Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb
|
|
|
- Claud Jones
- 9 years ago
- Views:
Transcription
1 Document: AND Lead Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., All rights reserved. Printed in USA.
2 Table of Contents 2 General Information Water Testing Guidelines 2.2 Sensor Bag: Cuvette and Sensor 2.3 Inserting Cuvette and Sensor 2.4 Sample Injection and Measurement 2.5 Pipetting Use Guidelines 3 On Site Calibration Required Materials 3.2 Preparing for On Site Calibration 3.3 On Site Calibration: Sample with Analyte Spike (Step 1 of 2) 3.4 On Site Calibration: Sample Test (Step 2 of 2) 3.5 Applying Results 4 Testing a Sample Required Materials 4.2 Testing a Sample 4.3 Results 5 Technical Specifications for ANDalyze Sensors Detection in Drinking Water 5.2 Interference 5.3 Product Accuracy Ranges 6 Consumables and Replacement Items 12 7 Customer Service Contact Information..13 Page 2 of 13
3 2 General Information Important: Before continuing, please read the entire AND1000 User Manual. Pay attention to all danger, warning and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment. Make sure that the safety features of this equipment are not impaired; do not use or install this equipment in a manner other than that specified in this manual. If the equipment is used in a manner not specified by the manufacturer, the safety features of this equipment may be impaired and injury to the operator or damage to the equipment may result. 2.1 Water Testing Guidelines Note: The below steps are for testing for dissolved, bio available metal ions in drinking water samples. These steps can also be used for other matrices such as surface, ground, industrial and wastewater which have been pre treated. Contact ANDalyze customer service for additional application notes on pre treatment methods. Water Sampling For best results use freshly collected sample (unpreserved in acid) for analysis. We recommend that you use the sample within 1 hour (maximum of 2 hours) of collection to minimize any metal loss to the walls of the sample container. This is particularly important for testing trace lead levels (Contact ANDalyze customer service for the citation to a peer reviewed publication about loss of lead to container walls). Large volumes (1L) may be stored up to 12 hours in HDPE containers in a refrigerator or cooler with ice packs if required. Once the sample is mixed with ANDalyze sample buffer, test within 15 minutes. Temperature Range ANDalyze test kits typically work with water samples at C (63 95 F). However, the most accurate and precise results are obtained when the sample is in the range of C (68 77 F). A change in sample temperature of several degrees will require an on site calibration to be performed. ph Range The ANDalyze sample buffer that is provided in the sample tubes brings the ph of the test solution to ph 7.0. Generally, the raw sample water can be in the range of ph 5 ph 8. If you have a sample which is acidic or basic, please check the ph first to confirm it is within the range prior to mixing with the ANDalyze buffer. The final buffered test solution should be ph 7.0 for best results. Note: Our tests have shown that environmental samples preserved in acid to a ph < 2 cannot be brought to a ph of 7.0 when mixed with the ANDalyze buffer. These samples have to be first neutralized with NaOH to a ph 5 before mixing with ANDalyze buffer. Contact ANDalyze customer service for additional application notes on pre treatment methods. Page 3 of 13
4 2.2 Sensor Bag: Cuvette and Sensor Sensor Bag: Each sensor bag contains a cuvette, a sensor and a desiccant. These are single use and must be discarded. The desiccant should be blue in color. If it has turned completely pink in color, the sensor may not perform well. 2.3 Inserting Cuvette and Sensor Cuvette: The cuvette has an arrow which should face you when inserted. Insert the cuvette completely so that the fluorimeter lid can close. The meter and cuvette design helps to prevent improper orientation. Sensor: The square portion of the sensor can be placed on the cuvette in any orientation with the round sections facing upward. Sensors can only be used once and should be disposed of immediately after use. 2.4 Sample Injection and Measurement The AND 1000 Fluorimeter should be laid flat on a stable surface during a measurement A buffered solution is prepared in a sample tube as described in the On Site Calibration section (3) and test section (4) and this is used for measurements A new 1 ml syringe should be used to withdraw 1 ml buffered sample water from a sample tube. This syringe can be attached to the top of the housing as shown in the picture. The sample should be injected through the housing into the cuvette at a constant speed of 3 5 seconds. The syringe and housing should be immediately removed and the sample door closed. The START button located just below the screen should be pressed to start any measurement. Important: After each analysis, discard all components used during the analysis including cuvette, sensor housing, sample tube, and syringes to avoid cross contamination. Page 4 of 13
