Radiation Safety Characteristics of the NOMAD Portable X-ray System
|
|
|
- Randell Gray
- 10 years ago
- Views:
Transcription
1 Radiation Safety Characteristics of the NOMAD Portable X-ray System D. Clark Turner 1, Donald K. Kloos 1, Robert Morton 2 1 Aribex, Inc., 754 South 400 East, Orem, UT USA, 2 Quality and Regulatory Services, Lincoln, CA USA I. Introduction An important part of a modern dental exam is radiographic imaging. While there are risks to the use of ionizing radiation, the diagnostic benefits have been determined to outweigh the risks; thus the procedures have now become routine during regular exams. In a recent publication on radiation safety in dentistry (NCRP 145, page 45), "It seems reasonable to conclude that radiation-related risks to dental patients and dental x-ray equipment operators are numerically very small and may be zero." There are many regulations and guidelines governing the development and use of diagnostic x-ray equipment. In the USA, regulations such as 21CFR , promulgated by the FDA, provide standards for equipment performance 1. Internationally, similar standards are provided in the IEC document 2. Voluntary guidelines have been provided by the National Council on Radiation Protection and Measurement (NCRP) for use of radiation protection in dentistry 3. Many states regulate the use of x-ray equipment per the CRCPD's Suggested State Regulations for the Control of Radiation 4. assembly, preferably behind a barrier (SSRCR, F7c.2 ). These concerns stem from three areas: A) Leakage radiation that is transmitted through the primary system shielding to irradiate the operator, B) Backscattered radiation from the patient (particularly the skin and teeth for dental x-rays) that can impinge on the operator, and C) Inadequate image quality will be obtained if the x-ray tube head moves during the exposure time, requiring a second irradiation of the patient. The intent of this paper is to show that the potential risks of radiation exposure to the operator and patient from NOMAD are sufficiently small as to more than meet the regulatory standards for stationary equipment, and to meet the intent of the "as low as reasonably achievable" (ALARA) principle when considering the economic and social factors. An important consideration is: are these older guidelines still applicable as x-ray equipment technology has continued to evolve and improve? For example, a new handheld, battery-powered, x-ray system has been developed for use in intra-oral radiographic imaging. The NOMAD system is shown in Figure 1 below. This technology is disruptive, as it seems on first consideration to be inconsistent with the regulations and guidelines mentioned above. There are several concerns in the guidelines that must be specifically addressed when considering the use of NOMAD as a diagnostic radiation : 1) Normally, the x-ray tube assembly must be able to be mounted so that it is not held in the operators' hand (SSRCR, F4g). 2) The operator must trigger the x-ray exposure from a distance of 2m from the x-ray tube Figure 1. NOMAD portable handheld x-ray system. II. Other Handheld X-ray Systems The idea of handheld portable x-ray equipment is not new. There are a number of other x-ray tube-based handheld systems. For example, tube-based x-ray fluorescence (XRF) analyzers have been developed in the last 10 years for use in materials analysis, scrap metal sorting, lead in paint analysis, and other applications. 5
2 An example of a battery-powered unit is shown in Figure 2. Systems have been developed for portable imaging, including one intra-oral dental system that is FDA cleared for marketing to the US military. 6 A system photo is shown in Figure 3. While portable, this system still requires connection to AC electrical supply. Figure 2. NITON handheld battery-powered XRF analyzer. Figure 3. Dental EZ Portable HDX Intra-oral X-ray System. III. NOMAD System Description The NOMAD system is a handheld, battery-operated, portable device. It is powered by a 14.4 volt rechargeable battery. The unit is designed with reduced weight and size for easy manipulation by the operator. The system specifications typical for the unit are given in Table 1 below. Table 1. Summary specifications for NOMAD. PARAMETER VALUE Anode voltage 60kV Anode current 2.3 ma Exposure time range s Focal spot size 0.4mm Minimum inherent filtration 1.5mm Al equivalent Source to skin distance 20cm X-ray field 60mm round Maximum duty cycle 1s:60s