USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
|
|
|
- Darrell Lee
- 9 years ago
- Views:
Transcription
1 Iron, Total DOC USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required for determining total iron. 1 USEPA approved for reporting wastewater analysis, Federal Register, June 27, 1980; 45 (126:43459). 2 Adapted from Standard Methods for the Examination of Water and Wastewater. Test preparation Instrument-specific information Table 1 shows sample cell and orientation requirements for reagent addition tests, such as powder pillow or bulk reagent tests. Table 2 shows sample cell and adapter requirements for AccuVac Ampul tests. The tables also show all of the instruments that have the program for this test. To use the table, select an instrument, then read across to find the applicable information for this test. Table 1 Instrument-specific information for reagent addition Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR 3900 The fill line is to the right The fill line is toward the user. DR 900 The orientation mark is toward the user Table 2 Instrument-specific information for AccuVac Ampuls Instrument Adapter Sample cell DR 6000 DR 5000 DR 900 DR 3900 DR 1900 DR 3800 DR 2800 DR LZV846 (A) or (C) LZV584 (C)
2 Before starting Install the instrument cap on the DR 900 cell holder before ZERO or READ is pushed. The reagent in this test procedure converts all soluble iron and most insoluble forms of iron in the sample to soluble ferrous iron for measurement. For regulatory reporting, however, the sample must be digested with heat and acid to make sure that all forms of the metal are measured. For the best results, measure the reagent blank value for each new lot of reagent. Replace the sample with deionized water in the test procedure to determine the reagent blank value. Subtract the reagent blank value from the sample results automatically with the reagent blank adjust option. For turbid samples, treat the blank with one 0.1-g scoop of RoVer Rust Remover. Swirl to dissolve. Review the Safety Data Sheets (MSDS/SDS) for the chemicals that are used. Use the recommended personal protective equipment. Dispose of reacted solutions according to local, state and federal regulations. Refer to the Safety Data Sheets for disposal information for unused reagents. Refer to the environmental, health and safety staff for your facility and/or local regulatory agencies for further disposal information. Items to collect Powder pillows Description Quantity FerroVer Iron Reagent Powder Pillows, 10-mL 1 1 Sample cells. (For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1.) 2 1 FerroVer is a registered trademark of Hach Company. Refer to Consumables and replacement items on page 7 for order information. AccuVac Ampuls Description Quantity FerroVer Iron Reagent AccuVac Ampul 1 Beaker, 50-mL 1 Sample cells (For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1.) 1 Stopper for 18-mm tubes and AccuVac Ampuls 1 Refer to Consumables and replacement items on page 7 for order information. Sample collection and storage Collect samples in clean glass or plastic bottles that have been cleaned with 6 N (1:1) hydrochloric acid and rinsed with deionized water. To measure only dissolved iron, filter the sample immediately after collection and before acidification. To preserve samples for later analysis, adjust the sample ph to less than 2 with concentrated nitric acid (approximately 2 ml per liter). No acid addition is necessary if the sample is tested immediately. Keep the preserved samples at room temperature for a maximum of 6 months. Before analysis, adjust the ph to 3 5 with 5 N sodium hydroxide solution. Correct the test result for the dilution caused by the volume additions. 2 Iron, Total, FerroVer Method (3.00 mg/l)
3 Powder pillow procedure Start 1. Start program 265 Iron, FerroVer. For information about sample cells, adapters or light shields, refer to Instrument-specific information on page Prepare the sample: Fill a sample cell with 10 ml of sample. 3. Add the contents of one FerroVer Iron Reagent Powder Pillow to the sample cell. 4. Swirl the sample cell to mix. Undissolved powder will not affect accuracy. Note: Although the program name can be different between instruments, the program number does not change. 5. Start the instrument timer. A 3-minute reaction time starts. 6. Prepare the blank: Fill a second sample cell with 10 ml of sample. 7. Clean the blank sample cell. 8. When the timer expires, insert the blank into the cell holder. An orange color will show if iron is present. Let samples that contain rust react for 5 minutes or more. Zero Read 9. Push ZERO. The display shows 0.00 mg/l Fe. 10. Clean the prepared sample cell. 11. Insert the prepared sample into the cell holder. 12. Push READ. Results show in mg/l Fe. Iron, Total, FerroVer Method (3.00 mg/l) 3