5 2.5 Pipette Use Guidelines 1. New Pipette Tip Attach a new tip by placing the end of the pipette into one of the available tips and pressing down on the pipette body. Note: Tips are disposable and should never be used more than once. Use of tips helps prevent contaminating the pipette. 2. Depress the plunger button on the top to the first stop (see photo at left). DO NOT depress all the way to the pipette body. 3. Immerse the clean tip into the solution to be withdrawn. 4. Release the pressure slowly to withdraw the solution into the tip. Note: Make sure that the pipette tip continues to be immersed in the solution during release so as to not expose the tip point to air. 5. Remove the pipette from the solution. 6. Immerse the tip into the liquid present in the sample tube where the withdrawn solution is to be dispensed. 7. Slowly depress the operating button ALL THE WAY to dispense the liquid contained in the pipette tip. (See photo at right) 8. Remove the pipette and discard the used tip. Note: Dispose of tips immediately after use to prevent possible contamination of the pipette. Page 5 of 13
6 3 On Site Calibration 3.1 Required Materials AND1000 Fluorimeter (2) Sensor Bags with Sensor & Cuvette (2) 1mL Syringe (2) Sample tubes (with buffer) Lead Standard Solution (1.1 ppm) 100 µl Fixed Volume Pipette and Tips (1) Clean sample collection cup (Not Provided) 3.2 Preparing for On Site Calibration When testing water at a new location, the instrument has to be calibrated for accurate readings. Completing the On Site Calibration requires two separate solutions and tests. At the end of the process, the fluorimeter is calibrated for testing water at that site and also provides the test results for the site. A calibration may only be accurate when originally performed, as changes in the sample matrix, temperature, and sensor lot will affect the results. Do not rely on a calibration to be accurate over days as environmental samples can vary greatly. If in doubt, perform On site Calibration Important: On Site Calibration must be performed if testing is being done: At a new location which has not been previously saved. To adjust for significant changes in sampling temperature. When beginning to use a new lot of sensors. 1. Site: Select the site to be tested or enter a new site as described in Section 5 of the AND1000 User Manual 2. Collect Sample Water In a clean sample cup, collect a small volume of water to be tested. 3. Prepare Calibration Solutions Slowly pour the test water into two sample tubes containing the liquid buffer up to the 5 ml mark. If required, use one of the provided disposable plastic pipettes to transfer water accurately, rinsing the pipette in the solution prior to use. Do not exceed the 5 ml mark for accurate results. 5 ml Important: Open tube carefully so that liquid buffer does not spill out. One tube will be used for analyzing a sample spiked with Lead and the other for analyzing an unspiked sample. Page 6 of 13
7 3.3 On Site Calibration: Sample with Lead Spike (Step 1 of 2) 1. Attach a new tip to the 100µL fixed volume pipette. 2. Withdraw 100µL of the Lead Standard Solution from the bottle using the pipette. Transferring Spike Solution to Sample Tube 3. Mix Transfer the 100µL spike into one of the sample tubes filled with buffered sample water. 4. Shake Close cap and shake well. For most accurate results, wait 5 minutes before testing. Metal standard spiked for on site calibration is required to incubate for all environmental water samples. Withdrawing Standard Solution Using Pipette Important: Open tube carefully so that liquid does not spill out. 5. Pouch Open the sensor bag which contains the sensor (colored plastic housing) and plastic cuvette. 6. Cuvette Place the plastic cuvette in the AND1000 Fluorimeter. 7. Sensor Place colored plastic sensor on the cuvette. (Any orientation). Make sure sensor is seated securely on the cuvette. 8. Enter the Site Calibration screen (as seen in photo to the right) which will appear once a new site has been entered (See Section 5 of AND1000 User Manual). Important: Do not press START until the sample is ready. Follow steps below. Note: If the unit is left on for more than a few minutes without any activity, a screen saver (black screen) will be activated; press any button to resume operation (Do not press and hold ON/OFF). The instrument automatically turns off if not used for more than 5 15 minutes (user settable). 9. Test With one of the provided 1 ml syringes, draw 1 ml buffered sample water from the spiked sample tube into the syringe. Attach the syringe to the housing over the cuvette in the instrument. Maintaining a constant speed (over 3 5 seconds), carefully squeeze the sample water through the housing into the cuvette. Quickly remove sensor housing and syringe and close sample chamber door. Press the START button located just below the screen. Remove the cuvette when complete. Page 7 of 13