Total weight <4kg There are several features of the NOMAD that address the safety concerns mentioned in the introduction. Refer to Figure 4 for graphical interpretation of these features as they are discussed. First, the high-voltage power supply is a highfrequency DC unit, which reduces the dose to the patient. More will be said about this later. Second, the x-ray tube is surrounded by compounds of heavy metals to reduce the leakage radiation from the tube. (These compounds are not Pb, but a proprietary mixture designed to provide equivalent or better shielding.) The beam limiting device (exit cone) is lined with Pb. This would not normally be required in a dental unit, as the exit cone is defined by Pb apertures closer to the tube. However, in our case, backscattered radiation from the patient could travel through the exit cone and impinge on the operators' hand. This backscattered radiation is absorbed in the lead-lined cone. Backscattered radiation is also absorbed in a Pb-filled acrylic shield attached at the end of the exit cone. This shield has a Pb-equivalent of 0.5mm thickness, and protects the operators' torso, hands, face, and gonads from backscattered radiation from the patient's face and teeth. NOMAD has the other beam quality and safety features normally specified for dental intra-oral systems. The minimum inherent filtration in the x-ray beam is at least 1.5mm Al equivalent. The -to-skin distance is a minimum of 20 cm. The radiographic technique factors of peak potential and anode current are both fixed at 60kV and 2.3mA respectively. The only operator-adjustable parameter is the irradiation time, which varies from 0.01 to 0.99 seconds. IV. Radiation Protection of the Patient The main consideration in any radiographic procedure involving live patients is to reduce the dose to the patient as much as possible while still achieving the diagnostic goals. Intra-oral dental radiography has been shown to be safe and effective, and the radiation risks are low relative to the patient benefits. 7 NOMAD was designed to operate with technique factors that help to maintain this safe operating region: High-frequency, constant potential, has been shown to reduce patient dose by up to 1/3 per image. 8 System is optimized for digital sensors at 60kV. Digital sensors reduce the dose per image by 1/4th to 1/10th compared with conventional film. Beam is limited to 6cm diameter, rather than the allowed 7cm. This smaller irradiation area reduces the patient dose by 25% without significantly compromising the beam aiming capability
3 X-RAY SOURCE FOCAL SPOT Fixed BEAM LIMITING DEVICE Backscatter Radiation Shielding GEOMETRICAL RADIATION FIELD due to BEAM LIMITING DEVICE PRIMARY PROTECTIVE SHIELDING PERMANENT FILTRATION SOURCE to Skin Distance PATIENT plane Figure 4. Cross-section of the NOMAD showing the relevant safety features. V. Radiation Protection of the Operator The effective radiation exposure to the operator is the sum of any radiation that is leaking from the x-ray tube assembly (Leakage Radiation) and any radiation that scatters from the patient or any objects in the room that are in the x-ray field (Backscattered Radiation) and gets back to the operator's body. The normal practice is to provide a means for the operator to trigger the x-ray irradiation from a position outside the room with the x-ray. Alternatively, another approach is to provide a physical barrier between the operator and the. These barriers are specified to be 2 meters in height to protect the tallest operators. a. Leakage Radiation This is the highest concern. Since the operator is holding the x-ray assembly the principle of "distance" as a safety factor cannot apply. The USFDA regulations state that the maximum permissible radiation leakage from the x-ray assembly is 100mR/hour measured at a distance of 1 meter from the x-ray. To find possible "hot spots" of leakage, the limit is specified to be averaged over an area of 100cm 2, and it should be measured at enough sites to approximate a sphere around the instrument. The IEC specification is even more rigorous at 0.25mGy/hr (25mR/hr) as the limit. To characterize the NOMAD, measurements of leakage radiation were made at 11 different sites to approximate a sphere around the instrument. These sites are shown in Figure 5 below. All exposure measurements were made using a RadCal MDH Model 1015 x-ray monitor S/N: 1535 (last NIST-traceable, MQSA required calibration on July 30, 2004). Leakage