is necessary to use a substantial number of activity meters in order to cover all NMS, and concerns about the development of quality control programs as well as radiological safety are inevitable. These two factors have often been studied, and they are a requisite for the national and international standards, since their main objective is to ensure the precise measurement of radiopharmaceuticals activity before being administered to patients. The activimeter needs to be in perfect operation conditions, otherwise it may present values smaller or larger than the real one, resulting in unnecessary high doses to patients without any benefit, or repeated procedures leading to radiation exposure to the patient as well as individuals involved in the process, thus failing to meet the radioprotection principles (1,2).
2 The activimeter has a unique feature compared to other systems which use ionization chambers, the electronic circuit which allows response to be showed directly in units of activity (3). Several factors influence the accuracy of the dose given by activimeters. Before becoming available for sale, the activimeter, after its manufacture, is submitted to calibration measurements using a set of certified radiation sources. The manufacturer limits its initial accuracy to a range between +/- 1% and +/- 5%. Over time this accuracy can vary due to pressure changes in the gas of its chamber and electronic attraction. To ensure the proper operation of the instrument, due to these variations, regular tests were established to be applied at specific intervals. There are three types of tests which are part of a quality control program: (I) acceptance tests, (II) reference tests, (III) operational tests. (I) Acceptance tests should be performed after the acquisition of a new activimeter by NMS. Some of the acceptance tests are: physical inspection (display lighting, battery, support, wiring, etc), precision, accuracy, activity response linearity and background radiation. (II) To verify the apparatus performance, reference tests are periodically performed, as listed in Table 1, including accuracy, precision, linearity and background radiation tests. Those tests must conform to acceptance limits established by the CNEN standard (1). TABLE 1: Tests established by CNEN for quality control of activimeters. Test Precision Accuracy Linearity Reproducibility Frequency Bianual Bianual Bianual Anual Recommended sources Acceptance limits 57 Co, 133 Ba or 137 Cs 10% 57 Co, 133 Ba or 137 Cs 5% 99m Tc 20% 57 Co or 133 Ba 5% (III) Operational tests include reproducibility and background radiation tests (4). 2. MATERIALS AND METHODS In this study the following activimeters were tested: the reference system, the secondary standard, NPL-CRC radionuclide calibrator, and the tertiary system, Capintec basic CRC - 15BT. They consist of two parts: the console in which the electrometer is situated, and the ionization chamber, both by Capintec, as presented in Figure 1. The reference system is a secondary standard, model NPL-CRC radionuclide calibrator, serial number , manufactured by Southern Scientific plc, with traceability to National Physical Laboratory (NPL). The system to be used as work standard is a model CRC -15BT,
3 serial number , calibrated by Accredited Dosimetry Calibration Laboratory of Medical Radiation Research Center University of Wisconsin, with traceability to National Institute of Standard and Technology (NIST). (a) (b) Figure 1. a) Reference system: NPL-CRC, b) Work system: CRC -15 BT. For the quality control tests, a set of standard sources were used as reference, provided by the Centro de Tecnologia das Radiações (CTR) of IPEN. Three sealed sources: 133 Ba, 57 Co e 137 Cs, are presented in Figure 2. Figure 2: Set of standard Eckert & Ziegler sources. Each standard source is supplied in a 27 ml polyethylene vials with uniform distribution of the active element in 20 ml of epoxy resin, composed of a density of approximately 1,0 g/cm 3. These sources were manufactured by the German company Eckert & Ziegler Isotope Products (E&Z), with traceability certificates to National Institute of Standards and Technology (NIST). In Table 2 its characteristics are related. Radionuclides TABLE 2: Main characteristics of sources E&Z. Primary photon energy (kev) Activity (MBq) Uncertainty (%) T 1/2 (days) Reference date 133 Ba 81; ± /11/09 57 Co ± /03/ Cs ± /11/09