8 3.4 Site Calibration: Unspiked Sample Test (Step 2 of 2) The Site Calibration screen (as seen in photo to the right) will appear once Step 1 of 2 has been completed for a new site. DO NOT press START until the second sample is ready for injection! 1. With one of the provided 1 ml syringes, draw 1 ml buffered sample water from the second unspiked sample tube into a new syringe. 2. Attach the syringe to the housing over the cuvette in the instrument. 3. Maintaining a constant speed (over 3 5 seconds), carefully squeeze the sample water through the housing into the cuvette. 4. Quickly remove sensor housing and syringe and close sample chamber door. 5. Press the START button located just below the screen. Remove the cuvette when complete. 3.5 Applying Results The results display the site name, a calibration factor associated with that site, and the Lead test results for the water sample used during this process. Note: The results displayed are for the sample water and not for the sample water containing the spike. The results should be treated as an actual test and can be seen in the results menu screen. To save the factor associated with this site, press APPLY. On Site Calibration from this site has been saved. You are now ready to test more samples from this site. Page 8 of 13
9 4 Testing a Sample This section will walk you through the process of testing a water sample for the desired analyte. Please read this entire section before beginning the test as parts of the test will be time sensitive. Be aware and pay attention to all notes. 4.1 Required Materials AND1000 Fluorimeter (1) Sensor Bag with Sensor & Cuvette (1) 1 ml Syringe (1) Sample tube (with buffer) (1) Clean sample collection cup (Not Provided) Note: Before using the AND1000 assure that the fluorimeter is charged. (See Section 4.1 in the AND1000 User Manual for Battery Charging Information) 4.2 Testing a Sample 1. Start Up Initialize the instrument by pressing the ON/OFF button. Instrument will initialize in about 1 2 seconds. 2. Sensor Type Ensure that the fluorimeter is on the correct sensor screen. 3. Site Confirm that the site being tested is correct. Unless the Instructional Screens are turned ON, Do Not press START until the sample is ready! 4. Pouch Open the sensor bag which contains the sensor (colored plastic housing) and plastic cuvette. Note: Unit may go into screen saver mode after a few minutes. Press any button to resume (Do not press and hold ON/OFF). Important: The bag contains a transparent desiccant pouch. This should be blue in color. If desiccant has turned completely pink in color, the sensor may not perform well. 5. Cuvette Place the plastic cuvette in the instrument. 6. Sensor Place colored plastic sensor on the cuvette. (Any orientation). Make sure the sensor is seated securely on the cuvette. Page 9 of 13
10 7. Sample Tube Collect water to be tested in a clean cup (not provided). 8. Slowly pour the test water into the provided sample tube containing liquid buffer up to the 5 ml mark. If required, use one of the provided disposable plastic pipettes to transfer water accurately without exceeding 5 ml volume mark. Close cap tightly and mix well by shaking. 5 ml Important: Open tube carefully so that liquid buffer does not fall out. Do not exceed the 5 ml mark for accurate results. 9. Syringe With one of the provided syringes, draw 1 ml of buffered sample water from the sample tube into syringe. 10. Attach the syringe to the housing over the cuvette in the instrument. (Syringe tip will fit into top of sensor) Note: If the unit is left on for more than a few minutes without any activity, a screen saver (black screen) will be activated; press any button to resume operation (Do not press and hold ON/OFF). The instrument automatically turns off if not used for more than 5 15 minutes (user settable). 11. Maintaining a constant speed (over 3 5 seconds), carefully squeeze the buffered sample water through the housing into the cuvette. 12. Quickly remove sensor housing and syringe and close sample chamber door. 13. Press the START button located just below the screen. 14. Remove cuvette when complete. 4.3 Results The sample will be tested, and results displayed in parts per billion (ppb) or parts per million (ppm) in under a minute. To save the results, press the SAVE button. Note: A reading of Below 2ppb Limit indicates the amount of Lead in the sample is below the minimum detectable levels of the instrument. Page 10 of 13
11 5 Technical Specifications for ANDalyze Sensors 5.1 Detection in Drinking Water ANDalyze s proprietary catalytic DNA sensor uses a DNAzyme reaction that fluoresces in the presence of the target contaminant (lead, uranium, copper, etc). The fluorescence of the reaction is measured using the AND 1000 fluorimeter to determine the concentration of the free analyte ion (Pb 2+, UO 2 2+, Cu 2+, etc.) in solution and is reported in parts per billion (ppb) or parts per million (ppm). Materials Used AND1000 Fluorimeter Lead Sensor Kit Analyte/Metal Standard Solution Performance Sensor Lead dilutions containing 0, 5, 10, 15, 25, 50, 75, 100, 150, in DI water. Five replicates were used for each test at each dilution. 200 ppb Pb 2+ were prepared Limit of Detection (LOD) 1 ppb Pb 2+ Based on 3 sigma method Limit of Quantification (LOQ) 2 ppb Pb+ Based on 10 sigma method Linear Detection Range ppb Pb 2+ Precision Standard: 15 ppb Pb 2+ 95% Confidence Limits: ppb Pb 2+ Coefficient of Variation (CV): ppb Pb 2+ ±15% Note: All specifications are subject to change without notice. 5.2 Interference Interference tests were done with a 30ppb Pb 2+ solution plus the potential interfering ion. The interference tolerance levels represent the concentration above which the lead concentration is changed to ±10%. Data represents an average of at least three replicates. For each interference test, an onsite calibration with the particular water matrix (containing the interfering ion) was performed. Interfering ion Calcium, Ca 2+ Magnesium, Mg 2+ Zinc, Zn 2+ Aluminum, Al 3+ Copper, Cu 2+ Iron, Fe 3+ Cadmium, Cd 2+ Mercury, Hg 2+ Manganese, Mn 2+ + Ammonium, NH 4 2 Carbonate, CO 3 3 Phosphate, PO 4 Chloride, Cl 2 Sulfate, SO 4 Nitrate, NO 3 Interference level 500 ppm 500 ppm 100 ppm 0.1 ppm 0.5 ppm 0.04 ppm 15 ppm 0.03 ppm 150 ppm 2000 ppm 100 ppm 50 ppm 2000 ppm 100 ppm 5000 ppm Page 11 of 13
