1 m a s s s p e c t r o m e t r y Finnigan ELEMENT2 High Performance High Resolution ICP-MS Analyze Detect Measure Control
2 Finnigan ELEMENT2 High Performance High Resolution ICP-MS Finnigan ELEMENT2 High Performance High Resolution ICP-MS An Instrument for every Application A success story around the world with over 3 instruments installed, for example: Belgium: OCAS. Research centre for the application of steel with New Wave laser ablation system. Sweden: Analytica AB. Scandinavia s leading contract lab with Emma Engström. Canada: Seastar Chemicals Inc. A manufacturer of high purity chemicals with Dr. Brad McKelvey. USA: Micron Technology Inc. Advanced semiconductor solutions with Kevin Coyle. Germany: Institute for Transuranium Elements. Shown with a glovebox for radioactive sample measurements. USA: Desert Research Institute. Hydrologic Sciences Division with Steve Lambert.
3 Multielement analysis across the periodic table covering a mg/l to sub pg/l concentration range Compatible with inorganic and organic solution matrices and solids High mass resolution to access spectrally interfered isotopes Produces unambiguous elemental spectra A multielemental detector for transient signals For example, CE, HPLC, GC, FFF and laser ablation High precision isotope ratios Independent of interferences or interfered isotopes Fully automated tuning and analysis In conjunction with a comprehensive, customizable quality control system Reliability and robustness to serve as a 24 / 7 production control tool Highest sample throughput Highest flexibility and accessibility to serve as an advanced research tool
4 Finnigan ELEMENT2 High Performance High Resolution ICP-MS High Mass Resolution The unequivocal separation of analyte ions from spectral interferences is a prerequisite of accurate and precise analysis. High mass resolution is the universal means for this separation. Spectral interferences are the main limitation of ICP-MS. The argon plasma gas, water, acid and the sample matrix itself can combine to introduce a wide range of polyatomic ion species. The resultant interfering species may have the same nominal mass as an analyte ion and thus return a falsely high value for the analyte. Numerous strategies have been used in order to try to minimize or circumvent the formation of these spectral interferences. These include mathematical corrections, special sample introduction systems, special plasma parameters and collision/dynamic reaction cells to neutralize part of the interferences. High resolution with a sector-field mass spectrometer simply distinguishes the analyte from interference by difference in mass. The capability of high mass resolution is a feature unique to the Finnigan ELEMENT2. This ability can be used for quantification and isotope ratio analysis for nearly the whole periodic table, and in almost all matrices. Even in a sample matrix as simple as ultra pure water (UPW), interferences do exist especially at low analyte concentration levels: Phosphorus in UPW, Medium Resolution Zoom in 28 Si 31 P 12 C 16 O 15 N 16 O 14 N 16 OH N Silicon in UPW, Medium Resolution 12 C 16 O2 28 Si 16 O 44 Ca Calcium in UPW, Medium Resolution
5 32 S 32 S Zoom in 64 Zn Zinc in H 2 SO 4 (1 % w/w), High Resolution 32 S 16 O The more complex the sample matrix, the wider the range of interferences that will occur. The main advantage of high mass resolution as a technique to remove spectral interferences is that it is not just limited to a particular type of interference. With high mass resolution, iron at m/z 56 is easily separated in a simple matrix from 4 Ar 16 O as well as in more complex matrices from, for example, 4 Ca 16 O. Arsenic at m/z 75 can be determined in HCl, separated from the 4 Ar 35 Cl interference, or, in a matrix containing Ca and chloride, from 4 Ca 35 Cl. Matrices such as mineral acids and organic solvents, which form a plethora of spectral interferences, can be easily analyzed using the high resolution mode. High resolution results in simple clear spectra and does not create new interferences. 4 Ar 16 O 12 C2 16 O2 56 Fe Iron in Liquid Crystal Matrix, Cold Plasma, Medium Resolution 44 Ca 16 O 44 Ca 16 OH Ni 43 Ca 16 OH 32 S 16 O 12 C Nickel Isotopes in Groundwater, Medium Resolution 61 Ni 62 Ni 46 Ca 16 O Resolution Specification (1 % peak valley definition) 3 fixed resolutions: Low Resolution > 3 Medium Resolution > 4 High Resolution > 1
6 Finnigan ELEMENT2 High Performance High Resolution ICP-MS Principle The Finnigan ELEMENT2 is a double focusing magnetic sector field ICP-MS. Plasma and Interface The argon plasma ion source and sampling interface of the Finnigan ELEMENT2 are at ground potential 1. This enables the straightforward coupling of peripherals like HPLC, CE, GC and laser ablation. The Finnigan ELEMENT2 interface reduces the initial kinetic energy spread from ~2 to ~5 ev by capacitively decoupling the plasma from the load coil, using a grounded guard electrode. This reduced energy spread increases the ion transmission and delivers superior sensitivities at all resolutions. The ion transfer optics focus the ions from the plasma interface on the entrance slit of the double focusing analyzer. The Finnigan ELEMENT2 ion transfer optics are designed for low background, highest sensitivity and minimum mass bias at maximum stability 2. Highly laminated, water cooled magnet, m/z: in < 15 ms 25 ppm/8 hour mass stability High Resolution The Finnigan ELEMENT2 is able to fully automatically change between three fixed resolutions by switching the positions of the entrance and exit slits in < 1 s. The patented design of the fixed slit mechanism 3 offers maximum stability and reproducibility of resolutions. MR LR HR 1 Patent issued: US , GB Patent issued: US Patent issued: US , GB fixed resolutions: R = 3 R = 4 R = 1 Ion Transfer Optics Guarantees a flat response curve, low background and high sensitivity
