VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES
|
|
- Brian King
- 8 years ago
- Views:
Transcription
1 2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 Associação Brasileira de Energia Nuclear - ABEN ISBN: VoxelDose: A SOFTWARE FOR DOSIMETRIC EVALUATIONS IN X-RAY DIAGNOSES Viriato Leal Neto 1,2, José W. Vieira 1,3, Borko Stosic 4 and Fernando R. de A. Lima 5 1 Centro Federal de Educação Tecnológica de Pernambuco Avenida Professor Luis Freire Cidade Universitária, Recife, PE viriatoleal@yahoo.com.br 2 Departamento de Energia Nuclear UFPE Avenida Professor Luis Freire Cidade Universitária, Recife, PE 3 Escola Politécnica de Pernambuco UPE Rua Benfica Madalena, Recife, PE jwvieira@br.inter.net 4 Departamento de Estatística e Informática UFRPE Dois Irmãos, s/n CEP , Recife, PE borko@ufrpe.br 5 Centro Regional de Ciências Nucleares CRCN Avenida Professor Luis Freire Cidade Universitária, Recife, PE falima@cnen.gov.br ABSTRACT In order to evaluate the dose absorbed by the patient in a series of examinations, it is necessary to accomplish simulations using an exposure computational model. Such models are fundamentally formed from a virtual representation of the human body (phantom) and a Monte Carlo code. There are several Monte Carlo codes available and also specific voxel phantoms for the two genders have been developed. However, the coupling of a voxel phantom to a Monte Carlo code is a complex process and it almost always results in the solution of a particular problem. This means that the use of these computational tools is not practicable on the routine of clinics and hospitals that carry out X-Ray examinations, because the simulations with computational models of exposition demand time, knowledge about the used code and several adjustments have to be implemented from one simulation to another. In such context, it was developed a GUI (Graphics User Interface) called VoxelDose, which contains a database with the simulation result of several X-Ray exams in C++. The database was built using the voxel phantoms MAX (Male Adult voxel) and FAX (Female Adult voxel) [1, 2], and the Monte Carlo code EGS4 (Electron Gamma Shower, version 4) [3]. The software has special menus to create databases, insert new examinations, visualize the phantom-field and the phantom-source, and obtain the summary of the doses in the organ and tissues on the screen of the monitor. If the user wants, he will be able to save or print the dosimetric results.
2 1. INTRODUCTION The present work developed a GUI (Graphics User Interface), called VoxelDose to facilitate the use of computational models of exposition based in phantoms of voxel and the Monte Code Carlo EGS4 (Electron Gamma Shower, version 4), for dosimetric evaluations with X- Ray diagnoses. VoxelDose software was developed using the Visual Microsoft C++.NET compiler and the resources of the tools package called.net Framework [4]. In the main window of the software, the user must choose the phantom, the type of examination, the type of projection, the spectre, the focus-film distance and the field to be used. This choice triggers two buttons: one to visualize the phantom-field and the phantom-source, making possible the saving or printing of images, and the other to calculate the dosimetric results, which can be saved or printed. 2. MATERIALS AND METHODS 2.1. Definition of the examinations and its parameters The Present work of thorax and abdomen examinations - AP (Anterior-Posterior), PA (Posterior-Anterior), LD/LE (Profile. LD = Lateral Direita; LE = Lateral Esquerda), was done to present the use and functionality of the VoxelDose. The parameters of the examinations were defined from visits made to Clinics and Hospitals in Recife, and Consultation to the Guidelines of direction and Recommendations for the definition of the user of the selection of parameters of the project DIMOND III image of the European Commission [5]. This brings tables with suggestions of parameters for thorax and abdomen examinations. Among these parameters the most used peak tensions in the examinations: 70 kv, 80 kv, 90 kv, 110 kv and 120 kv. To get the spectre of more probable energies for these tensions, an text file called mspectra.dat was appealed to it, that possesses 61 spectre of energies for diverse types of radioactive sources used in medicine and radioprotection. As to the fields, there is practically a standard of use and they approach to the dimensions of the film used in each case The necessary procedures to the simulation of VOXELS PHANTOM/EGS4 model To build the database it was necessary to execute the VOXEL PHANTOMS VOXELS/EGS4 computational model [6]. For this, it was necessary to develop and to connect geometry to be radiated to the EGS4, build a text file with the shock sessions for the radiation and the used ways, prepare operational data entrances, install the EGS4 and place the user s files, compile and execute the model Development and coupling of geometry
