Evaluation of the image quality in computed tomography: different phantoms
|
|
|
- Clifford Simon
- 10 years ago
- Views:
Transcription
1 Artigo Original Revista Brasileira de Física Médica.2011;5(1): Evaluation of the image quality in computed tomography: different phantoms Avaliação da qualidade de imagem na tomografia computadorizada: diferentes fantomas Vinicius C. Silveira 1, Larissa C. Oliveira 2, Rômulo S. Delduck 1, Simone Kodlulovich 1, Fernando A. Mecca³ and Humberto O. Silva 4 1 Instituto de Radioproteção e Dosimetria, Comissão Nacional de Energia Nuclear (CNEN), Rio de Janeiro (RJ), Brazil. 2 Nuclear Instrumentation Laboratory / Instituto Alberto Luiz Coimbra de Pós-graduação e pesquisa de Engenharia (COPPE), Universidade Federal do Rio de Janeiro (UFRJ), Rio de Janeiro (RJ), Brazil. 3 National Institute of Cancer (INCa), Rio de Janeiro (RJ), Brazil. 4 Copa D Or Hospital, Rede Labs D Or, Rio de Janeiro (RJ), Brazil. Abstract The aim of this paper was to compare the simulators provided by the CT manufactures and Catphan s Phantom with the American College of Radiology (ACR) computed tomography phantom. The image evaluation followed the protocols established by the manufactures of the phantoms. For slice thickness evaluation, the maximum percentage difference was 9% between the phantoms ACR and Siemens. In CT number accuracy test, the measurements of CT number of water showed a difference of 10 HU between the CT simulators. Comparing the uniformity results, the discrepancy was 11% and 55% for Siemens and Philips respectively in relation to the result obtained with the ACR phantom. The result of low contrast was the same for all phantoms. The MTF50 and MTF10 obtained with Siemens phantom was 4 and 8 pl/mm. For Catphan, 6 and 7 pl/mm. Results demonstrate that the ACR simulator was the most comprehensive and flexible to be used in several scanner models. Some simulators did not present all image quality indicators to perform a complete test. Keywords: computed tomography, image quality, phantoms. Resumo O objetivo deste trabalho foi comparar os simuladores fornecidos pelos fabricantes de tomógrafos e o fantoma Catphan com o fantoma de tomografia computadorizada do Colégio Americano de Radiologia (ACR). A avaliação da imagem seguiu os protocolos estabelecidos pelos fabricantes dos fantomas. Para a avaliação da espessura de corte, a maior diferença foi de 9% entre os fantomas ACR e Siemens. No teste de exatidão do número de CT, as medidas do número de CT da água mostraram uma diferença de 10 HU entre os fantomas de CT. Comparando os resultados de uniformidade, a discrepância foi de 11% e 55 % para os fantomas da Siemens e da Philips em relação ao valor obtido com o fantoma do ACR. O resultado de baixo contraste foi o mesmo para todos os fantomas. Os valores de MTF50 e MTF10 para a resolução de alto contraste do Siemens foram 4,2 e 7,6 pl/mm e para o Catphan, 6 e 7 pl/mm. Os resultados demonstraram que o simulador do ACR foi o mais compreensivo e flexível a ser usado em diversos modelos de tomógrafos. Alguns simuladores não apresentaram dados suficientes para realizar o teste completo. Palavras-chave: tomografia computadorizada, qualidade da imagem, fantomas. Introduction The optimization program includes an evaluation of the image quality. Each manufacture has developed a specific simulator for their computed tomography (CT) scanner. These phantoms present differences in the physical indicators to evaluate image quality, values of tolerance, and especially the procedure to carry out the tests. The American College of Radiology (ACR) CT phantoms have been used in an accreditation program in the USA. By applying a standard methodology, it is possible to evaluate and to compare scanners from all manufactures or models. The minimum physical indicators recommended for the evaluation of image quality are: positioning of coach, CT number accuracy, slice width, low contrast resolution, high contrast (spatial) resolution, CT number uniformity, and noise. However, several simulators do not have these indicators to conduct a full assessment. Despite of the increase of multi-slice scanners in Latin America, hospitals do not have personnel trained, Corresponding author: Vinicius da Costa Silveira Institute of Radioprotection and Dosimetry Av. Salvador Allende, s/n Recreio dos Bandeirantes Rio de Janeiro (RJ), Brazil [email protected] Associação Brasileira de Física Médica 67