7 Simultaneous measurement in analog and counting modes > 1 9 linear dynamic range <.2 cps dark noise Fully automatic cross calibration Interface at ground potential Easily changeable cones Resistant against chemicals Mass Separation The magnetic field disperses ions according to their mass and energy. The magnet used in the Finnigan ELEMENT2 is specifically designed for use in ICP-MS applications. It is relatively small (sufficient for the mass range 26 u), highly laminated and efficiently water-cooled for the highest mass stability. Changing of the magnetic field is controlled by a magnetic field regulator with a new high power stage, which delivers the fastest scan speed ever possible with a magnetic sector field instrument 4. After passing through the magnetic field the ions enter the electrostatic analyzer for energy focusing. The combination of the magnetic and electrostatic fields results in the double focusing, high resolution properties of the Finnigan ELEMENT2. Detection System The Finnigan ELEMENT2 is equipped with a discrete dynode detector system. Rather than having the ion beam directly strike the detector to initiate an electron cascade, the secondary electron multiplier implemented in the Finnigan ELEMENT2 uses a conversion dynode at -8 kv, producing a uniform response across the mass range. The detector is linear over nine orders of magnitude from ppq to ppm concentrations. The quantification of trace and major elements is therefore possible in a single analysis. 4 Thermo Electron Application Note AN311_E12/3C Torch with Guard Electrode (GE) The GE decreases the ion energy spread, thus increasing ion transmission. This and the high acceleration voltage used in a sector field ICP-MS results in increased sensitivity. The GE is also required for Cold Plasma measurements
8 Finnigan ELEMENT2 High Performance High Resolution ICP-MS Resolution Low Resolution Low resolution (R = 3) is used for the analysis of non-interfered isotopes. In this mode the Finnigan ELEMENT2 s sensitivity is the highest of all commercially available ICP-MS instruments. Additionally, the flat top peak shape is an advantage for high precision isotope ratio measurements. Medium Resolution Medium resolution (R = 4) guarantees interference-free analysis for most elements in the majority of sample matrices. For example, transition elements are routinely measured in medium resolution due to the formation of many interfering polyatomic species in the mass range 24 7 u. High Resolution High resolution (R = 1) is used for the analysis of elements in the most challenging sample matrices. For example, high resolution is used to separate As and Se from argon dimer interferences and argon chloride interferences in chlorine matrices, heavy rare earth elements from light rare earth element oxides in geological matrices, platinum group elements from argon-transition metal molecular species, and/or the oxides of Hf, Ta and W. Intensity [cps] ng/l 115 In, Low Resolution Intensity [cps] ng/l 115 In, Medium Resolution 6 5 Intensity [cps] ng/l 115 In, High Resolution
9 Intensity [cps] Hf 16 O Since the change in mass resolution is achieved by changing the width of the entrance and exit slits of the mass spectrometer, the instrumental sensitivity of a high resolution ICP-MS is dependent on the resolution mode used. Therefore, the Finnigan ELEMENT2 with three fixed resolutions has three sensitivities: the wider the slit, the higher the sensitivity Pt Fixed sensitivity ratio between resolutions: independent of mass and matrix µg/l Hafnium, 1 ng/l Platinum, High Resolution Even in high resolution mode, the intrinsic sensitivity of the Finnigan ELEMENT2 provides sub ppt detection limits Ar 12 C Intensity [cps] Cr 37 Cl 16 O Cr Chromium in Blood, Medium Resolution Eu 16 O Intensity [cps] Tm Thulium in Europium Matrix, High Resolution Sensitivity Specification Low Resolution (R = 3) 115 In > 1x1 6 cps/ppb Medium Resolution (R = 4) 115 In > 1x1 5 cps/ppb High Resolution (R = 1) 115 In > 1.5x1 4 cps/ppb Background in all 3 resolutions <.2 cps
10 Finnigan ELEMENT2 High Performance High Resolution ICP-MS Sensitivity and Stability Sensitivity and Low Background It is obvious that the highest instrumental sensitivity is essential to achieve the lowest detection limits. However, it is the signal to noise ratio that dictates the detection limit. The Finnigan ELEMENT2 guarantees an off-peak background of <.2 cps for all three resolutions. Detection limits in the fg/l range are possible. Detection Limits in Complex Matrices Even in cases where the lowest detection limits are not the goal, the exceptionally high sensitivity of the Finnigan ELEMENT2 offers an important advantage. With the higher sensitivity, higher dilution factors for complex matrices can be used without sacrificing detection limits. Intensity [cps] bkg<.2 cps 226 Ra bkg<.2 cps Mineral Water, undiluted (5 % HNO 3 ), ~2 fg/l Radium, Low Resolution Sensitivity and Dilute and Shoot The requirement for matrix separation for samples such as seawater, mineral acids or organic solvents is replaced by a simple dilution. Using this dilute and shoot approach, the matrix load on the sample introduction system, plasma and interface is reduced. This is also particularly important for the analysis of nuclear sample matrices where the minimization of waste is of paramount importance. 23 Na 9 Zr 17 Ag 151 Eu 47 Ti 52 Cr 63 Cu 66 Zn 79 Br 75 As 155 Gd Resolution Low Low Low Low Medium Medium Medium Medium Medium High High LoD [ng/l] in solution Detection Limits in 5 mg/l Uranium 6 Normalized Intensity [ppq] Concentration [ppq] 226 Ra, External Calibration, 1, 4, 6 pg/l, R 2 = 1.