4 AccuVac procedure Start 1. Start program 267 Iron, FerroVer AV. For information about sample cells, adapters or light shields, refer to Instrumentspecific information on page 1. Note: Although the program name can be different between instruments, the program number does not change. 2. Prepare the blank: Fill the sample cell with 10 ml of sample. 3. Prepare the sample: Collect at least 40 ml of sample in a 50-mL beaker. Fill the AccuVac Ampul with sample. Keep the tip immersed while the AccuVac Ampul fills completely. 4. Quickly invert the AccuVac Ampul several times to mix. Undissolved powder will not affect accuracy. Zero 5. Start the instrument timer. A 3-minute reaction time starts. 6. Clean the blank sample cell. 7. When the timer expires, insert the blank into the cell holder. 8. Push ZERO. The display shows 0.00 mg/l Fe. An orange color will show if iron is present. Let samples that contain rust react for 5 minutes or more. Read 9. Clean the AccuVac Ampul. 10. Insert the prepared sample AccuVac Ampul into the cell holder. 11. Push READ. Results show in mg/l Fe. 4 Iron, Total, FerroVer Method (3.00 mg/l)
5 Interferences Interfering substance Interference level Calcium, Ca 2+ No effect at less than 10,000 mg/l as CaCO 3. Chloride, Cl Copper, Cu 2+ High iron levels Iron oxide No effect at less than 185,000 mg/l. No effect. Masking agent is contained in FerroVer Reagent. Inhibit color development. Dilute sample and re-test to verify results. A mild, vigorous or Digesdahl digestion is necessary. After digestion, adjust the sample ph to 3 5 with sodium hydroxide, then analyze. Magnesium No effect at 100,000 mg/l as CaCO 3. Molybdate molybdenum High sulfide levels, S 2 No effect at 50 mg/l as Mo. Pretreat the sample in a fume hood or in an area with sufficient airflow before analysis: 1. Add 5 ml of 6.0 N (1:1) hydrochloric acid solution to 100 ml of sample in a 250-mL Erlenmeyer flask. 2. Boil for 20 minutes. 3. Let the solution cool to room temperature. 4. Adjust the ph to 3 5 with 5 N sodium hydroxide solution. 5. Add deionized water until the volume is 100 ml. 6. Use the treated sample in the test procedure. Turbidity Pre-treat the sample before analysis: 1. Add one 0.1-g scoop of RoVer Rust Remover to the blank. Swirl to mix. 2. If the sample remains turbid, add three 0.2-g scoops of RoVer Rust Remover to 75 ml of sample. Let stand 5 minutes. 3. Filter through a 0.45-micron membrane filter and filter holder. 4. Use the treated sample in the test procedure. Highly buffered samples or extreme sample ph Can prevent the correct ph adjustment of the sample by the reagents. Sample pre-treatment may be necessary. Adjust the ph to 3 5. Accuracy check Standard additions method (sample spike) Use the standard additions method (for applicable instruments) to validate the test procedure, reagents and instrument and to find if there is an interference in the sample. Items to collect: Iron Voluette Ampule Standard, 25 mg/l Ampule Breaker Pipet, TenSette, ml and tips 1. Use the test procedure to measure the concentration of the sample, then keep the (unspiked) sample in the instrument. 2. Go to the Standard Additions option in the instrument menu. 3. Select the values for standard concentration, sample volume and spike volumes. 4. Open the standard solution. 5. Prepare three spiked samples: use the TenSette pipet to add 0.1 ml, 0.2 ml and 0.3 ml of the standard solution, respectively, to three 10-mL portions of fresh sample. Mix well. Note: For AccuVac Ampuls, add 0.2 ml, 0.4 ml and 0.6 ml of the standard solution to three 50-mL portions of fresh sample. Iron, Total, FerroVer Method (3.00 mg/l) 5