3 The geometry to be radiated must be build in text files according to the decoding programmed in the EGS4. In this work, the task of creation of the files was carried through by software FANTOMAS [7]. Phantoms without the arms (maxsb and faxsb) had been developed from the original versions of MAX and FAX [1, 2] models and used in the simulations of the examinations of thorax and abdomen profile in this work. Once the geometries defined the user s files of the phantom.code.mor EGS4 were adapted to the thorax simulations and abdomen Building of file with the shock sessions The shock sessions for photons and electrons were obtained in the own EGS4, where data were supplied such as the energy interval and the elementary compositions of constituent means of phantoms The operational Entrance Data The FANTOMAS software [7] was also used to create the texts files with operational entrance data (one for each model), called Expo.input, for the Monte Carlo simulation to be carried through: The examination identification number (ID), Resolution of phantom voxels, the ID of the used spectre, the focus-film distance, the field dimensions (height and width), the point source ID - AP, PA, LD and LE -, the number of histories (10 8 ), the cut-off energies for electrons (200 kev) and photons (2 kev), and name of the exit file Compilation and execution of the model With the installed EGS4 and the user s files duly placed, the compiler FORTRAN PowerStation 4.0 of the package Microsoft Developer Studio was used, to tie to the mass of general data and specific data produced by the user in MORTRAN3 [8]. Thus, getting phantoms.exe. The simulation of each planned examination took 2.5 on average h, in a Pentium 4 of 3.0 MHz with 512 MB of memory. The exit of each simulation presents a form with diverse information on the dosimetric Monte Carlo simulation and results for all the radiosensitive organs and tissues. The gotten forms turn to be text files that the user can make use through the Interface. 3. RESULTS AND CONCLUSIONS The simulations results were organized in two text files named MaxBD.txt and FaxBD.txt. These files will used be to demonstrate the GUI of VoxelDose performance To create a data base and to insert an examination The Figure 1 presents the main window of VoxelDose software. To create a database, click in the menu File menu item Criar Novo Banco de Dados it is written the name of the new database to be saved, for example, MaxBD.txt Salvar.
4 Figure 1: Main window of VoxelDose software presenting the Sub-menus of the menu File. In case of insertion of a certain examination, it is clicked the menu Arquivo menu item Inserir Novo Exame to type a number in the text box Posição do Exame no BD... and the button Abrir arquivos do BD e do Exame Abra o arquivo txt do BD... Abrir Arquivo do BD e do Exame TXT... Abra o arquivo txt do Exame... Abrir um Exame... Salva o Novo BD como To visualize and to get the results of the examination In the main window of VoxelDose, it is clicked, in the sequence, the radio buttons Max Tórax AP Selecionar o Espectro, a Distância Foco-Filme e o Campo and it is chosen one of the available examinations, for example, Max_Torax_AP_Exame_7: Espectro: 100 KVP DIAGNOSTIC,2 MM AL,HWS: 3 MM AL, DFF = 100 cm, Largura: 35, Altura: 43 cm, Figure 2. Figure 2: Main window of the VoxelDose presenting the steps during the selection of an examination.
5 With the definition of the intended examination, the Visualizar a Região do Exame and Calcular buttons are qualified. By setting the button Visualizar a Região do Exame two secondary windows appear: Field and Field-Source, showing, respectively, the field object of the examination and the position of the source related to phantom, Figure 3. Figure 3: Images visualized after the trigging of the button To visualize the Region of the Examination. Finally, when clicking the button Calcular, a secondary window appears, Formulário, with the summary of the doses in the radiosensitive organs and tissues related to the selected examination, Figure 4. Figure 4: Window Form with the summary of the doses.