11 Intensity [cps] ng/l As Concentration [ppt] Signal Stability and Detection Limits Sector field ICP-MS offers superior ion transmission stability due to the high acceleration voltage (-8 kv) and excellent focusing properties. In combination with the high sensitivity, low background, interference-free measurements and an advanced sample introduction system, the Finnigan ELEMENT2 delivers the lowest detection limits independent of the sample matrix. The resulting low detection limits combined with highest stability at single digit ppt level enable quantification at the lowest concentrations. Standard Addition Analysis of Arsenic in HCl (1 % v/v) Recovery [%] Cu Zr Ag Ba Tl Cr Mn Li Na Fe element 1 ng/l Spike Recovery UPW concentration [ppt] Sn12(LR) 2.5% Fe56(MR) 2.2% As75(HR) 4.1% Signal Stability 2 Hours Low, Medium and High Resolution: 5 ng/l Stability Specifications < 1 % RSD in 1 minutes < 2 % RSD in 1 hour analysis #
12 Finnigan ELEMENT2 High Performance High Resolution ICP-MS Mass Stability The mature design of the Finnigan ELEMENT2 delivers the fastest scan speed ever for a magnetic sector field ICP-MS. With the Finnigan ELEMENT2 a jump from 7 to 24 to 7 is realized in < 15 ms. Scanning Routine scanning is performed by a combination of magnetic and electric jumps. The Finnigan ELEMENT2 has the unique capability to scan +3% from the magnet mass by decreasing the acceleration voltage. The magnetic field is kept constant while the acceleration voltage is varied. This ingenious combination of scan techniques delivers the fastest scanning ever realized by a sector field ICP-MS. Coil Current (A) m/z ms 5 ms 5 ms m/z ms m/z ms m/z ms 5 ms 5 ms 5 ms Time (ms) Timing for a Full Magnet Cycle Mass (m/z) Mass Stability The Finnigan ELEMENT2 guarantees the highest mass stability of any ICP-MS. This enables fast peak top jumping analyses rather than scanning across the whole peak, thus significantly decreasing analysis times. A combination of mature hardware and intelligent software guarantees mass stability in high resolution, which makes mass calibration a rare event. The synthesis of these characteristics combined with high sensitivity, allowing short integration times, makes the Finnigan ELEMENT2 the fastest sector field ICP-MS ever. Centroid Mass Analysis y-scale is equivalent to peak width in high resolutiom The scan speed of the Finnigan ELEMENT2 in combination with with the high sensitivity opens the dimension of its use as a detector for fast transient signals. 12-Hour Mass Stability for 78 Se in high resolution Specifications Mass stability: 25 ppm / 8 hours Magnetic scan speed: m/z 7 to 24 to 7 < 15 ms
13 Ion Detection System analog signal The goal of real life analysis is the simultaneous determination of major and trace elements. Conversion Dynode Electrons e - Ions m + analog counting SEM counting signal The Finnigan ELEMENT2 is equipped with a discrete dynode detection system that enables the quantification of both trace and major elements in a single analysis across a dynamic range of 1 9. The Finnigan ELEMENT2 detector system incorporates a conversion dynode at -8 kv. The high acceleration to -8 kv results in a mass independent detector response, enabling fully automatic cross calibration between the counting and analog modes. No user interaction is necessary. Finnigan ELEMENT2 Ion Detection System 2 1 The automatic cross calibration ensures a constantly updated detector response at all times ready for the unexpected sample! Additionally the detector enables the measurement of large isotope ratios. With the minor isotope in counting mode and the abundant isotope in analog mode the highest precision and accuracy can be obtained. 8 Intensity [cps] ng/L 25 TI Analog: red line Counting: blue line 6 4 Rel. Intensity [%] Simultaneous Measurement of the Counting and Analog Signals Isotopes of Uranium Determination of large Isotope ratios using the both mode blue signals = counting, red signals = analog Specifications Dynamic Range > 1 9 with automatic gain calibration
14 Finnigan ELEMENT2 High Performance High Resolution ICP-MS The Software Suite The Finnigan ELEMENT2 Software The Finnigan ELEMENT2 software controls and monitors all instrument functions for ICP-MS analysis. This includes data acquisition and auto-tuning of the Finnigan ELEMENT2. The software also provides the full range of quantification procedures required in elemental analysis (qualitative, quantitative, semi-quantitative, isotope dilution) as well as isotope ratio and time resolved analysis modes. Controlling and Tuning Autotuning of all parameters, including ICP parameters, torch position, lenses and multiplier voltage Fully automated and configurable plasma start and stop sequence Easy autosampler setup using a graphical display of the autosampler Setup of Methods Intuitive and easy selection of target isotopes in periodic table or spreadsheet display modes Automatic, customized isobaric interference correction Spreadsheet style click-and-drag cell fill down Creating and Running Sequences Intuitive and easy creation of sample analysis sequences in graphical or spreadsheet display modes Integrated powerful QA/QC package that meets internationally regulated requirements including US EPA 2.8, 62. The flexible editor included can be used to define specific QA/QC criteria in any laboratory. Remote Control & Diagnostic Displaying Results and Creation Displaying Results and Creation of Reports Real time display of spectra, calibration curves, fully quantitative results and time resolved analyses On-line export of time resolved data in several formats (ASCII, GRAMS, Spectacle, GLITTER, ANDI Xcalibur for further analysis in third party programs Setting & Read-Back of All Instrument Parameters Instrument Startup and Tune The Finnigan ELEMENT2 software is a state-of-the-art, simple to use software suite taking advantage of the reliability and stability of the modern operating system Microsoft Windows XP Professional. The software package is optimized for the needs of the routine analyst, providing stability and ease of use for basic operation of the Finnigan ELEMENT2, and yet retains the flexibility for advanced operation. Due to the use of the Microsoft Windows XP Professional operating system and standardized programming, the data system computer can be easily connected to a network, enabling data transfer and allowing remote control of the Finnigan ELEMENT2.