2 Silveira VC, Oliveira LC, Delduck RS, Kodlulovich S, Mecca FA, Silva HO instrumentation, and phantoms to implement the quality assurance program. Nowadays, the regulatory authority does not have any information about the performance of scanners of the services. Besides, countries of Latin America do not have a CT accreditation program. Consequently, there is no information about patient dose and image quality. The aim of this paper was to compare the simulators provided by the manufactures and the Catphan s Phantom with the ACR CT Phantom. The image evaluation followed the protocol established by the manufacture of each phantom. s and methods The phantoms evaluated were provided by General Electric (GE), Siemens, Philips, Catphan 500, and ACR CT Phantom. Image quality tests were performed in three scanners: Philips Brilliance 40, GE Light speed and Siemens Somaton, with their respective simulators and the CT ACR Phantom. Additional tests were carried out in the public hospital with Catphan 500 and ACR phantom on the scanner Picker. Figure 1. (a) Catphan 500; (b) ACR. Table 1. Quality assurance of each manufacturer Test QA CAT¹ ACR GE PHI² SI³ Collimation x x x x x Accuracy of # CT x x x* x x* Pixel Size x High Contrast x x x x x Low Contrast x x x x x Noise and Uniformity x x x x x Contrast scale x x x x x Alignment x x - - x Accuracy laser light - - x - - x* - only water and air QA: quality assurance; 1 CAT: Catphan; 2 PHI: Philips; 3 SI: Siemens. A Figure 2. (a) Phantom Philips / Brilliance 40; (b) Phantom GE/ Light speed). A B B Simulators characteristics The simulators have distinct characteristics, according to the specificity of the test performed. The Catphan (Figure 1a) is made in The Phantom Laboratory Incorporated, in New York. It is a solid Phantom containing four modules: CTP528, 21 line-pair high resolutions; CTP 515, sub-slice and supra-slice low contrast; CTP404, position verification, slice width, sensitometry and pixel size; CTP486, solid image uniformity module. The ACR 2 CT accreditation phantom (Figure 1b) is a solid phantom containing four modules, constructed primarily from solid water. There are external markings (BBs) on the first and last module to allow the alignment of the phantom in the axial, coronal, and sagittal directions. Using this phantom, it is possible to evaluate alignment, CT number accuracy, slice width, low and high contrast resolution, uniformity, and noise. The Philips Phantom 3 (Figure 2a) has two parts: head and body. The part of the head contain: physical layer, impulse response and slice width; water layer, noise and uniformity; multi-strip, contrast scale and sensitometry. The part of the body contain: a Teflon strip and water hole. GE s Phantom (Figure 2b) for scanners of light speed series can evaluate six quality image criteria: contrast scale, resolution of high and low contrast, noise, uniformity, slice width, accuracy of laser, and linearity of CT number 4. It is divided into three parts: resolution s block, contrast membrane, and water hole. The Siemens Phantom for Somaton scanners contains a number of modules suitable for testing different CT image quality characteristics, such as: slice thickness, impulse response, CT number accuracy (water and air), high and low contrast, noise and uniformity, and alignment. Table 1 presents the proposed tests by each manufacturer. For this study, the following were compared: collimation, accuracy and linearity of CT number, evaluation of high and low contrast, noise and uniformity, and contrast scale. Therefore, they demonstrate the adequacy to a proper image quality evaluation of each phantom. Experimental setup Simulators were placed and aligned using light beams of the scanners. All phantoms are cylindrical; the alignment in the gantry was performed considering sagittal and coronal projections. The position was determined by specifics marks of each simulator. The evaluation followed the respective manual of the manufacture. To compare the ACR and Catphan phantoms, the head routine protocol on axial acquisition was used. Results and discussion Siemens Somaton Phantom Table 2 shows the results of slice thickness to ACR and Siemens phantoms on Siemens/Somaton. 68 Revista Brasileira de Física Médica.2011;5(1):67-72.