6 6. Use the test procedure to measure the concentration of each of the spiked samples. Start with the smallest sample spike. Measure each of the spiked samples in the instrument. 7. Select Graph to compare the expected results to the actual results. Note: If the actual results are significantly different from the expected results, make sure that the sample volumes and sample spikes are measured accurately. The sample volumes and sample spikes that are used should agree with the selections in the standard additions menu. If the results are not within acceptable limits, the sample may contain an interference. Standard solution method Use the standard solution method to validate the test procedure, the reagents and the instrument. Items to collect: Iron standard solution, 100 mg/l 100-mL volumetric flask, Class A 2-mL volumetric pipet, Class A and pipet filler Deionized water 1. Prepare a 2.00-mg/L iron standard solution as follows: a. Use a pipet to add 2 ml of the 100-mg/L iron standard solution into the volumetric flask. b. Dilute to the mark with deionized water. Mix well. Prepare this solution daily. 2. Use the test procedure to measure the concentration of the prepared standard solution. 3. Compare the expected result to the actual result. Note: The factory calibration can be adjusted slightly with the standard adjust option so that the instrument shows the expected value of the standard solution. The adjusted calibration is then used for all test results. This adjustment can increase the test accuracy when there are slight variations in the reagents or instruments. Method performance The method performance data that follows was derived from laboratory tests that were measured on a spectrophotometer during ideal test conditions. Users can get different results under different test conditions. Program Standard Precision (95% Confidence Interval) Sensitivity Concentration change per Abs change mg/l Fe mg/l Fe mg/l Fe mg/l Fe mg/l Fe mg/l Fe Summary of method FerroVer Iron Reagent converts all soluble iron and most insoluble forms of iron in the sample to soluble ferrous iron. The ferrous iron reacts with the 1-10 phenanthroline indicator in the reagent to form an orange color in proportion to the iron concentration. The measurement wavelength is 510 nm for spectrophotometers or 520 nm for colorimeters. 6 Iron, Total, FerroVer Method (3.00 mg/l)
7 Consumables and replacement items Required reagents Description Quantity/Test Unit Item no. FerroVer Iron Reagent Powder Pillow 1, 10-mL 1 100/pkg OR FerroVer Iron Reagent AccuVac Ampul 1 25/pkg FerroVer is a registered trademark of Hach Company Required apparatus Description Quantity/Test Unit Item no. Beaker, 50-mL 1 each 50041H Stoppers for 18-mm tubes and AccuVac Ampuls 2 6/pkg Recommended standards and apparatus Description Unit Item no. Flask, volumetric, Class A, 100-mL glass each Iron Standard Solution, 100-mg/L Fe 100 ml Iron Standard Solution, 10-mL Voluette Ampule, 25-mg/L Fe 16/pkg Metals Drinking Water Standard, LR for Cu, Fe, Mn 500 ml Metals Drinking Water Standard, HR for Cu, Fe, Mn 500 ml Pipet filler, safety bulb each Pipet, TenSette, ml each Pipet tips for TenSette Pipet, ml 50/pkg Pipet tips for TenSette Pipet, ml 1000/pkg Pipet, volumetric, Class A, 2-mL each Water, deionized 4 L Optional reagents and apparatus Description Unit Item no. AccuVac Ampul Snapper each Mixing cylinder, graduated, 50-mL each Filter, glass fiber membrane, 1.5-micron, 47-mm 100/pkg Filter membrane filter holder, 47-mm each Hydrochloric Acid, concentrated 500 ml Nitric Acid, concentrated 500 ml RoVer Rust Remover 454 g Sodium Hydroxide Standard Solution, 5.0 N 100 ml MDB Spoon, measuring, 0.1-g each Iron, Total, FerroVer Method (3.00 mg/l) 7
8 FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING: In the U.S.A. Call toll-free Outside the U.S.A. Contact the HACH office or distributor serving you. On the Worldwide Web HACH COMPANY WORLD HEADQUARTERS Telephone: (970) FAX: (970) Hach Company/Hach Lange GmbH, All rights reserved. 01/2014, Edition 9
FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows
Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for
Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
Phosphorus, Reactive (Orthophosphate)
Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from
Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution
Volatile Acids DOC316.53.01144 Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution Scope and application: For digestor sludges. 1 Adapted from The Analyst, 87, 949 (1962).
LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506
Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking
Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
Quaternary Ammonium Compounds DOC316.53.01128
Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower
Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information
Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater
DR/4000 PROCEDURE. CHLORINE, Free
DR/4000 PROCEDURE CHLORINE, Free Method 10069 DPD method* UHR (0.1 10.0 mg/l as Cl 2 ) Scope and Application: For testing higher levels of free chlorine (hypochlorous acid and hypochlorite ion) in drinking
Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate
Oxygen Scavengers DOC316.53.01105 Iron Reduction Method Method 8140 3 to 450 µg/l DEHA; 5 to 600 µg/l carbohydrazide; 9 to 1000 µg/l hydroquinone; 13 to 1500 µg/l iso-ascorbic acid (ISA); 15 to 1000 µg/l
Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater
Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Salicylate Method * 1. Enter the stored program number for ammonia nitrogen (NH 3 -N). 2. Press: 64 ENTER The display
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
USEPA Electrode Method Method 8156
ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method
Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor
Oxygen, Dissolved DOC316.53.01241 Direct Measurement Method Method 8157 0 to 20.0 mg/l (or 0 to 200% saturation) O 2 Clark-type Amperometric Sensor Scope and application: For water, wastewater and process
An Introduction to Standards and Quality Control for the Laboratory Barbara Martin
An Introduction to Standards and Quality Control for the Laboratory Barbara Martin Hach Company, 2002. All rights reserved. Lit. No. 2426 eac/rb 8/02 1ed In memory of Clifford C. Hach (1919-1990) inventor,
Solids, Volatile Dissolved and Fixed Dissolved
, 8277 Solids, Volatile Dissolved and Fixed Dissolved Gravimetric Method 1 Scope and Application: For wastewater. 1 Adapted from Standard Methods for the Examination of Water and Wastewater DOC316.53.001206
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS Sl. No. Contents Preamble 5.1 Aim 5.2 Introduction 5.2.1 Environmental Significance 5.3 Principle 5.4 Materials Required 5.4.1 Apparatus Required 5.4.2
Total Aerobic Bacteria, Yeasts and Molds
Total Aerobic Bacteria, Yeasts and Molds DOC316.53.01223 Total Aerobic Bacteria (Amber), Yeast and Mold (Red), Total Coliform (Red) and Disinfection Control (Purple) Paddle Testers Scope and application:
Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents
Colorimetric Determination of Iron in Vitamin Tablets
Colorimetric Determination of Iron in Vitamin Tablets Big Picture Conceptual Approach Vitamin Tablet How much Fe? ph = 3.5 Vitamin Tablet How much Fe? Too difficult to eyeball so will have the colorimeter
4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES
4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES Sl. No. Contents Preamble 4.1 Aim 4.2 Introduction 4.2.1 Environmental Significance 4.3 Principle 4.4 Materials Required 4.4.1 Apparatus Required 4.4.2 Chemicals
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure
Recovery of Elemental Copper from Copper (II) Nitrate
Recovery of Elemental Copper from Copper (II) Nitrate Objectives: Challenge: Students should be able to - recognize evidence(s) of a chemical change - convert word equations into formula equations - perform
Chapter 16: Tests for ions and gases
The position of hydrogen in the reactivity series Hydrogen, although not a metal, is included in the reactivity series because it, like metals, can be displaced from aqueous solution, only this time the
Colorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
XI. Methods of Analysis DETERMINATION OF POTASSIUM CARBONATE CALCULATIONS REAGENTS PROCEDURE
XI. NOTE: An automatic titrator may be utilized for ease of analysis, especially if this test is performed often. Contact Technical Service for further information concerning automatic titration. Methods