6 The current version of the GUI VoxelDose allows the obtainment of dosimetric results for a predefined examination in a few seconds. It can be very useful in clinics and hospitals for the monitoring of the dosimetric distribution in patients who are submitted to the successive diagnoses involving the X-Rays. The current results are printed in the general format of conversion coefficients between, for example, dose/kerma in the air. In new versions of this GUI, it will also be possible to obtain dose/dose in the private dosimeter, most common coefficient of conversion in radiodiagnoses. It is also intended to extend the number of examinations and make the database available on the Internet. ACKNOWLEDGMENTS The authors are grateful to CNPQ and CEFET-PE for the financial support. REFERENCES 1. R. Kramer, J.W. Vieira, H. J. Khoury, F. R. A. Lima, D. Fuelle. All about MAX: a male adult voxel phantom for Monte Carlo calculations in radiation protection dosimetry, Phys. Med. Biol., Vol. 48, p , R. Kramer, J. W. Vieira, H. J. Khoury, F. R. A. Lima, E. C. M. Loureiro, V. J. M. Lima, G. Hoff. All about FAX: a Female Adult voxel Phantom for Monte Carlo Calculation in Radiation Protection Dosimetry, Phys. Med. Biol., 49, , EGS4 Source Code Availability and Distribution (2006). 4. J. Templeman, A. Olsen, Microsoft visual c++.net, step by step. Version 2003, Microsoft Press, USA, Projeto DIMOND III da European Commission (2004), (2007) 6. J. W. Vieira, Construção de um modelo computacional de exposição para cálculos dosimétricos utilizando o código Monte Carlo EGS4 e fantomas de voxels, Tese de Doutorado, DEN-UFPE, Recife-PE, Brasil, J. W. Vieira, B. Stosic, F. R. A. Lima, R. Kramer, A. M. Santos, V. J. M. Lima. Um software para editar fantomas de voxels e calcular coeficientes de conversão para a proteção radiológica, 1º Congresso Brasileiro de Proteção Radiológica, Rio de Janeiro, 02 a 05 de Novembro de W. R. Nelson, H. Hirayama, D. W. O Rogers. The EGS4 code system, Report SLAC- 265, Stanford Linear Accelerator Center, Stanford University, Stanford, California, CEMBER, H. Introduction to health physics, 3 rd edition, McGraw-Hill, M. H. Kalos, P. A. Whitlock. Monte Carlo methods, Vol. 1, Basics: Wiley & Sons, New York, USA, Comissão Nacional de Energia Nuclear, Diretrizes básicas de proteção radiológica NN (2006). 12. P.J. Collins, D. Gorbatkov, F. W. Shultz. A graphical user interface diagnostic radiology dosimetry using Monte Carlo (MCNP) simulation, 10 th International Congress of the International Radiation Protection Association, IRPA, Hiroshima, Japan, May 2000.
REMOTE SYSTEM FOR MONITORING AND CONTROL OF CONTROLLED AREA OF NUCLEAR INSTALLATION
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 REMOTE SYSTEM FOR MONITORING
More informationDevelopment of a pregnant woman phantom using polygonal mesh, for dosimetric evaluations
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-11 Development of a pregnant woman phantom using polygonal mesh, for dosimetric evaluations M.O.M. Cabral a ; J.W. Vieira b,c ; V. Leal Neto
More informationTA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1
TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT
More informationDOSIMETRIC CHARACTERIZATION OF DYED PMMA SOLID DOSIMETERS FOR GAMMA RADIATION
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DOSIMETRIC CHARACTERIZATION
More informationA COMPUTATIONAL PHANTOM OF HEAD AND NECK - SISCODES
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 A COMPUTATIONAL PHANTOM
More informationDATABASE IMPLEMENTATION TO A FLUIDIZED CRACKING CATALYTIC-FCC PROCESS
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 DATABASE IMPLEMENTATION
More informationImprovement and storage of conventional mammography images using techniques of digital image processing
BJRS BRAZILIAN JOURNAL OF RADIATION SCIENCES 03-1A (2015) 01-13 Improvement and storage of conventional mammography images using techniques of digital image processing M. M. A. Silva a ; J. W. Vieira b,c
More informationDEVELOPMENT AND IMPLEMENTATION OF AN AUTOMATED SYSTEM TO EXCHANGE ATTENUATORS OF THE OB85/1 GAMMA IRRADIATOR
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 DEVELOPMENT AND IMPLEMENTATION
More informationPERFORMANCE EVALUATION OF THE REFERENCE SYSTEM FOR CALIBRATION OF IPEN ACTIVIMETERS
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 PERFORMANCE EVALUATION
More informationIMPROVEMENT OF GAMMA CALIBRATION PROCEDURES WITH COMMERCIAL MANAGEMENT SOFTWARE
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 IMPROVEMENT OF GAMMA
More informationIrradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
More informationDetermination of the Effective Energy in X-rays Standard Beams, Mammography Level
Determination of the Effective Energy in X-rays Standard Beams, Mammography Level Eduardo de Lima Corrêa 1, Vitor Vivolo 1, Maria da Penha A. Potiens 1 1 Instituto de Pesquisas Energéticas e Nucleares
More informationESTIMATION OF EFFECTIVE DOSE FROM PEDIATRIC RADIOGRAPHY OF THE CHEST IN MEDELLÍN-COLOMBIA
Revista Facultad de Ciencias Universidad Nacional de Colombia, Sede Medellín V 4 N 1 enero-junio 2015 ISSN 0121-747X / ISSN-e 2357-5749 Research Paper Páginas 11 a 16 ESTIMATION OF EFFECTIVE DOSE FROM
More informationNATURAL RADIOACTIVITY FROM THE SOIL OF SÃO BERNARDO DO CAMPO
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 NATURAL RADIOACTIVITY
More informationTLD ALBEDO DOSIMETER PERFORMANCE ON A Hp(10) NEUTRON DOSE IAEA INTERCOMPARISON
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 TLD ALBEDO DOSIMETER
More informationEffects of Cosmic Radiation in Aircrafts: A Discussion about Aircrew over South America
doi: 10.5028/jatm.2012.04021012 Effects of Cosmic Radiation in Aircrafts: A Discussion about Aircrew over South America Claudio Antonio Federico 1,*, Odair Lélis Gonçalez 1, Gian Maria Agostino Angelo
More informationPortuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom
Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom Fernanda B. C. Nonato, Claudinei T. Cescon and Linda V. E. Caldas, Instituto de Pesquisas Energéticas e Nucleares,
More informationCFD SIMULATION OF IPR-R1 TRIGA SUBCHANNELS FLUID FLOW
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 CFD SIMULATION OF IPR-R1 TRIGA
More informationDiagnostic Exposure Tracking in the Medical Record
Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand
More information90 degrees Bremsstrahlung Source Term Produced in Thick Targets by 50 MeV to 10 GeV Electrons
SLAC-PUB-7722 January 9 degrees Bremsstrahlung Source Term Produced in Thick Targets by 5 MeV to GeV Electrons X. S. Mao et al. Presented at the Ninth International Conference on Radiation Shielding, Tsukuba,
More informationManual for simulation of EB processing. Software ModeRTL
1 Manual for simulation of EB processing Software ModeRTL How to get results. Software ModeRTL. Software ModeRTL consists of five thematic modules and service blocks. (See Fig.1). Analytic module is intended
More informationA MULTIELEMENT ANALYSIS OF NATURAL RADIONUCLIDES IN AN AREA OF PERNAMBUCO - BRAZIL
2007 International Nuclear Atlantic Conference - INAC 2007 Santos, SP, Brazil, September 30 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-02-1 A MULTIELEMENT ANALYSIS
More informationEFFECTS OF GAMMA IRRADIATION ON ELECTRIC CHARACTERISTICS OF POWER DIODES
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 EFFECTS OF GAMMA
More informationSTUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE
STUDIES FOR A MULTIPURPOSE RESEARCH REACTOR FOR THE CRCN/CNEN-PE Antônio C. O. Barroso*, José R. Maiorino, João M. L. Moreira **, José L. Bastos, José E. R. da Silva, Fernando R. de A. Lima, Mitsuo Yamaguchi,
More informationMastem: A Mathematics Tutoring Multi-Agent System
Mastem: A Mathematics Tutoring Multi-Agent System Jéssyka Vilela 1, Ricardo Ramos 2, Jaelson Castro 1 1 Universidade Federal de Pernambuco Centro de Informática Av. Jornalista Anibal Fernandes, S/N, Cidade
More informationMonte Carlo Simulations in Proton Dosimetry with Geant4
Monte Carlo Simulations in Proton Dosimetry with Geant4 Zdenek Moravek, Ludwig Bogner Klinik und Poliklinik für Strahlentherapie Universität Regensburg Objectives of the Study what particles and how much
More informationDesign and Implementation of an Institution-Wide Patient-Specific Radiation Dose
Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Monitoring Program for Computed Tomography, Digital Radiography, and Nuclear Medicine by Olav Christianson Medical Physics
More informationDEVELOPMENT OF A QUALITY MANAGEMENT SYSTEM FOR BRAZILIAN NUCLEAR INSTALLATIONS
2005 International Nuclear Atlantic Conference - INAC 2005 Santos, SP, Brazil, August 28 to September 2, 2005 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 85-99141-01-5 DEVELOPMENT OF A QUALITY
More informationDETERMINATION OF THE EFFECTIVE VOLUME OF AN EXTRAPOLATION CHAMBER FOR X-RAY DOSIMETRY
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
More informationRa, 232 Th and 40 K analysis in sand samples from some beaches of Great Vitória, Espírito Santo, Brazil: Preliminary results.