3 Evaluation of the image quality in tomography computed: different phantoms In Table 2, the maximum difference between the results obtained with the phantoms was 9% for the slice width with 10 mm. In relation to nominal slice, differences were lower than 3% for Siemens and 10% for ACR. Philips In this study, it was possible to compare the phantoms results only for 5 mm slice thickness (Table 3), in which the result obtained was equal for both phantoms and nominal slice thickness. For 10 mm, the percentage difference between the nominal and ACR was of 46%. CT GE Hi-Speed In the Table 4, for nominal 3 mm slice thickness, the Phantoms GE and ACR presented the same values. For the other thicknesses (5, 7 and 10), we could only obtain the comparison with nominal values. For ACR measurements, the values for 3 and 7 mm were the same of the selected. For GE, the difference between the measured and the nominal values was 5%. CT Picker Table 5 presents the slice thickness measurements using Catphan and ACR phantoms. For all nominal slice thickness, the differences between the phantoms results were 5%. Comparing with the nominal value of 3 mm, the percentage difference was approximately 40%. Table 2. Comparison of slice thickness measured with the CT Somaton phantom and ACR Nominal slice measured (mm) thickness (mm) Phantom ACR Phantom Siemens D (%) Table 3. for CT Brilliance, Philips selected measure (mm) selected (mm) Phantom ACR Phantom Philips Table 4. for CT Hi-Speed, GE selected measure (mm) selected (mm) Phantom ACR Phantom GE Table 5. for CT Picker selected measure (mm) selected (mm) Phantom ACR Phantom Catphan 3 2,2 2, , CT number accuracy Siemens phantoms Table 6 presents the accuracy values of the CT number carried out with ACR and Siemens Phantoms. The Siemens Phantom only has water and air inserts. Important structures, like soft tissue and bone, are not available. For all cases, the values are in the tolerance range. CT Brillance, Philips Table 7 presents the results of CT number accuracy with ACR and Philips Phantoms. For polyethylene and acrylic, the results showed a difference of 14 and 2%, respectively. For water, although the values were in accordance to the tolerance, the value measured with ACR phantom was approximately 7 HU. Philips Phantom does not have a material similar to air. Therefore, with the exception of acrylic, the CT number accuracy was adequate. CT Hi-Speed, GE In Table 8, the results of the CT number accuracy for CT Hi-Speed from GE are presented. In this case, water is the only common material in the phantoms. The result to the discrepancy was 43% between the simulators. Table 6. Accuracy of CT number for CT Somaton, Siemens Phantom ACR Phantom Siemens Reference (HU) Polyethylene and 87 Water and +7 Acrylic and 130 Bone and 970 Air ,005 and 970 Table 7. Accuracy of CT number for CT Brillance, Philips Phantom ACR Phantom Philips Reference (HU) Polyethylene and 87 Water and +7 Acrylic and 130 Bone and 970 Air ,005 and 970 Table 8. Accuracy of CT number for CT Hi-Speed, GE Phantom ACR Phantom GE Reference (HU) Polyethylene and 87 water and +7 Acrylic and 130 Bone and 970 Air ,005 and 970 Polyethyrene - -1,1 - Revista Brasileira de Física Médica.2011;5(1):