Thermo Scientific Orion AQUAfast Powder and Tablet Reagent Chemistry User Guide
Thermo Scientific Orion AQUAfast Powder and Tablet Reagent Chemistry User Guide for use with AquaMate 7000 Vis and AquaMate 8000 UV-Vis Spectrophotometers 2012 Thermo Fisher Scientific Inc. All rights
OXIDATION-REDUCTION TITRATIONS-Permanganometry
Experiment No. Date OXIDATION-REDUCTION TITRATIONS-Permanganometry INTRODUCTION Potassium permanganate, KMnO 4, is probably the most widely used of all volumetric oxidizing agents. It is a powerful oxidant
Determination of Citric Acid in Powdered Drink Mixes
Determination of Citric Acid in Powdered Drink Mixes Citric acid and its salts (sodium citrate and potassium citrate) are found in many foods, drinks, pharmaceuticals, shampoos, and cosmetics. The tartness
NITRIC OXIDE and NITROGEN DIOXIDE 6014
NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm
Pocket Colorimeter II
DOC022.53.80451 Pocket Colorimeter II User Manual 04/2014, Edition 1 Table of Contents Section 1 Specifications...3 Section 2 General information...5 2.1 Safety information...5 2.1.1 Use of hazard information...5
STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14
STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14 OBJECTIVE The objective of this experiment will be the standardization of sodium hydroxide using potassium hydrogen phthalate by the titration
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow
Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow Parameter and code: Phosphorus, total-in-bottom-material, dry weight, I-6600-88 (mg/kg as P): 00668 1. Application This method is used
Phosphates (ortho- and total)
INTRODUCTION Phosphates (ortho- and total) Phosphorus is an essential nutrient for all aquatic plants and algae. Only a very small amount is needed, however, so an excess of phosphorus can easily occur.
EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..
Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set
EXPERIMENT 12: Empirical Formula of a Compound
EXPERIMENT 12: Empirical Formula of a Compound INTRODUCTION Chemical formulas indicate the composition of compounds. A formula that gives only the simplest ratio of the relative number of atoms in a compound
COMMON LABORATORY APPARATUS
COMMON LABORATORY APPARATUS Beakers are useful as a reaction container or to hold liquid or solid samples. They are also used to catch liquids from titrations and filtrates from filtering operations. Bunsen
Syllabus OC18 Use litmus or a universal indicator to test a variety of solutions, and classify these as acidic, basic or neutral
Chemistry: 9. Acids and Bases Please remember to photocopy 4 pages onto one sheet by going A3 A4 and using back to back on the photocopier Syllabus OC18 Use litmus or a universal indicator to test a variety
Reaction Stoichiometry and the Formation of a Metal Ion Complex
Reaction Stoichiometry and the Formation of a Metal Ion Complex Objectives The objectives of this laboratory are as follows: To use the method of continuous variation to determine the reaction stoichiometry
Calcium Analysis by EDTA Titration
Calcium Analysis by EDTA Titration ne of the factors that establish the quality of a water supply is its degree of hardness. The hardness of water is defined in terms of its content of calcium and magnesium
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014 This experiment explores quality assurance practices which are commonly
Solutions and Dilutions
Learning Objectives Students should be able to: Content Design a procedure for making a particular solution and assess the advantages of different approaches. Choose the appropriate glassware to ensure
Preparation of frequently used solutions
Preparation of frequently used solutions Content 1. Diluting Concentrated Acids (Last Login: 08/08/2009) 2. Indicators (Last Login: 27/07/2009) 3. Standard Buffer Solutions (Last Login: 27/07/2009) 4.