4 Precision and accuracy tests were applied after stabilization of the device, 10 readings were performed with a 30 seconds time interval between them. The precision test was calculated using the percentage deviation (P) between individual activity measure (A i ) and the measurements mean ( A ) according to equation 1. ( A A P (%) = 100 x i ) A (1) In which: A = Arithmetic mean of the activity measures A i = Activity of the individual source The accuracy test (E) was calculated using the percentage deviation between the average of the activity measures ( A ) relative to the mean of the first ten measurements in the standard system with correction for the decay ( A ), given by the equation 2 (1,3,5,6). 10 E (%)= 100 x ( A A A ) (2) In which: A = Arithmetic mean of the activity measurements A 10 = Arithmetic mean of the first ten measurements in the standard system with correction for the decay on the date in which the test was realized. The repeatability and reproducibility tests were applied equally in both activimeters to ensure the quality control of each apparatus regarding stability, using a 137 Cs source. For the linearity test a 99m Tc source was used, in a homogeneous mixture of 4 ml of saline solution into a penicilin vial with maximum volume of 20 ml and initial activity of 5.1 GBq. 3. RESULTS Precision: the maximum percentage deviation between the individual activity measure (A i ) with respect to the measured mean activities (Ā) was less than 0.32% for all the tests performed, not exceeding the recommended limit of ± 5%, thus showing satisfactory results with reliability of 95%. Accuracy: the acceptance limit is ± 10% with a reliability of 90% recommended by CNEN standard (1). All tests performed were satisfactory, not exceeding the recommended limit. In Table 3, it is possible to observe the results of accuracy test, that is the percentage deviation between mean of activity measured (10 measurements) and the standard source activity with the application of the decay correction factor.
5 TABLE 3: Results obtained for the accuracy test with reference sources in both activimeters. NPL-CRC CRC -15BT Ā (MBq) A (MBq) Accuracy (%) Ā (MBq) A (MBq) Accuracy (%) Bário Césio Cobalto For the stability test a sequence of measurements was performed. 10 first measurements were employed to determine the value to be used as reference, the maximum coefficient of variation obtained was 1% (acceptable variation is 5%) (7). Figure 3 shows the reference and work system stability over time (standard deviation of the average), respectively. For the reference system, maximum variation of the results was 1% throughout the testing period and for the work system, 1.5%, both within the acceptance limits recommended by international protocols (8,9). ACKNOWLEDGMENTS Relative activity This template was adapted from the template for the American Nuclear Society Topical Meeting in Mathematics & Computations, Gatlinburg, TN, 2003 posted on the Internet. Acknowledge the help of colleagues, and sources of funding, if you wish. Relative activity Measurement number Measurement number Figure 3: Stability test using 137 Cs standard source. a) Reference system: NPL-CRC, b) Work system: CRC -15 BT. Linearity: activimeters response was accompanied by the decay of a 99m Tc source, with initial activity of 5,01 (GBq), as shown in Figure 4. Date and time of the measurements were registered in time intervals of one hour between them for a period of 56 hours.
6 Activity (Gbq) Linear (T) Linear (CRC) Linear (NPL) Time (h) Figure 4: Linearity test The reference system (NPL-CRC ) and the work system (CRC -15BT) presented results within the acceptance limit recommended by the manufacturer s manual (± 10%), so both activimeters revealed linear responses between the values of measured activity which is in accordance with the experimental values. 4. CONCLUSIONS For the precision, accuracy and reproducibility tests, the Brazilian standard for nuclear medicine services, CNEN-NE 3.05, was used as reference, which recommends variation limits of up to 10%, 5% and 5% respectively. Results obtained for repeatability presented a variation lower than 1,5%, within the recommendation of up to 3% of IEC international standard (7). Activimeters response regarding linearity showed concordance between measured activities and the theoretical curve for both activimeters. Therefore, the quality control program implemented for the LCI activimeters demonstrated that both instruments are in good operation condition and the CRC - 15 BT apparatus can be used as work standard, employing the appropriate corrections. ACKNOWLEDGMENTS The authors thank to Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq), to Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP), to Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES) and to Ministério de Ciência e Tecnologia (MCT), Project: Instituto Nacional de Ciência e Tecnologia (INCT) em Metrologia das Radiações na Medicina, the partial financial support.