4 Silveira VC, Oliveira LC, Delduck RS, Kodlulovich S, Mecca FA, Silva HO Picker The CT numbers obtained with ACR and Catphan Phantoms are presented in Table 9. For polyethylene and air, the discrepancy between results was of 4 and 1%, respectively. The reading of water had a very high discrepancy between phantoms. Low contrast resolution The Siemens and ACR Phantoms contain low contrast groups of objects inside a similar background with different sizes. For both phantoms, groups with 5 and 2 mm diameter were identified applying the manuals. For Philips Brilliance CT, we could not visualize the group of 5 mm as the manufacture in ACR and Philips. The GE Phantoms contain a polystyrene membrane suspended in water with holes with diameters of 10, 7.5, 5, 3 and 1 mm. It was observed holes of 3 mm. The differences between CT number of the membrane and water are equal to 10 (contrast level). In the Catphan 500, the contrast levels are measured marking region of interest (ROIs) over the largest target visualised in supra-slice, sub-slice, and in the background (Table 10). With the ACR s Phantom, inserts of 6 mm were visualized inserts by applying head s protocols. With the Catphan, the smallest diameter discernible was 5 mm for supra (0.3% contrast level) and 5 mm for sub-slice (1% contrast level). Uniformity and noise Table 10 presents the results of uniformity and noise according to ACR manual for all scanners. For Siemens CT scanner using its own phantom, it was possible to evaluate only uniformity. ACR showed a uniformity of -0.6 HU and Siemens, 7.3 HU. The results were satisfactory. The results to the Hi-Speed GE Scanner and Philips Brilliance by uniformity and noise using their own simulator were satisfactory. Nevertheless, when comparing to the ACR, the values for uniformity and noise were respectively 3.4 HU and 2.4 to GE and -1.4 HU and 4.8 to ACR. For Philips, the noise showed a difference of 16% in relation to the manufactures tolerance. Compared with the ACR, the # CT was 55% superior and the standard deviation was 20% lower. For the Picker, the value of uniformity obtained by Catphan was 8.7 HU and 10.4 with the ACR. The maximum values of noise were 9.1 with the Catphan and 7 with ACR phantom. High contrast resolution Table 11 shows the results to the test of high contrast carried out with the ACR Phantom on scanners: Philips, Siemens, and Picker. 1 Results to the CT GE using the own phantom of manufacturer presented a difference between the measured and reference value (18%) equal to 17%. To the Siemens and Catphan phantom, the MTF method was used to quantify the values of high resolution. The results obtained for Siemens and Picker scanners were according to the manufacturer s tolerances (Tables 12 and 13). Table 9. Accuracy of CT number for CT Picker Phantom ACR Phantom Catpham Reference (HU) Polyethylene -98,2-94,4-107 and 87 Water 10,1 0-7 and +7 Acrylic and 130 Bone and 970 Air -974, and 970 Table 10. Uniformity and noise Phantom ACR Parameter / Position ± standard deviation CT Siemens CT Philips CT GE CT Picker Center (C) -0.4± ± ± ±5.2 3h -1.0± ± ± ±6.3 6h -2.3± ± ±4.9 11,0±7,0 9h -1.0± ± ±4.3 10,8±4,7 12h 0.3± ± ± ±4.8 Table 11. High contrast resolution - Phantom ACR Technique MTF (u) Nominal Value (lp/cm) Spatial frequency (pl/mm) CT Siemens CT Philips CT Picker Reference Abdomen adult Chest Hi-Resolution Head Table 12. High contrast resolution CT Siemens Tolerance (lp/cm) Measured value (lp/cm) Conform 50% 4,50 0,45 4,2 y 10% 8,00 0,80 7,6 y 2% 10,00 1,00 9,6 y Table 13. High contrast resolution CT Picker MTF (u) (%) Value (lp/cm) Tolerance (lp/cm) 60 5 ±50 % 50 6 ±50 % 8 7 ±50 % 70 Revista Brasileira de Física Médica.2011;5(1):67-72.