15 Product Specification Specifications and Installation Requirements Sensitivity (Concentric Nebulizer) > 1 x 1 9 counts per second (cps)/ppm In Detection Power Dark Noise Dynamic Range Mass Resolution Signal Stability Scan Speed (magnetic) Scan Speed (electric) < 1 ppq for non-interfered nuclides <.2 cps > 1 9 linear with automatic gain calibration 3, 4, 1, (1 % valley, equivalent to 5 % height); 6, 8, 2, (FWHM) < 1 % RSD over 1 minutes < 2 % RSD over 1 hour m/z 7 to 24 to 7 < 15 ms 1 ms/jump, independent of mass range Oxide and Doubly Charged Ions ratio measured BaO + /Ba + <.2 Ba 2+ / Ba + <.3 Power 3-phase, 23/4 V ±1 %, 5/6 Hz fused 32 A per phase Power consumption: ~9 kva Environment Temperature C (64 75 F) Humidity 5 6 %, noncondensing, non-corrosive Cooling Water ~ 2 l/h Temperature 1 2 C 4 6 bar (43 65 psi) Argon Purity min. 18 L/min Regulated pressure 8 1 bar ( psi) Uninterrupted argon supply recommended Plasma Exhaust 1 x 6 cm Ø; 9 m 3 /h (Argon + suspended sample) Electronics Exhaust 2 x 15 cm Ø; 8 m 3 /h Finnigan ELEMENT2, Footprint and Dimensions (all dimensions in cm)
16 Plasma Capabilities from Thermo Electron In addition to these offices, Thermo Electron Corporation maintains a network of representative organizations throughout the world. The use of an Inductively Coupled Plasma source (ICP) is the accepted and most powerful technique for the analysis and quantification of trace elements in both solid and liquid samples. Its applications range from routine environmental analyses to the materials industry, geological applications to clinical research and from the food industry to the semiconductor industry. Thermo Electron Corporation is the only instrument manufacturer to offer the full range of Inductively Coupled Plasma Spectrometers (ICP-OES, Quadrupole and Sector ICP-MS) to satisfy every aspect of plasma spectrometry from routine to highly demanding research applications. IRIS Intrepid II Develop your lab from the easy-to-use IRIS Intrepid II ICP-OES to the high performance X Series Quadrupole ICP-MS and up to the ultra-sophisticated Finnigan ELEMENT2 and Finnigan NEPTUNE Sector ICP-MS instruments. Each instrument combines leading-edge technology, fit for purpose and affordability with a tradition of quality, longevity, accuracy and ease of use. X Series ICP-MS Australia [email protected] Austria [email protected] Belgium [email protected] Canada [email protected] China [email protected] France [email protected] Germany [email protected] Italy [email protected] Japan [email protected] Netherlands [email protected] Nordic [email protected] South Africa [email protected] Spain [email protected] Switzerland [email protected] UK [email protected] USA [email protected] Finnigan ELEMENT2 Finnigan NEPTUNE DIN EN ISO 91:2 Thermo Electron (Bremen) GmbH is certified DIN EN ISO 91:2 24 Thermo Electron Corporation. All rights reserved. Microsoft and Windows are registered trade marks of the Microsoft Corporation. All other trademarks are the property of Thermo Electron Corporation and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. BR327_E 5/4C
Thermo Scientific ConFlo IV Universal Interface
m a s s s p e c t r o m e t r y Thermo Scientific ConFlo IV Universal Interface The Universal Interface Part of Thermo Fisher Scientific ConFlo IV Universal Interface The Universal Interface The development
How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer
What is ICP-MS? and more importantly, what can it do? Inductively Coupled Plasma Mass Spectrometry or ICP-MS is an analytical technique used for elemental determinations. The technique was commercially
Thermo Scientific ExactFinder Software
mass spectrometry Thermo Scientific ExactFinder Software Workflow software for routine targeted and general unknown screening Unifies qualitative and quantitative high-resolution accurate mass workflows
aurora M90 Innovation with Integrity A new era in ICP-MS ICP-MS
aurora M90 A new era in ICP-MS Innovation with Integrity ICP-MS Your Partner in ICP-MS Solutions Bruker continues to find new and novel ways to meet your changing needs. As a leader in elemental analysis
Thermo Scientific SIEVE Software for Differential Expression Analysis
m a s s s p e c t r o m e t r y Thermo Scientific SIEVE Software for Differential Expression Analysis Automated, label-free, semi-quantitative analysis of proteins, peptides, and metabolites based on comparisons
Sample Analysis Design Isotope Dilution
Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber
Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber Application note Energy & chemicals, petrochemicals Authors Elizabeth
Thermo Scientific icap 7000 Series ICP-OES. Low cost ICP-OES analysis high quality data
Low cost ICP-OES analysis high quality data The NEW Thermo Scientific icap 7000 Series ICP-OES provides low cost multi-element analysis for measuring trace elements in a diverse sample range. The instrument