2009 International Nuclear Atlantic Conference - INAC 2009 Rio de Janeiro,RJ, Brazil, September27 to October 2, 2009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 226 Ra, and
More informationCalculation of Source-detector Solid Angle, Using Monte Carlo Method, for Radioactive Sources with Various Geometries and Cylindrical Detector
International Journal of Pure and Applied Physics ISSN 0973-1776 Volume 3, Number 2 (2007), pp. 201 208 Research India Publications http://www.ripublication.com/ijpap.htm Calculation of Source-detector
More informationQuality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme
Medical Exposure Directive 97/43 Euratom Quality Assurance Ministry of Health, Radiation Protection Department, Luxembourg Alexandra Schreiner Medical Physicist Quality Assurance (QA): All those planned
More informationDOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES
DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES Ausra Urboniene, Birute Griciene Lithuania Introduction Medical staff during interventional radiology procedures
More informationTHE AWARENESS OF THE FUNCTIONAL AND NEAR POPULATION WITH THE RELATION TO THE RESEARCH NUCLEAR REACTOR IEA-R1
11 International Nuclear Atlantic Conference - INAC 11 Belo Horizonte MG, Brazil, October 24-28, 11 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-8-99141-4- THE AWARENESS OF THE FUNCTIONAL
More informationEvaluation of the image quality in computed tomography: different phantoms
Artigo Original Revista Brasileira de Física Médica.2011;5(1):67-72. Evaluation of the image quality in computed tomography: different phantoms Avaliação da qualidade de imagem na tomografia computadorizada:
More informationSecondary Neutrons in Proton and Ion Therapy
Secondary Neutrons in Proton and Ion Therapy L. Stolarczyk Institute of Nuclear Physics PAN, Poland on behalf of WG9 EURADOS Acknowledgments EURADOS Workig Group 9 Roger Harrison Jean Marc Bordy Carles
More informationUsing Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues
Journal of mathematics and computer science 9 (2014), 41-45 Using Monte Carlo method for simulation of electromagnetic radiation penetration in body ' s tissues S. A. Mahdipour 1, H. R. Abdi Roknabadi
More informationMonte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration.
Monte Carlo simulation of a scanning whole body counter and the effect of BOMAB phantom size on the calibration. Gary H. Kramer, Linda C. Burns and Steven Guerriere Human Monitoring Laboratory, Radiation
More informationDose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
More informationPrepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
More informationUse of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
More informationQuality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
More informationAN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
More informationTowards a National Dose Registry in Portugal
Towards a National Dose Registry in Portugal M.B. Martins, J.G. Alves, A.R. Roda, J.N. Abrantes Instituto Tecnológico e Nuclear, Departamento de Protecção Radiológica e Segurança Nuclear, Estrada Nacional
More informationCT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
More informationTest Section for Experimental Simulation of Loss of Coolant Accident in an Instrumented Fuel Assembly Irradiated in the IEA-R1 Reactor
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 Test Section for Experimental
More informationLaue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
More informationEXTERNAL DOSE RATES IN COASTAL URBAN ENVIRONMENTS IN BRASIL
X Congreso Regional Latinoamericano IRPA de Protección y Seguridad Radiológica Radioprotección: Nuevos Desafíos para un Mundo en Evolución Buenos Aires, 12 al 17 de abril, 2015 SOCIEDAD ARGENTINA DE RADIOPROTECCIÓN
More informationQA of intensity-modulated beams using dynamic MLC log files
36 Original Article QA of intensity-modulated beams using dynamic MLC log files M. Dinesh Kumar, N. Thirumavalavan, D. Venugopal Krishna, M. Babaiah Department of Radiation Oncology, Yashoda Cancer Institute,