exposure measurements were made using a pancake probe S/N: 5464 with the probe placed at 1 meter from the focal spot of the x-ray. Measurements were made on 3 different NOMAD instruments. As can be seen from the first 3 columns in Table 2, no x-radiation could be detected at 1 meter distance. The probe was then positioned at 5 cm from the case housing, and the leakage radiation measurements were performed on two of the instruments. These data are reported in the next section of Table 2. The results are still significantly below the FDA and IEC limits (100mR/hr and 25mR/hr respectively), even at this short distance. Finally, to verify this result on a statistically significant sample, 10 NOMAD instruments were tested for leakage at the case with a Victoreen 451 Survey Meter with a 20cm 2 lead aperture over the detector. The results were normalized to 100cm 2 area, and the average and maximum at each of the 11 sites are reported in the last columns of Table 2. These results suggest that there is no significant leakage radiation from the NOMAD design
4 E H K A B L C D G I M Figure 5. Measurement sites for leakage radiation tests. Table 2. Radiation Leakage Measurements. (See Figure 2 for measurement sites.) Serial # Units - Site 100 cm from 100 cm from 100 cm from Distance from 5 cm from case 10 Units Maximum Average 5 cm from case At the case At the case mr/hr mr/hr mr/hr Units mr/hr mr/hr mr/hr mr/hr A ND ND ND ND ND B ND ND ND ND ND C ND ND ND D ND ND ND E ND ND ND G ND ND ND H ND ND ND I ND ND ND K ND ND ND 0.12 ND ND ND L ND ND ND ND ND ND ND M ND ND ND ND ND Max ND ND ND ND None detected. Detection limit is 0.001mR per exposure (0.06mR/hour at 1:60 duty cycle and maximum exposure time of 0.99 seconds.) Table 3. Radiation Leakage Measurements from traditional stationary dental systems. A A Unit # B B 100 cm from 5 cm from case Distance from 100 cm from 5 cm from Site case mr/hr mr/hr Units mr/hr mr/hr A B C D E G 24.0 H I K L 1.26 ND M Max
5 An interesting consideration is how NOMAD compares to conventional stationary x-ray systems with respect to leakage radiation. Two different stationary dental intra-oral x-ray systems were measured with the Victoreen survey meter at 5 cm from the case. The location showing the highest leakage was measured again at a distance of 1 meter from the focal spot. The results are reported in Table 3 above. It is interesting to note that the readings are generally higher than those of the NOMADs, and that the highest reading recorded, almost 40mR/hr at the case, was on one of the stationary systems. b. Backscatter Radiation An important consideration, not addressed in the FDA regulations, but included in the Annex of the IEC standard, is the level of backscatter radiation incident on the operator of an x-ray system. The IEC standard defines a "zone of significant occupancy" within which the operator can be standing. This standard is specifically designed for fluoroscopic applications where the operator would be standing next to the patient, but it can be applied to NOMAD since the operator will be holding the and standing next to the patient. Figure 6 shows the geometrical definition of the zone of occupancy. Based on the size of the backscatter shield, a cone of protected space envelopes the NOMAD instrument and the area where the operator is standing. To test this zone, measurements of the backscattered radiation were made at the plane defined by the back of the NOMAD instrument (marked with a scale in Figure 6.) The sample was a dental phantom prepared for the NEXT test protocol 9 by Cardinal Health, catalog # The density and composition of the plastic in this phantom is designed to permit an accurate simulated clinical image evaluation. Measurements were made with and without the backscatter shield in place. The chart in Figure 6 shows the recorded radiation measurements as a function of height from the floor, with the NOMAD positioned at 1 meter above the floor. The measurement probe was the Radcal MDH pancake probe with active area of 100cm 2. The scattered radiation recorded in a 1 second exposure was normalized to one hour of operation by multiplying the result by 60 exposures per hour at the maximum duty cycle of 1:60. It can be seen from Figure 6 that the backscatter shield reduces the exposure by almost 1 order of magnitude compared with the results without the shield. This result is consistent with the Pbequivalence of the plastic, which is 0.5mm thickness of