Standard Operation Procedure. Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry
Standard Operation Procedure Elemental Analysis of Solution samples with Inductively Coupled Plasma Mass Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
aurora ICP-MS Innovation with Integrity The Science of Sensitivity ICP-MS
aurora ICP-MS The Science of Sensitivity Innovation with Integrity ICP-MS Powers the Business of Quantification aurora ICP-MS enhanced value package Bruker s ICP-MS is famous for its innovative and unique
TargetQuan 3 Software. Leading the way in regulatory. POPs quantification. Bullet Bullet Bullet
TargetQuan 3 Software Leading the way in regulatory POPs quantification Bullet Bullet Bullet Leading the way in regulatory POPs quantification Analyse Samples Open Sequence Process Sequence Analyse Samples
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT
NANOCOMPOSIX'S GUIDE TO ICP-MS MEASUREMENT AND ANALYSIS SEPTEMBER 2012, V 1.1 4878 RONSON CT STE K SAN DIEGO, CA 92111 858-565 - 4227 NANOCOMPOSIX.COM Note to the Reader: We at nanocomposix have published
Thermo Scientific Prelude SPLC System FPO. Making LC/MS accessible. to clinical research and toxicology labs
Thermo Scientific Prelude SPLC System FPO Making LC/MS accessible to clinical research and toxicology labs Specifically designed to make the power of LC/MS accessible Analysis of target compounds in complex
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS
A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS Kent W. Warnken, Thermo Fisher Scientific, USA Patrick Messina, Quest Diagnostics,
How To Use The Nexion 300
APPLICATION NOTE ICP - Mass Spectrometry Authors Ewa Pruszkowski, Ph.D. Senior ICP-MS Application Scientist Cynthia P. Bosnak Senior Product Specialist PerkinElmer, Inc. Shelton, CT USA The Analysis of
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES
Determination of Metals in a 3% Sodium Chloride (NaCl) Matrix by Axially-Viewed ICP-OES Application Note Inductively Coupled Plasma-Optical Emission Spectrometers Author Deen Johnson Introduction The development
AMD Analysis & Technology AG
AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds
The 30-Minute Guide to ICP-MS
Technical Note The 30-Minute Guide to ICP-MS ICP-Mass Spectrometry A Worthy Member of the Inorganic Analysis Team For nearly 30 years, inductively coupled plasma mass spectrometry (ICP-MS) has been gaining
Getting Answers at the Source Requires a New Way of Thinking About Software
m o l e c u l a r s p e c t r o s c o p y RESULT Software Getting Answers at the Source Requires a New Way of Thinking About Software Method Development Workflow Development Deployment & Support Methodology
Regional Environmental Protection Agency- Veneto Region LABORATORY REGIONAL DEPARTMENT
Twinning Contract BG 06 IB EN 01 Ministry of Environment & Water ICP-MS methods for metals: in-house validation and extension of field of application of methods from inland waters to transitional and sea-water.
Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS
Application Note: 446 Selective Testosterone Analysis in Human Serum by LC-FAIMS-MS/MS Jonathan McNally and Michael Belford, Thermo Fisher Scientific, San Jose, CA James Kapron, Thermo Fisher Scientific,
How To Use An Ionsonic Microscope
Sample Analysis Design Element2 - Basic Software Concepts Scan Modes Magnetic Scan (BScan): the electric field is kept constant and the magnetic field is varied as a function of time the BScan is suitable
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Mass spectrometry. What are the principles behind MS? What do all MS instruments have in common?
Mass spectrometry What are the principles behind MS? What do all MS instruments have in common? What are the different types of MS? Lecture outline: 1) Introduction to mass spectrometry 2) sample introduction
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
Specifications. MassHunter WorkStation for Agilent 7700 Series ICP-MS. Specifications: MassHunter Workstation (G7211A) for the Agilent 7700 Series
Specifications MassHunter WorkStation for Agilent 7700 Series ICP-MS Specifications: MassHunter Workstation (G7211A) for the Agilent 7700 Series The ICP-MS MassHunter Workstation for the Agilent 7700 Series
Elemental Analyses by ICP-AES
Elemental Analyses by ICP-AES Henry Gong, Senior Analytical Chemist September 10, 2008 ICP-AES inductively coupled plasma atomic emission spectrophotometry Electrons of an atom absorb energy and jump to
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
OLIVÉR BÁNHIDI 1. Introduction
Materials Science and Engineering, Volume 39, No. 1 (2014), pp. 5 13. DETERMINATION OF THE ANTIMONY- AND STRONTIUM- CONTENT OF ALUMINIUM ALLOYS BY INDUCTIVELY COUPLED PLASMA ATOM EMISSION SPECTROMETRY
AutoSpec Premier. Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements?