More informationConversion coefficients from air kerma to personal dose equivalent H p (3) for eye-lens dosimetry
Conversion coefficients from air kerma to personal dose equivalent H p (3) for eye-lens dosimetry ORAMED project Rapport CEA J. Daures (1), J. Gouriou (1), JM Bordy (1) (1) CEA, LIST, Laboratoire National
More informationBANK BRADESCO FORM 6-K/A. (Amended Report of Foreign Issuer) Filed 07/18/13 for the Period Ending 09/30/13
BANK BRADESCO FORM 6-K/A (Amended Report of Foreign Issuer) Filed 07/18/13 for the Period Ending 09/30/13 CIK 0001160330 Symbol BBD SIC Code 6022 - State Commercial Banks Industry Regional Banks Sector
More informationDOSE RATES EVALUATION ON ENVIRONMENT AROUND ALMIRANTE ALVARO ALBERTO NUCLEAR POWER STATION AFTER ANGRA 2 NUCLEAR POWER PLANT S OPERATION
2007 International Nuclear Atlantic Conference INAC 2007 Santos, SP, Brazil, September 29 to October 5, 2007 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR ABEN DOSE RATES EVALUATION ON ENVIRONMENT AROUND ALMIRANTE
More informationJosé Egidio M.Tondin 1, Mario O. de Menezes 1, Marina B.A.Vasconcellos 1 and João A.Osso Jr. 1 ABSTRACT 1. INTRODUCTION
PROSPECTION OF IMPLEMENTATION OF DISTANCE LEARNING AT IPEN CNEN/SP FOR THE COURSE ON FUNDAMENTALS OF NUCLEAR PHYSICS USING INFRASTRUCTURE OF FREE SOFTWARE José Egidio M.Tondin 1, Mario O. de Menezes 1,
More informationX-ray Radiation Safety Course. James Kane & Rob Deters Office of Radiological Control 545-7581
X-ray Radiation Safety Course James Kane & Rob Deters Office of Radiological Control 545-7581 About the Course X-ray Radiation Safety X-ray radiation safety training is mandatory for radiation workers
More informationStudy the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
More informationCALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY
243 CALCULATION METHODS OF X-RAY SPECTRA: A COMPARATIVE STUDY B. Chyba, M. Mantler, H. Ebel, R. Svagera Technische Universit Vienna, Austria ABSTRACT The accurate characterization of the spectral distribution
More informationLasers and Specturbility of Light Scattering Liquids - A Review
Artigo Original Revista Brasileira de Física Médica.2011;5(1):57-62. Reproducibility of radiant energy measurements inside light scattering liquids Reprodutibilidade de medidas de energia radiante dentro
More informationTABLE OF CONTENTS INSTALLATION MANUAL MASTERSAF DW
TABLE OF CONTENTS IMPORTANT INFORMATION...1 INSTALLATION PROCEDURES FOR THE DATABASE... 3 PREREQUISITES... 3 GUIDELINES... 3 DATABASE VALIDATION (OBLIGATORY)...8 INSTALLATION (FOR CLIENT - APPLICATION)...
More informationMODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, j.setti@pucpr.br Pontifícia Universidade Católica do Paraná / Rua Imaculada
More informationNUCLEAR ENERGY EDUCATION SCENARIO AROUND THE WORLD
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 NUCLEAR ENERGY EDUCATION SCENARIO
More informationTreating Thyroid Cancer using I-131 Maximum Tolerable Dose Method
Treating Thyroid Cancer using I-131 Maximum Tolerable Dose Method Christopher Martel, M.Sc., CHP Lisa Thornhill,, NRRPT, RT(NM) Boston University Medical Center Thyroid Carcinoma New cases and deaths in
More informationBREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY
BREAST CHARACTERISTICS AND DOSIMETRIC DATA IN X RAY MAMMOGRAPHY - A LARGE SAMPLE WORLDWIDE SURVEY N. GEERAERT a,b,c, R. KLAUSZ a, S. MULLER a, I. BLOCH c, H. BOSMANS b a GE Healthcare, Buc, France b Department
More informationHow To Write A Solar Resource Map
BRAZILIAN SOLAR RESOURCE ATLAS CD-ROM Chigueru Tiba e Naum Fraidenraich Departamento de Energia Nuclear Universidade Federal de Pernambuco -UFPE Av. Prof. Luiz Freire, 1000 Cidade Universitária 50.740-540
More informationTHE RESTRUCTURING OF THE QUALITY MANAGEMENT SYSTEM OF CDTN
2011 International Nuclear Atlantic Conference - INAC 2011 Belo Horizonte,MG, Brazil, October 24-28, 2011 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-04-5 THE RESTRUCTURING OF THE
More informationVIRTUAL INSTRUMENTATION ON MOBILE DEVICES FOR DEPLOYMENT IN NUCLEAR INSTALLATIONS