12 5.3 Product Accuracy Ranges The graphs below depict the average measured Pb 2+ concentration as displayed on the AND1000 fluorimeter (y axis) vs. the known concentration of the lead standards (x axis). Error bars depict the standard deviation from five measurements. Figure 1 shows the linear range of ppb Pb 2+. For higher concentrations of Pb 2+ (tested up to 200 ppb Pb 2+ ), the accuracy decreases (Graph 2). Samples containing higher than 100 ppb Pb 2+ can be diluted 1:1 and re analyzed. Graph 1: Average Pb 2+ Conc. vs. Known Conc. Graph 2: Accuracy vs. Linear Detection Range 6 Consumables and Replacement Items Fluorimeter (Part Number: AND001) o Capable of measuring multiple metals. o Includes: Fluorimeter USB to MINI B Cable 100µL Fixed Volume Pipette and Tips ph Test Strips User Manual Sensor Kit (Part Number: AND010) o Equipment for (25) Tests and/or Calibrations o Kit Includes: (25) Sensor Bags with Sensor & Cuvette (25) Sample Tubes (with buffer) (25) 1 ml Syringes (15) Disposable Transfer Pipettes 8 ml Lead Standard Solution (1.1 ppm Pb 2+ ) Instruction Manuals Material Safety Data Sheets (MSDS) Page 12 of 13
13 7 Customer Service Contact Information Contact us by By Telephone: (Toll Free in US) or By Fax: Business Service Hours: 9:00am to 5:00pm Eastern Standard Time (USA) Business Service Hours: 9:00am to 5:00pm Central Standard Time (USA) Laboratory/Offices Address: ANDalyze Inc S Oak Street, Suite 102 Champaign, IL USA Sales Office Address: ANDalyze Inc. 800 Boylston Street Prudential Tower, Floor 16 Boston, MA USA Website: ANDalyze Inc S Oak Street, Suite 102 Champaign, IL USA [email protected] Telephone: or Page 13 of 13
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required
Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
Quaternary Ammonium Compounds DOC316.53.01128
Quaternary Ammonium Compounds DOC316.53.01128 Direct Binary Complex Method Method 8337 0.2 to 5.0 mg/l as CTAB (cetyl-trimethylammonium bromide) Powder Pillows Scope and application: For cooling tower
Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856
Nickel DOC316.53.01065 Dimethylglyoxime Method Method 10220 0.1 to 6.0 mg/l Ni TNTplus 856 Scope and application: For water and wastewater. Test preparation Instrument-specific information Table 1 shows
Phosphorus, Reactive (Orthophosphate)
Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from
Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy
Determining the Quantity of Iron in a Vitamin Tablet Computer 34 As biochemical research becomes more sophisticated, we are learning more about the role of metallic elements in the human body. For example,
FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows
Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for
LZV585 (B) The 1-cm path is aligned with the arrow on the adapter. 5940506
Chlorine, Total DOC316.53.01029 USEPA DPD Method 1 Method 10070 0.1 to 10.0 mg/l Cl 2 (HR) Powder Pillows Scope and application: For testing higher levels of total chlorine (free and combined) in drinking
Use of Micropipettes
Use of Micropipettes Prior to lab you should understand: The function of micropipettes in the laboratory Basic parts of micropipette What volumes are measured with P, P and P1 micopipettors How to read
Ion Selective Electrodes
Ion Selective Electrodes OAKTON has a large selection of Ion Selective Electrodes () to suit a wide variety of applications. Each electrode has a typical response time of 20 to 30 seconds but will vary
Mouse Insulin ELISA. For the quantitative determination of insulin in mouse serum and plasma
Mouse Insulin ELISA For the quantitative determination of insulin in mouse serum and plasma Please read carefully due to Critical Changes, e.g., Calculation of Results. For Research Use Only. Not For Use
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
The Determination of an Equilibrium Constant
The Determination of an Equilibrium Constant Computer 10 Chemical reactions occur to reach a state of equilibrium. The equilibrium state can be characterized by quantitatively defining its equilibrium
Chlorine, Total. DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials. Test preparation. Instrument-specific information
Chlorine, Total DOC316.53.01028 DPD Method 1 Method 10101 0.09 to 5.00 mg/l Cl 2 Test 'N Tube Vials Scope and application: For testing higher levels of total (free plus combined) chlorine in drinking water,
HS 1003 Part 2 HS 1003 Heavy Metals Test
HS 1003 Heavy Metals Test 1. Purpose This test method is used to analyse the heavy metal content in an aliquot portion of stabilised hot acetic acid extract by Atomic Absorption Spectroscopy (AAS). Note:
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)
Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method) Grace Analytical Lab 536 South Clark Street 10th Floor Chicago, IL 60605 April 15, 1994 Revision 2 Standard Operating Procedure
Determination of calcium by Standardized EDTA Solution
Determination of calcium by Standardized EDTA Solution Introduction The classic method of determining calcium and other suitable cations is titration with a standardized solution of ethylenediaminetetraacetic
DR/4000 PROCEDURE. CHLORINE, Free
DR/4000 PROCEDURE CHLORINE, Free Method 10069 DPD method* UHR (0.1 10.0 mg/l as Cl 2 ) Scope and Application: For testing higher levels of free chlorine (hypochlorous acid and hypochlorite ion) in drinking
Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE
Sodium DOC316.53.01240 Direct ISE Method Method 8359 10 to 1000 mg/l Na + Sodium ISE Scope and application: For drinking water and process water. Test preparation Instrument-specific information This procedure
OXIDATION-REDUCTION TITRATIONS-Permanganometry
Experiment No. Date OXIDATION-REDUCTION TITRATIONS-Permanganometry INTRODUCTION Potassium permanganate, KMnO 4, is probably the most widely used of all volumetric oxidizing agents. It is a powerful oxidant
Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater
Method 8155 NITROGEN, AMMONIA (0 to 0.50 mg/l NH 3 -N) For water, wastewater, seawater Salicylate Method * 1. Enter the stored program number for ammonia nitrogen (NH 3 -N). 2. Press: 64 ENTER The display
Luminol Test PROCESS SKILLS SCIENCE TOPICS VOCABULARY
EXPERIMENT: LUMINOL TEST Luminol Test Visitors mix a solution of luminol with fake blood (hydrogen peroxide) to produce a reaction that gives off blue light. OBJECTIVES: Visitors learn that some chemical
Determination of the Mass Percentage of Copper in a Penny. Introduction
Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny
SCP SCIENCE www.scpscience.com. Certified Reference Materials. Materials
Certified Reference 47 EnviroMAT and AgroMAT Certifi ed Reference (CRMs) are designed to complement existing Performance Evaluation Programs in environmental and agricultural analysis. EnviroMAT CRMs allow
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
A Potentiometric Analysis of Fluoride Ion in Toothpaste
CHEM 311L Quantitative Analysis Laboratory Revision 2.0 A Potentiometric Analysis of Fluoride Ion in Toothpaste In this laboratory exercise, we will analyze a toothpaste sample for its Fluoride Ion (F
Appendix C. Vernier Tutorial
C-1. Vernier Tutorial Introduction: In this lab course, you will collect, analyze and interpret data. The purpose of this tutorial is to teach you how to use the Vernier System to collect and transfer
Compact Water Quality Meter B-700 series
Compact Water Quality Meter B-700 series Actual Size Compact Water Quality Meter Debut! HORIBA s 60 years of sensor engineering realize accurate direct measurement from only a single drop on the unique
Experiment 16-Acids, Bases and ph
Definitions acid-an ionic compound that releases or reacts with water to form hydrogen ion (H + ) in aqueous solution. They taste sour and turn litmus red. Acids react with certain metals such as zinc,
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS Sl. No. Contents Preamble 5.1 Aim 5.2 Introduction 5.2.1 Environmental Significance 5.3 Principle 5.4 Materials Required 5.4.1 Apparatus Required 5.4.2
USEPA Electrode Method Method 8156
ph DOC316.53.01245 USEPA Electrode Method Method 8156 Scope and application: For drinking water 1, wastewater 2 and process water. 1 Based on Standard Method 4500-H+B, ASTM Method D1293-95 and USEPA Method
Decomposition. Composition
Decomposition 1. Solid ammonium carbonate is heated. 2. Solid calcium carbonate is heated. 3. Solid calcium sulfite is heated in a vacuum. Composition 1. Barium oxide is added to distilled water. 2. Phosphorus
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014
SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, [email protected] 2014 This experiment explores quality assurance practices which are commonly
Scope and application: For testing residual corrosion inhibitors (oxygen scavengers) in boiler feed water or condensate
Oxygen Scavengers DOC316.53.01105 Iron Reduction Method Method 8140 3 to 450 µg/l DEHA; 5 to 600 µg/l carbohydrazide; 9 to 1000 µg/l hydroquinone; 13 to 1500 µg/l iso-ascorbic acid (ISA); 15 to 1000 µg/l
EXPERIMENT 8: Activity Series (Single Displacement Reactions)
EPERIMENT 8: Activity Series (Single Displacement Reactions) PURPOSE a) Reactions of metals with acids and salt solutions b) Determine the activity of metals c) Write a balanced molecular equation, complete
Complexometric Titrations
Complexometric Titrations Complexometric titrations are based on the formation of a soluble complex upon the reaction of the species titrated with the titrant. M + L ML In the module Chemistry 2 you were
Experiment 18: ph Measurements of Common Substances. Experiment 17: Reactions of Acids with Common Substances
Experiment 18: ph Measurements of Common Substances and Experiment 17: Reactions of Acids with Common Substances What is this lab about? You mean what ARE THESE labs about? Ok, so what are THESE labs about?