7 REFERENCES 1. Comissão Nacional de Energia Nuclear. CNEN-NN Requisitos de radioproteção e segurança para serviços de medicina nuclear. Resolução CNEN-10/96. Diário Oficial da União, (1996). 2., A. C. M. Bessa, A. M. Costa, L. V. E Caldas. Levantamento do controle de qualidade de calibradores de dose de radiofármacos em serviços de medicina nuclear na cidade de São Paulo. Radiol. Bras. v. 51, p (2008). 3. A.M. Costa. Métodos de calibração e de intercomparação de calibradores de dose utilizados em serviços de medicina nuclear. São Paulo. Dissertação (Mestrado) Universidade de São Paulo. IPEN/USP (1999). 4. International Atomic Energy Agency. Quality control of nuclear medicine instruments. TecDoc-602, (1991). 5. S.G.P. Cecatti. Desenvolvimento de sistemas Tandem de activímetros e estabelecimento de dosimetria beta em serviços de medicina nuclear. São Paulo. Tese (Doutorado) Universidade de São Paulo. IPEN/USP (2004). 6. M.C.F Fragoso. Implantação de um programa de intercomparação de medidas de atividade de radiofármacos utilizados em serviços de medicina nuclear no Nordeste. Recife. Tese (Mestrado) Universidade Federal de Pernambuco (2010). 7. International Electrotechnical Comission. Medical electrical equipment Dosimeters with ionization chambers and or semi- conductor detectors as used in X-ray diagnostic imaging. Geneva. IEC (1997). 8. National Physical Laboratory. Protocol for establishing and maintaining the calibration of medical radionuclide calibrators and their quality control. Middlesex, United Kingdom. A National Measurement Good Practice Guide, n. 93, NPL (2006). 9. Laboratoire National Henri Becquerel. Guide d utilisation et de controle qualité dês activimètres. França. Societé française de radiopharmacie (2006).
DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 DEVELOPMENT AND IMPLEMENTATION
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level Eduardo de Lima Corrêa 1, Vitor Vivolo 1, Maria da Penha A. Potiens 1 1 Instituto de Pesquisas Energéticas e Nucleares
IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 IMPROVEMENT OF GAMMA
DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DOSIMETRIC CHARACTERIZATION
Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom
Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Fernanda B. C. Nonato, Claudinei T. Cescon and Linda V. E. Caldas, Instituto de Pesquisas Energéticas e Nucleares,
DEVELOPMENT OF A QUALITY MANAGEMENT SYSTEM FOR BRAZILIAN NUCLEAR INSTALLATIONS
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DEVELOPMENT OF A QUALITY
TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1
TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT
DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
Ra, 232 Th and 40 K analysis in sand samples from some beaches of Great Vitória, Espírito Santo, Brazil: Preliminary results.