2013 International Nuclear Atlantic Conference - INAC 2013 Recife, PE, Brazil, November 24-29, 2013 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-05-2 VIRTUAL INSTRUMENTATION ON MOBILE
More informationPROCESSING OF AISI M2 HSS WITH ADDITION OF NbC BY MECHANICAL ALLOYING USING TWO DIFFERENT TYPES OF ATTRITOR MILLS
Materials Science Forum Vols. 660-661 (2010) pp 17-22 Online available since 2010/Oct/25 at www.scientific.net (2010) Trans Tech Publications, Switzerland doi:10.4028/www.scientific.net/msf.660-661.17
More informationAPPLIED SOLAR ENERGY IN THE REDUCTION OF THE POISON OF THE RICINUS COMMUNIS PIE
APPLIED SOLAR ENERGY IN THE REDUCTION OF THE POISON OF THE RICINUS COMMUNIS PIE José Ubiragi de Lima Mendes Laboratório de Mecânica dos Fluidos - DEM / CT / NTI / UFRN Campus Universitário Lagoa Nova,
More informationarxiv:1401.3092v1 [physics.ins-det] 14 Jan 2014
Monte Carlo Simulation of Indoor External Exposure due to Gamma-emitting Radionuclides in Building Materials arxiv:1401.3092v1 [physics.ins-det] 14 Jan 2014 DENG Jun 1) CAO Lei SU Xu 2) Key Laboratory
More informationGuidelines for a risk management methodology for product design
Guidelines for a risk management methodology for product design Viviane Vasconcellos Ferreira Federal University of Santa Catarina viviane@nedip.ufsc.br André Ogliari Federal University of Santa Catarina
More informationPatient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationRADIATION THERAPY FACULTY TOBJECTIVES. Post-Baccalaureate Certificate. Marissa Johnson, M.S.R.S., R.T. (T)
R A D I A T I O N T H E R A P Y c l i n i c a l n u t r i t i o n e m e r g e n c y m e d i c i n e e d u c a t i o n h e a l t h c a r e s c i e n c e s p h y s i c a l t h e r a p y p h y s i c i a n
More informationASSESSMENT BASED ON NAIVE BAYES FOR TRAINING BASED ON VIRTUAL REALITY
ASSESSMENT BASED ON NAIVE BAYES FOR TRAINING BASED ON VIRTUAL REALITY Ronei Marcos de Moraes 1, Liliane dos Santos Machado 2 Abstract Nowadays several areas present training systems based on virtual reality.
More informationHUMAN FACTORS QUESTIONNAIRE AS A TOOL FOR RISK ASSESSMENT
009 International Nuclear Atlantic Conference - INAC 009 Rio de Janeiro,RJ, Brazil, September7 to October, 009 ASSOCIAÇÃO BRASILEIRA DE ENERGIA NUCLEAR - ABEN ISBN: 978-85-99141-03-8 HUMAN FACTORS QUESTIONNAIRE
More informationDedicated computer software to radiation dose optimization for the staff performing nuclear medicine procedures
NUKLEONIKA 2012;57(4):497 502 ORIGINAL PAPER Dedicated computer software to radiation dose optimization for the staff performing nuclear medicine procedures Katarzyna Matusiak, Jakub Kosek Abstract. Nuclear
More informationWhat is the Radiotherapy Quality Control Program (PQRT) of the National Cancer Institute Rio de Janeiro/Brazil?
Congreso Internacional Conjunto Cancún 2004 LAS/ANS-SNM-SMSR/International Joint Meeting Cancun 2004 LAS/ANS-SNM-SMSR XV Congreso Anual de la SNM y XXII Reunión Anual de la SMSR/XV SNM Annual Meeting and
More informationPavilhão de Verificação de Óbitos Source: Brazil Builds, de Philip Goodwin, 1943
Full Documentation Fiche 2003 composed by national/regional working party of: 0. Picture of building/ group of buildings/ urban scheme/ landscape/ garden depicted item: Pavilhão de Verificação de Óbitos
More informationDose enhancement near metal electrodes in diamond X- ray detectors. A. Lohstroh*, and D. Alamoudi
Dose enhancement near metal electrodes in diamond X- ray detectors Acknowledgements Surrey University: P.J. Sellin M. Abd-El Rahman P. Veeramani H. Al-Barakaty F. Schirru Mechanical Workshop A. Lohstroh*,
More informationVariance reduction techniques used in BEAMnrc
Variance reduction techniques used in BEAMnrc D.W.O. Rogers Carleton Laboratory for Radiotherapy Physics. Physics Dept, Carleton University Ottawa, Canada http://www.physics.carleton.ca/~drogers ICTP,Trieste,
More informationRequired RS Training Info
C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness
More informationRadiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment.