Appendix D lists the Field Services Standard Operating Procedures. Appendix E lists the Biological Monitoring Standard Operating Procedures.
Page 16 of 87 3.3 Sample Collection, Storage and Preservation Figure 3 details required containers, sample volumes, preservation techniques, and holding times for proper sample collection. A discussion
EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline
EXPERIMENT 5 Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline UNKNOWN Submit a clean, labeled 100-mL volumetric flask to the instructor so that your unknown iron solution
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
Experiment 8 - Double Displacement Reactions
Experiment 8 - Double Displacement Reactions A double displacement reaction involves two ionic compounds that are dissolved in water. In a double displacement reaction, it appears as though the ions are
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method
Care and use guide. Contents. a handy solution made easy. www.getbluelab.com. page
Care and use guide Contents page Features Quick guide Before use To operate Important - ph pen probe care 4 Cleaning 5 Battery replacement 5 Hydration 6 Calibration 6 Error messages 7 Troubleshooting guide
Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit
Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Catalog No: E0479r 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater
Agilent High Sensitivity DNA Kit Guide
Agilent High Sensitivity DNA Kit Guide Agilent High Sensitivity DNA Agilent Technologies Notices Agilent Technologies, Inc. 2009, 2013 No part of this manual may be reproduced in any form or by any means
Measuring ph in Water or CaCl2 Using a ph Meter. Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3
Measuring ph in Water or CaCl2 Using a ph Meter Compiled by Darren Murray, June 30, 2011 TABLE OF CONTENTS OVERVIEW 3 PREPARING THE ph METER FOR USE (Calibration and Buffer Solutions) 4 When is Calibration
Figure 1. A voltaic cell Cu,Cu 2+ Ag +, Ag. gas is, by convention, assigned a reduction potential of 0.00 V.
Voltaic Cells Introduction In this lab you will first prepare a set of simple standard half-cells and then measure the voltage between the half-cells with a voltmeter. From this data you will be able to
Experiment 17: Potentiometric Titration
1 Experiment 17: Potentiometric Titration Objective: In this experiment, you will use a ph meter to follow the course of acid-base titrations. From the resulting titration curves, you will determine the
Chloride and Salinity
INTRODUCTION Chloride Chloride and Chloride, in the form of the Cl ion, is one of the major inorganic anions, or negative ions, in saltwater and freshwater. It originates from the dissociation of salts,
Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA
KAMIYA BIOMEDICAL COMPANY Mouse Creatine Kinase MB isoenzyme (CKMB) ELISA For the quantitative determination of mouse CKMB in serum, plasma, cell culture fluid and other biological fluids Cat. No. KT-57681
Calibration of Volumetric Glassware
CHEM 311L Quantitative Analysis Laboratory Revision 2.3 Calibration of Volumetric Glassware In this laboratory exercise, we will calibrate the three types of glassware typically used by an analytical chemist;
Santa Monica College Chemistry 11
Types of Reactions Objectives The objectives of this laboratory are as follows: To perform and observe the results of a variety of chemical reactions. To become familiar with the observable signs of chemical
Experiment 5. Chemical Reactions A + X AX AX A + X A + BX AX + B AZ + BX AX + BZ
Experiment 5 Chemical Reactions OBJECTIVES 1. To observe the various criteria that are used to indicate that a chemical reaction has occurred. 2. To convert word equations into balanced inorganic chemical
Acid Base Titrations
Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually
IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.