Factors Affecting Enzyme Activity
INTRODUCTION Factors Affecting Enzyme Activity The chemical reactions occurring in living things are controlled by enzymes. An enzyme is a protein in the cell which lowers the activation energy of a catalyzed
DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included.
DYES AND DYEING 2003 by David A. Katz. All rights reserved. Permission for classroom use provided original copyright is included. Dyeing of textiles has been practiced for thousands of years with the first
Complexometric Titration Analysis of Ca 2+ and Mg 2+ in seawater
Complexometric Titration Analysis of Ca 2+ and Mg 2+ in seawater Introduction As the mountains on the continents are draped with snow, the mountains on the ocean floor are draped with sediment rich in
LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e
1 SIDE BY SIDE COMPARISON TABLE FOR THE DETERMINATION OF CYANIDE BY UV DIGESTION AND AMPEROMETRIC DETECTION FIA METHOD - 10-204-00-5-D (ASTM D7511-09 e ) TOPIC Method ASTM D7511-09e LACHAT METHOD NUMBER
DRINKING WATER - LAB EXPERIMENTS. Coagulation and flocculation LAB EXPERIMENTS. Jartest
DRINKING WATER - LAB EXPERIMENTS LAB EXPERIMENTS Coagulation and flocculation Jartest coagulation and flocculation - jartest lab experiments Framework This module explains the lab experiment on coagulation
Experiment 16-Acids, Bases and ph
Definitions acid-an ionic compound that releases or reacts with water to form hydrogen ion (H + ) in aqueous solution. They taste sour and turn litmus red. Acids react with certain metals such as zinc,
1.4 This method is capable of measuring DO in the range of 0.20 to 20 mg/l.
Luminescence Measurement of Dissolved Oxygen in Water and Wastewater and for Use in the Determination of BOD 5 and cbod 5 Revision 1.2 October 2011 1. Scope and Application 1.1 This method is for the measurement
Physical and Chemical Properties and Changes
Physical and Chemical Properties and Changes An understanding of material things requires an understanding of the physical and chemical characteristics of matter. A few planned experiments can help you
A Visual Method for the Detection of Arsenic 0 to 500 µg/l By Dan Kroll R&D Chemist. Hach Co., Loveland Colorado
Introduction: Arsenic is a common contaminant of ground water that has been found to adversely effect human health at levels as low as 10 µg/l. The current USEPA maximum contaminant level (MCL) for arsenic
OBJECTIVES: Visitors learn what an antioxidant is and how it behaves. They also learn how to test for the presence of vitamin C..
Vitamin C Visitors use iodine to compare the reactivity of two starch solutions one with vitamin C added, one without vitamin C. OBJECTIVES: Visitors learn what an antioxidant is and how it behaves. They
Solubility Product Constants
Solubility Product Constants PURPOSE To measure the solubility product constant (K sp ) of copper (II) iodate, Cu(IO 3 ) 2. GOALS 1 To measure the molar solubility of a sparingly soluble salt in water.
Determination of the Mass Percentage of Copper in a Penny. Introduction
Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny
Neutralizing an Acid and a Base
Balancing Act Teacher Information Objectives In this activity, students neutralize a base with an acid. Students determine the point of neutralization of an acid mixed with a base while they: Recognize
Total Water Hardness
Test 14 INTRODUCTION When water passes through or over deposits such as limestone, the levels of Ca 2+, Mg 2+, and HCO Ð 3 ions present in the water can greatly increase and Hard-Water Cations cause the
Experiment 12- Classification of Matter Experiment
Experiment 12- Classification of Matter Experiment Matter can be classified into two groups: mixtures and pure substances. Mixtures are the most common form of matter and consist of mixtures of pure substances.
Determination of calcium by Standardized EDTA Solution
Determination of calcium by Standardized EDTA Solution Introduction The classic method of determining calcium and other suitable cations is titration with a standardized solution of ethylenediaminetetraacetic
Appendix D lists the Field Services Standard Operating Procedures. Appendix E lists the Biological Monitoring Standard Operating Procedures.