Dictionary Radiation therapy involves using many terms you may have never heard before. Below is a list of words you could hear during your treatment. Applicator A device used to hold a radioactive source
More informationRecife, December 3 rd, 2008
Recife, December 3 rd, 2008 My name is Eduardo Araujo Oliveira. I graduated in Computer Science (B.Sc.) from the Catholic University of Pernambuco in 2005 and pursued further studies through a master degree
More informationClinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University
Clinical Physics Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University Physics of Radiotherapy using External Beam Dose distribution PHANTOMS Basic dose distribution data are usually
More informationConsequence for a dualband application
rel. bandwidth -6dB [%] DESIGN CONSIDERATIONS FOR INTEGRATED MOBILE PHONE ANTENNAS D. Manteuffel, A. Bahr, D. Heberling, I. Wolff IMST GmbH, Germany, e-mail: manteuffel@imst.de Abstract Based on the investigation
More informationRADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
More informationHow To Use A Proton For Radiation Therapy
Proton Cancer Therapy: Using Laser Accelerate Protons for Radiation Therapy A Thesis Submitted in Partial Fulfillment of the Requirements for Graduation with Research Distinction in Engineering Physics
More informationRecognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
More informationThe Challenge of CT Dose Records
The Challenge of CT Dose Records Kimberly Applegate, MD, MS, FACR Professor of Radiology and Pediatrics Emory University *financial disclosures: -Springer Textbook contracts -AIM advisory board for patient
More informationREPORT ORGANIZATION ABSTRACT I. INTRODUCTION CALCULATION OF EFFECTIVE DOSE PER IMAGE ACQUISITION CALCULATION OF EFFECTIVE DOSE PER TREATMENT COURSE
E S T I M A T I O N O F T H E I M A G I N G D O S E F O R T H E C Y B E R K N I F E R O B O T I C R A D I O S U R G E R Y S Y S T E M REPORT ORGANIZATION ABSTRACT I. INTRODUCTION II. III. IV. CALCULATION
More informationI-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit
I-Max Touch Range PAN / CEPH / 3D digital panoramic unit Evolutive 3 in 1 panoramic unit 3D A new dimension for a complete diagnosis I-Max Touch 3D Evolutive, simple, fast The panoramic unit realizes complete
More informationRadiologic Technology Training Program
Radiologic Technology Training Program 438 Radiologic Technology Training Program Radiologic Technology Training Program Program Coordinator: Mansour, Zepure General Information The Radiologic Technology
More informationVisualization of Output Data from Particle Transport Codes
Visualization of Output Data from Particle Transport Codes Phase 1 Final Report Instrument No: DE-FG02-07ER84735 June 20, 2007 through March 19, 2008 Recipient: Randolph Schwarz, Visual Editor Consultants
More informationIV WORKSHOP RNP2 (WRNP2)
IV WORKSHOP RNP2 (WRNP2) Hotel Pirâmide, Natal, RN, 19-20 de Maio de 2003 Advanced Network Engineering and Applications from Space Missions - AUTHORS - Eduardo W. Bergamini Serviços Aplicativos em Missões
More informationhttp://www.bls.gov/oco/ocos104.htm
http://www.bls.gov/oco/ocos104.htm Nuclear Medicine Technologists Nature of the Work Training, Other Qualifications, and Advancement Employment Job Outlook Projections Data Earnings OES Data Related Occupations
More informationOLTARIS: An Efficient Web-Based Tool for Analyzing Materials Exposed to Space Radiation
OLTARIS: An Efficient Web-Based Tool for Analyzing Materials Exposed to Space Radiation Tony C. Slaba *a, Amelia M. McMullen b, Sheila A. Thibeault a, Chris A. Sandridge a, Martha S. Clowdsley a, Steve
More informationTRAINING PROGRAMS IN MEDICAL PHYSICS IN THE UNITED STATES
11 I '* / TRAINING PROGRAMS IN MEDICAL PHYSICS IN THE UNITED STATES Lawrence H. Lanzl, Ph.D. Professor of Medical Physics Department of Radiology and The Franklin McLean a '" Memorial Research Institute,
More informationProtons vs. CyberKnife. Protons vs. CyberKnife. Page 1 UC SF. What are. Alexander R. Gottschalk, M.D., Ph.D.
Protons vs. CyberKnife UC SF Protons vs. CyberKnife UC SF Alexander R. Gottschalk, M.D., Ph.D. Associate Professor and Director of the CyberKnife Radiosurgery Program Department of Radiation Oncology University
More informationSAR Distribution in Test Animals Exposed to RF Radiation
German Mobile Telecommunication Research Programme, July 25th, 2006 SAR Distribution in Test Animals Exposed to RF Radiation Verónica Berdiñas Torres, Andreas Christ, Niels Kuster IT IS Foundation, ETH
More informationModel SL355 Noise Dosimeter/Datalogger
Model SL355 Noise Dosimeter/Datalogger Software Help Guide Software Introduction The SL355 Noise Dosimeter software allows for setup and retrieval of recorded data in the Noise Dosimeter. A total of 20
More information