AutoSpec Premier Are you? Analyzing dioxins and related compounds? Required to satisfy regulatory requirements? Looking for sensitivity, selectivity and dynamic range? The AutoSpec Premier is the latest
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES
ICP OPTICAL ATOMIC EMISSION SPECTROSCOPY Analysis of Chlorine, Bromine and Iodine in Water using ICP-AES HORIBA Scientific Longjumeau, France Keywords: metallurgy, non-ferrous, halogens 1 Introduction
Simultaneous Quantitation of 43 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method
Simultaneous Quantitation of 4 Drugs in Human Urine with a Dilute-and-Shoot LC-MS/MS Method Xiang He and Marta Kozak, Thermo Fisher Scientific, San Jose, CA Application Note 76 Key Words TSQ Quantum Access
Simplicity. Reliability. Performance
Simplicity Reliability Performance ICP Without Compromise At Teledyne Leeman Labs atomic spectroscopy is our business, our only business. We are deeply committed to providing you with technically superior
VALIDATION REPORT. Investigation of the major interferences for proposed Horizontal ICP AES method. ECN protocol
Page 1 van 36 pages VALIDATION REPORT Investigation of the major interferences for proposed Horizontal ICP AES method Identification code Validation in accordance with Reference method Scope Reference
Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping
Thermo Scientific PepFinder Software A New Paradigm for Peptide Mapping For Conclusive Characterization of Biologics Deep Protein Characterization Is Crucial Pharmaceuticals have historically been small
Thermo Scientific Qtegra Intelligent Scientific Data Solution Software. Delivering quality. Driving productivity
Thermo cientific Qtegra Intelligent cientific Data olution oftware Delivering quality Driving productivity There can be economy only where there is efficiency The Thermo cientific Qtegra Intelligent cientific
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
MODERN ATOMIC THEORY AND THE PERIODIC TABLE
CHAPTER 10 MODERN ATOMIC THEORY AND THE PERIODIC TABLE SOLUTIONS TO REVIEW QUESTIONS 1. Wavelength is defined as the distance between consecutive peaks in a wave. It is generally symbolized by the Greek
h e l p s y o u C O N T R O L
contamination analysis for compound semiconductors ANALYTICAL SERVICES B u r i e d d e f e c t s, E v a n s A n a l y t i c a l g r o u p h e l p s y o u C O N T R O L C O N T A M I N A T I O N Contamination
[ Care and Use Manual ]
PREP Calibration Mix DIOS Low i. Introduction Pre-packaged PREP Calibration Mixtures eliminate the need to purchase and store large quantities of the component calibration reagents, simplifying sample
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
Thermo Scientific Compound Discoverer Software. A New Generation. of integrated solutions for small molecule structure ID
Thermo Scientific Compound Discoverer Software A New Generation of integrated solutions for small molecule structure ID Transforming Small Molecule Research Solutions for Small Molecule Structural ID Thermo
Appendix 5 Overview of requirements in English
Appendix 5 Overview of requirements in English This document is a translation of Appendix 4 (Bilag 4) section 2. This translation is meant as a service for the bidder and in case of any differences between
Universal Data Acquisition (UDA)
Universal Data Acquisition (UDA) I C P - O P T I C A L E M I S S I O N P R O D U C T N O T E Introduction Historically, Inductively Coupled Plasma (ICP) spectroscopy has been used for multiple analyte
API 3200 LC/MS/MS SYSTEM. Performance, productivity and value combined
API 3200 LC/MS/MS SYSTEM Performance, productivity and value combined API 3200 LC/MS/MS SYSTEM Reliability, reproducibility, and confidence in your data. With an unmatched heritage of technological innovation
PosterREPRINT AN LC/MS ORTHOGONAL TOF (TIME OF FLIGHT) MASS SPECTROMETER WITH INCREASED TRANSMISSION, RESOLUTION, AND DYNAMIC RANGE OVERVIEW
OVERVIEW Exact mass LC/MS analysis using an orthogonal acceleration time of flight (oa-tof) mass spectrometer is a well-established technique with a broad range of applications. These include elemental
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence)
University of Wisconsin Chemistry 524 Spectroscopic Applications (GFAA, ICP, UV/Vis, Fluorescence) For this laboratory exercise, you will explore a variety of spectroscopic methods used in an analytical
13C NMR Spectroscopy
13 C NMR Spectroscopy Introduction Nuclear magnetic resonance spectroscopy (NMR) is the most powerful tool available for structural determination. A nucleus with an odd number of protons, an odd number
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and
Modern approaches to determination of toxic metals in marine environmental objects. Atomic absorption and inductively coupled plasma, advantages and disadvantages Atomic spectroscopy Atomic spectroscopy
Thermo Scientific irm-gc/ms Technology
m a s s s p e c t r o m e t r y Thermo Scientific irm-gc/ms Technology Get More Information from Your GC Samples Environmental Studies Forensics and Doping Control Ecology and Metabolic Studies Geochemistry
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Why? Chemists are concerned with mass relationships in chemical reactions, usually run on a macroscopic scale (grams, kilograms, etc.). To deal with
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
Standard Operation Procedure. Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry
Standard Operation Procedure Elemental Analysis of Solution Samples with Inductively Coupled Plasma Optical Emission Spectrometry Soil & Plant Analysis Laboratory University of Wisconsin Madison http://uwlab.soils.wisc.edu