IgM ELISA For the quantitative determination of IgM in human serum and plasma For Research Use Only. Not For Use In Diagnostic Procedures. Please read carefully due to Critical Changes, e.g., Calibrator
Determining the Free Chlorine Content of Swimming Pool Water. HOCl H + + OCl. Evaluation copy
Determining the Free Chlorine Content of Swimming Pool Water Computer 33 Physicians in the nineteenth century used chlorine water as a disinfectant. Upon the discovery that certain diseases were transmitted
Relative Humidity Calibration Kit
Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration
Solutions and Dilutions
Learning Objectives Students should be able to: Content Design a procedure for making a particular solution and assess the advantages of different approaches. Choose the appropriate glassware to ensure
USING TECHNOLOGY FOR DATA COLLECTION
USING TECHNOLOGY FOR DATA COLLECTION The advances in technology have provided a number of instruments for collecting and interpreting data. These include portable microscopes, electronic balances, ph meters,
Inc. Wuhan. Quantity Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4 Standard (liquid) 2
Uscn Life Science Inc. Wuhan Website: www.uscnk.com Phone: +86 27 84259552 Fax: +86 27 84259551 E-mail: [email protected] ELISA Kit for Human Prostaglandin E1(PG-E1) Instruction manual Cat. No.: E0904Hu
Syllabus OC18 Use litmus or a universal indicator to test a variety of solutions, and classify these as acidic, basic or neutral
Chemistry: 9. Acids and Bases Please remember to photocopy 4 pages onto one sheet by going A3 A4 and using back to back on the photocopier Syllabus OC18 Use litmus or a universal indicator to test a variety
Neutralizing an Acid and a Base
Balancing Act Teacher Information Objectives In this activity, students neutralize a base with an acid. Students determine the point of neutralization of an acid mixed with a base while they: Recognize
Dot Blot Analysis. Teacher s Guidebook. (Cat. # BE 502) think proteins! think G-Biosciences www.gbiosciences.com
PR110 G-Biosciences 1-800-628-7730 1-314-991-6034 [email protected] A Geno Technology, Inc. (USA) brand name Dot Blot Analysis Teacher s Guidebook (Cat. # BE 502) think proteins! think G-Biosciences
General Chemistry Lab Experiment 6 Types of Chemical Reaction
General Chemistry Lab Experiment 6 Types of Chemical Reaction Introduction Most ordinary chemical reactions can be classified as one of five basic types. The first type of reaction occurs when two or more
Human Free Testosterone(F-TESTO) ELISA Kit
Human Free Testosterone(F-TESTO) ELISA Kit Catalog Number. MBS700040 For the quantitative determination of human free testosterone(f-testo) concentrations in serum, plasma. This package insert must be
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit
Rat creatine kinase MM isoenzyme (CK-MM) ELISA Kit Catalog Number. CSB-E14405r For the quantitative determination of rat creatine kinase MM isoenzyme (CK-MM) concentrations in serum, plasma and tissue
Analytical Chemistry Lab Reports
Analytical Chemistry Lab Reports Format and Calculations John Collins [email protected] Measurement Analytical chemistry is entirely about measurement, what these measurements signify, and the understanding
ALS TRIBOLOGY Engine Coolant Reference Guide
ALS TRIBOLOGY Engine Coolant Reference Guide Engine coolants are a mixture of, inhibitors, and water. Each formula is designed for specific protection and engine requirements. Mixing different coolants
ph Measurements of Common Substances
Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the
KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)
Selenium Determination, Page 1 KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY I. BACKGROUND. (VERSION 1.8) The majority of reactions used in analytical chemistry possess the following characteristics:
4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES
4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES Sl. No. Contents Preamble 4.1 Aim 4.2 Introduction 4.2.1 Environmental Significance 4.3 Principle 4.4 Materials Required 4.4.1 Apparatus Required 4.4.2 Chemicals
Chapter 16: Tests for ions and gases
The position of hydrogen in the reactivity series Hydrogen, although not a metal, is included in the reactivity series because it, like metals, can be displaced from aqueous solution, only this time the
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
18 Conductometric Titration
Lab Activity 18 CONDUCTOMETRIC TITRATION LAB ACTIVITY 18 Conductometric Titration Background Titration is the a method of determining the concentration of an unknown solution (the analyte) by reacting
A Visual Method for the Detection of Arsenic 0 to 500 µg/l By Dan Kroll R&D Chemist. Hach Co., Loveland Colorado
Introduction: Arsenic is a common contaminant of ground water that has been found to adversely effect human health at levels as low as 10 µg/l. The current USEPA maximum contaminant level (MCL) for arsenic
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
Standard methods in water analysis
Branch General analytical laboratories; water analysis Keywords Water analysis; standard methods; ASTM; DIN; ISO; USP; EPA; SLMB; EN; SCA; titration; ion chromatography; voltammetry; branch 1; branch 2
Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist
Application of ICP-MS as a Tool for Environmental Monitoring Bill Spence Environmental Market Specialist Introduction Knowledge of the concentration of metallic species in the environment has been important
ENZYME KINETICS ENZYME-SUBSTRATE PRODUCTS
ENZYME KINETICS INTRODUCTION The study of reaction rates catalyzed by enzymes and the factors affecting them is generally referred to as enzyme kinetics. The basic components of an enzyme catalyzed reaction
6 H2O + 6 CO 2 (g) + energy
AEROBIC RESPIRATION LAB DO 2.CALC From Biology with Calculators, Vernier Software & Technology, 2000. INTRODUCTION Aerobic cellular respiration is the process of converting the chemical energy of organic
ELECTROCHEMICAL CELLS LAB
ELECTROCHEMICAL CELLS LAB Purpose: The purpose of this lab is to demonstrate the ability of chemistry to make electric current using oxidation/reduction (REDOX) reactions, and to measure the electric current
Mouse IgM ELISA. Cat. No. KT-407 K-ASSAY. For the quantitative determination of IgM in mouse biological samples. For Research Use Only. 1 Rev.