5 Evaluation of the image quality in tomography computed: different phantoms Conclusions Evaluation of slice thickness showed similar results for all phantoms. For the accuracy of CT number, the water s CT number showed a very large discrepancy for all simulators. The GE and Siemens Phantoms do not have the structures to simulate soft and high materials, which are necessary to image quality evaluation. For low contrast resolution, all phantoms showed equivalent results. For uniformity and noise, GE and Siemens phantoms presented results with a very large discrepancy in relation to ACR. However, Philips and Cathan Phantoms showed equivalent results. Results demonstrate that the ACR simulator was the most comprehensive and flexible for use in several scanner models. It also had all the tests recommend by the International Image Quality Assurance 5. References 1. Goodenough DJ. CatPhan 504 Manual Laboratories Incorporated, Copyright American College of Radiology. Instruction Manual for testing the ACR CT Phantom. Preston White Drive: Reston VA, Philips Medical System _D Volume Manual manufacture GE Rev. 5. European Commission. Quality Criteria for Computed Tomography. Working Document EUR (Brussels: EC) (1997). Revista Brasileira de Física Médica.2011;5(1):
6
Lasers and Specturbility of Light Scattering Liquids - A Review
Artigo Original Revista Brasileira de Física Médica.2011;5(1):57-62. Reproducibility of radiant energy measurements inside light scattering liquids Reprodutibilidade de medidas de energia radiante dentro
American College of Radiology CT Accreditation Program. Testing Instructions
American College of Radiology CT Accreditation Program Testing Instructions (Revised July 24, 2015) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission
PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
What is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?
Congreso Internacional Conjunto Cancún 2004 LAS/ANS-SNM-SMSR/International Joint Meeting Cancun 2004 LAS/ANS-SNM-SMSR XV Congreso Anual de la SNM y XXII Reunión Anual de la SMSR/XV SNM Annual Meeting and
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1
TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
How To Improve Your Ct Image Quality
Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo
Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905
The phantom portion of the American College of Radiology ACR Computed Tomography CT accreditation program: Practical tips, artifact examples, and pitfalls to avoid a Cynthia H. McCollough b) and Michael
2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography
Today s agenda Computed Tomography Presented by: Dina Hernandez, BSRT, RT (R), CT, QM Krista Bush, RT, MBA Leonard Lucey, JD ACR Quality & Safety MIPPA overview and CMS requirements CT accreditation How
PERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 PERFORMANCE EVALUATION
Purchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage
WHERE IN THE WORLD JILL LIPOTI?
WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain
DEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 DEVELOPMENT AND IMPLEMENTATION
The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
Quality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM [email protected] 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom
Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Fernanda B. C. Nonato, Claudinei T. Cescon and Linda V. E. Caldas, Instituto de Pesquisas Energéticas e Nucleares,
Moving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
Quality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level Eduardo de Lima Corrêa 1, Vitor Vivolo 1, Maria da Penha A. Potiens 1 1 Instituto de Pesquisas Energéticas e Nucleares
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Assessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
SUITABILITY OF RELATIVE HUMIDITY AS AN ESTIMATOR OF LEAF WETNESS DURATION
SUITABILITY OF RELATIVE HUMIDITY AS AN ESTIMATOR OF LEAF WETNESS DURATION PAULO C. SENTELHAS 1, ANNA DALLA MARTA 2, SIMONE ORLANDINI 2, EDUARDO A. SANTOS 3, TERRY J. GILLESPIE 3, MARK L. GLEASON 4 1 Departamento
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
A COMPUTATIONAL PHANTOM OF HEAD AND NECK - SISCODES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 A COMPUTATIONAL PHANTOM
Rem: Revista Escola de Minas ISSN: 0370-4467 [email protected] Escola de Minas Brasil
Rem: Revista Escola de Minas ISSN: 0370-4467 [email protected] Escola de Minas Brasil Queiroz, Gilson; Carvalho, Hermes; Rodrigues, Francisco; Pfeil, Michèle Estimation of friction contribution in the
TLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 TLD ALBEDO DOSIMETER
Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.
www.neurologica.com Critical Care Clinic Operating Room Radiology ED Trauma OR Neuro/Spine Pediatric Plastics Thoracic General Interventional Radiology ENT ED Trauma Trauma Stroke Critical Care Neuro Pediatric
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
SITE IMAGING MANUAL ACRIN 6698
SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012
3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
DETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
HARVESTING AND WOOD TRANSPORT PLANNING WITH SNAP III PROGRAM (Scheduling and Network Analysis Program) IN A PINE PLANTATION IN SOUTHEAST BRAZIL 1
Harvesting and Wood Transport Planning with SNAP III... 831 HARVESTING AND WOOD TRANSPORT PLANNING WITH SNAP III PROGRAM (Scheduling and Network Analysis Program) IN A PINE PLANTATION IN SOUTHEAST BRAZIL
Data. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE
STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE Antônio C. O. Barroso*, José R. Maiorino, João M. L. Moreira **, José L. Bastos, José E. R. da Silva, Fernando R. de A. Lima, Mitsuo Yamaguchi,
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
Multi-slice Helical CT Scanning of the Chest
Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.