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
amazon SL Innovation with Integrity Setting New Standards in Performance, Simplicity and Value Ion Trap MS
amazon SL Setting New Standards in Performance, Simplicity and Value Innovation with Integrity Ion Trap Best-In-Class Ion Trap Mass Spectrometer for Routine Analysis The amazon SL entry-level system is
Determination of elements by ICP-AES and ICP-MS
September 2003 HORIZONTAL - 19 Determination of elements by ICP-AES and ICP-MS Henk J van de Wiel National Institute of Public Health and the Environment (RIVM) Bilthoven, The Netherlands 1 HORIZONTAL
LC-MS/MS Solutions for the Pharmaceutical and CRO Industries TSQ Series -based Solutions for Drug Discovery and Development
m a s s s p e c t r o m e t r y LC-MS/MS Solutions for the Pharmaceutical and CRO Industries TSQ Series -based Solutions for Drug Discovery and Development Part of Thermo Fisher Scientific ACCELERATING
Elemental analyzer multi EA 5000
Elemental analyzer multi EA 5000 Analytik Jena AG Konrad-Zuse-Str. 1 07745 Jena/Germany www.analytik-jena.com [email protected] 1/6 multi EA 5000 The multi EA 5000 is a powerful elemental analyzer
Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS)
Application Note: 52020 Application of Automated Data Collection to Surface-Enhanced Raman Scattering (SERS) Timothy O. Deschaines, Ph.D., Thermo Fisher Scientific, Madison, WI, USA Key Words Array Automation
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity
Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use Confidence in Results With Data Integrity 2 Make the World Healthier With the LC-MS Tests You Run Confidence in Test Results With Data
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates
In-Depth Qualitative Analysis of Complex Proteomic Samples Using High Quality MS/MS at Fast Acquisition Rates Using the Explore Workflow on the AB SCIEX TripleTOF 5600 System A major challenge in proteomics
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays
Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative
HRMS in Clinical Research: from Targeted Quantification to Metabolomics
A sponsored whitepaper. HRMS in Clinical Research: from Targeted Quantification to Metabolomics By: Bertrand Rochat Ph. D., Research Project Leader, Faculté de Biologie et de Médecine of the Centre Hospitalier
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
Agilent Cary 60 UV-Vis
Agilent Cary 60 UV-Vis Efficient. Accurate. Flexible. Specifications Introduction The Agilent Cary 60 UV-Vis spectrophotometer is efficient, accurate and flexible, and is designed to meet both current
Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES. FIMS 100 and 400. Flow Injection Mercury Systems
Image Area EXCEPTIONAL PERFORMANCE FOR CHALLENGING MERCURY ANALYSES FIMS 100 and 400 Flow Injection Mercury Systems Delivering the Performance You Need The determination and monitoring of sub-ppb concentrations
SUS. Company Profile. Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0)208/658535 Fax 0049(0)208/658536
SUS Ulrich Nell, Feldstr.23, D - 46149 Oberhausen, Tel. 0049(0208/658535 Fax 0049(0208/658536 Company Profile SUS was founded in 1986 in Oberhausen in the Ruhr area (close to Düsseldorf, in order to meet
APPLICATION OF ICP-MS TECHNOLOGY FOR TRACE METALS ANALYSIS
MR. TELLIARD: Our next speaker is Paula Hogg. Paula is currently a lab manager at Hampton Roads Sanitation District s Central Environmental Laboratory. We would like to also thank Hampton Roads for sending
NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight.
NexION 350 ICP-MS Maximum throughput. Minimum downtime. Optimum insight. When your instrument is more efficient and productive, so too are your scientists and your lab. And nothing puts the flow in your
Application of ICP-MS as a Tool for Environmental Monitoring. Bill Spence Environmental Market Specialist
Application of ICP-MS as a Tool for Environmental Monitoring Bill Spence Environmental Market Specialist Introduction Knowledge of the concentration of metallic species in the environment has been important
QGA Quantitative Gas Analyser
QGA Quantitative Gas Analyser A compact bench-top system for real-time gas and vapour analysis Detailed product information / introduction catalysis studies environmental gas analysis fermentation off-gas
DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS
DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display
G4 ICARUS HF. Innovation with Integrity. Simultaneous Determination of Carbon and Sulfur in Solids. Combustion Method
G4 ICARUS HF Simultaneous Determination of Carbon and Sulfur in Solids Innovation with Integrity Combustion Method High Performance Carbon & Sulfur Analyzer Combustion Method with High-Frequency Furnace
High Sensitivity ICP-MS: Overcome the problem of complex samples
High Sensitivity ICP-MS: Overcome the problem of complex samples Meike Hamester, Andrew Toms and René Chemnitzer Bruker Daltonics Berlin, Germany - Milton, Canada Historical Product Roadmap 2003 2005 2010
Start of introducing ICP MS method for analyzing water samples for Strontium. November 27, 2014 Tokyo Electric Power Company
Start of introducing ICP MS method for analyzing water samples for Strontium November 27, 2014 Tokyo Electric Power Company 1. Current Strontium analysis and introduction of various methods Strontium analysis
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
CLASS TEST GRADE 11. PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change