K-ASSAY Mouse IgM ELISA For the quantitative determination of IgM in mouse biological samples Cat. No. KT-407 For Research Use Only. 1 Rev. 072309 K-ASSAY PRODUCT INFORMATION Mouse IgM ELISA Cat. No. KT-407
HOMEWORK 4A. Definitions. Oxidation-Reduction Reactions. Questions
HOMEWORK 4A Oxidation-Reduction Reactions 1. Indicate whether a reaction will occur or not in each of following. Wtiring a balcnced equation is not necessary. (a) Magnesium metal is added to hydrochloric
Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution
Volatile Acids DOC316.53.01144 Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution Scope and application: For digestor sludges. 1 Adapted from The Analyst, 87, 949 (1962).
2C: One in a Million. Part 1: Making solutions. Name: Section: Date: Materials
Name: Section: Date: 2C: One in a Million Drinking water can contain up to 1.3 parts per million (ppm) of copper and still be considered safe. What does parts per million mean? Both living things and the
ATOMS. Multiple Choice Questions
Chapter 3 ATOMS AND MOLECULES Multiple Choice Questions 1. Which of the following correctly represents 360 g of water? (i) 2 moles of H 2 0 (ii) 20 moles of water (iii) 6.022 10 23 molecules of water (iv)
TITRATION CURVES, INDICATORS, AND ACID DISSOCIATION CONSTANTS
TITRATION CURVES, INDICATORS, AND ACID DISSOCIATION CONSTANTS Adapted from "Chemistry with Computers" Vernier Software, Portland OR, 1997 INTRODUCTION Titration is the volumetric measurement of a solution
Periodic Table, Valency and Formula
Periodic Table, Valency and Formula Origins of the Periodic Table Mendelѐѐv in 1869 proposed that a relationship existed between the chemical properties of elements and their atomic masses. He noticed
Nomenclature and Formulas of Ionic Compounds. Section I: Writing the Name from the Formula
Purpose: Theory: Nomenclature and Formulas of Ionic Compounds 1. To become familiar with the rules of chemical nomenclature, based on the classification of compounds. 2. To write the proper name of the
Household Acids and Bases
Household Acids and Bases GRADE LEVEL INDICATORS Experiment Demonstrate that the ph scale (0-14) is used to measure acidity and classify substances or solutions as acidic, basic, or neutral. 21 Develop
Spectrophotometer - Milton Roy Spectronic 21D or equivalent.
3M COMPANY OCCUPATIONAL HEALTH AND ENVIRONMENTAL SAFETY DIVISION DETERMINATION OF FORMALDEHYDE VAPORS IN AIR USING 3M 3721 FORMALDEHYDE MONITORS May, 2002 SCOPE This procedure covers the method of collecting
Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution.
Chem 405 Biochemistry Lab I Experiment 2 Quantitation of an unknown protein solution. Introduction: The determination of protein concentration is frequently required in biochemical work. Several methods
Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*
Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU
Immunoglobulin E (IgE) concentrations in Human. Immunoglobulin E (IgE) Human ELISA Kit
ab108650 Immunoglobulin E (IgE) Human ELISA Kit Instructions for Use For the quantitative measurement of Immunoglobulin E (IgE) concentrations in Human serum. This product is for research use only and
Bluelab Soil ph Meter Instruction Manual
Instruction Manual www.getbluelab.com Table of Contents 1.0 Information about measuring the ph of soils/media. 2 2.0 Introduction to Bluelab Soil ph Meter 4 3.0 Preparing the Bluelab Soil ph Meter for
Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
Total Suspended Solids Total Dissolved Solids Hardness
Total Suspended Solids (TSS) are solids in water that can be trapped by a filter. TSS can include a wide variety of material, such as silt, decaying plant and animal matter, industrial wastes, and sewage.