ENERGETIC EFFICIENCY OF AN AGRICULTURAL TRACTOR IN FUNCTION OF TIRE INFLATION PRESSURE
ENERGETIC EFFICIENCY OF AN AGRICULTURAL TRACTOR IN FUNCTION OF TIRE INFLATION PRESSURE LEONARDO DE A. MONTEIRO 1, DANIEL ALBIERO 2, KLEBER P. LANÇAS 3, ANDRÉ V. BUENO 4, FABRICIO C. MASIERO 5 ABSTRACT:
HYBRID INTELLIGENT SUITE FOR DECISION SUPPORT IN SUGARCANE HARVEST
HYBRID INTELLIGENT SUITE FOR DECISION SUPPORT IN SUGARCANE HARVEST FLÁVIO ROSENDO DA SILVA OLIVEIRA 1 DIOGO FERREIRA PACHECO 2 FERNANDO BUARQUE DE LIMA NETO 3 ABSTRACT: This paper presents a hybrid approach
IGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
CT QC Under the ACR QC Manual. What? There s a Manual?? Learning Objectives 8/6/12! ! YES!!! Almost. ! Doesn t have a pretty cover yet
CT QC Under the ACR QC Manual Douglas Pfeiffer, MS, DABR Boulder Community Hospital What? There s a Manual??! YES!!! Almost! Doesn t have a pretty cover yet! Should be out by RSNA! Really seriously I mean
IMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 IMPROVEMENT OF GAMMA
Quality Assurance of Radiotherapy Equipment
Clinical Implementation of Technology & Quality Assurance in Radiation Oncology For the course: Survey of Clinical Radiation Oncology 1 Quality Assurance of Radiotherapy Equipment Outline: A. Criteria
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
Improvement and storage of conventional mammography images using techniques of digital image processing
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-13 Improvement and storage of conventional mammography images using techniques of digital image processing M. M. A. Silva a ; J. W. Vieira b,c
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment
Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Atomic Energy Regulatory Board Mumbai 400 094 1 DEFINITIONS 1. Quality Assurance (QA): Planned and systematic actions
CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
VIRTUAL INSTRUMENTATION ON MOBILE DEVICES FOR DEPLOYMENT IN NUCLEAR INSTALLATIONS
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 VIRTUAL INSTRUMENTATION ON MOBILE
Accreditation Is Coming
Accreditation Is Coming Accreditation Is Coming Accreditation Is Coming Yes, It Is! Currently, Accreditation is Voluntary However, Some of the Accrediting Organizations Are Lobbying the Congress to Adopt
Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
TEACHING CALCULUS USING E-LEARNING IN A MOODLE PLATFORM
TEACHING CALCULUS USING E-LEARNING IN A MOODLE PLATFORM Caridade C.M.R., Faulhaber M.C., Rosa P.M., Silva P.M., Baeta N.S. Coimbra Institute of Engineering [email protected]; [email protected]; [email protected];
Phantom Test Guidance
Phantom Test Guidance for the ACR MRI Accreditation Program The American College of Radiology 1891 Preston White Dr Reston, VA 20191-4397 www.acr.org Contents 0.0 INTRODUCTION............................................................................