Pb equivalent. Since the half-value thickness for 60keV x-rays in Pb is about 0.2 mm, the measured result is consistent with about 2.5 half layers, or 0.5mm thickness. It should also be noted that the level of backscattered radiation at the back of the instrument, with no shield, is on the same order as the leakage radiation at the case. It is also significantly below the 100 mr/hr "safe" limit. With the shield in place, the backscattered radiation is sufficiently attenuated that the leakage radiation becomes the primary contributor to the operator dose. c. Total Dose calculations Given the Leakage Radiation and Backscatter Radiation measurements at the instrument backplane, it is possible to estimate the maximum dose that would be received by an operator using the NOMAD. The typical whole body exposure to the operator can be estimated using the measured sum of backscatter and leakage radiation level at the control panel of approximately 1 mr/hr. This is at the maximum duty cycle of 60 images per hour. NCRP reports that a high-volume dental practice would take 300 images per week. This would then produce 5mR of operator exposure per week, if each image required a full 1 second dose of radiation. However, typical Kodak D speed film requires an irradiation time of only 0.5 seconds, reducing the weekly exposure to 2.5mR. If digital sensors are used, the typical exposure time is only 0.1 second, reducing the weekly exposure to 0.5mR. This suggests annual exposures of 130mR in an operatory using film, and 26 mr in an office using digital sensors, if only one operator takes all the images. In a typical practice this workload would be shared between 2 or more operators, so the results reported in Table 4 assume 2 operators in the practice. Exposure to the hand is even more benign relative to the recommended exposure limits. The highest leakage data was 3.6mR/hr. Table 4 shows the cumulative hand exposure for the same 300 films per week, at reduced exposure times for film and digital sensors and 2 operators sharing the load. NCRP gives a recommended annual occupational dose limit at 50mSv, with an exposure limit to the skin, hands, and feet of 500mSv. These levels are easily achieved with the NOMAD. Table 4. Total Annual Operator Exposures Hand/Extremity Exposure mr msv Recommended Occupational limit 50, Typical NOMAD used with film Typical NOMAD used with digital Whole Body Exposure Recommended Occupational limit 5, Typical NOMAD used with film Typical NOMAD used with digital Average annual background in USA
6 Height above floor (cm) Air kerma in one hour in mr at back plane of instrument (workload 138 mas) With Shield Without Shield ZONE OF SIGNIFICANT OCCUPANCY Figure 6. Example of the "zone of significant occupancy" for the NOMAD system. VI. Protection of the Public NOMAD is intended to be used in the same settings, and with the same protocols, as stationary dental x-ray equipment. Since the instrument is ambulatory, training must be provided about good radiation practice in areas outside the dental office. In general, if standard radiology protocols are utilized, there is no reason that this portable instrument should be less safe than a stationary unit. Figure 7. Postmortem image collected with a digital sensor
7 VII. Image Quality An important safety consideration is the ability to take a quality image with a handheld system. One of the reasons that existing standards require the x-ray tube assembly to be mounted in a stationary position is so that stability of the instrument can be assured. Any blurred or sloppy pictures require another irradiation of the patient. These should be avoided whenever possible. Figure 10. Root canal monitoring using D speed film. VIII. Summary Figure 8. Skull phantom imaged with a digital sensor. We have taken a significant number of images using both conventional dental film and several different brands of digital sensors. In all cases - with reasonable care to steady the instrument with both hands - the operator has been successful in creating dental images of comparable quality to stationary systems. Examples of digital and film-captured images are shown in Figures (These samples were primarily on full-head phantoms, although NOMAD units have been in use in dental clinics in Thailand since December, 2004.) Figure 9. Implant check using D speed film. We have shown that use of NOMAD in its intended applications presents no more risk to the operator or the patient than using