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 226 Ra, and
THE RESTRUCTURING OF THE QUALITY MANAGEMENT SYSTEM OF CDTN
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 THE RESTRUCTURING OF THE
Simone Alberigi and Brigitte Roxana Soreanu Pecequilo
27 International Nuclear Atlantic Conference - INAC 27 Santos, SP, Brazil, September 3 to October, 27 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-8-99141-2-1 CAVES OF PARQUE ESTADUAL TURÍSTICO
TLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 TLD ALBEDO DOSIMETER
IMPLEMENTATION OF A MANAGEMENT SYSTEM FOR OPERATING ORGANIZATIONS OF RESEARCH REACTORS
IMPLEMENTATION OF A MANAGEMENT SYSTEM FOR OPERATING ORGANIZATIONS OF RESEARCH REACTORS Eduardo Kibrit 1, Afonso Rodrigues de Aquino 2 and Désirée Moraes Zouain 3 1 Centro Tecnológico da Marinha em São
SHORT-TERM STABILITY TEST FOR URANIUM SOIL CANDIDATE A REFERENCE MATERIAL
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
NUCLEAR ENERGY EDUCATION SCENARIO AROUND THE WORLD
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 NUCLEAR ENERGY EDUCATION SCENARIO
STUDY OF MAGNETISM IN GDTIN (T = NI, PD, CU) COMPOUNDS BY PERTURBED GAMMA-GAMMA ANGULAR CORRELATION SPECTROSCOPY
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 STUDY OF MAGNETISM IN GDTIN
Test Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 Test Section for Experimental
ASPECT OF SUCROSE AND ITS MONOMERS FROM SUGARCANE JUICE SUBMITTED TO DIFFERENT DOSES OF COBALT-60 IRADDIATION
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 ASPECT OF SUCROSE AND ITS
Quality Assurance Testing of AtomLab Dose Calibrators
(Manufacturer s* Instructions) Introduction The following procedures are the manufacturer s instructions for qualifying AtomLab 100, 100+, 200, 300, 400, and 500 dose calibrators in accordance with NRC
CFD SIMULATION OF IPR-R1 TRIGA SUBCHANNELS FLUID FLOW
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 CFD SIMULATION OF IPR-R1 TRIGA
STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE
STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE Antônio C. O. Barroso*, José R. Maiorino, João M. L. Moreira **, José L. Bastos, José E. R. da Silva, Fernando R. de A. Lima, Mitsuo Yamaguchi,
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the
Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material)
Radiation Sciences Nuclear Medicine Technology Program Hot Lab Log Book (Material) Table of Content Receiving a Radioactive Package Constancy Test Daily Survey Weekly Wipes Accuracy Test Linearity Geometric
PROCESSING OF AISI M2 HSS WITH ADDITION OF NbC BY MECHANICAL ALLOYING USING TWO DIFFERENT TYPES OF ATTRITOR MILLS
Materials Science Forum Vols. 660-661 (2010) pp 17-22 Online available since 2010/Oct/25 at www.scientific.net (2010) Trans Tech Publications, Switzerland doi:10.4028/www.scientific.net/msf.660-661.17
Public Funding and the Beginning of a New Era in Higher Education in Brazil
80 Comparative & International Higher Education 5 (2013) Public Funding and the Beginning of a New Era in Higher Education in Brazil Danilo de Melo Costa a,* a Federal University of Minas Gerais, Brazil
EVALUATION OF ANALYSIS METHOD STANDARDLESS BY WDXRF AND EDXRF OF ALUMINUM POWDER USED IN MTR TYPE FUEL
EVALUATION OF ANALYSIS METHOD STANDARDLESS BY WDXRF AND EDXRF OF ALUMINUM POWDER USED IN MTR TYPE FUEL Valdirene O. Scapin 1 ; Vera L. R. Salvador 1 ; Marycel E. B. Cotrim 1 ; Maria A. F. Pires 1 ; Marcos
What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?