Page 16 of 87 3.3 Sample Collection, Storage and Preservation Figure 3 details required containers, sample volumes, preservation techniques, and holding times for proper sample collection. A discussion
Mixtures and Pure Substances
Unit 2 Mixtures and Pure Substances Matter can be classified into two groups: mixtures and pure substances. Mixtures are the most common form of matter and consist of mixtures of pure substances. They
(1) Hydrochloric acid reacts with sodium hypochlorite to form hypochlorous acid: NaOCl(aq) + HCl(aq) HOCl(aq) + NaCl(aq) hypochlorous acid
The Determination of Hypochlorite in Bleach Reading assignment: Chang, Chemistry 10 th edition, pages 156-159. We will study an example of a redox titration in order to determine the concentration of sodium
Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media
Enhanced Organic Precursor Removals Using Aged Media Page 1 Enhanced Organic Precursor Removals Using Aged Media O b j e c t i v e s The main goals of this project was to evaluate the dissolved organic
Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration
Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration GOAL AND OVERVIEW Antacids are bases that react stoichiometrically with acid. The number of moles of acid that
Reagents also compatible in Hach
Reagents also compatible in Hach VARIO Powder Packs (PP) and Reagents for Photometry Test Range Reagent Liquid Reagent Tube Tests Powder Pack VARIO Aluminium Reagent, Set F20 Aluminium 0 0.22 mg/l Al VARIO
The Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
Experiment 8 Preparation of Cyclohexanone by Hypochlorite Oxidation
Experiment 8 Preparation of Cyclohexanone by ypochlorite xidation In this experiment we will prepare cyclohexanone from cyclohexanol using hypochlorite oxidation. We will use common household bleach that
Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.
TITRATION: STANDARDIZATION OF A BASE AND ANALYSIS OF STOMACH ANTACID TABLETS 2009, 1996, 1973 by David A. Katz. All rights reserved. Reproduction permitted for education use provided original copyright
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
Simulation of the determination of lead azide content in waste water from explosives manufacture
Simulation of the determination of lead azide content in waste water from explosives manufacture Lead azide ranks in the category of intensive explosives, which may, even in an insignificant amount, initiate
Experiment 7: Titration of an Antacid
1 Experiment 7: Titration of an Antacid Objective: In this experiment, you will standardize a solution of base using the analytical technique known as titration. Using this standardized solution, you will
MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1
MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry EXPERIMENT #5 THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1 I. PURPOSE OF THE EXPERIMENT In this experiment
DETERMINING THE ENTHALPY OF FORMATION OF CaCO 3
DETERMINING THE ENTHALPY OF FORMATION OF CaCO 3 Standard Enthalpy Change Standard Enthalpy Change for a reaction, symbolized as H 0 298, is defined as The enthalpy change when the molar quantities of reactants
EXPERIMENT 2 THE HYDROLYSIS OF t-butyl CHLORIDE. PURPOSE: To verify a proposed mechanism for the hydrolysis of t-butyl Chloride.
PURPOSE: To verify a proposed mechanism for the hydrolysis of t-butyl Chloride. PRINCIPLES: Once the Rate Law for a reaction has been experimentally established the next step is its explanation in terms
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.
Titration Curves PURPOSE To determine the equivalence points of two titrations from plots of ph versus ml of titrant added. GOALS 1 To gain experience performing acid-base titrations with a ph meter. 2
Ascorbic Acid Titration of Vitamin C Tablets This lab will be completed individually! Make sure you come prepared!