stationary x-ray equipment. With moderate care, the operator can prevent any risk to the public. We have also shown that the image quality rivals that of stationary equipment, even when the x-ray is handheld particularly in cases using digital sensors with correspondingly short exposure times. IX. Acknowledgements The authors wish to thank Ann M. Jones of Medical Physics Consultants, Jason McCeighen, and Khoi Nguyen for the leakage radiation measurements and backscattered radiation measurements on the NOMAD devices. X. References 1. Code of Federal Regulations, Title 21, Vol. 8, April "General Requirements for Radiation Protection in Diagostic X-ray Equipment," European Committee for Electrotechnical Standardization, Brussesls, "Radiation Protection in Dentistry," National Council on Radiation Protection and Measurement, Bethesda, MD, Report #45, "Suggested State Regulations for the Control of Radiation," Council of Radiation Control Program Directors, S. Piorek, Field Anal. Chem. and Tech. 1(6), (1997). 6. "DentalEZ Portable HDX Intraoral X-ray," User's Manual, Flow X-ray Inc., 03/ NCRP, page Compendium Report, May 1993, Vol. XIV, No. 5, "X- Radiation: Potential Risks and Dose Reduction Mechanisms." 9. NEXT (Nationwide Evaluation of X-Ray Trends) is a committee of the Conference of Radiation Control Program Directors (CRCPD) that oversees quality control procedures for diagnostic radiology - 7 -
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
Elements of required physical infrastructures: space, schielding, and patient flow..
Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto
EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information
Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
X-RAY REGULATORY GUIDE
Minnesota Department of Health Radiation Control, X-ray Unit Protecting, maintaining and improving the health of all Minnesotans by promoting radiation safety through guidance and collaboration with the
Radiation safety in dental radiography
Radiation safety in dental radiography Dental Radiography Series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at minimum
X-Mind. Instinct for perfection
X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
X-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
Production of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards
Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards Presented: August 3, 2011 Daniel Kassiday United States Food and Drug Administration Center for Devices and Radiological
Digital Dental X-ray. Midmark Animal Health Products
Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION
CODE OF PRACTICE & SAFETY GUIDE
CODE OF PRACTICE & SAFETY GUIDE R a d i a t i o n P r o t e c t i o n i n D e n t i s t r y RADIATION PROTECTION SERIES Radiation The Radiation is published by the Australian Radiation and Nuclear Safety
Agency for Health Care Administration
Page 1 of 20 FED - H0000 - INITIAL COMMENTS Title INITIAL COMMENTS CFR Type Memo Tag FED - H0005 - COMPLIANCE WITH FED/STATE/LOCAL LAWS Title COMPLIANCE WITH FED/STATE/LOCAL LAWS CFR 486.100 Type Condition
X-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581
X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
portable x-ray survey report
department of HealtH and Human services centers for medicare & medicaid services form approved omb no. 0938-0027 portable x-ray survey report provider number H1 date surveyed (1) initial (2) survey W H2
NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF DENTAL RADIOLOGIC TECHNOLOGY
NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF DENTAL RADIOLOGIC TECHNOLOGY This document contains the Board s accreditation standards for schools of
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1
TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric
SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
SOP #: Revision #: Current Version Implementation Date: Page #: Page 1 of 10 Last Reviewed/Update Date: Expiration
Implementation Page #: Page 1 of 10 Last Reviewed/Update 1. Purpose and Scope The purpose of this document is to describe the Medical Physics and Radiation Safety program at Boston University (BU) and
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act
GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES. Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic)
GENERAL SHIELDING REQUIREMENTS FOR DIAGNOSTIC X-RAY FACILITIES 1. PROTECTIVE SCREENS (a) Major Medical (Radio/Fluoro/CT/Cardiac/Chiropractic) Typically, 15 kg m -2 (1.3 mm) sheet lead or its shielding
PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS
Sec. B.1 - B.4 SSRCR Volume I - February 2009 PART B REGISTRATION [LICENSURE] OF RADIATION MACHINE FACILITIES, [SERVICES] */ AND ASSOCIATED HEALTHCARE PROFESSIONALS Sec. B.1 - Purpose and Scope. a. This
Dose Tracking System:
Dose Tracking System: A Paradigm Shift in Patient Dose Monitoring Andrew Kuhls-Gilcrist, Ph.D. XRVL Clinical Science Manager Toshiba America Medical Systems, Inc. 2 Dose Tracking System: A Paradigm Shift
Radiation Safety Issues for Radiologic Technologists
Radiation Safety Issues for Radiologic Technologists Greg Sackett, M.S., CHP Medical Physicist Radiation Worker Risks? 1 Patient Risks? Acute Effects? Delayed Effects? Patient Questions? Radiation Dose
LOUISIANA STATE UNIVERSITY SCHOOL OF DENTISTRY INSTITUTIONAL POLICY FOR THE DIAGNOSTIC USE OF IONIZING RADIATION
LOUISIANA STATE UNIVERSITY SCHOOL OF DENTISTRY INSTITUTIONAL POLICY FOR THE DIAGNOSTIC USE OF IONIZING RADIATION August 2015 SECTION I INTRODUCTION The purpose of the following guidelines is to standardize
Required RS Training Info
C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions
QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES (a) Applicability Quality assurance programs as described in paragraph (c) of this section are recommended for all diagnostic radiology facilities.
420-3-26-.06 RADIATION SAFETY REQUIREMENTS FOR USERS OF X-RAY IN HEALING ARTS OR SERVICERS OF X-RAY EQUIPMENT
420-3-26-.06 RADIATION SAFETY REQUIREMENTS FOR USERS OF X-RAY IN HEALING ARTS OR SERVICERS OF X-RAY EQUIPMENT (1) Scope. Rule 420-3-26-.03 establishes standards for use of x-rays in the healing arts including
X-ray Imaging Systems
Principles of Imaging Science I (RAD 119) X-ray Tube & Equipment X-ray Imaging Systems Medical X-ray Equipment Classified by purpose or energy/current levels kvp, ma Radiographic Non-dynamic procedures
Standard Test Method for Classification of Film Systems for Industrial Radiography 1
Designation: E 1815 96 (Reapproved 2001) Standard Test Method for Classification of Film Systems for Industrial Radiography 1 This standard is issued under the fixed designation E 1815; the number immediately
The University of Southern Maine. XRF X-Ray Radiation Protection Program
The University of Southern Maine XRF X-Ray Radiation Protection Program 1 University of Southern Maine Department of Environmental Sciences 105 Bailey Gorham Hand Held XRF Analyzer X-Ray Radiation Protection
Radiation Safety In Dental Practice
Radiation Safety In Dental Practice A study guide and excerpts from The California Radiaton Control Regulations pertaining to dental practice Radiation Safety Protection Program template December 2014
Radiation safety in dental radiography
Radiation safety in dental radiography Kodak s dental radiograph series The goal of dental radiography is to obtain diagnostic information while keeping the exposure to the patient and dental staff at
X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
RADIATION PROTECTION PROGRAM (RPP) GUIDANCE
RADIATION PROTECTION PROGRAM (RPP) GUIDANCE The Colorado Rules and Regulations Pertaining to Radiation Control, (6 CCR 1007-1) are the regulations developed by the Colorado Department of Public Health
In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the
In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the discussion on the major components of the x-ray machine. Today
I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit
I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete
STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT
2010 STATE OF NEBRASKA STATUTES RELATING TO MEDICAL RADIOGRAPHY PRACTICE ACT Department of Health and Human Services Division of Public Health Licensure Unit 301 Centennial Mall South, Third Floor PO Box
Radiation Safety In Dental Practice
Radiation Safety In Dental Practice A study guide and excerpts from The California Radiaton Control Regulations pertaining to dental practice Radiation Safety Protection Program template December 2013
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,
Patient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University
Digital panoramic X-ray for practical diagnostics
ORTHOPHOS XG 3 instruments treatment centers Imaging systems cad/cam systems ORTHOPHOS XG 3 Digital panoramic X-ray for practical diagnostics System concept Standard panoramic X-ray with proven technology.
Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Acknowledgement. Diagnostic X-Ray Shielding. Nomenclature for Radiation Design Criteria. Shielding Design Goal (Air Kerma):
Diagnostic X-Ray Shielding Multi-Slice CT Scanners Using NCRP 47 Methodology Melissa C. Martin, M.S., FAAPM, FACR Therapy Physics Inc., Bellflower, CA AAPM Annual Meeting, Orlando, FL Refresher Course
Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X
DIGITAL X-RAY SYSTEM with ceiling mounted tube support Universal Digital Radiographic System Conventional system, full DR components Best-in-class image quality Flexible connectivity to PACS systems General
Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW
Qiang Lu Chapter 3. System Scanning Hardware Overview 79 Chapter 3 SYSTEM SCANNING HARDWARE OVERVIEW Since all the image data need in this research were collected from the highly modified AS&E 101ZZ system,
X-ray (Radiography), Chest
X-ray (Radiography), Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray makes images of the heart, lungs, airways,
Introduction. Chapter 15 Radiation Protection. Regulatory bodies. Dose Equivalent. Regulatory bodies. Main Principles of Radiation Protection
Introduction Chapter 15 Radiation Protection Radiation Dosimetry I Text: H.E Johns and J.R. Cunningham, The physics of radiology, 4 th ed. F.M. Khan, The Physics of Radiation Therapy, 4th ed., Chapter
X-ray (Radiography) - Chest
Scan for mobile link. X-ray (Radiography) - Chest What is a Chest X-ray (Chest Radiography)? The chest x-ray is the most commonly performed diagnostic x-ray examination. A chest x-ray produces images of
Staff Doses & Practical Radiation Protection in DEXA
Patient Xray X Doses Staff Doses & Practical Radiation Protection in DEXA Una O ConnorO Dept. of Medical Physics & Bioengineering, St. James s s Hospital. Examination Types General XrayX Fluoroscopy /
C-10 Radiation Monitoring Network
C-10 Radiation Monitoring Network Table of Contents C-10 Foundation History C-10 Mission & Goals History of C-10 Radiological Monitoring Network C-10 Radiation Monitoring Network (CRMN) CRMN System Architecture
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION
CONTENT SPECIFICATIONS FOR THE FLUOROSCOPY EXAMINATION Publication Date: November 2010 Implementation Date: March 2011 The purpose of the American Registry of Radiologic Technologists Fluoroscopy Examination
Veraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life.
Veraview IC5 HD High definition, digital imaging excellence Thinking ahead. Focused on life. Thinking ahead. Focused on life. 2 3 IC5 HD Super High-Speed with High Definition Clarity For dental radiology
JB-70 Dental X-Ray System
Dental X-Ray System User Manual 00-02-1568 Rev. U 0120 ECN: P3250 Attention: The equipment must only be installed and operated in accordance with the safety procedures and operating instructions given
Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
JEFFERSON COLLEGE. Radiographic Physics
JEFFERSON COLLEGE COURSE SYLLABUS RAD160 Radiographic Physics 3 Credit Hours Revised by: Janet E. Akers BS RT (R)(M) Date: October 3, 2013 Kenny Wilson, Director, Health Occupation Programs Dena McCaffrey,
Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition)
Compliance Guidance for QUALITY ASSURANCE MANUAL (2 nd Edition) New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ 08625 FAX 609-984-5811 Website: http://www.state.nj.us/dep/rpp
Personal Dosimetry in Dental Radiology
Personal Dosimetry in Dental Radiology Position Statement Personal Dosimetry in Dental Radiology Radiological Protection Institute of Ireland May 2011 3 Clonskeagh Square, Dublin 14 Background The RPII
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
What Parents Should Know about the Safety of Dental Radiology.