IAEA HUMAN HEALTH SERIES
IAEA HUMAN HEALTH SERIES No. 19 Quality Assurance Programme for Computed Tomography: Diagnostic and Therapy Applications IAEA HUMAN HEALTH SERIES PUBLICATIONS The mandate of the IAEA human health programme
How To Test Water Penetration On Concrete Block And Mortar
WATER PENETRATION TEST ON CONCRETE BLOCK MASONRY Vilató, Rolando Ramirez 1 1 PhD, Prof., Mackenzie Presbyterian University, School of Engineering, São Paulo, [email protected] This paper considers
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
The Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO
The Whys, Hows and Whats of the Noise Power Spectrum Helge Pettersen, Haukeland University Hospital, NO Introduction to the Noise Power Spectrum Before diving into NPS curves, we need Fourier transforms
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
ESCOAMENTO DO AR EM ARMAZÉM GRANELEIRO DE GRANDE PORTE COM SISTEMA DE AERAÇÃO SIMULATION OF AIR FLOW IN LARGE AERATED GRAIN STORAGE
ESCOAMENTO DO AR EM ARMAZÉM GRANELEIRO DE GRANDE PORTE COM SISTEMA DE AERAÇÃO Oleg Khatchatourian 1 Antonio Rodrigo Delepiane de Vit 2 João Carlos Cavalcanti da Silveira 3 Luciano Ferreira 4 RESUMO Foi
Quantitative Imaging In Clinical Trials Using PET/CT: Update
Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative
The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.
DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION
DOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DOSIMETRIC CHARACTERIZATION
How To Test The Quality Of A Radiography Display
Quality control ORIGINAL of diagnostic ARTICLEdisplays Display quality control based on luminance measurements in post-processing units at imaging diagnosis centers * Controle de qualidade de monitores
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,
Radiation Therapy. 1. Introduction. 2. Documentation of Compliance. 3. Didactic Competency Requirements. 4. Clinical Competency Requirements
PRIMARY CERTIFICATION DIDACTIC AND CLINICAL COMPETENCY REQUIREMENTS EFFECTIVE JANUARY 2014 Radiation Therapy 1. Introduction Candidates for certification and registration are required to meet the Professional
Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences [email protected] X-Ray
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School
Chest CT protocols Mannudeep K. Kalra, MD, DNB Dianna D. Cody, PhD Massachusetts General Hospital Harvard Medical School M.D. Anderson Cancer Center Specific principles Routine chest CT Lung nodule follow
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 Associação Brasileira de Energia Nuclear - ABEN ISBN: 978-85-99141-02-1 VoxelDose: A SOFTWARE
Practical exercise: Effective dose estimate in CT
Practical exercise: Effective dose estimate in CT TRAINING COURCE PROGRAM 19 20 May 2011, Sofia, Bulgaria Virginia Tsapaki Medical Physics Dpt Konstantopoulio General Hospital email: [email protected]
Concepts for High-Resolution Low-Dose CT of the Breast
RSNA 2012 Refresher Course 721B, Chicago, Nov. 30, 2012 Concepts for High-Resolution Low-Dose CT of the Breast Disclosures WAK is founder, shareholder and CEO of CT Imaging GmbH, Erlangen, Germany. Willi
Proton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
CURRICULUM VITAE. Kalpana M. Kanal, Ph.D., DABR
May 31, 2005 ADDRESS: PERSONAL: CURRICULUM VITAE Kalpana M. Kanal, Ph.D., DABR Department of Radiology University of Washington Medical Center 1959 NE Pacific Avenue Seattle, WA 98195-7115 (206) 616-1966
IQWorks progress to date and how you can get involved. Ed McDonagh, Andrew Reilly
IQWorks progress to date and how you can get involved Ed McDonagh, Andrew Reilly Progress since Nottingham Open source DICOM module Consolidation Evolution of algorithms Particularly edge MTF and NPS Display
QUESTIONÁRIOS DE AVALIAÇÃO: QUE INFORMAÇÕES ELES REALMENTE NOS FORNECEM?
QUESTIONÁRIOS DE AVALIAÇÃO: QUE INFORMAÇÕES ELES REALMENTE NOS FORNECEM? Grizendi, J. C. M [email protected] Universidade Estácio de Sá Av. Presidente João Goulart, 600 - Cruzeiro do Sul Juiz de Fora