Congreso Internacional Conjunto Cancún 2004 LAS/ANS-SNM-SMSR/International Joint Meeting Cancun 2004 LAS/ANS-SNM-SMSR XV Congreso Anual de la SNM y XXII Reunión Anual de la SMSR/XV SNM Annual Meeting and
Dedicated computer software to radiation dose optimization for the staff performing nuclear medicine procedures
NUKLEONIKA 2012;57(4):497 502 ORIGINAL PAPER Dedicated computer software to radiation dose optimization for the staff performing nuclear medicine procedures Katarzyna Matusiak, Jakub Kosek Abstract. Nuclear
A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS
A MTR FUEL ELEMENT FLOW DISTRIBUTION MEASUREMENT PRELIMINARY RESULTS W. M. Torres, P. E. Umbehaun, D. A. Andrade and J. A. B. Souza Centro de Engenharia Nuclear Instituto de Pesquisas Energéticas e Nucleares
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
DOSE RATES EVALUATION ON ENVIRONMENT AROUND ALMIRANTE ALVARO ALBERTO NUCLEAR POWER STATION AFTER ANGRA 2 NUCLEAR POWER PLANT S OPERATION
2007 International Nuclear Atlantic Conference INAC 2007 Santos, SP, Brazil, September 29 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR ABEN DOSE RATES EVALUATION ON ENVIRONMENT AROUND ALMIRANTE
VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 Associação Brasileira de Energia Nuclear - ABEN ISBN: 978-85-99141-02-1 VoxelDose: A SOFTWARE
VIRTUAL INSTRUMENTATION ON MOBILE DEVICES FOR DEPLOYMENT IN NUCLEAR INSTALLATIONS
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 VIRTUAL INSTRUMENTATION ON MOBILE
One approach to the use of the practices of CMMI- DEV V1.3 level 2 in a process of development of Embedded Systems
One approach to the use of the practices of CMMI- DEV V1.3 level 2 in a process of development of Embedded Systems Magda A. Silvério Miyashiro IMT Instituto Mauá de Tecnologia São Caetano do Sul, Brasil
PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT
RP&AD /Remote QA/ 01 PROFORMA FOR ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT BHABHA ATOMIC RESEARCH CENTRE Radiological Physics & Advisory Division CT & CRS Building,
How To Test Water Penetration On Concrete Block And Mortar
WATER PENETRATION TEST ON CONCRETE BLOCK MASONRY Vilató, Rolando Ramirez 1 1 PhD, Prof., Mackenzie Presbyterian University, School of Engineering, São Paulo, [email protected] This paper considers
The Calibration Chain: Role of BIPM,
The Calibration Chain: Role of BIPM, PSDLs and ADCLs J. Seuntjens and M. McEwen McGill University & National Research Council CANADA Clinical Dosimetry for Radiotherapy, 2009 AAPM Summer School PSDL ADCL
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
A MULTIELEMENT ANALYSIS OF NATURAL RADIONUCLIDES IN AN AREA OF PERNAMBUCO - BRAZIL
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 A MULTIELEMENT ANALYSIS
Radium-223 Therapy: Handling & Radiation Safety Issues
Nuclear Physics Enterprises Radium-223 Therapy: Handling & Radiation Safety Issues Jeffry A. Siegel, Ph.D. President & CEO 2014 Meeting Mid-Atlantic States Radiation Control Programs Malvern, PA NRC s
NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM
NEW YORK STATE DEPARTMENT OF HEALTH BUREAU OF ENVIRONMENTAL RADIATION PROTECTION EXTERNAL BEAM & BRACHYTHERAPY QUALITY ASSURANCE PROGRAM AUDIT FORM Purpose: To provide licensees and registrants with a
Research Topics in Renewable Energies in Brazil
WORKSHOP BRAZIL-ITALY ROME 29-31/03/2012 Research Topics in Renewable Energies in Brazil Prof. Dr.-Ing. João Tavares Pinho Full Professor at the Universidade Federal do Pará Head of the Grupo de Estudos
Evaluation of the image quality in computed tomography: different phantoms
Artigo Original Revista Brasileira de Física Médica.2011;5(1):67-72. Evaluation of the image quality in computed tomography: different phantoms Avaliação da qualidade de imagem na tomografia computadorizada:
STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION
Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration
Syringe Calibration Factors and Volume Correction Factors for the NPL Secondary Standard Radionuclide Calibrator.
Syringe Factors and Factors for the NPL Secondary Standard Radionuclide Calibrator. D K Tyler, M J Woods September 2002 Syringe Factors and Factors for the NPL Secondary Standard Radionuclide Calibrator.