What Parents Should Know about the Safety of Dental Radiology. There are many different types of x-ray images (pictures) that can be taken of children in the dental office to assist in diagnosis. These
7üVNORD. TÜV NORD Röntgentechnik
.. TÜV NORD Röntgentechnik Niedersachsen Nordrhein-Westfalen Hamburg Schleswig-Holstein Baden-Württemberg Mecklenburg-Vorpommern Bremen Sachsen-Anhalt Berlin Brandenburg Sachsen' Thüringen' Bayern' Hessen
APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS
APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL PROCEDURES USING X RAYS SAFETY REPORTS SERIES No. 39 APPLYING RADIATION SAFETY STANDARDS IN DIAGNOSTIC RADIOLOGY AND INTERVENTIONAL
Integrated Management & Educational Consultancy
Integrated Management & Educational Consultancy Services MANDEC Radiology Update Seminar for The Dental Team Dr. Richard DeCann & Mr. Tim Reynolds IMECS Radiography Consultants Tel:07855 183117 email:[email protected]
Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology
Easy Quality Control with - PTW Equipment Code of Practice - Quality control of X-ray equipment in diagnostic radiology Revised edition October 2010 Code of Practice- Quality control of X-ray equipment
NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF RADIATION THERAPY TECHNOLOGY
NEW JERSEY RADIOLOGIC TECHNOLOGY BOARD OF EXAMINERS (BOARD) ACCREDITATION STANDARDS FOR SCHOOLS OF RADIATION THERAPY TECHNOLOGY This document contains the Board s accreditation standards for schools of
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
The Field. Radiologic technologists take x-rays and administer nonradioactive materials into patients' bloodstreams for diagnostic purposes.
Radiologic Technologist Overview The Field - Specialty Areas - Preparation - Day in the Life - Earnings - Employment - Career Path Forecast - Professional Organizations The Field Radiologic technologists
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
X-ray Safety Manual 2013
X-ray Safety Manual 2013 X-ray Machines and Other Ionizing Radiation Producing Devices The guidelines in this manual, when followed will ensure that appropriate protective and regulatory actions required
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION
INFO-0545 RADIOISOTOPE SAFETY MONITORING FOR RADIOACTIVE CONTAMINATION 1. INTRODUCTION This document provides general guidance for monitoring and controlling radioactive contamination, and relating the
PERSONNEL MONITORING AND DOSIMETRY POLICIES
PERSONNEL MONITORING AND DOSIMETRY POLICIES All individuals who are required to have their exposure to ionizing radiation monitored must be trained prior to using the source(s) of radiation. The radioactive
Assessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY
GUIDE ST 5.3 / 4 MAY 2007 USE OF IONIZING RADIATION IN THE TEACHING OF PHYSICS AND CHEMISTRY 1 Ge n e r a l 3 2 Safety licences and exemption from licensing 3 3 Safety licence-exempt use of radiation 3
The Basics of Scanning Electron Microscopy
The Basics of Scanning Electron Microscopy The small scanning electron microscope is easy to use because almost every variable is pre-set: the acceleration voltage is always 15kV, it has only a single
Dental Radiography collimator Ionising radiation image radiolucent area radiopaque area controlled zone scatter radiation intraoral
Dental Radiography X-rays for dental radiography are produced by high voltages of electricity within an x-ray head and come out through a metal tube called a collimator. This ensures the x-rays only come
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
Quality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
X-ray Operating Procedures and Safety Manual
X-ray Operating Procedures and Safety Manual December 2014 Revision 6 TABLE OF CONTENTS PAGE # 1.0 Introduction 4 1.1 Purpose 4 1.2 Emergency Telephone Numbers 5 1.3 Responsibilities 5 1.3.1 Radiation
Radiologic Technology Program. Radiation Safety and Protection Program
Radiologic Technology Program Radiation Safety and Protection Program Name of Program: Charles R. Drew University of Medicine and Science College of Science and Health CA RHB Program Number: 1029 Name
I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS
I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS A. Slavtchev 1, I. Manolov 2 1 National Centre of Radiobiology and Radiation Protection 2 Medrom Ltd. Abstract In recent years the number