CLASS TEST GRADE PHYSICAL SCIENCES: CHEMISTRY Test 6: Chemical change MARKS: 45 TIME: hour INSTRUCTIONS AND INFORMATION. Answer ALL the questions. 2. You may use non-programmable calculators. 3. You may
SCION SQ Series GC-MS
SCION SQ Series GC-MS The Gas Chromatographers Detector SCION SQ TM GC-MS The SCION SQ series GC-MS systems combine innovative engineering with detailed cutomer requirements to produce a truly innovative
Simultaneous qualitative and quantitative analysis using the Agilent 6540 Accurate-Mass Q-TOF
Simultaneous qualitative and quantitative analysis using the Agilent 654 Accurate-Mass Q-TOF Technical Overview Authors Pat Perkins Anabel Fandino Lester Taylor Agilent Technologies, Inc. Santa Clara,
INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE. AxION iqt GC/MS/MS
AxION iqt GC/MS/MS INTEGRATED TECHNOLOGY TO SEE THE WHOLE PICTURE IDENTIFY WHAT YOU RE LOOKING FOR (AND EVEN WHAT YOU RE NOT) The GC/MS/MS Solution for the Work You Do The ideal solution for industrial,
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System
Quick and Sensitive Analysis of Multiclass Veterinary Drug Residues in Meat, Plasma, and Milk on a Q Exactive Focus LC-MS System Olaf Scheibner, Maciej Bromirski, Thermo Fisher Scientific, Bremen, Germany
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances
Accurate Mass Screening Workflows for the Analysis of Novel Psychoactive Substances TripleTOF 5600 + LC/MS/MS System with MasterView Software Adrian M. Taylor AB Sciex Concord, Ontario (Canada) Overview
FT-NIR for Online Analysis in Polyol Production
Application Note: 51594 FT-NIR for Online Analysis in Polyol Production Key Words Acid Number Ethylene Oxide FT-NIR Hydroxyl Value Polyester Polyols Abstract Hydroxyl value and other related parameters
Agilent Television Power Consumption Testing. Application Note
Agilent Television Power Consumption Testing Application Note Introduction Today, there are many types of televisions (TVs) on the market: the cathode ray tube (CRT) TV, liquid crystal display (LCD) TV,
Overview. Introduction. AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS System
Investigating the use of the AB SCIEX TripleTOF 4600 LC/MS/MS System for High Throughput Screening of Synthetic Cannabinoids/Metabolites in Human Urine AB SCIEX MPX -2 High Throughput TripleTOF 4600 LC/MS/MS
How To Build An Amd Quas 3 Ar Mass Spectrometer
Classical and special AMD Magnetic Sector mass spectrometers (extracted from power point presentations) High Resolution Mass Spectromter AMD 402/403 S 6 High Resolution Mass Spectrometer AMD 604 S 8 Version
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests
Effects of Intelligent Data Acquisition and Fast Laser Speed on Analysis of Complex Protein Digests AB SCIEX TOF/TOF 5800 System with DynamicExit Algorithm and ProteinPilot Software for Robust Protein
Amount of Substance. http://www.avogadro.co.uk/definitions/elemcompmix.htm
Page 1 of 14 Amount of Substance Key terms in this chapter are: Element Compound Mixture Atom Molecule Ion Relative Atomic Mass Avogadro constant Mole Isotope Relative Isotopic Mass Relative Molecular
Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers. Rob Grothe Thermo Fisher Scientific, San Jose, CA
Mass and Resolution Calibration for New Triple-Stage Quadrupole Mass Spectrometers Rob Grothe Thermo Fisher Scientific, San Jose, CA Overview Purpose: Robust and accurate calibration of mass positions
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics
MRMPilot Software: Accelerating MRM Assay Development for Targeted Quantitative Proteomics With Unique QTRAP and TripleTOF 5600 System Technology Targeted peptide quantification is a rapidly growing application
Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers
Key Questions & Exercises Chem 115 POGIL Worksheet - Week 4 Moles & Stoichiometry Answers 1. The atomic weight of carbon is 12.0107 u, so a mole of carbon has a mass of 12.0107 g. Why doesn t a mole of
Wipe Analysis to Determine Metal Contamination on Critical Surfaces
By Albert Dato, Ph.D., Warren York, Jennifer Jew, Laarni Huerta, Brice Norton, and Michael Coste On-wafer metallic contamination is detrimental to the fabrication and performance of semiconductor devices.
Flow Injection Analysis
Flow Injection Analysis Almost all reagent based assays can be downscaled and automated by Flow Injection Analysis. FIA was first described in a patent filed in Denmark by Ruzicka and Hansen in 1974. Since
Bruker ToxScreener TM. Innovation with Integrity. A Comprehensive Screening Solution for Forensic Toxicology UHR-TOF MS
Bruker ToxScreener TM A Comprehensive Screening Solution for Forensic Toxicology Innovation with Integrity UHR-TOF MS ToxScreener - Get the Complete Picture Forensic laboratories are frequently required
CHEM 1411 Chapter 5 Homework Answers
1 CHEM 1411 Chapter 5 Homework Answers 1. Which statement regarding the gold foil experiment is false? (a) It was performed by Rutherford and his research group early in the 20 th century. (b) Most of
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
GC-xt. The most flexible GC /GC-MS liquid sample injector
GC-xt Prep and Load Platform The most flexible GC /GC-MS liquid sample injector Environmental Foods/Beverages Consumer Products Forensics Petrochemicals/Polymers Pharmaceuticals GC-xt Liquid mode Prep
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
G8 GALILEO. Innovation with Integrity. High-End Melt-extraction Analyzer. Inert Gas Method
G8 GALILEO High-End Melt-extraction Analyzer Innovation with Integrity Inert Gas Method Determination of O, N and H The market demands Metals, minerals, and inorganic compound markets demand high-quality