A COMPUTATIONAL PHANTOM OF HEAD AND NECK - SISCODES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 A COMPUTATIONAL PHANTOM
Lasers and Specturbility of Light Scattering Liquids - A Review
Artigo Original Revista Brasileira de Física Médica.2011;5(1):57-62. Reproducibility of radiant energy measurements inside light scattering liquids Reprodutibilidade de medidas de energia radiante dentro
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
Study of the Natural Circulation Phenomenon for Nuclear Reactors
Study of the Natural Circulation Phenomenon for Nuclear Reactors Thadeu das Neves Conti, Gaianê Sabundjian, Walmir M. Torres, Luiz A. Macedo, Delvonei A. Andrade, Pedro E. Umbehaun and Roberto N. Mesquita
WebBee a Web-based Information Network on Bees
WebBee a Web-based Information Network on Bees A.M. Saraiva 1 V.L. Imperatriz-Fonseca 2 R.S. Cunha 1 E.A.Cartolano-Júnior 1 {antonio.saraiva, renato.cunha, etienne.cartolano}@poli.usp.br., [email protected]
Radiation Rule Guide. Veterinary Facilities. Hawaii Administrative Rules, Chapter 11-45, Radiation Control. Hawaii State Department of Health
Radiation Rule Guide Veterinary Facilities Hawaii Administrative Rules, Chapter 11-45, Radiation Control Hawaii State Department of Health MAY 2009 Radiation Rule Guide for Veterinary Facilities Hawaii
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
centre for radiation protection
centre for radiation protection To excel in radiation science so as to: enforce and promote the radiation safety of workers, the public and the environment; ensure that irradiating apparatus and nuclear
AP Physics 1 and 2 Lab Investigations
AP Physics 1 and 2 Lab Investigations Student Guide to Data Analysis New York, NY. College Board, Advanced Placement, Advanced Placement Program, AP, AP Central, and the acorn logo are registered trademarks
HEALTH PROFESSIONS ACT 56 OF 1974
HEALTH PROFESSIONS ACT 56 OF 1974 REGULATIONS DEFINING THE SCOPE OF THE PROFESSION OF RADIOGRAPHY Published under Government Notice R2326 in Government Gazette 5349 of 3 December 1976 and amended by GN
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy
An Innovative Method for Dead Time Correction in Nuclear Spectroscopy Upp, Daniel L.; Keyser, Ronald M.; Gedcke, Dale A.; Twomey, Timothy R.; and Bingham, Russell D. PerkinElmer Instruments, Inc. ORTEC,
Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY
Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY A. Mrozik, B. Marczewska, P. Bilski, W. Gieszczyk Institute of Nuclear Physics PAN,
The International classification for nursing practice: Participation of Brazilian Nurses in the Project of the International Council of Nurses*
Palestra The International classification for nursing practice: Participation of Brazilian Nurses in the Project of the International Council of Nurses* Classificação Internacional para a Prática de Enfermagem:
MAJOR AND TRACE ELEMENTS IN GEOLOGICAL SAMPLES FROM ITINGUSSÚ BASIN IN COROA-GRANDE, R J.
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 MAJOR AND TRACE ELEMENTS
Natural and Man-Made Radiation Sources
Natural and Man-Made Radiation Sources All living creatures, from the beginning of time, have been, and are still being, exposed to radiation. This chapter will discuss the sources of this radiation, which
DATABASE IMPLEMENTATION TO A FLUIDIZED CRACKING CATALYTIC-FCC PROCESS
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DATABASE IMPLEMENTATION
OPERATIONAL STANDARD
1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.
arxiv:1301.4944v1 [stat.ml] 21 Jan 2013
Evaluation of a Supervised Learning Approach for Stock Market Operations Marcelo S. Lauretto 1, Bárbara B. C. Silva 1 and Pablo M. Andrade 2 1 EACH USP, 2 IME USP. 1 Introduction arxiv:1301.4944v1 [stat.ml]
RESULTS OBTAINED IN MEASUREMENTS OF Rn-222 WITH THE ROMANIAN STANDARD SYSTEM*
NUCLEAR PHYSICS RESULTS OBTAINED IN MEASUREMENTS OF Rn-222 WITH THE ROMANIAN STANDARD SYSTEM* M. SAHAGIA 1, **, A. LUCA 1, A. C. WATJEN 1, A. ANTOHE 1, C. IVAN 1, D. STANGA 1, C.VARLAM 2, I. FAURESCU 2,
A RELAP5 STUDY TO IDENTIFY FLOW REGIME IN NATURAL CIRCULATION PHENOMENON
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte, MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 A RELAP5 STUDY TO IDENTIFY
Lake at IPEN location
Mission Our mission is to improve the Brazilian people quality of life by producing scientific knowledge, developing technologies and services and promoting human resources for nuclear and correlated areas.