Ascorbic Acid Titration of Vitamin C Tablets This lab will be completed individually! Make sure you come prepared! Introduction Vitamin C (also known as ascorbic acid, HC6H7O6) is a necessary ingredient
NNIN Nanotechnology Education
NNIN Nanotechnology Education Student Guide Part 1: Silver Nanoparticle Synthesis and Spectroscopy Introduction: In this lab you will synthesize silver nanoparticles one of the most commonly used nanoparticles
The Molar Mass of a Gas
The Molar Mass of a Gas Goals The purpose of this experiment is to determine the number of grams per mole of a gas by measuring the pressure, volume, temperature, and mass of a sample. Terms to Know Molar
Vitamin C Content of Foods
Vitamin C Content of Foods Experiment #11 Objective: To measure the heat and alkaline stability of vitamin C and its quantity in juices or tablets. Introduction Vitamin C is an essential component of the
EDTA Titrations 1: Standardization of EDTA and Analysis of Zinc in a Supplement Tablet. by Professor David Cash. September, 2008
CHEMICAL, ENVIRONMENTAL, AND BIOTECHNOLOGY DEPARTMENT EDTA Titrations 1: Standardization of EDTA and Analysis of Zinc in a Supplement Tablet by Professor David Cash September, 2008 Mohawk College is the
SODIUM CARBOXYMETHYL CELLULOSE
SODIUM CARBOXYMETHYL CELLULOSE Prepared at the 28th JECFA (1984), published in FNP 31/2 (1984) and in FNP 52 (1992). Metals and arsenic specifications revised at the 55 th JECFA (2000). An ADI not specified
Honors 227 Fall 2007 Laboratory with Ms. Clark. Enzymes, Reactions, Metabolism and Homeostasis
1 Name: Honors 227 Fall 2007 Laboratory with Ms. Clark Enzymes, Reactions, Metabolism and Homeostasis Background Enzymes, which are comprised of amino acids, are very important macromolecules found in
Analysis of Vitamin C Using Iodine. Introduction
Analysis of Vitamin C Using Iodine Introduction Vitamin C (ascorbic acid) is oxidized to dehydroascorbic acid using a mild oxidizing agent such as iodine. The oxidation is a two- electron process, following
FAJANS DETERMINATION OF CHLORIDE
EXPERIMENT 3 FAJANS DETERMINATION OF CHLORIDE Silver chloride is very insoluble in water. Addition of AgNO 3 to a solution containing chloride ions results in formation of a finely divided white precipitate
ENCM 31073-Environmental Impact Assessment and Environmental Monitoring. Determination of nitrate and phosphate levels in well water/surface water
ENCM 31073-Environmental Impact Assessment and Environmental Monitoring ntalmonitoringenvironmental Monitoring - Practical Number 4 Determination of nitrate and phosphate levels in well water/surface water
PART I: PREPARATION OF SOLUTIONS AND STANDARDIZATION OF A BASE
TITRATION: STANDARDIZATION OF A BASE AND ANALYSIS OF STOMACH ANTACID TABLETS 2009, 1996, 1973 by David A. Katz. All rights reserved. Reproduction permitted for education use provided original copyright
Determination of Aspirin using Back Titration
Determination of Aspirin using Back Titration This experiment is designed to illustrate techniques used in a typical indirect or back titration. You will use the NaH you standardized last week to back
NIKAL BP RTU ELECTROLYTIC NICKEL SOLUTION For Advanced Packaging Applications
NIKAL BP RTU ELECTROLYTIC NICKEL SOLUTION For Advanced Packaging Applications Regional Product Availability N.America Japan/Korea Asia Europe Bath Make-up Component Nikal BP RTU Solution Percent by Volume
Chlorine Residual Testing Fact Sheet
Chlorine Residual Testing Fact Sheet The presence of free chlorine residual in drinking water indicates that: 1) a sufficient amount of chlorine was added to the water to inactivate most of the bacteria
Phenolphthalein-NaOH Kinetics
Phenolphthalein-NaOH Kinetics Phenolphthalein is one of the most common acid-base indicators used to determine the end point in acid-base titrations. It is also the active ingredient in some laxatives.
Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07
Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #: 3460-07 TABLE OF CONTENTS GENERAL INFORMATION... 2 Product Description... 2 Procedure Overview... 2 Kit Contents, Storage and Shelf Life... 3