José Egidio M.Tondin 1, Mario O. de Menezes 1, Marina B.A.Vasconcellos 1 and João A.Osso Jr. 1 ABSTRACT 1. INTRODUCTION
PROSPECTION OF IMPLEMENTATION OF DISTANCE LEARNING AT IPEN CNEN/SP FOR THE COURSE ON FUNDAMENTALS OF NUCLEAR PHYSICS USING INFRASTRUCTURE OF FREE SOFTWARE José Egidio M.Tondin 1, Mario O. de Menezes 1,
PHYS 1624 University Physics I. PHYS 2644 University Physics II
PHYS 1624 Physics I An introduction to mechanics, heat, and wave motion. This is a calculus- based course for Scientists and Engineers. 4 hours (3 lecture/3 lab) Prerequisites: Credit for MATH 2413 (Calculus
THE EFFECT OF THE HDDR PROCESS ON THE PRODUCTION OF Pr-Fe-Co-B-Nb SINTERED MAGNETS
THE EFFECT OF THE HDDR PROCESS ON THE PRODUCTION OF Pr-Fe-Co-B-Nb SINTERED MAGNETS E. A. Ferreira, J. C. S. Casini, E. A. Périgo, R. N. Faria, H. Takiishi Av. Prof. Lineu Prestes, 2242 - CEP 05508-000,
THE IMPACT OF EXCHANGE RATE VOLATILITY ON BRAZILIAN MANUFACTURED EXPORTS
THE IMPACT OF EXCHANGE RATE VOLATILITY ON BRAZILIAN MANUFACTURED EXPORTS ANTONIO AGUIRRE UFMG / Department of Economics CEPE (Centre for Research in International Economics) Rua Curitiba, 832 Belo Horizonte
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
Improvement and storage of conventional mammography images using techniques of digital image processing
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-13 Improvement and storage of conventional mammography images using techniques of digital image processing M. M. A. Silva a ; J. W. Vieira b,c
APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMMAND: PERMIT NUMBER: TYPE OF ACTION:
APP C1 NAVAL RADIOACTIVE MATERIAL PERMIT REVIEW CHECKLIST NUCLEAR MEDICINE COMM: PERMIT NUMBER: TYPE OF ACTION: NAVY MARINE CPS PUBLIC HEALTH CENTER REVISED NOVEMBER 2007 REVIEWER/DATE: SECOND REVIEW/DATE:
Antoine Henri Becquerel was born in Paris on December 15, 1852
Discovery Antoine Henri Becquerel was born in Paris on December 15, 1852 Summit Environmental Technologies, Inc. Analytical Laboratories 3310 Win Street Cuyahoga Falls, Ohio 44223 Fax: 1-330-253-4489 Call
CHARACTERIZATION OF COLOR CENTERS IN QUARTZ INDUCED BY GAMMA IRRADIATION
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 CHARACTERIZATION
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration
Implementation Page #: Page 1 of 10 Last Reviewed/Update 1. Purpose and Scope The purpose of this document is to describe the Medical Physics and Radiation Safety program at Boston University (BU) and
TRAINING PROGRAMS IN MEDICAL PHYSICS IN THE UNITED STATES
11 I '* / TRAINING PROGRAMS IN MEDICAL PHYSICS IN THE UNITED STATES Lawrence H. Lanzl, Ph.D. Professor of Medical Physics Department of Radiology and The Franklin McLean a '" Memorial Research Institute,
Principles of dosimetry The ionization chamber
Principles of dosimetry The ionization chamber FYS-KJM 4710 Audun Sanderud Department of Physics Ionometry 1) Ionometry: the measurement of the number of ionizations in substance The number of ionizations